Abstract
Unfractionated heparin remains the standard anticoagulant during cardiopulmonary bypass (CPB) and extracorporeal membrane oxygenation (ECMO). Despite decades of clinical use, however, no universally validated reference method exists for measuring biologically active unfractionated heparin in these settings. Conventional assays such as the activated clotting time and the activated partial thromboplastin time lack specificity and standardisation. Consequently, anti–factor Xa (anti-Xa) assays are increasingly recommended as alternative methods (or even reference methods). Nevertheless, their validity as a true reference standard in CPB and ECMO has never been formally established. This situation creates a methodological paradox: an assay with context-dependent limitations is used to calibrate and validate other imperfect tests. Anti-Xa assays differ substantially in design, particularly regarding the inclusion of dextran sulphate. Dextran sulphate was introduced to prevent ex vivo neutralisation of heparin by platelet factor 4. However, platelet factor 4 release also occurs in vivo during extracorporeal circulation. Depending on assay configuration and clinical context, anti-Xa measurements may therefore overestimate or underestimate biologically active heparin—especially at low concentrations or following protamine reversal. As a result, anti-Xa assays cannot be assumed to be interchangeable in CPB and ECMO. We argue that heparin monitoring in extracorporeal circulation requires conceptual re-evaluation. Distinct clinical scenarios must be differentiated, and rigorous validation studies are essential before any assay can be regarded as a true reference standard.
Introduction
Unfractionated heparin (UFH) is a heterogeneous, sulphated glycosaminoglycan anticoagulant that exerts its effect primarily by potentiating antithrombin, thereby accelerating inhibition of thrombin and factor Xa (1). In cardiac surgery with cardiopulmonary bypass (CPB), UFH is the standard systemic anticoagulant because blood contact with the extracorporeal circuit activates coagulation, and UFH provides rapid, titratable anticoagulation that can be reversed with protamine at the end of bypass (2). For extracorporeal membrane oxygenation (ECMO), UFH is likewise widely used to mitigate circuit- and patient-related thrombotic complications during prolonged extracorporeal support (3).
UFH administration usually includes monitoring and dose adaptation because the required doses in CPB and ECMO show marked interpatient variability, and both under- and over-anticoagulation can lead to clinical complications (2). During CPB, monitoring is commonly performed using the activated clotting time (ACT) in a near-patient setting, sometimes supplemented by heparin concentration–based or anti–factor Xa (anti-Xa)–based approaches (2). In ECMO, either the activated partial thromboplastin time (aPTT), the ACT, or anti-Xa analysis is performed (4).
The limitations of the aPTT and ACT as heparin-monitoring assays are well known and generally accepted. Both assays show poor correlation with actual UFH concentrations as determined by anti-Xa analysis (5–9). This is not surprising given the multitude of factors involved in these assays (contact phase; clotting factors XII, XI, IX, VIII, X, V, and II); therefore, the assay reaction is relatively unspecific for the drug to be tested (heparin). In addition, standardisation of these assays is poor: different aPTT reagents have different sensitivities to UFH, and whole-blood point-of-care aPTT assays do not agree well with laboratory aPTT tests (10–12). Different ACT assays are likewise not interchangeable (6, 13–15). Thus, both assays are analytically weak tools for monitoring UFH and poorly standardised. On the positive side, they are widely available, rapid, and inexpensive.
Viscoelastic testing (VET) is another whole-blood–based coagulation test method that has been proposed as a tool to monitor unfractionated heparin (16–18). The VET assays used for this purpose (CK on TEG, in-tem on ROTEM, and IN-test on ClotPro) are—similar to aPTT and ACT—triggered via the contact phase and thus involve factors XII, XI, IX, VIII, X, V, and II, which renders them relatively unspecific for heparin. In several publications (16, 17), results are expressed as ratios of clotting times before and after inactivation of heparin by heparinase, which may improve assay performance. However, overall clinical experience with these VET-based approaches in ECMO and cardiac surgery remains limited, and current review recommendations are cautious regarding their use in these settings (19, 20). As noted, VET assays are activated via the intrinsic pathway and therefore remain susceptible to the same confounding factors that affect aPTT and ACT, including contact activation, factor deficiencies, haemodilution, and inflammatory alterations. Consequently, their ability to reliably quantify residual heparin effects—especially in the setting of protamine reversal—may be constrained.
Considering these limitations, anti-Xa testing—although more time-consuming and restricted to laboratory-based analysis—has been recommended as an alternative method for heparin monitoring in ECMO and, in selected cases, during cardiac surgery with CPB (2, 21). Given that UFH represents a heterogeneous mixture of sulphated glycosaminoglycan chains with variable lengths and molecular weights (1), direct measurement of the drug (e.g., by mass spectrometry) is not feasible as a reference method.
The 2022 ELSO Adult and Paediatric Anticoagulation Guidelines acknowledge several limitations of anti-Xa assays, including limited standardisation among commercially available tests, interference by dextran sulphate, and potential analytical sources of error. Nevertheless, they provide example target ranges based on anti-Xa activity (21). By contrast, the 2023 ISTH/SSC guideline on anticoagulation in adult patients supported by ECMO does not explicitly address these methodological limitations (22). Similarly, the STS/SCA/AmSECT Clinical Practice Guidelines on anticoagulation during CPB recommend anti-Xa testing as a validating assay for point-of-care tests reflecting anti-Xa activity, yet do not specify how anti-Xa testing itself was validated for this purpose in the CPB setting (2).
Despite these recommendations, a fundamental question remains insufficiently addressed: can anti-Xa assays be considered a validated reference standard for UFH monitoring in CPB and ECMO? If an assay with context-dependent methodological limitations is used to calibrate other imperfect tests, a conceptual circularity arises. This Perspective examines whether current monitoring strategies rest on sufficiently validated premises and outlines a framework for re-evaluating UFH monitoring in extracorporeal circulation.
Discussion
Are recommendations to use anti-Xa testing in cardiac surgery and ECMO justified?
A fundamental principle of epistemology is that conclusions cannot be more robust than the premises on which they are based. Assigning reference status to an assay that has not been adequately validated for a specific clinical context is problematic, particularly when subsequent recommendations and target ranges are derived from it.
Anti-Xa assays were not originally developed for monitoring UFH during cardiac surgery or ECMO, and several uncertainties remain regarding their suitability in these settings (23). Although anti-Xa assays are generally considered better standardised than aPTT or ACT (24), inter-assay variability between commercially available tests has been documented (25, 26). A key determinant of assay behaviour is the use of dextran sulphate (24), which is included in some reagents (HemosIL® Liquid Anti-Xa by Werfen, INNOVANCE® Heparin by Siemens, and BIOPHEN™ Heparin LRT by Hyphen) but not in others (STA® Liquid Anti-Xa by Stago, Technochrom® Anti-Xa by Technoclone).
The rationale for introducing dextran sulphate into certain anti-Xa assays was to prevent ex vivo inhibition of heparin during sample handling (24, 27). During blood transport, storage and centrifugation to obtain platelet-poor plasma, platelet activation may occur, resulting in the release of platelet factor 4 (PF4). PF4 is a platelet-derived protein capable of binding to and neutralising the biological activity of heparin (28). Dextran sulphate was therefore added to some anti-Xa reagents to displace heparin from PF4 and to avoid falsely low anti-Xa results caused by ex vivo heparin neutralisation (24, 27).
However, the use of dextran sulphate in anti-Xa testing has recently been questioned (23, 29), particularly because it may lead to overestimation of heparin concentrations, e.g., after protamine administration (29, 30). Importantly, PF4 release is not limited to ex vivo sample handling. Platelet activation with subsequent PF4 release occurs in vivo during CPB and ECMO support (31, 32). In these settings, dextran sulphate–containing anti-Xa assays may liberate heparin that has already been neutralised in vivo by PF4, or subsequently by protamine, and render it measurable in vitro. This mechanism may result in falsely elevated anti-Xa levels that do not necessarily reflect biologically active circulating heparin.
Conversely, anti-Xa assays that do not contain dextran sulphate may be susceptible to ex vivo PF4-mediated heparin neutralisation during sample transport and processing. This may lead to falsely low anti-Xa results, particularly when pneumatic tube systems, centrifugation delays, or prolonged pre-analytical intervals are involved. Thus, depending on assay composition, anti-Xa testing in cardiac surgery and ECMO may be prone either to overestimation or underestimation of clinically relevant heparin activity. Switching to currently available point-of-care assays such as ACT, whole-blood aPTT, or VET does not appear to resolve these fundamental methodological problems, for the reasons outlined above.
A mechanistic comparison of the assays applied for the monitoring of heparin is shown in Table 1.
Table 1
| Monitoring method | POC | Specificity for heparin | Sensitivity for heparin | Interference with protamine | Inactivation of heparin during sample transport and preanalytics |
|---|---|---|---|---|---|
| Anti-Xa with DS | No | High2 | High | Potentially yes6 | Unlikely (DS) |
| Anti-Xa without DS | No | Potentially yes7 | |||
| aPTT | Mostly no1 | Low3 | Moderate | Yes (prolongation when protamine is present in excess) | |
| ACT | Yes | Moderate/low 4 | No (POC) | ||
| VET in-tem/IN-test | Moderate | ||||
| VET CK (TEG) | High5 |
Mechanistic comparison of assays used for the monitoring of UFH.
1POC aPTT tests exist, most aPTT tests are however laboratory-based.
2Detection of UFH action via the inhibition of FXa or FIIa (thrombin).
3Detection of UFH action in an assay reaction mediated via factors XII, XI, IX, VIII, X, V, II.
4The ACT is usually optimized for a wide measuring range for the detection of the high heparin doses applied in cardiac surgery.
5The CK assay uses a much weaker intrinsic activation as compared to aPTT or in-tem/IN-test.
6Several publications claim a release of protamine-bound heparin by dextrane sulphate.
7Platelet activation during sample transport, centrifugation and waiting periods could lead to release of PF4 and subsequent ex vivo heparin inactivation.
What are the consequences?
If anti-Xa assays are to be used as reference methods in cardiac surgery and ECMO, their methodological limitations must be explicitly acknowledged.
First, no universally validated reference assay currently exists for the assessment of UFH activity during and after CPB or ECMO support. Second, pending appropriately designed comparative studies using samples derived specifically from CPB and / or ECMO patients, different anti-Xa assays cannot be considered interchangeable in these settings. Reported target ranges and inter-study comparisons must therefore always be interpreted considering the specific reagent and assay configuration applied. Third, the impact of dextran sulphate on anti-Xa measurements in cardiac surgery and ECMO requires systematic investigation, particularly in the context of protamine reversal and low residual heparin concentrations.
In theory, a potential reference strategy could involve the use of an anti-Xa assay without dextran sulphate in combination with blood collection in citrate–theophylline–adenosine–dipyridamole (CTAD) tubes, which may reduce ex vivo platelet activation and PF4 release (33). However, such an approach would itself require formal validation within the specific clinical contexts of CPB and ECMO before it could be considered a reliable standard.
Distinguishing clinical scenarios
Importantly, monitoring strategies for UFH cannot be discussed as a single homogeneous problem. At least four distinct clinical scenarios should be differentiated:
Standard-dose UFH outside extracorporeal circulation (e.g., treatment of deep vein thrombosis):
In this setting, significant in vivo PF4 release is unlikely and protamine is not administered. Dextran sulphate may therefore be appropriate because PF4-bound heparin would predominantly reflect ex vivo phenomena rather than in vivo (“true”) neutralisation.
High-dose UFH during CPB:
Here, very high heparin concentrations are present and in vivo PF4 release occurs. Given the magnitude of circulating heparin levels, the quantitative impact of dextran sulphate–mediated heparin release is likely to be negligible in relation to the total measurable activity.
Low- to moderate-dose UFH during ECMO:
In this scenario, in vivo PF4 release is expected, while heparin concentrations are substantially lower than during CPB. Under these conditions, dextran sulphate–mediated displacement of PF4-bound heparin may meaningfully influence (and potentially distort) measured anti-Xa levels.
Detection of residual UFH after protamine administration:
The suitability of the different monitoring methods for these clinical scenarios is summarized in Table 2.
Table 2
| Monitoring method | i.v. Heparin without bypass (e.g., ICU) | Cardiac surgery (before protamine) | Cardiac surgery (after protamine) | ECMO |
|---|---|---|---|---|
| Anti-Xa with DS | ++ DS reducing preanalytical errors due to heparin inactivated ex vivo | o POC ACT is an established procedure for this use, anti-Xa is mainly proposed as a tool for validating the ACT method applied, due to time delay and cost | o Interpretation may be limited due to potential interference of dextran sulphate with protamine–heparin and PF4–heparin complexes formed in vivo | o interpretation may be limited due to potential interference of dextran sulphate with protamine–heparin and PF4–heparin complexes formed in vivo |
| Anti-Xa without DS | + Potential inactivation of heparin during preanalytics | + potential inactivation of heparin during preanalytics | ||
| ACT | o Weak dose–response for heparin | + Established procedure | o Due to low expected residual heparin doses the assay specificity/sensitivity for heparin might be suboptimal | o the assay specificity/sensitivity for heparin might be suboptimal in ECMO |
| aPTT | − No clotting due to high heparin concentrations | |||
| VET in-tem/IN-test | ||||
| VET CK (TEG) | ||||
Suitability of heparin monitoring methods for different clinical scenarios.
DS, dextran sulphate; POC, point of care; PF4, platelet factor 4; ACT, activated clotting time; VET, viscoelastic testing; ECMO, extracorporeal membrane oxygenation.
++ = highly suitable; + = suitable with limitations; o = limited suitability; − = not suitable.
A research agenda
For each of these clinical indications, three fundamental steps are required:
Definition of an appropriate reference method for UFH measurement
Establishment of clinically meaningful acceptance criteria
Validation of candidate monitoring tools (including point-of-care assays) against these predefined standards
Once such suited heparin monitoring methods have been characterized, their results would have to be investigated in relation to the clinical course including thrombotic or hemorrhagic events, with particular emphasis on their performance during prolonged CPB or ECMO, conditions which are known to be associated with a progressing activation of coagulation.
This effort will require coordinated, multicenter clinical studies and close collaboration across disciplines. Six decades after the introduction of the ACT (34), reliable monitoring of UFH in extracorporeal circulation remains an unresolved challenge.
Statements
Data availability statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.
Author contributions
CW: Conceptualization, Investigation, Writing – original draft, Writing – review & editing. KZ: Conceptualization, Writing – review & editing. AC: Conceptualization, Investigation, Writing – review & editing. ML: Writing – review & editing.
Funding
The author(s) declared that financial support was not received for this work and/or its publication.
Conflict of interest
AC is a consultant to Apiro Diagnostics Kft, Hungary, a manufacturer of viscoelastometry devices including assays sensitive to unfractionated heparin.
The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that Generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1.
HirshJBauerKADonatiMBGouldMSamamaMMWeitzJI. Parenteral anticoagulants: American College of Chest Physicians evidence-based clinical practice guidelines (8th edition). Chest. (2008) 133:141S–59S. doi: 10.1378/chest.08-0689,
2.
Shore-LessersonLBakerRAFerrarisVGreilichPEFitzgeraldDRomanPet al. STS/SCA/AmSECT clinical practice guidelines: anticoagulation during cardiopulmonary bypass. J Extra Corpor Technol. (2018) 50:5–18. doi: 10.1051/ject/201850005
3.
ChlebowskiMMBaltagiSCarlsonMLevyJHSpinellaPC. Clinical controversies in anticoagulation monitoring and antithrombin supplementation for ECMO. Crit Care. (2020) 24:19. doi: 10.1186/s13054-020-2726-9,
4.
ProttiAIapichinoGEDi NardoMPanigadaMGattinoniL. Anticoagulation management and antithrombin supplementation practice during veno-venous extracorporeal membrane oxygenation: a worldwide survey. Anesthesiology. (2020) 132:562–70. doi: 10.1097/ALN.0000000000003044,
5.
DelmasCJacqueminAVardon-BounesFGeorgesBGuerreroFHernandezNet al. Anticoagulation monitoring under ECMO support: a comparative study between the activated coagulation time and the anti-Xa activity assay. J Intensive Care Med. (2020) 35:679–86. doi: 10.1177/0885066618776937,
6.
RussoFMArtoniACotzaMVeroneseGCenciSVillanoMet al. Activated clotting time measured by Hemochron signature elite in adult cardiac surgery: implications for clinical practice. J Cardiothorac Vasc Anesth. (2025) 39:2615–22. doi: 10.1053/j.jvca.2025.05.031,
7.
CarmonaCABainJKaramO. Concordance and discordance of anticoagulation assays in children supported by ECMO: the truth is out there. Perfusion. (2025) 40:1592–9. doi: 10.1177/02676591241309841,
8.
AubronCChapalainXBaileyMBoardJBuhrHCartwrightBet al. Anti-factor-Xa and activated partial thromboplastin time concordance and outcomes in adults undergoing extracorporeal membrane oxygenation: a secondary analysis of the pilot low-dose heparin in critically ill patients undergoing extracorporeal membrane oxygenation randomized trial. Crit Care Explor. (2023) 5:e0999. doi: 10.1097/CCE.0000000000000999,
9.
RajsicSTremlB. Anticoagulation monitoring during ECMO support: monitor or flip a coin?Clin Cardiol. (2024) 47:e70061. doi: 10.1002/clc.70061,
10.
TengYYanSLiuGLouSZhangYJiB. An agreement study between point-of-care and laboratory activated partial thromboplastin time for anticoagulation monitoring during extracorporeal membrane oxygenation. Front Med. (2022) 9:931863. doi: 10.3389/fmed.2022.931863,
11.
LardinoisBHardyMMichauxIHorlaitGRotensTJacqminHet al. Monitoring of unfractionated heparin therapy in the intensive care unit using a point-of-care aPTT: a comparative, longitudinal observational study with laboratory-based aPTT and anti-Xa activity measurement. J Clin Med. (2022) 11:1338. doi: 10.3390/jcm11051338,
12.
KarigowdaLDeshpandeKJonesSMillerJ. The accuracy of a point-of-care measurement of activated partial thromboplastin time in intensive care patients. Pathology. (2019) 51:628–33. doi: 10.1016/j.pathol.2019.05.002,
13.
LiHSerrickCRaoVYipPM. A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass. Perfusion. (2021) 36:610–9. doi: 10.1177/0267659120949351,
14.
LeeJMParkEYKimKMWonJCJungTKLeeSK. Comparison of activated clotting times measured using the Hemochron Jr. signature and Medtronic ACT plus during cardiopulmonary bypass with acute normovolemic haemodilution. J Int Med Res. (2018) 46:873–82. doi: 10.1177/0300060517731952,
15.
OjitoJWHannanRLBurgosMMLimHHuynhMVelisEet al. Comparison of point-of-care activated clotting time systems utilized in a single pediatric institution. J Extra Corpor Technol. (2012) 44:15–20. doi: 10.1051/ject/201244015,
16.
MirusMDietzeATiebelOBeyer-WestendorfJGrottkeOTeschFet al. Viscoelastic point-of-care testing for monitoring unfractionated heparin in critically ill patients: a prospective observational study comparing ClotPro® IN/HI-test, aPTT and anti-Xa activity. BMC Anesthesiol. (2026) 26:222. doi: 10.1186/s12871-026-03781-4,
17.
SchultingeLHulshofAMvan NeervenDMulderMMGSelsJEMHulseweHPMGet al. Applications of rotational thromboelastometry in heparin monitoring in critical COVID-19 disease: observations in the Maastricht intensive care COVID cohort. Thromb Update. (2023) 12:100140. doi: 10.1016/j.tru.2023.100140,
18.
DiasJDLopez-EspinaCGPanigadaMDaltonHJHartmannJAchneckHE. Cartridge-based Thromboelastography can be used to monitor and quantify the activity of unfractionated and low-molecular-weight heparins. TH Open. (2019) 3:e295–305. doi: 10.1055/s-0039-1696658,
19.
YinEBBraceyAWChatterjeeS. Thromboelastography versus thromboelastometry for unfractionated heparin monitoring in adult patients on extracorporeal membrane oxygenation. Perfusion. (2025) 40:235–42. doi: 10.1177/02676591241232513,
20.
ReglingKSainiACashenK. Viscoelastic testing in pediatric mechanical circulatory support. Front Med. (2022) 9:854258. doi: 10.3389/fmed.2022.854258
21.
McMichaelABVRyersonLMRatanoDFanEFaraoniDAnnichGM. 2021 ELSO adult and pediatric anticoagulation guidelines. ASAIO J. (2022) 68:303–10. doi: 10.1097/MAT.0000000000001652,
22.
HelmsJFrereCThieleTTanakaKANealMDSteinerMEet al. Anticoagulation in adult patients supported with extracorporeal membrane oxygenation: guidance from the scientific and standardization committees on perioperative and critical care Haemostasis and thrombosis of the international society on thrombosis and Haemostasis. J Thromb Haemost. (2023) 21:373–96. doi: 10.1016/j.jtha.2022.11.014,
23.
Gouin-ThibaultIMansourAHardyMGuéretPde MaistreESiguretVet al. Management of therapeutic-intensity unfractionated heparin: a narrative review on critical points. TH Open. (2024) 8:e297–307. doi: 10.1055/a-2359-0987,
24.
AmiralJAmiralCDunoisC. Optimization of heparin monitoring with anti-FXa assays and the impact of dextran sulfate for measuring all drug activity. Biomedicine. (2021) 9:700. doi: 10.3390/biomedicines9060700,
25.
ToulonPSmahiMDe PooterN. aPTT therapeutic range for monitoring unfractionated heparin therapy: significant impact of the anti-Xa reagent used for correlation. J Thromb Haemost. (2021) 19:2002–6. doi: 10.1111/jth.15264,
26.
SmahiMDe PooterNHollestelleMJToulonP. Monitoring unfractionated heparin therapy: lack of standardization of anti-Xa activity reagents. J Thromb Haemost. (2020) 18:2613–21. doi: 10.1111/jth.14969,
27.
LyonSGLasserECSteinR. Modification of an amidolytic heparin assay to express protein-bound heparin and to correct for the effect of antithrombin III concentration. Thromb Haemost. (1987) 58:884–7. doi: 10.1055/s-0038-1646008,
28.
MaccaranaMLindahlU. Mode of interaction between platelet factor 4 and heparin. Glycobiology. (1993) 3:271–7. doi: 10.1093/glycob/3.3.271,
29.
HardyMCaboJDeliègeADouxfilsJGouin-ThibaultILecompteTet al. Reassessment of dextran sulfate in anti-Xa assay for unfractionated heparin laboratory monitoring. Res Pract Thromb Haemost. (2023) 7:102257. doi: 10.1016/j.rpth.2023.102257,
30.
LasneDToussaint-HacquardMDelassasseigneCBautersAFlaujacCSavardPet al. Factors influencing anti-Xa assays: a multicenter prospective study in critically ill and noncritically ill patients receiving unfractionated heparin. Thromb Haemost. (2023) 123:1105–15. doi: 10.1055/s-0043-1770096,
31.
HatoriNYoshizuHHagaYKusamaYTakeshimaSSegawaDet al. Biocompatibility of heparin-coated membrane oxygenator during cardiopulmonary bypass. Artif Organs. (1994) 18:904–10. doi: 10.1111/j.1525-1594.1994.tb03342.x,
32.
MazzeffiMClarkMGrazioliADuganCRectorRDaltonHet al. Platelet factor-4 concentration in adult veno-arterial ECMO patients. Perfusion. (2021) 36:688–93. doi: 10.1177/0267659120965104,
33.
ContantGGouault-HeilmannMMartinoliJL. Heparin inactivation during blood storage: its prevention by blood collection in citric acid, theophylline, adenosine, dipyridamole-C.T.A.D. mixture. Thromb Res. (1983) 31:365–74. doi: 10.1016/0049-3848(83)90337-7,
34.
HattersleyPG. Activated coagulation time of whole blood. JAMA. (1966) 196:436–40. doi: 10.1001/jama.1966.03100180108036,
Summary
Keywords
activated clotting time, anti-Xa activity assessment, activated partial thromboplastin time, cardio-pulmonary bypass, dextran sulfate, extracorporeal membrane oxygenation, unfractionated heparin, viscoelastometry
Citation
Weber CF, Zacharowski K, Calatzis A and Lindner ML (2026) Monitoring unfractionated heparin in ECMO and cardiac surgery: the search for a valid reference standard. Front. Med. 13:1818646. doi: 10.3389/fmed.2026.1818646
Received
26 February 2026
Revised
13 April 2026
Accepted
21 April 2026
Published
14 May 2026
Volume
13 - 2026
Edited by
Paweł Kuca, Medical University of Warsaw, Poland
Reviewed by
Heidi J. Dalton, Inova Health System, United States
Anna Gkiouliava, G. Papanikolaou General Hospital, Greece
Janne Moilanen, Oulu University Hospital, Finland
Updates
Copyright
© 2026 Weber, Zacharowski, Calatzis and Lindner.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Andreas Calatzis, andreas@calatzis.net
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.