Abstract
Background:
Meniscal injury is a common musculoskeletal disease. An artificial meniscal prosthesis is an effective choice for replacing allograft transplantation of the meniscus; however, the best method of fixation is still controversial. This finite element study aims to investigate the biomechanical effect of various anterior and posterior horn suture stiffnesses on a medial meniscal prosthesis under different knee flexion angles.
Methods:
A validated three-dimensional finite element model of a human knee joint was used. A medial meniscal prosthesis of polycarbonate urethane (PCU) material was implanted, and spring elements were used to simulate the anterior and posterior horn fixation sutures and four levels of stiffness (13, 25, 62, and 200 N/mm) to represent available suture materials in the clinic. Knee flexion angles of 0°, 30°, and 60° were applied with combined anterior tibial load (134 N), compressive load (1150 N), and internal rotation moment (4 Nm). Tensile force in the sutures, meniscus translation (medial–lateral and anterior–posterior), and meniscus tibial contact stresses were extracted and analyzed.
Results:
Increasing the suture stiffness resulted in increased force in the anterior horn sutures and the force in posterior horn sutures being mainly influenced by joint loading and not by suture stiffness. At 60° flexion, posterior horn tensile forces (50.11–52.35 N) were about 3.4–4.1 times greater than anterior horn tensile forces, which indicated the posterior horn as the major load-bearing structure. Increasing suture stiffness progressively decreased the medial–lateral and anterior–posterior displacements of the medial meniscus, and the 200 N/mm group of suture stiffnesses most closely approximated native meniscal kinematics. Medial tibial cartilage contact stress decreased with the increase of suture stiffness. It was found that 200 N/mm of suture stiffness most closely simulated the contact stress of the native knee for most loading conditions. The lateral compartment was not influenced by medial suture stiffness changes.
Conclusion:
Suture stiffness is key to the overall biomechanical performance of artificial menisci. Stiffer fixation (200 N/mm) reduces the translation of the medial meniscal and tibial cartilage contact stress and restores the load distribution.
Introduction
The meniscus plays a number of important roles in knee joints, including transmitting loads, absorbing shock, and stabilizing and lubricating the joint (). It is greatly exposed to gravity, so it is the most commonly injured structure in sports. In 2006 alone, there were almost one million knee arthroscopy cases performed, nearly half of which were for treatment of meniscal tears (). Traumatic or postoperative meniscal defect or loss could be detrimental to or extinguish the ability of the meniscus to transmit loads and stresses, decrease the tibiofemoral contacted area, and lead to over-contact stresses, thus inducing the long-term development of osteoarthritis (; ; ; ; ; ).
For patients with traumatic or postoperative meniscal deficiency, existing and successful clinical treatment options include allograft meniscus transplantation; the implantation of artificial meniscal prostheses and tissue-engineered menisci are still under study (). Allograft meniscus transplantation suffers from many limitations, such as a shortage of donor meniscus supply, size matching between allograft and recipient, difficulty in tissue preservation, the risk of transmitting disease, risk of immune rejection, and high costs (). Artificial meniscal prosthesis can overcome some of these limitations. The artificial meniscal prosthesis NUsurface® developed by Active Implants has been on the market since 2008. This artificial meniscal prosthesis has a non-anatomical floating implantation that does not need to fix the implantation site. However, at least 14 clinical cases have been reported as being displaced or subluxed within 1 year after implantation, leading to patient pain, reduced joint motion, or even revision surgery (; ; ).
For other prostheses still in research, the most efficacious method is to directly suture anterior and posterior horns with suture threads and then to secure them by the bone tunnel (; ; ). used a prosthesis fixing with the interface screws that were put into bone tunnels and then fixed through built-in anterior and posterior horn ligament-like strap structures. used meniscal prostheses designed with anterior and posterior horns as allograft bone plugs, with a built-in bolt-like structure fixed in the tibial plateau. There is still no conclusive method for the fixation structure and for the anterior and posterior horns of the meniscal prostheses.
A suitable tension of fixation at the anterior and posterior horns of the meniscus is very important. A tense fixation guarantees the “hoop stress effect” of the meniscus that converts axial compressive forces into circumferential stresses and thereby can significantly alleviate stress on the tibial plateau (). Overly excessive tensed fixation may impair the motion of the meniscus during knee flexion and extension and decrease the meniscal contact area, further accelerating joint stress (). Therefore, the more securely fixed the anterior and posterior horns of meniscus are, the stronger the hoop stress effect of the meniscus is exerted, while, meniscal mobility is diminished. Conversely, if the horn fixation is too loose, too much meniscal movement will weaken the hoop stress effect.
A consideration of the optimal combination of meniscal mobility and the hoop stress effect is helpful for achieving the maximal effect of artificial menisci in reducing joint stress. This study will use finite element analysis to study the effect of meniscal prosthesis anterior and posterior horn suture stiffness on meniscus prosthesis mobility, horn strain, and joint stress under different knee motions.
Materials and methods
This study uses a previously developed and validated finite element model of a human knee joint (). A 3D model of the knee was reproduced from magnetic resonance images of a 45-year-old male subject captured on a SIEMENS MAGNETOM Skyra scanner (Siemens, Germany). The knee model includes the femur, tibia, articular cartilage, menisci, ACL, PCL, and medial (MCL) and lateral (LCL) collateral ligaments (MCL and LCL) (Figure 1).
FIGURE 1
The material properties of the parts are shown in Table 1. Bone and articular cartilage were modeled as an isotropic linear elastic material (). The menisci were modeled as transversely isotropic linear elastic materials with their material properties (). The ligaments were modeled as isotropic hyper-elastic materials described by the strain-energy density function (). The material properties of the ACL and PCL were described by the Veronda–Westmann constitutive model, while the MCL and LCL were described by the Mooney–Rivlin model (; ; ; ). The contact was assumed to be frictionless between femoral and tibial cartilage surfaces and between the femoral cartilage and menisci, providing relative freedom of movement but no penetration of surfaces. Tie constraints were used between the ligament insertion locations and related bony inserted locations and also between cartilage and bone surfaces beneath the cartilage. The model was meshed with four-node tetrahedron elements in HyperMesh 12.0. Mesh convergence was assessed by decreasing the element size until the AL force under a 2.5 mm translational load at full extension converged within 0.5 N. The validated element size was 1 mm. The total number of elements in the intact knee model was 659,251. Model validation was performed on the same cadaveric knee using a robotic/UFS testing system. Under a 134 N anterior tibial load at 30°, 60°, and 90° flexion, anterior tibial translation (ATT) and in situ ACL force were recorded. The same loading conditions were applied to the FE model. The ACL force between testing and simulation were compared to verify the model when the difference was less than 5%.
TABLE 1
| Tissue/structure | Material properties |
|---|---|
| Bone | E = 0.4 GPa, ν = 0.33 |
| Articular cartilage | E = 5 MPa, ν = 0.46 |
| Intact meniscus | Eθ = 125 MPa, ER = EZ = 27.5 MPa GθR = GθZ = 2 MPa, GRZ = 2 MPa VθR = VθZ = 0.1, VRZ = 0.33 |
| ACL | αACL = 0.3 MPa, βACL = 12.2 |
| PCL | αPCL = 0.18 MPa, βPCL = 12.2 |
| MCL/LCL | C1 = 30.1 MPa, C2 = −27.1 MPa |
| Artificial meniscus | E = 11 MPa, ν = 0.49 |
Material properties of the finite model.
From our surgical approach for medial artificial meniscus implantation, we attributed material properties of a 3D-printed meniscal implant fabricated from polycarbonate urethane (PCU; Bionate® Grade II, DSM Biomedical Inc., Berkeley, CA, United States) to the artificial meniscus (). Springs were used to model the anterior and posterior horn sutures that fixed the medial meniscal implant to the tibial plateau (Figure 2). Although these sutures were relatively stable when fixed inside the bone tunnels, only the portion of the meniscus attached to the tibial tunnel aperture was modeled. We modeled four values of spring stiffness (13, 25, 62, and 200 N/mm) which respectively correspond to suture materials widely used clinically: No. 2 Ethibond sutures (Ethicon, Somerville, NJ, United States), No. 5 Ethibond sutures, No. 5 FiberWire® high-strength sutures (Arthrex Inc., Mironomics International Ltd., Naples, FL, United States), and Dyneema Purity-based textile tape (DSM Biomedical, Geleen, Netherlands), (; ).
FIGURE 2
A Cartesian coordinate system was set up in which the origin is the center point between the medial and lateral femoral condylar rotational centers. The line between the two centers defines the x-axis, the direction perpendicular to the joint line defines the z-axis, and the anterior–posterior direction of the knee defines the y-axis. For the femoral side, only a flexion–extension rotation about the x-axis was allowed, and all other degrees of freedom were fixed. For the tibial side, only flexion–extension rotation was fixed; all other degrees of freedom were allowed. The femur was loaded with flexion angles of 0°, 30°, and 60°. In addition, an anterior tibial load of 134 N in the direction of y, a compressive load of 1150 N in the direction of z, and an internal rotation torque of 4 Nm were applied (; ; ).
Peak contact stresses of the tibial cartilage and menisci, tensile forces in the anterior and posterior horn fixation sutures, and meniscal displacements in the anterior–posterior and medial–lateral directions were extracted.
Results
Tensile forces in the anterior and posterior horn sutures
The tensile forces of the anterior and posterior horn fixation sutures are at different stiffnesses. Sutural pull in the anterior and posterior horn fixation sutures at different stiffnesses are shown in Figure 3. It can be observed that the tensile forces of both anterior horn and posterior horn sutures increased with increasing suture stiffness, but the magnitude of increase was different for the two sutures. The tensile force in the anterior horn sutures increased linearly with increasing stiffness while the increasing rate of posterior horn suture force was relatively limited under higher loading.
FIGURE 3
For the anterior horn sutures, tensile forces at 0°, 30°, and 60° flexion of the 13 N/mm group were 10.49 N, 10.85 N, and 12.24 N, respectively. As the suture stiffness increased to 25, 62, and 200 N/mm, the tensile force gradually increased to ranges of 11.03–11.67 N, 11.83–12.91 N, and 13.49–15.13 N, respectively. Compared with the 13 N/mm group, the 200 N/mm group increased 11.2%, 19.0%, and 23.6% at 0°, 30°, and 60° flexion, respectively, indicating a great influence of anterior horn loading on suture stiffness.
For the posterior horn, the tensile forces for 13 N/mm were 5.69 N, 32.51 N, and 50.11 N for 0°, 30°, and 60° flexion, respectively, which increased to 7.89 N, 33.67 N, and 52.35 N, respectively, for the 200 N/mm group. Compared with the 13 N/mm group, these increases were 38.7%, 3.6%, and 4.5%, respectively. The differences were small between stiffness groups at 30° and 60° flexion, and thus posterior horn loading was influenced most strongly by joint load transfer rather than suture stiffness.
Comparison of the anterior and posterior horn sutures revealed that the former was under greater tensile force than the latter at 0° flexion. However, the postflexion calculated tensile force in the posterior horn (tendon) increased dramatically and was bigger than that in the anterior horn (tendon) when flexed 30° and 60°. At 60° flexion, the tensile force in the posterior horn was 50.11–52.35 N, which was about 3.4–4.1 times of that in the anterior horn. We conclude that posterior horn fixation is the load-bearing part of the artificial meniscus fixation system and is crucial to the transmission of load.
Meniscal displacement
Figure 4 shows the medial–lateral and anterior–posterior meniscal displacements under different stiffness sutures.
FIGURE 4
For the medial–lateral motion of the medial meniscus, implantation of the artificial meniscus displayed displacement, which was strongly associated with the suture stiffness. At 0° flexion, the suture stiffness had very little effect on medial displacement but had a very strong effect on lateral displacement. At 30° and 60° flexion, the lowest stiffness group (13 N/mm) had the greatest medial displacement and the least lateral displacement. At 60° flexion, the medial displacement of the artificial meniscus reached 3.053 mm in the 13 N/mm group. With increases in suture stiffness, the medial displacement of the artificial meniscus decreased gradually and the lateral displacement increased. At suture stiffness of 200 N/mm, the medial displacement of the artificial meniscus was similar to that of the native meniscus. These results show that increasing suture stiffness can limit radial displacement of the artificial meniscus and maintain its position well.
Regarding the anterior–posterior motion of the medial meniscus, suture stiffness significantly affected meniscal mobility. For 0° flexion, anterior–posterior displacement decreased with increasing suture stiffness. The group with 13 N/mm suture presented with the largest posterior displacement under all loading conditions, being 7.395 mm and 10.330 mm at 30° and 60° flexion, respectively. Higher suture stiffness led to further decreases in the posterior displacement of the artificial meniscus. At 60° flexion, posterior displacement in the 25, 62, and 200 N/mm suture groups was 10.050 mm, 9.885 mm, and 9.841 mm, a decrease of 2.7%, 4.3%, and 4.7% of that in the 13 N/mm suture group, respectively. Greater suture stiffness thus restrains the posterior displacement of the artificial meniscus and maintains its good kinematic stability.
For the lateral meniscus, there was little difference for medial–lateral displacement among different stiffness groups. At 0° and 30° flexion, medial displacement was near 0.5 mm in all groups and increased to almost 1.4 mm at 60° flexion; lateral displacement increased with an increased angle of flexion; however, the maximum difference among different stiffness groups was less than 0.3 mm. This suggests that changes of the suture stiffness of a medial artificial meniscus have little influence on the radial motion of the lateral meniscus.
Similarly, only slight difference in the displacement of the lateral meniscus from anterior to posterior direction was observed. For the native knee, the posterior displacement was 5.249 mm and 8.270 mm at flexions of 30° and 60°, respectively. Although the displacement of artificial meniscus groups was slightly smaller than that of the native knee, the differences between the stiffness groups were all less than 0.5 mm, meaning that suture stiffness only had a slight effect on the anterior–posterior motion of the lateral meniscus.
Meniscal contact stress
Table 2 summarizes the contact stresses of the medial and lateral menisci under different flexion angles (0°, 30°, and 60°) and different suture stiffness conditions. For the same flexion angle, the contact stress of the medial and lateral menisci of the native knee was larger than that of implanted artificial meniscus models. For 0° flexion, the contact stress of medial meniscus increased with increasing suture stiffness. In contrast, for 30° and 60° flexion, the contact stress of the medial meniscus decreased with increasing suture stiffness.
TABLE 2
| Suture stiffness conditions | Contact stress of medial cartilage (MPa) | Contact stress of lateral cartilage (MPa) | ||||
|---|---|---|---|---|---|---|
| | 0° | 30° | 60° | 0° | 30° | 60° |
| Intact | 1.553 | 2.831 | 3.967 | 1.994 | 2.327 | 3.357 |
| 13 N/mm | 1.376 | 2.726 | 3.815 | 1.920 | 2.214 | 2.866 |
| 25 N/mm | 1.448 | 2.631 | 3.470 | 1.928 | 2.222 | 2.983 |
| 62 N/mm | 1.466 | 2.436 | 3.433 | 1.934 | 2.224 | 3.083 |
| 200 N/mm | 1.466 | 2.601 | 3.625 | 1.937 | 2.221 | 3.128 |
Peak contact stress of medial menisci under various suture stiffness conditions during flexion.
Compared with the medial compartment, there were only small variations of lateral meniscal contact stresses in different stiffness groups. The values were always close to the native knee value throughout all loading levels and showed that the stiffness of fixation sutures to the lateral artificial meniscus did not have an obvious influence on the mechanical environment of the lateral compartment.
Figure 5 shows the contact stress distributions of the menisci. At 0° flexion, the peak of meniscal contact stress of the medial meniscus was in the medial aspect of the anterior horn. At 30° and 60° flexion, it shifted to medial aspects of the meniscal body. For each flexion angle, the distribution pattern of meniscal contact stress was minimally affected by suture stiffness. For all loading conditions, the peak of the contact stress for the lateral meniscus was largely located in medial aspects of the meniscal body.
FIGURE 5
Tibial cartilage contact stress
Table 3 shows the contact stress of the medial tibial cartilage under various suture stiffness conditions. On low-stiffness fixation (13 N/mm), the cartilage of the medial tibia was generally of higher contact stress; with increasing suture stiffness, the contact stress of the medial cartilage decreased gradually; The stresses of the 200 N/mm group were closest to those of native knees under most loads. Conversely, for the cartilage of the lateral tibia, slight differences could be found in different stiffness groups but their magnitude was small, indicating that changes in suture stiffness affected the load borne mainly by the medial compartment and had little effect on the lateral compartment.
TABLE 3
| Suture stiffness conditions | Contact stress of medial cartilage (MPa) | Contact stress of lateral cartilage (MPa) | ||||
|---|---|---|---|---|---|---|
| | 0° | 30° | 60° | 0° | 30° | 60° |
| Intact | 2.741 | 3.667 | 4.078 | 2.699 | 3.463 | 2.955 |
| 13 N/mm | 2.757 | 3.811 | 3.954 | 2.696 | 3.530 | 3.026 |
| 25 N/mm | 2.754 | 3.807 | 3.911 | 2.696 | 3.523 | 3.021 |
| 62 N/mm | 2.752 | 3.780 | 3.929 | 2.695 | 3.516 | 3.017 |
| 200 N/mm | 2.751 | 3.760 | 3.983 | 2.694 | 3.515 | 3.015 |
Peak contact stress of medial tibial cartilage under various suture stiffness conditions during flexion.
The contact stress distributions of tibial cartilage are shown in Figure 6. Generally, fixing the artificial meniscus with sutures of different stiffness values kept the tibiofemoral contact patterns similar to those of the intact knee joint. In different loading modes, the domains of higher contact stress around the medial and lateral sides of the tibial cartilage were more concentrated in the whole contact region, and the overall stress patterns of the different suture stiffness groups were largely the same.
FIGURE 6
With increasing suture stiffness, only small variations could be observed in the range of the distribution and position of contact stress. No evident changes of the load-bearing area were seen in the medial and lateral parts of the tibial plateau. These observations suggest that suture stiffness has little influence on the tibiofemoral load-bearing area and that it changes little over the whole contact mechanics of the cartilage after artificial meniscus implantation.
Discussion
The most important finding in our study is that the suture stiffness of the medial meniscus prosthesis, especially of the posterior horn, has a large effect on the load-sharing, location, and contact mechanics of the artificial meniscus, and that generally a greater stiffness (200 N/mm) leads to a better biomechanical effect that is closer to the normal knee.
Sutures or tape have been used to fix the meniscus prosthesis to the anterior and posteriors horn in recent research (; ; ). Like the transtibial pullout technique of repairing meniscus root tear, the suture threads or tape of the posterior horns are passed through the bone tunnel of the tibia and fixation (suturing) (), so that the fixation strength influences the mobility and hoop stress effect of the meniscus prosthesis. The suture threads or tape used in studies or clinic include Ethibond sutures, Fiber wire® sutures, and Dyneema Purity-based textile tape, with stiffness values from 13 to 242 ± 33 N/mm (; ). To date, according to the literature summarized above, Dyneema Purity-based textile tape is the highest tensile thread that can fix the meniscus prosthesis, with stiffness values reaching 242 ± 33 N/mm ().
Our stiffness value is consistent with the measured biomechanical properties of the posterior horns of the human medial menisci: 207.2 ± 52.79 N/mm (). Consequently, four springs with different stiffness values were used in our finite element model; these can mimic the sutures or tape mentioned above—the largest is 200 N/mm. Furthermore, we investigated the biomechanical effects of the stiffness of anterior and posterior horn fixation on the performance of the medial meniscal prosthesis by systematically varying the stiffness of suture fixation to clarify the optimal design between securing the hoop stress effect and allowing sufficient mobility of the implant.
One of the main findings of our study is the different biomechanical role of anterior and posterior horn fixation sutures. The tensile force in the posterior horn suture was much larger than that in the anterior horn, especially at 60° flexion, where they were almost 3.4–4.1 times greater. Our finding is consistent with the measured biomechanical properties of the anterior and posterior horns of the human medial menisci, in which the linear stiffness of the anterior horn of the medial meniscus was 169.4 ± 24.19 N/mm, whereas that of the posterior horn was 207.2 ± 52.79 N/mm (). Moreover, the elastic modulus of the mid-anterior horn is 1.35 ± 0.76 MPa with an ultimate stress value of 0.34 ± 0.12 MPa, whereas that of the mid-posterior horn was 6.42 ± 0.78 MPa with an ultimate stress value of 1.73 ± 0.32 MPa ().
Overall, this implies that the posterior horn fixation is the major load-bearing element in the fixation mechanism. This finding is compatible with the biomechanics of the native meniscus: the posterior horn bore a greater load and is most important for load transmitting when the knee is in flexion (). The role of the posterior horn in resisting anterior tibial translation and generating hoop stress appears to be well replicated by the artificial meniscus, but a strong fixation on the posterior horn is very important. Interestingly, the tensile forces acting in the posterior horn were mainly affected by the applied joint load, with the flexion angle increasing rather than the stiffness of the suture; the only difference observed between stiffness groups was for 30° and 60° flexion.
These findings therefore lead us to consider that although secure fixation is important, what the challenge for the posterior horn suture might be, and that how it manages the loads transmitted from the joint and increasing stiffness beyond a certain point, would not necessarily proportionally reduce the loads to the suture.
We also found the influence of suture stiffness on meniscal displacement to be very significant in the medial compartment. On the basis of the measured mobility of the meniscus in fresh cadaveric knees, the mobility of the medial meniscus posteriorly displaced 5.1 ± 0.96 mm; the anterior horn translated by 7.0 ± 0.7 mm and the posterior horn by 3.2 ± 1.3 mm on extension to 120° flexion (). found with open MRI that meniscal motion was evaluated in the loaded and unloaded conditions. Under load on full extension to 90° flexion, the anterior horn of the medial meniscus posteromedially translated by 7.1 ± 2.49 mm, and the posterior horn by 3.9 ± 1.75 mm and laterally by 3.6 ± 2.32 mm; under no-load conditions, these values were 5.4 ± 2.53 mm, 3.8 ± 1.63 mm, and 3.3 ± 2.27 mm, respectively. By increasing the suture stiffness from 13 N/mm to 200 N/mm, we found both the ML and the AP displacements of the artificial meniscus progressively reduced. Although the ML and AP displacements could not be restored as well as those of native meniscus, the highest stiffness group (the set of 200 N/mm) was the closest. This result directly answers a significant concern in clinical practice of using floating or poorly fixed artefacts such as the NUsurface® device that has been associated with a high incidence of dislocation and subluxation (; ; ). By limiting excessive extrusion and translation, stiffer fixation may improve prosthesis stability and reduce abnormal meniscal movement. This improved stability is a prerequisite for the effective generation of the hoop stress effect, which converts axial loading into circumferential hoop stress (). The contact stress analysis added even more compelling evidence for the advantages of stiff fixation. We also observed that increased suture stiffness led to a decrease in contact stress on the medial tibial cartilage. That is important because the main purpose of meniscal replacement is to reduce cartilage contact stress on the articular cartilage and avoid the long-term development of osteoarthritis (). The high contact stresses on the cartilage observed for the low-stiffness group (13 N/mm) indicates that the prosthesis was unable to sufficiently share the load on account of its excessive mobility and inability to sufficiently induce hoop stress. For the 200 N/mm group, the contact stresses were always closest to the native knee for all flexion angles.
This also indicates the need for robust fixation of the anterior and posterior horns for an artificial meniscus to fulfill its load-distributing function and for its superior chondroprotective effect.
However, we need to take into account the possible trade-off between fixation stability and meniscal mobility. As mentioned in our background, while secure fixation enhances the hoop stress effect and load distribution, it has the potential to impair the meniscus motion during knee flexion and extension, which theoretically increases stresses elsewhere in the joint (). Our results, however, have not shown the adverse effects of the stiff fixation on the lateral compartment and also no paradoxical increment in the peak meniscal stress. Contact stress of the lateral meniscus and tibial cartilage was mostly unaffected by changes in the medial suture stiffness. Thus, the main biomechanical effect is more focused on the medial compartment. Furthermore, the pattern of contact stress distribution of the meniscus was barely affected by suture stiffness, which maintains physiological-like loading patterns.
It also showed that the increased fixation which we investigated stabilized the prosthesis without generating new abnormal stresses on the prosthesis, which has a good balance in this finite element model.
There are limitations to this study. First, our model simulates a simple and static loading situation and does not consider dynamic conditions, muscle forces, or the very complex kinematics of a natural gait cycle. The use of only a knee model from a 45-year-old man also makes our results less generalizable to different individuals, whose anatomy and tissue properties vary. Second, an artificial meniscus was modeled as a homogeneous transversely isotropic material, which does not provide for the complex, non-linear, viscoelastic behavior of a 3D-printed PCU implant. Similarly, we modeled the sutures as linear spring elements—a rough simplification which does not represent their non-linear stress–strain relationship or susceptibility to creep or failure with time. Finally, our model does not simulate the biological response to the implant or the long-term effects of wear and debris, which are major factors for success in clinical trials.
In conclusion, this finite element model shows that suture stiffness is an important factor in the design of meniscal prostheses. Within the tested range and under the specific loading conditions used in this model, higher suture stiffness generally reduced medial meniscal displacement and produced tibial cartilage contact stresses that were closer to those of the intact knee, while the effect on meniscal contact stress was more complex and dependent on flexion angle. Our study suggests that a higher (stiffer) fixation, such as that provided by high-strength sutures or textile tapes (such as 200 N/mm group), generally produces biomechanical responses closest to those of the native knee model under simulated loading conditions. This provides a strong biomechanical rationale for preferring a secured horn fixation in future designs of meniscal prostheses and in surgery to maximize clinical results and delay of the onset of osteoarthritis. Future studies should focus on corroborating our results in cadavers and clinical studies and exploring how these principles of fixation could be improved in combination with a more advanced range of prosthetic materials and designs.
Statements
Data availability statement
The original contributions presented in the study are included in the article/supplementary material; further inquiries can be directed to the corresponding author.
Author contributions
XL: Methodology, Validation, Visualization, Writing – original draft. YC: Formal analysis, Visualization, Writing – review and editing. WY: Supervision, Visualization, Writing – review and editing. XZ: Supervision, Visualization, Writing – review and editing. CF: Conceptualization, Funding acquisition, Supervision, Visualization, Writing – review and editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. Sponsored by the Fundamental Research Funds for the Central Universities (project number YG2024QNB21).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AbrahamA. C.MoyerJ. T.VillegasD. F.OdegardG. M.Haut DonahueT. L. (2011). Hyperelastic properties of human meniscal attachments. J. Biomech.44 (3), 413–418. 10.1016/j.jbiomech.2010.10.001
2
BartoloM. K.ProvaggiE.AthwalK. K.NewmanS.AccardiM. A.DiniD.et al (2022). Strength of interference screw fixation of meniscus prosthesis matches native meniscus attachments. Knee Surg. Sports Traumatol. Arthrosc.30 (7), 2259–2266. 10.1007/s00167-021-06772-9
3
BokshK.ShepherdD. E. T.EspinoD. M.PlakogiannisC.GhoshA.AujlaR.et al (2026). A single root repair and centralisation tunnel best restores tibiofemoral contact mechanics and extrusion following a medial meniscus posterior root tear: an in vitro biomechanical study. Knee Surg. Sports Traumatol. Arthrosc.34 (2), 459–471. 10.1002/ksa.12683
4
CaldwellG. L.AllenA. A.FuF. H. (1994). Functional anatomy and biomechanics of the meniscus. Oper. Tech. Sports Med.2 (3), 152–163. 10.1016/1060-1872(94)90013-2
5
D'LimaD. D.ChenP. C.KesslerO.HoeneckeH. R.ColwellC. W.Jr (2011). Effect of meniscus replacement fixation technique on restoration of knee contact mechanics and stability. Mol. Cell Biomech.8 (2), 123–134. 10.3970/MCB.2011.008.123
6
DhaherY. Y.KwonT. H.BarryM. (2010). The effect of connective tissue material uncertainties on knee joint mechanics under isolated loading conditions. J. Biomech.43 (16), 3118–3125. 10.1016/j.jbiomech.2010.08.005
7
HauchK. N.VillegasD. F.Haut DonahueT. L. (2010). Geometry, time-dependent and failure properties of human meniscal attachments. J. Biomech.43 (3), 463–468. 10.1016/j.jbiomech.2009.09.043
8
KobayashiM.ChangY. S.OkaM. (2005). A two year in vivo study of polyvinyl alcohol-hydrogel (PVA-H) artificial meniscus. Biomaterials26 (16), 3243–3248. 10.1016/j.biomaterials.2004.08.028
9
KurtC.GökA. (2026). Effect of knee joint meniscus tears on joint cartilage contact and pressure with finite element analysis. Biomedicines14 (4), 869. 10.3390/biomedicines14040869
10
KutznerI.HeinleinB.GraichenF.BenderA.RohlmannA.HalderA.et al (2010). Loading of the knee joint during activities of daily living measured in vivo in five subjects. J. Biomech.43 (11), 2164–2173. 10.1016/j.jbiomech.2010.03.046
11
LiG.GilJ.KanamoriA.WooS. L. (1999). A validated three-dimensional computational model of a human knee joint. J. Biomech. Eng.121 (6), 657–662. 10.1115/1.2800871
12
LiuC.TomaI. C.MastrogiacomoM.KrettekC.von LewinskiG.JagodzinskiM. (2013). Meniscus reconstruction: today's achievements and premises for the future. Arch. Orthop. Trauma Surg.133 (1), 95–109. 10.1007/s00402-012-1624-2
13
McKeonB. P.ZaslavK. R.AlfredR. H.AlleyR. M.EdelsonR. H.GersoffW. K.et al (2020). Preliminary results from a US clinical trial of a novel synthetic polymer meniscal implant. Orthop. J. Sports Med.8 (9), 232596712095241. 10.1177/2325967120952414
14
MittonG.EngelkeK.UkS.LaredoJ. D.ChappardC. (2020). A degenerative medial meniscus retains some protective effect against osteoarthritis-induced subchondral bone changes. Bone Rep.12, 100271. 10.1016/j.bonr.2020.100271
15
MohamadiA.MomenzadehK.MasoudiA.WalleyK. C.IerardiK.RamappaA.et al (2021). Evolution of knowledge on meniscal biomechanics: a 40 year perspective. BMC Musculoskelet. Disord.22 (1), 625. 10.1186/s12891-021-04492-2
16
PatschC.DirisamerF.ScheweB. (2021). Relevance of meniscus loss for the progression of osteoarthritis and treatment options for early arthritis. Orthopade50 (5), 366–372. 10.1007/s00132-021-04101-z
17
PenaE.CalvoB.MartinezM.DoblareM. (2006). A three-dimensional finite element analysis of the combined behavior of ligaments and menisci in the healthy human knee joint. J. Biomech.39 (9), 1686–1701. 10.1016/j.jbiomech.2005.04.030
18
PiolettiD. P.RakotomananaL.BenvenutiJ. F.LeyvrazP. F. (1998). Viscoelastic constitutive law in large deformations: application to human knee ligaments and tendons. J. Biomech.31 (8), 753–757. 10.1016/s0021-9290(98)00077-3
19
QuappK. M.WeissJ. A. (1998). Material characterization of human medial collateral ligament. J. Biomech. Eng.120 (6), 757–763. 10.1115/1.2834890
20
RhoJ. Y.Kuhn-SpearingL.ZiouposP. (1998). Mechanical properties and the hierarchical structure of bone. Med. Eng. Phys.20 (2), 92–102. 10.1016/s1350-4533(98)00007-1
21
SaengpetchN.WisitkamthornP.LertbutsayanukulC.ChitrapaztN.VijittrakarnrungC. (2026). Radiologic and clinical analysis of posterior medial meniscal root tears: a comparative study of all-suture anchor and transtibial pullout techniques. Orthop. J. Sports Med.14 (7), 23259671261464696. 10.1177/23259671261464696
22
ShriramD.YamakoG.ChosaE.SubburajK. (2019). Biomechanical evaluation of isotropic and shell-core composite meniscal implants for total meniscus replacement: a nonlinear finite element study. IEEE Access7, 140084–140101. 10.1109/ACCESS.2019.2943689
23
SmithH. E.LyonsM. M.PatelN. M. (2021). Epidemiology of meniscal allograft transplantation at children's hospitals in the United States. Orthop. J. Sports Med.9 (9), 23259671211034877. 10.1177/23259671211034877
24
SukoppM.SchallF.HackerS. P.IgnatiusA.DürselenL.SeitzA. M. (2021). Influence of menisci on tibiofemoral contact mechanics in human knees: a systematic review. Front. Bioeng. Biotechnol.9, 765596. 10.3389/fbioe.2021.765596
25
TuJ. F.WangX. Z.YanS. Y.WangY. r.YangJ. w.ShiG. x.et al (2025). Thermal sensitization of acupoints in patients with knee osteoarthritis: a cross-sectional case-control study. J. Integr. Med.23 (3), 289–296. 10.1016/j.joim.2025.03.004
26
VediV.WilliamsA.TennantS. J.SpouseE.HuntD. M.GedroycW. M. (1999). Meniscal movement. An in-vivo study using dynamic MRI. J. Bone Jt. Surg. Br.81 (1), 37–41. 10.1302/0301-620x.81b1.8928
27
VerhaegheL.BoerenK. (2018). A rare complication after synthetic meniscus replacement. J. Belg Soc. Radiol.102 (1), 63. 10.5334/jbsr.1601
28
VranckenA. C.EggermontF.van TienenT. G.HanninkG.BumaP.JanssenD.et al (2016). Functional biomechanical performance of a novel anatomically shaped polycarbonate urethane total meniscus replacement. Knee Surg. Sports Traumatol. Arthrosc.24 (5), 1485–1494. 10.1007/s00167-015-3632-6
29
VranckenA. C. T.HanninkG.MadejW.VerdonschotN.van TienenT. G.BumaP. (2017). In vivo performance of a novel, anatomically shaped, total meniscal prosthesis made of polycarbonate urethane: a 12-month evaluation in goats. Am. J. Sports Med.45 (12), 2824–2834. 10.1177/0363546517713687
30
WangB.LiuQ.WiseB. L.KeY.XingD.XuY.et al (2018). Valgus malalignment and prevalence of lateral compartmental radiographic knee osteoarthritis (OA): the wuchuan OA study. Int. J. Rheum. Dis.21 (7), 1385–1390. 10.1111/1756-185X.13079
31
WangH.ZhangB.ChengC. K. (2020). Stiffness and shape of the ACL graft affects tunnel enlargement and graft wear. Knee Surg. Sports Traumatol. Arthrosc.28 (7), 2184–2193. 10.1007/s00167-019-05772-0
32
WellsM. E.ScanaliatoJ. P.DunnJ. C.GarciaE. J. (2021). Meniscal injuries: mechanism and classification. Sports Med. Arthrosc. Rev.29 (3), 154–157. 10.1097/jsa.0000000000000311
33
WillingerL.FoehrP.AchtnichA.ForkelP.VossA.LiskaF.et al (2019). Effect of lower limb alignment in medial meniscus-deficient knees on tibiofemoral contact pressure. Orthop. J. Sports Med.7, 2325967118824611. 10.1177/2325967118824611
34
YaoJ.SnibbeJ.MaloneyM.LernerA. L. (2006). Stresses and strains in the medial meniscus of an ACL deficient knee under anterior loading: a finite element analysis with image-based experimental validation. J. Biomech. Eng.128 (1), 135–141. 10.1115/1.2132373
35
YaoJ.LancianeseS. L.HovingaK. R.LeeJ.LernerA. L. (2008). Magnetic resonance image analysis of meniscal translation and tibio-menisco-femoral contact in deep knee flexion. J. Orthop. Res.26 (5), 673–684. 10.1002/jor.20553
36
ZaslavK. R.FarrJ.AlfredR.AlleyR. M.DyleM.GomollA. H.et al (2022). Treatment of post-meniscectomy knee symptoms with medial meniscus replacement results in greater pain reduction and functional improvement than non-surgical care. Knee Surg. Sports Traumatol. Arthrosc.30 (4), 1325–1335. 10.1007/s00167-021-06573-0
37
ZhangX.YuanS.WangJ.LiaoB.LiangD. (2021). Biomechanical characteristics of tibio-femoral joint after partial medial meniscectomy in different flexion angles: a finite element analysis. BMC Musculoskelet. Disord.22 (1), 322. 10.1186/s12891-021-04187-8
38
ZurG.Linder-GanzE.ElsnerJ. J.ShaniJ.BrennerO.AgarG.et al (2011). Chondroprotective effects of a polycarbonate-urethane meniscal implant: histopathological results in a sheep model. Knee Surg. Sports Traumatol. Arthrosc.19 (2), 255–263. 10.1007/s00167-010-1210-5
Summary
Keywords
contact stress, finite element analysis, hoop stress, knee biomechanics, meniscal prosthesis, suture stiffness
Citation
Li X, Chen Y, Yu W, Zhou X and Fang C (2026) Effect of anterior and posterior horn suture stiffness on the biomechanical performance of a medial meniscal prosthesis: a finite element analysis. Front. Bioeng. Biotechnol. 14:1973619. doi: 10.3389/fbioe.2026.1973619
Received
20 August 2026
Revised
13 September 2026
Accepted
25 September 2026
Published
11 November 2026
Volume
14 - 2026
Edited by
Yanan Jiang, Harbin Medical University, China
Reviewed by
Phi Duong Nguyen, City Children’s Hospital, Vietnam
Arif Gök, Dumlupinar University, Türkiye
Updates
Copyright
© 2026 Li, Chen, Yu, Zhou and Fang.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Chaohua Fang, fangchaohua@sjtu.edu.cn
† These authors have contributed equally to this work and share first authorship
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.