Validation of Droplet Digital Polymerase Chain Reaction for Salmonella spp. Quantification

  • 1. Departamento de Química, Universidad Nacional de Colombia, Bogota, Colombia

  • 2. Grupo de Metrología en Bioanálisis, Instituto Nacional de Metrología, Bogota, Colombia

Abstract

Salmonellosis is a foodborne disease caused by Salmonella spp. Although cell culture is the gold standard for its identification, validated molecular methods are becoming an alternative, because of their rapidity, selectivity, and specificity. A simplex and duplex droplet digital polymerase chain reaction (ddPCR)-based method for the identification and quantification of Salmonella using ttr, invA, hilA, spaQ, and siiA gene sequences was validated. The method has high specificity, working interval between 8 and 8,000 cp/μL in ddPCR reaction, a limit of detection of 0.5 copies/μL, and precision ranging between 5 and 10% measured as a repeatability standard deviation. The relative standard measurement uncertainty was between 2 and 12%. This tool will improve food safety in national consumption products and will increase the competitiveness in agricultural product trade.

Introduction

Salmonella spp. is one of the most important pathogens and the leading cause of foodborne diseases (). Hence, the development and validation of biometric tools, such as measurement methods and reference materials, are important for making decisions related to the national and global trade of food products (), as well as for public health surveillance, among others.

With recent advances in molecular techniques, several rapid methodologies for the detection and quantification of pathogens based on specific genes and proteins have been developed. Besides their rapidity, these methods provide lower limits of detection and better specificity (). In particular, the detection of Salmonella spp. using polymerase chain reaction (PCR) is based on the presence of single genes in a large number of different serotypes and unique copies within the genome. Among the most commonly used ones are (i) invA, which encodes for invA, a protein required for invasion of the bacterium to epithelial cells (), (ii) hilA (hyperinvasive locus), which encodes for hilA, a transcriptional regulator of the OmpR/ToxR family that activates the expression of genes of the type III secretion system, required for bacterial invasion (), (iii) the ttr locus, which is required for respiration with tetrathionate and survival of the bacteria after infection (), (iv) spaQ, a part of the inv/spa complex, which is required for the entry of the bacteria into non-phagocytic host cells (), and (v) siiA, which produces a regulatory protein encoded by SP14, important for adhesion to epithelial cells during Salmonella invasion ().

Digital PCR is considered a potential primary method of measurement as it does not require standards of the same quantity to yield a measurement result (). It is based on microfluidics technology, which allows the generation of multiple reaction partitions (1,000–10 million, depending on the platform) that work as individual reactions. Based on the positive or negative fraction and following the Poisson distribution, it is possible to determine the absolute concentration of the target of interest in terms of the number of copies per microliter (cp/μL) in the dPCR reaction (; ; ). Besides, owing to the dilution and partitioning of the sample, the technique is less sensitive to inhibitors, which results in better precision and less uncertainty contribution (). In this study, an in-house validation has been described for droplet digital PCR (ddPCR) for quantitating genomic DNA from Salmonella spp. using five different targets, run in two duplex and one simplex form. The performance characteristics evaluated were specificity, working interval, precision, limit of detection (LOD), limit of quantification (LOQ), and measurement uncertainty, contributing to the strengthening of monitoring process of Salmonella in agricultural products in Colombia.

Materials and Methods

Bacterial Strains and Growth Conditions

Sixteen serotypes of Salmonella enterica subsp. enterica and other reference strains phylogenetically close to Salmonella spp., such as Escherichia coli, Klebsiella spp., and Shigella spp. (Table 1) were cultured in solid tryptone soya agar (TSA) or XLT4 (xylose, lysine, and tergitol 4) and liquid selective medium Rappaport Vassiliadis (RV). All strains were grown at 37°C for 12–18 h in liquid medium RV before DNA extraction.

TABLE 1

Strain descriptionSource
Salmonella enteritidisATCC® 13076
Salmonella typhimuriumClinical isolate
Salmonella typhiClinical isolate
Salmonella derbyClinical isolate
Salmonella dublinClinical isolate
Salmonella giveClinical isolate
Salmonella saintpaulClinical isolate
Salmonella hadarClinical isolate
Salmonella infantisClinical isolate
Salmonella anatumClinical isolate
Salmonella panamaClinical isolate
Salmonella typhimurium var5-Clinical isolate
Salmonella javianaClinical isolate
Salmonella braenderupClinical isolate
Salmonella muenchenClinical isolate
Salmonella paratyphi AClinical isolate
Bacillus cereusATCC® 10876
Enterococcus faecalisATCC® 14506
Escherichia coli O104:H4ATCC® BAA-2326
Escherichia coli O145:NMATCC® CDC99–3311
Escherichia coli O157:H7ATCC® 35150
Proteus mirabilisATCC® 12453
Proteus vulgarisATCC® 33420
Shigella boydiiATCC® 9207
Shigella sonneiATCC® 9290
Staphylococcus aureusATCC® 25923
Staphylococcus aureusATCC® 6538
Vibrio parahaemolyticusATCC® 17802

Enterobacteria strains used in this study.

ATCC, Reference strain from the American Type Culture Collection.

