Abstract
Posterior chamber implantable collamer lens (ICL) implantation provides additive refractive correction for selected phakic patients, especially those with high myopia and limited corneal laser options. This narrative Mini Review synthesizes literature published mainly from January 2021 through May 2026 to connect patient selection with measurable safety boundaries, separate lens power and size planning, dynamic vault assessment, and long-term surveillance. It also discusses hyperopic ICLs, non-EVO lenses, alternative phakic IOL platforms such as IPCL, and emerging artificial intelligence. True anterior chamber depth, age-appropriate endothelial reserve, open angles, crystalline lens status, and refractive stability define eligibility. Central vault is interpreted using pragmatic numerical categories alongside angle, pressure, and lens findings; no single vault threshold guarantees safety or mandates exchange. Refractive and vault prediction models have different inputs and outcomes, and favorable average accuracy does not establish performance in extreme or underrepresented anatomy. Modern central-port ICLs show good outcomes in appropriately selected myopic eyes, while evidence for hyperopic and alternative platforms is smaller and not interchangeable. The proposed clinical framework represents expert synthesis rather than a validated decision rule. Future priorities include externally validated multimodal prediction and prospective lifelong safety monitoring.
1 Introduction
Posterior chamber implantable collamer lens (ICL) surgery provides additive refractive correction while preserving the crystalline lens and corneal tissue. Central-port EVO/EVO+ designs have broadened its role in moderate to high myopia, particularly when corneal thickness, topography, ocular surface status, or optical demands limit laser surgery (ā). However, removability does not reverse every complication: cataract, endothelial loss, glaucoma, and additional intraocular surgery remain relevant risks.
This review links candidate selection to measurable safety boundaries, lens power and size selection, dynamic vault assessment, and lifelong surveillance. High myopia remains the principal focus, with hyperopic ICLs and alternative phakic intraocular lenses (pIOLs) included to clarify the limits of extrapolating EVO evidence. The proposed clinical framework is an expert narrative synthesis, not a prospectively validated decision tool.
2 Search strategy and scope
The narrative literature review prioritizes English-language articles published from January 2021 through May 2026, with no lower date restriction for foundational device studies and long-term reports. This is the publication eligibility window, rather than a period of patient data collection. PubMed/MEDLINE was the principal source, supplemented by reference lists and publisher pages, including Frontiers. Search terms included implantable collamer lens, EVO/EVO+, V4c/V5, toric ICL, vault, sizing, endothelial cell density, cataract, intraocular pressure, high myopia, SMILE, and LASIK. During revision, targeted searches added hyperopia, IPCL, other phakic IOLs, power calculation, predictive formula, artificial intelligence, and machine learning; the May 2026 publication cutoff was retained.
Clinical cohorts, comparative studies, prediction-model studies, systematic reviews, and meta-analyses were considered when they addressed selection, visual/refraction outcomes, sizing, vault, or complications. Case reports informed unusual safety mechanisms, not incidence estimates; regulatory documents clarified device-specific eligibility. Duplicate publications, nonclinical material without relevant design information, and reports outside scope were not retained. Evidence was synthesized by clinical domain, considering device generation, study design, follow-up, sample size, and validation setting. FigureĀ 1 describes this qualitative selection pathway. No formal PRISMA screening, reconstructed screening counts, quantitative pooling, or formal risk-of-bias assessment was undertaken.
FigureĀ 1
3 Why ICL is a distinct refractive category
SMILE, femtosecond laser-assisted LASIK (FS-LASIK), and transepithelial photorefractive keratectomy (TransPRK) alter corneal shape; ICL adds a posterior chamber optical element. Tissue preservation can be valuable in high corrections, but must be balanced against intraocular infection, inflammation, cataract, endothelial, and pressure-related risks (ā). The 0.36-mm central port of EVO/EVO+ permits aqueous flow and removes the routine need for prophylactic peripheral iridotomy for these models; it does not eliminate pupillary block or sizing-related complications (, , ).
4 Patient selection and preoperative safety
Suitable candidates have stable refraction, a clear crystalline lens, adequate true anterior chamber depth (ACD), open angles, and age-appropriate endothelial reserve. TableĀ 1 summarizes preoperative boundaries and postoperative monitoring. True ACD is measured from corneal endothelium to the anterior lens surface; an epithelial-reference measurement must not be used interchangeably. Thresholds depend on the model and jurisdiction. For US EVO labeling, true ACD must be at least 3.0Ā mm; minimum endothelial cell density (ECD) varies with both age and ACD ().
TableĀ 1
| Domain | Preoperative boundary | Postoperative assessment |
|---|---|---|
| Age and stability | Follow local labeling. US EVO age range: 21ā60 years; stable refraction within 0.50 D for at least 1 year (, ). | Review age-related lens change and refractive drift; counsel on presbyopia and future cataract surgery. |
| True ACD and angle | US EVO: ACD >=3.0 mm from endothelium to lens; gonioscopic angle at least grade III. Other device/jurisdiction thresholds differ (, ). | Assess angle crowding and peripheral clearance with gonioscopy/AS-OCT; a central vault alone does not establish safety. |
| Endothelial reserve | Use the model-specific age/ACD matrix. Example US EVO, age 26-30: minimum ECD 3,425, 3,375, and 2,900 cells/mm2 for true ACD >=3.0, >=3.2, and >=3.5 mm, respectively (). | Serial specular microscopy using comparable methods; investigate sustained decline, abnormal morphology, or edema (, ). |
| Crystalline lens | Clear lens; assess lens rise, configuration, and anticipated cataract risk (, ). | Monitor opacity and lens contact; evaluate symptoms and progression before deciding on exchange/explantation. |
| IOP and glaucoma | No uncontrolled glaucoma, narrow-angle disease, or unexplained ocular hypertension; normal IOP does not exclude glaucoma (). | Check IOP at 1ā6 hours per US directions; reassess early and during steroid treatment. New pressure rise, pain, or angle closure warrants prompt evaluation (, ). |
| Central vault | Compare multimodal measurements and plausible sizes; do not rely on a single nomogram (ā). | Low <250; common target 250-750; high >750-1,000; excessive >1,000 micrometers. <100 or lens contact is particularly concerning. Interpret with lens clarity, angle, and IOP (, , ). |
| Retina and follow-up | Document retinal/macular status, visual potential, and capacity for continued surveillance (, ). | Continue review of vault, ECD, IOP/angle, lens clarity, refraction, and symptoms. New flashes/floaters or field loss needs urgent retinal assessment. |
Preoperative and postoperative safety boundaries.
Regulatory examples refer to the US EVO platform; clinicians must use the current labeling for the actual model and jurisdiction. ECD examples are minimum eligibility values, not targets. Vault categories are pragmatic reporting ranges, not automatic intervention rules. Long-term visit frequency should be individualized to anatomy, symptoms, and observed trends.
US approval expanded the EVO/EVO+ ICL/TICL age range from 21ā45 to 21ā60 years in November 2025 (). This regulatory range does not establish suitability for every older patient. Lens opacity, lens rise, angle crowding, presbyopia, and the likely interval to cataract surgery require separate assessment. A fixed age threshold around 40 years should therefore not determine the choice between ICL and lens extraction. Patients must also understand persistent high-myopia retinal risk and commit to long-term follow-up (, , , ).
4.1 Hyperopic ICLs
Hyperopic ICL can be considered for selected accommodating adults with stable moderate to high hyperopia when corneal correction is unsuitable and anterior segment anatomy meets the relevant lens labeling. Cycloplegic refraction helps identify latent hyperopia. Available published designs include spherical ICH V3 and toric V4B lenses; availability and approval vary by country. These positive-power, non-EVO lenses should not inherit the central-port claims or myopic sizing assumptions of EVO. Peripheral iridotomy is required for the non-port designs (ā).
Hyperopic eyes often have shorter axial length, shallower chambers, and narrower angles; positive-power optics and crystalline lens configuration can further constrain space. Vault must therefore be interpreted alongside peripheral lens clearance and angle anatomy. There is no robust evidence that hyperopic vault is uniformly higher or lower than myopic vault: a small older-model cohort reported similar long-term central vault in the two groups (). Amblyopia, latent hyperopia, and limited toric availability can also constrain visual gains and predictability.
In 20 eyes receiving hyperopic toric ICL, mean vault decreased from approximately 650 micrometers at one month to 613 micrometers at one year (). A 2025 single-arm study of spherical ICL followed by LASIK in 48 hyperopic astigmatic eyes reported 93.8% within +/-0.75 D of target, but assessed the combined procedure over short follow-up (). These findings cannot establish equivalence to myopic EVO outcomes or long-term safety. A shallow or narrow-angle hyperopic eye remains unsuitable for pIOL implantation despite a large refractive error.
5 Non-EVO ICL and alternative phakic IOLs
Earlier non-port Collamer ICLs use posterior chamber plate-haptic support and require an alternative aqueous pathway through peripheral iridotomy. IPCL V2.0 instead uses a hydrophilic acrylic platform with a different haptic geometry and a central port in the studied myopic design; spherical, toric, hyperopic, and diffractive options are model- and market-dependent (, ā). Both posterior chamber platforms vault through the interaction of lens geometry and anatomical support. Identical lens lengths, powers, or vault nomograms cannot be assumed interchangeable.
A randomized comparative study of 60 myopic eyes found visual improvement with both central-port ICL and IPCL over 12 months, without a detected difference in reported adverse effects; its size and duration cannot exclude differences in rare or late complications (). Eyecryl is another hydrophilic acrylic posterior chamber platform, whereas Artisan/Artiflex lenses use anterior chamber iris fixation. Their endothelial clearance and fixation-related risks differ from posterior chamber vault concerns (, ). Supplementary Table S1 compares design, optics, indications, support, and evidence limitations. Available comparisons support individualized selection, not device equivalence.
6 Preoperative imaging and lens sizing
Traditional sizing uses white-to-white distance (WTW) and ACD as surrogates for posterior chamber anatomy. Anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) add anterior chamber width (ACW), angle-to-angle distance (ATA), sulcus-to-sulcus distance (STS), crystalline lens rise (CLR), and iridociliary morphology (, , ). Measurements from different devices may yield different size recommendations ().
The NK formula incorporates ACW and CLR with lens size; the KS approach uses ATA and lens size. In a 106-eye comparison, achieved-minus-predicted vault bias was -104.82 micrometers for NK and -68.55 micrometers for KS, with wide agreement limits and no significant overall difference in absolute prediction error (). Thus a smaller mean bias does not establish greater precision. Haptic placement, iris shape, and inter-eye differences warrant multimodal review, especially when two adjacent sizes appear plausible (ā).
7 ICL power calculation and refractive prediction
Power selection determines refractive correction; sizing and vault prediction determine anatomical fit. These tasks require distinct endpoints. Manufacturer power calculations use a modified vergence approach based on spectacle-plane refraction, vertex distance, keratometry, ACD, and target refraction; toric planning additionally considers cylinder and axis (, ). Repeatable manifest and cycloplegic measurements and correct vertex conversion are especially important at large refractive errors. Cataract IOL formulas should not be substituted directly because the natural lens remains in situ.
Jiang etĀ al. evaluated 2,767 eyes using stacked regression models with refraction, age, ocular biometry, and implanted lens parameters. The best spherical-equivalent prediction mean absolute errors were 0.339 D for non-toric ICL and 0.325 D for toric ICL; overall accuracy was comparable to the modified vergence formula, with possible advantages concentrated in low-to-moderate myopia (). This does not demonstrate superiority in extreme myopia, hyperopia, or atypical chambers. Supplementary Table S2 separates refractive and vault models, their inputs, and reported accuracy. No universally superior formula is established for these uncommon phenotypes.
8 Vault as a dynamic safety biomarker
Central vault is the perpendicular separation between the posterior ICL surface and anterior crystalline lens surface. For consistent terminology, this review defines low vault as <250 micrometers, a commonly targeted range as 250ā750 micrometers, high vault as >750 to 1,000 micrometers, and excessive vault as >1,000 micrometers; <100 micrometers or lens contact deserves particular concern (, , ). These pragmatic categories are not validated universal intervention thresholds, and published acceptable ranges differ ().
Low vault raises concern for lens contact and anterior subcapsular opacity; excessive vault may crowd the angle and elevate intraocular pressure (IOP). A low stable vault without contact or opacity, or a high vault with open angles and normal IOP, may be observed. Lens touch, progressive cataract, angle closure, pigment dispersion, or pressure elevation requires mechanism-based reassessment, potentially including repositioning, exchange, or explantation (, , ). Management must not be triggered by a central number alone.
Illumination, pupil diameter, accommodation, imaging device, and postoperative timing affect vault. Serial measurements should use comparable conditions and include angle assessment. Posterior chamber morphology, haptic position, lens aging, and rotation explain changes over time and differences between fellow eyes (, , , ).
9 Visual and refractive outcomes
Reviews and cohorts support good efficacy and predictability in selected high and super-high myopic eyes (, ). Comparisons with SMILE suggest possible optical advantages for ICL at higher corrections, but selection bias, baseline differences, and heterogeneous endpoints limit procedure rankings (, ). Corneal procedures avoid intraocular risks and may suit lower corrections with favorable corneal parameters. Glare, halos, mesopic pupil size, and subjective visual quality remain relevant even with excellent acuity.
Toric ICL offers correction of regular astigmatism, but residual cylinder depends on measurement repeatability, axis alignment, and postoperative rotation (). Vertical placement has been investigated to improve stability; it is not a universal recommendation and must respect model-specific axis planning ().
10 Safety of the endothelium crystalline lens and IOP
Safety assessment includes ECD, lens transparency, IOP, angle configuration, inflammation, infection, retinal status, and reoperation. Meta-analyses support favorable outcomes in selected eyes while demonstrating nonzero complication risks (, ). Endothelial loss may be clinically silent; baseline reserve, measurement reliability, surgical trauma, and longitudinal trends matter more than a single acceptable postoperative count (). Cataract remains relevant despite central-port design, particularly with older age, lens contact, or adverse lens anatomy (, ).
Early pressure elevation can reflect retained viscoelastic, pupillary block, or excessive vault, whereas later elevation may reflect steroid response or pigment-related mechanisms. FDA directions advise an initial IOP check within 1ā6 hours after implantation; subsequent surveillance should follow clinical findings and local protocols (, ). Adverse-event databases identify safety signals but cannot estimate incidence or compare devices reliably without denominators and control for reporting bias (, ).
Reports of late glaucoma with first-generation pIOLs and perioperative drug-related maculopathy illustrate surveillance and operative hazards; they should not be used to quantify current EVO risk (, ). Patients developing cataract may need ICL removal with phacoemulsification, an additional procedure that should be anticipated in lifetime counseling ().
11 Special clinical scenarios
In high and super-high myopia, retinal, macular, and optic nerve disease may limit visual potential independently of refraction. ICL corrects ametropia, not pathological myopia (). Concave iris or atypical posterior chamber morphology requires careful imaging because predicted support and vault may be unreliable; unfavorable anatomy may justify deferring surgery ().
Presbyopia-age candidates require discussion of reading needs, tolerance of monovision, and future cataract surgery. Small comparative studies of ICL and FS-LASIK monovision provide preliminary evidence, not a preferred strategy for all older patients (, ). Clear-lens status and anatomical eligibility remain necessary even within an approved age range.
12 Selection against corneal and lens alternatives
Thin corneas, high refractive correction, or ocular surface vulnerability may favor ICL, while shallow chambers, inadequate ECD, narrow angles, and lens opacity favor another pathway. FS-LASIK, SMILE, and TransPRK avoid intraocular implant risks but have different tissue, flap, surface-healing, and dry-eye trade-offs (, , ). Refractive lens exchange removes accommodation and needs particular caution in young high myopes. Patient priorities and suitability for each procedure should determine selection; tissue preservation alone cannot override pIOL safety boundaries.
13 Artificial intelligence in individualized planning
AI can combine refraction, lens parameters, biometry, and anterior segment imaging to predict postoperative refraction or vault and flag eyes at risk of an extreme vault. In 1,745 eyes, AS-OCT-based machine learning improved vault prediction over the conventional nomogram (). Another 707-eye study reported a test-set XGBoost mean absolute vault error of 121.70 micrometers, but 95% agreement limits remained approximately -307 to +257 micrometers (). Clinically important individual errors can therefore remain despite favorable average performance.
Potential applications include candidate size/power comparison, image-derived anatomical features, and risk alerts integrated into surgical planning (, , ). Current evidence is largely retrospective, with limited representation of hyperopic eyes, unusual anatomy, and underrepresented populations. Patient-level data splitting is necessary to avoid leakage between fellow eyes. Models require multicenter external validation, device-specific calibration, standardized timing and vault measurement, uncertainty reporting, and prospective evidence that their use reduces adverse outcomes. Human review remains essential when predictions conflict with imaging or manufacturer eligibility.
14 Discussion and future perspectives
Individualized sizing and power selection, refined haptic/optic designs, and AI-assisted planning should be evaluated against prospective refractive and safety endpoints, rather than predicted vault alone. Larger or presbyopia-oriented optics may improve particular visual needs, but their benefits, dysphotopsia, and anatomical effects require model-specific evaluation (, , ). Newer acrylic platforms and hyperopic designs need longer follow-up and adequately powered comparisons before safety equivalence can be assumed.
Registries should link baseline anatomy to vault trajectories, endothelial loss, cataract, IOP events, toric rotation, patient-reported symptoms, and explantation. Long-term endothelial reserve and subsequent cataract surgery are particularly important in young recipients (, , ). This narrative synthesis has selection and publication-bias limitations, does not formally grade evidence, and cannot validate the proposed framework. Numerical boundaries aid consistent reporting but must be interpreted with device labeling and individual anatomy.
15 Conclusion
ICL provides effective additive correction for appropriately selected phakic patients, especially those with high myopia and limited corneal options. Measurable eligibility, separate power and size planning, context-sensitive vault interpretation, and long-term monitoring define its safety boundaries. Hyperopic lenses and alternative pIOLs broaden the available options but have distinct anatomy, designs, and evidence gaps. AI may assist individualized planning, while external validation and prospective safety assessment remain necessary.
Statements
Author contributions
ZY: Investigation, Writing ā original draft. TQ: Conceptualization, Supervision, Writing ā review & editing.
Funding
The author(s) declared that financial support was not received for this work and/or its publication.
Acknowledgments
The authors thank colleagues involved in refractive surgery education and clinical discussion.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. During manuscript preparation, AI-assisted drafting and language polishing tools were used under author supervision; all scientific content, references, and final wording were reviewed and approved by the authors.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisherās note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fopht.2026.1908210/full#supplementary-material
References
1
ThompsonVCummingsABWangX. Implantable Collamer Lens procedure planning: a review of global approaches. Clin Ophthalmol. (2024) 18:1033ā43. doi:Ā 10.2147/OPTH.S456397
2
ReinsteinDZMacGregorCArcherTJGuptaRPotterJG. A review of posterior chamber phakic intraocular lenses. Curr Opin Ophthalmol. (2024) 35:138ā46. doi:Ā 10.1097/ICU.0000000000001018
3
Martinez-PlazaELopez-de la RosaALopez-MiguelAHolguerasAMaldonadoMJ. EVO/EVO+ Visian Implantable Collamer Lenses for the correction of myopia and myopia with astigmatism. Expert Rev Med Devices. (2023) 20:75ā83. doi:Ā 10.1080/17434440.2023.2174429
4
Montes-MicoRRuiz-MesaRRodriguez-PratsJLTana-RiveroP. Posterior-chamber phakic implantable collamer lenses with a central port: a review. Acta Ophthalmol. (2021) 99:e288ā301. doi:Ā 10.1111/aos.14599
5
AbingAAOhAOngLFMarvastiAHTranDBLeeJK. Surgical options and clinical outcomes for high myopia. Curr Opin Ophthalmol. (2024) 35:284ā91. doi:Ā 10.1097/ICU.0000000000001053
6
ZhangHGongRZhangXDengY. Analysis of perioperative problems related to intraocular Implantable Collamer Lens implantation. Int Ophthalmol. (2022) 42:3625ā41. doi:Ā 10.1007/s10792-022-02355-w
7
LarivoirNBRibeiroACSMorgadoCRSanthiagoMR. Efficacy, predictability, and safety of phakic Implantable Collamer Lenses V4c and V5: a systematic review and meta-analysis. J Refract Surg. (2025) 41:e1349ā59. doi:Ā 10.3928/1081597X-20251010-01
8
PassaroMLBuzziMQureshiRZouYChanVFMencucciRet al. Potential harms of posterior chamber phakic IOL: a systematic review and meta-analysis of complication incidence. Am J Ophthalmol. (2026) 285:73ā82. doi:Ā 10.1016/j.ajo.2026.01.028
9
STAAR Surgical Company. EVO/EVO+ Visian implantable collamer lens and EVO/EVO+ Visian toric implantable collamer lens. Directions for useUS Food and Drug Administration (2022). Available online at: https://www.accessdata.fda.gov/cdrh_docs/pdf3/P030016S035C.pdf (Accessed October 4, 2026).
10
US Food and Drug Administration. Premarket Approval P030016, Supplement S048: expansion of the EVO/EVO+ ICL/TICL indicated age range to 21-60 years. Decision date November 14, 2025 (2025). Available online at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?ID=P030016S048 (Accessed October 4, 2026).
11
YeYXianYLiuFZhangZNiuLShiWet al. Comparison of monovision surgery using ICL V4c or femtosecond laser LASIK for myopia correction in the presbyopia age patients. Sci Rep. (2025) 15:7629. doi:Ā 10.1038/s41598-025-88588-y
12
MaYYeYXianYNiuLShiWZhouXet al. The visual quality of patients with myopia and presbyopia after ICL V4c implantation and FS-LASIK using monovision design. Graefes Arch Clin Exp Ophthalmol. (2025) 263:3247ā56. doi:Ā 10.1007/s00417-025-06920-8
13
KocovaHVlkovaEMichalcovaLMotykaO. Implantation of posterior chamber phakic intraocular lens for myopia and hyperopia - long-term clinical outcomes. J Fr Ophtalmol. (2017) 40:215ā23. doi:Ā 10.1016/j.jfo.2016.10.009
14
CoskunsevenEKavadarliISahinOKayhanBPallikarisI. Refractive outcomes of 20 eyes undergoing ICL implantation for correction of hyperopic astigmatism. J Refract Surg. (2017) 33:604ā9. doi:Ā 10.3928/1081597X-20170504-06
15
LiDQGauvinMRochaGTrottierPRacineLWallersteinA. Spherical implantable collamer lens with postoperative adjunctive LASIK in the treatment of high compound hyperopic astigmatism. BMC Ophthalmol. (2025) 25:44. doi:Ā 10.1186/s12886-024-03828-5
16
Van HoeWTermoteKSaelensIDelbekeH. Implantable phakic contact lens: systematic review. J Cataract Refract Surg. (2025) 51:922ā32. doi:Ā 10.1097/j.jcrs.0000000000001686
17
RatebMGadAAMTohamyDElmohamadyMN. A prospective comparative study between implantable phakic intraocular contact lens and implantable collamer lens in treatment of myopia in adults. J Ophthalmol. (2022) 2022:9212253. doi:Ā 10.1155/2022/9212253
18
NaujokaitisTAuffarthGULabuzGKesslerLJKhoramniaR. Diagnostic techniques to increase the safety of phakic intraocular lenses. Diagnostics (Basel). (2023) 13:2503. doi:Ā 10.3390/diagnostics13152503
19
ParedesBMora de OnateJMartin SanchezDPineroDP. Characterization of vault with two models of posterior chamber phakic intraocular lenses. J Cataract Refract Surg. (2024) 50:1262ā9. doi:Ā 10.1097/j.jcrs.0000000000001536
20
NonpassoponMJongkhajornpongPPhimphoPCheewaruangrojNLekhanontKChuckpaiwongV. Agreement of implantable collamer lens sizes using parameters from different devices. BMJ Open Ophthalmol. (2022) 7:e000941. doi:Ā 10.1136/bmjophth-2021-000941
21
MoshirfarMSantosJMChaDSHerronMStoakesIMHoopesPC. Exploring nomograms for Implantable Collamer Lens size selection in myopia: a literature-based compilation. Clin Ophthalmol. (2023) 17:3307ā22. doi:Ā 10.2147/OPTH.S427815
22
ZhongXLiYLiYWangGDuYZhangM. Comparison of predictability in vault using NK formula and KS formula for the implantable collamer lens surgery. J Ophthalmol. (2024) 2024:4256371. doi:Ā 10.1155/2024/4256371
23
TanXLiuWChangYWuWYangLLiuJ. Analysis of inter-eye vault differences after Implantable Collamer Lens V4c implantation. Int J Gen Med. (2023) 16:2451ā9. doi:Ā 10.2147/IJGM.S384858
24
XiongLWuJDuHWangZ. Comparison of formulas in the implantable collamer lens vault prediction. Med (Baltimore). (2023) 102:e36542. doi:Ā 10.1097/MD.0000000000036542
25
TanWChenQYangRWangZZengQLeiXet al. Characteristics and factors associated with the position of the haptic after ICL V4c implantation. J Cataract Refract Surg. (2023) 49:416ā22. doi:Ā 10.1097/j.jcrs.0000000000001134
26
JiangYShenYChenXNiuLLiBChengMet al. Artificial intelligence-based refractive error prediction and EVO-implantable collamer lens power calculation for myopia correction. Eye Vis (Lond). (2023) 10:22. doi:Ā 10.1186/s40662-023-00338-1
27
WannapanichTKasetsuwanNReinprayoonU. Intraocular Implantable Collamer Lens with a central hole implantation: safety, efficacy, and patient outcomes. Clin Ophthalmol. (2023) 17:969ā80. doi:Ā 10.2147/OPTH.S379856
28
ZhuJChengDZhuXYLiFFYangYYeYF. Clinical prediction of insufficient vaults after implantable collamer lens implantation. Int J Ophthalmol. (2024) 17:1300ā6. doi:Ā 10.18240/ijo.2024.07.15
29
ChoiHRyuIHLeeISJungHLeeCEYooTKet al. Size exchange and vault response in posterior chamber phakic intraocular lens implantation: 12 years at a high-volume center. J Cataract Refract Surg. (2026) 52:476ā82. doi:Ā 10.1097/j.jcrs.0000000000001857
30
LinQYangDZhouX. Early outcomes of anterior segment parameters after implantable collamer lens V4c implantation. BMC Ophthalmol. (2022) 22:429. doi:Ā 10.1186/s12886-022-02656-9
31
WanQChenLHePWeiRMaKYinHet al. Five years comparation of efficacy and safety after ICL-V4c implantation for high and super high myopia correction. Ann Med. (2025) 57:2448282. doi:Ā 10.1080/07853890.2024.2448282
32
LuoWArumaALiMWangJXieJXiaoXet al. Four-year visual outcomes and optical quality of SMILE and implantable collamer lens V4c implantation for high myopia: a retrospective study. BMC Ophthalmol. (2023) 23:341. doi:Ā 10.1186/s12886-023-03050-9
33
ChenDZhaoXChouYLuoY. Comparison of visual outcomes and optical quality of SMILE and Visian Implantable Collamer Lens V4c implantation for moderate to high myopia: a meta-analysis. J Refract Surg. (2022) 38:332ā8. doi:Ā 10.3928/1081597X-20220411-01
34
Carreras-DiazHRenones de AbajoJCarreras-DiazMDRLorente-VelazquezA. Refractive outcomes after V4c Toric collamer lens implantation over 1 year of follow-up. Int J Ophthalmol. (2024) 17:1322ā30. doi:Ā 10.18240/ijo.2024.07.18
35
LeeYHanSBAuffarthGUSonHSKhoramniaRChoiCYet al. Vertical implantable collamer lens as a novel method to increase rotational stability. PloS One. (2024) 19:e0308830. doi:Ā 10.1371/journal.pone.0308830
36
KisielFBGurumurthyGJ. Endothelial cell loss post-implantable collamer lens V4c: meta-analysis. J Cataract Refract Surg. (2024) 50:420ā3. doi:Ā 10.1097/j.jcrs.0000000000001389
37
MoshirfarMMoinKAPandyaSKarimianFZauggBKhanSet al. Severe intraocular pressure rise after implantable collamer lens implantation. J Cataract Refract Surg. (2024) 50:985ā9. doi:Ā 10.1097/j.jcrs.0000000000001534
38
ChenQYanHChenLWangLZhengQRenY. Adverse events associated with Implantable Collamer Lens: insights from the FDA MAUDE database. Front Med (Lausanne). (2025) 12:1613060. doi:Ā 10.3389/fmed.2025.1613060
39
ChenQNiSYanHZhengQRenY. Real-world safety profile of EVO and EVO+ Implantable Collamer Lenses: a retrospective study based on FDA surveillance data. Ophthalmol Ther. (2025) 14:2753ā67. doi:Ā 10.1007/s40123-025-01222-y
40
Sarnat-KucharczykMHitnarowiczAZiembaAPojda-WilczekDMrukwa-KominekEPatelS. Advanced bilateral glaucoma decades after first-generation Fyodorov 'collar-button' anterior-posterior phakic intraocular lens implantation: a case report and clinical management considerations. Front Ophthalmol (Lausanne). (2026) 6:1677323. doi:Ā 10.3389/fopht.2026.1677323
41
LinZHuangLLiDYiSWengJ. Case report: maculopathy following standard dose intracameral cefuroxime injection during ICL surgery. Front Ophthalmol (Lausanne). (2026) 6:1778421. doi:Ā 10.3389/fopht.2026.1778421
42
FathimaTZTamilarasiSSachdevGDandapaniR. Concurrent phakic intraocular lens explantation and phacoemulsification with IOL implantation: a retrospective analysis of efficacy and safety. Indian J Ophthalmol. (2026) 74:S46ā51. doi:Ā 10.4103/IJO.IJO_1718_25
43
ZhangZNiuLLiuTShenYShangJZhaoJet al. Primary observations of EVO ICL implantation for high myopia with concave iris. Eye Vis (Lond). (2023) 10:18. doi:Ā 10.1186/s40662-023-00335-4
44
KamiyaKRyuIHYooTKKimJSLeeISKimJKet al. Prediction of phakic intraocular lens vault using machine learning of anterior segment optical coherence tomography metrics. Am J Ophthalmol. (2021) 226:90ā9. doi:Ā 10.1016/j.ajo.2021.02.006
45
DiYFangHLuoYLiYXuY. Predicting implantable collamer lens vault using machine learning based on various preoperative biometric factors. Transl Vis Sci Technol. (2024) 13:8. doi:Ā 10.1167/tvst.13.1.8
46
ZhuJLiFFLiGXJiangSYChengDBaoFJet al. Enhancing vault prediction and ICL sizing through advanced machine learning models. J Refract Surg. (2024) 40:e126ā32. doi:Ā 10.3928/1081597X-20240131-01
Summary
Keywords
artificial intelligence, EVO ICL, hyperopia, implantable collamer lens, IPCL, lens power calculation, phakic intraocular lens, vault
Citation
Yu Z and Qin T (2026) Implantable collamer lens surgery for high myopia: safety boundaries, vault management, and patient selection. Front. Ophthalmol. 6:1908210. doi: 10.3389/fopht.2026.1908210
Received
13 June 2026
Revised
04 October 2026
Accepted
05 October 2026
Published
27 October 2026
Volume
6 - 2026
Edited by
Ashok Sharma, Dr Ashok Sharmaās Cornea Centre, India
Reviewed by
Sonam Yangzes, Post Graduate Institute of Medical Education and Research (PGIMER), India
Neha Kumari, Sarvdrishti Eye Hospital, India
Updates
Copyright
Ā© 2026 Yu and Qin.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Tong Qin, qintong@sjtu.edu.cn
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.