Abstract
Electrical stimulation of the central and peripheral nervous systems - such as deep brain stimulation, spinal cord stimulation, and epidural cortical stimulation are common therapeutic options increasingly used to treat a large variety of neurological and psychiatric conditions. Despite their remarkable success, there are limitations which if overcome, could enhance outcomes and potentially reduce common side-effects. Micromagnetic stimulation (μMS) was introduced to address some of these limitations. One of the most remarkable properties is that μMS is theoretically capable of activating neurons with specific axonal orientations. Here, we used computational electromagnetic models of the μMS coils adjacent to neuronal tissue combined with axon cable models to investigate μMS orientation-specific properties. We found a 20-fold reduction in the stimulation threshold of the preferred axonal orientation compared to the orthogonal direction. We also studied the directional specificity of μMS coils by recording the responses evoked in the inferior colliculus of rodents when a pulsed magnetic stimulus was applied to the surface of the dorsal cochlear nucleus. The results confirmed that the neuronal responses were highly sensitive to changes in the μMS coil orientation. Accordingly, our results suggest that μMS has the potential of stimulating target nuclei in the brain without affecting the surrounding white matter tracts.
Introduction
Implanted medical devices based on electrical stimulation such as cardioverter-defibrillators and pacemakers (), spinal cord stimulation (), and deep brain stimulation (DBS) () devices have become well-accepted therapeutic options to treat a wide variety of medical conditions. Electrical stimulation has considerable clinical impact in alleviating symptoms of an increasingly diverse range of neurological and psychiatric disorders including for example, cochlear () and auditory brainstem implants for restoring hearing (), DBS to treat symptoms of Parkinsonism (; ), cortical stimulation for epilepsy and depression (; ; ), spinal cord stimulation for neuropathic pain (), and vagus nerve stimulation for epilepsy () and depression (; ), just to mention a few. More recently, electrical stimulation has also shown promise for the restoration of function of retinal implants to restore vision in the blind (; ; ; ; ).
Although electrical techniques for neuronal stimulation have proven quite useful, they have several limitations that can be overcome by micro magnetic stimulation (μMS) which uses sub-millimeter coils. For example, for an electrode pair to generate currents it needs to be placed in contact with a conductive media (e.g., excitable tissue). Electric currents that are delivered by these electrodes diffuse and can spread to undesired areas adjacent to the structures being targeted, leading to unintended side-effects (; ; ; ). A magnetic coil, on the other hand, can induce electric currents in the tissue from a distance (i.e., through an insulation layer). In nature these currents are closed-loop circular currents with a higher spatial focality (Figure 1). Furthermore, the fact that μMS coils can deliver stimulation while being insulated from the tissue increases their biocompatibility and compatibility with magnetic resonance imaging (considering no ferromagnetic material is present). Finally, as μMS coils can be positioned within or immediately adjacent to the neural tissue, the power needed to evoke neuronal activities is significantly reduced compared to techniques such as transcranial magnetic stimulation (TMS) which are designed to generate strong magnetic fields that pass through the skull and deliver stimulation to the cortical tissue.
FIGURE 1
Our group recently demonstrated the feasibility of using μMS to elicit neuronal activation in vitro (), as well as the activation of neuronal circuitry on the system level in vivo (). As μMS is a novel technology, its mechanism(s) of nerve activation, induced field characteristics, and optimum topological features are yet to be explored. In this work, we performed numerical simulations to provide an insight into spatial distribution of μMS-induced electric fields, which in turn dictate the dynamics of nerve stimulation threshold changes with different axonal directionalities. Electromagnetic simulations were performed to estimate the magnetic flux
and the electric field
and its spatial gradient at different distances from the coil. These simulations were based on the actual coil prototypes built (Figure 2) and utilized in our animal experiments (Figure 3). The estimated
fields were then used in conjunction with the NEURON cable model to investigate the directional sensitivity of μMS (Figures 4, 5). Finally, we performed in vivo experiments where we studied responses evoked in the inferior colliculus (IC) of rodents by applying μMS stimuli to the surface of animal’s dorsal cochlear nucleus (DCN). Specifically, we examined the IC responses to different coil orientations (Figure 6).
FIGURE 2
FIGURE 3
FIGURE 4
FIGURE 5

Simulations results. (Top) Computational results of the magnitude of Ex and Ey and their spatial derivatives (dEx/dx, dEy/dy) on transverse planes located at 20 μm below the surface of microcoil. (Bottom) NEURON estimated current threshold levels for the two orientations (orange-mediolateral and blue-rostrocaudal) at various distances from the axon.
FIGURE 6

Simulation results. Schematic comparison of the distribution of electric field produced during electric and magnetic nerve stimulation. (A) A conventional DBS electrode pair in bipolar configuration (electrode length, diameter, and spacing mimics the Medtronic Lead 3389). The differential voltage between anode and cathode is set to be at the lowest limit of typical values (ranging from 1 to 5 V). (B) A μMS coil with dimensions matching the DBS electrode. The coil is fed with 20A similar to those used in our experiments and is perpendicular to the surface of the tissue. The electric field is symmetrical, but is confined to an annulus region under the coil. (C) A μMS parallel to the surface of the tissue. The electric field is mostly confined to a region beneath the coil and it’s asymmetric, leading to different neural activation thresholds depending on neurons’ axonal direction.
Methods
Electromagnetic Simulations
Numerical modeling has been long used to understand the phenomenology of field-tissue interaction in a wide variety of medical and diagnostic applications. Examples include use of electrostatic finite element modeling to predict the volume of activated tissue in electrical brain stimulation (
to find current density and then subtracts all input currents and other sources. For a perfect solution, the result would be zero, whereas for a real finite mesh the result would include some amount of residual current density. An energy value calculated from this residual current density is then used as the criteria to refine the mesh. An iterative process then will follow, which refines the mesh in each step until the energy error is below a user-specified value (1% in our case). The final solution had ∼630,000 mesh elements with edge length varying from 9 μm inside the tissue to 2 mm at the outer boundary the air box. The simulations converged after two adaptive passes which completed in 17 h on a Dell PowerEdge R730 with 16x32GB = 512GB of RAM, an NVIDIA K80 GPU and 28 cores (2xIntel Xeon CPU with each 14 cores) running 64-bit Windows Server 2012. Electric field values were then exported to MATLAB (The Mathworks, Inc., Natick, MA, United States) for smoothing and were used to simulate the response of neurons with different orientations below the coil.
μMS Coil Orientations
In our previous work (
Neuron Modeling
A computational model of axons was built for simulation of neuronal activation for the three-dimensional electric fields obtained in the previous section. The parallel fiber axon model was assumed to have a diameter of 2 μm (
Microcoil Construction
All microcoils were constructed to keep overall resistances below 5 Ω and inductances below 150 nH (4263B, Keysight Technologies, Santa Rosa, CA, United States) in order to ensure high stimulation efficacy. A multilayer inductor (ELJ-RFR10JFB, Panasonic Electronic Devices Corporation of America, Knoxville, TN, United States) was attached by soldering to two 34-AWG copper wires (Philmore Mfg., Rockford, IL, United States) with polyimide enamel inner coat and polyurethane overcoat. To insulate the tissue from the voltage applied and to protect against moisture, the microcoils were coated with an acrylic conformal coating (419C, MG = Chemicals, Burlington, ON, Canada) that offered high dielectric strength. The 419C Technical Data Sheet reports a thickness of 25 μm with an estimated variability of the dielectric thickness to be ± 5 μm. After curing (24 h), the insulation of the microcoil was tested by immersing the coils in a saline solution (0.9% of NaCl) and verifying that the resistance between the microcoil and an EGG electrode also dipped in the saline solution was greater than 5 MΩ (TX3, Tektronix, Inc., Beaverton, OR, United States). The microscopic stimulator was placed on the tip of a 23 AWG needle (Becton Dickinson, Franklin Lakes, NJ, United States) and the two wires inserted in the shaft/hub of the needle, and through the tip and barrel of a 3 ml syringe with the plunger that was removed and the flange end piece of the syringe was attached to a BNC connector by means of a glue gun, and electrically connected.
Magnetic Stimulation
Two adult male Syrian golden hamsters were studied in this work. All procedures performed were approved by the Institutional Animal Care and Use Committee of the Cleveland Clinic, which adheres to the NIH Guide for the Care and Use of Laboratory Animals. In each experiment, μMS coils was mounted to a micromanipulator and manually positioned so that the coil was located just above the dorsal surface of the DCN, as described by
Electrophysiology
Recordings were conducted at multiple sites along the tonotopic axis of the central nucleus of the contralateral IC. This region was recognized by its sharp tuning properties and by the progression from high to low-frequency selectivities as the electrode was moved along the dorsoventral axis. Methods for recording and analyzing multiunit signals were similar to those described in previous studies (
Acoustic Stimulation
Acoustic stimuli were needed for the dual purposes of characterizing the frequency tuning properties of recorded neurons to determine tonotopic coordinates of the IC recording electrodes, and also to examine effects of changing the acoustic stimulus conditions on IC responses. For both measures, we used 40 ms tone bursts (5 ms rise/fall times, 40 ms interstimulus intervals). For testing the tuning properties, we used a battery of 800 tone bursts varied in frequency from 3 to 32 kHz and in intensity from 6 to 96 dB SPL as previously described (
Data Analysis
Samples of activity recorded from IC neurons were obtained from 100 to 150 repetitive stimuli and plotted as a function of time. The resulting time sweeps were used to compare responses during and following stimulation with the activity level recorded during the prestimulus period and to derive measures of several response characteristics, such as amplitude and magnetic coil orientation. Each of these measures was obtained for each of the parameters of stimulation that were tested and were compared with responses to baseline activities stimulation. Specifically, for each stimulation parameter (Amplitude and Orientation) baseline activities (15 ms before each stimulation pulse) were compared to stimulation response (15 ms following stimulation artifact; 17–32 ms). To facilitate comparisons, the absolute value of raw electrophysiological activities were summed for the baseline and stimulus–response periods for each stimulus delivered. Significant differences in responses for each parameter were compared using parametric ANOVA with Bonferroni corrections. In each analysis, responses were compared to their baseline activities and then tested relative to each amplitude or orientation.
Results
We have explored the effects of coil orientation on the resulting stimulation capabilities both with numerical simulations and with animal studies.
Numerical Simulations
The use of magnetic fields to induce electric fields or currents in the tissue from a distance is extremely inefficient from an energy standpoint. We hypothesize that much smaller energy than TMS may be required for neural stimulation at a microscopic level. One important difference between
and
is that the magnitude of the latter is well-known to fall much more rapidly in space (e.g., quadratic vs. cubic law for an electric vs. magnetic dipole in empty space). Our hypothesis is based on the prediction by various activation models (
The FEM simulations also predict that solenoidal μMS coils placed parallel to the surface of the tissue are capable of differentially activating neurons based on their axonal direction. It is established that neural activation function is proportional to the spatial derivative of the electric field along the axon’s axis (
Importantly, and as demonstrated here, μMS provides a unique opportunity over electrical stimulation techniques in that neural interfaces can be constructed that take advantage of the orientation properties provided by magnetic stimulation. Namely, the construction of brain stimulation leads that maximally active the target tissue while mitigating the activation of fibers in the passage would have a significant advantage in DBS therapies, as the activation of fibers of passage represents the greatest side-effects for patients. Likewise, μMS coils could be used to provide more spatial resolution over existing electrical stimulation strategies, by designing interfaces that account for the orientation of the coils relative to the target tissue to be activated. This can be better appreciated from Figure 6, illustrating the schematic of the electric field distribution produced by (A) a conventional DBS electrode pair in bipolar configuration and (B,C) same-sized μMS coils in perpendicular and parallel positions. It can be observed from the figure that even for a DBS voltage as low as 1 V (typical values range from 1 to 5 V) the electric field produced by the electrode pair covers a large symmetric area containing both electrode contacts. The electric field of the μMS coil on the other hand, is more confined to the edges at the periphery of the coil when the coil is perpendicular to the tissue, and to the center of the solenoid when the coil is parallel to the tissue. Specifically, when the coil is in parallel position the induced electric field is asymmetric, indicating different sensitivity for neuron activation depending on their axonal direction. Moreover, we speculate that our modeling also brings up the notion that the mechanisms of action of magnetic stimulation may be fundamentally different from that of electrical stimulation. Specifically, the ionic movement of charge that ultimately results in neuronal activation operates differently between magnetic and electrical stimulation. In electrical stimulation, the current flow from pole to pole of the electrical stimulator while in magnetic stimulation the induced current flows as eddy currents relative to the magnetic fields. Electrical stimulation activates neural elements by operating on the electric potential of the extracellular matrix and manipulating the transmembrane potentials. In contrast, eddy currents act not only upon the extracellular matrix but also on the intracellular matrix as the magnetic stimulation fields penetrate the cellular compartments.
In general, μMS operates similarly to TMS generating time-varying magnetic fields and inducing electric fields in the brain, which can stimulate surrounding cortical or subcortical neurons (
Animal Experiments
Recently it has been demonstrated that μMS is capable of eliciting neuronal activation in both retinal ganglion cells in vitro (
The microcoil prototypes we tested in this study reproducibly activated the brain, in vivo, in a tissue-appropriate manner consistent with the known microcircuitry of the DCN and projection patterns that link the DCN to the IC. Two adult male Syrian golden hamsters were studied, and all surgical procedures used to expose the DCN and IC were the same as previously described (
FIGURE 7

Animal experiments results. Top: responses of the IC to magnetic stimuli delivered to the DCN from microscopic stimulators with different orientations with respect to the long axis of the DCN. Blue lines: microcoil oriented parallel to the medial–lateral (tonotopic) axis of the DCN. Red lines: microcoil oriented perpendicular to the medial–lateral axis of the DCN. Recordings from two animals are shown for each orientation. Each curve represents the average of 100 trials performed for each coil orientation on each animal.
In a set of experiments, we examined the effect of coil orientation on neuronal activation properties. Two different orientations of the microcoils were studied, one with the long axis of the μMS parallel to the long (medial–lateral or tonotopic) axis of the DCN (Figure 3, bottom), and the other with the long axis of the μMS coil parallel to the rostrocaudal axis of the DCN (i.e., parallel to the isofrequency bands) (Figure 3, bottom). Strong IC responses were observed for microcoil orientations parallel to the medial–lateral axis of the DCN, while weaker or absent responses for the orthogonal orientations (Figure 7, top). In animal 2 (Figure 7, bottom), both the medial–lateral orientations were different from the rostral–caudal orientation but were not different from animal 1.
An important aspect of our results that was unexpected was the dependence of the strength of the IC response on the rotational angle of the microcoil above the DCN. IC responses were vigorous when the long axis of the microcoil was parallel to the medial–lateral axis of the DCN but weak for micro-stimulator orientations parallel to the rostrocaudal axis. This difference implies contrasting levels of efficacy in stimulus–response coupling between the different micro-stimulator orientations. The simplest mechanism to explain this results is that micro-stimulator orientations parallel to the medial–lateral axis of the DCN more effectively excite the main output neurons of the DCN, the fusiform cells, which project to the contralateral IC (
Discussion
There are some limitations that currently limit the efficacy and safety of electrical stimulation. First, electric currents delivered by microelectrodes can spread to undesired areas adjacent to the targeted structures, leading to unintended side effects (
are depolarized or hyperpolarized depending on the direction and strength of
, but the processes transverse to the
are not affected (
Second, unlike electrical stimulation, μMS does not require direct galvanic contact with the tissue. For an electrode pair (Figure 1A) to generate current, it needs to be placed in direct contact with a conductive media (e.g., excitable tissue). In a bipolar electrode pair, the ‘anode’ or source (Figure 1A, plus sign) injects a current and hyperpolarizes the neuronal membrane toward more negative potential which can arrest the neural action, whereas the ‘cathode’ acts as a sink and depolarizes the axon membrane which could trigger an action potential (Figure 1A, minus sign). However, the metal electrode implanted in the tissue may lead to oxidation–reduction reaction at the electrode-tissue interface changing the pH of surrounding tissue which may provoke an immune response. Histopathology analysis has shown gliosis and spongiosis around the stimulation electrode track (
inside and in the space surrounding the coil. In the conductive tissue, this time-varying magnetic field
, in turn, generates an orthogonal current density
capable of evoking neuronal action potentials (Figures 1B–D), according to Faraday’s Law [i.e., ∇×
= − in homogeneous isotropic medium where ∇× is the curl operator and σ is, the tissue conductivity, albeit the brain has tissues with anisotropic conductivities (
Finally, unlike electrical stimulation μMS does not require a charge-balanced stimulation waveform. In electrical stimulation, charge balancing is necessary to avoid excessive charge accumulation at the neural interface, and thus undesired stimulation and electroporation (
is induced by the time-varying magnetic field. The current density in the tissue
is a rotating field that mirrors the current direction in the coil (Figure 1B). Because the induced electric field is a solenoidal or incompressible vector field in three dimensions, μMS does not suffer from charge buildup (
Despite these specific limitations, electrically based DBS has been tremendously successful. However, the application of μMS could mitigate some of the challenges of these limitations. Theoretically, and supported by limited data (
Conclusion
Microscopic magnetic stimulation (μMS) could potentially become the pacemaker and brain stimulator of the future with their contactless ability to deliver the neuronal stimulation needed for therapeutic efficacy in patients with Parkinson’s disease, epilepsy, in need of implantable cardioverter-defibrillators or pacemakers, and so forth. Due to recent advancements in micro-machining technologies, we can now utilize manufactured inductors (or coils) constructed on the sub-millimeter scale to produce magnetic fields. Such coils would offer several advantages over classical electrical and TMS techniques. Unlike TMS coils, the coils are sub-millimeter in size and can be placed within or near a neuronal substrate, increasing spatial resolution and reducing the power needed to evoke neuronal activity. Moreover, because the coils are not in direct contact with the tissue and no current is directly injected into the tissue, they may overcome the inflammatory tissue encapsulation and mitigate charge buildup issues inherent in traditional electrical stimulation technologies. Our data indicate that these microcoils can activate neuronal activity with high degrees of spatial and temporal resolution and that the orientation of the coils relative to the tissue activated can be used to activate specific elements optimally and to avoid the activation of others. Future work will concentrate on developing specific neural models of the target structures to quantify the parameters of μMS for directional stimulation, and include more animal models to establish the statistical features of the neural response.
Statements
Ethics statement
This study was carried out in accordance with the recommendations of Public Health Service (PHS) Policy on Humane Care and Use of Laboratory Animals and the Animal Welfare Act. The protocol was approved by the institutional animal care and use committee (IACUC) Committee of the Cleveland Clinic.
Author contributions
LaG, JG, JK, and GB conceptualized the study; provided support and guidance with data interpretation. H-JP designed and performed the NEURON analyses with support from LaG and JK. NM and LyG designed and performed the animal experiments with support from JK and H-JP. LaG and GB contributed electromagnetic data analysis and visualization software. FH built the μMS coils. LaG wrote the manuscript, with contributions from GB and comments from all other authors.
Funding
This work was supported by the US National Institutes of Health (NIH) grants R43MH107037 (FH, JG, GB, and JK), R01MH111875 (GB), and K99EB021320 (LaG).
Conflict of interest
JG provided consulting services for Alpha Omega Co. USA, Inc. and author FH was employed by the company FHC, Inc. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
1
AstromM.DiczfalusyE.MartensH.WardellK. (2015). Relationship between neural activation and electric field distribution during deep brain stimulation.IEEE Trans. Biomed. Eng.62664–672. 10.1109/TBME.2014.2363494
2
AytonL. N.BlameyP. J.GuymerR. H.LuuC. D.NayagamD. A.SinclairN. C.et al (2014). First-in-human trial of a novel suprachoroidal retinal prosthesis.PLoS One9:e115239. 10.1371/journal.pone.0115239
3
BarkerA. T.JalinousR.FreestonI. L. (1985). Non-invasive magnetic stimulation of human motor cortex.Lancet11106–1107. 10.1016/S0140-6736(85)92413-4
4
BasserP. J.RothB. J. (1991). Stimulation of a myelinated nerve axon by electromagnetic induction.Med. Biol. Eng. Comput.29261–268. 10.1007/BF02446708
5
BehrendM. R.AhujaA. K.HumayunM. S.ChowR. H.WeilandJ. D. (2011). Resolution of the epiretinal prosthesis is not limited by electrode size.IEEE Trans. Neural Syst. Rehabil. Eng.19436–442. 10.1109/TNSRE.2011.2140132
6
BenabidA. L. (2003). Deep brain stimulation for Parkinson’s disease.Curr. Opin. Neurobiol.13696–706. 10.1016/j.conb.2003.11.001
7
BencsikM.BowtellR.BowleyR. (2007). Electric fields induced in the human body by time-varying magnetic field gradients in MRI: numerical calculations and correlation analysis.Phys. Med. Biol.522337–2353. 10.1088/0031-9155/52/9/001
8
BeurrierC.BioulacB.AudinJ.HammondC. (2001). High-frequency stimulation produces a transient blockade of voltage-gated currents in subthalamic neurons.J. Neurophysiol.851351–1356. 10.1152/jn.2001.85.4.1351
9
BeyerlB. D. (1978). Afferent projections to the central nucleus of the inferior colliculus in the rat.Brain Res.145209–223. 10.1016/0006-8993(78)90858-2
10
BiksonM.DmochowskiJ.RahmanA. (2013). The “quasi-uniform” assumption in animal and computational models of non-invasive electrical stimulation.Brain Stimul.6704–705. 10.1016/j.brs.2012.11.005
11
BlackstadT. W.OsenK. K.MugnainiE. (1984). Pyramidal neurones of the dorsal cochlear nucleus: a Golgi and computer reconstruction study in cat.Neuroscience13827–854. 10.1016/0306-4522(84)90099-X
12
BlakeD. T.BhattiP. T.Beek-KingJ. V.CrawfordJ.McKinnonB. (2014). “Micro magnetic stimulation of the feline cochlea,” inProceedings of the 13th International Conference on Cochlear Implants and Other Implantable Auditory Technologies, Munich, S4–S8.
13
BonmassarG.LeeS. W.FreemanD. K.PolasekM.FriedS. I.GaleJ. T. (2012). Microscopic magnetic stimulation of neural tissue.Nat. Commun.3:921. 10.1038/ncomms1914
14
ButsonC. R.McIntyreC. C. (2006). Role of electrode design on the volume of tissue activated during deep brain stimulation.J. Neural Eng.31–8. 10.1088/1741-2560/3/1/001
15
CantN. B.BensonC. G. (2003). Parallel auditory pathways: projection patterns of the different neuronal populations in the dorsal and ventral cochlear nuclei.Brain Res. Bull.60457–474. 10.1016/S0361-9230(03)00050-9
16
Caparros-LefebvreD.RuchouxM. M.BlondS.PetitH.PercheronG. (1994). Long-term thalamic stimulation in Parkinson’s disease: postmortem anatomoclinical study.Neurology441856–1860. 10.1212/WNL.44.10.1856
17
CarnevaleN. T.HinesM. L. (2006). The Neuron Book.Cambridge: Cambridge University Press. 10.1017/CBO9780511541612
18
ChenC.SmyeS. W.RobinsonM. P.EvansJ. A. (2006). Membrane electroporation theories: a review.Med. Biol. Eng. Comput.445–14. 10.1007/s11517-005-0020-2
19
Clare McElcheranL. G.GrahamS. (2014). “Reduced heating of implanted electrical conductors using parallel radiofrequency transmission,” inProceedings of the Joint Annual Meeting of the International Society of Magnetic Resonance in Medicine (ISMRM), Milan.
20
CondonB.HadleyD. M. (2000). Potential MR hazard to patients with metallic heart valves: the Lenz effect.J. Magn. Reson. Imaging12171–176. 10.1002/1522-2586(200007)12:1<171::AID-JMRI19>3.0.CO;2-W
21
CrowleyJ. M. (1973). Electrical breakdown of bimolecular lipid membranes as an electromechanical instability.Biophys. J.13711–724. 10.1016/S0006-3495(73)86017-5
22
DengZ.-D.LisanbyS. H.PeterchevA. V. (2013). Electric field depth–focality tradeoff in transcranial magnetic stimulation: simulation comparison of 50 coil designs.Brain Stimul.61–13. 10.1016/j.brs.2012.02.005
23
DeuschlG.Schade-BrittingerC.KrackP.VolkmannJ.SchäferH.BötzelK.et al (2006). A randomized trial of deep-brain stimulation for Parkinson’s disease.N. Engl. J. Med.355896–908. 10.1056/NEJMoa060281
24
EllenbogenK. A.WoodM. A. (2008). Cardiac Pacing and ICDs, 5th Edn, Vol. IX. Hoboken, NJ: Blackwell Publishing, 558.
25
FinlaysonP. G.KaltenbachJ. A. (2009). Alterations in the spontaneous discharge patterns of single units in the dorsal cochlear nucleus following intense sound exposure.Hear. Res.256104–117. 10.1016/j.heares.2009.07.006
26
FujinoK.OertelD. (2003). Bidirectional synaptic plasticity in the cerebellum-like mammalian dorsal cochlear nucleus.Proc. Natl. Acad. Sci. U.S.A.100265–270. 10.1073/pnas.0135345100
27
GiffordR. H. (2013). Cochlear Implant Patient Assessment: Evaluation of Candidacy, Performance, and Outcomes. Core Clinical Concepts in Audiology.San Diego, CA: Plural Publishing, 144.
28
Golestani-RadL.ElahiB.Rashed-MohasselJ. (2007). Investigating the effects of external fields polarization on the coupling of pure magnetic waves in the human body in very low frequencies.Biomagn. Res. Technol.5:3. 10.1186/1477-044X-5-3
29
GolestaniradL.DlalaE.WrightG.MosigJ. R.GrahamS. J. (2012a). Comprehensive analysis of Lenz Effect on the artificial heart valves during magnetic resonance imaging.Prog. Electromagn. Res.1281–17. 10.2528/PIER12031505
30
GolestaniradL.ElahiB.MolinaA.MosigJ. R.PolloC.ChenR.et al (2013). Analysis of fractal electrodes for efficient neural stimulation.Front. Neuroeng.6:3. 10.3389/fneng.2013.00003
31
GolestaniradL.IaconoM. I.KeilB.AngeloneL. M.BonmassarG.FoxM. D.et al (2017a). Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.Neuroimage147577–588. 10.1016/j.neuroimage.2016.12.056
32
GolestaniradL.IzquierdoA. P.GrahamS. J.MosigJ. R.PolloC. (2012b). Effect of realistic modeling of deep brain stimulation on the prediction of volume of activated tissue.Prog. Electromagn. Res.1261–16. 10.2528/PIER12013108
33
GolestaniradL.KeilB.AngeloneL. M.BonmassarG.MareyamA.WaldL. L. (2017b). Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants.Magn. Reson. Med.771701–1712. 10.1002/mrm.26220
34
GolestaniradL.MattesM.MosigJ. R.PolloC. (2010). Effect of model accuracy on the result of computed current densities in the simulation of transcranial magnetic stimulation.IEEE Trans. Magn.464046–4051. 10.1109/TMAG.2010.2082556
35
GolestaniradL.RouhaniH.ElahiB.ShahimK.ChenR.MosigJ. R.et al (2012c). Combined use of transcranial magnetic stimulation and metal electrode implants: a theoretical assessment of safety considerations.Phys. Med. Biol.57:7813. 10.1088/0031-9155/57/23/7813
36
HistedM. H.BoninV.ReidR. C. (2009). Direct activation of sparse, distributed populations of cortical neurons by electrical microstimulation.Neuron63508–522. 10.1016/j.neuron.2009.07.016
37
HowlandR. H. (2008). Neurosurgical approaches to therapeutic brain stimulation for treatment-resistant depression.J. Psychosoc. Nurs. Ment. Health Serv.4615–19.
38
HumayunM. S.DornJ. D.daCruz LDagnelieG.SahelJ. A.StangaP. E.et al (2012). Interim results from the international trial of Second Sight’s visual prosthesis.Ophthalmology119779–788. 10.1016/j.ophtha.2011.09.028
39
JohnsonM. D.McIntyreC. C. (2008). Quantifying the neural elements activated and inhibited by globus pallidus deep brain stimulation.J. Neurophysiol.1002549–2563. 10.1152/jn.90372.2008
40
KaltenbachJ. A.AfmanC. E. (2000). Hyperactivity in the dorsal cochlear nucleus after intense sound exposure and its resemblance to tone-evoked activity: a physiological model for tinnitus.Hear. Res.140165–172. 10.1016/S0378-5955(99)00197-5
41
KaltenbachJ. A.McCaslinD. L. (1996). Increases in spontaneous activity in the dorsal cochlear nucleus following exposure to high intensity sound: a possible neural correlate of tinnitus.Audit. Neurosci.357–78.
42
KanoldP. O.YoungE. D. (2001). Proprioceptive information from the pinna provides somatosensory input to cat dorsal cochlear nucleus.J. Neurosci.217848–7858. 10.1523/JNEUROSCI.21-19-07848.2001
43
Koshtoiants KhS. (1957). Enzymo-chemical principles of the action of the vagus nerve on the heart (role of potassium).Fiziol. Zh. SSSR Im. I M Sechenova43681–684.
44
KreisP.FishmanS. (2009). Spinal Cord Stimulation Implantation Percutaneous Implantation Techniques.Oxford: Oxford Press.
45
LeeS. W.FriedS. I. (2017). Enhanced control of cortical pyramidal neurons with micromagnetic stimulation.IEEE Trans. Neural Syst. Rehabil. Eng.251375–1386. 10.1109/TNSRE.2016.2631446
46
LiZ.ZhangJ. G.YeY.LiX. (2016). Review on factors affecting targeting accuracy of deep brain stimulation electrode implantation between 2001 and 2015.Stereotact. Funct. Neurosurg.94351–362. 10.1159/000449206
47
LicariF. G.ShkoukaniM.KaltenbachJ. A. (2011). Stimulus-dependent changes in optical responses of the dorsal cochlear nucleus using voltage-sensitive dye.J. Neurophysiol.106421–436. 10.1152/jn.00982.2010
48
LopezW. O.BarbosaD. C.TeixeraM. J.PaizM.MouraL.MonacoB. A.et al (2016). Pain relief in CRPS-II after spinal cord and motor cortex simultaneous dual stimulation.Pain Phys.19E631–E635.
49
MachereyO.CarlyonR. P. (2014). Cochlear implants.Curr. Biol.24R878–R884. 10.1016/j.cub.2014.06.053
50
ManisP. B. (1989). Responses to parallel fiber stimulation in the guinea pig dorsal cochlear nucleus in vitro.J. Neurophysiol.61149–161. 10.1152/jn.1989.61.1.149
51
ManzoorN. F.GaoY.LicariF.KaltenbachJ. A. (2013). Comparison and contrast of noise-induced hyperactivity in the dorsal cochlear nucleus and inferior colliculus.Hear. Res.295114–123. 10.1016/j.heares.2012.04.003
52
ManzoorN. F.LicariF. G.KlapcharM.ElkinR. L.GaoY.ChenG.et al (2012). Noise-induced hyperactivity in the inferior colliculus: its relationship with hyperactivity in the dorsal cochlear nucleus.J. Neurophysiol.108976–988. 10.1152/jn.00833.2011
53
MatteucciP. B.Barriga-RiveraA.EiberC. D.LovellN. H.MorleyJ. W.SuaningG. J. (2016). The effect of electric cross-talk in retinal neurostimulation.Invest. Ophthalmol. Vis. Sci.571031–1037. 10.1167/iovs.15-18400
54
McElcheranC. E.YangB.AndersonK. J. T.GolestaniradL.GrahamS. J. (2017). Parallel radiofrequency transmission at 3 tesla to improve safety in bilateral implanted wires in a heterogeneous model.Magn. Reson. Med.782406–2415. 10.1002/mrm.26622
55
McIntyreC. C.GrillW. M. (2001). Finite element analysis of the current-density and electric field generated by metal microelectrodes.Ann. Biomed. Eng.29227–235. 10.1114/1.1352640
56
McIntyreC. C.MoriS.ShermanD. L.ThakorN. V.VitekJ. L. (2004). Electric field and stimulating influence generated by deep brain stimulation of the subthalamic nucleus.Clin. Neurophysiol.115589–595. 10.1016/j.clinph.2003.10.033
57
McIntyreC. C.RichardsonA. G.GrillW. M. (2002). Modeling the excitability of mammalian nerve fibers: influence of afterpotentials on the recovery cycle.J. Neurophysiol.87995–1006. 10.1152/jn.00353.2001
58
MehtaA. H.OxenhamA. J. (2017). Vocoder simulations explain complex pitch perception limitations experienced by cochlear implant users.J. Assoc. Res. Otolaryngol.18789–802. 10.1007/s10162-017-0632-x
59
MøllerA. R. (2006). Cochlear and Brainstem Implants. Advances in Oto-Rhino-Laryngology, Vol. VIII. New York, NY: Karger, 228. 10.1159/isbn.978-3-318-01380-1
60
MontgomeryE. B. (2010). Deep Brain Stimulation Programming: Principles and Practice, Vol. XVII. Oxford: Oxford University Press, 179.
61
MoraceR.Di GennaroG.QuaratoP.D’AnielloA.AmasciaA.GrammaldoL.et al (2016). Deep brain stimulation for intractabile epilepsy.J. Neurosurg. Sci.60189–198.
62
MugnainiE.WarrW. B.OsenK. K. (1980). Distribution and light microscopic features of granule cells in the cochlear nuclei of cat, rat, and mouse.J. Comp. Neurol.191581–606. 10.1002/cne.901910406
63
NagarajanS. S.DurandD. M.WarmanE. N. (1993). Effects of induced electric fields on finite neuronal structures: a simulation study.IEEE Trans. Biomed. Eng.401175–1188. 10.1109/10.245636
64
NdukaC. C.PolandN.KennedyM.DyeJ.DarziA. (2017). The cutting mechanism of the electrosurgical scalpel.J. Phys. D Appl. Phys.50:025401.
65
NeumannE.RosenheckK. (1972). Permeability changes induced by electric impulses in vesicular membranes.J. Membr. Biol.10279–290. 10.1007/BF01867861
66
O’ReardonJ. P.CristanchoP.PeshekA. D. (2006). Vagus nerve stimulation (VNS) and treatment of depression: to the brainstem and beyond.Psychiatry354–63.
67
PanayiotopoulosC. P. (2011). Principles of Therapy in the Epilepsies, Vol. VIII. New York, NY: Springer, 99. 10.1007/978-0-85729-009-0
68
ParkH. J.BonmassarG.KaltenbachJ. A.MachadoA. G.ManzoorN. F.GaleJ. T. (2013). Activation of the central nervous system induced by micro-magnetic stimulation.Nat. Commun.4:2463. 10.1038/ncomms3463
69
PashutT.WolfusS.FriedmanA.LavidorM.Bar-GadI.YeshurunY.et al (2011). Mechanisms of magnetic stimulation of central nervous system neurons.PLoS Comput. Biol.7:e1002022. 10.1371/journal.pcbi.1002022
70
RenZ. (2002). T-Omega formulation for eddy-current problems in multiply connected regions.IEEE Trans. Magn.38557–560. 10.1109/20.996146
71
RothB. J.BasserP. J. (1990). A model of the stimulation of a nerve fiber by electromagnetic induction.IEEE Trans. Biomed. Eng.37588–597. 10.1109/10.55662
72
RyugoD. K.WillardF. H. (1985). The dorsal cochlear nucleus of the mouse: a light microscopic analysis of neurons that project to the inferior colliculus.J. Comp. Neurol.242381–396. 10.1002/cne.902420307
73
SaleA. J.HamiltonW. A. (1968). Effects of high electric fields on micro-organisms. 3. Lysis of erythrocytes and protoplasts.Biochim. Biophys. Acta16337–43. 10.1016/0005-2736(68)90030-8
74
SchaeferD.BourlandJ.NyenhuisJ. (2000). Review of patient safety in time-varying gradient fields.J. Magn. Reson. Imaging1220–29. 10.1002/1522-2586(200007)12:1<20::AID-JMRI3>3.0.CO;2-Y
75
SeranoP.AngeloneL. M.KatnaniH.EskandarE.BonmassarG. (2015). A novel brain stimulation technology provides compatibility with MRI.Sci. Rep.5:9805. 10.1038/srep09805
76
Sheĭkh-ZadeI. R.KruchininV. M.SukachL. I.PokrovskiĭM. V.UrmanchevaT. G. (1987). General principles of heart rate control during burst stimulation of the vagus nerve in various animals.Fiziol. Zh. SSSR Im. I M Sechenova731325–1330.
77
ShepherdR. K.ShivdasaniM. N.NayagamD. A.WilliamsC. E.BlameyP. J. (2013). Visual prostheses for the blind.Trends Biotechnol.31562–571. 10.1016/j.tibtech.2013.07.001
78
StinglK.Bartz-SchmidtK. U.BeschD.CheeC. K.CottriallC. L.GekelerF.et al (2015). Subretinal visual implant alpha IMS–clinical trial interim report.Vision Res.111(Pt B), 149–160. 10.1016/j.visres.2015.03.001
79
TolbertD. L.ConoyerB.ArielM. (2004). Quantitative analysis of granule cell axons and climbing fiber afferents in the turtle cerebellar cortex.Anat. Embryol.20949–58. 10.1007/s00429-004-0423-0
80
TuchD. S.WedeenV. J.DaleA. M.GeorgeJ. S.BelliveauJ. W. (1999). Conductivity mapping of biological tissue using diffusion MRI.Ann. N. Y. Acad. Sci.888314–316. 10.1111/j.1749-6632.1999.tb07965.x
81
TzounopoulosT.RubioM. E.KeenJ. E.TrussellL. O. (2007). Coactivation of pre- and postsynaptic signaling mechanisms determines cell-specific spike-timing-dependent plasticity.Neuron54291–301. 10.1016/j.neuron.2007.03.026
82
United States Congress Senate Committee on Finance (2006). Committee Staff Report to the Chairman and Ranking Member: Review of the FDA’s Approval Process for the Vagus Nerve Stimulation Therapy System for Treatment-Resistant Depression. S. Prt, Vol. IV. Washington, DC: U.S. Government Printing Office, 386.
83
WagnerT.GangitanoM.RomeroR.ThéoretH.KobayashiM.AnschelD.et al (2004). Intracranial measurement of current densities induced by transcranial magnetic stimulation in the human brain.Neurosci. Lett.35491–94. 10.1016/S0304-3940(03)00861-9
84
WagnerT. A.ZahnM.GrodzinskyA. J.Pascual-LeoneA. (2004). Three-dimensional head model simulation of transcranial magnetic stimulation.IEEE Trans. Biomed. Eng.511586–1598. 10.1109/TBME.2004.827925
85
WalshV.CoweyA. (2000). Transcranial magnetic stimulation and cognitive neuroscience.Nat. Rev. Neurosci.173–79. 10.1038/35036239
86
WangS. X.SunN. X.YamaguchiM.YabukamiS. (2000). Properties of a new soft magnetic material.Nature407150–151. 10.1038/35025142
87
WarmanE. N.GrillW. M.DurandD. (1992). Modeling the effects of electric fields on nerve fibers: determination of excitation thresholds.IEEE Trans. Biomed. Eng.391244–1254. 10.1109/10.184700
88
WeiX. F.GrillW. M. (2005). Current density distributions, field distributions and impedance analysis of segmented deep brain stimulation electrodes.J. Neural Eng.2139–147. 10.1088/1741-2560/2/4/010
89
WeitzA. C.NanduriD.BehrendM. R.Gonzalez-CalleA.GreenbergR. J.HumayunM. S.et al (2015). Improving the spatial resolution of epiretinal implants by increasing stimulus pulse duration.Sci. Transl. Med.7:318ra203. 10.1126/scitranslmed.aac4877
90
WilliamsN. R.ShortE. B.HopkinsT.BentzleyB. S.SahlemG. L.PannuJ.et al (2016). Five-year follow-up of bilateral epidural prefrontal cortical stimulation for treatment-resistant depression.Brain Stimul.9897–904. 10.1016/j.brs.2016.06.054
91
WoockJ. P.YooP. B.GrillW. M. (2010). Finite element modeling and in vivo analysis of electrode configurations for selective stimulation of pudendal afferent fibers.BMC Urol.10:11. 10.1186/1471-2490-10-11
92
ZrennerE. (2013). Fighting blindness with microelectronics.Sci. Transl. Med.5:210s16. 10.1126/scitranslmed.3007399
Summary
Keywords
eddy currents, TMS, finite element method, microcoils, inductive stimulation, numerical modeling, neurostimulation
Citation
Golestanirad L, Gale JT, Manzoor NF, Park H-J, Glait L, Haer F, Kaltenbach JA and Bonmassar G (2018) Solenoidal Micromagnetic Stimulation Enables Activation of Axons With Specific Orientation. Front. Physiol. 9:724. doi: 10.3389/fphys.2018.00724
Received
13 March 2018
Accepted
24 May 2018
Published
27 July 2018
Volume
9 - 2018
Edited by
Christopher Basciano, Becton Dickinson, United States
Reviewed by
Yael Yaniv, Technion – Israel Institute of Technology, Israel; Arun V. Holden, University of Leeds, United Kingdom
Updates

Check for updates
Copyright
© 2018 Golestanirad, Gale, Manzoor, Park, Glait, Haer, Kaltenbach and Bonmassar.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Giorgio Bonmassar, giorgio@nmr.mgh.harvard.edu
†These authors have contributed equally to this work.
This article was submitted to Computational Physiology and Medicine, a section of the journal Frontiers in Physiology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.