AUTHOR=Berti Francesca , Antonini Luca , Poletti Gianluca , Fiuza Constantino , Vaughan Ted J. , Migliavacca Francesco , Petrini Lorenza , Pennati Giancarlo TITLE=How to Validate in silico Deployment of Coronary Stents: Strategies and Limitations in the Choice of Comparator JOURNAL=Frontiers in Medical Technology VOLUME=Volume 3 - 2021 YEAR=2021 URL=https://www.frontiersin.org/journals/medical-technology/articles/10.3389/fmedt.2021.702656 DOI=10.3389/fmedt.2021.702656 ISSN=2673-3129 ABSTRACT=This work aims at proposing and discussing useful indications to all those who need to validate a numerical model of coronary stent deployment. The proof of the reliability of a numerical model is becoming of paramount importance in the era of in silico trials. Recently the ASME V&V Standard Committee for medical devices prepared the V&V 40 standard document that provides a framework that guides users in establishing and assessing the relevance and adequacy of verification and validation activities performed for proving models' credibility. To the authors' knowledge, only a few examples of the application of the V&V 40 framework are available in the literature applied to medical devices, but none about stents. Specifically, in this study, the authors wish to emphasize the choice of a relevant set of experimental activities to provide data for the validation of computational models aiming to predict coronary stent deployment. Attention is focused on the use of ad hoc 3D-printed mock vessels in the validation plan, which could allow evaluating aspects of clinical relevance in a representative but controlled environment.