Analysis of compressional acoustic wave responses in cell culture media for reliable QCM-based breast cancer cell monitoring

  • 1. Department of Biotechnology, Birla Insitute of Technology and Science, Pilani – Dubai Campus, Dubai International Academic City, Dubai, United Arab Emirates

  • 2. Department of Electrical and Electronics Engineering, Birla Insitute of Technology and Science, Pilani – Dubai Campus, Dubai International Academic City, Dubai, United Arab Emirates

Abstract

Quartz Crystal Microbalance (QCM) sensors are widely used to study biological and soft-matter interfaces due to their exceptional sensitivity to mass loading and interfacial mechanical properties. While classical QCM theory assumes predominantly shear-wave coupling into a semi-infinite Newtonian liquid, finite liquid thickness and acoustic reflections give rise to pronounced compressional (longitudinal) wave effects that strongly modulate both resonance frequency and motional resistance. Such compressional acoustic-wave responses should be properly accounted for when sensing in the liquid phase, for instance, when working with cell suspensions. In this work, we systematically investigate compressional-wave responses in typically used cell culture media for breast cancer cells including DMEM and RPMI-1640 across varying droplet volumes using a 5 MHz AT-cut QCM. Time-resolved measurements are analyzed using four parameters: the time period of compressional acoustic wave modulation , the time associated with a phase shift between resonance frequency and resistance oscillations , the peak-to-peak shifts in frequency and resistance . DMEM and RPMI-1640 both exhibit strong volume-dependent periodic oscillations. At lower volumes, they exhibit low-frequency oscillations of the time period minutes. However, as volume increases, the oscillations gradually evolve into high frequency oscillations of the time period minutes. The peak-to-peak shifts and are and , respectively. The resonance frequency and resistance oscillations also exhibit a phase shift of minutes. These results highlight that compressional-wave effects occur even in simple cell culture media, necessitating their explicit consideration when interpreting QCM data in the presence of cells.

1 Introduction

QCM sensors (Murrieta-Rico et al., 2022; Gouda et al., 2023; Carli et al., 2024; Chen et al., 2025; Wang et al., 2025) are widely used to investigate biological interfaces because of their exceptional sensitivity to changes in mass, viscoelasticity, and interfacial mechanical properties (Lim et al., 2020; Jandas et al., 2021; Alanazi et al., 2023; Ganesan et al., 2016). Recent developments have demonstrated the use of QCM sensors for monitoring cellular environments (Gouda et al., 2023). However, the acoustic response of a QCM in a liquid environment can contain contributions beyond the conventional shear-wave interaction. Classical QCM theory assumes that the vibrating quartz crystal generates predominantly shear waves that couple into a semi-infinite Newtonian liquid (Sauerbrey, 1959; Voinova et al., 1999; Johannsmann, 2015; Fort et al., 2022). However, in some experimental situations, there may also be secondary compressional wave effects due to the longitudinal propagation of acoustic waves into the liquid.

A limited number of studies have reported compressional wave propagation in QCM systems in the past. McKenna et al. (2001) showed that 2 L and 5 L microdroplets of water exhibit noticeable compressional waves. This study shows that droplet position and height strongly influence the acoustic response, with good agreement between theoretical predictions and experiments for both static and evaporating microdroplets. Zhang et al. (2010) demonstrated these effects in silicone oil droplets with viscosity in a range from to , resulting in cyclical variations in the frequency and resistance. The effects of interfacial slip and viscoelasticity have been discussed to qualitatively explain the observed positive frequency shift. More recently, Kowarsch et al. (2020) report that compressional-wave effects are stronger at the fundamental mode, with the normal motion amplitude (1% of transverse motion) decreasing for higher overtones. Experiments and analytical modeling show that compressional wave and standing-wave effects diminish at higher overtones, consistent with the small-load approximation.

Despite these studies, compressional acoustic wave effects have not yet been systematically studied in biological assays. This gap is particularly important, as biological assays often rely on liquid-phase environments where additional acoustic phenomena may arise and influence the sensor response. Prior to conducting cell-based experiments, such as cell counting or characterization (Eshun et al., 2023; Kowalczyk et al., 2023; Özdabak Sert et al., 2023; Palleschi et al., 2024; Rogala et al., 2025; Kandel, 2026), it is therefore essential to first isolate and account for compressional wave effects in the liquid media alone. Neglecting these contributions may lead to misinterpretation of the measured signals, especially when attributing frequency shifts or dissipation changes solely to cellular properties.

In this study, we systematically investigate and demonstrate the presence of compressional acoustic waves in commonly used cell culture media for breast cancer cells including Dulbecco’s Modified Eagle Medium (DMEM) and Roswell Park Memorial Institute (RPMI-1640). By establishing their contribution in the absence of breast cancer cells, we provide a necessary baseline for accurately interpreting QCM responses in more complex biological environments. This understanding is crucial for improving the reliability and reproducibility of QCM-based measurements in applications involving cell suspensions and other biological systems.

2 Theory for compressional acoustic wave generation in QCM crystals

In the analysis of QCM operation in liquids, the interaction between the resonator surface and the contacting medium is described through the load impedance , defined as the ratio of stress to particle velocity. Because the surface motion contains both tangential (shear) and normal (compressional) components, the load impedance naturally separates into two physically distinct contributions. The term represents the shear-wave impedance associated with tangential motion, while represents the compressional-wave impedance arising from normal motion that launches longitudinal waves into the liquid. Together, these components describe how the liquid loads the resonator within the small-load approximation.

The general tensorial form of the load impedance relates the surface velocity to the resulting stress through

Here, is the stress tensor, the velocity component at the surface, and the third-rank load-impedance tensor. For a thickness-shear resonator, only the components and are relevant.

The resulting complex frequency shift of the resonator is obtained by integrating the local contributions of shear and normal motion over the crystal surface:

In this expression, is the effective mass of the resonator, denotes a point on the surface, and is a reference velocity chosen such that the integral of equals unity.

When the small-load approximation is valid, the shear and compressional contributions combine linearly, yielding the simplified relation (Kowarsch et al., 2020)

For a Newtonian liquid, the shear-wave impedance is given bywith the angular frequency, the liquid density, and the dynamic viscosity.

If the liquid were semi-infinite with respect to compressional waves, the corresponding compressional-wave impedance would be approximated bywhere is the P-wave modulus and the speed of sound.

In practice, compressional waves reflect between the resonator surface and the opposite wall of the liquid cell, forming a Fabry–Perot-type cavity. The resulting impedance is described by

The impedance evolves as

Table 1 presents the description of the parameters mentioned in the Equations 1–7.

TABLE 1

SymbolDescription
Stress tensor component at the resonator surface
Velocity component at the surface
Load-impedance tensor relating stress and velocity
Shear-wave load impedance
Compressional-wave load impedance
Effective mass of the resonator
Position on the resonator surface
Reference velocity used for normalization
Fundamental resonance frequency
Shear impedance of quartz
Numerical factor from mode-shape integration
Angular frequency
Density of the liquid
Dynamic viscosity of the liquid
P-wave (longitudinal) modulus
Speed of sound in the liquid
Compressional-wave wavenumber
Thickness of the liquid layer
Acoustic reflectivities at the cavity boundaries
Time for liquid interface to move by half a wavelength
Time offset in the impedance oscillation
Phase associated with reflectivity product
Time-averaged electrical drive power
Motional resistance of the resonator
Amplitude of the electrical current
Transverse velocity amplitude
Thickness of the quartz plate
Effective active area of the resonator
Piezoelectric stress coefficient
Piezoelectric strain coefficient
Quality factor of the resonator
Overtone number
Transverse displacement amplitude

Parameter explanation.

3 Experimental methods

To study the compressional acoustic wave responses, the experimental investigations are conducted using a QCM based measurement system (QCM200 Stanford Research Systems) configured for real-time monitoring of mass variations at the sensor interface. The setup primarily comprised a QCM100 Digital Controller integrated with a QCM25 Crystal Oscillator and interfaced with a computer-based acquisition platform for simultaneous recording of resonance frequency and resistance responses. The system operated with a nominal resonance frequency of 5 MHz employing a 1-inch-diameter AT-cut quartz crystal with a 13 mm diameter active gold electrode, selected due to their superior temperature stability and high sensitivity toward micro-scale mass deposition in liquid and biological environments. Electrical connectivity between the controller and oscillator units is established through a shielded RJ-45 interface cable, while the frequency and resistance outputs are obtained through BNC terminals and relayed to external monitoring and logging instruments. Figure 1 shows the experimental setup.

FIGURE 1

The sensing element consisted of a gold-plated quartz crystal mounted securely within a dedicated crystal holder assembly. The holder incorporated spring-loaded contact pins to ensure uniform electrical contact with the crystal electrodes and to maintain mechanical stability during liquid loading conditions. Prior to each experimental run, the crystal surface is subjected to a standardized cleaning protocol to eliminate contamination and ensure repeatability of measurements. The sensor is first immersed in 70% ethanol to remove organic residues and microbial contaminants, followed by rinsing with deionized water to eliminate solvent traces. The crystal is then allowed to air dry under ambient laboratory conditions before being mounted within the holder.

Data acquisition is performed using dedicated QCM interfacing software installed on a laboratory workstation. After sensor mounting and completion of the electrical connections, the system is powered on and allowed to stabilize to obtain baseline frequency and resistance readings. Experimental measurements are initiated through the software interface, enabling continuous logging of sensor responses, which are subsequently exported in spreadsheet format for graphical and analytical processing. Sample introduction is carried out using a calibrated micropipette, wherein the test liquid is carefully dispensed onto the central active surface of the quartz crystal to ensure uniform coverage of the sensing region.

The experimental study included characterization of biologically relevant cell culture media used for breast cancer cells including DMEM (DMEM; high glucose formulation, sigma-Aldrich, cat no. D6429) and RPMI-1640 (Gibco, cat. no. 11875093). These culture media are prepared as a sterile 1X working solution prior to use. To generate complete culture media, the basal culture media are supplemented with 10% (v/v) fetal bovine serum (FBS) and 1% (v/v) penicillin-streptomycin solution derived from a 100x stock. All procedures are carried out under aseptic conditions using sterile pipetting techniques within a laminar flow hood. The supplemented media are gently mixed to ensure uniform distribution of components while minimizing froth formation. The prepared media are stored at 2°–8 °C (commonly maintained at 4 °C) and used within 4 weeks. To maintain media integrity and prevent degradation of temperature sensitive components, repeated cycles of warming are avoided.

Upon loading of DMEM or RPMI-1640 sample on QCM surface, the interaction between the deposited mass and the oscillating quartz surface resulted in measurable shifts in resonance frequency along with corresponding variations in motional resistance. These responses are recorded over defined time intervals to capture the behavior of the test media. The experiments are conducted on droplets with four different volumes of DMEM/RPMI-1640: 400 L, 600 L, 800 L and 1 mL.

For each droplet volume and liquid condition, three independent experimental runs were performed, with each measurement recorded continuously for 120 min at a sampling interval of 1.02 s. All experiments were conducted at room temperature (approximately 25 °C) under normal laboratory ambient conditions and atmospheric pressure; relative humidity and minor variations in ambient pressure were not independently monitored. The liquid was deposited with the crystal maintained in its standard horizontal orientation, producing an approximate liquid height of 2 mm above the sensing region. Measurements were performed outside the biological safety cabinet with the measurement chamber open. No significant evaporation was observed during the 120-min measurement period under the present experimental conditions because of the relatively large liquid volume used. Following liquid deposition, the sample was allowed to settle for approximately 1 min before data acquisition was initiated. Prior to liquid loading, the quartz crystal and QCM measurement system were allowed to temperature-equilibrate and stabilize.

Between successive measurements, the residual sample was removed and the QCM crystal was cleaned according to the procedure described above to minimize cross-contamination and carryover effects. The repeated measurements were used to assess experimental reproducibility and variability. For each condition, the reported values represent the mean of the independent measurements, and the error bars in Figure 2 represent the corresponding standard deviation (SD). No formal hypothesis test was performed; therefore, the repeated measurements are used to characterize measurement variability rather than to establish statistical significance.

FIGURE 2

After completion of each measurement cycle, the residual sample was carefully removed, and the crystal underwent the cleaning procedure prior to subsequent testing to prevent cross-contamination. All experiments were conducted under controlled laboratory conditions to minimize environmental disturbances such as temperature fluctuations and mechanical vibrations that could influence frequency stability. The system is allowed sufficient warm-up time to achieve oscillator equilibrium, and capacitance compensation procedures were performed before liquid measurements to ensure that the recorded frequency shifts accurately corresponded to mass loading effects at the crystal interface.

4 Results and discussion

Figure 3 shows how the compressional-wave response in DMEM changes as the droplet volume increases. At 400 L, the frequency and resistance traces remain smooth and slowly oscillating. Besides this, there is also a continuous decrease in the frequency. However, beyond 600 L, the oscillations of resonance frequency and resistance are relatively stable with minimal drift. When the volume reaches 800 L and 1 mL, the oscillations become more irregular and high-frequency fluctuations are evidenced. This transition from smooth to highly modulated behavior demonstrates that larger droplets support stronger internal reflections, making the acoustic cavity more active and less stable.

FIGURE 3

Figure 4 illustrates that RPMI-1640 exhibits a similar volume-dependent oscillatory behavior. However, the high-frequency fluctuations become noticeable even at 600 L. At 1 mL, the signal becomes highly irregular with strong fluctuations throughout the measurement. These results indicate that RPMI-1640 can generate strong compressional-wave activity even in smaller volumes.

FIGURE 4

Figures 3a,e

, and similarly

Figures 4a,e

indicate that the resonance frequency and resistance oscillations show different patterns as the compressional waves influence mass loading and dissipation through distinct physical pathways.

Figure 5

presents the detailed oscillatory behavior of a 600

L DMEM droplet. The

and

curves are clearly phase-shifted, confirming that compressional waves modulate mass loading and dissipation differently. The parameters—

,

,

, and

— are extracted to quantitatively understand the volume-dependent compressional acoustic-wave behavior.

  • — Compressional-wave-induced modulation period.

  • Represents the characteristic timescale of the low-frequency modulation observed in the QCM resonance frequency and resistance responses. It should not be interpreted as the propagation period of the underlying MHz longitudinal acoustic wave; rather, it characterizes the temporal evolution of the QCM response resulting from changes in the compressional-wave interference condition.

  • — Peak-to-peak frequency modulation amplitude

  • Quantifies the total excursion of the frequency shift during each oscillation cycle. A larger signifies stronger modulation of the resonator’s mass-loading response, typically associated with more pronounced compressional-wave activity.

  • — Peak-to-peak resistance modulation amplitude

  • Measures the total excursion of the resistance oscillations . This parameter captures the extent of energy dissipation changes induced by the compressional wave.

  • — Time associated with the phase shift

  • Denotes the time by which the resistance oscillations lead the resonance frequency oscillations. The observable phase offset between the and oscillations indicates that compressional waves influence mass loading and dissipation through distinct physical pathways.

FIGURE 5

Figure 2 provides a consolidated comparison of how the key compressional-wave parameters evolve with droplet volume for DMEM, RPMI-1640, and water. In DMEM, all parameters exhibit a strong dependence on volume. At lower volumes, low-frequency oscillatory signatures of minutes are observed, with , and minutes. As volume increases, the acoustic cavity supports more pronounced longitudinal-wave interference, resulting in high-frequency fluctuations corresponding to minutes. The error bars shown in Figure 2 represent the standard deviation across the repeated measurements. The relatively small SD in , compared with the larger variability observed in and , indicates that the characteristic oscillation period was more reproducible than the amplitudes of the frequency and resistance modulations under the present experimental conditions.

The observed dependence of the compressional-wave response on droplet volume can be understood from the evolution of the effective acoustic cavity formed between the QCM surface and the liquid–air interface. At lower volumes, the thinner liquid layer produces a relatively slowly varying acoustic response, and the measured oscillations are dominated by the longer timescale modulation with . As the volume increases, the thicker liquid layer provides a larger acoustic path length and changes the phase condition of the reflected longitudinal waves. The resulting change in the interference pattern produces stronger temporal modulation of the effective compressional-wave impedance, manifested experimentally as increasingly rapid fluctuations. Thus, the reduction in the observed characteristic period from approximately 40 min to approximately 5 min should be interpreted as a change in the temporal modulation of the cavity-mediated acoustic loading rather than as the direct acoustic propagation time across the liquid layer.

An increase in liquid volume does not necessarily correspond to a proportional increase in the effective cavity thickness, since the additional liquid may also spread laterally over the QCM surface. The present Fabry–Perot-type model presented in Equation 6 establishes how the compressional-wave impedance depends on the cavity thickness through the phase factor , while the volume-dependent experiments establish the corresponding experimental trend. A fully quantitative comparison between volume and S would require independent characterization of the liquid geometry and acoustic path length for each loading condition, which is beyond the scope of the present study.

RPMI-1640 displays a similar trend for all the parameters. Water exhibits the most pronounced compressional-wave behavior in comparison to DMEM and RPMI-1640. Across all volumes, and remain the largest, while and are consistently the shortest. Also, the high-frequency fluctuations of minutes are observed at 800 L itself. In summary, Figure 2 shows that compressional-wave behavior is governed by the interplay between fluid properties and droplet geometry. Larger droplets exhibit more pronounced high-frequency fluctuations, reflecting the stronger influence of cavity-mediated compressional-wave interference.

The comparatively larger and observed for water indicate a stronger modulation of both the effective mass-loading and dissipative response of the resonator. The shorter and further indicate that the acoustic loading evolves more rapidly in water under the present experimental conditions. In contrast, the dissolved salts, nutrients, proteins, and other constituents present in DMEM and RPMI-1640 modify their effective density, viscosity, acoustic impedance, and wave attenuation relative to water. These differences can alter the strength and phase of the reflected longitudinal waves and consequently the resulting cavity interference. Therefore, the observed differences among the three liquids are not solely a consequence of droplet volume but arise from the coupled influence of fluid properties and cavity geometry.

5 Conclusion

This study demonstrates that compressional acoustic-wave effects contribute significantly to QCM responses in liquid-phase measurements, including commonly used breast cancer cell-culture media such as DMEM and RPMI-1640. The observed dependence of the QCM response on liquid volume and fluid composition highlights the importance of considering the coupled effects of acoustic cavity geometry, fluid properties, and longitudinal-wave interference when interpreting frequency and resistance changes. These findings establish an important baseline for QCM-based measurements involving breast cancer cells, where compressional-wave contributions could otherwise be misinterpreted as cell-induced changes. Future experiments will therefore incorporate the identified compressional-wave response as a baseline correction when investigating breast cancer cell adhesion, growth, and other cellular dynamics, improving the reliability and interpretation of QCM-based biological measurements (Yılmaz et al., 2021; Adel et al., 2024; Izzo et al., 2026).

Statements

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Author contributions

HK: Formal Analysis, Investigation, Methodology, Software, Validation, Writing – original draft. RB: Writing – original draft. TG: Writing – original draft. AM: Methodology, Writing – original draft. MD: Resources, Supervision, Writing – review and editing. AG: Conceptualization, Funding acquisition, Project administration, Supervision, Writing – review and editing.

Funding

The author(s) declared that financial support was received for this work and/or its publication. The authors sincerely thank the School of Interdisciplinary Research and Entrepreneurship (SIRE), BITS Pilani for providing the funding support for this project through its SPARKLE program.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author MD declared that they were an editorial board member of Frontiers at the time of submission. This had no impact on the peer review process and the final decision.

Generative AI statement

The author(s) declared that generative AI was used in the creation of this manuscript. The authors declare the use of generative artificial intelligence (AI) and AI-assisted technologies to improve readability, grammar and spelling check in the writing process. The authors carefully reviewed and edited the results produced by AI and take full responsibility for the content of the publication.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fsens.2026.1949532/full#supplementary-material

References

  • 1

    AdelM.AllamA.SayourA. E.RagaiH. F.UmezuS.Fath El-BabA. M. (2024). Design and development of a portable low-cost QCM-based system for liquid biosensing. Biomed. Microdevices26 (1), 11. 10.1007/s10544-024-00696-0

  • 2

    AlanaziN.AlmutairiM.AlodhaybA. N. (2023). A review of quartz crystal microbalance for chemical and biological sensing applications. Sens. Imaging24 (1), 10. 10.1007/s11220-023-00413-w

  • 3

    CarliF.FortA.MichelettiF.MugnainiM.PanzardiE.VignoliV. (2024). A smart low-cost QCM-D based on lightweight frequency domain processing. IEEE Trans. Instrum. Meas.73, 1–13. 10.1109/tim.2023.3348895

  • 4

    ChenQ.YaoY.AoJ.YuX.WuD.ShouM.et al (2025). Advances in quartz crystal microbalance relative humidity sensors: a review. Measurement243, 116415. 10.1016/j.measurement.2024.116415

  • 5

    EshunG. B.CrapoH. A.YazganI.CronmillerL.SadikO. A. (2023). Sugar–lectin interactions for direct and selective detection of Escherichia coli bacteria using QCM biosensor. Biosensors13 (3), 337. 10.3390/bios13030337

  • 6

    FortA.LandiE.GrassoA. L.MugnainiM.PanzardiE.VaccarellaP.et al (2022). QCM measurements in Newtonian liquids: problems and performance analysis. IEEE Trans. Instrum. Meas.71, 1–13. 10.1109/tim.2022.3154837

  • 7

    GanesanA.RajendranG.ErcoleA.SeshiaA. A. (2016). Multifrequency acoustics as a probe of mesoscopic blood coagulation dynamics. Appl. Phys. Lett.109 (6), 063701. 10.1063/1.4960978

  • 8

    GoudaM.GhazzawyH. S.AlqahtaniN.LiX. (2023). The recent development of acoustic sensors as effective chemical detecting tools for biological cells and their bioactivities. Molecules28 (12), 4855. 10.3390/molecules28124855

  • 9

    IzzoF.MazumderA.De GannesF. P.HallilH. (2026). Gold nanoparticle–assisted multiharmonic and multivariate QCM sensing of miRNA-21. IEEE Sensors J. 26 (9), 12881–12888. 10.1109/JSEN.2026.3673862

  • 10

    JandasP. J.PrabakaranK.LuoJ.DH. M. G. (2021). Effective utilization of quartz crystal microbalance as a tool for biosensing applications. Sensors Actuators A Phys.331, 113020. 10.1016/j.sna.2021.113020

  • 11

    JohannsmannD. (2015). The quartz crystal microbalance in soft matter research. Soft Biol. Matter1, 143–168. 10.1007/978-3-319-07836-6

  • 12

    KandelR. (2026). BPS2026–Probing cytoskeletal dynamics on tunable heparin/collagen bio-interfaces using QCM-D. Biophysical J.125 (4), 414a. 10.1016/j.bpj.2025.11.2530

  • 13

    KowalczykA.Gajda-WalczakA.Ruzycka-AyoushM.TargonskaA.MosieniakG.GlogowskiM.et al (2023). Parallel SPR and QCM-D quantitative analysis of CD9, CD63, and CD81 tetraspanins: a simple and sensitive way to determine the concentration of extracellular vesicles isolated from human lung cancer cells. Anal. Chem.95 (25), 9520–9530. 10.1021/acs.analchem.3c00772

  • 14

    KowarschR.SuhakY.EduarteL. C.MansourM.MeyerF.PeschelA.et al (2020). Compressional-wave effects in the operation of a quartz crystal microbalance in liquids: dependence on overtone order. Sensors20 (9), 2535. 10.3390/s20092535

  • 15

    LimH. J.SahaT.TeyB. T.TanW. S.OoiC. W. (2020). Quartz crystal microbalance-based biosensors as rapid diagnostic devices for infectious diseases. Biosens. Bioelectron.168, 112513. 10.1016/j.bios.2020.112513

  • 16

    McKennaL.NewtonM. I.McHaleG.LucklumR.SchroederJ. (2001). Compressional acoustic wave generation in microdroplets of water in contact with quartz crystal resonators. J. Appl. Phys.89 (1), 676–680. 10.1063/1.1331337

  • 17

    Murrieta-RicoF. N.PetranovskiiV.GalvánD. H.Antúnez-GarcíaJ.SergiyenkoO.LindnerL.et al (2022). Basic aspects in the application of QCMs as sensors: a tutorial. IEEE Sensors J.22 (11), 10163–10172. 10.1109/jsen.2022.3148039

  • 18

    Özdabak SertA. B.BittrichE.UhlmannP.KokF. N.KılıçA. (2023). Monitoring cell adhesion on polycaprolactone–chitosan films with varying blend ratios by quartz crystal microbalance with dissipation. ACS Omega8 (19), 17017–17027. 10.1021/acsomega.3c01055

  • 19

    PalleschiS.SilvestroniL.RossiB.DinarelliS.MagiM.GiacomelliL.et al (2024). Simple and sensitive method for in vitro monitoring of red blood cell viscoelasticity by quartz crystal microbalance with dissipation monitoring (QCM-D). Biosens. Bioelectron. X21, 100554. 10.1016/j.biosx.2024.100554

  • 20

    RogalaA.TiainenH.Zaytseva-ZotovaD. (2025). Using QCM-D for real-time analysis of cell adhesion dynamics at biointerfaces. Adv. Mater. Interfaces12 (9), 2400811. 10.1002/admi.202400811

  • 21

    SauerbreyG. (1959). Use of quartz vibration for weighing thin films on a microbalance. Z. für Phys.155 (2), 206–222. 10.1007/BF01337937

  • 22

    VoinovaM. V.RodahlM.JonsonM.KasemoB. (1999). Viscoelastic acoustic response of layered polymer films at fluid–solid interfaces: continuum mechanics approach. Phys. Scr.59 (5), 391–396. 10.1238/physica.regular.059a00391

  • 23

    WangL.SongJ.YuC. (2025). Recent progress on mass-sensitive gas sensors for environmental and industrial applications. Measurement249, 117039. 10.1016/j.measurement.2025.117039

  • 24

    YılmazM.BakhshpourM.GöktürkI.PişkinA. K.DenizliA. (2021). Quartz crystal microbalance (QCM) based biosensor functionalized by HER2/Neu antibody for breast cancer cell detection. Chemosensors9 (4), 80. 10.3390/chemosensors9040080

  • 25

    ZhuangH.LeeH. P.LimS. P. (2010). “Dynamic response of quartz crystal microbalances in contact with silicone oil droplets,”, in Proc. Fourth Int. Conf. Exp. Mechx. SPIE, 7522, 834–841. 10.1117/12.848916

Summary

Keywords

biosensing, breast cancer cells, cell culture media, compressional acoustic waves, quartz crystal microbalance

Citation

Kannan H, Babu RP, Ghosh T, Mohapatra A, Dutta M and Ganesan A (2026) Analysis of compressional acoustic wave responses in cell culture media for reliable QCM-based breast cancer cell monitoring. Front. Sens. 7:1949532. doi: 10.3389/fsens.2026.1949532

Received

27 July 2026

Revised

02 September 2026

Accepted

09 September 2026

Published

20 October 2026

Volume

7 - 2026

Edited by

Arvind Mukundan, Sanjivani University, India

Reviewed by

Mostafa Gouda, National Research Centre, Egypt

Gabriel Guillermo Muñoz, National University of Entre Ríos, Argentina

Andrea Bettucci, Sapienza University of Rome, Italy

Updates

Copyright

*Correspondence: Adarsh Ganesan,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics