ORIGINAL RESEARCH article

Front. Chem., 05 August 2021
Sec. Analytical Chemistry
Volume 9 - 2021 | https://doi.org/10.3389/fchem.2021.730954

Development, Validation, and Application of a Novel Method for the Analysis of Vitamin E Acetate and Other Tocopherols in Aerosol Emissions of E-Cigarettes, or Vaping Products Associated With Lung Injury

  • Tobacco and Volatiles Branch, Division of Laboratory Sciences, National Center for Environmental Health, U.S. Centers for Disease Control and Prevention, Atlanta, GA, United States

E-cigarette, or vaping, product (EVP) use has increased dramatically in the United States over the last 4 years, particularly in youth and young adults. Little information is available on the chemical contents of these products. Typically, EVPs contain an active ingredient such as nicotine, CBD, or THC dissolved in a suitable solvent that facilitates aerosol generation. One EVP solvent, vitamin E acetate (VEA), has been measured in EVP liquids associated with lung injury. However, no validated analytical methods for measuring VEA in the aerosol from these devices was previously available. Therefore, we developed a high throughput isotope dilution LC-MS/MS method to simultaneously measure VEA and three other related tocopherols in aerosolized EVP samples. The assay was precise, with VEA repeatability ranging from 4.0 to 8.3% and intermediate precision ranging from 2.5 to 6.7%. Similar precision was obtained for the three other tocopherols measured. The LODs for the four analytes ranged from 8.85 × 10−6 to 2.28 × 10−5 μg analyte per mL of aerosol puff volume, and calibration curves were linear (R2 > 0.99). This method was used to analyze aerosol emissions of 147 EVPs associated with EVALI case patients. We detected VEA in 46% of the case-associated EVPs with a range of 1.87 × 10−4–74.1 µg per mL of aerosol puff volume and mean of 25.1 µg per mL of aerosol puff volume. Macro-levels of VEA (>0.1% w/w total aerosol particulate matter) were not detected in nicotine or cannabidiol (CBD) products; conversely 71% of the EVALI associated tetrahydrocannabinol (THC) products contained macro-levels of VEA. Trace levels of other tocopherol isoforms were detected at lower rates and concentrations (α-tocopherol: 41% detected, mean 0.095 µg analyte per mL of aerosol puff volume; γ-tocopherol: 5% detected, mean 0.0193 µg analyte per mL of aerosol puff volume; δ-tocopherol: not detected). Our results indicate that VEA can be efficiently transferred to aerosol by EVALI-associated EVPs vaped using a standardized protocol.

Introduction

In recent years the use of e-cigarettes, or vaping products (EVPs) have dramatically increased1. More than eight million U.S. adults reported using these products on a regular basis1 (Creamer MR, 2019). The use of nicotine containing EVPs could potentially benefit adult smokers if used as a complete substitute for traditional cigarettes and other combusted tobacco products rather than dual use (Shahab et al., 2017). However, EVP use increases disease risk for those who are not already using tobacco products (Orzabal and Ramadoss, 2019; Wang et al., 2020). Research is needed to better understand potential long-term health effects of inhaling EVP aerosols, including solvents, additives, and diluents (Ghosh and Drummond, 2017; Bhatnagar et al., 2019; Hajek et al., 2019). One of the main challenges for assessing the potential health impacts of EVP use is the accessibility to e-liquid components and formulations through “informal” and individualized marketing (Zhu et al., 2014) Decriminalization of cannabinoids for medical and non-medical purposes by some states facilitated a surge in use of EVPs for vaping cannabis as these products are more available in some markets2 (Pacula and Smart, 2017; Mcnamara, 2020).

From August 2019 to February 2020 an outbreak of e-cigarette, or vaping, product use associated lung injury (EVALI) occurred across the United States: at least 2,807 people were hospitalized, and 68 deaths reported in 29 states. Most EVALI cases were <35 years of age and previously healthy (Lozier et al., 2019; Moritz et al., 2019). Most patients that were hospitalized reported using e-liquid products containing THC (Butt et al., 2019; Lozier et al., 2019; Siegel et al., 2019). These lung injury cases have been associated to the use of vitamin E acetate (VEA)-containing EVPs (Duffy et al., 2020). Analysis of bronchoalveolar lavage fluid from EVALI patients identified VEA accumulation and thus implicated inhaled VEA as the likely cause of the 2019 EVALI lung injury outbreak (Blount et al., 2019; Blount et al., 2020). Multiple trade websites report the use of VEA and medium- chain triglycerides in THC products to enhance quality, appearance, and aroma, as well as a way to lower production cost (Downs, 2019a; Downs, 2019b; Zachary Eisenberg, 2019).

VEA is the shelf-stable synthetic form of vitamin E often used in skin care products and dietary supplements. Oral and topical administration of VEA has been used for years without significant adverse health effects. Because of the 2019 EVALI outbreak, the effects of inhaled VEA are starting to be evaluated. In fact, two recent studies find that mice exposed to VEA emissions develop lung injury and other pathologies similar to EVALI patients (Bhat et al., 2020; Matsumoto et al., 2020). Traditional analysis of VEA and other tocopherols are mainly performed through HPLC-UV for cosmetics and foods products intended for dermal application or ingestion (Cunha et al., 2006; Bustamante-Rangel et al., 2007; Almeida et al., 2009; Nada et al., 2010; Şeker et al., 2012; Cortés-Herrera et al., 2019; Sadrykia et al., 2019). Recently, the analysis of VEA in EVP liquids was reported using screening and targeted GC-MS and LC-MS/MS assays3 (Health, 2019). No methods existed for the analysis of VEA and other tocopherols in aerosol emissions of e-liquid products samples, thus here we report on the development and validation of an analytical method for VEA and other tocopherols in aerosol emissions samples. The efficacy of the method is subsequently demonstrated by analysis of aerosol emissions from EVPs associated with EVALI case patients.

Experimental

Materials and Methods

Methanol (CAS# 67-56-1; LC-MS/MS grade) purchased from Fischer Scientific was mixed with deionized water from an ultrapure water purifications system (Aqua Solutions model RODI-C-11BL, Jasper, GA, United States) and formic acid (CAS# 64-18-6; chemical purity: ≥ 98%; ACS reagent) to form a 90% methanol: 10% water and 0.1% formic acid mobile phase.

VEA, DL-alpha Tocopherol acetate (CAS# 7695-91-2, ≥99% purity), (+)-alpha Tocopherol (CAS# 59-02-9, ≥99% purity), (+)-gamma Tocopherol (CAS# 54-28-4, ≥96% purity), delta-Tocopherol (CAS# 119-13-1, ≥99% purity), alpha-Tocopherol-(phenyl-13C6) (≥99% atom purity, ≥96% compound purity), labeled vitamin E acetate - (trimethyl-d9) (≥ 98 atom %; chemical purity: ≥98%) were all purchased from Sigma Aldrich (St. Louis, Missouri, United States).

Stock solutions and calibrators of unlabeled VEA and other tocopherols were prepared by individually weighing neat compounds using a calibrated analytical balance and dissolving each in methanol. Multianalyte working solutions were prepared from individual stocks and stored at −20°C until use. The internal standard stock solutions were prepared in a similar manner using labeled alpha-tocopherol-(phenyl-13C6) and labeled VEA-(trimethyl-d9) in methanol. A working solution containing both labeled tocopherols was prepared in methanol and stored at −20°C.

Sample Preparation

Aerosol emissions were generated by vaping the EVP liquid samples on a Cerulean CETI-8 e-cigarette vaping machine equipped with button activation switches (Cerulean, Richmond, VA). The vaping machine puff volume was calibrated and verified using a soap bubble flow meter prior to use. Vaping conditions were adopted from the CORESTA Recommended Method No. 81 (CORESTA, Recommended Method Number 81: Routine analytical machine for e-cigarette aerosol generation and collection-definition and standard conditions). Freshly charged batteries were used in an EVP to vape each provided cartridge or liquid. Pre-filled cartridges provided without an EVP were vaped using a Honeystick (Fort Lauderdale, FL, United States) 510 Twist Vape Pen with the battery set at the highest voltage (4 V). When case-associated products included a compatible battery, the provided battery was charged and used at the highest voltage setting for vaping that case-associated cartridge. The aerosol from 15 consecutive puffs from vaped EVPs liquids was trapped on individual pre-conditioned Cambridge filter pads (CFPs; 44 mm) that were purchased from Thermo Fisher Scientific (Waltham, MA, United States) and housed in filter pad holders from Cerulean (Molins PLC, MK, United Kingdom). Custom-made connectors (“mouth pieces”) were fabricated in-house via 3D printing technology as needed for non-circular mouth piece geometries. The total particulate matter (TPM) was gravimetrically determined by mass difference of pre- and post-vaping CFP for each sample. EVPs that produced less than 6.5 mg TPM/15 puffs were considered unacceptably low, flagged as a QC failure, and not reportable. Post-vaped CFPs were individually placed into 16 ml amber vials for extraction. CFPs were extracted with 10 ml of methanol on an orbital shaker for 10 min at 160 rpm. Sample extracts were diluted 100-fold prior to tocopherol analysis. Prior to analysis, 100 µL of the dilute solution was spiked with labeled internal standard and diluted with methanol to 1 ml in an autosampler vial.

Instrumentation

A high-performance liquid chromatography (HPLC) system (Agilent Technologies, Santa Clara, CA, United States) coupled with electrospray tandem mass spectrometry (SCIEX 5500 Triple Quad Applied Biosystems, Foster City, CA, United States) was used to quantitatively measure vitamin E acetate and other tocopherols in trapped aerosol emissions of e-liquids. Chromatographic separation was achieved using isocratic elution at a flow rate of 0.75 ml/min on an XTerra MS C18 column 3.5 µm × 50 mm × 150 mm (Waters Corporation Milford, MA United States) with methanol, water, and formic acid (89.9:10:0.1) as the mobile phase. The eluent from the column was ionized using an electrospray interface to generate and transmit positive ions into the mass spectrometer for selective, quantitative analysis. Analyst software version 1.6.2 (Applied Biosystems, Foster City, CA, United States) was used to operate the HPLC and the 5500 Triple Quad. The mass spectrometer was operated in multiple reaction monitoring (MRM) mode for positive ions. The ion source temperature was set at 350°C and the electrospray ion voltage at 5,500 V. Table 1 presents the optimized MRM transitions used for quantification, confirmation, and internal standard.

TABLE 1
www.frontiersin.org

TABLE 1. MRM transitions and parameters for VEA and tocopherols at a dwell time of 250 ms for all analytes.

Quantification

Analyst software 1.6.2 was used for peak integration, calibration, and quantification. Analyte quantification was achieved using the ratio of relative peak area of the analyte to that of the labeled internal standard. Aerosol emissions samples results (instrument output in ng/mL) were normalized by aerosol puff volume to determine analyte yields per puff following the equation below.

µg per mL aerosol puff volume= tocopherolmeasured (ngmL)x 10 mL1000 ngµg x 15 puff x 55mLpuff

where tocopherol measured is the instrument calculated analyte concentration in ng/mL multiplied by the result of the division of the total sample extraction volume of 10 ml by the total puff aerosol volume defined as 15 puffs/pad × 55 ml/puff and a factor of 1,000 to convert ng to µg.

Quality Control Samples

EVPs are chemically diverse; therefore, we created a diverse surrogate matrix for preparing calibration curves, QC pools and blanks. The surrogate matrix was created by combining the aerosol extract from four different vape liquids, a commercial product (VUSE Solo Menthol), and three custom mixtures (25% squalene/25% squalane/50% mineral oil, 100% CDB oil, and 25% vitamin E acetate/25% medium chain triglycerides/50% hemp oils. The surrogate matrix was vaped using a Vaporin Presidential device and the aerosol trapped using the same technique as for unknown samples. Each pad was extracted with methanol for 10 min in an orbital shaker and combined to produce an 80 ml mixture. The vaped surrogate matrix extract was stored at −20°C. QC samples were prepared daily by individually spiking diluted vaped surrogate matrix extract with known amounts of mixed VEA and tocopherols. Two replicates of a low (QCL) and a high (QCH) level were analyzed per analytical batch. Characterization of each QC level was performed using 20 independent analyses to establish control limits. This QC characterization was subsequently used to evaluate assay performance for each analytical batch based on modified Westgard Rules as described by Caudill et al. (2008). If an analyte failed QC, then none of the results for that analyte in that analysis batch was reportable.

Calibration

Each analytical batch consisted of a set of seven calibration standards prepared in vaped surrogate matrix extract. The calibration was fit to a weighted 1/× least square model for all analytes generating linear curves with r2 > 0.9988. The limit of detection was defined as three times the standard deviation at zero concentration derived from the analysis of six replicates of the three lowest calibration standards (Taylor, 1987). Data are only reported that fall within the calibrated range. Samples exceeding the highest calibration point are diluted and reanalyzed.

Accuracy, Dynamic Range, Linearity, and Precision

Method accuracy was assessed by spiking the vaped surrogate matrix extract at three different levels of VEA and tocopherols. Six replicates of each level, 200, 400, and 600 ng/ml were used to calculate the analyte recovery. The dynamic range selected covers two orders of magnitude (10–1,000 ng/ml equivalent to 1.21 × 10−4–0.0121 µg per mL aerosol emission) to expand the screening capabilities of the assay. Linearity of the dynamic range was evaluated by residual analysis of seven independent curves. Method precision was evaluated as repeatability and intermediate precision of 20 independent QC samples results.

Method Application

EVP liquid samples were transferred to CDC by FDA and various state health departments for aerosols analysis. Samples that did not contain adequate liquid volume for the assay were not analyzed and the contents saved for liquid analyses. Strict chain of custody was maintained throughout the duration of the study. We applied the validated method to analyze aerosol emissions from 147 EVPs associated with the 2019 U.S. EVALI outbreak. EVP liquid samples were transferred to CDC by FDA and various state health departments for aerosols analysis. Strict chain of custody was maintained throughout the duration of the study. Of those 147 samples, a subset of 138 had reportable corresponding nicotine, CBD, and THC levels. These products were categorized as tetrahydrocannabinol (THC) products if THC ≥0.3% (w/w), nicotine products if nicotine >0.2% (w/w), and cannabidiol (CBD) products if CBD >1% (w/w) and THC <0.3%.

Results and Discussion

Method Validation

We developed a sensitive and quantitative method using LC-MSMS to detect VEA and other tocopherols in aerosol emissions of EVALI case-associated EVPs. Complete chromatographic separation was achieved for the tocopherols without any presence of potential matrix interferences as shown in Figure 1. Method specificity was attained by using isotopically labeled tocopherols to establish the presence of unlabeled tocopherols using both the LC retention time and MS/MS mass selection of the triple quad platform.

FIGURE 1
www.frontiersin.org

FIGURE 1. Representative chromatograms of vitamin E acetate (VEA) and other tocopherols spiked into emissions of simulated EVP liquid: (Top) VEA and other tocopherols combined quantitation transitions (500 ng/ml) (bottom) labeled α-tocopherol (200 ng/ml) and deuterated VEA (200 ng/ml).

Method accuracy was analyzed based on six replicates each of un-spiked and spiked vaped surrogate matrix extract at three different levels of VEA and other tocopherols. Spike recoveries (comparison of spiked calculated result to target concentration) and coefficients of variation (CV) were calculated for each spike level. The mean recoveries for VEA and other tocopherols ranged from 100 to 115% with overall CVs of 4–11% as shown in Table 2. We also evaluated matrix effects by analyzing calibration standards in both methanol and vaped surrogate matrix extract. The average slope of six independent calibration curves in solvent and vaped surrogate matrix extract showed a difference of less than 5% for all analytes. These measures document that the method accurately measures VEA and other tocopherols in aerosol emissions of simulated EVP liquids.

TABLE 2
www.frontiersin.org

TABLE 2. Analyte recovery in EVP vaped surrogate matrix extract at three spike levels.

Method precision was evaluated as repeatability and intermediate precision from the analysis of 20 independent results for each of two QC levels, QC low (150 ng/ml) and QC high (800 ng/ml), over 10 days (Table 3). Repeatability of both QC levels ranged from 3.96 to 8.32% for all the analytes. Intermediate precision ranged from 2.47 to 6.73% among all analytes for both QC levels. These data document the excellent precision of the method and the characterization of QCs to allow for evaluation of assay accuracy and precision for each analytical batch analyzed.

TABLE 3
www.frontiersin.org

TABLE 3. Method precision, range, and linearity for VEA and other tocopherols.

The method demonstrates excellent linearity R2 ≥ 0.99 for VEA and other tocopherols within the selected dynamic range of 10–1,000 ng/ml (Table 3). The sensitivity of the method is adequate to measure background levels of VEA (LOD = 1.85 × 10−5 μg per mL of aerosol puff volume (1.53 ng/ml methanol extract)) and other tocopherols (LODs: 8.85 × 10−6–2.28 × 10−5 μg per mL of aerosol puff volume (0.73–1.88 ng/ml methanol extract)) in aerosol emissions of EVPs liquid samples. The sensitivity for VEA detection was significantly better than previously published methods (LC-UV used to achieve an LOD of 580 ng/ml) (Brabcová et al., 2013). Our method was also 3–90 fold more sensitivity for α-tocopherol, δ-tocopherol, and γ-tocopherol compared with previously published methods (Bustamante-Rangel et al., 2007; Lanina et al., 2007; Cortés-Herrera et al., 2019).

Method Application

The analytical method was applied to aerosol emissions from 147 EVPs associated with EVALI cases (Table 4). VEA and α-tocopherol had the highest detection rates of 46 and 41% respectively. VEA content in aerosol emissions ranged from 1.87 × 10−4to 74.1 µg per mL of aerosol puff volume followed by α-tocopherol with a range of 1.47 × 10−2 – 0.908 µg per mL of aerosol puff volume. VEA levels were 264 times higher than α-tocopherol with a mean of 25.1 µg per mL of aerosol puff volume compared to mean of 0.095 µg per mL of aerosol puff volume for α-tocopherol. Further quantification of VEA and α-tocopherol in e-liquid and in vaped aerosol will help provide insight about possible VEA degradation to form reactive byproducts such as ethenone (Wu and O’Shea, 2020). Gamma-tocopherol was detected in five EVPs while δ-tocopherol was not present in any of the analyzed products.

TABLE 4
www.frontiersin.org

TABLE 4. VEA and other tocopherols concentrations and detection frequency in aerosol emissions of EVALI case-associated EVPs (µg per mL aerosol puff volume).

A subset of 139 products were stratified by active ingredient to further investigate the presence of macro-levels of VEA (>0.1%) in different product types. We evaluated macro-levels of VEA because VEA accumulation in the lungs could physically disrupt the tertiary structure of the alveolus, cause alveolar collapse, and subsequently lead to EVALI pathologies (Casals and Cañadas, 2012; Kamal and Raghunathan, 2012; Blount et al., 2020; Jonas and Raj, 2020). Products with higher VEA in aerosol emissions (>0.1% TPM) would deliver significant amounts of VEA to the lungs of people using the products. We show here that no nicotine or CBD products contain these high levels of VEA, and that 71% of case-associated THC products contained VEA as a macro-component (mean 32.0 µg per mL of aerosol puff volume). The high prevalence of VEA in THC products is consistent with the solubility of THC in VEA and the absence of VEA in nicotine products is consistent with the insolubility of nicotine in VEA. This result is also aligned with reported use of VEA as a diluent in the formulation of THC products (Downs, 2019a; Downs, 2019b; Zachary Eisenberg, 2019). VEA was also detected in two products with no THC, CBD, or nicotine with a mean level of 31.7 µg per mL of aerosol puff volume. One of these products was marketed as a THC-containing product by Dank Vapes but contained no detectable THC by our analysis. The high prevalence of macrolevel VEA in EVALI case-associated THC products further implicates VEA as a potential cause of vaping-associated lung injury (Cunha et al., 2006; Cilla et al., 2014).

Conclusion

A rapid, isotope dilution LC-MS/MS method was developed for the simultaneous analysis of VEA and other tocopherols in EVP aerosol emissions. The method demonstrated high accuracy, precision, and sensitivity. VEA and other tocopherols, except for δ-tocopherol, were detected in aerosol emissions from EVALI case-associated EVPs; the mean VEA concentration was several orders of magnitude higher than the mean α-tocopherol concentration. VEA was predominantly found in THC products, consistent with the reported use of VEA as a diluent in the formulation of these products. Our results also indicate that VEA can be efficiently transferred to aerosol by EVALI-associated devices vaped using a standardized protocol. This method can serve as a valuable tool to improve surveillance for the potentially harmful additive VEA in EVPs.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author Contributions

AP and MM contributed to the development and validation of the analytical method. AP wrote the first draft of the manuscript. CW, BB, and LV contributed to conception and design of the study. All authors contributed to manuscript revision, read, and approves the submitted version.

Author Disclaimer

The findings and conclusions in this report are those of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention. Use of trade names in for identification only and does not imply endorsement by the, Centers for Disease Control and Prevention the Public Health Service, or the U.S. Department of Health and Human Services.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

The authors would like to acknowledge Naudia Gray and Jose Perez for their contributions toward the inspection and evaluation, vaping, and sample preparation of EVP sample extracts; Carolina Fernandez for statistical evaluation of data for the CDC 2019 U.S. EVALI outbreak response; and Matt Karwowski, state health departments, EVALI clinicians, and EVALI patients for coordinating and providing EVP samples.

Footnotes

1https://www.cdc.gov/tobacco/basic_information/e-cigarettes

2https://www.ncsl.org/research/health/state-medical-marijuana-laws.aspx

3https://www.fda.gov/news-events/public-health-focus/lung-injuries-associated-use-vaping-products

References

Almeida, M. M., Alves, J. M. P., Patto, D. C. S., Lima, C. R. R. C., Quenca-Guillen, J. S., Santoro, M. I. R. M., et al. (2009). Determination of Tocopheryl Acetate and Ascorbyl Tetraisopalmitate in Cosmetic Formulations by HPLC. HPLC 31, 445–450. doi:10.1111/j.1468-2494.2009.00514.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Bhat, T. A., Kalathil, S. G., Bogner, P. N., Blount, B. C., Goniewicz, M. L., and Thanavala, Y. M. (2020). An Animal Model of Inhaled Vitamin E Acetate and EVALI-like Lung Injury. N. Engl. J. Med. 382, 1175–1177. doi:10.1056/nejmc2000231

PubMed Abstract | CrossRef Full Text | Google Scholar

Bhatnagar, A., Payne, T. J., and Robertson, R. M. (2019). Is There A Role for Electronic Cigarettes in Tobacco Cessation?, J Am Heart Assoc, 8. e012742 doi:10.1161/jaha.119.012742

PubMed Abstract | CrossRef Full Text | Google Scholar

Blount, B. C., Karwowski, M. P., Morel-Espinosa, M., Rees, J., Sosnoff, C., Cowan, E., et al. (2019). Evaluation of Bronchoalveolar Lavage Fluid from Patients in an Outbreak of E-Cigarette, or Vaping, Product Use-Associated Lung Injury - 10 States, August-October 2019. MMWR Morb. Mortal. Wkly. Rep. 68, 1040–1041. doi:10.15585/mmwr.mm6845e2

PubMed Abstract | CrossRef Full Text | Google Scholar

Blount, B. C., Karwowski, M. P., Shields, P. G., Morel-Espinosa, M., Valentin-Blasini, L., Gardner, M., et al. (2020). Vitamin E Acetate in Bronchoalveolar-Lavage Fluid Associated with EVALI. N. Engl. J. Med. 382, 697–705. doi:10.1056/nejmoa1916433

PubMed Abstract | CrossRef Full Text | Google Scholar

Brabcová, I., Kovářová, L., Satinsky, D., Havlíková, L., and Solich, P. (2013). A Fast HPLC Method for Determination of Vitamin E Acetate in Dietary Supplements Using Monolithic Column. Food Anal. Methods 6, 380–385. doi:10.1007/s12161-012-9452-0

CrossRef Full Text | Google Scholar

Bustamante-Rangel, M., Delgado-Zamarreño, M. M., Sánchez-Pérez, A., and Carabias-Martínez, R. (2007). Determination of Tocopherols and Tocotrienols in Cereals by Pressurized Liquid Extraction-Liquid Chromatography-Mass Spectrometry. Analytica Chim. Acta 587, 216–221. doi:10.1016/j.aca.2007.01.049

CrossRef Full Text | Google Scholar

Butt, Y. M., Smith, M. L., Tazelaar, H. D., Vaszar, L. T., Swanson, K. L., Cecchini, M. J., et al. (2019). Pathology of Vaping-Associated Lung Injury. N. Engl. J. Med. 381, 1780–1781. doi:10.1056/nejmc1913069

PubMed Abstract | CrossRef Full Text | Google Scholar

Casals, C., and Cañadas, O. (2012). Role of Lipid Ordered/disordered Phase Coexistence in Pulmonary Surfactant Function. Biochim. Biophys. Acta (Bba) - Biomembranes 1818, 2550–2562. doi:10.1016/j.bbamem.2012.05.024

CrossRef Full Text | Google Scholar

Caudill, S. P., Schleicher, R. L., and Pirkle, J. L. (2008). Multi-rule Quality Control for the Age-Related Eye Disease Study. Statist. Med. 27, 4094–4106. doi:10.1002/sim.3222

PubMed Abstract | CrossRef Full Text | Google Scholar

Centers for Disease Control and Prevention.Electronic Cigarettes (E-cigarettes) [Online]. Available: https://www.cdc.gov/tobacco/basic_information/e-cigarettes. [Accessed].

Google Scholar

Cilla, A., Zanirato, V., Rodriguez-Estrada, M. T., and Garcia-Llatas, G. (2014). “Nutriential Hazards: Micronutrients: Vitamins and Minerals,” in Encyclopedia of Food Safety. Editor Y. Motarjemi (Waltham: Academic Press), 86–94. doi:10.1016/b978-0-12-378612-8.00431-5

CrossRef Full Text | Google Scholar

Coresta Recommended Method Number 81: Routine analytical machine for e-cigarette aerosol generation and collection-definition and standard conditions.

Google Scholar

Cortés-Herrera, C., Chacón, A., Artavia, G., and Granados-Chinchilla, F. (2019). Simultaneous LC/MS Analysis of Carotenoids and Fat-Soluble Vitamins in Costa Rican Avocados (Persea Americana Mill.), Molecules 24. 4517, doi:10.3390/molecules24244517

CrossRef Full Text | Google Scholar

Creamer, M. R., Wang, T. W., Babb, S., Cullen, K. A., Day, H., Willis, G., et al. (2019). Tobacco Product Use and Cessation Indicators Among Adults - United States, 2018. MMWR Morb. Mortal. Wkly. Rep. 68, 1013–1019. doi:10.15585/mmwr.mm6845a2

PubMed Abstract | CrossRef Full Text | Google Scholar

Cunha, S. C., Amaral, J. S., Fernandes, J. O., and Oliveira, M. B. P. P. (2006). Quantification of Tocopherols and Tocotrienols in Portuguese Olive Oils Using HPLC with Three Different Detection Systems. J. Agric. Food Chem. 54, 3351–3356. doi:10.1021/jf053102n

PubMed Abstract | CrossRef Full Text | Google Scholar

Downs, D., David, H., and Barcott, Bruce. (2019b). Journey of a Tainted Cartridge; From China's Labs to Your Lungs [Online]. Leafly. Available: https://www.leafly.com/news/politics/vape-pen-injury-supply-chain-investigation-leafly.[Accessed].

Google Scholar

Downs, D. (2019a). Vape Pen Lung Injury: Here's what You Need to Know. [Online]. Leafly. Available: https://www.leafly.com/news/health/vape-pen-lung-disease-advice-consumers. [Accessed].

Google Scholar

Duffy, B., Li, L., Lu, S., Durocher, L., Dittmar, M., Delaney-Baldwin, E., et al. (2020). Analysis of Cannabinoid-Containing Fluids in Illicit Vaping Cartridges Recovered from Pulmonary Injury Patients: Identification of Vitamin E Acetate as a Major Diluent. Toxics 8. 8, doi:10.3390/toxics8010008

PubMed Abstract | CrossRef Full Text | Google Scholar

Ghosh, S., and Drummond, M. B. (2017). Electronic Cigarettes as Smoking Cessation Tool 23, 111–116. doi:10.1097/mcp.0000000000000348

CrossRef Full Text | Google Scholar

Hajek, P., Phillips-Waller, A., Przulj, D., Pesola, F., Myers Smith, K., Bisal, N., et al. (2019). A Randomized Trial of E-Cigarettes versus Nicotine-Replacement Therapy. N. Engl. J. Med. 380, 629–637. doi:10.1056/nejmoa1808779

PubMed Abstract | CrossRef Full Text | Google Scholar

Health, N. Y. S. D. O. (2019). New York State Department of Health announces update on investigation into vaping-associated pulmonary illnesses: department warns against use of black-market vaping products: lab test results show high levels of vitamin E acetate, now focus of investigation [Online]. Available: https://www.health.ny.gov/press/releases/2019/2019-09-05_vaping.htm. [Accessed].

Google Scholar

Jonas, A. M., and Raj, R. (2020). Vaping-Related Acute Parenchymal Lung Injury. Chest 158, 1555–1565. doi:10.1016/j.chest.2020.03.085

PubMed Abstract | CrossRef Full Text | Google Scholar

Kamal, M. A., and Raghunathan, V. A. (2012). Modulated Phases of Phospholipid Bilayers Induced by Tocopherols. Biochim. Biophys. Acta (Bba) - Biomembranes 1818, 2486–2493. doi:10.1016/j.bbamem.2012.06.016

CrossRef Full Text | Google Scholar

Lanina, S. A., Toledo, P., Sampels, S., Kamal-Eldin, A., and Jastrebova, J. A. (2007). Comparison of Reversed-phase Liquid Chromatography-Mass Spectrometry with Electrospray and Atmospheric Pressure Chemical Ionization for Analysis of Dietary Tocopherols. J. Chromatogr. A 1157, 159–170. doi:10.1016/j.chroma.2007.04.058

PubMed Abstract | CrossRef Full Text | Google Scholar

Lozier, M. J., Wallace, B., Anderson, K., Ellington, S., Jones, C. M., Rose, D., et al. (2019). Update: Demographic, Product, and Substance-Use Characteristics of Hospitalized Patients in a Nationwide Outbreak of E-Cigarette, or Vaping, Product Use-Associated Lung Injuries - United States, December 2019. MMWR Morb. Mortal. Wkly. Rep. 68, 1142–1148. doi:10.15585/mmwr.mm6849e1

PubMed Abstract | CrossRef Full Text | Google Scholar

Lung Injuries Associated with Use of Vaping Products. Food and Drug Administration. Available: https://www.fda.gov/news-events/public-health-focus/lung-injuries-associated-use-vaping-products. [Accessed].

Google Scholar

Matsumoto, S., Fang, X., Traber, M. G., Jones, K. D., Langelier, C., Hayakawa Serpa, P., et al. (2020). Dose-Dependent Pulmonary Toxicity of Aerosolized Vitamin E Acetate. Am. J. Respir. Cel Mol Biol 63, 748–757. doi:10.1165/rcmb.2020-0209oc

PubMed Abstract | CrossRef Full Text | Google Scholar

Mcnamara, A. (2020). CBS News, These states now have legal weed, and which states could follow suit in 2020 [Online]. Available: https://www.cbsnews.com/news/where-is-marijuana-legal-in-2020-illinois-joins-10-other-states-legalizing-recreational-pot-2020-01-01/. [Accessed 2020].

Google Scholar

Moritz, E. D., Zapata, L. B., Lekiachvili, A., Glidden, E., Annor, F. B., Werner, A. K., et al. (2019). Update: Characteristics of Patients in a National Outbreak of E-Cigarette, or Vaping, Product Use-Associated Lung Injuries - United States, October 2019. MMWR Morb Mortal Wkly Rep. 68, 985–989. doi:10.15585/mmwr.mm6843e1

PubMed Abstract | CrossRef Full Text | Google Scholar

Nada, A., Krishnaiah, Y. S., Zaghloul, A. A., and Khattab, I. (2010). Analysis of Vitamin E in Commercial Cosmetic Preparations by HPLC. J. Cosmet. Sci. 61, 353–365.

PubMed Abstract | Google Scholar

National Conference of States Legislature State Medical Marijuana Laws [Online]. Available: https://www.ncsl.org/research/health/state-medical-marijuana-laws.aspx. [Accessed October 12, 2020].

Google Scholar

Orzabal, M., and Ramadoss, J. (2019). Impact of Electronic Cigarette Aerosols on Pregnancy and Early Development. Curr. Opin. Toxicol. 14, 14–20. doi:10.1016/j.cotox.2019.05.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Pacula, R. L., and Smart, R. (2017). Medical Marijuana and Marijuana Legalization. Annu. Rev. Clin. Psychol. 13, 397–419. doi:10.1146/annurev-clinpsy-032816-045128

PubMed Abstract | CrossRef Full Text | Google Scholar

Sadrykia, F., Shayanfar, A., Valizadeh, H., and Nemati, M. (2019). A Fast and Simple Method for Determination of Vitamin E in Infant Formula by Dispersive Liquid-Liquid Microextraction Combined with HPLC-UV. Food Anal. Methods 12, 23–31. doi:10.1007/s12161-018-1331-x

CrossRef Full Text | Google Scholar

Seker, M. E., Çelik, A., and Dost, K. (2012). Determination of Vitamin E Isomers of Grape Seeds by High-Performance Liquid Chromatography-UV Detection. J. Chromatogr. Sci. 50, 97–101. doi:10.1093/chromsci/bmr032

PubMed Abstract | CrossRef Full Text | Google Scholar

Shahab, L., Goniewicz, M. L., Blount, B. C., Brown, J., Mcneill, A., Alwis, K. U., et al. (2017). Nicotine, Carcinogen, and Toxin Exposure in Long-Term E-Cigarette and Nicotine Replacement Therapy Users. Ann. Intern. Med. 166, 390–400. doi:10.7326/m16-1107

PubMed Abstract | CrossRef Full Text | Google Scholar

Siegel, D. A., Jatlaoui, T. C., Koumans, E. H., Kiernan, E. A., Layer, M., Cates, J. E., et al. (2019). Update: Interim Guidance for Health Care Providers Evaluating and Caring for Patients with Suspected E‐cigarette, or Vaping, Product Use Associated Lung Injury - United States, October 2019. Am. J. Transpl. 19, 3420–3428. doi:10.1111/ajt.15690

PubMed Abstract | CrossRef Full Text | Google Scholar

Taylor, J. K. (1987). Quality Assurance of Chemical Measurements. New York: Lewis Publishers.

Wang, T. W., Neff, L. J., Park-Lee, E., Ren, C., Cullen, K. A., and King, B. A. (2020). E-cigarette Use Among Middle and High School Students - United States, 2020. MMWR Morb. Mortal. Wkly. Rep. 69, 1310–1312. doi:10.15585/mmwr.mm6937e1

PubMed Abstract | CrossRef Full Text | Google Scholar

Wu, D., and O’Shea, D. F. (2020). Potential for Release of Pulmonary Toxic Ketene from Vaping Pyrolysis of Vitamin E Acetate. Proc. Natl. Acad. Sci. USA 117, 6349–6355. doi:10.1073/pnas.1920925117

PubMed Abstract | CrossRef Full Text | Google Scholar

Zachary Eisenberg, D. M., Lam, Vu., Cheng, Cathy., Richard, Josh., and Burack, Benjamin. (2019). Contaminant Analysis of Illicit vs Regulated Market Extracts. [Online]. anresco Laboratories. Available: https://cannabis.anresco.com/analysis-of-illicit-vs-regulated-market-extracts/. [Accessed 2020].

Google Scholar

Zhu, Sh., Bonnevie, E., Sun, J. Y., Cummins, S. E., Gamst, A., Yin, L., et al. (2014). Four Hundred and Sixty Brands of E-Cigarettes and Counting: Implications for Product Regulation. Tob. Control. 23, iii3-iii9. doi:10.1136/tobaccocontrol-2014-051670

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: vitamin E acetate, tocopherols, EVP aerosol, LC-MS/MS, EVALI

Citation: Puetz A, Morel Espinosa M, Watson C, Blount BC and Valentín-Blasini L (2021) Development, Validation, and Application of a Novel Method for the Analysis of Vitamin E Acetate and Other Tocopherols in Aerosol Emissions of E-Cigarettes, or Vaping Products Associated With Lung Injury. Front. Chem. 9:730954. doi: 10.3389/fchem.2021.730954

Received: 25 June 2021; Accepted: 26 July 2021;
Published: 05 August 2021.

Edited by:

Alberto Salomone, University of Turin, Italy

Reviewed by:

Camilla Montesano, Sapienza University of Rome, Italy
Gianpiero Adami, University of Trieste, Italy

Copyright © 2021 Puetz, Morel Espinosa, Watson, Blount and Valentín-Blasini. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Maria Morel Espinosa, MMorelEspinosa@cdc.gov

Download