<?xml version="1.0" encoding="utf-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article article-type="research-article" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xml:lang="EN">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Pediatr.</journal-id>
<journal-title>Frontiers in Pediatrics</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Pediatr.</abbrev-journal-title>
<issn pub-type="epub">2296-2360</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3389/fped.2022.1055212</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Pediatrics</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>Computational investigation of the haemodynamics shows criticalities of central venous lines used for chronic haemodialysis in children</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author"><name><surname>Bruno</surname><given-names>Claudia</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/2026154/overview"/></contrib>
<contrib contrib-type="author"><name><surname>Sauvage</surname><given-names>Emilie</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/527802/overview" /></contrib>
<contrib contrib-type="author"><name><surname>Simcock</surname><given-names>Ian</given-names></name>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref></contrib>
<contrib contrib-type="author"><name><surname>Redaelli</surname><given-names>Alberto</given-names></name>
<xref ref-type="aff" rid="aff4"><sup>4</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/394010/overview" /></contrib>
<contrib contrib-type="author"><name><surname>Schievano</surname><given-names>Silvia</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/68139/overview" /></contrib>
<contrib contrib-type="author"><name><surname>Shroff</surname><given-names>Rukshana</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/1238784/overview" /></contrib>
<contrib contrib-type="author" corresp="yes"><name><surname>Capelli</surname><given-names>Claudio</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref>
<xref ref-type="corresp" rid="cor1">&#x002A;</xref><uri xlink:href="https://loop.frontiersin.org/people/1244612/overview" /></contrib>
</contrib-group>
<aff id="aff1"><label><sup>1</sup></label><addr-line>Institute of Child Health</addr-line>, <institution>University College London</institution>, <addr-line>London</addr-line>, <country>United Kingdom</country></aff>
<aff id="aff2"><label><sup>2</sup></label><addr-line>Institute of Cardiovascular Science</addr-line>, <institution>University College London</institution>, <addr-line>London</addr-line>, <country>United Kingdom</country></aff>
<aff id="aff3"><label><sup>3</sup></label><addr-line>Great Ormond Street Hospital for Children</addr-line>, <institution>NHS Foundation Trust</institution>, <addr-line>London</addr-line>, <country>United Kingdom</country></aff>
<aff id="aff4"><label><sup>4</sup></label><addr-line>Department of Electronics, Information and Bioengineering</addr-line>, <institution>Politecnico di Milano</institution>, <addr-line>Milan</addr-line>, <country>Italy</country></aff>
<author-notes>
<fn fn-type="edited-by"><p><bold>Edited by:</bold> Enrico Vidal, University of Udine, Italy</p></fn>
<fn fn-type="edited-by"><p><bold>Reviewed by:</bold> Francesco Garzotto, University of Padua, Italy Rupesh Raina, Akron Children&#x0027;s Hospital, United States</p></fn>
<corresp id="cor1"><label>&#x002A;</label><bold>Correspondence:</bold> Claudio Capelli <email>c.capelli@ucl.ac.uk</email></corresp>
<fn fn-type="other" id="fn001"><p><bold>Specialty Section:</bold> This article was submitted to Pediatric Nephrology, a section of the journal Frontiers in Pediatrics</p></fn>
</author-notes>
<pub-date pub-type="epub"><day>31</day><month>10</month><year>2022</year></pub-date>
<pub-date pub-type="collection"><year>2022</year></pub-date>
<volume>10</volume><elocation-id>1055212</elocation-id>
<history>
<date date-type="received"><day>27</day><month>09</month><year>2022</year></date>
<date date-type="accepted"><day>10</day><month>10</month><year>2022</year></date>
</history>
<permissions>
<copyright-statement>&#x00A9; 2022 Bruno, Sauvage, Simcock, Redaelli, Schievano, Shroff and Capelli.</copyright-statement>
<copyright-year>2022</copyright-year><copyright-holder>BRUNO, Sauvage, Simcock, Redaelli, Schievano, Shroff and Capelli</copyright-holder><license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by/4.0/">
<p>This is an open-access article distributed under the terms of the <ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by/4.0/">Creative Commons Attribution License (CC BY)</ext-link>. The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p></license>
</permissions>
<abstract><sec>
<title>Background</title>
<p>Haemodialysis is a life-saving treatment for children with kidney failure. The majority of children have haemodialysis through central venous lines (CVLs). The use of CVLs in pediatric patients is often associated to complications which can lead to their replacement. The aim of this study is to investigate haemodynamics of pediatric CVLs to highlight the criticalities of different line designs.</p>
</sec>
<sec>
<title>Methods</title>
<p>Four models of CVLs for pediatric use were included in this study. The selected devices varied in terms of design and sizes (from 6.5&#x2005;Fr to 14&#x2005;Fr). Accurate 3D models of CVLs were reconstructed from high-resolution images including venous and arterial lumens, tips and side holes. Computational fluid dynamics (CFD) analyses were carried out to simulate pediatric working conditions of CVLs in ideal and anatomically relevant conditions.</p>
</sec>
<sec>
<title>Results</title>
<p>The arterial lumens of all tested CVLs showed the most critical conditions with the majority of blood flowing through the side-holes. A zone of low flow was identified at the lines&#x2019; tip. The highest shear stresses distribution (&#x003E;10&#x2005;Pa) was found in the 8&#x2005;Fr line while the highest platelet lysis index in the 10&#x2005;Fr model. The analysis on the anatomical geometry showed an increase in wall shear stress measured in the 10&#x2005;F model compared to the idealised configuration. Similarly, in anatomical models an increased disturbance and velocity of the flow was found inside the vein after line placement.</p>
</sec>
<sec>
<title>Conclusion</title>
<p>This study provided a numerical characterization of fluid dynamics in pediatric CVLs highlighting performance criticalities (i.e. high shear stresses and areas of stagnation) associated to specific sizes (8&#x2005;Fr and 10&#x2005;Fr) and conditions (i.e. anatomical test).</p>
</sec>
</abstract>
<kwd-group>
<kwd>computational fluid dynamics</kwd>
<kwd>paediatric catheters performance</kwd>
<kwd>dialysis</kwd>
<kwd>catheters design</kwd>
<kwd>bioengineering</kwd>
<kwd>central venous line (CVL)</kwd>
<kwd>central venous access</kwd>
</kwd-group><contract-num rid="cn001">Paed_IN_004_20190926</contract-num><contract-num rid="cn002">&#x00A0;</contract-num><contract-sponsor id="cn001">a Paediatric Innovation</contract-sponsor><contract-sponsor id="cn002">Kidney Research UK<named-content content-type="fundref-id">10.13039/501100000291</named-content></contract-sponsor><counts>
<fig-count count="7"/>
<table-count count="3"/><equation-count count="7"/><ref-count count="20"/><page-count count="0"/><word-count count="0"/></counts>
</article-meta>
</front>
<body>
<sec id="s1" sec-type="intro">
<title>Introduction</title>
<p>Paediatric kidney disease is a severe clinical condition affecting around 12&#x2013;20 cases per million age-related population (<xref ref-type="bibr" rid="B1">1</xref>) and requiring dialysis or kidney transplantation (<xref ref-type="bibr" rid="B2">2</xref>, <xref ref-type="bibr" rid="B3">3</xref>). Dialysis can be performed in the form of peritoneal or haemodialysis (HD), with HD performed as the initial treatment modality in about half of the patients younger than 20 years of age (<xref ref-type="bibr" rid="B1">1</xref>). HD is technically feasible in children of all ages, even in small neonates, but vascular access remains the &#x2018;Achille&#x2019;s heel&#x2019; of HD, accounting for significant morbidity and even mortality (<xref ref-type="bibr" rid="B4">4</xref>). Most of the children requiring the dialysis treatment are dialysed using a central venous line (CVL). According to the International Pediatric Hemodialysis Network (IPHN) Registry, 73&#x0025; of HD patients treated in paediatric units around the globe received a CVL as first dialysis access between December 2012 and September 2017. However, almost half of these lines required to be replaced (<xref ref-type="bibr" rid="B5">5</xref>). CVL complications include thrombosis and fibrin sheath obstruction with younger patients presenting the highest CVL malfunctioning and failure rate (<xref ref-type="bibr" rid="B6">6</xref>, <xref ref-type="bibr" rid="B7">7</xref>). Additionally, sluggish blood flow in CVLs leads to poor dialysis adequacy and deposition of biofilm, predisposing to line infections. Repeated CVL changes cause stenosis of the central veins, and this can permanently damage blood vessels. Investigating the performance of CVLs <italic>in vivo</italic> could be crucial to understand the origins of such problems but is technically challenging and most of the haemodynamics parameters remain not accessible.</p>
<p>Computational fluid dynamics (CFD) simulations have been widely used in the design and testing of medical devices. The study of CVL by means of CFD is not an exception to this. The efficacy of the exchange of blood between device and patient has been matter of several numerical investigations (<xref ref-type="bibr" rid="B8">8</xref>&#x2013;<xref ref-type="bibr" rid="B10">10</xref>). Previous studies showed how different tip designs can greatly affect the CVL performance in terms of thrombogenicity and recirculation (<xref ref-type="bibr" rid="B9">9</xref>). The presence of side-holes, placed in proximity of the tip to increase the efficiency of the blood exchange, can also promote blood clot formation and infection, impacting on the longevity of the CVL <italic>in vivo</italic> (<xref ref-type="bibr" rid="B11">11</xref>, <xref ref-type="bibr" rid="B12">12</xref>). On the patient side, modelling CVLs function in both simplified working conditions (<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B9">9</xref>, <xref ref-type="bibr" rid="B11">11</xref>, <xref ref-type="bibr" rid="B13">13</xref>) and replicated anatomical characteristics (<xref ref-type="bibr" rid="B10">10</xref>, <xref ref-type="bibr" rid="B14">14</xref>) showed the influence of the catheter insertion within the central veins (<xref ref-type="bibr" rid="B15">15</xref>, <xref ref-type="bibr" rid="B16">16</xref>). Considering the literature consulted, computational studies of CVLs for paediatric use have not been reported in the literature (source: Pubmed) and the reasons for such poor performance are still widely unknown.</p>
<p>The aim of this study was to identify potential fluid-dynamics criticalities which can explain the frequent dysfunctions of pediatric CVLs.</p>
</sec>
<sec id="s2" sec-type="methods">
<title>Methods</title>
<p>A computational framework to simulate the fluid dynamics of pediatric CVLs currently in use at our Centre was set up to compute haemodynamic performance under varying conditions.</p>
<p>Four models of CVLs (MedCOMP - Harleysville, PA, USA) of varying design and size were included in this study, namely:
<list list-type="simple">
<list-item><label>&#x2022;</label><p>Tesio 6.5&#x2005;F (<xref ref-type="fig" rid="F1">Figure&#x00A0;1A</xref>): round single lumen with cut straight tip; six side-holes arranged in spiral configuration at the tip of each line; indicated for infants up to 6&#x2005;kg.</p></list-item>
<list-item><label>&#x2022;</label><p>Hemo-Cath 8&#x2005;F (<xref ref-type="fig" rid="F1">Figure&#x00A0;1B</xref>): double lumen CVL, indicated for children between 6 and 12&#x2005;kg with step tip and circle-C profile:
<list list-type="simple">
<list-item><label>&#x2022;</label><p>venous tip (O profile) included two opposite side-holes</p></list-item>
<list-item><label>&#x2022;</label><p>arterial tip (C profile) included a parallel dual-hole configuration.</p></list-item>
</list></p></list-item>
<list-item><label>&#x2022;</label><p>Pediatric Split Cath 10&#x2005;F (<xref ref-type="fig" rid="F1">Figure&#x00A0;1C</xref>): double-D lumen configuration ending in two separate but identical distal tips, characterized by a step-down in diameter; six side-holes in each lumen divided in two sets of three; recommended for paediatric patients between 15 and 30&#x2005;kg.</p></list-item>
<list-item><label>&#x2022;</label><p>Split Cath III 14&#x2005;F (<xref ref-type="fig" rid="F1">Figure&#x00A0;1D</xref>): double-D lumen configuration ending in two different and separate distal tips: venous tip with a circular lumen and five side-holes and arterial tip with an oval profile and six side-holes. This CVL is indicated for children over 30&#x2005;kg.</p></list-item>
</list></p>
<fig id="F1" position="float"><label>Figure 1</label>
<caption><p>Photographs and CAD models of the CVLs included in this study with magnification on the details of the side-holes: (<bold>A</bold>) Tesio (6.5&#x2005;F); (<bold>B</bold>) Hemo-Cath (8&#x2005;F); (<bold>C</bold>) Pediatric Split Cath (10&#x2005;F); and (<bold>D</bold>) Split Cath III (14&#x2005;F).</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g001.tif"/>
</fig>
<sec id="s2a">
<title>High-resolution imaging of pediatric CVLs</title>
<p>The geometries of the pediatric CVLs were reconstructed from high-resolution imaging. The tips of the devices (ranging from 30 to 50&#x2005;mm in length) were scanned using microfocus computed tomography (micro-CT) (Med-X Alpha; Nikon Metrology, Tring, UK). During the scanning procedure, the tip was attached to a carbon fibre plate and secured in place with parafilm B (Bemis, Neenah, WI, USA) to ensure stability during image acquisition. Images were obtained with optimised factors of an x-ray energy of 100&#x2005;kVp, current of 100&#x2005;&#x00B5;A, exposure time of 500&#x2005;ms, with an optimised number of projections for each examination and 4 frames per projection using a Tungsten target. Voxel sizes ranged between 15.29 and 26.22&#x2005;&#x00B5;m, with each acquisition taking approximately 40&#x2005;min.</p>
</sec>
<sec id="s2b">
<title>CAD design of CVLs</title>
<p>Following calibration, the images acquired with micro-CT (TIFF format) were post-processed to derive geometrical measurements using the commercially available segmentation software Simpleware ScanIP (Synopsys Inc., Mountain View, CA, USA). Geometrical information and dimensions were used to build CAD models of the CVLs by means of SolidWorks 2018 (Dassault Syst&#x00E8;mes, V&#x00E9;lizy-Villacoublay, France). Models resembling the four pediatric CVLs included in this study are shown in <xref ref-type="fig" rid="F1">Figure&#x00A0;1</xref>.</p>
</sec>
<sec id="s2c">
<title>CFD analysis of CVLs</title>
<p>The CFD models were created in Ansys DesignModeler (Ansys Inc., Canonsburg, PA, USA). CFD analyses of CVLs were carried out on two main configurations: 1. an idealised geometry of superior vena cava (SVC) was used to isolate the CVLs performance; 2. an anatomical representation of SVC allowed to investigate the interaction CVL-vein and to evaluate possible differences in the CVL performance. More details of the CFD configurations are here presented.
<list>
<list-item><label>1.</label><p>Idealised model: CVLs were coaxially placed inside a straight rigid cylinder with diameter comparable to the SVC of the different patient groups: 6&#x2005;mm, 8.5&#x2005;mm, 15&#x2005;mm and 18&#x2005;mm (<xref ref-type="fig" rid="F2">Figure&#x00A0;2A</xref>). The length of each SVC-CVL model was chosen to ensure a fully developed velocity profile.</p></list-item>
<list-item><label>2.</label><p>Anatomical model: CVLs were simulated inside anatomical models of the right atrium (RA), SVC and inferior vena cava (IVC). Anatomical models resembled the average characteristics of four subjects of different sizes but were not directly derived from patients. The dimensions of the anatomical models were adapted to represent paediatric subjects of different weight (<xref ref-type="bibr" rid="B17">17</xref>, <xref ref-type="bibr" rid="B18">18</xref>). Following clinical guidelines, CVL models were placed inside the SVC, with the tip laying in the RA (<xref ref-type="fig" rid="F2">Figure&#x00A0;2B</xref>). CFD simulations were carried out in anatomical models with and without the CVL to explore the changes in flow in the presence of the CVL.</p></list-item>
</list></p>
<fig id="F2" position="float"><label>Figure 2</label>
<caption><p>(<bold>A</bold>) Simplified cylindrical model of the SVC with the Hemo-Cath catheter coaxially placed inside; schematic of the flow directions is shown in the ideal model with a close-up at the inlet of the model. V1 and V2 are the venous inlet and outlet, A1 and A2 are the arterial inlet and outlet, T1 and T2 are the SVC inlet and outlet. (<bold>B</bold>) Anatomical model of RA, SVC and IVC after CVL placement. On the right-hand side, a close-up of the mesh is shown. SVC&#x2009;&#x003D;&#x2009;Superior vena cava; RA&#x2009;&#x003D;&#x2009;Right atrium; IVC&#x2009;&#x003D;&#x2009;Inferior vena cava.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g002.tif"/>
</fig>
<p>All the models were meshed in Ansys Meshing module and Ansys Fluent Release 19.0 using polyhedral elements. Local refinement of the mesh was applied to the proximity of the CVL and at the wall of the anatomical models (<xref ref-type="fig" rid="F2">Figure&#x00A0;2B</xref>). A sensitivity analysis was carried out for each model to ensure mesh independent results in terms of velocity, pressure and wall shear stress. The final numbers of cells ranged between 1&#x2005;million and 10&#x2005;million. More details are provided in <xref ref-type="sec" rid="s10">Supplementary Appendix A1</xref>.</p>
<p>A flat velocity profile at the inlet of the SVC (T1 in the ideal model, <xref ref-type="fig" rid="F2">Figure&#x00A0;2A</xref>) and of the IVC in the anatomical model to match physiological flow rates for each patient group (<xref ref-type="bibr" rid="B18">18</xref>). At the venous and arterial lumens of each CVL (respectively V1 and A2 in <xref ref-type="fig" rid="F2">Figure&#x00A0;2A</xref>), two identical but opposite velocity profiles were applied. Each CVL was tested with three velocity conditions closely mimicking the continuous blood flow rates used in routine clinical practice (<xref ref-type="table" rid="T1">Table&#x00A0;1</xref>) to investigate their performance at different working condition. A reference pressure equal to zero was set at the outlet of the SVC cylinder in the idealised geometry (T2, <xref ref-type="fig" rid="F2">Figure&#x00A0;2A</xref>), and at the level of the tricuspid valve in the anatomical one. Walls were assumed to be rigid, and a no-slip condition was applied with steady state simulations carried out under the assumption of laminar flow.</p>
<table-wrap id="T1" position="float"><label>Table 1</label>
<caption><p>Values of flow-rate prescribed in each CVL - SVC model.</p></caption>
<table frame="hsides" rules="groups">
<colgroup>
<col align="left"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left">CVL model</th>
<th valign="top" align="center" colspan="3">CVL flow rates [ml/min] (for both idealised and anatomical models)</th>
<th valign="top" align="center">SVC flow rate [ml/min] (for both idealised and anatomical models)</th>
<th valign="top" align="center">IVC flow rate [ml/min] (for anatomical models only)</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Tesio (6.5&#x2005;F)</td>
<td valign="top" align="center">20</td>
<td valign="top" align="center">30</td>
<td valign="top" align="center">40</td>
<td valign="top" align="center">810</td>
<td valign="top" align="center">540</td>
</tr>
<tr>
<td valign="top" align="left">Hemo-Cath (8&#x2005;F)</td>
<td valign="top" align="center">50</td>
<td valign="top" align="center">60</td>
<td valign="top" align="center">70</td>
<td valign="top" align="center">1,450</td>
<td valign="top" align="center">967</td>
</tr>
<tr>
<td valign="top" align="left">Pediatric Split Cath (10&#x2005;F)</td>
<td valign="top" align="center">140</td>
<td valign="top" align="center">150</td>
<td valign="top" align="center">180</td>
<td valign="top" align="center">2,700</td>
<td valign="top" align="center">1,800</td>
</tr>
<tr>
<td valign="top" align="left">Split Cath III (14&#x2005;F)</td>
<td valign="top" align="center">220</td>
<td valign="top" align="center">250</td>
<td valign="top" align="center">280</td>
<td valign="top" align="center">3,000</td>
<td valign="top" align="center">2,000</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>The blood was modelled as an incompressible Newtonian fluid with a constant viscosity &#x00B5; of 0.0035&#x2005;Pa&#x00B7;s and density &#x03C1; of 1060&#x2005;kg/m<sup>3</sup>. The assumption of Newtonian blood was considered reasonable in light of preliminary simulations showing a constant viscosity throughout the volume of interest. For each CVL, a set of simulations with varying input parameters were performed in Ansys Workbench Release 19.0. Navier-Stokes equations were resolved within ANSYS Fluent using a pressure-based solver with SIMPLE scheme for pressure velocity coupling, and Least Squares Cell Based, Second Order and Second Order Upwind for gradient, pressure and momentum, respectively. Simulations were considered completed when the scaled residual sum of continuity and velocity decreased by more than 10&#x2013;5. Simulations ran in a workstation with Intel&#x00AE; Xeon&#x00AE; W-2275 CPU (3.30&#x2005;GHz) processor and 64.0 GB of RAM.</p>
<p>For each simulation, velocity fields were assessed inside the CVL lumen, at the tip of the CVL and in the region of side-holes. In order to identify the presence of asymmetric and skewed jets through the openings of the CVL and compare the different designs, the ratio between the average and the maximum velocity at the CVL tip and side-holes were computed for both lumens of each CVL model. High velocity ratio indicates the presence of a flat velocity profile as described by Cavazzuti et al. in 2020 (<xref ref-type="bibr" rid="B19">19</xref>), whilst values close to zero are representative of asymmetric and skewed jets. Reynold numbers were computed at each opening to validate the assumption of laminar flow. Split flow proportions were computed dividing the flow rate through each CVL opening by the total flow rate inside the lumen. Blood residence time (RT) was evaluated in both tips of the CVLs, specifically inside the side-holes. A pathlines-based approach was used. Pathlines were coloured by particle properties, specifically time, and plotted in all the areas of interest. The maximum value of blood residence time was exported for every side-hole. This step was conducted solely on simulations run at intermediate flow rate. In accordance to previous CFD studies (<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B11">11</xref>), shear stress levels&#x2014;levels of the frictional (tangential) force of the flowing blood - were evaluated on the volume of arterial tip of the CVL where blood is withdrawn. Here, shear stress values were computed according to:<disp-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="UDM1"><mml:mover><mml:mover><mml:mi>&#x03C4;</mml:mi><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo>=</mml:mo><mml:mrow><mml:mi>&#x03BC;</mml:mi></mml:mrow><mml:mrow><mml:mover><mml:mi>&#x03B3;</mml:mi><mml:mo>&#x02D9;</mml:mo></mml:mover></mml:mrow></mml:math></disp-formula>where &#x00B5; is the viscosity of blood and <inline-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="IM1"><mml:mrow><mml:mover><mml:mi>&#x03B3;</mml:mi><mml:mo>&#x02D9;</mml:mo></mml:mover></mml:mrow></mml:math></inline-formula> is the strain rate, <inline-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="IM2"><mml:mrow><mml:mover><mml:mi>&#x03B3;</mml:mi><mml:mo>&#x02D9;</mml:mo></mml:mover></mml:mrow><mml:mo>=</mml:mo><mml:msqrt><mml:mrow><mml:mfrac><mml:mn>1</mml:mn><mml:mn>2</mml:mn></mml:mfrac></mml:mrow><mml:mover><mml:mover><mml:mi>D</mml:mi><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo>&#x003A;</mml:mo><mml:mover><mml:mover><mml:mi>D</mml:mi><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover></mml:msqrt></mml:math></inline-formula>, with <inline-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="IM3"><mml:mover><mml:mover><mml:mi>D</mml:mi><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover></mml:math></inline-formula> defined as <inline-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="IM4"><mml:mover><mml:mover><mml:mi>D</mml:mi><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo accent="false">&#x00AF;</mml:mo></mml:mover><mml:mo>=</mml:mo><mml:mspace width="thickmathspace" /><mml:mrow><mml:mo>(</mml:mo><mml:mrow><mml:mrow><mml:mfrac><mml:mrow><mml:mi mathvariant="normal">&#x2202;</mml:mi><mml:msub><mml:mi>u</mml:mi><mml:mi>j</mml:mi></mml:msub></mml:mrow><mml:mrow><mml:mi mathvariant="normal">&#x2202;</mml:mi><mml:msub><mml:mi>x</mml:mi><mml:mi>i</mml:mi></mml:msub></mml:mrow></mml:mfrac></mml:mrow><mml:mo>+</mml:mo><mml:mspace width="thickmathspace" /><mml:mrow><mml:mfrac><mml:mrow><mml:mi mathvariant="normal">&#x2202;</mml:mi><mml:msub><mml:mi>u</mml:mi><mml:mi>i</mml:mi></mml:msub></mml:mrow><mml:mrow><mml:mi mathvariant="normal">&#x2202;</mml:mi><mml:msub><mml:mi>x</mml:mi><mml:mi>j</mml:mi></mml:msub></mml:mrow></mml:mfrac></mml:mrow></mml:mrow><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> (<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B20">20</xref>).</p>
<p>To compare different designs, average values of shear stress were assessed together with the percentage of elements subjected to a level of stress higher than 10&#x2005;Pa according to Mareels et al. (<xref ref-type="bibr" rid="B8">8</xref>).</p>
<p>Values of Platelet Lysis Index (PLI), indicator of platelet activation, were computed as it follows, using a pathlines-based method for each model (<xref ref-type="bibr" rid="B8">8</xref>):<disp-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="UDM2"><mml:mrow><mml:mi mathvariant="normal">PLI</mml:mi></mml:mrow><mml:mo>=</mml:mo><mml:mn>3.66</mml:mn><mml:mo>&#x22C5;</mml:mo><mml:mspace width="thickmathspace" /><mml:msup><mml:mn>10</mml:mn><mml:mrow><mml:mrow><mml:mo>&#x2212;</mml:mo></mml:mrow><mml:mn>6</mml:mn></mml:mrow></mml:msup><mml:mspace width="thickmathspace" /><mml:msup><mml:mi>t</mml:mi><mml:mrow><mml:mn>0.77</mml:mn></mml:mrow></mml:msup><mml:mspace width="thickmathspace" /><mml:msup><mml:mi>&#x03C4;</mml:mi><mml:mrow><mml:mn>3.075</mml:mn></mml:mrow></mml:msup><mml:mspace width="thickmathspace" /></mml:math></disp-formula>where <italic>t</italic> is the exposure time and <inline-formula><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="IM5"><mml:mi>&#x03C4;</mml:mi></mml:math></inline-formula> the magnitude of shear stress.</p>
<p>For anatomical models, maximum velocities were measured in planes along the SVC, with and without CVL. On the same planes presence of secondary flows was investigated. Lastly, wall shear stress (WSS) was considered prior and after CVL insertion. A schematic of the evaluated parameters is provided in <xref ref-type="sec" rid="s10">Supplementary Appendix A2</xref>.</p>
</sec>
</sec>
<sec id="s3" sec-type="results">
<title>Results</title>
<p>A total of 34 simulations were successfully run in this study, 15 for the idealised configuration and 19 for the anatomical one. The parameters of interests are presented first separately for each configuration and then in a comparative way.</p>
<sec id="s3a">
<title>Idealised geometry</title>
<p>The velocity profiles of the venous and arterial lumens are shown in <xref ref-type="fig" rid="F3">Figure&#x00A0;3</xref> where regions of low velocity are visualised within all arterial lumens of the CVLs.</p>
<fig id="F3" position="float"><label>Figure 3</label>
<caption><p>Velocity maps plotted for intermediate flow rates on the vertical section of the venous and the arterial lumen for the CVLs: (<bold>A</bold>) Tesio; (<bold>B</bold>) Hemo-Cath; (<bold>C</bold>) Pediatric Split Cath; (<bold>D</bold>) Split Cath III. White arrows indicate regions of low velocity inside the arterial lumen. CVL&#x2009;&#x003D;&#x2009;Central venous line.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g003.tif"/>
</fig>
<p>The average to maximum velocity ratio measured in the side-holes were consistently lower compared to the tip in all models, and for both the venous and the arterial configuration. The outlets of the arterial configurations showed a skewed jet-like behaviour. The venous tips were characterized by a flatter velocity profile, with values ranging from 0.398 to 0.591 for every model, except for the Pediatric Split Cath. This same trend was observed in the arterial tip of the CVLs (<xref ref-type="sec" rid="s10">Supplementary Appendix A3</xref>).</p>
<p>In all models, Reynolds numbers were below the threshold of 2,000, confirming the laminar nature of the flow analysed in all different configurations (<xref ref-type="sec" rid="s10">Supplementary Appendix A4</xref>).</p>
<p>Split flow proportion between the distal tip and the side-holes are shown in <xref ref-type="fig" rid="F4">Figure&#x00A0;4</xref>. In all models, the flow conditions vary between the venous and arterial lumens. The tip of the CVL plays a significant role when the blood is carried to the body (i.e., venous lumen). In the case of Tesio and Split Cath III, the flow repartition was almost equal between the tip and the side-holes. The Hemo-Cath was characterized by the highest percentage of blood through the tip for the 3 tested conditions (89.4&#x0025;, 90.7&#x0025; and 91.7&#x0025;), whilst the Pediatric Split Cath showed the opposite behaviour with the highest side-holes flow rate (77.2&#x0025;, 76.4&#x0025; and 74.5&#x0025;). The increase in the velocity imposed at the inlet reflected in an increase in the flow rate flowing through the tip for each CVL except for the Tesio (Tesio: 48.5&#x0025;, 42.8&#x0025; and 43.3&#x0025;; Hemo-Cath: 89.4&#x0025;, 90.7&#x0025; and 91.7&#x0025;; Pediatric Split Cath: 23.1&#x0025;, 23.8&#x0025; and 25.7&#x0025;; Split Cath III: 48.8&#x0025;, 50.3&#x0025; and 51.6&#x0025;). In the arterial configuration, the side-holes played a predominant role accounting for more than 90&#x0025; of the overall flow rate in all models. In the case of Tesio, a flow loss through the tip was recorded in arterial configuration. This was compensated by an increase of flow through the side-holes. The percentage of flow loss seemed to reduce with the 3 tested conditions: 29.8&#x0025;, 18.6&#x0025; and 13.1&#x0025;.</p>
<fig id="F4" position="float"><label>Figure 4</label>
<caption><p>Split flow proportion between the tip and the side-holes for both lumens (venous and arterial). CVL&#x2009;&#x003D;&#x2009;Central venous line; FR&#x2009;&#x003D;&#x2009;Flow rate.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g004.tif"/>
</fig>
<p>Values of residence times are reported in <xref ref-type="table" rid="T2">Table&#x00A0;2</xref>. The highest values of RT were found in the Tesio model, consistent with large areas of stagnation&#x2014;an area with zero or nearly zero velocity&#x2014;present inside the lumen. For this model, all side-holes except one showed a residence time above 0.030&#x2005;s. Low values of residence time were recorded in the Hemo-Cath and in the Pediatric Split Cath models.</p>
<table-wrap id="T2" position="float"><label>Table 2</label>
<caption><p>Rt values recorded at the side-holes of the CVLs.</p></caption>
<table frame="hsides" rules="groups">
<colgroup>
<col align="left"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left" rowspan="3">Model</th>
<th valign="top" align="center" colspan="4">Residence Time (s)</th>
</tr>
<tr>
<th valign="top" align="center" colspan="2">Venous lumen</th>
<th valign="top" align="center" colspan="2">Arterial lumen</th>
</tr>
<tr>
<th valign="top" align="center">Highest</th>
<th valign="top" align="center">Lowest</th>
<th valign="top" align="center">Highest</th>
<th valign="top" align="center">Lowest</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Tesio (6.5&#x2005;F)</td>
<td valign="top" align="center">0.241</td>
<td valign="top" align="center">0.092</td>
<td valign="top" align="center">0.241</td>
<td valign="top" align="center">0.024</td>
</tr>
<tr>
<td valign="top" align="left">Hemo-Cath (8&#x2005;F)</td>
<td valign="top" align="center">0.039</td>
<td valign="top" align="center">0.033</td>
<td valign="top" align="center">0.005</td>
<td valign="top" align="center">0.004</td>
</tr>
<tr>
<td valign="top" align="left">Pediatric Split Cath (10&#x2005;F)</td>
<td valign="top" align="center">0.022</td>
<td valign="top" align="center">0.006</td>
<td valign="top" align="center">0.032</td>
<td valign="top" align="center">0.001</td>
</tr>
<tr>
<td valign="top" align="left">Split Cath III (14&#x2005;F)</td>
<td valign="top" align="center">0.090</td>
<td valign="top" align="center">0.007</td>
<td valign="top" align="center">0.077</td>
<td valign="top" align="center">0.004</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>Values of average shear stress, percentage of shear stress above 10&#x2005;Pa and PLI are reported in <xref ref-type="table" rid="T3">Table&#x00A0;3</xref>. The Hemo-Cath was associated with the highest percentage of shear stress above the threshold, whilst the Tesio with the lowest percentage. Overall, for each CVL shear stress values increased with the imposed inlet flow rate.</p>
<table-wrap id="T3" position="float"><label>Table 3</label>
<caption><p>Shear stress levels and PLI values for all the four CVL models at each flow rate.</p></caption>
<table frame="hsides" rules="groups">
<colgroup>
<col align="left"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left" rowspan="3">Model</th>
<th valign="top" align="center" colspan="9">Shear stress levels</th>
</tr>
<tr>
<th valign="top" align="center" colspan="3">Minimum flow rate</th>
<th valign="top" align="center" colspan="3">Intermediate flow rate</th>
<th valign="top" align="center" colspan="3">Maximum flow rate</th>
</tr>
<tr>
<th valign="top" align="center">Average SS (Pa)</th>
<th valign="top" align="center">&#x0025; SS&#x2009;&#x003E;&#x2009;10&#x2005;Pa</th>
<th valign="top" align="center">PLI</th>
<th valign="top" align="center">Average SS (Pa)</th>
<th valign="top" align="center">&#x0025; SS&#x2009;&#x003E;&#x2009;10&#x2005;Pa</th>
<th valign="top" align="center">PLI</th>
<th valign="top" align="center">Average SS (Pa)</th>
<th valign="top" align="center">&#x0025; SS&#x2009;&#x003E;&#x2009;10&#x2005;Pa</th>
<th valign="top" align="center">PLI</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Tesio (6.5&#x2005;F)</td>
<td valign="top" align="center">4.73</td>
<td valign="top" align="center">6.24</td>
<td valign="top" align="center">0.01</td>
<td valign="top" align="center">4.87</td>
<td valign="top" align="center">8.80</td>
<td valign="top" align="center">0.01</td>
<td valign="top" align="center">5.03</td>
<td valign="top" align="center">10.65</td>
<td valign="top" align="center">0.07</td>
</tr>
<tr>
<td valign="top" align="left">Hemo-Cath (8&#x2005;F)</td>
<td valign="top" align="center">4.32</td>
<td valign="top" align="center">27.56</td>
<td valign="top" align="center">0.11</td>
<td valign="top" align="center">4.75</td>
<td valign="top" align="center">30.04</td>
<td valign="top" align="center">0.14</td>
<td valign="top" align="center">5.20</td>
<td valign="top" align="center">31.30</td>
<td valign="top" align="center">0.17</td>
</tr>
<tr>
<td valign="top" align="left">Pediatric Split Cath (10&#x2005;F)</td>
<td valign="top" align="center">4.20</td>
<td valign="top" align="center">23.75</td>
<td valign="top" align="center">0.68</td>
<td valign="top" align="center">4.45</td>
<td valign="top" align="center">24.49</td>
<td valign="top" align="center">0.85</td>
<td valign="top" align="center">5.23</td>
<td valign="top" align="center">26.46</td>
<td valign="top" align="center">1.34</td>
</tr>
<tr>
<td valign="top" align="left">Split Cath III (14&#x2005;F)</td>
<td valign="top" align="center">3.63</td>
<td valign="top" align="center">22.81</td>
<td valign="top" align="center">0.42</td>
<td valign="top" align="center">4.09</td>
<td valign="top" align="center">24.44</td>
<td valign="top" align="center">0.66</td>
<td valign="top" align="center">4.55</td>
<td valign="top" align="center">26.11</td>
<td valign="top" align="center">0.86</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<fn id="table-fn1"><p>Values refer to the arterial lumen.</p></fn>
<fn id="table-fn2"><p>SS, Shear stress; PLI, Platelet lysis index.</p></fn>
</table-wrap-foot>
</table-wrap>
<p>Peaks of shear stress were consistently recorded close to the proximal side-holes (<xref ref-type="fig" rid="F5">Figure&#x00A0;5</xref>) in the arterial configuration. The Tesio model displayed the lowest values of PLI; the Pediatric Split Cath displayed the highest PLI.</p>
<fig id="F5" position="float"><label>Figure 5</label>
<caption><p>Shear stress distribution plotted for the four CVLs: the Tesio (<bold>A</bold>), the Hemo-Cath (<bold>B</bold>), the Pediatric Split Cath (<bold>C</bold>) and the Split Cath III (<bold>D</bold>). Contours refer to simulations run at intermediate CVL flow rates.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g005.tif"/>
</fig>
</sec>
<sec id="s3b">
<title>Anatomical geometries</title>
<p>When CVLs were placed in the anatomical SVC, the cross section of the vessel was reduced, causing an expected increase in the blood velocity within the SVC, as shown in <xref ref-type="fig" rid="F6">Figure&#x00A0;6</xref>. Areas of high velocity non-symmetrically distributed were noted around the CVL. The highest increases of velocity were recorded in the Tesio and in the Hemo-Cath models: 34&#x0025; and 28&#x0025; difference in the maximum velocity respectively, and 15&#x0025; and 13&#x0025; in the average velocity, respectively. With the smaller CVLs (Tesio and Hemo-Cath), a pair of small asymmetric eddies appeared in the SVC close to the wall of the device (<xref ref-type="fig" rid="F6">Figure&#x00A0;6B</xref>). Large eddies were found in the SVC cross-section.</p>
<fig id="F6" position="float"><label>Figure 6</label>
<caption><p>Velocity field and vectors of the anatomical model with Tesio: (<bold>A</bold>) Velocity maps plotted at three different planes along the SVC without CVL; (<bold>B</bold>) Velocity maps and in-plane velocity vectors are plotted on planes after CVL insertion. CVL&#x2009;&#x003D;&#x2009;Central venous line.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g006.tif"/>
</fig>
<p>After insertion of the CVL, WSS increased in all four models along the SVC wall, where the presence of the CVL altered the normal distribution. The highest WSSs were found in the smallest anatomical model. The Pediatric Split Cath recorded the highest increase in WSS, going from an initial value of 0.34&#x2005;Pa to a final one of 1.48&#x2005;Pa after CVL insertion.</p>
<p>Regarding the CVL performance, small differences were found in the levels of shear stress above the 10&#x2005;Pa threshold for the bigger models (Pediatric Split Cath and Split Cath III), as shown in <xref ref-type="fig" rid="F7">Figure&#x00A0;7</xref>. On the contrary, smaller CVL models recorded a higher percentage of shear stress above 10&#x2005;Pa in the anatomical model. This difference is more evident in the Tesio 6.5&#x2005;F CVL in which values were more than doubled at low flow rates.</p>
<fig id="F7" position="float"><label>Figure 7</label>
<caption><p>Comparison of the shear stress levels between the ideal and the anatomical model. For each CVL model, the average of the three flow rates is reported. CVL&#x2009;&#x003D;&#x2009;Central venous line.</p></caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fped-10-1055212-g007.tif"/>
</fig>
</sec>
</sec>
<sec id="s4" sec-type="discussion">
<title>Discussion</title>
<p>Children with kidney failure have a lifetime of kidney replacement therapies ahead of them, and their vascular access that is required for HD is literally their lifeline. CVL designs that allow for optimal dialysis therapy with the least complications will help preserve their vessels. This study investigated the fluid dynamics of CVLs used in paediatric patients on HD <italic>via</italic> computational simulations that allowed testing of the CVL geometries under a wide range of haemodynamics conditions. This method provides systematic information that is not obtainable <italic>via</italic> in-vivo and in-vitro approaches. Several criticalities were identified for all CVL designs: areas of stagnation, unbalanced distribution of flow between side-holes and tips, and high shear stresses which in turn can create an environment predisposing to cell lysis. In addition, analysis of the anatomical models quantified how the presence of the CVL in the SVC alters its physiological state by creating eddies and increased wall shear stresses. While it is impossible to avoid such perturbation to blood flow with the insertion of a device, it is important to estimate this effect in order to influence future design improvements.</p>
<p>CFD analyses on paediatric CVLs allowed a controlled comparison of design features. Overall, smaller models (Tesio 6.5&#x2005;F and Hemo-Cath 8&#x2005;F) were associated with criticalities such as large stagnation regions and high average values of shear stress. These conditions may make the CVLs more prone to thrombus formation (<xref ref-type="bibr" rid="B8">8</xref>). Interestingly, the CVL diameter itself is not immediately related to the criticalities. The Tesio 6.5&#x2005;F recorded a lower percentage of shear stress above the 10&#x2005;Pa threshold compared to the Split Cath III 14&#x2005;F. The highest PLI was found in the 10&#x2005;F model.</p>
<p>Side-holes are one of the most important features in the CVL design. The side-holes indeed facilitate the exchange of blood by increasing the surface of exchange and by providing additional blood ports in case of obstruction of the lumen tip (<xref ref-type="bibr" rid="B12">12</xref>). However, results of this study showed how these are associated with critical conditions. Each CVL model displayed different side-holes configuration, sometime differing between venous and arterial lumens. This study allowed us to evaluate the pros and cons of these configurations. It was noted that a parallel dual-hole configuration, in the arterial lumen, and a step-down in diameter, in the venous one, can guarantee a good flow distribution among the side-holes (as in the Hemo-Cath 8&#x2005;F and in the Pediatric Split Cath 10&#x2005;F). A spiral arrangement of the side-holes (Tesio 6.5&#x2005;F) divides flow equally between the side-holes and the tip, but unbalances in favour of the most proximal holes were observed. The presence of an oval and large side-hole together with circular and smaller holes (Split Cath III 14&#x2005;F) does not seem to have a positive impact on the flow distribution.</p>
<p>The arterial lumens of the CVLs, where blood is withdrawn, consistently presented the most challenging working conditions. In none of the models studied the tip could aspirate a considerable percentage of blood. The distribution of flow between tip and different side-holes was always unbalanced. In return, significant recirculation and stagnation areas were found at the tip of each design. This might lead to blood stagnation and clotting.</p>
<p>The venous tips, used to return blood back to the patient, did not show major criticalities. In this case, the side-holes contribute to reduce the velocity of the outflowing blood. However, with the current side holes configurations, the flow distribution appears to be unbalanced.</p>
<p>Performance of the CVLs were compared at three different working conditions, i.e., minimum, intermediate and maximum flow rate. As expected, an increase in the flow rate corresponded to a linear increase of the levels of shear stress, independently from the CVL design. In all the models, increasing the flow rate also had a small impact on the flow repartition, in favour of the tip. Only the lumen of the Tesio 6.5&#x2005;F model showed the opposite behaviour when used in venous configuration. Additionally, the flow rate had a visible impact also on the arterial flow repartition of the Tesio 6.5&#x2005;F model. The results of paediatric CVLs were compared to the ones reported on devices for adult population (<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B11">11</xref>). Overall, paediatric models showed non-superior levels of shear stress, probably related to lower flow rates. However, models of paediatric devices showed much higher values of PLI compared to the adult counterparts (<xref ref-type="bibr" rid="B8">8</xref>).</p>
<p>Simulations on the anatomical model were useful to address the effects of even more realistic conditions. In this scenario, Tesio 6.5&#x2005;F and Hemo-Cath 8&#x2005;F showed higher levels of shear stress recorded. Anatomical conditions proved to be useful, also, in evaluating the impact of the CVL insertion on the flow conditions of the SVC. The increase in WSS values, recorded after the insertion of the CVL, may trigger the expression of procoagulant activity or cause vascular endothelial damage, leading to a higher risk of venous thrombosis in agreement with Park (<xref ref-type="bibr" rid="B15">15</xref>). When smaller models (Tesio 6.5&#x2005;F and Hemo-Cath 8&#x2005;F) were used, a more disturbed flow, characterized by asymmetric eddies, was observed in the vein cross-section. These flow disturbances could therefore initiate and accelerate thrombosis (<xref ref-type="bibr" rid="B15">15</xref>, <xref ref-type="bibr" rid="B16">16</xref>).</p>
<p>Despite the potential relevance of these findings, the approach presented in this study has some limitations which are here summarised. First, boundary conditions of both ideal and anatomical SVC models were generalised from the literature. Flow pulsatility was neglected and vessel walls were assumed rigid, which allowed to run simplified simulations in a reduced amount of time. In the future, information from imaging such as ultrasounds and magnetic resonance might help in studying the effect of more realistic velocity profiles. Second, CVLs can be inserted with different angles and orientations. These positional differences could influence the CVLs performance and the impact of its insertion on the SVC. Third, the mapping of the results of simulations to clinical data was not performed in this study.</p>
<p>Nevertheless, the findings of this work might be a first step to guide the development of new CVL design, optimised for paediatric patients. Indeed, at the present time only limited models of paediatric CVLs exist and these often look like miniaturisation of the adult ones without accounting for the specialised challenges of paediatric design, which may be the cause of the frequent malfunctions seen within clinical practice of up to 25&#x0025; (<xref ref-type="bibr" rid="B5">5</xref>). Using CFD tools, new tip geometries could be investigated to overcome the limitations of the current designs.</p>
</sec>
<sec id="s5" sec-type="conclusions">
<title>Conclusion</title>
<p>This study provides an understanding on the performance of commercially available paediatric CVL models currently used in children on HD.</p>
<p>Comparative CFD analyses demonstrated advantages and criticalities of each CVL. Various geometric features of a CVL design have an impact on the overall performance. Smaller lines were mostly associated with higher values of shear stress at lower flow rates, while higher PLI values were computed for the larger sized CVL designs. Large areas of blood stagnation were also identified in the arterial lumen of all the CVLs investigated. These finding might contribute to improve the design and therefore outcomes of CVLs used in children.</p>
</sec>
</body>
<back>
<sec id="s6" sec-type="data-availability">
<title>Data availability statement</title>
<p>The original contributions presented in the study are included in the article/<xref ref-type="sec" rid="s10">Supplementary Material</xref>, further inquiries can be directed to the corresponding author/s.</p>
</sec>
<sec id="s7">
<title>Author contributions</title>
<p>Computational study concept: CB, CC. Clinical advisor: RS. Medical imaging: IS. Data collection: CB. Data analysis: CB, ES, CC. Drafting article: CB, CC. Critical revision and approval of article: all authors. All authors contributed to the article and approved the submitted version.</p>
</sec>
<sec id="s8" sec-type="funding-information">
<title>Funding</title>
<p>This work is supported by a Paediatric Innovation grant (Paed_IN_004_20190926) from Kidney Research UK.</p>
</sec>
<sec id="s11" sec-type="COI-statement">
<title>Conflict of interest</title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec id="s9" sec-type="disclaimer">
<title>Publisher&#x0027;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<sec id="s10" sec-type="supplementary-material">
<title>Supplementary material</title>
<p>The Supplementary Material for this article can be found online at: <ext-link ext-link-type="uri" xlink:href="https://www.frontiersin.org/articles/10.3389/fped.2022.1055212/full#supplementary-material">https://www.frontiersin.org/articles/10.3389/fped.2022.1055212/full&#x0023;supplementary-material</ext-link>.</p>
<supplementary-material id="SD1" content-type="local-data">
<media mimetype="application" mime-subtype="pdf" xlink:href="Datasheet1.pdf"/>
</supplementary-material>
</sec>
<ref-list><title>References</title>
<ref id="B1"><label>1.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Harambat</surname><given-names>J</given-names></name><name><surname>Van Stralen</surname><given-names>KJ</given-names></name><name><surname>Kim</surname><given-names>JJ</given-names></name><name><surname>Tizard</surname><given-names>EJ</given-names></name></person-group>. <article-title>Epidemiology of chronic kidney disease in children</article-title>. <source>Pediatr Nephrol</source>. (<year>2012</year>) <volume>27</volume>(<issue>3</issue>):<fpage>363</fpage>&#x2013;<lpage>73</lpage>. <pub-id pub-id-type="doi">10.1007/s00467-011-1939-1</pub-id><pub-id pub-id-type="pmid">21713524</pub-id></citation></ref>
<ref id="B2"><label>2.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Rees</surname><given-names>L</given-names></name><name><surname>Schaefer</surname><given-names>F</given-names></name><name><surname>Schmitt</surname><given-names>CP</given-names></name><name><surname>Shroff</surname><given-names>R</given-names></name><name><surname>Warady</surname><given-names>BA</given-names></name></person-group>. <article-title>Chronic dialysis in children and adolescents: challenges and outcomes</article-title>. <source>Lancet Child Adolesc Heal</source>. (<year>2017</year>) <volume>1</volume>(<issue>1</issue>):<fpage>68</fpage>&#x2013;<lpage>77</lpage>. <pub-id pub-id-type="doi">10.1016/S2352-4642(17)30018-4</pub-id></citation></ref>
<ref id="B3"><label>3.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ma</surname><given-names>A</given-names></name><name><surname>Shroff</surname><given-names>R</given-names></name><name><surname>Hothi</surname><given-names>D</given-names></name><name><surname>Lopez</surname><given-names>MM</given-names></name><name><surname>Veligratli</surname><given-names>F</given-names></name><name><surname>Calder</surname><given-names>F</given-names></name><etal/></person-group> <article-title>A comparison of arteriovenous fistulas and central venous lines for long-term chronic haemodialysis</article-title>. <source>Pediatr Nephrol</source>. (<year>2013</year>) <volume>28</volume>(<issue>2</issue>):<fpage>321</fpage>&#x2013;<lpage>6</lpage>. <pub-id pub-id-type="doi">10.1007/s00467-012-2318-2</pub-id><pub-id pub-id-type="pmid">23052655</pub-id></citation></ref>
<ref id="B4"><label>4.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Shroff</surname><given-names>R</given-names></name><name><surname>Calder</surname><given-names>F</given-names></name><name><surname>Bakkalo&#x01E7;lu</surname><given-names>S</given-names></name><name><surname>Nagler</surname><given-names>EV</given-names></name><name><surname>Stuart</surname><given-names>S</given-names></name><name><surname>Stronach</surname><given-names>L</given-names></name><etal/></person-group> <article-title>Vascular access in children requiring maintenance haemodialysis: a consensus document by the European society for paediatric nephrology dialysis working group</article-title>. <source>Nephrol Dial Transplant</source>. (<year>2019</year>) <volume>34</volume>(<issue>10</issue>):<fpage>1746</fpage>&#x2013;<lpage>65</lpage>. <pub-id pub-id-type="doi">10.1093/ndt/gfz011</pub-id><pub-id pub-id-type="pmid">30859187</pub-id></citation></ref>
<ref id="B5"><label>5.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Borzych-Duzalka</surname><given-names>D</given-names></name><name><surname>Shroff</surname><given-names>R</given-names></name><name><surname>Ariceta</surname><given-names>G</given-names></name><name><surname>Yap</surname><given-names>YC</given-names></name><name><surname>Paglialonga</surname><given-names>F</given-names></name><name><surname>Xu</surname><given-names>H</given-names></name><etal/></person-group> <article-title>Vascular access choice, complications, and outcomes in children on maintenance hemodialysis: findings from the international pediatric hemodialysis network (IPHN) registry</article-title>. <source>Am J Kidney Dis</source>. (<year>2019</year>) <volume>74</volume>(<issue>2</issue>):<fpage>193</fpage>&#x2013;<lpage>202</lpage>. <pub-id pub-id-type="doi">10.1053/j.ajkd.2019.02.014</pub-id><pub-id pub-id-type="pmid">31010601</pub-id></citation></ref>
<ref id="B6"><label>6.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Barnacle</surname><given-names>A</given-names></name><name><surname>Arthurs</surname><given-names>OJ</given-names></name><name><surname>Roebuck</surname><given-names>D</given-names></name><name><surname>Hiorns</surname><given-names>MP</given-names></name></person-group>. <article-title>Malfunctioning central venous catheters in children: a diagnostic approach</article-title>. <source>Pediatr Radiol</source>. (<year>2008</year>) <volume>38</volume>(<issue>4</issue>):<fpage>363</fpage>&#x2013;<lpage>78</lpage>. <pub-id pub-id-type="doi">10.1007/s00247-007-0610-2</pub-id><pub-id pub-id-type="pmid">17932667</pub-id></citation></ref>
<ref id="B7"><label>7.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mandel-Shorer</surname><given-names>N</given-names></name><name><surname>Tzvi-Behr</surname><given-names>S</given-names></name><name><surname>Harvey</surname><given-names>E</given-names></name><name><surname>Revel-Vilk</surname><given-names>S</given-names></name></person-group>. <article-title>Central venous catheter-related venous thrombosis in children with end-stage renal disease undergoing hemodialysis</article-title>. <source>Thromb Res</source>. (<year>2018</year>) <volume>172</volume>(<issue>August</issue>):<fpage>150</fpage>&#x2013;<lpage>7</lpage>. <pub-id pub-id-type="doi">10.1016/j.thromres.2018.10.031</pub-id><pub-id pub-id-type="pmid">30414505</pub-id></citation></ref>
<ref id="B8"><label>8.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mareels</surname><given-names>G</given-names></name><name><surname>Kaminsky</surname><given-names>R</given-names></name><name><surname>Eloot</surname><given-names>S</given-names></name><name><surname>Verdonck</surname><given-names>PR</given-names></name></person-group>. <article-title>Particle image velocimetry-validated, computational fluid dynamics-based design to reduce shear stress and residence time in central venous hemodialysis catheters</article-title>. <source>ASAIO J</source>. (<year>2007</year>) <volume>53</volume>(<issue>4</issue>):<fpage>438</fpage>&#x2013;<lpage>46</lpage>. <pub-id pub-id-type="doi">10.1097/MAT.0b013e3180683b7c</pub-id><pub-id pub-id-type="pmid">17667228</pub-id></citation></ref>
<ref id="B9"><label>9.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Clark</surname><given-names>TWI</given-names></name><name><surname>Isu</surname><given-names>G</given-names></name><name><surname>Gallo</surname><given-names>D</given-names></name><name><surname>Verdonck</surname><given-names>P</given-names></name><name><surname>Morbiducci</surname><given-names>U</given-names></name></person-group>. <article-title>Comparison of symmetric hemodialysis catheters using computational fluid dynamics</article-title>. <source>J Vasc Interv Radiol</source>. (<year>2015</year>) <volume>26</volume>(<issue>2</issue>):<fpage>252</fpage>&#x2013;<lpage>259.e2</lpage>. <pub-id pub-id-type="doi">10.1016/j.jvir.2014.11.004</pub-id><pub-id pub-id-type="pmid">25645414</pub-id></citation></ref>
<ref id="B10"><label>10.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>De Oliveira</surname><given-names>DC</given-names></name><name><surname>Owen</surname><given-names>DG</given-names></name><name><surname>Qian</surname><given-names>S</given-names></name><name><surname>Green</surname><given-names>NC</given-names></name><name><surname>Espino</surname><given-names>DM</given-names></name><name><surname>Shepherd</surname><given-names>DET</given-names></name></person-group>. <article-title>Computational fluid dynamics of the right atrium: assessment of modelling criteria for the evaluation of dialysis catheters</article-title>. <source>PLoS One</source>. (<year>2021</year>) <volume>16</volume>(<issue>2 February</issue>):<fpage>1</fpage>&#x2013;<lpage>26</lpage>. <pub-id pub-id-type="doi">10.1371/journal.pone.0247438</pub-id></citation></ref>
<ref id="B11"><label>11.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Owen</surname><given-names>DG</given-names></name><name><surname>de Oliveira</surname><given-names>DC</given-names></name><name><surname>Qian</surname><given-names>S</given-names></name><name><surname>Green</surname><given-names>NC</given-names></name><name><surname>Shepherd</surname><given-names>DET</given-names></name><name><surname>Espino</surname><given-names>DM</given-names></name></person-group>. <article-title>Impact of side-hole geometry on the performance of hemodialysis catheter tips: a computational fluid dynamics assessment</article-title>. <source>PLoS One</source>. (<year>2020</year>) <volume>15</volume>(<issue>8 August</issue>):<fpage>1</fpage>&#x2013;<lpage>16</lpage>. <pub-id pub-id-type="doi">10.1371/journal.pone.0236946</pub-id></citation></ref>
<ref id="B12"><label>12.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ash</surname><given-names>SR</given-names></name></person-group>. <article-title>Advances in tunneled central venous catheters for dialysis: design and performance</article-title>. <source>Semin Dial</source>. (<year>2008</year>) <volume>21</volume>(<issue>6</issue>):<fpage>504</fpage>&#x2013;<lpage>15</lpage>. <pub-id pub-id-type="doi">10.1111/j.1525-139X.2008.00494.x</pub-id><pub-id pub-id-type="pmid">19000125</pub-id></citation></ref>
<ref id="B13"><label>13.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Haniel</surname><given-names>J</given-names></name><name><surname>Lucas</surname><given-names>TC</given-names></name><name><surname>Huebner</surname><given-names>R</given-names></name></person-group>. <article-title>Evaluation of thrombogenic potential by partial differential equations in the blood flow dynamics with central venous catheter</article-title>. <source>J Brazilian Soc Mech Sci Eng</source>. (<year>2019</year>) <volume>41</volume>(<issue>7</issue>):<fpage>1</fpage>&#x2013;<lpage>9</lpage>. <pub-id pub-id-type="doi">10.1007/s40430-019-1801-0</pub-id></citation></ref>
<ref id="B14"><label>14.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mareels</surname><given-names>G</given-names></name><name><surname>De Wachter</surname><given-names>DS</given-names></name><name><surname>Verdonck</surname><given-names>PR</given-names></name></person-group>. <article-title>Computational fluid dynamics-analysis of the Niagara hemodialysis catheter in a right heart model</article-title>. <source>Artif Organs</source>. (<year>2004</year>) <volume>28</volume>(<issue>7</issue>):<fpage>639</fpage>&#x2013;<lpage>48</lpage>. <pub-id pub-id-type="doi">10.1111/j.1525-1594.2004.07371.x</pub-id><pub-id pub-id-type="pmid">15209857</pub-id></citation></ref>
<ref id="B15"><label>15.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Park</surname><given-names>MH</given-names></name><name><surname>Qiu</surname><given-names>Y</given-names></name><name><surname>Cao</surname><given-names>H</given-names></name><name><surname>Yuan</surname><given-names>D</given-names></name><name><surname>Li</surname><given-names>D</given-names></name><name><surname>Jiang</surname><given-names>Y</given-names></name><etal/></person-group> <article-title>Influence of hemodialysis catheter insertion on hemodynamics in the central veins</article-title>. <source>J Biomech Eng</source>. (<year>2020</year>) <volume>142</volume>(<issue>9</issue>):<fpage>1</fpage>&#x2013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.1115/1.4046500</pub-id></citation></ref>
<ref id="B16"><label>16.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Peng</surname><given-names>L</given-names></name><name><surname>Qiu</surname><given-names>Y</given-names></name><name><surname>Huang</surname><given-names>Z</given-names></name><name><surname>Xia</surname><given-names>C</given-names></name><name><surname>Dai</surname><given-names>C</given-names></name><name><surname>Zheng</surname><given-names>T</given-names></name><etal/></person-group> <article-title>Numerical simulation of hemodynamic changes in central veins after tunneled cuffed central venous catheter placement in patients under hemodialysis</article-title>. <source>Sci Rep</source>. (<year>2017</year>) <volume>7</volume>(<issue>1</issue>):<fpage>3</fpage>&#x2013;<lpage>10</lpage>. <pub-id pub-id-type="doi">10.1038/s41598-017-00038-6</pub-id><pub-id pub-id-type="pmid">28127052</pub-id></citation></ref>
<ref id="B17"><label>17.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Cowie</surname><given-names>BS</given-names></name><name><surname>Kluger</surname><given-names>R</given-names></name><name><surname>Rex</surname><given-names>S</given-names></name><name><surname>Missant</surname><given-names>C</given-names></name></person-group>. <article-title>The relationship between superior vena cava diameter and collapsibility and central venous pressure</article-title>. <source>Anaesth Intensive Care</source>. (<year>2015</year>) <volume>43</volume>(<issue>3</issue>):<fpage>357</fpage>&#x2013;<lpage>60</lpage>. <pub-id pub-id-type="doi">10.1177/0310057X1504300312</pub-id><pub-id pub-id-type="pmid">25943610</pub-id></citation></ref>
<ref id="B18"><label>18.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Salim</surname><given-names>MA</given-names></name><name><surname>DiSessa</surname><given-names>TG</given-names></name><name><surname>Arheart</surname><given-names>KL</given-names></name><name><surname>Alpert</surname><given-names>BS</given-names></name></person-group>. <article-title>Contribution of superior vena caval flow to total cardiac output in children: a Doppler echocardiographic study</article-title>. <source>Circulation</source>. (<year>1995</year>) <volume>92</volume>(<issue>7</issue>):<fpage>1860</fpage>&#x2013;<lpage>5</lpage>. <pub-id pub-id-type="doi">10.1161/01.CIR.92.7.1860</pub-id><pub-id pub-id-type="pmid">7671370</pub-id></citation></ref>
<ref id="B19"><label>19.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Cavazzuti</surname><given-names>M</given-names></name><name><surname>Corticelli</surname><given-names>MA</given-names></name><name><surname>Karayiannis</surname><given-names>TG</given-names></name></person-group>. <article-title>Compressible fanno flows in micro-channels: an enhanced quasi-2D numerical model for turbulent flows</article-title>. <source>Int Commun Heat Mass Transf</source>. (<year>2020</year>) <volume>111</volume>(<issue>December 2019</issue>):<fpage>104448</fpage>. <pub-id pub-id-type="doi">10.1016/j.icheatmasstransfer.2019.104448</pub-id></citation></ref>
<ref id="B20"><label>20.</label><citation citation-type="book"><collab>ANSYS Inc</collab>. <source>ANSYS&#x00AE; fluent user&#x2019;s guide (release 15.0):section 7.4.5</source>. <publisher-loc>Canonsburg</publisher-loc>: <publisher-name>ANSYS Inc</publisher-name> (<year>2013</year>).</citation></ref></ref-list>
</back>
</article>