Abstract
Ultrasmall microelectrode arrays have the potential to improve the spatial resolution of microstimulation. Carbon fiber (CF) microelectrodes with cross-sections of less than 8 μm have been demonstrated to penetrate cortical tissue and evoke minimal scarring in chronic implant tests. In this study, we investigate the stability and performance of neural stimulation electrodes comprised of electrodeposited platinum-iridium (PtIr) on carbon fibers. We conducted pulse testing and characterized charge injection in vitro and recorded voltage transients in vitro and in vivo. Standard electrochemical measurements (impedance spectroscopy and cyclic voltammetry) and visual inspection (scanning electron microscopy) were used to assess changes due to pulsing. Similar to other studies, the application of pulses caused a decrease in impedance and a reduction in voltage transients, but analysis of the impedance data suggests that these changes are due to surface modification and not permanent changes to the electrode. Comparison of scanning electron microscope images before and after pulse testing confirmed electrode stability.
1 Introduction
Ultramicroelectrodes are a class of neural interfaces distinguished by cross-sectional dimensions of less than 15 microns. Ultramicroelectrodes include carbon fiber arrays (; ), the Neurolace and NET probes (), silicon carbide microelectrodes (SiC) (), high density microwires (), and 3D printed arrays (). These devices drastically reduce the foreign body reaction compared to penetrating arrays with larger features (). The small size of ultramicroelectrodes makes them challenging to utilize since they either lack the stiffness to penetrate the brain (and require a temporary stiffener) or are brittle and thus difficult to handle. Polyimide is an example of an electrode substrate that requires a stiffener while silicon is an example of an electrode that is stiff but brittle at dimensions below 15 microns. Carbon fiber (CF) ultramicroelectrodes have proven to be strong enough to penetrate the cortex (; ; ) and peripheral nerve (; ) yet small enough to evoke minimal foreign body reaction (; ; ; , ). Since carbon is also conductive, it can serve as both a mechanical substrate and electrical conductor for an ultramicroelectrode (; ; ). However, carbon is not efficient for neural stimulation. This limitation, along with the small dimensions of ultramicroelectrodes, requires enhancement of carbon by coating with an additional material suitable for neural stimulation.
Platinum and platinum-iridium are widely used for neural stimulation (). Due to their proven efficiency and stability, these materials are used in clinical devices such as deep brain stimulators, which use macroelectrodes (millimeter dimensions), and therefore can stay within charge density limits (30 μC/cm2 ()), and still provide therapeutic levels of stimulation. However, the smaller electrodes site sizes used by ultramicroelectrodes may require a charge density above this limit to evoke responses from cortical cells. Motivated by this need, several alternative materials with improved charge injection capability have been developed. Porous titanium nitride (TiN) has shown biocompatibility and stability along with large charge-injection capacities due to surface roughness. However, accessing the full charge storage capacity under the high current density of neural stimulation is limited by the pore resistance (; ; ). The porous platinum known as “Pt gray” can inject a charge density of 1.4 mC/cm2 (; ; ) without causing material damage or irreversible reactions. Nanostructered platinum (nanoPt) has been recently investigated (; ) through in vitro and in vivo testing as a potential electrochemical coating both for neural recording and stimulation. Reproducibility of the coating process and stimulation performance for small microelectrodes (cross-section 100 μm) has not been evaluated yet. Hydrated Ir oxide films as activated or sputtered iridium oxide (AIROF or SIROF respectively) have reported high charge injection capabilities (; ) in the range of 1–5 mC/cm2. Further, poly (3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT—PSS) has shown a charge injection capacity three times higher than AIROF or SIROF () although PEDOT stability and robustness remain questionable (; ). Electrodeposited platinum iridium (PtIr) has demonstrated reduced impedance (; ; ) and lower voltage transients during pulsed current stimulation (), compared to uncoated control electrodes. A recent study () made a direct comparison between electrodeposited PtIr, reduced graphene oxide, and conducting hydrogel electrodes, finding that PtIr retained low impedance after pulse testing, while graphene failed and conducting hydrogel impedance trended higher. In our previous work () we analyzed the coating morphology of PtIr coated CFs and demonstrated a good coating adhesion during accelerated soak testing.
Here we evaluate CF microelectrode arrays coated with electrodeposited PtIr (PtIr-CF) under conditions of in vitro pulse testing at rates of 300 Hz (Chen et al., 2020) and 10 kHz (), which are of interest in neural engineering, specifically for visual cortex stimulation and high frequency spinal cord stimulation respectively. We characterized PtIr-CF electrodes with electrochemical impedance spectroscopy (EIS), cyclic voltammetry (CV), voltage transients (VT), and scanning electron microscopy (SEM) before and after the coating as well as before and after in vitro pulsing tests. Voltage transients were stable during the pulsing process and no delamination was apparent in pulsed electrodes. We also report data from two pilot studies: in vitro pulse testing of cellular scale PtIr-CF electrode and in vivo pulsing conducted in the visual cortex of a single animal.
2 Materials and Methods
2.1 Carbon Fibers Microelectrode Arrays Fabrication
The carbon fiber electrode arrays (CFEA) consist of eight sharpened CFs coated with a Parylene C insulation layer (800 nm) and are attached to a small printed circuit board [ZIF Probe, ZIF, ()]. Insulation is removed at the tip using a blowtorching process () to expose 100–150 μm of carbon with sharpened tips. The final tip diameter tapers from 8.4 to ∼2 μm diameter at the tip.
One ZIF with 8 CFs with a small tip exposure was used for the pilot study of in vitro pulsing tests of cellular scale PtIr-CF. For the tip preparation of these CFs, instead of the blowtorching method, a scissor was used to expose the CF at the tip. Once cut, the CFs were coated with PtIr (area of ∼108 μm2, see inset of Figure 7A).
For in vivo pulsing tests, one high density carbon fiber (HDCF) array was used. The HDCF fabrication is described elsewhere (). Briefly, the array consisted of 16 silicon shanks of 6 mm length with approximately 250 μm of CF protruding from the end of each shank. Parylene-C was used as an insulation layer and the CF tips were blowtorched to expose and sharpen the CF tips, exposing ∼100–150 μm.
2.2 Electrochemical Impedance Spectroscopy and Cyclic Voltammetry
EIS was collected by applying a 10 mV (peak) sine wave in a frequency range of 1 MHz–10 Hz. All EIS and CV measurements were performed in 1x PBS (0.01 M phosphate buffered saline) solution in a three electrode configuration at open circuit potential using a PtIr wire electrode (∼70 μm) as counter and an Ag/AgCl as reference electrode (3M NaCl, BASi, West Lafayette, IN, United States). A Gamry 600+ potentiostat (Gamry Inc., Warminster, PA, United States) was used for measurement collection.
Cyclic voltammetry (CV) measurements were obtained by sweeping three times between −0.6 and 0.8 V versus Ag/AgCl at a scan rate of 500 mV/s. CVs were measured to establish the cathodal charge storage capacity (CSCC) (). We calculated CSCC from the time integral of the cathodic current.
2.3 Pt–Ir Electrochemical Deposition
PtIr coatings were electrodeposited using a potential cycling technique in a solution of 0.2 g/L of Na3IrCl6H2O and 0.186 g/L of Na2PtCl6H2O in 0.1 M of nitric acid (HNO3) (). A constant temperature of 56°C and a pulsed sonication at a power of 2 W ( = 1 min and = 30 sec.) were used. A 70 μm PtIr wire (A-M System, Sequim, WA, United States) electrode and an Ag/AgCl (3M NaCl, BASi, West Lafayette, IN, United States) were used as counter and reference electrode respectively. The potential range for the electrodeposition process was −0.1 to 0.1 V with 200 mV/s of scan rate for 1,200 cycles, which corresponds to a coating process time of 45 min. A Gamry 600+ potentiostat (Gamry Instruments, Warminster, PA, United States) was used to apply potential cycles and an A700 Qsonica (Qsonica L.L.C. Newtown, CT, United States) was used for sonication.
2.4 SEM Images Acquisition
Images of CFs were acquired with scanning electron microscopy (SEM) before coating, after coating, and after pulse testing, using a Tescan Rise SEM (Tescan Orsay Holding, Brno—Kohoutovice, Czech Republic) in low vacuum mode (LVSTD, low vacuum secondary electron Tescan detector). Low vacuum mode allows imaging without deposition of a conductive film (i.e., gold). Samples were mounted on SEM stubs with the conductive carbon tape without the CF touching the tape. An excitation voltage between 5 and 20 kV was used.
2.5 In-vitro Pulse Testing
2.5.1 300 Hz and 10 kHz in Vitro Pulsing Test
A PlexStim Electrical Stimulator System (Plexon, Dallas, TX, United States, software version 2.3) was used to deliver cathodic first biphasic electric pulses with a charge density of 1 mC/cm
2. A bipolar stimulation was done choosing as ground one PtIr-CF on the pulsed CFEA. Two different current amplitude and pulse duration were applied on the sharp carbon fibers:
1) Electric pulses with a duration of 170 μs, interphase of 60 μs at a frequency of 300 Hz. Pulsing tests were conducted for 12 h (∼13 million pulses);
2) Electric pulses with 30 μs of duration at a frequency of 10 kHz for 3 h (∼108 million pulses).
EIS measurements were taken before and after electric pulsing tests. All the experiments were conducted at room temperature. SEM images were collected before and at the end of pulsing tests to determine surface modifications due to the electric stimulation. The PlexStim provides single ended outputs of the current waveform (presented as a voltage across a resistor) and the VT across the test and counter electrodes. The current and the VT were monitored both for in vitro and in vivo pulsing tests, using a digital oscilloscope (Tektronix TBS 1032B, Beaverton, OR, United States). The electrochemical cell used consisted of a petri dish (Millipore Sigma, diameter of 80 mm and a volume of ∼75 ml) containing 1x of a PBS solution; the leads of the stimulator were connected to the pulsed carbon fiber, and the ground connected to one of the CFs of the pulsed ZIF board.
2.5.2 Electrode Polarization
A three electrode configuration method was used to measure the maximum cathodic polarization, of PtIr-CF electrode (). For that purpose, we used the chronoamperometry experiment of the Gamry potentiostat. An Ag/AgCl and a large Pt foil (5 × 9 mm2) were used as reference and counter electrodes respectively. Biphasic cathodic first current pulses of duration of 500 μs and charge intensity from 0.1 to 1.25 mC/cm2 were applied.
2.5.3 In vitro Cellular Scale PtIr-CF Pulsing Test
One ZIF board with eight small PtIr-CFs (section 2.1 for fabrication details) was characterized by means of pulsing tests. A PlexStim Electrical Stimulator System (Plexon, Dallas, TX, United States, software version 2.3) was used to deliver cathodic first biphasic electric pulses with a duration and interphase gap of 100 μs at 100 Hz. Current amplitudes ranged from 5 to 20 μA. The pulse was applied for the duration of 1 min per CF. A 70 μm PtIr wire was used as the counter electrode.
2.6 In vivo Pulsing Tests
2.6.1 High Density Carbon Fiber Array Fabrication
One HDCF (section 2.1 for details) array was used for in vivo visual cortex stimulation. The HDCF used for this experiment consisted of four CF electrodes, one coated CF for pulse testing and three for stab controls, with the remaining 12 fibers removed for easier insertion.
2.6.2 Rat Surgery
The surgical procedure used was approved by the Michigan State University Animal Care and Use Committee. One Sprague-Dawley rat was anesthetized with 3% isoflurane and maintained at 1–3%. Once hind-limb pinch response was muted, the rat’s head was shaven and placed on a stereotaxic with an anesthesia nosecone, and the head was fixed with ear bars. A longitudinal incision was made at the dorsal surface of the animal’s head, the fascia was removed, and the exposed skull was cleaned and dried. Excessive bleeding of the skull or surrounding skin was stopped with either absorbent spear or cauterization. A 3 × 3 mm craniotomy at the primary visual cortex (V1) was made using this coordinate: (Anterior-Posterior, Medial-Lateral) (−6.5, −3.5) mm from Bregma.
2.6.3 In vivo Stimulation Protocol
For in vivo stimulation, the dura was removed and the PtIr-HDCF was driven into the cortex until a depth of 300 μm was reached. An 8 cm stainless steel wire (A-M Systems) was used as a stimulation return electrode. One minute after insertion, the PtIr-HDCF was connected to the Plexon PlexStim system, and the stimulation return electrode was sutured under the loose skin at the incision site using vicryl sutures. Cathodic-first, symmetric-biphasic pulses with 200 μs duration per phase, 100 μs interphase gap, and 50 Hz pulse frequency were used with four amplitudes: 10, 15, 20, 25 μA (from 0.1 to 0.3 mC/cm2). 6,000 pulses were applied for each current amplitude. After collection of VT, the electrode was driven further into the tissue until 600 μm, and the tissue was stimulated for a total of 1 h (∼180k pulses) at 25 μA (0.3 mC/cm2) and 5 μs interphase gap. VTs were collected every 10 minutes.
The summary image shown in Figure 1A explains the in vivo setup, with the HDCF array penetrating the visual cortex and the stainless steel wire as the counter electrode.
FIGURE 1
2.7 Statistical Analysis and Electric Circuit Model
A paired t-test was used to compare impedance data for CF and PtIr-CF as well as to compare CSCC prior to and after the coating. A paired t-test was also applied to impedance data before and after pulsing both at 300 and 10 kHz. One paired t-test was done for comparing the VTpeak of the VT at the beginning and at the end of each pulse test.
Impedance before and after pulsing at both 300 Hz and 10 kHz were fitted to a circuit model (see inset of Figure 5A). The circuit model consisted of a reference electrode (r.e.), the electrolyte resistance (Ru), the admittance (Y0), and the exponent (α) of the constant phase element (CPE), the parasitic capacitance (C), and the working electrode (w.e.). The extracted parameters were statistically analyzed using a paired t-test. For all the statistical tests, Matlab R2021a was used.
3 Results
3.1 PtIr Coating Protocol Assessment
A total of 39 blowtorched CFs (distributed across five ZIF Boards) were coated with PtIr. The surface area was estimated by SEM inspection and ranged from 1900 to 3,600 μm2. The impedance data were normalized by the electrode area (Ω × cm2). An average 1 kHz normalized impedance and standard deviation of 1.74 ± 0.98 Ω × cm2 was measured for PtIr-CF. Figure 2A shows a representative impedance spectrum from before and after PtIr coating and average data at 1 kHz, normalized to surface area. A significant decrease (p-value of 1.02 × 10−18) in 1 kHz impedance is noted, while impedance above 100 kHz state is unchanged, indicating no change in geometric surface area.
FIGURE 2
In Figure 2B, the cyclic voltammetry (CV) is reported for one CF prior (red) and after the PtIr coating. From the CV, the CSCC () can be quantified. In the example reported in Figure 2B the PtIr enhanced the CSCC from 0.2 to 11 mC/cm2. CV was done for 21 CFs and the mean and standard deviation of the CSCC was of 0.25 ± 0.13 mC/cm2 and 13.1 ± 8.74 mC/cm2 for the CF and PtIr-CF respectively. A paired t-test revealed a statistically significant difference with a p-value of 1.62 × 10−6.
To determine the (), PtIr-CF were characterized as described in section 2.5.2. In Figure 3 the VT are reported with a Eipp (bias level ()) of 0.3 V and an ranging from −0.45 to 0.24 V in the range of the water electrolysis window for PtIr electrodes of −0.6 to 0.8 V ().
FIGURE 3
SEM images were collected before and after the coating process. One example of a CF and a PtIr-CF is reported in Figures 4A,B respectively. The PtIr-CF surface appears homogeneously coated. Detailed SEM of three PtIr-CFs are shown in Figure 4C highlighting a rough, continuous coating. Some surface irregularities are also noted (see yellow arrow of Figure 4C). Adjacent CFs can become temporarily in contact during the plating process, which results in a ridge of PtIr along one side.
FIGURE 4

SEM of PtIr-CF. SEM of a carbon fiber with an exposed surface of 109 μm before (A) and after the PtIr coating (B); (C) SEM images of three different PtIr-CF.
EDS was done to evaluate the Pt and Ir percentage of the PtIr-CF surface. A percentage of 66% for Pt and 34% was measured on one PtIr-CF, consistent with results obtained in a previous set of coated electrodes (
3.2 Electric Pulsing Tests
3.2.1 300 Hz Electric Pulsing Tests
PtIr-CFs (N = 10, area ranging from 1900 to 2,100 μm2) were tested by applying biphasic cathodic first electric pulses of 170 μs duration at a frequency of 300 Hz (Chen et al., 2020) with a charge density of 1 mC/cm2 for 12 h on two consecutive days (6 h each day). In Figure 5A the impedance for one PtIr-CF is reported before (red solid line) and after (blue solid line) pulsing test on Day 1. An impedance decrease after the pulsing test was obtained for all pulsed PtIr-CFs on both days. Similar decreases in impedance after pulsing have been attributed to the modification of the electrode surface (
TABLE 1
| xcm2) | Ru (Ω) | Y0 (S*sα) | α | C (F) | VTpeak | |
|---|---|---|---|---|---|---|
| BPDay1 | 0.69 ± 0.1 | 6.8 ± 2.1 × 103 | 2.13 ± 0.86 × 10−7 | 0.58 ± 0.05 | 5.01 ± 0.94 × 10−11 | 1.19 ± 0.08 |
| APDay1 | 0.27 ± 0.02 | 6.2 ± 0.74 × 103 | 2.9 ± 1.1 × 10−7 | 0.7 ± 0.05 | 5.48 ± 0.68 × 10−11 | 0.94 ± 0.1 |
| p-valueDay1 | 7.45 × 10−5 | 0.33 | 0.003 | 1.29 × 10−5 | 0.25 | 9 × 10−7 |
| BPDay2 | 0.6 ± 0.13 | 6.7 ± 1.5 × 103 | 2.7 ± 0.8 × 10−7 | 0.49 ± 0.05 | 5.45 ± 1 × 10−11 | 1.1 ± 0.11 |
| APDay2 | 0.25 ± 0.04 | 6 ± 1.1 × 103 | 3.63 ± 1.2 × 10−7 | 0.7 ± 0.05 | 5.4 ± 1.7 × 10−11 | 0.93 ± 0.1 |
| p-valueDay2 | 6.41 × 10−6 | 0.07 | 0.0026 | 2.15 × 10−5 | 0.97 | 0.0014 |
Mean and standard deviation of the normalized impedance at 1 kHz, list of all components extracted from the equivalent circuit model (see inset of Figure 5A), VTpeak for Day 1 and Day 2 before and after pulsing. The p-value from the paired t-test are reported with a significance difference for the constant phase element components as well as for normalized impedance at 1 kHz and the VTpeak before and after pulsing on Day 1 and Day 2.
FIGURE 5

300 Hz Electric pulse test. (A) EIS of one PtIr-CF before (red solid line) and after (blue solid line) pulsing on Day1; as inset the circuit model used for fitting impedance data is shown. (B) Voltage transient for one PtIr-CF in the beginning (red) and at the end of pulsing test (blue); a biphasic pulse of 170 μs duration and amplitude of 105 μA (1 mC/cm2) was applied. (C–E) SEM of three PtIr-CF before as inset and after 12 h of pulsing test.
VT were monitored during the pulsing test. One example is shown in Figure 5B, in red and blue are reported VT for Day 1 at the beginning and end of the pulsing tests (6 h) respectively. In Table 1 the mean and standard deviation of the VTpeak are reported for Day 1 and Day 2 of pulsing tests. The p-values from the paired t-test were statistically significant for both days (Table 1).
3.2.2 10 kHz Electric Pulse Tests
PtIr-CFs (N = 10, area ranging from 1,700 to 2,100 μm2) were tested at a frequency of 10 kHz (
In Figure 6A the EIS of one PtIr-CF before (red solid line) and after (blue solid line) pulsing is reported. As in the 300 Hz pulsing test, a significant reduction of the 1 kHz impedance was found after pulsing (see Table 2). A circuit model (see inset of Figure 5A) was fitted to the impedance data prior to and after pulsing tests. In Table 2 the mean and the standard deviation of the extracted parameters are reported, with a statistically significant difference for the α component of the CPE, which suggests increased capacitive behavior for the electrode surface. The increase in Y0 was not statistically significant.
FIGURE 6

10 kHz Electric pulse test. (A) EIS of one PtIr-CF before (red solid line) and after (blue solid line) the pulsing test. (B) Voltage transient for one PtIr-CF at the beginning (red) and at the end (blue) of pulsing test; a biphasic pulse of 30 μs duration and amplitude of 710 μA (1 mC/cm2) was applied. (C–E) SEM of three PtIr-CFs before (as inset) and after 3 h of electric pulsing, in (D) one EDS is reported as inset with highlighted (yellow box) on the tip the presence of Na (red) and Cl (yellow) molecules, in blue and cyan the Ir and Pt molecules are shown as well.
TABLE 2
| xcm2) | Ru (Ω) | Y0 (S*sα) | α | C (F) | VTpeak | |
|---|---|---|---|---|---|---|
| BP | 0.58 ± 0.21 | 6.1 ± 1.9 × 103 | 3.4 ± 1.77 × 10−7 | 0.58 ± 0.07 | 5.01 ± 0.94 × 10−11 | 2.99 ± 0.37 |
| AP | 0.21 ± 0.05 | 5.9 ± 0.6 × 103 | 4.3 ± 1.9 × 10−7 | 0.73 ± 0.09 | 5.48 ± 1.4 × 10−11 | 2.74 ± 0.33 |
| p-value | 0.00017 | 0.77 | 0.14 | 7 × 10−4 | 0.24 | 0.002 |
Mean and standard deviation of the normalized impedance at 1 kHz, list of all components extracted from the equivalent circuit model reported as inset in Figure 5A, VTpeak before and after pulsing. The p-values from the paired t-test are reported with a significance difference for the α value of the CPE and for both normalized impedance and VTpeak.
VT were monitored for the whole pulsing test and the mean of the VTpeak at the beginning and end of pulsing is reported in Table 2. The VTpeak variation before and after the pulsing test was statistically significant. SEM inspection showed no surface modifications (Figures 6C–E), an accumulation of salt was detected on the tip of one fiber, as shown in the yellow box in Figure 6D with the corresponding EDS, showing Na and Cl molecules in red and yellow as well as in blue and cyan Ir and Pt molecules.
3.2.3 In vitro Cellular Scale PtIr-CF Pulsing Test
In vitro pulsing characterization was done using one ZIF with eight small PtIr-CFs (section 2.5.3) to show how electric pulses can be applied using a cellular scale electrode size (less than 15 μm diameter, electrode area of ∼108 μm2). In Figure 7A the impedance before and after pulsing is shown. In Figure 7B the voltage transients are reported for each current amplitude applied with a max of 784 mV applied across the electrode surface. No structure modifications occurred as confirmed by the SEM (see inset of Figure 7A).
FIGURE 7

Pulsing test of cellular scale PtIr-CF. (A) Impedance before (red solid line) and after (blue solid line) pulsing for one PtIr-CF; as inset SEM image of the CF before and after pulsing test. (B) Voltage transients when applying a cathodic first biphasic pulse of 100 μs duration and gap interphase, with current amplitudes from 5 to 20 μA.
3.3 In vivo Electric Pulsing Test
One HDCF board with one coated CF (PtIr-HDCF, area–2000 μm2) was used to perform pilot tests of in vivo electric pulsing test in rat visual cortex. As explained in the method section 2.6.3, VT were first collected at 300 μm depth at four different current amplitudes as shown in Figure 8A.
FIGURE 8

In vivo pulsing tests. (A) VT recorded in rat visual cortex at depth of 300 μm, 6,000 cathodic first biphasic pulses of 200 μs duration, 100 μs gap interphase at 50 Hz were delivered at four different current amplitudes (10, 15, 20 and 25 μA). (B) VT in vivo pulsing test (1 h) at 600 μm depth recorded every 10 min (C) PtIr-HDCF impedance before (red solid line) and after (blue solid line) the in vivo pulsing test. (D) SEM of the PtIr-HDCF after the in vivo pulsing test acquired at 10 kV (left) and 20 kV (right).
Then the PtIr-HDCF was driven at 600 μm of depth into the visual cortex. The tissue was stimulated for a total of 1 h at 25 μA (0.3 mC/cm2) and 5 μs interphase gap. VTs were collected each 10 min as shown in Figure 8B. The mean of the VTpeak was of 2.19 ± 0.3 V and a lower VT was recorded at the end of the pulsing test.
In Figure 8C the PtIr-HDCF impedance before (red solid line) and after (blue solid line) the in vivo pulsing test are reported. After the in vivo pulsing test, the impedance magnitude of the Pt-IrHDCF was slightly increased possibly due to biological tissue adhering on the electrode surface.
SEM were acquired after the in vivo pulsing test. In Figure 8D, left panel the SEM at 10 kV is reported showing in black, low conductive material adhering on the surface (e.g., biological material). On the right panel of Figure 8D the SEM at 20 kV is reported and the PtIr is visible with a crack in the PtIr film near the tip. No SEM was acquired prior to stimulation, so it is unclear when the crack formed. Since the VTs during the in vivo experiments were stable, it is more likely that the crack was due to a defect during PtIr deposition or as a result of handling during insertion or SEM preparation.
4 Discussion
The development of small microelectrodes for localized neural stimulation is the key challenge facing neural interfaces. CFEA have proven to effectively provide neural recording from the motor cortex (
Finally, voltage transient data measured in vivo includes the voltage across both the PtIr-CF electrode and the return electrode. Thus, we did not measure electrode polarization across PtIr-CF in vivo. However, it is known that electrode polarization is generally greater in vivo (
Statements
Data availability statement
The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.
Ethics statement
The animal study was reviewed and approved by the Michigan State University Institutional Animal Care and Use Committee (IACUC).
Author contributions
JW contributed to the conception of the study. EDV and JW contributed to the design of the study. EDV performed in vitro experiments and data curation. PRP fabricated CF and HDCF arrays. BK and QW performed in vivo experiments. All authors contributed to manuscript revision, read, and approved the submitted version.
Funding
This work was financially supported by the National Science Foundation (DMR-1625671, NSF award CBET 2129817), by the National Institutes of Health National Institute of Neurological Disorders and Stroke (UF1NS107659), the Office of the Director National Institutes of Health (OT2OD024907), the NSF-funded NeuroNex MINT hub (Multimodal Integrated Neural Technologies) at the University of Michigan under NSF 1707316, the Kellogg Vision Research Core funded by P30 EY007003 from the National Eye Institute, The University of Michigan College of Engineering, the School of Medicine, and Research to Prevent Blindness.
Acknowledgments
The authors acknowledge the Michigan Center for Materials Characterization and the Lurie Nanofabrication Facility for use of their instruments and staff assistance. The authors acknowledge technical support from Julianna Richie from the Cortical Neural Prosthetics Lab at University of Michigan.
Conflict of interest
The author JW has a financial interest in PtIr coating.
The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
A. FomaniA.MansourR. R. (2011). Fabrication and Characterization of the Flexible Neural Microprobes with Improved Structural Design. Sensors Actuators A: Phys.168, 233–241. 10.1016/j.sna.2011.04.024
2
AliM. A.HuC.JahanS.YuanB.SalehM. S.JuE.et al (2021). Sensing of COVID‐19 Antibodies in Seconds via Aerosol Jet Nanoprinted Reduced‐Graphene‐Oxide‐Coated 3D Electrodes. Adv. Mater.33, 2006647. 10.1002/adma.202006647
3
BeebeX.RoseT. L. (1988). Charge Injection Limits of Activated Iridium Oxide Electrodes with 0.2 Ms Pulses in Bicarbonate Buffered saline (Neurological Stimulation Application). IEEE Trans. Biomed. Eng.35, 494–495. 10.1109/10.2122
4
BoehlerC.VieiraD. M.EgertU.AsplundM. (2020). NanoPt-A Nanostructured Electrode Coating for Neural Recording and Microstimulation. ACS Appl. Mater. Inter.12, 14855–14865. 10.1021/acsami.9b22798
5
CassarI. R.YuC.SambangiJ.LeeC. D.WhalenJ. J.IIIPetrossiansA.et al (2019). Electrodeposited Platinum-Iridium Coating Improves In Vivo Recording Performance of Chronically Implanted Microelectrode Arrays. Biomaterials205, 120–132. 10.1016/j.biomaterials.2019.03.017
6
ChenX.WangF.FernandezE.RoelfsemaP. R. (2020). Shape Perception via a High-Channel-Count Neuroprosthesis in Monkey Visual Cortex. Science370, 1191–1196. 10.1126/science.abd7435
7
CoganS. F.LudwigK. A.WelleC. G.TakmakovP. (2016). Tissue Damage Thresholds during Therapeutic Electrical Stimulation. J. Neural Eng.13, 021001. 10.1088/1741-2560/13/2/021001
8
CoganS. F. (2008). Neural Stimulation and Recording Electrodes. Annu. Rev. Biomed. Eng.10, 275–309. 10.1146/annurev.bioeng.10.061807.160518
9
CoganS.PeramunageD.SmirnovA.EhrlichJ.MccreeryD.ManoonkitiwongsaP. (2007). Polyethylenedioxythiophene (Pedot) Coatings for Neural Stimulation and Recording Electrodes. Mater. Res. Soc. Meet.Abstr. QQ2. 7.
10
DalrympleA. N.HuynhM.RoblesU. A.MarroquinJ. B.LeeC. D.PetrossiansA.et al (2019). Electrochemical and Mechanical Performance of Reduced Graphene Oxide, Conductive Hydrogel, and Electrodeposited Pt-Ir Coated Electrodes: an Activein Vitrostudy. J. Neural Eng.17, 016015. 10.1088/1741-2552/ab5163
11
DehdashtianA.UrsuD.PatelP. G.WelleE.ChestekC. A.CedernaP. S.et al (2020). Abstract 193. Plast. Reconstr. Surg. - Glob. Open8, 131–132. 10.1097/01.gox.0000667828.04397.6c
12
DekuF.CohenY.Joshi-ImreA.KannegantiA.GardnerT. J.CoganS. F. (2018). Amorphous Silicon Carbide Ultramicroelectrode Arrays for Neural Stimulation and Recording. J. Neural Eng.15, 016007. 10.1088/1741-2552/aa8f8b
13
Della ValleE.WelleE. J.ChestekC. A.WeilandJ. D. (2021). Compositional and Morphological Properties of Platinum-Iridium Electrodeposited on Carbon Fiber Microelectrodes. J. Neural Eng.8 (5). 10.1088/1741-2552/ac20bb
14
DupontS. R.NovoaF.VoroshaziE.DauskardtR. H. (2014). Decohesion Kinetics of PEDOT:PSS Conducting Polymer Films. Adv. Funct. Mater.24, 1325–1332. 10.1002/adfm.201302174
15
FanB.RodriguezA. V.VercosaD. G.KemereC.RobinsonJ. T. (2020). Sputtered Porous Pt for Wafer-Scale Manufacture of Low-Impedance Flexible Microelectrodes. J. Neural Eng.17, 036029. 10.1088/1741-2552/ab965c
16
GillisW. F.LissandrelloC. A.ShenJ.PearreB. W.MertiriA.DekuF.et al (2018). Carbon Fiber on Polyimide Ultra-microelectrodes. J. Neural Eng.15, 016010. 10.1088/1741-2552/aa8c88
17
GoldbergI. B.BardA. J.FeldbergS. W. (1972). Resistive Effects in Thin Electrochemical Cells. Digital Simulations of Electrochemistry in Electron Spin Resonance Cells. J. Phys. Chem.76, 2550–2559. 10.1021/j100662a013
18
HuanY.GillJ. P.FritzingerJ. B.PatelP. R.RichieJ. M.della ValleE.et al (2021). Carbon Fiber Electrodes for Intracellular Recording and Stimulation. bioRxiv.
19
JimanA. A.RatzeD. C.WelleE. J.PatelP. R.RichieJ. M.BottorffE. C.et al (2020). Multi-channel Intraneural Vagus Nerve Recordings with a Novel High-Density Carbon Fiber Microelectrode Array. BioRxiv10 (1). 10.1038/s41598-020-72512-7
20
KapuralL.YuC.DoustM. W.GlinerB. E.VallejoR.SitzmanB. T.et al (2015). Novel 10-kHz High-Frequency Therapy (HF10 Therapy) Is Superior to Traditional Low-Frequency Spinal Cord Stimulation for the Treatment of Chronic Back and Leg Pain. Anesthesiology123, 851–860. 10.1097/aln.0000000000000774
21
KleinJ. D.ClausonS. L.CoganS. F. (1989). Morphology and Charge Capacity of Sputtered Iridium Oxide Films. J. Vacuum Sci. Tech. A: Vacuum, Surf. Films7, 3043–3047. 10.1116/1.576313
22
KolloM.RaczR.HannaM.-E.ObaidA.AngleM. R.WrayW.et al (2020). Chime: Cmos-Hosted In Vivo Microelectrodes for Massively Scalable Neuronal Recordings. Front. Neurosci.14, 834. 10.3389/fnins.2020.00834
23
KozaiT. D. Y.LanghalsN. B.PatelP. R.DengX.ZhangH.SmithK. L.et al (2012). Ultrasmall Implantable Composite Microelectrodes with Bioactive Surfaces for Chronic Neural Interfaces. Nat. Mater11, 1065–1073. 10.1038/nmat3468
24
LeeC. D.HudakE. M.WhalenJ. J.IIIPetrossiansA.WeilandJ. D. (2018). Low-impedance, High Surface Area Pt-Ir Electrodeposited on Cochlear Implant Electrodes. J. Electrochem. Soc.165, G3015–G3017. 10.1149/2.0031812jes
25
LeungR. T.ShivdasaniM. N.NayagamD. A.ShepherdR. K. (2014). In Vivo and In Vitro Comparison of the Charge Injection Capacity of Platinum Macroelectrodes. IEEE Trans. Biomed. Eng.62, 849–857. 10.1109/TBME.2014.2366514
26
MasseyT. L.SantacruzS. R.HouJ. F.PisterK. S. J.CarmenaJ. M.MaharbizM. M. (2019). A High-Density Carbon Fiber Neural Recording Array Technology. J. Neural Eng.16, 016024. 10.1088/1741-2552/aae8d9
27
NorlinA.PanJ.LeygrafC. (2004). Investigation of Electrochemical Behavior of Stimulation/sensing Materials for Pacemaker Electrode Applications: I. Pt, Ti, and Tin Coated Electrodes. J. Electrochem. Soc.152, J7. 10.1149/1.1842092
28
PatelP. R.NaK.ZhangH.KozaiT. D. Y.KotovN. A.YoonE.et al (2015). Insertion of Linear 8.4μm Diameter 16 Channel Carbon Fiber Electrode Arrays for Single Unit Recordings. J. Neural Eng.12, 046009. 10.1088/1741-2560/12/4/046009
29
PatelP. R.ZhangH.RobbinsM. T.NofarJ. B.MarshallS. P.KobylarekM. J.et al (2016). Chronicin Vivostability Assessment of Carbon Fiber Microelectrode Arrays. J. Neural Eng.13, 066002. 10.1088/1741-2560/13/6/066002
30
PolikovV. S.TrescoP. A.ReichertW. M. (2005). Response of Brain Tissue to Chronically Implanted Neural Electrodes. J. Neurosci. Methods148, 1–18. 10.1016/j.jneumeth.2005.08.015
31
PoseyF. A.MorozumiT. (1966). Theory of Potentiostatic and Galvanostatic Charging of the Double Layer in Porous Electrodes. J. Electrochem. Soc.113, 176. 10.1149/1.2423897
32
RoseT. L.RobbleeL. S. (1990). Electrical Stimulation with Pt Electrodes. Viii. Electrochemically Safe Charge Injection Limits with 0.2 Ms Pulses (Neuronal Application). IEEE Trans. Biomed. Eng.37, 1118–1120. 10.1109/10.61038
33
SchmidtE. M.BakM. J.HambrechtF. T.KuftaC. V.O'RourkeD. K.VallabhanathP. (1996). Feasibility of a Visual Prosthesis for the Blind Based on Intracortical Micro Stimulation of the Visual Cortex. Brain119, 507–522. 10.1093/brain/119.2.507
34
VomeroM.CastagnolaE.CiarpellaF.MaggioliniE.GoshiN.ZucchiniE.et al (2017). Highly Stable Glassy Carbon Interfaces for Long-Term Neural Stimulation and Low-Noise Recording of Brain Activity. Sci. Rep.7, 1–14. 10.1038/srep40332
35
WangY.GrahamS.UnsworthC. P. (2021). Superior Galvanostatic Electrochemical Deposition of Platinum Nanograss Provides High Performance Planar Microelectrodes for In Vitro Neural Recording. J. Neural Eng.10.1088/1741-2552/ac1bc1
36
WelleE. J.PatelP. R.WoodsJ. E.PetrossiansA.Della ValleE.Vega-MedinaA.et al (2020). Ultra-small Carbon Fiber Electrode Recording Site Optimization and Improved In Vivo Chronic Recording Yield. J. Neural Eng.17, 026037. 10.1088/1741-2552/ab8343
37
WelleE. J.WoodsJ. E.JimanA. A.RichieJ. M.BottorffE. C.OuyangZ.et al (2021). Sharpened and Mechanically Durable Carbon Fiber Electrode Arrays for Neural Recording. IEEE Trans. Neural Syst. Rehabil. Eng.29, 993–1003. 10.1109/tnsre.2021.3082056
38
XieC.LiuJ.FuT.-M.DaiX.ZhouW.LieberC. M. (2015). Three-dimensional Macroporous Nanoelectronic Networks as Minimally Invasive Brain Probes. Nat. Mater14, 1286–1292. 10.1038/nmat4427
39
ZengQ.XiaK.SunB.YinY.WuT.HumayunM. S. (2017). Electrodeposited Iridium Oxide on Platinum Nanocones for Improving Neural Stimulation Microelectrodes. Electrochimica Acta237, 152–159. 10.1016/j.electacta.2017.03.213
40
ZhouD. M. (2005). Platinum Electrode and Method for Manufacturing the Same. US Patent6 (974), 533. [Dataset].
41
ZhouD. M. (2011). Platinum Electrode Surface Coating and Method for Manufacturing the Same. US Patent7 (887), 681. [Dataset].
Summary
Keywords
carbon fiber microelectrode (CFME), PtIr coating, microstimulation, in vivo stimulation, in vitro stimulation
Citation
della Valle E, Koo B, Patel PR, Whitsitt Q, Purcell EK, Chestek CA and Weiland JD (2021) Electrodeposited Platinum Iridium Enables Microstimulation With Carbon Fiber Electrodes. Front. Nanotechnol. 3:782883. doi: 10.3389/fnano.2021.782883
Received
24 September 2021
Accepted
10 November 2021
Published
02 December 2021
Volume
3 - 2021
Edited by
Hui Fang, Dartmouth College, United States
Reviewed by
Hyowon Lee, Purdue University, United States
Wonryung Lee, Korea Institute of Science and Technology, South Korea
Yongli Qi, Dartmouth College, United States
Updates

Check for updates
Copyright
© 2021 della Valle, Koo, Patel, Whitsitt, Purcell, Chestek and Weiland.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Elena della Valle, eldellav@umich.edu
This article was submitted to Nanodevices, a section of the journal Frontiers in Nanotechnology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.