Evaluating the accuracy of intraoral scanners across varied depths and widths preparation for CAD/CAM post and core

  • 1. Clinical Sciences Department, College of Dentistry, Ajman University, Ajman, United Arab Emirates

  • 2. Center of Medical and Bio-Allied Health Sciences Research, Ajman University, Ajman, United Arab Emirates

  • 3. Department of Prosthodontics, Gerodontology and Dental Materials, Greifswald University Medicine, Greifswald, Germany

Abstract

Objectives:

This study aimed to investigate the accuracy of two intraoral scanners at different post space preparation depths and widths for CAD/CAM fabricated post and core fabrication.

Methods:

A total of 36 extracted anterior teeth were used in this study, which were divided into 6 groups according to the different widths and depths of post space preparations. Two operators carried out the intraoral scanning using two scanners: Planmeca Emerald and 3Shape TRIOS 4. Furthermore, the inEos X5 extraoral laboratory scanner was used as a reference. 3D point cloud processing software was used to analyze the impact of different widths and depths on the scanners' accuracy.

Results:

After ensuring examiner blinding, it showed that there was no significant difference between the 3Shape and Planmeca scanners. Additionally, longer posts resulted in lower scanning accuracy across the board, whereas wider posts tended to produce results that were more accurate.

Conclusions:

This study found that there was no difference in accuracy between the 3Shape and Planmeca scanners. The distribution of scanning accuracy at different post space preparation depths and widths showed that longer posts resulted in lower scanning accuracy, whereas wider posts tended to produce results that were more accurate, which provides information for the CAD/CAM post and core fabrication.

1 Introduction

In recent years, digital dentistry has transformed traditional dental practices, presenting unmatched precision, efficiency, and patient comfort (). Improved therapeutic interventions, treatment planning, and diagnostics have become possible as a result of workflows in dentistry that incorporate state-of-the-art technologies (). Together with these developments, intraoral scanners (IOS) have become essential instruments that enable dentists to get precise, 3D images of the oral cavity (, ). IOS has improved treatment outcomes and patient satisfaction by replacing traditional impression-taking techniques and modernizing the production of various dental prostheses, such as crowns, bridges, and implants (, ). In addition, digital dentistry has gone further with the use of computer-aided design (CAD) and computer-aided manufacturing (CAM) technologies, making it easier to fabricate custom restorations with a high degree of accuracy and precision (). The accuracy of a scanner is the ability of a measurement to match the actual value, and the precision of a scanner is the ability of a measurement to be consistently reproduced (). The accuracy can also be termed as trueness. The IOS presents some distortion and most often presents stitching errors (, ). The distortion of the scan varies with the scanners, techniques, and pattern of scan. The full-arch intraoral scans can be obtained by various scan strategies, with the segmental scan and merge methods. The segmental scan and merge methods using two scan parts can be a reliable alternative to the single scan method for full-arch intraoral scans (). But when more than three segmental scans negatively affect the accuracy of the complete arch scan.

Following the endodontic treatment, an appropriate restoration of the tooth establishes a coronal seal and protects the remaining dental structure. For the teeth with significant damage, posts and cores serve to stabilize and retain the crowns (). Several digital techniques have been used in the making of the post and core, which not only reduce patient chair time but also eliminate the necessity for temporary restorations and repetitive adjustments, thus improving treatment efficiency and minimizing overall treatment duration (, , ). However, as with any technological evolution, the efficacy, accuracy, and precision of IOS require thorough evaluation, particularly in significant applications such as CAD/CAM post and core fabrication ().

Digital fabrication for the post can be done by various techniques: direct digital technique by directly scanning the post, semi-direct technique by scanning the root canal impression, semi-direct technique by scanning the resin pattern, and indirect fabrication of the post. Direct digital technique involves directly scanning the post space using an IOS after root canal treatment, and designing using the CAD software (). The semi-direct technique involves scanning the root canal by injecting polyvinyl siloxane into the prepared root canal and then making an impression with heavy-body impression material (, 19). In the semi-direct technique by scanning the resin pattern, the auto-polymerizing acrylic resin (GC Pattern Resin) post (20). Following the scanning and computer designing, the post-and-core can be milled using different materials, such as fiber-reinforced composite (21) and fiberglass (22). This is the most practical method of fabrication of the post.

The evaluation of marginal and internal adaptation in dental restorations is crucial for the long-term success and functionality. Different techniques in measuring the internal gap in post and core include the silicone replica technique, micro-computed tomography [micro-CT], digital technique, semi-digital technique, optical coherence tomography (OCT), and radiograph technique (2328). Among various methods, the silicone replica technique stands out as a reliable and commonly utilized approach. Similarly, the digital technique involves the fabrication of posts and cores using the digital workflow. The ruler tool in Adobe Photoshop can be used to analyze the images (26). In a semi-digital technique of fabricating customized posts and cores, digital data using light scanning is used to acquire data from a manually created wax pattern manufactured using the traditional lost-wax procedure.

With this amount of diversity and complexity, thorough assessments of intraoral scanner accuracy become vital, particularly in the setting of CAD/CAM post and core preparations. This study aimed to investigate the accuracy of two intraoral scanners through varied depths and widths of tooth preparations, offering valuable insights into their clinical performance and potential limitations.

2 Materials and methods

2.1 Study samples and study groups

The study protocol was approved by the Institutional Review Board of the Ajman University (protocol code D-H-S-2023-NOV-20-14 and approval date 7 December 2023). From a total of 150 extracted anterior teeth, 36 intact teeth were selected according to the selection criterion: sound teeth, intact root, straight canals, single canal, and fully formed apex. The teeth were examined radiographically to evaluate the canals, to assess the anatomy, and to identify the canal curvature. Additionally, the teeth were examined using a stereomicroscope under high magnification to evaluate each tooth according to the criteria and check for the presence of any fractures or cracks. Teeth were excluded due to curved canals, incomplete apex formation, or the presence of cracks/fractures observed under the stereomicroscope. Hand scaler was used for the root surface debridement of each sample, followed by disinfection with 5% NaOCl solution for 24 h at 4 °C, and afterward, they were stored in 0.1 thymol solution (29, 30).

The study groups consisted of 9 groups of different widths and depths of post space preparation. The three widths used were 1.8, 2.1, and 2.3 mm, and the lengths were 6, 8, and 10 mm. To classify the teeth, each tooth was evaluated radiographically using the digital ruler to measure the length and width of the canal; accordingly, the samples (teeth) were prepared.

2.2 Sample preparation

After classifying the teeth according to the radiograph, the anatomical crowns of the teeth were sectioned 2 mm coronal to the cementoenamel junction (CEJ), and only the roots were used for the study. Each tooth was sectioned to the specific length of the group it belonged to, and this step was done by using a micromotor with a straight handpiece and diamond-coated metal (double-sided) cutting discs. The fine diamond discs permit precise cuts without affecting the tooth's integrity or producing any irregularities. The Periodontal Williams probe was used to clinically check the canal orifice after discing the teeth.

2.3 Post-Space preparation

Following sample preparation, cleaning, and shaping of the canals was done. Canals were irrigated with 1.75% of sodium hypochlorite (NaOCl) and saline using endodontic irrigation needles (Endo-Top, Cerkamed, Stalowa Wola, Poland). Patency of the canals was checked with K-Files size 10. After establishing the working length with the initial apical file for each tooth (Figure 1A), the canal system was shaped and prepared using the WaveOne Gold single-file reciprocating system (WO; Dentsply Maillefer, Ballaigues, Switzerland) (Figure 1B). To standardize all canal systems, all teeth were instrumented with the same large-size file. We used an endodontic instrument of size 25 (ISO 025) K-file and a rotary file, which has a tip diameter of 0.25 mm.

Figure 1

After root canal instrumentation, all the canals were dried with absorbent paper points (Meta Biomed Co, Ltd, Chungbuk, Korea), then obturated with a single cone obturation technique using WaveOne Gold Gutta Percha Points (DentsplyMaillefer, Ballaigues, Switzerland) that match the size of the prepared canal (Figure 1C). In all the samples, AH Plus (Dentsply De Trey, Konstanz, Germany) was used as an endodontic sealer. After endodontic preparation and obturation, the gutta-percha cone was cut by using the Endodontic Gutta-percha Cutter (Elements™ IC Backfill Unit, Kerr, California, USA). Teeth were kept at 37 °C and 100% humidity for 14 days in order to give the sealer time to fully set (31).

To achieve the desired widths and lengths of the teeth, specific burs were used. Tapered diamond burs with widths of 1.8, 2.1, 2.3 mm and lengths of 6, 8, 10 mm were used (Komet, Germany). All burs had a taper of 4 degrees.

Hydrorise putty (Zhermack SpA, Badia Polesine, Italy) was mixed and adapted onto the modeling table MT3 of the milling machine (AF 350, Amann Girrbach, Koblach, Austria) with a metal mold placed in the center. The metal mold was used to achieve a standardized shape and size for the acrylic blocks. To minimize material wastage, a small amount of hydrorise putty was added to the metal mold, filling it halfway (Figure 2C).

Figure 2

Before commencing the acrylic block fabrication process, the tooth was positioned according to its long axis (Figure 2A). Size 2 (0.7 mm) Gates Glidden drills (Mani Inc. Japan) were inserted into the gutta-percha partially (not to the full targeted preparation depth) to ensure clear long-axis orientation. Sticky wax was heated and applied between the drill and tooth to create a tight seal, preventing movement or disengagement. The assembled Gates Glidden drill and tooth were then placed into the bur slot of the milling machine.

Before mixing the acrylic, a thin layer of separating medium was applied to the metal mold to facilitate easy removal of the acrylic after setting. The tooth was positioned in the metal mold with its surface raised 2–3 mm above the metal's surface, resembling the ferrule (Figure 2B). This adjustment was achieved by manipulating the “pivoted lever for drilling” and turning the “fixation screw for height adjustment,” followed by locking the desired height with the “locking screw for jointed arm” of the milling machine.

Subsequently, an appropriate amount of Orthoplast Liquid (Vertex-Dental B.V., Soesterberg, Netherlands) was mixed with Vertex Orthoplast Powder to achieve a thin consistency. This mixture was poured into the mold, and any excess was carefully removed. The tooth was positioned, and the acrylic block was left to set (32).

For accurate and precise post space preparation, the procedure was carried out using the precision milling machine to eliminate potential human errors, such as hand tilting, which could result in deviation from the long axis of the tooth and potential perforation.

The post-space preparation was performed using a standardized drill (Taper 4, angle of convergence 8°). A size 2 (0.7 mm) Gates Glidden drill was used to remove the gutta-percha of the obturated tooth, leaving 5 mm of apical gutta-percha. The micrometer of the milling machine was adjusted to achieve the desired depth of penetration. Gutta-percha was removed from the 2/3 length of the root, keeping a minimum of 5 mm at the apex. The final post-space exhibited a coronal diameter of 2.3 mm at the orifice and extended to a depth of 10 mm along the canal's long axis. Owing to the 4° taper per side, the apical diameter was approximately 0.9 mm.

Subsequently, peeso reamers (Azdent, Henan, China) were sequentially employed, up to the maximum size that was smaller apically than the apical diameter of the final bur, such that we ensured that the outline of the preparation was created by the final bur. This was determined by the formula: Diameter—tan (4) × Length, where the diameter is the diameter of the bur to be used, 4 represents the degree taper of the bur, and length represents the length to which the tooth is to be prepared. For instance, for a tooth with a diameter of 2.3 mm and a length of 10 mm, the maximum allowed peeso reamer size was calculated to be 1.6 mm. Consequently, only reamer sizes up to 1.5 mm were used. The micrometer was adjusted according to the post depth to maintain the desired apical gutta-percha length.

Saline irrigation was continuously directed onto the reamer to prevent overheating and ensure adequate lubrication. To finalize post space preparation with the final bur, a designated bur (with the aimed width and length) was selected and placed into the milling machine's bur slot. The bur tip was positioned precisely above the orifice center. After adjusting the micrometer to the required depth, the magnet of the modeling table was locked, and saline irrigation was applied. The bur was allowed to enter with light pressure and copious irrigation. At halfway, the bur was cleaned to remove debris that could hinder reaching the full length. The tooth was irrigated with saline midway to prevent overheating.

Multiple entries with the bur were avoided to prevent unnecessary widening of the post space. After preparation, an x-ray was taken to confirm proper alignment along the tooth's long axis. Following x-ray confirmation, three equal grooves were created on the tooth using a carbide straight fissure bur, each measuring 0.5 × 0.5 mm, to facilitate later superimposition in the 3D software.

2.4 Intraoral scanning

Two IOSs were used for this study: Planmeca Emerald intraoral scanner using Planmeca Romexis® software (Helsinki, Finland) (Figure 3A) and 3Shape TRIOS 4 Intraoral scanner (Copenhagen, Denmark) (Figure 3B). Both IOSs were calibrated in accordance with the manufacturer's instructions before scanning.

Figure 3

The study consisted of 9 groups of varying three post lengths and three post widths. Each tooth was scanned 4 times with each IOS (36 scans per scanner); Planmeca Emerald intraoral scanner using Planmeca Romexis® software (v6.4.5.202, Planmeca, Helsinki, Finland) and TRIOS 4 intraoral scanner (v23.1.4; 3Shape, Copenhagen, Denmark). A total of 72 scans were obtained, and they were exported as STL files to be used later in the 3D superimposition software.

The InEos X5 extraoral laboratory scanner (Dentsply Sirona, PA, USA) scanned the silicone replica. Afterward, to digitally compare between 2 different sets of the intraoral 3D scanners, a 3D point cloud processing software (Cloudcompare, Paris, France) program was used, which is able to handle 3D mesh.

Two trained operators took part in this step. Inter and intra-examiner reliability was estimated. To begin with the scans, standardized coding was used for both intraoral scanners and for both operators.

2.5 Silicone replica fabrication

After the completion of scanning, silicone replica fabrication was performed. Plastic posts (Uniclip 1 MM, DentsplySirona, Ballaigues, Switzerland) were utilized. Initially, the plastic post was shortened using a heated wax knife to create a stopper, and then it was roughened using a straight diamond bur (Figure 4A). Subsequently, with the plastic post inserted into the canal, a small amount of Hydrorise putty (Zhermack SpA, Badia Polesine, Italy) was mixed and applied in the canal using a size 25 lentirulo spiral, followed by inserting the post (Figure 4B). Then, the putty (Zhermack SpA, Badia Polesine, Italy) was applied to the head of the post.

Figure 4

Following this step, the putty and post assembly was removed from the canal as one piece. Upon removal, the replica was evaluated, ensuring the absence of voids, tears, and complete penetration of the impression material into the grooves (Figure 4C). The silicone replica allows for precise measurement and visualization of gaps at various interfaces, such as the buccal, palatal, mesial, and distal aspects, aiding in identifying areas of discrepancy or misfit.

The InEos X5 extraoral laboratory scanner (Dentsply Sirona, PA, USA) was used to scan the silicon replica, initially by placing the silicon replica in a magnetic circular mold and a catalyst putty was placed inside the magnetic mold to position the replica inside the mold, avoiding it from being displaced or falling from the mold during scanning. All samples were sprayed with Scan Spray (Easy Scan, Alpha Dent, Germany) before scanning to increase the quality and the accuracy of the scans by eliminating reflections and improving the appearance of the texture of the scanned replicas. All replicas were scanned and exported as STL files to be used in the 3d superimposition software.

2.6 Superimposition using 3D point cloud processing software

The STL files that were exported from the intraoral scanners (Planmeca and TRIOS) and the extraoral laboratory scanner (inEos X5) are then imported into the 3D point cloud processing software (Cloudcompare, EDF R&D, Paris, France). The CloudCompare software can handle both point clouds and 3D mesh files (e.g., STL, OBJ, PLY). The 3D scans were initially imported as STL mesh files, which were then converted into point clouds within CloudCompare to enable a precise, quantitative surface-to-surface comparison. This conversion facilitates the computation of cloud-to-mesh and cloud-to-cloud distances, including deviation maps and statistical parameters such as mean, standard deviation, and RMS values.

Each scan was superimposed with its matching replica tool and adjusted manually to reduce the error (Figure 5). The superimposition was based on the entire acrylic block with the grooves, while the analysis itself focused specifically on the internal surface of the post space after segmentation. Using rotation/translation mode, the two scans were superimposed according to the prepared grooves. Only the post space preparation was segmented. The mean of the distribution fitting, standard deviation, and RMS (Root Mean Square) of the difference between the two superimposed scans were exported. Additionally, the post space length of each scanned replica and the matching scan was measured from the most cervical part to the most apical part. After the data were collected, it was documented and exported into an Excel spreadsheet for data analysis.

Figure 5

2.7 Statistical analysis

Descriptive statistics were calculated, and the normality of the data was analyzed. The statistical analyses were performed using SPSS 23.0 (SPSS Inc.). A p-value of <0.05 was statistically significant. Comparison of discrepancies among various post lengths and post widths with reference was done using the Kruskal–Wallis Test. Multiple comparisons of discrepancies among various post lengths and post widths were done using the Mann–Whitney U Test. The unit of discrepancies is mm.

The dispersion of the 3D image analysis was performed by the two examiners at two times in different sessions, 1 week apart. The two examiners had been previously trained in the use of the 3D point cloud processing software (Cloudcompare, Paris, France) program. To ensure examiner blinding, the STL files (3D images) were identified by codes and were analyzed anonymously in a random order. The intraclass correlation coefficient (ICC) was calculated for both intra-examiner and inter-examiner reliability, which considers both examiners and images as random factors.

3 Results

3.1 Descriptive results

The descriptive statistics of the RMS for each case according to the overlap in the scanning first and second examiners are shown in Table 1. The results of the Shapiro–Wilk test showed that the p-value was <0.05, which signified that the data were not normally distributed. In addition, the Q-Q plot, histogram, skewness, and kurtosis reflected similar results.

Table 1

ExaminersStatisticStd. Error
First ExaminerMean0.090.005
95% Confidence Interval for MeanLower Bound0.08
Upper Bound0.10
5% Trimmed Mean0.08
Median0.07
Variance0.002
Std. Deviation0.043
Skewness1.1730.283
Kurtosis1.3260.559
Second ExaminerMean0.070.006
95% Confidence Interval for MeanLower Bound0.05
Upper Bound0.08
5% Trimmed Mean0.06
Median0.05
Variance0.003
Std. Deviation0.052
Skewness2.7210.283
Kurtosis11.0730.559

Descriptive statistics of the root mean square (RMS) value of each case according to the overlap in the scanning of the first and second examiners.

Furthermore, the intraclass correlation coefficient (ICC) in the main examiner (two times in different sessions, 1 week apart) and between the main examiner vs. second examiner is shown in Table 2. The main examiner's ICC was excellent; intra-examiner reliability ICC > 0.9, and inter-examiner reliability was acceptable ICC∼0.6. As the data were not normally distributed, nonparametric tests were used.

Table 2

SummaryIntraclass Correlation95% Confidence Interval
Lower BoundUpper BoundValue
Average Measures0.971a0.9240.98938.030
Average Measures0.565b0.2970.7302.492

Intraclass correlation coefficient (ICC) in the main examiner (two times in different sessions, 1 week apart) and between the main examiner vs second examiner.

a

Strong intraclass correlation between the two measures in the main examiner.

b

Acceptable intraclass correlation between the two examiners.

3.2 Impact of the post length

The rank distribution of various groups of various post lengths is shown in Table 3. Kruskal–Wallis Test showed that the discrepancy between the STL scanning files (Planmeca and Trios) and the STL generated from the replica (reference) was a significant difference (p-value = 0.002) and the longer the post, the higher the discrepancy (Table 4).

Table 3

RanksPost LengthMean Rank
Groups629.88
830.92
1048.71

Rank distribution of various groups of various post lengths.

Table 4

Test Statisticsa,bGroups
Kruskal–Wallis H12.280
Df2
P-value0.002*

Comparison of discrepancies among various post lengths with reference using the Kruskal–Wallis test.

a

Kruskal Wallis Test.

b

Grouping Variable: Post Length.

*

Statistically significant difference at p-value of 0.05.

Moreover, the Mann–Whitney Test revealed that there was a significant difference between the 6 mm vs. 10 mm (p-value = 0.001) and 8 mm vs. 10 mm (p-value = 0.006) lengths. However, there was no statistically significant difference between 6 mm and 8 mm (p-value = 0.967) lengths (Table 5).

Table 5

Post lengths6 mm vs. 8 mm6 mm vs. 10 mm8 mm vs. 10 mm
Mann–Whitney U286.000127.000156.000
Wilcoxon W586.000427.000456.000
Z−0.041−3.320−2.722
p-value0.9670.001*0.006*

Multiple comparisons of discrepancies (mm) among various post lengths using the Mann–Whitney U test.

*

Statistically significant difference at p-value of 0.05.

3.3 Impact of the post width

The rank distribution of various groups of various post widths is shown in Table 6. Kruskal–Wallis Test revealed that the discrepancy between the STL scanning files (Planmeca and Trios 4) and the STL generated from the replica (reference) was a significant difference (p-value = 0.001), and the wider the post, the smaller the discrepancy (Table 7).

Table 6

RanksPost widthsMean Rank
Groups1.847.71
2.136.96
2.324.83

Rank distribution of various groups of various post widths.

Table 7

Test Statisticsa,bGroups
Kruskal–Wallis H14.353
Df2
p-value0.001*

Comparison of discrepancies among various post widths with reference using the Kruskal-–Wallis test.

a

Kruskal Wallis Test.

b

Grouping Variable: Post Length.

*

Statistically significant difference at p-value of 0.05.

Moreover, the Mann–Whitney Test revealed that there was a significant difference between the 1.8 mm vs. 2.3 mm (p-value < 0.001) and 2.1 mm vs. 2.3 mm (p-value = 0.048) widths. However, there was no statistically significant difference between 1.8 mm vs. 2.1 mm (p-value = 0.080) (Table 8).

Table 8

Post widths1.8 mm vs. 2.1 mm1.8 mm vs. 2.3 mm2.1 mm vs. 2.3 mm
Mann–Whitney U203.000104.000192.000
Wilcoxon W503.000404.000492.000
Z−1.753−3.794−1.979
p-value0.080<0.0001*0.048

Multiple comparisons of discrepancies (mm) among various post widths using the Mann–Whitney U test.

*

Statistically significant difference at p-value of 0.05.

The unit of discrepancies is mm.

4 Discussion

This study investigated the accuracy of intraoral scanners at different post space preparation depths and widths for CAD/CAM fabricated post and core fabrication. The differences in deviations were precisely quantified when using the IOS technique as compared to the polyvinyl siloxane silicone replica, the traditional impression technique in post and core fabrication. The conventional dental impressions were considered the control group. In this research, we investigated the possibilities of using IOS technology to produce posts that are anatomically suitable (the least amount of preparation). These posts are recognized for several advantageous characteristics, including the preservation of a greater amount of the natural root dentin (33, 34), reducing the layer of the cement (35), minor air bubbles formation during the process (36), improved post retention (37, 38), and an increase in fracture resistance (39, 40). The results showed that the data are not normally distributed, as confirmed by the Shapiro–Wilk test. The ICC indicated excellent intra-examiner reliability (ICC > 0.9), indicating that the scanners stay consistent when used by the same operator, which is critical for clinical applications. However, the inter-examiner reliability was acceptable (ICC∼0.6), which reveals that results may vary when different operators operate the scanners. This emphasizes the significance of good training and standardization to minimize variability among operators. Statistically significant differences were observed in scanning accuracy between the group using 2.3 mm post width preparation and the other groups using 1.8 mm and 2.1 mm post width preparation, indicating variations in accuracy based on post width. However, there was no significant difference between the groups using 1.8 mm and 2.1 mm post widths. The observed discrepancies based on post length and width indicate that these factors can influence the accuracy of intraoral scanners. Longer posts may pose challenges for accurate scanning, possibly due to difficulty capturing the entire length. Conversely, wider posts may provide better surface area for scanning, leading to more accurate results. In addition, short and wide posts are generally not the most effective way to reduce the risk of vertical root fractures, as they can cause excessive dentin removal and increase stress concentration. Instead, conservative preparation is preferred, utilizing longer, narrower posts; ideally, fiber-reinforced composite (FRC) with an elasticity similar to dentin, to achieve better stress distribution (41).

In this study, 3Shape TRIOS 4 and Planmeca Emerald S intraoral scanners because they are widely used in clinical and research settings, representing two different scanning technologies and acquisition strategies. The 3Shape TRIOS employs a confocal imaging principle, while the Planmeca Emerald S uses a short-wavelength laser with active triangulation. This allowed us to compare scanners with distinct optical systems and data-processing algorithms, providing a broader and more meaningful evaluation of accuracy in post-space digitization.

The trueness (accuracy) of the TRIOS 4 scanner is approximately 20.8 ± 6.2 µm or 34.0 [14.8] µm, while its precision is around 19.75 µm (IQR 17.90–21.95 µm) (42). Another study (43) found that Planmeca Emerald had significantly lower trueness values than either the 3Shape A/S TRIOS 3 (p-value = 0.001). Still, the trueness (accuracy) and precision of Trios 4 and Planmeca Emerald are comparable. In this study, there was no significant difference between the Planmeca and Trios scanning systems. Clinicians should consider these factors when selecting posts for CAD/CAM direct technique procedures to ensure optimal accuracy. It is important to comprehend the elements that impact the accuracy and precision of scanners for making clinical decisions and planning treatments. Clinicians need to understand the possible limitations of intraoral scanning devices and consider variables like post widths and lengths when evaluating the appropriateness of a scanner for clinical cases. It was found that the newer scanners (e.g., TRIOS 4, Primescan) show greater resilience to distance/speed variations and enable flexible clinical workflows (44).

The incorporation of CAD-CAM technology enhances the restorative procedure, providing a safer and more effective approach, saves time, and reduces the risk of complications that may occur while preparing the root canal (4547). The application of the CAD/CAM approach can lead to improvements in the effectiveness, efficiency, and accuracy of workflows and accuracy in prosthesis therapy. It empowers clinicians to create post and core restorations using tooth-colored materials that offer enhanced physical properties and superior mechanical qualities. This enables achieving optimal results in the restoration of endodontically treated teeth (26).

Nulty (42) conducted a study that compared the complete arch trueness of nine different intraoral scanners and four lab light digital scanners, and he reported that Ineos X5 had the best overall trueness value of (0.0 ± 1.9). Consequently, compared to all intra-oral scanners, the Ineos X5 demonstrated statistically superior precision. Researchers frequently assess the accuracy of their findings by measuring the trueness and precision, which has been a common approach in multiple studies (48, 49).

Similarly, the ferrule can be placed 3 mm above the CEJ to simulate clinical settings, resulting in larger access, potentially making it less complicated to scan the post space, especially for deeper depths (50). In 2019, the CEREC Company stated that it was feasible to scan deep gaps ≤20 mm using Primescan (Primescan, Dentsply Sirona), enabling the scan of long post spaces and the manufacture of custom posts using CAD/CAM technique. The longest post space length prepared was 20 mm, which is more likely to be found in canine teeth and long roots. They reported median RMS (Root Mean Square) values increased from 10 mm (357.1 μm) to 20 mm (897.5 μm) (50). The acceptable RMS range is 250–500 µm, and it is consistent with the clinically acceptable cement thickness for post-and-core restorations (51). However, this range represents a broader tolerance. The specific parameters of post space length and diameter influence accuracy, with longer and narrower post spaces generally having higher (less accurate) RMS values. Furthermore, longer posts (e.g., 10 mm) make the procedure more technique-sensitive and increase the potential for biomechanical failures (52, 53). And sometimes, a longer post can affect the fit and long-term clinical performance of the final restoration.

Furthermore, all IOSs use optical systems and possess some limitations. This is because light must reach every aspect of the post space preparation to ensure accurate scanning. Therefore, the quality of the digital impression of the post is influenced by the hardware and software parameters of the IOS, as the depth of focus and optimal ambient light conditions vary among different systems (54). Due to the study's in vitro setting, which cannot be able to accurately reflect clinical settings, this study has limitations. The efficiency of the IOS may be affected by clinical factors related to patient cooperation, illumination, saliva, oral cavity temperature, gingival mucosa and fluids, blood, limited mouth opening and accessibility, considering the precision of the fabrication of posts and cores, and the presence of adjacent teeth. Moreover, the research concentrated on a particular type of IOS, meaning its findings might not translate to other scanner varieties. Additional studies are necessary to assess how IOSs can be clinically useful for post- and core procedures across various scanner models and real-life clinical scenarios.

5 Conclusions

Within the limits of this study, the following conclusion can be reported:

  • Increasing the length of the post preparation reduced the accuracy of the scanning, regardless of the scanner used.

  • When it comes to post width, increasing the diameter of the post tends to improve the scanning accuracy.

Statements

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Ethics Committee of Ajman University (protocol code D-H-S-2023-NOV-20-14). The study was conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

DR: Resources, Funding acquisition, Visualization, Project administration, Validation, Data curation, Writing – review & editing, Supervision, Conceptualization, Writing – original draft. AA: Data curation, Methodology, Conceptualization, Validation, Supervision, Writing – review & editing, Formal analysis, Visualization, Resources, Funding acquisition, Writing – original draft, Project administration. MM: Resources, Conceptualization, Writing – original draft, Validation, Writing – review & editing, Formal analysis, Investigation, Methodology, Software, Data curation, Visualization. NH: Resources, Conceptualization, Writing – original draft, Validation, Writing – review & editing, Formal analysis, Investigation, Methodology, Software, Data curation, Visualization. GS: Visualization, Data curation, Resources, Formal analysis, Validation, Methodology, Writing – review & editing, Software, Investigation, Writing – original draft, Conceptualization. TA: Formal analysis, Writing – original draft, Methodology, Software, Data curation, Writing – review & editing, Conceptualization, Visualization, Resources, Investigation, Validation.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author(s) declared that Dinesh Rokaya is an editorial board member of Frontiers.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Correction Note

This article has been corrected with minor changes. These changes do not impact the scientific content of the article.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1.

    JodaTBräggerU. Digital vs. Conventional implant prosthetic workflows: a cost/time analysis. Clin Oral Implants Res. (2015) 26(12):14305. 10.1111/clr.12476

  • 2.

    ManganoFGandolfiALuongoGLogozzoS. Intraoral scanners in dentistry: a review of the current literature. BMC Oral Health. (2017) 17:111. 10.1186/s12903-017-0442-x

  • 3.

    AtiehMARitterAVKoC-CDuqumI. Accuracy evaluation of intraoral optical impressions: a clinical study using a reference appliance. J Prosthet Dent. (2017) 118(3):4005. 10.1016/j.prosdent.2016.10.022

  • 4.

    SainiRSAlshadidiAAFRakhraJAldosariLINHassanSABQuadriSAet alText mining analysis of scientific literature on digital intraoral scanners in dentistry: bibliometric analysis. Digit Health. (2024) 10:20552076241260837. 10.1177/20552076241260837

  • 5.

    ChochlidakisKMPapaspyridakosPGeminianiAChenC-JFengIJErcoliC. Digital versus conventional impressions for fixed prosthodontics: a systematic review and meta-analysis. J Prosthet Dent. (2016) 116(2):184190.e12. 10.1016/j.prosdent.2015.12.017

  • 6.

    WangYFanWWuYSunHZhangY. Intracanal nickel element distribution after root canal preparation using five nickel-titanium instruments and its cytotoxicity on dental pulp cells. Clin Oral Investig. (2025) 29(6):321. 10.1007/s00784-025-06412-3

  • 7.

    NedelcuROlssonPNyströmIRydénJThorA. Accuracy and precision of 3 intraoral scanners and accuracy of conventional impressions: a novel in vivo analysis method. J Dent. (2018) 69:1108. 10.1016/j.jdent.2017.12.006

  • 8.

    VágJNagyZSimonBMikoliczAKövérEMennitoAet alA novel method for complex three-dimensional evaluation of intraoral scanner accuracy. Int J Comput Dent. (2019) 22(3):23949.

  • 9.

    StefanelliLVFranchinaAPrannoAPellegrinoGFerriAPrannoNet alUse of intraoral scanners for full dental arches: could different strategies or overlapping software affect accuracy?Int J Environ Res Public Health. (2021) 18(19):9946. 10.3390/ijerph18199946

  • 10.

    MaiHYMaiH-NLeeC-HLeeK-BKimS-yLeeJ-Met alImpact of scanning strategy on the accuracy of complete-arch intraoral scans: a preliminary study on segmental scans and merge methods. J Adv Prosthodont. (2022) 14(2):8895. 10.4047/jap.2022.14.2.88

  • 11.

    GuragainMMathemaSRokayaD. Evaluation of fracture resistance and sites of failure of different dowel core restorations: an in-vitro study. Open Dent J. (2019) 13:45461. 10.2174/1874210601913010454

  • 12.

    HumagainMRokayaD. Integrating digital technologies in dentistry to enhance the clinical success. Kathmandu Univ Med J. (2019) 17(68):2567.

  • 13.

    AbouseifAMAbdelkaderSHAbdelraheemIM. Accuracy of various intraoral scanners in scanning post space preparation: an in vitro study. J Dent. (2025) 155:105651. 10.1016/j.jdent.2025.105651

  • 14.

    LeeSJGallucciGO. Digital vs. conventional implant impressions: efficiency outcomes. Clin Oral Implants Res. (2013) 24(1):1115. 10.1111/j.1600-0501.2012.02430.x

  • 15.

    AmornvitPRokayaDSanohkanS. Comparison of accuracy of current ten intraoral scanners. BioMed Res Int. (2021) 2021:2673040. 10.1155/2021/2673040

  • 16.

    ImburgiaMLogozzoSHauschildUVeronesiGManganoCManganoFG. Accuracy of four intraoral scanners in oral implantology: a comparative in vitro study. BMC oral Health. (2017) 17:113. 10.1186/s12903-017-0383-4

  • 17.

    KandutiDKoratLKosecTLegatAOvsenikMKopačI. Comparison between accuracy of posts fabricated using a digital CAD/CAM technique and a conventional direct technique. Int J Prosthodont. (2021) 34:21220. 10.11607/ijp.6502

  • 18.

    SpinaDRFda CostaRGCorrerGMRachedRN. Scanning of root canal impression for the fabrication of a resin CAD-CAM-customized post-and-core. J Prosthet Dent. (2018) 120(2):2425. 10.1016/j.prosdent.2017.08.009

  • 19.

    FarahRAlorainiASAl-Haj AliSN. Fabrication of custom post-and-core using a directly fabricated silicone pattern and digital workflow. J Prosthodont. (2020) 29(7):6315. 10.1111/jopr.13218

  • 20.

    MorganoSMBrackettSE. Foundation restorations in fixed prosthodontics: current knowledge and future needs. J Prosthet Dent. (1999) 82(6):64357. 10.1016/S0022-3913(99)70005-3

  • 21.

    PangJFengCZhuXLiuBDengTGaoYet alFracture behaviors of maxillary central incisors with flared root canals restored with CAD/CAM integrated glass fiber post-and-core. Dent Mater J. (2019) 38(1):1149. 10.4012/dmj.2017-394

  • 22.

    LibonatiADi TarantoVGallusiGMontemurroECampanellaV. CAD/CAM customized glass fiber post and core with digital intraoral impression: a case report. Clin Cosmet Investig Dent. (2020) 12:1724. 10.2147/CCIDE.S237442

  • 23.

    AL-ZomurSAbo-MadinaMHassounaM. Influence of different marginal preparation designs and materials on the marginal integrity and internal adaptation of endocrown restorations. Egyptian Dent J. (2021) 67(4):3491500. 10.21608/edj.2021.85048.1710

  • 24.

    UzunİHMalkoçMAKeleşAÖğretenAT. 3D micro-CT analysis of void formations and push-out bonding strength of resin cements used for fiber post cementation. J Adv Prosthodont. (2016) 8(2):1019. 10.4047/jap.2016.8.2.101

  • 25.

    CaceresEASampaioCSAtriaPJMouraHGianniniMCoelhoPGet alVoid and gap evaluation using microcomputed tomography of different fiber post cementation techniques. J Prosthet Dent. (2018) 119(1):1037. 10.1016/j.prosdent.2017.01.015

  • 26.

    HendiARMoharramiMSiadatHHajmiraghaHAlikhasiM. The effect of conventional, half-digital, and full-digital fabrication techniques on the retention and apical gap of post and core restorations. J Prosthet Dent. (2019) 121(2):364.e1e6. 10.1016/j.prosdent.2018.09.014

  • 27.

    SilvaNRodriguesMBicalhoAADeusRSoaresPBFSoaresCJ. Effect of magnification during post space preparation on root cleanness and fiber post bond strength. Braz Dent J. (2019) 30:4917. 10.1590/0103-6440201902662

  • 28.

    MinaminoTMineAMatsumotoMSugawaYKabetaniTHigashiMet alNondestructive observation of teeth post core-space using optical coherence tomography: comparison with microcomputed tomography and live images. J Biomed Opt. (2015) 20(10):107001. 10.1117/1.JBO.20.10.107001

  • 29.

    DamadeYKabirRGaddalaySDeshpandeSGiteSBambaleSet alRoot canal debridement efficacy of heated sodium hypochlorite in conjunction with passive ultrasonic agitation: an ex vivo study. J Dent Res Dent Clin Dent Prospects. (2020) 14(4):2358. 10.34172/joddd.2020.040

  • 30.

    NawrockaAŁukomska-SzymańskaM. Extracted human teeth and their utility in dental research. Recommendations on proper preservation: a literature review. Dent Med Probl. (2019) 56(2):18590. 10.17219/dmp/105252

  • 31.

    SchäferEKösterMBürkleinS. Percentage of gutta-percha–filled areas in canals instrumented with nickel-titanium systems and obturated with matching single cones. J Endod. (2013) 39(7):9248. 10.1016/j.joen.2013.04.001

  • 32.

    PintoAArcuriLCarosiPNardiRLibonatiAOttriaLet alIn vitro evaluation of the post-space depth Reading with an intraoral scanner (IOS) compared to a traditional silicon impression. Oral Implantol. (2017) 10(4):360. 10.11138/orl/2017.10.4.360

  • 33.

    KuttlerSMcLEANADornSFischzangA. The impact of post space preparation with Gates-Glidden drills on residual dentin thickness in distal roots of mandibular molars. J Am Dent Assoc. (2004) 135(7):9039. 10.14219/jada.archive.2004.0336

  • 34.

    GrandiniSGoracciCMonticelliFBorracchiniAFerrariM. SEM evaluation of the cement layer thickness after luting two different posts. J Adhes Dent. (2005) 7(3):235.

  • 35.

    FerracaneJL. Developing a more complete understanding of stresses produced in dental composites during polymerization. Dent Mater. (2005) 21(1):3642. 10.1016/j.dental.2004.10.004

  • 36.

    GluskinAHAhmedIHerreroDB. The aesthetic post and core: unifying radicular form and structure. Pract Proced Aesthet Dent. (2002) 14(4):31321.

  • 37.

    FerrariMVichiAMannocciFMasonPN. Retrospective study of the clinical performance of fiber posts. Am J Dent. (2000) 13:9B13B.

  • 38.

    PiraniCChersoniSFoschiFPianaGLoushineRJTayFRet alDoes hybridization of intraradicular dentin really improve fiber post retention in endodontically treated teeth?J Endod. (2005) 31(12):8914. 10.1097/01.don.0000164853.92310.e7

  • 39.

    LassilaLVTannerJLe BellA-MNarvaKVallittuPK. Flexural properties of fiber reinforced root canal posts. Dent Mater. (2004) 20(1):2936. 10.1016/S0109-5641(03)00065-4

  • 40.

    Le BellA-MLassilaLVKangasniemiIVallittuPK. Bonding of fibre-reinforced composite post to root canal dentin. J Dent. (2005) 33(7):5339. 10.1016/j.jdent.2004.11.014

  • 41.

    OulghaziIEl YamaniAMorchadB. Factors influencing vertical radicular fractures in teeth supported by metallic dental core: a scoping review. Clin Cosmet Investig Dent. (2024) 16:10114. 10.2147/CCIDE.S458697

  • 42.

    NultyAB. A comparison of full arch trueness and precision of nine intra-oral digital scanners and four lab digital scanners. Dent J. (2021) 9(7):75. 10.3390/dj9070075

  • 43.

    MichelinakisGApostolakisDTsagarakisAKourakisGPavlakisE. A comparison of accuracy of 3 intraoral scanners: a single-blinded in vitro study. J Prosthet Dent. (2020) 124(5):5818. 10.1016/j.prosdent.2019.10.023

  • 44.

    SainiRSQuadriSAKanjiMAH BinduhayyimRIAvetisyanAMosaddadSAet alAnalyzing the influence of scanning speeds and distances on digital intraoral scans: a systematic review and meta-analysis. Evid Based Dent. (2025) 26(3):148. 10.1038/s41432-025-01135-6

  • 45.

    SaraoSKBerlin-BronerYLevinL. Occurrence and risk factors of dental root perforations: a systematic review. Int Dent J. (2021) 71(2):96105. 10.1111/idj.12602

  • 46.

    ConnertTKrugREggmannFEmsermannIElAyoutiAWeigerRet alGuided endodontics versus conventional access cavity preparation: a comparative study on substance loss using 3-dimensional–printed teeth. J Endod. (2019) 45(3):32731. 10.1016/j.joen.2018.11.006

  • 47.

    LinPXuZLuoYYinL. A digital technique for a prefabricated custom post-and-core restoration. J Prosthet Dent. (2023) 130(3):27883. 10.1016/j.prosdent.2021.10.001

  • 48.

    EnderAMehlA. In-vitro evaluation of the accuracy of conventional and digital methods of obtaining full-arch dental impressions. Quintessence Int. (2015) 46(1):9.

  • 49.

    AlzahraniSJHindE-HMorganoSMElkassabyH. Evaluation of the accuracy of 2 digital intraoral scanners: a 3D analysis study. J Prosthet Dent. (2021) 126(6):78792. 10.1016/j.prosdent.2020.10.004

  • 50.

    ElterBDikerBTakÖ. The trueness of an intraoral scanner in scanning different post space depths. J Dent. (2022) 127:104352. 10.1016/j.jdent.2022.104352

  • 51.

    MeshniAAJainSOsaysiHNMHezamKNAdlanSSG. The comparison of accuracy of post space digital impressions made by three different intraoral scanners: an in vitro study. Diagnostics. (2024) 14(24):2893. 10.3390/diagnostics14242893

  • 52.

    AlshabibAAbid AlthaqafiKAlMoharibHSMirahMAlFawazYFAlgamaiahH. Dental fiber-post systems: an in-depth review of their evolution, current practice and future directions. Bioengineering. (2023) 10(5):551. 10.3390/bioengineering10050551

  • 53.

    AmarnathGSSwethaMUMuddugangadharBCSonikaRGargARaoTR. Effect of post material and length on fracture resistance of endodontically treated premolars: an in-vitro study. J Int Oral Health. (2015) 7(7):228.

  • 54.

    Thongma-EngPAmornvitPSilthampitagPRokayaDPisitanusornA. Effect of ambient lights on the accuracy of a 3-dimensional optical scanner for face scans: an in vitro study. J Healthc Eng. (2022) 2022:1. 10.1155/2022/2637078

Summary

Keywords

accuracy, CAD/CAM, digital dentistry, digital scan, intraoral scanner, post and core

Citation

Rokaya D, Al Jaghsi A, Masri MA, Hussein NM, Said GT and Akhras T (2026) Evaluating the accuracy of intraoral scanners across varied depths and widths preparation for CAD/CAM post and core. Front. Dent. Med. 7:1691514. doi: 10.3389/fdmed.2026.1691514

Received

23 August 2025

Revised

04 April 2026

Accepted

13 April 2026

Published

29 April 2026

Corrected

19 May 2026

Volume

7 - 2026

Edited by

Naseer Ahmed, Altamash Institute of Dental Medicone, Pakistan

Reviewed by

Nicolas Decerle, Université de Clermont Auvergne, France

Gerardo Pellegrino, University of Bologna, Italy

Subhabrata Maiti, Saveetha Dental College and Hospitals, India

Updates

Copyright

*Correspondence: Ahmad Al Jaghsi

† These authors have contributed equally to this work

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics