Abstract
Seismocardiography (SCG) provides a non-invasive means of monitoring cardiac mechanical activity using wearable accelerometers. However, conventional SCG is typically recorded at a single chest location, limiting the spatial interpretation of the measured vibrations. In this study, we investigate whether multichannel SCG can be used to reconstruct the three-dimensional distribution of cardiac mechanical sources. We developed a physics-informed framework combining an electromechanical finite-element heart model, an attenuating transverse-shear radiation model, and a three-dimensional inverse reconstruction pipeline. Synthetic multichannel SCG signals without valve transients (hence not clinical SCG) were reconstructed using delay-and-sum (DAS), minimum variance distortionless response (MVDR), CLEAN, and sparse Bayesian learning (SBL). Three of the four recovered both position and shape on at least one array geometry. Our sparse Bayesian implementation localized the source but recovered no volume, collapsing onto a few voxels; thus, we report it as an approach we could not successfully adapt. Localization and robustness were evaluated against null controls and under varying noise levels, propagation uncertainties, and array geometries. With the source displaced 2 cm from the array center, a fully enclosing array, as a theoretical upper bound rather than a wearable configuration, recovered the centroid to 0.7–0.9 cm. On a more realizable one-sided array, the estimators separated: sparse deconvolution reached 1.0 cm and the adaptive beamformer 1.7 cm, both inside a 2.0 cm no-information baseline, while conventional delay-and-sum fell well outside it. Shape recovery cleared that baseline on both array geometries and collapsed when inter-channel phase was randomized. Reconstruction remained robust to measurement noise and moderate wave-speed mismatch but was more sensitive to geometric and registration errors. These findings indicate that multichannel seismocardiography, in principle, carries spatial information sufficient to localize cardiac mechanical activity in three dimensions. However, on a one-sided array, that recovery depends on the estimator, and centroid accuracy there requires an unmodeled local shear-speed contrast to stay within roughly 20– for the most tolerant estimator, and about for the adaptive one. Shape recovery is the more robust of the two. Forward and inverse models share the same physics, a limit that a structural mismatch test bounds but does not remove.
1 Introduction
Seismocardiography (SCG) records the subtle vibrations of the chest wall generated by the mechanical activity of the heart and has emerged as a promising technique for unobtrusive and wearable cardiovascular monitoring (; ). In contrast to electrocardiography, which reflects the electrical activation of the myocardium, SCG captures mechanical events associated with myocardial contraction, valve motion, and blood ejection, thereby providing complementary information on cardiac function (; ; ). Despite these advantages, conventional SCG is typically acquired at a single sternal site. The measured waveform therefore represents a superposition of vibrations originating from different regions of the heart that propagate through heterogeneous thoracic tissues before reaching the sensor. As a result, characteristic waveform features cannot be uniquely assigned to their anatomical origin, limiting the physiological interpretation of SCG recordings (; ). Recent advances in wearable sensor technology have enabled multichannel SCG systems that can record the spatial propagation of cardiac vibrations (; ). Differences in wave arrival times and amplitudes across multiple sensors contain information about the underlying mechanical sources, transforming SCG into an inverse source localization problem conceptually related to electrocardiographic imaging () and array signal processing (). The spatial information arises from differences in wave arrival times, amplitudes, and polarization across multiple sensors, which together encode the location of the underlying mechanical sources. While multichannel SCG has primarily been used to investigate spatial vibration patterns (), realistic electromechanical heart models now provide an opportunity to evaluate inverse reconstruction algorithms under controlled conditions (; ).
In this work, we present a physics-informed framework for three-dimensional localization of cardiac mechanical sources from multichannel SCG. A published electromechanical whole-heart model serves as the source of myocardial acceleration, while an attenuating transverse-shear radiation model generates synthetic multichannel SCG recordings. Source reconstruction is performed using representative beamforming approaches, including delay-and-sum (DAS), minimum variance distortionless response (MVDR), CLEAN, and sparse Bayesian learning (SBL). Localization performance is assessed using spatial overlap, localization accuracy, and robustness against realistic perturbations, including measurement noise, wave-speed mismatch, array geometry, and sensor registration errors.
This work links electromechanical heart simulation, elastic wave propagation, multichannel seismocardiography, and volumetric inverse reconstruction into a single pipeline and scores the reconstruction against the same finite-element ground truth that generated the data. Rather than proposing a new reconstruction algorithm, this study investigates whether multichannel SCG contains sufficient spatial information to recover localized three-dimensional cardiac mechanical sources. The selected reconstruction methods therefore serve as representative inverse approaches for evaluating this hypothesis under identical conditions.
2 Materials and methods
2.1 Overview of the proposed framework
Figure 1 summarizes the proposed physics-informed framework for three-dimensional localization of cardiac mechanical sources from multichannel seismocardiography. An electromechanical whole-heart model generates physiologically realistic myocardial acceleration fields, which are propagated through an elastic medium using an attenuating scalar propagation model with a transverse projection, approximating shear-wave radiation, to synthesize multichannel SCG recordings. The mechanical activity is estimated using four representative inverse methods on the simulated SCG signals: DAS, MVDR, CLEAN, and SBL. Localization performance is subsequently evaluated using spatial overlap, localization accuracy, and robustness under realistic perturbations, including measurement noise, wave-speed mismatch, array geometry, and sensor registration errors.
FIGURE 1
2.2 Electromechanical heart model
Cardiac mechanical activity was simulated using the open electromechanical whole-heart model developed by . The model combines a physiologically detailed finite-element representation of the human heart with electrophysiological excitation, active contraction, and passive myocardial mechanics to reproduce realistic cardiac motion throughout the cardiac cycle. For the present study, the nodal acceleration vectors obtained from the finite-element simulation were used as the mechanical source distribution. These acceleration fields served as the ground truth for all forward simulations and subsequent inverse localization experiments. The complete finite-element node set is used both to drive the forward model and to build the ground truth. The published simulation output is used as provided, at its 1 ms sampling interval over one cycle. No finite-element solve was carried out for this study. The model provides physiologically plausible spatial and temporal distributions of myocardial motion, enabling systematic evaluation of source localization methods under controlled and reproducible conditions.
2.3 Forward wave propagation model
2.3.1 Configuration and grid definition
Physical parameters governing wave propagation are specified. At seismocardiographic frequencies (the broader 1–80 Hz SCG band), the compressional wavelength ( m/s (), giving m below 60 Hz) exceeds the thoracic dimensions by orders of magnitude and therefore carries no localizable phase across a chest-scale array. Only the shear-wave field, with cm-scale wavelengths, is spatially informative. The forward model is therefore driven by the shear component, with reference shear speed m/s at Hz, slightly below the 3.2–3.6 m/s reported for healthy myocardium (; ). Attenuation follows a Szabo power law with prefactor Np/m/ and exponent (). is set conservatively and treated as a swept parameter. The phase-velocity dispersion scales as through a Kramers–Kronig-consistent relation continuous at .
A 3D volumetric grid is generated centered on the heart with half-width cm. Two grid resolutions are used: a refined 10 mm spacing ( voxels) for the headline benchmark of Table 2, and a coarser 20 mm spacing ( voxels) for the perturbation sweeps of Section 3.4, where per-trial cost dominates. Sensor positions are defined by the virtual accelerometer array (64 three-axis sensors, i.e., 192 channels, on a single ellipsoidal chest shell, semi-axes m), and signal processing parameters are set: 2–60 Hz analysis band, short-time Fourier transform (STFT) windows of 125 ms with 75% overlap. A shorter window helps the snapshot-starved vector covariance : it makes the non-stationary source appear more instantaneous, increasing the usable rank, and drops the sub-8 Hz band the chest-scale aperture cannot resolve (Section 4). Reconstruction voxels within a 1 cm keep-out radius of any sensor are excluded from the imaging grid, so that voxels sitting essentially on top of a sensor cannot trigger the near-field singularity of the steering vector. On the 10-mm grid, this removes 190 of 9261 voxels for the sphere and 208 for the hemisphere, all of them adjacent to the shell and none within the source region.
2.3.2 FEM data extraction
The ParaView FEM database is loaded, and acceleration vectors and nodal positions are extracted for every timestep , giving arrays of shape with time samples and nodes. For the forward model, an optional subset of nodes is retained by farthest-point sampling: a purely geometric coverage subsampling that thins the mesh while preserving the spatial extent of the activation region, without biasing the selection toward the most energetic nodes while reducing the computational footprint. The ground-truth reference (both the amplitude-weighted centroid and the source silhouette of Phase 5) is always built from the full FEM node set, decoupled from whatever subset is fed to the forward model, so the reported error and overlap are measured against the true source distribution rather than the reconstructed model. The full finite-element mesh carries 30,973 nodes, and the forward model uses all of them. The subsample size matters: reducing the forward set to a 4000-node farthest-point subsample moves Dice-AUC by up to 0.09 and the half-power centroid by up to 1.1 cm, whereas 12,000 nodes agree with the complete mesh to within 0.3 cm on every method. We therefore subsample only for the 20 mm-grid analyses. Each node’s acceleration is weighted by , the tissue mass it represents, so that the sum forms an equivalent body-force distribution rather than a set of equally weighted radiators. The same weight scales the ground-truth strength map. With this weighting, the synthesized amplitude is on the milli-g scale reported for recorded sternal SCG (), which an unweighted nodal sum exceeds by a factor of . The nodal acceleration fields are used at the native finite-element output rate of 1 kHz. No temporal up-sampling is applied at any stage.
Only bins within enter the covariance, and removes everything above the array spatial-Nyquist limit of 34.9 Hz. The single cardiac cycle is replicated four times, each replicate carrying an independent additive perturbation at 1% of signal root mean square (RMS), to populate the snapshot set from which the spatial covariance is estimated. The cycle is periodic to within the seam it creates: the step across the join is a typical within-cycle step, so the replication introduces no discontinuity larger than the signal’s own sample-to-sample variation. The replication is a numerical device for covariance estimation and is not presented as a model of physiological beat-to-beat variability. Its measured effect on the usable rank is reported below. Its effect on the reconstruction is smaller still (20-mm grid, forward nodes): removing the replication entirely leaves the DAS and CLEAN centroids unchanged (1.04 cm and 0.86 cm on the sphere, 4.87 cm and 1.26 cm on the one-sided array) and their Dice-AUC unchanged to within 0.002. Varying the perturbation from 0 to of signal RMS changes neither. Only the adaptive estimator responds, as only it inverts the covariance: its one-sided centroid moves between 1.83 cm and 2.25 cm across both sweeps.
2.3.3 Forward propagation
Waves are propagated from each finite-element method (FEM) source node to each sensor via the shear-wave Green’s function. Each node retains its full three-component acceleration (world ), and each physical sensor is a three-axis accelerometer, so the arriving fi eld is a 3-vector (Equation 1):
where is the vector source spectrum and is the inverse Fourier transform. The propagation is governed by the scalar shear-wave Green’s function , which analytically models spherical spreading , phase delay with wave-speed dispersion, frequency-dependent attenuation according to the power law , and the near-field coupling term (). The near-field term is explicitly retained because of the close distance to the wave source: at the mean sensor range, cm, and the low edge of the usable band, , so the term is roughly half the far-field contribution in magnitude, although its effect on amplitude is about . Removing it entirely moves the hemisphere MVDR centroid from 1.69 cm to 1.89 cm and the CLEAN centroid from 1.40 cm to 2.17 cm (20-mm grid, so not directly comparable with Table 2), so the term is not neutral for either estimator on the one-sided array. Modeling true elastodynamics would require that the near-field couples both compressional (P) and shear (S) waves, making it not purely transverse. However, as a theoretical simplification, our simulation only models transverse waves. This is a full three-component (vector) radiation model rather than the shared-dominant-axis scalar reduction of a single Green’s function. It makes each physical sensor contribute three channels for an effective channel count and directly addresses whether the recovery depends on collapsing the source onto one axis. The compressional field is spatially unresolvable in this band because soft tissue compressional wave speeds ( m/s) result in very long wavelengths ( m below 60 Hz) that manifest as uniform whole-body translation rather than resolvable spatial gradients (). While P-waves are unresolvable for beamforming, they will act as coherent interference on real sensors, meaning our shear-only assumption is an idealized limitation. Propagation is evaluated in short-time Fourier-transformed windows with overlapped reconstruction, and a cumulative signal is summed over all sources for each sensor individually.
Additive white Gaussian measurement noise is added to reach an array-global signal-to-noise ratio (SNR) of 30 dB (noise referenced to the global clean-signal RMS), consistent with modern microelectromechanical systems (MEMS) accelerometers on the chest. Referencing to a single global RMS makes the per-channel ratio uneven, 12.5–35.0 dB across the array, a 22 dB spread about a median of 29.7 dB, and because the analyzed band carries only of the signal power, the effective in-band ratio at the nominal operating point is 26.8 dB rather than 30 dB.
2.4 Inverse source localization
Rather than proposing a new reconstruction algorithm, this work investigates whether multichannel seismocardiography contains sufficient spatial information for three-dimensional localization of cardiac mechanical activity.
The methods are chosen to span a broad range of inversion methods while remaining sufficiently inexpensive for the per-trial perturbation sweeps of Section 3.4. Implementation parameters are listed in Table 1.
TABLE 1
| Parameter | Value |
|---|---|
| Wave physics | |
| P-wave speed (soft tissue) | 1,540 m/s |
| Shear speed at Hz | 2.8 m/s |
| Attenuation /exponent | 0.10 Np/m//1.0 |
| Radiation/source model | transv. shear/3-vector |
| Noise and snapshots | |
| Sensor SNR (array-global) | 30 dB |
| Cycle replications | 4 |
| Additive replicate perturbation | 1% of RMS |
| STFT snapshots bins | (4 pass on the sphere) |
| Effective rank of | of |
| Diagonal loading (relative, ) | 0.05 |
| Grid and frequency | |
| Grid half-width | 0.10 m |
| Grid step, Table 2 | 0.010 m ( voxels) |
| Grid step, sweeps (Section 3.4) | 0.020 m ( voxels) |
| Bandpass – | 2–60 Hz |
| Effective band, sphere/one-sided | 8–35/8–60 Hz |
| STFT window/overlap | 125 ms/75% |
| Array and reconstruction | |
| Sensors axes | (192 ch) |
| Forward/ground-truth nodes | 30,973/all |
| CLEAN iters (headline/sweep); gain | 150/60; 0.1 |
| CLEAN restoring beam | 2.3 cm |
| Centroid power threshold | dB |
Physical and simulation parameters.
Each method operates on per-window spatial covariance matrices formed from short-time Fourier snapshots, which are 125 ms wide with 75% overlap. The methods return a 3D energy map over the reconstruction grid, normalized to peak power. Processing proceeds as follows:
Bandpass restriction: only STFT bins inside the primary SCG band (2–60 Hz) contribute to the covariance and to the frequency sum of the beamformer. This boundary spans the band in which cardiac mechanical energy is concentrated (). The effective band is narrower on both sides and is what the estimator actually uses: the 125 ms window sets an 8-Hz lower bin edge, and zeroes every bin above the array spatial-Nyquist limit , which for the nominal 64-sensor sphere is 34.9 Hz. All reported results for the full-sphere array therefore rest on 8–35 Hz. The denser one-sided array raises its own limit to 60.8 Hz, set by its closest sensor pair, and so uses the full 8–60 Hz band. Its mean and worst spacings place the limit nearer 47 Hz (Section 4.3).
Covariance estimation: for each frequency bin, the spatial covariance is the average of over the 100 short-time windows of the replicated cycle and is regularized by diagonal loading alone, with . This does not make well conditioned: measured over the analysis band, its entropy (participation) rank is of the channels, and replication buys little: quadrupling the number of replicated cycles raises it by 0.7, and a further quadrupling raises it by 0.2. At that rank, the question is whether the adaptive estimator is using the data at all or instead is relying on the diagonal loading. Sweeping upward drives the Capon algorithm toward its heavily-loaded limit, where , and the map degenerates to the no-information case (at , Dice-AUC 0.18, and a 2.00 cm centroid, against the uniform baseline’s 0.20 and 2.01 cm), whereas at the operating , it reaches Dice-AUC 0.43 and 1.11 cm (20-mm grid, forward nodes, so not directly comparable with Table 2). The estimator is therefore adaptive at the operating point, although the margin over its own degenerate limit is what the rank permits. The covariance is thus strongly rank-deficient throughout.
Spatial power map: DAS, MVDR, CLEAN, or SBL evaluates a source-power score at every voxel and accumulates it across the band into a single volume. CLEAN additionally convolves its restored point list with a fixed-width restoring beam of width set by the array Rayleigh limit (Table 1), not chosen freely. Scaling between and its derived value leaves the enclosing array centroid between 0.56 cm and 0.86 cm (20-mm grid, forward nodes), so the reported centroid is not an artifact of that choice. On the one-sided array, it is more exposed, degrading from 1.26 cm to 3.08 cm as the beam is doubled. Shape overlap responds most strongly (Dice-AUC 0.09 to 0.72 on the sphere), and a beam scores higher than the derived width on both geometries. We retain the derived width rather than the best-scoring one.
TABLE 2
| Array | Method | Headline | Supplementary | ||||
|---|---|---|---|---|---|---|---|
| Dice-AUC | Centr (cm) | Peak (cm) | HD95 (cm) | ||||
| Sphere | uniform baseline | 0.20 | 2.01 | – | – | – | – |
| | DAS | 0.48 | 0.67 | 1.00 | 0.55 | 3.16 | 0.81 |
| | MVDR | 0.43 | 0.88 | 1.00 | 0.80 | 1.41 | 0.73 |
| | CLEAN | 0.48 | 0.72 | 1.00 | 0.82 | 8.12 | 0.73 |
| | SBL | 0.01 | 1.22 | 1.41 | 0.01 | 6.67 | 0.05 |
| Hemisphere | uniform baseline | 0.20 | 2.01 | – | – | – | – |
| | DAS | 0.26 | 4.49 | 10.63 | 0.43 | 7.07 | 0.39 |
| | MVDR | 0.38 | 1.70 | 1.00 | 0.47 | 5.10 | 0.59 |
| | CLEAN | 0.44 | 1.01 | 2.00 | 0.56 | 7.35 | 0.56 |
| | SBL | 0.00 | 1.73 | 1.73 | 0.01 | 8.06 | 0.06 |
Per-method performance of the ideal simulation by array geometry (64 three-axis sensors, 192 channels, 30 dB).
Threshold-free Dice-AUC (shape) and the two localization measures: the half-power power-weighted centroid error and the threshold-free peak-voxel error are the headline metrics. The best-threshold Dice (, an oracle upper bound), HD95, and intensity correlation are supplementary. The forward model uses the complete -node FEM mesh and the body-force weighting. The source is displaced 2 cm from the array center, and the one-sided array covers one hemisphere of the sensor shell. With the source at the center of the search volume, a uniform map scores 0.01 cm on the sphere (0.17 cm on the one-sided array), so the centroid there measures where the grid was placed rather than what was reconstructed. At 2 cm, the uniform baseline is itself 2 cm, and a method achieves localization only where it falls below that row. All values use the refined 10-mm grid ( voxels), whereas the sweeps of Figure 3 use the coarser 20-mm grid. Values are means over realizations, redrawing measurement noise and the replicate perturbation, and are near-deterministic at 30 dB (median headline per-cell standard error , largest 0.41 cm, on the SBL peak). Bold values mark the best headline value per array geometry, judged on unrounded values; the three-way tie in sphere peak error is not marked.
2.4.1 Steering matrix and band weighting
Both beamformers below act on the same two objects, defined here once. The steering matrix maps a unit source 3-vector at voxel onto the sensor channels: the block belonging to physical sensor is , that is, the same scalar Green’s function and transverse projector used in the forward model, so that channel carries world axis of sensor . The matrix is normalized per by its Frobenius norm, so DAS remains a coherence measure rather than an amplitude map.
The band weight restricts the frequency sum to bins the array can actually resolve; it is unity inside the analysis band and zero above the spatial-Nyquist limit set by the nearest-neighbor sensor spacing , which suppresses the aliased lobes that would otherwise fold energy back into the map.
2.4.2 Delay-and-sum beamforming
Conventional backprojection () sums phase-aligned, band-weighted contributions over the three-dimensional source subspace at each voxel (Equation 2),where the trace over the form sums the steered power over the source’s three world directions, giving a broad main lobe set by the array’s Rayleigh limit.
The DAS and MVDR/Capon beamformers are inherently narrowband: both the steering matrix and the covariance are thus defined per frequency bin of the simulated SCG channel and only then combined across the band.
2.4.3 Minimum variance distortionless response
We implement the adaptive minimum variance distortionless response (MVDR) beamformer (). For stability, we use the numerically stable sum form, governed by the largest (best-observed) eigenvalue (Equation 3),which restores the shape while retaining the sharp localization. Because the cubic grid’s corners protrude outside the sensor shell, where is unconstrained and the power estimate diverges, two guardrails are applied. Each frequency’s map is divided by its spatial mean, one scalar per frequency, which equalizes frequencies before the band sum and leaves the spatial pattern within each one untouched. The summed map is then clipped at three times its 99.9th percentile.
Both were swept on a 20-mm grid with forward nodes. The clip is inconsequential: removing it, moving it to the 99th percentile, or increasing it tenfold leaves every reported centroid unchanged. Normalization matters, but no variant dominates across geometries. The standard unit-noise-gain weighting, , collapses the MVDR shape score to the no-information level on both arrays, while omitting the normalization entirely is better on the enclosing sphere but loosens the one-sided centroid. We therefore retain the spatial mean form throughout, and the reported MVDR performance depends on this choice. The power is likewise taken as rather than the conventional per-voxel form . Implemented on identical data, the latter is dominated by the source direction the transverse projector nulls, the one the array cannot see, and it doubles the sphere centroid error ( cm), nearly triples the one-sided one, and drops shape recovery to the no-information level on both.
2.4.4 CLEAN beamforming
We apply iterative Högbom deconvolution () on the subspace-DAS dirty map to extract point sources, which are restored with a fixed-width Gaussian beam. Its one nonstandard element is a rank- deflation that accounts for the unknown 3-vector source direction: each iteration subtracts the matched source contribution , with and .
2.4.5 Sparse Bayesian learning
A Type-II ML/evidence-maximization alternative () assigns each voxel an independent non-negative power hyperparameter and maximizes the marginal likelihood by the standard fixed-point expectation–maximization (EM) iteration, here initialized from the DAS map. The Bayesian prior sparsifies automatically, without a preset number of sources. On this dictionary, the iteration is not globally stable: it decreases the Type-II cost for roughly ten steps and then diverges, driving the ratio of the largest to the second-largest up without bound until the map is one voxel wide, while itself increases again. A threshold never triggers in that regime, so iterating to a fixed cap returns a diverged state rather than an estimate. We therefore evaluate at every step and keep the attaining its minimum, a stopping rule that uses no ground truth. Stopped this way, the estimator is stable and places the source at 1.22 cm on the sphere and at 1.73 cm on the one-sided array, but its support collapses to roughly four voxels, so it recovers position without volume. With a full per voxel, estimated by a geometric-mean fixed point, the hemisphere centroid error falls from 1.73 cm to 0.70 cm while the peak error rises from 1.73 cm to 3.16 cm and the shape overlap stays low (0.004 against 0.008). Specifying the source covariance properly sharpens the peak but leaves the distribution unrecovered. We report these numbers as the behavior of our adaptation to this problem rather than as a result about sparse Bayesian inversion.
2.5 Performance evaluation
Two ground-truth representations are generated: the first is the Distributed (FEM nodal), which is the FEM-based source map created by amplitude RMS-weighted node positions in the voxel space (Equation 4):where is the spatial coordinate vector of the 3D reconstruction voxel being queried is the true physical position of the th FEM mesh node, is the node-wise RMS acceleration acting as the source weight, and mm is the Gaussian kernel width.
The second ground truth is the Centroid (FEM nodal), which is the amplitude-weighted mean position of the same nodal set, used as the reference point for the centroid and peak-voxel error metrics (Equation 5):with the same node weights as above.
Reconstruction accuracy is assessed via five metrics:
Dice Coefficient: Spatial overlap of the activation masks , , both binarized at a common fraction-of-maximum threshold (Equation 6),
A single fixed threshold is method-unfair: scored against the extended FEM source, a focal estimator (MVDR, CLEAN) is penalized at high
while a broad one (DAS) is favored, and conversely at low
. As the headline shape metric, we therefore report the threshold-free area under the Dice-versus-
curve over
(
Equation 7)
The per-map maximum
(best-threshold Dice) is an oracle upper bound, a parameter chosen after computation with knowledge of the ground truth per method. It is reported separately and only as a reference.
2. Hausdorff Distance (95th percentile): boundary separation in cm, taken as with the 95th-percentile directed surface distance, evaluated on masks binarized at each map’s best-Dice isolevel.
3. Spatial Correlation: the Pearson correlation between the normalized intensity maps.
4. Centroid Error: the 3D Euclidean norm in cm between the reference centroid and the estimate’s half-power (0.5 of peak) power-weighted centroid , that is, the intensity-weighted mean voxel position over all voxels within half-power of the map maximum.
5. Peak-Voxel Error: the 3D Euclidean norm in cm between and the single argmax voxel of . Being an argmax over a discrete lattice, it is quantized to the grid: on the 10-mm grid, the only attainable values are 0, 1.00 cm, 1.41 cm, 1.73 cm, and so on; that is, 0, 1, , voxels, and it should be read in voxel units. It is a threshold-free complement to the centroid error that is insensitive to how much below-peak mass surrounds the estimate.
Resolution versus localization precision. The measures above locate a single dominant lobe, which is a different quantity from the ability to resolve two nearby sources. With m/,s the shear wavelength falls from 36.7 cm at 8 Hz to 8.0 cm at 35 Hz, so for an aperture cm at a mean range cm, the classical resolution scale spans 17.4 cm to 3.8 cm across the usable band. A centroid may be placed better than that scale, so the two limits are reported separately and are not interchangeable. Separability is measured directly: two copies of one compact activation region (1.5 cm radius, 209 nodes) are imaged at a controlled separation and scored by the Sparrow criterion, a dip of at least between the two maxima (Section 3.3).
Statistical analysis. The four estimators represent the conventional, adaptive, sparse, and sparse-Bayesian families and are reported to show what the general approach recovers, not to establish one as superior; no pairwise tests are performed. Each headline metric in Table 2 is the mean over realizations that redraw the additive measurement noise and the replicate perturbation, reported as a point estimate: the median per-cell standard error is , the largest being 0.41 cm on the SBL peak, the four cells above 0.01 are the sphere SBL peak and centroid plus the hemisphere MVDR and CLEAN centroids (0.021 cm and 0.010 cm). The seven-panel sweep of Section 3.4 averages 30 realizations per operating point, where the spread is observable. The null-control sweep uses three, and the respiration sweep uses five. Those curves carry bands about the per-point mean, shown as the shaded regions in Figure 3. All dispersion reported in this article describes that measurement-noise ensemble and nothing else. A single cardiac cycle of a single heart model underlies every number, so between-beat, between-subject, and between-anatomy variability is not sampled at all. Therefore, only descriptive statistics are reported.
3 Results
The results aim toward feasibility rather than a method competition and are reported in three parts that follow the logic of the pipeline. We first show that the forward stage produces plausible, array-resolvable seismocardiography (SCG). Then we investigate what spatial information can be recovered by four representative beamformers. Finally, we test how that recovery degrades as idealizations are relaxed toward real-world measurement conditions.
3.1 Forward simulation
Figure 2 shows the forward stage. At the channel sitting at the pole of the one-sided array, the synthesized acceleration peaks at 47 mg with an RMS of 12 mg, and restricting it to the 8–35 Hz band the estimator uses concentrates it into a burst between the QRS and peak active stress. The model carries no valve events, but it does carry the transmembrane potential, and the extracellular pseudo-ECG that follows from it by the standard lead-field integral shows a clear P, QRS, and T (Figure 2C). The simulated trace can therefore be read against an electrical reference. However, it does not show an SCG1 and SCG2 pair. The used myocardium-only model with a lumped circulation does not resolve these, and the model output carries no valve-event reference to mark against them.
FIGURE 2
Before propagation, of the unfiltered nodal-acceleration power already lies inside 8–35 Hz (a further between 60 Hz and 200 Hz). Propagation is what attenuates it: the fraction of channel energy falling inside the analyzed band varies roughly elevenfold across the array at nearly equal source range, from to . The array-pole element carries , close to the array median of , while the best-placed channel carries (Figures 2B,C). The reconstruction results of Table 2 and Figure 3 use the nominal 30 dB operating point unless stated otherwise.
FIGURE 3
3.2 Reconstruction performance under ideal conditions
All four beamformers are reported side by side. Table 2 gives per-method results for the idealized full sphere and a realizable one-sided hemisphere at the nominal 64-sensor (192-channel), 30 dB operating point. The headline metrics are the threshold-free Dice-AUC (shape) and two localization measures: the half-power power-weighted centroid and the threshold-free peak-voxel error, which are reported together because, on the one-sided hemisphere, the diffuse below-peak energy of the MVDR/Capon beamformer separates where the estimator places the peak from how much mass surrounds it. Best-threshold Dice (, an oracle upper bound), HD95, and correlation are supplementary. The gap between the two Dice columns, such as MVDR’s 0.43 AUC against its 0.80 oracle, measures how much a per-map threshold choice would flatter each estimator.
Two findings dominate. First, array geometry has the largest influence and acts unevenly across methods. Second, the metrics are stable across repeated noise draws, but that is reproducibility across measurement realizations only. A single heart, anatomy, and array underlie every entry (Section 4).
On the full sphere, all three non-Bayesian beamformers place the peak within one voxel and recover the centroid to 0.67–0.88 cm, well inside the 2.01 cm baseline, and their shape scores span 0.43–0.48. The supplementary metrics separate them: MVDR posts the smallest boundary error (HD95 1.41 cm) from its single focal peak, the broad DAS lobe achieves the highest intensity correlation , and CLEAN posts the largest boundary error (8.12 cm), the sharper maps trading overlap for boundary precision.
3.3 Performance under a one-sided sensor configuration
Restricting the array to one hemisphere separates the estimators sharply from the baseline. DAS degrades, its centroid reaching 4.49 cm and its peak reaching 10.63 cm, more than twice the 2.01 cm baseline. CLEAN and MVDR both clear it: CLEAN reaches 1.01 cm, a factor of two inside the baseline, and MVDR reaches 1.70 cm, with its peak within one voxel. Shape separates the same way, CLEAN at a Dice-AUC of 0.44 and MVDR at 0.38 against a 0.20 baseline, while DAS manages 0.26. Boundary error increases for DAS and MVDR (HD95 7.1 cm and 5.1 cm) and stays high for CLEAN (7.4 cm), the counterpart of the below-peak skirt a one-sided aperture produces, and it is the supplementary metric that most clearly marks the cost of losing half the array.
For the estimators tested here, two-source separability is an order of magnitude coarser than this single-source precision. The smallest separation resolved for mutually coherent copies is 16 cm on the sphere and 12 cm on the hemisphere for CLEAN and 20 cm for DAS, while MVDR resolves none up to the 20 cm tested on either geometry. Decorrelating the two sources moves every limit to 8–16 cm, placing much of the coherent-case penalty on coherent-source cancellation. That is a property of the estimators rather than of the array, so the decorrelation remedies of Section 4 would be expected to recover part of it.
A dynamic, Kalman-smoothed variant of CLEAN was also evaluated during development. It is marginally better on shape and worse on centroid than the static per-cycle map and is not reported separately. The feasibility statement is thus bounded: with the right estimator, the one-sided array still localizes the source to roughly one voxel, although it no longer reconstructs a sub-centimeter centroid as a fully enclosing array does. These numbers hold under matched forward/inverse physics (Section 4). The perturbation study below is a first step toward a more realistic setting.
3.4 Robustness to noise and model uncertainty
The configuration results above are best-case: nominal noise, perfectly known sensor positions, and matched propagation physics. To test for more adverse measurement conditions, we perturb one factor at a time about a nominal operating point (64 sensors, 30 dB SNR, matched ) and re-estimate. The sweeps score all four beamformers (DAS, MVDR, CLEAN, SBL) on every trial. Shape is scored with the threshold-free Dice-AUC, alongside the centroid error. Both are averaged over 30 trials per point. Figure 3 reports seven sweeps spanning two qualitatively different stressors: signal quality limitations (sensor noise, signal attenuation, array size) and array-geometry errors (sensor distance, heart off-center, sensor misplacement, wave-speed mismatch). We use the fast reconstruction path here (coarser grid); the curves therefore characterize degradation behavior and method ordering rather than the absolute accuracies of Table 2.
The pipeline tolerates poor signal quality better than poor geometry. Across sensor noise from 40 dB to 5 dB, the centroid is flat for DAS (1.04–1.02 cm) and for CLEAN (0.85 cm throughout) and moves by 0.24 cm for MVDR. MVDR’s shape score falls from 0.41 to 0.37 (Figure 3A). A sixteenfold increase in shear-wave absorption tightens the DAS, MVDR, and CLEAN centroids, although it costs CLEAN shape (0.49–0.40), and a forward/inverse wave-speed mismatch moves none of the three by more than 0.36 cm, against 1.89 cm for SBL (Figures 3B,G). SBL’s shape score is near zero across most of the sweep (0.01–0.06). It increases only where the evidence minimum is reached before the map sparsifies, and the estimate therefore stays near its DAS initialization, at 8–16 sensors (0.34 and 0.40) and at a 2 cm sensor misplacement (0.43). The wave-speed result relaxes one form of the inverse-crime concern: an unknown shear speed in an inhomogeneous thorax does not by itself collapse the inversion, although it is a uniform scalar perturbation within the shared propagation model, and a spatially confined contrast is the more demanding test (below).
What does degrade performance is array geometry. Pulling the shell in too close degrades shape. At a 6 cm mean sensor-to-heart range, the DAS and MVDR shape scores fall to 0.16 and 0.17, against 0.51 for CLEAN, while the centroids hold near a centimeter (1.26 cm, 1.83 cm, and 0.84 cm). All three recover once the array sits at or beyond a 10 cm standoff for this array and source (Figure 3D). Displacing the source further from the shell center is primarily a localization failure: DAS and MVDR reach 3.96 cm and 4.08 cm at an additional 8 cm, with their shape scores falling to 0.27 and 0.20, whereas CLEAN stays between 0.7 cm and 1.6 cm and its shape-AUC rises as the source moves into a less symmetric part of the aperture (Figure 3E). Shape also falls once the array drops below roughly 16 sensors, again for DAS and MVDR rather than for CLEAN, which is flat across the whole range: over the full 8- to 128-element sweep, its centroid stays between 0.80 cm and 0.90 cm and its Dice-AUC between 0.49 and 0.56. The adaptive estimator moves the other way; its centroid expands from 0.78 cm to 1.48 cm as elements are added, and the covariance becomes harder to estimate at a fixed snapshot count (Figure 3C).
Sensor misplacement separates the methods differently from the other factors. An assumed-versus-true position error of 2 cm leaves DAS unchanged ( cm) and moves MVDR by 0.09 cm but degrades CLEAN by a factor of 2.0 ( cm, Figure 3F).
Because the reconstruction grid is built around the array, a source sitting at its center can be recovered by any roughly symmetric map. The source is therefore displaced from the shell center and re-scored against two baselines: a uniform map over the non-excluded grid and channels whose phase has been randomized, with their magnitude spectra preserved. Figure 4 plots the displacement sweep against the uniform baseline. The phase-randomized control is reported numerically below. On the sphere, all three separate from the uniform baseline from 1 cm of displacement onward. On the one-sided array, CLEAN separates from 1 cm and holds a margin out to 6 cm, MVDR separates from 2 cm, and DAS separates only at the largest displacement tested. At zero displacement, none of them separates; the uniform baseline itself scores 0.17 cm on the one-sided array and 0.01 cm on the sphere. Shape behaves differently again: Dice-AUC clears the uniform baseline (0.17–0.20) at every displacement for CLEAN and MVDR, and phase randomization collapses it for CLEAN, from 0.44 to 0.04 on the one-sided array, while MVDR falls exactly to the baseline. The centroid behaves the same way at 2 cm of displacement: phase randomization drives CLEAN to 9.91 cm and DAS to 6.44 cm on the one-sided array, while MVDR decreases to 1.95 cm, level with the 2.01 cm baseline rather than beating it. Shape recovery therefore rests on inter-channel phase throughout, whereas the centroid only becomes informative once the source is off-center.
FIGURE 4
Respiration acts differently from a static displacement because it modulates the source during the interval the covariance averages over. Chest wall and heart move partly together, so only the residual half of the excursion is relative motion at the array, and that residual is what the forward model applies. The excursion is sinusoidal and zero-mean across the replicate set, so the source’s mean position is unchanged and only its spread within the averaging interval increases. The uniform baseline is therefore the same 1.99 cm at every amplitude, unlike the displacement sweep of Figure 4, where it increases with the offset. Sweeping the excursion to 3 cm at a 2 cm offset, on the 20 mm sweep grid, leaves the sphere unaffected. The two estimators diverge on the one-sided array: MVDR degrades monotonically from 1.65 cm to 2.95 cm and crosses the 1.99 cm uniform baseline at about 0.6 cm of excursion, while CLEAN improves slightly, from 1.48 cm to 1.07 cm, and maintains its margin throughout.
None of the sweeps above alters the form of the propagation model, and only two of the seven vary a propagation parameter at all. As a structural test, scored on the 10-mm grid of Table 2, we placed a 3 cm low-shear inclusion between source and array. A sphere in which the shear-wave speed is halved, so its shear modulus is a quarter of the surrounding value, with attenuation tripled. The sphere is crossed by of node–sensor paths on the sphere and on the one-sided array and absent from the steering vectors, so that the inverse operator has no representation of it. The enclosing array is essentially unaffected for the three non-Bayesian methods (Dice-AUC within 0.031, centroids within 0.19 cm). On the one-sided array, shape recovery survives: Dice-AUC falls from 0.44 to 0.39 for CLEAN and from 0.38 to 0.32 for MVDR, both still well clear of the 0.20 no-information baseline. Localization does not survive: MVDR degrades from 1.70 cm to 3.90 cm and CLEAN from 1.01 cm to 3.52 cm, losing the margin each held, while DAS is unchanged at 4.5 cm.
Because a single operating point cannot show where the boundary lies, we swept the severity on the one-sided array (20-mm grid, forward nodes). The result is driven by the wave-speed contrast and not by the attenuation: holding at half and varying absorption over 1– moves the CLEAN centroid by 0.02 cm. Against the contrast, it degrades progressively: CLEAN reaches 1.75 cm at a speed reduction and 1.89 cm at , both still inside the baseline, and only loses the margin at and beyond (2.42 cm, then 4.92 cm at half speed). MVDR is the more fragile of the two and crosses at about . Shape recovery is flat across the entire sweep (0.39–0.41 for CLEAN). Two features of the setup set that tolerance: the array is one-sided, and the contrast is local and absent from the steering. A uniform speed error rescales every path delay together and costs the enclosing array almost nothing (Figure 3G), as does the halved-speed inclusion on that array. What the contrast removes is the thin margin the one-sided aperture holds over the no-information baseline. On the one-sided array, the assumed shear speed must be right to roughly 20– along the paths crossing the heterogeneity for sparse deconvolution to localize, and it must be right to about for the adaptive estimator, whereas shape recovery tolerates the full range tested.
Quality is not the binding constraint across the sweeps signal. Array geometry is the binding constraint, and unmodeled propagation structure sets the tightest requirement.
3.5 Illustrative reconstructions
Figure 5 renders the DAS, MVDR, CLEAN, and SBL reconstructions under the best-case 64-sensor full array as an isosurface beside orthogonal slabs, both overlaid on the FEM ground truth. To compare shape and placement on equal footing, each estimate is drawn at a volume-matched isolevel, that is, the level at which its rendered volume equals that of the ground-truth shell. DAS, MVDR, and CLEAN sit on the true centroid (sub-centimeter, Table 2), with the broad DAS map overlapping the extended ground truth most directly. SBL is the visible outlier: stopped at its evidence minimum, its support is only about four voxels. The isosurface panel understates this because matching its rendered volume forces its level to of peak (against 0.69, 0.39, and 0.21 for the others), drawing the surface from the map’s low-level skirt.
FIGURE 5
Figure 6 shows a time-resolved centroid estimate, in which the estimated center follows the source as it moves through the cardiac cycle. Using the full-sphere array (so the trajectory reflects tracking, not geometry loss), we localize each short-time window independently, take the half-power centroid of the per-window CLEAN map, and run a constant-velocity Kalman () smoother () on that 3D trajectory. The raw per-window centroid already follows the FEM source’s energy-weighted node centroid on all three axes, albeit noisily (mean error 1.10 cm). Smoothing reduces this to 0.76 cm (median 0.75 cm), preserving the beat-scale motion, for a target that migrates several centimeters along its dominant axis. Like Table 2, this analysis uses the refined 10-mm grid and the same 2-cm source displacement. The pipeline thus resolves not only the cycle-average position but also how the centroid moves within the cycle.
FIGURE 6
4 Discussion
4.1 Principal findings
The main finding of this study is that multichannel seismocardiography contains sufficient spatial information to localize cardiac mechanical activity in three dimensions when combined with physics-informed forward modeling and inverse reconstruction. This was consistently observed across several fundamentally different reconstruction methods with an enclosing aperture. On a one-sided array, it no longer holds: the estimators separate, and only CLEAN and MVDR clear the no-data baseline in both position and shape. Two qualifications bound that statement. First, the forward and inverse stages share the same propagation model, so every accuracy reported here is an upper bound on what an unmatched, anatomically heterogeneous thorax would permit. The low-shear inclusion of Section 3.4 is the one test that breaks that symmetry. It bounds how far the assumed physics may depart from the true physics before the centroid margin is lost on the one-sided array, which is roughly a 20–local shear-speed contrast for the most tolerant estimator and about for the adaptive one, with shape recovery holding throughout, whereas a uniform speed error of the same size costs the enclosing array almost nothing. Second, what is recovered is the center of mechanical activity, not the full three-dimensional source distribution: centroid errors near a centimeter coexist with Dice-AUC values of 0.44–0.48 against a 0.20 no-information baseline, so the extent of the active region is recovered only coarsely. Localization precision here should not be read as volumetric resolving power, which the two-source experiment in Section 3.3 places an order of magnitude coarser. The estimators differed most where the array is one-sided. On the enclosing sphere, all three non-Bayesian estimators recovered the centroid to under a centimeter, and their shape scores spanned only 0.05, so the choice of estimator was largely immaterial. Restricting to one hemisphere separated them: sparse deconvolution and the adaptive beamformer both stayed inside the no-data baseline while conventional delay-and-sum did not. Breathing-scale source motion brought the adaptive estimator’s result above the baseline, and an unmodeled low-shear inclusion removed the centroid margin of both, although shape recovery survived it at every severity tested. What the inclusion establishes is a tolerance rather than a failure, quantified in Section 3.4. The robustness analysis further showed that geometric factors, such as sensor placement and registration accuracy, have a greater impact on reconstruction quality than measurement noise or wave-speed mismatch. These observations indicate that future improvements in multichannel SCG imaging will depend not only on advances in reconstruction algorithms but also on accurate anatomical modeling, reliable sensor localization, and optimized sensor array design. Taken together, these findings suggest that multichannel SCG could evolve from a single-point monitoring modality toward a modality capable of spatial imaging of cardiac mechanical activity.
4.2 Comparison with previous work
The inverse perspective adopted here connects seismocardiography to a broader family of source reconstruction problems. Its closest analog is electrocardiographic imaging (ECGI), which reconstructs cardiac electrical sources from dense body-surface potential recordings (). The present framework pursues the mechanical counterpart of that idea: it estimates the three-dimensional distribution of myocardial mechanical activity from surface vibration measured across a sensor array. Within the mechanical-based signal literature, prior multichannel work has pursued related but distinct goals. Array SCG has been employed to characterize the spatial variation of the signal across the chest (), describing where surface vibration patterns differ rather than inverting for the position of their origin. Subspace localization has also been applied to heart sounds from a chest microphone array (). Electromechanical finite-element models such as the one used here have primarily served as forward generators of realistic SCG (; ) rather than as controlled ground truth for evaluating an inverse pipeline. The present study differs from both by coupling a published finite-element source, a forward-propagation model, and an array-processing inverse stage into a framework to quantify what spatial information a chest surface SCG array can recover. Three-dimensional localization of cardiac mechanical sources from multichannel SCG has not previously been demonstrated in this form. Taken together, these components establish an end-to-end physics-informed framework that links cardiac electromechanics, wave propagation, multichannel sensing, and volumetric inverse reconstruction. This integration, rather than the individual reconstruction algorithms themselves, represents the primary methodological contribution of the present work.
The failure modes are shared with electromagnetic source imaging, where they have been studied far longer. Depth bias, which motivates the per-frequency normalization used here, is treated there by unit-noise-gain and neural-activity-index weightings (). The collapse of MVDR on two coherent sources is the cancellation that spatial smoothing () and forward–backward averaging address, and the choice between scalar and vector source models, with the behavior of adaptive filters on correlated sources, is set out in . None of these remedies are applied here, which plausibly contributes to the adaptive estimator being fragile throughout the tests.
The two modalities also differ in kind. ECGI and magnetocardiography image a quasi-static field: propagation is effectively instantaneous. Seismocardiography images a propagating elastic wave with finite delay. That delay structure is what encodes position here, which is why randomizing inter-channel phase collapses shape recovery (Section 3.4). Binding limits follow from it: a wavelength and an aperture set the resolution scale , and the inter-sensor spacing caps the usable band at the array spatial-Nyquist limit.
4.3 Limitations
The framework was evaluated exclusively on simulated data, using the published output of an electromechanical model that we consume rather than re-run. Its nodal accelerations are taken as ground truth; we do not independently verify them against measured myocardial accelerations in the 8–35 Hz analysis band. The simulated trace is accordingly not a clinical seismocardiogram, and because the most spatially distinctive features of a recorded SCG are the valve-related ones, transfer of this result to recorded data is not established here.
The propagation model omits the traction-free skin boundary and the lungs, both first-order effects on amplitude and on channel-to-channel transfer. Forward and inverse operators otherwise share the same functional form: five of the seven sweeps hold them matched and vary the measurement conditions alone, while sensor misplacement and wave-speed mismatch detune the inverse operator against the forward one. Furthermore, the low-shear inclusion of Section 3.4 is absent from the steering vectors altogether, and it costs both MVDR and CLEAN their centroid margin over a no-data baseline on the one-sided array once the local shear-speed contrast exceeds roughly 20–, although their shape recovery survives it. It crosses of node–sensor paths on that array. The reported MVDR performance also depends on its normalization choice (Section 2.4.3). All reported variability comes from redrawing measurement noise and the replicate perturbation on a single heart, anatomy, and array geometry, so the reported performance remains an upper bound. On the one-sided hemisphere, the 64 elements sit at a minimum nearest-neighbor spacing of 2.3 cm, which lifts the array spatial-Nyquist limit to 60.8 Hz. That bound is set by the closest pair, and the mean and worst spacings (2.7 cm and 3.0 cm) place it at 52 Hz and 47 Hz, so the usable band is closer to 47 Hz than 60 Hz. Capping it there changes only MVDR, from 1.70 cm to 1.56 cm. That spacing also implies a denser patch of elements than a chest surface interrupted by ribs and sternum would readily carry. The half-shell is a geometric bound on one-sidedness rather than a sited chest array: it is the hemisphere of the sensor shell, and no anatomical placement is claimed for it. A realizable version would trade element count for spacing, and Figure 3C bounds that trade at roughly 16 sensors, although that bound is measured on the enclosing geometry, and its transfer to a one-sided array is untested here. Experimental validation with phantoms and multichannel recordings are thus necessary next steps.
4.4 Future perspectives
The limits reported here are boundaries this first study locates rather than fixed properties of the modality: localization on a one-sided array required an unmodeled local shear-speed contrast to stay within roughly 20– along the paths crossing it for the most tolerant estimator, and nearer for the adaptive one, which lost its margin at about 0.6 cm of respiratory excursion, which respiratory gating or motion-tracked steering would relax. While this study tested scalar perturbations, future research should encompass comprehensive structural model-mismatch studies to simulate things like a realistic thorax, accounting for guided-wave dispersion, lungs, ribs, and sensor-coupling. A forward operator that represents that heterogeneity, a design requirement given that myocardial shear-wave speed is about 3 m/s in healthy volunteers and up to about four times higher in diseased or stressed myocardium (; ), would remove the mismatch rather than simply survive it. Beyond forward modeling, subsequent steps could explore remedies for the sparse Bayesian approach we could not adapt here, evaluate alternative inverse estimators, and develop learned or diffusion-prior-based inversion methods. Shape recovery held wherever localization did not, so the two goals may reward different estimators rather than one. On the signal processing front, integrating dispersion-corrected (Lamb-A0) and polarization-sensitive beam steering using triaxial channels could prove promising. To support these advancements, a Cramér–Rao resolution-floor map could guide future array co-design, while validation efforts should optimally leverage physically complete synthetic multichannel SCG datasets paired with echo or MRI-derived reference sets. Ultimately, these steps lay the groundwork for time-resolved shape reconstruction.
5 Conclusion
This work shows that, in simulation, multichannel seismocardiography carries sufficient spatial information to enable three-dimensional localization of the center of cardiac mechanical activity when combined with physics-informed forward modeling and inverse reconstruction. Using a published electromechanical heart model, elastic wave propagation, and representative beamforming approaches, the framework recovers the position of localized mechanical sources under the simulated conditions examined here. Two limits should be carried with that result: the forward and inverse stages share a propagation model, so the reported accuracy is an upper bound rather than an expected field performance, and what is recovered is the center of activity rather than the full source distribution. Both are boundaries this first study locates rather than fixed limits of the modality, and Section 4.4 sets out possible routes to widen them. The result is accordingly a proof of concept obtained under largely controlled modeling conditions. Experimental validation with phantoms and with recorded multichannel SCG is the necessary next step, and wearable mechanical cardiac imaging remains a prospect these results motivate rather than a demonstrated capability.
Statements
Data availability statement
The simulation and reconstruction pipeline that produces every result and figure in this study is released under an open-source licence under Zenodo DOI 10.5281/zenodo.22256749. The electromechanical heart model used as ground truth is previously published and available from its original authors (). We redistribute no part of it.
Author contributions
RK: Methodology, Visualization, Writing – original draft, Conceptualization, Formal Analysis, Software, Writing – review and editing, Investigation. U-VA: Writing – review and editing, Methodology, Supervision, Conceptualization.
Funding
The author(s) declared that financial support was received for this work and/or its publication. The authors acknowledge financial support for the publication costs by the Open Access Publication Fund of Bielefeld University and the German Research Foundation (DFG).
Acknowledgments
The authors thank the developers of the open electromechanical heart model, Gerach et al., for providing the whole-heart model used as the basis for the simulations performed in this study.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. Generative AI tools were used in a limited capacity during manuscript preparation, solely to assist with language translation and refinement of selected text passages. All AI-assisted outputs were carefully reviewed, revised, and validated by the authors to ensure accuracy, clarity, and consistency with the intended scientific content. No generative AI tools were used for study design, data analysis, interpretation of results, or formulation of scientific conclusions.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AkiK.RichardsP. G. (2002). Quantitative Seismology. (2nd Edn.) Sausalito, CA: University Science Books.
2
BernardS.KazemiradS.CloutierG. (2017). A frequency-shift method to measure shear-wave attenuation in soft tissues. IEEE Trans. Ultrasonics, Ferroelectr. Freq. Control64, 514–524. 10.1109/TUFFC.2016.2634329
3
CaponJ. (1969). High-resolution frequency-wavenumber spectrum analysis. Proc. IEEE57, 1408–1418. 10.1109/proc.1969.7278
4
CrowR. S.HannanP.JacobsD.HedquistL.SalernoD. M. (1994). Relationship between seismocardiogram and echocardiogram for events in the cardiac cycle. Am. Journal Noninvasive Cardiology8, 39–46. 10.1159/000470156
5
DehkordiP.Khosrow-KhavarF.Di RienzoM.InanO. T.SchmidtS. E.BlaberA. P.et al (2019). Comparison of different methods for estimating cardiac timings: a comprehensive multimodal echocardiography investigation. Front. Physiol.10, 1057. 10.3389/fphys.2019.01057
6
Di RienzoM.VainiE.LombardiP. (2017). An algorithm for the beat-to-beat assessment of cardiac mechanics during sleep on Earth and in microgravity from the seismocardiogram. Sci. Rep.7, 15634. 10.1038/s41598-017-15829-0
7
Di RienzoM.RizzoG.IşılayZ. M.LombardiP. (2020). SeisMote: a multi-sensor wireless platform for cardiovascular monitoring in laboratory, daily life, and telemedicine. Sensors (Basel)20, 680. 10.3390/s20030680
8
GerachT.SchulerS.FröhlichJ.LindnerL.KovachevaE.MossR.et al (2021). Electro-mechanical whole-heart digital twins: a fully coupled multi-physics approach. Mathematics9, 1247. 10.3390/math9111247
9
GurevV.LeeT.ConstantinoJ.ArevaloH.TrayanovaN. A. (2011). Models of cardiac electromechanics based on individual hearts imaging data. Biomechanics Model. Mechanobiol.10, 295–306. 10.1007/s10237-010-0235-5
10
HögbomJ. A. (1974). Aperture synthesis with a non-regular distribution of interferometer baselines. Astronomy Astrophysics Suppl. Ser.15, 417.
11
InanO. T.MigeotteP.-F.ParkK.-S.EtemadiM.TavakolianK.CasanellaR.et al (2015). Ballistocardiography and seismocardiography: a review of recent advances. IEEE J. Biomed. Health Inf.19, 1414–1427. 10.1109/JBHI.2014.2361732
12
KálmánR. E. (1960). A new approach to linear filtering and prediction problems. J. Basic Eng.82, 35–45. 10.1115/1.3662552
13
KrimH.VibergM. (1996). Two decades of array signal processing research: the parametric approach. IEEE Signal Processing Magazine13, 67–94. 10.1109/79.526899
14
MoghaddasiH.AlmasganjF.ZoroufianA. (2017). Imaging of heart acoustic based on the sub-space methods using a microphone array. Comput. Methods Programs Biomed.146, 133–142. 10.1016/j.cmpb.2017.04.017
15
MunckK.SørensenK.StruijkJ. J.SchmidtS. E. (2020). Multichannel seismocardiography: an imaging modality for investigating heart vibrations. Physiol. Meas.41, 115001. 10.1088/1361-6579/abc0b7
16
RaiD.ThakkarH. K.RajputS. S.SantamariaJ.BhattC.RocaF. (2021). A comprehensive review on seismocardiogram: current advancements on acquisition, annotation, and applications. Mathematics9, 2243. 10.3390/math9182243
17
RamanathanC.GhanemR. N.JiaP.RyuK.RudyY. (2004). Noninvasive electrocardiographic imaging for cardiac electrophysiology and arrhythmia. Nat. Med.10, 422–428. 10.1038/nm1011
18
RauchH. E.TungF.StriebelC. T. (1965). Maximum likelihood estimates of linear dynamic systems. AIAA J.3, 1445–1450. 10.2514/3.3166
19
SantosP.PetrescuA. M.PedrosaJ.OrlowskaM.KominiV.VoigtJ.-U.et al (2019). Natural shear wave imaging in the human heart: normal values, feasibility, and reproducibility. IEEE Trans. Ultrasonics, Ferroelectr. Freq. Control66, 442–452. 10.1109/TUFFC.2018.2881493
20
SarvazyanA. P.SkovorodaA. R.EmelianovS. Y.FowlkesJ. B.PipeJ. G.AdlerR. S.et al (1995). Biophysical bases of elasticity imaging. Acoust. Imaging21, 223–240. 10.1007/978-1-4615-1943-0_23
21
SekiharaK.NagarajanS. S. (2008). Adaptive Spatial Filters for Electromagnetic Brain Imaging. Berlin: Springer. 10.1007/978-3-540-79370-0
22
ShanT.-J.WaxM.KailathT. (1985). On spatial smoothing for direction-of-arrival estimation of coherent signals. IEEE Trans. Acoust. Speech, Signal Process.33, 806–811. 10.1109/TASSP.1985.1164649
23
SørensenK.SchmidtS. E.JensenA. S.SøgaardP.StruijkJ. J. (2018). Definition of fiducial points in the normal seismocardiogram. Sci. Rep.8, 15455. 10.1038/s41598-018-33675-6
24
TroelstraM. A.RungeJ. H.BurnhopeE.PolcaroA.GuenthnerC.SchneiderT.et al (2021). Shear wave cardiovascular MR elastography using intrinsic cardiac motion for transducer-free non-invasive evaluation of myocardial shear wave velocity. Sci. Rep.11, 1403. 10.1038/s41598-020-79231-z
25
Van VeenB. D.van DrongelenW.YuchtmanM.SuzukiA. (1997). Localization of brain electrical activity via linearly constrained minimum variance spatial filtering. IEEE Trans. Biomed. Eng.44, 867–880. 10.1109/10.623056
26
WipfD. P.RaoB. D. (2004). Sparse bayesian learning for basis selection. IEEE Trans. Signal Processing52, 2153–2164. 10.1109/tsp.2004.831016
Summary
Keywords
beamforming, cardiac mechanics, cardiac source imaging, electromechanical heart model, inverse source localization, multichannel signal processing, physics-informed modeling, seismocardiography
Citation
Kuhlmann R and Albrecht U-V (2026) Physics-informed 3D source localization from multichannel seismocardiography. Front. Signal Process. 6:1937659. doi: 10.3389/frsip.2026.1937659
Received
14 July 2026
Revised
21 September 2026
Accepted
29 September 2026
Published
11 November 2026
Volume
6 - 2026
Edited by
Soroosh Solhjoo, F. Edward Hébert School of Medicine, United States
Reviewed by
Parastoo Dehkordi, Heart Force Medical Inc., Canada
Peshala Thibbotuwawa Gamage, Florida Institute of Technology, United States
Updates
Copyright
© 2026 Kuhlmann and Albrecht.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Urs-Vito Albrecht, urs-vito.albrecht@uni-bielefeld.de
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.