Skip to main content

ORIGINAL RESEARCH article

Front. Pediatr., 24 November 2023
Sec. Pediatric Pulmonology
Volume 11 - 2023 | https://doi.org/10.3389/fped.2023.1258846

Comparison of clinical outcomes following delivery of budesonide by both vibrating mesh nebulizer and jet nebulizer in premature infants with bronchopulmonary dysplasia

Jian-Fu Zhou1,2* Yi-Bing Zhang1,2 Zhi-Wei Zhang1,2
  • 1Fujian Medical University, Fuzhou, China
  • 2Department of Neonatology, The Affiliated Hospital of Putian University, Putian, China

Background: This study aimed to compare the efficacy of budesonide inhalation suspension administered via a vibrating mesh nebulizer vs. a jet nebulizer in the treatment of premature infants with bronchopulmonary dysplasia (BPD) undergoing high-frequency oscillatory ventilation (HFOV).

Methods: Between July 2020 and July 2022, we retrospectively analyzed the medical records of 36 preterm infants diagnosed with BPD who underwent HFOV. Based on the nebulizer type used, infants were categorized into the vibrating mesh nebulizer group (VMN group) or the jet nebulizer group (JN group). Post-nebulization outcomes, such as the duration of mechanical ventilation, length of stay in the neonatal intensive care unit (NICU), ventilator-associated parameters, and arterial blood gas metrics, were compared between the two groups. Treatment-associated complications were also documented.

Results: No significant differences were noted between the VMN and JN groups in terms of mechanical ventilation duration (p = 0.519), NICU length of stay (p = 0.112), ventilator-associated parameters, or complications (p = 0.700). However, after 2 weeks of treatment, the oxygenation index (p = 0.012) and arterial partial pressure of carbon dioxide (p = 0.006) were more favorable in the VMN group compared to the JN group.

Conclusion: Among premature infants with BPD on HFOV, for administration of budesonide inhalation suspension resulted in an improved oxygenation index and reduced arterial partial pressure of carbon dioxide when compared to a jet nebulizer, indicating superior therapeutic efficacy.

Introduction

With advancements in perinatal medicine and neonatal monitoring technologies, survival rates for premature and low birth weight infants have significantly improved. However, concurrently, we've observed a rise in complications like bronchopulmonary dysplasia (BPD) (13). BPD remains a predominant respiratory complication among premature infants, commonly observed in those subjected to prolonged oxygen therapy and/or mechanical ventilation (4, 5). Factors like reduced gestational age, low birth weight, mechanical ventilation, and exposure to high concentrations of inhaled oxygen contribute to lung tissue damage. This damage potentially hampers neonatal lung development and leads to decreased lung function (69). Infants diagnosed with BPD often demand extended hospital stays and might encounter respiratory and feeding challenges, cognitive impairments, and risks of cerebral palsy post-discharge. Hence, proactive prevention and treatment become crucial in managing this condition (10). Current therapeutic modalities for BPD include surfactant supplementation, non-invasive or invasive mechanical ventilation, infection prevention, nutritional management, and symptomatic treatments (1113).

In clinical settings, nebulized glucocorticoids have gained traction due to their efficacious outcomes and controlled dosing. They can swiftly inhibit inflammatory responses, alleviating clinical symptoms in patients (14, 15). Budesonide, with its potent anti-inflammatory properties and lung tissue-targeting capabilities, stands out as a preferred choice (14). Ehrmann's study showed that 99% of respondents working in intensive care units used aerosol therapy during mechanical ventilation, but this finding was not specific to premature infants (15). There's a paucity of research on nebulizer therapy during high-frequency oscillatory ventilation (HFOV), especially in the context of premature infants with BPD. From an aerosol delivery perspective, modeling these patients in bench studies to ascertain metrics like bias flow and drug deposition rate remains challenging. The clinical efficacy can vary significantly based on the aerosol generator device used. Specifically, in neonatal HFOV models, vibrating mesh nebulizers (VMNs) have demonstrated the delivery of inhaled doses over 80 times higher than jet nebulizers (JNs), highlighting a substantial difference in aerosol generation efficiency between these two devices (16). Historically, these devices struggled to accommodate the small tidal volumes, rapid breathing rates, and unfavorable I:E ratios. Thus, this study aims to discern any differences in therapeutic outcomes between the VMN and the JN when used during HFOV in preterm infants diagnosed with BPD.

Materials and methods

In this retrospective study, we enrolled 36 premature infants diagnosed with BPD who underwent HFOV at our institution between July 2020 and July 2022. The SLE6000 high-frequency oscillatory ventilator (SLE, Croydon, UK) was used for HFOV. All patients' families provided informed consent. The study adhered to the Declaration of Helsinki as set by the World Medical Association and received approval from the hospital's ethics committee.

Inclusion criteria: gestational age ≤32 weeks; immediate admission to the neonatal intensive care unit (NICU) post-birth; underwent HFOV with endotracheal intubation within the first 7 days post-birth; diagnosis consistent with BPD criteria; informed consent from the patient's family. Exclusion criteria: suspected chromosomal abnormalities; significant congenital malformations (e.g., cyanotic congenital heart disease, congenital diaphragmatic hernia); administration of neuromuscular relaxants or history of neuromuscular disease; presence of pulmonary hemorrhage or intracranial hemorrhage upon admission; administration of other types of glucocorticoids during treatment; unilateral discharge or discontinuation of treatment; refusal to participate in the study.

The primary outcome measure was the arterial partial pressure of carbon dioxide (PaCO2). Based on literature review findings, the PaCO2 for the VMN group was recorded as 44.8, and for the JN group, it was 49.2. With a two-sided α set at 0.05 and a power of 90%, the calculated sample size, using PASS 15.0 software, was N = 18 for each group, leading to a total sample size of 36 subjects.

From July 2020 to June 2021, 18 infants were treated using a jet nebulizer, designated as the JN group. The jet nebulizer (BaiRui, Inhaler pro compression nebulizer, Shanghai, China) was attached to the ventilator suction line via a T-tube for aerosol administration. From July 2021 to July 2022, another 18 infants received treatment via a vibrating mesh nebulizer, forming the VMN group. This vibrating mesh nebulizer (Aerogen Pro®, in combination with the Aerogen USB controller, Aerogen Ltd., Galway, Ireland) was connected to the ventilator suction line using a curved pipe for aerosol delivery. The nebulizer was placed between the endotracheal tube and the ventilator circuit (Figure 1). Patients were administered 1 ml of budesonide suspension (specification: 2 ml: 1 mg, AstraZeneca Co., Ltd., approval No. LOT325384) for inhalation twice daily, with a total treatment duration of two weeks.

FIGURE 1
www.frontiersin.org

Figure 1. The vibrating mesh nebulizer or jet nebulizer was placed between the endotracheal tube and the ventilator circuit.

Upon admission, all patients received standard care for preterm infants, encompassing core interventions like thermoregulation, nutritional support, infection prevention, airway patency, and water and electrolyte balance maintenance. Continuous monitoring of vital signs and blood gas analysis was implemented. The same team of NICU physicians and nurses oversaw the care for all patients. Decisions regarding the timing for endotracheal intubation and extubation were at the discretion of the attending NICU physician. Every patient was managed using the SLE 5,000 ventilator (Specialized Laboratory Equipment Ltd., South Croydon, UK) set in the HFOV mode. The ventilator's initial parameters were set as: FiO2 ranging from 50% to 100%, frequency between 8 and 12 Hz, an inhalation/exhalation ratio of 1:1, mean airway pressure (MAP) of 10–15 cmH2O, and amplitude within 30–40 cmH2O. The MAP was increased by 1 cmH2O every 2–3 min until there was no further improvement in arterial oxygen saturation. Subsequently, the MAP was decreased by 1–2 cmH2O every 2–3 min until a decline in arterial oxygen saturation was noted, then raised again by 1–2 cmH2O. Visible chest movement was confirmed by bedside chest x-ray conducted an hour post-HFOV, indicating the right diaphragm's positioning at the 8th–9th posterior rib level. Throughout HFOV, ventilator parameters were adjusted based on arterial blood gas analysis findings. Criteria for extubation included: Stable hemodynamics and adequate spontaneous respiratory function post-discontinuation of sedatives. Ventilator support parameters demonstrating FiO2 ≤ 40% and MAP ≤ 8 cmH2O. Arterial blood gas (ABG) results indicating PaCO2 < 50 mmHg, PaO2 > 70 mmHg, pH >7.30, and lactic acid levels below 2.0 mmol/L. Post-extubation, all patients transitioned to nasal continuous positive airway pressure (nCPAP) support.

The primary outcomes of interest were arterial blood gas measurements and the oxygenation index. Arterial blood samples were drawn from all patients both pre-treatment and at 2 weeks post-treatment, and then analyzed using the ABL90 FLEX blood gas analyzer from the USA. Results provided values for the arterial partial pressure of oxygen (PaO2) and arterial partial pressure of carbon dioxide (PaCO2), from which the oxygenation index (PaO2/FiO2) was calculated. Secondary outcomes comprised: (1) demographic data: gender, gestational age, birth weight, and the 5-min Apgar score were documented for both groups. (2) ventilator-related parameters: FiO2, tidal volume, frequency, and MAP. (3) clinical outcomes: both the duration of mechanical ventilation—from the initiation of endotracheal intubation to extubation—and the total length of NICU stay were tracked. (4) complications: incidence rates of ventilator-associated pneumonia (VAP), pneumothorax, retinopathy of prematurity (ROP), intraventricular hemorrhage (IVH), and necrotizing enterocolitis (NEC) were documented.

For BPD diagnosis, we adhered to the diagnostic criteria and severity classifications put forth by the National Institute of Child Health and Human Development in 2000. BPD was characterized by the need for oxygen concentrations >21% for a minimum of 28 days (17). The severity of BPD was determined either at 36 weeks postmenstrual age or upon discharge: mild BPD required no supplemental oxygen, moderate BPD necessitated <30% oxygen, and severe BPD required ≥30% oxygen and/or positive pressure ventilation. The diagnosis of VAP was grounded in the diagnostic criteria specified for infants under 1 year of age by the U.S. (18). Centers for Disease Control and Prevention. Pneumothorax was identified through bedside chest x-rays. The revised 2005 Classification of Retinopathy of Prematurity was used to diagnose ROP (19). IVH was detected via bedside cranial ultrasounds (20). Lastly, necrotizing enterocolitis was diagnosed based on the modified Bell's staging criteria (21).

Statistical analysis

Data were processed using SPSS 26.0 software. To determine the normality of data distribution, we carried out a Kolmogorov–Smirnov test before any subsequent analyses. For normally distributed data, t-test analyses were utilized; if not, the Wilcoxon-Mann–Whitney test was applied. Qualitative data between both groups were compared using the Chi-square test. When the conditions for the χ2 test weren't met, we employed Fisher's exact test. A P value less than 0.05 was deemed statistically significant.

Results

In this study, 36 neonates were included. They were categorized based on the nebulizer type into the VMN group (n = 18) and the JN group (n = 18). Pre-treatment, the PaO2 (p = 0.889), PaCO2 (p = 0.368), and oxygenation index (p = 0.170) didn't show significant differences between both groups. However, post two-week treatment, both groups saw significant improvements in PaO2 when compared to their initial values. Yet, there wasn't any noticeable difference in post-treatment PaO2 between the groups (p = 0.388). After the two-week treatment, the VMN group had a notably lower PaCO2 (p = 0.006) and a higher oxygenation index (p = 0.012) compared to the JN group (Table 1).

TABLE 1
www.frontiersin.org

Table 1. Primary outcome.

The gender (p = 0.738), gestational age (p = 0.213), birth weight (p = 0.138), and 5-min Apgar scores (p = 0.074) were similar between the two groups, ensuring the general characteristics of the cohorts were comparable (Table 2). When considering ventilator-related metrics like FiO2, tidal volume, frequency, and MAP, no significant difference was found between the two groups both before and after the two-week treatment (Table 3). Additionally, the duration of mechanical ventilation (p = 0.519) and the length of the NICU stay (p = 0.112) did not significantly differ between the two groups (Table 4).

TABLE 2
www.frontiersin.org

Table 2. Comparison of general data.

TABLE 3
www.frontiersin.org

Table 3. Comparison of ventilator related parameters.

TABLE 4
www.frontiersin.org

Table 4. Comparison of other outcomes.

Discussion

There is an inverse relationship between the occurrence of BPD and clinical parameters like gestational age and birth weight. Ultra-premature infants (those with a gestational age below 28 weeks) exhibit BPD incidence rates ranging from 48%–68% (22, 23). For preterm infants weighing between 500 and 1,500 g, the BPD incidence is approximately 22%, with those weighing between 500 and 750 g experiencing a staggering 42% incidence (24, 25). Our study encompassed 36 preterm infants with gestational ages spanning 23–35 weeks and weights ranging from 0.89–2.81 kg. Predominant causes leading to preterm births among these infants included maternal gestational diabetes mellitus, hypertension, and premature rupture of membranes, among others. Post-birth, all these infants necessitated endotracheal intubation and mechanical ventilation, exhibiting varying degrees of oxygen dependency. Based on the BPD diagnostic and severity criteria, all participants in this study were categorized as having severe BPD.

Glucocorticoids, due to their potent anti-inflammatory and immunomodulatory properties, have become a staple in BPD prevention and treatment. These properties allow glucocorticoids to minimize pulmonary edema, decrease the physiological demand for oxygen, and abbreviate the duration of mechanical ventilation (26, 27). The administration of glucocorticoid via aerosol inhalation, targeting the airway mucosa directly, accentuates its localized respiratory effects. Notably, it substantially curtails BPD incidence and the duration of oxygen utilization (28). Several factors can influence the efficacy of nebulization, with the choice of drug being paramount. For consistency, this study solely used budesonide suspension for aerosol inhalation in all participating infants. Previous research has consistently shown that budesonide is effective in preventing and treating BPD in premature infants (14, 29, 30). It not only reduces the incidence and complications of BPD but also improves blood gas metrics, curtails inflammation, and enhances pulmonary function in these infants. By exclusively using budesonide for nebulization, this study aimed to optimize drug-related efficacy while reducing potential confounding factors. Using a standardized therapeutic agent enabled a clearer evaluation of the nebulization's impact on patient outcomes.

The Aerogen Pro electronic nebulizer employs a sophisticated micropump system to produce precise aerosol generation. Its OnQTM aerosol generator features a vibrating membrane punctured by roughly 1,000 tapered apertures. When vibrated at ultrasonic frequencies, fluid is forced through these openings, creating controlled streams that fragment into optimized droplets. The diameter of the holes meticulously adjusts droplet size, yielding tailored particle distributions with minimal residual volumes in the microliter range. On the other hand, jet nebulizers, frequently used for aerosol delivery in intubated neonates, have shown inefficiencies. Research has highlighted notable performance disparities between different jet nebulizer brands (31). Inefficiencies largely arise from hygroscopic growth, where particles expand due to interactions with moist gas, and from drug losses in the expiratory limb during continuous flow. Furthermore, the narrow endotracheal tubes used in newborns present significant challenges to effective drug deposition.

Following two weeks of nebulizer treatment, there was a marked decrement in PaCO2 levels across both groups. The magnitude of PaCO2 decline in the VMN group surpassed that of the JN group. Concurrently, the oxygenation index was substantially elevated in the VMN group compared to the JN cohort. This insinuates that pulmonary ventilation function enhancement was more pronounced in the VMN group, thus facilitating more efficient carbon dioxide excretion. The enhanced performance of the vibrating mesh nebulizer may be attributed to its ability to produce a consistent, uniform aerosol with an average particle diameter ranging from 1–5 μm. This ensures efficient drug deposition deep into the respiratory tree, especially in the distal 5–6 airways, maximizing drug delivery efficacy (32). Numerous in vitro investigations have corroborated that the vibrating mesh nebulizers outperform both jet and ultrasonic nebulizers in terms of aerosol drug delivery rate and lung deposition (3335). Dubus et al.'s study on neonatal mechanically ventilated rhesus monkeys unveiled that lung deposition rate using the vibrating mesh nebulizer was an astounding 20-fold higher than with jet nebulizers (36). In our study, to eliminate potential variations from nebulizer placement, both the vibrating mesh and jet nebulizers were strategically positioned between the endotracheal tube and the ventilator circuit.

Pre-treatment comparative analysis in our study revealed no significant disparities between the two patient groups in terms of general data, initial therapeutic regimes, and requisite ventilator parameters. Furthermore, baseline ventilator parameters (encompassing FiO2, tidal volume, frequency, and MAP) demonstrated equivalence across the cohorts. Post two-week nebulizer intervention, there was a noticeable decline in the ventilator parameters for both groups. This decline, albeit more pronounced in the VMN group than the JN group, did not reach statistical significance. Yet, the data suggests a reduced dependence on ventilator pressure and oxygen in the VMN cohort. The implications of these findings underscore the potential benefits of the VMN group and beckon larger sample, extended duration studies in the future.

The duration of mechanical ventilation and the length of NICU stay in the VMN group were slightly shorter than those in the JN group, but the difference was not statistically significant. Mechanical ventilation is essential for the prevention and treatment of BPD in premature infants. However, its use can increase the risk of respiratory tract infection and potential brain tissue injury. A ventilator-associated infection, like VAP, can worsen lung function in premature infants within a year. In our study, the incidence of VAP between the two groups was low, with no significant difference. Additionally, the use of a vibrating mesh nebulizer did not elevate the risk of complications such as pneumothorax, ROP, IVH, and NEC. This is consistent with findings from Dugernier and Pitance et al. (37, 38).

A significant novel finding from our study is the evident superior clinical outcomes using the vibrating mesh nebulizer for budesonide aerosol delivery compared to jet nebulizer administration in patients with bronchopulmonary dysplasia. To the best of our knowledge, this represents the first direct evidence of improved patient outcomes with VMN-administered budesonide therapy. We theorize that the VMN's markedly increased aerosol output and inhaled dose delivery are foundational to its therapeutic advantages—a hypothesis aligned with bench studies that show an over 80-fold higher dose with VMN in mechanical ventilation settings. While previous data hinted at the superior efficacy of VMN, our research distinctively underscores the translational potential of VMN aerosol generators in enhancing clinical responses compared to jet nebulizers.

In addition, previous studies indicated that vibrating mesh nebulizers were more effective in treating children with asthma and severe pneumonia (31, 39). DiBlasi et al. reached a similar conclusion in a vitro study using an infant model (40). Future studies should investigate if these findings apply to other respiratory conditions. VMN might offer increased advantages for patients who need extended mechanical ventilation support, like infants with ARDS. Prior studies on VMN have incorporated in vitro models, rhesus monkey models, and infants. Future research could further explore the use of VMN in children.

However, it's essential to acknowledge the limitations inherent in our study. Being a retrospective investigation from a single center, it lacked randomization and involved a restricted patient cohort. We recognize the need for larger, multi-center, randomized studies to further validate and enhance the generalizability of our findings. Furthermore, the uniform dosage of budesonide across both groups, without accounting for possible variances in drug delivery due to the distinct inhalation mechanisms, might have influenced our results. Additionally, keeping the ventilation parameters consistent during nebulization might have influenced pulmonary drug deposition and the effects of treatment. Lastly, while our study shed light on short-term outcomes associated with the vibrating mesh nebulizer, the long-term clinical ramifications remain to be elucidated.

Conclusions

In sum, for premature infants with BPD on HFOV, the employment of a vibrating mesh nebulizer for budesonide inhalation suspension, relative to a jet nebulizer, seems to offer tangible therapeutic benefits by effectively lowering PaCO2, enhancing the oxygenation index, and thus presenting a preferable therapeutic strategy.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by the ethics committee of The Affiliated Hospital of Putian University. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation in this study was provided by the participants’ legal guardians/next of kin. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

JZ: Conceptualization, Investigation, Writing – original draft, Writing – review & editing. YZ: Data curation, Writing – review & editing. ZZ: Data curation, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Acknowledgments

We gratefully acknowledge all the pediatric doctors who contributed to this study.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Logan JW, Lynch SK, Curtiss J, Shepherd EG. Clinical phenotypes and management concepts for severe, established bronchopulmonary dysplasia. Paediatr Respir Rev. (2019) 31:58–63. doi: 10.1016/j.prrv.2018.10.004

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Hysinger EB, Friedman NL, Padula MA, Shinohara RT, Zhang H, Panitch HB, et al. Tracheobronchomalacia is associated with increased morbidity in bronchopulmonary dysplasia. Ann Am Thorac Soc. (2017) 14(9):1428–35. doi: 10.1513/AnnalsATS.201702-178OC

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Capasso L, Vento G, Loddo C, Tirone C, Iavarone F, Raimondi F, et al. Oxidative stress and bronchopulmonary dysplasia: evidences from microbiomics, metabolomics, and proteomics. Front Pediatr. (2019) 7:30. doi: 10.3389/fped.2019.00030

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Poindexter BB, Feng R, Schmidt B, Aschner JL, Ballard RA, Hamvas A, et al. Comparisons and limitations of current definitions of bronchopulmonary dysplasia for the prematurity and respiratory outcomes program. Ann Am Thorac Soc. (2015) 12(12):1822–30. doi: 10.1513/AnnalsATS.201504-218OC

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Collaco JM, McGrath-Morrow SA. Respiratory phenotypes for preterm infants, children, and adults: bronchopulmonary dysplasia and more. Ann Am Thorac Soc. (2018) 15(5):530–8. doi: 10.1513/AnnalsATS.201709-756FR

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Alonso AS, Diaz SP, Soto RS, Avila-Alvarez A. Epidemiology and risk factors for bronchopulmonary dysplasia in preterm infants born at or less than 32 weeks of gestation. An Pediatr (Engl Ed). (2022) 96(3):242–51. doi: 10.1016/j.anpedi.2021.03.002

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Speer CP. Pulmonary inflammation and bronchopulmonary dysplasia. J Perinatol. (2006) 26(Suppl 1):S57–62; discussion S63–4. doi: 10.1038/sj.jp.7211476

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Cokyaman T, Kavuncuoglu S. Bronchopulmonary dysplasia frequency and risk factors in very low birth weight infants: a 3-year retrospective study. North Clin Istanb. (2019) 7(2):124–30. doi: 10.14744/nci.2019.23427

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Brener Dik PH, Niño Gualdron YM, Galletti MF, Cribioli CM, Mariani GL. Bronchopulmonary dysplasia: incidence and risk factors. Arch Argent Pediatr. (2017) 115(5):476–82. doi: 10.5546/aap.2017.eng.476

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Wang C, Jiang ZD. Brainstem auditory abnormality in extremely premature babies and the impact of neonatal bronchopulmonary dysplasia. Acta Obstet Gynecol Scand. (2018) 97(5):545–51. doi: 10.1111/aogs.13312

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Baker CD. Chronic respiratory failure in bronchopulmonary dysplasia. Pediatr Pulmonol. (2021) 56(11):3490–8. doi: 10.1002/ppul.25360

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Sakaria RP, Dhanireddy R. Pharmacotherapy in bronchopulmonary dysplasia: what is the evidence? Front Pediatr. 2022;10:820259. doi: 10.3389/fped.2022.820259

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Venkataraman R, Kamaluddeen M, Hasan SU, Robertson HL, Lodha A. Intratracheal administration of budesonide-surfactant in prevention of bronchopulmonary dysplasia in very low birth weight infants: a systematic review and meta-analysis. Pediatr Pulmonol. (2017) 52(7):968–75. doi: 10.1002/ppul.23680

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Jimenez J, Richter J, Toelen J, Deprest J. Prenatal interventions to prevent bronchopulmonary dysplasia in animal models: a systematic review. J Matern Fetal Neonatal Med. (2016) 29(16):2555–62. doi: 10.3109/14767058.2015.1094789

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Ehrmann S, Roche-Campo F, Sferrazza Papa GF, Isabey D, Brochard L, Apiou-Sbirlea G, et al. Aerosol therapy during mechanical ventilation: an international survey. Intensive Care Med. (2013) 39(6):1048–56. doi: 10.1007/s00134-013-2872-5

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Fang TP, Ling HL, Chiu SH, Tsai YH, Fink JB. Aerosol delivery using jet nebulizer and vibrating mesh nebulizer during high frequency oscillatory ventilation: an in vitro comparison. JAMPDD. (2016) 29(5):447–53. doi: 10.1089/jamp.2015.1265

CrossRef Full Text | Google Scholar

17. Jobe AH, Bancalari E. Bronchopulmonary dysplasia. Am J Respir Crit Care Med. (2001) 163(7):1723–9. doi: 10.1164/ajrccm.163.7.2011060

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Cernada M, Brugada M, Golombek S, Vento M. Ventilator-associated pneumonia in neonatal patients: an update. Neonatology. (2014) 105(2):98–107. doi: 10.1159/000355539

PubMed Abstract | CrossRef Full Text | Google Scholar

19. International Committee for the Classification of Retinopathy of Prematurity. The international classification of retinopathy of prematurity revisited. Arch Ophthalmol. (2005) 123(7):991–9. doi: 10.1001/archopht.123.7.991

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Yogendrakumar V, Ramsay T, Fergusson DA, Demchuk AM, Aviv RI, Rodriguez-Luna D, et al. Redefining hematoma expansion with the inclusion of intraventricular hemorrhage growth. Stroke. (2020) 51(4):1120–7. doi: 10.1161/STROKEAHA.119.027451

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Heida FH, Stolwijk L, Loos M-LHJ, van den Ende SJ, Onland W, van den Dungen FAM, et al. Increased incidence of necrotizing enterocolitis in The Netherlands after implementation of the new Dutch guideline for active treatment in extremely preterm infants: results from three academic referral centers. J Pediatr Surg. (2017) 52(2):273–6. doi: 10.1016/j.jpedsurg.2016.11.024

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Stoll BJ, Hansen NI, Bell EF, Shankaran S, Laptook AR, Walsh MC, et al. Neonatal outcomes of extremely preterm infants from the NICHD neonatal research network. Pediatrics. (2010) 126(3):443–56. doi: 10.1542/peds.2009-2959

PubMed Abstract | CrossRef Full Text | Google Scholar

23. McEvoy CT, Jain L, Schmidt B, Abman S, Bancalari E, Aschner JL. Bronchopulmonary dysplasia: nHLBI workshop on the primary prevention of chronic lung diseases. Ann Am Thorac Soc. (2014) 11 Suppl 3(Suppl 3):S146–53. doi: 10.1513/AnnalsATS.201312-424LD

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Varga P, Berecz B, Pete B, Kollár T, Magyar Z, Jeager J, et al. Trends in mortality and morbidity in infants under 500 grams birthweight: observations from our neonatal intensive care unit (NICU). Med Sci Monit. (2018) 24:4474–80. doi: 10.12659/MSM.907652

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Fanaroff AA, Stoll BJ, Wright LL, Carlo WA, Ehrenkranz RA, Stark AR, et al. Trends in neonatal morbidity and mortality for very low birthweight infants. Am J Obstet Gynecol. (2007) 196(2):147.e1–e8. doi: 10.1016/j.ajog.2006.09.014

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Doyle LW, Davis PG, Morley CJ, McPhee A, Carlin JB, DART Study Investigators. Low-dose dexamethasone facilitates extubation among chronically ventilator-dependent infants: a multicenter, international, randomized, controlled trial. Pediatrics. (2006) 117(1):75–83. doi: 10.1542/peds.2004-2843

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Owen LS, Manley BJ, Davis PG, Doyle LW. The evolution of modern respiratory care for preterm infants. Lancet. (2017) 389(10079):1649–59. doi: 10.1016/S0140-6736(17)30312-4

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Nelin LD, Logan JW. The use of inhaled corticosteroids in chronically ventilated preterm infants. Semin Fetal Neonatal Med. (2017) 22(5):296–301. doi: 10.1016/j.siny.2017.07.005

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Yao Y, Zhang G, Wang F, Wang M. Efficacy of budesonide in the prevention and treatment of bronchopulmonary dysplasia in premature infants and its effect on pulmonary function. Am J Transl Res. (2021) 13(5):4949–58.34150079

PubMed Abstract | Google Scholar

30. Patel RM. Does early treatment with inhaled budesonide prevent death or bronchopulmonary dysplasia in extremely preterm infants? Acta Paediatr. (2017) 106:683. doi: 10.1111/apa.13718

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Moody GB, Luckett PM, Shockley CM, Huang R, Ari A. Clinical efficacy of vibrating mesh and jet nebulizers with different interfaces in pediatric subjects with asthma. Respir Care. (2020) 65(10):1451–63. doi: 10.4187/respcare.07538

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Galindo-Filho VC, Ramos ME, Rattes CSF, Barbosa AK, Brandão DC, Brandão SCS, et al. Radioaerosol pulmonary deposition using mesh and jet nebulizers during noninvasive ventilation in healthy subjects. Respir Care. (2015) 60(9):1238–46. doi: 10.4187/respcare.03667

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Sidler-Moix A-L, Di Paolo ER, Dolci U, Berger-Gryllaki M, Cotting J, Pannatier A. Physicochemical aspects and efficiency of albuterol nebulization: comparison of three aerosol types in an in vitro pediatric model. Respir Care. (2015) 60(1):38–46. doi: 10.4187/respcare.02490

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Sidler-Moix A-L, Dolci U, Berger-Gryllaki M, Pannatier A, Cotting J, Di Paolo ER. Albuterol delivery in an in vitro pediatric ventilator lung model: comparison of jet, ultrasonic, and mesh nebulizers. Pediatr Crit Care Med. (2013) 14(2):e98–102. doi: 10.1097/PCC.0b013e3182712783

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Michotte J-B, Jossen E, Roeseler J, Liistro G, Reychler G. In vitro comparison of five nebulizers during noninvasive ventilation: analysis of inhaled and lost doses. J Aerosol Med Pulm Drug Deliv. (2014) 27(6):430–40. doi: 10.1089/jamp.2013.1070

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Dubus JC, Vecellio L, De Monte M, Fink JB, Grimbert D, Montharu J, et al. Aerosol deposition in neonatal ventilation. Pediatr Res. (2005) 58(1):10–4. doi: 10.1203/01.PDR.0000156244.84422.55

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Dugernier J, Reychler G, Wittebole X, Roeseler J, Depoortere V, Sottiaux T, et al. Aerosol delivery with two ventilation modes during mechanical ventilation: a randomized study. Ann Intensive Care. (2016) 6(1):73. doi: 10.1186/s13613-016-0169-x

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Pitance L, Vecellio L, Leal T, Reychler G, Reychler H, Liistro G. Delivery efficacy of a vibrating mesh nebulizer and a jet nebulizer under different configurations. J Aerosol Med Pulm Drug Deliv. (2010) 23(6):389–96. doi: 10.1089/jamp.2010.0816

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Elnadoury O, Beattie J, Lubinsky AS. Uninterrupted continuous and intermittent nebulizer therapy in a COVID-19 patient using sequential vibratory mesh nebulizers: a case report. J Aerosol Med Pulm Drug Deliv. (2020) 33(6):357–60. doi: 10.1089/jamp.2020.1636

PubMed Abstract | CrossRef Full Text | Google Scholar

40. DiBlasi RM, Crotwell DN, Shen S, Zheng J, Fink JB, Yung D. Iloprost drug delivery during infant conventional and high-frequency oscillatory ventilation. Pulm Circ. (2016) 6(1):63–9. doi: 10.1086/685080

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: infant, bronchopulmonary dysplasia, vibrating mesh nebulizer, jet nebulizer, high-frequency oscillatory ventilation

Citation: Zhou J-F, Zhang Y-B and Zhang Z-W (2023) Comparison of clinical outcomes following delivery of budesonide by both vibrating mesh nebulizer and jet nebulizer in premature infants with bronchopulmonary dysplasia. Front. Pediatr. 11:1258846. doi: 10.3389/fped.2023.1258846

Received: 14 July 2023; Accepted: 14 November 2023;
Published: 24 November 2023.

Edited by:

Malcolm King, Saskatchewan Centre for Patient-Oriented Research, Canada

Reviewed by:

James B. Fink, Texas State University System, United States
Ronan MacLoughlin, Aerogen Ltd., Ireland

© 2023 Zhou, Zhang and Zhang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jian-Fu Zhou zhoujianfu2023@163.com

Download