AUTHOR=Zhao Wanyi , Zheng Dongmei , Shi Xianbao , Xu Feng , Chen Lin , Liu Xingmei , Jin Xiaolong , Guan Qingbo , Xu Chao TITLE=Clinical performance evaluation of the SiJoy GS1 continuous glucose monitor during oral glucose tolerance testing in healthy adults JOURNAL=Frontiers in Endocrinology VOLUME=Volume 16 - 2025 YEAR=2025 URL=https://www.frontiersin.org/journals/endocrinology/articles/10.3389/fendo.2025.1536292 DOI=10.3389/fendo.2025.1536292 ISSN=1664-2392 ABSTRACT=ObjectiveThis study evaluated the performance of the SiJoy GS1 Continuous Glucose Monitor (CGM) system by analyzing the time lag between plasma glucose (PG) and CGM measurements during an Oral Glucose Tolerance Test (OGTT) in healthy adults. This investigation would elucidate the implications of physiological delay time and optimize technical delays in populations.Research design and methodsA total of 129 participants wore SiJoy GS1 sensors on their posterior upper arms for at least 48 hours before undergoing an OGTT.ResultsTo minimize the Mean Absolute Relative Difference (MARD), two approaches were tested: MARD minimization and minimum deviation match. The demographic characteristics of the participants included a mean age of 37.62 (± 11.21) years, height of 169.84 (± 7.81) cm, and weight of 71.86 (± 18.0) kg. Among them 69.0% were healthy. SiJoy GS1 sensors exhibit an excellent performance of consistency with 96.6% at 20/20% and MARD of 8.01(± 4.9) % at the fasting phase. The consensus error grid results showed 89.22% of all values fell within Zone A, and 100% of values were in Zone A+B collectively. In terms of minimizing Mean Absolute Relative Difference (MARD), at 30 minutes of OGTT, the first method suggested a 15-minute delay while the second proposed a 10-minute average delay time. The latter approach was more suitable due to the less variability in the timing of glucose peaks during the OGTT.ConclusionsIn the study, the SiJoy GS1 sensor exhibited consistent performance. Its accuracy was unaffected by subject characteristics. The application of the minimum deviation match method proved advantageous in reducing the CGM delay time.