DNA Extraction

A protocol equally efficient for both Gram-positive and Gram-negative bacteria was used with some modifications (). Briefly, 3 mL cell culture in the exponential phase was centrifuged at 5,000 rpm for 5 min at 4°C. Then, 800 μL lysis buffer [CTAB 2% (w/v), Tris−HCl 100 mM, NaCl 1.4 M, EDTA 20 mM, and LiCl 0.2% (w/v), pH 8.0] was added to the bacterial pellet. The samples were incubated at 65°C for 30 min for Gram-negative bacteria and 2 h for Gram-positive bacteria. Then, these were centrifuged at 10,000 rpm for 5 min at 4°C. The supernatant was transferred to a new tube where 1 volume of chloroform-isoamyl alcohol (24:1) was added to precipitate the proteins. The samples were gently mixed and centrifuged at 12,000 rpm for 8 min at 4°C. Two further extractions with chloroform-isoamyl alcohol (24:1) were performed on the aqueous phases for each sample. From the pooled supernatants, the DNA was precipitated by the addition of 1 volume of cold isopropanol and 100 μL of 5 M sodium acetate after overnight incubation at −20°C. Subsequently, the samples were centrifuged at 12,000 rpm for 15 min at 4°C and washed with 1 mL of 70% ethanol by inverting the tubes. Finally, once the ethanol had evaporated, the DNA was resuspended in 100 μL 1X TE buffer (Tris–HCl 10 mM, EDTA 1 mM pH 8.0) and stored at −20°C. The quality of the extracted genomic DNA was evaluated using UV spectrophotometry by measuring the absorbance ratios at 260 nm/280 nm and 260 nm/230 nm; DNA integrity was assessed using electrophoresis on 1% agarose gel.

Primers and Probes

The primers and probes (Biosearch Technologies, Petaluma, CA, United States, purified by HPLC) for each target are shown in Table 2.

TABLE 2

GeneNameSequence 5′-3′Amplicon sequenceAmplicon size (pb)References
invAinvA_176FCAACGTTTCCTGCGGTACTGTCAACGTTTCCTGCGGTACTGTTAATTAC119
invA_291RCCCGAACGTGGCGATAATTCACGCTCTTTCGTCTGGCATTATCG
invA_Tx_208 TXa-FAM-CTCTTTCGTCTGGCATTATC GATCAGTACCA-BHQ1ATCAGTACCAGTCGTCTTATCTT GATTGAAGCCGATGCCGGTGAAAT TATCGCCACGTTCGGGCAA

ttrttr6F4287 – DirectoCTCACCAGGAGATTACAACATGGCTCACCAGGAGATTACAACATGG95;
ttr4R4381 – ReversoAGCTCAGACCAAAAGTGACCATCCTAATTTAACCCGTCGTCAGTGGCTA;
ttr5p4336 – SondaCFO-CACCGACGGCGAGACCG ACTTT-BHQ1AAAGTCGGTCTCGCCGTCGGTG GGATGGTCACTTTTGGTCTGAGCT

siiAsiiA FwACGACTGGGATATGAACGGGGAAACGACTGGGATATGAACGGGGAATT107
siiA RvTCGTTGTACTTGATGCTGCGGAGATTTTAATGAAAGAGATTAAG
siiaA TrFAM-ATCCTGATGTAGTTATTGAC ATGAG-BHQ1*AAGATATATCCTGATGTA GTTATTGACATGAGTGTTAACT CCGCAGCATCAAGTACAACGA

hilAhilA FACTGTACGGACAGGGCTAT*ACTGTACGGACAGGGCTATCGGTT129;
hilA RAGA CTC TCG GAT TGA ACC TGA-3′TAATCGTCCGGTCGTAGTGGTGT
hilA Lc640HEX-TCGTCCGGTCGTAGTGGTG TCTCC-BHQ1CTCCGCCAGCGCCGCAACCTACGACT CATACATTGGCGATACTTCCTTTTCAGAT GCAGGATCAGGTTCAATCCGAGAGTCT

spaQspaQ FwCCTGACGCCCGTAAGAGACCTGACGCCCGTAAGAGAGTAAAACTTA113;
spaQ RvGCAATTACAGGAACAGACGCTCGCCATACCAGCCAGACAGTAAA
spaQ PFAM-TAAAACTTCGCCATACCAGC CAGACA-BHQ1ACAAGCATAAACACACGCCA AGTAATTTAATGCCAAAAGG CAGCGTCTGTTCCTGTAATTGC

Primers and probes selected for the analysis of Salmonella spp.

FAM: 6-Carboxyfluorescein, HEX: Hexachlorofluorescein, CFO: Cal Fluor Orange 560, equivalent to HEX dye. BHQ1: Black Hole Quencher 1. *Modified in this study from references.

Quantitative PCR (qPCR)

Quantitative PCR was used as a preliminary test for evaluating the reaction efficiency of primers and probes for each target in simplex mode (CFX 96 Deep well- BioRad, cat. 1855196). Six dilutions of S. enteritidis reference strain ATCC 13076 DNA in 1X TE buffer were prepared (100–0.01 ng/μL). The 20 μL reaction contained 1X iTaqTM Universal Probes Supermix (BioRad cat. 1725131), 400 nM primers, 300 nM probes, and nuclease-free water. 1X TE was used as a non-template control (NTC). All samples were run in triplicate. The amplification cycle consisted of initial denaturation at 95°C for 10 min, followed by 40 cycles of denaturation at 95°C for 15 s and annealing-elongation at 60°C for 30 s, with a heating ramp of 2°C/s. Efficiency was evaluated using Eq. 1,

where E is the efficiency and m the slope of a linear regression between the amplification cycle and the log concentration of each sample. The acceptability criterion for PCR efficiency was 90–100%.

ddPCR

In order to establish the best conditions to perform the validation assay for every target sequence, the annealing temperature through a gradient from 55–63°C was evaluated. In addition, the primer concentration was also evaluated from 300 to 900 nM, and the ramp rate between 1 and 2°C/s. The aim was to establish general conditions to amplify the target sequences under the same experimental conditions. After optimization, the best amplification temperature was 60°C with a ramp rate of 2°C/s and 600 nM concentration of primers.

After optimization, according to MIQE guidelines (; Supplementary Table S8), ddPCR assay was performed in simplex (spaQ), and duplex (hilA-siiA and invA-ttr) form: (i) according to the number of total reactions, a stock master mix with 1X ddPCRTM Supermix for Probes (BioRad, CA, United States cat. 1863024), 600 nM primers, 300 nM probes, and nuclease-free water was prepared; (ii) for every triplicate 60 μL of the stock master mix was weighed, (iii) and finally 6 μL of DNA template (genomic DNA of S. enteritidis, measured gravimetrically) were added to complete 66 μL. Then to avoid pipetting bias, 21 μL PCR reaction mixture and 70 μL generator oil (BioRad cat. 183005) were loaded into an 8-well DG8 cartridge (BioRad cat. 1864008), for every replicate, in order to have the same dilution factor. Droplets were generated in a QX200 droplet generator, transferred to a 96-well plate (BioRad cat. 12001925), and sealed with a PX1 PCR plate sealer (BioRad cat. 1814000). Reactions were amplified in a CFX96 deep-well thermocycler (BioRad cat. 185–5196), and the cycling conditions were: 95°C for 10 min, 40 cycles at 95°C for 30 s, and 60°C for 1 min, with an overall ramp rate of 2°C/s. All the reactions were read in the QX200 droplet reader system (two detection channels FAM/EvaGreen and VIC/HEX) using the QuantasoftTM software V1.7 from BioRad.

Data Acquisition and Analysis

Droplets were classified as PCR-positive or PCR-negative according to a threshold fluorescence value set manually using the QuantasoftTM software V1.7. Technical reasons for excluding data from the analysis were: (i) the total number of droplets <12,000 () and (ii) wells with amplitude different from those of other wells with the same target. This variance indicated poor droplet generation, possibly due to poor handling or mixing of samples, and is likely to generate erroneous concentration values. Reference material ERM-AD623 () was used as the amplification control for assessing the performance of the thermocycling parameters.

Data were exported to a spreadsheet to calculate the concentration of the sample in cp/μL in the ddPCR reaction, with a droplet volume of 0.819 ± 0.017 nL (k = 2; Eqs 2, 3), as long as the confidence limits were associated with Poisson distribution.

where:

Csample: Sample concentration

λ : Copies/partition

N: Number of negative partitions

P: Number of total partitions

v: Droplet volume

D: Total dilution of the sample from the master mix.

The droplet volume was calculated as the mean of values reported previously (; ) and the uncertainty was estimated based on a rectangular distribution, to encompass any influence of droplet volume variability and to avoid underestimation of measurement uncertainty (; ).

Method Performance Characteristics

The performance characteristics evaluated were specificity, working interval, equivalence between simplex and duplex assay, precision as repeatability and intermediate precision, the LOD, the LOQ, and measurement uncertainty.

Results

qPCR Amplification Efficiency of DNA Targets

From the optimized extraction process, 1 mL of 4430 ng/μL of DNA was obtained, 1:4 dilution was prepared as a working solution (WS; see Supplementary Figure S1). Preliminary tests using qPCR were performed to determine whether all gene sequences amplified correctly and whether PCR inhibitors were present. The qPCR amplification efficiencies calculated for the five targets were between 90 and 99% (Table 3), indicating that inhibitors were not present in the starting genomic DNA. The dilutions and log concentration showed a good correlation, indicating the suitability of qPCR for the studied targets.

TABLE 3

GeneSlopeEfficiency (%)R2
invA–3.34699.01.000
ttr–3.35898.51.000
siiA–3.38397.51.000
hilA–3.58790.00.980
spaQ–3.36198.41.000

Amplification efficiency for target sequences under study.

Specificity

The DNA of sixteen serotypes of Salmonella enterica subsp. enterica (Table 1) was amplified using the optimized ddPCR method. On the other hand, the DNA of four enterobacteria groups closely related and not related to the species under study were also evaluated. Each group contained 1 ng/μL DNA of each species, and they were classified as (i) Gram-positive bacteria group: Staphylococcus aureus subsp. aureus, Bacillus cereus, and Enterococcus faecalis; (ii) Shigella group: Shigella sonnei, Shigella boydii, and (iii) SHEC Escherichia Group: E. coli O104: H4, E. coli O145: NM, E. coli O157: H7. As a positive control for this assay, a DNA mixture of 1 ng/μL of four serotypes of Salmonella spp., (S. typhimurium, S. typhi, S. derby, and S. paratyphi A.) was used.

The specificity of the selected sequences was evaluated initially using an in silico analysis. Then, sixteen serotypes of Salmonella spp. (Table 4) were evaluated using a ddPCR assay. The results indicated that a positive amplification response was generated for all Salmonella serotypes and that negative amplification results were generated for all Enterobacteria strains evaluated.

TABLE 4

SerovarResults for all targets
Bacillus cereus
Enterococcus faecalis
Escherichia coli O104:H4
Escherichia coli O145:NM
Escherichia. coli O157:H7
Proteus mirabilis
Proteus vulgaris
Shigella boydii
Shigella sonnei
Staphylococcus aureus subsp. aureus
Staphylococcus aureus subsp. aureus
Vibrio parahaemolyticus
Salmonella typhimurium+
Salmonella typhi+
Salmonella derby+
Salmonella enteritidis+
Salmonella dublin+
Salmonella give+
Salmonella saintpaul+
Salmonella hadar+
Salmonella infantis+
Salmonella anatum+
Salmonella panama+
Salmonella typhimurium var5-+
Salmonella javiana+
Salmonella braenderup+
Salmonella muenchen+
Salmonella paratyphi A+

Specificity analysis.

Working Interval

Using S. enteritidis ATCC 13076 DNA, five WS from 10 to 0.001 ng/μL (as nominal concentration), were prepared. Serial gravimetric dilution from WS to ddPCR master mix reaction were measured in triplicate on three different days (see Supplementary Tables S1S5). The working interval was determined using least squares regression analysis of the linear relationship between each dilution and their corresponding concentration (cp/μL) using the ddPCR method (Table 5). Dilution value is used to analyze the working interval as the nominal concentration is just an estimate of the real concentration. An example is presented in Figure 1. Grubbs’ test was performed to determine outliers on one hilA level replicate. The results for the NTC were negative.

TABLE 5

GeneSlopeInterceptR2
invA1.00946.83780.9996
ttr1.02006.84740.9990
spaQ0.97496.28320.9985
hilA0.97076.25280.9985
siiA*1.00636.39950.9996

Regression analysis for each target sequence evaluated at 95% confidence.

*siiA shows good linearity up to fourth Log dilution.

FIGURE 1

The acceptance criteria for linearity were as follows: a slope statistically different from zero and correlation coefficient >0.99 (). The method showed excellent linearity between 1 and 8,000 cp/μL in reaction (except for siiA, the interval of which was from 8 to 8,000 cp/μL). Therefore, the working interval for the whole method was established as 8–8,000 cp/μL.

As part of the validation process, the ability of the method to run in a duplex form (invA-ttr and hilA- siiA) was also evaluated through regression analysis. Both methods were compared over the working interval, using data of four concentration levels (8–8,000 cp/μL) in triplicate and on three different days (Figure 2; see Supplementary Table S6). A slope statistically equal to 1 and correlation coefficient >0.99 were set as acceptance criteria.

FIGURE 2

According to Figure 2, the results obtained did not differ when invA and ttr were measured independently or in combination. Identical results were obtained for hilA and siiA target sequences (see Supplementary Figure S2). This allowed simultaneous evaluation of four different targets in two duplex reactions with the same confidence.

In addition, the amplitude heat plots of 1D and 2D duplex reactions showed clear populations of droplet distribution in positive and negative partitions, minimizing misclassification of droplets (rain) and absence of cross-reaction (Figure 3). As DNA distribution follows a random pattern, in the 2D plot, droplets are cluster in four groups: (i) double negatives, (ii) FAM positives and HEX negatives, (iii) FAM negatives and HEX positives, or (iv) double positives; not all positive partitions are double-positive in the duplex assay (Figure 3C) due to some possible fragmentation process that affect DNA integrity (); Table 6 shows the variation of clusters fraction with DNA concentration.

TABLE 6

Concentration levelhilA+ siiA+clusterRelative fractionhilA+ siiA- clusterRelative fractionhilA- siiA+ clusterRelative fractionhilA- siiA- clusterRelative fractionTotal partitionsRatio hilA/siiA
150,16299,6%780,2%1250,2%00,0%50,3650.93
211,58222,5%12,48524,3%13,29025,8%14,07227,4%51,4290.96
32150,5%2,6806,0%2,8786,4%38,85187,1%44,6240.93
450,0%3570,7%3400,7%49,05498,6%49,7561.05

Relative fraction in respect to total partitions for every cluster in the duplex amplification for hilA-siiA targets.

FIGURE 3

Precision

Five concentration levels covering 1–8,000 cp/μL in ddPCR reaction, in triplicate reactions on three different days (see Supplementary Tables S1S5), relative repeatability standard deviation (Eq. 4), and relative intermediate standard deviation (Eq. 5), were calculated using one factor ANOVA were calculated using one-factor ANOVA ().

where

MSwitin run : Within run mean squares calculated using one-way ANOVA

MSbetween run : Between run mean squares calculated using one-way ANOVA

nreplicates: Number of replicates per run

Csample, mean: Averaged copy number concentration calculated over all runs.

Repeatability was below 5% for the three highest concentration levels and increased to 10% for the fourth one. However, for the lowest concentration (approximately 1 cp/μL in reaction), repeatability, was beyond 20% (Figure 4A), while intermediate precision was below 5% in almost the entire interval, except for ttr, and siiA (Figure 4B). When MS between <MS within, the Sinterm, rel was considered negligible with respect to Srepeat,rel.

FIGURE 4

LOQ

Limit of quantification represents the lowest copy number of the analyte that can be determined with acceptable performance. We have defined the performance criteria as relative repeatability standard deviation <20%, according to the precision study. The higher values represent a significant increase in the measurement uncertainty. Thus, the LOQ based on the response of all the genes in this study was established as 8 cp/μL (Figure 4A).

LOD

Limit of detection is defined as the lowest analyte concentration that can be distinguished from zero with a specified level of confidence. LOD was determined as the lowest concentration level, where at least three positive droplets were present in all three replicates (), or equivalent in at least nine positive partitions in the pooled replicates. Amplifications were performed in a simplex mode for spaQ and duplex mode for invA-ttr and hilA-siiA. Five gravimetric serial dilutions (1–0.1 cp/μL in the ddPCR reaction) were prepared and run in triplicate. Figure 5 presents the results in which each bar represents the total positive partitions obtained for three replicates compared to a total negative NTC. The estimated LOD was 0.5 cp/μL in the ddPCR reaction for three positive droplets (on average) in at least 12,000 total partitions.

FIGURE 5

Measurement Uncertainty

Based on precision data and considering the mathematical model, the measurement uncertainty estimation was established: first, all possible sources of uncertainty were considered (Figure 6), and they were grouped according to their relation and quantified. Then, all contributions were combined to have standard uncertainty (Eq. 6; ).

FIGURE 6

where

corresponds to the uncertainty contribution provided by the mathematical model (Eq. 2) of the measuring method and includes the uncertainty provided by the partitioning of the sample (λ factor), gravimetric dilution, and droplet volume (Eq. 7),

and , which corresponds to the precision uncertainty provided by the total variation of each concentration level, was defined as the highest degree of dispersion obtained between replicates and between different days (Eq. 8),

n: Days of measurement.

According to the validation data for all target sequences, the relative measurement uncertainty associated with the mathematical model varied between 1,6 and 9% (Table 7). The maximum contribution corresponded to λ factor (50–99%) at all levels (see Supplementary Figure S3). After combining the mathematical model and measurement uncertainty contributions (Eq. 6), the combined uncertainty was calculated for every target in every concentration level (Figure 7). The relative standard uncertainty value ranged between 2 and 12%, while its lower value was obtained at the concentration of 800 cp/μL in the ddPCR reaction.

TABLE 7

LevelConcentration (cp/uL)Relative uncertainty (%)
18,0003.6
28001.6
3803.8
489.0

Relative measurement uncertainty using the mathematical model for the five targets.

Data correspond to an average value.

FIGURE 7

Multiplex Correlation Analysis

The correlation between the targets invA-ttr and hilA-siiA was evaluated to determine the contribution source to the uncertainty associated with covariance between target sequences in duplex assays. Considering that all target sequences are single copy, the correlation was evaluated, keeping the primers concentration constant for one target sequence (600 nM), while this was varied for the other one as 0, 150, 300, 450, 600, 750, and 900 nM. In all cases, the probe concentration was 300 nM. The slope and correlation coefficients were calculated and their significance analyzed using the student’s t-test with a 95% confidence.

As the slopes were statistically equal to zero and the correlation were not significant (Figure 8 and Supplementary Figure S4) for any of the target sequences, then, there is not a covariance contribution to the combined uncertainty ().

FIGURE 8

Discussion

Digital PCR is a relatively new and promising technology, which is currently being used for detecting pathogens in food (). The molecular quantification of specific genes in food samples is a useful tool for evaluating the quality and safety of food products in time-, effort-, and cost-effective manner to generate reliable results (; ). At the same time, digital PCR is considered a potential primary method of measurement in chemical metrology, allowing among its multiple applications, developing reference materials in bioanalysis especially at the clinical and industrial level ().

The method reported here uses five targets of common use for the identification of Salmonella spp. using ddPCR. Two assays in duplex mode (invA-ttr and hilA-siiA) and one in simplex mode (spaQ) were used. The qPCR dilution curves showed that the DNA extraction and purification processes were suitable for obtaining DNA of good quality, which is necessary for validation. The amplification efficiency confirmed the absence of inhibitors that can affect the digital PCR validation process.

2D plots for duplex amplification (Figure 3), shows there is an apparent DNA integrity change, and some variation with DNA concentration. From Table 6, most positive droplets are double-positive for both target sequences (99.6%) at 8,000 cp/μL, but this proportion decrease with concentration level, passing to 22.5 and 0.5% in 2 log DNA concentrations (Table 6 and Supplementary Table S7), indicating a possible fragmentation process in the DNA. The observed clusters relative fraction depends on the fraction of positive droplets. For a fraction higher than 90%, according to Eq. 3, there will be 2 or more copies per partition while for lower ratios, there will be lower occupancy degree of DNA molecules. Droplets with less amplitude for the double positive cluster could indicate an imbalanced amplification process between hilA and siiA DNA target sequences for a non-competing duplex reaction (). However, the ration between hilA/siiA and invA/ttr target sequences keeps around 1, indicating that neither this fragmentation process, nor the imbalance amplification process are enough to affect the quantification in duplex form by digital PCR.

Although four targets (invA, ttr, hilA, and spaQ) showed good linearity over 5 log concentrations (covering 1 cp/μL, according to linear regression analysis), the working interval goes from 8,000 to 8 cp/μL in ddPCR reaction, where the lowest concentration corresponded to the LOQ, which was defined based on precision criteria of maximum 20% as a relative repeatability standard deviation. Precision values higher than 20% (as they were obtained for the lowest levels), contributed more to combined measurement uncertainty.

The LOD, based on the number of positive partitions (at least three in each of the three replicates) and required to differentiate a positive sample of low concentration from the blank, was established as 0.5 cp/μL in reaction, which equals 5 cp/μL, approximately in the stock/sample solution (according to the gravimetric dilution). This would be very important if the objective is diagnostic because it will allow taking decisions in advance. After all, the target sequences used would be enough to differentiate a positive from a negative sample at a very low concentration. Nevertheless, this fact confirms one of the most critical characteristics of dPCR, its sensitivity. Theoretically, it will allow the detection of as few as 5 live/dead cells in the sample. However, the sensitivity would be determined by other factors, especially the sample homogeneity and the DNA extraction efficiency.

The specificity test on 16 different Salmonella serotypes and other related microorganisms, indicated that these sequences could be used together as potential biomarkers for the detection and quantification of the genus. These five genes are of great importance for detecting Salmonella. In total, 329 isolates from environmental and food samples, containing 126 serovars belonging to all subspecies of Salmonella were identified using invA (; ). The ttr locus has also been used as an essential molecular marker for reducing false-positive results (; ).

As shown in Figure 4, the relative repeatability standard deviation of the assay varies from 2 to 12% and correlates with the uncertainty according to the mathematical model of the method. Concentration level 2 (800 cp/μL) presents the best precision and also has the lowest uncertainty associated with the model, indicating that it is the best concentration to measure. The mathematical model contribution (Eq. 7) is a function of the number of positive or negative fractions. Subsequently, once the assay has been optimized, it is possible to assign a relative uncertainty component to the sample based on its concentration (Figure 9A). uλ/λ is the predominant measurement uncertainty source within the mathematical model, ranging from 50 to 99%. The partition number can be increased to improve this estimation. In the QX200 platform (), the maximum partitions per well were 20,000; however, several wells in the QuantasoftTM software can be merged to increase the partition number. This will in turn reduce the uλ/λ contribution (Figure 9B), with consequent reduction in combined measurement uncertainty.

FIGURE 9

Measurement uncertainty contribution was not associated with the correlation between target amplification in the duplex assays. This can be attributed to the primer sequence, which amplifies different regions inside the genome and does not inter-hybridize. Therefore, each primer can function independently.

Conclusion

After a literature review and in silico analysis of the primary target sequences used for the quantification of Salmonella spp., five different amplification targets, namely invA, ttr, hilA, spaQ, and siiA, were selected, which belonged to five genes involved in pathogenicity and metabolism of the bacteria.

According to our results, the method presented here is suitable for estimating the genomic DNA content of Salmonella spp. with five different sequences in simplex and duplex form, from 8 to 8,000 cp/μL, with a detection limit of 0.5 copies/μL in the ddPCR reaction, LOQ of 8 cp/μL, precision between 5 and 10%, and combined relative standard measurement uncertainty between 2.0 and 12% over the working interval considering all sources that contribute to it. The most important factor contributing to measurement uncertainty comes from the mathematical model, specifically from the λ factor. To decrease the measurement uncertainty, some wells (at least two) can be merged using the QuantasoftTM software, which will increase the partition number, especially at low concentrations close to the LOQ.

Nevertheless, the unique feature of this assay that enables the detection of five genes instead of one or two ensures higher accuracy and confidence in the results. While a single gene can indicate the presence or absence of Salmonella spp., adding other molecular targets provides tools for more reliable quantification, especially for some specific applications such as reference material characterization.

This study contributes to the strengthening of monitoring processes and the increase in competitiveness in the marketing of agricultural industry via developing of metrological assurance tools as validated measurement methods for the identification and quantification of Salmonella spp. which is one of the most critical foodborne disease agents. This tool can also be used for the assignment of a reference value for a reference material based on the different sequences evaluated.

Statements

Data availability statement

All datasets generated for this study are included in the article/Supplementary Material.

Author contributions

CV performed the experiments and wrote the original draft. MC was involved in conceptualization, investigation, and funding acquisition. MA was involved in validation, writing, reviewing, and editing. JL designed the methodology, wrote the original draft, and was involved in project administration. All authors contributed to the article and approved the submitted version.

Funding

This work was supported by Departamento Administrativo de Ciencia, Tecnología e Innovación (Colciencias), (project 110174558815, contract 018:2017), and and División de Investigación Sede Bogotá, Universidad Nacional de Colombia (DIB; Call 2016–2018, project 37670).

Acknowledgments

We acknowledge the Instituto Nacional de Salud (INS) for kindly providing Salmonella references and clinical strains.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmicb.2020.01512/full#supplementary-material

References

  • 1

    AtashpazS.KhaniS.BarzegariA.BararJ.VahedS. Z.AzarbaijaniR.et al (2010). A robust universal method for extraction of genomic DNA from bacterial species.Microbiology79538542. 10.1134/S0026261710040168

  • 2

    Ben HassenaA.BarkallahM.FendriI.GrossetN.Ben NeilaI.GautierM.et al (2015). Real time PCR gene profiling and detection of Salmonella using a novel target: the siiA gene.J. Microbiol. Methods109915. 10.1016/j.mimet.2014.11.018

  • 3

    BioRad (2018). Droplet Digital PCR Applications Guide. Available online at: http://www.bio-rad.com/webroot/web/pdf/lsr/literature/Bulletin_6407.pdf(accessed April 18, 2020).

  • 4

    BoddickerJ. D.KnospB. M.JonesB. D. (2003). Transcription of the Salmonella invasion gene activator, hilA, requires HilD activation in the absence of negative regulators.J. Bacteriol.185525533. 10.1128/JB.185.2.525

  • 5

    BoerM. D.de BoerR. F.LameijerA.SterneE.SkidmoreB.SuijkerbuijkA. W. M.et al (2019). Selenite enrichment broth to improve the sensitivity in molecular diagnostics of Salmonella.J. Microbiol. Methods1575964. 10.1016/j.mimet.2018.12.018

  • 6

    CorbisierP.PinheiroL.MazouaS.KortekaasA.-M.ChungP. Y. J.GerganovaT.et al (2015). DNA copy number concentration measured by digital and droplet digital quantitative PCR using certified reference materials.Anal. Bioanal. Chem.40718311840. 10.1007/s00216-015-8458-z

  • 7

    DagataJ. A.FarkasN.KramerJ. A. (2016). Method for Measuring The Volume Of Nominally 100 μM Diameter Spherical Water-In-Oil Emulsion Droplets.Gaithersburg, MD: NIST.

  • 8

    DeprezL.CorbisierP.KortekaasA. M.MazouaS.Beaz HidalgoR.TrapmannS.et al (2016). Validation of a digital PCR method for quantification of DNA copy number concentrations by using a certified reference material.Biomol. Detect. Q.92939. 10.1016/j.bdq.2016.08.002

  • 9

    DeprezL.MazouaS.CorbisierP.TrapmannS.SchimmelH.WhiteH.et al (2012). The Certification Of The Copy Number Concentration Of Solutions Of Plasmid DNA Containing A BCR-ABL b3a2 Transcript Fragment.Luxemburgo: Publications Office of the EU.

  • 10

    EkiriA. B.LongM. T.HernandezJ. A. (2016). Diagnostic performance and application of a real-time PCR assay for the detection of Salmonella in fecal samples collected from hospitalized horses with or without signs of gastrointestinal tract disease.Vet. J.2082832. 10.1016/j.tvjl.2015.11.011

  • 11

    EmslieK. R.McLaughlinJ.GriffithsK.Forbes-SmithM.PinheiroL.BurkeD. (2019). Droplet volume variability and impact on digital PCR copy number concentration measurements.Anal. Chem.9141244131. 10.1021/acs.analchem.8b05828

  • 12

    European Comission (2007). Regulation (EC) No 2073/2005 Regarding Microbiological Criteria Specific To Food Products.Madrid: Official Journal of the European Union.

  • 13

    Furuta-HanawaB.YamaguchiT.UchidaE. (2019). Two-dimensional droplet digital pcr as a tool for titration and integrity evaluation of recombinant adeno-associated viral vectors.Hum. Gene. Ther. Methods.30127136. 10.1089/hgtb.2019.031

  • 14

    GalanJ. E.GinocchioC.CosteasP. (1992). Molecular and functional characterization of the Salmonella invasion gene inv a: homology of InvA to members of a new protein family.J. Bacteriol.17443384349. 10.1128/jb.174.13.4338-4349.1992

  • 15

    González-EscalonaN.BrownE. W.ZhangG. (2012). Development and evaluation of a multiplex real-time PCR (qPCR) assay targeting ttrRSBCA locus and invA gene for accurate detection of Salmonella spp. in fresh produce and eggs.Food Res. Int.48202208. 10.1016/j.foodres.2012.03.009

  • 16

    Gutierrez-AguirreI.RackiN.DreoT.RavnikarM. (2015). Droplet digital PCR for absolute quantification of pathogens.Methods Mol. Biol.1302331347. 10.1007/978-1-4939-2620-6_24

  • 17

    HuberL. (2007). Validation and Qualification In Analytical Laboratories.London: Informa Healthcare.

  • 18

    HuggettJ. F.FoyC. A.BenesV.EmslieK.GarsonJ. A.HaynesR.et al (2013). The digital MIQE guidelines: minimum information for publication of quantitative digital PCR experiments.Clin. Chem.59892902. 10.1373/clinchem.2013.206375

  • 19

    Joint Committee for Guides in Metrology [JCGM] (2008). ISO/IEC Guide 98-3. JCGM 100. Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement. Available online at: https://www.bipm.org/en/publications/guides/#gum(accessed April 10, 2020).

  • 20

    KasturiK. N.DrgonT. (2017). Real-time PCR method for detection of Salmonella spp. in environmental samples.Appl. Environ. Microbiol.83112. 10.1128/aem.00644-17

  • 21

    KoširA. B.DivietoC.PavšičJ.PavarelliE.DobnikD.DreoT.et al (2017). Droplet volume variability as a critical factor for accuracy of absolute quantification using droplet digital PCR.Anal. Bioanal. Chem.40966896697. 10.1007/s00216-017-0625-y

  • 22

    KurowskiB.Traub-dargatzJ. L.MorleyP. S.Gentry-weeksC. R. (2002). Detection of Salmonella spp. in fecal specimens by use of real-time polymerase chain reaction.Am J. Vet. Res.6312651268. 10.2460/ajvr.2002.63.1265

  • 23

    LawJ. W.-F.Ab MutalibN.-S.ChanK.-G.LeeL.-H. (2014). Rapid methods for the detection of foodborne bacterial pathogens: principles, applications, advantages and limitations.Front. Microbiol.5:770. 10.3389/fmicb.2014.00770

  • 24

    MagnussonB.ÖrnemarkU. (2014). Eurachem Guide: The Fitness for Purpose of Analytical Methods – A Laboratory Guide to Method Validation and Related Topics.London: Eurachem Guide.

  • 25

    MalornyB.BungeC.HelmuthR. (2007). A real-time PCR for the detection of Salmonella enteritidis in poultry meat and consumption eggs.J. Microbiol. Methods70245251. 10.1016/j.mimet.2007.04.013

  • 26

    McCabeE. M.BurgessC. M.O’ReganE.McGuinnessS.BarryT.FanningS.et al (2011). Development and evaluation of DNA and RNA real-time assays for food analysis using the hilA gene of Salmonella enterica subspecies enterica.Food Microbiol.28447456. 10.1016/j.fm.2010.10.012

  • 27

    MorissetD.StebihD.MilavecM.GrudenK.ZelJ. (2013). Quantitative analysis of food and feed samples with droplet digital PCR.PLoS One8:e062583. 10.1371/journal.pone.0062583

  • 28

    QuanP. L.SauzadeM.BrouzesE. (2018). dPCR: a technology review.Sensors1812711298. 10.3390/s18041271

  • 29

    RothrockM. J.HiettK. L.KiepperB. H.IngramK.HintonA. (2013). Quantification of zoonotic bacterial pathogens within commercial poultry processing water samples using droplet digital PCR.Adv. Microbiol.3403411. 10.4236/aim.2013.35055

  • 30

    SomanathB.KerryR. (2016). Digital polymerase chain reaction for characterization of DNA reference materials.BDQ104749. 10.1016/j.bdq.2016.04.001

  • 31

    SsemandaJ. N.ReijM. W.van MiddendorpG.BouwE.van der PlaatsR.FranzE.et al (2018). Foodborne pathogens and their risk exposure factors associated with farm vegetables in Rwanda.Food Control898696. 10.1016/j.foodcont.2017.12.034

  • 32

    WangH.JiangY.LiuX.QianW.XuX.ZhouG. (2016). Behavior variability of Salmonella enterica isolates from meat-related sources.LWT Food Sci. Technol.73375382. 10.1016/j.lwt.2016.06.036

  • 33

    WhaleA. S.HuggetJ. F.TzonevZ. (2016). Fundamentals of multiplexing with digital PCR.BDQ101523. 10.1016/j.bdq.2016.05.002

  • 34

    WhaleA. S.JonesG. M.PavšičJ.DreoT.RedshawN.AkyürekS.et al (2018). Assessment of digital PCR as a primary reference measurement procedure to support advances in precision medicine.Clin. Chem.6412961307. 10.1373/clinchem.2017.285478

  • 35

    WhiteH.DeprezL.CorbisierP.HallV.LinF.MazouaS.et al (2015). A certified plasmid reference material for the standardisation of BCR-ABL1 mRNA quantification by real-time quantitative PCR.Leukemia29369376. 10.1038/leu.2014.217

  • 36

    World Health Organization (2015). WHO Estimates Of The Global Burden Of Foodborne Diseases.Geneva: WHO.

  • 37

    ZhaoY.JiangX.QuY.PanR.PangX.JiangY.et al (2017). Salmonella detection in powdered dairy products using a novel molecular tool.J. Dairy Sci.10034803496. 10.3168/jds.2016-12535

Summary

Keywords

droplet digital PCR, Salmonella, validation, food safety, bioanalysis

Citation

Villamil C, Calderon MN, Arias MM and Leguizamon JE (2020) Validation of Droplet Digital Polymerase Chain Reaction for Salmonella spp. Quantification. Front. Microbiol. 11:1512. doi: 10.3389/fmicb.2020.01512

Received

09 March 2020

Accepted

10 June 2020

Published

07 July 2020

Volume

11 - 2020

Edited by

Dario De Medici, Istituto Superiore di Sanità (ISS), Italy

Reviewed by

David Dobnik, National Institute of Biology (NIB), Slovenia; Barbara Bertasi, Experimental Zooprophylactic Institute of Lombardy and Emilia Romagna (IZSLER), Italy

Updates

Copyright

*Correspondence: John Emerson Leguizamon,

This article was submitted to Food Microbiology, a section of the journal Frontiers in Microbiology

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics