Abstract
Proton therapy (PT) has the potential to deliver conformal doses to the tumor while sparing normal tissue, but is highly susceptible to treatment uncertainties. The occurrence of anatomical changes during PT treatments has a major impact on the delivered dose, often necessitating plan adaptations that are typically performed offline and require a few days before the adapted plan is ready. In order to react promptly to detected anatomical changes, online adaptive proton therapy (OAPT) has been proposed with the goal of adapting the plan while the patient is on the treatment couch. First OAPT workflows for daily plan adaptation that are effective against interfractional anatomical variations have reached clinical application. However, even faster OAPT workflows are needed to cope with faster anatomical changes. Near real-time adaptive PT (NAPT) relying on online in vivo treatment verification can be a potential solution for many tumor entities (e.g., thoraco-abdominal tumors), which would greatly benefit from the conformality of PT, but are presently challenging to treat with proton beams due to the influence of intrafractional variations. In addition, NAPT offers the opportunity to achieve the long-awaited closed PT feedback loop. In this paper we review the required tasks and necessary components in an OAPT workflow for the application of near real-time adaptation, proceeding sequentially from volumetric imaging for online plan adaptation up to online verification during delivery. Available technology and upcoming developments are discussed. Several aspects regarding regulatory approval, cost-benefit related issues and additional beyond-the-loop tasks are also addressed.
1 Introduction
Due to the characteristic behavior of proton beams and their energy release in tissues, proton therapy (PT) is particularly sensitive to anatomy variations occurring during the course of the treatment (). A modification of the patient anatomy may result in underdosage to the tumor, unnecessary high doses to the normal tissue or even overdosage to the organs-at-risk (OAR), potentially leading to severe side effects. PT remains constrained for certain entities (), particularly in the abdominal region. In parallel with the advent of latest technological developments, new radiotherapy (RT) treatment modalities have emerged that introduce an adaptive approach. The concept of adaptive radiotherapy (ART) based on a systematic feedback loop is well-known () and can be classified into three main time regimes: offline (adaptation performed between fractions), online (adaptation performed directly prior to delivery, while the patient is on the treatment couch) and real-time (adaptation performed within fraction delivery) (). Online ART has become a reality in photon-based radiotherapy (XRT) with the availability of hybrid systems of imaging plus linear accelerator (LINAC), like magnetic resonance (MR)-LINACs (), tomotherapy () and systems based on cone-beam computed tomography (CBCT) (). Within ART, positron emission tomography (PET)-integrated systems have also been recently proposed () that target biological changes in the tumor rather than anatomical variations, thereby propelling the concept of biology-driven adaptation ().
Robust optimization in PT, as implemented in commercial treatment planning systems (TPS) today, accounts only for positional shifts within an otherwise unchanged anatomy. However, anatomical changes, such as variations in cavity filling or tumor size, can lead to clinically significant differences. Given the large impact of anatomical variations on the conformity of the proton dose distribution, the spread of online ART has sparked interest for potential applications in PT (, ), with many PT centers actively working towards clinical implementation of this novel approach (–). The concept of online adaptive PT (OAPT) is not yet as established as its photon counterpart and refers generally to the opportunity of performing a plan adaptation (and related tasks) while the patient is on the treatment couch during the daily treatment session. This review focuses on online adaptations due to anatomical changes. There are different flavors of OAPT (Figure 1). They depend on the one hand on the timing of the addressed anatomical changes (and the necessity of potential plan adaptations per treatment and fraction) and on the other hand on the time needed to perform the adaptation itself ().
Figure 1
Interfractional changes like weight changes, tumor shrinkages or growth, relevant changes in cavity filling, and substantially different fill levels of the digestive organs with related shifts of nearby soft tissue happen on a timescale of hours or days (). These variations can be dealt with daily adaptive PT (DAPT), with the goal of adapting the plan to the daily anatomy. In a standard DAPT workflow (Figure 1A), the online plan adaptation follows the conventional setup and 3D imaging. This way the treatment plan is adapted to the current daily anatomy based on an image acquired prior the delivery. Attempts at implementing such a workflow have been recently proposed () and first clinical realizations have been reported (, ).
Generating the online adapted plan at each fraction places an undeniable burden on patient comfort and daily workload. To reduce the daily burden, an evaluation-based DAPT “on demand” is envisaged (Figure 1B), where the online adaptation is possibly performed after an assessment of its need based on the daily acquired 3D image and/or projected dose distribution or even on the information from pre-delivery range assessment like range probing (). If there is no need for adaptation, the workflow can continue with the delivery of the initial treatment plan.
However, in order to address faster changes in organ position and deformation, an even faster action is required that is capable of detecting and correcting for intrafractional variations. Such anatomical changes can happen on timescales of minutes (like urinary activity, organ drift), or seconds (like peristalsis, respiratory movements, cardiac motion) (–). In particular, respiratory movements are usually handled prospectively via techniques of motion management, like suppression of motion, 4D imaging and optimization, continuous motion surrogate monitoring, and appropriate changes in the beam delivery sequence (gating, repainting), etc. (, ). Motion management alone, however, is not able to compensate for the entire intrafractional range of movements, especially when dealing with irregular and/or unexpected variations, highlighting the need for a near real-time adaptive approach that includes online generation of an adapted plan (, ). For example, entities like thoraco-abdominal (e.g., lung, esophageal, liver, pancreas), gynecological and genitourinary tumors are greatly influenced by unexpected intrafractional variations and located in close position to gastrointestinal, cardiac, pulmonary and/or urinary OAR (, ), thus requiring steep dose distributions. For such entities a shift towards near real-time adaptation has already been observed in XRT after the advent of MR-guided radiotherapy (–). This emphasizes the necessity of adopting similar approaches for PT, which can offer superior conformality but is presently limited. With near real-time adaptation in PT, analogous to XRT, it would be possible to continuously monitor treatment delivery and promptly react to irregular intrafraction anatomical changes as they occur. This differs from standard or evaluation-triggered DAPT, in which the adaptation is only performed before the treatment delivery. Near real-time adaptive PT (NAPT) (Figure 1C) can address this issue by adapting the treatment plan using information on treatment deviations from online treatment verification systems that run concurrently with delivery. Upon identification of a deviation between the delivered and the expected proton dose, an online adaptive workflow is initiated with the acquisition of a 3D image and the assessment of the adaptation need. On the one hand, this approach would allow a rapid response to unforeseen anatomical changes and, on the other hand, it would provide a safety net to ensure that the planned dose is delivered as indicated and that further plan adaptation is only applied when necessary. This NAPT workflow, which involves imaging, contour generation, plan adaptation, and approval under stricter time constraints, is not yet available in a truly real-time setting due to the time currently required to complete these critical operations before resuming delivery.
In this sense, at the time of this review an actual “real-time” adaptive PT can only be represented by operations of motion management happening during the daily treatment session that do not require a beam delivery interruption or a treatment plan modification, leading instead to adjustments of the beam geometry or delivery sequence (, , ). These real-time operations have also been previously referred to as an in-line adaptive workflow ().
This review outlines step-by-step the necessary components within a daily treatment session to implement a NAPT closed feedback loop after standard patient setup, describing the current research status and highlighting areas that require further investigation. Each major step in the workflow (volumetric imaging for online treatment planning, online contour generation, decision on online adaptation need, online plan adaptation – optimization strategy and dose calculation, pre-delivery quality assurance of the online adapted plan, online treatment verification) will be discussed in dedicated sections. Additional aspects regarding integration, regulatory approval and tasks that are performed outside the loop will also be addressed.
2 Volumetric imaging for online treatment planning
The availability of 3D imaging in the treatment room, preferably keeping the patient in treatment position, is essential for identifying anatomical changes in the timeframe for OAPT. Given the stricter timeframe and potential for adaptation during paused/suspended delivery, 3D imaging in treatment position is especially relevant for in NAPT. The quality of the acquired images must be adequate for the online plan adaptation of proton-based treatments, i.e., it must ensure the correct retrieval of the stopping power ratio (SPR) for accurate range and dose calculation. Image-guided XRT has propelled the development of volumetric image guidance tools (), many of which are now being investigated for OAPT. Here pre-delivery imaging techniques based on in-room (fan-beam) CT, CBCT, MR imaging (MRI) and proton CT are discussed in terms of technological advancement, the ability to retrieve the current 3D patient’s anatomy and related SPR map for proton dose calculation, and the suitability for near real-time adaptation (Table 1). In-room imaging tools providing additional information for real-time motion management are briefly mentioned.
Table 1
| Imaging technique | SPR prediction | Imaging in treatment position (at isocenter) | Continuous imaging during delivery | Current application in PT | Strengths (+) | Limitations (-) |
|---|---|---|---|---|---|---|
| In-room (fan-beam) CT | Clinical gold standard, direct SPR prediction with spectral information (DECT or photon counting CT) | No for CT on rails; Yes for upright PT systems | No | For daily positioning and routine control CTs acquisition in clinical treatments; use as volumetric imaging for online adaptation in first clinical DAPT application has been reported () | + Established technique + High geometrical accuracy + Accurate quantitative information (CT number accuracy) + Low-dose imaging protocols available for online applications + 4D-CT available + Used in first clinical DAPT applications | − Imaging not in treatment position − Potentially complex integration into the workflow − Time required to move patient and CT system − No real-time monitoring during treatment |
| CBCT | Correction techniques required to overcome image artefacts | Yes, several configurations available: nozzle-mounted, gantry-mounted, ceiling-mounted, couch-mounted () | No | For patient set up and daily imaging in clinical treatments | + Imaging in treatment position + Already available for online adaptive XRT + Several vendor/configuration options + 4D-CBCT available | − Several image quality issues (field of view, CT number accuracy) currently limit its usage for proton treatment planning (not yet clinical) − No real-time monitoring |
| MRI | Conversion to CT numbers or SPR required | Potentially yes, in the foreseen in-room/integrated configuration for fixed beamlines (, ); | Potentially yes (foreseen for fixed beamline configurations); not currently available | Preclinical research and first clinical applications with shuttle-based systems (, ) | + No additional ionizing radiation + Enhanced resolution for soft tissue + Better delineation of tumor due to higher contrast + Potential real-time monitoring + First clinical experiences available (shuttle-based systems) | − Integration in the treatment room limited to fixed beamlines (not compatible with gantry) − Mutual interference of proton beam and magnetic field − Conversion of MR data to CT numbers for treatment planning required (additional uncertainties) − Inferior geometrical accuracy (distortion) |
| Proton CT | Direct SPR information | Yes, relying on the same system used for treatment delivery | In principle possible (mixed beam with different energies) | Preclinical research (, ) | + Direct retrieval of SPR | − Degraded information due to multiple Coulomb scattering − Limited to very few entities due to geometrical conditions (distance between detectors before and after patient) − Not available as medical product |
Overview of the main in-room pre-delivery volumetric imaging systems depicting their properties and suitability for online (near real-time) plan adaptation in PT.
In-room (fan-beam) CT is a well-known pre-treatment imaging solution for checking the anatomy in treatment position and for setup and positioning. Developments of mobile CT scanners (Figure 2A) to perform in-room CT for OAPT have been observed (, 38) and proposed for clinical use (39), although the integration in a fast and efficient workflow is challenging. CT scanners on rails have a fixed position in the room (Figure 2B). Despite the associated costs and impact on the overall available space in the treatment room, they are at present gaining traction for use in DAPT workflows (, ). In-room CT offers several advantages regarding image quality, the most accurate prediction of the SPR for PT planning including dual-energy CT (DECT) based SPR prediction (39, 40) and the option to perform 4D-imaging to monitor interfractional changes in the respiratory pattern (), however, it is currently not optimal for near real-time adaptation. In-room CT scans are acquired distant from the actual delivery position and usually incapable of returning images matched to the isocenter. Before proceeding further with the adaptation, the in-room CT images must be first registered to the reference image (e.g., planning CT), in order to share the same reference frame. Moreover, the reference frame of the in-room CT and its position within the treatment room must be also correctly related with the room coordinate system. In order to correctly locate at room isocenter a specific point of the treatment volume identified in the CT scan, a precise and reproducible position of the robot for patient positioning (or patient positioning system) is required. Securing sufficient accuracy of the patient table position over the large distance from the PT gantry to the CT gantry is paramount. The integration of in-room CT systems with the delivery system still represents a challenge and is heavily facility-dependent (). As a result, the insertion of such systems into a closed feedback loop for near real-time adaptation in PT is a complex undertaking. Still, in-room CTs are first choice for OAPT workflows as the fan-beam CT imaging is the gold-standard for accurate SPR prediction and dose calculation (39).
Figure 2
Over the past decade, CBCT has become available in newly built PT centers, with commercially available options from respective PT vendors seamlessly integrated into the treatment room. In detail, there are different configurations, like gantry-mounted (Figure 2C), mounted on robotic C-arms (Figure 2D), nozzle-mounted (Figure 2E), couch-mounted (Figure 2F), or ceiling-mounted (Figure 2G) (
The CT-based solutions discussed so far assume that the patient is imaged and treated in recumbent position. Recently, interest in upright treatment and imaging has increased significantly (53). The idea of treating patients in a seated position is not new (54), but has gained renewed attention due to the development of commercial systems for vertical CT imaging together with reported clinical applications (55–58). The main driving factor towards the investigation of upright CT imaging and treatment is the opportunity of reducing the costs of the expensive PT facilities. In fact, employment of upright positioning would allow for a gantry-less treatment room, where the patient is rotated relative to the fixed beam instead of moving the beam around the patient with a gantry (59). An additional aspect to consider is that upright CT imaging offers in principle high flexibility and the opportunity of imaging at the isocenter. This is relevant for OAPT treatments, especially when aiming at faster time regimes. For certain tumor entities, treatment in an upright patient position may offer potential benefits and, in some cases, could even be more comfortable for patients. However, given the limited experience with this emerging technology – including specialized treatment chairs and volumetric imaging systems (59, 60) – several challenges remain. These must be addressed before definitive conclusions can be drawn regarding the treatment quality and clinical outcomes of upright PT. The positioning accuracy and reproducibility need to be quantified for different treatment sites. Upright PT will very much rely on fast online adaptation capabilities in case the positioning accuracy is inferior to supine treatments and most importantly clinical evidence of safe and efficient treatment needs to be generated (61, 62).
Finally, MRI is noteworthy for its ability to provide high-quality soft-tissue imaging without exposing patients to additional ionizing radiation from imaging. In XRT, MRI has already been integrated into daily adaptive workflows using MR-LINACs, where it enables continuous anatomical monitoring during treatment. Although MRI offers superior soft-tissue contrast compared to CT, it is inferior in terms of geometrical accuracy (distortions) and dose calculation applicability. The enhanced soft-tissue contrast may be relevant for NAPT and for challenging cases affected by significant intra-fractional variations (
CT imaging with proton beams instead of X-rays is a promising technology to overcome the uncertainty in the conversion from X-ray attenuation to SPR (77–80). Despite ongoing research activities over the last decades, this technique is still in an early stage of development, i.e., first prototype proton CT scanners (Figure 2I) are constructed and experimentally validated, but a clinical system is currently not available (
Further non-volumetric imaging techniques may be combined with the above mentioned, in order to reconstruct the full range of intrafractional movements in the patient. Exemplary solutions are the visualization of external markers, the detection of internal markers or electronic transponders, online fluoroscopy for correlation of internal motion and external surrogate signal, ultrasound imaging and surface tracking (85). These motion management solutions are already quite established in image-guided routines and do not constitute per se an issue for their integration in a near real-time adaptive workflow, although stricter timing criteria and the accurate combination of multimodal information must be considered. A full review of motion management systems suitable in PT has been reported (
3 Online contour generation
One of the challenges associated with imaging for online plan adaptation is the rapid generation of contours on the 3D images, which provide the essential anatomy labels for the adapted plan and are crucial for performing dose-volume histogram (DVH)-based assessments. Despite being a long-known issue (86), this remains a major bottleneck for the development of an OAPT workflow (and online ART in general) (
Contour propagation based on (rigid or deformable) registration has already been investigated for clinical use in OAPT (
Auto-segmentation based on artificial intelligence (AI) is gaining increasing popularity and several approaches for auto-contouring are available or rapidly evolving. The development of deep learning (DL)-based segmentation in particular has driven a great number of studies lately (105). DL auto-contouring may in the future have the potential to outperform previous solutions, due to its capability to identify features in the data thanks to prior training and it may then be applicable to a broader range of tumor sites. However, current limitations stem primarily from the size and representativeness of the datasets used for training. Despite these limitations, many commercial software solutions offering DL-based auto-contouring are already available and are being extensively researched (106–108) and evaluated in clinical settings (109). Studies evaluating the impact of DL auto-contouring on contouring time have also been reported (110, 111), including comparisons with other auto-contouring techniques, such as atlas-based or DIR-based methods (112). Time savings compared to manual contouring have been shown, although no clear advantage for a specific method has been observed, and corrections are still required. Recently, strategies to account for the uncertainties related with the auto-contouring within the optimization process have also been proposed (113, 114). These methods could effectively mitigate the impact of contour uncertainties thus facilitating the adoption of automated contouring tools. However, further research is required at this stage.
Additional contouring best practices should also be considered. First and foremost, it is crucial to emphasize that contouring protocols must remain consistent between the planning CT and the additional images required for OAPT. Best practices include using pre-determined contouring strategies for specific structures, maintaining consistent naming conventions, and clearly differentiating between targets, OAR, and normal tissues. Further research is expected on the topic, with recent studies suggesting that manual corrections might be required only for the target contours and discarded for the OAR (115). A consensus has not been reached yet.
Moreover, the development of auto-contouring techniques for OAPT requires the definition of rigorous safety checks on the resulting anatomical structures. In fact, evaluating the acceptability of the contours and establishing suitable assessment metrics is another essential part of the process (95, 104, 116), and automatic safety protocols are of utmost importance for the application of auto-segmentation in OAPT. Final checks on the quality of the current structures can be considered as a part of the online quality assurance (QA) process and will be discussed in more details in the dedicated section.
4 Decision on online adaptation need
In conventional scheduled DAPT (Figure 1A), adaptation is performed at every treatment session. Alternatively, adaptation can be triggered (Figure 1B, C) while the patient is on the treatment couch, based on new 3D imaging, a re-calculated dose from the scheduled plan, or external signals such as treatment verification data. In this context, triggered NAPT (Figure 1C) can help reduce overall workload, since the adaptation would not be necessarily performed at every fraction. Online adaptation may be more efficient when performed only when clinically necessary or when a clear benefit is anticipated. This means, it is crucial to rapidly assess whether adaptation is deemed advantageous on the basis of the collected data (e.g., 3D image in treatment position, updated contours, re-calculated dose distribution and DVH parameters), before proceeding with the adaptive workflow. For photon-based ART, several workflows have been proposed depending on the timing of the adaptation, frequency of imaging and tumor characteristics (117). The approach of directly performing a plan adaptation, as exemplified in certain MR-LINAC workflows (
A straightforward approach to assess the need for adaptation is the use of pre-determined thresholds and checklists that include anatomical and dosimetric information, such as variations in structure volume, center of mass, and DVH properties for the scheduled plan on the updated anatomy (
Moreover, the development of automated decision support platforms for assessing the need for adaptation is expected to be greatly advanced thanks to AI. Several AI-based decision support systems have already been proposed in photon-based ART for different purposes, such as plan auditing (118), individualized treatment decisions for head and neck cancer (119, 120) and esophageal cancer (121), with the first commercial ART software platforms available for deciding when a plan adaptation is needed (122).
It’s important to note that the decision to adapt could be made at different levels depending on the availability of collected data, by extracting and processing relevant features. This could involve using only the anatomical information from 3D images, incorporating contour data, considering information from the dose distributions only, or a combination of these factors. Various AI techniques could be used to support this process. The capacity to extract pertinent information about the adaptation need from the images alone, without the necessity for initial plan recalculations or even updated contours, would facilitate a significantly more expedient decision-making process.
5 Online plan adaptation: optimization strategy and dose calculation
The online plan adaptation process involves two key aspects: the optimization algorithm used for the adapted plan generation and the dose engine used for the dose calculation. Relevant research in both areas is ongoing in the PT community. Several solutions to fasten the plan adaptation process have been proposed and comprehensively discussed (
In PT, the time required for standard plan optimization is largely determined by the number of robustness scenarios considered. In contrast, XRT can account for setup errors through an appropriate planning target volume (PTV) concept, and range uncertainties are negligible. Therefore, reducing optimization times is particularly critical for enabling fast online adaptations in PT compared with XRT. The main strategies for generating online adapted plans that align with the current anatomy can be broadly classified based on the underlying optimization technique and/or on the final goal of the adaptation (123). The principal plan adaptation strategies are: “standard” region-of-interest (ROI)-objective-based full re-optimization, constrained re-optimization retaining specific properties of the initial plan with or without the goal of reproducing the initial dose (dose restoration), and dose-mimicking-based full re-optimization (which can be tuned to mimic different dose distributions depending on the application). Furthermore, the use of plan libraries has been put forth as a potential strategy for OAPT (123). Additional adaptive strategies that could be applied after a partial delivery of the plan are particularly relevant for NAPT and will also be addressed down below. A detailed outline of the discussed strategies is presented in Table 2.
Table 2
| Adaptation strategy/technique | Implementation types | Suitable workflow application | Strengths (+) | Limitations (-) |
|---|---|---|---|---|
| Online ROI-objective-based “standard” full re-optimization Optimize energies, spot positions and spot weights with the goal that the resulting dose distribution fulfils requested DVH criteria | • Full new plan generation including beam setup according to the initial trial-and-error planning pipeline • Full re-optimization with pre-determined objectives for a daily template plan or, more specifically, with the initial plan optimization problem ( | Upfront online adaptation before starting daily delivery | + Clinical standard for offline adaptation + More flexibility in the optimization + Effective against drastic anatomy changes + Pre-established optimization templates enhance comparability and simplify both initial and adaptive planning + Clinically proven: first reported DAPT treatments based on the full re-optimization of a daily template plan | − Longer optimization process − High level of online QA required |
| Online constrained re-optimization retaining specific properties of the initial plan Adjust properties of given plan (e.g., modify spot position, energy and weight; add selected spots) to comply with new anatomy | • Dose restoration, when the goal is to reproduce the initial dose distribution (125–127) • Range shift compensation, with or without spot weight tuning and spot addition (e.g., for a subset of spots), not necessarily replicating the initial plan dose (96, 128–131) | Upfront online adaptation before starting daily delivery | + Related online QA expected to be less burdensome as specific plan properties (e.g., number of energy layers, dose distribution) are not/less modified + Keeping the initial dose distribution unchanged as in the dose restoration approach facilitates physician’s approval | − Unsuitable for drastic anatomy changes |
| Online dose-mimicking-based full re-optimization of a reference dose Optimize energies, spot positions and spot weights with the goal to deliver a given 3D reference dose distribution to the current anatomy | • Purpose defined by the reference dose to be mimicked ○ AI-predicted dose distribution (124, 132–134) • Rigid/Deformably propagated initial dose distribution (dose restoration approach) (124, 132, 135) | Upfront online adaptation before starting daily delivery | + The opportunity to select different reference doses grants a flexible approach + Less burdensome than the traditional optimization workflow (based on ROI-objectives) + Mimicking of the initial dose as in the dose restoration approach may facilitate physician’s approval | − Depends on the ability of the mimicking algorithm to accurately reproduce a reference dose − AI-prediction of a reference dose is limited by the training dataset − Mimicking of the initial dose as in the dose restoration approach is unsuitable against drastic anatomy changes − High level of online QA required |
| Plan library No plan optimization during the treatment session; just select the most suited plan | • Pre-treatment creation of a variety of plans to counteract potential anatomy changes (136–139) • Creation of the plan library alongside the treatment course (140) | Selection of the daily plan before starting daily delivery | + Potentially minor disruption to daily workflow as all plans are created beforehand and are not re-optimized online + Usual PSQA can be performed offline | − May not be representative of the current anatomy, ending up with further offline plan adjustments − Accurate plan-of-the-day selection criteria needed − Extra time and resources required for multiple plan creation and QA before the treatment |
| Online partial adaptation Adapt only a not-yet delivered part of a plan but consider the sub-optimal dose delivered by the non-adapted plan portion | • Dose mimicking of a reference dose considering the already delivered (background) dose (141, 142) • Range shift compensation considering delivered probing beam dose (143, 144) | Near real-time adjustments, e.g., triggered by treatment verification | + Allows to address detected treatment deviations during delivery (verification-triggered adaptation) + Adaptation may be performed only when necessary (e.g., in presence of a detected deviation), thus reducing the related burden on the daily workflow + Time-saving against standard DAPT, allowing to start treatment with non-adapted field and adapt the remaining portion of the plan in the meantime | − Specific online QA checks required to ensure safety of the partial plan irradiation |
Overview of the main proposed adaptive strategies for OAPT.
The generation of a fully new plan (which entails new energies, spot positions and weights, and potentially also beam angles and/or air gaps) following the initial (ROI-objective-based) pipeline and the conventional trial-and-error approach is still the gold standard for offline adaptation (
Differently from full re-optimization, online constrained re-optimization techniques enforce direct adjustments of plan parameters (96, 123, 128–131), such as the spot energy/range, isocenter position and/or tuning spot weights for a subset of spots, rather than ROI objectives, offering a faster and less time-intensive online plan adaptation. In fact, in these adaptive methods, the spot positions and energy layers shall remain unchanged compared to the initial plans. This strategy is subject to some limitations, in that it may prove inadequate for compensating for significant anatomical variations and deformations. In such cases, the addition of energy layers or other more substantial modifications to the original plan may be required. These techniques have been applied with or without the goal of restoring the initial planned dose and have been previously referred to as simply “re-optimization” (
Dose mimicking has been proposed (124, 132–134) as an alternative adaptive plan optimization strategy, where the optimizer aims at generating a completely new treatment plan (energies, spot positions and spot weights) that delivers a certain given reference dose distribution, similar as to what has been initially investigated in the context of XRT (145). The dose mimicking operation can have different goals, depending on the reference dose to be replicated. For example, the reference dose may be an AI-predicted dose on the new anatomy, in a “dose prediction plus mimicking” approach (132, 133, 146, 147). However, the reference dose may also be the dose distribution of the initial plan copied to the new image (132). This last approach aims at restoring the initial dose cloud and can therefore be regarded as another technical realization of dose restoration. Deformed dose restoration based on dose mimicking has been investigated as a potential solution to address some of the limitations of the rigid approach. The aim is to restore a dose cloud that has been adapted to the new anatomy through the use of guided dose deformation (135), where specific control structures serve as guiding reference for the deformation.
An explored alternative to online plan adaptation is the use of pre-determined plan libraries, where multiple treatment plans are created in advance to account for potential anatomical variations (e.g., organ filling changes). These plans undergo the elaborate routine QA before they can be selected for delivery based on the patient’s anatomy at each fraction, minimizing workflow disruption. This approach, which hence does not represent a truly online plan adaptation, is not expected to become standard clinical practice in PT (148, 149). Several retrospective planning studies across different tumor sites exist (136–139), and a first clinical application for an intra-thoracic tumor has been reported (140). In silico studies have explored various methods for generating plan libraries, such as motion modeling, marker tracking, and adjusting setup margins using retrospective control CTs. In the reported clinical case, the plan library was built progressively during treatment based on acquired control CTs. Inherent limitations of the strategy prevent a broader adoption of plan libraries. The plans available in the library may not include the full range of geometry variations and may end up being not suitable for the daily delivery; this is particularly relevant for NAPT aiming to tackle unexpected intrafractional changes.
Adaptive strategies as the ones described so far focus on the adaptation of the entire plan before starting any beam delivery in the daily treatment session. To streamline daily adaptive workflows, more flexible approaches are desirable – such as selectively adapting only portions of the plan, while a part of the fraction dose is already delivered. This can enable parallel operations, thereby improving overall efficiency (141). Dealing with a sudden change detected during delivery as in the foreseen near real-time adaptive setting is an additional challenge. The employment of online in vivo treatment verification will not only offer a tool to control the compliance of the delivered (adapted) treatment, but will also enable a transformation in the way the online adaptation is applied. In fact, the detection of a treatment deviation above an established threshold may serve as a signal to trigger a subsequent image acquisition in treatment position and, ultimately, an adaptation of the plan in the fully closed near real-time adaptive loop. This so-called adaptation triggered by treatment verification might in principle happen at any time during the delivery, propelling the need for plan adaptation strategies to be employed mid-delivery or even multiple times during delivery. In this context, the concept of a “partial adaptation”, whereby only a portion of the plan is adapted, is also of paramount importance. Without the ability to adapt only the undelivered portion of the scheduled plan in response to a detected treatment deviation and to consider the delivered dose up to the trigger, NAPT triggered by verification would be unfeasible. Planning strategies that introduce a partial adaptation for PT have recently been proposed (141, 143) and further investigated in combination with the potential trigger information from a treatment verification system (142, 144).
A fast dose calculation is essential within the iterative adaptation algorithms and beyond. Dose calculation plays also a key role in multiple steps of the online adaptive workflow, including initial plan recalculation on the current anatomy for adaptation decisions and independent dose verification for QA (
Finally, several studies propose novel AI-based approaches for proton dose computation, such as knowledge-based planning (160), long short-term memory networks (161) or transformers (162). Many of these approaches aim at achieving the same level of MC accuracy at shorter calculation time (157, 163–165). There are still limitations in applicability to different sites or different field arrangements than the ones included for the training of the AI algorithms. Hence recent works focus on generalizable algorithms that could return MC accuracy in sub-second speed, which could be potentially relevant for near real-time applications (166), if not during optimization then at least for fast secondary dose calculation in the online QA (see next section), including AI-based dose calculation models specifically designed for OAPT (167–170).
6 Pre-delivery QA of the online adapted plan
QA of the treatment plan, so-called patient-specific QA (PSQA), is an essential step to ensure the treatment safety. PSQA aims at checking 1) that the calculated dose distribution for the plan is correct, 2) the correct data transfer between the TPS and the delivery system, and 3) the deliverability of the plan (171). In most PT centers, PSQA still involves dosimetric measurements for plan deliveries to water and water-equivalent phantoms, which are time-consuming and limit the availability of the treatment room. Recently, the combination of log-file-based verification with an independent dose calculation became a valid alternative for the above-described three aims, that already found its way in clinical application in several centers (
For example, soft checks can be applied to evaluate the obtained structures on the actual anatomy, ensuring that the online contouring process has been performed correctly. This is particularly important when contouring is automated using dedicated software, which is desirable given the time constraints of OAPT. Exemplary structure parameters can be selected to be part of this structure pre-delivery QA to assess intensity and geometry, like the center of mass, the volume, the average CT numbers of specific structures, etc. (
Proton range assessment using range probing is an additional pre-delivery QA method, also applicable to OAPT workflows, and is currently under translational research. Its purpose is to verify the accuracy of the SPR prediction used in treatment planning (175, 176). Range probing utilizes high-energy, low-dose proton beams (177) to measure the remaining energy of protons after traversing the patient. This is achieved using either multi-layer ionizazion chambers (MLIC) (178) or flat panel detectors (179), allowing for a direct comparison between the measured energy loss and the simulated values from the planning CT or a pre-treatment image of the current fraction, when available. In the context of OAPT, range probing is primarily intended to validate CBCT-based SPR prediction (179–181). Following initial proof-of-principle and phantom-based validations, the first proof-of-concept clinical implementation has been performed (
7 Online treatment verification and detection of deviations during delivery
In vivo treatment verification during delivery has specifically two main goals: 1) to provide a safety net, checking that the adaptation worked out as intended; 2) to detect deviations due to inter-/intrafractional changes in the anatomy that may trigger NAPT. It is also highly desirable when aiming at reduced target margins. The first-to-clinic daily adaptive workflows do not include treatment verification, and are based solely on the detection of anatomical changes on the daily image taken before the delivery is started (
Table 3
| Verification technique | Flavors | Timeframe | Integration in the workflow | Clinical readiness | Strengths (+) | Limitations (-) |
|---|---|---|---|---|---|---|
| PET-based (PT-PET) | • In-beam PT-PET • In-room PT-PET • Offline PT-PET | A few minutes dependent on the activity build-up of positron-emitting isotopes | In-beam PET with smallest effect on clinical workflow (additional time need for acquisition after delivery), while in-room PET reduces patient throughput and offline PET requires additional resources (e.g., close-by PET scanner) | In clinical studies for treatment verification of proton and carbon ion treatments with in-beam detectors relying on common positron emitters. No medical product available | + Broad clinical experience (with hundreds of clinical applications) + Information available shortly after delivery + Commercially available detector technology from diagnostic PET can be utilized | − Information from different spots and fields not differentiable due to stochastic decay mechanism − No real-time capability − Bulky and complex instrumentation − Biological wash-out − Limited accuracy (also due to time until positron decay and low counting statistics) |
| Prompt gamma-based (PGTV) | • PGI • PGT • PGS • Multi-feature | Online; prompt gamma rays produced almost instantly (few ns) | Easily integrable with minimal effect on clinical workflow (acquisition during delivery) | In clinical studies for online verification (PGI and PGS). No medical product available | + Available during delivery + Produced quasi-instantaneous allowing spot-wise resolution and evaluation + Broad clinical experience (with hundreds of clinical applications) + Different evaluation approaches available (binary classification for relevant treatment deviation or eventually even 3D dose reconstruction) | − Complex and partly bulky instrumentation |
| Range probing | • MLIC-based • Flat-panel based | Before treatment delivery | Additional time needed to deliver and evaluated the range probing spots | In clinical study for online verification in at least one center. No medical product available | + No issues with multi Coulomb scattering (measurement compared with expectation) + Information available before treatment (e.g. on CBCT validation) + No advanced detector technology required | − Additional time required in treatment room. − No differentiation, if detected changes affect treatment field or not − Not during dose delivery (changes after range probing stay undetected) |
Overview of the close-to-clinic in vivo treatment verification approaches for PT in terms of their suitability for near real-time adaptation.
For completeness, range probing is included given its advanced status, even though it is generally applied before delivery of the therapeutic treatment and not during delivery, differently from PT-PET and PGTV.
One of the first in-vivo verification techniques studied was PET imaging during particle therapy delivery (previously also referred to as PT-PET) (182–184). Positron emitters (10C, 11C, 15O among the most commonly used) are produced via nuclear interactions between the incident protons and atoms of the patient’s tissue and PET-based treatment verification relies on the coincident detection of 511-keV photons produced by the annihilation of the emitted positrons with electrons. Unlike particle therapy with heavier ions, when using proton beams positron emitters can only originate from the target atoms, rather than from projectile fragmentation. Because the nuclear interactions and thereby the positron emission are closely related to the dose deposition, the positron activity can be used to approximate the proton beam range, usually by comparing the measured activity distribution with an expected one derived from the treatment plan (185) and deriving a range deviation. Given that PET imaging is a well-established diagnostic technique, there has been considerable interest in adapting it for the use in in vivo treatment verification (186). Several solutions have been proposed, each with their specific advantages and disadvantages: online in-beam PT-PET involves PET imaging directly after each field delivery, and sometimes during irradiation; online in-room PT-PET entails PET imaging with an in-room PET scanner after the patient is moved to the scanner position post-treatment; offline PT-PET requires transferring the patient to a nearby room, where PET imaging is performed using a standard diagnostic PET scanner (187). While offline PT-PET can rely on standard diagnostic systems, online PT-PET systems are usually dedicated in-house developed systems that can be used in the room while the patient is on the couch, hence they must respect certain geometrical constraints (188).
While many research groups have investigated PET-based range verification in clinical studies (184, 188, 189), and several hundred patients have been monitored at a few pioneering institutions, the method has not found its way into broad clinical application. This is probably due to the inherent limitations of the approach: due to the half-life time of the positron emitters in the range of minutes for the most prominent positron emitters (20 min for 11C, 2 min for 15O and 19 s for 10C) a spot-wise evaluation is not possible, as the signals from different pencil beam spots and even from subsequent treatment fields overlap each other. Additionally, the biological wash-out effect (190), which describes the movement of positron emitters between their generation and decay, leads to a blurring of the original activity distribution. Intrafractional movement needs to be monitored and considered correctly to correlate the locations of positron emitter generation and dose deposition with the location of positron emission (191–193). The positron activity produced is relatively low, around 6.6 kBq per Gy and cm³ (194), compared to 10–100 kBq/cm³ in diagnostic nuclear medicine. As a result, the derived signal is heavily influenced by statistical uncertainty, and interpreting clinically relevant changes at the dose distribution level remains challenging, particularly due to the differences in activity and dose distribution, especially when using proton beams.
Despite its limitations and restricted applicability, research in the field of PT-PET continues. An experimental intrafractional adaptive PT workflow based on PET verification has recently been proposed and investigated pre-clinically (144). Additionally, the use of short-lived isotopes, like 20N with a half-live of 11 ms (195, 196), has been suggested, however their activity concentration is about two orders of magnitude below that of conventional emitters mentioned above. Further developments of in-beam PET detectors (197), application of AI to the obtained PET images (198), time-of-flight PET (199, 200) and combined techniques (201) have also been reported.
Currently, the most promising in vivo treatment verification technique is the detection of prompt gamma (PG) rays, i.e., photons that are almost instantaneously (<ns) produced after nuclear interactions of the beam with the nuclei of the traversed tissues (
The use of PG rays for proton range verification was first proposed in 2003 (202). A range of different PGTV techniques is now actively studied, the main ones comprising PG imaging (PGI) (205), PG timing (PGT) (206) and PG spectroscopy (PGS) (207). PGI aims at recording the spatial distribution of the emitted PG rays, PGT is based on their temporal distribution, and PGS relies on the local energy spectrum of the emitted PG rays, relating the PG spectrum to the proton-energy-dependent cross section of the respective nuclear reactions. With improvements in detectors, different domains of information could also be combined, enabling so-called multi-feature treatment verification.
PGI is currently the most advanced of the available PGTV techniques. It is in clinical application since 2015 (208, 209) and has since been applied in more than 470 field deliveries within a retrospective clinical study (210), proving the feasibility under realistic conditions. PGS is in clinical application since 2020 (211). Recently, a multi-institutional comparison of PGI and PGS has been performed under controlled conditions, providing standard conditions and methodology that allow fair comparisons and performance characterization of different PGTV systems, which are hopefully used in future studies proposing novel PGTV techniques (212). Accuracy of PGTV systems on spot level is currently around 2–3 mm for range verification, given a high spot dose (205). While all approaches measure primarily the PG information per pencil beam spot, there are quite different approaches of evaluating this raw data to conclude on clinically relevant treatment deviations. Usually, first a spot-wise range shift is determined by comparing the measured PG signal with an expected PG signal (213, 214). However, in real-world application that spot-wise information has limited accuracy, especially for low weighted spots, but more importantly a conclusion on clinically relevant treatment deviation is difficult from range information at the spot-level. One proposal is to reconstruct the 3D dose distribution from PG information, while another approach is to condense the PG data into a classification that identifies relevant treatment changes (either binary, such as yes/no, or multi-class, distinguishing the source of error). The classification method has been successfully tested and validated on clinically acquired PGI data for prostate treatments (215) achieving very promising results. Furthermore, the classification approach has been explored on different complexity levels for head and neck treatments (216, 217). Yet another approach, called spot boosting, proposes to manipulate the clinical optimized treatment plan by boosting the dose for a limited number of pencil beam spots and to irradiate those spots first, to selectively use them for PG range verification at those selected spot locations (218).
Despite its proven clinical applicability, no medical products for PGTV are currently available. A significant technological challenge is the integration of the PGTV system near the isocenter (e.g., direct attachment to the nozzle or gantry). However, with several clinical studies (219) underway, focusing not just on technical proof-of-concept but also on clinical impact and treatment intervention – particularly in the context of OAPT – PGTV is well-positioned to play a key role in future OAPT-ready treatment systems. For example, an upcoming clinical interventional trial (DEPICT) planned at OncoRay, Dresden, aims to evaluate the capability of PGI to trigger an online intervention. In this case, the acquisition of a CT scan will be triggered if PGI detects a clinically relevant treatment deviation, based on a model (215) trained on paired PGI and ground-truth dose distribution data from a retrospective patient cohort (Figure 3). While this study may not involve online adaptation, it represents an important first step toward establishing a closed feedback loop reliant on treatment verification (in this case, PGI). This progress is also driving the investigation of novel PGTV techniques (220).
Figure 3

Schematic representation of the envisioned workflow for the DEPICT clinical trial. Treatment delivery is regularly monitored using PGI. If relevant treatment deviations are detected, a cCT scan acquisition is triggered to assess the need for plan adaptation immediately after the fraction dose delivery. If adaptation is required, an updated treatment plan is created offline for the next fraction. Otherwise, treatment proceeds with the next fraction as usual. (Courtesy of Jonathan Berthold).
Recently, treatment verification based on ionoacoustic imaging, which utilizes the generation of ultrasound waves from the periodic temperature increase in tissue caused by the pulsed proton beam, has been studied in simplified, non-clinical experimental settings (221–228). The localized energy deposition from a proton pulse around the Bragg peak causes a local increase in temperature and pressure. This generates an acoustic wave that propagates isotropically and that can be detected when it reaches the patient’s skin. By analyzing the time of flight of the ionoacoustic signal, the position of the Bragg peak can be determined. However, treatment verification using ultrasound detection is still in its early stages and faces several challenges, including the propagation of ultrasound through heterogeneous tissues with varying sound speeds, reflection and attenuation, transducer characteristics, and background noise. These factors significantly degrade the signal-to-noise ratio under realistic conditions, raising questions about its clinical applicability.
In addition to using dedicated systems during the delivery of treatment to monitor its compliance, log-file-based dose calculation offers the opportunity to check the actual delivered dose after the delivery is completed (229, 230), or even mid-treatment in recently reported investigations (231). Since log files are mostly not employed at the time of treatment and usual applications aim at reliable post-delivery quality controls and dose accumulation, they will be discussed in the next section dedicated to tasks “beyond” the closed OAPT feedback loop.
8 Beyond the loop
The advent of NAPT offers the potential to finally close the long-awaited feedback loop during the treatment session (
First, the technological realization of a NAPT solution with a closed feedback loop requires changes in soft- and hardware. While introducing new medical products is a standard procedure within the medical technology industry, this process is not typically applied to in-house developments in translational research. In these cases, institutions face significant challenges to be compliant with the Medical Device Regulation (MDR) in Europe or the regulations by the Food and Drug Administration (FDA) in the United States. Especially the MDR requirements not only demand substantial resources, but they can also delay the clinical application of new technologies. At this stage, it is highly advisable for pioneering institutions to seek structured guidance from MDR experts as early as possible to streamline the development process and ensure compliance. Additionally, adopting a pragmatic approach is essential in determining the depth of legal requirements to be met (e.g., risk analysis, documentation). Medical physicists are well-positioned to make these pragmatic decisions, as they have long been responsible for risk assessment and patient safety, well before the introduction of the MDR. For in-house developments, the goal should be to integrate this expertise and methodology into the MDR process as seamlessly as possible, while minimizing the burden of additional bureaucracy. In this context, it is important to recognize the priorities of the European Society for Radiotherapy and Oncology (ESTRO) within the framework of the European Alliance for Medical Radiation Protection Research (EURAMED) (232), which emphasize interdisciplinary collaboration to improve clinical outcomes and minimize side effects. ESTRO advocates for the clinical implementation of new radiotherapy technologies, as a means of advancing precision medicine. This focus aligns with ongoing research and regulatory efforts aimed at supporting clinical trials and ensuring that emerging technologies like OAPT are thoroughly evaluated for their clinical efficacy, time efficiency, and cost-effectiveness.
Secondly, as anticipated, the introduction of log-file-based dose recalculation allows to perform additional checks on the treatment compliance. They usually lie outside of the “online” loop, since they are mostly performed only after the clinical delivery is concluded. This “log-file-based QA” relies on treatment delivery log files from machine’s internal beam monitors to reconstruct the delivered dose under actual treatment conditions (gantry angle and patient-specific anatomy). This approach eliminates the necessity for phantom measurements at a fixed gantry angle of 0° and as such has already found its successful way in routine clinical application with available medical products (233). It has been investigated in numerous workflow scenarios in PT, including fully automated PSQA solutions (
Lastly, an efficient interconnection between the involved systems in the workflow is crucial for the implementation of the proposed near real-time adaptive closed feedback loop. Research to date has focused on the single tasks (
It is crucial to emphasize that the implementation of efficient data streams and the timely performance of tasks will be essential to achieve the overall objective of NAPT. For example, the possibility of performing parallel tasks is essential, such as initiating the assessment of adaptation needs while the contouring process is still ongoing, commencing the online plan adaptation process before the final contours have been approved, or sending the online adapted plan to the delivery machine before the final results of the online QA process have been obtained. Together with this, a thorough documentation of the online adaptive session is demanded which goes beyond the current standards for reporting and documenting PT treatments (244), with new guidelines currently under development. This documentation might take the form of extended plan reports, dedicated to specific tasks. Image-related documentation could include imaging protocols, the employed image registration (if images are not taken at the isocenter), contouring protocols documenting how different regions of interest were obtained and if manual adjustments were necessary. The reason for the online plan adaptation should also be reported, for example, by noting the DVH criteria that have triggered an automated adaptation due to exceeding predefined thresholds or other clinical motivations. The initial plan used for dose recalculation and assessment, details of the online plan adaptation (e.g., robustness settings, optimization settings) and the results of the DVH-metrics comparison could all be part of the online adapted plan documentation. Moreover, reports of the online QA containing the results of the performed sanity checks should be automatically provided. Finally, the timings for each step in the workflow, which would provide a basis for performance evaluations, should also be recorded. Whenever applicable, documentation should be appropriately stored in the OIS.
It is also important to recognize that while triggered NAPT has the potential to streamline certain aspects of the workflow, it may simultaneously introduce new complexities both inside and outside the treatment room. These include increased demands on staff for monitoring, decision-making, and QA, as well as the need for additional training to effectively operate new clinical tools and automation systems. All professional groups involved in the workflow would be affected, similar to current practices in online adaptive XRT: physicians would face additional workload for reviewing online-generated contours; medical physicists would take on responsibilities for online QA verification and offline review of the adapted plan; and radiotherapy therapists (RTTs) would need to manage a modified in-room workflow, potentially including increased responsibilities like in RTT-only workflows (245, 246). In this context, it is essential to establish clear intervention protocols that define decision rules, intervention thresholds, and corresponding actions. Such structured approaches enable efficient RTT-led workflows, ensuring that physicians and medical physicists are only required to intervene when unexpected or out-of-tolerance situations arise. Despite these initial challenges and the associated learning curve, it is expected that with growing clinical experience and ongoing technological advancements, the workflow will become more streamlined, thereby reducing the additional training and workload over time. In this regard, valuable insights can be drawn from the experience gained in implementing online adaptive XRT (247).
9 Summary and outlook
OAPT is on the path to becoming a clinical reality, but it still remains greatly limited at present. The driving force behind this progress lies in unmet clinical needs, such as the growing demand for more precise and timely treatments and the inclusion of previously undertreated entities. These needs are prompting ongoing research and technological advancements aimed at enabling faster and more efficient adaptations. This review provided a comprehensive overview of the key components and tasks essential for the successful implementation of an OAPT workflow, ranging from daily adaptive to near real-time, with main focus on the latter. First in-house clinical implementations of daily adaptive workflows have already been successfully carried out (
Figure 4

Timing reports for the first clinical applications of DAPT (A) and previous experimental tests of the same workflow (B), reproduced from (
At the same time, the requisite effort demands for the definition of clear guidelines and clinical approaches to be carefully redesigned. For a widespread use, several aspects of OAPT must be carefully assessed. This involves a multidimensional evaluation, considering factors such as clinical benefits, time efficiency, cost-effectiveness, and more.
The lack of evidence and the scarcity of clinical outcome data on PT represent in general a significant challenge to the establishment of solid indications for this modality, that needs to be overcome. The issue is thoroughly discussed in the PT community (248–251) and recently results of several completed clinical studies (including randomized clinical trials) have emerged with positive results showing improved clinical outcomes, reduction in toxicity, mitigation of acute and late effects, and consequential reduction in total cost burden (252–254). In this context, it is important that OAPT be introduced within clinical studies from the outset to assess its benefits across all dimensions. This includes evaluating its ability to enhance clinical outcomes, its competitiveness with XRT, its potential to enable hypofractionation, and its role in treating new tumor entities.
Economic factors are another critical consideration in the development of efficient OAPT and represent a significant limitation within the broader context of PT, especially when compared to XRT (250). While evaluations of opportunity costs for online adaptive treatments have already been presented for XRT (255), similar analyses are still lacking for PT. However, as the availability of PT centers continues to increase (252) and the number of patients grows, more clinical data will emerge, creating valuable opportunities for the development and effective use of OAPT.
The clinical benefit and cost must be related with other aspects to get a full evaluation, like the potentially prolonged time of treatment and overall available time in the treatment room. However, OAPT could also enable hypofractionation thereby leading to an overall reduction of treatment time per patient. Novel studies exploring possible scenarios for the introduction of OAPT in PT centers with focus on clinical outcome and time expenditures have been recently published (256, 257) and further similar investigations are expected. Borderías et al. (256) specifically highlighted that when online adaptation is time-consuming and results in some patients being treated with XRT instead, there is a trade-off between clinical benefit (expressed as reduced normal tissue complication probability, NTCP, for the whole cohort) and the time available for adaptation (Figure 5). While the study results are highly illustrative, one should keep in mind that the general transferability to other institutions might be limited due to the made assumptions on facility utilization. According to the study, the benefit of online adaptation is contingent on reducing the setup margin, and a maximum allowable adaptation time can be defined, beyond which the modelled clinical advantage is no longer maintained. Furthermore, the availability of online adaptive strategies for PT with reduced setup margins would favor the employment of hypofractionation or dose escalation, which are currently limited by the treatment-induced toxicities. This has already been observed in XRT (258–261). Likewise, NTCP-based adaptation would also be facilitated (130), as well as emerging approaches like simulation-free workflows, which are gaining significant interest in XRT (262–264). In this sense NAPT bringing forward innovative adaptive strategies would thus be an invaluable tool for the introduction and exploration of novel therapies, enabling the treatment of previously inaccessible tumor entities.
Figure 5

Variation in NTCP for 2-year mortality with three online adaptive proton therapy (OAPT) strategies using different setup margins for the lung cancer patient cohort investigated in (256) with institution-specific assumptions on facility occupancy and utilization. The dashed red line represents the NTCP gain for transitioning from conventional RT to non-adapted intensity-modulated PT. Points above this line indicate situations where OAPT strategies provide an NTCP gain across all patients. Seven different scenarios (Sideal, …,S6) are investigated, ranging from instantaneous OAPT for all patients (Sideal) to more realistic scenarios where an increasing patient proportion is treated with RT. For 2-year mortality, a maximum of 6 minutes of additional time per fraction per patient is allowed to maintain an overall NTCP gain without setup error reduction (blue line). With setup error reduction, 13.7 and 19 minutes of extra time per fraction can still provide an NTCP gain (orange and green lines). For details refer to (256). Reproduced from (256) with permission from Elsevier.
In order to ensure a careful and nuanced balance between the benefits of OAPT treatments and the present limitations, it is important to consider each patient indication carefully. Factors such as the potential for hypofractionation or dose escalation enabled by OAPT, treatment time, margin reduction, clinical outcomes, available clinical resources, and the comparison between DAPT, NAPT, and offline adaptive workflows must all be weighed. It is crucial to recognize that there is no one-size-fits-all approach, and each patient’s specific circumstances should guide the choice of the most appropriate adaptive treatment strategy.
Thus, in light of the growing range of available workflows and adaptive strategies in PT and the interest in the topic of OAPT from both vendors and clinical users, further clinical implementations towards faster time regimes are soon to be expected, leading to a multifaceted technological landscape and approach, with decisions dependent on the specifics of each patient case and available workflow.
Statements
Author contributions
VG: Methodology, Conceptualization, Writing – original draft, Investigation, Visualization, Writing – review & editing. KS: Visualization, Methodology, Writing – review & editing, Conceptualization, Supervision. CR: Resources, Methodology, Conceptualization, Funding acquisition, Supervision, Writing – review & editing.
Funding
The author(s) declare that no financial support was received for the research and/or publication of this article.
Acknowledgments
The authors would like to thank Dr. Lena Nenoff and Lukas Wolter for the fruitful discussions and feedback about specific aspects of the reviewed topic, Dr. Jonathan Berthold for discussing the upcoming clinical trial DEPICT and related figures and Dr. Fabian Hennings for discussing online adaptive workflows related time expenditures.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declare that no Generative AI was used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
KnopfA-CLomaxA. In vivo proton range verification: a review. Phys Med Biol. (2013) 58:R131–60. doi: 10.1088/0031-9155/58/15/r131
2
MohanR. A review of proton therapy – Current status and future directions. Precis Radiat Oncol. (2022) 6:164–76. doi: 10.1002/pro6.1149
3
YanDViciniFWongJMartinezA. Adaptive radiation therapy. Phys Med Biol. (1997) 42:123–32. doi: 10.1088/0031-9155/42/1/008
4
Glide-HurstCKLeePYockADOlsenJRCaoMSiddiquiFet al. Adaptive radiation therapy (ART) strategies and technical considerations: A state of the ART review from NRG oncology. Int J Radiat Oncol Biol Phys. (2021) 109:1054–75. doi: 10.1016/j.ijrobp.2020.10.021
5
NgJGregucciFPennellRTNagarHGoldenEBKniselyJPSet al. MRI-LINAC: A transformative technology in radiation oncology. Front Oncol. (2023) 13:1117874. doi: 10.3389/fonc.2023.1117874
6
TuK-YHuangY-SLauJLeeH-H. Adaptive Tomotherapy for locally advanced unresectable pancreatic neuroendocrine tumor: Case report and literature review. Front Oncol. (2022) 12:1045752. doi: 10.3389/fonc.2022.1045752
7
LiuHSchaalDCurryHClarkRMagliariAKupelianPet al. Review of cone beam computed tomography based online adaptive radiotherapy: current trend and future direction. Radiat Oncol. (2023) 18. doi: 10.1186/s13014-023-02340-2
8
OderindeOMShirvaniSMOlcottPDKuduvalliGMazinSLarkinD. The technical design and concept of a PET/CT linac for biology-guided radiotherapy. Clin Transl Radiat Oncol. (2021) 29:106–12. doi: 10.1016/j.ctro.2021.04.003
9
ThorwarthD. Biologically adapted radiation therapy. Z Med Phys. (2018) 28:177–83. doi: 10.1016/j.zemedi.2017.08.001
10
AlbertiniFMatterMNenoffLZhangYLomaxA. Online daily adaptive proton therapy. Br J Radiol. (2019) 93: 20190594. doi: 10.1259/bjr.20190594
11
PaganettiHBotasPSharpGCWineyB. Adaptive proton therapy. Phys Med Biol. (2021) 66:22TR01. doi: 10.1088/1361-6560/ac344f
12
NenoffLMatterMCharmillotMKrierSUherKWeberDCet al. Experimental validation of daily adaptive proton therapy. Phys Med Biol. (2021) 66:205010. doi: 10.1088/1361-6560/ac2b84
13
StockMGeorgDAbleitingerAZechnerAUtzAMumotMet al. The technological basis for adaptive ion beam therapy at MedAustron: Status and outlook. Z Med Phys. (2018) 28:196–210. doi: 10.1016/j.zemedi.2017.09.007
14
Guterres MarmittGPinANg Wei SiangKJanssensGSourisKCohilisMet al. Platform for automatic patient quality assurance via Monte Carlo simulations in proton therapy. Phys Med. (2020) 70:49–57. doi: 10.1016/j.ejmp.2019.12.018
15
GreenOLHenkeLEHugoGD. Practical clinical workflows for online and offline adaptive radiation therapy. Semin Radiat Oncol. (2019) 29:219–27. doi: 10.1016/j.semradonc.2019.02.004
16
PakelaJMKnopfADongLRucinskiAZouW. Management of motion and anatomical variations in charged particle therapy: past, present, and into the future. Front Oncol. (2022) 12:806153. doi: 10.3389/fonc.2022.806153
17
AlbertiniFCzerskaKVazquezMAndacaIBachtiaryBBessonRet al. First clinical implementation of a highly efficient daily online adapted proton therapy (DAPT) workflow. Phys Med Biol. (2024) 69:215030. doi: 10.1088/1361-6560/ad7cbd
18
National Library of Medicine (U.S.). Online proton adaptive radiotherapy (PARTy) (2024). Available online at: https://clinicaltrials.gov/ct2/show/NCT06310655 (Accessed December 5, 2025). Identifier NCT06310655.
19
MeijersASeller OriaCFreeJLangendijkJAKnopfACBothS. Technical Note: First report on an in vivo range probing quality control procedure for scanned proton beam therapy in head and neck cancer patients. Med Phys. (2021) 48:1372–80. doi: 10.1002/mp.14713
20
LangenKMJonesDTL. Organ motion and its management. Int J Radiat Oncol Biol Phys. (2001) 50:265–78. doi: 10.1016/s0360-3016(01)01453-5
21
DíezSGarcíaJSendraF. Analysis and evaluation of periodic physiological organ motion in radiotherapy treatments. Radiother Oncol. (2004) 73:325–9. doi: 10.1016/j.radonc.2004.08.023
22
MostafaeiFTaiAOmariESongYChristianJPaulsonEet al. Variations of MRI-assessed peristaltic motions during radiation therapy. PloS One. (2018) 13:e0205917. doi: 10.1371/journal.pone.0205917
23
von SiebenthalMSzékelyGLomaxAJCattinPC. Systematic errors in respiratory gating due to intrafraction deformations of the liver. Med Phys. (2007) 34:3620–9. doi: 10.1118/1.2767053
24
DaamenLAde Mol van OtterlooSRvan GoorIWJMEijkelenkampHEricksonBAHallWAet al. Online adaptive MR-guided stereotactic radiotherapy for unresectable Malignancies in the upper abdomen using a 1. 5T MR-linac. Acta Oncol. (2021) 61:111–5. doi: 10.1080/0284186x.2021.2012593
25
LeeSLBassettiMMeijerGJMookS. Review of MR-guided radiotherapy for esophageal cancer. Front Oncol. (2021) 11:628009. doi: 10.3389/fonc.2021.628009
26
PortelanceLCorradiniSEricksonBLalondrelleSPadgettKvan der LeijFet al. Online magnetic resonance-guided radiotherapy (oMRgRT) for gynecological cancers. Front Oncol. (2021) 11:628131. doi: 10.3389/fonc.2021.628131
27
BertholetJKnopfAEibenBMcClellandJGrimwoodAHarrisEet al. Real-time intrafraction motion monitoring in external beam radiotherapy. Phys Med Biol. (2019) 64:15TR01. doi: 10.1088/1361-6560/ab2ba8
28
RiboldiMOrecchiaRBaroniG. Real-time tumour tracking in particle therapy: technological developments and future perspectives. Lancet Oncol. (2012) 13:e383–91. doi: 10.1016/s1470-2045(12)70243-7
29
LandryGHuaC. Current state and future applications of radiological image guidance for particle therapy. Med Phys. (2018) 45: e1086-95. doi: 10.1002/mp.12744
30
HoffmannAObornBMoteabbedMYanSBortfeldTKnopfAet al. MR-guided proton therapy: a review and a preview. Radiat Oncol. (2020) 15. doi: 10.1186/s13014-020-01571-x
31
ObornBMDowdellSMetcalfePECrozierSMohanRKeallPJ. Future of medical physics: Real-time MRI-guided proton therapy. Med Phys. (2017) 44: e77-90. doi: 10.1002/mp.12371
32
BauerJBeyerCDorschSEllerbrockMHansmannTKlüterSet al. TOWARDS DAILY ONLINE MR-IMAGE-GUIDED PARTICLE THERAPY: FIRST CLINICAL APPLICATION OF A SHUTTLE-BASED WORKFLOW AT THE HEIDELBERG ION-BEAM THERAPY CENTER. Int J Part Ther. (2024) 12:100243. doi: 10.1016/j.ijpt.2024.100243
33
PestanaRSeidensaalKBeyerCHerfarthKDebusJKlüterSet al. TOWARDS MRI-GUIDED ONLINE ADAPTIVE PARTICLE THERAPY IN THE PELVIC REGION. Int J Part Ther. (2024) 12:100242. doi: 10.1016/j.ijpt.2024.100242
34
JohnsonRPBashkirovVDeWittLGiacomettiVHurleyRFPiersimoniPet al. A fast experimental scanner for proton CT: technical performance and first experience with phantom scans. IEEE Trans Nucl Sci. (2016) 63:52–60. doi: 10.1109/tns.2015.2491918
35
JohnsonRP. Meeting the detector challenges for pre-clinical proton and ion computed tomography. Phys Med Biol. (2024) 69:11TR02. doi: 10.1088/1361-6560/ad42fc
36
RossiMBelottiGPaganelliCPellaABarcelliniACerveriPet al. Image-based shading correction for narrow-FOV truncated pelvic CBCT with deep convolutional neural networks and transfer learning. Med Phys. (2021) 48:7112–26. doi: 10.1002/mp.15282
37
OliverJAZeidanOAMeeksSLShahAPPukalaJKellyPet al. The Mobius AIRO mobile CT for image-guided proton therapy: Characterization & commissioning. J Appl Clin Med Phys. (2017) 18:130–6. doi: 10.1002/acm2.12084
38
SunBYangDLamDZhangTDvergstenTBradleyJet al. Toward adaptive proton therapy guided with a mobile helical CT scanner. Radiother Oncol. (2018) 129:479–85. doi: 10.1016/j.radonc.2018.08.021
39
PetersNWolfahrtPRichterC. X-ray computed tomography for treatment planning: current status and innovations. In: PaganelliCGianoliCKnopfA, editors. Imaging in particle therapy: current practice and future trends. Bristol, UK: IOP Publishing (2024). p. 4–(1-18). doi: 10.1088/978-0-7503-5117-1
40
RichterCWolfahrtPRichterC. Dual-energy CT in radiation oncology. In: AlkadhiHEulerAMaintzDSahaniD, editors. Spectral imaging: dual-energy, multi-energy and photon-counting CT. Cham, Switzerland: Springer Cham (2022). p. 333–46. doi: 10.1007/978-3-030-96285-2
41
NesterukKPBobićMLalondeAWineyBALomaxAJPaganettiH. CT-on-rails versus in-room CBCT for online daily adaptive proton therapy of head-and-neck cancers. Cancers. (2021) 13:5991. doi: 10.3390/cancers13235991
42
ArchambaultYBoylanCBullockDMorgasTPeltolaJRuokokoskiEet al. Making on-line adaptive radiotherapy possible using artificial intelligence and machine learning for efficient daily re-planning. Med Phys Int. (2020) 8:77–86. Available online at http://mpijournal.org/pdf/2020-02/MPI-2020-02-p077.pdf (Accessed November 25, 2025)
43
StanleyDNHarmsJPogueJABelliveauJMarcromSRMcDonaldAMet al. A roadmap for implementation of kV-CBCT online adaptive radiation therapy and initial first year experiences. J Appl Clin Med Phys. (2023) 24. doi: 10.1002/acm2.13961
44
Groot KoerkampMLBolGHKroonPSKrikkeLLHarderwijkTZoeteliefAJet al. Bringing online adaptive radiotherapy to a standard C-arm linac. Phys Imaging Radiat Oncol. (2024) 31:100597. doi: 10.1016/j.phro.2024.100597
45
ElektaAB. Elekta launches AI-powered adaptive CT-Linac for next-level cancer care (2024). Available online at: https://ir.elekta.com/investors/press-releases/2024/elekta-launches-ai-powered-adaptive-ct-linac-for-next-level-cancer-care/ (Accessed Apr 10, 2025).
46
O’BrienRTDillonOLauBGeorgeASmithSWallisAet al. The first-in-human implementation of adaptive 4D cone beam CT for lung cancer radiotherapy: 4DCBCT in less time with less dose. Radiother Oncol. (2021) 161:29–34. doi: 10.1016/j.radonc.2021.05.021
47
MacKayRI. Image guidance for proton therapy. Clin Oncol. (2018) 30:293–8. doi: 10.1016/j.clon.2018.02.004
48
SchulzeRHeilUGroβDBruellmannDDranischnikowESchwaneckeUet al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. (2011) 40:265–73. doi: 10.1259/dmfr/30642039
49
HerrickMPenfoldSSantosAHicksonK. A systematic review of volumetric image guidance in proton therapy. Phys Eng Sci Med. (2023) 46:963–75. doi: 10.1007/s13246-023-01294-9
50
GiacomettiVHounsellARMcGarryCK. A review of dose calculation approaches with cone beam CT in photon and proton therapy. Phys Med. (2020) 76:243–76. doi: 10.1016/j.ejmp.2020.06.017
51
ReinersKDaganRHoltzmanABryantCAnderssonSNilssonRet al. CBCT-based dose monitoring and adaptive planning triggers in head and neck PBS proton therapy. Cancers. (2023) 15:3881. doi: 10.3390/cancers15153881
52
PengHWuCNguyenDSchuemannJMairaniAPuYet al. Recent advancements of artificial intelligence in particle therapy. IEEE Trans Radiat Plasma Med Sci. (2023) 7:213–24. doi: 10.1109/trpms.2023.3241102
53
Upright Radiotherapy Research Consortium. New research and training project launched — Upright Radiotherapy Research Consortium (2024). Available online at: https://www.uprightresearchconsortium.com/news/new-research-and-training-project-launched (Accessed December 9, 2025).
54
MunzenriderJEAustin-SeymourMBlitzerPJGentryRGoiteinMGragoudasESet al. Proton therapy at Harvard. Strahlentherapie. (1985) 161:756–63. Available online at https://pubmed.ncbi.nlm.nih.gov/3001976/ (Accessed November 25, 2025)
55
BoisbouvierSBoucaudATanguyRGrégoireV. Upright patient positioning for pelvic radiotherapy treatments. Tech Innov Patient Support Radiat Oncol. (2022) 24:124–30. doi: 10.1016/j.tipsro.2022.11.003
56
PryanichnikovAASokunovVVShemyakovAE. Some results of the clinical use of the proton therapy complex “Prometheus. Phys Part Nuclei Lett. (2018) 15:981–5. doi: 10.1134/s1547477118070592
57
GordonKGulidovISmykDSemenovAGolubevKLemaevaAet al. Upright proton therapy for esthesioneuroblastoma: a single-institution experience. Front Oncol. (2024) 14:1348291. doi: 10.3389/fonc.2024.1348291
58
LemaevaAGulidovISmykDAgapovaYKoryakinSEreminaIet al. A single-center experience of the upright proton therapy for skull-base chordomas and chondrosarcomas: Updated results. Clin Transl Radiat Oncol. (2024) 48:100814. doi: 10.1016/j.ctro.2024.100814
59
BoisbouvierSUnderwoodTMcNamaraJProbstH. Upright patient positioning for gantry-free breast radiotherapy: feasibility tests using a robotic chair and specialised bras. Front Oncol. (2023) 13:1250678. doi: 10.3389/fonc.2023.1250678
60
FeldmanJPryanichnikovAShwartzDHillmanYWygodaMBlumenfeldPet al. Study of upright patient positioning reproducibility in image-guided proton therapy for head and neck cancers. Radiother Oncol. (2024) 201:110572. doi: 10.1016/j.radonc.2024.110572
61
VolzLShengYDuranteMGraeffC. Considerations for upright particle therapy patient positioning and associated image guidance. Front Oncol. (2022) 12:930850. doi: 10.3389/fonc.2022.930850
62
VolzLKorteJMartireMCZhangYHardcastleNDuranteMet al. Opportunities and challenges of upright patient positioning in radiotherapy. Phys Med Biol. (2024) 69:18TR02. doi: 10.1088/1361-6560/ad70ee
63
MoteabbedMSmeetsJHongTSJanssensGLabarbeRWolfgangJAet al. Toward MR-integrated proton therapy: modeling the potential benefits for liver tumors. Phys Med Biol. (2021) 66:195004. doi: 10.1088/1361-6560/ac1ef2
64
YawsonAKPaulKMBeyerCDorschSKlüterSWelzelTet al. Pseudo-SPR map generation from MRI using U-net architecture for ion beam therapy application. Lect Notes Comput Sci. (2023) 14122: 257–67. doi: 10.1007/978-3-031-48593-0_19
65
GantzSHietscholdVHoffmannAL. Characterization of magnetic interference and image artefacts during simultaneous in-beam MR imaging and proton pencil beam scanning. Phys Med Biol. (2020) 65:215014. doi: 10.1088/1361-6560/abb16f
66
FuchsHPadilla-CabalFZimmermannLPalmansHGeorgD. MR-guided proton therapy: Impact of magnetic fields on the detector response. Med Phys. (2021) 48:2572–9. doi: 10.1002/mp.14660
67
FuchsHPalmansHHeilemannGZuschlagDGeorgDKuessP. Dosimetry in MRgPT: Impact of magnetic fields on TLD dose response during proton irradiation. Med Phys. (2024) 52:633–9. doi: 10.1002/mp.17454
68
CobanajMLebbinkFHorstFTraneusEHoffmannA. VERIFICATION OF PROTON DOSE CALCULATION IN THE MAGNETIC FIELD OF AN IN-BEAM MR SCANNER. Int J Part Ther. (2024) 12:100157. doi: 10.1016/j.ijpt.2024.100157
69
ObornBMSemioshkinaEvan der KraaijEHoffmannAL. Simulation and experimental benchmarking of a proton pencil beam scanning nozzle model for development of MR-integrated proton therapy. Med Phys. (2024) 51:6196–205. doi: 10.1002/mp.17279
70
SchiefereckeJGantzSHoffmannAPawelkeJ. Investigation of contrast mechanisms for MRI phase signal-based proton beam visualization in water phantoms. Magn Reson Med. (2023) 90:1776–88. doi: 10.1002/mrm.29752
71
GebauerBPawelkeJHoffmannALührA. Technical note: Experimental dosimetric characterization of proton pencil beam distortion in a perpendicular magnetic field of an in-beam MR scanner. Med Phys. (2023) 50:7294–303. doi: 10.1002/mp.16448
72
GantzSKarschLPawelkeJSchiefereckeJSchellhammerSSmeetsJet al. Direct visualization of proton beam irradiation effects in liquids by MRI. Proc Natl Acad Sci USA. (2023) 120(23):e2301160120. doi: 10.1073/pnas.2301160120
73
MarotMSurlaSBurkeEBronsSRunzAGreilichSet al. Proton beam dosimetry in the presence of magnetic fields using Farmer-type ionization chambers of different radii. Med Phys. (2023) 50:4590–9. doi: 10.1002/mp.16368
74
DietrichKAKlüterSDinkelFEchnerGBronsSOrzadaSet al. An essentially radiation-transparent body coil integrated with a patient rotation system for MR-guided particle therapy. Med Phys. (2024) 51:4028–43. doi: 10.1002/mp.17065
75
DorschSPaulKBeyerCKargerCPJäkelODebusJet al. Quality assurance and temporal stability of a 1.5 T MRI scanner for MR-guided Photon and Particle Therapy. Z Med Phys. (2023). 35:204–217. doi: 10.1016/j.zemedi.2023.04.004
76
FuchsHPadilla-CabalFObornBMGeorgD. Commissioning a beam line for MR-guided particle therapy assisted by in silico methods. Med Phys. (2022) 50:1019–28. doi: 10.1002/mp.16143
77
SchulteRBashkirovVLiTLiangZMuellerKHeimannJet al. Conceptual design of a proton computed tomography system for applications in proton radiation therapy. IEEE Trans Nucl Sci. (2004) 51:866–72. doi: 10.1109/tns.2004.829392
78
PenfoldSNRosenfeldABSchulteRWSchubertKE. A more accurate reconstruction system matrix for quantitative proton computed tomography. Med Phys. (2009) 36:4511–8. doi: 10.1118/1.3218759
79
PoludniowskiGAllinsonNMEvansPM. Proton radiography and tomography with application to proton therapy. Br J Radiol. (2015) 88:20150134. doi: 10.1259/bjr.20150134
80
PrallMDuranteMBergerTPrzybylaBGraeffCLangPMet al. High-energy proton imaging for biomedical applications. Sci Rep. (2016) 6: 27651. doi: 10.1038/srep27651
81
JohnsonRP. Review of medical radiography and tomography with proton beams. Rep Prog Phys. (2017) 81:016701. doi: 10.1088/1361-6633/aa8b1d
82
PryanichnikovAAShwartzDFeldmandJSecoJSpadeaMF. Dual-energy extraction for proton therapy and imaging: validation on a clinical synchrotron-based facility. arXiv [Preprint]. (2025). doi: 10.48550/ARXIV.2506.19736
83
FeldmanJPryanichnikovAAchkienasiAPolyanskyIHillmanYRaskinSet al. Commissioning of a novel gantry-less proton therapy system. Front Oncol. (2024) 14:1417393. doi: 10.3389/fonc.2024.1417393
84
DedesGDickmannJNiepelKWespPJohnsonRPPankuchMet al. Experimental comparison of proton CT and dual energy x-ray CT for relative stopping power estimation in proton therapy. Phys Med Biol. (2019) 64:165002. doi: 10.1088/1361-6560/ab2b72
85
LaneSASlaterJMYangGY. Image-guided proton therapy: A comprehensive review. Cancers. (2023) 15:2555. doi: 10.3390/cancers15092555
86
SharpGFritscherKDPekarVPeroniMShusharinaNVeeraraghavanHet al. Vision 20/20: Perspectives on automated image segmentation for radiotherapy. Med Phys. (2014) 41:050902. doi: 10.1118/1.4871620
87
HizamDATanLKSaadMMuaadzAUngNM. Comparison of commercial atlas-based automatic segmentation software for prostate radiotherapy treatment planning. Phys Eng Sci Med. (2024) 47:881–94. doi: 10.1007/s13246-024-01411-2
88
GibbonsEHoffmannMWesthuyzenJHodgsonAChickBLastA. Clinical evaluation of deep learning and atlas-based auto-segmentation for critical organs at risk in radiation therapy. J Med Radiat Sci. (2022) 70:15–25. doi: 10.1002/jmrs.618
89
RadiciLFerrarioSBorcaVCCanteDPaoliniMPivaCet al. Implementation of a commercial deep learning-based auto segmentation software in radiotherapy: evaluation of effectiveness and impact on workflow. Life. (2022) 12:2088. doi: 10.3390/life12122088
90
La MacchiaMFellinFAmichettiMCianchettiMGianoliniSPaolaVet al. Systematic evaluation of three different commercial software solutions for automatic segmentation for adaptive therapy in head-and-neck, prostate and pleural cancer. Radiat Oncol. (2012) 7. doi: 10.1186/1748-717x-7-160
91
VaassenFBoukerrouiDLooneyPCantersRVerhoevenKPeetersSet al. Real-world analysis of manual editing of deep learning contouring in the thorax region. Phys Imaging Radiat Oncol. (2022) 22:104–10. doi: 10.1016/j.phro.2022.04.008
92
BrouwerCLBoukerrouiDOliveiraJLooneyPSteenbakkersRJHMLangendijkJAet al. Assessment of manual adjustment performed in clinical practice following deep learning contouring for head and neck organs at risk in radiotherapy. Phys Imaging Radiat Oncol. (2020) 16:54–60. doi: 10.1016/j.phro.2020.10.001
93
RicciJCRineerJShahAPMeeksSLKellyP. Proposal and Evaluation of a Physician-Free, Real-Time On-Table Adaptive Radiotherapy (PF-ROAR) Workflow for the MRIdian MR-Guided LINAC. J Clin Med. (2022) 11:1189. doi: 10.3390/jcm11051189
94
HeilemannGBuschmannMLechnerWDickVEckertFHeilmannMet al. Clinical implementation and evaluation of auto-segmentation tools for multi-site contouring in radiotherapy. Phys Imaging Radiat Oncol. (2023) 28:100515. doi: 10.1016/j.phro.2023.100515
95
MackayKBernsteinDGlockerBKamnitsasKTaylorA. A review of the metrics used to assess auto-contouring systems in radiotherapy. Clin Oncol. (2023) 35:354–69. doi: 10.1016/j.clon.2023.01.016
96
BobićMChoulilitsaELeeHCzerskaKChristensenJBMayorAet al. Multi-institutional experimental validation of online adaptive proton therapy workflows. Phys Med Biol. (2024) 69:165021. doi: 10.1088/1361-6560/ad6527
97
QiaoYJagtTHoogemanMLelieveldtBPFStaringM. Evaluation of an open source registration package for automatic contour propagation in online adaptive intensity-modulated proton therapy of prostate cancer. Front Oncol. (2019) 9:1297. doi: 10.3389/fonc.2019.01297
98
NenoffLMatterMAmayaEJJosipovicMKnopfA-CLomaxAJet al. Dosimetric influence of deformable image registration uncertainties on propagated structures for online daily adaptive proton therapy of lung cancer patients. Radiother Oncol. (2021) 159:136–43. doi: 10.1016/j.radonc.2021.03.021
99
RiegelACAntoneJGZhangHJainPRainceJReaAet al. Deformable image registration and interobserver variation in contour propagation for radiation therapy planning. J Appl Clin Med Phys. (2016) 17:347–57. doi: 10.1120/jacmp.v17i3.6110
100
WoernerAJChoiMHarkenriderMMRoeskeJCSurucuM. Evaluation of deformable image registration-based contour propagation from planning CT to cone-beam CT. Technol Cancer Res Treat. (2017) 16:801–10. doi: 10.1177/1533034617697242
101
RigaudBSimonACastelliJLafondCAcostaOHaigronPet al. Deformable image registration for radiation therapy: principle, methods, applications and evaluation. Acta Oncol. (2019) 58:1225–37. doi: 10.1080/0284186x.2019.1620331
102
NashDPalmerALvan HerkMMcWilliamAVasquez OsorioE. Suitability of propagated contours for adaptive replanning for head and neck radiotherapy. Phys Med. (2022) 102:66–72. doi: 10.1016/j.ejmp.2022.09.002
103
PukalaJJohnsonPBShahAPLangenKMBovaFJStatonRJet al. Benchmarking of five commercial deformable image registration algorithms for head and neck patients. J Appl Clin Med Phys. (2016) 17:25–40. doi: 10.1120/jacmp.v17i3.5735
104
HvidCAElstrømUVJensenKAlberMGrauC. Accuracy of software-assisted contour propagation from planning CT to cone beam CT in head and neck radiotherapy. Acta Oncol. (2016) 55:1324–30. doi: 10.1080/0284186x.2016.1185149
105
FengMValdesGDixitNSolbergTD. Machine learning in radiation oncology: opportunities, requirements, and needs. Front Oncol. (2018) 8:110. doi: 10.3389/fonc.2018.00110
106
FinneganRNQuinnAHorsleyPChanJStewartMBromleyRet al. Geometric and dosimetric evaluation of a commercial AI auto-contouring tool on multiple anatomical sites in CT scans. J Appl Clin Med Phys. (2025). 26: e70067. doi: 10.1002/acm2.70067
107
FanMWangTLeiYPatelPRDresserSGhavidelBBet al. Evaluation and failure analysis of four commercial deep learning-based autosegmentation software for abdominal organs at risk. J Appl Clin Med Phys. (2025) 26: e70010. doi: 10.1002/acm2.70010
108
NgCKC. Performance of commercial deep learning-based auto-segmentation software for prostate cancer radiation therapy planning: A systematic review. Information. (2025) 16:215. doi: 10.3390/info16030215
109
LangmackKAAlexanderGGGardinerJMcKennaAShawcroftE. An audit of the impact of the introduction of a commercial artificial intelligence-driven auto-contouring tool into a radiotherapy department. Br J Radiol. (2024) 98:375–82. doi: 10.1093/bjr/tqae255
110
DoolanPJCharalambousSRoussakisYLeczynskiAPeratikouMBenjaminMet al. A clinical evaluation of the performance of five commercial artificial intelligence contouring systems for radiotherapy. Front Oncol. (2023) 13:1213068. doi: 10.3389/fonc.2023.1213068
111
GoddardLVeltenCTangJSkalinaKABoydRMartinWet al. Evaluation of multiple-vendor AI autocontouring solutions. Radiat Oncol. (2024) 19. doi: 10.1186/s13014-024-02451-4
112
NgCKCLeungVWSHungRHM. Clinical evaluation of deep learning and atlas-based auto-contouring for head and neck radiation therapy. Appl Sci. (2022) 12:11681. doi: 10.3390/app122211681
113
SmoldersABengtssonIForsgrenALomaxAWeberDCFredrikssonAet al. Robust optimization strategies for contour uncertainties in online adaptive radiation therapy. Phys Med Biol. (2024) 69:165001. doi: 10.1088/1361-6560/ad6526
114
NenoffLButiGBobićMLalondeANesterukKPWineyBet al. Integrating structure propagation uncertainties in the optimization of online adaptive proton therapy plans. Cancers. (2022) 14:3926. doi: 10.3390/cancers14163926
115
SmoldersAChoulilitsaECzerskaKBizzocchiNKrcekRLomaxAet al. Dosimetric comparison of autocontouring techniques for online adaptive proton therapy. Phys Med Biol. (2023) 68:175006. doi: 10.1088/1361-6560/ace307
116
BaroudiHBrockKKCaoWChenXChungCCourtLEet al. Automated contouring and planning in radiation therapy: what is ‘Clinically acceptable’? Diagnostics. (2023) 13:667. doi: 10.3390/diagnostics13040667
117
HeukelomJFullerCD. Head and neck cancer adaptive radiation therapy (ART): conceptual considerations for the informed clinician. Semin Radiat Oncol. (2019) 29:258–73. doi: 10.1016/j.semradonc.2019.02.008
118
LiuSChapmanKLBerrySLBertiniJMaRFuYet al. Implementation of a knowledge-based decision support system for treatment plan auditing through automation. Med Phys. (2023) 50:6978–89. doi: 10.1002/mp.16472
119
BenedickTLiMWangEPawooskar-AlmeidaSLeALiuB. Utilizing data-mining and deep learning methods on retrospective head and neck radiation therapy cases for decision support of individualized treatment planning. In: Medical imaging 2023: imaging informatics for healthcare, research, and applications. ellingham, Washington, USA: SPIE. (2023). 4. doi: 10.1117/12.2656180
120
RachiTArijiTTakahashiS. Development of machine-learning prediction programs for delivering adaptive radiation therapy with tumor geometry and body shape changes in head and neck volumetric modulated arc therapy. Adv Radiat Oncol. (2023) 8:101172. doi: 10.1016/j.adro.2023.101172
121
DraguetCBarragán-MonteroAMVeraMCThomasMPopulairePDefraeneGet al. Automated clinical decision support system with deep learning dose prediction and NTCP models to evaluate treatment complications in patients with esophageal cancer. Radiother Oncol. (2022) 176:101–7. doi: 10.1016/j.radonc.2022.08.031
122
GrosSAASanthanamAPBlockAMEmamiBLeeBHJoyceC. Retrospective clinical evaluation of a decision-support software for adaptive radiotherapy of head and neck cancer patients. Front Oncol. (2022) 12:7777933. doi: 10.3389/fonc.2022.7777933
123
OudMBreedveldSRojo-SantiagoJGiżyńskaMKKroesenMHabrakenSet al. A fast and robust constraint-based online re-optimization approach for automated online adaptive intensity modulated proton therapy in head and neck cancer. Phys Med Biol. (2024) 69:075007. doi: 10.1088/1361-6560/ad2a98
124
DraguetCPopulairePVeraMCFredrikssonAHaustermansKLeeJAet al. A comparative study on automatic treatment planning for online adaptive proton therapy of esophageal cancer: which combination of deformable registration and deep learning planning tools performs the best? Phys Med Biol. (2024) 69:205013. doi: 10.1088/1361-6560/ad80f6
125
JagtTBreedveldSVan De WaterSHeijmenBHoogemanM. Near real-time automated dose restoration in IMPT to compensate for daily tissue density variations in prostate cancer. Phys Med Biol. (2017) 62:4254–72. doi: 10.1088/1361-6560/aa5c12
126
BernatowiczKGeetsXBarraganAJanssensGSourisKSterpinE. Feasibility of online IMPT adaptation using fast, automatic and robust dose restoration. Phys Med Biol. (2018) 63:085018. doi: 10.1088/1361-6560/aaba8c
127
Borderías VillarroelEGeetsXSterpinE. Online adaptive dose restoration in intensity modulated proton therapy of lung cancer to account for inter-fractional density changes. Phys Imaging Radiat Oncol. (2020) 15:30–7. doi: 10.1016/j.phro.2020.06.004
128
BotasPKimJWineyBPaganettiH. Online adaption approaches for intensity modulated proton therapy for head and neck patients based on cone beam CTs and Monte Carlo simulations. Phys Med Biol. (2018) 64:015004. doi: 10.1088/1361-6560/aaf30b
129
LalondeABobićMWineyBVerburgJSharpGCPaganettiH. Anatomic changes in head and neck intensity-modulated proton therapy: Comparison between robust optimization and online adaptation. Radiother Oncol. (2021) 159:39–47. doi: 10.1016/j.radonc.2021.03.008
130
LalondeABobićMSharpGCChamseddineIWineyBPaganettiH. Evaluating the effect of setup uncertainty reduction and adaptation to geometric changes on normal tissue complication probability using online adaptive head and neck intensity modulated proton therapy. Phys Med Biol. (2023) 68:115018. doi: 10.1088/1361-6560/acd433
131
BobićMLalondeANesterukKPLeeHNenoffLGorissenBLet al. Large anatomical changes in head-and-neck cancers – A dosimetric comparison of online and offline adaptive proton therapy. Clin Transl Radiat Oncol. (2023) 40:100625. doi: 10.1016/j.ctro.2023.100625
132
Borderias-VillarroelEFredrikssonACvilicSDi PerriDLongtonEPierrardJet al. Dose mimicking based strategies for online adaptive proton therapy of head and neck cancer. Phys Med Biol. (2023) 68:105002. doi: 10.1088/1361-6560/accb38
133
MaesDHolmstromMHelanderRSainiJFangCBowenSR. Automated treatment planning for proton pencil beam scanning using deep learning dose prediction and dose-mimicking optimization. J Appl Clin Med Phys. (2023) 24:e14065. doi: 10.1002/acm2.14065
134
KaushikSStützerKÖdénJFredrikssonAToma-DasuI. Adaptive intensity modulated proton therapy using 4D robust planning: a proof-of-concept for the application of dose mimicking approach. Phys Med Biol. (2024) 69:185010. doi: 10.1088/1361-6560/ad75e0
135
MiyazakiKFujiiYYamadaTKanehiraTMiyamotoNMatsuuraTet al. Deformed dose restoration to account for tumor deformation and position changes for adaptive proton therapy. Med Phys. (2022) 50:675–87. doi: 10.1002/mp.16149
136
JagtTZBreedveldSvan HaverenRNoutRAAstreinidouEHeijmenBJMet al. Plan-library supported automated replanning for online-adaptive intensity-modulated proton therapy of cervical cancer. Acta Oncol. (2019) 58:1440–5. doi: 10.1080/0284186x.2019.1627414
137
PilskogSAbalBØvrelidKSEngesethGMYtre-HaugeKSHysingLB. Plan selection in proton therapy of locally advanced prostate cancer with simultaneous treatment of multiple targets. Int J Radiat Oncol Biol Phys. (2020) 106:630–8. doi: 10.1016/j.ijrobp.2019.11.007
138
OudMBreedveldSGiżyńskaMKroesenMHutschemaekersSHabrakenSet al. An online adaptive plan library approach for intensity modulated proton therapy for head and neck cancer. Radiother Oncol. (2022) 176:68–75. doi: 10.1016/j.radonc.2022.09.011
139
HamaideVSourisKDasnoyDGlineurFMacqB. Real-time image-guided treatment of mobile tumors in proton therapy by a library of treatment plans: a simulation study. Med Phys. (2023) 50:465–79. doi: 10.1002/mp.16084
140
TroostEGCMenkelSTschicheMThieleJJasterMHaakDet al. Towards online adaptive proton therapy: first report of plan-library-based plan-of-the-day approach. Acta Oncol. (2021) 61:231–4. doi: 10.1080/0284186x.2021.1994154
141
GambettaVFredrikssonAMenkelSRichterCStützerK. The partial adaptation strategy for online-adaptive proton therapy: A proof of concept study in head and neck cancer patients. Med Phys. (2024) 51:5572–81. doi: 10.1002/mp.17178
142
GambettaVPietaVBertholdJHölscherTFredrikssonARichterCet al. Partial adaptation for online-adaptive proton therapy triggered by during-delivery treatment verification: Feasibility study on prostate cancer treatments. Phys Imaging Radiat Oncol. (2025) 34:100755. doi: 10.1016/j.phro.2025.100755
143
ChenMZhongYShaoYJiangSLuW. Mid-range probing—towards range-guided particle therapy. Phys Med Biol. (2018) 63:13NT01. doi: 10.1088/1361-6560/aaca1b
144
ChenMYangDZhuXRMaLGrosshansDRShaoYet al. Investigation of intra-fractionated range guided adaptive proton therapy (RGAPT): II. Range-shift compensated on-line treatment adaptation and verification. Phys Med Biol. (2024) 69:155006. doi: 10.1088/1361-6560/ad56f2
145
McIntoshCWelchMMcNivenAJaffrayDAPurdieTG. Fully automated treatment planning for head and neck radiotherapy using a voxel-based dose prediction and dose mimicking method. Phys Med Biol. (2017) 62:5926–44. doi: 10.1088/1361-6560/aa71f8
146
ZhangLHolmesJMLiuZVoraSASioTTVargasCEet al. Beam mask and sliding window-facilitated deep learning-based accurate and efficient dose prediction for pencil beam scanning proton therapy. Med Phys. (2023) 51:1484–98. doi: 10.1002/mp.16758
147
VazquezILiangDSalazarRMGronbergMPSjogreenCWilliamsonTDet al. Deep learning techniques for proton dose prediction across multiple anatomical sites and variable beam configurations. Phys Med Biol. (2025) 70:075016. doi: 10.1088/1361-6560/adc236
148
TrnkovaPZhangYToshitoTHeijmenBRichterCAznarMCet al. A survey of practice patterns for adaptive particle therapy for interfractional changes. Phys Imaging Radiat Oncol. (2023) 26:100442. doi: 10.1016/j.phro.2023.100442
149
ZhangYTrnkovaPToshitoTHeijmenBRichterCAznarMet al. A survey of practice patterns for real-time intrafractional motion-management in particle therapy. Phys Imaging Radiat Oncol. (2023) 26:100439. doi: 10.1016/j.phro.2023.100439
150
SainiJTraneusEMaesDRegmiRBowenSRBlochCet al. Advanced Proton Beam Dosimetry Part I: review and performance evaluation of dose calculation algorithms. Transl Lung Cancer Res. (2018) 7:171–9. doi: 10.21037/tlcr.2018.04.05
151
De MartinoFClementeSGraeffCPalmaGCellaL. Dose calculation algorithms for external radiation therapy: an overview for practitioners. Appl Sci. (2021) 11:6806. doi: 10.3390/app11156806
152
SchaffnerBPedroniELomaxA. Dose calculation models for proton treatment planning using a dynamic beam delivery system: an attempt to include density heterogeneity effects in the analytical dose calculation. Phys Med Biol. (1999) 44:27–41. doi: 10.1088/0031-9155/44/1/004
153
NenoffLMatterMJarhallAGWinterhalterCGorgisyanJJosipovicMet al. Daily adaptive proton therapy: is it appropriate to use analytical dose calculations for plan adaption? Int J Radiat Oncol Biol Phys. (2020) 107:747–55. doi: 10.1016/j.ijrobp.2020.03.036
154
BurlacuTLathouwersDPerkóZ. A deterministic adjoint-based semi-analytical algorithm for fast response change computations in proton therapy. J Comput Theor Transp. (2023) 52:1–41. doi: 10.1080/23324309.2023.2166077
155
BurlacuTLathouwersDPerkóZ. Yet anOther Dose Algorithm (YODA) for independent computations of dose and dose changes due to anatomical changes. Phys Med Biol. (2024) 69:165003. doi: 10.1088/1361-6560/ad6373
156
TaylorPAKrySFFollowillDS. Pencil beam algorithms are unsuitable for proton dose calculations in lung. Int J Radiat Oncol Biol Phys. (2017) 99:750–6. doi: 10.1016/j.ijrobp.2017.06.003
157
HolmesJFengHZhangLFixMKJiangSBLiuW. Fast Monte Carlo dose calculation in proton therapy. Phys Med Biol. (2024) 69:17TR01. doi: 10.1088/1361-6560/ad67a7
158
FengHPatelSHWongWWYounkinJEPenoncelloGPMoralesDHet al. GPU-accelerated Monte Carlo-based online adaptive proton therapy: A feasibility study. Med Phys. (2022) 49:3550–63. doi: 10.1002/mp.15678
159
ShanJFengHMoralesDHPatelSHWongWWFatygaMet al. Virtual particle Monte Carlo: A new concept to avoid simulating secondary particles in proton therapy dose calculation. Med Phys. (2022) 49:6666–83. doi: 10.1002/mp.15913
160
XuYCyriacJDe OrnelasMBossartEPadgettKButkusMet al. Knowledge-based planning for robustly optimized intensity-modulated proton therapy of head and neck cancer patients. Front Oncol. (2021) 11:737901. doi: 10.3389/fonc.2021.737901
161
NeishabouriAWahlNMairaniAKötheUBangertM. Long short-term memory networks for proton dose calculation in highly heterogeneous tissues. Med Phys. (2021) 48:1893–908. doi: 10.1002/mp.14658
162
Pastor-SerranoODongPHuangCXingLPerkóZ. Sub-second photon dose prediction via transformer neural networks. Med Phys. (2023) 50:3159–71. doi: 10.1002/mp.16231
163
WuCNguyenDXingYMonteroABSchuemannJShangHet al. Improving proton dose calculation accuracy by using deep learning. Mach Learn Sci Technol. (2021) 2:015017. doi: 10.1088/2632-2153/abb6d5
164
SpezialettiMLapennaFCaianielloPFracchiollaFMuciacciaFPlacidiGet al. (2021). Using deep learning for fast dose refinement in proton therapy, in: 2021 IEEE International Conference on Systems, Man, and Cybernetics (SMC). doi: 10.1109/smc52423.2021.9658879
165
WangWChangYLiuYLiangZLiaoYQinBet al. Feasibility study of fast intensity-modulated proton therapy dose prediction method using deep neural networks for prostate cancer. Med Phys. (2022) 49:5451–63. doi: 10.1002/mp.15702
166
Pastor-SerranoOPerkóZ. Millisecond speed deep learning based proton dose calculation with Monte Carlo accuracy. Phys Med Biol. (2022) 67:105006. doi: 10.1088/1361-6560/ac692e
167
NeishabouriABauerJAbdollahiADebusJMairaniA. Real-time adaptive proton therapy: An AI-based spatio-temporal mono-energetic dose calculation model (CC-LSTM). Comput Biol Med. (2025) 188:109777. doi: 10.1016/j.compbiomed.2025.109777
168
PangBChenSZengYLiuMZhangQQuanHet al. Lightweight and universal deep learning model for fast proton spot dose calculation at arbitrary energies. Phys Med Biol. (2025) 70:105014. doi: 10.1088/1361-6560/add3b9
169
RadonicDXiaoFWahlNVossLNeishabouriADelopoulosNet al. Proton dose calculation with LSTM networks in presence of a magnetic field. Phys Med Biol. (2024) 69:215019. doi: 10.1088/1361-6560/ad7f1e
170
LiuYShangXLiNWangZFangCZouYet al. An AI dose-influence matrix engine for robust pencil beam scanning protons therapy. Med Phys. (2024) 52:1903–13. doi: 10.1002/mp.17602
171
MatterMNenoffLMeierGWeberDCLomaxAJAlbertiniF. Alternatives to patient specific verification measurements in proton therapy: a comparative experimental study with intentional errors. Phys Med Biol. (2018) 63:205014. doi: 10.1088/1361-6560/aae2f4
172
AtesOPirlepesovFZhaoLHuaCMerchantTE. Development of a log file analysis tool for proton patient QA, system performance tracking, and delivered dose reconstruction. J Appl Clin Med Phys. (2023) 24:e13972. doi: 10.1002/acm2.13972
173
BookbinderABobićMSharpGCNenoffL. An operator-independent quality assurance system for automatically generated structure sets. Phys Med Biol. (2024) 69:175003. doi: 10.1088/1361-6560/ad6742
174
MeierGBessonRNanzASafaiSLomaxAJ. Independent dose calculations for commissioning, quality assurance and dose reconstruction of PBS proton therapy. Phys Med Biol. (2015) 60:2819–36. doi: 10.1088/0031-9155/60/7/2819
175
SchneiderUPedroniELomaxA. The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol. (1996) 41:111–24. doi: 10.1088/0031-9155/41/1/009
176
FaracePRighettoRMeijersA. Pencil beam proton radiography using a multilayer ionization chamber. Phys Med Biol. (2016) 61:4078–87. doi: 10.1088/0031-9155/61/11/4078
177
ParodiKPolfJC. In vivo range verification in particle therapy. Med Phys. (2018) 45:e1036-50. doi: 10.1002/mp.12960
178
MumotMAlgranatiCHartmannMSchippersJMHugELomaxAJ. Proton range verification using a range probe: definition of concept and initial analysis. Phys Med Biol. (2010) 55:4771–82. doi: 10.1088/0031-9155/55/16/010
179
Seller OriaCFreeJMarmittGGKnäuslBBrandenburgSKnopfACet al. Technical note: Flat panel proton radiography with a patient specific imaging field for accurate WEPL assessment. Med Phys. (2023) 50:1756–65. doi: 10.1002/mp.16208
180
OriaCSMarmittGGBothSLangendijkJAKnopfACMeijersA. Classification of various sources of error in range assessment using proton radiography and neural networks in head and neck cancer patients. Phys Med Biol. (2020) 65:235009. doi: 10.1088/1361-6560/abc09c
181
Seller OriaCThummererAFreeJLangendijkJABothSKnopfACet al. Range probing as a quality control tool for CBCT-based synthetic CTs: In vivo application for head and neck cancer patients. Med Phys. (2021) 48:4498–505. doi: 10.1002/mp.15020
182
PawelkeJEnghardtWHabererTHaschBGHinzRKramerMet al. In-beam PET imaging for the control of heavy-ion tumour therapy. IEEE Trans Nucl Sci. (1997) 44:1492–8. doi: 10.1109/23.632694
183
ParodiKEnghardtWHabererT. In-beam PET measurements of β+radioactivity induced by proton beams. Phys Med Biol. (2001) 47:21–36. doi: 10.1088/0031-9155/47/1/302
184
ParodiKBortfeldTEnghardtWFiedlerFKnopfAPaganettiHet al. PET imaging for treatment verification of ion therapy: Implementation and experience at GSI Darmstadt and MGH Boston. Nucl Instrum Methods Phys Res A. (2008) 591:282–6. doi: 10.1016/j.nima.2008.03.075
185
ParodiKPaganettiHCascioEFlanzJBBonabAAAlpertNMet al. PET/CT imaging for treatment verification after proton therapy: A study with plastic phantoms and metallic implants. Med Phys. (2007) 34:419–35. doi: 10.1118/1.2401042
186
ParodiKYamayaTMoskalP. Experience and new prospects of PET imaging for ion beam therapy monitoring. Z Med Phys. (2023) 33:22–34. doi: 10.1016/j.zemedi.2022.11.001
187
ShakirinGBraessHFiedlerFKunathDLaubeKParodiKet al. Implementation and workflow for PET monitoring of therapeutic ion irradiation: a comparison of in-beam, in-room, and off-line techniques. Phys Med Biol. (2011) 56:1281–98. doi: 10.1088/0031-9155/56/5/004
188
FiorinaEFerreroVBaroniGBattistoniGBelcariNCamarlinghiNet al. Detection of interfractional morphological changes in proton therapy: A simulation and in vivo study with the INSIDE in-beam PET. Front Phys. (2021) 8:578388. doi: 10.3389/fphy.2020.578388
189
MoglioniMKraanACBaroniGBattistoniGBelcariNBertiAet al. In-vivo range verification analysis with in-beam PET data for patients treated with proton therapy at CNAO. Front Oncol. (2022) 12:929949. doi: 10.3389/fonc.2022.929949
190
ToramatsuCYoshidaEWakizakaHMohammadiAIkomaYTashimaHet al. Washout effect in rabbit brain: in-beam PET measurements using 10 C, 11 C and 15 O ion beams. BioMed Phys Eng Express. (2018) 4:035001. doi: 10.1088/2057-1976/aaade7
191
GianoliCDe BernardiERicottiRKurzCBauerJRiboldiMet al. First clinical investigation of a 4D maximum likelihood reconstruction for 4D PET-based treatment verification in ion beam therapy. Radiotherapy Oncol. (2017) 123:339–45. doi: 10.1016/j.radonc.2017.04.018
192
LaubeKMenkelSBertCEnghardtWHelmbrechtSSaitoNet al. 4D particle therapy PET simulation for moving targets irradiated with scanned ion beams. Phys Med Biol. (2013) 58:513–33. doi: 10.1088/0031-9155/58/3/513
193
StützerKBertCEnghardtWHelmbrechtSParodiKPriegnitzMet al. Experimental verification of a 4D MLEM reconstruction algorithm used for in-beam PET measurements in particle therapy. Phys Med Biol. (2013) 58:5085–111. doi: 10.1088/0031-9155/58/15/5085
194
PriegnitzMMöckelDParodiKSommererFFiedlerFEnghardtW. In-beam PET measurement of 7Li3+ irradiation induced β+-activity. Phys Med Biol. (2008) 53:4443–53. doi: 10.1088/0031-9155/53/16/015
195
DendoovenPBuitenhuisHJTDiblenFHeeresPNBiegunAKFiedlerFet al. Short-lived positron emitters in beam-on PET imaging during proton therapy. Phys Med Biol. (2015) 60:8923–47. doi: 10.1088/0031-9155/60/23/8923
196
CuccagnaCBattistoniGBisogniMGCerelloPDel GuerraAFerreroVet al. Few-seconds range verification with short-lived positron emitters in carbon ion therapy. Phys Med. (2024) 118:103209. doi: 10.1016/j.ejmp.2024.103209
197
BrzezińskiKBaranJBorysDGajewskiJChugNCoussatAet al. Detection of range shifts in proton beam therapy using the J-PET scanner: a patient simulation study. Phys Med Biol. (2023) 68:145016. doi: 10.1088/1361-6560/acdd4c
198
ZhangRMuDMaQWanLXiaoPQiPet al. Proton spot dose estimation based on positron activity distributions with neural network. Med Phys. (2024) 51:7226–39. doi: 10.1002/mp.17297
199
VandenbergheSMikhaylovaED’HoeEMolletPKarpJS. Recent developments in time-of-flight PET. EJNMMI Phys. (2016) 3. doi: 10.1186/s40658-016-0138-3
200
ContiMBendriemB. The new opportunities for high time resolution clinical TOF PET. Clin Transl Imaging. (2019) 7:139–47. doi: 10.1007/s40336-019-00316-5
201
YoshidaETashimaHNagatsuKTsujiABKamadaKParodiKet al. Whole gamma imaging: a new concept of PET combined with Compton imaging. Phys Med Biol. (2020) 65:125013. doi: 10.1088/1361-6560/ab8e89
202
StichelbautFJongenY. (2003). Verification of the proton beam position in the patient by the detection of prompt gamma-rays emission, in: 39th meeting of the Particle Therapy Co-Operative Group (PTCOG).
203
PintoM. Prompt-gamma imaging in particle therapy. Eur Phys J Plus. (2024) 139. doi: 10.1140/epjp/s13360-024-05664-4
204
KrimmerJDauvergneDLétangJMTestaÉ. Prompt-gamma monitoring in hadrontherapy: A review. Nucl Instrum Methods Phys Res A. (2018) 878:58–73. doi: 10.1016/j.nima.2017.07.063
205
CheonB-WMinCH. Prompt gamma imaging system in particle therapy: a mini-review. Front Phys. (2024) 12:1356572. doi: 10.3389/fphy.2024.1356572
206
WernerTBertholdJEnghardtWHueso-GonzalezFKoglerTPetzoldtJet al. (2017). Range verification in proton therapy by prompt gamma-ray timing (PGT): steps towards clinical implementation, in: 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). pp. 1–5. doi: 10.1109/nssmic.2017.8532807
207
VerburgJMSecoJ. Proton range verification through prompt gamma-ray spectroscopy. Phys Med Biol. (2014) 59:7089–106. doi: 10.1088/0031-9155/59/23/7089
208
RichterCPauschGBarczykSPriegnitzMKeitzIThieleJet al. First clinical application of a prompt gamma based in vivo proton range verification system. Radiother Oncol. (2016) 118:232–7. doi: 10.1016/j.radonc.2016.01.004
209
XieYBentefourEHJanssensGSmeetsJVander StappenFHotoiuLet al. Prompt gamma imaging for in vivo range verification of pencil beam scanning proton therapy. Int J Radiat Oncol Biol Phys. (2017) 99:210–8. doi: 10.1016/j.ijrobp.2017.04.027
210
Evaluation of prompt-gamma imaging for range verification for the proton irradiation of tumor patients (PRIMA. In: German clinical trial register. DRKS00009224.
211
VerburgJMHueso-GonzalezFTattenbergSWohlfahrtPRuggieriTBortfeldT. First-in-human use of prompt gamma-ray spectroscopy for proton range verification. In: 2020 Joint AAPM | COMP virtual meeting (2020). Hoboken, New Jersey, USA: Wiley. Available online at: https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.14316.(Accessed November 27, 2025)
212
Hueso-GonzálezFBertholdJWohlfahrtPBortfeldTKhamfongkhrueaCTattenbergSet al. Inter-center comparison of proton range verification prototypes with an anthropomorphic head phantom. Phys Med Biol. (2024) 69:225010. doi: 10.1088/1361-6560/ad8856
213
SterpinEJanssensGSmeetsJStappenFVPrieelsDPriegnitzMet al. Analytical computation of prompt gamma ray emission and detection for proton range verification. Phys Med Biol. (2015) 60:4915–46. doi: 10.1088/0031-9155/60/12/4915
214
HuangZTianLJanssensGSmeetsJXieYKevin TeoB-Ket al. An experimental validation of a filtering approach for prompt gamma prediction in a research proton treatment planning system. Phys Med Biol. (2024) 69:155025. doi: 10.1088/1361-6560/ad6116
215
BertholdJPietschJPiplackNKhamfongkhrueaCThieleJHölscherTet al. Detectability of anatomical changes with prompt-gamma imaging: first systematic evaluation of clinical application during prostate-cancer proton therapy. Int J Radiat Oncol Biol Phys. (2023) 117:718–29. doi: 10.1016/j.ijrobp.2023.05.002
216
PietschJKhamfongkhrueaCBertholdJJanssensGStützerKLöckSet al. Automatic detection and classification of treatment deviations in proton therapy from realistically simulated prompt gamma imaging data. Med Phys. (2022) 50:506–17. doi: 10.1002/mp.15975
217
KhamfongkhrueaCBertholdJJanssensGPetzoldtJSmeetsJPauschGet al. Classification of the source of treatment deviation in proton therapy using prompt-gamma imaging information. Med Phys. (2020) 47:5102–11. doi: 10.1002/mp.14393
218
TianLLandryGDedesGPintoMKampFBelkaCet al. A new treatment planning approach accounting for prompt gamma range verification and interfractional anatomical changes. Phys Med Biol. (2020) 65:095005. doi: 10.1088/1361-6560/ab7d15
219
BertholdJKhamfongkhrueaCPetzoldtJThieleJHölscherTWohlfahrtPet al. First-in-human validation of CT-based proton range prediction using prompt gamma imaging in prostate cancer treatments. Int J Radiat Oncol Biol Phys. (2021) 111:1033–43. doi: 10.1016/j.ijrobp.2021.06.036
220
EveraerePDauvergneDGallin-MartelM-LHéraultJKoudiaAKoumeirCet al. Prompt gamma energy integration: a new method for online-range verification in proton therapy with pulsed-beams. Front Phys. (2024) 12:1371015. doi: 10.3389/fphy.2024.1371015
221
ParodiKAssmannW. Ionoacoustics: A new direct method for range verification. Mod Phys Lett A. (2015) 30:1540025. doi: 10.1142/s0217732315400258
222
KirschLAssmannWGerlachSSchmidtA-KBenderMParodiKet al. Ionoacoustic monitoring of relativistic heavy ion beams. Nucl Instrum Methods Phys Res A. (2023) 1057:168755. doi: 10.1016/j.nima.2023.168755
223
SchauerJWieserHPHuangYRuserHLascaudJWürlMet al. Proton beam range verification by means of ionoacoustic measurements at clinically relevant doses using a correlation-based evaluation. Front Oncol. (2022) 12:925542. doi: 10.3389/fonc.2022.925542
224
SchauerJWieserHPLascaudJHuangYVidalMHeraultJet al. Range verification of a clinical proton beam in an abdominal phantom by co-registration of ionoacoustics and ultrasound. Phys Med Biol. (2023) 68:125009. doi: 10.1088/1361-6560/acd834
225
FreijoCHerraizJLSanchez-ParcerisaDUdiasJM. Dictionary-based protoacoustic dose map imaging for proton range verification. Photoacoustics. (2021) 21:100240. doi: 10.1016/j.pacs.2021.100240
226
CaronJGonzalezGPandeyPKWangSPratherKAhmadSet al. Single pulse protoacoustic range verification using a clinical synchrocyclotron. Phys Med Biol. (2023) 68:045011. doi: 10.1088/1361-6560/acb2ae
227
JiangZSunLYaoWWuQJXiangLRenL. 3D in vivo dose verification in prostate proton therapy with deep learning-based proton-acoustic imaging. Phys Med Biol. (2022) 67:215012. doi: 10.1088/1361-6560/ac9881
228
KellnbergerSAssmannWLehrackSReinhardtSThirolfPQueirósDet al. Ionoacoustic tomography of the proton Bragg peak in combination with ultrasound and optoacoustic imaging. Sci Rep. (2016) 6. doi: 10.1038/srep29305
229
LiHSahooNPoenischFSuzukiKLiYLiXet al. Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance. Med Phys. (2013) 40:021703. doi: 10.1118/1.4773312
230
ScandurraDAlbertiniFvan der MeerRMeierGWeberDCBolsiAet al. Assessing the quality of proton PBS treatment delivery using machine log files: comprehensive analysis of clinical treatments delivered at PSI Gantry 2. Phys Med Biol. (2016) 61:1171–81. doi: 10.1088/0031-9155/61/3/1171
231
GaleoneCSteinsbergerTDonettiMMartireMCMilianFMSacchiRet al. Real-time delivered dose assessment in carbon ion therapy of moving targets. Phys Med Biol. (2024) 69:205001. doi: 10.1088/1361-6560/ad7d59
232
ESTRO. Priorities of ESTRO for engagement within the EURAMED framework (2025). Available online at: https://www.estro.org/About/Newsroom/Newsletter/Research-Projects/Priorities-of-ESTRO-for-engagement-within-the-EURA (Accessed April 09, 2025).
233
SuZHsiWCO’neilMSabouriP. A comprehensive evaluation of an independent quality assurance system for intensity modulated proton therapy treatment plans and deliveries. Int J Radiat Oncol Biol Phys. (2024) 120:e589. doi: 10.1016/j.ijrobp.2024.07.1298
234
MeijersAJakobiAStützerKGuterres MarmittGBothSLangendijkJAet al. Log file-based dose reconstruction and accumulation for 4D adaptive pencil beam scanned proton therapy in a clinical treatment planning system: Implementation and proof-of-concept. Med Phys. (2019) 46:1140–9. doi: 10.1002/mp.13371
235
MurrMBrockKKFusellaMHardcastleNHusseinMJamesonMGet al. Applicability and usage of dose mapping/accumulation in radiotherapy. Radiother Oncol. (2023) 182:109527. doi: 10.1016/j.radonc.2023.109527
236
NenoffLAmstutzFMurrMArchibald-HeerenBFusellaMHusseinMet al. Review and recommendations on deformable image registration uncertainties for radiotherapy applications. Phys Med Biol. (2023) 68:24TR01. doi: 10.1088/1361-6560/ad0d8a
237
NenoffLRibeiroCOMatterMHafnerLJosipovicMLangendijkJAet al. Deformable image registration uncertainty for inter-fractional dose accumulation of lung cancer proton therapy. Radiother Oncol. (2020) 147:178–85. doi: 10.1016/j.radonc.2020.04.046
238
AmstutzFNenoffLAlbertiniFRibeiroCOKnopfACUnkelbachJet al. An approach for estimating dosimetric uncertainties in deformable dose accumulation in pencil beam scanning proton therapy for lung cancer. Phys Med Biol. (2021) 66:105007. doi: 10.1088/1361-6560/abf8f5
239
WuXAmstutzFWeberDCUnkelbachJLomaxAJZhangY. Patient-specific quality assurance for deformable IMRT/IMPT dose accumulation: Proposition and validation of energy conservation based validation criterion. Med Phys. (2023) 50:7130–8. doi: 10.1002/mp.16564
240
Ng Wei SiangKBothSOldehinkelELangendijkJAWagenaarD. Assessment of residual geometrical errors of clinical target volumes and their impact on dose accumulation for head and neck radiotherapy. Radiother Oncol. (2023) 188:109856. doi: 10.1016/j.radonc.2023.109856
241
SmoldersALomaxAWeberDCAlbertiniF. Deep learning based uncertainty prediction of deformable image registration for contour propagation and dose accumulation in online adaptive radiotherapy. Phys Med Biol. (2023) 68:245027. doi: 10.1088/1361-6560/ad0282
242
The RAPTOR consortium. Towards interventional particle therapy for advanced cancer care (2021). Available online at: https://raptor-consortium.com/raptor-white-paper/ (Accessed December 9, 2025).
243
ProtOnART – a new consortium for proton online adaptive radiation therapy (2022). Available online at: https://www.oncoray.de/news/article/protonart-a-new-consortium-for-proton-online-adaptive-radiation-therapy (Accessed December 9, 2025).
244
International Commission on Radiation Units and Measurements. ICRU report 78: Prescribing, recording, and reporting proton beam therapy (2007). Available online at: https://www.icru.org/report/prescribing-recording-and-reporting-proton-beam-therapy-icru-report-78/ (Accessed April 09, 2025).
245
GoudschaalKAzzaroualiSVisserJAdmiraalMWiersmaJvan WieringenNet al. Clinical implementation of RTT-only CBCT-guided online adaptive focal radiotherapy for bladder cancer. Clin Trans Radiat Oncol. (2025) 50:100884. doi: 10.1016/j.ctro.2024.100884
246
BeckertRSchiffJPMorrisESamsonPKimHLaugemanE. The impact of an Advanced Practice Radiation Therapist contouring for a CBCT-based adaptive radiotherapy program. Tech Innov Patient Support Radiat Oncol. (2024) 30:100242. doi: 10.1016/j.tipsro.2024.100242
247
ChettyIJCaiBChuongMDDawesSLHallWAHelmsARet al. Quality and safety considerations for adaptive radiation therapy: an ASTRO white paper. Int J Radiat OncologyBiologyPhysics. (2025) 122:838–64. doi: 10.1016/j.ijrobp.2024.10.011
248
OlsenDRBrulandØSFrykholmGNorderhaugIN. Proton therapy – A systematic review of clinical effectiveness. Radiother Oncol. (2007) 83:123–32. doi: 10.1016/j.radonc.2007.03.001
249
LiMLiXYaoLHanXYanWLiuYet al. Clinical efficacy and safety of proton and carbon ion radiotherapy for prostate cancer: A systematic review and meta-analysis. Front Oncol. (2021) 11:709530. doi: 10.3389/fonc.2021.709530
250
ChenZDominelloMMJoinerMCBurmeisterJW. Proton versus photon radiation therapy: A clinical review. Front Oncol. (2023) 13:1133909. doi: 10.3389/fonc.2023.1133909
251
ZhouPDuYZhangYZhuMLiTTianWet al. Efficacy and safety in proton therapy and photon therapy for patients with esophageal cancer. JAMA Netw Open. (2023) 6:e2328136. doi: 10.1001/jamanetworkopen.2023.28136
252
PressRHMehtaMP. Proton therapy: current status and controversies. JCO Oncol Pract. (2024) 20:747–9. doi: 10.1200/op.24.00132
253
EfstathiouJAYeapBYMichalskiJMHorickNZietmanALChristodouleasJPet al. Prostate advanced radiation technologies investigating quality of life (PARTIQoL): phase III randomized clinical trial of proton therapy vs. IMRT for localized prostate cancer. Int J Radiat Oncol Biol Phys. (2024) 120:S1. doi: 10.1016/j.ijrobp.2024.08.012
254
FrankSJBussePRosenthalDIHernandezMSwansonDMGardenASet al. Phase III randomized trial of intensity-modulated proton therapy (IMPT) versus intensity-modulated photon therapy (IMRT) for the treatment of head and neck oropharyngeal carcinoma (OPC). JCO. (2024) 42:6006–6. doi: 10.1200/jco.2024.42.16_suppl.6006
255
McComasKNYockADarrowKShinoharaET. Online adaptive radiation therapy and opportunity cost. Adv Radiat Oncol. (2023) 8:101034. doi: 10.1016/j.adro.2022.101034
256
Borderías-VillarroelEBarragán-MonteroASterpinE. Time is NTCP: Should we maximize patient throughput or perform online adaptation on proton therapy systems? Radiother Oncol. (2024) 198:110389. doi: 10.1016/j.radonc.2024.110389
257
OudMBreedveldSGiżyńskaMChenYHHabrakenSPerkóZet al. Dosimetric advantages of adaptive IMPT vs. Enhanced workload and treatment time – A need for automation. Radiother Oncol. (2024) 201:110548. doi: 10.1016/j.radonc.2024.110548
258
ThomsenSNMøllerDSKnapMMKhalilAAShcytteTHoffmannL. Daily CBCT-based dose calculations for enhancing the safety of dose-escalation in lung cancer radiotherapy. Radiother Oncol. (2024) 200:110506. doi: 10.1016/j.radonc.2024.110506
259
RheeDJBeddarSJaoudeJASawakuchiGMartinRPerlesLet al. Dose escalation for pancreas SBRT: potential and limitations of using daily online adaptive radiation therapy and an iterative isotoxicity automated planning approach. Adv Radiat Oncol. (2023) 8:101164. doi: 10.1016/j.adro.2022.101164
260
ErmongkonchaiTKhorRMuralidharanVTebbuttNLimKKutaibaNet al. Stereotactic radiotherapy and the potential role of magnetic resonance-guided adaptive techniques for pancreatic cancer. World J Gastroenterol. (2022) 28:745–54. doi: 10.3748/wjg.v28.i7.745
261
BainbridgeHEMentenMJFastMFNillSOelfkeUMcDonaldF. Treating locally advanced lung cancer with a 1.5 T MR-Linac – Effects of the magnetic field and irradiation geometry on conventionally fractionated and isotoxic dose-escalated radiotherapy. Radiother Oncol. (2017) 125:280–5. doi: 10.1016/j.radonc.2017.09.009
262
MacDonaldRLFalloneCChytyk-PraznikKRobarJCherpakA. The feasibility of CT simulation-free adaptive radiation therapy. J Appl Clin Med Phys. (2024) 25: e14438. doi: 10.1002/acm2.14438
263
de LeonJJelenUCarrMCrawfordDPictonMTranCet al. Adapting outside the box: Simulation-free MR-guided stereotactic ablative radiotherapy for prostate cancer. Radiother Oncol. (2024) 200:110527. doi: 10.1016/j.radonc.2024.110527
264
PriceATSchiffJPSilbersteinABeckertRZhaoTHugoGDet al. Feasibility of simulation free abdominal stereotactic adaptive radiotherapy using an expedited pre-plan workflow. Phys Imaging Radiat Oncol. (2024) 31:100611. doi: 10.1016/j.phro.2024.100611
Summary
Keywords
cancer, proton therapy, online adaptive proton therapy, near real-time adaptation, volumetric imaging, online plan optimization, online QA, treatment verification
Citation
Gambetta V, Stützer K and Richter C (2025) Current status and upcoming developments for online adaptive proton therapy enabling a closed feedback loop for near real-time adaptation. Front. Oncol. 15:1660605. doi: 10.3389/fonc.2025.1660605
Received
06 July 2025
Revised
10 November 2025
Accepted
17 November 2025
Published
17 December 2025
Volume
15 - 2025
Edited by
Konrad P. Nesteruk, Massachusetts General Hospital and Harvard Medical School, United States
Reviewed by
Mirek Fatyga, Mayo Clinic Arizona, United States
Peilin Liu, William Beaumont Hospital, United States
Mislav Bobić, Massachusetts General Hospital and Harvard Medical School, United States
Updates

Check for updates
Copyright
© 2025 Gambetta, Stützer and Richter.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Christian Richter, christian.richter@oncoray.de
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.