Abstract
Herein, advancements in electroanalytical devices for the simultaneous detection of diverse breast cancer (BC) markers are demonstrated. This article identifies several important areas of exploration for electrochemical diagnostics and highlights important factors that are pivotal for the successful deployment of novel bioanalytical devices. We have highlighted that the limits of detection (LOD) reported for the multiplex electrochemical biosensor can surpass the sensitivity displayed by current clinical standards such as ELISA, FISH, and PCR. HER-2; a breast cancer marker characterised by increased metastatic potential, more aggressive development, and poor clinical outcomes; can be sensed with a LOD of 0.5 ng/ml using electrochemical multiplex platforms, which falls within the range of that measured by ELISA (from picogram/ml to nanogram/ml). Electrochemical multiplex biosensors are reported with detection limits of 0.53 ng/ml and 0.21 U/ml for MUC-1 and CA 15-3, respectively, or 5.8 × 10−3 U/ml for CA 15-3 alone. The sensitivity of electrochemical assays is improved when compared to conventional analysis of MUC-1 protein which is detected at 11–12 ng/ml, and ≤30 U/ml for CA 15-3 in the current clinical blood tests. The LOD for micro-ribonucleic acid (miRNA) biomarkers analyzed by electrochemical multiplex assays were all notedly superior at 9.79 × 10−16 M, 3.58 × 10−15 M, and 2.54 × 10−16 M for miRNA-155, miRNA-21, and miRNA-16, respectively. The dogma in miRNA testing is the qRT-PCR method, which reports ranges in the ng/ml level for the same miRNAs. Breast cancer exosomes, which are being explored as a new frontier of biosensing, have been detected electrochemically with an LOD of 103–108 particles/mL and can exceed detection limits seen by the tracking and analysis of nanoparticles (∼ 107 particles/ml), flow cytometry, Western blotting and ELISA, etc. A range of concentration at 78–5,000 pg/ml for RANKL and 16–1,000 pg/ml for TNF is reported for ELISA assay while LOD values of 2.6 and 3.0 pg/ml for RANKL and TNF, respectively, are demonstrated by the electrochemical dual immunoassay platform. Finally, EGFR and VEGF markers can be quantified at much lower concentrations (0.01 and 0.005 pg/ml for EGFR and VEGF, respectively) as compared to their ELISA assays (EGRF at 0.31–20 ng/ml and VEGF at 31.3–2,000 pg/ml). In this study we hope to answer several questions: (1) Are the limits of detection (LODs) reported for multiplex electrochemical biosensors of clinical relevance and how do they compare to well-established methods like ELISA, FISH, or PCR? (2) Can a single sensor electrode be used for the detection of multiple markers from one blood drop? (3) What mechanism of electrochemical biosensing is the most promising, and what technological advancements are needed to utilize these devices for multiplex POC detection? (4) Can nanotechnology advance the sensitive and selective diagnostics of multiple BC biomarkers? (5) Are there preferred receptors (antibody, nucleic acid or their combinations) and preferred biosensor designs (complementary methods, sandwich-type protocols, antibody/aptamer concept, label-free protocol)? (6) Why are we still without FDA-approved electrochemical multiplex devices for BC screening?
1 Introduction
Accounting for a quarter of all cancers in women, breast cancer (BC), a notoriously heterogeneous disease, is the most commonly present cancer in women globally (). Currently, BC screening occurs using the gold standards of mammography and routine breast examination; each of which presents challenges in sensitivity and specificity. Unfortunately, overdiagnosis/false positives are a risk associated with mammography that can result in increased patient anxiety and, at times, unnecessary biopsy (). Additionally, mammography presents several confounding variables such as physician interpretation, poor positioning, and tumour subtypes that can result in as many as 30% of breast cancers going undetected (). Routine breast examinations on the other hand are limited to the ability of the physician and the patient to both identify and investigate changes to the breast over time. Those with dense breast tissue, approximately 43% of women, are at an even greater risk of diagnostic mishap and thus subsequent distress and harm caused by untimely intervention (, ). As such, the search for different screening methods with better specificity and lower false-positive rates is of great interest.
There are various risk factors associated with the incidence of BC including family history, age, geographic location, environmental pollutants, socioeconomic status, lifestyle (consumption of alcohol, diet, lack of physical activity) and increased breast density (). For women of all ages and at average risk, more frequent screening (mammography and clinical breast examination, CBE) could reduce mortality by 20% (). In fact, when caught in the early stages of disease BC presents with good clinical outcomes and is curable in 70%–80% of all cases (). Although the majority of patients diagnosed with new onset BC are postmenopausal (), young women (defined as <40 years old) and premenopausal women generally present with more aggressive disease and poorer outcomes than the older population. Although current screening standards have been integral to the successful identification and treatment of BCs, there are ways we may yet improve the efficiency of screening programs and reduce the burden of screening for those at an increased risk of BC. Current screening methods pose risks of both false positive and false negative results—especially in those at-risk who present with dense breast tissue (). These complications in detection may be supplemented using circulating biomarker detection panels which tend to be specific and accurate while allowing for a wide range array of serological differentiation.
It is well established in the literature that early detection is essential to improve outcomes and reduce mortality. Ensuring timely, affordable, and effective cancerous biomarker monitoring is the key to the successful implementation of cancer screening programs (). The concept of electrochemical biosensing can be utilized to accomplish each of these as testing is generally rapid (within minutes), sensitive (a single cell detection or atto-/femto-molar concentrations of biomarker), and portable (most electrochemical readers are battery-operated and miniaturized to a table or even the pocket-size). By measuring the change in electrochemical parameters such as resistivity (impedance spectroscopy) or current (amperometry and voltammetry) (–), redox reactions at the electrode or membrane surfaces can be detected and immediately interpreted for diagnostic results (, ). The core element of the electrochemical assay is its biorecognition element (receptor), which can be aptamers, antibodies, enzymes, or proteins, which are immobilized onto an electrical transducer (electrode). The receptor forms a three-dimensional structure that functions as a ligand which binds to its associated analyte (biomarker) with high specificity. Upon binding, the conformation-changing property of the receptor structure can be utilized in biosensors through a redox-active molecule (redox probe). This probe either covalently attaches to the receptor on the electrode surface or can be added to the measurement solution. Once it is presented with a biomarker-rich medium, binding occurs and the current passing through the transducer (electrode) changes proportionately in real-time (). Furthermore, the binding event is transduced into an electronic signal that quantifies the amount of analyte present in a sample and the electrochemical output can be immediately interpreted for diagnostic results.
Although effective, there is a distinct disadvantage in using a simple electrochemical sensor as it is designed to interact with, and therefore detect, only one biomolecule at a time. This is where the interesting nature of multiplex sensors can make an impact. Multiplex sensors use a variety and diversity of receptors immobilized on the single transducer platform composed of several electrodes (electrode array) to simultaneously detect a range of biological analytes taken from one sample. A screen-printed electrode array formed by several sensing electrodes is suitable for detecting multiple signals simultaneously, allowing for differential measurement of several analytes in the solution. A great example of the electrochemical multiplexer is Abbott's i-STAT blood tester equipped with cartridges supporting blood gas, chemistry, coagulation and traumatic brain injury (TBI) marker panels. (Datwyler et al., Patent WO-2018067468-A1) The Abbott i-STAT Alinity multiplex model with TBI Plasma cartridge was FDA-cleared via the 510(k) pathway in January 2021 (https://www.accessdata.fda.gov/cdrh_docs/reviews/K201778.pdf). The cartridge measures two TBI biomarkers (GFAP and UCH-L1) amperometrically using gold working electrodes and an Ag/AgCl reference electrode fabricated on a silicon substrate. (Datwyler et al., Patent WO-2018067468-A1) The i-STAT provides results 15 min after loading a plasma sample onto the cartridge, allowing for the test to be used clinically to assist in determining the need for a head CT scan (a positive result requires a head CT scan while a negative result rules out the need for a head CT). Figure 1 shows both Abbott's iSTAT electrochemical readers.
Figure 1
The key to multiplex testing lies in its functional simplicity; the multiplex sensor can in theory run a multitude of selective and specific tests on a single platform using only a single sample while simultaneously remaining low-cost and highly portable (, –). Hence, how can a multiplex assay be useful in early screening and diagnostics of BC? BC is a diverse disease with various subtypes, genetic mutations, and DNA damage (). For analytical scientists, BC is aptly categorized into five immunohistochemical subtypes based on the presence of human receptor tyrosine-protein kinase erbB-2 (HER2), antigen Ki-67, estrogen receptor (ER), and progesterone receptor (PR) (, ). These each contribute to the immunohistochemical subtypes: luminal A, luminal B (HER-2 negative), luminal B (HER-2 positive), over-expressed HER-2, and triple-negative BC (, ). Additionally, other forms of disease classification consider the stage-dependent biomarkers of BC and look to understand the overexpression of biomolecule-based biomarkers ().
Within the stage-dependent group, there are diagnostic (healthy vs. BC) (), prognostic (early BC vs. advanced BC) (), predictive (providing information regarding the treatment success probability) (), and therapeutic biomarkers (a target biomolecule for the therapeutic modalities) (). Biomolecule-based biomarkers are widely considered to be more apt targets for early diagnostics than prognostic or therapeutic biomarkers; however, there is a connection between the prognostic and diagnostic biomarkers which can be leveraged when designing a new BC diagnostic assay (). The connection between these biomarkers can be used for practical applications. For example, prognostic and predictive biomarkers are two categories that can be linked to therapy success in breast cancer subtypes. After the diagnosis of invasive breast cancer, decisions as to the choice of chemotherapy and adjuvant therapy use. These decisions can be aided by the analysis of prognostic and predictive biomarkers. The molecular markers such as estrogen/progesterone receptors (ER, PR), HER2, and the Mib1/Ki-67 proliferation index are each evaluated and play a role in the standard care of all primary, recurrent, and metastatic BC patients. CEA and CA15-3 antigens are important serum markers. Determining the series of these markers helps to monitor response to the treatment and early detection of recurrence or metastasis. Finally, BC biomarkers are also grouped as genomic (DNA), proteomic (protein-based molecules), transcriptomic (messenger RNA, micro-RNAs), and metabolomic (metabolites) based on their origin, function, and structure (). Therefore, together with the confirmation of BC, identifying the above-listed subcategories of BC would also represent an upgrade in breast cancer diagnostics and thus subsequent care.
A strategy to improve BC identification and differential diagnosis can be achieved by including additional biomarkers in the current practice. This calls for a diversity of multiplex analysis at a wide scope of molecular levels in contrast to conventional techniques. In this context, new miniaturized biosensing point-of-care (POC) prototypes are under development for fast multiplex monitoring of different biochemical and genetic pathways potentially associated with BC. There are several innovative electroanalytical platforms reported in the literature that are capable of multi-biomarker panel profiling (, ).
2 Research progress in review until the year 2018
The research aimed towards the development of multiplex electrochemical sensing devices for simultaneous sensing of BC markers published up until 2018 is summarized by Nasrollahpour et al. (published in 2022) (). This review article focused on the analysis of all types of BC electrochemical sensing approaches (that are for a single-analyte BC biomarker) and highlights the importance of nanomaterials in the design of efficient electrochemical monitoring systems. Although some multiplex electrochemical assays for BC are mentioned in this article (early work only), this technology is developing at a rapid pace. The authors of this review demonstrate the importance of nanostructured electrode components used in biosensor platforms, which allows for full utilization of the sensor's active surface area, leading to an improvement of the sensing signal and the sensor's overall performance. Lakhera and co-workers () in their recent review articles (year 2022) are specifically focused on the use of carbon electrodes (screen printed, glassy carbon) and their role as an affordable platform for detection of diverse BC markers (and potential for deployment towards their mass production for POC devices). The work presented here builds upon the analysis of a single-marker biosensor and multiplex BC electrochemical sensors are summarized with a focus on carbon electrodes used as the sensor panel.
Herein, we analyze in detail all research on the development of electrochemical multiplex (duplex, triplex, and quadruplex) BC biomarker sensing platforms. We strive to find the best approach in a multiplex design and to compare: (1) biorecognition elements that are antibody-based (immunosensors) with the nucleic acid aptamers (aptasensors); (2) the type of redox responder and its location (redox tag attached to the sensor probe or target analyte molecule) vs. redox active solutions (e.g., ferri/ferrous cyanides); (3) direct against indirect sensing mechanisms (e.g., use of sandwich-like design with/without enzymatic labels); (4) sensing signal amplification techniques; and finally (5) the reported statistical analysis and the validation of the newly developed electrochemical BC multiplex against the dogmas in BC biomarker assays.
2.1 Duplex
Cancer antigen 15-3 (CA 15-3) and HER2 extracellular domain (HER2-ECD) are independent biomarkers that when combined, can play a key role in high-risk BC diagnosis. For this reason, Marques and co-workers () developed a voltammetric sandwich-type immunosensor for the simultaneous determination of CA 15-3 and HER2-ECD. Dual screen-printed carbon electrodes (made via the printing of carbon on a ceramic platform) were coated with gold nanoparticles, followed by immobilization of captured antibodies by adsorption on each electrode (anti-CA 15-3, and anti-HER2-ECD), and backfilled by casein to minimize non-specific binding. Furthermore, a bi-immunosensor was incubated with a solution containing the antigens (HER2-ECD and CA 15-3) followed by a casting medium with anti-CA15-3- and anti-HER2- bio-antibodies. In such sandwich-type immunoassays, the antigen-antibody interaction is sensed using alkaline phosphatase (AP) as a label on detection antibodies in a mixture of 3-indoxyl phosphate with silver ions (3-IP/Ag+). The electrical signal in this system is the peak current intensity associated with the electrochemical activity of silver. This signal is captured by using linear sweep voltammetry, where the potential range is applied to the sensing platform to reach the standard electrochemical potential of silver redox probe (E0 of Ag+ + e− = Ag0), and the current associated with electron (e−) flow in this reaction is recorded. The maximum rate of this reaction is represented by a peak current on a voltammogram. The sensor signal results from metallic silver deposition on two individual electrodes (each has a separate electrical connection with the electrochemical reader), and it therefore avoids unwanted overlap of signals. This particular platform design uses the same redox label for the detection of both biomarkers (Figure 2). A sensor platform was inserted into an electrical connector, allowing for interfacing between working electrodes and a dual-channel electrochemical reader (bi-potentiostat) (). The reported LODs for the selected biomarkers were 5.0 U/ml for CA 15-3 and 2.9 ng/ml for HER2-ECD, which are comparable to the conventional assay measured by ELISA with sensitivity within 0–240 U/ml (Creative Diagnostic, Catalog # DEIA2145) and 10−12–10−9 g/ml (Bio-Techne, Catalog # DHER20) for CA 15-3 and HER2-ECD, respectively.
Figure 2
The simultaneous determination of Tumor Necrosis Factor-alpha (TNF-α), and Receptor Activator of Nuclear Factor-kB Ligand (RANKL) via a sandwich-type immunosensor built on a dual screen-printed carbon electrode platform was explored by Valverde and co-workers (
Figure 3

An electrochemical immunosensor designed on carbon screen-printed electrode (sPdCE) for the dual amperometric determination of RANKL and TNF biomarkers (
Another sandwich-type electrochemical biosensor was prototyped to detect protein MUC-1 biomarkers and MCF-7 BC cells—in which MUC-1 is overexpressed (
In brief, the sensor assembly starts with the coating of graphene/MUC-1 antibody bioconjugate on the glassy carbon disk. This step ensures sufficient surface area needed to accommodate antibodies (bare glassy carbon is a mirror-polished flat surface that is not suitable for any chemical modification needed to attach antibodies). For this reason, graphene was first mixed with EDC [N-ethyl-N'-(3-dimethylaminopropyl)carbodiimide] and NHS (N-Hydroxysuccinimide), and then with an anti-MUC-1 antibody to form an amide-type bond from amino-groups (present in antibody) and carboxyl moieties in graphene (Figure 4).
Figure 4

The formation of amide bond from the carboxylate moiety on the carbon with amine (NH2-R”) present in antibody catalyzed by EDC/NHS.
Afterwards, graphene-anti-MUC-1 was drop-cast onto the glassy carbon, dried, and exposed to analyte MUC-1 (or MUC-1 positive BC cells) for incubation (1 h). In parallel, the MUC-1 binding aptamer connected to 12-mer polycytosine (C12) was used as a template for Ag nanocluster (NCs) synthesis. Here, the researchers took advantage of the high affinity of Ag+ to negatively charged nucleic acids (from an array of phosphate groups) to control the growth and morphology of Ag NCs. The concept of using biopolymers as templating agents is particularly appealing due to the simplicity of integration of Ag nanostructures within the DNA's template (C12). The presence of four complementary bases (CGCG) at the ends of 12-mer polycytosine (C12) creates a stem-loop structure with a long C-loop (CGCGC12). The synthesis of silver nanostructures under CGCGC12 protection yields a well-designed Ag deposit on one end of this modified aptamer. In this work, the DNA strands contain MUC1 aptamer and C12 template (5′ GCAGTTGATCCTTTGGATACCCTGGC12-3′). The hairpin MUC-1 aptamer-Ag NCs were drop cast onto the MUC-1 (or MCF-7 cell) coated platform (1 h of incubation) and washed to remove unbound MUC-1 aptamer-Ag NCs. This assembly rules on specific MUC-1 biomarker-aptamer binding.
At this stage of sensor development, a square wave voltammogram (SWV) was recorded and the electrical signal corresponded to the reduction of silver originating from Ag NCs. Similar to the voltammetry method used by Marques et al. (
The sensor response was monitored by square wave voltammetry and LODs were reported as 0.5 nM, and 50 cells/ml for MUC-1 and MCF-7, respectively (
The same target cancer cells (MCF-7) and MUC-1 protein biomarkers were detected using silver-labelled biomolecules and a poly(glutamic acid)/carbon nanotube nanocomposite. MUC-1-binding aptamer was attached to the glassy carbon electrode and then modified with poly(glutamic acid)/carbon nanotube nanocomposite. This was followed by the immobilization of another Ag nanoparticle-labelled aptamer for recognition of MCF-7 cells. Such a sandwich design with an aptamer-cell-aptamer architecture has some similarities to the dual sandwich biosensor reported by Guo et. al (
In this biosensor design, Ag acts as a redox responder (similar to work by Guo et al.) (
This prototype demonstrated a limit of detection at 3 ng/ml and 25 cells/ml for MUC-1 and MCF-7, respectively (
Johari et al. designed an electrochemical biosensor platform made of a molecularly imprinted polymer that responded to epidermal growth factor receptor (EGFR) and vascular endothelial growth factor (VEGF) in serum samples (
Figure 5

An electrochemical duplex designed on a single gold screen-printed electrode for simultaneous detection of EGFR and VEGF BC biomarkers (
The first step was to modify the gold surface with thiol self-assembled monolayers (SAMs) to create a chemical bond (linker) with the gold surface at one end, and the -NH2 motifs needed for further conjugation with the target analyte molecule at the opposite end of SAMs. Afterwards, the platform was incubated with EGFR and VEGF proteins to form a covalent bond on the electrode surface via amide bond formation (Figure 4). Furthermore, molecularly imprinted polymer (MIP) was synthesized from acrylamide and N,N′-methylene-bis(acrylamide) around the EGFR and VEGF template. After polymerization was completed EGRF and VEGF templates were removed from the platform using oxalic acid (OXA). On one side, Cu(II) and Cd(II)-functionalized liposomes (Cu(II)@LP and Cd(II)@LP) were synthesized following an established protocol and resulted in nano-sized conjugates that were further modified with antibodies (AbEGFR and AbVEGF). These immunoconjugates were formed during several chemical steps which included a sulfo-SMCC crosslinking with liposomal' sulfhydryl moieties. This was followed by a washing and column separation of unbound antibodies from the antibody-tagged LPs (Cd(II)@LP and Cu(II)@LP). To assess the sensor, the MIP-SPE platform was immersed in standard solutions of VEGF and EGFR proteins, washed and incubated with AbEGFR-Cd(II)@LP and AbVEGF-Cu(II)@LP. The redox responders were Cd2+ and Cu2+ ions, and the concentration was measured by potentiometric stripping analysis (PSA) that was further correlated with EGFR and VEGF encapsulated within the MIP network. The electrochemical signal was generated using a constant current (flow of e−) through the electrode, resulting in the electrochemical reduction of cations: Cu2+ + 2e− ↔ Cu0 and Cd2+ + 2e− ↔ Cd0. The potential change associated with the electrochemical reduction of metal ions was measured (dt/dE). This indirect electrochemical test has a similar concept to biosensors discussed in the preceding section.
The set of experiments reported by this group was carried out in a standard solution of biomarkers, serum samples, and whole blood samples, and demonstrated promising selectivity and reported LODs of 0.01 and 0.005 pg/ml for EGFR and VEGF, respectively (
2.2 Triplex & quadruplex
There is limited work done for both triplex and quadruplex BC biosensors. Pioneering research on the simultaneous detection of carbohydrate antigens 153, 125, and 199 (CA 153, CA 125, CA 199) and carcinoembryonic antigen (CEA) markers was published by Wu and co-workers in 2008 (
Figure 6

A mechanism of electrochemical activity of HPR. Reproduced with permission from The catalytic pathway of horseradish peroxidase at high resolution by G. I. Berglund et al. (
It is important to note that the sensing event was enhanced by applying a driving potential, which produces an electric field at the electrode/electrolyte interface. This external electrophoretic force accelerates the transport of charged antigen molecules to electrodes and thus accelerates the binding of biomarkers. With an increasing amount of attached analyte, the immunocomplex blocks the direct electron transfer between HRP and the electrode. This was recorded as an increase in the total impedance of the sensor and corresponded to the decrease in recorded current projected on voltammograms. An increase in the concentration of proteins attached to the electrode surface caused a quenching of the electrochemical signal associated with the HRP's redox activity. This electrochemical signal was generated directly from the HRP centre without the help of any electron transfer mediator or enzymatic substrate (Figure 7).
Figure 7

Electrochemical immunosensor array (ECIA) with an electric field-driven incubation process for simultaneous detection of CA 153, CA 125, CA 199, and (CEA). (a) nylon sheet, (b) silver ink, (c) graphite auxiliary electrode, (d) Ag/AgCl reference electrode, (e) graphite working electrode, and (f) insulating dielectric. Adapted with permission from Electric Field-Driven Strategy for Multiplexed Detection of Protein Biomarkers Using a Disposable Reagentless Electrochemical Immunosensor Array by Jie Wu et al. (
The quadruplex platform showed detection limits of 0.04, 0.06, 0.1, and 0.03 U/ml for CEA, CA15-3, CA-19-9, and CA-125, respectively (
Finally, an electrochemical triplex biosensor for simultaneous identification of CA-125, CA-15-3, and CEA was constructed by immobilizing capture antibody (Ab1) on screen-printed carbon electrodes (
Table 1
| Year | Platform | BC Markers | LOD | Ref |
|---|---|---|---|---|
| 2015 | Duplex | MUC-1; MCF-7 | 0.5 nM; 50 cells/ml | ( |
| 2018 | Duplex | CA15-3; HER2-ECD | 5 U/ml; 2.9 ng/ml | ( |
| 2018 | Duplex | MUC-1; MCF-7 | 3 ng/mL; 25 cells/mL | ( |
| 2018 | Duplex | EGFR; VEGF | 0.01 pg/ml; 0.005 pg/ml | ( |
| 2020 | Duplex | TNF; RANKL | 3 pg/ml; 2.6 pg/ml | ( |
| 2020 | Duplex | CD24; CD340 | 1.94 × 105 exosomes/ul; 1.02 × 106 exosomes/ul | ( |
| 2021 | Duplex | EGFR; ICAM-1 | 103 to 105 cells | ( |
| 2021 | Duplex | CEA; CA15-3 | 3.9 pg/ml; 5.8 × 10−3 U/ml | ( |
| 2022 | Duplex | EpCAM; HER2-ECD | Single particle/ml | ( |
| 2022 | Duplex | miRNA-21; CA15-3 | 1.2 fM; 0.14 U/ml | ( |
| 2014 | Triplex | CEA; CA15-3; CA-125 | 7.0 pg/mL; 0.001 U/ml; 0.002 U/ml | ( |
| 2020 | Triplex | CEA; CA15-3; CA-125 | 0.04 pg/ml; 0.04 mU/ml; 0.04 mU/ml | ( |
| 2021 | Triplex | miRNA-155; miRNA-21; miRNA-16 | 9.79 × 10−16 M; 3.58 × 10−15 M; 2.54 × 10−16 M | ( |
| 2021 | Triplex | MUC-1; CA15-3; HER2-ECD | 0.53 ng/ml; 0.21 U/ml; 0.50 ng/ml | ( |
| 2023 | Triplex | CD63; HER2; EpCAM | 3.4 × 103–3.4 × 108 particles/ml | ( |
| 2008 | Quadruplex | CEA; CA15-3; CA19-9; CA-125 | 0.04 U/ml; 0.06 U/ml; 0.1 U/m; 0.03 U/ml | ( |
An overview of the electrochemical multiplex platforms targeting BC serum markers.
Similar to work reported by Wu et al. (
This multiplex device did not contain an enzymatic catalyst or any additional electrochemical redox mediator, it instead took advantage of the strong catalytic activity of platinum (Pt nanoparticles) toward the reduction of H2O2. The sensitivity of this multiplex platform was improved by the utilization of graphene nanosheets (NSs), which promoted fast electron transfer and expanded the surface area of the electrode, thus allowing for large loading of capture antibodies (Ab1). Taking advantage of high catalytic activity and large surface area of Pt nanoparticles, the reduction of H2O2 present in measurement solution (external redox responder) is accelerated. The triplex platform was also tested against serum samples containing known concentrations of each of the above markers.
The proposed immunoassay demonstrated very low detection limits at 0.001 U/ml, 0.002 U/ml, and 7.0 pg/ml for CA 15-3, CA-125, and CEA, respectively (
3 Recent progress: 2019–2023
This section highlights the most recent developments in electrochemical multiplex sensing of breast cancer biomarkers, with emphasis on the detection of new BC markers and new biomolecular modifications of sensing platforms (Table 1).
3.1 Duplex
Han et al. were able to prepare a set of DNA capture probes, each linked to one of two antibodies—epidermal growth factor receptor, EGFR or intercellular adhesion molecule-1, ICAM-1 (
Figure 8

Schematic illustration of in situ automatous DNA assembly for the discovery of dual therapeutic targets in breast cancer. Bing Han et al., Identification of dual therapeutic targets assisted by in situ automatous DNA assembly for combined therapy in breast cancer (
In the work completed by Hashkavayi et al., the researchers looked to fabricate a SPCE-based EpCAM and HER2 exosomal aptasensor with an added dual rolling circle amplification reaction initiator (an in-situ DNA amplification technique) thus allowing for single particle/ml detection limits (
Pothipor et al. looked to detect CA15-3 and miRNA-21 via the use of a poly(3-aminobenzylamine)/two-dimensional (2D) molybdenum selenide/ graphene oxide nanocomposite modified two-screen-printed carbon electrode array (
Shekari et al. fabricated a sandwich-type electrochemical as depicted in Figure 9 aptasensor for the simultaneous detection of CEA and CA 15-3 (
Figure 9

Schematic diagram of the different steps involved in the preparation of the sandwich–type aptasensor for the individual and simultaneous detection of CEA and CA 15–3 biomarkers (
3.2 Triplex devices
Electrochemistry can oftentimes be combined with a variety of varying sensing techniques to both increase the sensitivity of the prospective platform and allow for modularity of control over the platform parameters. Work published in 2020 by Cotchim et al. demonstrates the prototype of a multiplex electrochemically mediated immunosensor to detect breast cancer serum antigens CEA, CA153, and CA125 with anti-CEA, anti-CA153, and anti-CA125, respectively (
Pimalai et al. (
Kuntamung et al. were able to develop a label-free electrochemical breast cancer sensor that targeted MUC-1, CA 15-3, and HER2 (
Moura et al. leveraged the use of magnetic particles (MPs) to detect the presence of breast cancer exosomes in both lab-mixed and donor serums (
Figure 10

Confocal microscopy study for (A) MCF7 breast cancer cell lines and their corresponding exosomes covalently immobilized on MPs (exosomes-MPs), followed by indirect labelling with mouse antiCDX (being CDX either CD9, CD24, CD44, CD54, CD63, CD81, CD326 and CD340 biomarkers) and antimouse-Cy5. The concentration of exosomes was set at 4 × 109 per assay. The scale indicates the percentage of positive entities (cells and exosomes-coated MPs in panels A,B, respectively. Silio Lima Moura et al. Electrochemical immunosensing of nanovesicles as biomarkers for breast cancer (
The work completed by Zhang et al. targets the release of HER2-positive breast cancer cell (SK-BR-3) exosomes via the use of a multiplex CD63, HER2, and EpCAM aptasensor (
4 Conclusion
This perspective identifies several important areas of exploration for electrochemical diagnostics. With the prior developments summarized; we hope to answer the following questions:
- 1.
Are the limits of detection (LODs) reported for multiplex electrochemical biosensors of clinical relevance and how do they compare to well-established methods like ELISA, FISH, or PCR?
Yes, the LODs do display competitive sensitivity and specificity against the established methods. When exploring the variety of electrochemical platforms prototyped from a broad spectrum of molecular receptor capture probes (including nucleic acid aptamer and antibodies), these new electroanalytical methods demonstrate LODs at a similar range—or significantly lower—to conventional methods. For example, when comparing the LOD of 10−12–10−9 g/ml for HER2 biomarker measured by ELISA (in plasma serum, cell supernatant, or cell lysate) (taking as examples the Human ErbB2/HER2 ELISA Kit or MagVigen™—Human HER2 On-Bead ELISA Kit), the best performing HER2-responsive electrochemical multiplex platforms show a LOD of 0.5 ng/ml (
The LOD for micro-ribonucleic acid (miRNA) biomarkers were reported to be 9.79 × 10−16 M, 3.58 × 10−15 M, and 2.54 × 10−16 M for miRNA-155, miRNA-21, and miRNA-16, respectively (
Finally, the ELISA assay demonstrates LODs of 78–5,000 pg/ml for RANKL and 16–1,000 pg/ml for TNF, and low threshold detectable concentrations of 20 pg/ml and 5 pg/ml for RANKL and TNF, respectively. The reported LOD values of 2.6 and 3.0 pg/ml for RANKL and TNF, respectively, are comparatively superior as demonstrated by the electrochemical dual immunoassay platform (
- 2.
Can a single sensor electrode be used for the detection of multiple markers in a “one-blood-drop” fashion?
Not at the current state of development—particularly when nucleic acid aptamers are used as the sensor biorecognition elements. Many multiplex platforms use oligomers, such as aptamers, which are highly dynamic and versatile biorecognition elements that adopt a variety of 3-dimensional shapes and structures depending on the medium and matrix in which they are present. The sensitivity of aptamers to their environment and the variance in aptamer immobilization makes it challenging to achieve a uniform and steady state, sometimes leading to false positives caused by interferences. This is especially true after periods of electrode storage, and when interacting with high quantities of adjacent aptamer. Changes in the medium itself can additionally result in a more significant signal change than what might be caused by the target analyte. Even after more than twenty years of research on aptasensors, creating a single marketable electrochemical platform has proven to be a challenging endeavour. In addition to the complexity of aptamers, the target analyte biomolecules are also highly dynamic and exhibit an incredible range of functions and unique 3D conformations arising from their various primary, secondary, tertiary, and quaternary structures. These complex structures are often highly interdependent and are responsible for driving the molecule's function. Moreover, proteins exhibit complex inter- and intramolecular interactions that are difficult to study and quantify.
With this in mind, the most realistic solution for the electrochemical multiplex platform is to use electrode arrays in which each platform is engineered to detect a single analyte. Yet, it brings a question of cost and technical complexity. Fortunately, manufacturers of electrochemical accessories such as screen-printed electrode arrays (DropSense, ItalSense among many others) offer a good selection of array platforms that can be easily customized to adjust the number of markers being sensed. More importantly, these accessories are compatible with many electrochemical workstations allowing for simultaneous reading of electrical signals, all without sacrificing the size and adding to the cost of the electrochemical reader [MultiPalmSense4 or μStat-MultiX Multichannel (Bi)potentiostat/Galvanostat/Impedance Analyzer, MultiplEIS®]. Additionally, artificial intelligence-assisted machine learning algorithms are already allowing for simplified data processing (e.g., baseline correction, data standardization, and data compression) of electrochemical outputs. These tools will allow for the sensing of breast cancer biomarkers in real-time and could eventually help in delivering a POC single blood-drop analysis (
,
). With this in mind, as soon as the receptor-analyte affinity and binding are optimized, these multiplex biosensors have the potential to be very successful POC devices for the early screening of BC.
- 3.
What mechanisms of signal amplifications are most promising and what technological advancements are needed to utilize these devices for multiplex POC detection?
Any biorecognition element coupled with an amplification method, such as a dual rolling circle amplification (
), magnetic beads, metal nanoparticle, or enzymatic labels, has shown high potential for BC biomarker diagnoses and can allow for very low detection limits. However, these methods often require a multistep process for sample analysis; thus, they are more difficult to establish and are time-consuming. Though not as effective as molecular amplification, an appropriate and cost-effective alternative is magnetic enrichment which can be used for moderate levels of signal amplification.
- 4.
Can nanotechnology advance the sensitive and selective diagnostics of multiple BC biomarkers?
Nanomaterials-based electrochemical sensing
It is readily apparent that implementing nanomaterials in the design of a sensing platform has several benefits. By implementing changes to the electrode features themselves, nanomaterials (metal and carbon nanoparticles and quantum dots) provide an increased active surface area and improved electrical conductivity, and thus an improved charge transfer ability of the platform. This is of particular note when high quantities of poor-conducting biological species are immobilized on the sensor platform. Furthermore, as demonstrated in the works surveyed in this review, metal nanoparticles are often used as redox responders and can be easily conjugated with biomolecules, such as gold self-assemblies via thiol motifs or through the coupling of the carbon functionalities with amines via EDC/NHS coupling. H. Nasrollahpour and colleagues (
Micro-and nano-biosensors, microfluidics devices
The development of micro- and nano-biosensors gives us a unique lens into the possible future of cancer screening, diagnostics, and in-real-time treatment management. For instance, in microfluidic devices can take small quantities of human serum and pump them through a small array of tubes (
Without a doubt, non-specific binding of species present in biological fluids (serum, blood, plasma) is the main challenge that must be overcome to transfer electrochemical biosensors to the marketable product. The concentration of these interfering biomolecules is several orders of magnitude higher than that of target analytes. These non-target molecules simply adhere to the sensor surface without any specific receptor, thereby severely lowering the sensitivity and specificity of the assay. This problem must be solved before any further progress can be made towards clinical testing. One interesting concept that seeks for solution to the non-specific adsorption of proteins in a complex medium is referred to as nanoshearing (
,
). Aided by an alternating (AC) electric field, shear forces are generated near the electrode surface. Via manipulation of AC field and thus the vector of these forces, the selectivity of the assay can be improved. Specifically, selective displacement of weakly (non-specifically) bound proteins from the electrode surface is possible by tuning the fluid shear forces. Using this approach, the detection of HER2 BC marker and other multiple protein targets spiked in human serum can be significantly improved (1,000-fold sensitivity enhancement) in comparison to other flow-based assays for protein biomarker detection.
- 5.
Are there preferred receptors (antibody, nucleic acid or their combinations) and preferred biosensor designs (complementary methods, sandwich-type protocols, and antibody/aptamer concept, label-free protocol) that can aid in clearing the roadmap for early diagnosis of multiple BC markers?
In this review, we have surveyed different approaches, which include using diverse techniques, bioreceptors, bioassays, electrode types, and signal amplification strategies applied towards the robust sensing of BC. These methods range from simple, label-free tests to complex multistep procedures, but each provides attractive approaches that enable better biomarker recognition. However, it is not yet clear which of these approaches are reliable for sensitivity and selectivity since many of the models are not validated by others, nor are they tested clinically. More research needs to be carried out to cross-reference the fabrication and analytical response with minimum error for these multianalyte biosensors. Furthermore, there exists new innovative—but more expensive methods that have not yet been explored for multi-marker breast cancer detection. This includes the use of Slow Off-Rate Modified Aptamers (SOMAmers) as biorecognition elements, which could potentially improve the selectivity of the biorecognition element. Monomeric or zwitterionic peptides used as bio-recognition elements or electrode modifiers could effectively prevent unfavourable binding of non-specific proteins and cells. Several biosensing platforms which have been tested in human serum have been reported using the remarkable antifouling ability of peptides by Luo and co-workers with promising results (
–
). Even with these advancements, other technologies such as CRISPR-based electrochemical biosensing could lead to further advancement in BC and cancer recognition. The simultaneous detection of 37 RNAs with 95.2% specificity and 90.0% sensitivity from human serum of early-stage cell lung cancer patients has been shown by Sheng and colleagues (
). This approach could evolve as a next-generation diagnostic tool for screening serum BC biomarkers.
- 6.
Why are we still without FDA-approved electrochemical multiplex devices for BC screening?
There is only one FDA-approved, PCR-based multiplex testing kit, manufactured by Myriad Genetic Laboratories (BRACAnalysis CDx), which specifically targets BRCA1 and BRCA2 markers for breast, ovarian, and pancreatic cancers (
https://myriad.com/genetic-tests/bracanalysiscdx-germline-test/). If the above-discussed issues for the multiplex e-sensors are solved, these devices have the potential to meet the POC criteria, resulting in efficient, rapid, and smart multiplex serological tests for BC markers only.
In addition, to make strides towards commercialization and significant research advancement we must see a change in the reporting culture in sensing manuscripts. Currently, there is an overarching failure by researchers to utilize and report fundamental statistical concepts in every step of biosensor design and fabrication including the omission of error calculations; a lack of significant tests that demonstrate sensor validity and viability such as baseline charge-signal drift and the effects of long-term device storage on signal transduction; and a lack of a robust statistical design of experiments. As an example, the effectiveness of aptamer-based biosensors can significantly depend on experimental conditions such as salt concentration, buffer, and pH value, each of which has been found to affect the selectivity, affinity, and 3D structure of aptamers. Despite this, high-resolution structures of aptamer complexes have been determined under different experimental conditions with no apparent relationship between them. The DoE approach considers all factors simultaneously and provides an empirical correlation between these factors on the chosen response, unlike the conventional method of changing one factor at a time. The evaluation of experimental variables, stability of the assay, selectivity towards all possible interfering species, and finally testing in complex matrix (blood, serum, liquid biopsy) verified by both a newly developed electrochemical assay and tested against the gold standard of analytics (ELISA, PCR, FISH, etc.) are direly needed if these technologies are to make an impact in real-world settings.
Statements
Author contributions
CO: Writing – original draft. CK: Writing – review & editing, Investigation. SA: Writing – review & editing. AI: Writing – review & editing, Conceptualization, Formal Analysis, Funding acquisition, Project administration, Resources, Supervision, Visualization.
Funding
The author(s) declare financial support was received for the research, authorship, and/or publication of this article.
This work was carried out with financial support of the New Brunswick Innovation Foundation NBIF (Project No. EP-0000000155) & New Brunswick Health Research Fund NBHRF (Project No. 0000000045).
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1.
DeSantisCEBrayFFerlayJLortet-TieulentJAndersonBOJemalA. International variation in female breast cancer incidence and mortality rates. Cancer Epidemiol Biomarkers Prev. (2015) 24(10):1495–506. 10.1158/1055-9965.EPI-15-0535
2.
TanneJH. Breast cancer is overdiagnosed in one in six or seven cases, finds large US study. Br Med J. (2022) 376:o581. 10.1136/bmj.o581
3.
da Costa VieiraRABillerGUemuraGRuizCACuradoMP. Breast cancer screening in developing countries. Clinics (Sao Paulo). (2017) 72(4):244–53. 10.6061/clinics/2017(04)09
4.
NelsonHDPappasMCantorAGriffinJDaegesMHumphreyL. Harms of breast cancer screening: systematic review to update the 2009 U.S. Preventive services task force recommendation. Ann Intern Med. (2016) 164(4):256–67. 10.7326/M15-0970
5.
BrittKLCuzickJPhillipsKA. Key steps for effective breast cancer prevention. Nat Rev Cancer. (2020) 20(8):417–36. 10.1038/s41568-020-0266-x
6.
MomenimovahedZSalehiniyaH. Epidemiological characteristics of and risk factors for breast cancer in the world. Breast Cancer (Dove Med Press). (2019) 11:151–64. 10.2147/BCTT.S176070
7.
MyersERMoormanPGierischJMHavrileskyLJGrimmLJGhateSet alBenefits and Harms of breast cancer screening: a systematic review. JAMA. (2015) 314:1615–34. 10.1001/jama.2015.13183
8.
NaginiS. Breast cancer: current molecular therapeutic targets and new players. Anticancer Agents Med Chem. (2017) 17:152–63. 10.2174/1871520616666160502122724
9.
KudelaESamecMKubatkaPNachajovaMLaucekovaZLiskovaAet alBreast cancer in young women: status quo and advanced disease management by a predictive, preventive, and personalized approach. Cancers (Basel). (2019) 11:1791. 10.3390/cancers11111791
10.
DesreuxJAC. Breast cancer screening in young women. Eur J Obstet Gynecol Reprod Biol. (2018) 230:208–11. 10.1016/j.ejogrb.2018.05.018
11.
GinsburgOYipCHBrooksACabanesACaleffiMDunstan YatacoJAet alBreast cancer early detection: a phased approach to implementation. Cancer. (2020) 126 Suppl 10(Suppl 10):2379–93. 10.1002/cncr.32887
12.
HongPLiWLiJ. Applications of aptasensors in clinical diagnostics. Sensors. (2012) 12:1181–93. 10.3390/s120201181
13.
Díaz-FernándezALorenzo-GómezRMiranda-CastroRde-los-Santos-ÁlvarezNLobo-CastañónMJ. Electrochemical aptasensors for cancer diagnosis in biological fluids—a review. Anal Chim Acta. (2020) 1124:1–19. 10.1016/j.aca.2020.04.022
14.
LakheraPChaudharyVJhaASinghRKushPKumarP. Recent developments and fabrication of the different electrochemical biosensors based on modified screen printed and glassy carbon electrodes for the early diagnosis of diverse breast cancer biomarkers. Mater Today Chem. (2022) 26:101129. 10.1016/j.mtchem.2022.101129
15.
LiuLSWangFGeYLoPK. Recent developments in aptasensors for diagnostic applications. ACS Appl Mater Interfaces. (2021) 13:9329–58. 10.1021/acsami.0c14788
16.
HarbeckNPenault-LlorcaFCortesJGnantMHoussamiNPoortmansPet alBreast cancer. Nat Rev Dis Primers. (2019) 5(1):66. 10.1038/s41572-019-0111-2
17.
SharafeldinMRuslingJF. Multiplexed electrochemical assays for clinical applications. Curr Opin Electrochem. (2023) 39:101256. 10.1016/j.coelec.2023.101256
18.
MostafaIMTianYAnjumSHanifSHosseiniMLouBet alComprehensive review on the electrochemical biosensors of different breast cancer biomarkers. Sens Actuators B Chem. (2022) 365:131944. .1016/j.snb.2022.131944
19.
JingLXieCLiQYangMLiSLiHet alElectrochemical biosensors for the analysis of breast cancer biomarkers: from design to application. Anal Chem. (2022) 94(1):269–96. 10.1021/acs.analchem.1c04475
20.
ZubairMWangSAliN. Advanced approaches to breast cancer classification and diagnosis. Front Pharmacol. (2021) 11:632079. 10.3389/fphar.2020.632079
21.
ZhaoWLiXWangWChenBWangLZhangNet alAssociation of preoperative serum levels of CEA and CA15-3 with molecular subtypes of breast cancer. Dis Markers. (2021) 2021:5529106. 10.1155/2021/5529106
22.
YersalOBarutcaS. Biological subtypes of breast cancer: prognostic and therapeutic implications. World J Clin Oncol. (2014) 5:412–24. 10.5306/wjco.v5.i3.412
23.
SchickJRitchieRPRestiniC. Breast cancer therapeutics and biomarkers: past, present, and future approaches. Breast Cancer (Auckl). (2021) 15:1178223421995854. PMID: . PMCID: PMC8100889.
24.
AfzalSHassanMUllahSAbbasHTawakkalFKhanMA. Breast cancer; discovery of novel diagnostic biomarkers, drug resistance, and therapeutic implications. Front Mol Biosci. (2022) 9:783450. 10.3389/fmolb.2022.783450
25.
TarighatiEKeivanHMahaniH. A review of prognostic and predictive biomarkers in breast cancer. Clin Exp Med. (2023) 23(1):1–16. 10.1007/s10238-021-00781-1
26.
Rey-VargasLMejía-HenaoJCSanabria-SalasMCSerrano-GomezSJ. Effect of neoadjuvant therapy on breast cancer biomarker profile. BMC Cancer. (2020) 20(1):675. 10.1186/s12885-020-07179-4
27.
NasrollahpourHKhalilzadehBHasanzadehMRahbarghaziREstrelaPNaseriAet alNanotechnology-based electrochemical biosensors for monitoring breast cancer biomarkers. Med Res Rev. (2023) 43(3):464–569. 10.1002/med.21931
28.
MarquesRCVBCosta-RamaEViswanathanSNouwsHPACosta-GarciaADelerue-MatosCet alVoltammetric immunosensor for the simultaneous analysis of the breast cancer biomarkers CA 15-3 and HER2-ECD. Sens Actuators B Chem. (2018) 255:918–25. 10.1016/j.snb.2017.08.107
29.
ValverdeASerafinVGarozJMontero-CalleAGonzalez-CortesAArenasMet alElectrochemical immunoplatform to improve the reliability of breast cancer diagnosis through the simultaneous determination of RANKL and TNF in serum. Sens Actuators B Chem. (2020) 314:128096. 10.1016/j.snb.2020.128096
30.
GuoQLiXShenCZhangSQiHLiTet alElectrochemical immunoassay for the protein biomarker mucin 1 and for MCF-7 cancer cells based on signal enhancement by silver nanoclusters. Microchim Acta. (2015) 182:1483–9. 10.1007/s00604-015-1471-2
31.
YazdanparastSBenvidiABanaeiMNukukarHTezerjaniMDAzimzadehM. Dual-aptamer based electrochemical sandwich biosensor for MCF-7 human breast cancer cells using silver nanoparticle labels and a poly(glutamic acid)/MWNT nanocomposite. Microchim Acta. (2018) 185:405. 10.1007/s00604-018-2918-z
32.
Johari-AharMKaramiPGhaneiMAfkhamiABagheriH. Development of a molecularly imprinted polymer tailored on disposable screen-printed electrodes for dual detection of EGFR and VEGF using nano-liposomal amplification strategy. Biosens Bioelectron. (2018) 107:26–33. 10.1016/j.bios.2018.02.005
33.
WuJYanYYanFJuH. Electric field-driven strategy for multiplexed detection of protein biomarkers using a disposable reagentless electrochemical immunosensor array. Anal Chem. (2008) 80:6072–7. 10.1021/ac800905k
34.
BerglundGICarlssonGHSmithATSzökeHHenriksenAHajduJ. The catalytic pathway of horseradish peroxidase at high resolution. Nature. (2002) 417(6887):463–8. 10.1038/417463a
35.
CuiZWuDZhangYMaHLiHDuBet alUltrasensitive electrochemical immunosensors for multiplexed determination using mesoporous platinum nanoparticles as nonenzymatic labels. Anal Chim Acta. (2014) 807:44–50. 10.1016/j.aca.2013.11.025
36.
HanBShaLYuXYangMCaoYZhaoJ. Identification of dual therapeutic targets assisted by in situ automatous DNA assembly for combined therapy in breast cancer. Biosens Bioelectron. (2021) 176:112913. 10.1016/j.bios.2020.112913
37.
ZhuJBoškovićFNguyenB-NTNitschkeJRKeyserUF. Kinetics of toehold-mediated DNA strand displacement depend on FeII4L4 tetrahedron concentration. Nano Lett. (2021) 21:1368–74. 10.1021/acs.nanolett.0c04125
38.
HashkavayiABChaBSLeeESParkKS. Dual rolling circle amplification-enabled ultrasensitive multiplex detection of exosome biomarkers using electrochemical aptasensors. Anal Chim Acta. (2022) 1205:339762. 10.1016/j.aca.2022.339762
39.
PothiporCBamrungsapSJakmuneeJOunnunkadK. A gold nanoparticle-dye/poly(3-aminobenzylamine)/two dimensional MoSe2/graphene oxide electrode towards label-free electrochemical biosensor for simultaneous dual-mode detection of cancer antigen 15-3 and microRNA-21. Colloids Surf B Biointerfaces. (2022) 210:112260. 10.1016/j.colsurfb.2021.112260
40.
ShekariZZareHRFalahatiA. Dual assaying of breast cancer biomarkers by using a sandwich–type electrochemical aptasensor based on a gold nanoparticles–3D graphene hydrogel nanocomposite and redox probes labeled aptamers. Sens Actuators B Chem. (2021) 332:129515. 10.1016/j.snb.2021.129515
41.
CotchimSThavarungkulPKanatharanaPLimbutW. Multiplexed label-free electrochemical immunosensor for breast cancer precision medicine. Anal Chim Acta. (2020) 1130:60–71. 10.1016/j.aca.2020.07.021
42.
PimalaiDPutninTWaiwinyaWChotsuwanCAroonyadetNJaprungD. Development of electrochemical biosensors for simultaneous multiplex detection of microRNA for breast cancer screening. Mikrochim Acta. (2021) 188(10):329. 10.1007/s00604-021-04995-8
43.
KuntamungKJakmuneeJOunnunkadK. A label-free multiplex electrochemical biosensor for the detection of three breast cancer biomarker proteins employing dye/metal ion-loaded and antibody-conjugated polyethyleneimine-gold nanoparticles. J Mater Chem B. (2021) 9:6576–85. 10.1039/d1tb00940k
44.
MouraSLMartínCGMartíMPividoriMI. Electrochemical immunosensing of nanovesicles as biomarkers for breast cancer. Biosens Bioelectron. (2020) 150:111882. 10.1016/j.bios.2019.111882
45.
ZhangMXiaLMeiWZouQLiuHWangHet alOne-step multiplex analysis of breast cancer exosomes using an electrochemical strategy assisted by gold nanoparticles. Anal Chim Acta. (2023) 1254:341130. 10.1016/j.aca.2023.341130
46.
ChengJPYanYWangXYLuYLYuanYHJiaJet alMUC1-positive Circulating tumor cells and MUC1 protein predict chemotherapeutic efficacy in the treatment of metastatic breast cancer. Chin J Cancer. (2011) 30:54–61. 10.5732/cjc.010.10239
47.
TomlinsonIPMWhymanABarrettJAKremerJK. Tumour marker CA15-3: possible uses in the routine management of breast cancer. Eur J Cancer. (1995) 31:899–902. 10.1016/0959-8049(94)00447-1
48.
WangKYuanYChoJHMcClartySBaxterDGalasDJ. Comparing the MicroRNA spectrum between serum and plasma. PLoS One. (2012) 7(7):e41561. 10.1371/journal.pone.0041561
49.
LiJLiYChenSDuanWKongXWangYet alHighly sensitive exosome detection for early diagnosis of pancreatic cancer using immunoassay based on hierarchical surface-enhanced Raman scattering substrate. Small Methods. (2022) 6(6):e2200154. 10.1002/smtd.202200154
50.
TangQXiaoXLiRHeHLiSMaC. Recent advances in detection for breast-cancer-derived exosomes. Molecules. (2022) 27(19):6673. 10.3390/molecules27196673
51.
SinhaKUddinZKawsarHIIslamSDeenMHHowladerMMR. Analyzing chronic disease biomarkers using electrochemical sensors and artificial neural networks. TrAC, Trends Anal Chem. (2023) 158:116861. 10.1016/j.trac.2022.116861
52.
AmethiyaYPipariyaPPatelSShahM. Comparative analysis of breast cancer detection using machine learning and biosensors. Intel Med. (2022) 2:69–81. 10.1016/j.imed.2021.08.004
53.
WangYGaoWSunMFengBShenHZhuJet alA filter-electrochemical microfluidic chip for multiple surface protein analysis of exosomes to detect and classify breast cancer. Biosens Bioelectron. (2023) 239:115590. 10.1016/j.bios.2023.115590
54.
VaidyanathanRvan LeeuwenLMRaufSShiddikyMJATrauM. A multiplexed device based on tunable nanoshearing for specific detection of multiple protein biomarkers in serum. Sci Rep. (2015) 5:9756. 10.1038/srep09756
55.
ShiddikyMJAVaidyanathanRRaufSTayZTrauM. Molecular nanoshearing: an innovative approach to shear off molecules with AC-induced nanoscopic fluid flow. Sci Rep. (2014) 4:3716. 10.1038/srep03716
56.
XuYWangXDingCLuoX. Ratiometric antifouling electrochemical biosensors based on multifunctional peptides and MXene loaded with au nanoparticles and methylene blue. ACS Appl Mater Interfaces. (2021) 13:20388–96. 10.1021/acsami.1c04933
57.
YangXChenMZhangZLiYWangPLuoXet alAlpha-aminoisobutyric acid incorporated peptides for the construction of electrochemical biosensors with high stability and low fouling in serum. Anal Chim Acta. (2023) 1238:340646. 10.1016/j.aca.2022.340646
58.
ShengYZhangTZhangSJohnstonMZhengXShanYet alA CRISPR/Cas13a-powered catalytic electrochemical biosensor for successive and highly sensitive RNA diagnostics. Biosens Bioelectron. (2021) 178:113027. 10.1016/j.bios.2021.113027
Summary
Keywords
breast cancer, serum biomarker, multiplex, electrochemical biosensor, point-of-care POC
Citation
O’Brien C, Khor CK, Ardalan S and Ignaszak A (2024) Multiplex electrochemical sensing platforms for the detection of breast cancer biomarkers. Front. Med. Technol. 6:1360510. doi: 10.3389/fmedt.2024.1360510
Received
23 December 2023
Accepted
05 February 2024
Published
15 February 2024
Volume
6 - 2024
Edited by
William Facchinatto, University of Aveiro, Portugal
Reviewed by
Arunadevi Natarajan, Bharathiar University, India
Max M. Gong, Trine University, United States
Saroj Kumar, Apeejay Stya University, India
Updates

Check for updates
Copyright
© 2024 O'Brien, Khor, Ardalan and Ignaszak.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Anna Ignaszak aignasza@unb.ca
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.