<?xml version="1.0" encoding="UTF-8" standalone="no"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<?covid-19-tdm?>
<article xml:lang="EN" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" article-type="research-article">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Med. Technol.</journal-id>
<journal-title>Frontiers in Medical Technology</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Med. Technol.</abbrev-journal-title>
<issn pub-type="epub">2673-3129</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3389/fmedt.2022.899328</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Medical Technology</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>Testing of pandemic ventilators under early and agile development</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Tachatos</surname> <given-names>Nikolaos</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/862975/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Steffen</surname> <given-names>Nicola</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/1759498/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Zander</surname> <given-names>Mark</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/1843894/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Stankovic</surname> <given-names>Nikola</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Meboldt</surname> <given-names>Mirko</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Erb</surname> <given-names>Thomas O.</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Hammer</surname> <given-names>J&#x000FC;rg</given-names></name>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name><surname>Schmid Daners</surname> <given-names>Marianne</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="corresp" rid="c001"><sup>&#x0002A;</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/817398/overview"/>
</contrib>
</contrib-group>
<aff id="aff1"><sup>1</sup><institution>Product Development Group Zurich, Department of Mechanical and Process Engineering, ETH Z&#x000FC;rich</institution>, <addr-line>Zurich</addr-line>, <country>Switzerland</country></aff>
<aff id="aff2"><sup>2</sup><institution>Department of Anesthesiology, University Children&#x00027;s Hospital Basel, University of Basel</institution>, <addr-line>Basel</addr-line>, <country>Switzerland</country></aff>
<aff id="aff3"><sup>3</sup><institution>Division of Respiratory and Critical Care Medicine, University Children&#x00027;s Hospital Basel, University of Basel</institution>, <addr-line>Basel</addr-line>, <country>Switzerland</country></aff>
<author-notes>
<fn fn-type="edited-by"><p>Edited by: Michael Charles Stevens, University of New South Wales, Australia</p></fn>
<fn fn-type="edited-by"><p>Reviewed by: Holger Herff, Panklinik Cologne, Germany; Luke A. White, Louisiana State University Health Shreveport, United States</p></fn>
<corresp id="c001">&#x0002A;Correspondence: Marianne Schmid Daners <email>marischm&#x00040;ethz.ch</email></corresp>
<fn fn-type="other" id="fn001"><p>This article was submitted to Diagnostic and Therapeutic Devices, a section of the journal Frontiers in Medical Technology</p></fn></author-notes>
<pub-date pub-type="epub">
<day>16</day>
<month>08</month>
<year>2022</year>
</pub-date>
<pub-date pub-type="collection">
<year>2022</year>
</pub-date>
<volume>4</volume>
<elocation-id>899328</elocation-id>
<history>
<date date-type="received">
<day>18</day>
<month>03</month>
<year>2022</year>
</date>
<date date-type="accepted">
<day>11</day>
<month>07</month>
<year>2022</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#x000A9; 2022 Tachatos, Steffen, Zander, Stankovic, Meboldt, Erb, Hammer and Schmid Daners.</copyright-statement>
<copyright-year>2022</copyright-year>
<copyright-holder>Tachatos, Steffen, Zander, Stankovic, Meboldt, Erb, Hammer and Schmid Daners</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/"><p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p></license> </permissions>
<abstract>
<p>Aiming to address clinical requirements subsequent to SARS-CoV-2-related pulmonary disease, multiple research groups and industry groups carried out intensive studies to develop pandemic ventilators (PDVs). <italic>In vitro</italic> testing to critically evaluate the specific performance of the developed apparatuses is an essential requirement. This study presents a test protocol which promotes a test-oriented, iterative, and agile assessment and consecutive development of such PDVs. It allows for fast identification of specific characteristics of each PDV in the individual test features. The test protocol includes an evaluation of the accuracy of control systems and instruments at changing parameters, the oxygen dynamics, and the response to trigger signals. The test environment is a mechanical lung, which allows reproducing various lung mechanics and to simulate active breathing cycles. A total of three PDVs that are under development were iteratively tested, with a Hamilton T1 as a reference. Continuous testing of the PDVs under development enables quick identification of critical application aspects that deserve further improved. Based on the present test protocol, the ventilators demonstrate a promising performance justifying continued development.</p></abstract>
<kwd-group>
<kwd>respiratory devices</kwd>
<kwd>standardized testing</kwd>
<kwd>test protocol</kwd>
<kwd><italic>in vitro</italic> testing</kwd>
<kwd>SARS-CoV-2</kwd>
<kwd>COVID-19</kwd>
<kwd>pandemic ventilator</kwd>
</kwd-group>
<contract-sponsor id="cn001">Fondation Botnar<named-content content-type="fundref-id">10.13039/501100011318</named-content></contract-sponsor>
<counts>
<fig-count count="8"/>
<table-count count="3"/>
<equation-count count="0"/>
<ref-count count="25"/>
<page-count count="13"/>
<word-count count="7373"/>
</counts>
</article-meta>
</front>
<body>
<sec sec-type="intro" id="s1">
<title>Introduction</title>
<p>The SARS-CoV-2 pandemic caused an immediate shortage of mechanical ventilators and trained medical personnel to treat critically ill patients suffering from respiratory failure (<xref ref-type="bibr" rid="B1">1</xref>&#x02013;<xref ref-type="bibr" rid="B3">3</xref>). Numerous research groups initiated intensive efforts to develop pandemic ventilators (PDVs) that are cost-efficient, simple to use, and fast to produce (<xref ref-type="bibr" rid="B3">3</xref>, <xref ref-type="bibr" rid="B4">4</xref>).</p>
<p>The Medicines and Healthcare Products Regulatory Agency (MHRA) of the British government published a specification of the &#x0201C;minimally clinically acceptable&#x0201D; ventilator to be used in U.K. hospitals to treat patients requiring invasive ventilation due to respiratory failure caused by SARS-CoV-2 (<xref ref-type="bibr" rid="B5">5</xref>). These specifications describe minimal requirements for the hardware, the operational modes, the usability of alarms, and the safety features of PDVs.</p>
<p>In this context, a systematic test approach seems essential to support the development and fast-track assessment of low-cost PDVs. As of today, various teams have performed <italic>in vitro</italic> tests of their PDVs (<xref ref-type="bibr" rid="B6">6</xref>&#x02013;<xref ref-type="bibr" rid="B11">11</xref>). We established a process to iteratively test the performance and features of PDVs and to provide further guidance in the early development phase. The protocol covers parts of the norm ISO 80601-2-12:2020 (<xref ref-type="bibr" rid="B12">12</xref>) and the MHRA recommendation (<xref ref-type="bibr" rid="B5">5</xref>). It is mainly designed to allow fast assessment of the different ventilator characteristics and features with an automated evaluation procedure. The protocol includes (a) a defined test procedure to evaluate the performance of the PDVs, (b) the use of a mechanical lung system that can mimic basic lung mechanics, and (c) the use of a reference commercial high-end ventilator for comparison. This study describes the methodology of an iterative test-oriented development based on this protocol. In addition, our test process is used to iteratively assess the performance of two pressure-controlled and one volume-controlled PDV which are under development.</p>
</sec>
<sec sec-type="materials and methods" id="s2">
<title>Materials and methods</title>
<sec>
<title>Ventilators</title>
<p>In this study, the protocol was applied for testing three novel PDVs and one clinical approved ventilator as reference: the GirVent (Girtec AG, N&#x000E4;nikon, Switzerland), the High Energy Ventilator (HEV, CERN, Geneva, Switzerland) (<xref ref-type="bibr" rid="B8">8</xref>), and the breathe (ETH, Zurich, Switzerland) (<xref ref-type="bibr" rid="B9">9</xref>). The three PDVs selected represent the three different most commonly used functional principles. Each PDV tested in this study is developed by the corresponding team and is still under development. The test status presented here dates from December 2020. The GirVent is a pressure-controlled ventilator based on a turbine blower with two oxygen (O<sub>2</sub>) ports: one port is realized by a direct injection of O<sub>2</sub> into the patient tubing (LP1), while the other includes a 2-L balloon as reservoir (LP2). The HEV ventilator is based on a controlled valve system with a buffer reservoir in which the O<sub>2</sub> concentration of the supplied air can be adjusted. The volume-controlled breathe ventilator (<xref ref-type="bibr" rid="B13">13</xref>) is built on the concept of the MIT E-Vent (<xref ref-type="bibr" rid="B14">14</xref>) and consists of a resuscitator bag (AmbuBag, Ambu, Ballerup, Denmark), which is periodically compressed by two paddles. The Hamilton T1 (Hamilton, Bonaduz, Switzerland) was included as a reference ventilator and was tested using the same test protocol as the PDVs.</p>
<p><xref ref-type="table" rid="T1">Table 1</xref> depicts the relevant features of the ventilators tested in this study. Each ventilator system was tested using the intended auxiliary materials supplied by the developer, that is, tubing, sensors, high-efficiency particulate air filters, and PEEP valves. The ventilation modes tested were pressure-controlled continuous mandatory ventilation (PC-CMV) and volume-controlled continuous mandatory ventilation (VC-CMV). For the Hamilton T1, the volume-targeted (S)CMV&#x0002B; and pressure-controlled PCV&#x0002B; mode was tested. In each ventilator system, the flow and pressure sensors are placed at the distal end of the patient tubing. The breathe and GirVent ventilators do not have an O<sub>2</sub> sensor and do not include a closed-loop FiO<sub>2</sub> control system. Both ventilators operate with a low-pressure O<sub>2</sub> supply in feed-forward settings where the O<sub>2</sub> flow is adjusted directly on the O<sub>2</sub> source. The HEV ventilator operates with a high-pressure O<sub>2</sub> supply and comprises a sensor that measures the O<sub>2</sub> concentration in the buffer reservoir. Hence, in the HEV, FiO<sub>2</sub> can be controlled in a closed loop similarly to the Hamilton T1. The Hamilton T1 can be operated with a low- (feed-forward) or a high-pressure (closed-loop) supply.</p>
<table-wrap position="float" id="T1">
<label>Table 1</label>
<caption><p>Overview of the basic characteristics and auxiliary material of each ventilator tested.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th valign="top" align="left"><bold>Ventilator</bold></th>
<th valign="top" align="left"><bold>Modes tested</bold></th>
<th valign="top" align="left"><bold>Ventilation mechanism</bold></th>
<th valign="top" align="left"><bold>Flow and pressure sensor</bold></th>
<th valign="top" align="left"><bold>PEEP mechanism</bold></th>
<th valign="top" align="left"><bold>O<sub>2</sub> control</bold></th>
<th valign="top" align="left"><bold>Tubing system</bold></th>
<th valign="top" align="left"><bold>Trigger detection</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Girtec GirVent</td>
<td valign="top" align="left">PC-CMV</td>
<td valign="top" align="left">Blower</td>
<td valign="top" align="left">In house development</td>
<td valign="top" align="left">Active by blower</td>
<td valign="top" align="left">Low-pressure</td>
<td valign="top" align="left">Double lumen</td>
<td valign="top" align="left">Pressure</td>
</tr>
<tr>
<td valign="top" align="left">CERN HEV</td>
<td valign="top" align="left">PC-CMV</td>
<td valign="top" align="left">Pressurized valve system</td>
<td valign="top" align="left">Hamilton flow and pressure sensor</td>
<td valign="top" align="left">Passive valve</td>
<td valign="top" align="left">High-pressure</td>
<td valign="top" align="left">Coaxial tube</td>
<td valign="top" align="left">Flow</td>
</tr>
<tr>
<td valign="top" align="left">ETH Zurich breathe</td>
<td valign="top" align="left">VC-CMV</td>
<td valign="top" align="left">Resuscitator</td>
<td valign="top" align="left">SFM3019 and MFPX5010DP</td>
<td valign="top" align="left">Passive valve</td>
<td valign="top" align="left">Low-pressure</td>
<td valign="top" align="left">Single lumen tube</td>
<td valign="top" align="left">Flow</td>
</tr>
<tr>
<td valign="top" align="left">Hamilton T1</td>
<td valign="top" align="left">PCV&#x0002B;, (S)CMV&#x0002B;</td>
<td valign="top" align="left">Blower</td>
<td valign="top" align="left">Hamilton flow and pressure sensor</td>
<td valign="top" align="left">Active by blower</td>
<td valign="top" align="left">Low- and high-pressure</td>
<td valign="top" align="left">Coaxial tube</td>
<td valign="top" align="left">Flow</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p>The O<sub>2</sub> control system is either feed-forward defined by the low-pressure O<sub>2</sub> supply setting (flow rate) or controlled in a closed loop with a high-pressure supply.</p> 
<p>PEEP, positive end-expiratory pressure; PC-CMV, pressure-controlled continuous mandatory ventilation; VC-CMV, volume-controlled continuous mandatory ventilation; PCV&#x0002B;, pressure-controlled ventilation; (S)CMV&#x0002B;, synchronized controlled mandatory ventilation.</p>
</table-wrap-foot>
</table-wrap>
</sec>
<sec>
<title>Test environment</title>
<p>Each test described in this study was performed on a mechanical lung simulator (TestChest V3, Organis GmbH, Landquart, Switzerland). In brief, the TestChest can model different lung mechanics by varying airway resistance (R<sub>aw</sub>) and respiratory system compliance (C<sub>rs</sub>). Spontaneous breathing cycles can be set by means of a negative occlusion pressure (P<sub>0.1</sub>) and the respiratory rate (RR). The TestChest includes a pressure sensor measuring the airway pressure (P<sub>aw</sub>) and an O<sub>2</sub> sensor with a response time &#x0003C;6 s for 90% of the final value placed in the airway. A flow sensor SFM3019 (Sensirion AG, St&#x000E4;fa, Switzerland) was placed at the inlet of the TestChest to measure the resultant air flow. The ventilators were connected to the TestChest <italic>via</italic> an endotracheal tube (ETT) (SafetyClear I.D. 7.5, Teleflex Medical GmbH, Fellbach, Germany), as shown in <xref ref-type="fig" rid="F1">Figure 1</xref>. For the efficiency of the O<sub>2</sub> use and the response to a change of O<sub>2</sub> supply, a medical-grade O<sub>2</sub> tank was used, which includes an adjustable flow rate valve for low-pressure supply and a high-pressure port (four bar).</p>
<fig id="F1" position="float">
<label>Figure 1</label>
<caption><p>Schematic overview of the test environment for the experiments. Each ventilator system was attached via a tubing system and an endotracheal tube to the TestChest. The TestChest includes a pressure and an O<sub>2</sub> sensor. In addition, a flow sensor SFM3019 was attached to the inlet of the TestChest to measure the actual tidal volume. A medical-grade O<sub>2</sub> bottle (four bar) was used for the test of the O<sub>2</sub> dynamics.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0001.tif"/>
</fig>
</sec>
<sec>
<title>Test protocol and experiments</title>
<p>The test protocol is divided into three categories: testing of (1) the accuracy of control systems and instruments, (<xref ref-type="bibr" rid="B2">2</xref>) the use of O<sub>2</sub> and the response to a change in O<sub>2</sub> supply, and (3) the ventilator triggering. The TestChest was calibrated with its internal program prior to the initiation of the study.</p>
<sec>
<title>Accuracy of the control systems and instruments</title>
<p>The tests of the accuracy of the control systems and instruments assess the settings of the ventilators in the pressure- and volume-controlled modes. The TestChest was set to apnea, and the patient trigger signals of each ventilator were either switched off or set to the highest possible value. Each ventilator was tested at three different peak inspiratory pressure (PIP) or tidal volume (V<sub>t</sub>) settings with successively varying parameters to identify potential discontinuities and correlations. In total, 33 different settings were tested for the accuracy of the control systems and instruments. The base settings of the mechanical lung simulate a standard male adult (body height of 175 cm and weight of 68 kg) (<xref ref-type="bibr" rid="B15">15</xref>). The detailed settings are listed in <xref ref-type="table" rid="T2">Table 2</xref>. All measurements were recorded at steady state for 60 s.</p>
<table-wrap position="float" id="T2">
<label>Table 2</label>
<caption><p>Ventilator settings for the individual tests performed.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th valign="top" align="center" colspan="5" style="border-bottom: thin solid #000000;"><bold>Ventilator settings</bold></th>
<th valign="top" align="center" colspan="2" style="border-bottom: thin solid #000000;"><bold>TestChest parameters</bold></th>
</tr>
<tr>
<th valign="top" align="left"><bold>PIP&#x0002A;(cmH<sub><bold>2</bold></sub>O)</bold></th>
<th valign="top" align="center"><bold>Vt&#x0002A;&#x0002A;(mL)</bold></th>
<th valign="top" align="center"><bold>RR (bpm)</bold></th>
<th valign="top" align="center"><bold>PEEP (cmH<sub><bold>2</bold></sub>O)</bold></th>
<th valign="top" align="center"><bold>I:E ratio</bold></th>
<th valign="top" align="center"><bold>C<sub><bold>rs</bold></sub> (mL cmH<sub><bold>2</bold></sub>O<sup><bold>&#x02212;1</bold></sup>)</bold></th>
<th valign="top" align="center"><bold>R<sub><bold>aw</bold></sub> (cmH<sub><bold>2</bold></sub>O s L<sup><bold>&#x02212;1</bold></sup>)</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center" style="background-color:#bfbfc0">10 | <bold>15</bold> | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center" style="background-color:#bfbfc0">10 | 15 | <bold>20</bold></td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center" style="background-color:#bfbfc0"><bold>5</bold> |10 | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center" style="background-color:#bfbfc0">5 |10 | <bold>15</bold></td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center" style="background-color:#bfbfc0"><bold>1:1</bold> |1:2 |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center" style="background-color:#bfbfc0">1:1 |1:2 |<bold>1:3</bold></td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center" style="background-color:#bfbfc0">15 | <bold>30</bold> | 54</td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center" style="background-color:#bfbfc0"><bold>15</bold> | 30 | 54</td>
<td valign="top" align="center"><bold>5</bold> | 20 | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center" style="background-color:#bfbfc0">5 | <bold>20</bold> | 50</td>
</tr>
<tr>
<td valign="top" align="left">20 | 25 | 30</td>
<td valign="top" align="center">350 | 400 | 450</td>
<td valign="top" align="center"><bold>10</bold> | 15 | 20</td>
<td valign="top" align="center">5 |<bold>10</bold> | 15</td>
<td valign="top" align="center">1:1 |<bold>1:2</bold> |1:3</td>
<td valign="top" align="center">15 | 30 | <bold>54</bold></td>
<td valign="top" align="center" style="background-color:#bfbfc0">5 | 20 | <bold>50</bold></td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p>The gray-shaded settings represent the parameters changed with regard to the base setting of the ventilator (RR = 10 bpm, PEEP = 10 cmH<sub>2</sub>O, I:E = 1:2) and the test lung setting (C<sub>rs</sub> = 54 mL/cmH<sub>2</sub>O, R<sub>aw</sub> = 5 cmH<sub>2</sub>O /(L/s)). Each test was performed with a PIP of 20 cmH<sub>2</sub>O, 25 cmH<sub>2</sub>O, and 30 cmH<sub>2</sub>O and a Vt of 350 mL, 400 mL, and 450 mL for the pressure-controlled (<sup>&#x0002A;</sup>) and volume-controlled modes (<sup>&#x0002A;&#x0002A;</sup>), respectively. In total, 33 different settings were tested. The bold values in the table denote the set value of the parameter in the corresponding test scenario.</p>
<p>PIP, peak inspiratory pressure; Vt, tidal volume; PEEP, positive end expiratory pressure; RR, respiratory rate; I:E, inspiratory to expiratory ratio; Crs, respiratory compliance; Raw, airway resistance.</p>
</table-wrap-foot>
</table-wrap>
</sec>
<sec>
<title>Efficiency of oxygen use and response to a change in oxygen supply</title>
<p>The O<sub>2</sub> dynamic tests are divided depending on the intended use of the ventilator either into low- or high-pressure supply. The V<sub>t</sub> of each ventilator was set to 450 mL (6&#x02013;8 mL per kilogram of body weight) (<xref ref-type="bibr" rid="B16">16</xref>), or the PIP was adjusted accordingly for the pressure-controlled ventilators. The RR was set to 10 bpm, the PEEP to 10 cmH<sub>2</sub>O, and the I:E to 1:2. For the low-pressure tests, the O<sub>2</sub> supply flow was set to 2 L min<sup>&#x02212;1</sup>, 4 L min<sup>&#x02212;1</sup>, and 6 L min<sup>&#x02212;1</sup>, respectively, and for the high-pressure tests, the desired O<sub>2</sub> concentration was set in the user interface of the ventilator (40, 60, 80, and 100%). The resulting O<sub>2</sub> concentration was measured <italic>via</italic> the integrated O<sub>2</sub> sensor of the TestChest at steady state.</p>
<p>In addition, the t<sub>90</sub> values of each ventilator were tested. It is defined as the time required for the O<sub>2</sub> supply to reach 90% of the set O<sub>2</sub> concentration measured by the O<sub>2</sub> sensor in the TestChest. The TestChest was initially ventilated with ambient air. Then, the O<sub>2</sub> supply was either set to a flow rate of 12 L min<sup>&#x02212;1</sup> (low pressure) or to an O<sub>2</sub> concentration of 100% (high pressure) in the ventilator UI.</p>
</sec>
<sec>
<title>Ventilator trigger signals</title>
<p>Spontaneous breathing efforts were simulated by the TestChest by generating negative inspiratory pressures of P<sub>0.1</sub> = 2 cmH<sub>2</sub>O/100 ms and P<sub>0.1</sub> = 4 cmH<sub>2</sub>O/100 ms, respectively, at a respiratory rate of 10 bpm. Ventilators were set to a PEEP of 10 cmH<sub>2</sub>O, an I:E of 1:2, an RR of 8 bpm, and a PIP of 20 cmH<sub>2</sub>O for the pressure-controlled ventilators and 450 mL for the volume-controlled ventilator. The trigger signal was set to the lowest sensitivity threshold without causing auto triggering (<xref ref-type="bibr" rid="B17">17</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>). Flow trigger sensitivity for the Hamilton T1 was set at 1 L min<sup>&#x02212;1</sup>, for the HEV at 0.5 L min<sup>&#x02212;1</sup>, and for the breathe at 1 L min<sup>&#x02212;1</sup>. For the GirVent, a pressure trigger sensitivity of 0.1 cmH<sub>2</sub>O was set.</p>
<p>Triggering performance is characterized by four defined parameters, as shown in <xref ref-type="fig" rid="F2">Figure 2</xref>: (1) The maximum pressure drop (PD) from the PEEP indicates the inspiratory effort required to trigger ventilation. (2) The time from the beginning of the patient breath to the pressure minimum (TPM) represents the response time of the ventilator. (3) The trigger delay time (TDT) is the time required from the beginning of the patient breath until the P<sub>aw</sub> recovers to the PEEP value. (4) The pressure time product (PTP) is the integral of the pressure curve during the TDT (<xref ref-type="bibr" rid="B17">17</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>). The pressurization capacity of a ventilator is described by the ratio of the ideal PTP (iPTP) to the actual PTP during the first 300 ms (PTP300) or 500 ms (PTP500) after the P<sub>aw</sub> exceeds the PEEP level. The ratios are denoted as PC300 and PC500, respectively.</p>
<fig id="F2" position="float">
<label>Figure 2</label>
<caption><p>Illustration of the trigger response and the pressurization capacity parameters, where PD is the pressure drop of the patient effort. TPM is the time between the start of the patient inspiratory to the pressure minimum and TDT the time when the inspiratory pressure equals the positive end expiratory pressure (PEEP) value. PTP is the pressure time capacity defined as the area enclosed by the PD and the TDT. It describes the inspiratory patient effort prior to the ventilator support. The PTP300 and PTP500 values denote the pressurization capacity of the ventilator for the first 300 ms and 500 ms of the ventilation support cycle, respectively. The corresponding ideal areas of the pressurization capacity are denoted as iPTP300 and iPTP500.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0002.tif"/>
</fig>
</sec>
</sec>
<sec>
<title>Signal processing and statistics</title>
<p>Signals were recorded <italic>via</italic> the TestChest V3 and the SFM3019 mass flow sensor at 100 Hz and processed in MATLAB R2020a (The Mathworks Inc., Natick, USA). The data of each experiment were shifted and windowed to a length of 50 s to eliminate phase shift differences between the single measurement sets, as shown in <xref ref-type="table" rid="T2">Table 2</xref>. The signal of P<sub>aw</sub> and the flow measured in the TestChest (Q<sub>TestChest</sub>) were processed using a Gaussian filter with, a window length of 200 ms (20 samples). The RR was calculated using the time difference between two onsets of the inspiratory flow. The onset of the inspiratory flow starts when Q<sub>TestChest</sub> reached a specified threshold. This threshold was defined per ventilator as the most sensitive value at which the fluctuation of Q<sub>TestChest</sub> does not cause an onset detection. The inspiratory time (TI) and expiratory time (TE) are defined as the time during which Q<sub>TestChest</sub> is above the positive threshold (inspiratory flow) and below the negative threshold (expiratory flow), respectively.</p>
<p>For each test performed, P<sub>aw</sub> was measured inside the TestChest (<xref ref-type="fig" rid="F1">Figure 1</xref>); hence, the PIP and PEEP displayed in the results report the pressure recorded downstream to the ETT. The PIP was detected as the peak value of P<sub>aw</sub> averaged over 40 ms (four samples) symmetrically around the detected point signal. The PEEP value was determined as the minimum value of P<sub>aw</sub> in the time interval of two consecutive breaths that was closer to the inspiration of the latter breath. The V<sub>t</sub> was calculated by integrating the flow sensor signal. I:E is defined as the ratio of inspiratory time (T<sub>I</sub>) to expiratory time (T<sub>E</sub>) (T<sub>I</sub>:T<sub>E</sub>).</p>
<p>A Gaussian filter with a window length of 30 s (3,000 samples) was applied to the FiO<sub>2</sub> signal recorded by the TestChest. The filtered signal averages the FiO<sub>2</sub> values, while the unfiltered signal shows the FiO<sub>2</sub> fluctuations during each breath cycle. The starting time of the t<sub>90</sub> test was set when the unfiltered FiO<sub>2</sub> signal exceeded an FiO<sub>2</sub> of 22%. Maximum O<sub>2</sub> step response and the 90% O<sub>2</sub> step response were determined using the filtered signal.</p>
<p>Parameters for the trigger signals (PD, TPM, TDT, PTP) were evaluated from the P<sub>aw</sub> signal. The start of inspiration is defined when exceeding 10% of the maximum gradient recorded during the pressure drop in the inspiratory phase.</p>
</sec>
</sec>
<sec sec-type="results" id="s3">
<title>Results</title>
<sec>
<title>Accuracy of the control systems and instruments</title>
<p>The overall relative errors of individual parameters recorded over all experiments (<xref ref-type="table" rid="T2">Table 2</xref>) are depicted in <xref ref-type="fig" rid="F3">Figure 3</xref>. The relative error was calculated as the ratio of the measured error to the set value with the measured error being the difference between the value measured and the set value on the ventilator. <xref ref-type="table" rid="T3">Table 3</xref> lists all the measured mean values, including standard deviation (SD). The PDVs showed generally larger relative mean errors and SD than the Hamilton T1 (PCV&#x0002B; mode). In addition, more pronounced outliers were observed at individual measurement settings. Only the V<sub>t</sub> measured in the breathe and the RR in all PDVs were in a range similar to those of the Hamilton T1, with no parameter changes having a substantial effect on the accuracy. The two passive PEEP valves used in the HEV and the breathe showed a larger variation than the actively applied PEEP in the GirVent.</p><fig id="F3" position="float">
<label>Figure 3</label>
<caption><p>Boxplot representation of the relative error over all means of the individual measurement sets listed in <xref ref-type="table" rid="T3">Table 3</xref>. PIP, V<sub>t</sub>, RR, PEEP, and I:E are shown at varying RR, PEEP, I:E, C<sub>rs</sub>, and R<sub>aw</sub> for the pressure- and volume-controlled ventilators. <bold>(A)</bold> Hamilton T1 (PCV&#x0002B;), <bold>(B)</bold> HEV, <bold>(C)</bold> GirVent, <bold>(D)</bold> Hamilton T1 ((S)CMV&#x0002B;), and <bold>(E)</bold> breathe. Each point represents the mean value of one experiment.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0003.tif"/>
</fig>
<table-wrap position="float" id="T3">
<label>Table 3</label>
<caption><p>Mean and standard deviation (SD) of the relative error over all experiments and respiratory cycles for RR, PEEP, I:E, and PIP or Vt for the pressure-controlled and volume-controlled ventilators, respectively.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th/>
<th valign="top" align="center" colspan="3" style="border-bottom: thin solid #000000;"><bold>Pressure-controlled mode</bold></th>
<th valign="top" align="center" colspan="2" style="border-bottom: thin solid #000000;"><bold>Volume-controlled mode</bold></th>
</tr>
<tr>
<th valign="top" align="left"><bold>Ventilator /parameter</bold></th>
<th valign="top" align="center" style="border-bottom: thin solid #000000;"><bold>Hamilton T1 (PCV&#x0002B;)</bold></th>
<th valign="top" align="center" style="border-bottom: thin solid #000000;"><bold>GirVent</bold></th>
<th valign="top" align="center" style="border-bottom: thin solid #000000;"><bold>HEV CERN</bold></th>
<th valign="top" align="center" style="border-bottom: thin solid #000000;"><bold>Hamilton T1 ((S)CMV&#x0002B;)</bold></th>
<th valign="top" align="center" style="border-bottom: thin solid #000000;"><bold>breathe</bold></th>
</tr>
<tr>
<th/>
<th valign="top" align="center"><bold>Mean &#x000B1;SD (%)</bold></th>
<th valign="top" align="center"><bold>Mean &#x000B1;SD (%)</bold></th>
<th valign="top" align="center"><bold>Mean &#x000B1;SD (%)</bold></th>
<th valign="top" align="center"><bold>Mean &#x000B1;SD (%)</bold></th>
<th valign="top" align="center"><bold>Mean &#x000B1;SD (%)</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">PIP (cmH<sub>2</sub>O)</td>
<td valign="top" align="center">0.26, 1.12</td>
<td valign="top" align="center">&#x02212;1.89, 4.08</td>
<td valign="top" align="center">&#x02212;2.19, 3.94</td>
<td valign="top" align="center">&#x02013;</td>
<td valign="top" align="center">&#x02013;</td>
</tr>
<tr>
<td valign="top" align="left">V<sub>t</sub> (mL)</td>
<td valign="top" align="center">&#x02013;</td>
<td valign="top" align="center">&#x02013;</td>
<td valign="top" align="center">&#x02013;</td>
<td valign="top" align="center">&#x02212;4.72, 1.63</td>
<td valign="top" align="center">2.27, 0.89</td>
</tr>
<tr>
<td valign="top" align="left">RR (bpm)</td>
<td valign="top" align="center">0.02, 0.08</td>
<td valign="top" align="center">&#x02212;2.84, 1.26</td>
<td valign="top" align="center">0.01, 0.09</td>
<td valign="top" align="center">&#x02212;0.02, 0.17</td>
<td valign="top" align="center">&#x02212;0.89, 0.32</td>
</tr>
<tr>
<td valign="top" align="left">PEEP (cmH<sub>2</sub>O)</td>
<td valign="top" align="center">&#x02212;1.15, 1.16</td>
<td valign="top" align="center">&#x02212;6.34, 1.33</td>
<td valign="top" align="center">&#x02212;19.51, 3.04</td>
<td valign="top" align="center">&#x02212;0.25, 1.44</td>
<td valign="top" align="center">&#x02212;0.73, 4.00</td>
</tr>
<tr>
<td valign="top" align="left">I:E (-)</td>
<td valign="top" align="center">1.42, 1.05</td>
<td valign="top" align="center">&#x02212;8.28, 2.91</td>
<td valign="top" align="center">2.55, 1.91</td>
<td valign="top" align="center">1.64, 0.93</td>
<td valign="top" align="center">5.76, 4.20</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p>PCV&#x0002B;, pressure-controlled ventilation; (S)CMV&#x0002B;, synchronized controlled mandatory ventilation.</p>
</table-wrap-foot>
</table-wrap>
<p><xref ref-type="fig" rid="F4">Figures 4</xref>, <xref ref-type="fig" rid="F5">5</xref> show the effects of the varying parameters on the PIP and the V<sub>t</sub>. An increase in the RR, a decrease in the inspiratory time, and an increase in R<sub>aw</sub> showed the largest influence on the accuracy of the PIP. Moreover, the largest PIP error in each pressure-controlled ventilator was observed in the same setting of a PIP of 30 cmH<sub>2</sub>O and an RR of 20 bpm. By contrast, in the volume-controlled mode, only increasing the RR (breathe) or increasing R<sub>aw</sub> (Hamilton T1 (S)CMV&#x0002B; mode) had a prominent influence on the V<sub>t</sub> error. Other parameter changes had only a marginal impact on the accuracy of the parameters tested. Detailed plots and data of each measurement and parameter change, as well as the flow profiles, recorded for the base setting for each ventilator are provided in the <xref ref-type="supplementary-material" rid="SM1">Supplementary material</xref>.</p>
<fig id="F4" position="float">
<label>Figure 4</label>
<caption><p>Measured PIP error in the Hamilton T1 (PCV&#x0002B; mode), the HEV, and the GirVent: effects of altering PIP (cmH<sub>2</sub>O), RR (bpm), PEEP (cmH<sub>2</sub>O), I:E (ratio), Crs (mL cmH<sub>2</sub>O<sup>&#x02212;1</sup>), and Raw (cmH<sub>2</sub>O s L<sup>&#x02212;1</sup>). The base setting is indicated by <sup>&#x0002A;</sup>. The markers and lines denote the mean error and the maximum and minimum errors, respectively. The grayed-out area indicates the acceptable performance tolerance according to the ISO 80601-2-12 norm and the MHRA specification.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0004.tif"/>
</fig>
<fig id="F5" position="float">
<label>Figure 5</label>
<caption><p>Measured Vt error in the Hamilton T1 ((S)CMV&#x0002B; mode) and the breathe: effects of altering Vt (mL), RR (bpm), PEEP (cmH<sub>2</sub>O), I:E (ratio), Crs (mL cmH<sub>2</sub>O<sup>&#x02212;1</sup>), and Raw (cmH<sub>2</sub>O s L<sup>&#x02212;1</sup>). The base setting is indicated by <sup>&#x0002A;</sup>. The markers and lines denote the mean error and the maximum and minimum errors, respectively. The grayed-out area indicates the acceptable performance tolerance according to the ISO 80601-2-12 norm and the MHRA specification.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0005.tif"/>
</fig>
<p>The ISO 80601-2-12 norm and the MHRA specify an acceptable performance tolerance of &#x000B1; <italic>(2</italic> &#x0002B; <italic>(4 % of the actual reading)) cmH</italic><sub>2</sub><italic>O</italic> for the pressure value and &#x000B1; <italic>(4,0</italic> &#x0002B; <italic>(15% of the actual volume expired through the patient connection port)) mL</italic>. The acceptable tolerance is given in <xref ref-type="fig" rid="F4">Figures 4</xref>, <xref ref-type="fig" rid="F5">5</xref>. For the respiratory rate and I:E, no performance tolerance was stated in the norm. Therefore, for the tests performed in this study, we defined a tolerance of 5 and 10%, respectively.</p>
</sec>
<sec>
<title>Oxygen dynamics</title>
<p>The achieved fractions of inspired O<sub>2</sub> (FiO<sub>2</sub>) at three O<sub>2</sub> flow rates (2 L min<sup>&#x02212;1</sup>, 4 L min<sup>&#x02212;1</sup>, and 6 L min<sup>&#x02212;1</sup>) for the low- and high-pressure supply (four bar) of each ventilator are presented in <xref ref-type="fig" rid="F6">Figure 6</xref>. Fluctuations of FiO<sub>2</sub> were observed in each ventilator and are indicated by error bars (minimum and maximum values). Generally, since the results are all relative values obtained by using the O<sub>2</sub> sensor in steady state and the tests were all performed with the same O<sub>2</sub> sensor, the delay of the sensor was neglectable.</p>
<fig id="F6" position="float">
<label>Figure 6</label>
<caption><p><bold>(A)</bold> Mean FiO<sub>2</sub> at three O<sub>2</sub> flow rates (2 L min<sup>&#x02212;1</sup>, 4 L min<sup>&#x02212;1</sup>, and 6 L min<sup>&#x02212;1</sup>) for the ventilators with low-pressure O<sub>2</sub> supply, depicted over the last 30 s of steady state. <bold>(B)</bold> Measured &#x00394;FiO<sub>2</sub> in the Hamilton T1 and HEV (high-pressure port) to the set FiO<sub>2</sub> concentrations in the UI (0.40, 0.60, 0.80 and 1.00). The measured &#x00394;FiO<sub>2</sub> with an FiO<sub>2</sub> setting of 1.00 in the UI is indicated by <sup>&#x0002A;</sup>. The error bars indicate the fluctuation (minimum and maximum) observed at steady state. <bold>(C)</bold> Achieved FiO<sub>2</sub>, and t<sub>90</sub> values at an O<sub>2</sub> flow rate of 12 L/min (low pressure) or at a set O<sub>2</sub> concentration of 100% (high pressure) of each ventilator.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0006.tif"/>
</fig>
<p>The achieved mean FiO<sub>2</sub> at steady state with the breathe was in a similar range or higher than that achieved with the Hamilton T1 with the low-pressure supply. By contrast, FiO<sub>2</sub> measured with the GirVent was considerably low in the current design stage. In the tests with the high-pressure O<sub>2</sub> supply, the Hamilton T1 achieved the intended O<sub>2</sub> concentration with minor deviations from the set value. FiO<sub>2</sub> provided by the HEV was constantly higher than the set concentration in the graphic UI.</p>
<p>The t<sub>90</sub> values for the low-pressure supply (12 L min<sup>&#x02212;1</sup>) and the high-pressure supply (4 bar, O<sub>2</sub> at 100%), as well as the FiO<sub>2</sub> achieved in the airway (FiO<sub>2, t90</sub>), are shown in <xref ref-type="fig" rid="F6">Figure 6</xref>. GirVent showed for both O<sub>2</sub> ports (LP-1 and LP-2) t<sub>90</sub> values similar to those of the Hamilton T1; however, the achieved FiO<sub>2, t90</sub> values were also lower than those of the other PDVs. For the breathe (LP O<sub>2</sub> port) and the HEV (HP O<sub>2</sub> port), the t<sub>90</sub> values were higher than the values of the corresponding ports in the Hamilton T1, but both ventilators achieved FiO<sub>2, t90</sub> of 1.00.</p>
</sec>
<sec>
<title>Trigger signals</title>
<p>The trigger signals and pressurization characteristics of each ventilator for the two P<sub>0.1</sub> level set (2 cmH<sub>2</sub>O/100 ms and 4 cmH<sub>2</sub>O/100 ms) are shown in <xref ref-type="fig" rid="F7">Figure 7</xref>. Exact values for all measurements are depicted in the <xref ref-type="supplementary-material" rid="SM1">Supplementary material</xref>.</p>
<fig id="F7" position="float">
<label>Figure 7</label>
<caption><p>Results of the trigger characteristics for occlusion pressure values (P0.1) of 2 cmH<sub>2</sub>O/100 ms and 4 cmH<sub>2</sub>O/100 ms, indicated by the subscript in the corresponding legend. Error bars indicate the standard deviation. <bold>(A)</bold> pressure drop (PD), <bold>(B)</bold> time to pressure minimum (TPM) and trigger delay time (TDT), <bold>(C)</bold> time pressure product (TPP), and <bold>(D)</bold> pressurization capacity after 300 ms and 500 ms (PC300 and PC500).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0007.tif"/>
</fig>
<p>For both values of P<sub>0.1</sub>, PD and TDT values observed in the PDVs were generally larger than those of the reference ventilator. The HEV presented the smallest difference to the reference ventilator and achieved faster TDT at a P<sub>0.1</sub> of 2 cmH<sub>2</sub>O/100 ms.</p>
<p>The pressurization capacity of the PDVs was notably lower than that of the reference ventilator, which achieved values between 51.5 and 80.1%. The HEV presented values between 24.7% (PC300) and 72.5% (PC500), while those of the GirVent ranged between 40.4 and 64.0%, respectively. The values observed in the breathe ranged between 5.4 and 15.6%.</p>
</sec>
</sec>
<sec sec-type="discussion" id="s4">
<title>Discussion</title>
<p>The test protocol presented was designed over several iterations in conjunction with the PDVs. It covers the most essential aspects for the first development phases of PDVs according to the MHRA specification. The chosen testing aspects&#x02014;the accuracy of controls, the oxygen dynamics, and the trigger signals&#x02014;are, from a technical point of view, the most crucial aspects in the first development phases. Additional features such as the basic alarms and monitoring are also of importance, although these can be addressed in the later stages of development.</p>
<p>The test protocol presented offers a compact and comprehensible evaluation of ventilators compared to the current guidelines and norms (<xref ref-type="bibr" rid="B5">5</xref>, <xref ref-type="bibr" rid="B12">12</xref>). It serves as a complementary aid to the standard and covers the relevant aspects for the early development phases of ventilators. The automated plotting of the individual test scenarios allowed to see specific trends and provide conclusions for improvement. Various characteristics and the key differences of PDVs with different functional principles can be distinguished. The use of a reference ventilator allowed continuous comparison of the PDVs with a state-of-the-art ventilator. Overall, the most prominent differences and characteristics of the PDVs tested could be elaborated.</p>
<sec>
<title>Accuracy of the control systems and instruments</title>
<p>Generally, the design stage tested of the PDVs showed satisfactory accuracy of the control systems and instruments (<xref ref-type="bibr" rid="B12">12</xref>). Nonetheless, individual mechanisms of the PDVs resulted in some strong outliers and in increased inaccuracies at individual test settings compared to those of the reference ventilator.</p>
<p>All ventilators showed a decelerating flow profile, except the breathe which showed an ascending flow profile. Hence, a direct comparison between the ventilators is not desirable. However, the results presented evaluate their accuracy regarding their main control variables, independent of their different functional principles.</p>
<p>The applied tests resulted in a critical differentiation of functional performance of the evaluated PDVs, such as the passive PEEP valves (HEV and breathe) showed larger relative errors and variations compared to the actively controlled PEEP by blowers (Hamilton T1 and GirVent). The PEEP generated by a passive PEEP valve is more prone to errors and potentially less stable under clinical conditions of respiratory diseases.</p>
<p>Each pressure-controlled ventilator displayed maximum PIP error at a PIP of 30 cm H<sub>2</sub>O, with an RR of 20 bpm. As shown in <xref ref-type="fig" rid="F4">Figure 4</xref>, the PIP error increases when inspiratory time is decreased. This effect is mainly caused by the resistance of the ETT used in the experiments. <xref ref-type="fig" rid="F8">Figure 8</xref> shows an example of the pressure profile (HEV ventilator) measured with and without an ETT in the setting PIP = 30 cm H<sub>2</sub>O and RR = 20 bpm.</p>
<fig id="F8" position="float">
<label>Figure 8</label>
<caption><p>Influence of the endotracheal tube (ETT) on the measured airway pressure shown for the HEV ventilator. The ETT introduces additional flow resistance, resulting in a considerably lower peak pressure. This effect was evident in each ventilator tested.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmedt-04-899328-g0008.tif"/>
</fig>
</sec>
<sec>
<title>Oxygen dynamics</title>
<p>The main target regions for PDVs are environments with low or limited access to a supply of O<sub>2</sub>. The O<sub>2</sub> performance with a limited flow feed-forward control is highly relevant for PDVs, and the economic use of O<sub>2</sub> is of great importance. However, concerning clinical implications, it is crucial that a PDV is also capable of delivering a high O<sub>2</sub> concentration.</p>
<p>According to the ISO 80601-2-12 norm and the MHRA specification, the accepted accuracy for the oxygen concentrations is &#x000B1; 5% of the set value. As such, this can only be evaluated for the feedback control of the high-pressure O<sub>2</sub> supply tests. However, the norm and the MHRA specification do not state any accepted performance for the low-pressure O<sub>2</sub> efficiency.</p>
<p>In the low-pressure O<sub>2</sub> supply mode (2, 4, 6 L min<sup>&#x02212;1</sup>), the breathe achieved higher FiO<sub>2</sub> at steady state than the Hamilton T1. The resuscitator bag serves as a large reservoir, which premixes air and O<sub>2</sub>, thereby allowing a homogeneous gas mix. However, this system has no other form of O<sub>2</sub> control than the supply flow set. In sharp contrast, the GirVent achieved considerably lower FiO<sub>2</sub> with both O<sub>2</sub> port modes implemented. A solenoid valve is used here to limit the O<sub>2</sub> injection exclusively to the inspiratory phase. This design results in a sparing use of O<sub>2</sub>; however, high FiO<sub>2</sub> cannot be achieved. The HEV ventilator with the high-pressure O<sub>2</sub> port showed a larger error than 5%, with continuous overshoot to the set FiO<sub>2</sub>. Thus, it requires additional tuning of the software and calibration of the O<sub>2</sub> sensor. By contrast, the Hamilton T1 showed high accuracy in the high-pressure O<sub>2</sub> tests, ranging within the 5% tolerance.</p>
<p>The t<sub>90</sub> value represents the response to an FiO<sub>2</sub> change. A limited increase in maximal FiO<sub>2</sub> may result in a low t<sub>90</sub> value, but it also indicates that the ventilator system does not manage to provide the patient with a high FiO<sub>2</sub>. For instance, a very low t<sub>90</sub>value was observed in the GirVent, but the maximal FiO<sub>2</sub> achieved is likewise very low, whereas with the breathe and the HEV, a long t<sub>90</sub> value was observed, but the achieved FiO<sub>2</sub> concentrations approach 1.00. In general, the O<sub>2</sub> operation modes of the HEV and the breathe show that premixing O<sub>2</sub> in a reservoir is beneficial for a homogeneous O<sub>2</sub> supply. In addition, the implementation of an O<sub>2</sub> sensor is recommendable even if FiO<sub>2</sub> can be regulated with the setting of the O<sub>2</sub> source.</p>
</sec>
<sec>
<title>Ventilator trigger signals</title>
<p>The correct detection of and fast reaction to a patient&#x00027;s breathing efforts is crucial to prevent patient&#x02013;ventilator asynchrony, which may cause discomfort or even lung injury (<xref ref-type="bibr" rid="B21">21</xref>, <xref ref-type="bibr" rid="B22">22</xref>). A higher initial flow after triggering results in a better pressurization capacity of the ventilator and in decreased patient effort (<xref ref-type="bibr" rid="B23">23</xref>, <xref ref-type="bibr" rid="B24">24</xref>). Overall, the PDVs tested here detected the applied P<sub>0.1</sub> levels and suitably triggered breathing cycles generated by the TestChest. Concerning the acceptable performance for the trigger detection and reaction, neither the ISO 80601-2-12 standard nor the MHRA specification specifies any tolerances.</p>
<p>Relative to results reported in commercial ventilators, our measurements showed generally higher TDT, including the Hamilton T1 (<xref ref-type="bibr" rid="B17">17</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>). This may be explained by the additional resistance imposed by the ETT, which causes a restricted and delayed flow signal, which in turn requires more sensitive triggering. Compared to studies that used a PEEP level of 0 cmH<sub>2</sub>O or 5 cmH<sub>2</sub>O (<xref ref-type="bibr" rid="B17">17</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>), the use of an increased PEEP level (10 cmH<sub>2</sub>O) requires a less sensitive trigger to avoid auto-triggering (<xref ref-type="bibr" rid="B17">17</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>). This directly results in a slower detection and a longer TDT.</p>
<p>Although the HEV and the GirVent display slightly higher PD and TDT levels and lower pressurization capacities than the reference ventilator, the results are encouraging, also considering the fact that the devices are still in an early development state. Furthermore, Delgado et al. (<xref ref-type="bibr" rid="B18">18</xref>) reported pressurization capacities of different commercial ventilators in a similar magnitude to those we observed in the HEV and the GirVent.</p>
</sec>
<sec>
<title>Strengths and limitations</title>
<p>In this study, we showed that the presented test protocol enables rapid and conclusive evaluations of PDVs during their early development phase. The test protocol and the evaluations of the individual parameters were compared with the performance of the Hamilton T1 as a reference system. The simplicity of the protocol in combination with the automated detection and evaluation algorithm, fast quantification of the performance, as well as iterative and test-oriented development of PDVs, is supported. Safety features required according to the norm ISO 80601-2-12:2020 (<xref ref-type="bibr" rid="B12">12</xref>) and the MHRA recommendation (<xref ref-type="bibr" rid="B5">5</xref>) were considered but are not comprehensively incorporated in the present study. The objective of this simplified protocol was to enable data-driven development based on realistic <italic>in vitro</italic> tests. However, our test protocol does not replace the tests required for certification.</p>
<p>The norm ISO 80601-2-12:2020 (<xref ref-type="bibr" rid="B12">12</xref>) states to use a linear resistance that accounts for the resistance induced by the ETT and the pulmonary system. However, the restricting cross-sectional area of an ETT introduces a relevant parabolic resistance and flow disturbance that might not be accounted for solely with a linear resistance implemented in the mechanical lung, especially at high flow rates. The impact of the ETT is also evident in the trigger signal evaluation. Overall, the use of ETTs is of critical importance as this represents clinical reality of the gold standard therapeutic strategy, and it is beneficial to investigate representative clinical scenarios.</p>
<p>Other limitations of the test protocol presented include the lack of testing of the efficiency of O<sub>2</sub> use of each ventilator. To determine their actual efficiency, the amount of O<sub>2</sub> used from the source to achieve the desired FiO<sub>2</sub> needs to be measured. Also, the robustness of software and hardware of each ventilator ought to be assessed under high load. Last, the patient cycling that describes the transition from inspiration to expiration during assisted ventilation (<xref ref-type="bibr" rid="B25">25</xref>) was not measured in this study.</p>
<p>In comparison to <italic>in vivo</italic> testing, our test approach provides simplicity and complete experimental control. Nonetheless, it is physiologically limited, whereas <italic>in vivo</italic> tests provide more specific and reliable observation of biological effects but with strict regulations and compliance standards.</p>
<p>In conclusion, we presented a systematic test approach, which helps develop pandemic ventilators in a short time, independent of their ventilation principle. The test protocol covers the essential functional aspects for the initial development phase of PDVs. In addition, the automated and modular evaluation scripts allows for performing iterative investigations and drawing conclusions regarding specific ventilator characteristics.</p>
</sec>
</sec>
<sec sec-type="data-availability" id="s5">
<title>Data availability statement</title>
<p>The raw data of the Hamilton T1 supporting the conclusions of this article will be made available by the authors, without undue reservation.</p>
</sec>
<sec id="s6">
<title>Ethics statement</title>
<p>Ethical review and approval was not required for this study in accordance with the local legislation and institutional requirements.</p>
</sec>
<sec id="s7">
<title>Author contributions</title>
<p>The experiments were conceived and designed by NT and NSte. Data collection and analysis was performed by NT, NSte, and MZ. Literature search and study discussions were conducted by NT, NSte, MZ, MS, NSta, TE, and JH. The manuscript was prepared by NT and reviewed by NSte, MZ, NSta, MM, TE, JH, and MS. All authors contributed to the article and approved the submitted version.</p>
</sec>
<sec sec-type="funding-information" id="s8">
<title>Funding</title>
<p>This project was financed by ETH Z&#x000FC;rich and the Botnar Research Center for Child Health (COVent project). The sponsor had no role in the design or conduct of this research. Open access funding provided by ETH Z&#x000FC;rich.</p>
</sec>
<sec sec-type="COI-statement" id="conf1">
<title>Conflict of interest</title>
<p>Authors NT, NSte, MZ, MM, and MS with the Product Development Group Zurich, ETH Z&#x000FC;rich, declare that they have been involved in the development of each PDV presented in this study by actively testing several iterations and providing the teams with feedback. No funding has been received by any of the teams associated to the PDVs. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="disclaimer" id="s9">
<title>Publisher&#x00027;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
</body>
<back> 
<ack><p>We gratefully thank Adrian Marty and the team at the Simulation Center of the University Hospital Zurich for providing medical knowledge, the testing platform TestChest, and the Hamilton T1, as well as Josef Brunner (neosim AG) for his technical support. We thank the teams of the breathe, GirVent, and HEV ventilators for the productive discussions, as well as ETH Z&#x000FC;rich and the Botnar Research Center for Child Health (COVent project) for their financial support.</p>
</ack>
<sec sec-type="supplementary-material" id="s10">
<title>Supplementary material</title>
<p>The Supplementary Material for this article can be found online at: <ext-link ext-link-type="uri" xlink:href="https://www.frontiersin.org/articles/10.3389/fmedt.2022.899328/full#supplementary-material">https://www.frontiersin.org/articles/10.3389/fmedt.2022.899328/full#supplementary-material</ext-link></p>
<supplementary-material xlink:href="Table_1.pdf" id="SM1" mimetype="application/pdf" xmlns:xlink="http://www.w3.org/1999/xlink"/>
<supplementary-material xlink:href="Data_Sheet_1.PDF" id="SM2" mimetype="application/pdf" xmlns:xlink="http://www.w3.org/1999/xlink"/>
<supplementary-material xlink:href="Data_Sheet_2.ZIP" id="SM3" mimetype="application/zip" xmlns:xlink="http://www.w3.org/1999/xlink"/>
<supplementary-material xlink:href="Data_Sheet_3.ZIP" id="SM4" mimetype="application/zip" xmlns:xlink="http://www.w3.org/1999/xlink"/>
<supplementary-material xlink:href="Data_Sheet_4.ZIP" id="SM5" mimetype="application/zip" xmlns:xlink="http://www.w3.org/1999/xlink"/>
</sec>
<ref-list>
<title>References</title>
<ref id="B1">
<label>1.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Weng</surname> <given-names>L</given-names></name> <name><surname>Phua</surname> <given-names>J</given-names></name> <name><surname>Weng</surname> <given-names>L</given-names></name> <name><surname>Ling</surname> <given-names>L</given-names></name> <name><surname>Egi</surname> <given-names>M</given-names></name> <name><surname>Lim</surname> <given-names>C-M</given-names></name> <etal/></person-group>. <article-title>Fast and Chronic Programmes Intensive care management of coronavirus disease 2019 (COVID-19): challenges and recommendations</article-title>. <source>Lancet Respir Med.</source> (<year>2020</year>) <volume>8</volume>:<fpage>506</fpage>&#x02013;<lpage>23</lpage>. <pub-id pub-id-type="doi">10.1016/S2213-2600(20)30161-2</pub-id><pub-id pub-id-type="pmid">32272080</pub-id></citation></ref>
<ref id="B2">
<label>2.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wilcox</surname> <given-names>SR</given-names></name></person-group>. <article-title>Management of respiratory failure due to covid-19 Pathology and management are similar to acute respiratory distress syndrome</article-title>. <source>BMJ</source>. (<year>2020</year>) <volume>369</volume>:<fpage>m1786</fpage>. <pub-id pub-id-type="doi">10.1136/bmj.m1786</pub-id><pub-id pub-id-type="pmid">32366375</pub-id></citation></ref>
<ref id="B3">
<label>3.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ranney</surname> <given-names>ML</given-names></name> <name><surname>Griffeth</surname> <given-names>V</given-names></name> <name><surname>Jha</surname> <given-names>AK</given-names></name></person-group>. <article-title>Critical supply shortages &#x02014; the need for ventilators and personal protective equipment during the COVID-19 pandemic</article-title>. <source>N Engl J Med.</source> (<year>2020</year>) <volume>382</volume>:<fpage>e41</fpage>. <pub-id pub-id-type="doi">10.1056/NEJMp2006141</pub-id><pub-id pub-id-type="pmid">32212516</pub-id></citation></ref>
<ref id="B4">
<label>4.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Dhanani</surname> <given-names>J</given-names></name> <name><surname>Pang</surname> <given-names>G</given-names></name> <name><surname>Pincus</surname> <given-names>J</given-names></name> <name><surname>Ahern</surname> <given-names>B</given-names></name> <name><surname>Goodwin</surname> <given-names>W</given-names></name> <name><surname>Cowling</surname> <given-names>N</given-names></name> <etal/></person-group>. <article-title>Increasing ventilator surge capacity in COVID 19 pandemic: design, manufacture and in vitro-in vivo testing in anaesthetized healthy pigs of a rapid prototyped mechanical ventilator</article-title>. <source>BMC Res Notes.</source> (<year>2020</year>) <volume>13</volume>:<fpage>421</fpage>. <pub-id pub-id-type="doi">10.1186/s13104-020-05259-z</pub-id><pub-id pub-id-type="pmid">32894167</pub-id></citation></ref>
<ref id="B5">
<label>5.</label>
<citation citation-type="web"><person-group person-group-type="author"><collab>Medicines Healthcare products Regulatory Agency. Specification for Ventilators to be Used in UK Hospitals During the Coronavirus (COVID-19) Outbreak.</collab></person-group> (<year>2020</year>). Available online at: <ext-link ext-link-type="uri" xlink:href="https://www.gov.uk/government/publications/specification-for-ventilators-to-be-used-in-uk-hospitals-during-the-coronavirus-covid-19-outbreak">https://www.gov.uk/government/publications/specification-for-ventilators-to-be-used-in-uk-hospitals-during-the-coronavirus-covid-19-outbreak</ext-link> (accessed January 14, 2021)</citation>
</ref>
<ref id="B6">
<label>6.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Li</surname> <given-names>H</given-names></name> <name><surname>Li</surname> <given-names>E</given-names></name> <name><surname>Krishnamurthy</surname> <given-names>D</given-names></name> <name><surname>Kolbay</surname> <given-names>P</given-names></name> <name><surname>Chacin</surname> <given-names>B</given-names></name> <name><surname>Hoehne</surname> <given-names>S</given-names></name> <etal/></person-group>. <article-title>Utah-Stanford Ventilator (Vent4US): developing a rapidly scalable ventilator for COVID-19 patients with ARDS</article-title>. <source>medRxiv.</source> (<year>2020</year>). <pub-id pub-id-type="doi">10.1101/2020.04.18.20070367</pub-id></citation>
</ref>
<ref id="B7">
<label>7.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Galbiati</surname> <given-names>C</given-names></name> <name><surname>Bonivento</surname> <given-names>W</given-names></name> <name><surname>Caravati</surname> <given-names>M</given-names></name> <name><surname>De Cecco</surname> <given-names>S</given-names></name> <name><surname>Dinon</surname> <given-names>T</given-names></name> <name><surname>Fiorillo</surname> <given-names>G</given-names></name> <etal/></person-group>. <article-title>Mechanical Ventilator Milano (MVM): a novel mechanical ventilator designed for mass scale production in response to the COVID-19 pandemics</article-title>. <source>medRxiv.</source> (<year>2020</year>). <pub-id pub-id-type="doi">10.1101/2020.03.24.20042234</pub-id></citation>
</ref>
<ref id="B8">
<label>8.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Buytaert</surname> <given-names>J</given-names></name> <name><surname>Collins</surname> <given-names>P</given-names></name> <name><surname>Abed Abud</surname> <given-names>A</given-names></name> <name><surname>Allport</surname> <given-names>P</given-names></name> <name><surname>Pazos &#x000C1;lvarez</surname> <given-names>A</given-names></name> <name><surname>Akiba</surname> <given-names>K</given-names></name> <etal/></person-group>. <article-title>The HEV Ventilator: at the interface between particle physics and biomedical engineering</article-title>. <source>R Soc Open Sci.</source> (<year>2022</year>) <volume>9</volume>:<fpage>211519</fpage>. <pub-id pub-id-type="doi">10.1098/rsos.211519</pub-id><pub-id pub-id-type="pmid">35308626</pub-id></citation></ref>
<ref id="B9">
<label>9.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>du Pasquier</surname> <given-names>C</given-names></name> <name><surname>Hewing</surname> <given-names>L</given-names></name> <name><surname>Steffen</surname> <given-names>N</given-names></name> <name><surname>Lumpe</surname> <given-names>TS</given-names></name> <name><surname>Tachatos</surname> <given-names>N</given-names></name> <name><surname>Menner</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Systematic design, control, and parametric testing of an automated resuscitator bag mechanical ventilator</article-title>. <source>J Mech Des.</source> (<year>2021</year>) <volume>143</volume>:<fpage>125001</fpage>. <pub-id pub-id-type="doi">10.1115/1.4051290</pub-id></citation>
</ref>
<ref id="B10">
<label>10.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Madekurozwa</surname> <given-names>M</given-names></name> <name><surname>Bonneuil W</surname> <given-names>V</given-names></name> <name><surname>Frattolin</surname> <given-names>J</given-names></name> <name><surname>Watson</surname> <given-names>DJ</given-names></name> <name><surname>Moore</surname> <given-names>AC</given-names></name> <name><surname>Stevens</surname> <given-names>MM</given-names></name> <etal/></person-group>. <article-title>A novel ventilator design for COVID-19 and resource-limited settings</article-title>. <source>Front Med</source>. (<year>2021</year>) <volume>3</volume>:<fpage>707826</fpage>. <pub-id pub-id-type="doi">10.3389/fmedt.2021.707826</pub-id><pub-id pub-id-type="pmid">35047946</pub-id></citation></ref>
<ref id="B11">
<label>11.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Fang</surname> <given-names>Z</given-names></name> <name><surname>Li</surname> <given-names>AI</given-names></name> <name><surname>Wang</surname> <given-names>H</given-names></name> <name><surname>Zhang</surname> <given-names>R</given-names></name> <name><surname>Mai</surname> <given-names>X</given-names></name> <name><surname>Pan</surname> <given-names>T</given-names></name></person-group>. <article-title>AmbuBox: a fast-deployable low-cost ventilator for COVID-19 emergent care</article-title>. <source>SLAS Technol.</source> (<year>2020</year>) <volume>25</volume>:<fpage>573</fpage>&#x02013;<lpage>84</lpage>. <pub-id pub-id-type="doi">10.1177/2472630320953801</pub-id><pub-id pub-id-type="pmid">32882150</pub-id></citation></ref>
<ref id="B12">
<label>12.</label>
<citation citation-type="web"><person-group person-group-type="author"><collab>International Organization for Standardization</collab></person-group>. <article-title>ISO 80601-2-12:2020 - Medical electrical equipment &#x02014; Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators</article-title>. Available online at: <ext-link ext-link-type="uri" xlink:href="https://www.iso.org/standard/72069.html">https://www.iso.org/standard/72069.html</ext-link> (accessed January 16, 2021).</citation>
</ref>
<ref id="B13">
<label>13.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Hewing</surname> <given-names>L</given-names></name> <name><surname>Menner</surname> <given-names>M</given-names></name> <name><surname>Tachatos</surname> <given-names>N</given-names></name> <name><surname>Daners</surname> <given-names>M</given-names></name> <name><surname>Pasquier C</surname> <given-names>Du</given-names></name> <name><surname>Lumpe</surname> <given-names>TS</given-names></name> <etal/></person-group>. <article-title>Volume control of low-cost ventilator with automatic set-point adaptation</article-title>. in <source>European Control Conference (ECC).</source> (<year>2021</year>) <fpage>781</fpage>&#x02013;<lpage>6</lpage>. <pub-id pub-id-type="doi">10.23919/ECC54610.2021.9655092</pub-id></citation>
</ref>
<ref id="B14">
<label>14.</label>
<citation citation-type="web"><person-group person-group-type="author"><collab>MIT, E-Vent. Emergency Ventilator Project</collab></person-group>. Available online at: <ext-link ext-link-type="uri" xlink:href="https://emergency-vent.mit.edu/">https://emergency-vent.mit.edu/</ext-link> (accessed January 15, 2021).</citation>
</ref>
<ref id="B15">
<label>15.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Arnal</surname> <given-names>J-M</given-names></name> <name><surname>Garnero</surname> <given-names>A</given-names></name> <name><surname>Saoli</surname> <given-names>M</given-names></name> <name><surname>Chatburn</surname> <given-names>RL</given-names></name></person-group>. <article-title>Parameters for simulation of adult subjects during mechanical ventilation</article-title>. <source>Respir Care.</source> (<year>2018</year>) <volume>63</volume>:<fpage>158</fpage>&#x02013;<lpage>68</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.05775</pub-id><pub-id pub-id-type="pmid">29042486</pub-id></citation></ref>
<ref id="B16">
<label>16.</label>
<citation citation-type="journal"><person-group person-group-type="author"><collab>The Acute Respiratory Distress Syndrome Network</collab></person-group>. <article-title>Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome</article-title>. <source>N Engl J Med</source>. (<year>2000</year>) <volume>342</volume>:<fpage>1301</fpage>&#x02013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.1056/NEJM200005043421801</pub-id><pub-id pub-id-type="pmid">10793162</pub-id></citation></ref>
<ref id="B17">
<label>17.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Blakeman</surname> <given-names>TC</given-names></name> <name><surname>Branson</surname> <given-names>RD</given-names></name></person-group>. <article-title>Evaluation of 4 new generation portable ventilators</article-title>. <source>Respir Care.</source> (<year>2013</year>) <volume>58</volume>:<fpage>264</fpage>&#x02013;<lpage>72</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.01994</pub-id><pub-id pub-id-type="pmid">22781705</pub-id></citation></ref>
<ref id="B18">
<label>18.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Delgado</surname> <given-names>C</given-names></name> <name><surname>Romero</surname> <given-names>JE</given-names></name> <name><surname>Puig</surname> <given-names>J</given-names></name> <name><surname>Izquierdo</surname> <given-names>A</given-names></name> <name><surname>Ferrando</surname> <given-names>C</given-names></name> <name><surname>Belda</surname> <given-names>FJ</given-names></name> <etal/></person-group>. <article-title>Performance of the new turbine mid-level critical care ventilators</article-title>. <source>Respir Care.</source> (<year>2017</year>) <volume>62</volume>:<fpage>34</fpage>&#x02013;<lpage>41</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.04938</pub-id><pub-id pub-id-type="pmid">28003552</pub-id></citation></ref>
<ref id="B19">
<label>19.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Thille</surname> <given-names>AW</given-names></name> <name><surname>Lyazidi</surname> <given-names>A</given-names></name> <name><surname>Richard</surname> <given-names>J-CM</given-names></name> <name><surname>Galia</surname> <given-names>F</given-names></name> <name><surname>Brochard</surname> <given-names>L</given-names></name></person-group>. <article-title>A bench study of intensive-care-unit ventilators: new vs. old and turbine-based vs compressed gas-based ventilators</article-title>. <source>Intensive Care Med.</source> (<year>2009</year>) <volume>35</volume>:<fpage>1368</fpage>&#x02013;<lpage>76</lpage>. <pub-id pub-id-type="doi">10.1007/s00134-009-1467-7</pub-id><pub-id pub-id-type="pmid">19352622</pub-id></citation></ref>
<ref id="B20">
<label>20.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sassoon</surname> <given-names>CSH</given-names></name></person-group>. <article-title>Triggering of the ventilator in patient-ventilator interactions</article-title>. <source>Respir Care.</source> (<year>2011</year>) <volume>56</volume>:<fpage>39</fpage>&#x02013;<lpage>51</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.01006</pub-id><pub-id pub-id-type="pmid">21235837</pub-id></citation></ref>
<ref id="B21">
<label>21.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mirabella</surname> <given-names>L</given-names></name> <name><surname>Cinnella</surname> <given-names>G</given-names></name> <name><surname>Costa</surname> <given-names>R</given-names></name> <name><surname>Cortegiani</surname> <given-names>A</given-names></name> <name><surname>Tullo</surname> <given-names>L</given-names></name> <name><surname>Rauseo</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Patient-ventilator asynchronies: clinical implications and practical solutions</article-title>. <source>Respir Care.</source> (<year>2020</year>) <volume>65</volume>:<fpage>1751</fpage>&#x02013;<lpage>66</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.07284</pub-id><pub-id pub-id-type="pmid">32665426</pub-id></citation></ref>
<ref id="B22">
<label>22.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sottile</surname> <given-names>PD</given-names></name> <name><surname>Albers</surname> <given-names>D</given-names></name> <name><surname>Smith</surname> <given-names>BJ</given-names></name> <name><surname>Moss</surname> <given-names>MM</given-names></name></person-group>. <article-title>Ventilator dyssynchrony - detection, pathophysiology, and clinical relevance: a narrative review</article-title>. <source>Ann Thorac Med.</source> (<year>2020</year>) <volume>15</volume>:<fpage>190</fpage>&#x02013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.4103/atm.ATM_63_20</pub-id><pub-id pub-id-type="pmid">33381233</pub-id></citation></ref>
<ref id="B23">
<label>23.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Bonmarchand</surname> <given-names>G</given-names></name> <name><surname>Chevron</surname> <given-names>V</given-names></name> <name><surname>M&#x000E9;nard</surname> <given-names>JF</given-names></name> <name><surname>Girault</surname> <given-names>C</given-names></name> <name><surname>Moritz-Berthelot</surname> <given-names>F</given-names></name> <name><surname>Pasquis</surname> <given-names>P</given-names></name> <etal/></person-group>. <article-title>Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients</article-title>. <source>Crit Care Med.</source> (<year>1999</year>) <volume>27</volume>:<fpage>715</fpage>&#x02013;<lpage>22</lpage>. <pub-id pub-id-type="doi">10.1097/00003246-199904000-00023</pub-id><pub-id pub-id-type="pmid">10321660</pub-id></citation></ref>
<ref id="B24">
<label>24.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Chiumello</surname> <given-names>D</given-names></name> <name><surname>Pelosi</surname> <given-names>P</given-names></name> <name><surname>Croci</surname> <given-names>M</given-names></name> <name><surname>Bigatello</surname> <given-names>LM</given-names></name> <name><surname>Gattinoni</surname> <given-names>L</given-names></name></person-group>. <article-title>The effects of pressurization rate on breathing pattern, work of breathing, gas exchange and patient comfort in pressure support ventilation</article-title>. <source>Eur Respir J.</source> (<year>2001</year>) <volume>18</volume>:<fpage>107</fpage>&#x02013;<lpage>14</lpage>. <pub-id pub-id-type="doi">10.1183/09031936.01.00083901</pub-id><pub-id pub-id-type="pmid">11510780</pub-id></citation></ref>
<ref id="B25">
<label>25.</label>
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gentile</surname> <given-names>MA</given-names></name></person-group>. <article-title>Cycling of the mechanical ventilator breath</article-title>. <source>Respir Care.</source> (<year>2011</year>) <volume>56</volume>:<fpage>52</fpage>&#x02013;<lpage>60</lpage>. <pub-id pub-id-type="doi">10.4187/respcare.01088</pub-id><pub-id pub-id-type="pmid">21235838</pub-id></citation></ref>
</ref-list>
<glossary>
<def-list>
<title>Abbreviations</title>
<def-item><term>PDV</term>
<def><p>pandemic ventilators</p></def></def-item>
<def-item><term>PC-CMV</term>
<def><p>pressure-controlled continuous mandatory ventilation</p></def></def-item>
<def-item><term>VC-CMV</term>
<def><p>volume-controlled continuous mandatory ventilation</p></def></def-item>
<def-item><term>MHRA</term>
<def><p>Medicines and Healthcare Products Regulatory Agency</p></def></def-item>
<def-item><term>ETT</term>
<def><p>endotracheal tube</p></def></def-item>
<def-item><term>PEEP</term>
<def><p>positive end-expiratory pressure</p></def></def-item>
<def-item><term>PIP</term>
<def><p>peak inspiratory pressure</p></def></def-item>
<def-item><term>RR</term>
<def><p>respiratory rate</p></def></def-item>
<def-item><term>V<sub>t</sub></term>
<def><p>tidal volume</p></def></def-item>
<def-item><term>R<sub>aw</sub></term>
<def><p>airway resistance</p></def></def-item>
<def-item><term>C<sub>rs</sub></term>
<def><p>respiratory compliance</p></def></def-item>
<def-item><term>P<sub>aw</sub></term>
<def><p>airway pressure</p></def></def-item>
<def-item><term>Q<sub>TestChest</sub></term>
<def><p>airflow into TestChest</p></def></def-item>
<def-item><term>T<sub>I</sub></term>
<def><p>inspiratory time</p></def></def-item>
<def-item><term>T<sub>E</sub></term>
<def><p>expiratory time</p></def></def-item>
<def-item><term>I:E</term>
<def><p>inspiratory to expiratory time ratio</p></def></def-item>
<def-item><term>FiO<sub>2</sub></term>
<def><p>fraction of inspired O<sub>2</sub></p></def></def-item>
<def-item><term>HP</term>
<def><p>high pressure</p></def></def-item>
<def-item><term>LP</term>
<def><p>low pressure</p></def></def-item>
<def-item><term>PD</term>
<def><p>pressure drop</p></def></def-item>
<def-item><term>TPM</term>
<def><p>time to pressure minimum</p></def></def-item>
<def-item><term>TDT</term>
<def><p>trigger delay time</p></def></def-item>
<def-item><term>PTP</term>
<def><p>pressure time product</p></def></def-item>
<def-item><term>PC</term>
<def><p>pressurization capacity.</p></def></def-item>
</def-list>
</glossary> 
</back>
</article> 