<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article article-type="research-article" dtd-version="2.3" xml:lang="EN" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Robot. AI</journal-id>
<journal-title>Frontiers in Robotics and AI</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Robot. AI</abbrev-journal-title>
<issn pub-type="epub">2296-9144</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="publisher-id">778594</article-id>
<article-id pub-id-type="doi">10.3389/frobt.2022.778594</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Robotics and AI</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>Development of a Real-Time Thermoplastic Mask Compression Force Monitoring System Using Capacitive Force Sensor</article-title>
<alt-title alt-title-type="left-running-head">Kim et al.</alt-title>
<alt-title alt-title-type="right-running-head">Thermoplastic Mask Setup Monitoring System</alt-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname>Kim</surname>
<given-names>Tae-Ho</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1429218/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Cho</surname>
<given-names>Min-Seok</given-names>
</name>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Shin</surname>
<given-names>Dong-Seok</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Shin</surname>
<given-names>Dong Ho</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="corresp" rid="c001">&#x2a;</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1430543/overview"/>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Kim</surname>
<given-names>Siyong</given-names>
</name>
<xref ref-type="aff" rid="aff3">
<sup>3</sup>
</xref>
<xref ref-type="corresp" rid="c001">&#x2a;</xref>
</contrib>
</contrib-group>
<aff id="aff1">
<sup>1</sup>
<institution>Proton Therapy Center</institution>, <institution>National Cancer Center</institution>, <addr-line>Goyang</addr-line>, <country>South Korea</country>
</aff>
<aff id="aff2">
<sup>2</sup>
<institution>Department of Radiation Oncology</institution>, <institution>Yongin Severance Hospital</institution>, <addr-line>Yongin</addr-line>, <country>South Korea</country>
</aff>
<aff id="aff3">
<sup>3</sup>
<institution>Department of Radiation Oncology</institution>, <institution>Virginia Commonwealth University</institution>, <addr-line>Virginia</addr-line>, <addr-line>VA</addr-line>, <country>United States</country>
</aff>
<author-notes>
<fn fn-type="edited-by">
<p>
<bold>Edited by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/136503/overview">Jungwon Yoon</ext-link>, Gwangju Institute of Science and Technology, South Korea</p>
</fn>
<fn fn-type="edited-by">
<p>
<bold>Reviewed by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/1610395/overview">Amre Eizad</ext-link>, Gwangju Institute of Science and Technology, South Korea</p>
<p>
<ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/832001/overview">Emilio Andreozzi</ext-link>, University of Naples Federico II, Italy</p>
</fn>
<corresp id="c001">&#x2a;Correspondence: Dong Ho Shin, <email>dongho@ncc.re.kr</email>; Siyong Kim, <email>siyongkim@gmail.com</email>
</corresp>
<fn fn-type="other">
<p>This article was submitted to Biomedical Robotics, a section of the journal Frontiers in Robotics and AI</p>
</fn>
</author-notes>
<pub-date pub-type="epub">
<day>06</day>
<month>07</month>
<year>2022</year>
</pub-date>
<pub-date pub-type="collection">
<year>2022</year>
</pub-date>
<volume>9</volume>
<elocation-id>778594</elocation-id>
<history>
<date date-type="received">
<day>20</day>
<month>09</month>
<year>2021</year>
</date>
<date date-type="accepted">
<day>31</day>
<month>05</month>
<year>2022</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#xa9; 2022 Kim, Cho, Shin, Shin and Kim.</copyright-statement>
<copyright-year>2022</copyright-year>
<copyright-holder>Kim, Cho, Shin, Shin and Kim</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/">
<p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p>
</license>
</permissions>
<abstract>
<p>
<bold>Purpose:</bold> Thermoplastic masks keep patients in an appropriate position to ensure accurate radiation delivery. For a thermoplastic mask to maintain clinical efficacy, the mask should wrap the patient&#x27;s surface properly and provide uniform pressure to all areas. However, to our best knowledge, no explicit method for achieving such a goal currently exists. Therefore, in this study, we intended to develop a real-time thermoplastic mask compression force (TMCF) monitoring system to measure compression force quantitatively. A prototype system was fabricated, and the feasibility of the proposed method was evaluated.</p>
<p>
<bold>Methods:</bold> The real-time TMCF monitoring system basically consists of four force sensor units, a microcontroller board (Arduino Bluno Mega 2560), a control PC, and an in-house software program. To evaluate the reproducibility of the TMCF monitoring system, both a reproducibility test using a micrometer and a setup reproducibility test using a head phantom were performed. Additionally, the reproducibility tests of mask setup and motion detection tests were carried out with a cohort of six volunteers.</p>
<p>
<bold>Results:</bold> The system provided stable pressure readings in all 10 trials during the sensor unit reproducibility test. The largest standard deviation (SD) among trials was about 36&#xa0;gf/cm<sup>2</sup> (&#x223c;2.4% of the full-scale range). For five repeated mask setups on the phantom, the compression force variation of the mask was less than 39&#xa0;gf/cm<sup>2</sup> (2.6% of the full-scale range). We were successful in making masks together with the monitoring system connected and demonstrated feasible utilization of the system. Compression force variations were observed among the volunteers and according to the location of the sensor (among forehead, both cheekbones, and chin). The TMCF monitoring system provided the information in real time on whether the mask was properly pressing the human subject as an immobilization tool.</p>
<p>
<bold>Conclusion:</bold> With the developed system, it is possible to monitor the effectiveness of the mask in real time by continuously measuring the compression force between the mask and patient during the treatment. The graphical user interface (GUI) of the monitoring system developed provides a warning signal when the compression force of the mask is insufficient. Although the number of volunteers participated in the study was small, the obtained preliminary results suggest that the system could ostensibly improve the setup accuracy of a thermoplastic mask.</p>
</abstract>
<kwd-group>
<kwd>thermoplastic mask head fixation</kwd>
<kwd>real-time motion control</kwd>
<kwd>head and neck cancer</kwd>
<kwd>motion prediction</kwd>
<kwd>capacitive force sensor</kwd>
</kwd-group>
<contract-num rid="cn001">2021M2E8A1049195</contract-num>
<contract-num rid="cn002">2010140-2 2110610-1</contract-num>
<contract-sponsor id="cn001">National Research Foundation of Korea<named-content content-type="fundref-id">10.13039/501100003725</named-content>
</contract-sponsor>
<contract-sponsor id="cn002">National Cancer Center<named-content content-type="fundref-id">10.13039/100008746</named-content>
</contract-sponsor>
</article-meta>
</front>
<body>
<sec id="s1">
<title>1 Introduction</title>
<p>For accurate delivery of radiation therapy, daily setup accuracy and reproducibility are critical, especially for high-precision techniques like intensity-modulated radiotherapy (<xref ref-type="bibr" rid="B17">Purdy, 2000</xref>; <xref ref-type="bibr" rid="B10">Karger et al., 2001</xref>). In head and neck radiation therapy, thermoplastic mask systems have been widely used, and there are ample studies on their effectiveness and efficiency in reducing patient setup errors (<xref ref-type="bibr" rid="B25">Willner et al., 1997</xref>; <xref ref-type="bibr" rid="B18">Sweeney et al., 1998</xref>; <xref ref-type="bibr" rid="B22">Tsai et al., 1999</xref>; <xref ref-type="bibr" rid="B7">Gilbeau et al., 2001</xref>; <xref ref-type="bibr" rid="B6">Fuss et al., 2004</xref>).</p>
<p>A thermoplastic mask is made of a plastic material and is mesh-like. The mask holds the head and neck still and in the exact position needed for radiation treatment. It makes the patient in a position to ensure accurate delivery of the radiation beam and helps ensure that the setup position is as accurate and as effective as possible. A typical procedure for using the thermoplastic mask is as follows:<list list-type="simple">
<list-item>
<p>1) Place an unformed thermoplastic mask in a heater (for example, a wet or dry bath) with the set temperature and time to make the mask flexible.</p>
</list-item>
<list-item>
<p>2) Take the mask out of the bath. In case of using a water bath quickly remove excess water for patient comfort.</p>
</list-item>
<list-item>
<p>3) Check if the temperature of the mask is not too hot so it can be safely placed on the patient skin.</p>
</list-item>
<list-item>
<p>4) Position the mask on the patient face. Align the nose hole of the mask with the patient nose.</p>
</list-item>
<list-item>
<p>5) Mold the mask precisely around the nose bridge, chin, and eye sockets of the patient.</p>
</list-item>
<list-item>
<p>6) When the facial features have been precisely molded into the mask, start cooling down the mask.</p>
</list-item>
<list-item>
<p>7) When completely cooled, remove the mask.</p>
</list-item>
</list>
</p>
<p>Although these systems provide effective immobilization capabilities to a certain extent (<xref ref-type="bibr" rid="B4">Boda-Heggemann et al., 2006</xref>; <xref ref-type="bibr" rid="B21">Tryggestad et al., 2011</xref>), it is not unusual for them to suffer from noticeable daily variations in target alignment, especially when there are patient contour changes. Many studies, for this reason, have pointed out that it is not easy to perform accurate patient setups with a thermoplastic mask only. Therefore, it is recommended to use additional monitoring mechanisms such as x-ray-based guidance systems (for example, electronic portal imaging device, kV planar imaging, and/or cone-beam computed tomography) to reduce setup errors (<xref ref-type="bibr" rid="B10">Karger et al., 2001</xref>; <xref ref-type="bibr" rid="B9">Hong et al., 2005</xref>; <xref ref-type="bibr" rid="B19">Tae-Ho et al., 2012</xref>).</p>
<p>However, x-ray-based monitoring has a potential risk from excessive radiation exposure and no feasibility of monitoring a patient during the treatment (<xref ref-type="bibr" rid="B12">Li et al., 2006</xref>). In order to avoid excessive radiation exposure, optical camera-based monitoring systems have been developed (<xref ref-type="bibr" rid="B5">Buatti et al., 1998</xref>; <xref ref-type="bibr" rid="B23">Wagner et al., 2007</xref>), but such systems actually monitor the thermoplastic mask&#x2013;&#x2013;not the patient skin directly (<xref ref-type="bibr" rid="B11">Kim et al., 2004</xref>; <xref ref-type="bibr" rid="B13">Linthout et al., 2006</xref>).</p>
<p>For thermoplastic masks to maintain clinical efficacy, they should wrap the patient surface well and apply appropriate pressure to all areas (<xref ref-type="bibr" rid="B14">Orfit Industries, 2022</xref>, Instructions for Use). Also, if possible, the compression force of the mask is desired not to change through the whole treatment period. However, it may become inconsistent due to various causes such as deformation of the mask itself, patient contour change, and motion. These changes can reduce the ability of the mask as an immobilization tool and increase the magnitude of patient positioning error. Therefore, it is desirable to quantify the compression force of a mask and maintain proper compression force throughout the entire procedure to our best knowledge; however, no explicit and widely accepted method for achieving such a goal currently exists.</p>
<p>In a previous study, our research group performed basic tests on whether real-time compression force monitoring could be useful for quality assurance (QA) of immobilization tools used in radiation therapy. Through the study, a new monitoring method was proposed (a method different from the existing x-ray- or camera-based monitoring), and the possibility of improving patient setup accuracy was presented (<xref ref-type="bibr" rid="B26">Cho et al., 2016</xref>). However, force sensors made of polymer materials that had been used in the previous study turned out being suffered from the drift that could last several minutes (increasing/decreasing trends of sensor output in response to constant stimulus).</p>
<p>In this study, therefore, we intended to develop a real-time thermoplastic mask compression force (TMCF) monitoring system using a capacitive force sensor that showed a small drift and was suitable for force monitoring in the human body (<xref ref-type="bibr" rid="B20">Tang et al., 2020</xref>). A prototype system was fabricated, and the feasibility of the proposed method was evaluated.</p>
</sec>
<sec id="s2">
<title>2 Materials and Methods</title>
<sec id="s2-1">
<title>2.1 Real-Time TMCF Monitoring System</title>
<p>The real-time TMCF monitoring system basically consists of four force sensor units, a microcontroller board (Arduino Bluno Mega 2560), a control PC, and an in-house software program (<xref ref-type="fig" rid="F1">Figure 1</xref>). In this study, a SingleTact&#x2122; capacitive force sensor manufactured by Pressure Profile System, Inc. (Glasgow, UK) was used. SingleTact&#x2122; is a capacitive force sensor with a parallel capacitive configuration, and it is considered to be suitable for this application due to its high accuracy and repeatability (<xref ref-type="bibr" rid="B20">Tang et al., 2020</xref>). SingleTact&#x2122; consists of two thin round polyimide electrode plates separated by a dielectric. The working principle of the sensor is based on flexible interlayer compression as a result of the force that arises from distance change between two electrodes causing capacitance variation (<xref ref-type="bibr" rid="B8">Graudone et al., 2020</xref>). SingleTact&#x2122; includes electronics providing an interface to the main controller. The supply voltage needs to be between 3.7&#xa0;V and 12&#xa0;V with an input current of 2.7&#xa0;mA. The range of the analog output voltage is between 0.5&#xa0;V and 1.5&#xa0;V. In terms of data transfer, SingleTact&#x2122; can accommodate more than 100&#xa0;Hz (<xref ref-type="bibr" rid="B24">Warsito et al., 2020</xref>). For real-time TMCF monitoring, a calibrated sensor CS8-10N with a diameter of 8&#xa0;mm was used. Calibrated sensors offer improved accuracy and linearity over standard sensors and come as a matched sensor plus electronics interface board providing a pre-configured system in terms of the electronic interface (<xref ref-type="fig" rid="F2">Figure 2</xref>.) and with calibration carried out for linear output. Pre-calibrated sensors can measure pressure up to over 1500&#xa0;gf/cm<sup>2</sup> with a resolution of about 0.2% of the full scale. The response time of the system is 1&#xa0;ms (<xref ref-type="bibr" rid="B16">Pressure Prole Syst., Inc, 2022a</xref>, User Manual).</p>
<fig id="F1" position="float">
<label>FIGURE 1</label>
<caption>
<p>Real-time TMCF monitoring system includes a force sensor, a microcontroller board, a control system PC, and data acquisition software.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g001.tif"/>
</fig>
<fig id="F2" position="float">
<label>FIGURE 2</label>
<caption>
<p>Sensor unit consists of a force sensor and pressure absorption layer (PAL). The total thickness of the sensor unit in this study is 3.0&#x2013;3.5&#xa0;mm.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g002.tif"/>
</fig>
<p>While the thickness of the sensor itself is 0.35&#xa0;mm, two sponge pads of about 3&#xa0;mm (that is, 1.5&#xa0;mm each), called pressure absorption layer (PAL), are added, one above and the other below, to improve contact sensitivity between the mask and patient skin as shown in <xref ref-type="fig" rid="F2">Figure 2</xref>. A thermoplastic mask, in an ideal situation, should be formed tightly to the patient skin without gaps. In reality, however, there are often small gaps (about 1&#xa0;mm). Therefore, an appropriate thickness of PAL can help improve the contact sensitivity between the mask and the patient skin. The PAL thickness needs to be adjusted depending on the position to be monitored and the degree of contact between the mask and the face. It is also instrumental to select an appropriate material for PAL.</p>
<p>Initially, we tried to use silicon material (Dragon Skin TM 30) with stiffness similar to that of skin (30 Shore A). However, its stiffness turned out too strong to be used for mask force monitoring (the sensor signal got saturated easily). Therefore, the PAL material was changed to a soft sponge (about 60 Shore OO, stiffness similar to that of sliced bread). In order to find an appropriate PAL thickness that can reliably measure the mask force within the active range of the sensor, the thickness of the sponge was increased by a 1-mm interval.</p>
<p>When the PAL is too thin (1&#xa0;mm), as compared to the optimal value (3&#xa0;mm), movements inside the mask would not be monitored effectively. When the PAL was 1&#xa0;mm, TMCF baseline values were observed between 80 and 120&#xa0;gf/cm<sup>2</sup>. As can be seen in <xref ref-type="fig" rid="F5">Figure 5</xref>, this value is the force when the soft sponge (PAL) is compressed by about 0.1&#xa0;mm. Considering the stiffness of the soft sponge, it can be seen that the value is an insufficient force to keep the TMCF stable.</p>
<p>Actually, in real-time monitoring conducted on six volunteers, it was confirmed that the lower the baseline of TMCF, the more difficult it is to maintain the compression force stable. In addition, it has been confirmed that if the TMCF falls below 100&#xa0;gf/cm<sup>2</sup> (that is, when TMCF is too low), the compression force can no longer be maintained and often drops to zero. In other words, 100&#xa0;gf/cm<sup>2</sup> seemed a threshold value when the mask was about to lose meaningful contact with the surface of the volunteer. Therefore, the value of 100&#xa0;gf/cm<sup>2</sup> was applied as the first tolerance. The current version of GUI displays a yellow light warning to alert the user when the real-time TMCF falls below 100&#xa0;gf/cm<sup>2</sup> (first tolerance). Also, if TMCF drops to &#x2018;0&#x27; (2nd tolerance), it is designed to give a red-light warning.</p>
<p>In this study, every mask was made by a single experienced radiation therapy technologist to keep consistency and about 3&#xa0;mm of PAL was found appropriate in effectively monitoring mask force. When the PAL is too thick (over 5&#xa0;mm PAL), as compared to the optimal value (3&#xa0;mm), the sensor would encounter frequent signal saturation. Thus, the total thickness of the force sensor unit (sensor and the optimal value of PAL) is about 3.35&#xa0;mm (0.35 &#x2b; 3&#xa0;mm). Note, in actual practice, the optimal PAL thickness may vary depending on what type of mask is used and how it is formed.</p>
<p>The sensor units were placed between the thermoplastic mask and the patient face at four sensing points (forehead, both cheeks, and chin). It is recommended to avoid bending a sensor if possible because when bent, one electrode would be in compression while the other in tension, which can cause shearing to occur between the two electrodes. When bending is unavoidable, it is recommended to keep the circular portion (the active sensing region) as flat as possible while applying the bend along the tail end of the sensor. It is also necessary to keep a bend radius greater than 3&#xa0;mm (<xref ref-type="bibr" rid="B15">Pressure Prole Syst., Inc, 2022b</xref>, SingleTact Mounting Methods). Therefore, caution is needed when placing sensors, especially on the curved parts of the patient face.</p>
<p>The analog signal produced by the capacitive sensor is digitized with an Arduino Bluno Mega 2560 and connected to a PC for data processing and analysis (<xref ref-type="fig" rid="F1">Figure 1</xref>). The in-house developed software program, written in LabView, continuously monitors compression forces at four points and performs data processing to indicate whether the pressure at each point is within tolerance or not.</p>
</sec>
<sec id="s2-2">
<title>2.2 Reproducibility Test of the Sensor Unit</title>
<p>In this test, we tried to evaluate whether the same pressure values were obtained repeatedly when the sensor unit was pressed with known forces. A micrometer with a resolution of 0.001&#xa0;mm was used to compress the PAL consistently. <xref ref-type="fig" rid="F3">Figure 3</xref> shows the experimental setting for the sensor unit reproducibility test. The procedure is as follows:<list list-type="simple">
<list-item>
<p>1) Place a piece of mask and the sensor unit between the anvil and spindle of the micrometer (<xref ref-type="fig" rid="F3">Figures 3A,B</xref>).</p>
</list-item>
<list-item>
<p>2) Adjust the distance between the anvil and spindle of the micrometer to the extent that the sensor signal is not measured.</p>
</list-item>
<list-item>
<p>3) Set the micrometer scale to zero (<xref ref-type="fig" rid="F3">Figure 3B</xref>).</p>
</list-item>
<list-item>
<p>4) Obtain data (sensor output values) according to different compressing distances with the interval of 0.1&#xa0;mm (<xref ref-type="fig" rid="F3">Figure 3C</xref>).</p>
</list-item>
</list>
</p>
<fig id="F3" position="float">
<label>FIGURE 3</label>
<caption>
<p>Experimental setting for the reproducibility test of the sensor unit. <bold>(A)</bold> Prepare materials for the test, the sensor unit, a piece of thermoplastic mask, and a micrometer; <bold>(B)</bold> place the sensor unit and the piece of mask between the anvil and spindle of the micrometer, and then set the scale to zero; <bold>(C)</bold> obtain sensor output values according to different compressing distances with the interval of 0.1&#xa0;mm.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g003.tif"/>
</fig>
<p>Sensor output was recorded every 100 milliseconds for 1&#xa0;min in each trial, and a total of 10 trials were made for all four sensor units.</p>
</sec>
<sec id="s2-3">
<title>2.3 Effect of Time on the Measurement Accuracy Test and Reproducibility Test of the Real Time TMCF Monitoring System With a Head Phantom</title>
<p>Constant compression was given for 20&#xa0;min (a time interval comparable to a typical radiation therapy session), and pressure change with time was measured.</p>
<p>In addition, a reproducibility test was performed to evaluate the reliability of the developed monitoring system prior to a volunteer test. Under an ideal condition, compression force inside the mask is expected not to vary in multiple uses. To closely simulate an ideal situation, a stability test was performed with a rigid head phantom. <xref ref-type="fig" rid="F4">Figure 4</xref> shows the experimental setting for the system reproducibility test. Through the test, it was tried to confirm whether the same compression force was obtained repeatedly when the rigid head phantom was immobilized with the system in multiple trials. In each trial, the data were acquired every 100 milliseconds for 3&#xa0;min (that is, 1800 samples for each trial), and then the mean value was recorded. This process was repeated for a total of five times, and the mean SD value was obtained.</p>
<fig id="F4" position="float">
<label>FIGURE 4</label>
<caption>
<p>Experimental setting for both the time effect and reproducibility test of the real-time TMCF monitoring system. A total of four sensors were attached to the face of a head phantom (forehead, both cheeks, and chin), and then, the phantom was immobilized by a thermoplastic mask.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g004.tif"/>
</fig>
</sec>
<sec id="s2-4">
<title>2.4 Thermoplastic Mask Setup Reproducibility Study Using the Developed Real-Time TMCF Monitoring System With Volunteers</title>
<p>A total of six healthy volunteers (five males and one female) were enrolled in this simulation test. The volunteer group included two medical physicists, three graduate students, and one from the general public. The age ranged from 26 to 41&#xa0;yrs.</p>
<p>For each volunteer, a thermoplastic mask was formed, and four force sensor units were attached to the protrusions of the face (that is, forehead, both cheekbones, and chin) using a transparent adhesive film (Tegaderm Film, 3M). After completing the mask setup, compression force values were monitored for 3&#xa0;min in real time, which was considered one session. Each volunteer took four sessions, and, in between, the mask was off and an interval of approximately 10&#xa0;min was given.</p>
</sec>
<sec id="s2-5">
<title>2.5 Compression Force Variation due to Motion Inside the Thermoplastic Mask (Simulation Study: Intended Motion Test)</title>
<p>To investigate the effectiveness of the system for detecting abrupt force/motion, volunteers were instructed to try to move intentionally to simulate situations where a relatively large force of motion and/or variation occurred within the mask. Attempts of intentional movement (for example, nod motion with short bursts of force) were made at 3 specific times (100, 300, and 500&#xa0;msec after monitoring started). Compression force values were observed for 1&#xa0;min and compared with the results of the previous test (that is, <xref ref-type="sec" rid="s2-4">Section 2.4</xref>).</p>
</sec>
</sec>
<sec id="s3">
<title>3 Results</title>
<sec id="s3-1">
<title>3.1 Reproducibility Test of the Sensor Unit</title>
<p>
<xref ref-type="table" rid="T1">Table 1</xref> shows pressure values according to the compressed distance of the force sensor unit. In this test, we evaluated how consistent the monitored pressure values were among 10 trials. While absolute SD values tended to increase as the units were compressed more, relative ones decreased down to about 2.4% (that is, about 36&#xa0;gf/cm<sup>2</sup>) of 1500&#xa0;gf/cm<sup>2</sup> (the max of full scale). Overall, less than about 5% of SD was observed with pressure values larger than approximately 300&#xa0;gf/cm<sup>2</sup>. It was also found that pressure reading got almost saturated when the units were compressed to 1.1&#xa0;mm or more.</p>
<table-wrap id="T1" position="float">
<label>TABLE 1</label>
<caption>
<p>Summary of pressure values according to the compressed distance of force sensor units.</p>
</caption>
<table>
<thead valign="top">
<tr>
<th rowspan="2" align="left">Compressed distance [mm]</th>
<th colspan="4" align="center">Mean &#xb1; SD [gf/cm<sup>2</sup>]</th>
</tr>
<tr>
<th align="center">Sensor &#x23; 1</th>
<th align="center">Sensor &#x23; 2</th>
<th align="center">Sensor &#x23; 3</th>
<th align="center">Sensor &#x23; 4</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">0</td>
<td align="center">0</td>
<td align="center">0</td>
<td align="center">0</td>
<td align="center">0</td>
</tr>
<tr>
<td align="left">0.1</td>
<td align="char" char="plusmn">87 &#xb1; 8</td>
<td align="char" char="plusmn">87 &#xb1; 9</td>
<td align="char" char="plusmn">85 &#xb1; 8</td>
<td align="char" char="plusmn">99 &#xb1; 7</td>
</tr>
<tr>
<td align="left">0.2</td>
<td align="char" char="plusmn">152 &#xb1; 8</td>
<td align="char" char="plusmn">160 &#xb1; 10</td>
<td align="char" char="plusmn">162 &#xb1; 15</td>
<td align="char" char="plusmn">160 &#xb1; 9</td>
</tr>
<tr>
<td align="left">0.3</td>
<td align="char" char="plusmn">250 &#xb1; 10</td>
<td align="char" char="plusmn">264 &#xb1; 11</td>
<td align="char" char="plusmn">259 &#xb1; 13</td>
<td align="char" char="plusmn">257 &#xb1; 15</td>
</tr>
<tr>
<td align="left">0.4</td>
<td align="char" char="plusmn">378 &#xb1; 9</td>
<td align="char" char="plusmn">395 &#xb1; 16</td>
<td align="char" char="plusmn">377 &#xb1; 11</td>
<td align="char" char="plusmn">385 &#xb1; 14</td>
</tr>
<tr>
<td align="left">0.5</td>
<td align="char" char="plusmn">493 &#xb1; 16</td>
<td align="char" char="plusmn">529 &#xb1; 15</td>
<td align="char" char="plusmn">515 &#xb1; 17</td>
<td align="char" char="plusmn">503 &#xb1; 20</td>
</tr>
<tr>
<td align="left">0.6</td>
<td align="char" char="plusmn">650 &#xb1; 24</td>
<td align="char" char="plusmn">703 &#xb1; 19</td>
<td align="char" char="plusmn">694 &#xb1; 27</td>
<td align="char" char="plusmn">673 &#xb1; 21</td>
</tr>
<tr>
<td align="left">0.7</td>
<td align="char" char="plusmn">849 &#xb1; 30</td>
<td align="char" char="plusmn">899 &#xb1; 21</td>
<td align="char" char="plusmn">874 &#xb1; 26</td>
<td align="char" char="plusmn">889 &#xb1; 27</td>
</tr>
<tr>
<td align="left">0.8</td>
<td align="char" char="plusmn">1045 &#xb1; 35</td>
<td align="char" char="plusmn">1089 &#xb1; 28</td>
<td align="char" char="plusmn">1055 &#xb1; 30</td>
<td align="char" char="plusmn">1098 &#xb1; 33</td>
</tr>
<tr>
<td align="left">0.9</td>
<td align="char" char="plusmn">1253 &#xb1; 32</td>
<td align="char" char="plusmn">1292 &#xb1; 30</td>
<td align="char" char="plusmn">1270 &#xb1; 29</td>
<td align="char" char="plusmn">1324 &#xb1; 31</td>
</tr>
<tr>
<td align="left">1.0</td>
<td align="char" char="plusmn">1441 &#xb1; 36</td>
<td align="char" char="plusmn">1514 &#xb1; 27</td>
<td align="char" char="plusmn">1456 &#xb1; 33</td>
<td align="char" char="plusmn">1513 &#xb1; 36</td>
</tr>
<tr>
<td align="left">&#x3e; 1.1</td>
<td align="center">1534.2</td>
<td align="center">1586</td>
<td align="center">1552</td>
<td align="center">1579</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>Mean values in <xref ref-type="table" rid="T1">Table 1</xref> are plotted according to compression distances in <xref ref-type="fig" rid="F5">Figure 5</xref>. As it is shown in the figure, it could be observed that the differences between the measured forces for corresponding compressions increase with the compression since the stiffness of the PAL sponge increases as it is compressed (even if the PAL was compressed evenly to 1&#xa0;mm, the pressure applied to the sensor gradually increased). However, as illustrated, a clear linearity between the obtained pressure and applied compression was obtained (R<sup>2</sup> value over 0.98 obtained for all sensors).</p>
<fig id="F5" position="float">
<label>FIGURE 5</label>
<caption>
<p>Plot of pressure values according to compression distances. A linear trend was observed in all four sensors within 1&#xa0;mm compression.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g005.tif"/>
</fig>
</sec>
<sec id="s3-2">
<title>3.2 Effect of Time on the Measurement Accuracy Test and Reproducibility Test of the Real-Time TMCF Monitoring System With a Head Phantom</title>
<p>The pressure drift of the SingleTact sensor fluctuates between the zero points, thus giving positive and negative results intermittently. No obvious trend of the pressure drift can be observed (<xref ref-type="fig" rid="F4">Figure 4</xref>).</p>
<p>In the time effect test with the head phantom, since there was neither contour change nor motion, constant compression forces of the mask were expected, and the system showed stable readings as shown in <xref ref-type="fig" rid="F6">Figure 6</xref>.</p>
<fig id="F6" position="float">
<label>FIGURE 6</label>
<caption>
<p>Compression force variations at four measuring spots inside the thermoplastic mask with a rigid head phantom for a 3-min trial. The output signal was maintained stable in all of the sensors.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g006.tif"/>
</fig>
<p>
<xref ref-type="table" rid="T2">Table 2</xref> shows the results of the mask setup reproducibility test in which a total of five 3-min trials were repeated using the rigid head phantom. Because setups were attempted with the rigid object, if a precise mask setup was performed, compression force variation should be minimal. As summarized in <xref ref-type="table" rid="T2">Table 2</xref>, the standard deviation of each measuring spot did not exceed 39 gf/cm<sup>2</sup> (2.6% of 1500&#xa0;gf/cm<sup>2</sup>).</p>
<table-wrap id="T2" position="float">
<label>TABLE 2</label>
<caption>
<p>Result summary of the thermoplastic mask setup reproducibility test with a rigid head phantom. In each trial, the data were acquired at every 100&#xa0;ms for 3&#xa0;min, and the data represent the mean values of 1,800 samples for each fraction.</p>
</caption>
<table>
<thead valign="top">
<tr>
<th rowspan="2" align="left">Fraction</th>
<th colspan="4" align="center">Pressure value [gf/cm<sup>2</sup>]</th>
</tr>
<tr>
<th align="center">Forehead</th>
<th align="center">Left cheekbone</th>
<th align="center">Right cheekbone</th>
<th align="center">Chin</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">1</td>
<td align="center">876</td>
<td align="center">715</td>
<td align="center">814</td>
<td align="center">494</td>
</tr>
<tr>
<td align="left">2</td>
<td align="center">893</td>
<td align="center">698</td>
<td align="center">731</td>
<td align="center">454</td>
</tr>
<tr>
<td align="left">3</td>
<td align="center">834</td>
<td align="center">641</td>
<td align="center">792</td>
<td align="center">417</td>
</tr>
<tr>
<td align="left">4</td>
<td align="center">818</td>
<td align="center">731</td>
<td align="center">756</td>
<td align="center">506</td>
</tr>
<tr>
<td align="left">5</td>
<td align="center">802</td>
<td align="center">655</td>
<td align="center">824</td>
<td align="center">503</td>
</tr>
<tr>
<td align="left">Mean &#xb1; SD</td>
<td align="center">844.6 &#xb1; 38</td>
<td align="center">688 &#xb1; 39</td>
<td align="center">783.4 &#xb1; 39</td>
<td align="center">474.8 &#xb1; 38</td>
</tr>
</tbody>
</table>
</table-wrap>
</sec>
<sec id="s3-3">
<title>3.3 Thermoplastic Mask Setup Reproducibility Study Using the Developed Real-Time TMCF Monitoring System With Volunteers</title>
<p>
<xref ref-type="fig" rid="F7">Figure 7</xref> shows the compression force variation of each volunteer (result of the first session). Volunteers were pressed using a customized mask, and then the compression force variation was monitored for 3&#xa0;min per session for four sessions. As can be seen in the figure, it was observed that the compression force varied in a noticeable amount with time in general unlike the cases of the previous test with the head phantom (<xref ref-type="fig" rid="F6">Figure 6</xref>). The compression force of the forehead was generally higher and more stable than that of the other parts in most volunteers. Compression force values for both cheekbones were also generally stable in all volunteers. However, in the case of the chin, the overall compression force was lower than that in other regions. In addition, its variation was more dramatic and sometimes reached down to zero (0).</p>
<fig id="F7" position="float">
<label>FIGURE 7</label>
<caption>
<p>Compression force variation of each volunteer in the first 3-min session. Various compression force variations were observed depending on the location of the sensor (forehead, both cheekbones, and chin) and volunteers.</p>
</caption>
<graphic xlink:href="frobt-09-778594-g007.tif"/>
</fig>
<p>
<xref ref-type="fig" rid="F8">Figure 8</xref> shows the variation range and mean value of compression force for each session. Volunteers 1, 2, 4, and 6 maintained high compression forces (&#x3e;800&#xa0;gf/cm<sup>2</sup>) at all sessions, and the compression forces of volunteers 3 and 5 also did not drop below 600&#xa0;gf/cm<sup>2</sup>. In the case of both cheekbones, the variation ranges were slightly larger than those of the forehead case, but all volunteers maintained high compression forces in general (over 500 gf/cm<sup>2</sup>) in all sessions. In the chin case, the mean values were lower than those of other sites, and the ranges were noticeably larger in general. However, variations among sessions were not significant.</p>
<fig id="F8" position="float">
<label>FIGURE 8</label>
<caption>
<p>Compression force variation range and mean value were acquired to evaluate thermoplastic mask setup reproducibility. Each volunteer took four sessions (S1&#x2013;S4 means each session number).</p>
</caption>
<graphic xlink:href="frobt-09-778594-g008.tif"/>
</fig>
</sec>
<sec id="s3-4">
<title>3.4 Compression Force Variation due to Motion Inside the Thermoplastic Mask (Simulation Study: Intended Motion Test)</title>
<p>
<xref ref-type="fig" rid="F9">Figure 9</xref> shows compression force variation due to intended motion inside the thermoplastic mask. As shown, noticeable changes in the compression force were observed at times when movement was instructed (that is, at 100, 300, and 500&#xa0;ms), indicating the capability of the system to monitor possible patient motion and/or displacement inside the mask.</p>
<fig id="F9" position="float">
<label>FIGURE 9</label>
<caption>
<p>Compression force variation due to intended motion inside the thermoplastic mask. The gray vertical dotted line indicates the time at which the intended motion was instructed (10, 30, and 50&#xa0;s after monitoring starts).</p>
</caption>
<graphic xlink:href="frobt-09-778594-g009.tif"/>
</fig>
<p>As mentioned previously, the GUI of the monitoring system is designed to give a yellow light warning signal when the real-time compression force value falls below 100&#xa0;gf/cm<sup>2</sup>. Also, if the compression force value drops to &#x201c;0&#x201d;, it is designed to give a red-light warning signal. Therefore, if the yellow light is on, the patient should be observed more carefully, and if the red light is on, a readjustment is recommended because it can be assumed that the contact between the mask and face is not appropriate.</p>
</sec>
</sec>
<sec id="s4">
<title>4 Discussion</title>
<p>The thermoplastic mask is considered an effective immobilization tool that can prevent patients from movement (<xref ref-type="bibr" rid="B25">Willner et al., 1997</xref>; <xref ref-type="bibr" rid="B18">Sweeney et al., 1998</xref>; <xref ref-type="bibr" rid="B22">Tsai et al., 1999</xref>; <xref ref-type="bibr" rid="B7">Gilbeau et al., 2001</xref>; <xref ref-type="bibr" rid="B6">Fuss et al., 2004</xref>). This clinical utility assumes that the compression force inside the mask is maintained without significant change during the treatment period. On the other hand, it is commonly agreed that the compression force inside the mask may vary during the treatment period due to various causes such as patient contour change, patient motion, and deformation of the mask itself. Nevertheless, such assumption that consistent compression force is maintained inside the mask has never been actively confirmed on a real-time basis. We believe this is mainly because there has been no system available that can provide relevant information (for example, compression force) in real time. We took a challenge to develop such a system and performed a feasibility study of the system with a cohort of volunteers in this study.</p>
<p>In this study, PAL was added above and below the sensor itself to suit better the compression force monitoring inside the thermoplastic mask. As PAL was added, it was confirmed that the linearity of the sensor was slightly lower than the inherent performance of the sensor. Although there was concern that PAL would adversely affect the inherent performance of the sensor, as shown in <xref ref-type="fig" rid="F5">Figure 5</xref>, it seems that there is no problem in monitoring the compression force of the mask (R<sup>2</sup> value over 0.98 was obtained for all sensors within 1&#xa0;mm compression).</p>
<p>However, since the thermoplastic mask is manufactured differently for each patient, in order to further improve the accuracy and usability of the system, further study about the optimization of PAL material, shape, and thickness may be necessary. For example, using a dome-shaped mechanical coupler attached to the sensor improved the transfer of forces from multiple directions to the active area of the sensor (<xref ref-type="bibr" rid="B3">Andreozzi et al., 2020</xref>; <xref ref-type="bibr" rid="B2">Andreozzi et al., 2021a</xref>; <xref ref-type="bibr" rid="B1">Andreozzi et al., 2021b</xref>). A design (for example, truncated cone shape) that can effectively place compression force to the center of the sensor area may also be a good solution.</p>
<p>Through the thermoplastic mask setup reproducibility test, we confirmed the reliability of the developed monitoring system (<xref ref-type="table" rid="T2">Table 2</xref>) prior to a volunteer test. We also confirmed that the developed monitoring system could keep the output signal stable if there was no change inside the mask (<xref ref-type="fig" rid="F6">Figure 6</xref>).</p>
<p>As shown in <xref ref-type="fig" rid="F7">Figure 7</xref>, various compression force changes were observed depending on the location of the sensor (that is, forehead, both cheekbones, and chin) and volunteer. In case of forehead and both cheekbones, the compression force was generally stable in all volunteers. This means that the mask limits the movement of the forehead and both cheekbones by relatively constant and large forces (forehead &#x2265;600&#xa0;gf/cm2, chin &#x2265;500&#xa0;gf/cm2). Such variation trend was observed in all four sessions (<xref ref-type="fig" rid="F8">Figure 8</xref>). Also, in the simulation test performed assuming significant moving forces inside the mask, the baseline of the compression force did not significantly change or drop to zero before and after the peak value. In the case of the chin, the baseline of the compression force was lower than that of the other sites, and especially with volunteers 4, 5, and 6, the compression force often dropped close to zero (<xref ref-type="fig" rid="F7">Figure 7</xref>, chin case). A &#x201c;zero&#x201d; compression force means that there is a gap between the mask and the patient surface, and the mask could no longer play a proper role as an immobilization tool. Such behaviors were consistent through multiple sessions as shown in <xref ref-type="fig" rid="F8">Figure 8</xref>. Compression force dropping to &#x201c;0&#x201d; was more prominent in the intended motion test (<xref ref-type="fig" rid="F9">Figure 9</xref>), implying that either the chin area is a weak point in mask fabrication or more moving force is given to the chin area by patients or both. Therefore, using more caution on the chin area may be necessary, and close monitoring would be highly beneficial.</p>
<p>The aforementioned TMCF variation trend was limited to the volunteers who participated in this study. It was difficult to generalize these results since the number of volunteers participating in the study was not large enough. The TMCF baseline and variation trend can be affected by various factors such as the skill of the radiologist making the mask, the characteristics of the patient surface, whether the patient can maintain the setup without moving, and the location of the sensor. In addition, since the patient is not a rigid object, it is natural for TMCF to change in real time.</p>
<p>TMCF variation does not necessarily mean that the patient setup position has changed. If the TMCF baseline is maintained above a certain value (100&#xa0;gf/cm<sup>2</sup>), the mask performs an appropriate role as an immobilization tool. Therefore, in this study, we focused on checking whether the mask force is too weak (falls below 100&#xa0;gf/cm<sup>2</sup>) or falls to &#x2018;0&#x2032; to assure the mask is performing an appropriate role as an immobilization tool. However, with existing monitoring equipment, it is not easy to check the compression force inside the mask.</p>
<p>The monitoring system developed in this study can provide information in real time on whether the thermoplastic mask is properly pressing the patient as an immobilization tool. With such information, users can take appropriate actions both in real time and patient-specific level. The GUI of the monitoring system is designed to give a warning signal when there is a problem with the compression force of the mask. For instance, if the yellow light is on, the patient should be observed more carefully, and if the red light is on, a readjustment is recommended.</p>
<p>The ultimate goal of an immobilization tool is complete prevention of patient positioning errors due to mobility. The real-time TMCF monitoring system developed in this study is expected to be of great help in maintaining the clinical utility of the immobilization tool. This study assumed that if it is possible to keep the compression force of the mask consistent throughout the whole treatment course, it would help improve the accuracy of head and neck cancer radiation treatment. In this early study, we verified the feasibility of the system with a cohort of six volunteers only. In further studies, we are planning to investigate the functionality and effectiveness of the system with a cohort of actual patients.</p>
</sec>
<sec id="s5">
<title>5 Conclusion</title>
<p>We developed a real-time TMCF monitoring system. With the developed system, it is possible to monitor the effectiveness of the mask in real time by continuously measuring the compression force between the mask and patient during the treatment. The monitoring system can provide information in real time on whether the thermoplastic mask is properly pressing the patient as an immobilization tool. Although the number of volunteers participating in the study was small, these preliminary results suggest that the system could ostensibly improve the setup accuracy of thermoplastic masks.</p>
</sec>
</body>
<back>
<sec id="s6" sec-type="data-availability">
<title>Data Availability Statement</title>
<p>The original contributions presented in the study are included in the article/Supplementary Material; further inquiries can be directed to the corresponding authors.</p>
</sec>
<sec id="s7">
<title>Ethics Statement</title>
<p>The studies involving human participants were reviewed and approved by the Catholic University of Korea&#x2019;s CMC Clinical Research Coordinating Center. The patients/participants provided their written informed consent to participate in this study.</p>
</sec>
<sec id="s8">
<title>Author Contributions</title>
<p>Study concept and design: T-HK, DS, and SK. Data acquisition: T-HK, M-SC, and D-SS. Quality control of data: T-HK, M-SC, and DS. Data analysis: T-HK, D-SS, DS, and SK. Manuscript preparation and editing: T-HK and SK. Manuscript review: all authors. All authors read and approved the final manuscript.</p>
</sec>
<sec id="s9">
<title>Funding</title>
<p>This research was funded by the National Research Foundation of the Korean National Cancer Center Fund (Grant Nos. 2010140-2 and 2110610-1) and supported by the National Research Foundation of Korea (NRF) funded by the Ministry of Science and ICT (No. 2021M2E8A1049195).</p>
</sec>
<sec sec-type="COI-statement" id="s10">
<title>Conflict of Interest</title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="disclaimer" id="s11">
<title>Publisher&#x2019;s Note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<ref-list>
<title>References</title>
<ref id="B1">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andreozzi</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Gargiulo</surname>
<given-names>G. D.</given-names>
</name>
<name>
<surname>Esposito</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Bifulco</surname>
<given-names>P.</given-names>
</name>
</person-group> (<year>2021b</year>). <article-title>A Novel Broadband Forcecardiography Sensor for Simultaneous Monitoring of Respiration, Infrasonic Cardiac Vibrations and Heart Sounds</article-title>. <source>Front. Physiol.</source> <volume>12</volume>, <fpage>725716</fpage>. <pub-id pub-id-type="doi">10.3389/fphys.2021.725716</pub-id> </citation>
</ref>
<ref id="B2">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andreozzi</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Centracchio</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Punzo</surname>
<given-names>V.</given-names>
</name>
<name>
<surname>Esposito</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Polley</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Gargiulo</surname>
<given-names>G. D.</given-names>
</name>
<etal/>
</person-group> (<year>2021a</year>). <article-title>Respiration Monitoring via Forcecardiography Sensors</article-title>. <source>Sensors</source> <volume>21</volume>, <fpage>3996</fpage>. <pub-id pub-id-type="doi">10.3390/s21123996</pub-id> </citation>
</ref>
<ref id="B3">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andreozzi</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Fratini</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Esposito</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Naik</surname>
<given-names>G.</given-names>
</name>
<name>
<surname>Polley</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Gargiulo</surname>
<given-names>G. D.</given-names>
</name>
<etal/>
</person-group> (<year>2020</year>). <article-title>Forcecardiography: A Novel Technique to Measure Heart Mechanical Vibrations onto the Chest Wall</article-title>. <source>Sensors</source> <volume>20</volume>, <fpage>3885</fpage>. <pub-id pub-id-type="doi">10.3390/s20143885</pub-id> </citation>
</ref>
<ref id="B4">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Boda-Heggemann</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Walter</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Rahn</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Wertz</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Loeb</surname>
<given-names>I.</given-names>
</name>
<name>
<surname>Lohr</surname>
<given-names>F.</given-names>
</name>
<etal/>
</person-group> (<year>2006</year>). <article-title>Repositioning Accuracy of Two Different Mask systems&#x2014;3D Revisited: Comparison Using True 3D/3D Matching with Cone-Beam CT</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>66</volume>, <fpage>1568</fpage>&#x2013;<lpage>1575</lpage>. <pub-id pub-id-type="doi">10.1016/j.ijrobp.2006.08.054</pub-id> </citation>
</ref>
<ref id="B5">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Buatti</surname>
<given-names>J. M.</given-names>
</name>
<name>
<surname>Bova</surname>
<given-names>F. J.</given-names>
</name>
<name>
<surname>Friedman</surname>
<given-names>W. A.</given-names>
</name>
<name>
<surname>Meeks</surname>
<given-names>S. L.</given-names>
</name>
<name>
<surname>Marcus</surname>
<given-names>R. B.</given-names>
<suffix>Jr.</suffix>
</name>
<name>
<surname>Mickle</surname>
<given-names>J. P.</given-names>
</name>
<etal/>
</person-group> (<year>1998</year>). <article-title>Preliminary Experience with Frameless Stereotactic Radiotherapy</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>42</volume>, <fpage>591</fpage>&#x2013;<lpage>599</lpage>. <pub-id pub-id-type="doi">10.1016/s0360-3016(98)00276-4</pub-id> </citation>
</ref>
<ref id="B26">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cho</surname>
<given-names>M.-S.</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>T.-H.</given-names>
</name>
<name>
<surname>Kang</surname>
<given-names>S.-H.</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>D.-S.</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>K.-H.</given-names>
</name>
<name>
<surname>Shin</surname>
<given-names>D.-S.</given-names>
</name>
<etal/>
</person-group> (<year>2016</year>). <article-title>A Noble Technique a Using Force-Sensing Resistor for Immobilization-Device Quality Assurance: A Feasibility Study</article-title>. <source>J. Korean Phys. Soc.</source> <volume>68</volume>, <fpage>803</fpage>&#x2013;<lpage>809</lpage>. <pub-id pub-id-type="doi">10.3938/jkps.68.803</pub-id> </citation>
</ref>
<ref id="B6">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Fuss</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Salter</surname>
<given-names>B. J.</given-names>
</name>
<name>
<surname>Cheek</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Sadeghi</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Hevezi</surname>
<given-names>J. M.</given-names>
</name>
<name>
<surname>Herman</surname>
<given-names>T. S.</given-names>
</name>
</person-group> (<year>2004</year>). <article-title>Repositioning Accuracy of a Commercially Available Thermoplastic Mask System</article-title>. <source>Radiother. Oncol.</source> <volume>71</volume>, <fpage>339</fpage>&#x2013;<lpage>345</lpage>. <pub-id pub-id-type="doi">10.1016/j.radonc.2004.03.003</pub-id> </citation>
</ref>
<ref id="B7">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Gilbeau</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Octave-Prignot</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Loncol</surname>
<given-names>T.</given-names>
</name>
<name>
<surname>Renard</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Scalliet</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Gr&#xe9;goire</surname>
<given-names>V.</given-names>
</name>
</person-group> (<year>2001</year>). <article-title>Comparison of Setup Accuracy of Three Different Thermoplastic Masks for the Treatment of Brain and Head and Neck Tumors</article-title>. <source>Radiother. Oncol.</source> <volume>58</volume>, <fpage>155</fpage>&#x2013;<lpage>162</lpage>. <pub-id pub-id-type="doi">10.1016/s0167-8140(00)00280-2</pub-id> </citation>
</ref>
<ref id="B8">
<citation citation-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Graudone</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Apse-Apsitis</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Dabolina</surname>
<given-names>I.</given-names>
</name>
<name>
<surname>Porins</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Silina</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Lapkovska</surname>
<given-names>E.</given-names>
</name>
</person-group> (<year>2020</year>). &#x201c;<article-title>Research and Development of Clothing Fit by Sensor Matrix</article-title>,&#x201d; in <conf-name>2020 IEEE 61th International Scientific Conference on Power and Electrical Engineering of Riga Technical University (RTUCON)</conf-name>, <conf-loc>Riga, Latvia</conf-loc>, <conf-date>05-07 November 2020</conf-date>, <fpage>1</fpage>&#x2013;<lpage>4</lpage>. </citation>
</ref>
<ref id="B9">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hong</surname>
<given-names>T. S.</given-names>
</name>
<name>
<surname>Tom&#xe9;</surname>
<given-names>W. A.</given-names>
</name>
<name>
<surname>Chappell</surname>
<given-names>R. J.</given-names>
</name>
<name>
<surname>Chinnaiyan</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Mehta</surname>
<given-names>M. P.</given-names>
</name>
<name>
<surname>Harari</surname>
<given-names>P. M.</given-names>
</name>
</person-group> (<year>2005</year>). <article-title>The Impact of Daily Setup Variations on Head-And-Neck Intensity-Modulated Radiation Therapy</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>61</volume>, <fpage>779</fpage>&#x2013;<lpage>788</lpage>. <pub-id pub-id-type="doi">10.1016/j.ijrobp.2004.07.696</pub-id> </citation>
</ref>
<ref id="B10">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Karger</surname>
<given-names>C. P.</given-names>
</name>
<name>
<surname>J&#xe4;kel</surname>
<given-names>O.</given-names>
</name>
<name>
<surname>Debus</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Kuhn</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Hartmann</surname>
<given-names>G. H.</given-names>
</name>
</person-group> (<year>2001</year>). <article-title>Three-dimensional Accuracy and Interfractional Reproducibility of Patient Fixation and Positioning Using a Stereotactic Head Mask System</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>49</volume>, <fpage>1493</fpage>&#x2013;<lpage>1504</lpage>. <pub-id pub-id-type="doi">10.1016/s0360-3016(00)01562-5</pub-id> </citation>
</ref>
<ref id="B11">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kim</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Akpati</surname>
<given-names>H. C.</given-names>
</name>
<name>
<surname>Kielbasa</surname>
<given-names>J. E.</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>J. G.</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Amdur</surname>
<given-names>R. J.</given-names>
</name>
<etal/>
</person-group> (<year>2004</year>). <article-title>Evaluation of Intrafraction Patient Movement for CNS and Head &#x26; Neck IMRT</article-title>. <source>Med. Phys.</source> <volume>31</volume>, <fpage>500</fpage>&#x2013;<lpage>506</lpage>. <pub-id pub-id-type="doi">10.1118/1.1644641</pub-id> </citation>
</ref>
<ref id="B12">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Li</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Yin</surname>
<given-names>G.</given-names>
</name>
<name>
<surname>Zhuang</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Geng</surname>
<given-names>J.</given-names>
</name>
</person-group> (<year>2006</year>). <article-title>Real-Time 3D-Surface-Guided Head Refixation Useful for Fractionated Stereotactic Radiotherapy</article-title>. <source>Med. Phys.</source> <volume>33</volume>, <fpage>492</fpage>&#x2013;<lpage>503</lpage>. <pub-id pub-id-type="doi">10.1118/1.2150778</pub-id> </citation>
</ref>
<ref id="B13">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Linthout</surname>
<given-names>N.</given-names>
</name>
<name>
<surname>Verellen</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Tournel</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Storme</surname>
<given-names>G.</given-names>
</name>
</person-group> (<year>2006</year>). <article-title>Six Dimensional Analysis with Daily Stereoscopic x-Ray Imaging of Intrafraction Patient Motion in Head and Neck Treatments Using Five Points Fixation Masks</article-title>. <source>Med. Phys.</source> <volume>33</volume>, <fpage>504</fpage>&#x2013;<lpage>513</lpage>. <pub-id pub-id-type="doi">10.1118/1.2165417</pub-id> </citation>
</ref>
<ref id="B14">
<citation citation-type="web">
<collab>Orfit Industries</collab> (<year>2022</year>). <article-title>Instructions for Use</article-title>. <comment>Available at: <ext-link ext-link-type="uri" xlink:href="https://www.orfit.com/app/uploads/50025_PDF.pdf">https://www.orfit.com/app/uploads/50025_PDF.pdf</ext-link> (Accessed on April 21, 2022)</comment>. </citation>
</ref>
<ref id="B15">
<citation citation-type="web">
<collab>Pressure Prole Syst., Inc</collab> (<year>2022b</year>). <article-title>SingleTact Mounting Methods</article-title>. <comment>Available online: <ext-link ext-link-type="uri" xlink:href="https://www.singletact.com/wp-content/uploads/ST_AN_003_MountingMethods.pdf">https://www.singletact.com/wp-content/uploads/ST_AN_003_MountingMethods.pdf</ext-link> (Accessed on April 21, 2022)</comment>. </citation>
</ref>
<ref id="B16">
<citation citation-type="web">
<collab>Pressure Prole Syst., Inc</collab> (<year>2022a</year>). <article-title>SingleTact User Manual</article-title>. <comment>Available online: <ext-link ext-link-type="uri" xlink:href="https://www.singletact.com/SingleTact_Manual.pdf">https://www.singletact.com/SingleTact_Manual.pdf</ext-link> (Accessed on April 21, 2022)</comment>. </citation>
</ref>
<ref id="B17">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Purdy</surname>
<given-names>J. A.</given-names>
</name>
</person-group> (<year>2000</year>). <article-title>Future Directions in 3-d Treatment Planning and Delivery: A Physicist&#x2019;s Perspective</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>46</volume>, <fpage>3</fpage>&#x2013;<lpage>6</lpage>. <pub-id pub-id-type="doi">10.1016/s0360-3016(99)00509-x</pub-id> </citation>
</ref>
<ref id="B18">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sweeney</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Bale</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Vogele</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Nevinny-Stickel</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Bluhm</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Auer</surname>
<given-names>T.</given-names>
</name>
<etal/>
</person-group> (<year>1998</year>). <article-title>Repositioning Accuracy: Comparison of a Noninvasive Head Holder with Thermoplastic Mask for Fractionated Radiotherapy and a Case Report</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>41</volume>, <fpage>475</fpage>&#x2013;<lpage>483</lpage>. <pub-id pub-id-type="doi">10.1016/s0360-3016(98)00064-9</pub-id> </citation>
</ref>
<ref id="B19">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tae-Ho</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Jai-Woong</surname>
<given-names>Y.</given-names>
</name>
<name>
<surname>Seong-Hee</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Tae-Suk</surname>
<given-names>S.</given-names>
</name>
</person-group> (<year>2012</year>). <article-title>A Theoretical Model for the Analysis of Residual Motion Artifacts in 4D CT Scans</article-title>. <source>Prog. Med. Phys.</source> <volume>23</volume>, <fpage>145</fpage>&#x2013;<lpage>153</lpage>. <pub-id pub-id-type="doi">10.14316/pmp.2012.23.3.145</pub-id> </citation>
</ref>
<ref id="B20">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tang</surname>
<given-names>K. P. M.</given-names>
</name>
<name>
<surname>Yick</surname>
<given-names>K. L.</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>P. L.</given-names>
</name>
<name>
<surname>Yip</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Or</surname>
<given-names>K. H.</given-names>
</name>
<name>
<surname>Chau</surname>
<given-names>K. H.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Effect of Contacting Surface on the Performance of Thin-Film Force and Pressure Sensors</article-title>. <source>Sensors (Basel)</source> <volume>20</volume>, <fpage>6863</fpage>. <pub-id pub-id-type="doi">10.3390/s20236863</pub-id> </citation>
</ref>
<ref id="B21">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tryggestad</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Christian</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Ford</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Kut</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Le</surname>
<given-names>Y.</given-names>
</name>
<name>
<surname>Sanguineti</surname>
<given-names>G.</given-names>
</name>
<etal/>
</person-group> (<year>2011</year>). <article-title>Inter- and Intrafraction Patient Positioning Uncertainties for Intracranial Radiotherapy: A Study of Four Frameless, Thermoplastic Mask-Based Immobilization Strategies Using Daily Cone-Beam CT</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>80</volume>, <fpage>281</fpage>&#x2013;<lpage>290</lpage>. <pub-id pub-id-type="doi">10.1016/j.ijrobp.2010.06.022</pub-id> </citation>
</ref>
<ref id="B22">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tsai</surname>
<given-names>J.-S.</given-names>
</name>
<name>
<surname>Engler</surname>
<given-names>M. J.</given-names>
</name>
<name>
<surname>Ling</surname>
<given-names>M. N.</given-names>
</name>
<name>
<surname>Wu</surname>
<given-names>J. K.</given-names>
</name>
<name>
<surname>Kramer</surname>
<given-names>B.</given-names>
</name>
<name>
<surname>Dipetrillo</surname>
<given-names>T.</given-names>
</name>
<etal/>
</person-group> (<year>1999</year>). <article-title>A Non-invasive Immobilization System and Related Quality Assurance for Dynamic Intensity Modulated Radiation Therapy of Intracranial and Head and Neck Disease</article-title>. <source>Int. J. Radiat. Oncol. Biol. Phys.</source> <volume>43</volume>, <fpage>455</fpage>&#x2013;<lpage>467</lpage>. <pub-id pub-id-type="doi">10.1016/s0360-3016(98)00398-8</pub-id> </citation>
</ref>
<ref id="B23">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Wagner</surname>
<given-names>T. H.</given-names>
</name>
<name>
<surname>Meeks</surname>
<given-names>S. L.</given-names>
</name>
<name>
<surname>Bova</surname>
<given-names>F. J.</given-names>
</name>
<name>
<surname>Friedman</surname>
<given-names>W. A.</given-names>
</name>
<name>
<surname>Willoughby</surname>
<given-names>T. R.</given-names>
</name>
<name>
<surname>Kupelian</surname>
<given-names>P. A.</given-names>
</name>
<etal/>
</person-group> (<year>2007</year>). <article-title>Optical Tracking Technology in Stereotactic Radiation Therapy</article-title>. <source>Med. Dosim.</source> <volume>32</volume>, <fpage>111</fpage>&#x2013;<lpage>120</lpage>. <pub-id pub-id-type="doi">10.1016/j.meddos.2007.01.008</pub-id> </citation>
</ref>
<ref id="B24">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Warsito</surname>
<given-names>I. F.</given-names>
</name>
<name>
<surname>Hunold</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Haueisen</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Supriyanto</surname>
<given-names>E.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Performance Evaluation of Capacitive Based Force Sensor for Electroencephalography Head Caps</article-title>. <source>Int. J. Robot. Autom. Sci.</source> <volume>2</volume>, <fpage>4</fpage>&#x2013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.33093/ijoras.2020.2.1</pub-id> </citation>
</ref>
<ref id="B25">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Willner</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Flentje</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Bratengeier</surname>
<given-names>K.</given-names>
</name>
</person-group> (<year>1997</year>). <article-title>CT Simulation in Stereotactic Brain Radiotherapy &#x2014; Analysis of Isocenter Reproducibility with Mask Fixation</article-title>. <source>Radiother. Oncol.</source> <volume>45</volume>, <fpage>83</fpage>&#x2013;<lpage>88</lpage>. <pub-id pub-id-type="doi">10.1016/s0167-8140(97)00135-7</pub-id> </citation>
</ref>
</ref-list>
</back>
</article>