REVIEW article

Front. Sens., 27 October 2026

Sec. Biosensors

Volume 7 - 2026 | https://doi.org/10.3389/fsens.2026.1933492

Enzymatic biosensors: biocatalytic mechanisms, immobilization strategies, and signal amplification approaches

  • 1. Universidad Autónoma de Nuevo León, Facultad de Ciencias Químicas, Av. Universidad S/N Ciudad Universitaria, San Nicolás de los Garza, Mexico

  • 2. Centro de Investigación en Biotecnología y Nanotecnología (CIByN), Facultad de Ciencias Químicas, Universidad Autónoma de Nuevo León. Parque de Investigación e Innovación Tecnológica, Apodaca, Mexico

  • 3. Pelucco R&D Division, Apodaca, Mexico

  • 4. Instituto Nacional de Astrofísica, Óptica y Electrónica, Coordinación de Ciencias Computacionales, Puebla, Mexico

Abstract

Enzymatic biosensors are a crucial component of modern analytical science, offering exceptional molecular specificity across clinical, environmental, and industrial applications. Despite their widespread use, the development of reliable biosensor platforms continues to face challenges, particularly regarding enzymatic stability, reproducibility, and long-term operational lifespan. This review critically examines the core principles of high-performance enzymatic biosensors, integrating biochemical fundamentals with advances in materials. Our analysis begins with the kinetic and thermodynamic factors that govern sensor sensitivity and dynamic range, focusing on how immobilization alters Michaelis–Menten parameters and how diffusional and mass-transport limitations impact traditional kinetic models. We also examine environmental factors such as pH, temperature, and cofactor regeneration. Special attention is given to interfacial engineering, comparing oriented and random immobilization, considering factors such as fabrication complexity, cost, enzyme-specific applicability, and activity losses due to modifications. The review evaluates various signal transduction mechanisms, including electrochemical, optical, semiconductor-based, wearable, and emerging in-sensor computing architectures. Using gas-phase biosensing of volatile organic compounds as a case study, we highlight how this niche stresses every design parameter, revealing it to be smaller and more chemically constrained than is often implied. Finally, we discuss major unresolved challenges, especially the lack of standardized benchmarking methods that would permit fair comparison between competing strategies.

1 Introduction

Biosensors have become essential tools in modern analytical science, enabling the rapid and selective detection of biologically significant molecules in clinical, environmental, and industrial settings. Among these, enzymatic biosensors are notable for their inherent molecular specificity and well-characterized catalytic mechanisms (Ghosh et al., 2019; Coelho et al., 2019). By combining enzyme-catalyzed reactions with electrochemical, optical, thermal, or piezoelectric transducers, these systems transform biochemical events into quantifiable analytical signals (Kaur and Singh, 2020).

Since Clark and Lyons described the first enzyme electrode in 1962, in which glucose oxidase was held against an oxygen electrode by a semipermeable membrane (Clark and Lyons, 1962), enzymatic biosensors have advanced considerably. However, key engineering parameters still govern their analytical performance, including enzyme immobilization methods, interfacial electron-transfer efficiency, signal-amplification techniques, and material selection (Anjum et al., 2022; Nguyen et al., 2019). Challenges related to enzyme stability, reproducibility, and long-term operational performance remain, driving ongoing development of advanced immobilization techniques and nanostructured interfaces (Singh, 2011).

Recent progress in nanomaterials, surface functionalization, and bioelectronic integration has improved sensitivity, stability, and miniaturization. These innovations are transforming enzymatic biosensor design, making integration into portable and intelligent analytical platforms easier.

This review critically examines the engineering strategies and material innovations impacting enzymatic biosensor performance. Our focus includes immobilization techniques, nano-enabled architectures, transduction efficiency, and the current challenges shaping the future development of robust, scalable biosensing systems.

1.1 Literature search strategy

This narrative review was based on a structured search of Scopus, Web of Science, and PubMed, complemented by targeted searches in Google Scholar. The search covered publications from January 2015 to 2026, with particular emphasis on studies published during the last 5 years (2021–2026). Earlier foundational studies were also included when they described concepts that remain relevant to current biosensor research. Search strategies combined the terms enzymatic biosensor, enzyme immobilization, direct electron transfer, signal amplification, nanomaterial, and volatile organic compound using Boolean operators, together with enzyme- and technique-specific terms such as glucose oxidase, laccase, alcohol dehydrogenase, memristor, field-effect transistor, bio-sniffer, and apparent Michaelis constant. Peer-reviewed original research articles and review papers published in English were considered. Conference abstracts without full peer review were excluded, as were studies focused on non-enzymatic recognition elements, including antibody-, aptamer-, purely chemical-, and ionophore-based sensors, except when these systems were included for comparison. Reference lists from key reviews were also examined to identify relevant studies not captured by the initial searches. For quantitative comparisons, we included only studies for which the reported numerical values, units, and measurement conditions could be verified in the original publication.

2 Kinetic and thermodynamic foundations of enzymatic biosensing

2.1 Michaelis–Menten description of the free enzyme

Enzymatic biosensor performance depends fundamentally on enzyme–substrate kinetics and thermodynamic interactions. Most enzyme-based sensing platforms are described, at least nominally, by Michaelis-Menten kinetics, which relates substrate concentration to reaction rate (Equation 1):

In biosensing, the Michaelis constant (Km) indicates enzyme–substrate affinity and directly affects sensitivity, especially at low analyte concentrations. A lower Km typically enhances detection capability in diluted systems. Meanwhile, Vmax establishes the upper limit of response and helps define the sensor’s dynamic range (Khan et al., 2021; Kirthiga and Rajendran, 2015).

To estimate kinetic parameters experimentally, researchers often use linear transformations such as the Lineweaver–Burk plot. Although this double-reciprocal method can amplify errors at low substrate concentrations, it remains widely employed for comparative analyses in biosensor development (Silverstein and Goodney, 2010). Section 2.2 addresses its limitations for immobilized systems, which are more serious than the error-weighting issue.

In practical operation, enzymatic biosensors exhibit a limited linear response range. At low analyte concentrations, reaction kinetics may constrain the signal, while at high concentrations, enzyme saturation causes the signal to plateau.

2.2 Apparent kinetics, mass transport, and the limits of the classical model

The description above applies to enzyme freely dissolved in a well-mixed solution. Few biosensors operate in this regime. Once an enzyme is confined to an electrode surface, a porous carrier, or a polymer film, the measured response is no longer a direct readout of catalysis but a convolution of catalysis, substrate partitioning between phases, and diffusion through the immobilization layer and the adjacent solution boundary layer (Baronas, 2025; Baronas and Petrauskas, 2026). The parameters extracted from such a device are therefore apparent Km,app and Vmax,app, and reflect the complete sensing architecture rather than the enzyme itself (Figure 1).

FIGURE 1

The magnitude of this effect is not marginal. For lactate oxidase immobilized by the same chemistry on two different carbon matrices, apparent Michaelis constants of 73.7 mM and 8.25 mM were reported, a spread approaching an order of magnitude arising solely from the choice of electrode matrix (Jakubow-Piotrowska and Kowalewska, 2025). Laccase covalently attached to a lignocellulosic support through glutaraldehyde showed Km rising from 0.53 ± 0.032 mM in free solution to 0.72 ± 0.042 mM after immobilization, an increase the authors attributed explicitly to diffusional limitation within the porous carrier, steric hindrance near the active site, and conformational constraint from multipoint attachment (Koç and Işık, 2026). The shift is not always in the same direction: horseradish peroxidase immobilized on magnetic hybrid nanoflowers showed Km falling from 15.55 ± 0.059 mM to 7.67 ± 0.033 mM while Vmax fell roughly eighteen-fold over the same comparison (Bakar et al., 2024), demonstrating that Km,app and Vmax,app can move independently and that an improvement in apparent affinity may accompany a substantial loss of catalytic throughput.

Two transport regimes must be distinguished. External limitation arises in the stagnant solution film adjacent to the sensing layer and is sensitive to convection and cell geometry; internal limitation arises within the carrier pores and is sensitive to loading density and layer thickness. These two limitations can be experimentally distinguished, as shown for an immobilized flavin oxidase, where intraparticle oxygen sensing demonstrated the superior performance of enzyme immobilized at the outer particle surface compared with enzyme distributed throughout the pore volume; engineering carrier chemistry, functionalization, enzyme distribution, and cofactor co-immobilization together raised activity by up to 650% relative to suboptimal immobilization, with no change in the enzyme itself (Benítez-Mateos et al., 2020). Similarly, effectiveness factors between 6% and 14% have been reported for enzymes entrapped in printed hydrogel carriers, and counterintuitively, the effectiveness factor decreased as the absolute activity of the entrapped enzyme increased because a faster enzyme consumes the local substrate supply sooner (Schmieg et al., 2019).

Whether a given device is kinetically or diffusionally controlled can be established rather than assumed. Dimensionless treatments based on the Damköhler number, which compares the characteristic reaction rate to the characteristic transport rate, together with current-control coefficients that apportion the limitation among individual steps, have been applied directly to enzyme electrodes for this purpose (Rafat et al., 2021); the Thiele modulus and the associated effectiveness factor serve the same function for entrapped systems (Schmieg et al., 2020). Reaction–diffusion treatments of enzyme electrodes, including analyses of mono-enzymatic and micro-disk geometries (Khan et al., 2021; Kirthiga and Rajendran, 2015), provide the formal basis for this determination. Models incorporating interlayer partitioning show that mass-transport resistance and partitioning can amplify apparent signal gain (Baronas and Petrauskas, 2026), and that diffusion limitation alone can generate non-monotonic transients that the classical model would misread as anomalous kinetics (Baronas, 2025).

Three practical consequences follow, and they bear directly on how results should be reported. First, Lineweaver–Burk linearization is not merely statistically inefficient for immobilized systems but can be conceptually misleading, since the underlying rate law may no longer be adequately represented by the linearized form; this was recognized for insolubilized enzyme particles half a century ago (Gondo et al., 1973) and remains widely overlooked. Second, a Km,app. quoted without the accompanying immobilization chemistry, layer thickness, loading, and hydrodynamic conditions is difficult to interpret and cannot be reliably compared across studies. Third, comparisons of “enzyme performance” between platforms often reflect differences in transport geometry rather than intrinsic enzyme performance broader critique of the field has made a related point: immobilization outcomes are often evaluated based on a small number of experiments that are not always well designed, sustaining a literature in which reported benefits are difficult to reproduce (Bolivar et al., 2022). We return to this problem in Section 10, as it underlies the benchmarking gap we regard as the field’s most consequential unresolved issue.

3 Physicochemical factors affecting enzymatic biosensor performance

Physicochemical parameters such as pH, temperature, and cofactor availability strongly influence enzymatic biosensor performance and are closely related to the kinetic behavior described above. These factors affect catalytic efficiency, enzyme stability, and the resulting analytical response. Table 1 summarizes representative enzymes, their target analytes, and the pH, temperature, and cofactor conditions reported for common enzymatic biosensor platforms.

TABLE 1

EnzymeRepresentative applicationTarget analytepH rangeTemperature range (°C)Cofactor/prosthetic groupRef.
Glucose oxidase (GOx)Glucose monitoring; food analysis; cellular monitoringGlucose5.5–6.530–45FAD(Bollella and Katz, 2020)
Lactate oxidase (LOx)Clinical diagnostics; sports medicine; beverage analysisLactate6.5–7.525–40FAD(Sainz et al., 2022)
Horseradish peroxidase (HRP)Colorimetric and electrochemical sensingH2O26.0–7.020–45Heme (Fe3+)(Wang H.-R. et al., 2024)
UreaseRenal diagnostics; wastewater monitoringUrea6.5–8.030–60Ni2+(Li et al., 2022)
Alcohol dehydrogenase (ADH)Ethanol detection; fermentation monitoringEthanol7.0–9.030–45NAD+/NADH(Zan et al., 2023)
Xanthine oxidase (XOD)Gout diagnostics; purine metabolismXanthine7.0–8.530–50Mo-cofactor; FAD; Fe–S clusters(Sun et al., 2024)
Cholesterol oxidase (ChOx)Cholesterol monitoring in serum and foodCholesterol6.0–7.035–45FAD(Jayanthi and S.K, 2020)

Representative enzymes, target analytes, and operating conditions reported for enzymatic biosensors.

Temperature influences enzymatic reaction rates by affecting molecular motion and enzyme structural stability. Moderate temperature increases can enhance catalytic rates, whereas excessive heating can disrupt tertiary and quaternary structures and cause activity loss. For commonly used enzymes such as glucose oxidase (GOx), optimal activity is typically observed between 35 °C and 40 °C. Temperatures above 50 °C may promote structural instability, including changes around the flavin adenine dinucleotide (FAD)-containing active site (Sakalauskiene et al., 2022; Sainz et al., 2022; Liu S. et al., 2025; Wang et al., 2022).

To mitigate this sensitivity, researchers have employed several strategies, including enzyme immobilization on thermally stable supports, nanostructured materials such as graphene and silica nanoparticles, and advances in protein engineering. Enzymes derived from thermophilic microorganisms have received particular attention for high-temperature settings (Brytan et al., 2024).

These strategies are not without cost, and the trade-off is worth stating plainly because it is often omitted. Stabilization generally works by restricting conformational freedom, and the same restriction that resists thermal unfolding also constrains the conformational motions that catalysis requires. The literature on immobilization for industrial biocatalysis is explicit that modifications introduced to improve robustness are commonly associated with decreased activity or detrimental changes in catalytic properties (Bernal et al., 2018). Engineering campaigns that succeed on one axis may therefore lose ground on another: rational interface engineering of cellobiose dehydrogenase shifted its pH optimum from 4.5 to 7.0, appropriate for physiological measurement, and increased interdomain electron transfer at pH 7.5 more than twelve-fold, yet the accumulated positive charge responsible for that gain simultaneously decreased direct electron transfer to the electrode (Reichhart et al., 2023). Reports of enhanced stability should accordingly be read alongside the activity data, and, where a review such as this one summarizes them, both should be reported together.

4 Role of cofactors and electron transfer mediation

Beyond pH and temperature, many enzymatic biosensors depend on cofactors to support catalytic turnover and electron transfer. These cofactors include metal ions such as Mg2+, Zn2+, Fe2+/3+, and Ni2+, as well as organic coenzymes such as FAD and nicotinamide adenine dinucleotide in its oxidized and reduced forms (NAD+/NADH) (Doshi et al., 2025; Nielsen et al., 2023).

Cofactor stability and regeneration are important for maintaining sensor reliability. GOx uses FAD as a redox-active cofactor, whereas dehydrogenases typically depend on NAD+/NADH cycling. Because NAD+/NADH is consumed during catalysis, continuous operation often requires mediator systems or enzymatic cascades to regenerate the cofactor (Zayats et al., 2000). In complex matrices, changes in pH, metal ion concentrations, or competing ligands can further affect catalytic activity and reproducibility (Ye et al., 2025).

Recent approaches to cofactor engineering, including covalent attachment, artificial cofactor design, and nanostructured carrier systems, have improved cofactor stability, reduced leaching, and increased the potential for long-term reuse (Reus et al., 2024).

An important difference between these two cofactor systems is their mode of association with the enzyme. FAD is tightly, and often covalently, associated with its apoenzyme and is generally retained during operation. In contrast, NAD+/NADH is a diffusible cosubstrate that must be continuously supplied or regenerated in situ. This difference helps explain why FAD-dependent oxidases are more widely used in practical enzymatic sensing. In contrast, dehydrogenase-based devices, despite their broader substrate scope, remain more commonly studied at the laboratory scale. Two approaches are helping to address this limitation: hyperthermophilic dehydrogenases fused to cytochrome domains, which can convert an otherwise mediator-dependent enzyme into a direct electron-transfer system while retaining more than 80% of the initial current response after 2 months of storage at 4 °C (Maeno et al., 2025); and surface mediator modification, which has been used to convert NAD(P)+− dependent dehydrogenases into quasi-direct-transfer enzymes, with reported detection limits of 0.044–0.15 mM across engineered variants in buffer and artificial interstitial fluid (Ikegai et al., 2024).

5 Signal amplification strategies in enzymatic biosensors

Optimizing kinetics and controlling environmental conditions are necessary but not sufficient to achieve ultra-low detection limits; additional signal enhancement is often required. Enzymatic amplification strategies exploit enzyme catalytic turnover to convert minimal analyte concentrations into measurable signals. Unlike direct detection systems, in which the signal generally increases proportionally with analyte concentration, enzymatic amplification allows a single enzyme molecule to catalyze multiple reaction cycles, thereby enhancing the analytical response (Miao et al., 2015; Wang Z. et al., 2025). An illustrative case is substrate turnover-based amplification, where electroactive, fluorescent, or chromogenic reaction products accumulate over time.

One commonly employed strategy is the cascade enzyme system, in which the product of one enzymatic reaction serves as the substrate for a second enzyme. For example, glucose oxidase (GOx) converts glucose into hydrogen peroxide (H2O2), which horseradish peroxidase (HRP) then uses to oxidize a secondary substrate, enhancing signal generation (Sakalauskiene et al., 2022; Wang T. et al., 2024). Such coupled reactions improve both sensitivity and specificity. Co-encapsulation of the two enzymes within a hierarchically porous zeolitic imidazolate framework retained approximately 90% of initial activity at 60 °C and showed a 14.1% activity enhancement in 10% (v/v) methanol, outperforming the free enzymes under the same conditions (Gao et al., 2026); a related GOx–HRP system encapsulated in hydrophilic ZIF-90 reached a detection limit of 0.034 μg mL-1 for glucose (Chen H. et al., 2025). Confinement in these systems both organizes the cascade spatially and protects the enzymes from proteolytic and thermal degradation.

Cofactor recycling provides another amplification pathway, particularly in dehydrogenase-based systems, where continuous NAD+/NADH regeneration sustains turnover beyond individual reaction events (Schachinger et al., 2021).

Nano-enabled amplification strategies incorporate enzyme nanoconjugates, multi-enzyme assemblies, and DNA–enzyme hybrids that spatially organize catalytic components and reduce diffusion limitations. Nanocarriers such as liposomes and metal–organic frameworks (MOFs) can encapsulate enzymes or cofactors, allowing controlled release and localized signal enhancement (Liu et al., 2018). Nanozyme-assisted architectures follow the same logic with an artificial catalyst: carbon microfibers modified with hemin and gold nanoparticles, coupled to alcohol oxidase or glucose oxidase, delivered a 1.3- to 2.6-fold sensitivity increase relative to the nearest carbon-derived analogs, and were validated for ethanol and glucose in grape must and wine against enzymatic reference kits (Smutok et al., 2021). Figure 2 summarizes the main enzymatic amplification mechanisms discussed above, illustrating substrate turnover through enzyme cascades, cofactor recycling, and nanostructure-assisted signal enhancement.

FIGURE 2

6 Immobilization engineering strategies for performance optimization

Enzyme immobilization plays a crucial role in biosensor performance, significantly affecting catalytic activity, stability, reproducibility, and operational lifespan. The choice of strategy directly impacts enzyme orientation, substrate accessibility, and electron-transfer efficiency.

6.1 Conventional immobilization approaches

Traditional immobilization methods fall into two main categories: physical entrapment and covalent attachment.

Physical entrapment involves embedding enzymes within porous matrices such as sol–gel silica, hydrogels, chitosan, alginate, or polyacrylamide (Küchler et al., 2016; Ganonyan et al., 2020; Verma et al., 2020; Deng et al., 2019; Temoçin et al., 2018). This method helps maintain the native structure of enzymes but can result in diffusion limitations and enzyme leakage, affecting long-term stability and reproducibility (Bié et al., 2022). As noted in Section 2.2, entrapment is also the approach most likely to place a device firmly in the diffusion-controlled regime.

Covalent immobilization forms stable chemical bonds between enzyme functional groups and activated surfaces, using glutaraldehyde, EDC/NHS coupling, or epoxy-silane chemistry. It significantly reduces enzyme leaching and enhances mechanical stability, especially under continuous-flow or harsh conditions (Thangaraj and Solomon, 2019). However, if enzyme orientation is not controlled, rigid attachment may compromise catalytic efficiency. Spacer molecules and site-directed strategies often address steric hindrance (Guisan et al., 2022).

6.2 Oriented versus random immobilization

Enzyme orientation has emerged as a factor influencing performance beyond attachment chemistry alone. Random immobilization produces a heterogeneous enzyme arrangement that can block active sites and hinder electron transfer. Oriented strategies, including affinity tags, site-directed mutagenesis, ligand-mediated binding, and bioorthogonal chemistry, enable precise spatial arrangement of the catalytic site (Hou et al., 2014; Daronch et al., 2020; Martinez and Alvarez, 2018; Ukirade et al., 2025; Baskin and Bertozzi, 2007; Hernandez and Fernandez-Lafuente, 2011).

Where it succeeds, the effect is substantial and measurable. A glucose dehydrogenase fused to a minimal cytochrome and wired site-specifically through an unnatural amino acid showed more than fifteen times higher catalytic current and ten-fold higher sensitivity than the same enzyme wired non-specifically (Algov et al., 2021). An engineered binding peptide directing the orientation of the multicopper oxidase CueO produced a 1.38-fold increase in current density and a 21.2-fold improvement in kinetic current density relative to the wild-type enzyme (Zhang et al., 2025). Improved substrate accessibility and direct electron transfer (DET) matter most in high-sensitivity applications such as clinical diagnostics (Shimomura et al., 2008). Table 2 summarizes the comparison.

TABLE 2

ParameterRandom immobilizationOriented immobilizationRefs
Activity retentionModerate to low; heterogeneous enzyme orientation may partially block active sitesGenerally higher when the attachment site preserves active-site accessibility; site-directed immobilization does not inherently guarantee activity retention(Hou et al., 2014; Daronch et al., 2020; Martinez and Alvarez, 2018; Ukirade et al., 2025; Baskin and Bertozzi, 2007; Hernandez and Fernandez-Lafuente, 2011; Gao et al., 2021)
ReproducibilityVariable between preparations due to heterogeneous enzyme orientationGenerally higher because enzyme orientation and attachment sites are more controlled(Hernandez and Fernandez-Lafuente, 2011; Shimomura et al., 2008; Algov et al., 2021)
ImplementationRelatively simple; compatible with conventional coupling chemistriesMore complex; may require affinity tags, site-specific chemistry, or protein engineering(Zhao et al., 2025; Zhang et al., 2025; Patti et al., 2025)
Cost and scalabilityLow cost; readily compatible with large-scale fabricationModerate to high; may require purified or engineered enzymes and additional processing steps(Zhao et al., 2025; Bernal et al., 2018; Patti et al., 2025)
Direct electron transfer (DET)Usually low to moderate, depending on enzyme orientation and distance from the electrodePotentially higher due to improved orientation of redox centers toward the electrode(Algov et al., 2021; Zhang et al., 2025)
Main limitationActive-site occlusion, heterogeneous orientation, and batch-to-batch variabilityLoss of activity or stability caused by chemical modification or site-specific engineering(Gao et al., 2021; Mišković et al., 2024; Ngo et al., 2022; Bernal et al., 2018)

Comparison between oriented and random immobilization. Entries reflect the studies discussed in Sections 6.1 and 6.2 and are referenced accordingly.

6.2.1 Limitations of oriented immobilization

The advantages above are real but conditional, and the conditions are demanding enough that oriented immobilization is not a general-purpose replacement for random attachment. Four constraints warrant explicit statement.

First, oriented strategies are enzyme-specific rather than generic. Each target typically requires its own affinity tag, engineered cysteine, or bioorthogonal handle, positioned based on knowledge of the enzyme’s structure and which surface regions can be modified without disturbing catalysis. The engineered-peptide result cited above required semi-rational design of a thirteen-residue peptide followed by site-saturation mutagenesis with electrochemical screening, from which only three usable variants emerged (Zhang et al., 2025); that is a full protein-engineering campaign expended on a single enzyme. Reviews of immobilization for flow biocatalysis continue to describe genetic fusion technologies as promising but not yet mature, and call explicitly for work to address their current limitations before industrial deployment (Patti et al., 2025).

Second, the enabling modification is not activity-neutral. Adding a His-tag, the most routine step in oriented immobilization, can be sufficient to compromise an enzyme. The effect depends on position in ways structural methods may not reveal; for adenylosuccinate synthetase, a C-terminal tag was essentially benign, whereas the N-terminal construct aggregated strongly and showed reduced activity, with circular dichroism and X-ray diffraction detecting no structural change to explain it (Mišković et al., 2024). Genetic fusion carries the same risk at larger scale, with a twenty-fold activity decrease reported for an azoreductase depending on its position within the fusion protein (Ngo et al., 2022). The underlying difficulty, as noted in Section 3, is that enzyme engineering and immobilization are still typically pursued as separate workflows, so the consequences of one for the other are rarely characterized in advance (Bernal et al., 2018).

Third, site-directed immobilization does not necessarily lead to better performance than random attachment; the outcome depends strongly on where the enzyme is attached. In a systematic comparison of attachment chemistries conducted with an antibody rather than an enzyme, but still relevant to immobilization strategies, selective conjugation through the N-terminus retained only 10% functionality, compared with 50% for a randomly oriented control. In contrast, selective attachment through a different surface region preserved more than 75% (Gao et al., 2021). This example highlights that controlling orientation is not sufficient on its own, as an unsuitable attachment site can be more detrimental than random immobilization. Thus, the benefits of oriented immobilization should be considered in relation to the specific attachment site and enzyme rather than treated as a general advantage of orientation control.

Fourth, cost and scalability remain important barriers. Site-specific approaches typically require purified enzymes rather than crude preparations, while the protein tags used in some strategies can introduce additional steric constraints (Zhao et al., 2025). We found no published quantitative comparisons of cost or production throughput, so this limitation is discussed qualitatively rather than assigned a numerical estimate. In practice, oriented immobilization remains more common in high-value clinical and research applications, whereas low-cost, disposable, and point-of-care formats still rely mainly on random attachment. Recent guidance supports this view, emphasizing that activity losses caused by conformational changes and mass-transfer limitations remain common across immobilization methods and that protocols should be adapted to the specific application rather than selected according to a general ranking (Robescu and Bavaro, 2025).

6.3 Nanoengineered immobilization platforms

Nanomaterials enhance enzyme loading, control the local microenvironment, and facilitate interfacial electron transfer. Nanocarriers including mesoporous silica nanoparticles, MOFs, carbon nanotubes (CNTs), graphene oxide (GO), and polymeric nanogels provide high surface-to-volume ratios and customizable chemistries (Seenivasan et al., 2017; Shortall et al., 2022; Sales et al., 2023).

Carbon-based nanomaterials such as CNTs (Kumar et al., 2019) and GO (Ukirade et al., 2025) improve electrical conductivity and enable DET between the enzyme’s redox centre and the transducer surface. The effect on the enzyme itself is not uniform, however, and the distinction is easily lost in general statements about conductivity. In a structure–function comparison of glucose oxidase in the presence of two carbon nanomaterials, multi-walled carbon nanotubes induced conformational change and reduced enzyme activity by 7%, whereas bio-graphene preserved enzyme function; the bio-graphene biosensors correspondingly showed 1.33-fold higher sensitivity and improved reproducibility relative to the MWCNT-based devices (Alatzoglou et al., 2024). Nanostructure geometry matters as much as composition: wet-spun, glucose-reactive enzyme-functionalized semiconducting single-walled carbon nanotube fibers were configured into fiber-type sensors exhibiting field effects (Lee et al., 2020), illustrating how the same material class can be routed toward either amperometric or transistor-based transduction.

MOFs possess adjustable pore structures that can encapsulate enzymes and protect them from denaturation. Glucose oxidase immobilized on ZIF-7-III retained 98% relative catalytic activity at maximum enzyme loading, remained stable across a broad pH range and up to 80 °C, and preserved 66% of activity after 60 days of storage, with repeatability below 6% RSD (Aghayani et al., 2025). Co-immobilizing laccase with its mediator in Fe-doped ZIF-8 improved catalytic efficiency 2.6-fold over free laccase while enhancing thermal stability and reusability (Li et al., 2024), showing that the framework can house a complete catalytic system rather than the enzyme alone. The principal limitation of MOF supports is electrical: their intrinsically poor conductivity has restricted bioelectrochemical use, motivating two-dimensional metal–organic layers with ferrocene modification as a more conductive alternative, reported with a glucose detection limit of 3.9 μM at S/N = 3 (Dong et al., 2025).

These platforms can reduce diffusion limitations and improve enzyme stability while also facilitating miniaturization, making immobilization an important part of the overall sensor design rather than simply an attachment step. However, these benefits come with greater fabrication complexity, and reproducibility of nanostructured interfaces across batches remains less well documented than reported peak performance.

7 Signal transduction architectures in enzymatic biosensors

Signal transduction converts enzymatic catalytic events into measurable outputs. The choice of transduction strategy strongly influences sensitivity, response time, integration capacity, and applicability in biological contexts (Bharti et al., 2020; Senthilkumar et al., 2008).

7.1 Electrochemical platforms

Electrochemical biosensors excel in enzymatic sensing due to their sensitivity, scalability, and compatibility with portable electronics. They convert enzymatic redox reactions into electrical signals and are classified as amperometric, potentiometric, or impedimetric (Artigues et al., 2017).

Amperometric sensors measure the current generated by oxidation–reduction reactions at fixed potentials, and are commonly used with oxidoreductases such as glucose oxidase and lactate oxidase. First-generation sensors relied on oxygen as an electron acceptor; second-generation devices use redox mediators such as ferrocene derivatives, quinones, Prussian Blue, and osmium polymers to improve signal stability and reduce operating potential (Monkrathok et al., 2024).

Mediator-free systems leveraging DET eliminate soluble mediators, improving biocompatibility and simplifying design (Ghindilis et al., 1997). DET remains challenging because the redox centers of many enzymes are buried within the protein matrix, and only a limited number of oxidoreductases achieve it natively (Sowa et al., 2024). Nanostructured electrodes incorporating carbon nanotubes, graphene, and gold nanoparticles improve electronic coupling and tunneling efficiency; laser-induced graphene on a paper substrate, for example, has been used to support fructose dehydrogenase in a third-generation configuration for the determination of inulin, with reported detection limits of 0.3 mg L-1 in urine and 1.0 mg L-1 in serum at a relative standard deviation (RSD) ≤ 2.0% (Paolini et al., 2025). Laccase immobilized on gold nanoparticles deposited on screen-printed carbon electrodes provides a further amperometric example, reaching a detection limit of 0.83 µM as caffeic acid equivalents with a 1–250 µM linear range for polyphenols in ethanolic extracts of propolis, and reducing analysis time from 85 to 15 min relative to the Folin–Ciocalteu reference method (Mohtar et al., 2019).

Potentiometric biosensors detect changes in ion activity near the electrode surface under open-circuit conditions, making them well-suited to enzymatic reactions that produce ionic species. Impedimetric systems monitor changes in interfacial impedance, making them effective for label-free detection and surface inhibition assays (Abrar et al., 2016).

7.2 Optical platforms

Optical biosensors convert enzymatic reactions into measurable optical changes such as fluorescence, absorbance, or refractive index shifts. Fluorescence-based systems offer high sensitivity and real-time monitoring using fluorogenic substrates or nanomaterial-enhanced amplification (Qin et al., 2022). Surface plasmon resonance (SPR) biosensors detect refractive index changes at metal interfaces, enabling label-free monitoring of enzymatic interactions, and integration into fiber optics and wearable photonic platforms is expanding their use in non-invasive diagnostics. Colorimetric methods remain attractive for low-cost point-of-care applications owing to their simplicity and visual readout (Germain and Knapp, 2009; Wang et al., 2018). As Section 8 describes fluorometric detection of NAD(P)H autofluorescence has proven the dominant readout for gas-phase enzymatic sensing.

7.3 Semiconductor and hybrid platforms

Piezoelectric sensors measure mass changes on quartz surfaces, enabling label-free detection of enzymatic inhibition or product formation. At the same time, thermal biosensors detect heat variations from enzymatic reactions, providing universal but less selective readouts (Skládal, 2024).

Enzymatic field-effect transistors (ENFETs) combine enzymatic specificity with semiconductor amplification. By coupling enzymatic reactions to gate-potential modulation, ENFETs enable rapid, sensitive detection, and integrating graphene, MoS2, and flexible substrates positions them for wearable and implantable applications (Zhang et al., 2019; Wang Q. et al., 2025). Recent solution-gated graphene devices illustrate the amplification available from this architecture: a photoelectrochemical solution-gated graphene FET functionalized with an acetylcholinesterase–nanozyme cascade reported a thousand-fold current gain and a detection limit of 0.05 pM for organophosphates (Wang H.-R. et al., 2024), with a companion quantum-dot configuration reaching 0.21 pM across a 0.1 nM to 1 mM linear range (Wang H. et al., 2025). Transition metal dichalcogenides including MoS2 and WS2 continue to be developed as FET channel materials for biosensing (M et al., 2025). The principal unresolved issues for this class are drift and reference stability during extended operation, which bear directly on the continuous-monitoring applications for which these devices are proposed.

7.4 Wearable enzymatic platforms

Wearable formats have become one of the most active translation routes for enzymatic biosensing, since enzymes supply the selectivity required to quantify specific metabolites in complex biofluids without sample preparation. Sweat is the most accessible target. A fully inkjet-printed, integrated multiplexed patch has been reported for simultaneous glucose and alcohol determination in perspiration with minimal drift over 30 h of operation (Ma et al., 2024), and a single platform based on platinum nanoparticles on nitrogen-doped mesoporous carbon and reduced graphene oxide has been used to support three different immobilized enzymes for glucose, uric acid, and lactate, validated during exercise and dietary challenge against blood measurements (Chen M. et al., 2025).

Enzyme distribution within the wearable matrix, rather than enzyme identity, is often what limits performance. Magnetically regulated orientation of glucose oxidase within a porous hydrogel, adopted specifically to prevent the aggregation and leaching that distort sweat glucose signals, achieved a detection limit of 0.25 μmol L-1 with stable performance over 30 days (Chu et al., 2025), which is among the more informative operational-lifetime figures available in this area and illustrates the value of reporting stability as retained performance over a stated interval. For interstitial fluid, microneedle formats bring enzymatic sensing closer to continuous glucose monitoring: glucose oxidase interfaced with o-phenylenediamine on dendritic platinum microneedles has been reported with a sensitivity of 137.7 nA (mg dL-1)−1, linearity across 82–211 mg dL-1, and interferent-to-glucose current ratios below 4.0% in artificial interstitial fluid (Kim et al., 2026). Nanomaterial support in these devices serves a dual role: increasing enzyme loading and facilitating DET (Jamshidnejad-Tosaramandani et al., 2025).

Three constraints recur across this literature and, in our view, are more limiting than sensitivity. Sweat metabolite concentrations correlate imperfectly with blood values and depend on sweat rate; on-body operation exposes the enzyme to mechanical deformation and to a variable, often dehydrating microenvironment; and calibration drift over multi-day wear remains inadequately characterized, since operational lifetimes are frequently reported in hours of continuous measurement rather than days of realistic use.

7.5 Emerging in-memory and in-sensor computing architectures

Beyond conventional transduction, a growing body of work integrates sensing and computation within a single nanoscale element, an approach termed in-memory or in-sensor computing. Memristive and resistive random-access memory (RRAM) devices can store and process signals at the same physical location at which the enzymatic event is transduced, reducing the need for separate analog-to-digital conversion and external signal processing. A complementary metal–oxide–semiconductor (CMOS)-compatible memristor-based electrochemical biosensor transducer has been demonstrated that performs threshold sensing directly within the device rather than in downstream signal-processing modules, lowering power consumption and system complexity for point-of-care use (Kim et al., 2025). Nanomaterial-engineered memristive crossbar architectures have been reviewed for biosensing and neural-prosthetic applications, with device-to-device variability and signal integrity identified as the principal remaining obstacles (Bouzouita et al., 2025). At the algorithmic level, researchers have implemented analog processor-in-memory convolutional neural networks on-chip to classify biosensor signals at 19 mW during multiply-accumulate operations, corresponding to 5.38 TOPS/W (Byun et al., 2022).

For enzymatic biosensors, these architectures matter less for raw computational throughput than for what they enable at the sensing interface. Real-time drift compensation and adaptive recalibration performed locally would address the enzyme stability and reproducibility limitations discussed in Sections 2–6 without requiring the enzyme itself to be more stable, which is precisely the constraint that has proved most resistant to materials-based solutions. This remains an aspiration rather than a demonstrated capability for enzymatic devices, and the variability of memristive elements themselves must be resolved before it can be realized; we identify it in Section 10 as an emerging priority rather than an established solution.

8 Gas-phase enzymatic biosensing of volatile organic compounds: a case study

Having established the kinetic, immobilization, and transduction principles that govern enzymatic biosensor performance (Sections 2–7), we now examine how those principles behave under a particularly demanding set of constraints. Gas-phase detection of volatile organic compounds (VOCs) was selected as a focal case study because it requires the simultaneous optimization of every parameter discussed above: the enzyme must remain hydrated and catalytically competent outside bulk solution, the analyte must partition efficiently from air into the sensing phase, transport limitations are imposed by a membrane rather than by a solution boundary layer, and the transduction chain must resolve analyte concentrations in the parts-per-billion range. It therefore functions as a stress test for the principles developed in Sections 2–7 and synthesized into a design framework in Section 9, and its failures are as informative as its successes.

8.1 Scope and maturity of the field

An accurate account of this subfield must begin with its size. In our search (Section 1.1), queries combining gas-phase biosensing terms with enzymatic recognition returned on the order of tens of records in total, against a liquid-phase enzymatic biosensor literature numbering in the tens of thousands. The great majority of true gas-phase enzymatic devices originate from a single research group, and nearly all share one architecture: an NAD(P)-dependent dehydrogenase held in a hydrated polymer on a porous hydrophilic membrane, with fluorometric detection of NADH autofluorescence through an optical fiber (Chien et al., 2020; Toma et al., 2021; Iitani et al., 2023). Reported analytes are correspondingly narrow. The large majority are short-chain alcohols, aldehydes, and ketones, including ethanol, methanol, isopropanol, acetone, formaldehyde, and 2-nonenal, because these are the substrates that alcohol oxidases and NAD(P)-dependent dehydrogenases accept. The exceptions are instructive rather than numerous: a tyrosinase-based device extends the approach to a volatile phenol (Portugal-Gómez et al., 2025), and a cholinesterase-inhibition device addresses a semi-volatile organophosphate (Edwards et al., 2019), both using enzyme classes outside the dehydrogenase family. We identified no verified enzymatic gas-phase biosensor for benzene, toluene, ethylbenzene, or xylenes (BTEX aromatics). This is important because it indicates that current primary literature does not strongly support claims of broad enzymatic VOC-sensing capability, implying that the field’s apparent methodological uniformity reflects the state of the art rather than selective citation.

A second observation follows from the devices themselves. In every verified example, the enzyme operates in a hydrated phase: a poly (2-methacryloyloxyethyl phosphorylcholine-co-2-ethylhexyl methacrylate) (PMEH) film on porous polytetrafluoroethylene (PTFE) (Chien et al., 2020; Iitani et al., 2023), a deep eutectic solvent absorbed into filter paper (Portugal-Gómez et al., 2025), a poly (vinyl alcohol) (PVA) gel electrolyte (Sun et al., 2021), or an electrolyte-filled chamber behind a gas-permeable film (Shitanda et al., 2026). Gas-phase enzymatic biosensing is therefore more accurately described as vapor partitioning into a thin condensed phase followed by conventional aqueous biocatalysis, rather than as biocatalysis in the dry state. Genuinely dry-form devices, such as enzyme-and-cofactor-embedded electrospun fiber sensors, are presented as novelties for exactly this reason (Iitani et al., 2022). This reframing matters in practice: the rate-limiting step is often partitioning and diffusion across the membrane rather than turnover, placing these devices firmly in the diffusion-controlled regime described in Section 2.2.

8.2 Material platforms

Within these constraints, material selection is the designer’s principal lever, since it governs both analyte preconcentration and the stability of the enzyme microenvironment; sustainability considerations in material choice have also begun to receive attention (Cova et al., 2022).

8.2.1 Metal nanoparticles

Noble metal nanoparticles improve electron-transfer kinetics and permit high-density enzyme loading. Gold nanoparticles in particular facilitate DET and offer versatile surface chemistries for both covalent and electrostatic immobilization; combined with conductive polymers, they produce synergistic increases in catalytic current and reduced response times (Machado et al., 2025).

8.2.2 Carbon-based nanostructures

Carbon nanotubes and graphene-based materials create conductive networks that facilitate charge transport and electronic coupling to enzyme active sites. Their surface area supports stable immobilization of oxidases and dehydrogenases relevant to aldehyde and alcohol detection, and π–π interactions and linker-assisted anchoring further enhance enzyme stability and orientation (Kim et al., 2024).

8.2.3 Conducting polymers and molecularly imprinted polymers

Conducting polymers and molecularly imprinted polymers (MIPs) add selectivity. Beyond serving as immobilization matrices, they preconcentrate target VOCs near the catalytic site, enhancing sensitivity at low concentrations; MIP-based architectures also discriminate structurally similar volatiles in complex settings (Macagnano et al., 2025). Electrospun polystyrene–poly (amidoamine) (PAMAM) dendritic polymer nanofibers grafted with alcohol oxidase illustrate the polymer-nanofibre route, with a reported conductometric sensitivity of 150.53 μS cm-1 (v/v) and a gas-phase methanol detection limit of 100 ppm (Razmshoar et al., 2023).

8.2.4 Hybrid nanocomposites

Hybrid systems combining metals, carbon nanostructures, and polymeric matrices generally outperform single materials, offering improved catalytic activity, conductivity, and selective adsorption within hierarchical structures. These multifunctional interfaces help maintain enzymatic activity in gas-phase operation while enabling rapid amperometric or impedimetric measurement.

8.3 Reported performance

Table 3 summarizes verified gas-phase enzymatic VOC biosensors. Two conventions are applied deliberately. First, figures of merit are reported as the primary sources report them, distinguishing limits of detection from working or dynamic ranges; conflating the two is a common source of error in summary tables, since several widely cited devices report only a validated range. Second, the measurement phase is stated explicitly for every entry, since a device may be exposed to a vapor while expressing its figure of merit in liquid-sample units.

TABLE 3

Enzyme/sensing platformImmobilization strategyTarget analyte/applicationTransduction/readoutAnalytical performanceRef.
Secondary alcohol dehydrogenase/hydrophilic PTFE membraneEntrapment in PMEH polymerIsopropanol; exhaled breathBiofluorometric (NADH)LOD: 0.5 ppb(Chien et al., 2020)
Alcohol oxidase + formaldehyde dehydrogenase/hydrophilic PTFE membraneCross-linking to polyethyleneimine using glutaraldehydeMethanol; exhaled breathBiofluorometric (NADH)Dynamic range: 0.32–20 ppm; LOD not reported(Toma et al., 2021)
Aldehyde dehydrogenase/hydrophilic PTFE membraneEntrapment in PMEH polymertrans-2-Nonenal; body odorBiofluorometric (NAD(P)H)LOD: 0.23 ppm; LOQ: 0.26 ppm(Iitani et al., 2023)
Secondary alcohol dehydrogenase/MgO-templated carbon–porous polyimideCovalent epoxy–amine couplingAcetoneMediated amperometricLOD: 75 ppb; LOQ: 250 ppb(Shitanda et al., 2026)
Formaldehyde dehydrogenase/poly (methylene green)–buckypaperNot specifiedFormaldehyde; indoor airSelf-powered electrochromic, naked-eye readoutDynamic range: 80–3,000 ppb; LOD not reported(Sun et al., 2021)
Alcohol oxidase/polystyrene–PAMAM nanofibersGrafting onto dendrimer nanofibersMethanolConductometricLOD: 100 ppm(Razmshoar et al., 2023)
Tyrosinase/choline chloride–glycerol DES/screen-printed carbonDrop-casting4-Ethylguaiacol; wine headspaceVoltammetric/amperometricCCβ: 175 μg L-1; reported as liquid-sample equivalent(Portugal-Gómez et al., 2025)
Butyrylcholinesterase + HRP/gold screen-printed electrodeEncapsulation in peptide nanotubesMalathion vapor; semi-volatileInhibition-based cyclic voltammetryMeasurement range: 6–25 ppbv(Edwards et al., 2019)
Alcohol dehydrogenase/cotton mesh–CMOS headsetGlutaraldehyde cross-linkingEthanol; transcutaneous/ear canalBiofluorometric (NADH)Dynamic range: 10 ppb–10 ppm(Zhang et al., 2026)

Verified enzymatic biosensors for gas-phase detection of volatile organic compounds.

LOD, limit of detection; LOQ, limit of quantification; DR, dynamic or working range as reported by the authors; CCβ, detection capability; ppbv, parts per billion by volume; PMEH, poly (2-methacryloyloxyethyl phosphorylcholine-co-2-ethylhexyl methacrylate); PTFE, polytetrafluoroethylene. All entries are gas-phase measurements unless otherwise noted.

The dominant architecture is evident from the table: biofluorometric detection of NAD(P)H, generated by a dehydrogenase held in a hydrated polymer film, accounts for four of the nine entries and for the largest share of the sub-ppm performance, including the most sensitive device reported (0.5 ppb for isopropanol (Chien et al., 2020)). Sensitivity is not, however, the exclusive property of that architecture. The amperometric acetone sensor reaches 75 ppb using mediated electrochemical transduction (Shitanda et al., 2026), and the inhibition-based cholinesterase device operates in the ppbv range (Edwards et al., 2019), indicating that electrochemical readout is competitive where the enzyme chemistry permits it. Conversely, the conductometric nanofibre device is more than two orders of magnitude less sensitive than the fluorometric methanol sensor addressing the same analyte (Toma et al., 2021; Razmshoar et al., 2023), illustrating how strongly the readout constrains achievable performance for a given enzyme.

8.4 What this case study reveals

Three key elements emerge for this case, each extending beyond VOC sensing. First, the enzyme’s microenvironment, rather than the transducer, governs the binding constraint. Every device listed in Table 3 focuses on maintaining a hydrated environment around the enzyme, and their analytical performance depends more closely on this effort than on readout sophistication. Second, the scope of detectable analytes is limited by enzyme availability, not transducer capability. The focus on alcohols, aldehydes, and ketones results from the substrates accepted by NAD(P)-dependent dehydrogenases and alcohol oxidases. The two departures from that pattern in Table 3 are precisely the two devices built on a different enzyme class, including a phenol oxidase (Portugal-Gómez et al., 2025) and a cholinesterase (Edwards et al., 2019). Therefore, extending the field to aromatic VOCs is primarily an enzyme-discovery and enzyme-engineering challenge, not a material problem, which is not how the challenge is usually framed. Lastly, reporting conventions require careful attention. Within this small literature, we encountered limits of detection, limits of quantification, and decision limits. We validated dynamic ranges used side by side, occasionally with liquid-referenced units for vapor-phase measurements. These various metrics are difficult to compare reliably, illustrating the broader benchmarking challenge discussed in Section 10.

9 Design guidelines for high-performance enzymatic biosensors

Developing high-performance enzymatic biosensors requires a comprehensive engineering approach that optimizes biochemical, interfacial, and electronic parameters together. Enzyme selection, immobilization, nanomaterial integration, and signal transduction should not be treated as isolated components; next-generation biosensors are designed as integrated systems in which each component contributes to overall analytical performance (Mao et al., 2024).

Enzymatic reaction kinetics fundamentally affect sensitivity, dynamic range, and response time. Catalytic turnover rates and substrate affinity must be considered to ensure measurable responses within the relevant analyte concentration range while minimizing premature saturation. Protein engineering, cofactor regeneration, and cascade amplification can enhance catalytic efficiency, provided that enzymatic activity is effectively coupled to the selected transduction mechanism (Li and Cirino, 2024; Liu X. et al., 2025). As Section 2.2 established, the parameters that matter for design are the apparent ones exhibited by the assembled device, not the intrinsic constants of the enzyme in solution.

Enzyme immobilization and orientation at the sensing interface must be controlled with equal care. Beyond attachment, active-site accessibility must be maintained, and steric hindrance minimized. Surface functionalization, spacer molecules, and site-directed coupling improve enzyme stability and, in electrochemical systems, facilitate electron transfer, subject to enzyme-specific applicability, modification-induced activity losses, and the cost constraints set out in Section 6.2.

Nanomaterial integration can further enhance performance by increasing surface area and conductivity and by stabilizing the microenvironment. Well-designed nanoarchitectures promote charge transport, improve substrate diffusion, and protect against denaturation. The added structural complexity must be balanced against reproducible fabrication and long-term operational stability.

The transduction mechanism should be selected according to enzyme chemistry and the requirements of the intended application. Electrochemical platforms offer high sensitivity and miniaturization potential; optical systems enable sensitive, multiplexed detection and currently dominate gas-phase enzymatic sensing; semiconductor-based devices provide intrinsic signal amplification; and in-sensor computing architectures offer local signal processing, although their application to enzymatic biosensors remains in the early stages. Taken together, these considerations indicate that no single biosensor architecture is optimal. The appropriate configuration depends on the operational context, enzyme properties, required analytical performance, and practical application constraints. Figure 3 summarizes this relationship as a decision framework, linking these factors to the selection of enzyme, immobilization strategy, nanomaterial interface, and transduction mechanism while making the associated trade-offs explicit.

FIGURE 3

Biosensor optimization is inherently application-specific. Clinical diagnostics, environmental monitoring, wearable sensing, and gas-phase detection impose distinct constraints on stability, selectivity, portability, and power consumption. High-performance enzymatic biosensors thus arise from the coordinated integration of kinetic control, interfacial engineering, nanomaterial design, and targeted transduction within a clearly defined operational context, moving biosensor development from empirical assembly toward rational, performance-driven engineering.

10 Challenges and future perspectives

Despite substantial progress in enzyme engineering, immobilization, nanostructured interfaces, signal amplification, and miniaturized transduction, several challenges continue to limit the broader development of enzymatic biosensors. These challenges are interconnected: improving one component can limit another, and the performance of the assembled device cannot always be predicted from the properties of its individual components. Based on the evidence discussed throughout this review, we identify five priorities for future development.

The benchmarking gap. The most consequential challenge is not a lack of new materials or immobilization strategies, but the absence of standardized, quantitative benchmarking. As discussed in Section 2.2, apparent kinetic parameters such as Km,app and Vmax,app depend on immobilization chemistry, enzyme loading, layer thickness, partitioning, and mass transport. Reporting these values without the conditions that determine them makes comparisons between platforms difficult. The same problem affects stability and analytical figures of merit, which are often reported using different definitions, matrices, and operating conditions. Future studies should therefore report apparent kinetic parameters together with the relevant architectural and hydrodynamic conditions, and should describe stability as retained performance over a defined period. Without these practices, the literature will continue to accumulate results that are individually impressive but difficult to compare or reproduce.

Stability, transport, and scalability. Enzyme stability remains a major limitation, particularly in wearable, continuous, and gas-phase applications. As shown throughout Sections 2 and 6, stabilization, catalytic activity, substrate transport, and electron transfer are closely coupled and often involve trade-offs. At the same time, the strategies that provide the highest analytical performance—such as site-specific immobilization, engineered enzymes, and complex nanoarchitectures—are often the most difficult to scale and reproduce. Future work should therefore evaluate not only activity and sensitivity but also mass-transfer limitations, batch-to-batch reproducibility, storage stability, operational lifetime, and fabrication complexity.

Validation in realistic conditions. High performance in buffered solutions does not necessarily translate to reliable sensing in biological, environmental, food, or gas-phase samples. Interferents, humidity, temperature fluctuations, mechanical stress, and long-term operation can substantially alter sensor behavior. More rigorous validation in realistic matrices and comparison with established reference methods will be essential, particularly for wearable and continuous-monitoring applications. The limited substrate scope of gas-phase enzymatic biosensing further illustrates the need for enzyme discovery and engineering, rather than relying exclusively on new sensing materials.

Toward integrated and application-specific design. The evidence reviewed here supports a shift from empirical assembly toward integrated sensor design. Optimize enzyme selection, immobilization, mass transport, nanomaterial architecture, transduction, and signal processing as a coupled system rather than independent components. Emerging strategies such as multi-enzyme cascades, engineered enzymes, hybrid bio–nano interfaces, and in-sensor computing offer promising opportunities. Still, their value will ultimately depend on whether they improve overall device performance without adding disproportionate complexity or reducing reproducibility.

The future of enzymatic biosensing will therefore depend not only on developing new enzymes and materials, but also on establishing clearer standards to demonstrate genuine improvement. Closing the benchmarking gap may be as important for advancing the field as the next-generation of sensing architectures themselves.

11 Conclusion

Enzymatic biosensors remain central to bioanalytical research because enzymatic recognition provides a level of selectivity that is difficult to achieve by other approaches. At the same time, enzymes can be readily integrated with a wide range of transduction technologies. This review shows that high-performance sensing depends on integrating kinetic control, appropriate immobilization, nanoengineered interfaces, and compatible signal transduction strategies. Importantly, this performance should be evaluated based on the behavior of the complete device rather than on the properties of individual components considered separately.

Three main conclusions emerge. First, a major barrier to progress is methodological rather than purely technical. Without consistent reporting of apparent kinetic parameters, operating conditions, and stability, it is difficult to determine which strategies provide genuine improvements, and meaningful cross-study comparisons remain limited. Second, performance trade-offs are unavoidable and should be clearly recognized. Oriented immobilization, enzyme stabilization, and increased nanostructural complexity can improve performance, but often at the expense of general applicability, catalytic activity, or experimental simplicity. Reviews that focus only on performance gains may therefore provide an incomplete view of these systems. Third, some limitations commonly attributed to materials are instead related to the underlying enzyme chemistry. In gas-phase enzymatic sensing, for example, the limited range of detectable analytes is strongly influenced by cofactor requirements and enzyme properties, suggesting that enzyme discovery and engineering may be more effective solutions than further support optimization alone.

The transition from empirical assembly to more systematic, design-oriented development is already underway, and the framework presented in Section 9 contributes to this direction. Further progress will depend on adopting reporting practices that allow meaningful comparison across studies, together with continued interdisciplinary efforts to improve enzymes, interfaces, and overall sensor architectures.

Statements

Author contributions

IA-D: Conceptualization, Writing – original draft. GQ-V: Writing – original draft. EC-G: Writing – review and editing. JM-B: Writing – review and editing, Investigation. JV-C: Writing – review and editing, Writing – original draft, Conceptualization. MR-D: Writing – review and editing, Investigation.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. GQ-V was supported by a postdoctoral fellowship from the Secretaría de Ciencia, Humanidades, Tecnología e Innovación (SECIHTI), México, under the programme Estancias Posdoctorales por México (grant number 740156). No additional external funding was received for this work or its publication.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author JV-C declared that they were an editorial board member of Frontiers at the time of submission. This had no impact on the peer review process and the final decision.

Generative AI statement

The author(s) declared that generative AI was used in the creation of this manuscript. During the preparation of this manuscript, the authors used a large language model assistant (Claude, Anthropic; model claude-opus-5, accessed 2026) to assist with drafting and language editing of the revised text, with literature searching, and with cross-checking the correspondence between cited sources and the claims attributed to them. All literature retrieved in this way was independently verified by the authors against the primary sources, and all interpretation, analysis and scientific conclusions are the authors’ own. No content was generated without author review. The authors take full responsibility for the accuracy and integrity of the entire manuscript, including all data, citations and figures.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1

    AbrarM. A.DongY.LeeP. K.KimW. S. (2016). Bendable electro-chemical lactate sensor printed with silver nano-particles. Sci. Rep.6, 1–9. 10.1038/srep30565

  • 2

    AghayaniS.TangestaninejadS.StockN.AchenbachB.BahadoriM.MoghadamM.et al (2025). Glucose oxidase immobilized on ZIF-7-III: composite formation, optimization and integration in an electrochemical biosensor for selective glucose detection. J. Mater. Chem. B13, 11739–11749. 10.1039/d5tb01119a

  • 3

    AlatzoglouC.TzianniE. I.PatilaM.TrachiotiM. G.ProdromidisM. I.StamatisH. (2024). Structure-function studies of glucose oxidase in the presence of carbon nanotubes and bio-graphene for the development of electrochemical glucose biosensors. Nanomaterials14, 85. 10.3390/nano14010085

  • 4

    AlgovI.FeiertagA.AlfontaL. (2021). Site-specifically wired and oriented glucose dehydrogenase fused to a minimal cytochrome with high glucose sensing sensitivity. Biosens. Bioelectron.180, 113117. 10.1016/j.bios.2021.113117

  • 5

    AnjumT.HussainN.BilalM.IqbalH. M. N. (2022). Enzymatic biosensors, biocatal. Immobil. Found. Appl.32, 341–363. 10.1016/B978-0-323-91317-1.00006-2

  • 6

    ArtiguesM.AbellàJ.ColominasS. (2017). Analytical parameters of an amperometric glucose biosensor for fast analysis in food samples. Sensors Switz.17, 2620. 10.3390/s17112620

  • 7

    BakarB.AkbulutM.UlusalF.UluA.ÖzdemirN.AteşB. (2024). Horseradish peroxidase immobilized onto mesoporous magnetic hybrid nanoflowers for enzymatic decolorization of textile dyes: a highly robust bioreactor and boosted enzyme stability. ACS Omega9, 24558–24573. 10.1021/acsomega.4c00703

  • 8

    BaronasR. (2025). Non-monotonic effect of substrate inhibition in conjunction with diffusion limitation on the response of amperometric biosensors. Biosensors15, 441. 10.3390/bios15070441

  • 9

    BaronasR.PetrauskasK. (2026). Effects of diffusion limitations and partitioning on signal amplification and sensitivity in bienzyme electrochemical biosensors employing cyclic product conversion. Appl. Sci.16, 1171. 10.3390/app16031171

  • 10

    BaskinJ. M.BertozziC. R. (2007). Bioorthogonal click chemistry: covalent labeling in living systems, QSAR comb. Sci26, 1211–1219. 10.1002/qsar.200740086

  • 11

    Benítez-MateosA. I.HuberC.NidetzkyB.BolivarJ. M.López-GallegoF. (2020). Design of the enzyme–carrier interface to overcome the O2 and NADH mass transfer limitations of an immobilized flavin oxidase. ACS Appl. Mater. Interfaces.12, 56027–56038. 10.1021/acsami.0c17568

  • 12

    BernalC.RodríguezK.MartínezR. (2018). Integrating enzyme immobilization and protein engineering: an alternative path for the development of novel and improved industrial biocatalysts. Biotechnol. Adv.36, 1470–1480. 10.1016/j.biotechadv.2018.06.002

  • 13

    BhartiA.MittalS.RanaS.DahiyaD.AgnihotriN.PrabhakarN. (2020). Electrochemical biosensor for miRNA-21 based on gold-platinum bimetallic nanoparticles coated 3-aminopropyltriethoxy silane. Anal. Biochem.609, 113908. 10.1016/j.ab.2020.113908

  • 14

    BiéJ.SepodesB.FernandesP. C. B.RibeiroM. H. L. (2022). Enzyme Immobilization and Co-immobilization: Main Framework, Advances and Some Applications. 10.3390/pr10030494

  • 15

    BolivarJ. M.WoodleyJ. M.Fernandez-LafuenteR. (2022). Is enzyme immobilization a mature discipline? Some critical considerations to capitalize on the benefits of immobilization. Chem. Soc. Rev.51, 6251–6290. 10.1039/d2cs00083k

  • 16

    BollellaP.KatzE. (2020). Enzyme-based biosensors: tackling electron transfer issues. Sensors Switz.20, 1–32. 10.3390/s20123517

  • 17

    BouzouitaM.PathakS.ZayerF.BelgacemH.TzouvadakiI. (2025). Advanced memristive architectures based on nanomaterials for biomedical applications: a mini review. Front. Nanotechnol.7, 1558743. 10.3389/fnano.2025.1558743

  • 18

    BrytanW.ShortallK.DuarteF.SoulimaneT.PadrelaL. (2024). Contribution of a C-Terminal extension to the substrate affinity and oligomeric stability of aldehyde dehydrogenase from Thermus thermophilus HB27. Biochemistry63, 1075–1088. 10.1021/acs.biochem.3c00698

  • 19

    ByunS.-J.KimD.-G.ParkK.-D.ChoiY.-J.KumarP.AliI.et al (2022). A low-power analog processor-in-memory-based convolutional neural network for biosensor applications. Sensors22, 4555. 10.3390/s22124555

  • 20

    ChenM.ZhangJ.JiG.WangH.ZhuB.ChenC.et al (2025). Universal flexible wearable biosensors for noninvasive health monitoring. ACS Appl. Mater. Interfaces.17, 20741–20755. 10.1021/acsami.4c22623

  • 21

    ChenH.YanZ.NingJ.BaoX.DingL.ShuC. (2025). Increasing the dual-enzyme Cascade biocatalysis efficiency and stability of metal-organic frameworks via one-step coimmobilization for visual detection of glucose. Nanoscale17, 2174–2183. 10.1039/d4nr04156a

  • 22

    ChienP.-J.SuzukiT.YeM.TomaK.ArakawaT.IwasakiY.et al (2020). Ultra-sensitive isopropanol biochemical gas sensor (bio-sniffer) for monitoring of human volatiles. Sensors20, 6827. 10.3390/s20236827

  • 23

    ChuY.ChenY.GuZ.LiuX.ChengT.ShenY.et al (2025). Regulated enzyme distribution for enhanced wearable sweat glucose monitoring. Anal. Chim. Acta.1382, 344823. 10.1016/j.aca.2025.344823

  • 24

    ClarkL. C.LyonsC. (1962). Electrode systems for continuous monitoring in cardiovascular surgery. Ann. N. Y. Acad. Sci.102, 29–45. 10.1111/j.1749-6632.1962.tb13623.x

  • 25

    CoelhoJ. H.EiseleA. P. P.ValeziC. F.MattosG. J.SchirmannJ. G.DekkerR. F. H.et al (2019). Exploring the exocellular fungal biopolymer botryosphaeran for laccase-biosensor architecture and application to determine dopamine and spironolactone. Talanta204, 475–483. 10.1016/j.talanta.2019.06.033

  • 26

    CovaC. M.RincónE.EspinosaE.SerranoL.ZulianiA. (2022). Paving the way for a green transition in the design of sensors and biosensors for the detection of volatile organic compounds (VOCs). Biosensors12, 51. 10.3390/bios12020051

  • 27

    DaronchN. A.KelbertM.PereiraC. S.de AraújoP. H. H.de OliveiraD. (2020). Elucidating the choice for a precise matrix for laccase immobilization: a review. Chem. Eng. J.397, 125506. 10.1016/j.cej.2020.125506

  • 28

    DengZ.WangF.ZhouB.LiJ.LiB.LiangH. (2019). Immobilization of pectinases into calcium alginate microspheres for fruit juice application. Food Hydrocoll.89, 691–699. 10.1016/j.foodhyd.2018.11.031

  • 29

    DongL.ZengX.XiongY.XiaoX.ZhanD.WangS. (2025). Enzymatic bioelectrodes based on ferrocene-modified metal-organic layers for electrochemical glucose detection. Anal. Bioanal. Chem.417, 2217–2224. 10.1007/s00216-025-05808-0

  • 30

    DoshiD.KumarS.PatelB.ChaudhariD.PatelS.HiraniT.et al (2025). Micronutrients and the periodontium: a narrative review. Cureus17 (4), e81694. 10.7759/cureus.81694

  • 31

    EdwardsC.DuanghathaipornsukS.GoltzM.KanelS.KimD. S. (2019). Peptide nanotube encapsulated enzyme biosensor for vapor phase detection of malathion, an organophosphorus compound. Sensors19, 3856. 10.3390/s19183856

  • 32

    GanonyanN.BenmelechN.BarG.GvishiR.AvnirD. (2020). Entrapment of enzymes in silica aerogels. Mater. Today.33, 24–35. 10.1016/j.mattod.2019.09.021

  • 33

    GaoS.Rojas-VegaF.Rocha-MartinJ.GuisánJ. M. (2021). Oriented immobilization of antibodies through different surface regions containing amino groups: selective immobilization through the bottom of the Fc region. Int. J. Biol. Macromol.177, 19–28. 10.1016/j.ijbiomac.2021.02.103

  • 34

    GaoX.PanH.DangW.XvT.GuoX.QiaoC.et al (2026). Co-immobilization of dual-enzyme within hierarchical porous ZIF-8 for sensitive glucose colorimetric sensing. Talanta308, 129837. 10.1016/j.talanta.2026.129837

  • 35

    GermainM. E.KnappM. J. (2009). Optical explosives detection: from color changes to fluorescence turn-on. Chem. Soc. Rev.38, 2543–2555. 10.1039/b809631g

  • 36

    GhindilisA. L.AtanasovP.WilkinsE. (1997). Enzyme-catalyzed direct electron transfer: fundamentals and analytical applications. Electroanalysis9, 661–674. 10.1002/elan.1140090902

  • 37

    GhoshB.SahaR.BhattacharyaD.MukhopadhyayM. (2019). Laccase and its source of sustainability in an enzymatic biofuel cell. Bioresour. Technol. Rep.6, 268–278. 10.1016/j.biteb.2019.03.013

  • 38

    GondoS.SatoT.KusunokiK. (1973). Note on the lineweaver-burk plots for the insolubilized enzyme particle. Chem. Eng. Sci.28, 1773–1774. 10.1016/0009-2509(73)80032-6

  • 39

    GuisanJ. M.Fernandez-LorenteG.Rocha-MartinJ.Moreno-GameroD. (2022). Enzyme immobilization strategies for the design of robust and efficient biocatalysts. Curr. Opin. Green Sustain. Chem.35, 100593. 10.1016/j.cogsc.2022.100593

  • 40

    HernandezK.Fernandez-LafuenteR. (2011). Control of protein immobilization: coupling immobilization and site-directed mutagenesis to improve biocatalyst or biosensor performance. Enzyme Microb. Technol.48, 107–122. 10.1016/j.enzmictec.2010.10.003

  • 41

    HouJ.DongG.YeY.ChenV. (2014). Enzymatic degradation of bisphenol-A with immobilized laccase on TiO2 sol-gel coated PVDF membrane. J. Memb. Sci.469, 19–30. 10.1016/j.memsci.2014.06.027

  • 42

    IkegaiK.Okuda-ShimazakiJ.TranT. T.HatadaM.AsanoR.IkebukuroK.et al (2024). The 2.5th generation enzymatic sensors based on the construction of quasi-direct electron transfer type NAD(P)-dependent dehydrogenases. Biosens. Bioelectron.255, 116219. 10.1016/j.bios.2024.116219

  • 43

    IitaniK.NakayaM.TomonoT.TomaK.ArakawaT.TsuchidoY.et al (2022). Enzyme-embedded electrospun fiber sensor of hydrophilic polymer for fluorometric ethanol gas imaging in vapor phase. Biosens. Bioelectron.213, 114453. 10.1016/j.bios.2022.114453

  • 44

    IitaniK.MoriH.IchikawaK.TomaK.ArakawaT.IwasakiY.et al (2023). Gas-phase biosensors (bio-sniffers) for measurement of 2-nonenal, the causative volatile molecule of human aging-related body odor. Sensors23, 5857. 10.3390/s23135857

  • 45

    Jakubow-PiotrowskaK.KowalewskaB. (2025). Hybrid carbon nanostructures as efficient electron transfer platforms for lactate biosensing. Bioelectrochemistry167, 109080. 10.1016/j.bioelechem.2025.109080

  • 46

    Jamshidnejad-TosaramandaniT.KashanianS.OmidfarK.SchiöthH. B. (2025). The role of nanomaterials in the wearable electrochemical glucose biosensors for diabetes management. Biosensors15, 451. 10.3390/bios15070451

  • 47

    JayanthiK. G.S.KS. (2020). Cholesterol oxidase immobilized inulin based nanocomposite as the sensing material for cholesterol in biological and food samples. Enzyme Microb. Technol.140, 109631. 10.1016/j.enzmictec.2020.109631

  • 48

    KaurJ.SinghP. K. (2020). Enzyme-based optical biosensors for organophosphate class of pesticide detection. Phys. Chem. Chem. Phys.22, 15105–15119. 10.1039/d0cp01647k

  • 49

    KhanN. A.AlshammariF. S.Tavera RomeroC. A.SulaimanM.LaouiniG. (2021). Mathematical analysis of reaction–diffusion equations modeling the michaelis–menten kinetics in a micro-disk biosensor. Molecules26, 7310. 10.3390/molecules26237310

  • 50

    KimD. S.SobhanA.OhJ. H.LeeJ.ParkC.LeeJ. (2024). Development of single-walled carbon nanotube-based electrodes with enhanced dispersion and electrochemical properties for blood glucose monitoring. Biosensors14, 630. 10.3390/bios14120630

  • 51

    KimY.-J.KwonY.YooY.TheyagarajanK.SaikrithikaS.LeeA.et al (2025). Demonstration of CMOS-compatible memristor-based electrochemical biosensor transducer with threshold-sensing functionality. Nat. Commun.16, 10851. 10.1038/s41467-025-66372-w

  • 52

    KimH. J.SoB. J.ChoiY.NamE. J.ShinG.LeeJ. H.et al (2026). Glucose oxidase/o-phenylenediamine interfaces on dendritic Pt microneedles for highly sensitive and selective continuous glucose monitoring sensors. ACS Appl. Mater. Interfaces18, 4749–4759. 10.1021/acsami.5c19085

  • 53

    KirthigaO. M.RajendranL. (2015). Approximate analytical solution for non-linear reaction diffusion equations in a mono-enzymatic biosensor involving michaelis-menten kinetics. J. Electroanal. Chem.751, 119–127. 10.1016/j.jelechem.2015.05.036

  • 54

    KoçM. U.IşıkC. (2026). Laccase immobilization on a natural lignocellulosic biomass support: optimization, characterization, and bisphenol A degradation. Biotech16, 229. 10.1007/s13205-026-04862-4

  • 55

    KüchlerA.YoshimotoM.LuginbühlS.MavelliF.WaldeP. (2016). Enzymatic reactions in confined environments. Nat. Nanotechnol.11, 409–420. 10.1038/nnano.2016.54

  • 56

    KumarA.ParkG. D.PatelS. K. S.KondaveetiS.OtariS.AnwarM. Z.et al (2019). SiO2 microparticles with carbon nanotube-derived mesopores as an efficient support for enzyme immobilization. Chem. Eng. J.359, 1252–1264. 10.1016/j.cej.2018.11.052

  • 57

    LeeJ. Y.guk ChoD.ChoS. P.ChoiJ. H.SungS. J.HongS.et al (2020). Semiconducting carbon nanotube fibers for electrochemical biosensor platforms. Mater. Des.192, 108740. 10.1016/j.matdes.2020.108740

  • 58

    LiY.CirinoP. C. (2024). Recent advances in engineering proteins for biocatalysis. Biotechnol. Bioeng.111, 1273–1287. 10.1002/bit.25240

  • 59

    LiY.ZhaoA.WangJ.YuJ.XiaoF.SunH. (2022). Highly bright gold nanowires arrays for sensitive detection of urea and urease. Nanomaterials12, 4023. 10.3390/nano12224023

  • 60

    LiZ.ShiQ.DongX.SunY. (2024). Co-Immobilization of laccase and mediator into Fe-Doped ZIF-8 significantly enhances the degradation of organic pollutants. Molecules29, 307. 10.3390/molecules29020307

  • 61

    LiuY.LiuY.QiaoL.LiuY.LiuB. (2018). Advances in signal amplification strategies for electrochemical biosensing. Curr. Opin. Electrochem.12, 5–12. 10.1016/j.coelec.2018.05.001

  • 62

    LiuS.ChaoH.HeD.WangY.YangY. (2025). Biomimetic co-immobilization of β-glucosidase, glucose oxidase, and horseradish peroxidase to construct a multi-enzyme biosensor for determination of amygdalin. Int. J. Biol. Macromol.297, 139868. 10.1016/j.ijbiomac.2025.139868

  • 63

    LiuX.GuoS.XuY.ZhangZ.LiuY.ShaoY.et al (2025). Development of a portable biosensor for glyphosate detection using bacterial surface-displayed glyphosate oxidase. Biosens. Bioelectron.292, 117974. 10.1016/j.bios.2025.117974

  • 64

    MasudJ. S.AhmedF.KimJ. (2025). 2D materials: from design and synthesis to applications in electrical and electrochemical biosensors. Small21, e04955. 10.1002/smll.202504955

  • 65

    MaS.WanZ.WangC.SongZ.DingY.ZhangD.et al (2024). Ultra-sensitive and stable multiplexed biosensors array in fully printed and integrated platforms for reliable perspiration analysis. Adv. Mater.36, e2311106. 10.1002/adma.202311106

  • 66

    MacagnanoA.MolinariF. N.SerrecchiaS.PapaP.TaddeiA. R.De CesareF. (2025). Tuning nanofibrous sensor performance in selective detection of B-VOCs by MIP-NP loading. Nanomaterials15, 1–23. 10.3390/nano15161220

  • 67

    MachadoA. C.Salvo-CominoC.Rodriguez-MendezM. L.CaseliL. (2025). Enhancing biosensor capabilities through laccase immobilization in nanostructured langmuir-blodgett films for phenol detection. Results Surfaces Interfaces18, 100366. 10.1016/j.rsurfi.2024.100366

  • 68

    MaenoM.SakurabaH.OhshimaT.SuyeS.-I.SatomuraT. (2025). Creation of a highly stable direct electron transfer-type enzyme sensor by combining a hyperthermophilic dehydrogenase and natural electron mediator. Biotechnol. Lett.47, 45. 10.1007/s10529-025-03587-3

  • 69

    MaoS.JiangJ.XiongK.ChenY.YaoY.LiuL.et al (2024). Enzyme Engineering: Performance Optimization, Novel Sources, and Applications in the Food Industry. 10.3390/foods13233846

  • 70

    MartinezC. M.AlvarezL. H. (2018). Application of redox mediators in bioelectrochemical systems. Biotechnol. Adv.36, 1412–1423. 10.1016/j.biotechadv.2018.05.005

  • 71

    MiaoP.TangY.WangB.YinJ.NingL. (2015). Signal amplification by enzymatic tools for nucleic acids. Trac. - Trends Anal. Chem.67, 1–15. 10.1016/j.trac.2014.12.006

  • 72

    MiškovićM. Z.WojtyśM.Winiewska-SzajewskaM.Wielgus-KutrowskaB.MatkovićM.Domazet JurašinD.et al (2024). Location is everything: influence of his-tag fusion site on properties of adenylosuccinate synthetase from Helicobacter pylori. Int. J. Mol. Sci.25, 7613. 10.3390/ijms25147613

  • 73

    MohtarL. G.ArandaP.MessinaG. A.NazarenoM. A.PereiraS. V.RabaJ.et al (2019). Amperometric biosensor based on laccase immobilized onto a nanostructured screen-printed electrode for determination of polyphenols in propolis. Microchem. J.144, 13–18. 10.1016/j.microc.2018.08.038

  • 74

    MonkrathokJ.JanphuangP.SuphachiaraphanS.KampaengsriS.KamkaewA.ChansaenpakK.et al (2024). Enhancing glucose biosensing with graphene oxide and ferrocene-modified linear poly(ethylenimine). Biosensors14, 161. 10.3390/bios14040161

  • 75

    NgoT. A.SchultesE.MaierA.HadewigS. R.TischlerD. (2022). Improving biocatalytic properties of an azoreductase via the N-terminal fusion of formate dehydrogenase. ChemBioChem23, e202100643. 10.1002/cbic.202100643

  • 76

    NguyenH. H.LeeS. H.LeeU. J.FerminC. D.KimM. (2019). Immobilized enzymes in biosensor applications. Mater. (Basel)12, 1–34. 10.3390/ma12010121

  • 77

    NielsenJ. R.WeusthuisR. A.HuangW. E. (2023). Growth-coupled enzyme engineering through manipulation of redox cofactor regeneration. Biotechnol. Adv.63, 108102. 10.1016/j.biotechadv.2023.108102

  • 78

    PaoliniD.SilveriF.Della PelleF.Di CristoforoI. V.ScroccarelloA.BollellaP.et al (2025). Paper-based graphene bioelectrode enabling third-generation fructose dehydrogenase sensing of inulin. Biosens. Bioelectron.297, 118347. 10.1016/j.bios.2025.118347

  • 79

    PattiS.Magrini AlunnoI.PedroniS.RivaS.FerrandiE. E.MontiD. (2025). Advances and challenges in the development of immobilized enzymes for batch and flow biocatalyzed processes. ChemSusChem18, e202402007. 10.1002/cssc.202402007

  • 80

    Portugal-GómezP.SvigeljR.ZanetteF.TonioloR.Domínguez-RenedoO.Alonso-LomilloM. A. (2025). Gas sensor for 4-ethylguaiacol detection based on tyrosinase enzymatic activity in a deep eutectic solvent. Microchim. Acta.192, 364. 10.1007/s00604-025-07218-6

  • 81

    QinJ.WangW.GaoL.YaoS. Q. (2022). Emerging biosensing and transducing techniques for potential applications in point-of-care diagnostics. Chem. Sci.13, 2857–2876. 10.1039/d1sc06269g

  • 82

    RafatN.SatohP.WordenR. M. (2021). Electrochemical biosensor for markers of neurological esterase inhibition. Biosensors11, 459. 10.3390/bios11110459

  • 83

    RazmshoarP.BesbesF.MadaciA.MlikaR.BahramiS. H.RabieeM.et al (2023). A conductometric enzymatic methanol sensor based on polystyrene - PAMAM dendritic polymer electrospun nanofibers. Talanta260, 124630. 10.1016/j.talanta.2023.124630

  • 84

    ReichhartT. M. B.ScheiblbrandnerS.SygmundC.HarreitherW.SchenkenfelderJ.SchulzC.et al (2023). Interface engineering of cellobiose dehydrogenase improves interdomain electron transfer. Protein Sci.32, e4702. 10.1002/pro.4702

  • 85

    ReusB.DamianM.MuttiF. G. (2024). Advances in cofactor immobilization for enhanced continuous-flow biocatalysis. J. Flow. Chem.14, 219–238. 10.1007/s41981-024-00315-2

  • 86

    RobescuM. S.BavaroT. (2025). A comprehensive guide to enzyme immobilization: all you need to know. Molecules30, 939. 10.3390/molecules30040939

  • 87

    SainzR.del PozoM.VázquezL.Vilas-VarelaM.Castro-EstebanJ.BlancoE.et al (2022). Lactate biosensing based on covalent immobilization of lactate oxidase onto chevron-like graphene nanoribbons via diazotization-coupling reaction. Anal. Chim. Acta.1208, 339851. 10.1016/j.aca.2022.339851

  • 88

    SakalauskieneL.PopovA.Kausaite-MinkstimieneA.RamanaviciusA.RamanavicieneA. (2022). The impact of glucose oxidase immobilization on dendritic gold nanostructures on the performance of glucose biosensors. Biosensors12, 1–14. 10.3390/bios12050320

  • 89

    SalesM. B.NetoJ. G. L.De Sousa BrazA. K.De Sousa JuniorP. G.MeloR. L. F.ValérioR. B. R.et al (2023). Trends and opportunities in enzyme biosensors coupled to metal-organic frameworks (MOFs): an advanced bibliometric analysis. Electrochem4, 181–211. 10.3390/electrochem4020014

  • 90

    SchachingerF.ChangH.ScheiblbrandnerS.LudwigR. (2021). Amperometric biosensors based on direct electron transfer enzymes. Molecules26, 4525. 10.3390/molecules26154525

  • 91

    SchmiegB.DöbberJ.KirschhöferF.PohlM.FranzrebM. (2019). Advantages of hydrogel-based 3D-printed enzyme reactors and their limitations for biocatalysis. Front. Bioeng. Biotechnol.6, 211. 10.3389/fbioe.2018.00211

  • 92

    SchmiegB.NguyenM.FranzrebM. (2020). Simulative minimization of mass transfer limitations within hydrogel-based 3D-printed enzyme carriers. Front. Bioeng. Biotechnol.8, 365. 10.3389/fbioe.2020.00365

  • 93

    SeenivasanR.KolodziejC.KarunakaranC.BurdaC. (2017). Nanotechnology for electroanalytical biosensors of reactive oxygen and nitrogen species. Chem. Rec.17, 886–901. 10.1002/tcr.201600143

  • 94

    SenthilkumarM.MathiyarasuJ.JosephJ.PhaniK. L. N.YegnaramanV. (2008). Electrochemical instability of indium tin oxide (ITO) glass in acidic pH range during cathodic polarization. Mater. Chem. Phys.108, 403–407. 10.1016/j.matchemphys.2007.10.030

  • 95

    ShimomuraT.ItohT.SumiyaT.MizukamiF.OnoM. (2008). Electrochemical biosensor for the detection of formaldehyde based on enzyme immobilization in mesoporous silica materials. Sensors Actuators, B Chem.135, 268–275. 10.1016/j.snb.2008.08.025

  • 96

    ShitandaI.HayaseK.LoewN.WatanabeH.ItagakiM. (2026). Development of an enzyme-based electrochemical acetone gas sensor printed on a porous polyimide film. ACS Omega11, 31398–31404. 10.1021/acsomega.6c01572

  • 97

    ShortallK.OteroF.BendlS.SoulimaneT.MagnerE. (2022). Enzyme immobilization on metal organic frameworks: the effect of buffer on the stability of the support. Langmuir38, 13382–13391. 10.1021/acs.langmuir.2c01630

  • 98

    SilversteinT. P.GoodneyD. E. (2010). Enzyme-linked biosensors: michaelis-menten kinetics need not apply. J. Chem. Educ.87, 905–907. 10.1021/ed100381r

  • 99

    SinghR. P. (2011). Prospects of nanobiomaterials for biosensing. Int. J. Electrochem.2011, 1–30. 10.4061/2011/125487

  • 100

    SkládalP. (2024). Piezoelectric biosensors: shedding light on principles and applications. Microchim. Acta.191, 184. 10.1007/s00604-024-06257-9

  • 101

    SmutokO.KavetskyyT.ProkopivT.SerkizR.Wojnarowska-NowakR.ŠaušaO.et al (2021). New micro/nanocomposite with peroxidase-like activity in construction of oxidases-based amperometric biosensors for ethanol and glucose analysis. Anal. Chim. Acta.1143, 201–209. 10.1016/j.aca.2020.11.052

  • 102

    SowaK.Okuda-ShimazakiJ.FukawaE.SodeK. (2024). Direct electron transfer-type oxidoreductases for biomedical applications. Annu. Rev. Biomed. Eng.26, 357–382. 10.1146/annurev-bioeng-110222-101926

  • 103

    SunX.ZhangH.HuangL.HaoS.ZhaiJ.DongS. (2021). A naked-eye readout self-powered electrochemical biosensor toward indoor formaldehyde: on-site detection and exposure risk warning. Biosens. Bioelectron.177, 112975. 10.1016/j.bios.2021.112975

  • 104

    SunQ.YuW.GongM.MaJ.LiuG.MeiT.et al (2024). Xanthine oxidase immobilized cellulose membrane-based colorimetric biosensor for screening and detecting the bioactivity of xanthine oxidase inhibitors. Int. J. Biol. Macromol.275, 133450. 10.1016/j.ijbiomac.2024.133450

  • 105

    TemoçinZ.İnalM.GökgözM.YiğitoğluM. (2018). Immobilization of horseradish peroxidase on electrospun poly(vinyl alcohol)–polyacrylamide blend nanofiber membrane and its use in the conversion of phenol. Polym. Bull.75, 1843–1865. 10.1007/s00289-017-2129-5

  • 106

    ThangarajB.SolomonP. R. (2019). Immobilization of lipases – a review. Part I: enzyme immobilization. ChemBioEng Rev.6, 157–166. 10.1002/cben.201900016

  • 107

    TomaK.IwasakiK.ZhangG.IitaniK.ArakawaT.IwasakiY.et al (2021). Biochemical methanol gas sensor (MeOH bio-sniffer) for non-invasive assessment of intestinal flora from breath methanol. Sensors21, 4897. 10.3390/s21144897

  • 108

    UkiradeN.ChoudhariU.BhapkarS.JadhavU.JagtapS.RaneS. (2025). Enzyme-immobilized graphene oxide-based electrochemical biosensor for glutathione detection. RSC Adv.15, 12987–12996. 10.1039/d4ra09033k

  • 109

    VermaM. L.KumarS.DasA.RandhawaJ. S.ChamundeeswariM. (2020). Chitin and chitosan-based support materials for enzyme immobilization and biotechnological applications. Environ. Chem. Lett.18, 315–323. 10.1007/s10311-019-00942-5

  • 110

    WangK.RenH.LiN.TanX.DangF. (2018). Ratiometric fluorescence sensor based on cholesterol oxidase-functionalized mesoporous silica nanoparticle@ZIF-8 core-shell nanocomposites for detection of cholesterol. Talanta188, 708–713. 10.1016/j.talanta.2018.06.019

  • 111

    WangF.ChenX.WangY.LiX.WanM.ZhangG.et al (2022). Insights into the structures, inhibitors, and improvement strategies of glucose oxidase. Int. J. Mol. Sci.23, 9841. 10.3390/ijms23179841

  • 112

    WangT.FanX.MaY.LiZ.HeC. (2024). Dual enzymes proteinosome with cascade activity for ultrasensitive glucose biosensing. Sensors Actuators B Chem.402, 135067. 10.1016/j.snb.2023.135067

  • 113

    WangH.-R.HouE.-H.XuN.ZhangY.-F.WuJ.-F.YuanW.-J.et al (2024). Photoelectrochemical solution gated graphene field-effect transistor functionalized by enzymatic cascade reaction for organophosphate detection. Small20, e2402655. 10.1002/smll.202402655

  • 114

    WangQ.ZhaoZ. A.YaoK. Y.ChengY. L.WongD. S. H.WongD. W. C.et al (2025). The versatility of biological field-effect transistor-based biosensors (BioFETs) in point-of-care diagnostics: applications and future directions for peritoneal dialysis monitoring. Biosensors15, 193. 10.3390/bios15030193

  • 115

    WangH.HouE.XuN.WuJ.GaoS.NieP.et al (2025). Acetylcholinesterase-assisted photoelectrochemical solution gated graphene field-effect transistor for organophosphates pesticide detection. Food Chem.486, 144631. 10.1016/j.foodchem.2025.144631

  • 116

    WangZ.LiuW-J.TaoJ.HuJ.ZhangC-Y. (2025). Enzymatic cascade amplification-modulated Thermus thermophilus Argonaute biosensor for simultaneous monitoring of multiple Piwi-interacting RNAs. Biosens. Bioelectron.276, 117261. 10.1016/j.bios.2025.117261

  • 117

    YeJ.MiaoB.XiongY.GuanY.LuY.JiaZ.et al (2025). 3D printed porous magnesium metal scaffolds with bioactive coating for bone defect repair: enhancing angiogenesis and osteogenesis. J. Nanobiotechnology.23, 160. 10.1186/s12951-025-03222-3

  • 118

    ZanR.ZhuL.WuG.ZhangH. (2023). Identification of novel peptides with alcohol dehydrogenase (ADH) activating ability in chickpea protein hydrolysates. Foods12, 1574. 10.3390/foods12081574

  • 119

    ZayatsM.KharitonovA. B.KatzE.BuückmannA. F.WillnerI. (2000). An integrated NAD+-dependent enzyme-functionalized field-effect transistor (ENFET) system: development of a lactate biosensor. Biosens. Bioelectron.15, 671–680. 10.1016/S0956-5663(00)00120-2

  • 120

    ZhangJ.WangB.LiY.ShuW.HuH.YangL. (2019). An acetylcholinesterase biosensor with high stability and sensitivity based on silver nanowire-graphene-TiO2 for the detection of organophosphate pesticides. RSC Adv.9, 25248–25256. 10.1039/c9ra02140j

  • 121

    ZhangM.WangX.LiuW.CuiX.WangY.FanL.et al (2025). Engineering a binding peptide for oriented immobilization and efficient bioelectrocatalytic oxygen reduction of multicopper oxidases. ACS Appl. Mater. Interfaces17, 2355–2364. 10.1021/acsami.4c12970

  • 122

    ZhangG.HuangD.IchikawaK.IitaniK.NakajimaY.MitsubayashiK. (2026). Headset-type biofluorometric gas sensor with CMOS for transcutaneous ethanol from the ear canal. Sensors26, 2817. 10.3390/s26092817

  • 123

    ZhaoX.OuY.BaiR.YinJ.WangG.WangJ.et al (2025). In situ cleavage-based sortase A-mediated site-specific immobilization of beta2-adrenoceptor on gold surface for surface plasmon resonance measurement. Biosens. Bioelectron.281, 117452. 10.1016/j.bios.2025.117452

Summary

Keywords

apparent kinetics, conventional and oriented immobilization, direct electron transfer, enzymatic biosensors, gas-phase biosensing, signal amplification

Citation

Arvizu-De León IC, Quintanilla-Villanueva GE, Cedillo-González EI, Muñoz-Benítez J, Villarreal-Chiu JF and Rodríguez-Delgado MM (2026) Enzymatic biosensors: biocatalytic mechanisms, immobilization strategies, and signal amplification approaches. Front. Sens. 7:1933492. doi: 10.3389/fsens.2026.1933492

Received

10 July 2026

Revised

15 August 2026

Accepted

05 October 2026

Published

27 October 2026

Volume

7 - 2026

Edited by

Carlo Massaroni, Campus Bio-Medico University, Italy

Reviewed by

Andres Vercik, University of São Paulo, Brazil

Tahereh Jamshidnejad, Razi University, Iran

Updates

Copyright

*Correspondence: Melissa Marlene Rodríguez-Delgado, ; Juan Francisco Villarreal-Chiu,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics