Patient-Specific Simulation of Thoracic Endovascular Aortic Repair for Stanford Type B Aortic Dissection: Development and Validation of a Personalized Simulation System
- YZ
Yazheng Zhang 1
- ZZ
Zhijun Zheng 1
- LM
Lilong Ma 1
- YT
Yujun Tian 1
- PC
Pinghong Chen 2
- TN
Tao Niu 1
- YL
Youjin Li 1
1. People’s Hospital of Ningxia Hui Autonomous Region, Ningxia Medical University, NingXia, China
2. The Third Clinical College of Ningxia Medical University, Ningxia, China
Select one of your emails
You have multiple emails registered with Frontiers:
Notify me on publication
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Abstract
Background TEVAR is a primary treatment modality for Stanford Type B aortic dissection (TBAD). However, patient-specific anatomical variations and high post-procedural complication rates demand rigorous preprocedural planning and refined operative skills. Objectives This study aims to assess the preclinical utility and training effectiveness of a 3D-printed TEVAR simulator. Methods A total of 42 cardiovascular or vascular surgeons participated in this study. A patient-specific flexible 3D-printed model integrated into a pulsatile hemodynamic circuit was developed, incorporating a flexible dissection flap and all major branch vessels. Key anatomical parameters were measured and validated against original imaging data. Surgeons performed core procedural steps under fluoroscopy, with deployment and stent implantation times recorded. Finally, the simulator was evaluated using a Likert-scale questionnaire. Results The 42 participants had a mean age of 34.7 ± 5.2 years and an average clinical experience of 6.6 ± 3.5 years. Aortic models were fabricated based on original imaging data from 20 patients, and the measured parameters showed high consistency with real anatomical structures. Clinicians demonstrated significant improvement across four successive simulation sessions: component deployment time decreased from 6.1 ± 2.4 min to 2.9 ± 1.1 min, and stent implantation time reduced from 13.4 ± 5.9 min to 5.9 ± 2.5 min (both P < 0.001). Furthermore, all young clinicians expressed overall satisfaction with the simulator's comprehensive performance, procedural details, and material properties after simulations. Conclusions This integrated hemodynamic simulation system offers high precision and fidelity, shortening clinicians' learning curves and offering potential advantages for TEVAR training targeting TBAD.
Summary
Keywords
3D printing, aortic dissection, simulator, Stanford type B, Thoracic endovascular aortic repair, training
Received
21 April 2026
Accepted
27 September 2026
Copyright
© 2026 Zhang, Zheng, Ma, Tian, Chen, Niu and Li. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Youjin Li
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.