Abstract
This systematic review focuses on the latest developments and challenges in cardiorespiratory monitoring technology, capable of functioning in the unique conditions of space to ensure the health and safety of astronauts during prolonged missions. The review highlights development of unobtrusive wearable devices that can monitor vital functions continuously without impacting astronaut performance based on different principles, such as electrical or mechano-acoustic. The article discusses the technical requirements, design considerations, and the operational challenges these devices face, such as electromagnetic interference, physical durability, and the need for non-intrusiveness. Future directions emphasize advancements in textile-integrated sensors and nano-engineered materials or integrating machine learning algorithms for real-time analysis of the multimodal data. This review serves as a critical resource for researchers and engineers in the field of aerospace medicine and bioengineering, providing insights into the evolution of cardiorespiratory monitoring techniques from Earth to the extreme environment of space.
1 Introduction
Since the first human left Earth’s bounds in 1961, more than 1,200 crewed space missions have been conducted (Smith et al., 2020). The spaceflight environment—particularly microgravity—induces well-documented alterations in cardiovascular structure and function, including cardiovascular unloading, fluid redistribution, and cardiopulmonary deconditioning (Jirak et al., 2022). Additional environmental stressors such as exposure to cosmic radiation and variations in cabin pressure, temperature, and humidity further influence physiological responses (Di Rienzo et al., 2021). These adaptations underscore the need for reliable methods capable of capturing dynamic cardiorespiratory responses during daily activities in space and throughout long-duration missions.
Continuous monitoring of vital signs has therefore been an essential part of astronaut healthcare since the earliest space missions. Modern research continues to innovate and evaluate wearable or otherwise ambulatory devices designed specifically for the constraints of spaceflight (Di Rienzo et al., 2021). Such systems must operate accurately in microgravity, meet rigorous the National Aeronautics and Space Administration (NASA) safety requirements, integrate with International Space Station (ISS) hardware, and remain non-intrusive so astronauts can work and sleep without disruption (Di Rienzo et al., 2021; Kaufmann et al., 2020). Cardiac and respiratory monitoring, in particular, provides essential insight into physiological status and early indicators of pathological change (Kaufmann et al., 2020).
Accordingly, this review focuses exclusively on monitoring of cardiac and breathing activity. Three research questions (Q1–Q3) were formulated for this review. The main research question (Q1) addresses which sensors are used for in-flight monitoring; the search strategy and keywords were therefore constructed to maximize retrieval of studies reporting cardiorespiratory sensing technologies in human spaceflight and related contexts. Based on the body of literature identified for Q1, two additional research questions are then addressed: (Q2) How do sensing modalities differ between space and terrestrial conditions? (Q3) How are these monitoring systems validated?
Previous reviews and summary reports have addressed important but distinct aspects of astronaut physiological monitoring (Di Rienzo et al., 2021; Scalia et al., 2022; ; Gupta et al., 2025; Wang et al., 2025; ), as summarized in Table 1. Earlier works focused mainly on ballistocardiography (BCG) measurements during parabolic flights and orbital missions, including historical experiments from the 1970s to 1990s and related signal-analysis principles (; ). Later reviews expanded the scope toward wearable systems for ISS and Extravehicular Activity (EVA) applications, cardiovascular-monitoring patents, and technology transfer between space and terrestrial medicine (Di Rienzo et al., 2021; Scalia et al., 2022). Most recent reviews further emphasize flexible, wearable, and multiparametric platforms for astronaut health surveillance during prolonged missions (Gupta et al., 2025; Wang et al., 2025). However, these works either focus on selected technologies or address cardiorespiratory monitoring only as part of broader health-monitoring concepts. A focused synthesis of cardiorespiratory sensing principles, signal-quality limitations, validation strategies, and translation challenges across spaceflight, microgravity, and terrestrial analogue environments is still lacking.
Table 1
| Author, source, year | Paper type | Time period | Monitoring techniques | Technical requir. | Signal processing | Signal analysis | Physiology alterations | Laboratory, simulation | Microgravity, space |
|---|---|---|---|---|---|---|---|---|---|
| , 2011 | Literature review | Up to 2011 | BCG | No | No | Yes | No | No | Yes |
| , 2013 | Experimental summary report | 1970–1990 | BCG | No | No | Yes | Yes | No | Yes |
| Di Rienzo et al. (2021), 2021 | Literature review | Up to 2019 | ECG, ICG BCG, SCG PW, RESP, PPG | Yes | No | No | Yes | No | Yes |
| Scalia et al. (2022), 2022 | Literature review Patent search | 2010–2020 | ECG, BCG PPG, RESP | No | Yes | No | Yes | Yes | Yes |
| Gupta et al. (2025), 2025 | Literature review | 1965–2023 | ECG, PPG BP, RESP TEMP, GSR, activity | Yes | Yes | No | No | No | Yes |
| Wang et al. (2025), 2025 | Literature review | Not specified | ECG, PPG BP, EMG IMU, EEG | Yes | Yes | No | No | No | Yes |
| Present paper | Systematic review | Up to 2026 | ECG, ICG, BCG, SCG GCG, BP, PCG, PPG, PW, ECHO, RESP | Yes | Yes | Yes | Yes | Yes | Yes |
Area of interest of the related review/report papers and comparison with the present paper.
BCG, ballistocardiography; BP, blood pressure; ECHO, echocardiography; ECG, electrocardiography; ICG, impedance cardiography; GSR, galvanic skin response; PCG, phonocardiography; PPG, photopletyhsmograph; PW, pulse wave; RESP, respiration; SCG, seismocardiography; TEMP, body temperature; EEG, electroencephalography; IMU, inertial measurement unit.
This systematic review addresses that gap by providing a focused synthesis of cardiorespiratory monitoring approaches across these environments (see Figure 1). It summarizes sensing principles, technical requirements, and physiological and operational challenges, while also comparing the advantages, limitations, validation gaps, and signal-quality issues of individual approaches. Given that most evidence still comes from ISS-based studies, these findings are timely for guiding the development and validation of monitoring technologies for future platforms, including the Lunar Gateway, Artemis missions, and commercial space stations.
Figure 1
2 Literature selection methodology
2.1 Literature selection criteria
To ensure comprehensive coverage, databases Web of Science and Scopus were utilized, supplemented by additional search results from Google Scholar. Keywords such as “Vital signs,” “Monitoring,” “Cardiac,” “Respiratory,” “ECG,” “BCG,” and “SCG” were combined with terms like “Astronaut,” “Space flight,” “ISS,” and “Microgravity” to refine the search.
Filters were applied to include only journal articles and conference papers, excluding reports and research protocols. Additional filters were not used due to limitations in scope and resources for conducting experiments in such unique settings. The last systematic literature search was conducted on June 18, 2026.
The literature search is centered around studies concerning the monitoring of cardiac and respiratory activities, particularly in space or simulated conditions. Consequently, inclusion criteria prioritize research that specifically addresses the monitoring of cardiac and/or respiratory activity, including blood pressure, especially in contexts relevant to astronauts, such as space flights or simulations thereof. Exclusion criteria encompass studies that solely focus on measuring other physiological parameters, such as temperature or muscle activity, or discuss monitoring without considering the feasibility of application in space environments.
2.2 Literature organization
The searched literature was collected and organized using the Rayyan online tool by M.L. and K.B. independently. In cases of conflict, decisions were reached through discussion with R.V.K. and R.J. The structure and grouping of studies are outlined to ensure clarity and organization for further synthesis. The studies are categorized based on sensing principles, study observations, and the environment in which the experiments were conducted. These categories include the following grouping keywords: Sensing principles – electrodes, accelerometers, gyroscopes, scale, impedance, pressure, piezoelectric, thermistors, cuffs, photodetection; Study observations – device design, signal quality, physiology, signal and big data processing, wearability; Experimental setup – laboratory, simulated conditions, microgravity, and space flight.
This structured approach aids in understanding the breadth of research conducted and facilitates the easier identification of relevant studies based on specific criteria. Through this selection process, 56 relevant papers were identified from the total of 894 collected. The exclusion steps involved removing 129 duplicates, excluding 673 papers based on title/abstract screening, and excluding 36 papers based on full-text screening.
2.3 Selected literature summary
In total, 56 studies met the inclusion criteria, representing various physiological monitoring approaches used in both spaceflight missions and ground-based analogs when recruiting up to 148 subjects. Specifically, 12 studies were conducted during actual space missions, while the remaining were performed under simulated conditions or controlled laboratory environments. The most frequently monitored signals were electrocardiography (ECG), respiration (RESP), and blood pressure (BP), often complemented by ballistocardiography (BCG), impedance cardiography (ICG), or seismocardiography (SCG). These signals were primarily acquired using wearable or unobtrusive technologies, incorporating Ag/AgCl electrodes, textile-based sensors, and accelerometers. The dominant sensing principles aimed to ensure wearability, signal quality, and robustness in extreme or motion-prone conditions. Most studies focused on evaluating physiological changes during altered gravity exposure, assessing the accuracy and reliability of monitoring systems, or analyzing signal morphology and noise sensitivity. Given the unique opportunity to collect physiological data in these highly specific environments, many studies combined multiple sensing modalities and analytical approaches to validate their devices and extract meaningful insights from these multi-modal recordings.
An overview of the study characteristics, including experimental setups, measured signals, and key research objectives, is presented in Table 2. The detailed summary of evidence is then depicted in Figure 2 to illustrate the possible integration of the considered solutions.
Table 2
| Author, year | Experimental setup/ Space Mission | Subjects | Record | Signals | Sensing principle | Keynotes |
|---|---|---|---|---|---|---|
| Tahvanainen et al. (Tahvanainen et al., 1992), 1991 | Head down tilt | – | – | ECG ICG BP PCG | – – – | Sensing accuracy |
| Gallasch et al. (Gallasch et al., 1996), 1996 | Austromir-91 mission | 1 | – | BCG ECG PW | Accelerometer Ag/AgCl ELCs PPG | Unobtrusiveness Easy-to-use instrumentation |
| Kuipers (Kuipers, 1996), 1996 | STS-55 mission | 4 | – | ECHO BP ECG ICG RESP | Piezoelectric transducers – – – Inductance ELCs | Physiological changes |
| Gopalsamy et al. (Gopalsamy et al., 1999), 1999 | LAB | – | – | ECG RESP | – – | Wearability/Unobtrusiveness Reliability Easy-to-use instrumentation |
| Prisk et al. (Prisk et al., 2001), 2001 | STS-55 mission | 1 | 15 min | BCG ECG RESP | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs | Signal morphology Sensing accuracy |
| Montgomery et al. (Montgomery et al., 2004), 2004 | Parabolic flight | 6 | 26 s | ECG RESP BP | Ag/AgCl ELCs Ag/AgCl ELCs Auscultation | Wearability/Unobtrusiveness Reliability |
| Mundt et al. (Mundt et al., 2005), 2005 | Underwater analog | – | – | ECG RESP BP | Ag/AgCl ELCs Ag/AgCl ELCs Auscultation | Wearability/Unobtrusiveness Reliability |
| Castiglioni et al. (), 2006 | Parabolic flight | 23 | – | ECG RESP | Textile ELCs Textile piezoresistive ELCs | Signal quality |
| Baevsky et al. (), 2007 | Unspecified mission | 8 | 13 min | ECG PW BP RESP | – PPG Oscillometry – | Signal quality Level of noise/artifacts |
| Di Rienzo et al. (), 2007 | STS-107 mission | 4 | 15 + 5 min | ECG BP RESP | – – – | Physiological changes |
| Fei et al. (Fei et al., 2010), 2010 | Terrain analog | 2 | 4 days | PW | PPG | Testing in extreme conditions Sensing accuracy |
| Vigo et al. (Vigo et al., 2012), 2012 | Isolation analog | 6 | 24 h | ECG | Ag/AgCl ELCs | Physiological changes |
| Arai et al. (), 2013 | Parabolic flight | 8 | Few seconds | BP ECG | Volume clamp – | Signal quality |
| Deliere et al. (), 2013 | Parabolic flight | 5 | 20 s | BCG ECG ICG RESP | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs Nasal thermistor | Physiological changes Signal morphology |
| Luchitskaya et al. (Luchitskaya et al., 2013), 2013 | Parabolic flight | 5 | 20 s | BCG ECG ICG RESP SCG | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs – – | Signal delays |
| Wiard et al. (Wiard et al., 2013), 2013 | Parabolic flight | 1 | 17 s | BCG BCG ECG | Electronic scale Accelerometer Ag/AgCl ELCs | Signal analysis |
| Migeotte et al. (Migeotte et al., 2013), 2013 | Parabolic flight | 5 | 20 s | BCG ECG ICG RESP | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs Nasal thermistor | Signal analysis Level of noise/artifacts |
| Vigo et al. (Vigo et al., 2013), 2013 | Isolation analog | 6 | 24 h | ECG | Ag/AgCl ELCs | Physiological changes |
| McCall et al. (McCall et al., 2014), 2014 | Parabolic flight | 10 | 17 s | BCG BCG ECG | Electronic scale Accelerometer Ag/AgCl ELCs | Signal quality Level of noise/artifacts |
| Baldi et al. (), 2014 | Anechoic chamber | 1 | 60 s | RESP | UWB | Sensing accuracy Signal quality |
| Di Rienzo et al. (), 2014 | LAB (sleeping) | 1 | 24 h | ECG SCG RESP | Textile electrodes Accelerometer Textile piezoresistive ELCs | Signal quality Physiological changes |
| Arbeille et al. (), 2014 | Isolation analog | 6 | – | ECHO | Piezoelectric transducers | Physiological changes |
| Di Rienzo et al. (), 2015 | Futura mission | 1 | – | ECG SCG RESP | Textile ELCs Accelerometer Textile piezoresistive ELCs | Signal quality |
| Przybylski et al. (Przybylski et al., 2015), 2015 | LAB | 1 | – | ECG | – | Wearability/Unobtrusiveness Reliability |
| Roveda et al. (Roveda et al., 2016), 2016 | LAB (sleeping) | 1 | – | ECG RESP | Ag/AgCl ELCs Ag/AgCl ELCs | Stress strategies |
| Di Rienzo et al. (), 2017 | Futura mission | 1 | > 42 h | ECG SCG RESP | Textile ELCs Accelerometer Textile piezoresistive ELCs | Wearability/Unobtrusiveness |
| Maiolo et al. (Maiolo et al., 2017), 2017 | LAB (exercise) | 10 | 30 min | ECG | PVDF-TrFE | Movement tracking Device validation |
| Di Rienzo et al. (), 2018 | Futura mission | 1 | > 42 h | ECG SCG RESP | Textile ELCs Accelerometer Textile piezoresistive ELCs | Signal quality |
| Hill et al. (Hill et al., 2018), 2018 | Terrain analog | 2 | 137 h | ECG RESP | – – | COTS |
| Villa et al. (Villa-Colin et al., 2018), 2018 | LAB (exercise) | 4 | 120 min | ECG BP RESP | – – – | Sensing accuracy Level of noise/artifacts |
| Mlynczak et al. (Mlynczak et al., 2019), 2019 | Isolation analog | 6 | 14 days | ECG RESP | Ag/AgCl ELCs Ag/AgCl ELCs | Physiological changes Wearability/Unobtrusiveness |
| Rabineau et al. (Rabineau et al., 2020), 2020 | Head down tilt | 23 | 60 days | BCG ECG ICG | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs | Effects of exercise on BCG |
| Glazkova et al. (Glazkova et al., 2020), 2020 | LAB | – | – | RESP ECG | Nasal thermistor Ag/AgCl ELCs | Device validation |
| Borovik et al. (), 2020 | Dry immersion | 10 | 21 days | ECG BP RESP | Ag/AgCl ELCs Volume clamp Nasal thermistor | Physiological changes |
| Baevsky et al. (), 2021 | Unspecified mission | 12 | approx 10 min | BCG ECG ICG SCG | Accelerometer Ag/AgCl ELCs Ag/AgCl ELCs – | Signal morphology |
| Zhang et al. (Zhang et al., 2016), 2021 | LAB (sitting) | 10 | 1 min | RESP ECG ECG | Pneumotachograph Capacitively coupled ELCs Ag/AgCl ELCs | Sensing accuracy |
| Hoffmann et al. (Hoffmann et al., 2022), 2022 | Head down tilt | 20 | 60 days | ECG SCG BP | Ag/AgCl ELCs Accelerometer Volume clamp | Sensing accuracy Level of noise/artifacts Easy-to-use instrumentation |
| Shankhwar et al. (Shankhwar et al., 2022), 2022 | Head down tilt | 30 | 50 min | ECG BP RESP | – Oscillometry – | Physiological changes |
| Smith et al. (Smith et al., 2022), 2022 | High-G centrifuge | 24 | – | ECG RESP BP | – – – | Physiological changes |
| Whittle et al. (Whittle et al., 2022), 2022 | Head down tilt | 7 | 36 min | BP | Volume clamp | Sensing accuracy |
| Benítez-Salgado et al. (), 2022 | Terrain analog | 5 | 5 days/5 min | ECG | – | Physiological changes |
| Drobczyk et al. (Drobczyk et al., 2022), 2022 | LAB (sitting) | 2 | – | BCG | Accelerometer | Device validation Reliability |
| Zhedyaev et al. (Zhedyaev et al., 2023), 2023 | Dry immersion | 9 | 7 days | BP | Volume clamp | Physiological changes |
| Hoenemann et al. (Hoenemann et al., 2023), 2023 | Head down tilt | 24 | 60 days | ECG BP | Ag/AgCl ELCs Volume clamp | Physiology changes |
| Jagtap et al. (Jagtap et al., 2023), 2023 | 360° rotating platform | 18 | 3 min/angle | PW BP | Wrist piezoelectric transducer – | Physiology changes Device validation |
| Shibata et al. (Shibata et al., 2023), 2023 | Unspecified mission | 13 | 6 mo/24 h | ECG BP | Ag/AgCl ELCs Volume clamp | Physiology changes |
| Benítez-Salgado et al. (), 2024 | Terrain analog | 5 | 4 days/5 min | ECG | Ag/AgCl ELCs | Physiology changes Device validation |
| Balali et al. (), 2025 | Unspecified mission | 17 | 6–12mo/5 min | ECG ICG SCG BP RESP | Ag/AgCl ELCs Ag/AgCl ELCs Accelerometer – Nasal thermistor | Physiology changes |
| Derobertmasure et al. (), 2025 | Mars analog | 6 | 12 days | PW BP | PPG Oscillometry | Wearability/Unobtrusiveness Sensing accuracy |
| Kundu et al. (Kundu et al., 2025), 2025 | Head down tilt | 27 | 30 min | ECG BP | Textile ELCs Oscillometry | Physiology changes |
| Mastrandrea et al. (Mastrandrea et al., 2025), 2025 | Unspecified mission | 1 | 6 mo/5×48 h | ECG RESP PW | Textile ELCs Inductance ELCs PPG | Physiology changes Signal analysis |
| Moller et al. (Moller et al., 2025), 2025 | Isolation analog | 11 | 120 days | ECG ICG RESP | – – – | Physiology changes Device validation |
| Solbiati et al. (Solbiati et al., 2025), 2025 | Head down tilt | 148 | 5/21/60 days | ECG | Ag/AgCl ELCs | Device validation Signal analysis |
| Solbiati et al. (Solbiati and Caiani, 2025), 2025 | Head down tilt | 66 | 5/21/60 days | ECG | Ag/AgCl ELCs | Physiology changes Signal analysis |
| Sokol et al. (Sokol et al., 2025), 2025 | Underwater analog | 6 | 8 days | ECG RESP | Ag/AgCl ELCs – | Physiology changes Device validation |
| Serrano et al. (Serrano et al., 2026), 2026 | LAB (posture change) | 1 | 2 min | ECG | Textile ELCs | Physiology changes Device validation |
Study characteristics and objectives related to astronaut cardiorespiratory monitoring.
BCG, ballistocardiography; BP, blood pressure; ECHO, echocardiography; ECG, electrocardiography; ELCs, electrodes; GCG, gyrocardiography; ICG, impedance cardiography; LAB, laboratory; PCG, phonocardiography; PPG, photopletyhsmograph; PW, pulse wave; RESP, respiration; SCG, seismocardiography.
Figure 2
3 Q1: sensing principles
3.1 Overview of commercial devices on the ISS
Beyond the scope of the systematic literature search, it is crucial to establish the baseline of currently operational hardware used for standard health monitoring aboard the ISS. For over a decade, standard physiological monitoring has relied on established commercial devices modified for spaceflight (Di Rienzo et al., 2021; Scalia et al., 2022). For cardiac electrical activity, the NASA Holter Monitor 2—a modified Mortara H12+ (Sarapa, 2005) (Mortara Instrument, Milwaukee, WI, USA) serves as the gold standard for 12-lead ECG clinical data collection during periodic health evaluations. For daily exercise monitoring, astronauts utilize commercially available heart rate monitors, such as Polar (Giles et al., 2016) (Polar Electro, Kempele, Finland) chest straps. Long-term activity and sleep-wake patterns are tracked using wrist-worn actigraphy devices like the Actiwatch () (Philips Respironics, Murrysville, PA, USA). Regarding respiratory and metabolic monitoring, the Portable Pulmonary Function System (Moore et al., 2018) (Danish Aerospace Company, Odense, Denmark) represents the operational standard for gas exchange analysis and cardiopulmonary evaluation. This system utilizes photoacoustic gas analysis to provide breath-by-breath metabolic data. More recently, the Bio-Monitor system, utilizing the Astroskin (MacQuarrie et al., 2024; Mastrandrea et al., 2025), (Carré Technologies Inc., Montreal, Canada), has been deployed to provide continuous, multi-parameter monitoring. Beyond ECG, this wearable system employs respiratory inductance plethysmography with thoracic and abdominal bands to measure respiratory rate and volume, complemented by standard pulse oximetry for SpO2 monitoring. Additionally, continuous blood pressure monitoring for research often employs the volume-clamp method via devices like the Portapres or Finapres systems (Imholz et al., 1998) (Finapres Medical Systems, Enschede, The Netherlands).
3.2 Cardiac electrical signals
Electrocardiography (ECG) or impedance cardiography (ICG) are commonly employed to measure the electrical heart activity of astronauts, like beat-to-beat intervals and heart rate variability (HRV), particularly related to cardiac pathologies, electrolyte abnormalities or oxygenation problems (Van Mieghem et al., 2004).
The ECG served as the primary measurement technique in large amount of studies and also served as a gold standard, providing a reference for evaluating the accuracy of alternative techniques, particularly those measuring mechano-acoustic signals (Prisk et al., 2001; ; Wiard et al., 2013; Migeotte et al., 2013; ). Standard Ag/AgCl electrodes are commonly utilized for measuring astronauts’ ECG signals (Gallasch et al., 1996; Prisk et al., 2001; Montgomery et al., 2004; Mundt et al., 2005; ; Luchitskaya et al., 2013; Wiard et al., 2013; Migeotte et al., 2013; McCall et al., 2014; Roveda et al., 2016; Mlynczak et al., 2019; Rabineau et al., 2020; Glazkova et al., 2020; ; Zhang et al., 2016; Hoffmann et al., 2022; Sokol et al., 2025). Alternatively, textile electrodes (; ; ; ; ), capacitively coupled electrodes (Zhang et al., 2016), or piezoelectric copolymer sensors (Maiolo et al., 2017) can also be employed for this purpose.
In the selected studies, recording the ECG using standard Ag/AgCl electrodes typically required only 2 electrodes (Roveda et al., 2016) or 3 electrodes (Gallasch et al., 1996; Glazkova et al., 2020; Hoffmann et al., 2022), placed on the astronaut’s chest or limbs (). When electrodes were used for both ECG and respiration recording, successful outcomes were achieved with 6 electrodes (Montgomery et al., 2004; Mundt et al., 2005). While measurements using standard electrodes provided accurate information on cardiac activity, some experiments were invalidated by electrode disconnection, often due to the inappropriate adhesion of electrodes to sweaty skin (Mlynczak et al., 2019). Hence, textile electrodes made of conductive fibers incorporated into wearable cotton vests, such as MagIC-Space, were designed and tested for space monitoring (; ; ). One advantage of ECG measurement using textile electrodes, which are part of clothing, is the possibility of continuous unobtrusive scanning during various activities without the need for lengthy skin surface preparation and application of conductive paste. However, textile electrodes may suffer from poorer skin contact, potentially leading to low-quality signal recording. To mitigate this, it is advisable to design a wearable vest with a compression zone made of cotton and elastane in the area where the electrodes are placed to ensure good electrode-skin contact (). Beyond conventional single-lead monitoring, high-density textile arrays may enable spatial mapping of cardiac electrical activity (Serrano et al., 2026).
An alternative approach that can be used to achieve unobtrusive and accurate measurement of an astronaut’s ECG is the utilization of non-contact sensors based on ultra-thin flexible PVDF-TrFE (polyvinylidene fluoride-cotrifluoroethylene) pressure technology. In this scenario, the sensor is integrated into a wearable wristband and detects cardiac activity with high sensitivity and without requiring close contact with the astronaut’s skin (Maiolo et al., 2017). Another noncontact solution involves integrating a capacitively coupled electrode into a cotton vest. Capacitance electrodes can detect the P wave, QRS complex, and T wave with comparable accuracy to classical wet electrodes (similarity of waveform morphology is > 95%). However, capacitive electrodes require careful consideration of various parameters and requirements during the design phase to ensure high gain and low noise level of the input signal amplifier, high input impedance, and common mode rejection ratio, as well as incorporating shielding and anti-jamming capabilities into the electrodes (Zhang et al., 2016).
For combined ECG and ICG recording, a total of 8 electrodes were employed (; Luchitskaya et al., 2013; Migeotte et al., 2013; Rabineau et al., 2020; ). Two prototypes were designed for ICG measurement in space missions, parabolic flights, head-down tilt and dry immersion experiments using 8 standard electrodes, which corresponds to the so-called tetrapolar configuration: the Pneumocard device (; Luchitskaya et al., 2013; Migeotte et al., 2013; ), and the Cardiovector device (Rabineau et al., 2020; ). During ICG measurement, 4 electrodes are placed on the neck and 4 electrodes on the chest, positioned on both mid-axillary lines at the level of the processus xiphoideus.
3.3 Cardiac mechano-acoustic signals
Mechano-acoustic cardiac signals, manifesting as vibrations, can be captured by various sensors, either directly, by placing them in close proximity of the heart (e.g. on subject’s chest), or indirectly from sensory surfaces upon which the body rests (Tavakolian et al., 2022). Methods such as BCG, SCG, ECHO, PCG, photoplethysmography (PPG), and blood pressure (BP) measurement have been explored for monitoring astronauts’ mechano-acoustic signals and related cardiac parameters.
3.3.1 Ballistocardiography
A single three-axis accelerometer was used to record BCG signals in the majority of investigated studies (Prisk et al., 2001; ; Luchitskaya et al., 2013; Migeotte et al., 2013; Wiard et al., 2013; McCall et al., 2014; Rabineau et al., 2020; ; Drobczyk et al., 2022). Other studies employed two piezoresistive accelerometers to measure two-dimensional BCG (Gallasch et al., 1996), while some utilized two three-axis accelerometers arranged differentially to enhance the signal-to-noise ratio (SNR) (Drobczyk et al., 2022). In such cases, one sensor is positioned near the carotid artery, and the other is placed at the apex cordis. The maximum amplitude of the HIJK complex was observed in space along the head-to-foot axis (y) (Prisk et al., 2001; ), but the dorso-ventral axis (z) also proved to be relevant (Prisk et al., 2001; Drobczyk et al., 2022). That can be conducted by positioning the accelerometer in the lumbar region (Prisk et al., 2001), between the second and third lumbar vertebrae (Rabineau et al., 2020), or near the center of mass (; Luchitskaya et al., 2013; Migeotte et al., 2013).
When measuring in microgravity, various approaches have been employed to record BCG signals. One method involves using an electronic scale-based approach (Wiard et al., 2013; McCall et al., 2014), where the astronaut’s feet are secured to the scale using foot restraints. This approach allows for achieving a higher average SNR value compared to using a classic accelerometer. Additionally, it enables monitoring in fractional-g environments (such as the Moon or Mars) and eliminates the need to avoid collisions caused by free-floating. However, a limitation of the scale-based BCG method is that it can only measure the signal along one longitudinal axis.
3.3.2 Seismocardiography
A single-axis accelerometer was utilized for SCG measuring during space flights (Luchitskaya et al., 2013; Rabineau et al., 2020; ; Hoffmann et al., 2022), specifically along the dorso-ventral axis (z) (Rabineau et al., 2020; Hoffmann et al., 2022). Additionally, a triaxial micro electro mechanical system accelerometer, integrated into a wearable vest known as MagIC-Space, was tested for SCG measurement (; ; ; ). The positioning and secure attachment of the sensor are vital for accurate SCG measurement, with the area near the cardiac apex considered the most optimal. Consequently, probes are frequently placed in the sternum region and fastened securely with elastic straps. To mitigate the influence of sensor location on SCG signal quality, employing a larger sensor area has been suggested (Rabineau et al., 2020). Longitudinal ISS recordings showed that SCG can track cardiac time intervals and kinetic-energy features across months of spaceflight. Yet, single-axis apex SCG remains vulnerable to altered heart orientation and sensor alignment in microgravity ().
One study (Rabineau et al., 2020) used combined inertial setup to acquire 1-axis dorsoventral SCG from cardiac apex and 6-degreeof-freedom BCG (3-axis linear accelerations and 3-axis angular velocities) from lumbar region, so that three kineticenergy signals were derived: (a) linear BCG (x,y,z) kinetic energy, (b) rotational BCG (x,y,z) kinetic energy, and (c) SCG kinetic energy along the linear z-axis.
3.3.3 Echocardiography
Given by the overall nature of ECHO examination, it is employed for periodic physiological assessments and reference measurements, rather than for continuous unobtrusive monitoring. During the space mission, ECHO was obtained with the ultrasound probe and successfully utilized to measure parameters such as left ventricular enddiastolic volume, systolic volumes, stroke volume, and cardiac output (Kuipers, 1996). Another study used this facility to observe changes in vein diameter and thickness ().
3.3.4 Phonocardiography
Cardiac acoustic monitoring captures vibrations generated by mechanical heart activity, mainly valve closure and blood-flow-related events (Jaros et al., 2023). In wearable PCG implementations, these signals can be acquired using miniature electret microphones, micro-electromechanical systems (MEMS) microphones (Torabi et al., 2024), contact microphones (Gupta et al., 2020), or piezoelectric transducers (Qu et al., 2022) coupled to the chest wall. However, PCG performance depends strongly on sensor, stable mechanical coupling, contact pressure, and suppression of motion, garment-friction, respiratory, and environmental noise (Jaros et al., 2023; Gupta et al., 2020; Liu et al., 2016). One study involved PCG recordings in analogue space conditions, where the signals were captured from the aortic valve area (Tahvanainen et al., 1992) but without sensor technology specified.
3.3.5 Photoplethysmography
In both space and terrestrial missions, an infrared photoelectric sensor positioned on a fingertip, was employed (; Fei et al., 2010). As reported, less than 2% of artifacts were recorded in the measured data (). A 12-day Marsanalog study further demonstrated the operational feasibility of finger-to-toe PPG for repeated pulse-wave-velocity assessment during daily activities and extravehicular operations. However, cold-induced peripheral vasoconstriction limited data acquisition in some participants, indicating that peripheral PPG performance may depend strongly on local perfusion ().
3.3.6 Blood pressure
The studies focusing on BP measurement in astronauts employed so-called motion-tolerant devices including the Accutracker II (Montgomery et al., 2004; Mundt et al., 2005), (SunTech Medical, Morrisville, NC, USA), utilizing the auscultation method and a brachial cuff, or ones relying on the oscillometric method, such as TensoPlus () (Speidel & Keller, Jungingen, Germany), and NIBP 100B (Shankhwar et al., 2022) (BIOPAC Systems, Inc., Goleta, CA, USA). In contrast, continuous non-invasive hemodynamic monitoring in space often relies on the volume clamp principle. While devices like the Nexfin monitor (Whittle et al., 2022) (BMEYE B.V., Amsterdam, The Netherlands) have been utilized in ground-based simulations, the modified FMS Portapres (Hoffmann et al., 2022; Tahvanainen et al., 1992; ) and Finometer systems (; Zhedyaev et al., 2023), (Finapres Medical Systems, Enschede, The Netherlands) available on the ISS utilizes finger cuffs only. However, it is important to note limitations regarding the algorithms derived from these measurements in microgravity. Recent findings (Hughson et al., 2017) indicate that cardiac output estimated by pulse contour analysis from these devices does not match the values measured by rebreathing techniques during spaceflight. This suggests that the modeling algorithms may be violated under 0G conditions.
3.4 Respiratory signals
In addition to monitoring heart signals, it is crucial to monitor the respiratory activity of astronauts, focusing on parameters such as respiratory rate, volumes, or flows (; ). Methods for measuring these parameters can be broadly categorized into contact and non-contact methods. Contact devices may utilize acoustic measurement, airflow-based measurement, chest and abdominal movement detection, or oxygen saturation measurement. On the other hand, non-contact measurement methods may employ optical methods, radar-based approaches, or thermal sensors ().
3.4.1 Pneumotachography
Pneumotachography measures respiratory airflow from the pressure drop generated across a calibrated resistive element, typically a fine mesh screen or capillary tube bundle. The differential pressure signal is converted into inspiratory and expiratory flow and can be integrated to derive volume-related parameters such as tidal volume and minute ventilation (Massaroni et al., 2019). Although pneumotachography provides direct airflow information, its wearable applicability is limited by the need for a mouthpiece, mask, tubing, or flow head, as well as by calibration requirements, added dead space, breathing resistance, condensation, leaks, and interface displacement (Massaroni et al., 2019; Graham et al., 2019). A fixed pneumatic resistance element for capturing respiratory flow and volume was shown to be viable for space-relevant measurements in one study, when implemented in Cardiovector device ().
3.4.2 Plethysmography
Two types of plethysmography has been explored in the context of space missions: inductance (Kuipers, 1996), which relies on measuring changes in inductance of wires encircling the abdominal and thoracic regions, and impedance (Prisk et al., 2001; Montgomery et al., 2004; Mundt et al., 2005). The advantage of this approach lies in simultaneous acquisition of ECG and respiration using the same electrodes. Examples of impedance pneumographs that were used in space setting are Pneumonitor 2 (Mlynczak et al., 2019) and wearable BioPatch system (Roveda et al., 2016). Despite providing comfortable and accurate breath monitoring even under dynamic conditions, the standard electrodes used were found to be unsuitable for sweaty skin (Mlynczak et al., 2019). Furthermore, incorporating a textile piezoresistive plethysmograph into a wearable MagIC-Space vest proved functional (; ; ; ; ). This plethysmograph utilized a flexible conductive fabric placed around the subject’s abdominal area, specifically in the lower part of the sternum. As the fabric expanded and contracted during inhalation and exhalation, the electrical resistance of the piezoelectric materials changed in accordance with alterations in lung volume. The flexibility of the fabric facilitated comfortable wear, enhancing user experience.
3.4.3 Nasal thermistors
Nasal thermistors detect respiration from temperature changes near the nostrils caused by alternating inspired and expired airflow (Massaroni et al., 2019). During inspiration, cooler ambient air decreases the sensor temperature, whereas expiration exposes the sensor to warmer humidified air from the respiratory tract, producing a respiratory waveform suitable mainly for estimating breathing timing and respiratory rate (Massaroni et al., 2019; Sabil et al., 2019). These sensors are typically based on temperature-sensitive resistive semiconductor materials, often negative-temperature-coefficient metal-oxide ceramics (Feteira, 2009). Although nasal thermistors are compact, low-power, and simple to integrate, their signal is an indirect airflow surrogate and is sensitive to sensor position, mouth breathing, nasal obstruction, condensation, humidity, and ambient temperature changes (Massaroni et al., 2019; Sabil et al., 2019; Hussain et al., 2023). These resistive semiconductor sensors are commonly employed in medical settings and have also been successfully utilized in monitoring respiratory activity of astronauts (; Migeotte et al., 2013; Rabineau et al., 2020) when positioned in front of the nostrils.
3.4.4 Radar-based system
Non-contact ultra-wideband (UWB) technology operates by transmitting pulses with a sufficiently high repetition frequency from the transmitter. This approach offers several advantages, including high spatial resolution, sensitivity, low power spectral density, minimal risk of unwanted biological effects, and low cost. Moreover, the measurements are unaffected by the subject’s clothing ().
4 Q2: challenges of monitoring in space
4.1 Operational challenges
Operational challenges primarily revolve around ensuring the comfort of the subject being monitored, encompassing both the wearability of the monitoring system and its ease of handling. Additionally, this subsection delves into the physiological alterations experienced by astronauts, which can significantly impact the signals being sensed and processed. These factors should be carefully considered when designing both the hardware and software components.
4.1.1 Wearability and unobtrusiveness
The measuring devices are engineered to be compact, lightweight, with minimal cables, and featuring low-energy wireless data transmission (Montgomery et al., 2004; Mundt et al., 2005; Roveda et al., 2016; Glazkova et al., 2020; Gopalsamy et al., 1999; Przybylski et al., 2015). Due to concerns about classic Ag/AgCl electrodes being unsuitable for sweaty skin and requiring time-consuming preparation, experiments have focused on capacitive electrodes. These electrodes enable the detection of ECG signals through clothing without direct skin contact (Mlynczak et al., 2019; Zhang et al., 2016). Authors have also explored the design of textile electrodes or Smart textiles (; Villa-Colin et al., 2018). Examples include the MagIC-Space smart cotton vest (; ; ; ; ) and the SpaceTex Smart-Shirt (Drobczyk et al., 2022), which integrate wiring into seams for a nearly invisible appearance, and sensing socks that provide a comfortable fabric interface for physiological sensors (; ; ; ; ; Drobczyk et al., 2022; Fei et al., 2010). However, the prolonged use of textile electrodes, particularly during exercise, raises concerns regarding their hygienic and microbiological safety. Although several studies have investigated long-term monitoring (; ; ), none specifically assessed the effects of sweat exposure, microbial stability, or the feasibility of sanitation procedures. Durability testing was reported only in (Serrano et al., 2026), where the authors focused on the mechanical robustness of the textile electrode array. After 100 bending cycles at a radius of 3 mm, electrical continuity was preserved, with no observed cracking or delamination. On the other hand, wearable systems are not limited to textiles; flexible bracelet designs, such as those incorporating PVDF-TrFE pressure sensors, offer soft, skin-friendly options that can be worn below gloves without hindering arm movement (Maiolo et al., 2017). Non-contact UWB measurement further enhances unobtrusiveness with minimal risk of biological effects and low costs, unaffected by clothing ().
4.1.2 Ease of handling
Although ease of handling should be given a high priority, it was assessed in limited studies (Luchitskaya et al., 2013; ; Gallasch et al., 1996; Sokol et al., 2025; ), where participants reported finding the devices easy to use, thanks to an intuitive user interface. Additionally, the time required for setup and the complexity of experiment preparation within the ISS system were reported to be reduced in (; ). However, specific assessments were not provided in these studies.
4.2 Physiology alterations
Physiology alterations are a frequent topic of interest in the studies. Research during spaceflights, as suggested by Rabineau et al (Rabineau et al., 2020), should take into account several observations. Besides the known cardiovascular deconditioning of astronauts, there’s a risk of inadequate sensing, such as incorrect ranges, or misinterpretation of data due to changes in physiology. However, a significant limitation in interpreting spaceflight data is the relatively small sample size of individuals who have undergone spaceflight experiences (Hoffmann et al., 2022).
Many studies primarily focus on changes in heart rate (HR), respiratory rate (RR), heart rate variability (HRV), cardiac and respiratory output, or BP. However, the observed changes in these parameters vary between studies. For example, an increase in HR during space missions was noted in (), while other studies reported a decrease in simulated conditions (Rabineau et al., 2020; Shankhwar et al., 2022). This inconsistency may partly reflect differences in the terrestrial reference condition rather than true physiological disagreement. In a longitudinal ISS cohort, heart rate (HR), pre-ejection period (PEP) and the ratio of PEP to left ventricular ejection time (PEP/LVET) showed different in-flight trajectories when compared with sitting versus supine baseline values, demonstrating that baseline posture can alter the inferred direction of cardiovascular adaptation ().
In controlled breathing experiments, HRV increased in some cosmonauts but decreased in others, suggesting a reduced parasympathetic reserve (). Conversely, the dominance of the parasympathetic system and an increase in RR interval (although not significant) were observed during head-down tilt (HDT) experiments (Shankhwar et al., 2022). The initial decrease in supine HR followed by a return to baseline values later during HDT could be attributed to a reduction in vagal modulation and an increase in the sympathovagal balance. In control groups, a transient decrease in HR was observed, along with a decrease in precordial movements throughout the cardiac cycle (Rabineau et al., 2020). Morning short-term HRV collection in Mars-analog environments also has shown the significant changes throughout a mission ()(with the increase on the second and third day and decrease on the fourth day) but without a wider context of astronauts’ activities. Moreover, reduced daylight exposure in MARS-500 simulated mission could account for sleep-wake differences of long-term HRV recordings (Vigo et al., 2013; Vigo et al., 2012). In this study, wake HRV showed a progressive increase in mean RR interval and amplitude in very low and high frequency components, but during sleep, the relative amplitude of the high frequency component of HRV decreased (Vigo et al., 2013).
An important finding highlighted that HRV data in space cannot be properly interpreted without RR measurements or controlled breathing experiments. The difference in HRV between controlled and spontaneous breathing significantly increased in space and was primarily associated with reduced RR. The decrease in HRV during controlled respiration (6 breaths/min) indicates a diminished parasympathetic reserve, which may contribute to postflight disturbances (). Consistently, after 60 days of HDT bed rest, indices of parasympathetic cardiac modulation and baroreflex sensitivity decreased, whereas low-frequency systolic BP variability increased, collectively indicating altered autonomic cardiovascular control (Hoenemann et al., 2023). Pooled 24-h Holter ECG data from six HDT campaigns showed reductions in time-domain, frequency-domain, and nonlinear HRV indices during bed rest, with larger disturbances after HDT discontinuation in longer campaigns (Solbiati and Caiani, 2025). These findings indicate altered cardiovascular variability during simulated microgravity and re-exposure to orthostatic stress, rather than a direct measure of autonomic tone.
Time-dependent hemodynamic adaptations have been observed during prolonged exposure to dry immersion. In a 7-day dry-immersion protocol (Zhedyaev et al., 2023), only modest cardiovascular changes were observed: systolic arterial pressure remained unchanged during and after immersion and small increases in diastolic and mean arterial pressure occurred on the last immersion day and on day 3 of recovery. Pulse pressure, stroke volume, and HR were reduced only on the last immersion day, and no significant differences in stroke volume or HR responses were detected across experimental stages. By contrast, during longer dry-immersion exposures (), systolic BP and MAP reduced during orthostasis by day 14 but not day 19, and HR elevated at rest and sharply increased from day 7 onward, after which these responses appear to stabilize.
Similarly, the essence of mechano-acoustic methods, such as BCG, SCG, and PPG, facilitates the easy monitoring of changes in hemodynamics, primarily attributed to alterations in gravity levels. These changes can result in hypovolemia in the upright position (Kuipers, 1996) or a decrease in blood pressure (). Inadequate vasoconstriction may also lead to an excessive decrease in filling pressure and stroke volume (Kuipers, 1996). In microgravity conditions, the hemodynamic profile adopts a new pattern, characterized by increased filling of the pulmonary circulation, especially during the initial month of flight. Consequently, the systolic volume of the right ventricle does not expand with inspiration, while systolic ejection from this ventricle decelerates (). Changes in peak morphologies of SCG signals (Hoffmann et al., 2022) may also indicate chronic decreases in left ventricular mass, with reported mass losses of 9—12% during spaceflight and 8—16% during HDT studies.However, this structural change is strongly countermeasure-dependent: 13 astronauts assessed by cardiac magnetic resonance imaging preserved both left and right ventricular mass after 5-month ISS missions despite a 12% drop in 24-hour cardiac work, owing to daily in-flight exercise (Shibata et al., 2023).
Other parameters that can be derived using BCG and SCG include the timings of ejection, such as the PEP and LVET. In microgravity environments, an increase in LVET and a decrease in PEP have been observed, indicating an increase in stroke volume (). The acceleration of LVET appears to be correlated with reductions in the mechanical strength of the left ventricle, possibly stemming from changes in blood volume and ventricular mass loss associated with prolonged bed rest; in this study, the timing interval of pulse transition time decreased similarly to that observed during 6-months of spaceflight (Hoffmann et al., 2022). These changes need to be considered when (re)setting the parameters of the parameters of the corresponding peak detectors.
Additionally, the shortening of pulse transit time may reflect cardiac output redistribution in space (). The rapid decline in pulse transit time suggests that immediate fluid shifts lead to faster blood distribution transition times to the finger, potentially resulting in increased arterial stiffness and decreased arterial compliance. For example, significant increase in intima media thickness of the carotid and femoral arteries along and 2 days after the MARS-500 mission period was observed.
Moreover, changes in the timings also appear to be related to alterations in the time interval between the R peak in the ECG and the J peak of the BCG signal (i.e. RJ interval). A consistent decrease in the scale-based RJ interval after transitioning from a ground to a microgravity environment has been reported (Wiard et al., 2013; McCall et al., 2014), likely due to the transient increase in venous return and consequent decrease in the pre-ejection period experienced in microgravity. The RI time interval measured in the BCGy (foot-to-head axis) was consistently shorter compared to the R-dZ/dtmax time interval in microgravity, which may be partly explained by the higher heart rate (Luchitskaya et al., 2013). Moreover, changes in the delay of the maximum of ICG ranged between 23 and 28 ms at rest, being lowest under microgravity conditions (12 ± 4 ms) (Luchitskaya et al., 2013).
Changes in hemodynamics are closely linked to alterations in respiratory activity, particularly changes in breathing rhythm observed during flight, such as reductions in respiratory rate () or worsened regularity of breathing (). Suborbital acceleration has been associated with frequent respiratory and visual symptoms, with 88% of subjects reporting cataracts, 29% experiencing blackouts, and one subject losing consciousness (Smith et al., 2022). In microgravity conditions, a 45% reduction in cardiogenic oscillations has been observed (Kuipers, 1996), although there was no positive end-expired concentration deviation. Measurements of chest wall mechanics have revealed that the mechanical behavior of the respiratory system differs significantly in long-term microgravity compared to short periods of microgravity. These aspects should also be considered when designing the monitoring device and associated signal processing tools.
4.3 Technical requirements
Monitoring the vital signs of astronauts requires robust technical solutions that ensure high reliability and data quality under challenging operational conditions. Key considerations include efficient power management to ensure minimal power consumption, electromagnetic compatibility (EMC), and advanced signal processing to maintain accurate physiological measurements in the demanding space environment.
4.3.1 Reliability and data quality
The proposed measurement systems were assessed in various studies (Prisk et al., 2001; ; Mundt et al., 2005; Mlynczak et al., 2019; Glazkova et al., 2020; ; Drobczyk et al., 2022; ; ). However, to our knowledge, specific results regarding data detection accuracy evaluation are presented in only one study (), where the authors used the Positive Predictive Value to assess the effectiveness of their algorithm. The results indicate that the SCG quality remained consistently high in space (98.2%), with a slight reduction to 91% on the ground. This decrease in quality on Earth was primarily attributed to artifacts induced by changes in body position during sleep and contact with the bed. Conversely, in space, astronauts sleep inside a loosely tied sleeping bag, leading to minimal body movement and a consequent drastic reduction in artifact rates. In a follow-up study, the data demonstrated good quality with an accuracy exceeding 96% (). Additionally, several studies have reported a limited range of usable data for analysis (approximately 5%) due to movement artifacts (; Migeotte et al., 2013). In (), highquality ECG recordings were obtained in more than 99% of cases at rest and in 97% of cases during exercise, enabling the derivation of comprehensive tachograms and the detection of atrial fibrillation and ectopic beats.
4.3.2 Power management and consumption
As an essential part of the equipment design for on-board use, battery safety and energy consumption must be evaluated. Lithium batteries, particularly Lithium-polymer (Glazkova et al., 2020; Maiolo et al., 2017; Przybylski et al., 2015), or Li-ion (), are commonly employed to power non-thrown devices in various biomedical monitoring applications in space. These batteries offer advantages in power management, especially when coupled with transmission modes. For instance, the BioPatch device provides three transmission modes: fast, nominal, and slow, with the slow mode significantly extending battery life by up to 35% compared to the fast mode (Roveda et al., 2016). However, studies often lack reporting on the duration of continuous monitoring, with wide-ranging durations observed, such as over 6 hours (Maiolo et al., 2017), 24 hours (Roveda et al., 2016), 30 hours (Sokol et al., 2025), over 36 hours (Fei et al., 2010), and 7 days (Glazkova et al., 2020). To enable ultra-low-power operation, a 12-bit low-power chip by Analog Devices for the microcontroller unit and a Bluetooth Low Energy module from Microchip for wireless transmission were utilized in certain cases (Glazkova et al., 2020). A non-contact UWB radar system () or Smart Shirt, described as a low-power device, incorporates an Energy Harvesting Mode that utilizes internal solar cells to harvest energy from the light sources within the Columbus module (Drobczyk et al., 2022).
Among the studies reviewed, only one () outlines a safety measure in detail. In this study, part of the control circuitry inside the unit regulated voltage in the measurement unit, while another part was dedicated to safeguarding the batteries against discharge resulting from polarity reversal and external short circuits. The system monitored battery temperature, voltage, and current, and would shut down the entire system in case of abnormal values. To mitigate potential hazards, the batteries were encased in a special non-flammable textile material designed to absorb any leaking electrolyte and placed within a separate aluminum casing to prevent fire hazards. Additionally, a series of small ventilation holes were drilled into the aluminum casing, covered by a gas-permeable membrane, to allow gases produced during battery failure to escape and maintain safe pressure levels inside the case.
Another issue for the implementation of battery devices and electronic circuits in a space suit is the high oxygen content within the suit environment. Among the reviewed studies, only one (Fei et al., 2010) addressed this concern. In this study, the sensors of the Space Sock system were connected to the exterior of the suit using standard data cables, potentially mitigating the risks associated with electronic components inside the high-oxygen environment of the space suit.
4.3.3 Electromagnetic compatibility and related signal quality insurance
Since the device must operate effectively amidst surrounding electromagnetic noise from various environmental sources, EMC is another necessary presumption of the on-board devices. For electromagnetic modalities (ECG, ICG), this interference is dictated by the coupling of ambient electromagnetic fields generated by onboard systems. The measured voltage Vreccan be approximated as shown in Equation 1:
where is the biopotential, and the interference term is a function of the environmental flux change (magnetic induction) or capacitive coupling voltage, scaled by the coupling impedance Zc (Webster, 2009). This impedance significantly increases with poor electrode contact or textile drying in low humidity environments (). To mitigate these risks and ensure data integrity, a comprehensive verification strategy is required. Apart from standard EMC tests, ensuring compatibility with the conditions aboard the ISS is crucial. For instance, capacitive ECG signals are susceptible to interference from electromagnetic fields generated by AC conductors in spacecraft environments (Zhang et al., 2016). Compliance with the Standard for Emissions and Susceptibility, as per MIL-STD-461 and 462, is imperative for monitoring electromagnetic interference (). Tests specified by space-related standards ECSS-E-ST 07C and MIL-STD-461x, covering EMC, mechanical vibrations, and data integrity, were conducted to validate compliance (Drobczyk et al., 2022). These tests were conducted in the EMC chamber of the DLR in Bremen, Germany, ensuring emissions remained within the required limits for space-relevant frequency ranges.
The presence of high electromagnetic fields, flicker noise, and white noise in the environment can significantly interfere with the weak ECG signals (Roveda et al., 2016; Zhang et al., 2016). Moreover, the intrinsic noise of circuits and electromagnetic fields from AC conductors aboard spacecraft further compound interference issues (Zhang et al., 2016). To address these challenges, certain technical requirements must be met in system design, including using a differential signal amplification method with high gain and low noise levels, ensuring high input impedance and common-mode rejection ratio, employing electrodes with shielding and anti-jamming capabilities, selecting components with low noise and drift, implementing proper grounding techniques, and incorporating decoupling capacitors at the power output end (Zhang et al., 2016).
Beyond interference with signal acquisition itself, the spacecraft-like environment may also compromise wireless data transmission. During the DIANA underwater analog mission, signal reflections and multipath propagation within the metallic habitat affected 2.4-GHz Wi-Fi communication, with signal loss observed at distances as short as 3 m under experimental conditions, which was mitigated by using an external antenna (Sokol et al., 2025). Alternatively, UWB radars offer a different approach, capable of penetrating obstacles such as clothing while providing high resolution and low power spectral density. Their ultra-short pulse transmission and immunity to multipath interference make them particularly suitable for environments where conventional wireless and sensing technologies may be affected by electromagnetic interference and signal reflections ().
4.4 Signal processing
Signal processing is a final critical component of the measurement system, ensuring the quality of measured signals and the accuracy of evaluated parameters. The quality of data can be affected by factors of various origins, broadly categorized into ambient noise from the environment, interference from the measured subject, and the use of suitable algorithms for parameter detection. To systematically address these challenges and quantify the impact of these factors, a simplified superposition model can be proposed, where the recorded signal is defined as the sum of the physiological component and the environmental interference (Sornmo and Laguna, 2005), see Equation 2:
Depending on the spectral characteristics of the noise component, different processing strategies are required. Interference out of frequency range of the measured signal can be filtered by simple band-pass filtration () to enhance proper signal quality. Robust algorithms to cancel out interfering signals are mandatory also for breathing frequency detection (Zhang et al., 2016; ). The issue concerns the assessment of the procedures and the characteristics of the signal extraction algorithms to retreive the respiration waveform correctly in the presence of surrounding noise from various sources in the environment ().
The most challenging, however, is the elimination of movement artifacts as a result of tissue/body fluid moving during the measurement and sensor-body interface noise (e.g., offset voltages occurring in the electrodes for ECG) (Roveda et al., 2016) or directly artifacts caused by human body activities (Zhang et al., 2016; Sokol et al., 2025). For inertial sensing modalities (such as BCG and SCG), the accelerometer measures the total specific force acting on the proof mass. According to Newton’s second law, the recorded acceleration vector representing the superposition of signal and noise, can be expressed according to Equation 3:
where represents the micro-vibrations of the heart, denotes external body or environmental accelerations (e.g., launch vibrations, vehicle maneuvering, or station background noise), and is the local gravitational field (Titterton and Weston, 2004). In microgravity (), the static bias is removed, altering the sensor’s operating point, while high-magnitude can saturate the sensor dynamic range, masking the delicate signal. In this context, the sensor attachment site plays a crucial role in the SNR (Inan et al., 2014). It turned out that the classical placement of the SCG sensor in the sternum area led to the susceptibility of the measured signal to noise artifacts. Vibrational signals are dependent on mass components of the system, defining the equations of motion (Hoffmann et al., 2022). Some studies report that the motion artifacts are the main reason for very limited amount of suitable data for precise analysis (; Migeotte et al., 2013).
The development of adequate signal processing algorithms is necessary not only to filter ambient noise, or motionrelated artifacts, but also to adapt the analysis of signals exposed to physiological changes during the flight or stay on the ISS. The advantage of sensing mechanical-acoustic cardiac signals (BCG, SCG) is the decreased complexity of the 3D displacement curve and the result of the measurement is a more stable and less disturbed signal (Migeotte et al., 2013) compared to measurements on Earth, but there are other challenges for signal analysis and processing methods. For example, a specific algorithm was developed for the removal of artifacts of the SCG signal and identification of four fiducial points during sleep on earth and in microgravity (). Handling SCG sleep recordings implies the analysis of a large number of heart beats and the frequent and marked variations of cardiac rhythm that characterize the different sleep stages.
Amplitude changes, arising mainly from changes in the level of gravity, also have a significant effect on the quality of signal detection. Systemic loss of cardiovascular mass can contribute to the lowered mechanical peak performance (Hoffmann et al., 2022) and SCG peak complex amplitudes is decreasing towards the end of bed rest (Hoffmann et al., 2022). When measuring BCG signals, it is not only necessary to select the highest quality sensing axis, but also the amplitudes of the IJK complex are strongly dependent on respiratory activity, when a change in amplitude can be observed in exhalation compared to inhalation (Prisk et al., 2001) – however, this ratio increases with longer stay in microgravity (). It is therefore necessary to adequately prepare the detection algorithms so that they are able to reflect these changes even during the entire mission and work with the updated signal morphology.
The final step of data processing is its analysis and evaluation. An important insight is that when analyzing the SCG and BCG signals, it is important to distinguish different cardiac phases, but also to record precordial and global movements of the body. However, there is a delay between the Q wave of the ECG and the characteristic systolic complex that can be observed on BCG. Because of this phase offset, it is possible that some of the systolic activity is recorded at the beginning of the diastolic phase defined using the ECG (Rabineau et al., 2020). Also, BCG peak complexes do not seem to be perfectly related to the classical cardiac ejection timings from ICG (Migeotte et al., 2013).
The new trends in analysis of medical data show prediction models eg. based on neural networks. However, only one of the reviewed studies dealt with analytical model for Sleep-HRV-Emotion evaluation and prediction (Roveda et al., 2016). These interventions include the time to rest/sleep, hours of sleep, ultrasound pulses, and Yoga breathing. As such, these help recover, preserve, or improve pilot and astronaut performance, especially during long-duration space. A combination of several data categories (i.e., HR, RR, R-R intervals, and HRV) may provide an indicator for the quality of sleep and thus may be a way to track sleep-stress-emotion changes. Machine learning has also been applied to a different target—predicting maximal aerobic power (VO2max) after head-down-tilt bed rest from non-exercise 24h Holter ECG features, with an accuracy comparable to exercise-based testing (root mean square error, RMSE < 5 mL/kg/min) (Solbiati et al., 2025). This points toward passively estimating fitness decline from a routine ECG, reducing the need for dedicated in-flight testing.
5 Q3: from laboratory to space
5.1 Laboratory environments
An essential initial step in validating a space monitoring device is to conduct laboratory tests (Roveda et al., 2016; Glazkova et al., 2020; Zhang et al., 2016; ; Maiolo et al., 2017; Drobczyk et al., 2022; Gopalsamy et al., 1999; Przybylski et al., 2015; Villa-Colin et al., 2018). These tests primarily aim to identify errors, design shortcomings, or minor issues in the device’s functionality. The overarching goal of these preliminary tests is to ascertain whether the device operates as intended and whether the obtained results are reliable. Moreover, laboratory testing offers a cost-effective and time-efficient means of validation (Villa-Colin et al., 2018). Additionally, laboratory tests offer the significant advantage of verifying the device’s functionality across various scenarios, including measurements in a resting position (sitting) (Zhang et al., 2016; Drobczyk et al., 2022), during controlled posture changes (Serrano et al., 2026), or during physical activities (Maiolo et al., 2017; Villa-Colin et al., 2018). Given that astronauts in space must engage in regular exercise of varying intensity to prevent muscle loss and cardiovascular deconditioning, it is pertinent to test the device during both low-intensity activities (such as running in place or hand movements) (Maiolo et al., 2017) and high-intensity physical exercises (such as treadmill workouts, cycling, or rowing) (Villa-Colin et al., 2018). While the required daily exercise is generally considered essential, the actual average aerobic time spent by astronauts may vary, with some studies (Hughson et al., 2016) reporting less than 30 minutes per day on average. Equally important is testing the device during sleep to enable continuous monitoring of the astronaut’s health and to explore sleep physiology in microgravity (), as well as for monitoring sleep-stress-emotion changes (Roveda et al., 2016).
5.2 Simulated environments
Despite the numerous advantages of laboratory testing, there is unfortunately no guarantee that a device designed will perform equally well under extreme space conditions. Therefore, it is imperative to further validate the device in conditions that simulate the effects experienced during a stay in space (Mlynczak et al., 2019; Rabineau et al., 2020; Hoffmann et al., 2022; Tahvanainen et al., 1992; Fei et al., 2010; Shankhwar et al., 2022; Whittle et al., 2022; ; Hill et al., 2018). The advantage of simulated environments lies in their ability to test many factors and scenarios that are difficult or impossible to replicate in a laboratory setting. However, it is crucial to select the most appropriate type of simulation for the specific research issue at hand (Ngo-Anh et al., 2020).
To address this need, NASA has developed a series of space analogs, which are environments designed to simulate some of the conditions and effects experienced during space flight. These analogs primarily include terrestrial simulations that replicate the surface conditions of other planets (often focusing on physical and chemical properties), isolation simulations simulating the experiences of a group of astronauts, simulations of altered gravity effects, simulations of space radiation, and simulations of the distance from Earth (). Furthermore, an analog that simulates the electromagnetic environment of space can be created using an anechoic environment. This setup is suitable for testing monitoring equipment utilizing satellite systems, antennas, and antenna components (). Utilizing terrestrial analogs to simulate space conditions enables the monitoring of cardiorespiratory and other physiological functions of the crew during exploration activities similar to those anticipated during planetary terrain exploration, such as field traverses and sample collection (Fei et al., 2010; Hill et al., 2018). Examples of such terrain simulations include the Mars analog field sites at NASA’s Biologic Analog Science Associated with Lava Terrains (BASALT) in Idaho and Hawaii, Meteor Crater in Flagstaff, Arizona, USA (Fei et al., 2010; Hill et al., 2018), Mojave Desert, California, USA (; ), or the Mars Desert Research Station in Utah, USA ().
Cardiorespiratory signals are also valuable for monitoring when simulating astronaut isolation from sunlight during their stay at the Lunares Analog Research Station in Pila, Poland (Mlynczak et al., 2019), MARS-500 analog mission in Moscow, Russia (Russian Institute for Biomedical Problems) (; Vigo et al., 2013; Vigo et al., 2012), or the SIRIUS isolation campaigns at the Institute of Biomedical Problems, Moscow, Russia (Moller et al., 2025). The data collected can be utilized to monitor the subjective perception of time and the biological clock during isolation from sunlight, which can lead to sleep disorders, changes in circadian rhythms, and concentration problems (Mlynczak et al., 2019; Vigo et al., 2013; Vigo et al., 2012).
Furthermore, it is crucial to measure and analyze cardiovascular and respiratory variables during simulations of high-G acceleration experienced during launch and re-entry of suborbital spaceflights. This simulation can be conducted using a 7.5-meter radius centrifuge at the Royal Air Force High G Training and Test Facility (RAF Cranwell, UK), where a representative F-35 Lightning II cockpit is installed in the nacelle with a seatback angle of 22°. Dynamic combinations of acceleration and air breathing or 15% oxygen to simulate a cabin pressure altitude of 2438 meters can be tested in this setup (Smith et al., 2022). A lower-intensity rotational protocol recorded wrist pulse-wave morphology during stepwise 0°–360° body rotation (Jagtap et al., 2023). This setup may be useful for studying acute posture- and hydrostatic-gradientdependent pulse-wave changes, but it does not reproduce high-G exposure or sustained microgravity.
5.3 Gravitational unloading environments
One of the most frequently used analogs for studying the effects of altered gravity on the human body’s cardiorespiratory system is the bed rest simulation, where the subject lies in a head-down tilt position (Rabineau et al., 2020; Hoffmann et al., 2022; Tahvanainen et al., 1992; ; Shankhwar et al., 2022; Whittle et al., 2022; Solbiati and Caiani, 2025; Solbiati et al., 2025; Kundu et al., 2025). The commonly used angles are -6° or -7° head-down tilt, representing microgravity. Additionally, a 22.2° head-up tilt can simulate Martian gravity, while a 9.5° head-up tilt is used for Lunar gravity simulation. During this supine position, subjects engage in various activities such as reading, eating, personal hygiene, or exercise, while cardiovascular deconditioning is closely monitored and evaluated throughout the experiments (Rabineau et al., 2020; Hoffmann et al., 2022; Tahvanainen et al., 1992; Shankhwar et al., 2022; Whittle et al., 2022).
Similarly, dry immersion is used as a ground-based analog to simulate key unloading aspects of weightlessness and to study cardiovascular adaptations under controlled conditions. In this model, the participant is placed in thermoneutral water at approximately 33 °C while separated from direct contact with the water by a flexible elastic waterproof sheet, which eliminates almost all solid support and enables investigation of physiological effects of body unloading. Dry-immersion campaigns typically last 7–21 days and are often combined with head-up tilt protocols to assess cardiovascular responses to orthostatic challenge, with participants briefly removed from the bath once per day to allow for personal hygiene (; Zhedyaev et al., 2023).
Underwater analogs, such as the NASA Extreme Environment Mission Operations (NEEMO) facility in Key Largo, Florida, USA (Mundt et al., 2005) or Hydronaut H03 DeepLab in Jesenný, Czech Republic (Sokol et al., 2025), are used as space mission simulations that combine astronaut training with research in a body-unloading environment mimicking key operational aspects of EVAs, using the space station environment located at a depth of up to 18 meters. Besides that, underwater analogs can be valuable experimental scenario to evaluate ease of use, wearability, or trasmission reliability of health monitoring devices in isolated, confined, and extreme environment or during a simulated medical emergency (Mundt et al., 2005; Sokol et al., 2025).
For the most accurate simulation of the space environment, especially concerning the need to test equipment in a weightless state, experiments can be conducted in true microgravity environments (Montgomery et al., 2004; Mundt et al., 2005; ; ; ; Luchitskaya et al., 2013; Wiard et al., 2013; Migeotte et al., 2013; McCall et al., 2014). These tests can also assess the suitability of equipment for use in fractional-g environments such as the Moon or Mars (Wiard et al., 2013; McCall et al., 2014). Parabolic flights (; Luchitskaya et al., 2013; Migeotte et al., 2013; Montgomery et al., 2004; Wiard et al., 2013; McCall et al., 2014; ; ), are commonly used to create a microgravity environment. During parabolic flights, a state of weightlessness is achieved for about 20 seconds as a specially modified aircraft follows a parabola-shaped trajectory (Ngo-Anh et al., 2020). Various levels of gravity, including 1g, 0g, and 1.8g (), can be tested during these flights. However, the drawback of this approach is that weightlessness only lasts for a few seconds (typically 17 to 26 seconds) (; Luchitskaya et al., 2013; Migeotte et al., 2013; Montgomery et al., 2004; Wiard et al., 2013; McCall et al., 2014; ; ). Parabolic flights are conducted using aircraft such as the NASA KC-135 (Montgomery et al., 2004), a modified Boeing 727–200 (Wiard et al., 2013; McCall et al., 2014; ) or theAirbus A300-ZéroG operated by NOVESPACE (; Luchitskaya et al., 2013; Migeotte et al., 2013; ).
5.4 Space missions
After conducting a series of experiments in simulated conditions or in microgravity and meeting safety requirements, the equipment can be submitted for consideration to be sent to ISS. The ISS, the only permanently inhabited space station, saw its first module launched into orbit in 1998. Since November 2, 2000, when the first crew entered the station, the ISS has been continuously occupied, with crew rotations occurring every 6 months, and hosting numerous missions (Smith et al., 2020; ). Scientific experiments conducted on the ISS cover a wide range of areas, including investigations into plants, microorganisms, materials, and, notably, the physiology and health status of astronauts (). Cardiorespiratory signals, which have been the focus of several experiments in the past (Mastrandrea et al., 2025; Prisk et al., 2001; ; Gallasch et al., 1996; ; ; ; ; Kuipers, 1996; ; Shibata et al., 2023; ), are crucial for monitoring astronauts’ health and ongoing physiological changes in space.
Early spaceflight studies of cardiorespiratory function were conducted during the Austromir-91 mission aboard MIR, where experiments examined human physiological responses to short-term microgravity (Gallasch et al., 1996; Riedler, 1993). Further measurements were performed during the STS-55 mission in the Spacelab D-2 Anthrorack experiment, which focused on respiratory, circulatory, and endocrine responses in space (Prisk et al., 2001; Kuipers, 1996; Thomeczek et al., 1994). The STS-107 mission later extended this work by investigating microgravity-induced changes in cardiovascular and musculoskeletal function using experiments conducted in the Spacehab-RDM module (; Over et al., 2004). Long-duration monitoring was subsequently addressed during the Futura mission on the ISS, where wearable technologies were used to study physiological adaptation over nearly 200 days in orbit (; ; ; O’Sullivan, 2020).
6 Discussion
This review highlights both the progress and remaining limitations of cardiorespiratory monitoring technologies for space environments, where systems must provide reliable physiological information while remaining unobtrusive, robust, and compatible with operational constraints such as microgravity, electromagnetic interference, limited resources, and long-duration use. Although established methods such as ECG or BCG remain central, their future applicability depends on advances in wearable integration, comfort, noise suppression, and artifact-resistant signal processing. Several more advanced unobtrusive sensing approaches are being explored, but most have not yet been validated in actual space missions or terrestrial analog environments. Their gel-free contact, low power requirements, light weight, and breathability support integration into garments or suits for continuous monitoring during prolonged missions, as well as in terrestrial healthcare.
For example, textile and flexible sensing technologies offer promising routes toward more comfortable and unobtrusive cardiorespiratory monitoring. Textile-based ECG sensors, including metal-coated fabrics and graphene-based structures, can improve comfort, signal quality, and long-term wearability (Yang et al., 2022; Nigusse et al., 2021), while recent e-textile developments support fully integrated sensing garments for free-moving health and sport monitoring (Fu et al., 2024; Xu et al., 2025). Nanoengineered sensors based on conductive nanoparticles, carbon materials, flexible polymers, and emerging composite structures further enable multimodal pressure and strain sensing relevant to respiration, blood-flow-related signals, movement, and bending (Zazoum et al., 2022; Liu et al., 2023; ). In parallel, fiber-optic sensors provide an attractive alternative to conventional accelerometer-based BCG and SCG due to their electromagnetic immunity, high sensitivity, multiplexing capability, and potential integration into textiles or flexible substrates for unobtrusive cardiorespiratory monitoring (Zhang et al., 2023; Jha et al., 2024; Leal-Junior et al., 2024).
Looking ahead, generative AI and multimodal learning may further reduce hardware complexity by enabling reconstruction or estimation of multiple physiological signals from fewer sensing inputs. Recent studies have shown that ECG, ICG, or BP waveforms can be reconstructed from vibrational cardiography or PPG signals using generative models (Skoric et al., 2025; Sarkar and Etemad, 2021; Fang et al., 2025), potentially limiting the need for complex multi-lead systems. For spaceflight, however, such approaches must move beyond cloud-dependent processing toward autonomous, low-resource edge computing suitable for bandwidth-limited and delayed communication scenarios (; Polanco-Segovia et al., 2025; Lange et al., 2024). Future systems should therefore prioritize efficient onboard inference, adaptation to individual crew baselines, and interpretable outputs that can support medical decision-making without continuous ground support (Scott et al., 2023; Khan et al., 2025; Mani et al., 2024; ).
A limitation of this review is that the search strategy was primarily optimized for Q1, focusing on studies implementing cardiorespiratory sensing and monitoring technologies. The included studies were then further classified for Q2 and Q3; therefore, Q3 reflects only the analog environments represented within these monitoring studies rather than a systematic mapping of all space-analog platforms. Consequently, relevant studies indexed primarily by mission-, facility-, or program-specific terminology may not have been captured by the search strategy. However, several established analog environments not represented in the present review may still provide valuable testbeds for the development and validation of physiological monitoring technologies intended for spaceflight. These include, for example, Concordia Station at Dome C, Antarctica; Jules’ Undersea Lodge in Key Largo, Florida, USA; and the Human Exploration Research Analog (HERA) in Houston, Texas, USA (; Vannini and Vannini, 2025). A targeted evaluation of physiological monitoring across these and other analog platforms could therefore complement the present review and represents a relevant direction for future work.
Nevertheless, the review still reveals an important translational gap based on ISS-centered evidence, which may not fully represent the requirements of future commercial and deep-space missions. To our knowledge, relevant cardiorespiratory datasets from current commercial or exploration-class flight profiles remain unavailable, creating an important validation gap as the ISS approaches its planned decommissioning. Given the long development pathway from laboratory prototypes to space-qualified hardware, monitoring technologies for the post-ISS era should be advanced and tested under conditions reflecting the operational demands of emerging mission scenarios. Deepspace platforms such as Artemis and the Lunar Gateway will require greater autonomy, including edge-computing approaches capable of supporting onboard interpretation and decision-making under communication delays. Commercial stations and exploration missions may introduce new interoperability and infrastructure constraints, including higher radiation exposure and partial-gravity conditions (Okamura et al., 2024).
In conclusion, future research should therefore prioritize extensive validation of these emergent sensor technologies under realistic mission simulations, addressing concerns such as biocompatibility, mechanical durability, and seamless integration with spacecraft systems. Collaborative efforts among aerospace, biomedical, and materials science communities will be crucial to accelerate these developments, ensuring robust and reliable cardiorespiratory monitoring solutions. Together, these sensor innovations represent a crucial step toward proactive health surveillance in space, supporting longer and safer human spaceflight.
Statements
Author contributions
RM: Conceptualization, Methodology, Validation, Investigation, Resources, Writing – original draft, Funding acquisition. ML: Conceptualization, Methodology, Validation, Investigation, Writing – original draft, Visualization. KB: Conceptualization, Methodology, Validation, Investigation, Writing – original draft, Visualization. RJ: Conceptualization, Validation, Writing – original draft, Visualization. DV: Validation, Writing – original draft. AS: Validation, Writing – original draft. JB: Writing – review & editing. JK: Writing – review & editing. LD: Writing – review & editing. JP: Writing – review & editing. RVK: Conceptualization, Methodology, Validation, Writing – original draft, Supervision.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This article was co-funded by the European Union under the REFRESH – Research Excellence For REgion Sustainability and High-tech Industries project number CZ.10.03.01/00/22_003/0000048 via the Operational Programme Just Transition. This article was supported by the Ministry of Education of the Czechia (Project No. SP2026/012).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. The authors used OpenAI’s ChatGPT (GPT-5.5) solely for English language editing and phrasing improvements. All AI-assisted text was reviewed and edited by the authors, who take full responsibility for the manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AL-KhalidiF. Q.SaatchiR.BurkeD.ElphickH.TanS. (2011). Respiration rate monitoring methods: A review. Pediatr. Pulmonol.46, 523–529. doi: 10.1002/ppul.21416
2
AraiT.LimperU.GaugerP.BeckL. (2013). Pulse contour methods to estimate cardiovascular indices in micro-and hypergravity. Aviat. Space. Environ. Med.84, 1178–1185. doi: 10.3357/asem.3683.2013
3
ArbeilleP.ProvostR.VincentN.AubertA. (2014). Adaptation of the main peripheral artery and vein to long term confinement (mars 500). PloS One9, 1–7. doi: 10.1371/journal.pone.0083063
4
ArzoS. T.SikeridisD.DevetsikiotisM.GranelliF.FierroR.EsmaeiliM.et al. (2023). Essential technologies and concepts for massive space exploration: Challenges and opportunities. IEEE Trans. Aerosp. Electron. Syst.59, 3–29. doi: 10.1109/TAES.2022.3169126
5
AtesH. C.NguyenP. Q.Gonzalez-MaciaL.Morales-NarvaezE.GuderF.CollinsJ. J.et al. (2022). End-to-end design of wearable sensors. Nat. Rev. Mater.7, 887–907. doi: 10.1038/s41578-022-00460-x
6
BaevskyR. M.BaranovV. M.FuntovaI. I.DiedrichA.PashenkoA. V.ChernikovaA. G.et al. (2007). Autonomic cardiovascular and respiratory control during prolonged spaceflights aboard the international space station. J. Appl. Physiol.103, 156–161. doi: 10.1152/japplphysiol.00137.2007
7
BaevskyR. M.FuntovaI. I.LuchitskayaE. S.TankJ.BaevskyR. M.FuntovaI. I.et al. (2013). Microgravity: An ideal environment for cardiac force measuring. Cardiometry, 100–117. doi: 10.12710/cardiometry.2013.3.100117
8
BaevskyR. M.FuntovaI. I.LuchitskayaE. S. BaevskyR. M.FuntovaI. I.LuchitskayaE. S.. (2021). Role of the right and left parts of the heart in mechanisms of body adaptation to the conditions of long term space flight according to longitudinal ballistocardiography. Acta Astronaut.178, 894–899. doi: 10.1016/j.actaastro.2020.10.001
9
BalaliP.LuchitskayaE.HosseinA.AbdessaterE.FaoroV.DebeirO.et al. (2025). Evidence of effective cardiovascular countermeasures during spaceflights: Insights from wearable monitoring. NPJ Microgravity.11, 68. doi: 10.1038/s41526-025-00522-8
10
BaldiM.CerriG.ChiaraluceF.EusebiL.RussoP. (2014). Non-invasive uwb sensing of astronauts’ breathing activity. Sensors15, 565–591. doi: 10.3390/s150100565
11
BargerL. K.Flynn-EvansE. E.KubeyA.WalshL.RondaJ. M.WangW.et al. (2014). Prevalence of sleep deficiency and use of hypnotic drugs in astronauts before, during, and after spaceflight: An observational study. Lancet Neurol.13, 904–912. doi: 10.1016/S1474-4422(14)70122-X
12
Benitez-SalgadoA.Pena-CastilloM. A.Santiago-FuentesL. M.Zuniga-AvilesL. A.Abarca-CastroE. A.Talavera-PenaA. K.et al. (2024). Variations in heart rate variability and physiological responses during analog space missions: An exploratory study. Aerospace11, 833. doi: 10.3390/aerospace11100833
13
Benitez-SalgadoaA.Reyes-LagosaJ. J.Garduno-RodriguezcA. (2022). “ Analysis of short-term heart rate variability during training adaptation to mars-analog environment,” in 73rd International Astronautical Congress (IAC) (Paris: IAF), 1–4.
14
BorovikA. S.OrlovaE. A.TomilovskayaE. S.TarasovaO. S.VinogradovaO. L. (2020). Phase coupling between baroreflex oscillations of blood pressure and heart rate changes in 21-day dry immersion. Front. Physiol.11, 455. doi: 10.3389/fphys.2020.00455
15
BuchananG. F. (2013). Timing, sleep, and respiration in health and disease. Prog. Mol. Biol. Transl. Sci.119, 191–219. doi: 10.1016/B978-0-12-396971-2.00008-7
16
CastiglioniP.MeratiG.RizzoF.FainiA.ParatiG.FerratiniM.et al. (2006). A new wearable system for cardiorespiratory monitoring during gravitational stress. J. Gravitational. Physiol.13, 163–164.
17
CatchpoleJ. E. (2008). The International Space Station: Building for the Future (New York, NY: Springer Science & Business Media).
18
ChenX.ManshaiiF.TioranK.WangS.ZhouY.ZhaoJ.et al. (2024). Wearable biosensors for cardiovascular monitoring leveraging nanomaterials. Adv. Compos. Hybrid. Mater.7, 1–22. doi: 10.1007/s42114-024-00906-6
19
ChiY. M.JungT.-P.CauwenberghsG. (2010). Dry-contact and noncontact biopotential electrodes: Methodological review. IEEE Rev. Biomed. Eng.3, 106–119. doi: 10.1109/RBME.2010.2084078
20
CromwellR. L.HuffJ. L.SimonsenL. C.PatelZ. S. (2021). Earth-based research analogs to investigate space-based health risks. New Space.9, 204–216. doi: 10.1089/space.2020.0048
21
De la TorreG. G.GroemerG.Diaz-ArtilesA.PattynN.Van CutsemJ.MusilovaM.et al. (2024). Space analogs and behavioral health performance research review and recommendations checklist from esa topical team. NPJ Microgravity.10, 98. doi: 10.1038/s41526-024-00437-w
22
DeliereQ.MigeotteP.-F.NeytX.FuntovaI.BaevskyR. M.TankJ.et al. (2013). “ Cardiovascular changes in parabolic flights assessed by ballistocardiography,” in 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (Osaka: IEEE), 3801–3804. doi: 10.1109/EMBC.2013.6610372
23
De RidderS.MigeotteP.-F.NeytX.PattynN.PriskG. K.. (2011). “ Three-dimensional ballistocardiography in microgravity: A review of past research,” in 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Boston: IEEE), 4267–4270. doi: 10.1109/IEMBS.2011.6091059
24
DerobertmasureA.TohL. S.KafiL.ObeidH.BurkJ.DecadiA.et al. (2025). Exploring cardiovascular health and alternative biological fluid sampling methods in extreme environments: Lessons from a 12-day mars analog mission. Acta Astronaut.234, 194–206. doi: 10.1016/j.actaastro.2025.03.003
25
Di RienzoM.ParatiG.RizzoF.MeriggiP.MeratiG.FainiA.et al. (2007). “ Heart rate monitoring and control in altered gravity conditions,” in 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Lyon: IEEE), 6681–6684. doi: 10.1109/IEMBS.2007.4353893
26
Di RienzoM.VainiE.BrunoB.CastiglioniP.LombardiP.ParatiG.et al. (2014). “ Wearable seismocardiography: Towards the beat-to-beat assessment of cardiac mechanics during sleep in microgravity,” in 2014 8th Conference of the European Study Group on Cardiovascular Oscillations (ESGCO) (Trento: IEEE), 239–240. doi: 10.1109/ESGCO.2014.6847608
27
Di RienzoM.VainiE.LombardiP. (2015). “ Wearable monitoring: A project for the unobtrusive investigation of sleep physiology aboard the international space station,” in 2015 Computing in Cardiology Conference (CinC) (Nice: IEEE), 125–128. doi: 10.1109/CIC.2015.7408602
28
Di RienzoM.VainiE.LombardiP. (2017). An algorithm for the beat-to-beat assessment of cardiac mechanics during sleep on earth and in microgravity from the seismocardiogram. Sci. Rep.7, 1–12. doi: 10.1038/s41598-017-15829-0
29
Di RienzoM.VainiE.LombardiP.Di RienzoM.VainiE.LombardiP. (2018). Development of a smart garment for the assessment of cardiac mechanical performance and other vital signs during sleep in microgravity. Sensors. Actuators. A. Phys.274, 19–27. doi: 10.1016/j.sna.2018.02.034
30
Di RienzoM.PiccirilloS. (2021). “ Wearables for life in space,” in Wearable Sensors (London: Academic Press), 463–486. doi: 10.1016/B978-0-12-819246-7.00017-6
31
DrobczykM.LubkenA.StrowikC.AlbrechtU.-V.RustJ.BeringerJ.et al. (2022). A wireless communication network with a ballistocardiography experiment on the iss: Scenario, components and preflight demonstration. IEEE J. Radio. Frequency. Identification.6, 258–268. doi: 10.1109/JRFID.2022.3166026
32
ESAEsa Images. Available online at: https://www.esa.int/ESA_Multimedia/Images (Accessed (December 5, 2025).
33
FangX.JinJ.WangH.LiuC.CaiJ.NieG.et al. (2025). “ PPGFlowECG: Latent rectified flow with cross-modal encoding for PPG-guided ECG generation and cardiovascular disease detection,” in Arxiv. (Ithaca, NY: arXiv) Version Number: 1. doi: 10.48550/ARXIV.2509.19774
34
FeiD.-Y.ZhaoX.BoancaC.HughesE.BaiO.MerrellR.et al. (2010). A biomedical sensor system for real-time monitoring of astronauts’ physiological parameters during extra-vehicular activities. Comput. Biol. Med.40, 635–642. doi: 10.1016/j.compbiomed.2010.05.001
35
FeteiraA. (2009). Negative temperature coefficient resistance (ntcr) ceramic thermistors: An industrial perspective. J. Am. Ceram. Soc92, 967–983. doi: 10.1111/j.1551-2916.2009.02990.x
36
FuW.YuB.JiD.ZhouZ.LiX.WangR.et al. (2024). Intelligent fibers and textiles for wearable biosensors. Responsive. Materials.2, 1–34. doi: 10.1002/rpm.20240018
37
GallaschE.RafoltD.MoserM.HindingerJ.EderH.WiespeinerG.et al. (1996). Instrumentation for assessment of tremor, skin vibrations, and cardiovascular variables in mir space missions. IEEE Trans. Biomed. Eng.43, 328–333. doi: 10.1109/10.486291
38
GilesD.DraperN.NeilW. (2016). Validity of the polar v800 heart rate monitor to measure rr intervals at rest. Eur. J. Appl. Physiol.116, 563–571. doi: 10.1007/s00421-015-3303-9
39
GlazkovaN.PodladchikovaT.GerzerR.StepanovaD. (2020). Non-invasive wearable ecg-patch system for astronauts and patients on earth. Acta Astronaut.166, 613–618. doi: 10.1016/j.actaastro.2019.01.036
40
GopalsamyC.ParkS.RajamanickamR.JayaramanS. (1999). The wearable motherboard™: The first generation of adaptive and responsive textile structures (arts) for medical applications. Virtual. Reality.4, 152–168. doi: 10.1007/BF01418152
41
GrahamB. L.SteenbruggenI.MillerM. R.BarjaktarevicI. Z.CooperB. G.HallG. L.et al. (2019). Standardization of spirometry 2019 update. an official american thoracic society and european respiratory society technical statement. Am. J. Respir. Crit. Care Med.200, e70–e88. doi: 10.1164/rccm.201908-1590st
42
GuptaP.MoghimiM. J.JeongY.GuptaD.InanO. T.AyaziF. (2020). Precision wearable accelerometer contact microphones for longitudinal monitoring of mechano-acoustic cardiopulmonary signals. NPJ Digital. Med.3, 19. doi: 10.1038/s41746-020-0225-7
43
GuptaR.GhoshP. S. (2025). Advancements in health monitoring technologies for astronauts in deep space missions: A review. Life Sci. Space. Res.47, 190–196. doi: 10.1016/j.lssr.2025.06.011
44
HillJ. R.CaldwellB. S.DownsM.MillerM. J.LimD. S. S. (2018). Remote physiological monitoring in a mars analog field setting. IISE. Trans. Healthcare. Syst. Eng.8, 227–236. doi: 10.1080/24725579.2018.1501624
45
HoenemannJ.-N.MoestlS.DiedrichA.MulderE.FrettT.PetratG.et al. (2023). Impact of daily artificial gravity on autonomic cardiovascular control following 60-day head-down tilt bed rest. Front. Cardiovasc. Med.10, 1250727. doi: 10.3389/fcvm.2023.1250727
46
HoffmannB.DehkordiP.Khosrow-KhavarF.GoswamiN.BlaberA. P.TavakolianK. (2022). Mechanical deconditioning of the heart due to long-term bed rest as observed on seismocardiogram morphology. NPJ Microgravity.8, 1–9. doi: 10.1038/s41526-022-00206-7
47
HughsonR. L.RobertsonA. D.ArbeilleP.ShoemakerJ. K.RushJ. W. E.FraserK. S.et al. (2016). Increased postflight carotid artery stiffness and inflight insulin resistance resulting from 6-mo spaceflight in male and female astronauts. Am. J. Physiol. Heart Circ. Physiol.310, 628–638. doi: 10.1152/ajpheart.00802.2015
48
HughsonR. L.PetersonS. D.YeeN. J.GreavesD. K. (2017). Cardiac output by pulse contour analysis does not match the increase measured by rebreathing during human spaceflight. J. Appl. Physiol.123, 1145–1149. doi: 10.1152/japplphysiol.00651.2017
49
HussainT.UllahS.Fernandez-GarciaR.GilI. (2023). Wearable sensors for respiration monitoring: A review. Sensors23, 7518. doi: 10.3390/s23177518
50
ImholzB. P. M.WielingW.van MontfransG. A.WesselingK. H. (1998). Fifteen years experience with finger arterial pressure monitoring: Assessment of the technology. Cardiovasc. Res.38, 605–616. doi: 10.1016/S0008-6363(98)00067-4
51
InanO. T.MigeotteP.-F.ParkK.-S.EtemadiM.TavakolianK.CasanellaR.et al. (2014). Ballistocardiography and seismocardiography: A review of recent advances. IEEE J. Biomed. Health Inf.19, 1414–1427. doi: 10.1109/JBHI.2014.2361732
52
JagtapS.KumarA.MahaleB.DixitJ.KalangeA. E.KanawadeR.et al. (2023). Response of cardiac pulse parameters in humans at various inclinations via 360 rotating platform for simulated microgravity perspective. NPJ Microgravity.9, 54. doi: 10.1038/s41526-023-00301-3
53
JarosR.KoutnyJ.LadrovaM.MartinekR. (2023). Novel phonocardiography system for heartbeat detection from various locations. Sci. Rep.13, 14392. doi: 10.1038/s41598-023-41102-8
54
JhaR.MishraP.KumarS. (2024). Advancements in optical fiber-based wearable sensors for smart health monitoring. Biosens. Bioelectron.254, 1–16. doi: 10.1016/j.bios.2024.116232
55
JirakP.MirnaM.RezarR.MotlochL. J.LichtenauerM.JordanJ.et al. (2022). How spaceflight challenges human cardiovascular health. Eur. J. Prev. Cardiol.29, 1399–1411. doi: 10.1093/eurjpc/zwac029
56
KaufmannI.VeresT.BuchheimJ.-I.ChoukerA. (2020). “ Considerations for development and application of health monitoring tools in space,” in Stress Challenges and Immunity in Space: From Mechanisms to Monitoring and Preventive Strategies (Cham: Springer Cham), 407–420. doi: 10.1007/978-3-030-16996-1_21
57
KhanS. U. E.VargheseR. J.KassanosP.FarinaD.BurdetE. (2025). Space physiology and technology: Adaptations, countermeasures, and opportunity for wearable systems. Cyborg. Bionic. Syst.7, cbsystems.0477. doi: 10.34133/cbsystems.0477
58
KuipersA. (1996). First results from experiments performed with the esa anthrorack during the d-2 spacelab mission. Acta Astronaut.38, 865–875. doi: 10.1016/S0094-5765(96)00072-0
59
KunduC.SharmaS.GhoshT.DeepakK. K.DasA. (2025). Short-term 6° head-down tilt bed rest simulating microgravity ameliorates blood pressure and heart rate variability: A quasi-experimental study on healthy adults. Life Sci. Space. Res. 48, 225–231. doi: 10.1016/j.lssr.2025.11.014
60
LangeK.FontanaF.RossiF.VarileM.ApruzzeseG. (2024). “ Machine learning in space: Surveying the robustness of on-board ML models to radiation,” in 2024 IEEE Space Computing Conference (SCC) (Mountain View, CA: IEEE), 51–64. doi: 10.1109/SCC61854.2024.00012
61
Leal-JuniorA.SilvaJ.MacedoL.MarchesiA.MorauS.ValentinoJ.et al. (2024). The role of optical fiber sensors in the new generation of healthcare devices: A review. Sensors. Diag.3, 1135–1158. doi: 10.1039/D4SD00032C
62
LiuY.NortonJ. J. S.QaziR.ZouZ.AmmannK. R.LiuH.et al. (2016). Epidermal mechano-acoustic sensing electronics for cardiovascular diagnostics and human-machine interfaces. Sci. Adv.2, e1601185. doi: 10.1126/sciadv.1601185
63
LiuE.CaiZ.YeY.ZhouM.LiaoH.YiY. (2023). An overview of flexible sensors: Development, application, and challenges. Sensors23, 1–18. doi: 10.3390/s23020817
64
LuchitskayaE.DeliereQ.DiedrichA.PattynN.AlmoradA.BeckL.et al. (2013). “ Timing and source of the maximum of the transthoracic impedance cardiogram (dz/dt) in relation to the hij complex of the longitudinal ballistocardiogram under gravity and microgravity conditions,” in 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (Osaka: IEEE), 7294–7297. doi: 10.1109/EMBC.2013.6611242
65
MacQuarrieA.SidhuJ.DeetlefsC.WhitfieldS.StainerM. (2024). Establishing the validity and reliability of the astroskin® biometric shirt. Topics. Exercise Sci. Kinesiol.5, 1–12. doi: 10.9741/2766-2187.1078
66
MaioloL.MaitaF.CastielloA.MinottiA.PecoraA.. (2017). “ Highly wearable wireless wristband for monitoring pilot cardiac activity and muscle fine movements,” in 2017 IEEE international workshop on metrology for aerospace (metroaerospace) (Padua: IEEE), 271–275. doi: 10.1109/MetroAeroSpace.2017.7999578
67
ManiC.PaulT. S.ArchambaultP. M.MaroisA. (2024). Machine learning workflow for edge computed arrhythmia detection in exploration class missions. NPJ Microgravity.10, 71. doi: 10.1038/s41526-024-00409-0
68
MassaroniC.NicoloA.Lo PrestiD.SacchettiM.SilvestriS.SchenaE. (2019). Contact-based methods for measuring respiratory rate. Sensors19, 908. doi: 10.3390/s19040908
69
MastrandreaC. J.RabineauJ.GreavesD.HughsonR. L. (2025). Bio-monitor detects reduced obstructive sleep apnea and susceptibility to arrhythmia in spaceflight. Aerosp. Med. Hum. Perform.96, 1084–1089. doi: 10.3357/amhp.6745.2025
70
McCallC.StuartZ.WiardR. M.InanO. T.GiovangrandiL.CuttinoC. M.et al. (2014). “ Standing ballistocardiography measurements in microgravity,” in 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Chicago: IEEE), 5180–5183. doi: 10.1109/EMBC.2014.6944792
71
MigeotteP.-F.DeliereQ.TankJ.FuntovaI.BaevskyR.NeytX.et al. (2013). “ 3d-ballistocardiography in microgravity: Comparison with ground based recordings,” in 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (Osaka: IEEE), 7012–7016. doi: 10.1109/EMBC.2013.6611172
72
MlynczakM.KolodziejczykA.KrysztofiakH.AmbroszkiewiczG.ZylinskiM.CybulskiG. (2019). Cardiorespiratory profiling during simulated lunar mission using impedance pneumography. Biomed. Signal Process. Control.51, 216–221. doi: 10.1016/j.bspc.2019.02.015
73
MollerF. N.HoffmannU.FominaE.SteinbergF.Koschate-StormJ.PetersenL. G. (2025). Monitoring cardiorespiratory and cognitive performance during simulated spaceflight. Acta Astronaut.233, 242–249. doi: 10.1016/j.actaastro.2025.04.052
74
MontgomeryK.MundtC.ThonierG.TellierA.UdohU.BarkerV.et al. (2004). “ Lifeguard-a personal physiological monitor for extreme environments,” in The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society ( IEEE, San Francisco, USA), 2192–2195. doi: 10.1109/IEMBS.2004.1403640
75
MooreA. D.JrDownsM. E.FeivesonA. H.GuinedJ. R.LeeS. (2018). Evaluation of the Danish Aerospace Corporation Portable Pulmonary Function System. NASA Technical Memorandum NASA/TM-2018-220244. (Houston, TX: NASA Johnson Space Center).
76
MundtC. W.MontgomeryK. N.UdohU. E.BarkerV. N.ThonierG. C.TellierA. M.et al. (2005). A multiparameter wearable physiologic monitoring system for space and terrestrial applications. IEEE Trans. Inf. Technol. Biomed.9, 382–391. doi: 10.1109/TITB.2005.854509
77
NASANasa Image and Video Library. Available online at: https://images.nasa.gov (Accessed December 5, 2025).
78
Ngo-AnhT. J.RossiterA. (2020). “ Platforms for stress and immune research in preparation for long-duration space exploration missions,” in Stress Challenges and Immunity in Space: From Mechanisms to Monitoring and Preventive Strategies (Cham: Springer), 661–676. doi: 10.1007/978-3-030-16996-1_36
79
NigusseA. B.PurushothamanA. C.SelvamA.KumarC. (2021). Wearable smart textiles for long-term electrocardiography monitoring—a review. Sensors21, 1–24. doi: 10.3390/s21124174
80
O’SullivanJ.. (2020). “ Futura,” in European Missions to the International Space Station: 2013 to 2019 (Cham: Springer), 109–134. doi: 10.1007/978-3-030-30326-6_5
81
OkamuraY.GochiK.IshikawaT.HayashiT.FuseyaS.SuzukiR.et al. (2024). Impact of microgravity and lunar gravity on murine skeletal and immune systems during space travel. Sci. Rep.14, 28774. doi: 10.1038/s41598-024-79315-0
82
OverA. P.CassantoJ. M.CassantoV. A.DelucasL. J.ReichertP.OtilS. M.et al (2004). “ STS-107 mission after the mission: Recovery of data from the debris of Columbia,” in 42nd AIAA Aerospace Sciences Meeting and Exhibit. (Reston, VA: American Institute of Aeronautics and Astronautics), 1–9.
83
Polanco-SegoviaR.ZemerickS. A.MehtaP. M. (2025). A generalized software framework for deep neural network inference in space applications. IEEE Aerosp. Electron. Syst. Mag.40 (12), 1–11. doi: 10.1109/MAES.2025.3574111
84
PriskG. K.VerhaegheS.PadekenD.HamacherH.PaivaM. (2001). Three-dimensional ballistocardiography and respiratory motion in sustained microgravity. Aviat. Space. Environ. Med.72, 1067–1074.
85
PrzybylskiT.FroehleP.McDonaldC.MirzaeeM.NoghanianS.Fazel-RezaiR. (2015). “ Wearable wireless body area network for aeronautical applications,” in 2015 IEEE International Conference on Electro/Information Technology (EIT). (Dekalb: IEEE), 563–568. doi: 10.1109/EIT.2015.7293398
86
QuM.ChenX.YangD.LiD.ZhuK.GuoX.et al. (2022). Monitoring of physiological sounds with wearable device based on piezoelectric mems acoustic sensor. J. Micromech. Microeng.32, 014001. doi: 10.1088/1361-6439/ac371e
87
RabineauJ.HosseinA.LandreaniF.HautB.MulderE.LuchitskayaE.et al. (2020). Cardiovascular adaptation to simulated microgravity and countermeasure efficacy assessed by ballistocardiography and seismocardiography. Sci. Rep.10, 1–13. doi: 10.1038/s41598-020-74150-5
88
RiedlerW. (1993). The joint austro-soviet space project austromir-91. Space. Technol. Ind. Commercial. Appl.13, 175–180.
89
RovedaJ. M.FinkW.ChenK.WuW.-T. (2016). “ Psychological health monitoring for pilots and astronauts by tracking sleep-stress-emotion changes,” in 2016 IEEE Aerospace Conference (Big Sky: IEEE), 1–9. doi: 10.1109/AERO.2016.7500908
90
SabilA.GlosM.GuntherA.SchobelC.VeauthierC.FietzeI.et al. (2019). Comparison of apnea detection using oronasal thermal airflow sensor, nasal pressure transducer, respiratory inductance plethysmography and tracheal sound sensor. J. Clin. Sleep. Med.15, 285–292. doi: 10.5664/jcsm.7634
91
SarapaN. (2005). Digital 12-lead holter in the assessment of drug effects on cardiac repolarization. J. Electrocardiol.38, 1. doi: 10.1016/j.jelectrocard.2004.08.048
92
SarkarP.EtemadA. (2021). CardioGAN: Attentive generative adversarial network with dual discriminators for synthesis of ECG from PPG. Proc. AAAI. Conf. Artif. Intell.35, 488–496. doi: 10.1609/aaai.v35i1.16126
93
ScaliaT.BonventreL.TerranovaM. L. (2022). Devices for cardiovascular control: When space and earth tackle common challenges. Acta Astronaut.198, 660–668. doi: 10.1016/j.actaastro.2022.07.008
94
ScottR. T.SandersL. M.AntonsenE. L.HastingsJ. J. A.ParkS.-M.MackintoshG.et al. (2023). Biomonitoring and precision health in deep space supported by artificial intelligence. Nat. Mach. Intell.5, 196–207. doi: 10.1038/s42256-023-00617-5
95
SerranoR. R.-M.PicchioM. L.MantioneD.MecerreyesD.MalliarasG. G. (2026). Direct gravity-induced modulation of cardiac conduction pathways evidenced through propagation features in electrophysiological mapping. Adv. Mater. Technol.11 (12), e02642. doi: 10.1002/admt.202502642
96
ShankhwarV.SinghD.DeepakK. K. (2022). Effect of countermeasure bodygear on cardiac-vascular-respiratory coupling during 6-degree head-down tilt: An earth-based microgravity study. Life Sci. Space. Res.32, 45–53. doi: 10.1016/j.lssr.2021.10.004
97
ShibataS.WakehamD. J.ThomasJ. D.AbdullahS. M.PlattsS.BungoM. W.et al. (2023). Cardiac effects of long-duration space flight. J. Am. Coll. Cardiol.82, 674–684. doi: 10.1016/j.jacc.2023.05.058
98
SkoricJ.D’MelloY.PlantD. V. (2025). Generative reconstruction of multimodal cardiac waveforms from a single vibrational cardiography sensor. IEEE J. Biomed. Health Inf.29, 6576–6587. doi: 10.1109/JBHI.2025.3561071
99
SmithM. G.KelleyM.BasnerM. (2020). A brief history of spaceflight from 1961 to 2020: An analysis of missions and astronaut demographics. Acta Astronaut.175, 290–299. doi: 10.1016/j.actaastro.2020.06.004
100
SmithT. G.PollockR. D.BrittonJ. K.GreenN. D. C.HodkinsonP. D.MitchellS. J.et al. (2022). Physiological effects of centrifuge-simulated suborbital spaceflight. Aerosp. Med. Hum. Perform.93, 830–839. doi: 10.3357/AMHP.6153.2022
101
SokolM.VolfP.HejdaJ.LeovaL.H’yblJ.SchmirlerM.et al. (2025). Diana: An underwater analog space mission. Acta Astronaut.226, 349–360. doi: 10.1016/j.actaastro.2024.10.044
102
SolbiatiS.CaianiE. G. (2025). Heart rate variability adaptation due to head-down tilt bed rest: Insights from pooled data of six campaigns. NPJ Microgravity.11, 48. doi: 10.1038/s41526-025-00509-5
103
SolbiatiS.FiorentinoM. C.BendandiR.MocciaS.CaianiE. G. (2025). Ai-based prediction of vo2 max from 24-h holter ecg recording. NPJ Microgravity.11, 89. doi: 10.1038/s41526-025-00542-4
104
SörnmoL.LagunaP. (2005). Bioelectrical Signal Processing in Cardiac and Neurological Applications (Amsterdam: Academic Press).
105
TahvanainenK.LansimiesE.TikkanenP.HartikainenJ.KarkiT.LyyraT.et al. (1992). Microcomputer-based monitoring of cardiovascular functions in simulated microgravity. Adv. Space. Res.12, 227–236. doi: 10.1016/0273-1177(92)90287-8
106
TavakolianK.InanO. T.HahnJ.-O.Di RienzoM. (2022). Cardiac vibration signals: Old techniques, new tricks, and applications. Front. Physiol.13. doi: 10.3389/fphys.2022.931362
107
ThomeczekC.WilkeD.EsserP.WirtzM.PadekenD. (1994). The esa human physiology lab anthrorack: An overview on procedure development and crew training for the sts55/spacelab-d2 mission. Life. Sci. Res. Space.366, 497–501.
108
TittertonD. H.WestonJ. L. (2004). Strapdown Inertial Navigation Technology (Stevenage: Institution of Engineering and Technology).
109
TorabiY.ShiraniS.ReillyJ. P.GauvreauG. M. (2024). Mems and ecm sensor technologies for cardiorespiratory sound monitoring—a comprehensive review. Sensors24, 7036. doi: 10.3390/s24217036
110
Van MieghemC.SabbeM.KnockaertD. (2004). The clinical value of the ecg in noncardiac conditions. Chest125, 1561–1576. doi: 10.1378/chest.125.4.1561
111
VanniniP.VanniniA. (2025). Im-possible moorings and mobilities: The story of jules’ undersea lodge, the world’s only fully submerged hotel. Mobilities20, 719–733. doi: 10.1080/17450101.2024.2409639
112
VigoD. E.OgrinzB.WanL.BersenevE.TuerlinckxF.Van den BerghO.et al. (2012). Sleep-wake differences in heart rate variability during a 105-day simulated mission to mars. Aviat. Space. Environ. Med.83, 125–130. doi: 10.3357/ASEM.3120.2012
113
VigoD. E.TuerlinckxF.OgrinzB.WanL.SimonelliG.BersenevE.et al. (2013). Circadian rhythm of autonomic cardiovascular control during mars500 simulated mission to mars. Aviat. Space. Environ. Med.84, 1023–1028. doi: 10.3357/ASEM.3612.2013
114
Villa-ColinJ.ShawT.ToscanoW.CowingsP. (2018). Evaluation of astroskin bio-monitor during high intensity physical activities. Memorias. del. Congreso. Nacional. Ingenieria. Biomedica.5, 262–265. doi: 10.24254/CNIB.18.49
115
WangY.DongD.ZhouW.HeJ. (2025). Flexible wearable device applications for monitoring astronaut health: Current status and challenges. Wearable. Electron.2, 77–84. doi: 10.1016/j.wees.2024.12.007
116
WebsterJ. G. (2009). Medical Instrumentation: Application and Design (Hoboken, NJ: John Wiley & Sons).
117
WhittleR. S.StapletonL. M.PetersenL. G.Diaz-ArtilesA. (2022). Indirect measurement of absolute cardiac output during exercise in simulated altered gravity is highly dependent on the method. J. Clin. Monit. Comput.36, 1355–1366. doi: 10.1007/s10877-021-00769-y
118
WiardR. M.InanO. T.GiovangrandiL.CuttinoC. M.KovacsG. T. A. (2013). “ Preliminary results from standing ballistocardiography measurements in microgravity,” in 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (Osaka: IEEE), 7290–7293. doi: 10.1109/EMBC.2013.6611241
119
XuZ.ZhangC.WangF.YuJ.YangG.SurmenevR. A.et al. (2025). Smart textiles for personalized sports and healthcare. Nano. Micro. Lett.17, 1–39. doi: 10.1007/s40820-025-01749-6
120
YangY.WeiY.GuoZ.HouW.LiuY.TianH.et al. (2022). From materials to devices: Graphene toward practical applications. Small. Methods6, 1–27. doi: 10.1002/smtd.202200671
121
ZazoumB.BatooK. M.KhanM. A. A. (2022). Recent advances in flexible sensors and their applications. Sensors22, 1–17. doi: 10.3390/s22124653
122
ZhangL.ZhaiG. J.YanH.XuZ.LiangZ. G.SongJ. Z.et al. (2016). The research of application of noncontact ecg detection system in manned spaceflight. Man-Machine-Environment. System. Eng.406, 211–219. doi: 10.1007/978-981-10-2323-1_24
123
ZhangX.WangC.ZhengT.WuH.WuQ.WangY. (2023). Wearable optical fiber sensors in medical monitoring applications: A review. Sensors23, 1–39. doi: 10.3390/s23156671
124
ZhedyaevR. Y.TarasovaO. S.SharovaA. P.TomilovskayaE. S.VinogradovaO. L.BorovikA. S. (2023). Diverse effects of seven-day dry immersion on hemodynamic responses in head-up tilt and lower body negative pressure tests. Acta Astronaut.208, 105–110. doi: 10.1016/j.actaastro.2023.04.005
Summary
Keywords
astronaut health monitoring, cardiorespiratory signals, human spaceflight, microgravity, wearable sensors
Citation
Martinek R, Ladrova M, Barnova K, Jaros R, Vilimek D, Svoboda A, Brablik J, Kolarik J, Danys L, Platos J and Vilimkova Kahankova R (2026) Cardiorespiratory monitoring: from laboratory to space. Front. Physiol. 17:1956853. doi: 10.3389/fphys.2026.1956853
Received
03 August 2026
Revised
02 September 2026
Accepted
09 September 2026
Published
03 November 2026
Volume
17 - 2026
Edited by
Costantino Balestra, Haute École Bruxelles-Brabant (HE2B), Belgium
Reviewed by
Miroslav Rozloznik, University of Ostrava, Czechia
Riccardo Pelliccia, Dan Europe Foundation, Italy
Updates
Copyright
© 2026 Martinek, Ladrova, Barnova, Jaros, Vilimek, Svoboda, Brablik, Kolarik, Danys, Platos and Vilimkova Kahankova.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Martina Ladrova, martina.ladrova@vsb.cz
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.