Abstract
Introduction:
Current Toxoplasma gondii (TOX), Epstein–Barr virus (EBV), rubella virus, human cytomegalovirus (HCMV), and herpes simplex virus types I and II (HSV I and II) (TORCHes) diagnostic methods are limited by challenges, such as multi-step workflows and cold-chain dependency. This study aimed to develop an integrated “sample-in-result-out” system combining rapid nucleic acid release, room-temperature-stable multiplex polymerase chain reaction (PCR), and automated interpretation to improve diagnostic accuracy.
Methods:
We utilized a self-prepared lysis buffer for fast nucleic acid extraction from diverse sample types. Primers, probes, and reaction components for six TORCHes pathogens were fabricated into freeze-dried microspheres via vacuum freeze-drying, enabling single-tube multicolor melting curve analysis (MMCA). The companion software, TORCHes-MCAv1.0, was used to automate pathogen typing and result reporting. Evaluation metrics encompassed sensitivity, specificity, precision, stability, and anti-interference ability. The system was validated using 210 clinical samples compared to commercial single-target qPCR, with Sanger sequencing resolving any discrepancies.
Results:
Nucleic acid extraction and release were completed within 1 min. Freeze-dried microspheres demonstrated exceptional repeatability, with Tm intra- and inter-batch CVs of 0.03–0.18%. Limits of detection (LoD) were 200 copies/mL for HSV-I/II, HCMV, and TOX, and 500 copies/mL for EBV and RV. Specificity analysis showed no cross-reactivity with non-target pathogens. Stability testing confirmed the microspheres could be stored stably at room temperature for 1 year. Clinical validation showed near-perfect agreement with qPCR (Kappa = 0.965–1.000), with all seven discrepant results confirmed correct via sequencing. The TORCHes-MCAv1.0 software achieved 100% concordance with manual interpretation.
Conclusion:
We developed a robust, streamlined integrated TORCHes detection system. Its simplicity, stability, and high-throughput nature make it a valuable complementary tool for confirming active infection and differentiating pathogen subtypes in complex clinical scenarios.
Introduction
TORCH is an umbrella term for a group of perinatal pathogens—Toxoplasma gondii (TOX), Rubella virus (RV), Human Cytomegalovirus (HCMV), Herpes Simplex Virus types I and II (HSV-I/II), and Epstein–Barr virus (EBV)—known for their ability to traverse the placental barrier (Neu et al., 2015; Melvin et al., 2022; Hutton et al., 2014). The term “TORCHes” is used herein to specifically denote this expanded panel, which includes EBV in addition to the classic TORCH pathogens (Toxoplasma, Rubella, Cytomegalovirus, and Herpes simplex). Although EBV is not part of the original TORCH acronym, it is explicitly included under the “others (O)” category in the widely adopted Chinese Expert Consensus on TORCH detection, acknowledging its potential for adverse pregnancy outcomes, such as miscarriage (Zhu et al., 2020). Supporting its relevance, studies have demonstrated a significant association between EBV infection and miscarriage in later pregnancy, and EBV DNA has been detected in neonatal cord blood, confirming the occurrence of maternal-fetal transmission (Mitsoda et al., 2014; Sloan et al., 2022). As a member of the TORCHes pathogen group, EBV shares the broader clinical risks characteristic of these infections: maternal infection, often asymptomatic or presenting with nonspecific symptoms, can lead to devastating consequences for the fetus or newborn, including spontaneous abortion, stillbirth, premature delivery, and a spectrum of congenital anomalies and neuro developmental disorders. Given the limited therapeutic options after infection, accurate and timely pathogen detection is paramount for risk assessment and clinical management (Paquet and Yudin, 2018; Saldan et al., 2017; Kilby et al., 2019).
Serological testing for pathogen-specific IgM and IgG antibodies serves as the cornerstone for TORCHes screening in pregnant women, as affirmed by professional guidelines including the Chinese Expert Consensus (Zhu et al., 2020). Nevertheless, interpreting serological results presents considerable complexities, largely stemming from the unavoidable window period before antibody seroconversion occurs, frequent cross-reactivity contributing to false-positive IgM readings, and the challenge in reliably differentiating active from past infections based on static antibody profiles alone (Wandinger et al., 2011; Chen et al., 2019). Against this backdrop, nucleic acid amplification tests (NAATs) play an indispensable complementary role. As emphasized in the Expert Consensus, NAATs provide critical diagnostic value in specific situations: verifying active infection in cases of inconclusive serology, identifying the causative pathogen in symptomatic neonates, and especially in differentiating between HSV-I and HSV-II, a distinction with substantial implications for clinical counselling and therapeutic decision-making. Quantitative real-time PCR (qPCR) offers high analytical sensitivity and specificity. However, conventional single conventional single-target qPCR faces notable constraints in comprehensively detecting all six TORCHes pathogens, primarily due to limited fluorescence channels requiring multiple separate reactions. Additionally, the strict dependence of liquid reagents on uninterrupted cold-chain logistics poses a fundamental barrier to wider dissemination, particularly beyond centralized laboratory environments (Xiang et al., 2020; Kang et al., 2004; Vinjé, 2015).
To bridge this technological gap, the integration of lyophilized reagent technology with multicolour melting curve analysis (MMCA) emerges as a promising strategy (Abla and Mehanna, 2022). Lyophilization enhances reagent stability, allowing for ambient temperature storage. MMCA, utilizing differentially labelled probes and Tm analysis, holds the potential for simultaneous detection and differentiation of multiple pathogens within a single closed tube. This significantly streamlines the workflow and reduces contamination risk (Fu et al., 2012; Li et al., 2022). Nonetheless, despite these clear benefits, the application of this combined methodology to comprehensive TORCHes panel detection remains relatively unexplored. A persistent obstacle lies in the interpretation of intricate multi-target melting curves, which currently depends heavily on manual assessment. Such subjectivity introduces the possibility of inconsistency and misjudgment, highlighting an area in urgent need of standardization.
Therefore, we aimed to develop a novel “Streamlined Integrated TORCHes Detection System.” The rationale of this system was to seamlessly integrate rapid chemical lysis (a manual sample processing step) for nucleic acid release, ready-to-use stabilized freeze-dried multiplex PCR master mix, and dedicated software for entirely automated, objective result interpretation (Figure 1). By combining simplified manual preparation with fully automated analysis, this integrated approach aimed to significantly streamline the workflow and reduce operator dependency. Our study does not aim to replace serological screening but rather provide a powerful adjunct tool. It is designed to facilitate the confirmation of active infection and enable precise pathogen subtyping in complex clinical situations where serology alone proves inadequate, ultimately aiming to enhance diagnostic confidence in TORCHes-related cases.
Figure 1
Methods
Study population and specimens
Patients presenting to the Affiliated Jinhua Hospital of Zhejiang University School of Medicine, Jinhua Maternal and Child Health Hospital, and Jinhua Municipal People’s Hospital between April 2024 and October 2025 with clinically suspected TORCHes infections (TOX, RV, HSV-I/II, EBV, or HCMV) were recruited. A total of 210 clinical specimens were included for validation, comprising 157 whole-blood samples, 17 urine samples, 13 herpes fluid samples, 6 cerebrospinal fluid samples, 5 amniotic fluid samples, and 12 swab samples (e.g., throat and urethral swabs). The study was approved by the hospital ethics committee (Research 2024-Ethical Review-71) and conducted in accordance with the Declaration of Helsinki and its later amendments.
Rapid nucleic acid release and assay principle
A self-prepared lysis buffer (Table 1) was used for rapid, instrument-free nucleic acid release. Specimens were standardized to a 300-μL input volume for lysis. When body-fluid volume was <300 μL (e.g., herpes fluid, cerebrospinal fluid, and amniotic fluid), sterile normal saline was added to achieve a final volume of 300 μL. Swabs were placed in 500 μL of sterile physiological saline and agitated for 30 s, after which 300 μL of the eluate was used for processing. For lysis, 300 μL of the specimen (or processed eluate) was combined with 1,000 μL lysis buffer, mixed for 1 min, and briefly centrifuged (16,000 g, 10 s) to pellet debris; the supernatant was used as the nucleic acid-containing lysate for downstream amplification.
Table 1
| Name | Concentration | Function |
|---|---|---|
| Tris-hydrochloride (Tris–HCl) | 50 mM | Buffer |
| Sodium hydroxide (NaOH) | 150 mM | Osmotic agent |
| Phenylmethylsulfonyl fluoride (PMSF) | 1 mM | Protease inhibitor |
| Ethylenediaminetetraacetic acid (EDTA) | 1 mM | Stabilizer |
| Nonidet P-40 (NP-40) | 1% (v/v) | Detergent |
Cell lysis solution formula.
The assay employed a probe-consumption readout integrated with multicolor melting curve analysis. During amplification, target-specific probes hybridized to complementary amplicon regions; in the presence of target, the 5′ nuclease activity of Taq polymerase hydrolyzed the bound probe during extension, preventing generation of its characteristic post-amplification melting signal. Accordingly, target positivity was defined by the disappearance (loss) of the expected melting peak for that probe, whereas target-negative samples retained intact probes that yielded distinct melting peaks, enabling objective calling based on predefined peak patterns across fluorescence channels.
Target selection and multiplex design
Targets were selected based on diagnostic relevance, sequence conservation for robust amplification, and compatibility with multiplexing and subtype discrimination. TOX targeted the B1 gene per WS/T 486-2015 (National Health and Family Planning Commission of the People’s Republic of China, 2015). RV targeted the E1 gene due to its higher conservation relative to alternative structural gene options (e.g., E2), supporting stable detection across variants. For HSV-I/II, EBV, and HCMV, we followed the multiplex PCR strategy described by Xu et al. (2024), selecting UL30 (HSV-I/II), BKRF1 (EBV), and UL123 (HCMV). Leveraging sequence homology among these targets, multiplex detection and HSV subtyping were performed using a primer architecture consisting of a single universal upstream primer combined with two downstream primers for discriminatory amplification. An internal control was incorporated using a GAPDH-targeting probe (Cy5 at the 5′ end and BHQ3 at the 3′ end) to monitor overall process integrity. Sample validity was assessed based on both amplification behavior (Ct <38) and the presence of the expected internal-control melting peak, ensuring reliable detection and minimizing false-negative reporting.
Primers and probes were designed for stable multiplex amplification and channel-resolved melting analysis while minimizing peak overlap and cross-interactions; beacon fragments were incorporated at both the 5′ and 3′ ends of probes to fine-tune Tm separation. Oligonucleotides were synthesized commercially (Shanghai Shenggong Biotechnology Co., Ltd.). The sequences are provided in Table 2. Reaction composition was iteratively optimized using a commercial PCR buffer/enzyme mixture (Shanghai Janzy Biotechnology Co., Ltd.) by adjusting primer/probe concentrations and evaluating peak placement and separation across channels. The final optimized concentrations in the reaction system were as follows: 600 nM (1245F), 400 nM (4R), 400 nM (125R), 80 nM (HSV I-1P), 65 nM (HSV II-2P), 90 nM (EBV-4P), 70 nM (HCMV-5P), 400 nM (Tox F1), 400 nM (Tox R2), 90 nM (Tox P1), 500 nM (RV-F1), 500 nM (RV-R1), 100 nM (RV-P1), 200 nM (GAPDH-F1), 200 nM (GAPDH-R1), and 100 nM (GAPDH-P1).
Table 2
| Pathogen | Primer/Probe | Sequence (5′–3′) |
|---|---|---|
| Herpes virus | 1245F | TCATCTACGGGGACACGGAC |
| 4R | AAACTTGCAGGCCGTCTTCC | |
| 125R | TGCGCACCARATCCACG | |
| HSV I-1P | FAM-CCGCGCCATATTTGTGCTGTGCCGCGGCGCGG-BHQ1 | |
| HSV II-2P | FAM-CGCGCGCGCATTTTCGTTTTGTGCCGCGGCGCGCGCGCG-BHQ1 | |
| EBV-4P | VIC-CGCGCGCTGATATGTGGGGGTGCTGACGGACGGCAGCGCGCG-BHQ1 | |
| HCMV-5P | VIC-CGCGCGCGCCCTCCACGTTACATCGGCAAAGTGGAGGGCGCGCGCG-BHQ1 | |
| Toxoplasma gondii | Tox-F1 | GGTCGACTTCATGGGACGATAT |
| Tox-R2 | TCCGTCGTAATATCAGGCC | |
| Tox-P1 | ROX-CGCGCGCGCCGCTGTCTGTCTAGGGCACGCGCGCGCG-BHQ2 | |
| Rubella virus | RV-F1 | GATGAGGTTCTTGCCCCCG |
| RV-R1 | ACAACATCGCGCACTTCCC | |
| RV-P1 | ROX-CCGCGCCCCGTCGGCAGTTGGGTAAGAGACCAGGCGCGG-BHQ2 | |
| GAPDH | GAPDH-F1 | TCGCTCAGACACCATGGG |
| GAPDH-R1 | CTACGACTGCAAAGACCCG | |
| GAPDH-P1 | CY5-TGAAGGTCGGAGTCAACGGGTG-BHQ3 |
Sequencing primers and probes for TORCHes pathogenes.
Freeze-dried microspheres and PCR/MMCA conditions
To eliminate cold-chain dependence and develop a ready-to-use multiplex format, reaction components were fabricated as freeze-dried microspheres using vacuum lyophilization. A lyoprotectant system consisting of trehalose (25% w/v), dithiothreitol (0.8 mol/L), bovine serum albumin (0.5% v/v), and mannitol (25% w/v) was incorporated to preserve enzyme activity and probe stability during drying and storage. Reaction premix aliquots (20 μL per microsphere) were dispensed into a cryogenic medium to form uniform spherical structures, followed by staged vacuum freeze-drying under controlled temperature and pressure conditions to ensure structural stability and functional preservation. After drying, microspheres were sealed in dry PCR eight-tube strips for ambient storage (Figure 2). Freeze-drying was performed by Shanghai Janzy Biotechnology Co., Ltd.
Figure 2
For detection, each reaction was initiated by rehydrating a single freeze-dried microsphere with 20 μL nucleic acid lysate supernatant. Amplification and melting analysis were performed on the SLAN-96P platform (HONGSHI, China) using the following program: 50 °C × 10 min; 95 °C × 5 min (1 cycle); 95 °C × 10 s and 60 °C × 40 s (40 cycles); followed by melting at 95 °C × 10 s and 20–60 °C at 0.03 °C/s (1 cycle). Multi-channel melting curves were generated automatically for downstream interpretation.
Reference materials and analytical validation
Synthetic plasmids containing each target were produced commercially (Shanghai Shenggong Biotechnology Co., Ltd.) and used as reference materials. Plasmid concentrations were determined using an aminobenzyl-factor quantitative detection kit (PCR method). A standard curve was generated from serial dilutions of kit standards using qPCR measurements targeting the shared ampicillin resistance gene. Plasmid Ct values were interpolated from the regression equation to obtain concentrations for subsequent LoD, stability, and interference studies.
Analytical sensitivity, specificity, precision, stability, and interference tolerance were evaluated in a predefined framework. For sensitivity, plasmid standards for each pathogen were serially diluted in pooled pathogen-negative human serum to generate graded concentrations and processed using the same lysis and amplification workflow as clinical specimens; testing was performed in 20 replicates per concentration level, and LoD was defined as the lowest concentration achieving a detection rate ≥95%. To corroborate performance in real specimens, 20 clinically confirmed positive samples per pathogen were quantified/assigned concentrations using commercial assays (described below), adjusted to the target concentration series, and re-tested using the same workflow; LoD confirmation required ≥95% positivity (19/20) at the proposed threshold. Specificity was examined using clinically confirmed TORCHes-positive samples (HCMV, TOX, RV, HSV-I/II, EBV), samples positive for other pathogens (HIV, syphilis, HBV, HCV, Neisseria gonorrhoeae, Chlamydia trachomatis, and HPV), and DNA from healthy donors to assess cross-reactivity and non-specific signals. Precision was assessed based on reproducibility of Tm values after rehydrating microspheres with deionized water and directly performing melting analysis. Intra- and inter-batch precision were evaluated across five production batches (eight replicates per batch) and expressed as CVs for each target peak. Accelerated stability testing was performed by storing microspheres at 37 °C and evaluating performance at 6, 9, and 12 months using plasmids at low (LoD), medium (5 × LoD), and high (10 × LoD) concentrations (20 replicates per level); stability was considered acceptable if detection remained 100% at medium/high concentrations and ≥95% at the LoD level throughout testing.
