Abstract
Objective:
Identification of fitting and programming considerations for sound therapy in hearing aids based on electroacoustic characterization across brands, including assessment of interactions between sound therapy and other digital signal processing (DSP) features.
Methods:
The study was conducted in two parts: (1) electroacoustic characterization of default sound therapy settings, and (2) analysis of interactions between the sound generator, feedback cancellation, and noise reduction features. Five receiver-in-the-ear hearing aids from different manufacturers were assessed using the Ahead Simulations Clinical Assistant for Research and Learning (CARL) manikin and the Audioscan Verifit 2 hearing aid analyzer. Devices were programmed for two standard audiograms using both dome and custom earmold couplings.
Results:
Default sound therapy outputs varied significantly across brands, influenced primarily by audiogram and coupling type. Feedback testing revealed that enabling sound therapy caused devices to exceed their stable gain limits, resulting in audible feedback. Noise reduction algorithms reduced sound therapy signals in open-fit configurations.
Conclusions:
Sound therapy features differ across hearing aids and can interact with other DSP functions counterproductively. These interactions are measurable using standard clinical verification tools, highlighting the importance of verification of hearing aid programs. Five clinical fitting practices were developed to address these considerations.
1 Introduction
Chronic subjective tinnitus is a condition in which individuals experience the sensation of sound without an external source, and which can vary in degree of intrusiveness (). Although the pathophysiology of tinnitus is still disputed, tinnitus is thought to involve altered auditory or somatosensory inputs, and abnormal activity in central nervous structures (; ). Sound therapy is considered a potential treatment for individuals with chronic subjective tinnitus, including those with concurrent hearing loss (; ; ). Sound therapy can include providing sounds to the patient that either reduce the contrast of the tinnitus, distract attention from tinnitus, or are subjectively soothing (). In modern hearing aids, these may be available within the hearing aid itself or through connected software, reducing the need to provide sound therapy through specialized devices. The sound therapy approaches may aim to decrease perception of tinnitus, promote patient control and emotional regulation, enable habituation to tinnitus, and/or contribute to changes in the tinnitus neural pathways (). Sound therapy can be delivered in different forms and has been shown to provide effective intervention for tinnitus (; ; ).
Tinnitus is frequently associated with sensorineural hearing loss and steeply sloping audiograms (; ); as such, the use of hearing aids as an external sound generator is common in audiological care for tinnitus. Alongside standard amplification, modern hearing aids often include an onboard sound generator that can be programmed to deliver broadband noise as a background sound for tinnitus masking—one form of sound therapy for tinnitus intervention (; ). This feature is increasingly available in modern hearing aid technology, with devices that have the capability to provide both environmental amplification and sound therapy being referred to as combination devices. Recommended practice includes the individualized fitting and verification of hearing aids to meet patients' needs (). However, the availability, options, and customization capabilities of these sound-generating features vary across manufacturers, and this variability has not been well characterized. To address this, the primary aims of this study are to (1) investigate the electroacoustic characterizations of default sound therapy settings across different hearing aid brands, hearing loss severity, and coupling types, and (2) develop a concise and actionable set of considerations for fitting, programming, and verification of hearing aid tinnitus sound therapy (; ).
The purpose of the following investigations is to address two primary aims: Aim 1: Investigating between-product variability in sound therapy and interactions with digital signal processing features. Variability in the implementation of sound therapy has been demonstrated by ), who examined the white-noise characteristics across three different hearing aids. The researchers found notable differences in both intensity and frequency response across brands. They also observed that software settings for broadband noise often did not match the hearing aid's actual output as measured by verification in an Audioscan Verifit 2 hearing instrument test box (). This study was limited in scale to a small selection of hearing aids with dome coupling and a single hearing loss configuration, leaving unanswered questions about how default sound settings vary across different sloping hearing loss profiles and coupling types. Additionally, there is little research on how broadband noise–based sound therapy interacts with other digital signal processing (DSP) features of hearing aids. Manufacturer instructional guides either do not specify whether such interactions exist or outline other programming considerations, leaving clinicians to navigate these complexities independently. In a review article, ) developed a framework to objectively compare hearing aids regardless of brand or style. The framework details a set of essential hearing aid features that belong to domains such as signal processing, comfort, and adaptation (). Of these domains, sound therapy was not specifically categorized. As a result, we selected features from this framework and applied them in this study to assess their potential interactions with sound therapy, specifically noise reduction and an active feedback manager. Noise reduction (NR) was selected because we hypothesized that sound leakage from the ear canal during sound therapy may be captured by the hearing aid microphones and processed by the NR algorithm. As the sound therapy is a steady-state broadband noise, it is possible that basic NR algorithms are capable of reducing this type of signal (). This interaction has not been well documented in the literature. Furthermore, as the ear canal sound pressure level is a major contributor to audible feedback, we hypothesized that the additive effects of sound therapy with environmental sound amplification may exceed stable gain limits (). This warrants investigation, as audible feedback is a deterrent to consistent hearing aid use (). The present study addressed the following research questions: (a) How do the default intensity levels of manufacturer-programmed sound therapy vary across coupling types and hearing loss configurations? (b) Does digital noise reduction reduce the output level of sound therapy in the ear canal? (c) Does sound therapy delivered adjacent to amplification exceed maximal stable gain and result in audible feedback? Aim 2: Adapting Existing Fitting Guidelines for Tinnitus Sound Therapy. Despite the current evidence that combination devices are effective for tinnitus management, it has been suggested that too few patients receive sound therapy, leaving unmet needs (, ). One way to enhance consistent provision of care may be to further specify the clinical care pathway for providing sound therapy in hearing aids. There are several tinnitus management protocols, such as Progressive Tinnitus Management and Tinnitus Retraining Therapy, which outline the goals and general provision of sound therapy (; ). However, these guidelines may not provide specific steps on how to combine hearing aid provision with sound therapy. In 2020, the British Society of Audiology published an audiological practice guide for fitting combination hearing aids for subjects with tinnitus; in this guide, the authors emphasize the device user's preferences, highlighting the individualized approach required for tinnitus management. As with other tinnitus management guidelines, the clinical procedures for fitting, programming, and verifying sound therapy are not specified (). This contrasts with hearing aid fitting guidelines, which typically recommend using prescriptive targets and electroacoustic verification to provide person-specific sound levels (). The findings of this study should be interpreted within the context of hearing aid fittings for individuals with hearing loss and may not generalize to sound therapy delivered by hearing aids in individuals with normal hearing sensitivity.
2 Materials and methods
2.1 Study design
An electroacoustic investigation was conducted on commercially available hearing aids released from 2020 to 2023, using Ahead Simulations' Clinical Assistant for Research and Learning (CARL) basic manikin as a simulated ear (; ). An Audioscan Verifit 2 hearing aid analyzer was used to collect on-ear measures and verification of settings ().
2.2 Simulated hearing loss
Two standard audiograms were selected: N4, a moderate sloping to moderately-severe hearing loss; and S2, a normal hearing sensitivity sloping to profound hearing loss (). These hearing losses were selected because individuals with sloping hearing losses are likely to have tinnitus (). These hearing losses also represent clinically challenging fittings due to their combined candidacy for both venting and high gain, which can create the potential for feedback. For a sloping hearing loss, a typical hearing aid fitting would provide greater gain in the high-frequency regions where hearing sensitivity is reduced. Because audible feedback is commonly associated with frequencies between 2 and 4 kHz, feedback-management features are incorporated into the fitting to minimize feedback while maintaining adequate high-frequency amplification. Furthermore, individuals with sloping losses have better sensitivity in low frequencies compared to high frequencies, larger venting options are typically selected to improve comfort and sound quality. However, larger vents may also allow excessive unwanted noise into the ear canal. As a result, noise reduction features are often incorporated into these fittings to help mitigate the unwanted noise. The left ear hearing aid was programmed to the N4 hearing loss, and the right ear hearing aid was programmed to the S2 hearing loss.
2.3 Hearing aid fittings
Receiver-in-the-canal hearing aids from five different manufacturers were selected for this electro-acoustic analysis using medium-powered receivers. The hearing aids were selected to represent independent technology offerings: Resound, Starkey, Oticon, Widex, and Phonak. Brand-specific program settings and hearing aid models are listed in Table 1; performance data will be anonymized for the remainder of this paper. Two coupling methods were also used for each hearing aid: an open-fitting with a single-walled dome and a closed-fitting with a custom earmold tip, made for each brand of hearing aid from ear impressions taken from the CARL manikin's ear (size large for the right, and bendy for the left).The different types of ears used on CARL are representative of the anatomical variability observed across individuals, both types are validated against a sample of human participants (). Differences between ear canal anatomies is compensated with individually measured real ear coupler differences (RECD) for this study. The monaural hearing thresholds were entered into the Verifit system in dB HL, which were then converted to dB SPL (eardrum) using individualized manikin RECD values in combination with the appropriate reference equivalent threshold sound pressure level (RETSPL) values for insert earphones at each test frequency (; ). The manikin was mounted on a swing arm that was positioned 50 cm from the Verifit 2 speakers, as per positioning recommendations (). The position of the swing arm did not change between measures or when changing hearing aid to ensure uniformity of recording environment.
Table 1
| Brand | Model | Noise reduction settings | Feedback manager settings | Sound therapy settings | |
|---|---|---|---|---|---|
| Resound | Nexia | Per environment | Mild | Shaping setting: High frequency noise | |
| Quiet: - × 3 | Software level | ||||
| Speech (soft): - × 3 | Left: 55 dB SPL | Right: 55 dB SPL | |||
| Speech (loud): - × 4 | |||||
| Speech in noise (moderate): - × 6 | |||||
| Speech in noise (loud): - × 6 | |||||
| Noise (moderate): - × 8 | |||||
| Noise (loud): - × 9 | |||||
| Starkey | Genesis AI | Consonant Enhancement: 2 | Subtle adaptation | Shaping setting: Audiogram | |
| Sound Enhancement: Speech in Noise: 3, Quiet: | Software level | ||||
| Situational Sound Management: | Left: | Right: | |||
| Transients: 3, Binaural Wind: 3; Binaural machine: 3 | Low Fq: 65 | Low Fq: 41 | |||
| Mid Fq: 66 | Mid Fq: 41 | ||||
| High Fq: 63 | High Fq: 41 | ||||
| Oticon | More | Transient noise management: Medium | Normal | Shaping setting: Shaped | |
| Spatial noise Management: ON | Software level | ||||
| Binaural BroadBand: ON | Left: | Right: | |||
| Low Fq: 70 dB SPL Mid Fq: 75 dB SPL High Fq: 86 dB SPL | Low Fq: 37 dB SPL Mid Fq: 60 dB SPL High Fq: 86 dB SPL | ||||
| Widex | Moment | RT speech enhancer: ON | Midway between Less risk and more gain | Shaping setting: Noise | |
| Max soft noise reduction: ON | Software level | ||||
| TruSound softener: ON | Left: | Right: | |||
| Wind noise attenuation: ON | 12 steps from “minimum” | 12 steps from “minimum” | |||
| Phonak | Lumity | Auto programs (calm, speech in noise, speech in loud noise, speech in car): | Whistle block: 13 (moderate) | Shaping setting: Hearing loss | |
| Sound Relax: 8 (weak) | Software Level | ||||
| Noise Block: 8 (weak) | Left: | Right: | |||
| Windblock: 12 (moderate) | 77 dB A | 72 dB A | |||
| Auto program Comfort in Noise: | |||||
| Sound Relax: 8 (weak) | |||||
| Noise Block: 12 (moderate) | |||||
| Windblock: 16 (moderate) | |||||
Brand-specific model and programming settings, where the left hearing aid was programmed to the N4 Audiogram and the right hearing aid was programmed to the S2 audiogram ().
Values and units are reported as they appear on the software
A single dome type was selected for use on all hearing aids as a partial control condition. Unlike custom earmolds, which are receiver-specific and would not universally fit each manufacturer's receiver type, the dome we selected fit every receiver reasonably. This approach was not ideal and could not be a complete control condition, as small deviations in the fit around the receiver could have caused slit leaks for sound to escape. This dome was selected to minimize extraneous variability associated with manufacturer-specific dome designs. Specifically, the selected dome (1) represents the average vented dome available across manufacturers, as compared to the REOIG measures from ) and average insertion loss from Winther ), (2) is a size and material that is compatible with the manikin ear and (3) which could be reliably and consistently fit to the same position inside the manikin ear. For each manufacturer's software, we selected the corresponding single-vented dome type. Although using a non-brand-specific dome may have altered the frequency response of some hearing aid receivers, amplification gain was verified to be clinically appropriate for a DSL v5 adult fitting. The differences observed across brands reflect the influence of coupling type itself on sound therapy and were therefore considered relevant to the study objectives. The earmold vent size was selected based on the manufacturer's software recommendations, which ranged from completely occluded to 3 mm vents. The openness of fit was measured by insertion gain, specifically insertion loss due to the attenuation of the coupling (Real ear occluded gain, REOG), calculated by comparing the Real Ear Occluded Response (REOR) to the Real Ear Unaided Response (REUR; REOG = REOG–REUG). All measurements were made using pink noise at 65 dB SPL. Measures were reviewed for accuracy in probe tube depth and acoustic seal prior to taking final measurements. On average, the custom earmolds reduced the gain from 250 to 8,000 Hz by 13.7 dB (range: −1.27 to −21.55 dB) dB for the S2 audiogram and 7.4 dB (range: −0.61 to −13.48 dB) for the N4 audiogram, exceeding the attenuation of the domes, which attenuated about 6.7 dB and 3.6 dB for the S2 and N4 fittings, respectively. Overall, the earmolds attenuated more sound compared to the dome except for in Brand D (Table 2), which was a more open fit than the dome, with average attenuation of 1.3 and 0.6 dB for S2 and N4 fittings, respectively. A contributing factor to Brand B appearing more open than the dome condition may be slit leakage created by the fit of the receiver within the earmold. A number of factors can influence the effective acoustic openness of the fit, including the insertion depth and positioning of the receiver within the ear canal, venting characteristics of the coupling, and the placement of the probe tube during measurement (). Collectively, these factors may have contributed to increased acoustic leakage and a more open REUR/REOR response in this condition.
Table 2
| Fitting | N4 audiogram | S2 audiogram | ||
|---|---|---|---|---|
| Vent size | Real ear occluded insertion gain (dB) | Vent Size | Real ear occluded insertion gain (dB) | |
| Single-wall dome | N/A | −3.6 | N/A | −6.73 |
| Brand A EM | 1.8 mm | −9.01 | 2.4 mm | −8.51 |
| Brand B EM | No vent | −10.44 | No vent | −21.55 |
| Brand C EM | No vent | −13.48 | 3 mm | −16.63 |
| Brand D EM | No vent | −0.61 | No vent | −1.27 |
| Brand E EM | 0.7 mm | −3.51 | No vent | −20.52 |
Openness of fit for a single-walled dome compared to custom RIC earmolds (EM).
Occlusion values were computed as the change from the Real Ear Unaided Response to the Real Ear Occluded Response, averaged across 250 through 8,000 Hz (REOIG = REOG –REUG).
All hearing aids were programmed to binaural DSL v5 adult prescriptive targets (). The ear canal acoustics of the CARL manikin were measured using the real-ear-to-coupler difference (RECD) with a foam tip, as incorporated into the prescription. The DSL prescription was selected as it was found to be preferred by hearing aid users with tinnitus compared to other prescriptive methods (; ). Hearing aid output was measured on-ear using the Audioscan Verifit 2, and the fit to target was verified to be within 5 dB root mean square error (RMSE; ; ). Probe tube placement was verified to be within 5 mm of the manikin eardrum prior to each hearing aid fitting using the Probe Tube Guide feature on the Verifit 2.
2.4 Sound therapy programming
The hearing aids offer a variety of broadband noise stimuli, most of which are generated using Gaussian noise and can be spectrally shaped (e.g., white noise, pink noise, or speech-shaped noise). For this study, the noise stimulus was selected to approximate participants' hearing thresholds whenever this option was available within the manufacturer's fitting software. This feature had different names, such as “Audiogram shaped”, “Hearing loss shaped”, “High frequency noise”, or simply “Shaped”. This shaping was selected as ) had previously investigated white noise sound therapy, and to assess the effects on intensity levels in higher frequencies of the sound due to the sloping hearing loss. However, Brand E did not provide a hearing-threshold-shaped noise option; therefore, white noise was selected as the closest available alternative. Other than enabling these programs and selecting this custom shaping, no other parameters of the sound therapy were modified from the software's pre-selected settings, such as intensity and modulation. A list of software and programming parameters for sound therapy and digital sound processing features can be found in Table 1.
2.5 Electroacoustic measures of default sound therapy settings
The Speech-live feature on the Audioscan Verifit 2 was used to capture the frequency response of the default sound therapy. Speech-live is a measure that records the hearing aid output in the absence of an external signal for a duration of approximately 12 s (). A baseline recording was made with the hearing aids set to the default sound therapy program with noise reduction disabled. The overall level of the sound therapy was then calculated from 1/3 octave bands using power-sum analysis from 200 to 8,000 Hz, in dB SPL (eardrum).
2.6 Noise reduction measures
To evaluate the potential interaction between NR algorithms and sound therapy, Speech-live was used to record the frequency response of the sound therapy under various noise reduction conditions. The first condition was sound therapy with the noise reduction setting disabled, followed by sound therapy with the noise reduction feature enabled at default settings. For the second condition, the NR was enabled prior to recording. This was to ensure that the algorithm had time to activate and reach an asymptotic performance before the Speech-live recording (). A timer was used to standardize this interval across recordings, at approximately 20 s per recording. The recordings were made multiple times for each hearing aid to ensure stability, but only the final recording was used for data analysis. Both power sum averaging and arithmetic averaging across spectral level differences. Overall levels were calculated for conditions from 1/3-octave bands. The calculations together offered a more comprehensive characterization of the effects of NR, capturing both overall level changes and frequency-specific effects. This difference between conditions represents the reduction of sound therapy caused by the NR algorithm in dB SPL (ear drum). This measurement was possible in all hearing aids except for Brand E, where the software prevented enabling noise reduction settings while sound therapy was active.
2.7 Active feedback manager measures
To assess active feedback manager capability, the maximum stable gain was measured for each hearing loss configuration and coupling type. Similar to the experiment by ), a baseline measure was first obtained without sound therapy (MSG Baseline); the global gain was increased in 1 dB increments while a researcher monitored the hearing aid output in CARL's ear canal via headphones connected to the Verifit. To simulate a typical clinical listening check, researchers held their hands near the hearing aids, altering the feedback path, in order to detect sustained squealing while a 65 dB SPL speech passage played continuously from the Verifit. Once audible feedback was detected, the global gain was then reduced until feedback ceased, and the level was lowered an additional 2 dB to determine the maximal stable gain. The output was recorded using the same speech passage. In the follow-up measure, sound therapy was enabled (MSGST), and the same procedure as at baseline was performed, ensuring that only the amplification gain was manipulated and the sound therapy level remained constant. Maximum stable gain for 2,000, 3,000, and 4,000 Hz was calculated by subtracting Real-Ear Unaided Response from the Real-Ear Aided Response obtained (i.e., Real-ear Insertion Gain; ) for both conditions.
To quantify the effect of sound therapy on the presence of audible feedback, the Additional Gain Before Feedback (AGBF) measure was computed, following published procedures (). The AGBF is simply a calculation of the difference between the maximum stable gain measured with the tinnitus sound therapy and the unadjusted hearing aid gain set to DSLv5 adult targets alongside sound therapy. Negative values indicate that amplification reductions were required to maintain feedback stability when sound therapy was enabled. The effect of sound therapy on available headroom was evaluated across the full frequency range, as the sound therapy stimuli contained low-frequency energy in addition to mid- and high-frequency components.
2.8 Analysis of results
Recall that the aims of the study was to explore the range of between-product variability of interactions between signal processing and tinnitus sound therapy, to better inform fitting guideline modifications. As such, a small illustrative sample of fully-verified fittings was collected. This sample does not support inferential statistics, so descriptive analyses of range of variation will be presented below. Where variations exceed commonly-accepted clinical tolerances, clinical strategies in the form of modified fitting protocols will be developed.
3 Results
3.1 Variability of default sound therapy settings
Default sound therapy levels varied across brands, depending on the coupling type and hearing-loss configuration. Figure 1 presents the 1/3-octave band output levels for each sound therapy stimulus. To facilitate comparison with audiometric thresholds, the frequency axis is labeled using the standard audiometric octave frequencies in accordance with recommendations from the (). Between-brand variability was substantial, with differences in default levels (lowest to highest) of 18.7 dB SPL (ear drum) for S2 dome coupling (Figure 1, panel A), 25.0 dB SPL (ear drum) for S2 earmold coupling (Figure 1, panel B), 22.1 dB SPL (ear drum) for dome-fit N4 (Figure 1, panel C), and 22.3 dB SPL (ear drum) for earmold-fit N4 (Figure 1, panel D).
Figure 1
For the S2 audiogram, dome couplings yielded intensities between 59.8 and 78.5 dB SPL (eardrum; mean = 71.6 dB SPL, SD = 9.4), while earmold couplings ranged from 59.6 to 84.6 dB SPL (eardrum; mean = 70.8 dB SPL, SD = 11.3). For the N4 audiogram, dome couplings produced sound therapy intensities ranging from 69.2 to 91.3 dB SPL (eardrum; mean = 83.4, SD =8.8), whereas earmold couplings ranged from 65.4 to 88.1 dB SPL (eardrum; mean = 79.1 dB SPL, SD = 8.7).
Overall, the coupling that produced higher in-ear sound therapy levels depended on both brand and audiometric configuration. For the N4 audiogram, dome fittings yielded, on average, 4.3 dB SPL (eardrum) higher sound therapy levels than earmold fittings, with differences ranging from −1.3 to 14.1 dB SPL (eardrum; SD = 5.9). For S2 audiograms, results were mixed across brands, with an average difference between dome and earmold couplings of 0.8 dB SPL (eardrum), but a range of −6.1 to 10.3 dB SPL (eardrum; SD = 6.3).
3.2 Interactions between noise reduction and sound therapy
Differences between power-sum averaging and arithmetic averaging of spectral level differences were present when analyzing the effects of NR on sound therapy. Each approach has distinct advantages and limitations: power-sum averaging places greater emphasis on larger spectral differences and is more representative of perceived overall level changes, whereas arithmetic averaging is more sensitive to localized spectral differences but may be influenced by low-level spectral regions that contribute less to perceived loudness. Both metrics were presented in Figure 2. Positive values reflected a reduction in sound therapy output due to the NR algorithm, whereas negative values indicated an increase in sound therapy level, and likely minimal interaction between the noise reduction algorithm and sound therapy.
Figure 2

Difference in sound therapy levels calculated using both power-sum averaging and spectral averaging, with and without noise reduction algorithms enabled. Positive values indicate that the sound therapy was reduced in level. Data is shown by brand, coupling type, and audiogram. *Note: these data exclude Brand E which would not allow the noise reduction to be enabled while the sound therapy was enabled.
Examining the interaction through power-sum averaging revealed that Brands A and D were least affected, with less than 2 dB change in sound therapy levels between conditions and Brands B and C were observed to have varying effects on the sound therapy level, depending on both the audiogram and the coupling type. Across all brands, the N4 audiogram overall level changes ranged from −0.1 to 7.2 dB (mean = 1.6, SD = 3.2) for earmold fittings and from −0.1 to 9.4 dB (mean = 2.3, SD = 4.1) for dome fittings. For the S2 audiogram, changes ranged from −0.3 to 5.7 dB (mean = 1.7, SD = 2.6) for dome coupling and from 0.00 to 6.2 dB (mean = 1.7, SD = 2.7) for earmold fittings. In Brand B specifically, S2 audiograms were more susceptible to NR effects, with earmold coupling sound therapy reduced by 5.7 dB compared to dome coupling being reduced by 2.1 dB. N4 audiograms in Brand B were not as affected with a reduction in sound therapy of 0.8 and 2.2 dB for earmold and dome couplings, respectively. In Brand C, dome coupling types were most susceptible to NR effects regardless of audiogram, with reductions of up to 9.4 dB for N4 dome coupling and 6.2 dB S2 dome coupling, compared to 7.2 dB and 2.9 dB for N4 and S2 earmold coupling, respectively. Similarly, in the 1/3 octave band artimetic averaging, Brands A and D were least effected by the NR algorithm, with less than 2 dB of change between conditions. In general, Brand B sound therapy was reduced by the NR across all coupling types and audiograms, and for Brand C both audiogram and coupling type had varied effects of the NR. Across all brands, the N4 audiogram overall level changes ranged from −0.1 to 6.4 dB (mean = 1.8, SD =2.8) for earmold couplings, and 0 to 6.9 dB (mean = 1.9, SD = 3.0) for dome coupling. For the S2 audiogram, earmold coupling ranged from 0.6 to 4.6 dB (mean = 1.4, SD = 2.0), and −0.1 to 2.8 dB (mean = 1.0, SD = 1.5) for dome couplings. Brand B sound therapy was reduced by 2.5 dB on average across coupling types and audiograms. In Brand C, sound therapy was reduced by NR by an average of 6.7 dB for the N4 audiogram across dome and earmold coupling, and for the S2 audiogram, the sound therapy was reduced by an average of 3.7 dB across coupling types. Taken together, both the power-sum and arithmetic averaging approaches demonstrated that the impact of noise reduction on sound therapy output was highly dependent on hearing aid manufacturer, audiogram, and coupling configuration. While Brands A and D showed minimal changes in sound therapy levels, Brands B and C exhibited greater reductions, with the magnitude of these effects varying according to the measurement method use.
3.3 Maximum stable gain and additional gain before feedback with sound therapy
When sound therapy was enabled, audible feedback was present at DSL v5 adult prescriptive gain levels measured using a 65 dB SPL speech passage. The effect of sound therapy on feedback suppression capability was minimal. When pooled by coupling type across audiograms and frequencies (2,000–4,000 Hz), the mean difference in MSGBaseline and MSGST was 1.6 dB (SD = 1.2) for dome coupling and 0.8 dB (SD = 1.1) for earmold coupling. Per frequency, changes were smaller for the N4 audiogram and larger for the S2 audiogram, with dome coupling showing relatively stable values across frequency (Figure 3). For the N4 audiogram, individual frequency differences ranged from −0.3 to 1.5 dB for dome coupling (Figure 3, panel A) and from 0.0 to 0.6 dB for earmold coupling (Figure 3, panel B). In contrast, the S2 audiogram exhibited larger changes, with dome coupling remaining relatively stable across frequency, ranging from 2.0 to 2.7 dB (Figure 3, panel C) and earmold coupling varying from 0.7 to 2.7 dB (Figure 3, panel D).
Figure 3

Maximum Stable Gain of amplification with default sound therapy (MSGST) shown along side amplification gain set to DSL v5 adult targets with default sound therapy (DSL v5 Adult Gain + ST), recorded using a 65 dB SPL speech passage. Bars represent the minimum and maximum values observed across brands at each frequency. Additional Gain Before Feedback (AGBF) is the difference between the MSGST and DSL v5 Adult Gain + ST curves.
Additional Gain Before Feedback varied across devices with frequency and coupling condition. Across conditions, acoustic feedback occurred at DSL v5 adult gain levels adjacent to default sound therapy settings. In general, dome coupling provided greater AGBF than earmold coupling at low-to-mid frequencies (250–1,500 Hz), whereas differences between coupling conditions became more variable at higher frequencies, with several device configurations exceeding the MSG. In the low- to mid-frequency range (250–1,500 Hz), there was more headroom before the MSG was reached for dome coupling than for earmolds. N4 dome coupling had an average AGBF ranging from 7.5 to 11.1 dB (SD = 1.8), and 2.2–9.5 dB (SD = 3.1) for N4 earmold coupling. The S2 dome coupling had an average AGBF ranging from 3.5 to 11.4 dB (SD = 3.3), and the S2 earmold coupling had an average AGBF ranging from 0.6 to −0.5 dB (SD =1.1), which exceeded stable gain limits. For the N4 audiogram, dome coupling yielded minimal additional gain in the mid- to high-frequency range (2,000–4,000 Hz), with average AGBF values ranging from 2.3 to 1.1 dB before the MSG was reached. In contrast, earmold coupling demonstrated greater variability, with AGBF values spanning from 8.7 to −3.1 dB, indicating a gain reduction at higher frequencies where the MSG was exceeded. For the S2 audiogram, dome coupling produced AGBF values ranging from 8.4 to −1.9 dB across frequencies, while earmold coupling ranged from 5.8 to −3.3 dB, similarly reflecting reduced gain margins at higher frequencies. In the high frequencies (6,000–8,000 Hz), there was less available gain across conditions. Average AGBF for N4 dome coupling ranged from −4.0 to 2.8 dB (SD = 4.9). N4 earmold AGBF ranged from −6.3 to −2.1 dB (SD = 2.9). S2 dome coupling AGBF ranged from −4.0 to 2.2 dB (SD = 4.4), and earmold coupling ranged from 1.5 to 8.5 dB (SD = 4.9). A corresponding figure showing the per-brand and per-frequency AGBF values, along with the overall average, is provided in the Supplementary Material (Figure S1).
4 Discussion
In summary, we examined the electroacoustic characteristics of default sound therapy settings in receiver-in-the-ear hearing aids. The primary aim was to determine how these settings vary across manufacturers, as the lack of standardized guidelines may lead clinicians to rely on software-recommended presets during programming. Because each manufacturer uses its own sound therapy algorithms, these presets can differ substantially, contributing to variability in clinical practice. Additionally, the literature on electroacoustic verification of sound therapy signals is sparse, highlighting the need to investigate how verification can identify both the provided levels of sound therapy and any unwanted interactions between digital signal-processing features and tinnitus sound therapy outputs. Such interactions may complicate programming and troubleshooting, ultimately making the initial fitting of sound therapy less reliable and more variable. The results of this study revealed both between-brand variability in the levels of sound therapy provided by default, as well as interactions between the sound therapy and use of other signal processing features, summarized below.
4.1 Physical coupling of hearing aids for sound therapy and default settings
Electroacoustic characterization demonstrated that hearing aid coupling type influenced the intensity of default sound therapy, with the effect varying by brand and audiometric configuration. For the N4 audiogram, most brands showed higher in-ear sound therapy levels with dome couplings, whereas for the S2 audiogram, earmold couplings generally produced higher default sound therapy levels. These findings suggest that physical coupling should be considered not only for amplification requirements but also for its impact on sound therapy delivery.
The variability of default intensity settings across brands was evident, with large differences between the highest and lowest overall levels. This variability is important to understand as the intensity of sound therapy is related to therapeutic benefit (
Systematic differences were observed across sound therapy levels. Although the sample size was too small to make definitive conclusions regarding the source of this variability, several factors may have contributed to the overall differences in output levels. The spectral shaping of the sound therapy may have influenced the overall energetic content of the signal. In this study, audiogram- or hearing loss-based shaping was selected when available. For manufacturers offering this feature, it is possible that manufacturer-specific software settings related to RECD implementation influenced the shaping of the sound therapy and contributed to the variability observed between brands. Depending on whether the fitting software was based on KEMAR-derived RECD values or average RECD values from older adult populations, identical sound therapy and insertion gain settings could produce different acoustic outputs. The effect of implementing individually measured CARL on-ear RECD values within the manufacturer's software was not investigated in the present study and may represent an important consideration for future research. Another factor which can influence hearing aid output is the potential for interactions with gain corrections for use of binaural hearing aids. Binaural corrections are adjustments applied to prescriptive targets to compensate for delivery of amplified sound to both ears simultaneously. The DSL v5 prescription applies a −3 dB correction regardless of hearing loss configuration; this was applied to the asymmetrical loss that was evaluated in this paper. Currently, it is unknown whether manufacturers provide or apply a binaural correction to the delivery of on-board sound therapy in hearing aids. The programming software used in this study did not offer an option to select such a feature.
In addition to broadband noise sound therapy signals evaluated in the present study, manufacturers also provide alternative sound therapy stimuli, including notch therapy (e.g., Signia), fractal or musical tones (e.g., Widex Zen Therapy), and nature-like sounds (e.g., Oticon, Resound). These sound therapies differ substantially in their spectral and temporal characteristics and may therefore interact differently with hearing aid amplification, coupling acoustics, and digital sound processing features. The present study focused specifically on steady-state broadband sound therapy approaches, which are available across different brands; however, future research should investigate whether similar acoustic and fitting considerations apply to other sound therapy paradigms.
4.2 Programming sound therapy and digital signal processing features
The present study demonstrates that sound therapy and noise reduction algorithms can interact counterproductively; this interaction may become more apparent as sound therapy levels and noise reduction strength increase. For at least one brand, the overall sound therapy intensity was reduced by 9 dB due to the NR algorithm, exceeding the 6 dB JMD benchmark. This indicates that NR has the potential to meaningfully influence patients' perceived benefit of their sound therapy. Power-sum averaging, which is more closely related to the overall perceived level, was more sensitive to the effects of noise reduction than arithmetic averaging. In contrast, arithmetic averaging provided greater sensitivity to localized spectral differences but may overemphasize changes in frequency regions that contribute less to perceived loudness. The findings indicate that the interaction between noise reduction and sound therapy is not consistent across devices and fitting conditions, warranting further consideration of the clinical implications of these differences.
Furthermore, current tinnitus management protocols suggest that sound therapy intensity should be selected by the user, with perceptual targets such as a “comfortable background level”, or more fixed points such as 6 dB sensation level (
The maximum stable gain for the devices was minimally impacted with the presence of sound therapy, varying by only by 1–2 dB across the susceptible feedback frequencies (2,000–4,000 Hz). This suggests that, once the prescribed amplification is taken into account, the hearing aid system maintains a relatively stable functional feedback limit under each fitting configuration. The differences observed in MSGST values between the N4 and S2 conditions are likely not due to changes in the intrinsic maximum stable gain of the hearing aid system itself, rather, these differences reflect the influence of the programmed fitting parameters associated with each audiogram configuration. In the present study, measurements were conducted using the prescribed amplification, frequency shaping, and sound therapy settings specific to the N4 and S2 fittings. As a result, the gain distribution across frequencies differed between conditions, altering the amount of additional gain available before acoustic feedback occurred. The addition of sound therapy may have further contributed to these effects by increasing overall acoustic energy at frequencies susceptible to instability (2–4 kHz). Therefore, the observed variability in MSGST likely reflects the interaction between prescribed amplification, spectral shaping, and sound therapy characteristics, rather than changes in the underlying acoustic stability limit of the device itself.
The AGBF measure revealed a significant gain penalty associated with amplification to maintain gain stability and avoid acoustic feedback in the high-frequency regions (4,000–8,000 Hz). Across all conditions, the addition of sound therapy caused acoustic feedback to occur when amplification gain was set to prescriptive targets. This indicates that during programming, there will be a compromise among providing adequate levels of sound therapy, allowing audible levels of sound amplification, and avoiding audible feedback. In this regard, a secondary program specifically for tinnitus sound therapy is recommended in order to reduce the gain in high-frequency regions associated with feedback. Additionally, feedback reduction programs should be set to their maximum capability in the sound therapy program to minimize audible feedback.
4.3 Verification of sound therapy settings
Verification is an electroacoustic analysis of the hearing aid output; typically performed in the ear of the hearing aid wearer to assess the adequacy of the hearing aid output at the individual level. The Audiology Practice Standards Organization (
4.4 Adapting APSO standards to sound therapy
Knowledge translation and clinical practice guidelines in audiology are imperative for delivering evidence-based practice. In the past, it has been noted that despite growing interest and efforts to produce high-quality practice guidelines, the lack of coordination among multiple practice organizations can be a barrier to the adoption of new practices (
Table 3
| APSO standard | Feature | Adapted recommendation | Justification | Verification method |
|---|---|---|---|---|
| Standard 6: The hearing aid style and the ear coupling are chosen to be appropriate for the degree and configuration of the hearing loss. Style and coupling should reflect any physical limitations of the patient. Patient input regarding acceptable styles is taken into account. | Coupling of Hearing Aid to the Ear | When appropriate, consideration should be given to custom-made receiver-in-canal (RIC) sleeves or earmolds to optimize fit, comfort, and acoustic performance of both the amplification and sound therapy provided by hearing aids. | Custom coupling improves sound isolation and retention, helping reduce slit leaks that can cause feedback, and noise reduction (NR) interference with sound therapy (ST). | Fitting check: Assess for comfort and use. Although not necessary, the REOR can be measured to quantify the openness of fit. |
| Standard 7: The recommended hearing aids include signal processing and features that support the patient's listening needs. They have the appropriate gain and output, including reserve gain, to meet frequency-specific fitting targets as defined by a validated prescriptive method. | Noise Reduction | Disable noise reduction on the sound therapy program. | Open-fit hearing aids and loud-level sound therapy may be detected by the noise reduction algorithm because they are steady broadband noise. | Speech-live: Record with ST on and NR off using Speech Live. Then record again with both ST and NR on. A drop in the overall level indicates NR is affecting ST. |
| Adaptive Feedback Manager | Enable maximum feedback suppression settings for the sound therapy program. | Just as in audibility-based hearing aid fittings, feedback is a deterrent to using the sound therapy program. | Listening check: Move your hands near the hearing aids to check for audible feedback. If present, reduce gain and repeat until feedback is eliminated. | |
| Overall Gain Adjustment | Reduced overall gain in the sound therapy program, prioritized an appropriate level of ST instead of audibility. | The maximum stable gain may be exceeded if both amplification and sound therapy contribute to ear canal SPL. | ||
| Standard 11: Following individualized verification of hearing aid gain and output, if the fitting is not acceptable to the patient, minor deviations in gain and output may be necessary. | Overall verification of Sound Therapy | Verify that sound therapy output matches clinical intent. | Coupling type and severity of hearing loss influence the default sound settings of sound therapy. | Speech Live: Record with Sound Therapy on and assess the RMS level, and MAOF. Make adjustments as needed based on patient feedback. |
Considerations for fitting, programming and verifying sound therapy.
Adapted from
5 Limitations
A manikin was used for on-ear measures in this study, but the manufacturer-specific programming software is designed for use with human ears. Evaluation on human ears may therefore serve to extend the generalizability of these findings. The nature of this study was exploratory; no repeat measures were conducted, which potentially impacted the replicability of results and generalizability of findings. However, the measurement procedures and hearing aid features used in this study have previously been independently evaluated and shown to demonstrate acceptable reliability in prior research. Overall, real ear measures are found to be reliable measures of sound delivered in the ear canal, up to and beyond 8,000 Hz with appropriate insertion depth of probe tube (
Another limitation of this study is that only an asymmetrical hearing loss configuration was evaluated, with the N4 audiogram simulated in the left ear and the S2 audiogram simulated in the right ear. Although none of the manufacturer's fitting software provided explicit options for binaural correction of sound therapy, it remains possible that proprietary binaural processing or device-to-device communication influenced the delivery of sound therapy in ways that were not apparent to the researchers and were therefore not assessed in this study. Consequently, the observed outputs may reflect interactions specific to the asymmetrical fitting configuration; the findings may not be directly generalizable to listeners with symmetrical hearing losses. Furthermore, the default sound therapy levels demonstrated limited audibility across frequencies, with most brands producing levels below hearing loss thresholds. It may be that these sounds are audible, given that the sound therapy was broadband and the hearing thresholds were based on narrowband signals. However, because a manikin was used for data collection, subjective ratings, such as the audibility, loudness, or sensation level of the sound therapy, were not captured. Further assessments on human subjects are recommended. Finally, the small sample size prevented inferential data analysis, limiting the presented results to descriptive summaries for two standard audiograms. Further study in larger samples would clarify the generalizability of these results.
6 Conclusions
Sound therapy delivered via a hearing aid for tinnitus management is affected by the hearing aid's physical coupling and signal-processing settings during fitting. As a result, hearing aid verification procedures are useful for characterizing sound therapy within the broader context of hearing aid fitting; this study has revealed that sound therapy levels vary across brands, hearing loss configurations, and coupling types. In some cases, these default settings may also interact counterproductively with core digital processing features. This paper outlines key considerations for fitting, programming, and verification of sound therapy with the aim of supporting more consistent clinical practice and improving patient access to effective care.
Statements
Data availability statement
The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.
Author contributions
AH: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing, Funding acquisition, Project administration. SH: Writing – review & editing, Funding acquisition. JS: Writing – review & editing, Funding acquisition. VP: Conceptualization, Methodology, Writing – review & editing. SS: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Writing – review & editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This research was funded by the American Tinnitus Association Innovative Tinnitus Research Grant (ATA grant ID: 1507095). Grant funding supported costs associated with the project and provided resources to facilitate future studies that will build on the findings of this work.
Acknowledgments
The authors would like to thank the American Tinnitus Association and our colleagues at the National Center for Audiology for your continuous encouragement and support.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that Generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fauot.2026.1861736/full#supplementary-material
References
1
AgnewJ. (1996). Acoustic feedback and other audible artifacts in hearing aids. Trends Amplif.1, 45–82. doi: 10.1177/108471389600100202
2
AHead Simulations. (2024). CARL Basic. Available online at: https://aheadsimulations.com/collections/carl-units/products/carl-basic (Accessed May 31, 2026).
3
AlmufarrijI.DillonH.MunroK. J. (2021). Does probe-tube verification of real-ear hearing aid amplification characteristics improve outcomes in adults? A systematic review and meta-analysis. Trends Hear.25:2331216521999563. doi: 10.1177/2331216521999563
4
American Speech-Language-Hearing Association. (1990). Audiometric Symbols. American Speech-Language-Hearing Association.Available online at: https://www.asha.org/policy/gl1990-00006/?srsltid=AfmBOopTs92bH8xOo0uuR2nar_l41nqDZXVipqIy5KfGtbFiU70waC2Y (Accessed May 31, 2026).
5
APSO. (2021). Hearing Aid Fitting Standard for Adult and Geriatric Patients. Audiology Practice Standards Organization. Available online at: https://www.audiologystandards.org/standards/publications/APSO_S2.1_Adult-HAF_20210502.pdf (Accessed May 31, 2026).
6
Audioscan. (2023). Verifit 2 Advanced Verification (Version 2023.2.5; Issue Version 2023.2.5). Etymonic Design Incorporated. Available online at: https://docs.audioscan.com/userguides/vf2manual.pdf (Accessed May 31, 2026).
7
Audioscan. (2026). Verifit®VF-2 User's Guide 2025.1. Available online at: https://docs.audioscan.com/userguides/vf2manual.pdf (Accessed May 31, 2026).
8
BagattoM.MoodieS.ScollieS.SeewaldR.MoodieS.PlumfordJ.et al. (2005). Clinical protocols for hearing instrument fitting in the desired sensation level method. Trends Amplif.9, 199–226. doi: 10.1177/108471380500900404
9
BallingL. W.JensenN.CaporaliS.CubickJ.SwitalskiW. (2019, December 18). Challenges of instant-fit ear tips: what happens at the eardrum? The Hearing Review. Available online at: https://hearingreview.com/hearing-loss/patient-care/hearing-fittings/challenges-of-instant-fit-ear-tips-what-happens-at-the-eardrum (Accessed May 31, 2026).
10
BisgaardN.VlamingM. S.DahlquistM. (2010). Standard audiograms for the IEC 60118-15 measurement procedure. Trends Amplif.14, 113–120. doi: 10.1177/1084713810379609
11
BlauM.SankowskyT.StirnemannA.OberdannerH.SchmittN. (2008). Acoustics of open fittings. J. Acoust. Soc. Am. 123, 3011–3011. doi: 10.1121/1.2932603
12
BoisvertI.ClemeshaJ.LundmarkE.CromeE.BarrC.McMahonC. M. (2017). Decision-making in audiology: balancing evidence-based practice and patient-centered care. Trends Hear.21:2331216517706397. doi: 10.1177/2331216517706397
13
British Society of Audiology. (2020). Fitting of Combination Hearing Aids for Subjects with Tinnitus. Available online at: https://www.thebsa.org.uk/wp-content/uploads/2020/10/Fitting-of-combination-hearing-aids-for-subjects-with-tinnitus.pdf (Accessed May 31, 2026).
14
ChungD.GannonP.MasonK. (1984). Factors affecting the prevalence of tinnitus. Audiology23, 441–452. doi: 10.3109/00206098409070084
15
CimaR. F. F.MazurekB.HaiderH.KikidisD.LapiraA.NoreñaA.HoareD. J. (2019). A multidisciplinary European guideline for tinnitus: diagnostics, assessment, and treatment. HNO67, 10–42. doi: 10.1007/s00106-019-0633-7
16
DaoA.FolkeardP.BakerS.PumfordJ.ScollieS. (2021). Fit-to-targets and aided speech intelligibility index values for hearing aids fitted to the DSL v5-adult prescription. J. Am. Acad. Audiol.32, 090–098. doi: 10.1055/s-0040-1718707
17
dos SantosG. M.BentoR. F.de MedeirosI. R. T.OiticcicaJ.da SilvaE. C.PenteadoS. (2014). The influence of sound generator associated with conventional amplification for tinnitus control: randomized blind clinical trial. Trends Hear. 18:2331216514542657. doi: 10.1055/s-0034-1388704
18
FolkeardP.AlamM.KochR.AbbasalipourP.ScollieS. (2023). An evaluation of the CARL manikin for use in “patient-free” real ear measurement: consistency and comparison to normative data. Int. J. Audiol.62, 376–382. doi: 10.1080/14992027.2022.2048104
19
HenryJ. A. (2022). Sound therapy to reduce auditory gain for hyperacusis and tinnitus. Am. J. Audiol.31, 1067–1077. doi: 10.1044/2022_AJA-22-00127
20
HenryJ. A.CarlsonK. F.TheodoroffS.FolmerR. L. (2022). Reevaluating the use of sound therapy for tinnitus management: perspectives on relevant systematic reviews. J. Speech Lang. Hear. Res.65, 2327–2342. doi: 10.1044/2022_JSLHR-21-00668
21
HenryJ. A.SonstroemA.SmithB.GrushL. (2021). Survey of audiology graduate programs: training students in tinnitus management. Am. J. Audiol.30, 22–27. doi: 10.1044/2020_AJA-20-00140
22
HenryJ. A.ZauggT. L.MyersP. J.SchechterM. A. (2008). Using therapeutic sound with progressive audiologic tinnitus management. Trends Amplif.12, 188–209. doi: 10.1177/1084713808321184
23
HongT.RalliM.StockingC.SheppardA. (2020). Acoustic analysis of hearing aid sound therapy programs. Hear. Bal. Commun.18, 8–15. doi: 10.1080/21695717.2019.1692592
24
JastreboffP. J.HazellJ. W. P. (2004). Tinnitus Retraining Therapy: Implementing the Neurophysiological Model, 1st Edn. Cambridge, MA: Cambridge University Press. doi: 10.1017/CBO9780511544989
25
JastreboffP. J.JastreboffM. M. (2000). Tinnitus retraining therapy (TRT) as a method for treatment of tinnitus and hyperacusis patients. J. Am. Acad. Audiol.11, 162–177. doi: 10.1055/s-0042-1748042
26
JastreboffP. J.JastreboffM. M. (2024). “Tinnitus retraining therapy,” in Textbook of Tinnitus, W. Schlee, B. Langguth, D. De Ridder, S. Vanneste, T. Kleinjung, and A. R. Møller, Eds (New York, NY: Springer International Publishing), pp. 589–616. doi: 10.1007/978-3-031-35647-6_47
27
KochkinS. (2007). MarkeTrak VII: obstacles to adult non-user adoption of hearing aids. Hear. J.60:24. doi: 10.1097/01.HJ.0000285745.08599.7f
28
KönigO.SchaetteR.KempterR.GrossM. (2006). Course of hearing loss and occurrence of tinnitus. Hear. Res.221, 59–64. doi: 10.1016/j.heares.2006.07.007
29
KutybaJ.GosE.JedrzejczakW. W.Raj-KoziakD.KarpieszL.NiedziałekI.et al. (2022). Effectiveness of tinnitus therapy using a mobile application. Eur. Arch. Oto-Rhino-Laryngol.279, 1257–1267. doi: 10.1007/s00405-021-06767-9
30
LangguthB.KreuzerP. M.KleinjungT.De RidderD. (2013). Tinnitus: causes and clinical management. Lancet Neurol.12, 920–930. doi: 10.1016/S1474-4422(13)70160-1
31
LansbergenS.DreschlerW. A. (2020). Classification of hearing aids into feature profiles using hierarchical latent class analysis applied to a large dataset of hearing aids. Ear Hear.41, 1619–1634. doi: 10.1097/AUD.0000000000000410
32
LelicD.CaporaliS.ParkerD.NielsenJ.BallingL. W. (2024). Impact of a combination sound therapy on tinnitus distress: an exploratory one-year longitudinal study. Front. Audiol. Otol. 2:2596. doi: 10.3389/fauot.2024.1322596
33
LiuH.ZhangJ.YangS.WangX.ZhangW.LiJ.YangT. (2021). Efficacy of sound therapy interventions for tinnitus management. Medicine.100:e27509. doi: 10.1097/MD.0000000000027509
34
MarcrumS. C.PicouE. M.BohrC.SteffensT. (2018). Feedback reduction system influence on additional gain before feedback and maximum stable gain in open-fitted hearing aids. Int. J. Audiol.57, 737–745. doi: 10.1080/14992027.2018.1487088
35
McNeillC.Távora-VieiraD.AlnafjanF.SearchfieldG. D.WelchD. (2012). Tinnitus pitch, masking, and the effectiveness of hearing aids for tinnitus therapy. Int. J. Audiol.51, 914–919. doi: 10.3109/14992027.2012.721934
36
McSheffertyD.WhitmerW. M.AkeroydM. A. (2016). The just-meaningful difference in speech-to-noise ratio. Trends Hear. 20:2331216515626570. doi: 10.1177/2331216515626570
37
MoodieS. T.KothariA.BagattoM. P.SeewaldR.MillerL. T.ScollieS. D. (2011). Knowledge translation in audiology: promoting the clinical application of best evidence. Trends Amplif.15, 5–22. doi: 10.1177/1084713811420740
38
MyersP. J.GriestS.KaelinC.LegroM. W.SchmidtC. J.ZauggT. L.HenryJ. A. (2014). Development of a progressive audiologic tinnitus management program for veterans with tinnitus. J. Rehabil. Res. Dev.51, 609–622. doi: 10.1682/JRRD.2013.08.0189
39
NewmanC. W.SandridgeS. A.MeitS. S.CherianN. (2008). Strategies for managing patients with tinnitus: a clinical pathway model. Semin. Hear.29, 300–309. doi: 10.1055/s-0028-1082035
40
RickettsT.JohnsonE.FedermanJ. (2008). Individual differences within and across feedback suppression hearing aids. J. Am. Acad. Audiol.19, 748–757. doi: 10.3766/jaaa.19.10.3
41
RickettsT. A.BentlerR.MuellerH. G. (2017). Essentials of Modern Hearing Aids: Selection, Fitting, and Verification. San Diego, CA: Plural Publishing.
42
RochaA. V.MondelliM. F. C. G. (2020). Applicability of the real ear measurement for audiological intervention of tinnitus,. Brazil. J. Otorhinolaryngol.86, 14–22. doi: 10.1016/j.bjorl.2018.07.010
43
SchleeW.LangguthB.De RidderD.VannesteS.KleinjungT.MøllerA. R. (Eds.). (2024). Textbook of Tinnitus. New York, NY: Springer International Publishing. doi: 10.1007/978-3-031-35647-6
44
ScollieS.LevyC.PourmandN.AbbasalipourP.BagattoM.RichertF.et al. (2016). Fitting noise management signal processing applying the American Academy of Audiology pediatric amplification guideline: verification protocols. J. Am. Acad. Audiol.27, 237–251. doi: 10.3766/jaaa.15060
45
ScollieS.SeewaldR.CornelisseL.MoodieS.BagattoM.LaurnagarayD.et al. (2005). The desired sensation level multistage input/output algorithm. Trends Amplif.9, 159–197. doi: 10.1177/108471380500900403
46
ShekhawatG. S.SearchfieldG. D.KobayashiK.StinearC. M. (2013). Prescription of hearing-aid output for tinnitus relief. Int. J. Audiol.52, 617–625. doi: 10.3109/14992027.2013.799787
47
ShettyH. N.BasavarajV. (2022). Amplification strategies to reduce tinnitus: a paired comparison method. J. Int. Adv. Otol. 18, 8–13. doi: 10.5152/iao.2022.20057
48
TylerR. S.PerreauA.PowersT.WattsA.OwenR.JiH.ManciniP. C. (2020). Tinnitus sound therapy trial shows effectiveness for those with tinnitus. J. Am. Acad. Audiol.31, 6–16. doi: 10.3766/jaaa.18027
49
VaisbergJ. M.MacphersonE. A.ScollieS. D. (2016). Extended bandwidth real-ear measurement accuracy and repeatability to 10 kHz. Int. J. Audiol.55, 580–586. doi: 10.1080/14992027.2016.1197427
50
WaechterS.JönssonA. (2022). Hearing aids mitigate tinnitus, but does it matter if the patient receives amplification in accordance with their hearing impairment or not? a meta-analysis. Am. J. Audiol.31, 789–818. doi: 10.1044/2022_AJA-22-00004
Summary
Keywords
audiological care, combination devices, hearing aids, management, sound therapy, tinnitus, verification
Citation
Hajas A, Hayes SH, Scott J, Parsa V and Scollie S (2026) Considerations for fitting, programming, and verification of sound therapy delivered by hearing aids. Front. Audiol. Otol. 4:1861736. doi: 10.3389/fauot.2026.1861736
Received
21 April 2026
Revised
28 June 2026
Accepted
03 July 2026
Published
24 July 2026
Volume
4 - 2026
Edited by
Grant Searchfield, The University of Auckland, New Zealand
Updates

Check for updates
Copyright
© 2026 Hajas, Hayes, Scott, Parsa and Scollie.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Susan Scollie, scollie@nca.uwo.ca
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.