AUTHOR=Kahrovic Amila , Angleitner Philipp , Herkner Harald , Werner Paul , Poschner Thomas , Alajbegovic Leila , Kocher Alfred , Ehrlich Marek , Laufer Günther , Andreas Martin TITLE=Automated titanium fastener vs. hand-tied knots for prosthesis fixation in infective endocarditis JOURNAL=Frontiers in Cardiovascular Medicine VOLUME=Volume 11 - 2024 YEAR=2024 URL=https://www.frontiersin.org/journals/cardiovascular-medicine/articles/10.3389/fcvm.2024.1363336 DOI=10.3389/fcvm.2024.1363336 ISSN=2297-055X ABSTRACT=To date, there is no evidence regarding the safety of automated titanium fastener compared with hand-tied knots for prosthesis fixation in infective endocarditis.Between January 2016 and December 2022, a total of 220 patients requiring surgery for infective endocarditis were included in this retrospective analysis. The primary study endpoint was re-endocarditis during follow-up. The secondary study endpoints included stroke onset, allcause mortality, and a composite outcome of either re-endocarditis, stroke, or all-cause mortality during follow-up.Suture-securing with an automated titanium fastener was performed in 114 (51.8%) patients, whereas the conventional technique of hand knot-tying was used in 106 (48.2%) patients. The risk of re-endocarditis was significantly lower in the automated titanium fastener group, as shown in a multivariable proportional competing risk regression model (adjusted sub-hazard ratio 0.33, 95% confidence interval 0.11 -0.99, p = 0.048). The multivariable Cox proportional hazards regression analysis showed that the automated titanium fastener group was not associated with an increased risk of stroke-onset or attaining the composite outcome, respectively, (adjusted hazard ratio 0.54, 95% confidence interval 0.27 -1.08, p = 0.082), (adjusted hazard ratio 0.65, 95% confidence interval 0.42 -1.02, p = 0.061). Also, this group was not associated with an increased risk of all-cause mortality, as demonstrated in the multivariable Poisson regression analysis (adjusted incidence-rate ratio 1.42, 95% confidence interval 0.83 -2.42, p = 0.202).The use of automated titanium fastener device seems to be safe for infective endocarditis. Analyses of larger cohorts are required.