Abstract
Abnormal cardiac electromechanical coupling (cEMC) is associated with exercise-induced arrhythmias, yet monitoring cEMC during unrestricted movement remains challenging. While electrocardiogram (ECG) and phonocardiogram (PCG) recordings provide indirect measures of cardiac electrical and mechanical activity, each represents only a one-dimensional scalar observation of the cardiac system, and the system’s latent state-space dynamics response to exercise intensity remains inaccessible. Therefore, this study employs Takens’ delayed embedding phase-space reconstruction method to reconstruct the latent state trajectories and then combines them into a joint ECG-PCG phase space. Building on this representation, this study proposes dynamic graph vector autoregression (DG-VAR), a data-driven analytical approach for characterizing latent cEMC from the joint phase space of ECG and PCG during exercise. Benchmarking with an open-access database demonstrated that DG-VAR achieved the highest performance in predicting the joint phase space of ECG and PCG during exercise. The method yielded an average root mean square error (RMSE) of 0.0186 and mean absolute percentage error (MAPE) of 7.76 for ECG phase space predictions, and an average RMSE of 0.0085 and MAPE of 4.45 for PCG phase space predictions. Furthermore, matrix-derived parameters, including stationarity, coupling strength, and dynamic graph topology, showed significant correlations with cardiac electromechanical intervals. Analysis of these parameters revealed an intensity-dependent regulatory pattern in cardiac electromechanical dynamics: forward electromechanical coupling increases monotonically, the feedback mechanism rises initially and then stabilizes, and the network dynamics topology evolves towards a state of greater density and smoothness. This non-invasive, label-free, data-driven analytical framework offers a novel, physiologically interpretable approach to dynamic monitoring of exercise-induced cEMC responses.
1 Introduction
Exercise-induced arrhythmias represent a significant concern in sports medicine and exercise physiology, especially among individuals participating in competitive sports or recreational training (; ). Evidence suggests that these arrhythmias may result from transient destabilization of cardiac electromechanical coupling (cEMC) during exercise stress (). Given that exercise-induced cardiac risk manifests during physical exercise, further investigation is required to develop monitoring approaches for cEMC dynamics during exercise.
Echocardiography serves as the clinical standard for evaluating myocardial mechanical function, offering functional indicators such as ventricular wall motion, ejection fraction, and strain. When combined with electrocardiography (ECG), it enables assessment of cEMC parameters, including the time to peak during the systolic phase (; ). However, echocardiography requires the probe to be held firmly against the chest wall, which complicates data acquisition and limits its application for studying cEMC during exercise (). In contrast, the phonocardiogram (PCG) offers a noninvasive marker of mechanical contraction by capturing low-frequency acoustic signals generated by valvular closure and myocardial vibration (). The first (S1) and second (S2) heart sounds indicate key mechanical events: the onset of ventricular contraction and the closure of the aortic valve, respectively. Therefore, combining ECG and PCG enables inference of cardiac electromechanical intervals, including Q-S1, Q-S2, and S1-S2 intervals (; ). Multimodal fusion of synchronously recorded ECG and PCG has been successfully applied to cardiovascular monitoring and disease classification (; ). Nevertheless, both signals represent indirect observations and cannot independently reconstruct the coupling between electrical activity and mechanical contraction (). Thus, acquiring cEMC dynamic characteristics from ECG and PCG signals remains a significant challenge.
From a systems science perspective, cEMC can be characterized as a dynamical system that integrates electrical excitation with mechanical contraction. In this view, state-space modeling provides a natural framework for describing such systems by inferring latent-state trajectories from observed sequences (). Within this framework, ECG and PCG serve as noisy, partial observations of the system, whereas critical determinants, including conduction velocity, calcium transients, and ventricular pressure, constitute latent states that are not directly measurable and must be inferred (). However, classical state-space models typically rely on predefined transition functions with fixed parameters, which may be inadequate for describing the high-order and time-varying dynamics of cEMC during incremental exercise (, ). Consequently, reconstructing the dynamic structure of cEMC from limited observations without explicit physiological constraints is necessary.
Takens’ delay embedding theorem provides a theoretical foundation for reconstructing latent dynamical structures from observed signals. For physiological systems governed by deterministic nonlinear dynamics, delay-coordinate embedding has demonstrated that a phase space equivalent to the original system can be reconstructed from time-delayed observations of the measured variables (). Although physiological systems are inevitably influenced by noise, perturbations, and regulatory variability, ECG and PCG signals may still preserve essential dynamical information about cardiac electromechanical activity, which can approximate the latent phase-space trajectory of cEMC through delay-coordinate reconstruction (). Therefore, integrating Takens’ delay embedding with state-transition models offers a potential approach to characterizing cEMC dynamics using non-invasive physiological measurements. However, traditional linear state-transition models, such as vector autoregression (VAR), typically assume time-invariant coefficient matrices, which may not adequately describe the continuously adapting nature of cardiovascular regulation during incremental exercise (). Therefore, it is necessary to utilize parameter estimation approaches that enable the matrix coefficients to evolve over time to capture the dynamic characteristics of cEMC. Recent advances in artificial intelligence techniques have enabled the modeling of time-varying dynamical systems using data-driven approaches (). Neural network-based methods estimate nonlinear state-transition functions through flexible function approximation, providing powerful representations of complex temporal dynamics (). However, such approaches generally focus on predictive performance and provide limited insight into the dynamic structure among physiological subsystems. By contrast, graph-based modelling offers a structurally interpretable framework by representing interacting variables as nodes and their time-varying dependencies as evolving edges, thereby enabling the direct characterization of dynamic coupling patterns (). Nonetheless, dynamic graph approaches have not yet been explored for modelling cardiac electromechanical coupling during exercise.
Building on these considerations, this study attempts to investigate a data-driven approach to extract the latent dynamical characteristics of cEMC from limited observations during incremental exercise. First, considering the physiological characteristics of cEMC during incremental exercise, the Takens delayed embedding method is used to reconstruct the phase space of the ECG and PCG. Then, the joint phase-space dynamics of the ECG and PCG are treated as state variables, and a VAR model is used to characterize local dynamics as a locally linear approximation of the nonlinear system. Given the time-varying nature of cEMC during exercise, a time-varying parameter estimation strategy is incorporated, informed by dynamic graph-based modeling. The findings are expected to facilitate the extraction of dynamical features characterizing cEMC responses to incremental exercise, thereby providing a methodological basis for assessing exercise-related cardiovascular risk.
2 Materials and methods
2.1 Datasets
This study used the EPHNOGRAM dataset to evaluate the effectiveness of the proposed analysis approach (). This dataset includes synchronized ECG and PCG data from 24 healthy subjects (age: 25.4 ± 1.9 years) during rest, bicycle ergometer testing, and Bruce protocol treadmill incremental exercise testing. All signals were simultaneously recorded using a three-lead ECG device and a single PCG stethoscope at a sampling frequency of 8000 Hz. This dataset was approved by Shiraz University’s Biomedical Engineering Review Board. Since the ECG and PCG signals from the Bruce protocol treadmill stress test contained numerous artifacts and exhibited poor data quality, this study selected only data from bicycle ergometer tests of 10 subjects for subsequent analysis, with each subject contributing a 30-minute recording.
Previous studies have demonstrated that by extracting the R-peak to R-peak interval (RRI) from a 2-minute ECG segment and applying detrended fluctuation analysis (DFA), the DFA α1 index can be obtained and used to evaluate exercise intensity (). Specifically, during low-, moderate-, and high-intensity exercise, the DFA α1 values are approximately 1.0, 0.75, and ≤0.5, respectively (). Therefore, this study stratified DFA α1 into four grades (Zone 0, Zone 1, Zone 2, Zone 3) using the aforementioned critical thresholds.
2.2 Signal processing
Given the significant differences in the physical characteristics of ECG and PCG signals, and the widespread presence of noise, including baseline drift, electromyographic interference, and motion artifacts, in the raw signals, the signals were preprocessed before modelling. The preprocessing procedures primarily include sliding-window segmentation, signal denoising, envelope extraction, outlier cleaning, extraction of cardiac electromechanical intervals, and DFA analysis. The detail can be found in the Supplementary File.
2.3 Phase space reconstruction
Since raw ECG and PCG recordings, as one-dimensional scalar time series, do not fully represent the nonlinear dynamics of the cardiac dynamical system, the recordings are resampled to 1000 Hz and their phase spaces are reconstructed using Takens’ delay-embedding theorem, resulting in state vectors . The delay time τ is determined by the first local minimum of the mutual information function, thereby minimizing redundancy across embedding dimensions (). The embedding dimension m is selected using the false nearest neighbor (FNN) algorithm as the smallest m for which the FNN ratio falls below 1% (). Optimal values of τ () and m () for each signal are initially identified through a grid search across all segments. Finally, the individual ECG and PCG state vectors are concatenated to form a single ns-dimensional (ns is the sum of mecg and mpeg) joint state vector sk for subsequent modeling analysis.
2.4 Model architectures
The VAR and dynamic graph were integrated to construct a dynamic graph-based VAR (DG-VAR) model for dynamic analysis of cEMC. The effectiveness of this modeling approach was assessed through a comparative analysis of the DG-VAR model, traditional VAR models, and the time-varying extension (TV-VAR).
2.4.1 VAR
The dynamic model of the joint state vector sk was constructed using the VAR. The mathematical expression of the model is as shown in Equation 1:
Where is the time-invariant coefficient matrix, held constant within each sliding window. ϵk represents independently and identically distributed random noise terms. Takens embedding supplies a locally linear, noise-projected representation of the cardiac dynamics, but it does not render the observed series weakly stationary. Since the VAR coefficient estimation requires weak stationarity to avoid spurious regression bias, this study applies differencing operations to the input dataset. To distinguish the self-evolution patterns of ECG and PCG as well as their coupling relationship, the coefficient matrix is decomposed into blocks as shown in the Equation 2
where and describe the autoregressive inertia of PCG and ECG. denotes the electromechanical forward pathway by which cardiac electrical activity influences the subsequent contraction of the myocardium. In contrast, corresponds to the feedback pathway in which myocardial mechanical contraction affects cardiac electrical activity. Given the state sequence of the training set be , the lag and the target matrices are as shown in Equation 3:
The coefficient matrix was estimated using ridge regression, which avoids singularity and suppresses overfitting. The loss function is as shown in Equations 4-6:
Where denoted the reconstruction term and denoted the L2 regulaization. The L2 regularization coefficient was set to 10–4 as determined by cross validation, which corresponds to the optimal intersection point of the prediction error plateau and the dynamic stability.
2.4.2 TV-VAR
This study extends the VAR model by incorporating time-varying parameter estimation. Specifically, the coefficient matrix in the VAR model is replaced with a sequence of time-varying matrices , while the disturbance covariance matrix is also allowed to be time-varying. The mathematical expression of the model is shown in Equation 7:
Both and are generated using a two-layer fully connected neural network with a local history window of length 5 and the hyperbolic tangent activation function. Meanwhile, a time-varying smoothness regularization term is embedded into the loss function to penalize abrupt parameter variations, thereby ensuring the continuous evolution of cardiac dynamic. The loss function is as shown in the Equation 8.
where is the L2 regularization coefficient and is the time-varying smoothness regularization coefficient, both set to 10-4. The network is trained using the Adam optimizer with a learning rate of 10–2 for 100 epochs.
2.4.3 DG-VAR
Since the TV-VAR model employs a fully dense coefficient matrix that assumes default coupling among all state nodes, it fails to distinguish genuine physiological causal connections from noise-induced spurious correlations (). To address this limitation, this study introduces a range constraint and L1 sparsity regularization to transform the coefficient matrix into the adjacency matrix of a dynamic graph, thereby enabling the automatic selection of valid causal pathways (). The mathematical expression of the model is as shown in Equations 9 and Equation 10:
Where Ak denotes the time-varying adjacency matrix at time k, Wadj and badj represent the output-layer parameters, and hk is the hidden-layer feature representation. The hyperbolic tangent activation constrains matrix elements to [−1,1], where positive, negative, and zero values indicate promoting, inhibitory, and absent connections, respectively. An additional L1 regularization term is incorporated into the loss function to promote sparse coupling patterns by shrinking insignificant connections toward zero. The loss function is shown in Equation 11.
Where is the graph sparsity regularization coefficient (In this study, was set to 0.01). The is the L1 sparsity regularization term, which drives the weights of ineffective couplings toward zero by computing the mean of the L1 norms of the adjacency matrices Ak over all time steps. The calculation equation for is shown in Equation 12.
In Equation 12, the L1 regularization is equivalent to imposing a Laplace prior on the adjacency matrix, enabling automatic selection of sparse causal pathways.
2.5 Model evaluation
Each signal segment was split chronologically into training and test sets, with the initial 80% allocated to training and the remaining 20% retained for testing. The root mean square error (RMSE) and mean absolute percentage error (MAPE) were subsequently computed on test-set samples to quantitatively evaluate the model’s predictive performance.
2.6 Dynamics characteristics
The cEMC dynamics were characterized from three perspectives: stationarity, coupling strength, and dynamic graph topology.
2.6.1 Stationarity indicators
Stationarity was assessed by calculating the spectral radius of the coefficient submatrix (). The spectral radius of any given submatrix A is defined in Equation 13:
where denotes the i-th eigenvalue of submatrix A. Temporal variations in subsystem stability were evaluated by the average of the spectral radius sequence. The subsystem spectral radius was independently calculated for both ECG and PCG to characterize their intrinsic autoregressive dynamics. A spectral radius approaching 1 indicates a stronger autoregressive effect, while values exceeding 1 suggest potential local divergence.
2.6.2 Coupling strength indicators
Metrics were derived from the cross-channel coefficient submatrices () to quantify the strength of directional coupling. First, the average coupling magnitude was calculated as the mean absolute value of all matrix elements, representing the overall interaction intensity (). Second, the cross-channel synergistic strength was defined as the mean absolute value of the off-diagonal coefficients, which specifically reflects information exchange between different signal dimensions (). Taking as an example, the average coupling magnitude, and cross-channel synergistic strength are calculated using Equations 14, 15.
Where |O| denotes the cardinality of set O, corresponding to the total count of off-diagonal entries in the matrix. The index set collects all ordered row-column index pairs associated with the matrix’s off-diagonal elements. The bidirectional coupling strength was determined by integrating both forward and feedback coupling pathways. It is defined as Equation 16:
Where is the Frobenius norm. E is the bidirectional coupling energy.
2.6.3 Dynamic graph topology indicators
Based on the sparse causal graph adjacency matrix sequence , graph sparsity, average edge strength, and graph evolution smoothness were extracted. Graph sparsity was defined as the proportion of weak connections below a predefined threshold (), reflecting the structural sparsity of the causal network (). Average edge strength quantified the global interaction intensity among state variables (). Graph evolution smoothness measured temporal variations in network topology and coupling weights between adjacent time points, with lower values indicating more stable network evolution (). The above dynamic graph topology metrics were calculated using Equations 17–19.
Where T’ is the total number of effective time windows. ns is the dimension of the joint state vector. is the coupling weight in the i-th row and j-th column of the causal matrix at time k, and is the weak connection threshold. denotes the Frobenius norm, and is the number of elements satisfying the condition.
2.7 Statistical analysis
In this study, continuous data were described statistically using the median and the 25th and 75th percentiles. Welch’s one-way analysis of variance (ANOVA) was used to analyze differences in model parameters across different exercise intensity levels. Generalized mixed-effects regression was performed to analysis model parameters associated with cEMC parameters. All statistical computations were conducted via Jamovi software (version 1.8.1). A two-tailed p-value < 0.05 was considered statistically significant for all tests.
3 Results
This study demonstrates that the DG-VAR model exhibits significantly superior fitting and prediction performance for the ECG and PCG phase spaces than the VAR and TV-VAR models. The error analysis shows that in ECG phase-space forecasting, DG-VAR achieves a 58.8% reduction in median RMSE and a 60.5% reduction in median MAPE compared with VAR, and a 17.0% reduction in median RMSE and a 23.6% reduction in median MAPE relative to TV-VAR. Similarly, DG-VAR reduces the median RMSE and median MAPE for PCG prediction by 74.2% and 74.8% relative to VAR, and by 37.5% and 43.0% relative to TV-VAR. Further analysis stratified by exercise intensity shows that DG-VAR yields the lowest error in predicting both ECG and PCG signals, and its accuracy remains robust across all exercise intensity zones (Table 1). Although DG-VAR exhibits elevated errors for both signals in Zone 3, these errors are still smaller than those of VAR and TV-VAR. This finding suggests that DG-VAR exhibits robust performance advantages across different exercise intensities.
Table 1
| Signal | Exercise intensity | VAR | TV-BAR | DG-VAR | |||
|---|---|---|---|---|---|---|---|
| RMSE | MAPE | RMSE | MAPE | RMSE | MAPE | ||
| ECG | Overall | 0.0452 (0.0355, 0.0532) | 19.67 (14.50, 27.01) | 0.0224 (0.0145, 0.0289) | 10.16 (6.67, 14.35) | 0.0186 (0.0096, 0.0252) | 7.76 (4.63, 11.69) |
| Zone 0 | 0.0452 (0.0357, 0.0535) | 18.77 (14.46, 26.59) | 0.0225 (0.0151, 0.0289) | 10.12 (6.70, 14.05) | 0.0187 (0.0106, 0.0254) | 7.81 (4.83, 11.19) | |
| Zone 1 | 0.0356 (0.0251, 0.0469) | 18.07 (12.96, 26.30) | 0.0146 (0.0092, 0.0207) | 7.19 (4.15, 12.43) | 0.0087 (0.0036, 0.0190) | 3.38 (1.66, 8.89) | |
| Zone 2 | 0.0517 (0.0427, 0.0523) | 20.73 (17.71, 25.69) | 0.0273 (0.0192, 0.0303) | 10.16 (7.41, 15.73) | 0.0236 (0.0129, 0.0265) | 7.56 (4.57, 12.89) | |
| Zone 3 | 0.0478 (0.0436, 0.0553) | 29.28 (20.85, 35.15) | 0.0264 (0.0234, 0.0325) | 17.93 (12.47, 20.60) | 0.0197 (0.0170, 0.0282) | 15.00 (8.09, 18.13) | |
| PCG | Overall | 0.0329 (0.0267, 0.0417) | 17.65 (13.23, 22.43) | 0.0136 (0.0112, 0.0185) | 7.81 (5.68, 9.86) | 0.0085 (0.0059, 0.0116) | 4.45 (2.69, 6.27) |
| Zone 0 | 0.0312 (0.0260, 0.0397) | 17.61 (12.88, 22.83) | 0.0132 (0.0109, 0.0178) | 7.77 (5.49, 9.72) | 0.008 (0.0058, 0.0110) | 4.21 (2.52, 6.22) | |
| Zone 1 | 0.0395 (0.0375, 0.0442) | 16.64 (14.60, 19.60) | 0.0169 (0.0155, 0.0191) | 6.78 (6.44, 9.47) | 0.0114 (0.0094, 0.0128) | 4.15 (4.03, 6.10) | |
| Zone 2 | 0.0389 (0.0325, 0.0418) | 20.42 (13.90, 21.13) | 0.0166 (0.0147, 0.0193) | 7.85 (7.03, 10.73) | 0.0095 (0.009, 0.0122) | 4.77 (4.06, 6.43) | |
| Zone 3 | 0.0395 (0.0348, 0.0500) | 17.92 (16.33, 23.58) | 0.0187 (0.0155, 0.0248) | 9.47 (9.02, 13.05) | 0.0115 (0.0085, 0.0141) | 4.97 (4.56, 6.74) | |
The predictive performance of different models in estimating ECG and PCG.
Welch’s one-way ANOVA identified significant differences among exercise intensity levels for all matrix-derived parameters (all p < 0.001). The intra-individual variations of the model-derived parameters are presented on the secondary axis of Figure 1. The spectral radius of Axx demonstrates a J-shaped increase with exercise intensity, while the spectral radius of Ayy follows an inverse pattern. Forward coupling submatrix parameters, including the spectral radius, average magnitude, and synergistic strength of Ayz, increase in a consistent manner with exercise intensity. In contrast, the spectral radius of the feedback coupling submatrix exhibits a logarithmic growth curve trend. The bidirectional coupling energy increases monotonically across exercise intensity zones. Graph sparsity exhibits an initial increase followed by a decrease as exercise intensity. The average edge weight of dynamic graphs follows a J-shaped growth trajectory, whereas smoothness displays an exponential growth trend. The relationship between the model-derived and cEMC parameters as shown in Table 2.
Figure 1
Table 2
| Predictors | Q-S1 interval | Q-S2 interval | S1-S2 interval | |||
|---|---|---|---|---|---|---|
| β (95% CI) | p-value | β (95% CI) | p-value | β (95% CI) | p-value | |
| Spectral Radius of Ayy | 0.112 (0.042, 0.183) | 0.002 | 0.574 (0.509, 0.639) | <0.001 | 0.706 (0.640, 0.773) | <0.001 |
| Spectral Radius of Axx | 0.600 (0.488, 0.712) | <0.001 | 1.597 (1.492, 1.701) | <0.001 | 1.779 (1.673, 1.886) | <0.001 |
| Spectral Radius of Ayz | -0.254 (-0.376, -0.131) | <0.001 | -0.075 (-0.189, 0.039) | 0.196 | 0.074 (-0.042, 0.191) | 0.212 |
| Spectral Radius of Axy | -0.095 (-0.207, 0.017) | 0.098 | 0.093 (-0.011, 0.197) | 0.080 | 0.096 (-0.010, 0.203) | 0.077 |
| Average Coupling Magnitude of Ayz | 0.060 (-0.171, 0.291) | 0.612 | 0.004 (-0.210, 0.218) | 0.971 | -0.128 (-0.347, 0.091) | 0.252 |
| Average Coupling Magnitude of Axy | -0.487 (-0.697, -0.276) | <0.001 | 0.102 (-0.094, 0.297) | 0.307 | 0.557 (0.358, 0.757) | <0.001 |
| Synergistic Strength of Ayz | 0.341 (0.166, 0.516) | <0.001 | 0.609 (0.446, 0.771) | <0.001 | 0.629 (0.463, 0.795) | <0.001 |
| Synergistic Strength of Axy | 0.759 (0.602, 0.915) | <0.001 | 0.394 (0.248, 0.539) | <0.001 | 0.002 (-0.146, 0.151) | 0.975 |
| Bidirectional Coupling Energy | -0.844 (-1.017, -0.670) | <0.001 | -1.742 (-1.903, -1.581) | <0.001 | -1.807 (-1.972, -1.643) | <0.001 |
| Median of Graph Sparsity | -0.302 (-0.356, -0.248) | <0.001 | -0.394 (-0.444, -0.344) | <0.001 | -0.313 (-0.363, -0.262) | <0.001 |
| Median of Graph Edge | -0.910 (-1.036, -0.784) | <0.001 | -1.805 (-1.922, -1.688) | <0.001 | -1.853 (-1.973, -1.734) | <0.001 |
| Median of Graph Smoothness | 0.197 (0.165, 0.230) | <0.001 | 0.089 (0.059, 0.119) | <0.001 | 0.004 (-0.027, 0.034) | 0.809 |
The generalized mixed-effects regression results between model-derived parameters and cEMC parameters.
4 Discussion
In state-space modeling, system dynamics are represented by a coefficient matrix that functions as a transition operator, describing the evolution of system states over time (). VAR, as a simplified variant of these models, is widely used for physiological signal analysis. However, conventional VAR assumes a constant transition operator, resulting in a stationary linear mapping of system interactions (). This assumption may not adequately represent cEMC during exercise, in which autonomic regulation, hemodynamic responses, and myocardial contractility continuously adapt to increasing physiological demands (). Consequently, the coupling dynamics are expected to evolve, requiring temporal adaptation of the linear operator. Time-varying coefficient matrices allow the operator to adapt dynamically (). However, without structural constraints, temporal variation in the coefficient matrix introduces excessive degrees of freedom, increasing the risk of fitting non-physiological fluctuations as coupling alterations, especially in the presence of motion artefacts or measurement noise, which can compromise model generalizability. Dynamic graph modeling provides a potential solution to address these limitations by introducing structural constraints into the evolution of the coefficient matrix. Specifically, the distribution of non-zero coefficients is modeled as a time-varying network topology, representing the changing organization of effective interactions among system states (). Instead of assuming full connectivity, this approach assumes that only a limited subset of interactions contributes to system evolution at each time point. Such sparsity constraints reduce the effective dimensionality of the parameter space and restrict parameter estimation to informative connections, thereby mitigating the influence of non-physiological fluctuations (). Following this principle, the present study integrates dynamic graph topology constraints into time-varying VAR structures to develop DG-VAR. Error analysis demonstrates the effectiveness of this modeling approach, showing that the model simultaneously supports adaptive temporal coupling evolution and topology-driven sparse regularization, thereby overcoming the dual limitations of time-invariance and lack of structural constraints inherent in traditional fixed VAR transfer operators.
Rather than treating cardiac electrical and mechanical activity as independent processes, DG-VAR characterizes cEMC as an evolving bidirectional interaction within their joint phase space, with the time-varying cross-block coefficients capturing the dynamic influences from electrical to mechanical activity and vice versa. In this study, the matrix-derived parameters of DG-VAR exhibited significant correlations with the cardiac electromechanical interval, suggesting their ability to reflect exercise-induced variations in cEMC. From a dynamical systems perspective, these parameters describe cEMC adaptation, including the persistence of system-state evolution, the strength of electromechanical interactions, and the adaptive organization of cEMC network dynamics. The spectral radius characterizes the persistence of system-state evolution and reflects the dynamic stability of cEMC in response to exercise intensity (). Decomposition of the coefficient matrix revealed distinct adaptive patterns among functional pathways. The spectral radius of the electromechanical forward coupling submatrix increased progressively with exercise intensity, which suggests that transmission from electrical activation to mechanical contraction was enhanced during periods of increased cardiac demand (). Conversely, the spectral radius of the mechanoelectrical feedback coupling submatrix appears to exhibit a saturation effect. Although feedback modulation increases within the range of low to moderate exercise intensity, the coupling strength does not increase correspondingly under higher-intensity stimulation (). Beyond the state stability, the coupling strength and graph topology revealed the progressive reorganization of cEMC interactions. Increased bidirectional coupling energy reflected enhanced information exchange between electrical and mechanical activity, which was further supported by their inverse associations with the Q-S1, Q-S2, and S1-S2 intervals. Meanwhile, graph-based analysis demonstrated decreased network sparsity and increased edge strength with increasing exercise intensity, suggesting a transition toward a more densely interconnected cEMC. The exponential increase in the smoothness of graph evolution further suggested enhanced temporal organization of cEMC under elevated physiological demands (). Collectively, these findings suggest that exercise-induced cEMC adaptation emerges from coordinated adjustments in state persistence, electromechanical interactions, and adaptive organization of network dynamics, rather than isolated alterations in individual cardiac components.
Despite these findings being promising, several limitations should be acknowledged. Firstly, the identified cEMC characteristics are mathematical latent features derived from computational models rather than direct physiological measurements. The EPHNOGRAM dataset lacks relevant echocardiographic measures, such as electromechanical intervals and ventricular mechanical synchrony, precluding direct validation against myocardial electrical activation and mechanical contraction. Therefore, the physiological significance of these model-derived features requires validation using independent physiological measurements in future studies. Secondly, the limited sample size and population homogeneity may restrict the generalizability of the findings. The EPHNOGRAM dataset exclusively comprises young, healthy male individuals, and further research is necessary to ascertain whether the identified cEMC dynamics are consistent across populations. Thirdly, the evaluation of exercise intensity was constrained by the lack of metabolic measurements. While DFA α1 reflects non-linear autonomic dynamics associated with exercise stress, it does not directly correspond to metabolic load as measured by VO2, which is considered the gold standard for assessing exercise intensity (). The absence of VO2 data in the EPHNOGRAM dataset precludes independent calibration of exercise intensity based on metabolic demand. Future research could validate the relationship between exercise intensity and cEMC dynamic characteristics further by incorporating metabolic load measures. Finally, the selection of delay estimation strategy can affect the dynamics of the reconstructed phase space, which is a methodological limitation of this study. The sensitivity of cEMC dynamic reconstruction to different time delay estimation strategies is an important question for future research.
5 Conclusion
In summary, the DG-VAR model proposed in this study achieves outstanding performance in predicting the joint phase space of ECG and PCG signals during incremental exercise. Across all exercise intensity levels, the model consistently achieves superior prediction accuracy and robustness compared to conventional VAR and its time-varying extension. Analysis of matrix-derived parameters reveals an intensity-dependent regulatory pattern in cardiac electromechanical dynamics: the forward electromechanical coupling increases monotonically, the feedback mechanism exhibits a logarithmic growth curve trajectory, and the network dynamics topology progressively evolves toward denser and smoother configurations. These dynamic characteristics show significant correlations with the Q-S1, Q-S2, and S1-S2 intervals. This non-invasive, label-free, data-driven analytical framework offers a novel, physiologically interpretable approach to dynamic monitoring of exercise-induced cEMC responses.
Statements
Data availability statement
Publicly available datasets were analyzed in this study. This data can be found here: https://physionet.org/content/ephnogram/1.0.0/.
Author contributions
XY: Project administration, Methodology, Formal analysis, Conceptualization, Software, Writing – original draft, Funding acquisition, Writing – review & editing. QL: Writing – review & editing, Formal analysis, Software. QYL: Formal analysis, Software, Writing – review & editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This work was supported by the Natural Science Foundation of Xiamen, China (Grant No. 3502Z202473079).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. During the preparation of this work the author(s) used Doubao and ChatGPT in order to search for information, and polish language. After using this tool/service, the author(s) reviewed and edited the content as needed and take(s) full responsibility for the content of the publication.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fphys.2026.1946312/full#supplementary-material
References
1
AbbasF.McclellandV.CvetkovicZ.DaiW. (2025). Stationary and sparse denoising approach for corticomuscular causality estimation. IEEE Trans. Biomed. Eng.72, 1697–1707. doi: 10.1109/tbme.2024.3518602
2
AntonacciY.BaràC.ZaccaroA.FerriF.PerniceR.FaesL. (2023). Time-varying information measures: an adaptive estimation of information storage with application to brain-heart interactions. Front. Network. Physiol.3, 1242505. doi: 10.3389/fnetp.2023.1242505
3
AntonacciY.BaráC.SparacinoL.MijatovicG.MinatiL.FaesL. (2025). A method for the time-frequency analysis of high-order interactions in non-stationary physiological networks. J. Neural Eng.22, 066001. doi: 10.1088/1741-2552/ae15c0
4
BarnettL.SethA. K. (2015). Granger causality for state-space models. Phys. Rev. E. – Statistical. Nonlinear. Soft. Matter. Physics.91, 40101. doi: 10.1103/PhysRevE.91.040101
5
CharisopoulouD.KoulaouzidisG.RydbergA.HeneinM. Y. (2019). Exercise worsening of electromechanical disturbances: a predictor of arrhythmia in long QT syndrome. Clin. Cardiol.42, 701. doi: 10.1002/clc.23216
6
ChenZ.BrownE. N.BarbieriR. (2010a). Characterizing nonlinear heartbeat dynamics within a point process framework. IEEE Trans. Biomed. Eng.57, 1335–1347. doi: 10.1109/tbme.2010.2041002
7
ChenZ.PurdonP. L.BrownE. N.BarbieriR. (2010b). A differential autoregressive modeling approach within a point process framework for non-stationary heartbeat intervals analysis. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc2010, 3567–3570. doi: 10.1109/iembs.2010.5627462
8
ChristiniD. J.KulkarniA.RaoS.StutmanE. R.BennettF. M.HausdorffJ. M.et al. (1995). Influence of autoregressive model parameter uncertainty on spectral estimates of heart rate dynamics. Ann. Biomed. Eng.23, 127–134. doi: 10.1007/bf02368320
9
D'ascenziF.Sanz-De La GarzaM.AnselmiF.NunnoL.ArbeloE.JordàP.et al. (2020). Electromechanical delay by speckle-tracking echocardiography: a novel tool to distinguish between Brugada syndrome and isolated right bundle branch block. Int. J. Cardiol.320, 161–167. doi: 10.1016/j.ijcard.2020.06.029
10
FaesL.ErlaS.NolloG. (2012). Measuring connectivity in linear multivariate processes: definitions, interpretation, and practical analysis. Comput. Math. Methods Med.2012, 140513. doi: 10.1155/2012/140513
11
FraserA. M.SwinneyH. L. (1986). Independent coordinates for strange attractors from mutual information. Phys. Rev. A. Gen. Phys.33, 1134–1140. doi: 10.1103/physreva.33.1134
12
Garcia-RetortilloS.Ch IvanovP. (2025). Dynamics of cardio-muscular networks in exercise and fatigue. J. Physiol.603, 5121–5147. doi: 10.1113/jp286963
13
GottwaldG. A.ReichS. (2021). Combining machine learning and data assimilation to forecast dynamical systems from noisy partial observations. Chaos.31, 101103. doi: 10.1063/5.0066080
14
GronwaldT.RogersB.HoosO. (2020). Fractal correlation properties of heart rate variability: a new biomarker for intensity distribution in endurance exercise and training prescription? Front. Physiol.11, 550572. doi: 10.3389/fphys.2020.550572
15
GuoY.YangH.GuoT.PanJ.WangW. (2022). A novel heart sound segmentation algorithm via multi-feature input and neural network with attention mechanism. Biomed. Phys. Eng. Express.9, 015012. doi: 10.1088/2057-1976/ac9da6
16
HallE. C.RaskuttiG.WillettR. M. (2019). Learning high-dimensional generalized linear autoregressive models. IEEE Trans. Inf. Theory65, 2401–2422. doi: 10.1109/tit.2018.2884673
17
HarrisK. M.Mackey-BojackS.BennettM.NwaudoD.DuncansonE.MaronB. J. (2021). Sudden unexpected death due to myocarditis in young people, including athletes. Am. J. Cardiol.143, 131–134. doi: 10.1016/j.amjcard.2020.12.028
18
HatanoA.IzuL. T.Chen-IzuY.SatoD. (2025). Modeling autoregulation of cardiac excitation-Ca-contraction and arrhythmogenic activities in response to mechanical load changes. iScience.28, 111788. doi: 10.1016/j.isci.2025.111788
19
KazemnejadA.KarimiS.GordanyP.CliffordG. D.SameniR. (2024). An open-access simultaneous electrocardiogram and phonocardiogram database. Physiol. Meas.45, 055005. doi: 10.1088/1361-6579/ad43af
20
KennelM. B.BrownR.AbarbanelH. D. (1992). Determining embedding dimension for phase-space reconstruction using a geometrical construction. Phys. Rev. A.45, 3403–3411. doi: 10.1103/physreva.45.3403
21
KoppeG.ToutounjiH.KirschP.LisS.DurstewitzD. (2019). Identifying nonlinear dynamical systems via generative recurrent neural networks with applications to fMRI. PloS Comput. Biol.15, e1007263. doi: 10.1371/journal.pcbi.1007263
22
KostoglouK.RobertsonA. D.MacintoshB. J.MitsisG. D. (2019). A novel framework for estimating time-varying multivariate autoregressive models and application to cardiovascular responses to acute exercise. IEEE Trans. Biomed. Eng.66, 3257–3266. doi: 10.1109/tbme.2019.2903012
23
LangR. M.BadanoL. P.Mor-AviV.AfilaloJ.ArmstrongA.ErnandeL.et al. (2015). Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J. Am. Soc. Echocardiography.28, 1–39.e14. doi: 10.1016/j.echo.2014.10.003
24
LiX.HouY.ShiK.CaiY. (2025). TF-crossnet: a cross-modal attention fusion network for cardiovascular disease classification using pcg and ecg signals. Biomed. Phys. Eng. Express.11, 065048. doi: 10.1088/2057-1976/ae1a8b
25
MadhavK. V.KrishnaE. H.ReddyK. A. (2024). Improved bivariate-VAR model for extraction of respiratory information from artifact corrupted ECG and PPG signals. New Gener. Comput.42, 859–877. doi: 10.1007/s00354-024-00265-2
26
MarzbanradF.YaghmaieN.JelinekH. F. (2020). A framework to quantify controlled directed interactions in network physiology applied to cognitive function assessment. Sci. Rep.10, 18505. doi: 10.1038/s41598-020-75466-y
27
MeiH.WangZ.YangH.LiX.XuY. (2025). Network analysis of multivariate time series data in biological systems: methods and applications. Briefings Bioinf.26, bbaf223. doi: 10.1093/bib/bbaf223
28
MeyerT.LucíaA.EarnestC. P.KindermannW. (2005). A conceptual framework for performance diagnosis and training prescription from submaximal gas exchange parameters--theory and application. Int. J. Sports. Med.26, S38–S48. doi: 10.1055/s-2004-830514
29
ModyS. K.RangarajanG. (2022). Sparse representations of high dimensional neural data. Sci. Rep.12, 7295. doi: 10.1038/s41598-022-10459-7
30
PolyákA.TopalL.Zombori-TóthN.TóthN.ProrokJ.KohajdaZ.et al. (2023). Cardiac electrophysiological remodeling associated with enhanced arrhythmia susceptibility in a canine model of elite exercise. Elife.12, e80710. doi: 10.7554/eLife.80710
31
SawadaS. G.ThomaidesA. (2009). Three-dimensional stress echocardiography: the promise and limitations of volumetric imaging. Curr. Opin. Cardiol.24, 426–432. doi: 10.1097/HCO.0b013e32832c76db
32
SchiattiL.NolloG.RossatoG.FaesL. (2015). Extended Granger causality: a new tool to identify the structure of physiological networks. Physiol. Meas.36, 827–843. doi: 10.1088/0967-3334/36/4/827
33
ShojaieA.FoxE. B. (2022). Granger causality: a review and recent advances. Annu. Rev. Stat. Appl.9, 289–319. doi: 10.1146/annurev-statistics-040120-010930
34
SpringerD. B.TarassenkoL.CliffordG. D. (2016). Logistic regression-HSMM-based heart sound segmentation. IEEE Trans. Biomed. Eng.63, 822–832. doi: 10.1109/tbme.2015.2475278
35
SunC.LiuC.WangX.LiuY.ZhaoS. (2024). Coronary artery disease detection based on a novel multi-modal deep-coding method using ECG and PCG signals. Sensors. (Basel).24, 6939. doi: 10.3390/s24216939
36
TibbitsG. F.BarnardR. J.BaldwinK. M.CugaljN.RobertsN. K. (1981). Influence of exercise on excitation-contraction coupling in rat myocardium. Am. J. Physiol.240, H472–H480. doi: 10.1152/ajpheart.1981.240.4.H472
37
TranK. T.TranT. N.HuynhD. N.LeN. K.LeC. D.MaiH. X.et al. (2025). A multimodal system for comprehensive cardiovascular monitoring using ECG, PCG, and PPG signal fusion. Sensors. (Basel).25, 6708. doi: 10.3390/s25216708
38
WanH.-Y.BunsawatK.AmannM. (2023). Autonomic cardiovascular control during exercise. Am. J. Physiol. Heart Circ. Physiol.325, H675–H686. doi: 10.1152/ajpheart.00303.2023
39
WangW.YuanJ.WangB.FangY.ZhengY.HuX. (2022). Robust classification of heart valve sound based on adaptive EMD and feature fusion. PloS One17, e0276264. doi: 10.1371/journal.pone.0276264
40
WehmuthK.ZivianiA.FleuryE. (2015). “ A unifying model for representing time-varying graphs”, in: 2015 IEEE International Conference on Data Science and Advanced Analytics (DSAA). (Paris: IEEE (Institute of Electrical and Electronics Engineers)), 1–10. doi: 10.1109/DSAA.2015.7344810
41
YuR.WangR. (2024). Learning dynamical systems from data: an introduction to physics-guided deep learning. Proc. Natl. Acad. Sci.121, e2311808121. doi: 10.1073/pnas.2311808121
42
ZiaJ.KimballJ.RozellC.InanO. T. (2021). Harnessing the manifold structure of cardiomechanical signals for physiological monitoring during hemorrhage. IEEE Trans. Biomed. Eng.68, 1759–1767. doi: 10.1109/tbme.2020.3014040
Summary
Keywords
cardiac electromechanical coupling, dynamic graph, electrocardiogram, incremental exercise, phonocardiogram, vector autoregression
Citation
Ye X, Lu Q and Lin Q (2026) Dynamic graph vector autoregression for characterizing cardiac electromechanical dynamics during incremental exercise from the joint phase space of electrocardiogram and phonocardiogram. Front. Physiol. 17:1946312. doi: 10.3389/fphys.2026.1946312
Received
23 July 2026
Revised
03 September 2026
Accepted
14 September 2026
Published
29 September 2026
Volume
17 - 2026
Edited by
Fan Yang, University of Illinois at Urbana-Champaign, United States
Reviewed by
Soheil Keshmiri, Okinawa Institute of Science and Technology Graduate University, Japan
Andrii Sverstiuk, Ternopil State Medical University, Ukraine
Updates
Copyright
© 2026 Ye, Lu and Lin.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Xiaohong Ye, xiaohongye@jmu.edu.cn
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.