CASE REPORT article

Front. Surg., 05 April 2024
Sec. Heart Surgery
Volume 11 - 2024 | https://doi.org/10.3389/fsurg.2024.1290574

Atrial lead perforation early after device implantation—a case report and literature review

Zefeng Wang1,† Zhongyu Yuan1,† Haiwei Li1 Ke Zhang2 Hongkai Zhang3 Xiaoyan Li4 Changhua Wang1 Yilun Tian1 Yiqing Shen1 Xiaoping Zhang4* Yongquan Wu1*
  • 1Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China
  • 2Cardiovascular Surgery Center, Beijing Anzhen Hospital, Capital Medical University, Beijing, China
  • 3Department of Radiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China
  • 4Beijing Anzhen Hospital, Capital Medical University, The Key Laboratory of Remodelling-Related Cardiovascular Diseases, Ministry of Education, Beijing Institute of Heart, Lung and Blood Vessel Diseases, Beijing, China

We report three patients with screw-in lead perforation in the right atrial free wall not long after device implantation. All the patients complained of intermittent stabbing chest pain associated with deep breathing during the implantation. The “dry” epicardial puncture was utilized to avoid hemopericardium during lead extraction in the first case. The atrial electrode was repositioned in all cases and replaced by a new passive fixation lead in two patients with resolution of the pneumothorax or pericardial effusion. A literature review of 50 reported cases of atrial lead perforation was added to the findings in our case report.

Background

Lead perforations associated with the implantation of cardiovascular electronic devices are rare, with published rates of 0.1%–0.8% for permanent pacemakers and 0.14%–5.2% for implantable cardioverter defibrillators (ICDs) (1). The majority of cases occurred during implantation or within the first 24 h post procedure, but numerous instances of lead perforation occurring within days or months after procedure have also been reported (2). A wide range of typical and atypical symptoms can be observed, including pericardiac or pleuritic chest pain, hiccups, dyspnea, shortness of breath, hemoptysis, and even hemodynamic instability. Postoperative changes in electrical parameters may be a clue in asymptomatic patients (3). Transthoracic echocardiography (TTE) may reveal a hemodynamically relevant pericardial effusion, and the perforation site may be confirmed by computed tomography (CT). Detection of pneumothorax or hematopneumothorax on chest radiography (CXR) could also have auxiliary diagnostic value (2, 3). Furthermore, acute progression to cardiac tamponade is probably the most catastrophic consequence of a lead perforation, hence rapid management is crucial. Multiple studies in the literature have reported ventricular lead perforations induced by active, ICD, or even passive electrodes, regardless of the duration after the procedure. However, atrial lead perforations were less mentioned and mainly consisted of case reports. Herein we aimed to analyze the underlying mechanism of atrial lead perforation in the aspect of anatomy and provide the readers with some lessons learned for the diagnosis and treatment strategies of atrial perforation through this case series combined with a literature review.

Case presentation

Case 1: atrial lead perforation with right-sided hemothorax

A 71-year-old male patient was readmitted to our hospital on a complaint of right thoracic stabbing pain with alerting of a previously implanted cardiac resynchronization therapy defibrillator. He had non-ischemic dilated cardiomyopathy (DCM), left ventricular (LV) ejection fraction of 20%, left bundle branch block with a QRS duration (QRSd) of 175 ms, and New York Heart Association Class III symptoms despite optimal medical treatment. The procedure was then performed via the left axillary vein with a bipolar active fixation lead (Medtronic, CapSureFix Novus 5076, Minneapolis, MN, USA) positioned in the right atrial (RA) free wall (Figures 1A,B), a single coil lead (Medtronic, Sprint Quatro 6935M) screwed in the right ventricular (RV) anterior-mid septum, and a quadripolar lead (Medtronic, Attain Performa 4298) placed in the left ventricular anterior lateral wall successfully. At the time of implantation, the pacing threshold of the atrial lead was 0.8 V at 0.42 ms, the P-wave amplitude was 3 mV, and the impedance was 560 Ω. The pacing threshold of the shock lead was 0.9 V at 0.42 ms, the R-wave amplitude was 9 mV, and the impedance was 580 Ω. The left ventricular lead pacing threshold was 0.8 V without phrenic nerve stimulation. The postoperative electrocardiogram (ECG) showed a narrowed QRSd of 107 ms. The patient was discharged free of symptoms 2 days later on beta-receptor blocker, angiotensin receptor blocker, and diuretic medication.

Figure 1
www.frontiersin.org

Figure 1. Fluoroscopy images of the initial active fixation lead in RA free wall of case 1 [(A) RAO view and (B) LAO view]. Imaging examinations of case 1 obtained on readmission (C–F). Chest radiograph indicating the atrial lead protruding outside the contours of RA [(C) AP view, (D) RL view, and (E) LL view]. Chest CT scan showing the tip of the atrial electrode touching the right upper lobe of the lung and causing pneumothorax (F). AP, anterior posterior; CT, computed tomography; LAO, left anterior oblique; LL, left lateral; RA, right atrium; RAO, right anterior oblique; RL, right lateral.

On readmission, his vital signs were stable and within the normal range except for oxygen saturation of 93% in the room air. The physical examination revealed decreased respiratory sounds on the right side of the chest. Laboratory tests, including high-sensitive troponin T and serum myocardial enzyme levels, were within normal limits, and the ECG showed no ST-T changes or new Q wave, but a reprolonged QRSd of 203 ms excluded myocardial ischemia. Device interrogation revealed undersensing and loss of capture at maximal pacing output of the atrial lead. CXR showed an enlarged cardiac silhouette with the atrial lead protruding outside the contours of the RA (Figures 1C–E). However, CT visualized the tip of the helix of the atrial screw-in electrode touching the right upper lobe of the lung and causing pneumothorax (Figure 1F). As the TTE did not show any pericardial effusion, we considered that it was a pneumothorax that caused the chest pain.

Given that the helix tip of the protruding atrial lead bouncing with the heartbeat would probably cause further damage to the lung tissue, and that the patient was hemodynamically stable without a tendency to cardiac tamponade inclination, retention of the original atrial electrode or mini-thoracotomy was not considered as an optimal option. With ECG, blood pressure, oxygen saturation monitoring, prophylactic antibiotic therapy, and the pacemaker programmed for VVI ventricular sensing tracking pacing mode, the “dry” epicardial puncture was performed (4). The needle tip was initially directed to the middle point between the ventricular apex and annulus in the right anterior oblique (RAO) fluoroscopic projection, while the anterior–posterior angle was adjusted under the left lateral view. The J-tipped guidewire was placed through the needle into the pericardial space using a left anterior oblique (LAO) view (Figure 2A) and was left in situ in case of future cardiac effusion caused by electrode extraction. A new tined lead (Medtronic, CapSure Sense 4574) was implanted in the proximal right atrial appendage (RAA) (Figure 2B). The gentle traction of the target atrial lead was carefully applied using a LAO view under TTE monitoring, with an urgent thoracotomy planned in the event of retraction failure. Fluoroscopy visualized no pericardial effusion and the left ventricular lead remained in the appropriate position at various angles (Figures 2C–E). After 30 min of observation, TTE showed no pericardial effusion in the subxiphoid view. No complications occurred during or after the operative procedure. Iterative bedside TTEs during the postoperative 48 h showed no pericardial effusion and the pacing parameters were satisfactory, therefore the residual J-tipped guidewire was extracted. CXR showed spontaneous resolution of the pneumothorax. The patient was discharged after 7 days with a narrowed QRSd of 118 ms. During the 3-month follow-up, the patient remained asymptomatic and continued to show satisfactory device readings.

Figure 2
www.frontiersin.org

Figure 2. In case 1, a J-tipped guidewire was placed into the pericardial space using an LAO view (A) and was left in situ in case of future cardiac effusion caused by electrode extraction. A new passive fixation lead was implanted in proximal RAA in the LAO view (B). After complete retraction of the target atrial lead, fluoroscopy visualized no pericardial effusion and the left ventricular lead maintaining in the appropriate position was found in different views [(C) LAO view, (D) AP view, and (E), RAO view]. AP, anterior posterior; LAO, left anterior oblique; RAA, right atrial appendage; RAO, right anterior oblique.

Case 2: atrial lead perforation with pericardial effusion

A 78-year-old scrawny female patient with a symptomatic sick sinus syndrome (SSS) and amaurosis fugax was scheduled for elective implantation of a dual-chamber pacemaker implantation. Despite a history of paroxysmal atrial fibrillation, the patient was not regularly taking oral anticoagulants. A screw-in atrial lead (Boston Scientific, Ingevity 7741, North Charleston, SC, USA) was inserted from the left axillary vein and attached to the RA free wall (Figure 3A). Similarly, a screw-in ventricular lead (Boston Scientific, Ingevity 7742) was positioned in the RV septum. The pacing threshold of atrial lead was 0.8 V at 0.48 ms, the P-wave amplitude was 4.6 mV, and the impedance was 435 Ω. The ventricular lead pacing threshold was 0.9 V at 0.48 ms, the R-wave amplitude was 8.6 mV, and the impedance was 622 Ω.

Figure 3
www.frontiersin.org

Figure 3. A fluoroscopy image of case 2 showing the initial active fixation lead in RA free wall [(A) AP view]. A bright band at the bottom of the heart suggesting pericardial effusion in the LAO view [(B) white arrows]. After relief of chest pain by drainage through a pigtail catheter, a new passive fixation lead was placed in the distal RAA [(C) LAO view and (D) RAO view]. AP, anterior posterior; LAO, left anterior oblique; RA, right atrium; RAA, right atrial appendage; RAO, right anterior oblique.

A few minutes after being transferred from the operating table to the flat car, the patient began to complain of continuous left-sided chest pain with nausea. Of note, her blood pressure suddenly dropped to 100/60 mmHg. Immediate fluoroscopy demonstrated weakened cardiac pulsation and moderate pericardial effusion (Figure 3B). Pacemaker interrogation revealed loss of both atrial sensing and capture. With intravenous infusion of dopamine to improve the blood pressure to 105/62 mmHg and 0.9% sodium chloride for rapid rehydration and hemodynamic stability, x-ray guided pericardiocentesis was performed in the catheterization (cath) lab under local anesthesia. After drainage of incoagulable blood through a pigtail catheter, the patient reported remission of chest discomfort, and fluoroscopy visualized rehabilitative cardiac impulse and pericardial effusion much less than before. During a 30-min observation, her blood pressure fluctuated around 125/70 mmHg. A new tined lead (Boston Scientific, Fineline II Sterox 4480) was then placed in the distal RAA (Figures 3C,D) with pacing threshold of 1 V, lead impedance of 598 Ω, and P-wave amplitude of 2.6 mV. Reiterant bedside TTEs revealed no pericardial effusion and the patient was discharged 3 days after the procedure. The patient made a full recovery with no procedure-related complications at the 18-month follow-up.

Case 3: atrial lead perforation with pericardial effusion

A 60-year-old underweight female patient presented with symptomatic SSS and multiple syncope, which led to an uncomplicated implantation of a dual-chamber pacemaker. The ECG Holter showed Mobitz type II second-degree sinoatrial block with accidental junctional escapes and atrial premature beats. Implantation was uneventful with both leads inserted via the left axillary vein. The atrial lead (Biotronik, Solia S 53, Berlin, Germany) was actively fixed in the RA free wall (Figure 4A) and the ventricular lead (Biotronik, Solia S 60) was fixed in the RV septum. The atrial lead pacing threshold was 2 V at 0.48 ms, the P-wave amplitude was 3.3 mV, and the impedance was 532 Ω. The pacing threshold of the ventricular lead was 1.5 V at 0.48 ms, R-wave amplitude was 12.5 mV, and the impedance was 757 Ω.

Figure 4
www.frontiersin.org

Figure 4. A fluoroscopy image of case 3 showing initial active fixation lead in lateral RA and moderate pericardial effusion [(A) LAO view]. X-ray guided pericardiocentesis was performed using an LAO view and a pigtail catheter was inserted into the epidural space for continuous drainage, thus hemodynamic stability was obtained (B). Then a new active fixation electrode was implanted in RA septum [(C) RAO view and (D) LAO view]. LAO, left anterior oblique; RA, right atrium; RAO, right anterior oblique.

Approximately 30 min after returning to the ward, the patient complained of severe breath-related left-sided pain with nausea and emesis. Pacemaker interrogation revealed the failure of atrial pacing and sensing and decreased lead impedance. Blood pressure monitoring revealed a sudden drop to 70/40 mmHg and distant heart sounds were auscultated. An intravenous infusion of 0.9% sodium chloride 500 ml for rapid rehydration and an intravenous injection of dopamine to improve hemodynamics were immediately administered. With a heart rate of 76 bpm, atropine was also injected intravenously, followed by continuous dopamine infusion at 8 μg/kg/min, with blood pressure maintained at 100/50 mmHg and heart rate at 68 bpm. An emergency bedside TTE defined a mild to moderate pericardial effusion (anterior 9 mm, posterior 4 mm, and RV free wall 12 mm), confirming the suspicion of atrial lead perforation. X-ray guided pericardiocentesis was performed and a pigtail catheter was inserted into the epidural space for continuous drainage of 400 ml of incoagulable blood (Figure 4B). The patient reported significant amelioration of chest pain. Afterward, the atrial lead was removed transvenously and a new screw-in lead (Medtronic, Select-Secure 3830) was implanted in the RA septum (Figures 4C,D) with a pacing threshold of 0.8 V at 0.48 ms, P-wave amplitude of 3.1 mV, and impedance of 783 Ω. The patient was observed for 6 days in the ward postoperatively. The patient had no recurrent pericardial effusion during 6 months of clinical and echocardiographic follow-up, and pacemaker interrogation was unremarkable.

Review of the literature

We searched PUBMED for articles on atrial lead perforations from the period between 1986 and 2023. The inclusion criterion was myocardial or vascular injury due to atrial lead dislocation. The exclusion criteria were as follows: perforation caused by (i) sheath or catheter intraoperatively, (ii) erosive ventricular lead, (iii) lead extraction, and (iv) the piercing site not verified by any imaging examination or direct visualization during thoracotomy. As a result, 50 patients, including our patients, were enrolled (Table 1). There were 21 men and 29 women, ranging in age from 23 to 88 years (average 66.8 ± 14.0).

Table 1
www.frontiersin.org

Table 1. Published cases of atrial lead perforation between 1986 and 2021.

Manifestation

Lead perforations occurred within and over 1 month of the implantable procedures in 34 (68%) and 16 (32%) patients, respectively. Electrical performance such as capture, sensing, and lead impedance of the atrial lead was not reported in eight patients. Besides 18 patients with no change, the main malfunction was loss of capture in 15 patients, of which 14 were associated with sensing problems, 5 with elevated threshold, and 3 with impedance changes. Therefore, it is important to pay close attention to any discomfort within the first month after pacemaker implantation, even if the changes in pacing parameters are inconspicuous.

A total of 41 patients presented with clinical symptoms, predominantly chest pain in 25 patients. Other concomitant symptoms covered syncope, dyspnea, shortness of breath, hemoptysis, back pain, hiccups, nausea, and emesis. Of these, 12 patients were hemodynamically unstable at the time of detection of the perforations. The perforations were detected in two asymptomatic patients with normal electrical parameters on routine CXR (33, 41). Coronary angiography in one patient with prior coronary artery bypass grafting demonstrated that the atrial lead was protruding into an unusual filling mass communicating with the right coronary artery (RCA) through a fistula (22). In a total of 50 patients, 29 had CT evidence of lead perforation, while CXR illustrated the piercing site in 26 patients with 16 overlapping. Totally nine cases were investigated by TTE alone before intervention.

Position of the lead tip

In 40 (80%) patients, screw-in leads were perforated, 4 (8%) had passive fixation leads, and the fixation mechanism was unknown in 6 cases. The use of active fixation leads reduces the risk of lead displacement but may increase the risk of perforation. An observational study found that eight patients (42%) experienced subsequent lead displacement, while no patient experienced displacement of an active fixation lead in the coronary sinus for bi-atrial pacing (46). A recent study found that active fixation leads had a considerably reduced rate of displacement within 6 months compared with passive fixation quadripolar leads (1/135, 0.74%, vs. 16/341, 4.69%) for cardiac resynchronization therapy (47). Concerning a total of 40 active fixation leads, 16 were placed in the RA free wall, 9 in the RAA, 2 in the anterolateral RA, and 1 in the anterior RA with 12 not reported. Regarding the 4 cases utilizing tined leads, 1 reported the original lead position in the RAA, while the others were unknown.

In the total cohort of 50 patients, the lead perforated into the pleura in 25 patients, and pericardiac perforations in situ occurred in 22 patients. Bleeding from a hole injury in the atrial wall at the site of the RCA was observed in two patients during thoracotomy (31, 43). In another patient, mechanical irritation led to a fistulous connection with the RCA (22). One patient had a hemorrhagic ulcer of the aortic root in the region of the non-coronary sinus caused by friction from the atrial lead (27). In one patient, the dislodged atrial lead exited the superior vena cava (SVC) through a perforation, passed through the dorsal transverse sinus of the ascending aorta, and connected to the roof of the left atrium (28).

Management

Surgical interventions including thoracotomy, mini-thoracotomy, thoracoscope, and da Vinci robotic system were utilized in 23 patients. Among the rest of the cases, pericardiocentesis was required in seven patients and chest tube insertion in three patients. Repositioning of the initial atrial lead was achieved in 13 patients. In eight patients the protruding lead was removed. For the 13 patients with lead extraction and new lead implantation, 6 were replaced with passive fixation leads, 4 remained with active fixation leads, and 3 had epicardial leads placed. In a patient with hemoptysis, the application of local catecholamines under bronchoscopy decreased bleeding in the right middle lobe (36). Expectant management was continued due to the substantial high risk of lead revision in one patient with underlying severe chronic obstructive pulmonary disease (13) and spontaneous resolution of the pneumothorax in three patients each (30, 33). The postoperative course was uncomplicated in all but one patient (27).

Discussion

Backtracking along the intraoperative situation, we found that all three patients suffered from breath-dependent chest prickling pain to varying degrees during the procedure. Concerning the readmitted patients with atrial lead perforation, their vital signs were relatively stable. As chest pain should be differentiated from myocardial infarction, aortic dissection, and pulmonary embolism, the diagnosis was established by electrical parameters and imaging examinations. Hemodynamically irrelevant pericardial effusion on TTE should be identified with pericarditis. Significantly, we found that atrial electrode perforation occurring before discharge was generally associated with pericardial effusion and presented with more severe symptoms such as chest distress, spontaneous sweating, and subsequent hemorrhagic shock.

Using screw-in leads, improper fixation strength and excessive attempt times were all closely correlated with the high incidence of cardiac perforation. Furthermore, the influence of the implantation site cannot be ignored. Zoppo et al. assessed the morphology of the RAA in relation to pacing and concluded that the proximal antral RAA is potentially a dangerous region for lead anchorage, primarily because of its proximity to the aortic root and the paper-thin RAA wall in the interpectoral muscular spaces, compared with the safer distal saccular RAA beyond the tenia sagittalis (48). The first patient described validated this conclusion. The thickness of the atrial wall was proportional to the different heights of the RAA isthmuses, as well as the greatest near the valve annulus and the thinnest near the RAA orifice (49). Consistent with literature review, leads implanted in the anterior wall of the RA or the lateral wall of the RAA root have a higher risk of perforation. Based on our experience, if the active lead has been tilted to the aforementioned site while the patient, particularly elderly female patients with a low body mass index (BMI), complains of breath-dependent chest prickling pain during the implantation, the possibility of the lead penetrating the visceral pericardium should be taken into consideration. On this occasion, close attention should be paid to any change in the patient's condition, including symptoms, vital signs, and size of the cardiac silhouette. When there were persistent symptoms or altered electrical parameters, especially notably elevated threshold and decreased impedance, or progressive pericardial effusion, close monitoring following the lead replacement was recommended. Definitely, future studies are needed to elucidate the relationship between atrial electrode perforation and atrial anatomy to further reduce its incidence.

For patients who had hemopericardium and hypotension, urgent pericardiocentesis and fluid resuscitation should be performed. Pericardiocentesis under fluoroscopy was superior to bedside when there was only a minimal volume of pericardial effusion. The observation period before the end of intervention might be appropriately extended. After the hemodynamic stability was obtained, it was time for electrophysiologists to decide how the protruding electrode should be tackled on the basis of electrical parameters and imaging examinations. Patients who do not respond to the aforementioned treatment or who have a cardiac rupture should undergo surgical intervention. Other than ventricular perforation, atrial lead perforation can be safely managed by experienced electrophysiologists in the cath lab.

The “dry” epicardial puncture introduced by Sosa Scanavacca et al. was an established approach for epicardial mapping in patients with ventricular tachycardia after failed endocardial ablation (50). Since then its application has been broadened to ligation of the left atrial appendage, ablation of atrial fibrillation, and implantation of pacemaker leads (5153). We utilized this method with J-tipped guidewire insertion in the first patient in the event that lead extraction generated hemopericardium. This prophylactic pericardiocentesis was recommended in hemodynamically stable patients with no or merely mild pericardial effusion, especially in patients with heart failure and DCM who could not tolerate the thoracotomy.

Once in-hospital lead perforation was confirmed, right back to the cath lab withdrawing the screw-in lead and replacing with a tined one were recommended. Also, lead repositioning to the atrial septum was generally feasible. Alteration of both the type and position of the atrial lead transvenously was safe and reliable, because in the acute period the fibrous tissue around the protruding electrode has not formed yet, thus its retention could cause further damage to the surrounding tissue and pericardial effusion or tamponade during extraction.

Last but not the least, we should focus on the prevention of the lead perforation, especially when an active fixation lead is arranged to be inserted into the RA free wall in elderly women with low BMI who are quite likely to have thinned out myocardia. The procedure should be gentle and the pacing wire should be reserved of moderate length to avoid excessive tension at the distal tip.

Conclusions

In summary, atrial lead perforation is likely to occur when an active fixation lead is placed in the RA free wall in elderly patients with low BMI. Our experience is that if the patient has intermittent stabbing chest pain associated with deep breathing during the implantation, atrial perforation should be suspected, which could be confirmed by electrical parameters and imaging examinations. Other than ventricular perforation, the atrial lead perforation in the cath lab can be safely handled by experienced electrophysiologists.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.

Ethics statement

Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

ZW: Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Writing – original draft. ZY: . HL: Investigation, Validation, Visualization, Writing – review & editing. KZ: Project administration, Writing – original draft. HZ: Resources, Software, Visualization, Writing – original draft. XL: Project administration, Writing – review & editing. CW: Data curation, Writing – review & editing. YT: Data curation, Writing – review & editing. YS: Data curation, Writing – original draft. XZ: Funding acquisition, Supervision, Writing – review & editing. YW: Funding acquisition, Supervision, Writing – review & editing.

Funding

The authors declare financial support was received for the research, authorship, and/or publication of this article.

This study was supported by Beijing Hospitals Authority Clinical Medicine Development of Special Funding Support (grant no. XMLX202112), the National Natural Science Foundation of China (grant no. 62272415), and the Capital Medical University Major Science and Technology Innovation Research and Development Special Fund (grant no. KCZD202201).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Adamczyk M, Wasilewski J, Niedziela J, Rozentryt P, Gasior M. Pericardial tamponade as a complication of invasive cardiac procedures: a review of the literature. Postepy Kardiol Interwencyjnej. (2019) 15(4):394–403. doi: 10.5114/aic.2019.90213

PubMed Abstract | Crossref Full Text | Google Scholar

2. Van Gelder BM, Verberkmoes N, Nathoe R, Bracke FA. Late asymptomatic atrial lead perforation, a fortuitous finding during lead extraction using thoracoscopic surveillance: a case report and review of the literature. Europace. (2016) 18(12):1773–8. doi: 10.5114/aic.2019.90213

PubMed Abstract | Crossref Full Text | Google Scholar

3. Vamos M, Erath JW, Benz AP, Bari Z, Duray GZ, Hohnloser SH. Incidence of cardiac perforation with conventional and with leadless pacemaker systems: a systematic review and meta-analysis. J Cardiovasc Electrophysiol. (2017) 28(3):336–46. doi: 10.1111/jce.13140

PubMed Abstract | Crossref Full Text | Google Scholar

4. Long DY, Sun LP, Sang CH, Jiang CX, Guo XY, Li MM, et al. Pericardial access via wire-guided puncture without contrast: the feasibility and safety of a modified approach. J Cardiovasc Electrophysiol. (2020) 31(1):30–7. doi: 10.1111/jce.14237

PubMed Abstract | Crossref Full Text | Google Scholar

5. Sussman Chiles SK, Cooper C, Lowe JE. CT demonstration of myocardial perforation by a pacemaker lead. J Comput Assist Tomogr. (1986) 10(4):670–2. doi: 10.1097/00004728-198607000-00024

PubMed Abstract | Crossref Full Text | Google Scholar

6. Irwin JM, Greer GS, Lowe JE, German LD, Gilbert MR. Atrial lead perforation: a case report. Pacing Clin Electrophysiol. (1987) 10(6):1378–81. doi: 10.1111/j.1540-8159.1987.tb04973.x

PubMed Abstract | Crossref Full Text | Google Scholar

7. Ho WJ, Kuo CT, Lin KH. Right pneumothorax resulting from an endocardial screw-in atrial lead. Chest. (1999) 116(4):1133–4. doi: 10.1378/chest.116.4.1133

PubMed Abstract | Crossref Full Text | Google Scholar

8. Aizawa K, Kaneko Y, Yamagishi T, Utsugi T, Suzuki T, Ishikawa S, et al. Oozing from the pericardium as an etiology of cardiac tamponade associated with screw-in atrial leads. Pacing Clin Electrophysiol. (2001) 24(3):381–3. doi: 10.1046/j.1460-9592.2001.00381.x

PubMed Abstract | Crossref Full Text | Google Scholar

9. Tran, NT, Zivin A, Mozaffarian D, Karmy-Jones R. Right atrial perforation secondary to implantable cardioverter defibrillator insertion. Can Respir J. (2001) 8(4):283–5. doi: 10.1155/2001/257641

PubMed Abstract | Crossref Full Text | Google Scholar

10. Dilling-Boer D. Pericardial effusion and right-sided pneumothorax resulting from an atrial active-fixation lead. Europace. (2003) 5(4):419–23. doi: 10.1016/S1099-5129(03)00079-5

PubMed Abstract | Crossref Full Text | Google Scholar

11. Velavan P, Chauhan A. An unusual presentation of delayed cardiac perforation caused by atrial screw-in lead. Heart. (2003) 89(4):364. doi: 10.1136/heart.89.4.364

PubMed Abstract | Crossref Full Text | Google Scholar

12. Howell C, Bergin JD. A case report of pacemaker lead perforation causing late pericardial effusion and subacute cardiac tamponade. J Cardiovasc Nurs. (2005) 20(4):271–5. doi: 10.1097/00005082-200507000-00014

PubMed Abstract | Crossref Full Text | Google Scholar

13. Henrikson CA, Leng CT, Yuh DD, Brinker JA. Computed tomography to assess possible cardiac lead perforation. Pacing Clin Electrophysiol. (2006) 29(5):509–11. doi: 10.1111/j.1540-8159.2006.00385.x

PubMed Abstract | Crossref Full Text | Google Scholar

14. Geyfman V, Storm RH, Lico SC, Oren JW 4th. Cardiac tamponade as complication of active-fixation atrial lead perforations: proposed mechanism and management algorithm. Pacing Clin Electrophysiol. (2007) 30(4):498–501. doi: 10.1111/j.1540-8159.2007.00699.x

PubMed Abstract | Crossref Full Text | Google Scholar

15. Spencker S, Mueller D, Marek A, Zabel M. Severe pacemaker lead perforation detected by an automatic home-monitoring system. Eur Heart J. (2007) 28(12):1432. doi: 10.1093/eurheartj/ehl495

PubMed Abstract | Crossref Full Text | Google Scholar

16. Namazi MH, Karbasi-Afshar R, Safi M, Serati AR. Diaphragmatic stimulation: a case of atrial lead dislodgement and right atrium perforation. Indian Pacing Electrophysiol J. (2008) 8(2):133–6.18379658

PubMed Abstract | Google Scholar

17. Sadamatsu K, Enomoto N, Tsuji M, Tashiro H. Progressive atrial lead perforation developed 5 years after pacemaker replacement. J Cardiol. (2009) 53(1):150–3. doi: 10.1016/j.jjcc.2008.07.006

PubMed Abstract | Crossref Full Text | Google Scholar

18. Schiariti M, Cacciola MT, Pangallo A, Ciancia F, Puddu PE. Delayed pericarditis and cardiac tamponade associated with active-fixation lead pacemaker in the presence of mitochondrial myopathy and Ockham’s razor. J Cardiovasc Med. (2009) 10(11):879–82. doi: 10.2459/JCM.0b013e32832e6499

Crossref Full Text | Google Scholar

19. Peddi P, Vodnala D, Kalavakunta JK, Thakur RK. Acute chest pain: acute coronary syndrome versus lead perforation: a case report. Int Arch Med. (2010) 3:13. doi: 10.1186/1755-7682-3-13

PubMed Abstract | Crossref Full Text | Google Scholar

20. Antonelli D, Feldman A, Schliamser JE, Militianu A, Turgeman Y. Acute pericardial tamponade due to screw-in atrial lead heart perforation. Europace. (2012) 14(3):453–5. doi: 10.1093/europace/eur294

PubMed Abstract | Crossref Full Text | Google Scholar

21. Hussain S, Adams C, Mechulan A, Leong-Sit P, Kiaii B. Minimally invasive robotically assisted repair of atrial perforation from a pacemaker lead. Int J Med Robot. (2012) 8(2):243–6. doi: 10.1002/rcs.459

PubMed Abstract | Crossref Full Text | Google Scholar

22. Khoueiry G, Lakhani M, Abi Rafeh N, Azab B, Schwartz C, Kowalski M, et al. Right coronary artery fistula as a result of delayed right atrial perforation by a passive fixation lead. Circ Arrhythm Electrophysiol. (2012) 5(2):e46–7. doi: 10.1161/CIRCEP.111.970376

PubMed Abstract | Crossref Full Text | Google Scholar

23. Nakanishi H, Kashiwase K, Nishio M, Wada M, Hirata A, Ueda Y. Recurrent pericardial effusion caused by pacemaker lead perforation and warfarin therapy at seven years after implantation. Europace. (2012) 14(2):297. doi: 10.1093/europace/eur268

PubMed Abstract | Crossref Full Text | Google Scholar

24. O'Neill R, Silver M, Khorfan F. Pneumopericardium with cardiac tamponade as a complication of cardiac pacemaker insertion one year after procedure. J Emerg Med. (2012) 43(4):641–4. doi: 10.1016/j.jemermed.2010.04.026

Crossref Full Text | Google Scholar

25. Souretis G, Kotoulas C, Xapladeris P, Flessas D, Laschos N, Lazarides K. Late atrial appendage perforation from a pacemaker lead pacemaker complication. Asian Cardiovasc Thorac Ann. (2014) 22(1):83–5. doi: 10.1177/0218492312467541

PubMed Abstract | Crossref Full Text | Google Scholar

26. Higny J, Blommaert D. Pneumopericardium: an uncommon complication of atrial lead perforation following pacemaker implantation. Acta Cardiol. (2014) 69(2):213–5. doi: 10.1080/AC.69.2.3017308

PubMed Abstract | Crossref Full Text | Google Scholar

27. Di Marco A, Nuñez E, Osorio K, Dallaglio P, Anguera I, Toscano J, et al. Aortic perforation by active-fixation atrial pacing lead: an unusual but serious complication. Tex Heart Inst J. (2014) 41(3):327–8. doi: 10.14503/THIJ-13-3219

PubMed Abstract | Crossref Full Text | Google Scholar

28. Rath C, Andreas M, Khazen C, Wiedemann D, Habertheuer A, Kocher A. Pacemaker lead malposition in the left atrial roof is masked by normal pacing thresholds. J Cardiothorac Surg. (2014) 9:53. doi: 10.1186/1749-8090-9-53

PubMed Abstract | Crossref Full Text | Google Scholar

29. De Schryver N, Marchandise S, Colin GC, Ghaye B, le Polain de Waroux JB. Asymptomatic late migration of an atrial pacemaker lead into the right lung. Case Rep Cardiol. (2014) 2014:145917. doi: 10.1155/2014/145917

PubMed Abstract | Crossref Full Text | Google Scholar

30. Baird A, Gandhi M. Pneumopericardium and pneumothorax due to right atrial permanent pacemaker lead perforation. J Med Imaging Radiat Oncol. (2015) 59(1):74–6. doi: 10.1111/1754-9485.12200

PubMed Abstract | Crossref Full Text | Google Scholar

31. Nakagawa E, Abe Y, Komatsu R, Naruko T, Itoh A. Right coronary artery perforation by an active-fixation atrial pacing lead resulting in life-threatening tamponade. J Arrhythm. (2015) 31(5):313–5. doi: 10.1016/j.joa.2015.01.002

PubMed Abstract | Crossref Full Text | Google Scholar

32. Rali AS, Manyam H. Bilateral pneumothoraces following BiV ICD placement: a case of Buffalo chest syndrome. Am J Case Rep. (2015) 16:703–6. doi: 10.12659/AJCR.894671

PubMed Abstract | Crossref Full Text | Google Scholar

33. Ishizue N, Kishihara J, Niwano S, Ako J. Subacute pneumothorax contralateral to the venous access site associated with atrial lead perforation in a patient who was receiving corticosteroid therapy. J Arrhythm. (2017) 33(4):335–7. doi: 10.1016/j.joa.2017.03.003

PubMed Abstract | Crossref Full Text | Google Scholar

34. Saradna A, Sinha A, Abduraimova M, Rodriguez D, Yang F. Tale of a wandering lead: late atrial lead perforation into right lung following pacemaker implantation. Cureus. (2017) 9(11):e1865. doi: 10.7759/cureus.1865

PubMed Abstract | Crossref Full Text | Google Scholar

35. Aktaa S, Fatania K, Gains C, White H. Chest pain following permanent pacemaker insertion … a case of pneumopericardium due to atrial lead perforation. BMJ Case Rep. (2018) 1–3. doi: 10.1136/bcr-2018-226318

Crossref Full Text | Google Scholar

36. Schreiber T, Tscholl V, Niehues SM, Nagel P, Starck CT, Landmesser U, et al. A combination of rare complications 3 years after a dual-chamber pacemaker implantation. Clin Res Cardiol. (2019) 108(5):465–7. doi: 10.1007/s00392-018-1375-3

PubMed Abstract | Crossref Full Text | Google Scholar

37. Shenthar J, Singh B, Banavalikar B, Chakali SS, Delhaas T, Shivkumar K, et al. Cardiac perforation complicating cardiac electrophysiology procedures: value of angiography and use of a closure device to avoid cardiac surgery. J Interv Card Electrophysiol. (2020) 58(2):193–201. doi: 10.1007/s10840-019-00577-0

PubMed Abstract | Crossref Full Text | Google Scholar

38. Zhou X, Ze F, Li D, Li XB. Percutaneous management of atrium and lung perforation: a case report. World J Clin Cases. (2019) 7(24):4327–33. doi: 10.12998/wjcc.v7.i24.4327

PubMed Abstract | Crossref Full Text | Google Scholar

39. Futami M, Komaki T, Arinaga T, Morii J, Sugihara M, Ogawa M, et al. Postural conversion computed tomography for the diagnosis of pneumopericardium due to perforation by the active atrial lead. Intern Med. (2020) 59(4):541–4. doi: 10.2169/internalmedicine.3729-19

PubMed Abstract | Crossref Full Text | Google Scholar

40. Gianni C, Canby RC, Anannab A, Della Rocca DG, Natale A, Al-Ahmad A. The subtle journey of a right atrial lead: hemopneumothorax due to subacute pacemaker lead perforation. JACC Case Rep. (2020) 2(6):902–6. doi: 10.1016/j.jaccas.2020.05.018

PubMed Abstract | Crossref Full Text | Google Scholar

41. Olesen LL. Bilateral pneumothorax complicating pacemaker implantation, due to puncture of the left subclavian vein and electrode perforation of the right atrium. Cureus. (2020) 12(11):e11302. doi: 10.7759/cureus.11302

PubMed Abstract | Crossref Full Text | Google Scholar

42. Grebmer C, Russi I, Kobza R, Berte B. Atrial lead perforation early after device implantation: a case series. HeartRhythm Case Rep. (2021) 7(2):100–5. doi: 10.1016/j.hrcr.2020.11.012

PubMed Abstract | Crossref Full Text | Google Scholar

43. Tsuji Y, Miki T, Sato Y, Kojima A, Imai M, Kuroyanagi A, et al. A case of cardiac tamponade due to coronary artery injury caused by atrial screw-in lead. J Cardiol Cases. (2021) 23(2):80–2. doi: 10.1016/j.jccase.2020.09.011

PubMed Abstract | Crossref Full Text | Google Scholar

44. Yilancioglu RY, Turan OE, Baskurt AA, Ozcan EE. Lead related atrial perforation mimicking inferior ST elevation myocardial infarction. Pacing Clin Electrophysiol. (2022) 46:710–3. doi: 10.1111/pace.14641

PubMed Abstract | Crossref Full Text | Google Scholar

45. Enokizono K, Kamakura T, Kotoku A, Nakata J, Matama H, Kusano K. Very late-onset atrial lead perforation leading to pneumopericardium. J Cardiovasc Electrophysiol. (2023) 34:1473–6. doi: 10.1111/jce.15930

PubMed Abstract | Crossref Full Text | Google Scholar

46. Levy T. A comparison between passive and active fixation leads in the coronary sinus for biatrial pacing initial experience. Europace. (2000) 2(3):228–32. doi: 10.1053/eupc.2000.0105

PubMed Abstract | Crossref Full Text | Google Scholar

47. Robertson C, Duffey O, Tang PT, Fairhurst N, Monteiro C, Green P, et al. An active fixation quadripolar left ventricular lead for cardiac resynchronization therapy with reduced postoperative complication rates. J Cardiovasc Electrophysiol. (2022) 33(3):458–63. doi: 10.1111/jce.15346

PubMed Abstract | Crossref Full Text | Google Scholar

48. Zoppo F, Rizzo S, Corrado A, Bertaglia E, Buja G, Thiene G, et al. Morphology of right atrial appendage for permanent atrial pacing and risk of iatrogenic perforation of the aorta by active fixation lead. Heart Rhythm. (2015) 12(4):744–50. doi: 10.1016/j.hrthm.2014.12.023

PubMed Abstract | Crossref Full Text | Google Scholar

49. Holda J, Slodowska K, Tyrak K, Bolechala F, Jasinska KA, Koziej M, et al. Topographical anatomy of the right atrial appendage vestibule and its isthmuses. J Cardiovasc Electrophysiol. (2020) 31(12):3199–206. doi: 10.1111/jce.14767

PubMed Abstract | Crossref Full Text | Google Scholar

50. Sosa Scanavacca E, D'Avila A, Pilleggi F. A new technique to perform epicardial mapping in the electrophysiology laboratory. J Cardiovasc Electrophysiol. (1996) 7(6):531–6. doi: 10.1111/j.1540-8167.1996.tb00559.x

PubMed Abstract | Crossref Full Text | Google Scholar

51. Valderrabano M. Pericardial access for LARIAT left atrial appendage closure. Interv Cardiol Clin. (2014) 3(2):281–9. doi: 10.1016/j.iccl.2013.12.003

PubMed Abstract | Crossref Full Text | Google Scholar

52. Buch E, Shivkumar K. Epicardial catheter ablation of atrial fibrillation. Minerva Med. (2009) 100(2):151–7.19390501

PubMed Abstract | Google Scholar

53. Mickelsen SR, Ashikaga H, DeSilva R, Raval AN, McVeigh E, Kusumoto F. Transvenous access to the pericardial space: an approach to epicardial lead implantation for cardiac resynchronization therapy. Pacing Clin Electrophysiol. (2005) 28(10):1018–24. doi: 10.1111/j.1540-8159.2005.00236.x

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: pacemaker complications, atrial perforation, active fixation, cardiac perforation, active fixation lead

Citation: Wang Z, Yuan Z, Li H, Zhang K, Zhang H, Li X, Wang C, Tian Y, Shen Y, Zhang X and Wu Y (2024) Atrial lead perforation early after device implantation—a case report and literature review. Front. Surg. 11:1290574. doi: 10.3389/fsurg.2024.1290574

Received: 7 September 2023; Accepted: 19 March 2024;
Published: 5 April 2024.

Edited by:

Mark Gallagher, St George's University Hospitals NHS Foundation Trust, United Kingdom

Reviewed by:

Francesco Patanè, Azienda Ospedaliera Ospedali Riuniti Papardo Piemonte, Italy
Nachiket Apte, University of Louisville, United States

© 2024 Wang, Yuan, Li, Zhang, Zhang, Li, Wang, Tian, Shen, Zhang and Wu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Xiaoping Zhang xiaoping_zhang2004@ccmu.edu.cn Yongquan Wu wuyongquan67@126.com

These authors have contributed equally to this work and share first authorship

Abbreviations RA, right atrial; ICDs, implantable cardioverter defibrillators; TTE, transthoracic echocardiography; CT, computed tomography; CXR, chest radiography; DCM, dilated cardiomyopathy; QRSd, QRS duration; ECG, electrocardiogram; RAO, right anterior oblique; LAO, left anterior oblique; RAA, right atrial appendage; SSS, sick sinus syndrome; RV, right ventricular; RCA, right coronary artery; SVC, superior vena cava; BMI, body mass index.