Interference tolerance was assessed by spiking endogenous interferents (hemoglobin ≤5 g/L, triglycerides ≤20 mmol/L, IgG/IgM ≤50 g/L) and exogenous interferents (heparin ≤100 IU/mL, acyclovir/ganciclovir ≤1 μg/mL) into negative serum or whole-blood lysate matrices. Pathogen plasmids were added at 3 × LoD to generate interference-positive samples; interference-free positive controls, interference-matrix negative controls, and water blanks were included, each in triplicate, and all samples were processed using the established workflow. Resistance to interference was defined as correct positivity in interference-positive samples and minimal Tm deviation in interference-negative samples (≤1 °C vs. non-interference negative controls).
Clinical comparison, discrepancy adjudication, and automated interpretation
Clinical performance was assessed by parallel testing of all 210 suspected clinical samples using the microsphere assay and commercial single-target qPCR reagents as comparators, including TOX and RV kits (Janzy Biotechnology Co., Ltd., Shanghai), HCMV and EBV kits (Shengxiang Biotechnology Co., Ltd., Hunan), and an HSV-I/II typing kit (PCR-fluorescence probe method; Prologg Bioproducts Co., Ltd., Shanghai). Discordant results between methods were adjudicated by Sanger sequencing using the primer sequences listed in Table 2.
To standardize melting-curve interpretation and eliminate subjectivity, TORCHes-MCA v1.0 was developed in Python 3.10 using NumPy, SciPy, and PyQt5. The software converted raw melting curves via a negative first-order derivative transformation (−dF/dT), identified candidate peaks, and applied Gaussian fitting (μ as Tm; σ as peak width) optimized by least squares to resolve adjacent peaks. Pathogen calls were produced by matching fitted peaks against channel-specific Tm thresholds, and outputs included pathogen typing and internal control results. Input files were supported from SLAN-96P/S and 7500 exports (Ct and melting peak data in Excel format). A quality-control module denoted “Extraction failed” when the internal control peak was absent or Ct >38. Baseline correction was implemented to reduce background noise. Multiprocessing enabled batch processing, and a fault-tolerant mechanism triggered secondary fitting when fit confidence was insufficient (R2 < 0.95) to prompt manual review.
Statistical analysis
Agreement between methods was assessed using Cohen’s Kappa (SPSSAU online analysis platform). Kappa values were interpreted as follows: 0.8–1.0, strong agreement; 0.6–0.8, relatively strong; 0.4–0.6, moderate; 0.2–0.4, fair; < 0.2, poor.
Results
Establishment of the concentration of plasmid standards
The concentrations of the six synthetic plasmid standards, as determined by the amino benzyl factor quantitative kit, were as follows: HSV-I, 49,339 copies/mL; HSV-II, 37,329 copies/mL; EBV, 122,857 copies/mL; HCMV, 91,491 copies/mL; RV, 89,144 copies/mL; TOX, 51,301 copies/mL. The standard curve exhibited excellent linearity with a correlation coefficient (R2) of 0.998 (Figure 3).
Figure 3
Single-tube multiplex detection performance
The freeze-dried microsphere-based MMCA assay enabled simultaneous differentiation of all six TORCHes pathogens within a single reaction. Representative melting curve analyses are shown in Figure 4. In the FAM channel, HSV-I and HSV-II yielded distinct melting peaks at 36.8 ± 1 °C and 46.2 ± 1 °C, respectively. In the VIC channel, EBV and HCMV were identified by peaks at 40.8 ± 1 °C and 55.8 ± 1 °C, respectively. In the ROX channel, RV and TOX were distinguished by peaks at 37.0 ± 1 °C and 46.2 ± 1 °C, respectively. A positive result was indicated by the disappearance of the corresponding pathogen-specific peak.
Figure 4
Sensitivity and specificity verification
The LoDs were 200 copies/mL for HSV-I, HSV-II, HCMV, and TOX, and 500 copies/mL for EBV and RV (Figure 5 and Table 3). These LoDs were confirmed using 20 clinical positive samples for each pathogen, with detection rates of ≥95% at the respective LoD concentrations. Specificity testing showed that all TORCHes-positive samples were correctly detected and accurately typed. In contrast, samples positive for other pathogens and DNA samples from healthy donors were negative (Supplementary Table 1).
Figure 5
Table 3
| Clinical samples concentration (copies/mL) | Relevance ratio | |||||
|---|---|---|---|---|---|---|
| HSV-I | HSV-II | EBV | HCMV | RV | TOX | |
| 1,000 | 100% | 100% | 100% | 100% | 100% | 100% |
| 500 | 100% | 100% | 100% | 100% | 100% | 100% |
| 200 | 100% | 100% | 90% | 100% | 90% | 100% |
| 100 | 0 | 0 | 0 | 0 | 0 | 0 |
Positive detection rates of serially diluted clinical samples.
Precision and stability verification
The intra-batch and inter-batch CVs for the Tm values of all targets ranged from 0.03 to 0.18% (Figure 6 and Table 4). Under accelerated storage at 37 °C, the freeze-dried microspheres maintained performance for at least 12 months, with detection rates remaining 100% for high (10 × LoD) and medium (5 × LoD) concentration samples. At the LoD level, detection rates met the acceptance criterion of ≥95% across the entire testing period (Figure 7). The finished product (see Figure 2) was designed for long-term storage at room temperature, eliminating the need for cold-chain logistics.
Figure 6
Table 4
| Batches | HSV-I | HSV-II | EBV | CMV | RV | TOX |
|---|---|---|---|---|---|---|
| 1 | 36.83 | 46.11 | 40.78 | 55.81 | 36.98 | 46.25 |
| 2 | 36.67 | 46.14 | 40.83 | 55.85 | 37.01 | 46.33 |
| 3 | 36.86 | 46.15 | 40.84 | 55.81 | 36.96 | 46.19 |
| 4 | 36.78 | 46.13 | 40.84 | 55.84 | 36.97 | 46.24 |
| 5 | 36.80 | 46.22 | 40.80 | 55.80 | 36.96 | 46.20 |
| CV% | 0.18 | 0.08 | 0.05 | 0.03 | 0.06 | 0.11 |
Coefficient of variation (CV) of Tm values across five lyophilized microsphere batches.
Figure 7
Anti-interference study
Common endogenous (hemoglobin ≤5 g/L, triglycerides ≤20 mmol/L, immunoglobulins ≤50 g/L) and exogenous (heparin ≤100 IU/mL, acyclovir/ganciclovir ≤1 μg/mL) interferents did not affect assay performance. All positive samples containing interferents were correctly detected, and Tm values for negative samples containing interferents deviated by less than 0.5 °C from controls (Table 5 and Supplementary Figure S5).
Table 5
| Interfering substances | Concentration | Tm shift (°C) | False positive rate |
|---|---|---|---|
| Hemoglobin | 5 g/L | ≤0.3 | 0% |
| Triglycerides | 20 mmol/L | ≤0.2 | 0% |
| Immunoglobulin IgG/IgM | 50 g/L | ≤0.4 | 0% |
| Heparin | 100 IU/mL | ≤0.3 | 0% |
| Acyclovir | 1 μg/mL | ≤0.2 | 0% |
Tm value shifts of target melting peaks in the presence of interferents.
Comparison of methodology
Among the 210 clinical samples, Kappa values for each pathogen ranged from 0.965 to 1.000, indicating high agreement with the comparator assay (Table 6). TOX results were completely consistent (Kappa = 1.000). The Kappa values for HSV-I, HSV-II, EBV, HCMV, and RV were 0.986, 0.967, 0.986, 0.984, and 0.965, respectively. Results for 16 mixed-infection samples were concordant between methods; the most frequent co-detections were EBV and RV (68.75%, 11/16). Seven discordant cases were resolved by Sanger sequencing, and in all instances, the results generated by this method were confirmed as correct (Supplementary Figure S6). Notably, across diverse clinical sample types (whole blood, urine, herpes fluid, cerebrospinal fluid, amniotic fluid, and swabs), our method maintained high overall agreement (96.67%, 203/210) with fluorescent PCR, with 100% concordance for herpes fluid, cerebrospinal fluid, and amniotic fluid, and >91% for blood, urine, and swabs (Supplementary Table 2).
Table 6
| New TORCHes typing method | Fluorescent PCR method analysis | ||||
|---|---|---|---|---|---|
| Pathogen types | + | − | Kappa values | 95% CI | |
| HSV-I | + | 41 | 0 | 0.986 | [0.958, 1.000] |
| − | 1 | 168 | |||
| HSV-II | + | 34 | 0 | 0.967 | [0.922, 1.000] |
| − | 2 | 174 | |||
| EBV | + | 43 | 0 | 0.986 | [0.959, 1.000] |
| − | 1 | 166 | |||
| CMV | + | 37 | 1 | 0.984 | [0.954, 1.000] |
| − | 0 | 172 | |||
| RV | + | 32 | 1 | 0.965 | [0.918, 1.000] |
| − | 1 | 176 | |||
| TOX | + | 12 | 0 | 1.000 | [1.000, 1.000] |
| − | 0 | 198 | |||
Agreement analysis between the developed method and reference method for 210 clinical samples.
Melting curve interpretation software
TORCHes-MCA v1.0 was used to automatically analyze raw melting curve data and generate result outputs, with the operational workflow shown in Figure 8. The software automatically reported pathogen typing results and internal control Ct values (Figure 9) and flagged “extraction failure” when the internal standard was abnormal (peak missing or Ct >38). Using the dataset from the 210 clinical samples processed by cell lysate and tested with freeze-dried microspheres, the results showed that the software outputs were fully concordant with manual interpretation, yielding a coincidence rate of 100%.
Figure 8
Figure 9
Discussion
At present, the clinical detection of TORCHes pathogens mainly relies on ELISA, chemiluminescence immunoassay (CLIA) and single fluorescent quantitative PCR technology. Despite the high sensitivity of ELISA and CLIA, serological methods have a high false positive rate due to cross-reaction of antibodies. One study of prepregnant women showed that ELISA IgM had a false positive rate of 67.3% (Paquet and Yudin, 2018). CLIA is limited by technical bottlenecks such as short luminescence time of signal molecules (Sun et al., 2021; Kemppainen et al., 2018; Cook et al., 2021). Although single fluorescent quantitative PCR can achieve early nucleic acid diagnosis, it has problems such as low throughput (multiple tubes of reaction are required to cover six pathogens) and liquid reagents rely on strict cold chain storage and transportation (−20 °C), which means its application in remote areas is limited (Vinjé, 2015). The shortcomings in operation complexity, stability and accuracy of these methods severely limit the efficiency of large-scale screening.
In this study, a streamlined technical system of “nucleic acid extraction of cell lysate—multiple detection of freeze-dried microspheres—intelligent interpretation of melting curve” was successfully constructed, which effectively solved the three bottlenecks of traditional detection, such as complicated operation, poor stability and subjectivity of interpretation: (1) Operation convenience: By optimizing the components, the lysate efficiently released nucleic acids within 1 min and was stable in hemolytic/lipid blood samples, significantly simplifying the pretreatment; (2) Stability and integration: the freeze-dried microspheres were protected by trehalose-DTT system and stable at 37 °C for 1 year. Based on the freeze-dried microspheres and MMCA technology, six TORCHes pathogens could be detected simultaneously in a single tube, completely eliminating the dependence on cold chain; (3) Intelligent interpretation: the matching TORCHes-MCA v1.0 software improved the typing efficiency through algorithm optimization and eliminates manual interpretation errors. The system showed excellent performance including high sensitivity (LoD = 200 copies/mL for HSV-I/II/HCMV/TOX, LoD = 500 copies/mL for EBV/RV), high specificity, good precision (both intra-batch and inter-batch CV <5%) and strong tolerance to specific interferences. In validating 210 clinical samples, our method showed high consistency with control qPCR (Kappa 0.965–1.000; 100% coincidence) and strong agreement across diverse specimen types, including 100% concordance for herpes fluid, CSF, and amniotic fluid, supporting broad clinical utility. This indicated that the closed-loop system constructed in this study significantly improved the detection convenience, stability and multi-target detection ability, and provided an efficient and reliable new tool for high-throughput TORCHes pathogen identification.
The core of this stability breakthrough lied in the innovative application of vacuum freeze-drying technology and its successful integration with multiple detection systems. By effectively removing water, this technology significantly improved the stability of reagents and reduced the dependence of storage and transportation temperature, thereby prolonging the shelf life while maintaining the reaction activity and sensitivity. The key breakthrough was that the complex multiple primer and probe system for six TORCHes pathogens was successfully lyophilized into ready-to-use microspheres, eliminating the tedious and error-prone operation steps of traditional liquid reagents such as frequent preparation, packaging, and premix. Through the optimization of protective agents (such as trehalose and DTT), the challenge of maintaining the multiple amplification efficiency and melting curve accuracy after freeze-drying was resolved. Thus, freeze-dried microspheres not only showed excellent performance (HSV-I/II/HCMV/TOX LoD = 200 copies/mL) comparable to or even better than single-weight qPCR (Shanghai ProMag Biotechnology Co., Ltd., 2023) but also maintained 100% detection of LoD-concentration plasmids after 1 year of storage at 37 °C. Notably, the absence of a cold chain requirement significantly enhances operational efficiency. It transformed the testing process into a more streamlined and robust procedure, ensuring reliable performance in daily laboratory practice.
Notably, our self-prepared lysate nucleic acid extraction protocol, optimized through the synergistic effect of osmotic regulator and detergent, enabled efficient nucleic acid release from a wide range of sample types—including whole blood, herpes fluid, urine, cerebrospinal fluid, amniotic fluid, and multiple swabs—in under 1 min, without relying on commercial extraction instruments. This one-step lysis protocol was designed for rapid nucleic acid release rather than purification and concentration. While commercial extraction kits may provide higher nucleic acid yields via multi-step enrichment processes, our simplified approach delivered nucleic acid of sufficient quality and quantity, as evidenced by the high analytical sensitivity (LoD: 200–500 copies/mL), robust inhibitor tolerance, and excellent clinical agreement achieved in this study. Compared to traditional magnetic bead-based nucleic acid extraction kits—which typically require six core manual steps (lysis, binding, two washes, elution, and collection)—our one-step direct lysis protocol reduced the key processing operations by approximately 83% (5 out of 6 steps). This substantial simplification significantly reduced hands-on time and lowered the risk of cross-contamination, rendering the method particularly suitable for settings prioritizing workflow simplicity and speed over maximum nucleic acid recovery. Such a streamlined, instrument-free workflow is especially well-suited for primary healthcare settings with limited access to specialized laboratory equipment, enhancing the feasibility of molecular testing in resource-constrained environments. At the same time, the designed freeze-dried microspheres exhibited excellent platform compatibility, supporting both crude nucleic acid from our direct lysis and high-purity nucleic acid extracted by commercial kits, thereby providing users with simple and high-sensitivity dual-track detection paths. This flexibility can accommodates resource differences across healthcare institutions—enabling primary hospitals to perform “sample direct detection” while allowing tertiary hospitals to integrate with existing automated platforms to improve throughput, thus offering technical support for hierarchical healthcare systems.
In terms of primer design, we adopted the strategy of “single universal upstream primer + specific downstream primer” (applied to HSV-I/II, EBV and HCMV) to accurately screen conserved gene regions, thereby reducing the total number of primers from conventional 24 to 12 (a 50% reduction) (Abla and Mehanna, 2022). This simple architecture effectively reduced the secondary structure interference and cross-reaction risk between primers, improved the amplification efficiency of multiplex PCR, and provided a key guarantee for the system to achieve high sensitivity (HSV-I/II LoD = 200 copies/mL) and stability. This strategy provided an optimized paradigm for the development of primer design for higher throughput multiplex pathogen nucleic acid detection.
In order to completely eliminate manual interpretation errors and improve efficiency, the TORCHes-MCA v1.0 software was developed to integrate the core algorithm and intelligent quality control function. Based on negative first derivative transformation and Gaussian fitting algorithm, the melting peak was accurately identified (temperature difference resolution ≤0.5 °C). Clinical verification showed that the coincidence rate with manual interpretation was 100%, which ensured the objectivity of typing. The software incorporated a dynamic background calibration function to effectively eliminate interference from factors, such as hemolysis (≤5 g/L Hb) and antibiotics, (acyclovir ≤1 μg/mL), thereby ensuring zero false-positive results. Simultaneously, an internal control mechanism flagged extraction failure in real time when the internal standard GAPDH melting peak was missing or Ct >38, thereby avoiding the risk of false negative. The software worked seamlessly with the freeze-dried microspheres to form a closed loop of “reagent-data-algorithm,” and the standardized output (pathogen typing) could be directly linked to the LIS system. This design realized the automatic process from “sample in” to “result out,” which was especially suitable for large-scale screening in resource-poor settings and significantly improved the throughput and reliability of test results. While our system is qualitative in nature (reporting presence/absence of pathogen nucleic acid), it provides direct evidence of active viral replication. By detecting the pathogen’s genetic material in clinical samples, it directly confirms the presence of the infectious agent itself, circumventing the inherent lag and ambiguity of indirect serological markers. Its result, when integrated into the clinical diagnostic pathway diagnostic pathway recommended by the Expert Consensus, is pivotal for concluding whether an active infection is occurring, especially in clarifying equivocal serology and determining the HSV subtype.
This study successfully established a streamlined technology system integrating “cell lysate extraction, freeze-dried microsphere multiplex detection, and intelligent interpretation software.” This integrated system demonstrated significant advantages in terms of procedural simplicity, enhanced reagent stability, and objective result reporting, offering a practical alternative for decentralized testing environments, particularly those lacking stable cold-chain infrastructure. However, the generalizability of the system across larger populations and diverse sample variations needs further validation through multicenter studies, and the long-term practical effectiveness and cost-efficiency require additional evaluation in real-world clinical practice.
Limitations
This study has some limitations. First, the clinical validation, while demonstrating excellent agreement with reference methods, was performed primarily at a single center with a sample size that yielded limited positive cases for some lower-prevalence pathogens (e.g., TOX). Second, the uneven distribution of sample types means that performance in certain matrices (e.g., CSF, amniotic fluid) was assessed in a smaller number of specimens. Future multi-center studies with larger, prospectively enrolled cohorts are warranted to confirm the robustness and clinical utility of this system across diverse populations and specimen types, and to obtain more precise performance estimates for each individual pathogen.
Conclusion
This study successfully established a streamlined, integrated TORCHes detection system featuring “one-minute lysate-based nucleic acid liberation”, “cold chain-independent multiplex PCR via freeze-dried microspheres”, and “algorithm-driven interpretation.” This system effectively overcame the key limitations of traditional molecular methods, namely operational complexity, reagent instability, and interpretive subjectivity. With its advantages of extreme simplicity, high stability, and excellent multiplexing capability, the system provides an efficient and reliable novel tool for large-scale TORCHes screening, demonstrating significant potential for application in primary healthcare settings.
Statements
Data availability statement
The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding authors.
Ethics statement
The studies involving humans were approved by the Jinhua Hospital Ethics Committee (Research 2024-Ethical Review-71). The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation in this study was provided by the participants’ legal guardians/next of kin.
Author contributions
MF: Methodology, Data curation, Validation, Writing – review & editing, Conceptualization, Writing – original draft, Funding acquisition. SH: Writing – review & editing, Methodology. YW: Methodology, Writing – review & editing, Formal analysis. SM: Investigation, Writing – review & editing, Methodology, Formal analysis. WC: Writing – review & editing, Methodology. NZ: Formal analysis, Methodology, Writing – review & editing. XC: Formal analysis, Writing – review & editing, Methodology. JWa: Investigation, Writing – review & editing, Methodology, Funding acquisition. LZ: Writing – review & editing, Methodology. XX: Methodology, Writing – review & editing. JWu: Conceptualization, Supervision, Resources, Funding acquisition, Project administration, Writing – review & editing. HW: Project administration, Writing – review & editing, Methodology. YM: Writing – review & editing, Project administration, Investigation.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This study was supported by Provincial Public Welfare Fund Research Project Analysis Test Project (No. LTGC24H200004), Jinhua Municipal Science and Technology Plan Project Social Development Key Project (Nos. 2023-3-105 and 2024-3-069), Jinhua Municipal Science and Technology Project Public Welfare Project (No. 2022-4-088), Jinhua Municipal Science and Technology Plan Project Social Development Major Projects (No. 2023-3-067), Project for Basic Research in Jinhua Central Hospital (No. JY2022-6-03).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that Generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmicb.2026.1788209/full#supplementary-material
References
1
AblaK. K.MehannaM. M. (2022). Freeze-drying: a flourishing strategy to fabricate stable pharmaceutical and biological products. Int. J. Pharm.628:122233. doi: 10.1016/j.ijpharm.2022.122233,
2
ChenL.LiuJ.ShiL.SongY.SongY.GaoY.et al. (2019). Seasonal influence on TORCHes infection and analysis of multi-positive samples with indirect immunofluorescence assay. J. Clin. Lab. Anal.33:e22828. doi: 10.1002/jcla.22828,
3
CookN.XuL.HegazyS.WheelerB. J.AndersonA. R.CritelliN.et al. (2021). Multiplex assessment of SARS-CoV-2 antibodies improves assay sensitivity and correlation with neutralizing antibodies. Clin. Biochem.97, 54–61. doi: 10.1016/j.clinbiochem.2021.08.006,
4
FuG.MilesA.AlpheyL. (2012). Multiplex detection and SNP genotyping in a single fluorescence channel. PLoS One7:e30340. doi: 10.1371/journal.pone.0030340
5
HuttonJ.RowanP.GreisingerA.MouzoonM. (2014). Rubella monitoring in pregnancy as a means for evaluating a possible reemergence of rubella. Am. J. Obstet. Gynecol.211, 534.e1–534.e4. doi: 10.1016/j.ajog.2014.05.046,
6
KangG.Iturriza-GomaraM.WheelerJ. G.CrystalP.MonicaB.RamaniS.et al. (2004). Quantitation of group A rotavirus by real-time reverse-transcription-polymerase chain reaction: correlation with clinical severity in children in South India. J. Med. Virol.73, 118–122. doi: 10.1002/jmv.20053,
7
KemppainenR. J.BehrendE. N.CarterS. F.ColeJ. E. (2018). Excess EDTA interferes with cortisol measurement using a solid-phase, chemiluminescent enzyme immunoassay. J. Vet. Diagn. Invest.30, 438–441. doi: 10.1177/1040638718763474
8
KilbyM. D.VilleY.AcharyaG. (2019). Screening for cytomegalovirus infection in pregnancy. BMJ367:l6507. doi: 10.1136/bmj.l6507
9
LiW.LiW.LiL.GuoY.ChenJ.ShangS.et al. (2022). Multiplex detection of eight different viral enteropathogens in clinical samples, combining RT-PCR technology with melting curve analysis. Virol. J.19:61. doi: 10.1186/s12985-022-01789-z
10
MelvinA. J.MohanK. M.VoraS. B.SelkeS.SullivanE.WaldA. (2022). Neonatal herpes simplex virus infection: epidemiology and outcomes in the modern era. J. Pediatric Infect. Dis. Soc.11, 94–101. doi: 10.1093/jpids/piab105,
11
MitsodaR. M.SadygovI. M.PekarA. I. (2014). Role of virus infection as etiological factors of miscarriage. Wiad. Lek.67, 230–231.,
12
National Health and Family Planning Commission of the People’s Republic of China (2015). Diagnosis of Toxoplasmosis. Beijing: Standards Press of China.
13
NeuN.DuchonJ.ZachariahP. (2015). TORCHes infections. Clin. Perinatol.42, 77–103. doi: 10.1016/j.clp.2014.11.001
14
PaquetC.YudinM. H. (2018). No. 285-toxoplasmosis in pregnancy: prevention, screening, and treatment. J. Obstet. Gynaecol. Can.40, e687–e693. doi: 10.1016/j.jogc.2018.05.036
15
SaldanA.FornerG.MengoliC.GussettiN.PalùG.AbateD. (2017). Testing for cytomegalovirus in pregnancy. J. Clin. Microbiol.55, 693–702. doi: 10.1128/JCM.01868-16,
16
Shanghai ProMag Biotechnology Co., Ltd. (2023). Herpes Simplex Virus Type I and II Nucleic Acid Genotyping Detection Kit (PCR-Fluorescent probe Method). Shanghai: Shanghai ZJ Bio-Tech Co., Ltd. (Liferiver).
17
SloanP. E.RodriguezC.HoltzL. R. (2022). Viral prevalence by gestational age and season in a large neonatal cord blood cohort. J. Matern. Fetal Neonatal Med.35, 8482–8487. doi: 10.1080/14767058.2021.1983537,
18
SunT.QuS.HuangT.PingY.LinQ.CaoY.et al. (2021). Rapid and sensitive detection of L-FABP for prediction and diagnosis of acute kidney injury in critically ill patients by chemiluminescent immunoassay. J. Clin. Lab. Anal.35:e24051. doi: 10.1002/jcla.24051,
19
VinjéJ. (2015). Advances in laboratory methods for detection and typing of norovirus. J. Clin. Microbiol.53, 373–381. doi: 10.1128/JCM.01535-14
20
WandingerK. P.SaschenbreckerS.SteinhagenK.ScheperT.MeyerW.BarteltU.et al. (2011). Diagnosis of recent primary rubella virus infections: significance of glycoprotein-based IgM serology, IgG avidity and immunoblot analysis. J. Virol. Methods174, 85–93. doi: 10.1016/j.jviromet.2011.04.001,
21
XiangW.PengZ.XuJ.ShenH.LiW. (2020). Evaluation of a commercial latex agglutination test for detecting rotavirus A and human adenovirus in children’s stool specimens. J. Clin. Lab. Anal.34:e23208. doi: 10.1002/jcla.23208,
22
XuY.LvY.LinM.LiM.CuiD.WangY.et al. (2024). A novel multiplex real-time PCR assay for the detection of cytomegalovirus, Epstein-Barr virus, herpes simplex virus 1/2 and strategies for application to blood screening. Diagn. Microbiol. Infect. Dis.109:116234. doi: 10.1016/j.diagmicrobio.2024.116234,
23
ZhuY.ShangS.ChenY.ChenD.JiaL.QuW.et al. (2020). Expert consensus on standardized TORCH laboratory detection and clinical application. Chin. J. Lab. Med.43, 553–561. doi: 10.3760/cma.j.cn114452-20191020-00597
Summary
Keywords
analysis of melting curves of multi-color probes, freeze-dried microspheres, intelligent interpretation, lysate, TORCHes pathogen
Citation
Fu M, He S, Wu Y, Mao S, Chen W, Zhang N, Chen X, Wang J, Zhou L, Xi X, Wu J, Wang H and Ma Y (2026) A streamlined integrated system integrating lysate release, freeze-dried reagents for multiplex polymerase chain reaction, and intelligent analysis for TORCHes pathogen identification. Front. Microbiol. 17:1788209. doi: 10.3389/fmicb.2026.1788209
Received
15 January 2026
Revised
05 March 2026
Accepted
09 March 2026
Published
24 March 2026
Volume
17 - 2026
Edited by
Alexandro Guterres, Universidade de São Paulo, Brazil
Reviewed by
Soegianto Ali, Atma Jaya Catholic University of Indonesia, Indonesia
Shailabh Kumar, Samsung Semiconductor, United States
Updates
Copyright
© 2026 Fu, He, Wu, Mao, Chen, Zhang, Chen, Wang, Zhou, Xi, Wu, Wang and Ma.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Junqi Wu, jhwujunqi@sina.com; Huabin Wang, whb798183844@126.com; Yongjun Ma, jhyymyj@126.com
† These authors have contributed equally to this work
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.