AUTHOR=Kassi Mahwash , Filippini Stefano , Avenatti Eleonora , Xu Susan , El-Tallawi Kinan Carlos , Angulo Clara I. , Vukicevic Marija , Little Stephen H. TITLE=Patient-specific, echocardiography compatible flow loop model of aortic valve regurgitation in the setting of a mechanical assist device JOURNAL=Frontiers in Cardiovascular Medicine VOLUME=Volume 10 - 2023 YEAR=2023 URL=https://www.frontiersin.org/journals/cardiovascular-medicine/articles/10.3389/fcvm.2023.994431 DOI=10.3389/fcvm.2023.994431 ISSN=2297-055X ABSTRACT=Background - Aortic regurgitation (AR) occurs commonly in patients with continuous-flow left ventricular assist devices (LVAD). No gold standard is available to assess AR severity in this setting. Aim of this study was to create a patient-specific model of AR-LVAD with tailored AR flow assessed by Doppler echocardiography. Methods – An echo-compatible flow loop incorporating a 3D printed left heart of a Heart Mate II (HMII) recipient with known significant AR was created. Forward flow and LVAD flow at different LVAD speed were directly measured and AR regurgitant volume (RegVol) obtained by subtraction. Doppler parameters of AR were simultaneously measured at each LVAD speed. Results - We reproduced hemodynamics in a LVAD recipient with AR. AR in the model replicated accurately the AR in the index patient by comparable Color Doppler assessment. Forward flow increased from 4.09 to 5.61 L/min with LVAD speed increasing from 8800 to 11000 RPM while RegVol increased by 0.5 L/min (2.01 L/min to 2.5 L/min). Conclusions - Our circulatory flow loop was able to accurately replicate AR severity and flow hemodynamics in an LVAD recipient. This model can be reliably used to study echo parameters and aid clinical management of patients with LVAD.