ORIGINAL RESEARCH article

Front. Sens., 26 October 2026

Sec. Biosensors

Volume 7 - 2026 | https://doi.org/10.3389/fsens.2026.1864839

Toward a modular label-free electrochemical biosensing platform based on a one-pot avidin functionalization of reduced graphene oxide

  • Instituto de Investigaciones en Físico-Química de Córdoba (INFIQC-CONICET), Departamento de Fisicoquímica, Facultad de Ciencias Químicas, Universidad Nacional de Córdoba, Ciudad Universitaria, Córdoba, Argentina

Abstract

A one-pot, ultrasound-assisted and environmentally responsible strategy is proposed for the simultaneous partial reduction and biofunctionalization of graphene oxide (GO), using L-tyrosine (L-tyr) and ascorbic acid (AA) as mild and complementary reducing agents. This synergistic approach enables controlled partial restoration of the sp2 carbon framework while preserving oxygen functionalities required for biomolecular anchoring. Avidin is incorporated in situ during the reduction process, eliminating post-synthetic modification steps while enabling its functional role in biotinylated recognition element immobilization. The resulting avidin-functionalized reduced graphene oxide (rGO) provides a functional electrochemical interface for the oriented immobilization of biotinylated recognition elements through the avidin-biotin affinity interaction. This immobilization strategy was conceived to facilitate the assembly of different biosensing interfaces without requiring surface re-engineering. The platform was implemented as a label-free electrochemical interface and characterized by electrochemical impedance spectroscopy to monitor interfacial assembly and biorecognition events. As a proof-of-concept, the proposed immobilization strategy was experimentally validated using the thrombin-binding aptamer as a representative biotinylated recognition element for the detection of human α-thrombin (α-th). The platform exhibited a linear response from 1.34 × 10−9 M to 2.67 × 10−6 M (R2 = 0.987) and a detection limit of 4.4 × 10−10 M, demonstrating selective response in the presence of potentially interfering proteins and enriched human serum samples. These results demonstrate a sustainable one-pot strategy for producing avidin-functionalized rGO and establish a proof-of-concept for an avidin-based electrochemical biointerface. Although the present study experimentally validates the approach using a single aptamer-target pair, the proposed immobilization strategy may enable the future development of adaptable electrochemical biosensing platforms employing different biotinylated recognition elements without modifying the underlying transducer.

Graphical Abstract

1 Introduction

Electrochemical biosensors based on graphene-derived materials have attracted significant attention due to their high sensitivity, tunable surface chemistry, and suitability for miniaturized and point-of-care applications (Korucu, 2022; Palomba et al., 2022). Among these materials, reduced graphene oxide (rGO) has emerged as an attractive electrochemical transducer, as the partial restoration of its sp2 carbon network enhances charge transfer rates, while residual oxygen-containing functional groups provide anchoring sites for the immobilization of biorecognition elements (Wang et al., 2014; Chu et al., 2023).

Despite these advantages, rGO-based biointerfaces are commonly fabricated through chemical or thermal reduction processes followed by post-synthetic biofunctionalization. These approaches often involve harsh conditions, multistep protocols, and extended processing times, which may compromise interfacial control, limit reproducibility, and adversely affect the bioactivity of immobilized biomolecules (Cai et al., 2022; Park et al., 2022). As a result, material reduction and biofunctionalization are typically treated as separate stages, increasing fabrication complexity and restricting the robustness of the resulting bioelectrochemical interfaces.

In this context, mild and sustainable reduction strategies based on biomolecules have emerged as attractive alternatives (Utkan et al., 2023). Amino acids and ascorbic acid have been explored as eco−friendly reducing agents capable of partially restoring conjugated carbon domains while preserving oxygenated functionalities, particularly at sheet edges. Under ultrasound-assisted treatment, these approaches enable controlled reduction while maintaining surface chemistries compatible with biomolecular immobilization (Tewatia et al., 2021; Korucu, 2022; Lenin et al., 2025). However, the integration of material reduction and biofunctionalization into a single processing step remains largely unexplored. Moreover, most graphene-based biosensing platforms are designed for single-analyte detection, lacking the flexibility required for rapid adaptation to different analytical targets.

Aptamer−based biosensors offer a versatile alternative to antibody-based systems, enabling selective and label-free detection of a wide range of analytes. Their synthetic accessibility, chemical stability, and tunable structure make them particularly compatible with graphene-based platforms. Nevertheless, most reported graphene−based aptasensors rely on multistep fabrication strategies in which material preparation and biofunctionalization are performed independently, resulting in labor-intensive, target-specific sensing interfaces that are difficult to reconfigure for each new analyte target (Yang and Zhang, 2025).

In this regard, the avidin-biotin system provides a robust and widely applicable bioaffinity framework that can serve as a molecular adaptor, enabling the immobilization of a broad range of biotinylated biomolecules while preserving their functionality. Leveraging this interaction at the material synthesis stage offers a versatile strategy for developing modular sensing platforms, where the analytical target can be readily altered through substitution of the biotinylated recognition element without modifying the underlying transducer.

Herein, we report a one-pot and environmentally responsible strategy for the simultaneous reduction and in situ biofunctionalization of GO using L-tyrosine (L-tyr) and ascorbic acid (AA) as complementary and mild reducing agents under ultrasound-assisted conditions. This approach enables the incorporation of avidin (Av) during the reduction process, yielding a bioactive rGO platform for the immobilization of biotinylated recognition elements through the avidin-biotin interaction. By embedding the bioaffinity layer directly into the material, the proposed strategy is intended to decouple surface chemistry from target recognition, potentially facilitating the assembly of biosensors by simply exchanging the biotinylated recognition element.

As a proof−of−concept, the proposed immobilization strategy was experimentally validated through the label-free electrochemical detection of human α−thrombin (α−th) using the thrombin-binding aptamer as a representative biotinylated recognition element. Therefore, this application was intended to demonstrate the feasibility of the proposed biofunctionalization strategy rather than its full analytical versatility. Human α−thrombin, a clinically relevant protein biomarker (Zhang et al., 2021; Hardy et al., 2024; Gil M. R., 2025), was selected to evaluate the analytical performance of the proposed biointerface. The interfacial assembly, analytical response, and robustness of the platform were systematically evaluated, and the environmental impact of the fabrication process was assessed using the AGREE metric (Pena-Pereira et al., 2022; Wojnowski et al., 2022).

2 Materials and methods

2.1 Reagents

Graphene oxide aqueous dispersion (0.4 wt%) was purchased from Graphenea. L-tyrosine (L-tyr), L-cysteine (L-cys), L-tryptophan (L-trp), L-histidine (L-hist), bovine serum albumin (BSA), hydroquinone (HQ), benzoquinone (BQ), lysozyme (Lys) from chicken egg white, cytochrome C (Cyt C) and myoglobin (Myo) from equine heart, uric acid, glucose, dopamine, potassium chloride (KCl), sodium chloride (NaCl), magnesium chloride (MgCl2) and sodium carbonate (Na2CO3) were obtained from Sigma–Aldrich. Human α-thrombin (α-th; Lot N°: Tbd 2,208, 541 IU mg-1), a commercially available human blood-derived therapeutic reagent intended exclusively for in vitro use, was supplied by the Public Hemoderivatives Laboratory of the National University of Córdoba (Laboratorio de Hemoderivados, UNC). This material does not involve the use of clinical human samples. Ascorbic acid (AA) and ethyl alcohol absolute (anhydrous, 99.88%) were purchased from J.T. Baker. Polysorbate 20 (Tween 20) was purchased from Parafarm. TRIS buffer was obtained from Biopack. Standatrol S-E (control Level 1 and Level 2, catalog N° N1-263940) a commercially available homogenized and lyophilized human serum for in vitro analytical use, was obtained from Wiener Lab. Avidin (Av) from egg white (catalog N° A-2667) and the biotinylated DNA thrombin-binding aptamer sequence (b-TBA: 5′-biotin-TTTTTTGGTTGGTGTGGTTGG-3′), were supplied by Invitrogen Life Technologies. The thrombin-binding aptamer (TBA), a well-established 15-mer DNA sequence originally reported by Bock et al., 1992, was employed as the biorecognition element. A biotinylated TBA (b-TBA) bearing a poly-T spacer was used to improve aptamer accessibility and surface-oriented recognition, as previously reported for label-free surface plasmon resonance-based anti-α-th aptasensing platforms (Jalit et al., 2013). This modification does not alter the specific recognition motif of the aptamer but facilitates its functional performance at the bioelectrochemical interface.

A 0.020 M phosphate buffer solution (PBS, pH 7.40) was used as supporting electrolyte for the electrochemical experiments. The following buffer solutions were used for specific purposes during the aptasensor construction, facilitating molecular interactions, removal of unbound species, and the detection process:

Dispersion Buffer (DB): 0.010 M TRIS (2-amino-2-hydroxymethyl-1,3-propanediol) buffer solution, pH 8.60.

Interaction Buffer (IB): 0.010 M TRIS buffer solution, pH 8.60, including 0.010 M NaCl, 5.0 × 10−3 M KCl, 1.0 × 10−3 M MgCl2.

Washing Buffer (WB): 0.010 M TRIS buffer solution, pH 8.60, including 0.010 M NaCl, 5.0 × 10−3 M KCl, 1.0 × 10−3 M MgCl2, 0.010% Tween-20.

All the solutions were prepared using ultrapure water (ρ = 18.0 MΩ cm) obtained from a Millipore-MilliQ system. All experiments were performed at room temperature.

2.2 Apparatus

Electrochemical impedance spectroscopy (EIS) was carried out using an Autolab PGSTAT 30 potentiostat (Metrohm) equipped with an ADC750 and a frequency response analyzer (FRA) hardware module and software. The EIS parameters were the following: amplitude: 0.010 V, frequency range: from 1.0 × 10−2 to 1.0 × 106 Hz, and working potential: 0.050 V. All EIS spectra were analyzed and fitted using the Z-view software. The impedance spectra were fitted using a Randles-type equivalent circuit consisting of the solution resistance (Rs), charge-transfer resistance (Rct), a constant phase element (CPE1) accounting for the non-ideal interfacial capacitive behavior, and a second constant phase element (CPE2) representing the diffusional (Warburg-type) impedance contribution. This model was used to describe the electrochemical response of the modified electrodes and to extract Rct values employed throughout the study.

A conventional three-electrode system was employed for all the electrochemical measurements, consisting of a glassy carbon electrode (GCE, Model 104, CH Instruments, 3.0 mm diameter) as the working electrode, a platinum wire as counter electrode, and an Ag/AgCl (3.0 M NaCl) as reference electrode. The redox probe solution 0.020 M HQ/BQ (1:1 M ratio) was prepared in a 0.020 M PBS (pH 7.40) as supporting electrolyte. Unless otherwise specified, all electrochemical measurements were performed in five independent replicates (n = 5), and the reported errors correspond to the standard deviation of the independent replicates.

Ultrasound-assisted reduction and biofunctionalization were carried out using a TB04TA Testlab ultrasonic cleaner operating at 40 kHz and 160 W ultrasonic power. UV-Vis spectra were recorded with a Shimadzu UV-1700 Pharma spectrometer. Fourier-transform infrared (FTIR) spectra were obtained using a Nicolet 5-SXC spectrometer. Samples for FTIR analysis were prepared by drop-casting the corresponding solution or dispersion onto a ZnSe disk, followed by solvent evaporation.

SEM images were acquired using a Crossbeam 550 field-emission scanning electron microscope (Carl Zeiss, Germany) at the Laboratorio de Análisis de Materiales por Espectrometría de Rayos X (LAMARX), Facultad de Matemática, Astronomía, Física y Computación (FAMAF), Universidad Nacional de Córdoba (UNC), Córdoba, Argentina.

2.3 Preparation of the working electrodes

2.3.1 Preparation of functionalized reduced graphene oxide-avidin (f-rGO) dispersion

The functionalized reduced graphene oxide-avidin (f-rGO) dispersion was prepared by mixing graphene oxide (GO, 1.00 mg mL-1) with L-tyrosine (L-tyr, 0.50 mg mL-1) and ascorbic acid (AA, 0.25 mg mL-1) as reducing agents. This dispersion was prepared in DB: ethanol (1:2, v/v). Avidin (Av, 1.00 mg mL-1) was incorporated directly into the reaction mixture during the reduction process. The resulting dispersion was subjected to ultrasound-assisted reduction for 30 min.

The selected experimental conditions were chosen based on preliminary optimization studies aimed at achieving efficient reduction, stable dispersion, and effective biofunctionalization under mild conditions. In particular, the ultrasonication time was limited to 30 min and mild reducing agents (L-tyrosine and ascorbic acid) were employed to minimize potential protein denaturation and preserve biomolecular functionality (Gutierrez et al., 2019).

2.3.2 Construction of the GCE/(f-rGO/BSA/b-TBA) platform

Prior to surface modification, the glassy carbon electrode (GCE) was mechanically polished using 0.05 µm alumina slurry (Micropolish, Buehler) for 1 min. The GCE surface was thoroughly rinsed with deionized water, sonicated in ultrapure water for 10 s to remove residual alumina particles, and dried under a nitrogen stream. Subsequently, a 10 µL aliquot of the f-rGO dispersion was drop-cast onto the GCE surface and dried at 50 °C for 15 min.

All subsequent modification steps were carried out in a humidity-controlled chamber to ensure reproducible surface assembly.

2.3.2.1 Pretreatment of the GCE/f-rGO surface to prevent non-specific binding

To minimize non-specific binding on the platform, the GCE/f-rGO surface was incubated with a 10 μL droplet of bovine serum albumin (BSA, 1.00 mg mL-1), prepared in interaction buffer (IB), for 30 min. After the blocking step, the electrodes were sequentially rinsed with WB and IB for 20 s each.

2.3.2.2 Immobilization of b-TBA on the GCE/f-rGO/BSA surface

Prior to immobilization, the biotinylated thrombin-binding aptamer (b-TBA) was pretreated by heating at 90 °C for 5 min and subsequently cooled to room temperature for 10 min to promote proper folding. The aptamer was diluted in IB, and a 10 μL aliquot was incubated on the GCE/f-rGO/BSA surface for 60 min. After incubation, the electrodes were thoroughly rinsed with WB and IB (20 s each) to remove unbound aptamer.

2.3.2.3 Detection step of human α-thrombin (α-th)

For the detection step, the prepared aptasensing platform was incubated with 10 μL aliquots of α-th solutions at concentrations ranging from 1.34 × 10−9 M to 2.67 × 10−6 M for 30 min, prior to electrochemical measurements.

3 Results and discussion

3.1 Spectroscopic characterization

Spectroscopic techniques were used to elucidate the chemical and structural transformations of graphene oxide induced by ultrasound in the presence of L-tyr and AA during the one-pot reduction and in situ biofunctionalization process. UV-Vis and FTIR spectroscopy (Figures 1A,B) were employed as complementary techniques to evaluate the extent of reduction and the partial recovery of sp2 carbon framework, providing direct insight into the synergistic action of both reducing agents on the GO structure.

FIGURE 1

3.1.1 UV-vis analysis

Figure 1A shows the UV-Vis absorption spectra of commercial GO dispersion before (a) and after partial reduction induced by ultrasound in the presence of L-tyr (b) and AA (c). The GO sample (a) exhibits a broad absorption band centered at approximately 230 nm, together with a weak shoulder around 300 nm. These spectral characteristics are commonly attributed to the π-π* transition of conjugated C=C bonds within sp2-hybridized carbon domains and to n-π* transitions associated with carbonyl-related oxygenated functionalities, respectively (Wang et al., 2014). After the incorporation of L-tyr (rGO-L-tyr) (b), noticeable changes in the absorption spectrum are observed. In particular, an enhanced absorption contribution in the 210–230 nm region becomes evident, which can be associated with π-π* electronic transitions arising from the aromatic ring of the tyrosine molecule, superimposed on the intrinsic absorption band of GO (Wang et al., 2014). The exact position and intensity of this contribution may depend on experimental parameters such as pH, solvent polarity, and molecular concentration (Salmahaminati and Roca-Sanjuan, 2024). Concomitantly, the main π-π* absorption band of GO undergoes a marked bathochromic shift accompanied by spectral broadening toward longer wavelengths (approximately 250–300 nm), which is consistent with a partial reduction of GO involving the removal of oxygen-containing functionalities and a progressive restoration of π-electron conjugation within the sp2-carbon framework (Wang et al., 2014). Upon incorporation of AA (rGO-L-tyr-AA) (c), a further increase in absorption intensity and an additional extension of the π-π* band toward longer wavelengths are observed, together with a relative attenuation of n–π* contributions associated with carbonyl-related oxygen groups. This spectral behavior indicates a more effective, still partial, restoration of electronic conjugation within the sp2-carbon network, promoted by the synergistic reducing action of L-tyr and AA under ultrasound-assisted conditions (Le et al., 2019; Bakry et al., 2024; Zafar et al., 2025).

Overall, the UV-Vis spectra provide evidence supporting a controlled and progressive partial reduction of GO, in which electronic conjugation is enhanced while residual oxygenated functionalities are preserved. This balance is particularly advantageous for electrochemical biosensing applications, as it improves interfacial electron transfer without compromising the surface chemistry required for subsequent biofunctionalization.

3.1.2 FTIR analysis

Figure 1B shows the FTIR spectra obtained from the same samples analyzed by UV-Vis spectroscopy. Commercial GO (a) exhibits characteristic bands associated with hydroxyl, carbonyl, and epoxy groups, consistent with its highly oxidized structure. A broad band at approximately 3,400 cm-1 is assigned to O-H stretching vibrations, while the band at 1725 cm-1 corresponds to the C=O stretching of carboxyl and carbonyl functionalities. The band at ∼1,580 cm-1 is attributed to the C=C stretching vibrations of the sp2-hybridized carbon domains. Additional features in the 1,050–1,240 cm-1 region arise from C-O-C, C-O, and C-OH stretching vibrations, indicative of oxygenated functional groups typically associated with GO sheet-edges (Xue et al., 2019; Abdelhalim et al., 2020; Das et al., 2023).

Following ultrasound-assisted treatment in the presence of L-tyr, the FTIR spectrum of rGO-L-tyr (b) displays noticeable changes. The bands in the 3,200–3,400 cm-1 region reflect overlapped contributions from N-H stretching vibrations of amino groups and residual O-H groups, suggesting successful incorporation of L-tyr onto the GO surface. The appearance of bands around 2,900 cm-1, assigned to stretching vibrations of methylene groups (-CH2-), is consistent with L-tyr-derived surface functionalization promoted by the ultrasound-assisted processing. In addition, changes in the relative intensity of the C=O (at 1725 cm-1) and C=C (at 1,580–1,620 cm-1) bands are consistent with partial removal of oxygenated functionalities and concomitant restoration of sp2-hybridized carbon domains (Abdelhalim et al., 2020; Tarighat et al., 2025). A more pronounced reduction effect is observed upon incorporation of AA (rGO-L-tyr-AA). In this case, the FTIR spectrum (c) shows a marked decrease in the intensity of the carbonyl band at 1725 cm-1, suggesting enhanced removal of oxygenated edge groups, including lactone and carboxyl functionalities. Concurrent changes in the 1,050–1,240 cm-1 region are attributed to modifications of residual C-O, C-N, and C-OH functionalities. Moreover, the increased definition and relative intensity of the C=C stretching bands in the 1,560–1,640 cm-1 region suggest a more effective restoration of the sp2 carbon framework (Le et al., 2019; Xue et al., 2019; Das et al., 2023).

Altogether, the combined UV-Vis and FTIR analyses demonstrate that the synergistic action of L-tyr and AA under ultrasound-assisted conditions promotes both partial reduction and functionalization of GO. While L-tyr contributes primarily to basal-plane modification through polymerization-assisted reduction, AA contributes to the removal of oxygenated groups, particularly those located at sheet edges, facilitating mild conditions and controlled electronic restoration (Korucu, 2022). Importantly, this process preserves surface functionalities, which is particularly advantageous for electrochemical biosensing applications.

3.2 Morphological characterization: scanning electron microscopy (SEM)

SEM was employed to examine the morphological evolution of GO throughout the one-pot reduction and biofunctionalization process as well as to correlate structural features with spectroscopic findings.

Figure 2 presents representative SEM micrographs of pristine GO (Figure 2A) and the materials obtained after partial reduction and biofunctionalization (Figures 2B,C). The pristine GO sample (Figure 2A) exhibits the characteristic morphology of exfoliated graphene oxide sheets, consisting of thin, paper-like lamellae with pronounced folds, corrugations, and overlapping regions. This highly wrinkled and flexible morphology is associated with the presence of abundant oxygen-containing functional groups and weak interlayer interactions, which promote sheet deformation and restacking behavior (Aliyev et al., 2019). After ultrasound-assisted treatment in the presence of L-tyr and AA (Figure 2B), the material undergoes a clear morphological reorganization. The partially reduced GO displays a more compact and interconnected lamellar architecture, characterized by thicker and less translucent sheets, smoother basal-plane regions interspersed with residual wrinkles, and more defined edge features. This evolution reflects the partial restoration of sp2 carbon domains and enhanced intersheet interactions, arising from the controlled removal of oxygen-containing groups while preserving the laminar structure of the material (Stobinski et al., 2014). Upon incorporation of Av during the one-pot process (Figure 2C), the resulting biofunctional material maintains its overall lamellar architecture while displaying increased surface heterogeneity and a more textured morphology. These features are consistent with biomolecule incorporation within the nanostructured matrix, leading to localized variations in surface roughness and interlayer interactions without compromising the structural integrity of the rGO framework (Saputra et al., 2025). In this sense, the SEM analysis provides direct morphological evidence that supports the spectroscopic findings obtained by UV-Vis and FTIR studies. The transition from highly corrugated and flexible GO sheets to a more compact, interconnected lamellar architecture after ultrasound-assisted treatment in the presence of L-tyr and AA is consistent with partial reduction and controlled structural reorganization, while preserving its intrinsic sheet-like morphology. Importantly, the retention of laminar domains together with the emergence of surface heterogeneity upon avidin incorporation is consistent with the successful integration of the biomolecule into the graphene-based framework, without inducing structural collapse or excessive aggregation. Taken together, these morphological features are consistent with a mild and controlled reduction strategy that preserves accessible basal planes and edge regions, yielding a structurally stable and chemically versatile interface suitable for subsequent sensing applications.

FIGURE 2

3.3 Electrochemical characterization

Electrochemical impedance spectroscopy (EIS) was employed to investigate the interfacial properties of the modified electrodes during the different stages of aptasensing platform construction, as schematically illustrated in Scheme 1. EIS is particularly suitable in the equivalent circuit shown in Scheme 1, CPE1 accounts for the non-ideal capacitive behavior of the electrode/electrolyte interface, whereas CPE2 represents the diffusional impedance contribution referred to here as the Warburg element (Gutierrez et al., 2019). Variations in Rct, obtained from fitting the impedance spectra using the equivalent circuit described in Section 2.2, were used as the analytical parameter to assess (i) the optimization of GO reduction under ultrasound-assisted conditions, (ii) the efficiency of the one-pot biofunctionalization strategy, and (iii) the subsequent biomolecular recognition events leading to α-th detection.

SCHEME 1

The following sections describe the EIS response associated with each step of the platform construction and biorecognition process.

3.3.1 Optimization of GO reduction under ultrasound-assisted conditions

Neither complete reduction nor extensive exfoliation of GO is required or desirable for electrochemical sensing applications, as both processes may compromise surface functionality. Instead, an optimal balance between electron transfer efficiency and the availability of surface functional groups is critical to achieve high electrochemical performance while enabling effective biofunctionalization. In this work, ultrasound-assisted conditions were employed to promote efficient interaction between GO sheets and mild reducing agents, enabling controlled structural modification under environmentally friendly conditions.

The influence of L-tyr concentration on the GO reduction process was first evaluated over the range from 0.00 to 4.00 mg mL-1. For this purpose, L-tyr solutions were mixed with 1.00 mg mL-1 of GO dispersed in a DB/ethanol solution (1:2 v/v), and the ultrasonication time was fixed at 30 min. In parallel, the electrochemical response of GO dispersion subjected to ultrasound-assisted treatment in the absence of L-tyr was also evaluated for comparison (Supplementary Figure S1A). The EIS results showed no significant variation in Rct values with increasing L-tyr concentration. This behavior reflects a balance between partial restoration of the sp2 carbon network, which favors electron transfer, and the presence of oxidized L-tyr-derived species interacting with oxygenated groups on GO, which may hinder charge transfer as previously reported (Wang et al., 2014). Based on these results, 0.50 mg mL-1 L-tyr was selected as the optimal concentration, since no significant improvement was observed at higher concentrations and the lowest concentration providing a stable electrochemical response was selected.

The effect of ultrasonication time on the GO reduction process was subsequently evaluated using dispersions containing 1.00 mg mL-1 GO and 0.50 mg mL-1 L-tyr (Supplementary Figure S1B). Short ultrasonication times resulted in poorly stable dispersions, indicating insufficient interaction between GO sheets and the reducing agent. In contrast, prolonged treatment (60 min) led to a significant increase in Rct, likely associated with excessive exfoliation and partial sheet fragmentation (Gutierrez et al., 2019; Ye and Feng, 2016; Wang et al., 2020a; Wang et al., 2020b), which reduces effective electron transfer pathways. A compromise between electrochemical response, dispersion stability, and reproducibility was achieved at 30 min, which was therefore selected as the optimal condition for subsequent experiments.

To further enhance the reduction process under mild conditions, ascorbic acid (AA) was incorporated together with L-tyr (Supplementary Figure S1C). The addition of AA produced a slight but reproducible decrease in Rct, indicating additional partial restoration of the sp2 carbon framework. No significant differences were observed between AA concentrations of 0.25 and 0.50 mg mL-1, suggesting that the reduction process reaches a plateau within this concentration range. Accordingly, 0.25 mg mL-1 AA was selected as the optimal concentration, minimizing reagent consumption while maintaining the electrochemical performance of the dispersion.

Finally, the effect of ultrasonication time was evaluated in the presence of both reducing agents, L-tyr (0.50 mg mL-1) and AA (0.25 mg mL-1) (Supplementary Figure S1D). Under these conditions, the evolution of Rct as a function of ultrasonication time followed a trend similar to that observed in the absence of AA (Supplementary Figure S1B). Therefore, 30 min remained the condition that provides the best compromise between partial electronic restoration, dispersion stability, and reproducibility (Tewatia et al., 2021; Korucu, 2022). These results indicate that AA enhances the reduction efficiency while preserving the ultrasound-assisted structural balance established for the GO/L-tyr system.

3.3.2 Optimization of one-pot biofunctionalization and biointerface assembly on f-rGO

Once the optimal conditions for GO reduction were established, avidin (Av) was incorporated during the ultrasound-assisted treatment to generate a biofunctional platform through a single-step reduction and functionalization process. This strategy enables the direct formation of a bioactive interface while avoiding post-synthesis immobilization procedures, thereby simplifying fabrication and minimizing interfacial perturbations.

Since the successive modification steps involve distinct physicochemical and biointerfacial processes, the optimization criteria were not based exclusively on maximizing or minimizing Rct values. Instead, changes in Rct were interpreted according to the expected interfacial behavior associated with each stage. Therefore, the selected conditions were established considering the overall electrochemical response, biofunctionalization efficiency, interface stability, and practical aspects related to platform fabrication.

It should be noted that harsh pretreatment processes, including ultrasound-assisted processing, may compromise the biological activity of biomolecules. Therefore, assessing the preservation of Av functionality after the one-pot ultrasound-assisted process is a critical requirement to ensure effective immobilization of the biotinylated thrombin-binding aptamer (b-TBA) and, consequently, reliable target biorecognition (Kim and Yang, 2024).

Supplementary Figure S2 illustrates the different stages of platform modification (bars a–d). The biofunctionalization experiments shown in Supplementary Figure S2 were performed using a preliminary avidin concentration of 0.10 mg mL-1 as part of the platform optimization. The transition from GCE/rGO (Supplementary Figure S2a) to GCE/f-rGO (Supplementary Figure S2b) resulted in a clear increase in Rct, attributable to the incorporation of Av and the insulating character of the protein layer at the electrode surface, thus confirming successful biofunctionalization. Subsequent BSA blocking (Supplementary Figure S2c) produced an additional increase in Rct, consistent with the formation of a passivating layer that reduces non-specific adsorption. Finally, immobilization of the biotinylated aptamer (Supplementary Figure S2d) generated a further increase in Rct, supporting successful formation of the avidin-biotin recognition interface. The preservation of avidin functionality following the ultrasound-assisted one-pot process was indirectly supported by the efficient immobilization of b-TBA and the subsequent selective electrochemical response of the resulting biosensing interface. Notably, b-TBA coupling produced an additional increase in Rct, supporting the retention of avidin-biotin binding capability after the ultrasound-assisted treatment (Gutierrez et al., 2019; Wang et al., 2020a; Wang et al., 2020b).

The effect of surface blocking with bovine serum albumin (BSA) used to suppress non-specific adsorption was subsequently evaluated (Supplementary Figure S3). Blocking times shorter than 30 min (Supplementary Figure S3A) were insufficient to increase the value of Rct, reflecting the inability to effectively minimize non-specific interactions. In contrast, incubation for 30 min (Supplementary Figure S3B) led to a clear increase in Rct, indicative of effective surface blocking. Proper BSA blocking was therefore essential to reduce interfacial variability and to ensure selective b-TBA immobilization (Kim and Yang, 2024).

BSA adsorption is expected to occur preferentially at surface sites not occupied by Av on the GCE/f-rGO platform. These regions are primarily associated with methylene-rich (-CH2-) domains located on the basal planes of rGO. The affinity of BSA toward hydrophobic, methylene-containing domains has been previously reported (Contreras-Naranjo and Aguilar, 2019), while the presence of such groups in the proposed rGO nanohybrid was confirmed by FTIR spectroscopy (Figure 1B). The selected sequence of surface functionalization was established based on preliminary optimization experiments, in which reversing the order of BSA blocking and b-TBA immobilization resulted in a lower analytical response (data not shown). This design was guided by the architecture of the proposed sensing platform, which involves two consecutive affinity-driven processes: aptamer immobilization through the avidin-biotin interaction followed by target recognition by the immobilized aptamer. Importantly, this blocking strategy does not appear to interfere with subsequent b-TBA immobilization, since BSA passivates non-specific adsorption sites on the carbonaceous surface, whereas aptamer attachment occurs through the specific and high-affinity biotin−avidin interaction. Therefore, BSA blocking prior to b-TBA immobilization allows minimizing non-specific interactions while preserving the ability of the platform to support subsequent aptamer immobilization through the avidin-biotin interaction.

Finally, the influence of Av concentration on the biofunctionalization process was investigated (Supplementary Figure S4). Comparison of Supplementary Figure S4A,B shows that platforms prepared with 1.00 mg mL-1 Av exhibited higher absolute Rct values after sequential functionalization than those prepared with 0.10 mg mL-1 Av, indicating a more pronounced modification of the electrode interface under the selected conditions. Moreover, both avidin concentrations enabled successful formation of the avidin-biotin recognition interface, as evidenced by the additional increase in Rct observed after aptamer coupling. Therefore, the selection of 1.00 mg mL-1 Av was not based on a markedly different relative Rct variation after aptamer immobilization, but rather on the higher absolute electrochemical response obtained after complete biofunctionalization. The increase in Rct can be interpreted as an indicator of successful interface modification, reflecting the formation of a protein layer that acts as a physical barrier to electron transfer at the electrode/electrolyte interface (Gutierrez et al., 2019; Nemčeková and Labuda, 2021). Accordingly, the selected avidin concentration was considered suitable for the construction of the biofunctional interface employed in the subsequent aptasensing experiments.

Supplementary Table S1 summarizes the experimental conditions evaluated during the optimization of GO reduction, ultrasound-assisted processing time, and biofunctionalization parameters, together with the selected optimal conditions employed for construction of the final sensing platforms. Under these conditions, a stable and reproducible biofunctional interface was obtained, combining mild GO reduction, effective surface blocking, and efficient b-TBA immobilization while preserving the avidin-mediated immobilization capability required for biosensor assembly, thereby supporting the subsequent construction of the aptasensing platform.

3.3.3 Effect of b-TBA immobilization time

Once the optimal biofunctionalization and blocking conditions (GCE/f-rGO/BSA) were established, the incubation time for b-TBA immobilization (1.45 × 10−5 M) was optimized. As shown in Supplementary Figure S5, Rct increased progressively with increasing incubation time, indicating a time-dependent surface coverage of the aptamer. A noticeable increase in Rct, accompanied by a consistent electrochemical response, was observed after 30 min of incubation, reflecting the immobilization of b-TBA on the avidin-functionalized surface. Extending the incubation time to 60 min resulted in a significantly higher Rct increment, consistent with increased aptamer immobilization and a more effective occupation of available avidin binding sites.

After exposure to α-th (2.67 × 10−6 M), platforms prepared using 60 min of aptamer immobilization exhibited a substantially larger Rct increase compared to those incubated for 30 min. This behavior confirms that longer incubation times favour the formation of a functional recognition layer, leading to enhanced sensitivity of the aptasensing interface. Based on these results, 60 min was selected as the optimal condition for b-TBA immobilization.

The specificity of the biosensing platform was subsequently evaluated through control experiments, as summarized in Supplementary Figure S6. As shown in Supplementary Figure S6a, the GCE/f-rGO/BSA platform exhibited the baseline Rct value prior to aptamer immobilization. Following b-TBA immobilization, an increase in Rct was observed for the GCE/f-rGO/BSA/b-TBA platform (Supplementary Figure S6b), consistent with the successful incorporation of the biotinylated aptamer onto the avidin-functionalized interface. In contrast, when the GCE/f-rGO/BSA platform was directly exposed to α-thrombin (2.67 × 10−6 M) in the absence of the aptamer, no significant change in Rct was observed (Supplementary Figure S6c). This result indicates that direct interaction of α-thrombin with the avidin/BSA-modified interface is negligible under the experimental conditions employed.

Altogether, these results demonstrate that (i) efficient b-TBA immobilization is achieved within 60 min of incubation, (ii) the non-specific interaction of the target with the sensing interface is negligible, indicating that the blocking strategy effectively suppresses non-specific binding, thereby supporting the selectivity of the proposed aptasensor. Overall, the stepwise construction of the aptasensing platform and the subsequent biorecognition event are validated by systematic changes in charge transfer resistance, as shown in Figure 3A, a–e and Figure 3B, a–e.

FIGURE 3

The Nyquist plots in Figure 3A, a–e illustrate the successive stages of aptasensor assembly, beginning with the bare GCE (a), followed by its modification with reduced and biofunctionalized graphene oxide (GCE/f-rGO, b), BSA blocking (GCE/f-rGO/BSA, c), immobilization of the biotinylated thrombin-binding aptamer (GCE/f-rGO/BSA/b-TBA, d), and, finally, α-th recognition (GCE/f-rGO/BSA/b-TBA/α-th, e). The progressive evolution of the impedance response reflects the sequential modification of the electrode/electrolyte interface during platform fabrication and target recognition. The corresponding Rct values extracted from the impedance spectra are summarized in Figure 3B, a–e.

3.4 Analytical performance of GCE/f-rGO/BSA/b-TBA interface

The physiological activity of α-th is generally quantified in NIH units/mL (National Institutes of Health per milliliter), where 1 NIH unit is defined as the amount of thrombin required to clot 1 mL of a standard fibrinogen solution under specified conditions. In human plasma, α-th concentrations are generally reported within the nanomolar-to-micromolar range. In healthy individuals, plasma α-th concentration typically ranges from 4.70 × 10−9 M to 1.00 × 10−7 M (Yang and Ellington, 2008; Bravo et al., 2020), while elevated levels have been associated with several pathological conditions.

Accordingly, the analytical performance of the GCE/f-rGO/BSA/b-TBA platform for α-th detection was evaluated by EIS using increasing target concentrations. Representative Nyquist plots recorded after incubation with different α-th concentrations are shown in Figure 4A, a–i, while the target-free condition (0 M) is shown in the inset of Figure 4A. The corresponding charge transfer resistance (Rct) values as a function of α-th concentration are shown in Figure 4B. The target-free condition corresponds to the GCE/f-rGO/BSA/b-TBA interface prior to α-th exposure and is shown to facilitate visualization of the effect of target interaction on the EIS response. The Rct values obtained after target interaction were used directly for calibration, without additional normalization or baseline correction. The Rct response increased systematically with increasing α-th concentration within the investigated concentration series, consistent with the progressive formation of the aptamer-target complex at the electrode surface and the resulting physical barrier to electron transfer between the redox probe and the electrode surface (Bao et al., 2025). At the lowest α-th concentrations, Rct values lower than the Rct value of the target-free condition were observed. This behavior may be associated with target-induced conformational rearrangement of the immobilized aptamer upon binding, which may alter the spatial organization of the interfacial recognition layer and consequently modify the accessibility of the redox probe to the electrode surface (Gu et al., 2018). A similar behavior has been previously observed during the development of this sensing platform and was interpreted as being associated with conformational reorganization of the aptamer following target binding, potentially facilitating access of the redox probe to the electrode surface.

FIGURE 4

When Rct was plotted as a function of –log[α-th], a linear relationship was obtained over the concentration range from 1.34 × 10−9 M to 2.67 × 10−6 M (R2 = 0.987), as shown in the calibration curve (inset) of Figure 4B. This relationship demonstrates the suitability of the proposed platform for quantitative α-th detection under label-free conditions. The linear range obtained encompasses α-th concentration levels reported under both physiological and pathological conditions, supporting the potential applicability of the proposed platform in clinically relevant concentration domains (Yang and Ellington, 2008; Bravo et al., 2020).

The LOQ was determined experimentally as the lowest concentration that could be quantified with acceptable analytical performance, in accordance with the current IUPAC definition, which defines the LOQ as the smallest measured quantity value fulfilling predefined fitness-for-purpose requirements (Hibbert et al., 2021). Because this established LOQ represents the quantification capability of the proposed analytical platform, the LOD was estimated from the conventional ICH/IUPAC relationship between LOQ (10σ/S) and LOD (3.3σ/S). This approach preserves the accepted statistical relationship between both limits while anchoring the calculation to the experimentally verified quantification limit of the method. Considering the fixed microliter-scale sample volume employed (10 μL), the achieved LOQ corresponds to an absolute amount of 13.4 fmol of α-th, highlighting the ability of the platform to detect extremely low amounts of analyte. These results demonstrate that the proposed biofunctionalized rGO platform enables α-th detection at the femtomole level without the need for labelling or signal amplification strategies.

3.5 Aptasensor selectivity

The selectivity of the aptasensor was evaluated by comparing the EIS response of the target-free GCE/f-rGO/BSA/b-TBA platform (Figure 5a) with the responses obtained after separate exposure to human α-thrombin (2.67 × 10−7 M) (b), cytochrome C (Cyt C, 8.10 × 10−7 M) (c), myoglobin (Myo, 5.90 × 10−7 M) (d), lysozyme (Lys, 7.10 × 10−7 M) (e), human serum samples (Standatrol S-E Level 1, tenfold dilution) (f), and human serum samples supplemented with Cyt C, Myo, and Lys at the same concentrations (g). Conditions (c–g) were evaluated without added α-thrombin. The selected proteins (Cyt C, Myo, and Lys) were employed as representative non-target biomacromolecules with different molecular sizes, structures, and charge distributions to challenge the specificity of the aptamer-based recognition interface.

FIGURE 5

Additionally, experiments were conducted at three α-th concentrations under different conditions: (1) α-th alone, (2) α-th in the presence of interfering proteins, and (3) α-th in enriched commercial serum samples (Standatrol S-E Level 1, 10x dilution) containing the same interferents (Supplementary Table S2). The impedance response toward α-th remained essentially unchanged in the presence of human serum and human serum with an excess of these interferents, indicating negligible non-specific adsorption. In all cases, the aptasensor exhibited a consistent, concentration-dependent electrochemical response, enabling reliable detection and quantification under complex matrix conditions. The relative standard deviation (RSD) values obtained for different conditions were within the expected range for electrochemical assays, with α-th recoveries between 96% and 104% (Supplementary Table S2).

To further challenge the sensing interface, additional experiments were performed using a broad panel of individual interferents, including amino acids (L-cysteine, L-tyrosine, L-tryptophan, and L-histidine), electroactive biomolecules (uric acid, glucose, and dopamine), inorganic salts (KCl, NaCl, MgCl2, and Na2CO3), and commercial human serum samples containing elevated concentrations of serum components (Standatrol S-E Level 2, 10x dilution) (Supplementary Figure S7A). In addition, mixtures of amino acids, electroactive biomolecules, inorganic salts, and all tested species simultaneously were evaluated (Supplementary Figure S7B), representing a comprehensive evaluation of potential interferences under the conditions investigated. Negligible interference was observed under all tested conditions, confirming the high selectivity of the proposed aptasensor toward α-th.

It should be noted that the selection of interferents varies among reported graphene-based thrombin aptasensing platforms (Table 1) depending on the sensing architecture and the intended application. Therefore, direct comparison of selectivity studies should consider the different interferent panels and experimental conditions employed in each case.

TABLE 1

PlatformOne pot fabricationLabel-free detectionTransductionLOD (nM)Linear rangeSelectivityReal SampleStabilityReference
Au/nafion@graphene/MB/ nano-Au/BSA/TBAYesMB reporterDPV0.006 (6.0 pM)(0.01 – 50) nMBSA, IgG, & HBNR20 days at 4 °CSun et al. (2011)
Fc-TBA/SH-β-CD-Gr/AuNPs/GCEYesFc-labelledDPV0.000001 (1.0 × 10-15 M)(1.6 × 10-15 - 8.0 × 10-13) MBSA, HSA, LYS & INSNR2 weeks at 4 °CXue et al. (2015)
TBA/GO/PGENoYesDPV0.07(0.1-10) nMHSA, BSA, Tyr, PH-A & HBDiluted serum samples (100×)1 week at 4 °CAhour and Ahsani (2016)
TA/Fe2O3@graphene/GCENoYesDPV0.001 (1.0 × 10−12 M)(1.0 × 10-11–4.0 × 10-9) MBSA, HSA, LYS & INSNR15 days at 4 °CZhang et al. (2017)
TBA/FGO@Fe-MOG/DEPNoYesEIS0.06 (2.2 ng mL−1)(2–14) ng mL-1BSA, IgG, & STREPDiluted serum samples (10×)NRUrbanová et al. (2018)
AuUME /TBA2/thrombin/TBA1-AgNP-GONoSandwich approach (label: TBA1-AgNP-GO)SWV0.03 (0.03 nmol L−1)(0.05–5) nmol L-1AFP, CEA & IgGDiluted serum samples (100×)NRQin and Yang (2018)
GCE/rGO/PTCA/TBANoYesEIS0.0002 (0.2 pM)1 pM-100 nMBSANRNRWu et al. (2020)
TBA/AuNPs-ERGO/GCENoMB reporterDPV0.17(0.5–10) nMBSA & Lys in bufferNR1 week at 4 °CYang et al. (2021)
TBA/BSA/GNPs/CNOs/CS/GCENoYesDPV0.0000086 (8.61 fM)26.74 pM - 26.74 fMApoA1, CRP, VEGF, LCN2, LEPDiluted serum samples (Dilution NR)14 days (Storage conditions NR)Ansari et al. (2023)
GCE/f-rGO/BSA/b-TBAYesYesEIS0.441.34 nM- 2.67 µMCyt C, Myo, and Lys, L-cys, L-tyr, L-trp, L-hist, uric acid, glucose, dopamine, KCl, NaCl, MgCl2, and Na2CO3Diluted serum samples (10×) Level 1 and Level 2 enriched with Lys, Cyt C & Myo9 days at 4 °CThis work

Relevant electrochemical thrombin aptasensors based on graphene derivatives.

Abbreviations: AFP: alpha fetoprotein; AgNP: silver nanoparticles; ApoA1:apolipoprotein A1; AuNPs: gold nanoparticles; AuUME: gold ultramicroelectrode; BSA: bovine serum albumin; b-TBA: biotinylated-thrombin binding aptamer; CEA: carcinoembryonic antigen; CNOs: carbon nano onions; CRP: C-reactive protein; CS: chitosan; CytC: cytochrome C; DEP: disposable electrically printed electrodes; DPV: differential pulse voltammetry; EIS: electrochemical impedance spectroscopy; ERGO: electroreduced graphene oxide, Fc: ferrocene; Fe2O3@graphene: Fe2O3-coated graphene nanostructure; FGO@Fe-MOG: fluorinated graphene oxide@metal-organic gel composite; f-rGO: functionalized-reduced graphene oxide; GCE: glassy carbon electrodes; GNPs: graphene nanoplatelets; GO: graphene oxide; Gr:graphene; HB: haemoglobin; HSA: human serum albumin; IgG: immunoglobulin G; INS: insulin; LCN2:lipocalin-2; LEP: leptin; LYS: lysozyme; MB: methylene blue ; MYO: myoglobin; NR: not reported; PGE: pencil graphite electrode; PH-A: phenyl alanine; PTCA: 3, 4, 9, 10-perylenetetracarboxylic acid; rGO: reduced graphene oxide; SH-β-CD: thio-β-cyclodextrin; STREP: streptavidin; SWV: square wave voltammetry; TA: thrombin aptamer; TBA: thrombin binding aptamer; TYR: tyrosine, VEGF: vascular endothelial growth factor. Molar conversions were calculated using the molecular weight of human α-thrombin.

In summary, these results demonstrate that the GCE/f-rGO/BSA/b-TBA aptasensor enables sensitive, selective, and reproducible detection of α-th under label-free conditions, even in complex biological matrices. The analytical performance achieved highlights the effectiveness of the one-pot reduction and biofunctionalization strategy as a robust and practical platform for electrochemical aptasensing.

3.6 Aptasensor stability

The storage stability of the fabricated GCE/f-rGO/BSA/b-TBA aptasensor was evaluated by monitoring the Rct response of the platform by EIS over time under two storage conditions: at room temperature (RT, 25 °C) and at 4 °C. The evolution of the Rct was normalized and expressed as percentage signal retention relative to the initial response (Supplementary Figure S8A,B respectively). All measurements were performed in triplicate to ensure reproducibility. Electrodes stored at 4 °C exhibited stable Rct values and preserved their electrochemical response up to 9 days, suggesting effective preservation of the biofunctional interface during refrigerated storage. In contrast, electrodes stored at RT showed a gradual increase in signal variability over time, suggesting progressive alteration of the biofunctional interface, possibly associated with conformational rearrangements or desorption processes occurring at the electrode surface. Beyond 9 days of refrigerated storage, the impedance response deviated from its initial value, suggesting partial aptamer desorption or progressive degradation of the biorecognition layer. To further evaluate the preservation of sensing capability during storage, aptasensors stored at 4 °C for different periods were subsequently challenged with α-th (2.67 × 10−6 M), and the analytical response was compared with that of freshly prepared platforms (Supplementary Figure S8C). Approximately 70% of the initial analytical signal was retained after 14 days, confirming the partial preservation of sensing capability over extended storage periods. Therefore, these results demonstrate that the proposed aptasensor maintains its electrochemical performance for at least 9 days when stored under refrigerated conditions. Although longer storage periods have been reported for some electrochemical aptasensing platforms, the stability achieved in the present work appears compatible with batch preparation, short-term storage and routine analytical applications (Hianik et al., 2007; Fallah et al., 2024). The observed stability is consistent with previously reported graphene-based aptasensing platforms and suggests that the immobilization strategy employed provides a robust biofunctional interface (Aye et al., 2021; Radi and Abd-Ellatief, 2021).

Table 1 summarizes representative graphene-based electrochemical aptasensors reported for detection of α-th in the last years (Ahour and Ahsani, 2016; Xue et al., 2015; Sun et al., 2011; Zhang et al., 2017; Urbanová et al., 2018; Qin and Yang, 2018; Wu et al., 2020; Yang et al., 2021; Ansari et al., 2023). Although some graphene-based platforms achieve lower limits of detection, most reported systems rely on multistep fabrication procedures, harsh chemical reducing agents, labeled detection schemes, or complex sandwich-type and nanocomposite architectures, which may increase fabrication complexity and limit practical applicability. In contrast, the proposed platform integrates mild and environmentally benign chemical reduction with in situ biofunctionalization in a single-step, one-pot process, while maintaining high sensitivity, selectivity, and a simplified fabrication workflow.

This strategy enables a controlled partial restoration of the sp2 carbon network while preserving oxygen-containing functionalities that contribute to the formation of a stable biofunctional interface for aptamer immobilization.

From an analytical and biomedical perspective, the proposed aptasensor operates over a broad dynamic range spanning the nM-to-μM levels, which is particularly relevant for monitoring α-th in physiological and pathological contexts, including hemostasis, thrombotic disorders, neurodegenerative diseases, atherosclerosis, and cancer-related processes (Urbanová et al., 2018; Paimard et al., 2021; Park, 2022). Compared to most graphene-based thrombin aptasensors summarized in Table 1, the wider dynamic range may facilitate monitoring of α-th across concentration levels relevant to different physiological and pathological conditions.

3.7 Sustainability assessment

The sustainability of the proposed aptasensor was assessed using the AGREE (Analytical GREEnnes Evaluation) metric (Pena-Pereira et al., 2022; Wojnowski et al., 2022), yielding an overall score of 0.75, which indicates a high level of compliance with the principles of green analytical chemistry (Supplementary Figure S9).

This favorable sustainability profile is primarily attributed to the use of low-toxicity reducing agents (L-tyr and AA), low reagent consumption, short ultrasound-assisted processing times, and the integration of reduction and biofunctionalization within a single-step process. Collectively, these features contribute to reduced material usage, lower energy demand, and decreased procedural complexity.

Minor limitations are mainly related to the semi-automated nature of the procedure, which still requires some operator intervention and therefore limits full automation. Nevertheless, these aspects do not significantly compromise the overall greenness of the approach and could be addressed in future developments through increased automation. In addition, future adaptation of the platform toward multiplexed detection schemes could further improve resource efficiency by reducing reagent consumption, sample requirements, and waste generation per analytical determination. Although the AGREE metric provides a comprehensive assessment of the greenness of the analytical methodology, a complete life-cycle assessment and a quantitative analysis of energy consumption were beyond the scope of the present study. Future investigations addressing these aspects would provide a more comprehensive evaluation of the overall environmental impact of the proposed approach.

Overall, the AGREE evaluation supports that the proposed aptasensor constitutes an environmentally responsible and analytically efficient platform, in line with current sustainability requirements for electrochemical biosensing.

4 Conclusion

In this work, a one-pot strategy for graphene oxide processing was developed and systematically evaluated as a versatile route for the fabrication of biofunctionalized graphene-based interfaces. The proposed approach enables the generation of partially reduced, biofunctionalized graphene platforms that combine controlled surface chemistry with favourable electrochemical properties, offering a simplified and sustainable alternative to conventional multistep graphene processing routes for electrochemical biosensing applications.

Multimodal spectroscopic and morphological characterization demonstrated that the synergistic action of mild and complementary reducing agents under ultrasound-assisted conditions enables controlled partial restoration of the sp2 carbon framework while preserving oxygen-containing surface functionalities required for efficient biomolecular anchoring. This balance between electronic conductivity and surface chemistry results in robust interfaces with suitable electrochemical performance.

Electrochemical impedance analysis revealed a progressive and reproducible evolution of interfacial properties during biofunctionalization and target recognition, highlighting the capability of the platform to support selective molecular interactions in a label-free configuration. The sensing performance obtained in protein-enriched human serum samples further illustrates the robustness and analytical applicability of the proposed interface design.

Sustainability assessment using the AGREE metric underscored the favourable green profile of the proposed methodology, reflecting the use of low-toxicity reagents, mild processing conditions, reduced procedural complexity, and minimized reagent consumption.

In summary, these results validate the proposed integrated approach, demonstrating that controlled partial reduction combined with simultaneous biointerface formation constitutes an effective strategy for engineering stable and electrochemically active graphene-based bioelectrochemical interfaces. Notably, the demonstrated 9-day storage stability under refrigerated conditions, coupled with the one-pot protocol, highlights the platform’s potential for batch preparation and streamlined fabrication workflows. Further studies should address the long-term integrity of the biofunctional layer, including possible effects of protein denaturation or biomolecule desorption and their impact on sensor reliability over extended storage periods. Although reproducible results were obtained under the experimental conditions employed, future studies should further investigate process standardization and potential batch-to-batch variations associated with the partial reduction process.

Beyond the specific thrombin aptasensing application demonstrated here, the modular nature of the proposed biointerface engineering strategy provides a versatile foundation for the development of next-generation graphene-based electrochemical biosensors and integrated analytical platforms targeting a broad range of biomolecular analytes. Furthermore, the versatility of the avidin-mediated biofunctionalization approach may facilitate future integration into portable, microfluidic, and point-of-care sensing platforms.

Statements

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.

Author contributions

DFR: Data curation, Visualization, Investigation, Writing – original draft, Validation, Formal Analysis. GAR: Funding acquisition, Resources, Formal Analysis, Methodology, Writing – review and editing. MDR: Writing – review and editing, Methodology, Conceptualization, Resources, Formal Analysis, Funding acquisition. MCR: Funding acquisition, Supervision, Writing – review and editing, Formal Analysis, Project administration, Methodology, Resources, Conceptualization.

Funding

The author(s) declared that financial support was received for this work and/or its publication. The authors acknowledge the National Agency of Promotion of Research, Technological Development and Innovation (ANPCyT, Argentina), PICT No. 2018–01177 and PICT No. 2020-SERIE A-1523, and the Secretariat of Science and Technology of the National University of Córdoba (SECyT-UNC), Res. No. 411/18, for financial support.

Acknowledgments

The authors wish to thank MSc María E. Bernardi and MSc Ricardo E. M. Lombardi from Laboratorio de Hemoderivados (Universidad Nacional de Córdoba) for providing human α-thrombin. DFR acknowledges CONICET for the fellowship received.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author MCR declared that they were an editorial board member of Frontiers at the time of submission. This had no impact on the peer review process and the final decision.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fsens.2026.1864839/full#supplementary-material

References

  • 1

    AbdelhalimA. O.SharoykoV. V.MeshcheriakovA. A.MartynovaS. D.AgeevS. V.IurevG. O.et al (2020). Reduction and functionalization of graphene oxide with L-cysteine: synthesis, characterization and biocompatibility. Nanomedicine Nanotechnol. Biol. Med.29, 102284. 10.1016/j.nano.2020.102284

  • 2

    AhourF.AhsaniM. K. (2016). An electrochemical label-free and sensitive thrombin aptasensor based on graphene oxide modified pencil graphite electrode. Biosens. Bioelectron.86, 764–769. 10.1016/j.bios.2016.07.053

  • 3

    AliyevE.FilizV.KhanM. M.LeeY. J.AbetzC.AbetzV. (2019). Structural characterization of graphene oxide: surface functional groups and fractionated oxidative debris. Nanomaterials9 (8), 1180. 10.3390/nano9081180

  • 4

    AnsariM. A.LiewW. J.KurupC. P.AhmedM. U. (2023). Label-free electrochemical aptasensor for ultrasensitive thrombin detection using graphene nanoplatelets and carbon nano onion-based nanocomposite. J. Electroanal. Chem.937, 117422. 10.1016/j.jelechem.2023.117422

  • 5

    AyeN. N. S.MaramingP.TavichakorntrakoolR.ChaibunruangA.BoonsiriP.DaduangS.et al (2021). A simple graphene functionalized electrochemical aptasensor for the sensitive and selective detection of glycated albumin. Appl. Sci.11 (21), 10315. 10.3390/app112110315

  • 6

    BakryA. M.AlamierW. M.AbdallahA. B.El-ReashY. G. A.El-ShallM. S.AwadF. S. (2024). Enhanced performance of amine and thiol chemically modified graphene oxide for effective removal of Hg (II), Pb (II), and Cr (VI) from aqueous solutions. Appl. Water Sci.14 (8), 179. 10.1007/s13201-024-02234-y

  • 7

    BaoW.AodengG.GaL.AiJ. (2025). Aptamer-based electrochemical biosensors: signal transduction mechanisms, application progress, and future trends. Sensors Actuators Rep.10, 100366. 10.1016/j.snr.2025.100366

  • 8

    BockL. C.GriffinL. C.LathamJ. A.VermaasE. H.TooleJ. J. (1992). Selection of single-stranded DNA molecules that bind and inhibit human thrombin. Nature355 (6360), 564–566. 10.1038/355564a0

  • 9

    BravoM. I.RaventósA.PérezA.CostaM.WillisT. (2020). Non-additive effect on thrombin generation when a plasma-derived factor VIII/von Willebrand factor (FVIII/VWF) is combined with emicizumab in vitro. J. Thromb. Haemost. 18 (8), 1934–1939. 10.1111/jth.14887

  • 10

    CaiW.MaW.ChenW.LiuP.LiuY.LiuZ.et al (2022). Microwave-assisted reduction and sintering to construct hybrid networks of reduced graphene oxide and MXene for electromagnetic interference shielding. Compos. Part A Appl. Sci. Manuf.157, 106928. 10.1016/j.compositesa.2022.106928

  • 11

    ChuZ.ZhuangS.LuJ.LiJ.WangC.WangT. (2023). In-situ electro-polymerization of L-tyrosine enables ultrafast, long cycle life for lithium metal battery. Chin. Chem. Lett.34 (5), 107563. 10.1016/j.cclet.2022.05.077

  • 12

    Contreras-NaranjoJ. E.AguilarO. (2019). Suppressing non-specific binding of proteins onto electrode surfaces in the development of electrochemical immunosensors. Biosensors9 (1), 15. 10.3390/bios9010015

  • 13

    DasN. S.GogoiK. K.ChowdhuryA.RoyA. (2023). Investigation of optical and structural properties of synthesized graphene oxide and thermally reduced graphene oxide. Mater. Today Proc.76, 160–165. 10.1016/j.matpr.2022.10.295

  • 14

    FallahA.FooladiA. A. I.HavaeiS. A.MahboobiM.SedighianH. (2024). Recent advances in aptamer discovery, modification and improving performance. Biochem. Biophysics Rep.40, 101852. 10.1016/j.bbrep.2024.101852

  • 15

    GilM. R. (2025). “Thrombin generation assays,” in Transfusion Medicine and Hemostasis. Editors ShazB. H.HillyerC. D.SchwartzJ.Reyes GilM.Fourth Ed. (New York: Elsevier), 825–826.

  • 16

    GuQ.NanneyW.CaoH. H.WangH.YeT. (2018). Single molecule profiling of molecular recognition at a model electrochemical biosensor. J. Am. Chem. Soc. 140 (43), 14134–14143. 10.1021/jacs.8b07325

  • 17

    GutierrezF. A.RubianesM. D.RivasG. A. (2019). New bioanalytical platform based on the use of avidin for the successful exfoliation of multi-walled carbon nanotubes and the robust anchoring of biomolecules. Application for hydrogen peroxide biosensing. Anal. Chim. Acta1065, 12–20. 10.1016/j.aca.2019.03.022

  • 18

    HardyM.DouxfilsJ.XhaetO.RobayeB.LessireS.LecompteT.et al (2024). Development and validation of an in vitro model to study thrombin generation on the surface of catheters in platelet-poor and platelet-rich plasma. Thrombosis Research244, 109194. 10.1016/j.thromres.2024.109194

  • 19

    HianikT.OstatnáV.SonlajtnerovaM.GrmanI. (2007). Influence of ionic strength, pH and aptamer configuration for binding affinity to thrombin. Bioelectrochemistry70 (1), 127–133. 10.1016/j.bioelechem.2006.03.012

  • 20

    HibbertD. B.KorteE. H.ÖrnemarkU. (2021). Metrological and quality concepts in analytical chemistry (IUPAC recommendations 2021). Pure Appl. Chem.93 (9), 997–1048. 10.1515/pac-2019-0819

  • 21

    JalitY.GutierrezF. A.DubachevaG.GoyerC.Coche-GuerenteL.DefrancqE.et al (2013). Characterization of a modified gold platform for the development of a label-free anti-thrombin aptasensor. Biosens. Bioelectron.41, 424–429. 10.1016/j.bios.2012.08.061

  • 22

    KimG.YangH. (2024). Electrochemical biosensor using direct electron transfer and an antibody–aptamer hybrid sandwich for target detection in complex biological samples. Biosens. Bioelectron.253, 116184. 10.1016/j.bios.2024.116184

  • 23

    KorucuH. (2022). Evaluation of the performance on reduced graphene oxide synthesized using ascorbic acid and sodium borohydride: experimental designs‐based multi‐response optimization application. J. Mol. Struct.1268, 133715. 10.1016/j.molstruc.2022.133715

  • 24

    LeG. T.ChanlekN.ManyamJ.OpaprakasitP.GrisdanurakN.SreearunothaiP. (2019). Insight into the ultrasonication of graphene oxide with strong changes in its properties and performance for adsorption applications. Chem. Eng. J.373, 1212–1222. 10.1016/j.cej.2019.05.108

  • 25

    LeninR. A.NagaraniS.KumarM.RameshkumarS.ChenC. L.KhanF. S.et al (2025). Ecofriendly fabrication and theoretical insights of ascorbic acid assisted rGO electrodes for high performance solid state supercapacitors. Sci. Rep.15 (1), 26318. 10.1038/s41598-025-11896-w

  • 26

    NemčekováK.LabudaJ. (2021). Advanced materials-integrated electrochemical sensors as promising medical diagnostics tools: a review. Mater. Sci. Eng. C120, 111751. 10.1016/j.msec.2020.111751

  • 27

    PaimardG.GholivandM. B.ShamsipurM.AhmadiE.ShahlaeiM. (2021). Introduction of a thrombin sensor based on its interaction with dabigatran as an oral direct thrombin inhibitor. Mater. Sci. Eng. C119, 111417. 10.1016/j.msec.2020.111417

  • 28

    PalombaM.CarotenutoG.LongoA. (2022). A brief review: the use of L-ascorbic acid as a green reducing agent of graphene oxide. Materials15 (18), 6456. 10.3390/ma15186456

  • 29

    ParkK. (2022). Impedance technique-based label-free electrochemical aptasensor for thrombin using single-walled carbon nanotubes-casted screen-printed carbon electrode. Sensors22 (7), 2699. 10.3390/s22072699

  • 30

    ParkO. K.KimN. H.LeeJ. H. (2022). Rapid effective reduction by microwave-irradiated thermal reaction for large-scale production of high-quality reduced graphene oxide. Carbon187, 330–337. 10.1016/j.carbon.2021.11.018

  • 31

    Pena-PereiraF. J.TobiszewskiM.WojnowskiW.PsillakisE. (2022). A tutorial on AGREEprep an analytical greenness metric for sample preparation. Adv. Sample Prep.3, 100025. 10.1016/j.sampre.2022.100025

  • 32

    QinB.YangK. (2018). Voltammetric aptasensor for thrombin by using a gold microelectrode modified with graphene oxide decorated with silver nanoparticles. Microchim. Acta185 (9), 407. 10.1007/s00604-018-2924-1

  • 33

    RadiA. E.Abd-EllatiefM. R. (2021). Electrochemical aptasensors: current status and future perspectives. Diagnostics11 (1), 104. 10.3390/diagnostics11010104

  • 34

    SalmahaminatiRoca-SanjuanD. (2024). The photophysics and photochemistry of phenylalanine, tyrosine, and tryptophan: a CASSCF/CASPT2 study. ACS Omega9 (33), 35356–35363. 10.1021/acsomega.4c00875

  • 35

    SaputraA. M. A.MarpongahtunA.GoeiR.TokA. I. Y.GoutianosS.GeaS.et al (2025). Facile synthesis of reduced graphene oxide using natural plant extracts as green reducing agents. J. King Saud University–Engineering Sci.37 (3), 5. 10.1007/s44444-025-00007-9

  • 36

    StobinskiL.LesiakB.MalolepszyA.MazurkiewiczM.MierzwaB.ZemekJ.et al (2014). Graphene oxide and reduced graphene oxide studied by the XRD, TEM and electron spectroscopy methods. J. Electron Spectrosc. Relat. Phenom.195, 145–154. 10.1016/j.elspec.2014.07.003

  • 37

    SunT.WangL.LiN.GanX. (2011). Label-free electrochemical aptasensor for thrombin detection based on the nafion@ graphene as platform. Bioprocess Biosystems Engineering34 (9), 1081–1085. 10.1007/s00449-011-0558-3

  • 38

    TarighatM. A.BarghandanF.HashemifardS. A.AbdiG. (2025). Development and characterization of PAN/GO-tyrosine hollow fiber membranes for enhanced heavy metal adsorption and SPME-Spectrophotometric detection. RSC Advances15 (5), 3721–3737. 10.1039/D4RA08423C

  • 39

    TewatiaK.SharmaA.SharmaM.KumarA. (2021). Synthesis of graphene oxide and its reduction by green reducing agent. Mater. Today Proc.44, 3933–3938. 10.1016/j.matpr.2020.09.294

  • 40

    UrbanováV.JayaramuluK.SchneemannA.KmentS.FischerR. A.ZbořilR. (2018). Hierarchical porous fluorinated graphene oxide@ metal–organic gel composite: label-free electrochemical aptasensor for selective detection of thrombin. ACS Applied Materials and Interfaces10 (48), 41089–41097. 10.1021/acsami.8b14344

  • 41

    UtkanG.YumusakG.TunaliB. C.OzturkT.TurkM. (2023). Production of reduced graphene oxide by using three different microorganisms and investigation of their cell interactions. ACS Omega8 (34), 31188–31200. 10.1021/acsomega.3c03213

  • 42

    WangQ.LiM.SzuneritsS.BoukherroubR. (2014). Environmentally friendly reduction of graphene oxide using tyrosine for nonenzymatic amperometric H2O2 detection. Electroanalysis26 (1), 156–163. 10.1002/elan.201300356

  • 43

    WangS.HossainM. Z.HanT.ShinozukaK.SuzukiT.KuwanaA.et al (2020a). Avidin–biotin technology in gold nanoparticle-decorated graphene field effect transistors for detection of biotinylated macromolecules with ultrahigh sensitivity and specificity. ACS Omega5 (46), 30037–30046. 10.1021/acsomega.0c04429

  • 44

    WangS.HossainM. Z.ShinozukaK.ShimizuN.KitadaS.SuzukiT.et al (2020b). Graphene field-effect transistor biosensor for detection of biotin with ultrahigh sensitivity and specificity. Biosens. Bioelectron.165, 112363. 10.1016/j.bios.2020.112363

  • 45

    WojnowskiW.TobiszewskiM.Pena-PereiraF.PsillakisE. (2022). AGREEprep – analytical greenness metric for sample preparation. Trac. Trends Anal. Chem.149, 116553. 10.1016/j.trac.2022.116553

  • 46

    WuH.XiK.XiaoS.NgaiS.ZhouC.HeM.et al (2020). Self-assembled perylenetetracarboxylic acid-reduced graphene oxide film for high-sensitive impedimetric determination of thrombin. Surf. Coatings Technol.402, 126491. 10.1016/j.surfcoat.2020.126491

  • 47

    XueQ.LiuZ.GuoY.GuoS. (2015). Cyclodextrin functionalized graphene–gold nanoparticle hybrids with strong supramolecular capability for electrochemical thrombin aptasensor. Biosens. Bioelectron.68, 429–436. 10.1016/j.bios.2015.01.025

  • 48

    XueC.GaoH.HuY.HuG. (2019). Hyperelastic characteristics of graphene natural rubber composites and reinforcement and toughening mechanisms at multi-scale. Compos. Struct.228, 111365. 10.1016/j.compstruct.2019.111365

  • 49

    YangL.EllingtonA. D. (2008). Real-time PCR detection of protein analytes with conformation‐switching aptamers. Anal. BioChem. 380, 164–173. 10.1016/j.ab.2008.05.018

  • 50

    YangJ.ZhangY. (2025). Research progress on aptamer electrochemical biosensors based on signal amplification strategy. Sensors25 (14), 4367. 10.3390/s25144367

  • 51

    YangH.HuP.TangJ.ChengY.WangF.ChenZ. (2021). A bifunctional electrochemical aptasensor based on AuNPs-coated ERGO nanosheets for sensitive detection of adenosine and thrombin. J. Solid State Electrochem.25 (4), 1383–1391. 10.1007/s10008-021-04916-y

  • 52

    YeS.FengJ. (2016). The effect of sonication treatment of graphene oxide on the mechanical properties of the assembled films. RSC Advances6 (46), 39681–39687. 10.1039/C6RA03996K

  • 53

    ZafarS.ManiS. K.MonishaM.LochabB. (2025). Concentration-dependent kinetic study of graphene oxide (GO) reduction using biophenol and electrochemical analysis. RSC Sustain.3 (1), 365–375. 10.1039/D4SU00430B

  • 54

    ZhangH.ZhangB.ChenA.QinY. (2017). Controllable n-Fe2O3@graphene nanomaterials by ALD applied in an aptasensor with enhanced electrochemical performance for thrombin detection. Dalton Trans.46 (23), 7434–7440. 10.1039/C7DT01184A

  • 55

    ZhangH.QuH.HeQ.GaoL.ZhangH.WangY.et al (2021). Thrombus-targeted nanoparticles for thrombin-triggered thrombolysis and local inflammatory microenvironment regulation. J. Control. Release339, 195–207. 10.1016/j.jconrel.2021.06.043

Summary

Keywords

avidin-biotin interaction, human α-thrombin, label-free electrochemical sensing, modular biosensing platform, one-pot biofunctionalization, reduced graphene oxide

Citation

Reartes DF, Rivas GA, Rubianes MD and Rodríguez MC (2026) Toward a modular label-free electrochemical biosensing platform based on a one-pot avidin functionalization of reduced graphene oxide. Front. Sens. 7:1864839. doi: 10.3389/fsens.2026.1864839

Received

24 April 2026

Revised

09 August 2026

Accepted

02 September 2026

Published

26 October 2026

Volume

7 - 2026

Edited by

Dermot Diamond, Dublin City University, Ireland

Reviewed by

Shivaji D. Pawar, Universal AI University, India

Diana Raquel De Lima Cunha, University of Porto, Portugal

Updates

Copyright

*Correspondence: Marcela Cecilia Rodríguez, ; María Dolores Rubianes,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics