AUTHOR=Cao Jiajun , Yuan Chang , Zhang Yukun , Quan Yue , Chang Peipei , Yang Jing , Song Qingwei , Miao Yanwei TITLE=Observation of intracranial artery and venous sinus hemodynamics using compressed sensing-accelerated 4D flow MRI: performance at different acceleration factors JOURNAL=Frontiers in Neuroscience VOLUME=Volume 18 - 2024 YEAR=2024 URL=https://www.frontiersin.org/journals/neuroscience/articles/10.3389/fnins.2024.1438003 DOI=10.3389/fnins.2024.1438003 ISSN=1662-453X ABSTRACT=Objective: To investigate the feasibility and performance of 4D flow MRI accelerated by compressed sensing (CS) for the hemodynamic quantification of intracranial artery and venous sinus.: Forty healthy volunteers were prospectively recruited, and 20 volunteers underwent 4D flow MRI of cerebral artery, and the remaining volunteers underwent 4D flow MRI of venous sinus. A series of 4D flow MRI was acquired with different acceleration factors (AFs), including sensitivity encoding (SENSE, AF=4) and CS (AF=CS4, CS6, CS8 and CS10) at a 3.0T MRI scanner. The hemodynamic parameters, including flow rate, mean velocity, peak velocity, max axial wall shear stress (WSS), average axial WSS, max circumferential WSS, average circumferential WSS and 3D WSS, were calculated at the internal carotid artery (ICA), transverse sinus (TS), straight sinus (SS) and superior sagittal sinus (SSS).Results: Compared to the SENSE4 scan, fFor the left ICA C2, mean velocity measured by CS8 and CS10 groups, and 3D WSS measured by CS6, CS8, and CS10 groups were underestimated; ; for the right ICA C2, mean velocity measured by CS10 group, and 3D WSS measured by CS8 and CS10 groups were underestimated; for the right ICA C4, mean velocity measured by CS10 group, and 3D WSS measured by CS8 and CS10 groups were underestimated; and for the right ICA C7, mean velocity and 3D WSS measured by CS8 and CS10 groups, and average axial WSS measured by CS8 group were also statistically underestimated compared to the SENSE4 scan (all P < 0.05). For the left TS, max axial WSS and 3D WSS measured by CS10 group were significantly underestimated (P= 0.032 and 0.003). Similarly, for SS, mean velocity, peak velocity, average axial WSS measured by the CS8 and CS10 groups, max axial WSS measured by CS6, CS8, and CS10 groups, and 3D WSS measured by CS10 group were significantly underestimated compared to the SENSE4 scan (P = 0.000 to 0.021). The hemodynamic parameters measured by CS4 group had only minimal bias and great limits of agreement compared to conventional 4D flow (SENSE4) in the ICA and every venous sinus .