<article-meta>
<article-id pub-id-type="doi">10.3389/fmed.2022.868217</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Medicine</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>Characterization and Analysis of Retinal Axial Motion at High Spatiotemporal Resolution and Its Implication for Real-Time Correction in Human Retinal Imaging</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Cai</surname> <given-names>Yao</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/1649860/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Grieve</surname> <given-names>Kate</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/1413974/overview"/>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name><surname>Mec&#x000EA;</surname> <given-names>Pedro</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="aff" rid="aff4"><sup>4</sup></xref>
<xref ref-type="corresp" rid="c001"><sup>&#x0002A;</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/1836091/overview"/>
</contrib>
</contrib-group>
<aff id="aff1"><sup>1</sup><institution>Institut Langevin, ESPCI Paris, CNRS, PSL University</institution>, <addr-line>Paris</addr-line>, <country>France</country></aff>
<aff id="aff2"><sup>2</sup><institution>Sorbonne Universit&#x000E9;, INSERM, CNRS, Institut de la Vision</institution>, <addr-line>Paris</addr-line>, <country>France</country></aff>
<aff id="aff3"><sup>3</sup><institution>CHNO des Quinze-Vingts, INSERM-DGOS CIC 1423</institution>, <addr-line>Paris</addr-line>, <country>France</country></aff>
<aff id="aff4"><sup>4</sup><institution>DOTA, ONERA, Universit&#x000E9; Paris Saclay</institution>, <addr-line>Palaiseau</addr-line>, <country>France</country></aff>
<author-notes>
<fn fn-type="edited-by"><p>Edited by: Hossein Rabbani, Isfahan University of Medical Sciences, Iran</p></fn>
<fn fn-type="edited-by"><p>Reviewed by: Qinqin Zhang, University of Washington, United States; Yousef Ahmed Fouad, Ain Shams University, Egypt</p></fn>
<corresp id="c001">&#x0002A;Correspondence: Pedro Mec&#x000EA; <email>pedro.mece&#x00040;onera.fr</email></corresp>
<fn fn-type="other" id="fn001"><p>This article was submitted to Ophthalmology, a section of the journal Frontiers in Medicine</p></fn></author-notes>
<pub-date pub-type="epub">
<day>12</day>
<month>07</month>
<year>2022</year>
</pub-date>
<pub-date pub-type="collection">
<year>2022</year>
</pub-date>
<volume>9</volume>
<elocation-id>868217</elocation-id>
<history>
<date date-type="received">
<day>02</day>
<month>02</month>
<year>2022</year>
</date>
<date date-type="accepted">
<day>20</day>
<month>06</month>
<year>2022</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#x000A9; 2022 Cai, Grieve and Mec&#x000EA;.</copyright-statement>
<copyright-year>2022</copyright-year>
<copyright-holder>Cai, Grieve and Mec&#x000EA;</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/"><p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p></license>
</permissions>
<abstract>
<p>High-resolution ophthalmic imaging devices including spectral-domain and full-field optical coherence tomography (SDOCT and FFOCT) are adversely affected by the presence of continuous involuntary retinal axial motion. Here, we thoroughly quantify and characterize retinal axial motion with both high temporal resolution (200,000 A-scans/s) and high axial resolution (4.5 &#x003BC;m), recorded over a typical data acquisition duration of 3 s with an SDOCT device over 14 subjects. We demonstrate that although breath-holding can help decrease large-and-slow drifts, it increases small-and-fast fluctuations, which is not ideal when motion compensation is desired. Finally, by simulating the action of an axial motion stabilization control loop, we show that a loop rate of 1.2 kHz is ideal to achieve 100% robust clinical <italic>in-vivo</italic> retinal imaging.</p></abstract>
<kwd-group>
<kwd>optical coherence tomography</kwd>
<kwd>eye motion</kwd>
<kwd>high-resolution retinal imaging</kwd>
<kwd>eye tracking</kwd>
<kwd>OCT angiography</kwd>
</kwd-group>
<contract-num rid="cn001">101001841</contract-num>
<contract-num rid="cn001">610110</contract-num>
<contract-num rid="cn002">x16-CARN 0029-01</contract-num>
<contract-num rid="cn003">IHU FOReSIGHT ANR-18-IAHU-0001</contract-num>
<contract-num rid="cn004">Sesame 4D-EYE EX047007</contract-num>
<contract-sponsor id="cn001">European Research Council<named-content content-type="fundref-id">10.13039/501100000781</named-content></contract-sponsor>
<contract-sponsor id="cn002">Fondation Voir et Entendre<named-content content-type="fundref-id">10.13039/501100013121</named-content></contract-sponsor>
<contract-sponsor id="cn003">Agence Nationale de la Recherche<named-content content-type="fundref-id">10.13039/501100001665</named-content></contract-sponsor>
<contract-sponsor id="cn004">Conseil R&#x000E9;gional, &#x000CE;le-de-France<named-content content-type="fundref-id">10.13039/501100003990</named-content></contract-sponsor>
<counts>
<fig-count count="6"/>
<table-count count="1"/>
<equation-count count="0"/>
<ref-count count="24"/>
<page-count count="8"/>
<word-count count="4611"/>
</counts>
</article-meta>
</front>
<body>
<sec sec-type="intro" id="s1">
<title>1. Introduction</title>
<p>A major challenge for <italic>in vivo</italic> retinal high resolution ophthalmic imaging with optical coherence tomography (OCT) is posed by artifacts caused by constant involuntary axial eye motion, from sources including head and body motion, physiological phenomena such as cardiac cycle, blood flow, pulsation, muscular, and/or respiratory activity (<xref ref-type="bibr" rid="B1">1</xref>, <xref ref-type="bibr" rid="B2">2</xref>). For simple B-scan acquisition with SD-OCT devices, motion related artifacts can be automatically corrected during post-processing, but with the growing popularity of imaging methods based on volumetric SDOCT acquisitions with <italic>en face</italic> image reconstruction such as motion contrast-based OCT-Angiography (OCTA) (<xref ref-type="bibr" rid="B3">3</xref>), axial eye motion may provoke enough decorrelation to cause the false positive appearance of flow, significantly degrade the image quality, and add distracting artifacts in reconstructed <italic>en face</italic> slices (<xref ref-type="bibr" rid="B4">4</xref>).</p>
<p>Axial eye motion is also an important consideration for camera-based OCT techniques which acquire directly in the <italic>en face</italic> direction, such as time-domain full-field optical coherence tomography (FFOCT). FFOCT is an imaging modality capable of recording high-speed en-face sections of a sample at a given depth (<xref ref-type="bibr" rid="B5">5</xref>). One very attractive characteristic of FFOCT is the use of spatially incoherent illumination to make the FFOCT lateral resolution robust to symmetric optical aberrations (<xref ref-type="bibr" rid="B6">6</xref>), which are the dominant aberrations in the eye (around 90%) (<xref ref-type="bibr" rid="B7">7</xref>). By exploiting this phenomenon, we were able to apply FFOCT for high 3D resolution imaging of the living human retina. This achievement was possible owing to recent advances in FFOCT, including shaping the temporal coherence gate to match the retina curvature, achieving a wide field-of-view (5 &#x000D7; 5&#x000B0;) (<xref ref-type="bibr" rid="B8">8</xref>), stabilizing the axial retinal motion in real-time (<xref ref-type="bibr" rid="B9">9</xref>), and increasing signal-to-noise ratio (SNR) using the adaptive-glasses approach for ocular aberration correction (<xref ref-type="bibr" rid="B10">10</xref>). Given its relative simplicity and small footprint, FFOCT holds promise for adoption by clinicians (<xref ref-type="bibr" rid="B11">11</xref>).</p>
<p>When acquiring a single en-face image in FFOCT, a micrometer stabilization precision can be obtained. However, longer acquisition durations on the order of a few seconds may be necessary in the clinical setting in order to allow image averaging to improve SNR. Therefore, it is necessary to accurately compensate for axial motion during image acquisition by moving the reference or the sample arm of FFOCT accordingly (<xref ref-type="bibr" rid="B9">9</xref>). A few methods exist, but they are either used in conventional OCT, thus not suitable for FFOCT en-face imaging (<xref ref-type="bibr" rid="B12">12</xref>, <xref ref-type="bibr" rid="B13">13</xref>), or are not sufficiently stable to remain within the coherence gate for acquisition durations over several seconds (<xref ref-type="bibr" rid="B9">9</xref>). Indeed, our previous work (<xref ref-type="bibr" rid="B9">9</xref>) achieved an RMS error of around 10 &#x003BC;m, i.e., greater than the coherence gate width of 8 &#x003BC;m, which was nevertheless sufficient for a satisfactory 60% imaging success rate in subjects with good fixation. However, for imaging in patients who may have poorer fixation, and eventual extension to <italic>in vivo</italic> functional imaging with dynamic FFOCT (<xref ref-type="bibr" rid="B14">14</xref>), a greater precision of axial tracking is desirable. Although the causes of axial retinal motion are well understood, the literature on axial retinal motion is still incomplete, as collected data had low temporal and axial resolution (<xref ref-type="bibr" rid="B1">1</xref>), and tested relatively few subjects (<xref ref-type="bibr" rid="B1">1</xref>, <xref ref-type="bibr" rid="B9">9</xref>, <xref ref-type="bibr" rid="B13">13</xref>), so that general conclusions on the most suitable design of a retinal axial motion stabilization control-loop could not be drawn.</p>
<p>In this article, we present the first characterization of the axial retinal motion with both high temporal resolution (200,000 A-scans/s) and high axial resolution (4.5 &#x003BC;m in water), over a typical data acquisition duration of 3 s. Characterization is performed on a 14-subject population, making it possible to come up with a statistical description reflecting the inter-subject variability. We also evaluate for the first time the direct influence of breathing on axial motion in our population, as this is a parameter that is clinically feasible to control during short image acquisition durations and which may affect stability. Finally, we draw conclusions on design considerations of axial retinal motion stabilization for robust clinical <italic>in vivo</italic> retinal OCTA and FFOCT imaging.</p>
</sec>
<sec sec-type="methods" id="s2">
<title>2. Methods</title>
<p>To measure the retinal axial position, we used a commercial spectral domain OCT system (SD-OCT; GAN611, Thorlabs). The SD-OCT system comprises a broadband superluminescent diode with 930 nm central wavelength and 100 nm bandwidth, providing a theoretical axial resolution of 4 &#x003BC;m in water. It presents an A-scan rate up to 248 kHz with a sensitivity of 84 dB and 1,024 axial pixels, comprising an axial range of 2.2 mm in water. The galvanometer scanners of the SD-OCT were conjugated to the eye&#x00027;s pupil with a 4-f system. The SD-OCT light beam arrives in the eye&#x00027;s pupil with a 4 mm diameter. The experimental setup is shown in <xref ref-type="fig" rid="F1">Figure 1</xref>.</p>
<fig id="F1" position="float">
<label>Figure 1</label>
<caption><p>Schematic drawing of the experimental setup used for retinal axial motion measurement, consisting of a commercial SD-OCT, a 4-f system conjugating the scanner in SD-OCT to the eye&#x00027;s pupil, and a LCD projector to display the target for eye fixation. The SD-OCT light beam arrives in the eye&#x00027;s pupil with a 4 mm diameter.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0001.tif"/>
</fig>
<p>To obtain statistics on axial retinal motion, 14 healthy subjects (age 22&#x02013;80 years) free of ocular disease were invited to participate in the study. Research procedures followed the tenets of the Declaration of Helsinki. Prior to data collection, the nature and possible consequences of the study were explained, and informed consent was obtained from all subjects. This study was authorized by the appropriate ethics review boards [CPP and ANSM (IDRCB number: 2019-A00942-55)]. Subjects were seated in front of the system and stabilized with a chin and forehead rest, without any pupil dilation nor cycloplegia. Image acquisition was realized in a dark room, maximizing the pupil dilation. They were asked to gaze at a yellow fixation cross displayed on a black screen using an LCD projector, maintaining fixation for about 4 s until acquisition was complete. To provide a good trade-off between acquisition speed and SNR, we chose to scan 1&#x000B0; field of view (FOV) of the retina with 256 A-scans at 200 kHz A-scan rate, corresponding to 1.28-ms exposure time for a single B-scan. With a 2-ms fly-back time, we were able to achieve a B-scan rate of 300 Hz. During image acquisition, the output power measured at the eye pupil plane was 850 &#x003BC;W, which is below the ocular safety limits established by the ISO standards for group 1 devices. To study the influence of breathing on axial retinal motion, two groups of data were obtained from our 14-subject population: while subjects held their breath for around 5 s or while they were breathing normally. The participants would indicate if they succeeded in holding their breath during acquisition. In the case of ambiguous data, a new dataset would be acquired.</p>
<p>After image acquisition, SD-OCT B-scan images were used to measure the retinal axial position using a previously introduced algorithm based on normalized cross-correlation and parabolic fitting to obtain a subpixel precision (<xref ref-type="bibr" rid="B9">9</xref>). Blinks were automatically detected by an intensity-based algorithm (<xref ref-type="bibr" rid="B9">9</xref>). We extracted blink-free B-scan sequences, resulting in 3-s time series of retinal axial position data for all 14 subjects. Temporal power spectral densities (PSD) were calculated by a FFT routine.</p>
</sec>
<sec id="s3">
<title>3. Results and Discussion</title>
<sec>
<title>3.1. Temporal Fluctuation of Retinal Axial Position</title>
<p><xref ref-type="fig" rid="F2">Figures 2A&#x02013;D</xref> presents four typical examples of retinal axial position as a function of time. A good superposition of measured axial position (red line) and averaged A-scan time series has validated the proposed data processing method. As exemplified here, slow drifts and relative fast oscillations could be observed in all 14 subjects. Still, the slow and fast motion amplitude and frequency varied among subjects. <xref ref-type="fig" rid="F2">Figures 2E&#x02013;H</xref> present the corresponding axial motion power spectra density (PSD) for the same subjects in <xref ref-type="fig" rid="F2">Figures 2A&#x02013;D</xref>. All PSDs present similar behavior, with an energy decay in a <italic>f</italic><sup>&#x02212;<italic>p</italic></sup> power-law, where f is the frequency, before reaching the noise-level plateau around 150 Hz. Note that the presence of pronounced peaks between 1 and 4 Hz varied in the population.</p>
<fig id="F2" position="float">
<label>Figure 2</label>
<caption><p>Dynamics of retinal axial position. <bold>(A&#x02013;D)</bold> Four examples of acquired averaged A-scan time series over 3 s without blinks. The red line indicates the measured axial position temporal evolution. <bold>(E&#x02013;H)</bold> Corresponding temporal PSD computed for the retinal axial position. Note that subjects 1 and 12 held their breath, while subjects 4 and 9 breathed normally.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0002.tif"/>
</fig>
<p>To draw statistics of retinal axial motion, and to investigate the effect of breathing on motion, we computed the standard deviation of each axial position time series for each of 14 subjects. <xref ref-type="fig" rid="F3">Figure 3</xref> shows temporal fluctuation of axial position under breath holding (blue bars) and normal breathing (red bars) conditions. Note that 9 subjects out of 14 presented a lower temporal fluctuation while holding their breath. Greater variations observed in subjects 12 and 13 are caused by larger drifts, most probably due to head motion. Two participants (subjects 1 and 6) were trained for eye fixation tasks. In contrast to retinal lateral motion, where experience in fixation can play an important role in the temporal fluctuation (<xref ref-type="bibr" rid="B15">15</xref>, <xref ref-type="bibr" rid="B16">16</xref>), here no significant difference was noted, apart from a decrease of slow-and-large drifts linked to breathing and head motion. Additionally, although fixational microsaccades, i.e., fast and large amplitude lateral motion (<xref ref-type="bibr" rid="B15">15</xref>), took place during acquisition, axial retinal motion is not impacted by microsaccades.</p>
<fig id="F3" position="float">
<label>Figure 3</label>
<caption><p>Fluctuations of axial position for 14 subjects tested during breath-holding (blue) and normal breathing (red), calculated as the standard deviation of axial position.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0003.tif"/>
</fig>
<p>We computed the median &#x000B1; standard deviation over the population of peak-to-valley (PV) amplitude, mean deviation from zero, root mean square (RMS) value and speed of 3-s time series of retina axial positions for both breath-holding and normal breathing conditions. Results of this statistical analysis is summarized in <xref ref-type="table" rid="T1">Table 1</xref>. During normal breathing, retinal axial motion presents higher PV amplitude, meaning that slow drifts are more important compared with the breath-holding condition. Indeed, it has been shown that breathing may cause slow and large drifts due to the movement of the subject&#x00027;s head (<xref ref-type="bibr" rid="B1">1</xref>). Not surprisingly, after filtering out slow drifts corresponding to &#x0003C;2 Hz temporal frequency, we found similar PV amplitude value for both conditions, 158 &#x000B1; 52 &#x003BC;m for normal breathing compared to 160 &#x000B1; 46 &#x003BC;m for breath-holding. Similar values of mean deviation from zero (26 &#x000B1; 15 against 25 &#x000B1; 11 &#x003BC;m) and RMS (32 &#x000B1; 18 against 31 &#x000B1; 12 &#x003BC;m) were also obtained for normal breathing and breath-holding conditions respectively. These results suggest that respiration induces axial motion at temporal frequencies up to 2 Hz, whereas respiration rate is around 0.2 Hz (<xref ref-type="bibr" rid="B1">1</xref>). On the other hand, the breath-holding condition exhibited a faster axial movement of 988 &#x000B1; 209 &#x003BC;m/s, compared to 941 &#x000B1; 194 &#x003BC;m/s in the case of normal breathing, for which the causes are still unclear. In both cases, computed speeds are in line with 958 &#x003BC;m/s median speed previously reported (<xref ref-type="bibr" rid="B2">2</xref>), and are assumed to be related to the systolic phase of the heart pulse.</p>
<table-wrap position="float" id="T1">
<label>Table 1</label>
<caption><p>Axial motion statistics (median &#x000B1; standard deviation) over the 14-subject population.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th/>
<th valign="top" align="center"><bold>PV (&#x003BC;m)</bold></th>
<th valign="top" align="center"><bold>Mean deviation (&#x003BC;m)</bold></th>
<th valign="top" align="center"><bold>RMS (&#x003BC;m)</bold></th>
<th valign="top" align="center"><bold>Speed (&#x003BC;m/s)</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Normal breathing</td>
<td valign="top" align="center">276 &#x000B1; 154</td>
<td valign="top" align="center">57 &#x000B1; 36</td>
<td valign="top" align="center">68 &#x000B1; 42</td>
<td valign="top" align="center">941 &#x000B1; 194</td>
</tr>
<tr>
<td valign="top" align="left">Holding breath</td>
<td valign="top" align="center">230 &#x000B1; 98</td>
<td valign="top" align="center">42 &#x000B1; 24</td>
<td valign="top" align="center">52 &#x000B1; 28</td>
<td valign="top" align="center">988 &#x000B1; 209</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p><italic>Statistics were drawn in both normal breathing and breath-holding conditions in time series of 3-s duration. PV, peak-to-valley amplitude</italic>.</p>
</table-wrap-foot>
</table-wrap>
<p><xref ref-type="fig" rid="F4">Figure 4</xref> presents the averaged PSD over all 14 subjects in both breath-holding and normal breathing conditions. The averaged PSD has a <italic>f</italic><sup>&#x02212;2.2</sup> mean power-law before reaching the experimental noise-level at 150 Hz. Temporal frequencies below 1Hz present a factor of two lower PSD in the breath-holding condition than the normal breathing condition, highlighting the influence of respiration on axial motion. Remarkably, this difference is more pronounced at temporal frequencies associated to heartbeat (1&#x02013;2 Hz), showing a link between breathing and heartbeat on retinal axial motion. Indeed, apnea can induce a reduction in cardiac output and peripheral vasoconstriction (<xref ref-type="bibr" rid="B17">17</xref>, <xref ref-type="bibr" rid="B18">18</xref>). Both normal breathing and breath-holding conditions presented a similar behavior in the range of 3&#x02013;6 Hz, with some marked peak rises around 3&#x02013;4 Hz, which may be a high order harmonic of heartbeat rate (<xref ref-type="bibr" rid="B1">1</xref>) or related to pulsatile blood flow in the retina or in the choroid (<xref ref-type="bibr" rid="B19">19</xref>). Meanwhile, similar to the computed speed behavior, the PSD in the range of 7&#x02013;150 Hz is more important in the breath-holding condition than under normal breathing.</p>
<fig id="F4" position="float">
<label>Figure 4</label>
<caption><p>Averaged PSD over 14 subjects in breath-holding (blue trace) and normal breathing (red trace) conditions. The black dashed line gives a linear regression of the PSD following a <italic>f</italic><sup>&#x02212;2.2</sup> behavior.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0004.tif"/>
</fig>
</sec>
<sec>
<title>3.2. Axial Tracking Loop Simulation</title>
<p>Having characterized retinal axial motion, and knowing that breathing, heartbeat and pulsatile blood flow each play an important role, we next addressed the question of how fast the stabilization control-loop should run to precisely compensate for retinal axial motion in real-time for <italic>in vivo</italic> retinal imaging. To answer this question, we simulated the action of a retinal axial motion stabilization loop and integrator-based control scheme with a 0.5 gain (ensuring 45&#x000B0; stability margin) (<xref ref-type="bibr" rid="B20">20</xref>), assuming a two-frame delay (<xref ref-type="bibr" rid="B21">21</xref>). <xref ref-type="fig" rid="F5">Figure 5</xref> displays the overall distribution of axial RMS tracking errors among all participants and for different stabilization control loop rates. This result provides a valuable tool for selecting the acquisition frequency for retinal axial motion measurement, and consequently the control-loop rate for stabilization. For instance, it indicates that at 150 Hz loop-rate, the residual axial motion should be &#x0007E;10 &#x003BC;m rms in a normal breathing condition.</p>
<fig id="F5" position="float">
<label>Figure 5</label>
<caption><p>Box plot representation of residual RMS motion for different acquisition frequencies. The line within the box identifies the median. The blue and red traces indicate median RMS tracking errors during breath-holding and normal breathing, respectively. The length of the box represents the interquartile range (IQR), with the edges being the 25th (lower quartile) and 75th (upper quartile) percentiles. The ends of the whiskers represent minimum and maximum values. Values more than 1.5 IQRs are labeled as outliers (&#x0002B;). Note that the figure is plotted in a logarithmic scale.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0005.tif"/>
</fig>
<p>In the context of applying dynamic (D-) FFOCT <italic>in vivo</italic>, the use of singular value decomposition (SVD) processing was recently proposed to filter out the axial displacement of the sample from the local fluctuations linked to intracellular motility (<xref ref-type="bibr" rid="B22">22</xref>). This study showed that axial motion suppression works as long as residual axial motion after correction is smaller than the temporal coherence gate width (<xref ref-type="bibr" rid="B22">22</xref>, <xref ref-type="bibr" rid="B23">23</xref>). In FFOCT high-resolution <italic>in vivo</italic> retinal imaging, a temporal coherence gate of 8 &#x003BC;m is used (<xref ref-type="bibr" rid="B10">10</xref>). Therefore, if we approximate the temporal fluctuation of retinal axial motion as a Gaussian distribution, one can establish that an RMS error below 4 &#x003BC;m will ensure a residual axial motion of amplitude lower than the 8 &#x003BC;m coherence gate width more than 95% of the time. An RMS error of 4 &#x003BC;m can be achieved on average with a stabilization control-loop rate of 400 Hz under normal breathing conditions. The same precision can be obtained under the breath-holding condition, but with a higher loop rate, here around 600 Hz, showing that holding breath does not necessarily help axial motion stabilization but may in fact hinder it. Indeed, although breath holding decreases low temporal frequencies, these are the frequencies that are better managed by an integrator-based control scheme (<xref ref-type="bibr" rid="B9">9</xref>, <xref ref-type="bibr" rid="B20">20</xref>). On the other hand, breath holding amplifies high temporal frequencies, which are also amplified by an integrator-based control scheme (<xref ref-type="bibr" rid="B9">9</xref>, <xref ref-type="bibr" rid="B20">20</xref>).</p>
<p>In clinical application of FFOCT, one needs to consider the temporal fluctuation variability among subjects and ensure that the stabilization precision is enough to achieve robust imaging for 100% of the patients. <xref ref-type="fig" rid="F6">Figure 6</xref> highlights the proportion of subjects (of the 14 in total) where stabilization precision below 4 &#x003BC;m was obtained throughout the entire time series (i.e., 3 s) for different stabilization control-loop rates. To ensure robust FFOCT acquisition in all subjects, a control-loop rate of at least 1.2 kHz is necessary. Enticingly, this level of stability would also theoretically allow the first <italic>in vivo</italic> acquisition of dynamic FFOCT datasets, as axial motion would remain within the coherence gate width over the 3 s acquisition time required for a dynamic FFOCT image stack to probe temporal frequencies carrying intracellular dynamic signal (typically between 1 and 10 Hz) (<xref ref-type="bibr" rid="B14">14</xref>). This finding further supports the work of Pircher et al. (<xref ref-type="bibr" rid="B13">13</xref>) who found that increased high loop rate of 1 kHz enabled the reduction of the residual axial motion below 5 &#x003BC;m in transverse-scanning OCT. As commonly done in OCTA, the reduction of the axial resolution can help to decrease the sensitivity to axial motion, thus relaxing the control loop rate (<xref ref-type="bibr" rid="B4">4</xref>). For instance, a 600 Hz loop rate would be sufficient to ensure a residual RMS motion less than 8 &#x003BC;m in 100% of the patients for a 16 &#x003BC;m axial resolution (<xref ref-type="fig" rid="F6">Figure 6</xref>). Another strategy to alleviate the control loop rate is the use of Linear-quadratic-Gaussian (LQG) control law (<xref ref-type="bibr" rid="B24">24</xref>). Indeed, such an LQG approach could incorporate in the control strategy the specific spatio-temporal statistical structure of the axial retinal motion. The implementation of the LQG control strategy will be explored during future system optimization. An interesting way to reduce axial motion dynamics would be to perform measurements when the subject is lying down. Indeed, Kinkelder et al. (<xref ref-type="bibr" rid="B1">1</xref>) showed that the movements of the subject&#x00027;s head were reduced in a supine position. However, this posture is impractical for patients in clinical applications and so was not considered in our study. Correcting the axial motion in real-time in volumetric SDOCT and OCTA en-face projections can also present advantages such as artifact suppression, optimization of SNR&#x02014;given that the sensitivity decay with depth position would be avoided&#x02014;access to the highest axial resolution, and decrease in the computational complexity of 3D registration algorithms.</p>
<fig id="F6" position="float">
<label>Figure 6</label>
<caption><p>Proportion of data with residual RMS motion below 4 &#x003BC;m with respect to 8 &#x003BC;m axial resolution for <italic>in vivo</italic> FFOCT imaging, and below 8 &#x003BC;m with respect to 16 &#x003BC;m axial resolution for different acquisition frequencies. Blue and green traces for breath-holding, red and black traces for normal breathing.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-09-868217-g0006.tif"/>
</fig>
</sec>
</sec>
<sec sec-type="conclusions" id="s4">
<title>4. Conclusion</title>
<p>In conclusion, this study has been one of the first attempts to thoroughly quantify and characterize retinal axial motion with high temporal resolution. The results of this experiment show that head motion during breathing, heartbeat and pulsatile blood flow significantly account for involuntary retinal axial displacement in the low-frequency domain below 5 Hz. We also demonstrated that breath holding can help decrease large-and-slow drifts. However, contrary to expectations, when holding breath, axial motion presented a higher speed and higher PSD for temporal frequencies from 7 to 150 Hz, which is not ideal when motion compensation is desired. Further investigation is necessary to understand the mechanism behind this phenomenon. Finally, by simulating the action of an axial motion stabilization control loop, we showed that a loop rate of 1.2 kHz is ideal to achieve 100% robust FFOCT in <italic>in vivo</italic> human retina, and could even pave the way to acquiring dynamic FFOCT <italic>in vivo</italic>. To relax the loop-rate constraint, a practical solution can be the combination of lower axial resolution and LQG control law.</p>
</sec>
<sec sec-type="data-availability" id="s5">
<title>Data Availability Statement</title>
<p>Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.</p>
</sec>
<sec id="s6">
<title>Ethics Statement</title>
<p>The studies involving human participants were reviewed and approved by CPP and ANSM (IDRCB number: 2019-A00942-55). The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.</p>
</sec>
<sec id="s7">
<title>Author Contributions</title>
<p>YC constructed the optical system, collected the data, and drafted the manuscript. YC and PM analyzed the results. PM designed the experiment and developed the algorithm to measure the retinal axial motion. KG and PM provided overall guidance to the project. KG obtained funding to support this research. All authors discussed the results and contributed to the manuscript.</p>
</sec>
<sec sec-type="funding-information" id="s8">
<title>Funding</title>
<p>This work was supported by OPTORETINA [European Research Council (ERC)] (&#x00023;101001841), HELMHOLTZ [European Research Council (ERC)] (&#x00023;610110), IHU FOReSIGHT [ANR-18-IAHU-0001], Region Ile-De-France fund SESAME 4D-EYE [EX047007], Fondation Visio, French state fund CARNOT VOIR ET ENTENDRE [x16-CARN 0029-01].</p>
</sec>
<sec sec-type="COI-statement" id="conf1">
<title>Conflict of Interest</title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="disclaimer" id="s9">
<title>Publisher&#x00027;s Note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
</body>
<back>
<ack><p>The authors thank all participants for their support, and Olivier Thouvenin and Yan Liu for fruitful discussions.</p>
</ack>
<ref-list>
<title>References</title>
<ref id="B1">
<label>1.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>de Kinkelder</surname> <given-names>R</given-names></name> <name><surname>Kalkman</surname> <given-names>J</given-names></name> <name><surname>Faber</surname> <given-names>DJ</given-names></name> <name><surname>Schraa</surname> <given-names>O</given-names></name> <name><surname>Kok</surname> <given-names>PHB</given-names></name> <name><surname>Verbraak</surname> <given-names>FD</given-names></name> <etal/></person-group>. <article-title>Heartbeat-induced axial motion artifacts in optical coherence tomography measurements of the retina</article-title>. <source>Invest Ophthalmol Vis Sci</source>. (<year>2011</year>) <volume>52</volume>:<fpage>3908</fpage>&#x02013;<lpage>13</lpage>. <pub-id pub-id-type="doi">10.1167/iovs.10-6738</pub-id><pub-id pub-id-type="pmid">21467182</pub-id></citation></ref>
<ref id="B2">
<label>2.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ginner</surname> <given-names>L</given-names></name> <name><surname>Kumar</surname> <given-names>A</given-names></name> <name><surname>Fechtig</surname> <given-names>D</given-names></name> <name><surname>Wurster</surname> <given-names>LM</given-names></name> <name><surname>Salas</surname> <given-names>M</given-names></name> <name><surname>Pircher</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Noniterative digital aberration correction for cellular resolution retinal optical coherence tomography <italic>in vivo</italic></article-title>. <italic>Optica</italic>. (<year>2017</year>) <volume>4</volume>:<fpage>924</fpage>&#x02013;<lpage>31</lpage>. <pub-id pub-id-type="doi">10.1364/OPTICA.4.000924</pub-id></citation>
</ref>
<ref id="B3">
<label>3.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Lavia</surname> <given-names>C</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Nassisi</surname> <given-names>M</given-names></name> <name><surname>Bonnin</surname> <given-names>S</given-names></name> <name><surname>Marie-Louise</surname> <given-names>J</given-names></name> <name><surname>Couturier</surname> <given-names>A</given-names></name> <etal/></person-group>. <article-title>Retinal capillary plexus pattern and density from fovea to periphery measured in healthy eyes with swept-source optical coherence tomography angiography</article-title>. <source>Sci Rep</source>. (<year>2020</year>) <volume>10</volume>:<fpage>1</fpage>&#x02013;<lpage>11</lpage>. <pub-id pub-id-type="doi">10.1038/s41598-020-58359-y</pub-id><pub-id pub-id-type="pmid">32001769</pub-id></citation></ref>
<ref id="B4">
<label>4.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Spaide</surname> <given-names>RF</given-names></name> <name><surname>Fujimoto</surname> <given-names>JG</given-names></name> <name><surname>Waheed</surname> <given-names>NK</given-names></name></person-group>. <article-title>Image artifacts in optical coherence angiography</article-title>. <source>Retina</source>. (<year>2015</year>) <volume>35</volume>:<fpage>2163</fpage>. <pub-id pub-id-type="doi">10.1097/IAE.0000000000000765</pub-id><pub-id pub-id-type="pmid">26428607</pub-id></citation></ref>
<ref id="B5">
<label>5.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Beaurepaire</surname> <given-names>E</given-names></name> <name><surname>Boccara</surname> <given-names>AC</given-names></name> <name><surname>Lebec</surname> <given-names>M</given-names></name> <name><surname>Blanchot</surname> <given-names>L</given-names></name> <name><surname>Saint-Jalmes</surname> <given-names>H</given-names></name></person-group>. <article-title>Full-field optical coherence microscopy</article-title>. <source>Opt Lett</source>. (<year>1998</year>) <volume>23</volume>:<fpage>244</fpage>&#x02013;<lpage>6</lpage>. <pub-id pub-id-type="doi">10.1364/OL.23.000244</pub-id><pub-id pub-id-type="pmid">18084473</pub-id></citation></ref>
<ref id="B6">
<label>6.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Barolle</surname> <given-names>V</given-names></name> <name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Chassot</surname> <given-names>JM</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Fink</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Manifestation of aberrations in full-field optical coherence tomography</article-title>. <source>Opt Exp</source>. (<year>2021</year>) <volume>29</volume>:<fpage>22044</fpage>&#x02013;<lpage>65</lpage>. <pub-id pub-id-type="doi">10.1364/OE.419963</pub-id><pub-id pub-id-type="pmid">34265978</pub-id></citation></ref>
<ref id="B7">
<label>7.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jarosz</surname> <given-names>J</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Conan</surname> <given-names>JM</given-names></name> <name><surname>Petit</surname> <given-names>C</given-names></name> <name><surname>Paques</surname> <given-names>M</given-names></name> <name><surname>Meimon</surname> <given-names>S</given-names></name></person-group>. <article-title>High temporal resolution aberrometry in a 50-eye population and implications for adaptive optics error budget</article-title>. <source>Biomed Opt Exp</source>. (<year>2017</year>) <volume>8</volume>:<fpage>2088</fpage>&#x02013;<lpage>105</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.8.002088</pub-id><pub-id pub-id-type="pmid">28736657</pub-id></citation></ref>
<ref id="B8">
<label>8.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Thouvenin</surname> <given-names>O</given-names></name> <name><surname>Fink</surname> <given-names>M</given-names></name> <name><surname>Grieve</surname> <given-names>K</given-names></name> <etal/></person-group>. <article-title>Coherence gate shaping for wide field high-resolution <italic>in vivo</italic> retinal imaging with full-field OCT</article-title>. <source>Biomed Opt Exp</source>. (<year>2020</year>) <volume>11</volume>:<fpage>4928</fpage>&#x02013;<lpage>41</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.400522</pub-id><pub-id pub-id-type="pmid">33014591</pub-id></citation></ref>
<ref id="B9">
<label>9.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Boccara</surname> <given-names>C</given-names></name></person-group>. <article-title>High-resolution <italic>in-vivo</italic> human retinal imaging using full-field OCT with optical stabilization of axial motion</article-title>. <source>Biomed Opt Exp</source>. (<year>2020</year>) <volume>11</volume>:<fpage>492</fpage>&#x02013;<lpage>504</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.381398</pub-id><pub-id pub-id-type="pmid">32010530</pub-id></citation></ref>
<ref id="B10">
<label>10.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Grieve</surname> <given-names>K</given-names></name> <name><surname>Boccara</surname> <given-names>C</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name></person-group>. <article-title>Adaptive-glasses time-domain FFOCT for wide-field high-resolution retinal imaging with increased SNR</article-title>. <source>Opt Lett</source>. (<year>2020</year>) <volume>45</volume>:<fpage>5901</fpage>&#x02013;<lpage>4</lpage>. <pub-id pub-id-type="doi">10.1364/OL.403135</pub-id><pub-id pub-id-type="pmid">33137028</pub-id></citation></ref>
<ref id="B11">
<label>11.</label>
<citation citation-type="book"><person-group person-group-type="author"><name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Grieve</surname> <given-names>K</given-names></name> <name><surname>Boccara</surname> <given-names>C</given-names></name></person-group>. <article-title>Adaptive glasses wavefront sensorless Full-Field OCT for high-resolution retinal imaging over a wide field-of-view</article-title>. In: <source>Ophthalmic Technologies XXXI</source>. Vol. 11623. International Society for Optics and Photonics (<year>2021</year>). p. 1162306. <pub-id pub-id-type="doi">10.1117/12.2576689</pub-id></citation>
</ref>
<ref id="B12">
<label>12.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Baghaie</surname> <given-names>A</given-names></name> <name><surname>Yu</surname> <given-names>Z</given-names></name> <name><surname>D&#x00027;Souza</surname> <given-names>RM</given-names></name></person-group>. <article-title>Involuntary eye motion correction in retinal optical coherence tomography: hardware or software solution?</article-title> <source>Med Image Anal</source>. (<year>2017</year>) <volume>37</volume>:<fpage>129</fpage>&#x02013;<lpage>45</lpage>. <pub-id pub-id-type="doi">10.1016/j.media.2017.02.002</pub-id><pub-id pub-id-type="pmid">28208100</pub-id></citation></ref>
<ref id="B13">
<label>13.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Pircher</surname> <given-names>M</given-names></name> <name><surname>G&#x000F6;tzinger</surname> <given-names>E</given-names></name> <name><surname>Sattmann</surname> <given-names>H</given-names></name> <name><surname>Leitgeb</surname> <given-names>RA</given-names></name> <name><surname>Hitzenberger</surname> <given-names>CK</given-names></name></person-group>. <article-title><italic>In vivo</italic> investigation of human cone photoreceptors with SLO/OCT in combination with 3D motion correction on a cellular level</article-title>. <source>Opt Exp</source>. (<year>2010</year>) <volume>18</volume>:<fpage>13935</fpage>&#x02013;<lpage>44</lpage>. <pub-id pub-id-type="doi">10.1364/OE.18.013935</pub-id><pub-id pub-id-type="pmid">20588526</pub-id></citation></ref>
<ref id="B14">
<label>14.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Goureau</surname> <given-names>O</given-names></name> <name><surname>Sahel</surname> <given-names>JA</given-names></name> <name><surname>Fink</surname> <given-names>M</given-names></name> <name><surname>Reichman</surname> <given-names>S</given-names></name> <etal/></person-group>. <article-title>Dynamic full-field optical coherence tomography: 3D live-imaging of retinal organoids</article-title>. <source>Light Sci Appl</source>. (<year>2020</year>) <volume>9</volume>:<fpage>1</fpage>&#x02013;<lpage>9</lpage>. <pub-id pub-id-type="doi">10.1038/s41377-020-00375-8</pub-id><pub-id pub-id-type="pmid">32864115</pub-id></citation></ref>
<ref id="B15">
<label>15.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Jarosz</surname> <given-names>J</given-names></name> <name><surname>Conan</surname> <given-names>JM</given-names></name> <name><surname>Petit</surname> <given-names>C</given-names></name> <name><surname>Grieve</surname> <given-names>K</given-names></name> <name><surname>Paques</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Fixational eye movement: a negligible source of dynamic aberration</article-title>. <source>Biomed Opt Exp</source>. (<year>2018</year>) <volume>9</volume>:<fpage>717</fpage>&#x02013;<lpage>27</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.9.000717</pub-id><pub-id pub-id-type="pmid">29552407</pub-id></citation></ref>
<ref id="B16">
<label>16.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Zhang</surname> <given-names>M</given-names></name> <name><surname>Gofas-Salas</surname> <given-names>E</given-names></name> <name><surname>Leonard</surname> <given-names>BT</given-names></name> <name><surname>Rui</surname> <given-names>Y</given-names></name> <name><surname>Snyder</surname> <given-names>VC</given-names></name> <name><surname>Reecher</surname> <given-names>HM</given-names></name> <etal/></person-group>. <article-title>Strip-based digital image registration for distortion minimization and robust eye motion measurement from scanned ophthalmic imaging systems</article-title>. <source>Biomed Opt Exp</source>. (<year>2021</year>) <volume>12</volume>:<fpage>2353</fpage>&#x02013;<lpage>72</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.418070</pub-id><pub-id pub-id-type="pmid">33996234</pub-id></citation></ref>
<ref id="B17">
<label>17.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Schuitema</surname> <given-names>K</given-names></name> <name><surname>Holm</surname> <given-names>B</given-names></name></person-group>. <article-title>The role of different facial areas in eliciting human diving bradycardia</article-title>. <source>Acta Physiol Scand</source>. (<year>1988</year>) <volume>132</volume>:<fpage>119</fpage>&#x02013;<lpage>20</lpage>. <pub-id pub-id-type="doi">10.1111/j.1748-1716.1988.tb08306.x</pub-id><pub-id pub-id-type="pmid">3223302</pub-id></citation></ref>
<ref id="B18">
<label>18.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Delahoche</surname> <given-names>J</given-names></name> <name><surname>Delapille</surname> <given-names>P</given-names></name> <name><surname>Lema&#x000EE;tre</surname> <given-names>F</given-names></name> <name><surname>Verin</surname> <given-names>E</given-names></name> <name><surname>Tourny-Chollet</surname> <given-names>C</given-names></name></person-group>. <article-title>The role of different facial areas in eliciting human diving bradycardia</article-title>. <source>Int J Sports Med</source>. (<year>2005</year>) <volume>26</volume>:<fpage>177</fpage>&#x02013;<lpage>81</lpage>. <pub-id pub-id-type="doi">10.1055/s-2004-820976</pub-id><pub-id pub-id-type="pmid">3223302</pub-id></citation></ref>
<ref id="B19">
<label>19.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Puyo</surname> <given-names>L</given-names></name> <name><surname>Paques</surname> <given-names>M</given-names></name> <name><surname>Fink</surname> <given-names>M</given-names></name> <name><surname>Sahel</surname> <given-names>JA</given-names></name> <name><surname>Atlan</surname> <given-names>M</given-names></name></person-group>. <article-title>Waveform analysis of human retinal and choroidal blood flow with laser Doppler holography</article-title>. <source>Biomed Opt Exp</source>. (<year>2019</year>) <volume>10</volume>:<fpage>4942</fpage>&#x02013;<lpage>63</lpage>. <pub-id pub-id-type="doi">10.1364/BOE.10.004942</pub-id><pub-id pub-id-type="pmid">31646021</pub-id></citation></ref>
<ref id="B20">
<label>20.</label>
<citation citation-type="book"><person-group person-group-type="author"><name><surname>Roddier</surname> <given-names>F</given-names></name></person-group>. <source>Adaptive Optics in Astronomy</source>. Cambridge University Press (<year>1999</year>). <pub-id pub-id-type="doi">10.1017/CBO9780511525179</pub-id></citation>
</ref>
<ref id="B21">
<label>21.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gofas-Salas</surname> <given-names>E</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Petit</surname> <given-names>C</given-names></name> <name><surname>Jarosz</surname> <given-names>J</given-names></name> <name><surname>Mugnier</surname> <given-names>LM</given-names></name> <name><surname>Bonnefois</surname> <given-names>AM</given-names></name> <etal/></person-group>. <article-title>High loop rate adaptive optics flood illumination ophthalmoscope with structured illumination capability</article-title>. <source>Appl Opt</source>. (<year>2018</year>) <volume>57</volume>:<fpage>5635</fpage>&#x02013;<lpage>42</lpage>. <pub-id pub-id-type="doi">10.1364/AO.57.005635</pub-id><pub-id pub-id-type="pmid">30118075</pub-id></citation></ref>
<ref id="B22">
<label>22.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Scholler</surname> <given-names>J</given-names></name></person-group>. <article-title>Motion artifact removal and signal enhancement to achieve <italic>in vivo</italic> dynamic full field OCT</article-title>. <source>Opt Exp</source>. (<year>2019</year>) <volume>27</volume>:<fpage>19562</fpage>&#x02013;<lpage>72</lpage>. <pub-id pub-id-type="doi">10.1364/OE.27.019562</pub-id><pub-id pub-id-type="pmid">31503714</pub-id></citation></ref>
<ref id="B23">
<label>23.</label>
<citation citation-type="book"><person-group person-group-type="author"><name><surname>Scholler</surname> <given-names>J</given-names></name> <name><surname>Mec&#x000EA;</surname> <given-names>P</given-names></name> <name><surname>Groux</surname> <given-names>K</given-names></name> <name><surname>Mazlin</surname> <given-names>V</given-names></name> <name><surname>Boccara</surname> <given-names>C</given-names></name> <name><surname>Grieve</surname> <given-names>K</given-names></name></person-group>. <article-title>Motion artifact removal and signal enhancement to achieve <italic>in vivo</italic> dynamic full field OCT (Conference Presentation)</article-title>. In: <source>Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIV</source>. <publisher-loc>Vol. 11228. San Francisco, CA</publisher-loc>: <publisher-name>International Society for Optics and Photonics</publisher-name> (<year>2020</year>). p. 112281N. <pub-id pub-id-type="doi">10.1117/12.2542635</pub-id></citation>
</ref>
<ref id="B24">
<label>24.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sivo</surname> <given-names>G</given-names></name> <name><surname>Kulcs&#x000E1;r</surname> <given-names>C</given-names></name> <name><surname>Conan</surname> <given-names>JM</given-names></name> <name><surname>Raynaud</surname> <given-names>HF</given-names></name> <name><surname>Gendron</surname> <given-names>&#x000C9;</given-names></name> <name><surname>Basden</surname> <given-names>A</given-names></name> <etal/></person-group>. <article-title>First on-sky SCAO validation of full LQG control with vibration mitigation on the CANARY pathfinder</article-title>. <source>Opt Exp</source>. (<year>2014</year>) <volume>22</volume>:<fpage>23565</fpage>&#x02013;<lpage>91</lpage>. <pub-id pub-id-type="doi">10.1364/OE.22.023565</pub-id><pub-id pub-id-type="pmid">25321824</pub-id></citation></ref>
</ref-list> 
</back>