SYSTEMATIC REVIEW article

Front. Digit. Health, 29 February 2024

Sec. Health Technology Implementation

Volume 6 - 2024 | https://doi.org/10.3389/fdgth.2024.1326511

Validity of resting heart rate derived from contact-based smartphone photoplethysmography compared with electrocardiography: a scoping review and checklist for optimal acquisition and reporting

  • 1. Sport and Physical Activity Research Institute, School of Health and Life Sciences, University of the West of Scotland, Glasgow, United Kingdom

  • 2. Future Health Technologies, Singapore-ETH Centre, Campus for Research Excellence and Technological Enterprise (CREATE), Singapore, Singapore

  • 3. Saw Swee Hock School of Public Health, National University of Singapore, Singapore, Singapore

Abstract

Background:

With the rise of smartphone ownership and increasing evidence to support the suitability of smartphone usage in healthcare, the light source and smartphone camera could be utilized to perform photoplethysmography (PPG) for the assessment of vital signs, such as heart rate (HR). However, until rigorous validity assessment has been conducted, PPG will have limited use in clinical settings.

Objective:

We aimed to conduct a scoping review assessing the validity of resting heart rate (RHR) acquisition from PPG utilizing contact-based smartphone devices. Our four specific objectives of this scoping review were to (1) conduct a systematic search of the published literature concerning contact-based smartphone device-derived PPG, (2) map study characteristics and methodologies, (3) identify if methodological and technological advancements have been made, and (4) provide recommendations for the advancement of the investigative area.

Methods:

ScienceDirect, PubMed and SPORTDiscus were searched for relevant studies between January 1st, 2007, and November 6th, 2022. Filters were applied to ensure only literature written in English were included. Reference lists of included studies were manually searched for additional eligible studies.

Results:

In total 10 articles were included. Articles varied in terms of methodology including study characteristics, index measurement characteristics, criterion measurement characteristics, and experimental procedure. Additionally, there were variations in reporting details including primary outcome measure and measure of validity. However, all studies reached the same conclusion, with agreement ranging between good to very strong and correlations ranging from r = .98 to 1.

Conclusions:

Smartphone applications measuring RHR derived from contact-based smartphone PPG appear to agree with gold standard electrocardiography (ECG) in healthy subjects. However, agreement was established under highly controlled conditions. Future research could investigate their validity and consider effective approaches that transfer these methods from laboratory conditions into the “real-world”, in both healthy and clinical populations.

Introduction

Rationale

Photoplethysmography (PPG) can provide important clinical outcome measures and has been used for the diagnosis, monitoring, and screening of various diseases and disorders (). “Photoplethysmography” consists of “photo,” meaning light; “plethysmo,” meaning volume; and “graphy” meaning recording (). PPG was first suggested as a technique for measuring blood volume changes by Hertzman in 1937 (, ). PPG is a measurement of light either absorbed (transmissive photoplethysmography) or reflected (reflective photoplethysmography) by human tissues (), and is based on optical properties such as absorption, scattering and transmission (). Transmissive PPG measures light that passes through the various human tissues and is mainly used at the distal parts of the body where those tissues are thin, for example at the fingers, toes, and earlobes. Reflective PPG measures scattered light that irradiates skin tissue and produces a reduced light intensity (). While transmissive PPG exhibits more stable PPG performance () since the reflective type of signal is degraded, the latter has the advantage of a greater number of measurement sites such as the forehead, wrist, carotid artery, and esophagus, where transmissive PPG would be difficult (, ).

As such, PPG data is explained by Beer-Lambert's Law which defines resultant light intensity by the extinction coefficient, concentration, and optical path length of a medium when light passes through it (). The Beer–Lambert law can be described by:

Whereby: the transmitted light intensity (I) through a medium will decrease exponentially in irradiated light intensity (I0) in relation to the absorption coefficient (ε), where (λ) is the specific absorptivity, characteristic of the traversed tissue and dependent on the light wavelength λ, ρ is the density of the tissue, and d is the light pathlength ().

Since most of these factors are constant for a given tissue the signal quality is mainly impacted through the later part of the equation, through manipulation of λ, ρ and d, which can be modified through measurement site selection, wavelength selection and contact pressure, resulting in a reduced ε, which could explain why fingers and earlobes are preferred.

Various PPG devices have been utilized in clinical practice (). However, since the release of the first iPhone in 2007, smartphones have been widely adopted globally () and are now considered a tool with high utility, avoiding some major pitfalls of traditional data collection techniques. The traditional approach where an individual's health is monitored periodically, often by appointment, may not be an accurate representation of the possible variations in physiological measurements that occur longitudinally (, ). Moreover, smartphone technology and embedded cameras allow PPG acquisition without the need for additional, potentially costly, external devices () and could be suitable for targeting populations in traditionally underserved groups () particularly those whose demographic, geographic, or economic characteristics negatively affect health care access and delivery (, ). Therefore, telemedicine technologies are becoming more widely adopted in practice, especially since the recent COVID-19 pandemic, which highlighted the need for vital signs evaluated using telemonitoring (, , ). As a result, the proliferation of smartphone-based telemedicine appears to be here to stay and could address the United Nations Sustainable Development Goals (UN SDGs) (), in particular UN SDG 3 ().

Smartphone PPG has been previously utilized to estimate resting heart rate (RHR) through the measurement of distal pulse rate (PR) at rest, during exercise, and whilst completing mental tasks (). However, at the time of writing, there is no consensus on what metric should be used to establish the validity of smartphone-based PPG or under what conditions. Another issue is that to convert the PPG signal, a mathematical algorithm is required which not only affects smartphone performance but also validity and reliability. This is problematic given the proliferation of telemedicine, and it is therefore essential mobile health (mHealth) technologies are considered reliable and valid compared to gold standard measurements before universal adoption (). In this context, De Ridder et al. () conducted a meta-analysis of articles published between 1st January 2009 and 7th December 2016 investigating the use of smartphones to measure PR by performing PPG in comparison with a range of methods, including ECG, pulse oximetry and radial pulse. Although these methods suffer various pitfalls, comparisons with multiple validation methods could strengthen smartphone device and application validity. Results revealed good agreement between smartphone-derived (HR-PPG) and validated method-derived RHR. These authors therefore concluded that RHR obtained from a smartphone PPG signal could be used as an alternative to traditional methods, such as ECG, in an adult population, in the right context. However, De Ridder et al. () highlighted several limitations to the included studies. Firstly, there was high statistical heterogeneity between studies, ostensibly due to participant characteristics, measurement conditions, and the smartphone devices utilized (). Secondly, the latest IOS device reviewed was the iPhone 5 (released 2012) and the latest android was the Samsung Galaxy S4 (released 2013). Emerging evidence suggests advancements in technology, such as the availability of various camera positions (i.e., front-facing vs. rear-facing) and the advent of multiple lenses, could result in improvements in PPG acquisition ().

These technological enhancements are promising for the telemedicine sphere as HR-PPG could be considered a population-level biomarker, utilized for screening, surveillance, and to monitor responses to policy interventions in epidemiology and public health. Population-level biomarkers are easy to measure in the real-world, low-cost and scalable (). RHR has considerable population-level applicability and can predict adverse outcomes and the development of disease. As smartphone ownership is increasing [80% of over 65-year-olds own a smartphone in the UK (, )], and smartphone HR-PPG removes the barrier to scalability of “wearable” ownership, valid contact-based HR-PPG from a smartphone device has significant scope for public health surveillance. However, before that goal is reached, it is imperative to consider the existing literature in terms of HR-PPG validity.

Two approaches of measuring PR via PPG are known: contact and non-contact. With contact PPG, PR is measured by placing a finger on the phone rear camera, while in non-contact, imaging photoplethysmography (iPPG) is extracted from the face, without the need for direct skin contact. iPGG has some advantages over contact-based PPG, such as detecting PR in crowds and at long-distance (, ). However, in general, contact PPG exhibits better accuracy than non-contact PPG (). Considering that contact-based PPG is generally more accurate than non-contact PPG, and our group's interest in this methodology, we were interested in the validity of RHR acquisition from PPG utilizing contact-based smartphone devices.

Objectives

As a result of the importance of using validated PPG for telemedicine, and the rapidly improving technology, we aimed to conduct a scoping review assessing the validity of RHR acquisition from PPG (referred to as HR-PPG) utilizing contact-based smartphone devices against gold standard ECG (referred to as HR-ECG). Our four specific objectives of this scoping review were to 1) conduct a systematic search of the published literature concerning contact-based smartphone device-derived PPG, 2) map study characteristics and methodologies, 3) identify if methodological and technological advancements have been made, and 4) provide recommendations for the advancement of the investigative area.

Methods

Protocol and registration

The review was not preregistered, as scoping reviews are not. This review was conducted and reported in accordance with the Preferred reporting items for systematic reviews and meta-analyses extension for scoping reviews (PRISMA-ScR) guidelines ().

Eligibility criteria

Studies were included if the measurement of HR-PPG was conducted via the front or rear facing camera of a smartphone by contact-based PPG. Only studies compared with the gold standard measurement [electrocardiography (ECG)], were included. Studies were excluded if the index measurement was conducted with a device connected to a smartphone, such as a mobile sensor, medical device or wearable device; the paper did not include validity assessment of HR-PPG and HR-ECG as an outcome measurement; the study used a clinical population (we assumed healthy population unless stated otherwise); the paper was not an original article (i.e., utilized a database from a secondary source); the paper was a review; there was no abstract or full text available.

Literature search

We conducted a systematic literature search of ScienceDirect, PubMed and SPORTDiscus from January 1st, 2007, to November 6th, 2022, with the following search key: (((((“validity”) AND (“mobile”)) AND (“photoplethysmography”)) OR (“PPG”)) AND (“heart rate”)) NOT (“wearable”) AND [2007:2022(pdat)], which were developed through examination of previously published original and review articles. Filters were applied to ensure only literature written in English were included. Reference lists of included studies were manually searched for additional eligible studies.

Study selection

Studies were identified by the first author and evaluated by JDM and LDH independently and compared in an unblinded and standardized manner. Once database searches were complete, all studies were downloaded to a single reference list [utilizing Zotero software (version 6.0.26)] and duplicates were removed. First, titles and abstracts were screened for eligibility (JDM). Full text articles were then read and coded in relation to exclusion criteria, utilizing “tags” in Zotero [version 6.0.26], which was reviewed by the second author (LDH). This process involved a thorough assessment of all eligibility criteria with authors JDM and LDH confirming inclusion and exclusion. Additionally, disagreements were addressed by a third reviewer (NFS).

Data extraction

Data extracted from each study included author(s), sample size, participant sex, country of study, age, skin pigmentation, if participants were considered healthy, smartphone model, name of application utilized, whether the application was commercially available, index measurement sampling rate, camera position and resolution, flash (torch) settings, channel used for computations, ECG device utilized, electrode placement, ECG processing information, instructions given to participants, dietary control, participant posture, region of interest, breathing pattern, environmental conditions, stabilization period, duration of measurement, number of attempts or trials, primary outcome measures and measures of validity.

Outcome measures

Our primary interests were measurements of validity and mean differences between heart rate via gold standard ECG measurement (HR-ECG), and pulse rate measured by contact-based smartphone PPG (HR-PPG). Additionally, issues that arose regarding the reporting and conducting of HR-PPG validity assessment were compiled into a checklist (Table 1).

Table 1

DomainItemDescriptionTick
Target population1BMI (kg/m2)Ο
2Body height (m or cm)Ο
3Skin tone (State scale utilized and distribution)Ο
4Sample size (number of subjects)Ο
5Participant age (years)Ο
6Participant sex (n = male, n = female)Ο
7Healthy vs. clinical [if clinical report condition and medication(s)]Ο
8Sampling method (random, convenient etc.)Ο
Criterion measure9ECG manufacturer's details (model and brand)Ο
10Number of leadsΟ
11Number of channelsΟ
12Pre-measurement preparation (i.e., skin preparation procedure)Ο
13Placement according to manufacturer's details or state if otherwiseΟ
14ECG sampling frequency (Hz)Ο
15Type of electrode (make, wet vs. dry)Ο
Index measure16Device manufacturer's detailsΟ
17Application name, version and commercial availabilityΟ
18Sampling rate (Hz)Ο
19Camera(s) utilized (i.e., front- and/or rear-facing)Ο
20Camera resolution (pixels)Ο
21Torch/flash setting during measurementsΟ
22Wavelength channel used for computations (RGB)Ο
Testing conditions23Clear and concise participant instructionsΟ
24State dietary control (duration, hours prior to testing)Ο
25Report medication (including dosage)Ο
26Any physical exercise restrictions imposed (report in hours prior)Ο
27Participant posture(s)Ο
28Body region(s) measuredΟ
29Breathing pattern (spontaneous vs. metronome rate)Ο
30Environmental conditions (Environmental noise, temperature, ambient lighting conditions, indoors, outdoors, laboratory vs. free-living environment)Ο
31Stabilization period (minutes or seconds)Ο
32Duration of measurement (minutes or seconds)Ο
33Number of attemptsΟ
34Artificially induced motion artifact (MA) should describe the method used to induce the MA (i.e., shaking the device) and the frequency of the MA induced (Hz)Ο
35Define mental stress tasks (if any) (i.e., mental arithmetic/mirror tracing)Ο
36State if criterion and index measurements were simultaneousΟ
37Testing conditions reviewed with expert input (i.e., physiologist)Ο
Data processing38State PPG noise removal (motion artifact, baseline wandering and hypoperfusion) technique(s) (i.e., frequency domain filter, high-pass filter etc.)Ο
39State cut off frequencies for noise removal.Ο
40State pre-processing techniques (frequency filtering, empirical mode decomposition, wavelet transform etc.).Ο
41State method of peak detection (zero-crossing, local maxima or minima, adaptive threshold, machine learning etc.).Ο
42Report any PPG waveform reconstruction.Ο
Statistical analysis43Report correlation coefficient results utilizing guidelines proposed by Vincent (1999) or justify otherwise.Ο
44Report Post-hoc comparisons utilized.Ο
45Utilize inferential statistics for sample sizes >30 participants.Ο

Items to consider when reporting validity protocols for the acquisition of RHR via contact-based PPG, using smartphone devices.

Results

Study selection

Following initial database searches, 1,401 articles were identified, and 1,365 titles and abstracts were screened once duplicates (n = 36) were removed. These were screened for inclusion, resulting in 251 full text articles being screened. Of these 247 were excluded and four remained. A further six articles were manually identified by consulting reference lists of the included four articles, resulting in a further six articles, and therefore a total of 10 articles were included in analysis (Figure 1).

Figure 1

Study characteristics

Of the ten studies included in the review, all (100%) reported the country of study, which were upper-middle to high income countries. Sample sizes were reported in all studies (100%) (, –39) and ranged from one to 50 participants. Seven (70%) reported the number of male and female participants, of which most were male, and seven (70%) studies reported participant age (–39). Only one study (10%) reported participant skin colour (36) and five (50%) reported participants’ health status (, 34–37) (Table 2).

Table 2

ReferenceSample sizeSex (M/F)Country of studyParticipant age (mean ± SD or range)Skin colorHealthy population
Bánhalmi et al., (37)5039/11Hungary27Not reportedY
Bolkhovsky, Scully and Chon ()22Not reportedUnited StatesNot reportedNot reportedNot reported
Drijkoningen et al., ()28Not reportedBelgiumNot reportedNot reportedY
Matsumura and Yamakoshi ()127/5Japan21–24Not reportedNot reported
Matsumura et al., (34)1212/0Japan20.6 ± 0.76Not reportedY
Nam et al., (35)119/2Korea, United States and China20–40Not reportedY
Nemcova et al., (38)229/13Czech Republic18–78Not reportedNot reported
Nemcova et al., (39)126/6Czech Republic21–61Not reportedNot reported
Scully et al., ()1Not reportedUnited StatesNot reportedNot reportedNot reported
Yan et al., (36)4020/20China24.7 ± 5.2von Luschan skin color, median [IQR (Interquartile range)] Male: 23.5 (22–24), Female: 19 (18–25.75), All: 23 (19–25), P = .19.Y

General study information of investigations concerning smartphone rear-facing PPG measurement and ECG for the determination of heart rate (pulse rate) and descriptive statistics of participants.

Index measurement characteristics

Index measurement characteristics are displayed in Figure 2A and Table 3 (measurement settings) and Table 4 (device hardware specifications). Eight studies (80%) used a single smartphone for data collection (, –37, 39) and two studies (20%) utilized two or more devices (, 38). Four articles (40%) stated the name of the smartphone application (, 34, 36, 38), three (30%) of which were commercially available (, 34, 38). In the remaining seven studies (70%), commercial availability was not reported (, , , 35–37, 39). No studies (0%) reported beat detection algorithm. Eight studies (80%) reported which camera recorded smartphone PPG measurements (–39) of which the rear-facing camera was utilized for all with torch (flash) turned on. Two studies (20%) failed to report camera location and torch (flash) settings (, ). Camera resolution was reported in six studies (60%) (–35, 37, 39) and varied in resolution. In the remaining studies resolution was not reported (40%) (, , 36, 38). Smartphone sampling rate was reported in nine studies (90%) (, –35, 37–39) and one study (10%) did not report sampling rate (36). Of the nine studies that did report sampling rate six (60%) recorded at 30 Hz (, –35, 38, 39) one (10%) recorded at 30 Hz and 20 Hz, which was dependent on smartphone device (), one study (10%) recorded at 24.99 Hz () and the final study (10%) that reported sampling rate recorded in “slow-motion” capture mode at 240 Hz (37). Nine studies (90%) reported the color channel used during analysis (, –35, 37–39), and in one study (10%) it was not reported (36). Of these studies two studies (20%) utilized red, green and blue color channels (34, 38). Three studies (30%) utilized green only (, , 35). One study (10%) utilized red and green depending on smartphone used (). Two studies (20%) utilized red only (37, 39). Finally, one study (10%) converted to a single grey value ().

Figure 2

Table 3

ReferenceSmartphone model(s) utilizedName of mobile applicationCommercially available (Y/N)Sampling rateCamera(s) utilized and resolutionTorch/Flash during measurement (Y/N)Channel used for computationsPeak detection method
Bánhalmi et al., (37)iPhone 6Not reportedNot reported240 fps (Hz) (“Slow-motion” capture mode)Rear-facing (720 pixels)Y [Flash (torch) was set to “maximum”]RedIdentifying maxima points of the PPG signal
Bolkhovsky, Scully and Chon ()iPhone 4S (n = 9) and Motorola Droid (n = 13)Not reportedNot reportediPhone: 30 fps (Hz), Motorola Droid: 20 fps (Hz)Not reportedNot reportediPhone (green band) and Motorola Droid (red band)Identifying maxima points of the PPG signal
Drijkoningen et al., ()Samsung Galaxy S4Not reportedNot reported30 fps (Hz)Not reportedNot reportedConverted to a single grey value.Identifying maxima points of the PPG signal
Matsumura and Yamakoshi ()iPhone 4SiPhysioMeterY30 fps (Hz)Rear-facing (192 × 144 pixels)YGreenIdentifying maxima and minima points of the PPG signal
Matsumura et al., (34)iPhone 4SiPhysioMeterY30 fps (Hz)Rear-facing (192 × 144 pixels)YRed, green, and blueIdentifying maxima and minima points of the PPG signal combined with the detection of peaks using thresholds
Nam et al., (35)HTC One M8Not reportedNot reported30 fps (Hz) (effective frame rate 20–25 Hz)Rear-facing (176 × 144 pixels)YGreen (region closest to the flash of 176 × 72)Power spectral analysis over a 120 s sample.
Nemcova et al., (38)(12 smartphone models) The Lenovo Vibe S1 was utilized alongside the participants personal smartphone device. Other smartphones utilized include Honor 7 Lite, Apple iPhone SE, Lenovo S60, Xiaomi Redmi 3, Samsung Galaxy S4, Samsung Galaxy J5, Lenovo Vibeshot, Lenovo S750, Huawei P10, Samsung Galaxy A3 and Apple iPhone 6S.BP Data Acquisition AppY30 fps (Hz)Rear-facing (Not reported)YRed, green, and blueIdentifying maxima points of the PPG signal
Nemcova et al., (39)Xiaomi Mi9Not reportedNot reported30 fps (Hz)Rear-facing (720 × 1,280 pixels)YRedIdentifying maxima and minima points of the PPG signal
Scully et al., ()Motorola Droid RNot reportedNot reported24.99 fps (Hz)Rear-facing (720 × 480 pixels)YGreen (50 × 50-pixel average of a region)Identifying maxima and minima points of the PPG signal combined with the detection of peaks using thresholds
Yan et al., (36)iPhone 6SCardiioNot reportedNot reportedRear-facing (Not reported)YNot reportedProprietary commercially available app (Cardiio app).

Index measurement (smartphone) descriptive statistics for included studies.

fps, frames per second; Hz, hertz.

Table 4

ReferenceModelDisplayChipCameraVideo Recording
Bánhalmi et al., (37)iPhone 6Retina HD displayA8 chip with 64-bit architecture8-megapixel iSight camera with 1.5 µ pixels1080 p HD video recording (30 fps or 60 fps)
4.7-inch (diagonal) LED-backlit widescreen Multi-Touch display with IPS technology
True tone flash
Autofocus with Focus PixelsSlo-mo video (120 fps or 240 fps)
ƒ/2.2 apertureTime-lapse video with stabilization
Optical image stabilization (iPhone 6 Plus only)Cinematic video stabilization
Continuous autofocus video
Take still photos while recording video
M8 motion coprocessor
Improved face detection
True tone flash3x zoom
Five-element lensVideo geotagging
1,334-by-750-pixel resolution at 326 ppiHybrid IR filter
1,400:1 contrast ratio (typical)Backside illumination sensor
500 cd/m2 max brightness (typical)Sapphire crystal lens cover
Full sRGB standardAuto image stabilization
Dual-domain pixels for wide viewing anglesAuto HDR for photos
Face detection
Exposure control
Panorama (up to 43 megapixels)
Burst mode
Tap to focus
Fingerprint-resistant oleophobic coating on frontPhoto geotagging
Timer mode
Support for display of multiple languages and characters simultaneously
Display zoom
Reachability
Bolkhovsky, Scully and Chon ()iPhone 4SRetina display8-megapixel camera
3.5-inch (diagonal) widescreen Multi-Touch displayAutofocus
Tap to focus
Face detection in still images
LED flash
960-by-640-pixel resolution at 326 ppiVideo recording, HD (1080 p) up to 30 frames per second with audio
800:1 contrast ratio (typical)
500 cd/m2 max brightness (typical)
Fingerprint-resistant oleophobic coating on front and backVideo stabilization
Front camera with VGA-quality photos and video at up to 30 frames per second
Support for display of multiple languages and characters simultaneously
Photo and video geotagging
Bolkhovsky, Scully and Chon ()Motorola DroidInformation not availableInformation not availableInformation not availableInformation not available
Drijkoningen et al., ()Samsung Galaxy S4TechnologyCPU TypeCamera Resolution (Rear)Video Codec
FHD sAMOLEDQuadCMOS, 13MP Camera Resolution (Front)MPEG4, H.263, H.264, DivX, DivX3.11, VC-1, VP8, WMV7 / 8, Sorenson Spark, HEVC
Colour DepthCPU Speed
16M1.9GHzCMOS, 2MP Flash
SizePower LED (1EA) Auto Focus YesVideo Resolution
5″Full HD (1080 p) Video Playback
ResolutionVideo Frame rate
1,920 × 1,08030fps
Audio Codec
MP3, AAC, AAC+, eAAC+, AMR-NB / WB, OGG, FLAC, AC-3, apt-X (Bluetooth)
Matsumura and Yamakoshi ()iPhone 4SRetina display8-megapixel cameraInformation not available
3.5-inch (diagonal) widescreen Multi-Touch displayAutofocus
Tap to focus
960-by-640-pixel resolution at 326 ppiFace detection in still images
800:1 contrast ratio (typical)LED flash
500 cd/m2 max brightness (typical)Video recording, HD (1080 p) up to 30 frames per second with audio
Fingerprint-resistant oleophobic coating on front and back
Video stabilization
Front camera with VGA-quality photos and video at up to 30 frames per second
Support for display of multiple languages and characters simultaneously
Photo and video geotagging
Matsumura et al., (34)iPhone 4SRetina display8-megapixel cameraInformation not available
3.5-inch (diagonal) widescreen Multi-Touch displayAutofocus
Tap to focus
Face detection in still images
960-by-640-pixel resolution at 326 ppi
LED flash
Video recording, HD (1080 p) up to 30 frames per second with audio
800:1 contrast ratio (typical)
500 cd/m2 max brightness (typical)
Video stabilization
Front camera with VGA-quality photos and video at up to 30 frames per second
Fingerprint-resistant oleophobic coating on front and back
Photo and video geotagging
Support for display of multiple languages and characters simultaneously
Nam et al., (35)HTC One M85.0 inchQualcomm® Snapdragon™ 801, quad-core CPUsPrimary camera:Information not available
Full HD 1080 pHTC UltraPixel™ camera
BSI sensor
pixel size 2.0 um
sensor size 1/3"
ƒ/2.0
28 mm lens
HTC ImageChip 2.
1080 p Full HD video recording with HDR video
Secondary camera:
Capture depth information
Nemcova et al., (38)Lenovo Vibe S1Capacitive touchscreen,MT6752 64-bit 1.7 GHz Octa-CoreRear:Information not available
16M colors, 5-point multitouch 5.0" (1,920 × 1,080) Full HD13MP AF with dual-color flash,
PDAF, BSI sensor
IPS display @ 440 ppi
Honor 7 LiteSizeCPU ModelTriple Rear CameraInformation not available
6.5 inches, Aspect RatioQualcomm Snapdragon 480 Plus50 MP camera (f/1.8)+depth camera (f/2.4)+Macro camera (f/2.4)
20:9, Colour
16.7 million colours, TypeCPU Type
TFTLCD, ResolutionOcta-core processors
1,600*720, GesturesCPU Dominant FrequencyVideo Shooting
Multi-touch geatures, up to 10 touch points supported2*A76*2.2GHz + 6*A55*1.9GHzSupport 1080 P video shooting
GPU
Focus Mode
Up to 8x digital zoom.
Image Resolution
Support up to 4,096 × 3,072 pixels
Adreno™ 619
*The actual image resolution may vary depending on the shooting mode.
Keyboard Type
Gestures, Three-key navigation, Navigation dock
Video Resolution
Support up to 1,920 × 1,080 pixels
Features
*The actual video resolution may vary depending on the shooting mode.
Face Recognition/One-Handed mode/App Lock/App Twin
Rear Flashlight
Support
Capture Mode
Rear Camera: Aperture, Night, Portrait(including beauty mode and bokeh), Pro, Panorama, HDR, Stickers, time-lapse photography, Super macro, High-res, dual-view, story, Capture smile
iPhone SERetina HD displayA15 Bionic chip12MP Main camera4K video recording at 24 fps, 25 fps, 30 fps or 60 fps
4.7-inch (diagonal) widescreen LCD Multi-Touch display with IPS technology6-core CPU with 2 performance and 4 efficiency coresƒ/1.8 aperture
Digital zoom up to 5x1080 p HD video recording at 25 fps, 30 fps or 60 fps
4-core GPUPortrait mode with Depth Control
720 p HD video recording at 30 fps
16-core Neural EngineExtended dynamic range for video up to 30 fps
Portrait Lighting with six effectsOptical image stabilisation for video
1334 × 750-pixel resolution at 326 ppi
Digital zoom up to 3x
LED true tone flash
Optical image stabilisationQuickTake video
Slo-mo video support for 1080 p at 120 fps or 240 fps
1,400:1 contrast ratio (typical)True tone flash with slow sync
Time-lapse video with stabilisation
True tone display
Night mode Time-lapse
Cinematic video stabilisation (4K, 1080 p and 720 p)
Wide colour display (P3)Panorama (up to 63MP)
Haptic TouchSapphire crystal lens cover
Continuous autofocus video
625 nits max brightness (typical)Autofocus with Focus Pixels
Take 8MP still photos while recording 4K video
Fingerprint-resistant oleophobic coatingWide colour capture for photos and Live PhotosPlayback zoom
Video formats recorded: HEVC and H.264
Display zoomDeep Fusion
Stereo recording
ReachabilitySmart HDR 4
Photographic Styles
Advanced red-eye correction
Auto image stabilisation
Burst mode
Photo geotagging
Image formats captured: HEIF and JPEG
Lenovo S60Capacitive touchscreen,Qualcomm® Snapdragon™Rear:Information not available
16M colors,MSM8916 1.2 GHz 64-bit Quad Core13MP auto-focus, LED Flash
5-point multitouch 5.0" HD (1280 × 720)
IPS display
Xiaomi Redmi 3Information not availableInformation not availableInformation not availableInformation not available
Samsung Galaxy S4TechnologyCPU TypeCamera Resolution(Rear)Video Codec
FHD sAMOLEDQuadCMOS, 13MP Camera Resolution(Front)MPEG4, H.263, H.264, DivX, DivX3.11, VC-1, VP8, WMV7 / 8, Sorenson Spark, HEVC
Colour DepthCPU Speed
16M1.9GHzCMOS, 2MP Flash
SizePower LED (1EA) Auto Focus Yes
5"Video Resolution
ResolutionFull HD (1080 p) Video Playback
1,920 × 1,080
Video Frame rate
30fps
Audio Codec
MP3, AAC, AAC+, eAAC+, AMR-NB / WB, OGG, FLAC, AC-3, apt-X (Bluetooth)
Samsung Galaxy J5Size (Main Display)CPU SpeedMain Camera—ResolutionInformation not available
5.2” (131.8 mm)1.6GHz
Resolution (Main Display)CPU TypeCMOS 13.0 MP
720 × 1,280 (HD)Octa-CoreMain Camera—F Number
F1.7
Technology (Main Display)Main Camera—Auto Focus
Super AMOLED
Color Depth (Main Display)Yes
16MFront Camera—Resolution
S Pen Support
NoCMOS 13.0 MP
Front Camera—F Number
F1.9
Main Camera—Flash
Yes
Video Recording Resolution
FHD (1,920 × 1,080) @30fps
Lenovo VibeshotCapacitive touchscreen,64-bit Qualcomm® Snapdragon™ 615 1.7 GHzRear:Information not available
16M colors, 5-point multitouch 5.0" (1,920 × 1,080) Full HD16MP AF with true 16:9 BSI
Octa Coresensor, tricolor flash, OIS,
IR sensor, 6P lens with blue glass
IPS display @ 440 ppifilter and sapphire cover
Lenovo S750Information not availableInformation not availableInformation not availableInformation not available
Huawei P10Information not availableInformation not availableInformation not availableInformation not available
Samsung Galaxy A3Size (Main Display)CPU SpeedMain Camera—ResolutionInformation not available
4.7" (120.4 mm)1.5GHzCMOS 13.0 MP
Resolution (Main Display)CPU TypeMain Camera—F Number
720 × 1,280 (HD)Quad-Coref/1.9
Main Camera—Auto Focus
Yes
Front Camera—Resolution
CMOS 5.0 MP
Technology (Main Display)Front Camera—F Number
Super AMOLEDf/1.9
Color Depth (Main Display)Main Camera—Flash
16MYes
S Pen SupportVideo Recording Resolution
No
FHD (1,920 × 1,080) @30fps
iPhone 6SRetina HD display with 3D TouchA9 chip with 64-bit architecture12-megapixel camera4K video recording at 30 fps
Embedded M9 motion coprocessorLive Photos with stabilization
1080 p HD video recording at 30 fps or 60 fps
4.7-inch (diagonal) widescreen LCD Multi-Touch display with IPS technology720 p HD video recording at 30 fps
Autofocus with Focus Pixels
Optical image stabilization for video (iPhone 6s Plus only)
Optical image stabilization (iPhone 6s Plus only)
1,334-by-750-pixel resolution at 326 ppiTrue tone flash
Slo-mo video support for 1080 p at 120 fps and 720 p at 240 fps
True tone flash
Panorama (up to 63 megapixels)
Time-lapse video with stabilization
1,400:1 contrast ratio (typical)
Cinematic video stabilization (1080 p and 720 p)
Auto HDR for photos
Exposure control
Continuous autofocus video
Burst modeNoise reduction
Take 8-megapixel still photos while recording 4K video
500 cd/m2 max brightness (typical)Timer mode
ƒ/2.2 aperturePlayback zoom
Five-element lens3x digital zoom
5x digital zoomFace detection
Hybrid IR filterVideo geotagging
Backside illumination sensor
Full sRGB standard
Sapphire crystal lens cover
Dual-domain pixels for wide viewing anglesAuto image stabilization
Local tone mapping
Noise reduction
Face detection
Photo geotagging
Fingerprint-resistant oleophobic coating on front
Support for display of multiple languages and characters simultaneously
Display zoom
Reachability
Nemcova et al., (39)Xiaomi Mi9Information not availableInformation not availableInformation not availableInformation not available
Scully et al., ()Motorola DroidInformation not availableInformation not availableInformation not availableInformation not available
Yan et al., (36)iPhone 6SRetina HD display with 3D TouchA9 chip with 64-bit architecture12-megapixel camera4K video recording at 30 fps
Embedded M9 motion coprocessorLive Photos with stabilization
1080 p HD video recording at 30 fps or 60 fps
4.7-inch (diagonal) widescreen LCD Multi-Touch display with IPS technology
720 p HD video recording at 30 fps
Autofocus with Focus Pixels
Optical image stabilization for video (iPhone 6s Plus only)
Optical image stabilization (iPhone 6s Plus only)
1,334-by-750-pixel resolution at 326 ppiTrue tone flash
Slo-mo video support for 1080 p at 120 fps and 720 p at 240 fps
True tone flash
Panorama (up to 63 megapixels)
Time-lapse video with stabilization
1,400:1 contrast ratio (typical)Cinematic video stabilization (1080 p and 720 p)
Auto HDR for photos
500 cd/m2 max brightness (typical)Exposure controlContinuous autofocus video
Burst mode
Timer modeNoise reduction
ƒ/2.2 apertureTake 8-megapixel still photos while recording 4K video
Five-element lens
5x digital zoom
Hybrid IR filterPlayback zoom
Backside illumination sensor3x digital zoom
Full sRGB standardFace detection
Sapphire crystal lens cover
Auto image stabilization
Dual-domain pixels for wide viewing anglesLocal tone mappingVideo geotagging
Noise reduction
Fingerprint-resistant oleophobic coating on frontFace detection
Photo geotagging
Support for display of multiple languages and characters simultaneously
Display zoom
Reachability

Index device hardware technical specifications for included studies.

Criterion measurement characteristics

Criterion measurement characteristics are reported in Figure 2B and Table 5. 12-Lead ECG was used in two studies (20%) (, 36), three studies (30%) used 5-lead (, , 35), one study (10%) used 4-lead (37), two studies (20%) used 2-lead (, 34), and two studies (20%) used 1-lead (38, 39). Six studies (60%) reported ECG electrode placement and sampling frequency (, 34, 35, 37–39).

Table 5

ReferenceElectrocardiogram (ECG) utilizedElectrode placementECG Processing information
Bánhalmi et al., (37)4-lead Cardiax PC-ECG deviceFour electrodes connected to the four limbs of the participants (3 channel data)∼500 Hz
Bolkhovsky, Scully and Chon ()5-lead ECG HP 78354A systemNot reportedNot reported
Drijkoningen et al., ()12-lead ECGNot reportedNot reported
Matsumura and Yamakoshi ()2-lead ECGNot reportedNot reported
Matsumura et al., (34)2-lead ECG (Kanazawa University)Spot electrode at the wrist, left leg and body earth.All signals were sampled using an A/D converter at a rate of 1 kHz with a resolution of 16 bits, and stored digitally in a computer
Nam et al., (35)5-lead ECG HP 78354A systemStandard 5-lead configuration1,000 Hz
Nemcova et al., (38)1-lead Bittium Faros 180 ECGRecording of one bipolar lead of ECG signal from the chestSampling frequency of up to 1,000 Hz
Nemcova et al., (39)1-lead Bittium Faros 360 ECGElectrodes attached to the chest according to device manual1,000 Hz
Scully et al., ()5-lead ECG HP 78354A systemStandard 5-lead configuration400 Hz
Yan et al., (36)12-lead ECG (GE Series 2,000)Not reportedNot reported

Methodology of included studies.

Hz, hertz.

Experimental procedure characteristics

Environmental procedure characteristics are reported in Figure 2C and Table 6. Nine studies (90%) provided participant instructions (, –38) and two studies (20%) had dietary restrictions (, 34). Participant postures were stated in six studies (60%) (, –35, 37, 39). Of these studies four (40%) were measured in seated posture (–35, 37) and two studies (20%) were measured participants in two or more postures (, 39). All studies (100%) reported measurement site (, –39). Of these studies four (40%) were measured at the index finger (left) (–35), two (20%) at the index finger (right) (, ), and four (40%) at the index finger (left or right not reported) (36–39). Two studies (20%) reported breathing pattern (, 35) and participants were instructed to breathe at various metronome rates. Environmental conditions were reported in three studies (30%) (, 34, 36) and not reported in the remaining seven (70%) (, , , 35, 37–39). Stabilization period was reported in five studies (50%) (–37). Of these studies one (10%) allowed participants 10 min for stabilization (), two studies (20%) allowed participants 5 min (34, 36), one (10%) was permitted 20 s (37), and one (10%) was given an unspecified stabilization period (35). All other studies (50%) did not report stabilization period (, , , 38, 39). Measurement duration was reported in all ten studies (100%) (, –39). Finally, the number of attempts allowed per participant was reported in five studies (50%) (–35). No study reported skin temperature and one study (38) measured contact pressure using the strain gauge array under the screen (3D Touch) but this was for the estimate of BP, and the actual force values were not reported.

Table 6

ReferenceParticipant instructionsDietary controlParticipant postureRegion of interest (right or left)Breathing patternEnvironmental conditionsStabilization periodDuration of measurementNumber of attempts
Bánhalmi et al., (37)Participants were instructed to sit in a relaxed position without talking during measurementsNot reportedSeatedIndex finger (Not reported)Not reportedNot reported20 s ‘practice’ to locate rear facing camera with index finger.5 minNot reported
Bolkhovsky, Scully and Chon ()Participants were in a supine position and were instructed to place their right index finger on the camera lens of either the iPhone or Motorola smartphones. Participants were then instructed to sit up in the chair in the tilt position where measurements were repeated.Not reportedSeated and supineIndex finger (right)Not reportedNot reportedNot reportediPhone: 2 min, and Motorola: 5 min2 (1 measurement per posture)
Drijkoningen et al., ()Asked to keep the smartphone in right hand during measurement.Not reportedNot reportedIndex finger (right)Not reportedNot reportedNot reported1 minNot reported
Matsumura and Yamakoshi ()Participants were instructed to remain as still as possiblePrior to testing participants abstained from medication for 24 h, the consumption of food and caffeinated substances, and from intense physical activity for 2 hSeatedIndex finger (left)Not reportedThe experiment was conducted in a sound attenuated room, maintained at a temperature of 27–28 °C10 minBL: final 3 min of 10-minute stabilization3 (one BL, on MA, and one MT)
MA: 3-minutes
MT: 3-minutes
Measurements separated by 5 min rest interval.
Matsumura et al., (34)Participants were instructed to sit quietly with the smartphone positioned in their left hand, placed on a cushion on their knee. They were instructed to grip the smartphone firmly, despite the existence of motion artifactParticipants were asked in advance to refrain from any medication from the previous day of laboratory testing and, for 2 h before laboratory testing, to avoid consumption of food and caffeine-containing substance, intense physical activity, and smoking.SeatedIndex finger (Not reported)Not reportedSound-attenuated experimental room, maintained at a temperature of 24–26°C5 min3 min (20 s sets, 10 s rest between sets)2 sets of 3 x (20s + 10s)
Nam et al., (35)Participants were instructed to breathe at a metronome rate, in an upright seated position, whilst placing their index finger on the rear-facing camera lens of the smartphone, which was positioned on a table. The front facing camera recorded chest/abdominal movements.Not reportedSeatedIndex finger (left)Instructed to breathe at a metronome rate whilst the front facing camera recorded chest/abdominal movementsNot reportedAll subjects were acclimated to different metronome breathing rates2 min7 per subject
Nemcova et al., (38)The smartphone provided audio-visual feedback to the measured subject, indicating whether the placement of the smartphone was adequate to provide signals of sufficient quality. The participant adjusted the position of the smartphone based on the application feedback (particularly by the identification of quasi-periodic peaks/spikes seen in the signals). A flat signal or a signal with many peaks/spikes with the absence of quasi-periodicity represented a low-quality signal.Not reportedNot reportedIndex finger (Not reported)Not reportedNot reportedNot reported20 and 15 s (training and testing data, respectively)Not reported
Nemcova et al., (39)Not reportedNot reportedMeasurements 1, 2 & 3 seated. Measurement 4, walking/movingIndex finger (Not reported)Not reportedNot reportedNot reported30 sNot reported
Scully et al., ()Participants were instructed to place their left index finger over the camera lens, with the flash on, without pressing down with additional forceNot reportedNot reportedIndex finger (left)Participants inhaled with each beat of the metronome at 12, 18 and 24 beats per minute. Each metronome recording was 2-minutes in length.Not reportedNot reported2 min1 spontaneous (single subject). 3 metronome rates
Yan et al., (36)Participants were instructed to not speak and remain still during measurementsNot reportedNot reportedIndex finger (Not reported)Not reportedThe median backdrop and background light intensity during signal acquisition was 199 lux (IQR 127–249)5 min20 sNot reported

Methodology of included studies.

Primary outcome measure(s) and results of included studies

Seven studies (70%) reported mean and standard deviation for HR acquisition via criterion ECG (–36, 38) and eight studies (80%) reported mean and standard deviation for HR acquisition via smartphone PPG measurement (–36, 38, 39). Only one study (10%) reported the mean difference and its significance between criterion and index measurement (36) (Table 7).

Table 7

ReferenceDevice/conditionsNECG (Mean ± SD)PPG (Mean ± SD)Mean Diff.P-Value
Bánhalmi et al., (37)iPhone 650Not reportedNot reportedNot reportedNot reported
Bolkhovsky, Scully and Chon ()iPhone 4s supine:970.8 ± 12.270.7 ± 12.1Not reportedNot reported
iPhone 4s tilt:975.8 ± 12.0*75.8 ± 11.9*
Droid supine:1371.9 ± 7.971.7 ± 7.9
Droid tilt:1377.4 ± 6.9*77.1 ± 7.3*
Drijkoningen et al., ()Samsung Galaxy S428Not reportedNot reportedNot reportedNot reported
Matsumura and Yamakoshi ()Rest1271.0 ± 9.671.2 ± 9.8Not reportedNot reported
MA1286.7 ± 14.786.8 ± 14.6
MT1275.1 ± 12.375.4 ± 12.1
Matsumura et al., (34)HR-BL1269.8 ± 7.7Red: 69.8 ± 7.7, Green: 70.0 ± 7.8, Blue: 69.9 ± 7.9Not reportedNot reported
HR-HMA1270.4 ± 8.2Red: 70.2 ± 8.3, Green: 70.5 ± 8.3, Blue: 70.4 ± 8.6
HR-VMA1270.3 ± 9.3Red: 70.3 ± 9.1, Green: 70.4 ± 9.2, Blue: 70.5 ± 9.2
Nam et al., (35)HTC One M8 (All BR at rest)1174.9 ± 7.474.8 ± 8.0Not reportedNot reported
Nemcova et al., (38)Training dataset3083.9783.4Not reportedNot reported
Testing dataset (Lenovo Vibe S1)1071.371.8
1070.769.3
Testing dataset (Various smartphones)
Nemcova et al., (39)Xiaomi Mi912Not reportedSWT -2nd Band (3.8–7.5 Hz): 119Not reportedNot reported
SWT -3rd Band (1.9–3.8 Hz): 117
SWT -4th Band (0.94–1.9 Hz): 96
SWT -5th Band (0.47–0.94 Hz): 54
SWT -6th Band (0.23–0.47 Hz): 30
Scully et al., ()Motorola Droid R192.2 ± 5.392.3 ± 5.9Not reportedNot reported
Yan et al., (36)iPhone 6S4073.46 ± 12.7473.41 ± 12.60−0.05 (1.03)0.69

Primary outcome measure(s) and results of included studies. .

*

Represents significant difference (p < 0.05) between supine and tilt position with paired samples t-test: BL, baseline; HMA, horizontal motion artifact; VMA, vertical motion artifact; SWT, stationary wavelet transforms; Hz, hertz; MA, mental arithmetic; MT, mirror tracing.

Measures of validity

Correlations were reported in seven studies (70%) (, , 34–38), Limits of agreement (LoA) (Bland-Altman method) were calculated in seven studies (70%) (, , –37), ANOVA, Tukey HSD and geometric mean regression (GMR) were utilized concurrently in one study (10%) (34), Wilcoxon ranked sum test, were reported in two studies (20%) (36, 38), paired student t test were reported in one study (10%) (36), one study (10%) reported mean ± SD only () and one study (10%) utilized a non-numeric technical validation method of five expert reviewers (39). All studies (100%) (, , –39) reported agreement, ranging from good to very strong and correlations ranging from r = .98 to 1, between HR-PPG and HR-ECG utilizing the methods outlined above (Table 8).

Table 8

 ReferenceResultsMeasure of validitySummary of result
Bánhalmi et al., (37)HR (b/min) results:Pearson correlation, Lin. m and b represent the coefficients for the linear regression on HRV (PRV) with the corresponding mean error (err) (MSE), R2 is the coefficient of determination, and bias, SD, and BAR values are the results of the Bland-Altman analysis.Significant very strong correlation between smartphone device/application and Cardiax PC-ECG device (r = 1, P = <0.001).
Pearson correlation: 1
P value: <10−23
Lin. m: 1.00
Lin. b: −0.12
Lin. err (MSE): 0.011
Lin. R2: 1
Bias: 0.032
SD: 0.110
BAR: <0.001
Bolkhovsky, Scully and Chon ()iPhone 4s supine: Pearson correlation: >.99Pearson correlation and Bland-Altman method to calculate 95% LoA.Very strong correlation between both smartphone devices and 5-lead ECG HP 78354A system in both postures (supine, tilt) (Sig. not reported).
LoA: 0.29
iPhone 4s tilt: Pearson correlation:>.99
LoA: 0.29
Droid supine: Pearson correlation:.98
LoA: 3.20
Droid tilt: Pearson correlation: >.99
LoA: 1.40
Drijkoningen et al., ()Correlation:Correlation coefficient and Bland-Altman ratio (BAR).Significant very strong correlation between smartphone device/application and 12-lead ECG device (r = .98, P = <.001) (De Ridder et al., 2018). BAR indicated no significant changes (results not reported).
R2 (%): 95.7
P-Value: <.001
Matsumura and Yamakoshi ()Correlation:ANOVA, Tukey HSD, GMR [95% CI], and Bland-Altman method to calculate 95% LoA.Very strong correlation (r = .999) between iPhysioMeter and 2-lead ECG measured by geometric mean regressions and r = .060 measured by Bland-Altman method (Sig. not reported).
GMR = .999 [.9985, .9993],
BAP = .060 [−.131, .246]
Bias: −0.20
SD: 0.63
LoA: −1.43, 1.03
Matsumura et al., (34)Red:Mean (SD), Pearson correlation [95% CI], ANOVA, Tukey HSD, GMR [95% CI], and Bland-Altman method to calculate 95% LoA.Very strong agreement for R, G and B channels measured with iPhone 4s (r = .9960, .9991 and.9975, respectively) in comparison with 2-lead ECG. For HR, the repeated-measures ANOVA did not reveal any significant main effects of measurement, F(3, 33) = 2.39, p = 0.119, ε = 0.63, ηp2 = 0.18, and condition, F(2, 22) = 0.25, p =thinsp;0.783, ηp2 = 0.02, and measurement × condition interaction, F(6, 66) = 0.76, p = 0.502, ε = 0.40, ηp2 = 0.06.
GMR-R: 0.9960 [.9935, .9975]
BAP-R: 0.03 [−0.20, 0.26]
Bias: 0.10
SD: 0.74
LoA: −1.36, 1.56
Green:
GMR-G: 0.9991 [.9985, .9994]
BAP-G: −0.14 [−0.36, 0.10]
Bias: −0.12
SD: 0.36
LoA: −0.83, 0.58
Blue:
GMR-B: 0.9975 [.9961, .9985]
BAP-B: −0.29 [−0.48, −0.06]
Bias: −0.07
SD: 0.61
LoA: −1.27, 1.13
Nam et al., (35)Bias: 0.12Pearson correlation [95% CI], LoA, and Bland-Altman plot method to calculate 95% LoA.Bland-Altman and correlation plots, both show good agreement with non-statistically significant bias, in HR between the green color band of the HTC camera and ECG measurements.
LoA: −5.58, 5.52
Nemcova et al., (38)Training dataset:Wilcoxon rank-sum test, Pearson and Spearman correlation [95% CI].Wilcoxon p-value is higher than α = 0.05 for both datasets. The null hypothesis (H0: there is no relationship between estimated and reference values) was rejected (p < 0.05) in all cases.
Pearson R: 0.9844
Spearman P: 0.9796
Wilcoxon test: 0.8298
Pearson: 1.14E-22
Spearman: 4.71E-21
MAE: 1.3 bpm (1.61%)
MAX: 4 bpm
Testing dataset:
Pearson R: 0.9907
Spearman P: 0.9902
Wilcoxon test: 0.9136
Pearson: 4.95E-17
Spearman: 7.92E-17
MAE: 1.4 bpm (1.89%)
P < 0.05 [all values]
Nemcova et al., (39)Total 48 signals.Technical validation of waveforms from five expert annotators.31 of 48 signals were regarded “good quality” and had a HR error equal or lower than 5 bpm for each expert.
Of 48 signals, 31 were regarded “good quality” according to all annotators.
Scully et al., ()The mean ± SD was 92.2 ± 5.3 bpm for HR-ECG and 92.3 ± 5.9 bpm for HR-GREEN.Mean (SD) Diff.Authors confirmed the accuracy of HR-GREEN vs. HR-ECG (Sig. not reported).
Yan et al., (36)Pearson R:.997 (P < .001)Mean (SD) Diff., Pearson R, Wilcoxon rank-sum test, Paired student t test, Bland-Altman plots.Significant very strong correlation between smartphone device/application and 12-lead ECG device (r = .99, P = <.001).
R2 (%): 99.3
Wilcoxon test: 0.53
Paired student t test: 0.69
Bias: 0.046
LoA: −1.98, 2.07

Results for heart rate: correlations, measures of validity and summary of results.

Discussion

Principle findings

This scoping review provided an overview of existing literature regarding the acquisition and validity of HR-PPG, in healthy subjects at rest utilizing smartphone devices, with the aim of facilitating improvements in future research and clinical practice. In relation to our objective of assessing the validity of HR-PPG acquisition from PPG measurement utilizing contact-based smartphone devices, this review highlighted several methodological and reporting discrepancies between studies which can lead to different results that do not reflect outcome of comparison (). As there is currently no consensus on what metric should be used to establish the validity of smartphone-based PPG or under what conditions, the reviewed research appears to have utilized an exploratory approach. However, with the rapid development in technology and an improved understanding of this research area, we have highlighted key considerations for reporting contact-based PPG RHR acquisition with smartphones (Table 1).

Target population considerations

With regards to the general study information reported (Table 2) results revealed only one study (10%) (37) met the suggested guidelines for validating heart rate devices (albeit wearables) suggested by Mühlen et al. (40). Overall reporting was poor with small and unjustified sample sizes, and few studies adequately reported sex, skin color, or age of participants. An expert consensus suggested that studies validating HR-PPG should determine sample size based on an expected mean absolute difference, expected SD of differences and a pre-defined clinical maximum difference needed to obtain a power of 80% or 90% to assess agreement with sufficient precision (41). If no a priori level of “in agreement” is specified a sample size of 45 is recommended (42). However, all but one study had a sample size of n < 45, and therefore results could be under powered (43). We suggest sample sizes should be carefully calculated during study design utilising current guidelines (43) and these calculations should be presented in the methods section.

Comorbidities were poorly reported, particularly those that might affect pulse rate or amplitude, such as arterial stiffening or conditions affecting cardiac electrophysiology (, 44). At a minimum, studies should report either that participants were free from such health conditions, or clearly state their health conditions, if their aim is to validate HR-PPG in a particular population.

Additionally, there was inadequate reporting of participant skin color within this review. Felix von Luschan chromatic scale (VLCS) (range 1–36) was utilized, which is a validated method of skin color evaluation (45). Skin color is an important consideration during PPG acquisition as skin tone may affect the accuracy of measurements (40). However, a recent systematic review of wrist-word devices, which utilize reflective PPG, stated evidence is inconclusive possibly due to small sample sizes and the requirement for a more objective way of identifying participants' skin tone (46). Nevertheless, authors suggested HR-PPG detection may be less accurate in darker skin tones (46). Since the papers in this review failed to adequately report skin tone this cannot be corroborated with regards to camera-based methods and although Yan at al. (36) measured skin tone (Table 2), participants fell within the mid-range of the skin tone spectrum (range 19–25), with one representing light skin and 36 representing dark skin (45). Consequently, it is not clear if participants' skin tone influenced the results of the studies in the present review. Therefore, human factors such as skin color should be recorded (47) and appropriate light wavelength should be selected (48). Moreover, it is evident that more research is required investigating the effect of darker skin tones on signal quality.

Index measurement considerations

Interestingly, the majority of studies reported the use of a single smartphone device (, –37, 39). This of course maximizes internal validity of each study, but does somewhat hamper ecological validity and generalizability, given the vast options in terms of smartphones at the time of writing. Additionally, since a major advantage of mHealth technologies are their reach, it is advisable to assess the index measurements validity cross-platform, at a minimum of one phone from each. Moreover, the most recent article was Nemcova et al. published in 2021 (), suggesting future measurements could improve through the utilization of newer technology ().

As highlighted in the results, heterogeneity existed between smartphone model and application utilized and although authors reported the name of the smartphone application, zero studies reported the specific programming code utilized for beat detection. This could be due to financial, security and/or privacy reasons, as some applications were commercially available. This makes direct comparisons between apps and devices difficult as there is no guarantee two apps used the same code. Additionally, around half of the studies stated the application utilized was developed specifically for the intended research, therefore the algorithm could have been described or the code made available. Consequently, validation of specific algorithms within this review was not possible, this could be feasible in future if algorithms and build versions were explicit (49). Moreover, it is difficult to extrapolate these data to the real world without testing the efficacy of those applications outside stringent conditions of a laboratory. Identification of certain smartphones or applications which produce better PPG signals could lead to improvements in HR measures (). However, this is difficult as there is currently no consensus on what metric should be used to establish the validity of smartphone-based PPG or under what conditions, therefore protocols vary dramatically. Identifying optimal device(s) and application(s) is difficult. Therefore, we present a checklist (Table 1) to facilitate superior acquisition of HR-PPG via smartphone devices.

Although there has been a considerable increase in the number of mobile apps, many have been designed without regulation regarding development, risk mitigation, and quality control. Therefore, we advise future developers to adhere to the guidelines proposed by Llorens-Vernet and Miro (50), which consist of 36 important criteria and outline standards for mobile health-related applications. These criteria are grouped into eight categories including usability, privacy, security, appropriateness and suitability, transparency and content, safety, technical support and updates, and technology.

Most studies reported which camera recorded smartphone PPG measurements (–39) of which the rear-facing camera was utilized for all with torch (flash) turned on. However, recent research investigating rear- vs. front-facing PPG smartphone measurement revealed the front-facing camera to be more advantageous when considering greater control over the emitted light and finger detection. It is possible that previous research has not utilized this method as smartphone devices with front-facing camera capabilities are a newer technology that is still under development (). However, regardless of the camera selected it is advisable to state this as camera selection clearly influences PPG signal quality.

Over half of the studies reported camera resolution (–35, 37, 39). However, it was not clear if the reported resolution was referring to the smartphone cameras hardware settings or if the resolution was selected through the applications capture settings. Raposo et al., () suggest resolution should be set to its minimum value to reduce computational load. Moreover, implementation of interpolation techniques can be used to increase fiducial point detection through improvements in temporal resolution (51). This could influence device selection as future research could utilize devices with theoretically suboptimal resolution. For example, a device that, without adjustment of capture resolution would have high computational load, yet have other PPG performance advantages, we could then manually determine the resolution to the desired level within capture settings (i.e., reducing the capture resolution within the app), potentially improving PPG signal quality, and reducing computational load. For this reason, it is important to report what the resolution is and how it was acquired since newer devices often provide multiple rear-facing lenses, of which some have “slow-motion” technology, providing potentially enhanced sampling rate capabilities.

Smartphone sampling rate was reported in most studies (, –35, 37–39). Sampling rate can be as high as 1,000 Hz for medical equipment (52) however, for most smartphone cameras, it is typically less than 30 Hz (53), which can result in inaccurate waveform analysis (54). As outlined in our results, sampling rate was generally 20–30 Hz. For context, the latest smartphone model in the reviewed studies was the iPhone 6s (released 2015), which has a sampling rate of 30, 60 or 240 Hz, depending on the resolution settings during recording. Implications of inappropriate sampling frequency selection could result in inaccurate waveform analysis (54) and HR-PPG determination. Beres and Hejjel (51) investigated the minimum sampling frequency requirements for HR-PPG parameters in healthy individuals and concluded a minimum of 5 Hz is sufficient without interpolation, for pulse rate determination. However, although lower sampling frequencies minimize the computational load and, as a result, the power consumption consequently extending battery life (51), they can also deteriorate the accuracy of fiducial point detection in HR-PPG and/or HR-ECG, decreasing signal accuracy. Moreover, applications intending on measuring other parameters, for example those related to HRV, would require higher sampling rates with possible interpolation (51). As sampling rate is largely determined by smartphone make/model, we advise future research to utilise devices with higher sampling rate capabilities and/or implement interpolation techniques. When designing an application, it is important to consider the parameter being measured (higher sampling rates required for HRV in comparison to HR analysis) and the target demographic, as applications that are compatible with newer and older smartphone models could provide for broader scope, especially for those in low- and middle-income countries (LMIC) that may not have access to adequate healthcare.

As various wavelengths interact differently with blood and tissues (55), important consideration must be had with regards to wavelength selection (56) (i.e., red, green or blue colour channels). Emerging research suggests green wavelengths demonstrated stronger cardiac pulse signals in comparison with red or blue bands during remote PPG imaging (37). However, this was demonstrated in wrist-worn devices and more research is required in smartphone-derived PPG. Finally, improvements in pulse signal could be attained through optimization of the pixel averaging region (), whereby the video area closest to the light source is analysed increasing the overall gain of the signal and therefore improving signal quality ().

Experimental procedure considerations

Firstly, when describing the experimental procedure, studies described the technical computer science methods well. However, their relationship to physiology (i.e., what variable they are measuring and the relationship between the signal capture and the underpinning physiology) was not described in as much detail. Nearly all studies provided participant instructions (, –38), however, some study designs were hard to follow and not enough detail was provided to allow accurate replication. Studies that provided sufficient detail utilized schematic diagrams and detailed subsections within the methods as to index and criterion measurements, experimental procedure, and participant instructions.

Over half the studies reported participant postures with the seated posture being the most frequently utilized measurement position. Postural changes can result in deviations in cardiovascular measurements, such as HR (57, 58). Therefore, participant measurement posture should be reported when describing the experimental procedure. In addition to measurement posture, measurement site is also an important consideration. Hartmann et al. (59) investigated the effect of measurement site on HR-PPG waveform characteristics utilizing a reflective PPG sensor with a peak wavelength of 880 nm, comparable with reflective wavelengths utilized in smartphone devices that utilize an infrared light wavelength (880–940 nm) (). Authors determined that under normal and deep breathing conditions the finger produced the most analyzable waveforms (95% and 86% analyzable, respectively) in terms of mean amplitude, pulse peak time (Tp), dicrotic notch time (Tn), and the reflection index (RI) (all p < 0.001), which could be due to higher sensitivity to volumetric fluctuations in the cutaneous vascular walls of the finger compared with other measurement sites (59).

The application of pressure at the measurement site is something to be considered, as this is the fundamental of blood pressure measurement (i.e., an increase in pressure eventually results in occlusion). Variations in contact pressure can result in changes in several waveform characteristics (60). Increased contact pressure decreases the optical path length through the tissue, increasing AC amplitude. AC amplitude reaches its maximum when transmural pressure, defined as the difference between intraarterial pressure on the vessel wall and contact pressure, reach zero (61, 62). Additional pressure beyond this begins to occlude the vessel reducing amplitude until no signal is visible. Conversely, contact pressure applied too softly increases the optical path length through the tissue, decreasing AC amplitude. Considering this, applying enough pressure to create conditions where transmural pressure is zero could be beneficial for RHR determination, as this could make peaks more easily identifiable. While this paragraph briefly outlines the underlying physiology and AC amplitude changes form varying contact pressures, from a technical standpoint, Apple stopped incorporating the strain gauge array under the screen (3D Touch) from ∼2017 onwards. Therefore, no force measures can be obtained directly from the device. For this reason, our in-house pilot testing has suggested that providing the app user with the real-time PPG signal (i.e., visual feedback) can enhance the quality of the PPG signal. This approach has been previously conducted by Nemcova et al. (38) who reported they provided app feedback (visual peaks presented on the smartphone display) to enhance signal quality during measurement conditions. These authors stated that quality was evaluated visually by the users; quasi-periodic peaks/spikes must be seen in the signals. A flat signal or a signal with many peaks/spikes with the absence of quasi-periodicity represents a low-quality signal. The user should iteratively change the position of the smartphone according to the feedback of the application. Therefore, applying contact pressure which allows a signal which displays key pulse wave fiducial points, that has many quasi-periodic peaks would be considered ideal.

Previous research stated environmental conditions such as ambient light or motion can influence HR measurement (49, 63). In addition, careful consideration of the environmental temperature has the benefit of reducing possible HR-ECG and HR-PPG noise due to shivering (64). Of course, these environmental conditions ultimately influence participant temperature, and temperature of the measurement site (i.e., skin temperature). However, no study included in this review reported skin temperature. From a technical standpoint, the device temperature sensors are only designed for management of the CPU and battery, so measurement of environmental or skin temperature is beyond the scope of those sensors. Thus, skin temperature reporting would require an additional device such as a skin thermometer. From our in-house pilot testing, we have observed that having cold hands can reduce the quality of the PPG signal (by “quality” we mean a signal which displays key pulse wave fiducial points, that has many quasi-periodic peaks would be considered ideal). This in-house pilot testing in our lab is supported by previous work suggesting that both increased and decreased skin temperature can alter the increased PPG amplitude and total signal, PPG waveform amplitude, and PPT (60, 65–68). Research suggests ambient light may also affect light sensitive diodes; however, the size of the effect is currently unknown (40). Allen (47) suggests correct positioning of the device and the use of light modulation filtering can reduce ambient light interference.

We identified HR-ECG and HR-PPG were generally recorded simultaneously for short durations (<3 min), which is acceptable. Nemcova et al. (39) suggest ultrashort- (< 5 min) and short-term (∼ 5 min) measurements have several advantages over longer term measurements, including minimal risk of data loss during measurement, subject comfort (including flash/torch burn risk) and reduced computational demands that influence battery capacity and memory. Definitions of short- and ultra-short vary depending on the intended research, 10 s duration is commonly cited as the most appropriate duration within the literature, for HR-PPG acquisition. However, although all studies in the current review reported measurement duration (Table 6), no study compared the effect of increased or reduced measurement duration on signal quality. Therefore, the impact of measurement duration in the present review is unclear.

The time taken for a pulse wave to travel along a fixed arterial length is considered the pulse transit time (PTT). When that pulse arrives, known as pulse arrival time (PAT), it is represented by a peak in the HR-PPG signal, however due to the PTT, there is misalignment, or “time lag”, between the R wave of the HR-ECG signal and the HR-PPG peak (69). A recent review of open-source beat detection algorithms describes a method of time alignment where HR-ECG and HR-PPG derived beats within the range of <150 ms were determined to be correctly identified. The time lag between beats was manipulated by offsetting the beats in increments of 20 ms. The time lag that resulted in the most correctly identified beats (the most HR-ECG and HR-PPG beats within the range of <150 ms) was considered the “true lag” (70). Time alignment allows for direct beat comparison and ensures that not only are the same time frames are being analysed but also the same beats, improving validity assessment. However, only Bolkhovsky et al. () explicitly stated that HR-ECG and HR-PPG were aligned during post-recording data analysis.

Finally, the number of attempts allowed per participant was inadequately reported (Table 6). Holmes et al. (71) suggest number of attempts should be limited to three as additional measurements would counteract the advantages of ultrashort-term measures outlined above. We argue that there is a compromise to be made between end-user burden/acceptability and reliability/precision. Whilst it is likely that more trials per participant will increase the chances of acquiring a good signal and therefore improve validity, the more trials a user completes the greater the data entry burden (72), which could reduce usability and adherence.

Primary outcome and statistical measures of validity considerations

Results of this scoping review highlight the agreement between HR-PPG and HR-ECG (Table 8). In this scoping review CIs, LoA, or bias [from which LoA can be derived (LoA = bias ± 1.96 SD)] were not reported in all studies. Yet Mühlen et al. (40) state 95% confidence intervals (CIs) and LoAs should be provided for between-device comparisons. Interestingly, given the large number of studies reporting correlation coefficients, zero papers defined guidelines utilized to determine strength of coefficients (73, 74). We conducted post hoc interpretation and six articles (, , 34, 36–38) exceeded the minimum requirements for “high” or “strong” correlation using previously reported guidelines (73, 74). However, it was unclear whether these studies examined mean HR agreement, rather than time alignment and beat to beat agreement.

Nam et al. (35) stated PPG measured from the green wavelength (HR-Green) demonstrated “good agreement” (Table 8) in comparison with HR-ECG, however, neither the coefficient itself, nor the criteria for this qualitative assessment was provided. In summary, statistical interpretation could be improved in future research, utilizing the Bland-Altman method (75) for testing agreement between HR-ECG and HR-PPG, rather than relationship between the two (as agreement and relationship are different concepts). We also propose greater transparency in statistical reporting, including precise coefficients, a priori thresholds for interpretation (i.e., “poor”, “good”, “very good”) etc.

Effects of mobile platform (iOS and android)

It is worth noting that there are technical and practical issues related to the platform (iOS vs. Android) used to collect PPG data. A common approach is splitting the captured image into its primary colour components since red or green channels often provide a better signal. This technique largely hinges on the sensor's colour sensitivity and its colour filter array (CFA) precision. Given the stricter manufacturing control, Apple's iOS devices have a more uniform sensor technology and will likely offer a consistent baseline for PPG measurements between devices. The wider range of Android manufacturers means devices will use sensors from different producers, resulting in a broader range of sensor metrics between devices. The distinct approaches to sensor integration and image processing algorithms may further compound these differences. Apple's control over hardware and software typically results in predictable sensor performance. In contrast, Android devices might exhibit significant variability in sensor behaviour, potentially impacting homogeneity of PPG measurements across devices.

While these differences mean different phones may have different magnitudes of sensor values, the degree to which this impacts peak detection or other variables (such as frequency domain HRV) has yet to be widely investigated. More broadly, both platforms offer frame rates that enable sampling at 30 Hz. While this is suitable for peak detection, more nuanced analyses, such as wave morphology and feature identification, are challenging at this frame rate. Both manufacturers have started to include higher framerate video capture, such as “slow motion” modes with frame rates between 120 and 240 Hz. While these modes may reveal more significant detail in the collected wave data, the validity and reliability of this higher framerate regarding critical variables (e.g., the consistency of the period between frames) is not known.

There are other more practical issues regarding using phones for widescale HR monitoring. While not a significant feature of this review, in our testing (our unpublished observation) we have found that the flash on some Android phones gets uncomfortably hot when used in bulb mode (necessary to generate the PPG data). Similarly, the trend for integrating more lenses into the phone has, in some cases, moved the lenses further from the flash, resulting in less consistent lighting across the tissue in contact with the lens (our unpublished observation).

Conclusions and practical recommendations

To ensure validity and comparability with previous research, we have proposed a framework for optimal reporting (Table 1). This was based on the “Towards Intelligent Health and Well-Being: Network of Physical Activity Assessment” (INTERLIVE) best-practice recommendations (40). We took the INTERLIVE statement for wearable devices and adapted it for phone camera-based PPG. The validation process should consider six domains: the target population, criterion measure, index measure, testing conditions, data processing and the statistical analysis (40). Adherence to the checklist will result in superior acquisition of HR-PPG via smartphone devices, facilitating improvements in research and clinical practice. Future research could investigate validity with consideration towards effective approaches that transfer these methods from laboratory conditions into the “real-world”, in both healthy and clinical populations.

Statements

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Author contributions

JDM: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Project administration, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing. LDH: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Project administration, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing. JLM: Writing – review & editing. NFS: Conceptualization, Data curation, Formal Analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article.

This research was partially funded by the National Research Foundation, Prime Minister's Office, Singapore under its Campus for Research Excellence and Technological Enterprise (CREATE) program and by Future Health Technologies grant money.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1.

    AlmarshadMAIslamMSAl-AhmadiSBaHammamAS. Diagnostic features and potential applications of PPG signal in healthcare: a systematic review. Healthcare. (2022) 10:547. 10.3390/healthcare10030547

  • 2.

    AlnaebMECrabtreeVPBoutinAMikhailidisDPSeifalianAMHamiltonG. Prospective assessment of lower-extremity peripheral arterial disease in diabetic patients using a novel automated optical device. Angiology. (2007) 58:579–85. 10.1177/0003319707305685

  • 3.

    HertzmanAB. Photoelectric plethysmography of the fingers and toes in man. Exp Biol Med. (1937) 37:529–34. 10.3181/00379727-37-9630

  • 4.

    HertzmanAB. The blood supply of various skin areas as estimated by the photoelectric plethysmograph. Am J Physiol-Leg Content. (1938) 124:328–40. 10.1152/ajplegacy.1938.124.2.328

  • 5.

    ChallonerAVJRamsayCA. A photoelectric plethysmograph for the measurement of cutaneous blood flow. Phys Med Biol. (1974) 19:317–28. 10.1088/0031-9155/19/3/003

  • 6.

    ParkJSeokHSKimS-SShinH. Photoplethysmogram analysis and applications: an integrative review. Front Physiol. (2022) 12:808451. 10.3389/fphys.2021.808451

  • 7.

    LiKZhangSYangLJiangHChiZWangAet alChanges of arterial pulse waveform characteristics with gestational age during normal pregnancy. Sci Rep. (2018) 8:15571. 10.1038/s41598-018-33890-1

  • 8.

    VenemaBBlanikNBlazekVGehringHOppALeonhardtS. Advances in reflective oxygen saturation monitoring with a novel in-ear sensor system: results of a human hypoxia study. IEEE Trans Biomed Eng. (2012) 59:2003–10. 10.1109/TBME.2012.2196276

  • 9.

    WannenburgJMalekianR. Body sensor network for mobile health monitoring, a diagnosis and anticipating system. IEEE Sens J. (2015) 15:6839–52. 10.1109/JSEN.2015.2464773

  • 10.

    MayerhöferTGPahlowSPoppJ. The bouguer-beer-lambert law: shining light on the obscure. ChemPhysChem. (2020) 21:2029–46. 10.1002/cphc.202000464

  • 11.

    DrijkoningenLLenaertsFVandervoortPGrietenL. Validation of a smartphone based photoplethysmographic beat detection algorithm for normal and ectopic complexes. a smartphone based photoplethysmographic beat.

  • 12.

    GoërtzYMJBraamseAMJSpruitMAJanssenDJAEbadiZVan HerckMet alFatigue in patients with chronic disease: results from the population-based lifelines cohort study. Sci Rep. (2021) 11:20977. 10.1038/s41598-021-00337-z

  • 13.

    MairJLHayesLDCampbellAKSculthorpeN. Should we use activity tracker data from smartphones and wearables to understand population physical activity patterns?J Meas Phys Behav. (2022) 1:1–5. 10.1123/jmpb.2021-0012

  • 14.

    RaposoAda SilvaHPSanchesJ. Camera-based photoplethysmography (cbPPG) using smartphone rear and frontal cameras: an experimental study. 2021 43rd Annu Int Conf IEEE Eng Med Biol Soc EMBC. Mexico: IEEE; 2021. p.7091–4. DOI: 10.1109/EMBC46164.2021.9630847.

  • 15.

    KimBHGlanzK. Text messaging to motivate walking in older African Americans. Am J Prev Med. (2013) 44:71–5. 10.1016/j.amepre.2012.09.050

  • 16.

    BlumenthalDMortE. The Efficacy of Primary Care for Vulnerable Population Groups.

  • 17.

    WeitzTAFreundKMWrightL. Identifying and caring for underserved populations: experience of the national centers of excellence in women’s health. J Womens Health Gend Based Med. (2001) 10:937–52. 10.1089/152460901317193521

  • 18.

    HayesLDIngramJSculthorpeNF. More than 100 persistent symptoms of SARS-CoV-2 (long COVID): a scoping review. Front Med. (2021) 8. 10.3389/fmed.2021.750378

  • 19.

    McLaughlinMCerexheCMacdonaldEIngramJSanal-HayesNEMHayesLDet alA cross-sectional study of symptom prevalence, frequency, severity, and impact of long-COVID in Scotland: part I. Am J Med. (2023). 10.1016/j.amjmed.2023.07.004

  • 20.

    Goal 3: Department of Economic and Social Affairs. Available online at:https://sdgs.un.org/goals/goal3(cited Aug 19, 2023).

  • 21.

    AsiYMWilliamsC. The role of digital health in making progress toward sustainable development goal (SDG) 3 in conflict-affected populations. Int J Med Inf. (2018) 114:114–20. 10.1016/j.ijmedinf.2017.11.003

  • 22.

    VandenberkTStansJMortelmansCVan HaelstRVan SchelvergemGPelckmansCet alClinical validation of heart rate apps: mixed-methods evaluation study. JMIR MHealth UHealth. (2017) 5:e129. 10.2196/mhealth.7254

  • 23.

    De RidderBVan RompaeyBKampenJKHaineSDillesT. Smartphone apps using photoplethysmography for heart rate monitoring: meta-analysis. JMIR Cardio. (2018) 2:e4. 10.2196/cardio.8802

  • 24.

    GonzalesTIJeonJYLindsayTWestgateKPerez-PozueloIHollidgeSet alResting heart rate is a population-level biomarker of cardiorespiratory fitness: the fenland study. PloS One. (2023). 10.1371/journal.pone.0285272

  • 25.

    Smartphone ownership by age 2012-2023. Statista. Available online at:https://www.statista.com/statistics/271851/smartphone-owners-in-the-united-kingdom-uk-by-age/(cited Jan 25, 2024).

  • 26.

    SilverL. Smartphone ownership is growing rapidly around the world, but not always equally. Pew Res Cent Glob Attitudes Proj. (2019). Available online at: https://www.pewresearch.org/global/2019/02/05/smartphone-ownership-is-growing-rapidly-around-the-world-but-not-always-equally/

  • 27.

    MercuriMLoratoIRLiuY-HWieringaFHoofCVTorfsT. Vital-sign monitoring and spatial tracking of multiple people using a contactless radar-based sensor. Nat Electron. (2019) 2:252–62. 10.1038/s41928-019-0258-6

  • 28.

    AmelardRScharfenbergerCKazemzadehFPfistererKJLinBSClausiDAet alFeasibility of long-distance heart rate monitoring using transmittance photoplethysmographic imaging (PPGI). Sci Rep. (2015) 5:14637. 10.1038/srep14637

  • 29.

    CoppettiTBrauchlinAMügglerSAttinger-TollerATemplinCSchönrathFet alAccuracy of smartphone apps for heart rate measurement. Eur J Prev Cardiol. (2017) 24:1287–93. 10.1177/2047487317702044

  • 30.

    TriccoACLillieEZarinWO’BrienKKColquhounHLevacDet alPRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med. (2018) 169:467–73. 10.7326/M18-0850

  • 31.

    BolkhovskyJBScullyCGChonKH. Statistical analysis of heart rate and heart rate variability monitoring through the use of smart phone cameras. 2012 Annu Int Conf IEEE Eng Med Biol Soc. San Diego, CA: IEEE; 2012. p. 1610–3. DOI: 10.1109/EMBC.2012.6346253.

  • 32.

    ScullyCGSelvarajNRombergFWWardhanRRyanJFlorianJPet alUsing time-frequency analysis of the photoplethysmographic waveform to detect the withdrawal of 900 ml of blood. Anesth Analg. (2012) 115:74–81. 10.1213/ANE.0b013e318256486c

  • 33.

    MatsumuraKYamakoshiT. Iphysiometer: a new approach for measuring heart rate and normalized pulse volume using only a smartphone. Behav Res Methods. (2013) 45:1272–8. 10.3758/s13428-012-0312-z

  • 34.

    MatsumuraKRolfePLeeJYamakoshiT. Iphone 4s photoplethysmography: which light color yields the most accurate heart rate and normalized pulse volume using the iPhysioMeter application in the presence of motion artifact?PloS One. (2014) 9:e91205. 10.1371/journal.pone.0091205

  • 35.

    NamYKongYReyesBReljinNChonKH. Monitoring of heart and breathing rates using dual cameras on a smartphone. PloS One. (2016) 11:e0151013. 10.1016/j.ifacol.2015.10.166

  • 36.

    YanBPChanCKLiCKToOTLaiWHTseGet alResting and postexercise heart rate detection from fingertip and facial photoplethysmography using a smartphone camera: a validation study. JMIR MHealth UHealth. (2017) 5:e33. 10.2196/mhealth.7275

  • 37.

    BánhalmiABorbásJFidrichMBilickiVGinglZRudasL. Analysis of a pulse rate variability measurement using a smartphone camera. J Healthc Eng. (2018) 2018:4038034. 10.1155/2018/4038034

  • 38.

    NemcovaAJordanovaIVareckaMSmisekRMarsanovaLSmitalLet alMonitoring of heart rate, blood oxygen saturation, and blood pressure using a smartphone. Biomed Signal Process Control. (2020) 59:101928. 10.1016/j.bspc.2020.101928

  • 39.

    NemcovaAVargovaESmisekRMarsanovaLSmitalLVitekM. Brno university of technology smartphone PPG database (BUT PPG): annotated dataset for PPG quality assessment and heart rate estimation. BioMed Res Int. (2021) 2021:3453007. 10.1155/2021/3453007

  • 40.

    MühlenJMStangJLykke SkovgaardEJudicePBMolina-GarciaPJohnstonWet alRecommendations for determining the validity of consumer wearable heart rate devices: expert statement and checklist of the INTERLIVE network. Br J Sports Med. (2021) 55:767–79. 10.1136/bjsports-2020-103148

  • 41.

    LuM-JZhongW-HLiuY-XMiaoH-ZLiY-CJiM-H. Sample size for assessing agreement between two methods of measurement by bland−altman method. Int J Biostat. (2016) 12:20150039. 10.1515/ijb-2015-0039

  • 42.

    CarstensenB. Comparing Clinical Measurement Methods: A Practical Guide. Chichester: Wiley (2010).

  • 43.

    SerdarCCCihanMYücelDSerdarMA. Sample size, power and effect size revisited: simplified and practical approaches in pre-clinical, clinical and laboratory studies. Biochem Medica. (2021) 31:27–53. 10.11613/BM.2021.010502

  • 44.

    AllenJMurrayA. Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes. J Hum Hypertens. (2002) 16:711–7. 10.1038/sj.jhh.1001478

  • 45.

    TreesirichodAChansakulpornSWattanapanP. Correlation between skin color evaluation by skin color scale chart and narrowband reflectance spectrophotometer. Indian J Dermatol. (2014) 59:339–42. 10.4103/0019-5154.135476

  • 46.

    KoerberDKhanSShamsheriTKirubarajanAMehtaS. Accuracy of heart rate measurement with wrist-worn wearable devices in Various skin tones: a systematic review. J Racial Ethn Health Disparities. (2022) 10:2676–84. 10.1007/s40615-022-01446-9

  • 47.

    AllenJ. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. (2007) 28:R1–39. 10.1088/0967-3334/28/3/R01

  • 48.

    FallowBATarumiTTanakaH. Influence of skin type and wavelength on light wave reflectance. J Clin Monit Comput. (2013) 27:313–7. 10.1007/s10877-013-9436-7

  • 49.

    BruiningNCaianiEChronakiCGuzikPVan Der VeldeE. Acquisition and analysis of cardiovascular signals on smartphones: potential, pitfalls and perspectives: by the task force of the e-cardiology working group of European society of cardiology. Eur J Prev Cardiol. (2014) 21:4–13. 10.1177/2047487314552604

  • 50.

    Llorens-VernetPMiróJ. Standards for mobile health–related apps: systematic review and development of a guide. JMIR MHealth UHealth. (2020) 8:e13057. 10.2196/13057

  • 51.

    BéresSHejjelL. The minimal sampling frequency of the photoplethysmogram for accurate pulse rate variability parameters in healthy volunteers. Biomed Signal Process Control. (2021) 68:102589. 10.1016/j.bspc.2021.102589

  • 52.

    KokBECoffeyKACohnMACatalinoLIVacharkulksemsukTAlgoeSBet alHow positive emotions build physical health: perceived positive social connections account for the upward spiral between positive emotions and vagal tone. Psychol Sci. (2013) 24:1123–32. 10.1177/0956797612470827

  • 53.

    PengR-CZhouX-LLinW-HZhangY-T. Extraction of heart rate variability from smartphone photoplethysmograms. Comput Math Methods Med. (2015) 2015:516826.

  • 54.

    LiuINiSPengK. Enhancing the robustness of smartphone photoplethysmography: a signal quality Index approach. Sensors. (2020) 20:E1923. 10.3390/s20071923

  • 55.

    BiswasDSimoes-CapelaNVan HoofCVan HelleputteN. Heart rate estimation from wrist-worn photoplethysmography: a review. IEEE Sens J. (2019) 19:6560–70. 10.1109/JSEN.2019.2914166

  • 56.

    VizbaraV. Comparison of green, blue and infrared light in wrist and forehead photoplethysmography. Biomed Eng OnLine. (2013) 17.

  • 57.

    VerkruysseWSvaasandLONelsonJS. Remote plethysmographic imaging using ambient light. Opt Express. (2008) 16:21434. 10.1364/OE.16.021434

  • 58.

    WangWden BrinkerAC. Camera-based respiration monitoring. In: Wang W, Wang X, editors. Contactless Vital Signs Monitoring. Amsterdam: Elsevier (2022). p. 79–97. 10.1016/B978-0-12-822281-2.00012-3

  • 59.

    HartmannVLiuHChenFQiuQHughesSZhengD. Quantitative comparison of photoplethysmographic waveform characteristics: effect of measurement site. Front Physiol. (2019) 10:00198. 10.3389/fphys.2019.00198

  • 60.

    FineJBrananKLRodriguezAJBoonya-anantaTAjmal Ramella-RomanJCet alSources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors. (2021) 11:126. 10.3390/bios11040126

  • 61.

    YamakoshiKShimazuHShibataMKamiyaA. New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 1: model experiment. Med Biol Eng Comput. (1982) 20:307–13. 10.1007/BF02442797

  • 62.

    ForsterFKTurneyD. Oscillometric determination of diastolic, mean and systolic blood pressure–a numerical model. J Biomech Eng. (1986) 108:359–64. 10.1115/1.3138629

  • 63.

    BoulosMWheelerSTavaresCJonesR. How smartphones are changing the face of mobile and participatory healthcare: an overview, with example from eCAALYX. Biomed Eng OnLine. (2011) 10:24. 10.1186/1475-925X-10-24

  • 64.

    SattarYChhabraL. Electrocardiogram. Treasure Island (FL): StatPearls Publishing (2023). Available online at: http://www.ncbi.nlm.nih.gov/books/NBK549803/

  • 65.

    AllenJFrameJRMurrayA. Microvascular blood flow and skin temperature changes in the fingers following a deep inspiratory gasp. Physiol Meas. (2002) 23:365. 10.1088/0967-3334/23/2/312

  • 66.

    LindbergLGTamuraTÖbergPÅ. Photoplethysmography. Med Biol Eng Comput. (1991) 29:40–7. 10.1007/BF02446294

  • 67.

    KhanMPrettyCGAmiesACElliottRShawGMChaseJG. Investigating the effects of temperature on photoplethysmography. IFAC-Pap. (2015) 48:360–5. 10.1016/j.ifacol.2015.10.166

  • 68.

    ZhangX-YZhangY-T. The effect of local mild cold exposure on pulse transit time. Physiol Meas. (2006) 27:649. 10.1088/0967-3334/27/7/008

  • 69.

    ZhouZ-BCuiT-RLiDJianJ-MLiZJiS-Ret alWearable continuous blood pressure monitoring devices based on pulse wave transit time and pulse arrival time: a review. Materials (Basel). (2023) 16:2133. 10.3390/ma16062133

  • 70.

    CharltonPHKotzenKMejía-MejíaEAstonPJBudidhaKMantJet alDetecting beats in the photoplethysmogram: benchmarking open-source algorithms. Physiol Meas. (2022) 43:085007. 10.1088/1361-6579/ac826d

  • 71.

    HolmesCJFedewaMVWinchesterLJMacDonaldHVWindSAEscoMR. Validity of smartphone heart rate variability pre- and post-resistance exercise. Sensors. (2020) 20:E5738. 10.3390/s20205738

  • 72.

    MathewsSCMcSheaMJHanleyCLRavitzALabriqueABCohenAB. Digital health: a path to validation. Npj Digit Med. (2019) 2:38. 10.1038/s41746-019-0111-3

  • 73.

    VincentWJ. Statistics in kinesiology. 2nd ed.Champaign, IL: Human Kinetics Publishers (1999).

  • 74.

    FallowfieldJLHaleBJWilkinsonDM. Using Statistics in Sport and Exercise Science Research. Chichester: Lotus Publishing (2005).

  • 75.

    BlandJMAltmanDG. Statistical Methods for Assessing Agreement between Two Methods of Clinical Measurement.

Summary

Keywords

photoplethysmography, PPG, mobile, heart rate, validity

Citation

Mather JD, Hayes LD, Mair JL and Sculthorpe NF (2024) Validity of resting heart rate derived from contact-based smartphone photoplethysmography compared with electrocardiography: a scoping review and checklist for optimal acquisition and reporting. Front. Digit. Health 6:1326511. doi: 10.3389/fdgth.2024.1326511

Received

23 October 2023

Accepted

08 February 2024

Published

29 February 2024

Volume

6 - 2024

Edited by

Francesco Onorati, Takeda Pharmaceuticals, United States

Reviewed by

Kausik Basak, JIS Institute of Advanced Studies and Research, India

Xiaoman Xing, Chinese Academy of Sciences (CAS), China

Updates

Copyright

*Correspondence: Jacqueline L. Mair

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.