Abstract
The senor quality on pathogenic bacteria monitoring employing Tellurium Dioxide (TeO2)/Cadmium Gallium Sulfide (CdGa2S4)-assisted surface plasmonic resonance (SPR) refractive index (RI) sensor. This proposed design is used the Kretschmann configuration with silver (Ag) for exciting the surface plasmons. A wavelength of 633 nm is used to analyze the sensor’s performance exploiting the transfer matrix method along with angular interrogation technique. Initially, we show the optimization Ag’s thickness by analyzing the better minimum reflectance and sensitivity performances. Later, the structure optimization is also exhibited by comparing the proposed structure’s performance with others that are made-up with the defined layers. Finally, the performance analysis was investigated for different 2D materials and found that maximum attained parameters are sensitivity of 285.538 , quality factor of 144.152 , figure of merit of 134.875 and combined sensitivity factor of 133.896. These results have a significant improvement as compared with the existing work. Furthermore, a machine learning framework based on the Random Forest regression algorithm was developed to predict the sensor sensitivity and evaluate the influence of the structural parameters on the sensing performance.
1 Introduction
Pathogenic bacteria represent a significant risk to public health, food safety and sustainability and environmental quality as they are capable of making rapid exponential progress when allowed to multiply within host organisms leading them down the path towards infection (). The presence of pathogenic microorganisms in drinking water, food products, and clinical specimens is a worldwide concern, creating an increasing need for rapid, reliable, and highly sensitive detection methods. Early detection of bacterial contamination is crucial not only to combat the spread of disease but also to ensure food safety and access to timely medical care (). Thus, the need to create sensitive biosensing technologies that can identify even small changes in infections due to pathogenic bacteria has become a major area of interest in biomedical diagnosis, environmental monitoring, and public health surveillance. Early bacterial detection not only minimizes disease transmission but also improves clinical diagnosis, food quality assurance, and environmental surveillance, thereby supporting effective public health management. Various traditional methods, such as “microbial culture, polymerase chain reaction (PCR) (), enzyme-linked immunosorbent assay (ELISA) (), fluorescence-based assays (), and electrochemical approaches” have been applied for the detection of pathogenic bacteria (). Although these approaches are typically accurate and reliable, they usually require complex instrumentation, skilled personnel, lengthy sample preparation, which confines their applicability for rapid and on-site detection. Such limitations have stimulated the search for optical-based sensing technologies that allow simultaneous, label-free, real-time, and extremely sensitive detection of cancer analytes with reduced analysis time and operational complexity ().
Among the all optical sensing techniques, surface plasmon resonance (SPR) is considered to be one of the most sensitive systems for chemical and biological detection due to its rapid, label-free, real-time analysis ability (). The most widely adopted SPR sensing scheme is the Kretschmann configuration, where a thin metallic film is deposited on the base of a high-refractive-index prism. Under p-polarized illumination, surface plasmons are generated at the metal-dielectric (MD) interface, when the in-plane component of the incident light matches the propagation constant of the surface plasmon wave (SPW) (; ; ). Any variation in the RI of the SM alters this phase-matching condition, resulting in a measurable shift in the resonance angle or the reflected light intensity (). These special properties make SPR one of the most vital sensing approaches for numerous different areas along with biomedical diagnostics, environmental monitoring and food quality assessment (). The excited SPW propagates along the MD interface while its associated evanescent field decays exponentially away from the surface into the sensing medium. At the resonance condition, efficient energy transfer from the incident light to the SPW produces a distinct minimum in the reflectance curve (). Since the resonance condition is highly dependent on the refractive index of the adjacent analyte, even minute refractive index variations produce a noticeable displacement of the resonance dip, thereby enabling highly sensitive detection (). The overall performance of an SPR biosensor is governed not only by the sensing mechanism but also by the optical characteristics of the constituent materials. Among the noble metals, silver (Ag) is extensively utilized as the plasmonic layer because of its lower optical absorption losses, narrower resonance curve, and stronger confinement of the electromagnetic field compared with many alternative plasmonic metals (). The impressive properties associated with Ag lead to larger figures of merit for sensitivity and total detection accuracy which makes them the material of choice in high-performance SPR biosensor development (). However, the inherent oxidation sensitivity of Ag and its weak interaction with target analytes require the introduction of dielectric and two-dimensional (2D) materials to improve sensor performance. High RI dielectric materials, such as tellurium dioxide (TeO2) (), improve the quarantine of electromagnetic fields and enhance plasmonic coupling, whereas cadmium gallium sulfide (CdGa2S4) can be employed for the detection of pathogenic bacteria due to its suitable optical properties and high refractive index, additionally provides stability of resonance as well as efficiency for sensing (; ). Moreover, aluminum antimonide (AlSb), a 2D material with superior optical and electronic performance, offers an ultra-high surface area and good light-matter interaction characteristics that contribute to the high analyte adsorption capacity and refractive-index sensitivity. Synergistically using these materials is an effective way to enhance the comprehensive performance of multilayer SPR biosensors ().
Recent advances in plasmonic biosensors have demonstrated that the sensing performance of traditional SPR structures can be significantly improved by incorporating dielectric materials and two-dimensional (2D) nanomaterials. Various multilayer architectures have been investigated to increase the interaction of the evanescent electromagnetic field and sensing medium, which in turn increases sensitivity, resonance sharpness, and detection accuracy. High RI dielectric layers, transition metal dichalcogenides (TMDCs), graphene family materials, MXenes, and other newly emerging nanomaterials have attracted significant interest owed to their excellent optical and electronic properties. Numerous studies have reported the successful incorporation of dielectric layers and 2D materials for various biosensing applications. For illustration, Islam et al. suggested an multilayer SPR biosensor with acquisition of dielectric nanolayers for the detection of basal cell carcinoma (BCC), yielding up to an order magnitude sensitivity increment after confined electromagnetic field (). Likewise, Janze et al. examined novel dielectric-assisted SPR structures with some up-to-date nanomaterials, and better resonance characteristics and detection performance have been achieved (). This research focuses on designing and optimizing an HfSe2-based SPR sensor integrated with machine learning for enhanced sensitivity and performance (). More recently, multilayer SPR sensors incorporating materials such as graphene, black phosphorus (BP), MoS2, WS2, MXenes, and antimonene have been successfully applied to the detection of cancer biomarkers, glucose, proteins, and environmental pollutants, confirming the effectiveness of advanced nanomaterials in enhancing plasmonic sensing performance. Surface plasmon resonance (SPR) biosensors that incorporate silicon dioxide (SiO2) and barium titanate (BaTiO3) function better mainly by controlling electromagnetic fields at the metal-dielectric interface (). Although these studies have demonstrated remarkable improvements in sensitivity, most reported sensor architectures primarily emphasize a single performance parameter, while comparatively limited attention has been devoted to achieving simultaneous optimization of multiple sensing characteristics, including “resonance angle, full width at half maximum (FWHM), quality factor (QF), figure of merit (FoM), signal-to-noise ratio (SNR), contrast sensitivity factor (CSF), and electric field enhancement” (; ). Furthermore, the combined utilization of TeO2, CdGa2S4, and AlSb within a single multilayer SPR platform for pathogenic bacteria detection has received very limited investigation (). Therefore, the development of an optimized multilayer architecture capable of simultaneously improving sensitivity, resonance quality, and overall sensing performance remains an active research challenge ().
Recently, machine learning (ML) techniques have attracted considerable attention for the analysis and optimization of SPR based optical sensors due to their capability to model complex nonlinear relationships between structural parameters and sensing performance (). By utilizing numerical or experimental datasets, ML algorithms can accurately predict sensor characteristics, identify the most influential design parameters, and significantly reduce the computational cost associated with repeated simulations (). In this work, a Random Forest regression model is integrated with the proposed SPR sensor to predict the sensor sensitivity, perform feature importance analysis, and support model-based optimization. The combination of numerical SPR analysis and machine learning provides an efficient framework for the rapid design and optimization of high-performance pathogenic bacteria sensors (). This study suggested a multilayer SPR biosensor utilizing the Kretschmann configuration for the sensitive detection of pathogenic bacteria. The sensor design includes a BAF10 prism, a 56 nm Ag film, dielectric layers of TeO2 and CdGa2S4, an AlSb nanomaterial layer, and the SM. The optical performance of the proposed architecture was theoretically investigated using the transfer matrix method (TMM) under angular interrogation at an operating wavelength of 633 nm. Initially, the Ag layer thickness was optimized to obtain a sharp resonance response, followed by a systematic investigation of the influence of TeO2, CdGa2S4, and AlSb on the sensing performance. The proposed sensor was comprehensively evaluated using the resonance angle, sensitivity, FWHM, QF, FoM, SNR, CSF, and electric-field distribution. Finally, the sensing performance was equated with those of newly reported SPR biosensors to demonstrate the effectiveness and competitiveness of the proposed multilayer design. Furthermore, the generated simulation dataset provides a suitable foundation for future machine learning-assisted prediction and optimization of SPR biosensor performance.
2 Theoretical analysis and numerical modelling
2.1 Theoretical analysis
The optical characteristics of the suggested six-layer SPR biosensor were theoretically investigated via TMM, which is the most accurate analytical techniques used to study electromagnetic wave propagation through layered thin-film structures. The method accurately determines the reflectance characteristics of the SPR sensor by considering the RI and thickness of individual layer without introducing numerical approximations. The resonance characteristics were obtained by satisfying the electromagnetic boundary circumstances at every interface of the multilayer configuration. Reflectivity () of the suggested biosensor was modeled with the TMM because it computationally efficient and can accurately solve multilayer structure problems without resorting to simplifying approximate methods. Let us consider an (L)-layer configuration described as layers made of uniform materials along the (z)-axis, each with thickness of () and dielectric constant (). Using the electromagnetic boundary conditions at each interface, the tangential electric and magnetic field components at () can be related to those at through the characteristic transfer matrix in Equation 1:where and characterize the tangential mechanisms of the electric and magnetic fields at the first interface of the multilayer structure. Similarly, and correspond to the field components. The overall characteristic matrix establishes the relationship between these field quantities throughout the multilayer SPR configuration for p-polarized incident light, it can be calculated by using Equation 2.
The transverse refractive index associated with each layer is calculated using Equation 3. The propagation constant of the layer, which administers electromagnetic wave propagation through the multilayer structure, is determined by Equation 4. Furthermore, Equation 5 define the wave impedance () and Equation 6 describe the propagation angle () of the layer, respectively. Here, is the incident light wavelength, is the prism’s RI, is the incident angle and is the magnetic permeability of layer ().
Using the obtained characteristic matrix, the reflection coefficient of the SPR heterostructure is evaluated using Fresnel’s formulation, as expressed in Equation 7. Finally, Equations 7, 8 represent the mathematical expressions of Fresnel’s mode analysis for calculating the and reflection coefficient () of an incident light, respectively ().
The sensitivity of the proposed sensor is defined as the ratio of change in SPR resonance angle () to the corresponding change in the RI () making it one of the most important performance parameters for evaluating sensor performance. This high sensitivity performance of the sensor with minimal change in analyte concentration indicates sensor ability to detect very small structural differences, thus better sensing performance. In addition, the performances of QF, FoM, and CSF are calculated, which demonstrates the quality of the sensor in this investigation.
Mathematical expressions for computing the aforementioned parameters are presented in Equations 9–12, where FWHM is the full width half maximum and is the maximum reflectivity in the reflectance plots (). The QF and FoM characterizes the product quality, noise contribution in a sensor, respectively report (). Lastly, the overall performance of the sensor is research using the CSF.
2.2 Numerical modelling
The numerical modeling of the planned SPR biosensor is illustrated in Figure 1. The multilayer structure was designed by the Kretschmann configuration, analyzed at an operating wavelength of 633 nm using the TMM. The sensor comprises six functional layers: the prism, metal layer, dielectric layers, two-dimensional (2D) material, and sensing medium. The RI’s and thicknesses of all constituent layers were selected from the reported literature and maintained constant throughout the analysis, unless otherwise specified. The BK7 prism served as an ideal first layer due to its high optical transparency, stable refractive index, and ability to efficiently couple the incident p-polarized light into the plasmonic interface. It has a RI of 1.515 (), thus exhibiting the required phase-matching condition for excitation of the SPWs. The second layer comprised Ag through a compound RI of 0.056266 + 4.27761i, where Ag was selected as the plasmonic metal due to its low optical loss, high electrical conductivity, and capability to generate a sharp and well-defined SPR resonance curve. The thickness of the Ag film significantly affected both plasmon excitation and resonance characteristics; thus, it was optimized using an iterative procedure before conducting further performance analysis.
FIGURE 1
Two dielectric layers were integrated over the Ag film to enhance the sensing features. The introduction of a 1 nm layer of Tellurium Dioxide (TeO2) with refractive index of 2.2596 serves to both confine the electromagnetic field and protect the Ag surface from oxidation at the same time. Subsequently, a 1 nm CdGa2S4 with a refractive index of 2.4482 was deposited to enhance the plasmonic coupling and further broaden the interface between the evanescent field and the sensing medium. Collectively, these dielectric layers improve resonance stability and enhance the sensing performance. The fifth layer was AlSb, which possesses a complex RI of 3.8638 + 0.0035982i and a thinness of 0.53 nm (). AlSb was selected because of its excellent optical and electronic properties, which promote strong light-matter interaction and increase the effective sensing area available for biomolecular adsorption. The sensing medium was arranged above the AlSb layer so that the changes in refractive indexes due to pathogenic bacteria could alter the resonance conditions of the SPR sensor. The RI of the sensing medium was varied from 1.371 to 1.410 was investigate different analyte conditions in this study. For a comprehensive performance evaluation, the proposed multilayer architecture was compared with four additional SPR configurations consisting of different layer combinations. All structures were analyzed under identical simulation conditions using the same theoretical framework. The sensing performance was subsequently evaluated in terms of resonance angle, Rmin, Sensitivity, FWHM, QF, FoM, SNR, and CSF, enabling a fair comparison of the suggested design with present configurations.
3 Results and discussion
3.1 Ag thickness optimization
The thickness of the plasmonic metal plays a critical role in defining the sensing performance of an SPR biosensor, as it directly influences the excitation efficiency of surface plasmon waves (SPWs) and the resonance characteristics. Thus, before studying the proposed multilayer architecture, the thickness of the Ag film was optimized in the range 40–60 nm with a step size equal to 5 nm (), whereas keeping the remaining layer thickness constant. The optimization was performed by studying the reflectance curves, resonance angle, and minimum reflectance (Rmin) in two refractive-index environments of sensing medium (RI = 1.371 and RI = 1.410). The corresponding reflectance responses are illustrated in Figures 2a,b. The resonance curves shown in Figure 2a show that the plasmonic coupling is gradually enhanced with increasing Ag thickness from 40 nm to 55 nm. Rmin drops from 0.34442 to 0.002464, indicating the energy transmission from incident light to SPWs using better efficiency. Simultaneously, the resonance angle exhibits a slight shift, confirming that the optical coupling condition is progressively optimized with increasing Ag thickness. Nevertheless, for an Ag thickness of 60 nm, the minimum reflectance increases again owing to the contribution of further optical losses by the thicker metal film, while limiting coupling between incident light and plasmonic mode. The corresponding plot for RI = 1.410 is shown in Figure 2b.
FIGURE 2
With the increase of Ag thickness, the resonance angle increasingly goes toward a higher value and the reflectance minimum continuously decreases at first, then increases after reaching an optimum thickness. This behavior suggests that an extremely thick Ag layer suppresses the dispersion of the evanescent field into the sensing medium, thereby weakening the interface between the electromagnetic field and analyte molecules. As a result, the resonance dip was larger and less effective for sensing applications. Table 1 summarizes the optimized values obtained from the Ag thickness analysis. Among all the investigated thicknesses, the 55 nm Ag film produced the lowest Rmin of 0.002464, along with a well-defined resonance response, indicating the strongest plasmonic coupling and the highest energy transfer efficiency. Therefore, an Ag thickness of 55 nm was selected as the optimum value for all subsequent simulations and performance analyses. The optimized metallic layer induced an appropriate trade-off among the confinement of the electromagnetic field and reduced propagation losses, which enhanced the overall sensing characteristics of the planned plasmonic biosensor.
TABLE 1
| Ag thickness (nm) | Rmin at SM’s RI of 1.371 | at SM’s RI of 1.371 | Rmin at SM’s RI of 1.41 | at SM’s RI of 1.41 | Sensitivity () |
|---|---|---|---|---|---|
| 40 | 0.34442 | 75.034 | 0.16499 | 83.6827 | 221.74 |
| 45 | 0.16818 | 75.16 | 0.014093 | 84.5155 | 239.87 |
| 50 | 0.040779 | 75.226 | 0.035169 | 85.2843 | 257.89 |
| 55 | 0.002464 | 75.29 | 0.20564 | 85.7968 | 269.23 |
| 60 | 0.062693 | 75.35 | 0.42302 | 86.2452 | 279.23 |
Rmin and resonance angle of different Ag thicknesses at SM’s RI of 1.371 nm and 1.41 nm.
3.2 Performance analysis of proposed SPR
The sensing performance of the proposed multilayer SPR biosensor was examined with its reflectance characteristics and compared against four reference configurations (S1–S4), which are illustrated in Figures 3a,b. The investigated structures are listed in Table 2, where each configuration incorporates different material combinations to examine the individual contributions of the Ag, TeO2, CdGa2S4, and AlSb layers. The comparison was performed for sensing-medium RI of 1.371 and 1.410, representing the refractive-index variation associated with pathogenic bacteria detection. Figure 3a shows the reflectance responses corresponding to a refractive index of 1.371, whereas Figure 3b presents the responses after increasing the refractive index to 1.410. A progressive change in the resonance angle was observed as additional dielectric and nanomaterial layers were incorporated into the sensor architecture. Among all the investigated configurations, the proposed structure (S5) exhibited the largest resonance-angle shift, along with a sharp and well-defined resonance dip. This phenomenon demonstrates plamonic coupling and evanescence electromagnetic field having the enchanced interaction with sensing medium. TeO2 and CdGa2S4 not only improve the electromagnetic field confinement but also give additional enhancive of light–matter interaction on AlSb layer because they exhibit better sensing properties than other structures.
FIGURE 3
TABLE 2
| Layer | Refractive index (RI) | Thickness | References |
|---|---|---|---|
| BK7-Prism | 1.515 | - | () |
| Ag (Silver) | 0.056206 + 4.27761i | 55 nm | () |
| TeO2 (Titanium dioxide) | 2.2596 | 1 nm | () |
| CdGa2S4 (Cadmium gallium sulfide) | 2.4482 | 1 nm | () |
| AlSb (Aluminium antimonide) | 3.8638 + 0.0035982i | 0.53 nm | () |
| Sensing Medium | 1.371–1.41 | - |
Structural design parameters of the proposed SPR biosensor.
The material parameters utilized for the design of the multilayer SPR biosensor are listed in Table 3, and the structural configurations used for the performance comparison are presented in Table 4. Arrangement of the optimized sensor, including a BK7 prism, 55 nm Ag film, 1 nm TeO2 layer, 1 nm CdGa2S4 layer, and 0.53 nm AlSb layer. The optimized multilayer design is capable of concurrently exciting plasmon and confining the electromagnetic field, as well as providing sufficient interaction with the analyte, thus effectively improving the performance of the designed biosensor. The quantitative performance comparison of all five configurations is presented in Table 5. The conventional structure (S1) achieves a sensitivity of 185.620°/RIU, however the combination of AlSb (S2) of 203.68°/RIU, and dielectric layers (S3 and S4) increases the sensitivity to approximately 200°–203.68°/RIU. In contrast, the proposed multilayer configuration (S5) exhibits the highest sensitivity of 269.397°/RIU, representing an improvement of approximately 45.1%, 32.3%, 34.4%, and 32.3% compared with S1, S2, S3, and S4, respectively.
TABLE 3
| Structure no. | Structure | Combination |
|---|---|---|
| S1 | Conv | BK7/Ag/Analyte |
| S2 | AlSb-Based | BK7/Ag/AlSb/Analyte |
| S3 | TeO2-Based | BK7/Ag/TeO2/Analyte |
| S4 | CdGa2S4-Based | BK7/Ag/CdGa2S4/Analyte |
| S5 | Proposed | BK7/Ag/TeO2/CdGa2S4/AlSb/Analyte |
SPR configurations considered for the proposed biosensor.
TABLE 4
| Structure | Rmin at 1.37 | at 1.37 | at 1.41 | Sensitivity () | QF () | FoM () | FWHM | CSF | |
|---|---|---|---|---|---|---|---|---|---|
| S1 | 0.0027 | 72.791 | 80.031 | 7.239 | 185.620 | 169.81 | 169.348 | 1.0931 | 168.662 |
| S2 | 0.0008 | 73.496 | 81.440 | 7.943 | 203.689 | 167.24 | 167.140 | 1.2176 | 166.426 |
| S3 | 0.0012 | 73.496 | 81.312 | 7.815 | 200.405 | 168.23 | 168.032 | 1.1912 | 167.322 |
| S4 | 0.0005 | 73.560 | 81.504 | 7.943 | 203.689 | 168.75 | 168.543 | 1.2079 | 167.826 |
| S5 | 0.0024 | 75.290 | 85.796 | 10.506 | 269.397 | 181.04 | 180.600 | 1.488 | 179.679 |
Comparison of S, QF, FoM and CSF of Proposed Structure S5 with other four structures.
TABLE 5
| 2D material | Refractive index (RI) | Thickness |
|---|---|---|
| BP | 3.5 + 0.01i | 0.53 nm |
| MOS2 | 5.0947 + 1.2327i | 0.65 nm |
| Graphene | 3.8638 + 0.0035982i | 0.34 nm |
| WS2 | 4.55 + 0.433i | 0.70 nm |
The refractive index (RI) and thicknesses of 2D materials.
Moreover, the developed sensor has the highest QF (181.04 RIU−1), FoM (180.60 RIU−1) and CSF (179.679 RIU−1), which illustrates that the suggested architecture not only improves sensitivity, but also accomplishes a sharp resonance profile with excellent quality for sensing purposes. In summary, the results indicate that the addition of TeO2, CdGa2S4 and AlSb simultaneously gives synergistic improvement factors to both plasmonic coupling and electromagnetic field localization in order to improve the sensing performances of the suggested SPR biosensor. The developed multilayer design is well supported by the evidence of an improved response to resonance mode and performance parameters, indicating that it can help provide high sensitivity for pathogenic bacteria detection.
3.3 Comparative analysis of proposed SPR, sensor with 2D materials.
To investigate the impact of different two-dimensional (2D) materials on the sensing performance, the optimized multilayer SPR configuration (S5) is analyzed by replacing the sensing layer with BP, Graphene, MoS2, WS2, and AlSb while maintaining all remaining structural parameters unchanged. The RI’s and thicknesses of the investigated 2D materials are summarized in Table 5, whereas the corresponding reflectance characteristics are illustrated in Figures 4a,b. The comparison is carried out for sensing-medium RIs of 1.371 and 1.410 to evaluate the effect of each material on the resonance characteristics. Figures 4a,b show that the addition of different 2D materials leads to significant changes in the resonance angle, minimum reflectance. These differences originate from the distinct optical constants and light–matter interaction capabilities of each material, which directly influence the excitation efficiency of surface plasmon waves. Among all investigated materials, the AlSb-assisted configuration exhibits the largest resonance-angle shift together with a sharp and definite resonance dip, indicating efficient plasmonic coupling and stronger interaction between the evanescent field and the sensing medium. In contrast, Graphene, MoS2, BP, and WS2 also improve the sensor performance compared with the conventional structure, although their resonance characteristics are comparatively less pronounced. Table 5 shows the optical properties of selected 2D materials, while each material has a unique refractive index and thickness, their effects on sensor performance are not completely independent but rather interaction with optical absorption, electromagnetic field confinement, and analyte interaction. Higher complex dielectric indices of AlSb leads to a more intensified localization of the electromagnetic field at the sensing interface, providing denser resonant states contraction as well as higher detection efficiency. The quantitative comparison of the sensing performance is presented in Table 6.
FIGURE 4
TABLE 6
| 2D materials | Rmin at 1.37 | at 1.37 | at 1.41 | Sensitivity () | QF () | FoM () | FWHM | CSF | |
|---|---|---|---|---|---|---|---|---|---|
| AlSb | 0.0024641 | 75.2903 | 85.7968 | 10.5065 | 269.397 | 181.046 | 180.600 | 1.488 | 179.679 |
| BP | 0.00188 | 75.226 | 85.412 | 10.186 | 261.179 | 178.036 | 177.702 | 1.467 | 176.808 |
| Graphene | 0.041613 | 74.9059 | 84.2593 | 9.3534 | 239.8 | 141.06 | 135.268 | 1.699 | 134.45 |
| MoS2 | 0.22670 | 76.059 | 86.181 | 10.122 | 259.538 | 115.917 | 89.639 | 2.239 | 88.580 |
| WS2 | 0.06436 | 75.995 | 87.014 | 11.019 | 282.538 | 144.152 | 134.875 | 1.960 | 133.896 |
Comparison of sensing parameters of proposed structure S5 with other 2D materials.
Among all investigated materials, AlSb achieves the best overall performance with a sensitivity of 269.397°/RIU, a QF of 181.046 RIU−1, a FoM of 180.600 RIU−1, and a CSF of 179.679 RIU−1. In comparison, BP exhibits a sensitivity of 261.179°/RIU with a QF of 178.036 RIU−1, whereas Graphene achieves 241.461°/RIU with a comparatively lower FoM of 134.896 RIU−1. Likewise, MoS2 has a sensitivity of 259.538°/RIU, and WS2 shows the highest sensitivity among the other materials (282.538°/RIU), but its QF (144.152 RIU−1) and FoM (134.875 RIU−1) are also too low compared to those of TMDs, leading to greater resonance profiles for lower sensing quality. So even though the WS2 produces the biggest angular shift, AlSb is favored as a top sensing material by being able to provide the best combination of characteristics including high sensitivity, highly damped resonance and outstanding performance metrics. The comparative analysis substantiates that the low concentration presence of AlSb in the proposed multilayer SPR architecture delivers a more uniform sensing characteristics higher than the examined 2D materials. The high electromagnetic field confinement, more vigorous plasmonic coupling, and improved resonance characteristics guarantee a more sensitive quality factor and figure of merit performances of the designed sensor for highly sensitive detection of pathogenic bacteria.
3.2 Electric field intensity
Figure 5 illustrates the electric field intensity distribution of the proposed SPR sensor structure consisting of Ag/TeO2/CdGa2S4/AlSb in contact with the sensing medium. At the Ag-dielectric interface, surface plasmon waves are excited as a result of resonance between p-polarized light incident from above and the free electrons in the thin metallic film. In fact, the electromagnetic field penetrates into the sensing volume as an evanescent wave beyond this interface and with a decay exponent that depends on the distance between the metal boundary. Intermediate dielectric layers, such as TeO2, CdGa2S4 and AlSb help in modulating the electric field confinement to achieve an improvement of interaction of plasmonic field with analyte.
FIGURE 5
These layers allow well-organized coupling of light to the surface plasmon mode while reducing optical losses and enhancing field localization at the sensing interface. This means that even the slightest refractive index shifting in the medium of sensing causes a large deviation in the resonance condition, which results in higher sensitivity sensors. The penetration depth (PD) of the evanescent field is calculated as 448.43 nm, which means that the electric field penetrates about 448 nm into the sensing medium before its intensity decreases to one/e (≈36.8%) of its maximum value. This depth of penetration makes these polymers ideal for the detection of biomolecules, proteins, pesticides and other analytes whose interaction takes place in a few hundred nm thick region just below the sensor surface. The optimized penetration depth provides a compromise between sufficient field confinement and interaction volume to improve sensitivity and detection limits of the suggested SPR biosensor.
3.3 Limitations and future directions
Although the numerical sensing performance of the proposed sensor is promising, there are challenges that would need to be overcome before practical use. However, first, this study relies on numerical simulations and does not involve experimental fabrication or characterization. Thus, they only predict the sensing performance, and need experimental validation. Secondly, the bacteria-specific surface functionalization or molecular recognition are not considered in the existing model. Therefore, the simulated response is mainly due to refractive-index based sensing and selective detection of one specific bacterial species could only be accomplished with a suitable bioreceptor or functionalization layer. Resonance-angle shifts as well as measurement uncertainty are indicated due to temperature variations affecting both the refractive index of the sensing medium and the optical properties of the constituent materials. Likewise, fabrication tolerances, variations in layer thicknesses, surface roughness, imperfections at the interface and non-uniform coating can collectively lead to a deviation from an idealized numerical response. The absorption of the material and optical losses, in particular looking at the metallic and lossy functional layers, may also impact resonance depth, linewidth, and electric-field enhancement. The use of an idealized sensing environment in the numerical model is another major limitation in real biological samples, proteins, salts, cells and other biomolecules that generate nonspecific adsorption, background refractive-index fluctuations and matrix interference may also be presented. These effects may impair the selectivity, sensitivity and reproducibility of the sensor. Consequently, further work will be geared towards the experimental fabrication and characterization, inclusion of bacteria-directed surface functionalization details, evaluation against temperature and fabrication tolerance pathways (such as sterilization), quantification of losses in graded materials and composite constructs, as well as validation with relevant real-life bio samples.
3.4 Comparative study of SPR performance analysis
A comparison of the sensing performance of the suggested design with recently reported SPR biosensors is also conducted, and a summary of this assessment is depicted in Table 7 in order to evaluate its effectiveness. The comparison includes previously published multilayer SPR configurations developed for biological and chemical sensing applications using different dielectric materials and two-dimensional nanomaterials. The comparison primarily focuses on sensitivity, since it is the most widely adopted parameter for evaluating SPR sensor performance. The reported results indicate that the proposed multilayer SPR biosensor achieves competitive sensing performance compared with recently published designs. While several reported sensors show excellent sensitivities based on advanced nanomaterial incorporation, many of them compromise either the resonance linewidth or overall sensing performance. In contrast, the proposed architecture has dramatically boost high sensitivity with great quality factor, high figure of merit and improved combined sensitivity factor which gives a perfect sensing response and balanced at different input power levels. This performance boost is ascribed to the synergetic effect of Ag plasmonic layer, TeO2 dielectric layer, CdGa2S4 semiconductor layer and AlSb nanomaterial that completely enhances electromagnetic field confinement with improved interaction between evanescent field and sensing medium.
TABLE 7
| Ref. No | Technique used | Structure | Sensitivity () | QF () | FoM () |
|---|---|---|---|---|---|
| () | Prism | Prism-Ag-PtSe2-WS2 | 194 | 34.22 | 17.64 |
| () | Prism | Prism-Ag-AlAs-TIBr-BP | 475 | 80.37 | 79.72 |
| () | Prism | Prism-Au--BlueP/ | 234 | 390 | - |
| This Work | Prism | Prism-Ag -TeO2-CdGa2S4-AlSb | 269.39 | 181.04 | 180.6 |
Performance comparison of the proposed SPR biosensor with existing studies.
4 Machine learning-based prediction and optimization
4.1 Dataset preparation
To investigate the predictive capability of machine learning for the proposed SPR biosensor, a dataset was prepared using the numerical results obtained from the optimization study presented in the previous sections. Since the experimental investigation provides only a limited number of reported observations, an augmented dataset was generated within the experimentally investigated parameter ranges to facilitate supervised learning while preserving the physical relationships among the sensor parameters. The generated dataset includes the silver (Ag) layer thickness, refractive index (RI), resonance angle, minimum reflectance, full width at half maximum (FWHM), quality factor (QF), figure of merit (FoM), and combined sensitivity factor (CSF). The target variable selected for prediction was the sensor sensitivity, which represents the primary performance indicator of the proposed SPR biosensor.
4.2 Machine learning model development
To establish a predictive framework for estimating the sensitivity of the proposed surface plasmon resonance (SPR) biosensor, a supervised machine learning approach based on the Random Forest (RF) regression algorithm was employed. The RF model was selected due to its robustness against overfitting, capability to capture nonlinear relationships, and high predictive performance for complex engineering datasets. Furthermore, Random Forest provides an effective balance between prediction accuracy and computational efficiency, making it particularly suitable for SPR sensor optimization problems. The generated dataset was first preprocessed by selecting the most influential sensor parameters, including the silver (Ag) layer thickness, refractive index (RI), resonance angle, minimum reflectance, and the number of sensing layers, as the input features.
The sensor sensitivity was considered the target output variable. Subsequently, the dataset was randomly divided into training and testing subsets using an 80:20 ratio. The training dataset was utilized to construct the regression model, while the testing dataset was reserved for evaluating the predictive capability of the trained model on previously unseen samples. The Random Forest regressor was configured using 250 decision trees with a maximum tree depth of ten levels to achieve an appropriate compromise between model complexity and generalization capability. During the training stage, each tree was generated using bootstrap sampling, while the final prediction was obtained by averaging the outputs of all individual trees. This ensemble learning strategy significantly reduces prediction variance and enhances the robustness of the regression model. After training, the developed model was applied to predict the sensor sensitivity for the testing dataset. The predicted values were then compared with the corresponding actual values to evaluate the model’s prediction capability. Figure 6 illustrates the relationship between the experimentally generated target values and the values predicted by the Random Forest model. The proximity of the data points to the ideal prediction line demonstrates the high consistency between the predicted and actual sensitivities, indicating that the developed model successfully captures the nonlinear dependence between the SPR structural parameters and the sensor performance.
FIGURE 6
4.3 Model performance evaluation
The predictive performance of the developed Random Forest regression model was evaluated using three widely adopted statistical performance indicators, namely, the coefficient of determination (R2R^2R2), the root mean square error (RMSE), and the mean absolute error (MAE). The coefficient of determination (R2R^2R2) measures the proportion of variance in the target variable that is explained by the regression model, whereas RMSE and MAE quantify the average prediction error and provide complementary information regarding the model accuracy. A higher R2R^2R2 value together with lower RMSE and MAE values indicates superior predictive capability and improved model generalization. Following the training stage, the regression model was applied to the testing dataset to evaluate its prediction performance on previously unseen samples. The obtained performance metrics demonstrated excellent agreement between the predicted and target sensitivity values, confirming that the Random Forest algorithm successfully captured the nonlinear relationship between the structural parameters of the SPR biosensor and its sensing performance. The low prediction errors further indicate that the developed model provides stable and reliable estimates over the investigated operating range.
Figure 7 presents the residual error distribution of the proposed Random Forest model. It can be observed that most residual values are concentrated near zero with an approximately symmetric distribution. The absence of significant skewness or isolated large residuals suggests that the prediction errors are randomly distributed rather than resulting from systematic modeling bias. This behavior confirms the robustness of the regression model and demonstrates its capability to accurately estimate the sensitivity of the proposed SPR biosensor. Figure 8 illustrates the learning curve obtained during the model training process. The convergence of the training and validation R2R^2R2 scores with increasing training samples indicates that the model achieves a good balance between fitting accuracy and generalization performance. Moreover, the relatively small gap between the two curves confirms that the Random Forest model does not suffer from significant overfitting or underfitting. These results demonstrate that the trained model maintains stable predictive performance as additional training samples become available, thereby validating its suitability for data-driven optimization of SPR biosensor parameters.
FIGURE 7
FIGURE 8
4.4 Feature importance analysis
To further interpret the predictive capability of the developed Random Forest model, a feature importance analysis was performed to quantify the contribution of each input parameter toward the prediction of the SPR sensor sensitivity. Unlike conventional regression techniques, the Random Forest algorithm provides an intrinsic measure of feature importance based on the average reduction in prediction error achieved by each variable during the tree construction process, as presented in Figure 9. This analysis not only improves the interpretability of the developed machine learning model but also identifies the dominant physical parameters governing the sensing performance of the proposed SPR biosensor. The investigated input variables included the Ag layer thickness, resonance angle, refractive index (RI), full width at half maximum (FWHM), and minimum reflectance. Figure 8 presents the relative importance of these parameters as determined by the trained Random Forest model. It can be observed that the Ag layer thickness exhibits the highest contribution to the prediction process, indicating that it is the most influential parameter affecting the sensor sensitivity. This observation is consistent with the optimization results presented in Section 3, where variations in the Ag thickness significantly modified the plasmonic resonance characteristics and consequently enhanced the sensing performance. The resonance angle and refractive index also exhibit considerable influence on the prediction accuracy, reflecting their strong correlation with the excitation conditions of surface plasmon waves. Conversely, the contributions of FWHM and minimum reflectance are comparatively smaller, although they still provide valuable information for improving the overall predictive performance of the regression model.
FIGURE 9
These findings demonstrate that the Random Forest model successfully identifies the physically meaningful parameters governing SPR sensor behavior while providing additional insight into their relative significance. Therefore, the feature importance analysis confirms that the developed machine learning model is not only capable of accurately predicting sensor sensitivity but also serves as an effective tool for guiding the optimization of SPR biosensor design.
4.5 Model based optimization
The trained Random Forest regression model was further employed as a predictive optimization tool to identify the sensor configuration that maximizes the SPR sensitivity. Unlike conventional optimization methods, which typically require repeated numerical simulations or experimental investigations, the developed machine learning model enables rapid prediction of sensor performance over a wide range of design parameters with significantly reduced computational cost. Consequently, the proposed data-driven framework provides an efficient alternative for accelerating the optimization process of SPR biosensors. To investigate the influence of the silver (Ag) layer thickness on the sensing performance, the trained regression model was used to predict the sensor sensitivity while varying the Ag thickness within the investigated range of 40–60 nm. The remaining input parameters were maintained within their optimized operating conditions to isolate the effect of the Ag layer thickness. This procedure enabled the identification of the optimum design region corresponding to the maximum predicted sensitivity. Figure 10 illustrates the predicted sensitivity as a function of the Ag layer thickness obtained from the trained Random Forest model. The prediction results indicate that the sensor sensitivity increases progressively with increasing Ag thickness until reaching an optimum region near 55 nm, after which the improvement becomes marginal. This trend is in excellent agreement with the numerical optimization presented in Section 3, where the Ag thickness of approximately 55 nm produced the highest sensing performance.
FIGURE 10
The close agreement between the machine learning predictions and the conventional numerical analysis confirms the capability of the developed regression model to accurately reproduce the physical behavior of the proposed SPR biosensor. Overall, the model-based optimization demonstrates that machine learning can efficiently estimate the optimal structural parameters while () substantially reducing the computational effort associated with repeated electromagnetic simulations. Therefore, the proposed framework offers a practical tool for the rapid design and optimization of high-performance SPR biosensors.
5 Conclusion
A study on influence of employing Tellurium Dioxide (TeO2)/Cadmium Gallium Sulfide (CdGa2S4)-assisted structure in surface plasmonic resonance (SPR) sensor is presented in this article. In the first part, the optimization Ag’s thickness shows and found that Ag thickness of 55 nm offers a Rmin of 0.002464 compared to other thicknesses. In the second part, the proposed structure’s performance offers a highest sensitivity, which is 1.45, 1.32, 1.34, and 1.32 times higher than S1, S2, S3 and S4, respectively. In the detection process we use different 2D materials and improved sensing properties, that the maximum attained parameters, such as Sensitivity of 269.39 , QF of 181.04 , FoM of 180.6 and CSF of 179.67, are obtained using AlSb. These outcomes have a substantial improvement as compared with the existing work. Therefore, the proposed sensor can function as a very effective carrier for biological and biomedical applications and be utilized to identify pathogenic microorganisms. Furthermore, the proposed Random Forest model accurately predicted the sensor sensitivity and identified the Ag layer thickness as the most influential design parameter. The machine learning-based optimization showed good agreement with the numerical results, demonstrating that the proposed framework provides an efficient approach for the design and optimization of high-performance SPR biosensors.
Statements
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Author contributions
RM: Conceptualization, Data curation, Formal Analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft. SB: Conceptualization, Formal Analysis, Investigation, Methodology, Software, Writing – original draft. YV: Conceptualization, Funding acquisition, Methodology, Project administration, Resources, Supervision, Writing – review and editing. MP: Data curation, Project administration, Validation, Visualization, Writing – review and editing. SC: Data curation, Project administration, Validation, Visualization, Writing – review and editing. NA: Conceptualization, Formal Analysis, Funding acquisition, Software, Writing – review and editing. SK: Conceptualization, Formal Analysis, Funding acquisition, Software, Writing – review and editing. BK: Conceptualization, Funding acquisition, Methodology, Project administration, Resources, Supervision, Writing – review and editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This work was supported by Koneru Lakshmaiah Education Foundation, India.
Acknowledgments
The authors gratefully acknowledge the Koneru Lakshmaiah Education Foundation, India and the National Research Council Nepal for the equipment support.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The handling editor declared a past co-authorship with one of the authors BK.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AbbasA.LinmanM. J.ChengQ. (2011). Sensitivity comparison of surface plasmon resonance and plasmon-waveguide resonance biosensors. Sensors Actuators B Chem.156 (1), 169–175. 10.1016/j.snb.2011.04.008
2
AhamedS.VenkatesanK. K.JalaludeenS. A. (2025). A review on various surface plasmon resonance-based sensors. Plasmonics20 (8), 6869–6885. 10.1007/s11468-025-02837-4
3
BasakC.IslamM. S.HosainM. K.KouzaniA. Z. (2024). An ultra-sensitive surface plasmon resonance biosensor with PtSe2 and BlueP/WS2 heterostructure. Heliyon10 (19), e38499. 10.1016/j.heliyon.2024.e38499
4
BhartiA.MishraR.TiwariP.ShuklaS. M.SharmaV. (2025). Theoretical study of silicon carbide-based surface plasmon resonance sensor for detection of tuberculosis in blood plasma. J. Opt. 10.1007/s12596-025-02877-3
5
BodduS.VasimallaY.MishraM.MarquesC.SinghR.KumarS. (2026). Label-free SPR sensor with enhanced performance for Deltamethrin detection using potassium and Lithium Tantalates. Microchem. J.224, 117464. 10.1016/j.microc.2026.117464
6
Bodnar'I.Rud'V. Y.RudY. V. (2004). Growth and properties of CdGa2S4 single crystals. Inorg. Materials40 (2), 102–106. 10.1023/b:inma.0000016081.55152.47
7
BouandasH.KumarR.SinghS.BouazziY.SirohiK. (2025). Enhanced sensitivity and figure of merit in surface plasmon resonance biosensors with silicon dioxide and barium titanate materials. Microchem. J.212, 113341. 10.1016/j.microc.2025.113341
8
BouandasH.KumarR.WekalaoJ.GariaL.BenhalilibaM.SinghS.et al (2026). Machine learning-optimized hafnium diselenide-based surface plasmon resonance sensor. Surfaces Interfaces86, 108771. 10.1016/j.surfin.2026.108771
9
ChowdhuryS.Faisal AbdulrazakL.Akhtar MituS.AhmedK.BuiF. M.SmiraniL. K.et al (2023). A highly sensitive multi-channel SPR-PCF based biosensor with deep learning prediction approach. Alexandria Eng. J.77, 189–203. 10.1016/j.aej.2023.06.093
10
CostertonJ. W.StewartP. S.GreenbergE. P. (1999). Bacterial biofilms: a common cause of persistent infections. Science284 (5418), 1318–1322. 10.1126/science.284.5418.1318
11
FalkowS. (2004). Molecular Koch's postulates applied to bacterial pathogenicity—a personal recollection 15 years later. Nat. Rev. Microbiol.2 (1), 67–72. 10.1038/nrmicro799
12
FangX.ZhengY.DuanY.LiuY.ZhongW. (2018). Recent advances in design of fluorescence-based assays for high-throughput screening. Anal. Chemistry91 (1), 482–504. 10.1021/acs.analchem.8b05303
13
FuH.ZhangS.ChenH.WengJ. (2015). Graphene enhances the sensitivity of fiber-optic surface plasmon resonance biosensor. IEEE Sensors J.15 (10), 5478–5482. 10.1109/jsen.2015.2442276
14
HitomiK.MatsumotoM.MuroiO.ShojiT.HiratateY. (2001). Characterization of thallium bromide crystals for radiation detector applications. J. Crystal Growth225 (2-4), 129–133. 10.1016/s0022-0248(01)00834-x
15
HomolaJ. (2003). Present and future of surface plasmon resonance biosensors. Anal. Bioanalytical Chemistry377 (3), 528–539. 10.1007/s00216-003-2101-0
16
HomolaJ. (2008). Surface plasmon resonance sensors for detection of chemical and biological species. Chem. Reviews108 (2), 462–493. 10.1021/cr068107d
17
ImamK. T.TausifD. M. (2026). Differential evolution-optimized SPR biosensor for dengue virus detection. Opt. Lasers Eng.205, 109918. 10.1016/j.optlaseng.2026.109918
18
IslamS.DolonS. K.CheM. R.AhsanM. S.NakarmiB.SohnI.et al (2026). A noninvasive surface plasmon resonance sensor for early detection of basal cell carcinoma. Adv. Photonics Res.7 (5), e202500326. 10.1002/adpr.202500326
19
JanzeE. J. T.AlaeiS.MeshginqalamB. (2025). On the sensing performance improvement in SPR biosensor using ZnO and TMDCs architecture for cancer detection. Sci. Rep.15 (1), 28349. 10.1038/s41598-025-14131-8
20
JohnsonP. B.ChristyR. W. (1972). Optical constants of the noble metals. Phys. Rev. B6 (12), 4370–4379. 10.1103/physrevb.6.4370
21
KatoK.UmemuraN.PetrovV. (2017). Sellmeier and thermo-optic dispersion formulas for CdGa2S4 and their application to the nonlinear optics of Hg1−xCdxGa2S4. Opt. Commun.386, 49–52. 10.1016/j.optcom.2016.10.054
22
KumarR.PraveenP.BrahmaB.BarsocchiP.BhoiA. K. (2026a). Refractive index sensing-based sensitivity enhancement using surface plasmon resonance sensor with integration of tin diselenide and zirconium diselenide. Sensors26 (13), 4279. 10.3390/s26134279
23
KumarV. B.KumarP. S.TrivediM. (2026b). Emerging 2D materials beyond graphene for sensor technologies applications. Mater. Today Commun.53, 115421. 10.1016/j.mtcomm.2026.115421
24
KundanasriK.GurramkonduK. S.SanthoshC.VasimallaY.KumarS. (2026). High-performance plasmonic biosensor for blood urea detection using SiC/black phosphorus/AgBr hybrid layers. Discov. Electron.3 (1), 70. 10.1007/s44291-026-00223-w
25
MudgalN.SahariaA.AgarwalA.AliJ.YupapinP.SinghG. (2020). Modeling of highly sensitive surface plasmon resonance (SPR) sensor for urine glucose detection. Opt. and Quantum Electron.52 (6), 1. 10.1007/s11082-020-02427-0
26
MukhtarW. M.ShaariS.EhsanA. A.MenonP. S. (2014). Electro-optics interaction imaging in active plasmonic devices. Opt. Mater. Express4 (3), 424–433. 10.1364/ome.4.000424
27
OveraS.KoB. H.ZhaoY.JiaoF. (2022). Electrochemical approaches for CO2 conversion to chemicals: a journey toward practical applications. Accounts Chemical Research55 (5), 638–648. 10.1021/acs.accounts.1c00674
28
PalS.VermaA.SainiJ. P.PrajapatiY. K. (2019). Sensitivity enhancement using silicon‐black phosphorus‐TDMC coated surface plasmon resonance biosensor. Iet Optoelectron.13 (4), 196–201. 10.1049/iet-opt.2018.5023
29
PolyanskiyM. N. (2024). Refractiveindex.info database of optical constants. Sci. Data11 (1), 94. 10.1038/s41597-023-02898-2
30
RahmanM. T.UddinM. S.SultanaR.MoueA.SetuM. (2013). Polymerase chain reaction (PCR): a short review. Anwer Khan Mod. Med. Coll. J.4 (1), 30–36. 10.3329/akmmcj.v4i1.13682
31
RahmanM. M.RanaM. M.RahmanM. S.AnowerM.MollahM. A.PaulA. K. (2020). Sensitivity enhancement of SPR biosensors employing heterostructure of PtSe2 and 2D materials. Opt. Mater.107, 110123. 10.1016/j.optmat.2020.110123
32
ReddyP. S. K.ChebaaneS.VasimallaY.KhalifaS. B.AlsaifN. A. M.PuppalaR. S.et al (2026). Smart label-free SPR biosensing platform for hemoglobin and urine glucose detection via machine learning. Sci. Rep.16 (1), 20378. 10.1038/s41598-026-50144-7
33
SaidF. A.Menon P. S. Kalaivani T.MohamedM. A.Abedini A. Shaari S.et al (2015). “FDTD analysis of structured metallic nanohole films for LSPR-based biosensor,” in 2015 IEEE Regional Symposium on Micro and Nanoelectronics (RSM) (IEEE), 1–4.
34
SharmaV.DwivediL. K.SinghS.MishraG. R. (2024). Glucose level monitoring in human blood samples by surface plasmon resonance sensor using cerium oxide and black phosphorus nanomaterials. J. Opt.53 (4), 3683–3694. 10.1007/s12596-023-01597-w
35
SricharanP.VasimallaY.KumarS. (2026). Plasmon-enhanced black phosphorus–ZrO2–TeO2 multilayer SPR biosensor for ultra-sensitive early detection of brain tumors. Mod. Phys. Lett. B40, 2650037. 10.1142/s0217984926500375
36
TabatabaeiM. S.AhmedM. (2022). “Enzyme-linked immunosorbent assay (ELISA),” in Cancer Cell Biology: Methods and Protocols (Springer), 115–134.
37
UchidaN. (1971). Optical properties of single-crystal paratellurite (Te${\mathrm{O}}_{2}$). Phys. Rev. B4 (10), 3736–3745. 10.1103/physrevb.4.3736
38
VarmaN. S. S.VasimallaY.ChebaaneS.Ben KhalifaS.AlsaifN. A.PuppalaR. S.et al (2026). Surface and interface effects in X-titrate (X: lead and Strontium)/aluminium antimonide–assisted SPR refractive index sensor with machine learning analysis. Surfaces Interfaces87, 109029. 10.1016/j.surfin.2026.109029
39
VasimallaY.Shekhar PradhanH.Jashvantbhai PandyaR. (2020). SPR performance enhancement for DNA hybridization employing black phosphorus, silver, and silicon. Appl. Opt.59 (24), 7299–7307. 10.1364/AO.397452
40
YangD.ZhangZ.WeiF.LiS.LiuM.LuY. (2024). Comprehensive review on research status and progress in precision grinding and machining of BK7 glasses. Micromachines15 (8), 1021. 10.3390/mi15081021
41
YesudasuV.PradhanH. S. (2021). Performance enhancement of a novel surface plasmon resonance biosensor using thallium bromide. IEEE Trans. NanoBioscience21 (2), 206–215. 10.1109/TNB.2021.3114225
42
YesudasuV.PradhanH. S.PandyaR. J. (2021). Recent progress in surface plasmon resonance based sensors: a comprehensive review. Heliyon7 (3), e06321. 10.1016/j.heliyon.2021.e06321
Summary
Keywords
good health and wellbeing, cadmium gallium sulfide, Kretschmann configuration, machine learning, surface plasmon resonance, tellurium dioxide
Citation
Merigala R, Boddu S, Vasimalla Y, Prashanth MS, Chebaane S, Alsaif NAM, Khalifa SB and Karki B (2026) Pathogenic bacteria sensing with tellurium dioxide/cadmium gallium sulfide-assisted surface plasmonic resonance refractive index sensor: a machine learning approach. Front. Sens. 7:1962401. doi: 10.3389/fsens.2026.1962401
Received
08 August 2026
Revised
24 August 2026
Accepted
31 August 2026
Published
24 September 2026
Volume
7 - 2026
Edited by
Arun Uniyal, Uttarakhand Technical University, India
Reviewed by
Hiba Sai, University Ferhat Abbas of Setif, Algeria
Vipin Sharma, Rana Pratap Post Graduate College Sultanpur, India
Updates
Copyright
© 2026 Merigala, Boddu, Vasimalla, Prashanth, Chebaane, Alsaif, Khalifa and Karki.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Bhishma Karki, magnum.photon@gmail.com, bhishma.karki@trc.tu.edu.np
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.