FG-DPUNet: a feature-aware gating dual-path U-Net for ECG motion artifact removal with waveform integrity

  • 1. Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, China

  • 2. University of Science and Technology of China, Hefei, China

Abstract

Wearable electrocardiography (ECG) has significant clinical value in ambulatory monitoring. However, motion artifacts during daily activities frequently distort faint P/T waves, compromising diagnostic interpretation. Meanwhile, single-branch or static fusion U-Net-based networks struggle to balance effective noise suppression with waveform integrity. Existing denoising methods primarily optimize global metrics such as signal-to-noise ratio improvement (), percentage root-mean-square difference (PRD), and root mean square error (RMSE), overlooking that denoising distorts P-QRS-T waveform and thus affects clinical diagnosis. This paper presents FG-DPUNet (Feature-aware Gating Dual-Path U-Net), a framework tailored to address P/T diagnostic waveform distortion induced by motion artifacts in wearable ambulatory ECG monitoring. The framework employs heterogeneous dual branches with kernel sizes of and to extract multi-scale ECG features in parallel, and adaptively adjusts fusion weights via a dynamic gating mechanism that operates on shared shallow features to produce temporal weights. To quantitatively assess waveform recovery quality, we introduce the P-QRS-T weighted correlation degradation index () and P/T wave peak deviation metrics (, ), providing a waveform-aware quantitative evaluation approach. On the MIT-BIH dataset under standard additive noise testing, FG-DPUNet improves the average by 5.94 dB and reduces the average PRD from 63.87% to 10.07% compared with state-of-the-art (SOTA) methods. On the BUT real-world motion artifact dataset, cross-subject experiments demonstrate that our model improves by approximately 16% over static dual-branch structures, demonstrating improved morphological preservation under motion artifacts.

1 Introduction

Wearable ECG devices, as a key technology for long-term ambulatory cardiac monitoring, have demonstrated significant clinical value in early screening of arrhythmias, alerting to myocardial ischemia, and remote postoperative follow-up (Choi et al., 2025; Arif et al., 2025; Zhang et al., 2025). Nevertheless, daily activities inevitably introduce severe motion artifacts that corrupt clinically vital ECG waveforms (Zhang et al., 2025; Suliman et al., 2025). Differing from static hospital-recorded ECG, wearable recordings suffer from continuous electrode-skin impedance variation and skeletal muscle contraction, yielding complex mixed noise with both additive and non-additive characteristic (Yu and Li, 2022; Prakash and Thomas, 2025; Midani et al., 2025). Such noise distorts delicate P/T wave morphologies, degrading signal fidelity and increasing the risk of clinical misdiagnosis. This restricts large-scale clinical deployment and weakens the utility of long-term wearable cardiac monitoring. To address the challenges posed by motion artifacts, a variety of ECG denoising methods have been developed over the past decades.

Traditional ECG denoising methods mainly rely on linear and adaptive filtering techniques. Least mean squares (LMS) and recursive least squares (RLS) adaptive filters achieve signal denoising through iterative optimization with noise references (Galdos et al., 2024; Wu et al., 2022; An et al., 2022; Marzog et al., 2022), featuring low computational overhead and real-time processing capability. However, their performance heavily relies on precisely synchronized noise references. Although such reference signals can be collected, they do not maintain a strict correlation with motion artifacts embedded in ECG recordings, which restricts practical deployment on wearable devices. Time-frequency analysis methods, such as discrete wavelet transform (DWT) and empirical mode decomposition (EMD), eliminate the demand for predefined prior noise assumptions and separate signal and noise via multi-scale component decomposition (Hussein et al., 2022; Skoric et al., 2024; Zhu et al., 2025). Nevertheless, their static thresholding schemes exhibit weak robustness against complicated motion artifacts and frequently over-suppress faint P/T wave features. This results in irreversible loss of clinically critical diagnostic information. Overall, traditional filtering techniques lag behind deep learning-based algorithms in handling non-stationary motion contamination (Zhang et al., 2025).

Beyond these conventional approaches, recent deep learning paradigms have also been extensively explored for ECG denoising. Attention-based networks (Wang S. et al., 2025) and Transformers (Cui et al., 2026; Wang and Li, 2025) have been introduced to capture long-range dependencies in ECG signals, leveraging self-attention mechanisms to model global temporal contexts. Diffusion models (Fang et al., 2025; Hassoon et al., 2025) have demonstrated promising results in generative signal reconstruction through iterative refinement, while state-space models (Parmonangan et al., 2025) offer alternative frameworks for sequential signal modeling. Despite their promising denoising performance on public benchmark datasets, these advanced architectures typically involve complex model structures and relatively high computational overheads. For instance, self-attention in Transformers scales quadratically with sequence length (Vaswani et al., 2017), and diffusion models involve multi-step iterative sampling (Ho et al., 2020), both of which make them less amenable to resource-constrained wearable scenarios. In such devices, lightweight architectures are critical yet often trade off denoising performance for computational efficiency.

Within this landscape, denoising autoencoders, Convolutional Neural Networks (CNNs), and U-Net-based models have been widely adopted for ECG signal enhancement (Ullah et al., 2025; Shaheen et al., 2025; Yuen et al., 2020). Among these, U-Net has emerged as a mainstream baseline framework owing to its encoder-decoder architecture and skip connections, which strike a favorable balance between noise suppression and the preservation of underlying waveform features (Hu et al., 2024; Li et al., 2025; Wang K. et al., 2025). However, single-branch U-Net architectures face an intrinsic receptive field trade-off. Small convolution kernels excel at retaining local P-QRS-T waveforms yet lack the capacity to suppress large-scale baseline wander and myoelectric interference. Conversely, large convolution kernels effectively remove low-frequency artifacts but tend to blur and attenuate faint diagnostic components such as P/T waves. This receptive field limitation hinders single-branch models from simultaneously fulfilling the dual demands of powerful noise suppression and high-fidelity recovery of ECG waveform.

To alleviate the limitation of single-scale receptive fields, existing studies have constructed multi-branch parallel architectures to enhance denoising performance through complementary multi-scale features (Zhang et al., 2024; Xiang et al., 2024). However, current multi-branch methods generally adopt fixed averaging or static convolutional fusion mechanisms, where all temporal samples share unified fusion weights, lacking the ability to adapt at each time step and per sample. In wearable scenarios, motion artifacts intensity fluctuates dynamically. Static fusion strategies can only achieve suboptimal compromises, failing to dynamically adjust the contribution of each branch according to real-time noise intensity. To address this issue, a dynamic gating mechanism has been adopted in ECG denoising autoencoders, where gate masks are adaptively generated from the input to control feature flow (Shaheen et al., 2025). Their experiments demonstrated that FGDAE’s per-layer dynamic gating achieves strong performance compared with static fusion schemes. Nevertheless, this architecture relies on duplicated serial gating sub-modules deployed at nearly every encoder-decoder and skip-connection layer. Despite its reduced parameter count, such duplicated gating branches introduce non-negligible inference overhead, which limits its deployment potential for resource-constrained applications.

The quantitative evaluation frameworks adopted in existing ECG denoising studies commonly rely on statistical metrics such as and RMSE, alongside distance-based metrics including PRD and maximum absolute deviation (MAD) (Zhang et al., 2025). While these metrics reflect overall amplitude errors, they are unable to capture morphological distortions within diagnostically vital waveforms such as P/T waves. In clinical practice, diagnosis heavily depends on P-wave morphology, PR interval, T-wave morphology, and QT interval. Even minor distortions in these features can impair diagnostic accuracy for pathologies including arrhythmias and myocardial ischemia (Fang et al., 2024; Rautaharju et al., 2009; Kligfield et al., 2007). Consequently, optimizing only global metrics may produce favorable numerical outcomes yet mask morphologically meaningful waveform distortions. This demonstrates that current evaluation paradigms are inadequate for quantifying the clinical diagnostic utility of ECG denoising algorithms.

Based on the above analysis, existing ECG denoising methods still present two practical limitations for wearable artifact suppression. On one hand, current representative network architectures struggle to balance noise suppression and waveform morphological preservation. On the other hand, conventional global metrics such as

, RMSE, and PRD only reflect overall amplitude errors. They are unable to quantify subtle distortions in diagnostic P-QRS-T waveforms, leading to a lack of waveform-aware quantitative evaluation criteria for waveform fidelity. To address these limitations, this paper presents FG-DPUNet, an adaptive dual-branch ECG denoising framework with a compact architecture, whose core design goal is to preserve P-QRS-T waveform morphology. The main contributions of this study can be summarized as follows:

  • FG-DPUNet is a compact dual-branch fusion framework designed to adaptively balance the short-receptive-field and long-receptive-field branches. It employs cross-branch feature-aware gating for dual-branch interaction to enable such a compact architecture, and addresses the key challenge of balancing noise suppression and waveform fidelity observed in existing methods.

  • A set of morphological-temporal fidelity metrics (, , ) is introduced to quantify morphological preservation of the P-QRS-T waveform. These metrics complement traditional indices such as , PRD, and RMSE, providing a more comprehensive performance evaluation of wearable ECG denoising algorithms.

  • The presented FG-DPUNet is specifically validated on a cross-subject wearable dataset that captures real-world motion artifacts. This validation demonstrates the model’s capability to handle non-stationary, non-additive interference in ambulatory settings, suggesting its potential for wearable ECG monitoring application.

The remainder of this paper is organized as follows. Section 2 describes the architecture of FG-DPUNet, including the dual-branch structure and the gating mechanism, and introduces the morphological-temporal fidelity metrics used for morphological evaluation. Section 3 details the experimental setup, including datasets, evaluation metrics, and implementation details. Section 4 presents the experimental results on both synthetic and real-world datasets. Section 5 discusses the implications of the findings, limitations, and future work. Section 6 concludes the paper.

2 Methods

2.1 Overall architecture

FG-DPUNet is presented in this work as a compact dual-branch fusion framework that generates fusion weights from shared shallow features to adaptively balance the two branches. These branches are designed with different receptive fields to capture multi-scale ECG features. The overall architecture of the presented framework is shown in Figure 1.

FIGURE 1

Given a noisy single-lead ECG input (, the number of samples per segment, corresponding to 5 s at 250 Hz), FG-DPUNet consists of three core components: a shared shallow convolution layer, two heterogeneous U-Net branches with different kernel sizes, and a lightweight feature-aware gating module. The shared shallow convolution first extracts low-level unified features (, the number of feature channels) from the input signal, which are then delivered in parallel to the dual-branch structure and the gating module. The dual-branch structure, with two parallel U-Net branches using kernel sizes of 7 and 21, extracts complementary fine temporal detail and broad temporal context. The gating module subsequently aggregates channel-wise information from the shared features to produce a temporal gate vector, which dynamically modulates the fusion weights of the two branch outputs. The final denoised ECG is obtained via element-wise weighted aggregation.

2.2 Heterogeneous dual-branch structure

The detailed internal encoder-decoder layout with skip concatenation layers of the dual-path U-Net is presented in Figure 2. Both branches adopt symmetric encoder-decoder architectures consisting of multi-scale downsampling and upsampling stages. Within each encoder, stride-based convolutions progressively compress feature resolution to capture multi-scale noise patterns. The corresponding decoder recovers the original temporal resolution via transposed convolutions. Channel-wise skip concatenation recovers fine-grained temporal details lost during downsampling, enabling accurate ECG waveform reconstruction.

FIGURE 2

The receptive field of a convolutional kernel can be approximated as in the time domain, where is the kernel size and is the sampling interval. At a sampling rate of 250 Hz (), a kernel of size covers approximately 28 ms, and a kernel of size covers approximately 84 ms. Based on this, the short-receptive-field branch () captures fine temporal details, while the long-receptive-field branch () provides a broader temporal context. The short-receptive-field branch is designed to capture rapid morphological fluctuations, including sharp QRS complexes and fine P/T wave details, enabling precise localization of abrupt ECG changes and preserving diagnostically critical features. Nevertheless, the branch is primarily optimized for fine-scale noise suppression, and its ability to handle broad-scale motion interference is relatively limited. In contrast, the long-receptive-field branch targets slow waveform drifts such as baseline wander and electrode motion artifacts, suppressing persistent motion contamination inherent to wearable ECG recordings via its expanded temporal receptive field. However, this wide window tends to over-smooth delicate waveform transitions and weaken faint clinical cardiac signatures. In practice, skip connections in both branches help retain fine details, so the difference in QRS preservation between the two branches is relatively modest.

The outputs of the two branches are formulated as:where and denote the reconstructed signals from the short-receptive-field and long-receptive-field branches, respectively, and and represent the full encoder-decoder transformation pipelines of the short-receptive-field and long-receptive-field branches, respectively.

This dual-branch framework assigns scale-specific feature extraction tasks to different branches, establishing complementary temporal-scale characteristics that provide a structural prerequisite for the subsequent gating fusion.

2.3 Feature-aware gating

To adaptively integrate complementary denoising outputs from the two branches, this work designs a lightweight feature-aware gating module that generates temporally dynamic fusion weights in an end-to-end manner without additional supervision. As shown in the middle of Figure 1, driven by real-time noise intensity encoded in the shared shallow features, the gating module first aggregates channel-wise information derived from and normalizes the aggregated representation through a sigmoid function to produce a temporal weight vector bounded within . This lightweight design, employing only average pooling and sigmoid activation, is intentionally chosen to maintain the compactness of the overall architecture. Unlike existing gating mechanisms for ECG denoising, which mainly perform spatial- or channel-wise attention and per-layer feature modulation, our gating module serves as a lightweight temporal fusion-weight generator. It does not build pairwise temporal dependencies as in conventional temporal attention. Built upon global average pooling and a subsequent sigmoid activation, our module directly outputs cross-branch fusion weights rather than performing feature re-calibration inside individual branches.where denotes the cth channel of shared features, and refers to the Sigmoid activation function.

Based on the derived gate vector, element-wise adaptive weighted aggregation is performed on the reconstructed signals from two branches, formulated as:where denotes the Hadamard (element-wise) product operation.

This adaptive fusion mechanism enables the network to dynamically adjust the contribution of each branch according to instantaneous input signal quality, addressing the key challenge of balancing noise suppression and waveform fidelity. For low signal-to-noise ratio (SNR) segments corrupted by severe baseline wander and electrode motion, a larger gate value boosts the long-receptive-field branch’s contribution to suppress broad-scale motion interference while preserving the morphological fidelity of P-QRS-T waveform. In contrast, a smaller gate value prevails for clean high-SNR segments, enabling the short-receptive-field branch to dominate waveform reconstruction and preserve fine-grained morphological details of P-QRS-T waveform without over-smoothing. Notably, the gating module incurs negligible computational overhead and supports fully end-to-end joint optimization of the entire network.

2.4 Metrics for ECG morphological and temporal fidelity

ECG morphological preservation is reflected in both morphological fidelity and temporal accuracy. To quantitatively assess these aspects, we introduce three metrics: , and .

2.4.1 Weighted correlation degradation index ()

The weighted correlation degradation index quantifies the loss of morphological fidelity, with a particular emphasis on P-QRS-T complex. It is defined as:where is the weighted Pearson correlation coefficient between the denoised ECG signal and the clean reference :

To emphasize the clinical importance of the P-QRS-T complex, higher weight coefficients are assigned to P-wave and T-wave regions, where diagnostic information is most concentrated. Specifically, is set to five within the P-wave region (200 ms–50 ms before the R peak) and the T-wave region (150 ms–400 ms after the R peak), and set to one in all other regions.

2.4.2 Peak interval deviations ( and )

Two absolute error metrics are defined to quantify the distortion of ECG characteristic intervals, denoted as and (unit: ms):where , , and , , denote the peak positions of P wave, R wave, and T wave in the clean reference and denoised signal, respectively. Smaller and values indicate better preservation of PR and RT intervals.

To improve the temporal resolution of peak localization, the original 250 Hz signals are upsampled to 1000 Hz using cubic interpolation to reduce discrete sampling errors. The P wave is searched for within a 150 ms window preceding each detected R peak, and the T wave is searched for within a 300 ms window following each R peak. Local maxima within the corresponding windows are identified as P and T wave peaks, respectively.

3 Experiments

3.1 Datasets

Two datasets are employed with distinct evaluation protocols to comprehensively assess the presented method. All ECG signals from both datasets were resampled to 250 Hz to ensure a consistent sampling rate across experiments. The original sampling rates and lead configurations are detailed in Table 1. The Massachusetts Institute of Technology-Beth Israel Hospital (MIT-BIH) dataset provides controlled additive noise conditions for standardized benchmarking, while the Brno University of Technology Smartphone PPG Database (BUT) offers realistic motion artifacts to assess generalization under ambulatory settings.

TABLE 1

FeatureMIT-BIHBUT
Acquisition scenarioClinical (static)Wearable (ambulatory)
Original sampling rate360 Hz1000 Hz
Resampled rate250 Hz250 Hz
Available raw leadsDual channels (MLII + auxiliary V1/V2/V5)Lead I
Selected leads for experimentsML IILead I
Noise typeArtificially added (BW/EM/MA)Real-world motion artifacts
PurposeSOTA quantitative comparison, ablationAblation, gating adaptivity and morphological-temporal fidelity metrics analysis

Key characteristics of the MIT-BIH and BUT ECG datasets used in this study.

3.1.1 MIT-BIH arrhythmia database

Raw ECG signals from the MIT-BIH Arrhythmia Database were preprocessed using a median filter with a window size of 200 to remove baseline wander (Moody and Mark, 2001). Three types of real noise, including baseline wander (BW), electrode motion (EM), and muscle artifact (MA), were sourced from the MIT-BIH Noise Stress Test Database (NSTDB) (Goldberger et al., 2000). Following the protocol in 2D-CEDNet (Ullah et al., 2025), six recordings (118, 119, 16265, 16272, 16273, and 16420) were selected for experiments. For each recording, 30 min of clean ECG and 30 min of mixed noise were used. The noisy signal was synthesized by adding noise to the clean signal in 5-s segments, with the scaling factor determined by ), where and are the standard deviations of the clean and noise segments, respectively. Six input SNR levels () range from −6 dB to 24 dB were synthesized.

3.1.2 BUT wearable ECG dataset

The BUT dataset contains real-world wearable ECG recordings with motion artifacts collected during daily activities (walking, laughing, talking, coughing, and resting) (Nemcova et al., 2021). The dataset comprises recordings from 50 subjects (25 females, 25 males), ranging in age from 19 to 76 years. Recordings were acquired under both resting and multiple motion conditions. For our experiments, we selected subjects with IDs from 112 to 149, excluding 115, 121, and 143. This dataset is used to validate denoising performance and cross-subject generalization under ambulatory conditions. The clean reference ECG for each subject was constructed based on the real-time RR intervals: P-QRS-T waveforms are generated from resting-state ECG signals, with event widths and amplitudes adaptively adjusted according to heart rate variations (Agostinelli et al., 2016; Nasim et al., 2019; Altamira-Colado et al., 2021).

3.1.3 Evaluation protocols

The MIT-BIH dataset is used for standardized SOTA comparison under a random split protocol consistent with previous studies. The BUT dataset is used for cross-subject evaluation, where 35 eligible subjects are divided into 27 for training (with 3 of these used for validation) and 8 for testing, to evaluate the model’s generalization ability toward unseen individuals.

3.2 Evaluation metrics

The denoising performance is evaluated using six metrics: three conventional metrics (, PRD, RMSE) and three morphological-temporal fidelity metrics (, ΔPR, ). The conventional metrics follow the standard formulations in (Zhang et al., 2025); the morphological-temporal fidelity metrics are defined in Section 2.4.

3.3 Implementation details

All experiments were conducted on a 64-bit x64 workstation with an Intel i9-14900KF CPU (3.20 GHz), 64 GB RAM, and an NVIDIA RTX 4080 SUPER GPU (16 GB VRAM), with CUDA 12.6 enabled for GPU acceleration. Each input segment consisted of 1250 samples, corresponding to a 5 s ECG signal sampled at 250 Hz.

Training was performed using the AdamW optimizer with , , and . A ReduceLROnPlateau learning rate scheduler (; ) and an early stopping strategy () were adopted throughout the training process. The loss function MorphologyLoss was utilized with weighing coefficients set as: , , . Only the checkpoint with optimal validation performance was retained for subsequent testing. All hyperparameters were optimized solely on the validation set, and the test set was used only for final performance evaluation.

The model contains 2.6 M trainable parameters with a storage size of 9.93 MB. The total computational cost is 1012.60 MMACs per 1250-point segment (5 s at 250 Hz), with the long-receptive-field branch accounting for approximately 72% of the total. On a single CPU core, the average inference latency is 16.58 ms per segment, confirming the feasibility of real-time inference on wearable devices.

3.4 Experiment 1: MIT-BIH graded additive noise

3.4.1 Experimental protocols

All samples were randomly shuffled and split into training, validation, and test sets at an 80%/10%/10% ratio. This partitioning scheme was exclusively applied to the evaluation of the presented model and its ablation variants in this experiment.

3.4.2 Comparison with SOTA methods

Under the random-split protocol, FG-DPUNet was compared with four SOTA methods. Comparisons were conducted in two stages. First, FG-DPUNet and 2D-CEDNet were evaluated under six levels using the three conventional metrics. Second, the average performance across six SNR levels was compared between FG-DPUNet and four SOTA methods (2D-CEDNet, 2D-SCEDNet, MrSeNet, and TCDAE) based on the same metrics.

3.4.3 Ablation study on MIT-BIH

To assess the contribution of each architectural component, five model variants were evaluated under the identical random-split protocol using the three conventional metrics and the morphological-temporal fidelity metrics:

SUNet-k7: Single-branch U-Net with kernel size

SUNet-k21: Single-branch U-Net with kernel size

AvgFusion: Dual-branch network with fixed averaging fusion

DP-UNet: Dual-branch network with concatenation-based learnable fusion

FG-DPUNet: The presented full model

3.5 Experiment 2: BUT real-world evaluation

3.5.1 Cross-subject validation protocol

To evaluate generalization to unseen individuals, the BUT dataset was split on a per-subject basis: 27 subjects for training (with three of these used for validation) and eight for testing, which eliminated any subject overlap between training and test sets.

3.5.2 Quantitative comparison of ablation variants

The same five model variants introduced in Section 3.4.3 were evaluated on the BUT dataset following the cross-subject protocol, with performance metrics reported separately for each activity category (walking, coughing, laughing, talking, and resting).

3.5.3 Gating adaptivity analysis

To verify whether the feature-aware gating module dynamically adjusted dual-branch fusion weights according to motion artifact severity, we extracted gate outputs of FG-DPUNet from the BUT test set and grouped all samples into five activity groups. Spearman’s rank correlation between instantaneous input SNR and gate values was computed within each group to quantify the monotonic relationship between noise intensity and gating responses. Non-parametric Mann-Whitney U tests were further employed to compare gate value distributions between resting-state baseline and each activity category, to assess whether motion artifacts induce significant gate value increases.

Furthermore, three representative recordings covering resting, mild walking, and heavy walking scenarios were visualized, including raw noisy ECG traces, clean reference signals, denoised waveforms, real-time gate responses, and local SNR temporal curves, to intuitively demonstrate the adaptive gating behavior of the presented module.

3.5.4 Analysis protocol of morphological-temporal fidelity metrics

To verify the complementary value of , ΔPR, and , a pairwise correlation analysis of six evaluation indices was performed on FG-DPUNet denoised outputs from the BUT test set. A full Pearson correlation matrix and three scatter plots were computed to analyze the associations between traditional amplitude metrics and the presented waveform indices, as well as the relationships of with PRD, , and .

4 Results

4.1 MIT-BIH graded noise experiment

4.1.1 Comparison with SOTA methods

Performance comparisons between FG-DPUNet and 2D-CEDNet under six input SNR () levels are shown in Table 2. The presented FG-DPUNet outperforms 2D-CEDNet across all noise levels in terms of and PRD. Under the most challenging low condition (−6 dB), FG-DPUNet achieves an of 21.61 dB compared with 19.54 dB for 2D-CEDNet, and reduces PRD from 89.59% to 18.91%, corresponding to performance improvements of 2.07 dB and 70.68%, respectively. At 24 dB, where the input signal is already relatively clean, FG-DPUNet yields a positive of 1.85 dB, whereas 2D-CEDNet yields a negative of −1.66 dB. Meanwhile, FG-DPUNet achieves a considerably lower PRD of 5.77% compared with 77.20% for 2D-CEDNet, demonstrating that the presented method avoids excessive signal oversmoothing even under mild noise conditions.

TABLE 2

Metric2D-CEDNetFG-DPUNet
-6dB (dB)19.5421.61 ± 2.64
PRD (%)89.5918.91 ± 6.37
RMSE (mV)0.080.08 ± 0.03
0dB (dB)16.6819.03 ± 2.37
PRD (%)86.5912.41 ± 3.59
RMSE (mV)0.080.05 ± 0.02
6dB (dB)12.1415.4 ± 2.32
PRD (%)84.229.42 ± 2.68
RMSE (mV)0.060.04 ± 0.01
12dB (dB)7.3811.25 ± 2.35
PRD (%)81.287.69 ± 2.58
RMSE (mV)0.050.03 ± 0.01
18dB (dB)2.526.99 ± 2.29
PRD (%)78.596.22 ± 1.74
RMSE (mV)0.040.02 ± 0.01
24dB (dB)−1.661.85 ± 2.68
PRD (%)77.205.77 ± 2.56
RMSE (mV)0.040.02 ± 0.01

Performance comparison under different SNRin levels (−6 dB–24 dB) on MIT-BIH.

Table 3 presents the average performance of FG-DPUNet and four SOTA methods across all levels for comprehensive comparison. The presented FG-DPUNet achieves the highest average of 12.69 dB and the lowest average PRD of 10.07%, outperforming all competing methods. In terms of RMSE, FG-DPUNet obtains a value of 0.04 mV, which is identical to 2D-SCEDNet and lower than that of the other compared models. Notably, several SOTA methods lack partial metric results: MrSeNet does not report the PRD value, and TCDAE provides no result. In contrast, FG-DPUNet yields complete and optimal results across all three conventional metrics. Although FG-DPUNet achieves significant advantages in and PRD, its RMSE is comparable to the RMSE values of 2D-SCEDNet and 2D-CEDNet. This phenomenon is mainly attributed to the distinct amplitude scaling strategy adopted in 2D-CEDNet, which involves converting 1D ECG signals to RGB images for processing and then mapping the network output back to the 1D domain.

TABLE 3

Method (dB)PRD (%)RMSE (mV)
2D-CEDNet (Ullah et al., 2025)6.7563.870.05
2D-SCEDNet (Ullah et al., 2025)7.1062.960.04
MrSeNet (Wang Z. et al., 2025)8.71--0.28
TCDAE (Chen et al., 2024)--22.6230.06
FG-DPUNet12.6910.070.04

Comprehensive comparison with SOTA methods.

Bold values indicate the best results achieved by competing methods.

4.1.2 Ablation study on MIT-BIH

To quantify the independent contribution of each core module, controlled ablation experiments are conducted with five incrementally modified variants: single-branch U-Net with kernel size (SUNet-k7), single-branch U-Net with kernel size (SUNet-k21), dual-branch network with fixed averaging fusion (AvgFusion), dual-branch network with concatenation-based learnable fusion (DP-UNet), and the presented full model (FG-DPUNet). Figure 3 illustrates six metrics covering ranging from −6 dB to 24 dB, including three standard denoising metrics (RMSE, PRD, ) and three morphological-temporal fidelity metrics (, ΔPR, ). Lower RMSE, PRD, , , and correspond to better waveform preservation; a higher denotes stronger noise suppression.

FIGURE 3

Single-branch U-Nets show inferior performance across all levels. SUNet-k7 yields the worst overall metrics: at dB, dB, mV. Its short receptive field is unable to capture broad-scale motion P/T wave features under noise, producing larger waveform distortion (; ms) than all dual-branch models. SUNet-k21 reduces P/T waves distortion to some extent (; ms at 0 dB) but delivers weaker noise suppression, with worse RMSE, PRD and relative to dual-branch architectures.

All dual-branch models outperform single-branch baselines on all measured metrics. AvgFusion with fixed equal-weight fusion exhibits higher reconstruction error and waveform distortion than learnable fusion variants at every SNR. DP-UNet achieves better conventional denoising metrics under the lowest (−6 dB): dB and , slightly superior to FG-DPUNet’s dB and . For from 0 dB to 24 dB, FG-DPUNet obtains competitive or better RMSE, PRD and . At 0 dB, FG-DPUNet reaches dB and , outperforming SUNet-k7, SUNet-k21 and AvgFusion. In terms of waveform morphological-temporal fidelity metrics, FG-DPUNet records smaller ΔPR and ΔRT across nearly all points. At −6 dB, FG-DPUNet’s is marginally higher than DP-UNet’s , yet its interval deviations are lower ( ms vs. ms; ms vs. ms).

Overall ablation observations confirm that dual-branch structures outperform single-scale U-Nets. Among dual-branch schemes, FG-DPUNet delivers superior waveform interval fidelity across most conditions, despite minor disadvantages in standard denoising metrics and under heavy synthetic noise at −6 dB.

4.2 BUT real-world evaluation

4.2.1 Quantitative comparison of ablation variants

The five model variants are evaluated on the BUT dataset under a cross-subject partition protocol (27 training, three validation, and eight unseen test subjects), with performance across five daily activity categories visualized in Figure 4. Evaluation metrics follow prior definitions: smaller RMSE, PRD, , ΔPR, and represent better waveform fidelity, while a higher indicates stronger noise suppression.

FIGURE 4

Single-branch SUNet-k7 and SUNet-k21 exhibit inferior performance across all five activity groups. Their fixed single receptive field cannot separate compound wearable motion artifacts and multi-scale ECG features, producing larger reconstruction errors and more severe P/T waves distortion relative to all dual-branch variants. All dual-branch architectures outperform single-branch baselines on all metrics. AvgFusion with fixed equal weighting underperforms learnable fusion variants for every activity. DP-UNet surpasses FG-DPUNet on conventional denoising metrics for walking samples, with mV vs. mV and dB vs. dB. For coughing, laughing, talking and resting, FG-DPUNet achieves lower RMSE and PRD as well as higher .

In terms of morphological-temporal fidelity metrics, FG-DPUNet yields the minimum across all motion categories. Subject-aggregated paired t-tests were performed, where metric values were first averaged across all segments per subject prior to pairwise comparison. No statistically significant inter-model differences were observed: (p = 0.366), (p = 0.484), and (p = 0.490). Minor gaps relative to DP-UNet are observed in partial interval metrics: FG-DPUNet has slightly larger ΔPR under walking (1.61 ms vs. 1.38 ms) and marginally higher ΔRT during coughing (14.51 ms vs. 14.04 ms). For laughing, talking and resting, FG-DPUNet attains the smallest ΔPR and ΔRT among all tested models.

Unlike the wide-ranging SNR trade-off observed under synthetic heavy noise on the MIT-BIH dataset, minor weaknesses in standard denoising metrics only emerge for walking samples in real wearable recordings from the BUT dataset. FG-DPUNet maintains competitive overall waveform fidelity across all ambulatory activities. These cross-subject ablation results verify the effectiveness of the dual-branch multi-scale design and adaptive gating module for preserving ECG waveforms, despite slight deficits in partial denoising and interval metrics during high-intensity walking and coughing.

4.2.2 Quantitative validation of gating adaptivity

To quantitatively validate the overall noise-adaptive characteristic of the feature-aware gating module, the correlation between mean gate values and across all test ECG samples from the BUT dataset is analyzed and visualized in Figure 5. For each noisy ECG segment, the mean gate value is obtained by averaging the temporal gate vector, while is calculated using paired noisy and clean ECG signals. Both Pearson linear correlation and Spearman rank correlation are adopted to quantify the association.

FIGURE 5

A prominent statistically significant negative correlation is observed on the aggregate sample set, with Pearson () and Spearman (). The observed negative correlation across all recordings indicates that the learned gate value declines as input signal quality improves. When ECG segments are contaminated by severe broad-scale motion interference with low , the model generates relatively higher gate values to strengthen the contribution of the long-receptive-field branch for artifact suppression. For high-SNR clean waveforms with minor interference, the gate value decreases accordingly, allowing the short-receptive-field branch to dominate reconstruction and retain complete clinical cardiac morphological details.

To quantitatively verify the noise-adaptive property of the feature-aware gating module, correlations between mean gate values and across five daily activity categories are analyzed on the BUT test set, with results visualized in Figure 6. For each noisy ECG segment, the mean temporal gate value and corresponding are calculated, and both Pearson linear correlation and Spearman rank correlation are evaluated for each activity group separately.

FIGURE 6

Strong statistically significant negative correlations (Pearson to and Spearman to , all ) are detected for high-motion scenarios (walking, coughing, laughing). Laughing yields the largest correlation magnitude ( and , ), revealing near-linear weight modulation of dual branches against noise severity within valid ranges. By contrast, talking exhibits a weak non-significant negative correlation ( and , ), and resting shows a trivial non-significant positive correlation ( and , ), indicating stable gate outputs under nearly noise-free static conditions.

In summary, these statistical outcomes verify that the feature-aware gating module dynamically adjusts dual-branch fusion weights according to real-time motion artifact intensity.

4.2.3 Typical waveform and gating visualization

Representative ECG segments from walking, laughing, and resting subsets of the BUT test set are selected for waveform and gate visualization in Figure 7. Each activity category occupies one column with paired subplots: the upper panel overlays the raw noisy ECG, clean reference signal, and denoised output generated by FG-DPUNet; the lower panel illustrates real-time gate values along with their temporal mean.

FIGURE 7

For walking and laughing scenarios with severe motion artifacts, the model yields higher mean gate values (0.57 and 0.55, respectively) accompanied by substantial real-time gate fluctuations. The denoised ECG waveforms closely follow the clean reference, with well-preserved P/T waves. For resting signals with slight motion interference, the gate response remains stable around a lower mean value of 0.54 with negligible fluctuations. Notably, gate values during QRS complexes remain close to 1, indicating a higher contribution from the long-receptive-field branch. The skip connections in the short-receptive-field branch preserve sufficient fine temporal QRS details, allowing the branch to retain adequate QRS fidelity. However, P/T wave morphology is more susceptible to the smoothing effect of the long receptive field, as these waveforms primarily occupy broad-scale temporal components and exhibit slower variations. The long-receptive-field branch thus remains less effective at preserving P/T wave details compared to the short-receptive-field branch.

These visual observations align with the previous correlation results: intensified motion artifacts increase gate magnitudes, thereby strengthening long-receptive-field feature contributions and enabling adaptive multi-scale fusion.

4.2.4 Correlation results of morphological-temporal fidelity metrics

The full Pearson correlation matrix of six evaluation metrics calculated on the FG-DPUNet denoised outputs is presented in Figure 8. The introduced shows a strong correlation with PRD (; ) and (; ), while exhibiting weak correlations with (; ) and (; ). These results reveal that mainly characterizes T-wave and RT-segment distortion, a feature not fully captured by conventional amplitude metrics. Notably, and are weakly correlated (; ), which shows that PR-interval and RT-interval deviations represent relatively independent patterns of distortion.

FIGURE 8

The scatter plots in Figure 9 further illustrate the intrinsic association between and other core metrics. Although and PRD are strongly correlated, substantial dispersion in values is observed at fixed PRD levels, reflecting distinct morphological information extracted by the morphological-temporal fidelity metric. Its tight correlation with confirms that RT-interval distortion dominates variations in , while the weak association with suggests that PR-interval deformation contributes only modestly to overall waveform distortion.

FIGURE 9

5 Discussion

5.1 Implications of this study

FG-DPUNet is designed to preserve P-QRS-T waveform morphology while suppressing motion artifacts in wearable ECG signals, rather than pursuing maximum SNR improvement and/or minimum RMSE. Its objective is to achieve a practical balance between noise removal and waveform fidelity.

The core structural advancement lies in the feature-aware dynamic gating module. Previous dual-branch designs generally rely on fixed-weight allocation or simple feature concatenation, which cannot adapt to time-varying motion noise and thus struggle to balance artifact suppression and waveform preservation. This motivates our adoption of a lightweight gating mechanism that provides a minimal-cost solution to the adaptive fusion problem, avoiding the computational overhead of more expressive alternatives for resource-constrained deployment. The presented mechanism adaptively prioritizes the long-receptive-field branch under heavy motion artifacts and the short-receptive-field branch under low-noise conditions. This adaptivity is statistically validated on the BUT dataset, where gate values show strong negative correlations with SNR in high-motion scenarios ( to , ) and remain stable at rest (; ). While this confirms the adaptivity of the gating module, it does not by itself demonstrate the adequacy or necessity of this specific gating formulation against more complex alternatives. Compared with static DP-UNet, the dynamic fusion mitigates the trade-off between noise removal and detail retention, yielding more robust performance on real wearable data. These results provide evidence that the lightweight gating design can improve waveform-preservation performance relative to the static and non-gated fusion strategies, as reflected by generally lower across all motion categories. A minor trade-off is a ∼0.23 ms increase in P-wave deviation, which stays within a set tolerance threshold ( ms). Meanwhile, FG-DPUNet shows improvement over DP-UNet in and in terms of overall morphological-temporal fidelity. It should also be noted that DP-UNet yields slightly lower RMSE and higher in the walking scenario. Conversely, FG-DPUNet exhibits marginally larger during walking and higher during coughing. These observations suggest that the gating mechanism, while beneficial for overall waveform fidelity, may introduce minor trade-offs under specific high-motion conditions.

Beyond network architecture optimization, this study further highlights the necessity of waveform-fidelity-oriented evaluation in wearable ECG denoising. Traditional metrics including , PRD, and RMSE only reflect global amplitude errors while ignoring subtle morphological distortions, such as those in diagnostically vital P and T waves. Such evaluation mechanisms tend to produce numerically high SNR but morphologically distorted denoised results. This greatly limits the practical deployment of existing denoising methods. By introducing morphological-temporal fidelity metrics , , and , this work provides a waveform-aware evaluation approach. As shown in Figures 3, 4, FG-DPUNet achieves lower waveform distortion than DP-UNet across most motion scenarios, particularly in resting conditions where conventional metrics show only modest differences. For instance, in resting scenarios, conventional metrics place the two models on par (: 3.36 vs. 4.19 dB; PRD: 16.21% vs. 14.20%). Morphological-temporal fidelity metrics, however, highlight distinct performance differences: FG-DPUNet achieves a 28.0% reduction in (from 0.0429 to 0.0309) and a 2.77 ms shorter (from 13.32 to 10.55 ms). The correlation analysis in Section 4.2.4 further substantiates this argument. is strongly correlated with (; ) but weakly correlated with (; ). This confirms that captures T-wave morphological fidelity independently of SNR improvement. However, it is important to note that this correlation analysis was performed at the segment level. For inter-model comparison, subject-aggregated evaluation accounting for within-subject dependence revealed no statistically significant differences for , , or . While these metrics are sensitive to waveform distortions, their clinical validity and diagnostic utility remain unestablished in this study. The lack of detectable statistical separation between models may be related to moderate inter-subject performance variability and the small cohort of test subjects. Meanwhile, the weak correlation between and (; ) suggests that P-wave and T-wave distortions are independent phenomena, necessitating separate evaluation. These results indicate that conventional global metrics can create a misleading impression of equivalent performance. In contrast, morphological-temporal fidelity metrics capture differences that are relevant to the assessment of morphological preservation.

In terms of cross-domain generalization, this analysis explains why state-of-the-art models suffer performance degradation in practical wearable scenarios. Most existing methods are optimized and validated on datasets where noisy signals are synthesized by adding real noise sources from the MIT-BIH NSTDB to clean ECG recordings. This linear and additive construction fundamentally differs from the non-stationary, non-linear, and non-additive nature of real-world motion artifacts. Under ideal additive noise conditions, fixed-weight dual-branch structures achieve performance comparable to FG-DPUNet. However, on the complex BUT dataset containing authentic human motion interference, static fusion fails to dynamically adjust branch contributions according to instantaneous noise intensity, leading to an inevitable trade-off between artifact removal and waveform fidelity. Equipped with adaptive gating modulation, FG-DPUNet achieves an optimal balance between global noise suppression and subtle waveform preservation. This makes it a promising candidate for long-term ambulatory ECG monitoring application.

In summary, FG-DPUNet is a scenario-oriented denoising model tailored for irregular human motion artifacts in wearable ECG signals. Different from general models optimized only for synthetic benchmarks, it exhibits distinct advantages in handling real-world dynamic interference. This study not only verifies the effectiveness of dynamic gated dual-branch fusion but also offers a practical paradigm for high-fidelity, waveform-aware wearable ECG denoising research.

5.2 Limitations and future work

Despite its promising performance, FG-DPUNet has several limitations that suggest future research directions. The model is optimized for motion artifact removal but has not been validated on extreme noise such as electrode detachment or sensor contact noise. This leaves room for robustness improvement under complex clinical conditions. Future work will focus on two directions: (1) enriching the dataset with multi-category mixed noises and conducting cardiologist validation of the presented morphological-temporal fidelity metrics to further assess their clinical relevance; and (2) applying quantization and pruning to the current lightweight design for deployment on low-power embedded hardware with real-time denoising validation.

6 Conclusion

FG-DPUNet is presented to tackle ECG waveform distortion induced by motion artifacts in wearable ambulatory ECG monitoring. The framework integrates heterogeneous dual branches to extract multi-scale ECG features in parallel. It also adopts a feature-aware dynamic gating mechanism built on shared shallow features to balance noise suppression and waveform integrity. Complementarily, we define , and , providing a waveform-aware quantitative evaluation approach for diagnostic waveform recovery. On the MIT-BIH dataset under standard additive noise conditions, FG-DPUNet achieves superior overall performance against SOTA models. Cross-subject experiments on the BUT dataset with authentic motion artifacts demonstrate that the dynamic gating structure improves the reconstruction accuracy of P-QRS-T complex morphology. The model contains only 2.6 M parameters, enabling lightweight edge deployment. In summary, this work delivers an integrated solution for wearable ECG denoising that prioritizes the preservation of diagnostic waveform morphology. Future research will expand datasets covering diverse mixed noise categories and realize full deployment on embedded systems.

Statements

Data availability statement

Publicly available datasets were analyzed in this study. This data can be found here: https://doi.org/10.13026/C2F305https://doi.org/10.13026/C2HS3Thttps://doi.org/10.13026/gxwp-p940.

Author contributions

BL: Methodology, Visualization, Formal Analysis, Writing – original draft, Data curation, Software, Conceptualization, Validation. SW: Software, Validation, Writing – review and editing, Formal Analysis. CY: Methodology, Writing – review and editing, Validation. YX: Resources, Conceptualization, Project administration, Writing – review and editing. YS: Funding acquisition, Supervision, Writing – review and editing. XY: Methodology, Conceptualization, Supervision, Writing – review and editing, Funding acquisition, Resources.

Funding

The author(s) declared that financial support was received for this work and/or its publication. This work was supported by the Prevention and Control of Emerging and Major Infectious Diseases-National Science and Technology Major Project (2025ZD01902300) and the Anhui Province Science and Technology Innovation Program (202423l10050047).

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1

    AgostinelliAngelaSbrolliniA.GiulianiC.FiorettiS.Di NardoF.et al (2016). Segmented beat modulation method for electrocardiogram estimation from noisy recordings. Med. Eng. and Phys.38 (6), 560–568. 10.1016/j.medengphy.2016.03.011

  • 2

    Altamira-ColadoE.Bravo-ZanogueraM.Cuevas-GonzálezD.Reyna-CarranzaM.López-AvitiaR. (2021). Removal of ECG baseline wander using a resting cycle template. Eng. Proc.10, 11289. 10.3390/ecsa-8-11289

  • 3

    AnX.LiuY.ZhaoY.LuS.StyliosG. K.LiuQ. (2022). Adaptive motion artifact reduction in wearable ECG measurements using impedance pneumography signal. Sensors Basel, Switz.22 (15), 5493. 10.3390/s22155493

  • 4

    ArifS.SiddiquiM. M.AlqahtaniN.ShahK.KhanM. A.KraiemN. (2025). Advanced signal processing algorithm-based state estimation and denoising of ECG signals for biomedical applications. IEEE Access13, 137240–137254. 10.1109/ACCESS.2025.3595453

  • 5

    ChenM.LiY.ZhangL.LiuL.HanB.ShiW.et al (2024). Elimination of random mixed noise in ECG using convolutional denoising autoencoder with transformer encoder. IEEE J. Biomed. Health Inf.28 (4), 1993–2004. 10.1109/JBHI.2024.3355960

  • 6

    ChoiS. H.KangM.KwonK.-W.KimJ. S.JungJ.-M.LeeH.-M. (2025). An energy-efficient real-time wearable ECG acquisition system with near-data prediagnosis for Mobile applications. IEEE Trans. Instrum. Meas.74, 1–11. 10.1109/TIM.2025.3547110

  • 7

    CuiM.GanY.ChenJ.ZhengK.XieY.HuD.et al (2026). A frequency-guided denoising framework based on convolutional transformer for electrocardiogram signals. Eng. Appl. Artif. Intell.175 (July), 114570. 10.1016/j.engappai.2026.114570

  • 8

    FangY.WangC.RenY.XuF. (2024). ECG signal fusion reconstruction via hash autoencoder and margin semantic reinforcement. J. King Saud Univ. - Comput. Inf. Sci.36 (6), 102124. 10.1016/j.jksuci.2024.102124

  • 9

    FangS.ZhangH.WangZ.LiuY.LiQ.ZhangP. (2025). “TDD: task-aware diffusion dual-matching for robust electrocardiogram denoising,” in 2025 IEEE International Conference on Bioinformatics and Biomedicine (BIBM), December, 4993–5000. 10.1109/BIBM66473.2025.11356935

  • 10

    GaldosJ.Lopez ColqueN.Medina RodirguezA.Huarca QuispeJ.RendulichJ.EspinozaE. S. (2024). Comparison and evaluation of lms-derived algorithms applied on ecg signals contaminated with motion artifact during physical activities. Appl. Comput. Sci.20 (1), 157–172. 10.35784/acs-2024-10

  • 11

    GoldbergerA. L.AmaralL. A. N.GlassL.HausdorffJ. M.IvanovP. C.MarkR. G.et al (2000). PhysioBank, PhysioToolkit, and PhysioNet. Circulation101 (23), e215–e220. 10.1161/01.CIR.101.23.e215

  • 12

    HassoonB. A.XiongS.HassonM. A.AbdulsalamiA. O. (2025). Leveraging conditional diffusion and pruning for optimal ECG signal denoising. Comput. Biol. Med.193 (July), 110341. 10.1016/j.compbiomed.2025.110341

  • 13

    HoJ.JainA.AbbeelP. (2020). Denoising diffusion probabilistic models. Adv. Neural Inf. Process. Syst.33, 6840–6851. 10.48550/arXiv.2006.11239

  • 14

    HuL.CaiW.ChenZ.WangM. (2024). A lightweight U-Net model for denoising and noise localization of ECG signals. Biomed. Signal Process. Control88 (February), 105504. 10.1016/j.bspc.2023.105504

  • 15

    HusseinA. F.MohammedW. R.JaberM. M.KhalafO. I. (2022). An adaptive ECG noise removal process based on empirical mode decomposition (EMD). Contrast Media and Mol. Imaging2022 (1), 3346055. 10.1155/2022/3346055

  • 16

    KligfieldP.GettesL. S.BaileyJ. J.ChildersR.DealB. J.HancockE. W.et al (2007). Recommendations for the standardization and interpretation of the electrocardiogram. JACC49 (10), 1109–1127. 10.1016/j.jacc.2007.01.024

  • 17

    LiZ.TianY.JinY.WeiX.WangM.LiuJ.et al (2025). EDDM: a novel ECG denoising method using dual-path diffusion model. IEEE Trans. Instrum. Meas.74, 1–15. 10.1109/TIM.2025.3542875

  • 18

    MarzogH. A.AbdH. J.YonisA. Z. (2022). “Noise removal of ECG signal using multi-techniques,” in 2022 IEEE Integrated STEM Education Conference (ISEC), March, 397–403. 10.1109/ISEC54952.2022.10025094

  • 19

    MidaniW.LtifH.AyedM. B. (2025). “A comprehensive review of noise removal techniques in ECG signal processing,” in Procedia Computer Science, 29th International Conference on Knowledge-Based and Intelligent Information and Engineering Systems (KES 2025), 4134–4143. 10.1016/j.procs.2025.09.538

  • 20

    MoodyG. B.MarkR. G. (2001). The impact of the MIT-BIH arrhythmia database. IEEE Eng. Med. Biol. Mag.20 (3), 45–50. 10.1109/51.932724

  • 21

    NasimA.PintiF.GentiliA.BelliA.PalmaL.PierleoniP. (2019). “Dynamic segmented beat modulation method for denoising ECG data from wearable sensors,” in 2019 International Conference on Sensing and Instrumentation in IoT Era (ISSI), August, 1–4. 10.1109/ISSI47111.2019.9043650

  • 22

    NemcovaA.SmisekR.VargovaE.MaršánováL.VitekM.SmitalL. (2021). Brno university of technology smartphone PPG database (BUT PPG). PhysioNet. 10.13026/7vy8-av04

  • 23

    ParmonanganI. H.FernandoT.DenmanS. (2025). “Robust ECG classification using mamba and self-supervised representation learning,” in 2025 47th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), July 14, 1–7. 10.1109/EMBC58623.2025.11253901

  • 24

    PrakashG.ThomasL. (2025). “Investigating slow multiplicative noise and its influence on ECG signals,” in 2025 International Conference on Emerging Trends in Signal Processing and Computational Intelligence (ICCSPCI), December, 1–5. 10.1109/ICCSPCI68199.2025.11451674

  • 25

    RautaharjuP. M.SurawiczB.GettesL. S. (2009). AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram. JACC53 (11), 982–991. 10.1016/j.jacc.2008.12.014

  • 26

    ShaheenA.YeL.KarunaratneC.SeppänenT. (2025). Fully-gated denoising auto-encoder for artifact reduction in ECG signals. Sensors25 (3), 801. 10.3390/s25030801

  • 27

    SkoricJ.D’MelloY.PlantD. V. (2024). A wavelet-based approach for motion artifact reduction in ambulatory seismocardiography. IEEE J. Transl. Eng. Health Med.12, 348–358. 10.1109/JTEHM.2024.3368291

  • 28

    SulimanA.FarahmandM.GaleottiL.ScullyC. G. (2025). Clinical evaluation of the ECG noise extraction tool as a component of ECG analysis algorithms evaluation. IEEE Trans. Biomed. Eng.72 (7), 2062–2071. 10.1109/TBME.2024.3386493

  • 29

    UllahH.HeyatMd B. B.BiswasT.NehaN. I.RaihanMd. M. S.LaiD. (2025). An end-to-end motion artifacts reduction method with 2D convolutional de-Noising auto-encoders on ECG signals of wearable flexible biosensors. Digit. Signal Process.160 (May), 105053. 10.1016/j.dsp.2025.105053

  • 30

    VaswaniA.ShazeerN.ParmarN.UszkoreitJ.JonesL.GomezA. N.et al (2017). Attention is all you need. Adv. Neural Inf. Process. Syst.30. 10.48550/arXiv.1706.03762

  • 31

    WangS.LeiLi (2025). TF-TransUNet1D: time-frequency guided transformer U-Net for robust ECG denoising in digital twin. arXiv:2508.20398. 10.48550/arXiv.2508.20398

  • 32

    WangK.FengB.ZhouF.ZhongC.QiaoJ.LuoJ.et al (2025). QRS complexes detection using UNet-BiLSTM model combined with adaptive post-processing for multilead wearable ECG smart T-Shirt system. IEEE Internet Things J.12 (16), 33437–33454. 10.1109/JIOT.2025.3574944

  • 33

    WangS.TangH.HimenoR.Solé-CasalsJ.CaiafaC. F.AokiS.et al (2025). ECGDenoiser: a magnitude-aware deep learning framework with phase retrieval for electrocardiogram signal enhancement. Circuits, Syst. Signal Process.44 (10), 7526–7548. 10.1007/s00034-025-03142-5

  • 34

    WangZ.XieH.LiuY.ZhuH.ZhangH.WangZ.et al (2025). MrSeNet: electrocardiogram signal denoising based on multi-resolution residual attention network. J. Electrocardiol.89 (March), 153858. 10.1016/j.jelectrocard.2024.153858

  • 35

    WuY.XieY. X.GeH. (2022). “An improved adaptive filtering algorithm for ECG signals,” in 2022 7th International Conference on Computer and Communication Systems (ICCCS), April, 354–357. 10.1109/ICCCS55155.2022.9846659

  • 36

    XiangX.ChenC.LiuR.LiuZ. (2024). “ECG signal denoising based on multi-scale residual dense U-Net,” in Proceedings of the 2023 4th International Symposium on Artificial Intelligence for Medicine Science (New York, NY, USA: ISAIMS ’23), 213–219. 10.1145/3644116.3644155

  • 37

    YuX.LiJ. (2022). Adaptive kalman filtering for recursive both additive noise and multiplicative noise. IEEE Trans. Aerosp. Electron. Syst.58 (3), 1634–1649. 10.1109/TAES.2021.3117896

  • 38

    YuenB.DongX.LuT. (2020). Detecting noisy ECG QRS complexes using WaveletCNN autoencoder and ConvLSTM. IEEE Access8, 143802–143817. 10.1109/ACCESS.2020.3012904

  • 39

    ZhangX.ChenM.LiY.GaoJ.SunY.LiuF.et al (2024). “A dual-path interactive ECG denoising algorithm,” in 2024 17th International Congress on Image and Signal Processing, BioMedical Engineering and Informatics (CISP-BMEI), October, 1–6. 10.1109/CISP-BMEI64163.2024.10906137

  • 40

    ZhangJ.GuoYuDongX.WangT.WangJ.MaX.et al (2025). Opportunities and challenges of noise interference suppression algorithms for dynamic ECG signals in wearable devices: a review. Measurement250 (June), 117067. 10.1016/j.measurement.2025.117067

  • 41

    ZhuP.FengL.YuK.ZhangY.WenC.JianzhongH. (2025). Enhanced DWT for denoising heartbeat signal in non-invasive detection. Sensors Basel, Switz.25 (6), 1743. 10.3390/s25061743

Summary

Keywords

dual-path U-Net, ECG denoising, feature gating, motion artifact, waveform integrity

Citation

Lv B, Wang S, Yu C, Xu Y, Sun Y and Yang X (2026) FG-DPUNet: a feature-aware gating dual-path U-Net for ECG motion artifact removal with waveform integrity. Front. Signal Process. 6:1945432. doi: 10.3389/frsip.2026.1945432

Received

22 July 2026

Revised

24 August 2026

Accepted

21 September 2026

Published

04 November 2026

Volume

6 - 2026

Edited by

Amine NAIT-ALI, Université Paris-Est Créteil Val de Marne, France

Reviewed by

Somayeh Afrasiabi, Shiraz University, Iran

Muhammad Fauzan Edy Purnomo, Universitas Brawijaya Jurusan Teknik Elektro, Indonesia

Updates

Copyright

*Correspondence: Xianjun Yang,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics