Abstract
Introduction:
Current clinical practice suggests using patellar and quadriceps tendon autografts with a 10ā mm diameter for ACL reconstruction. This can be problematic for patients with smaller body frames. Our study objective was to determine the minimum diameter required for these grafts. We hypothesize that given the strength and stiffness of these respective tissues, they can withstand a significant decrease in diameter before demonstrating mechanical strength unviable for recreating the knee's stability.
Methods:
We created a finite element model of the human knee with boundary conditions characteristic of the Lachman test, a passive accessory movement test of the knee performed to identify the integrity of the anterior cruciate ligament (ACL). The Mechanical properties of the model's grafts were directly obtained from cadaveric testing and the literature. Our model estimated the forces required to displace the tibia from the femur with varying graft diameters.
Results:
The 7ā mm diameter patellar and quadriceps tendon grafts could withstand 55ā60ā N of force before induced tibial displacement. However, grafts of 5.34- and 3.76-mm diameters could only withstand upwards of 47ā N and 40ā N, respectively. Additionally, at a graft diameter of 3.76ā mm, the patellar tendon experienced 234% greater stiffness than the quadriceps tendon, with similar excesses of stiffness demonstrated for the 5.34- and 7-mm diameter grafts.
Conclusions:
The patellar tendon provided a stronger graft for knee reconstruction at all diameter sizes. Additionally, it experienced higher maximum stress, meaning it dissociates force better across the graft than the quadriceps tendon. Significantly lower amounts of force were required to displace the tibia for the patellar and quadriceps tendon grafts at 3.76- and 5.34-mm graft diameters. Based on this point, we conclude that grafts below the 7ā mm diameter have a higher chance of failure regardless of graft selection.
Introduction
The ACL originates at the medial side of the lateral femoral condyle and runs an oblique course through the intercondylar fossa. It is the primary restraint to anterior tibial translation and secondary restraint to internal rotation of the weight-bearing and non-weight-bearing knee (1). An anterior cruciate ligament (ACL) injury can result in significant impairment, especially in highly active individuals. Numerous graft sources have been described during the evolution of ACL reconstruction surgery (2, 3). Traditionally, the surgical approach to ACL reconstruction favored the use of allografts or autografts from the patellar tendon, although the hamstrings, and to a lesser degree, the quadriceps tendon have become popular grafts amongst many surgeons and institutions throughout the years (4). Many studies have been conducted to investigate the biomechanical properties of these grafts to guide surgeons in choosing the graft most suitable for their patients (5, 6). More specifically, these studies define biomechanical properties such as tensile strength, elasticity, stiffness, stress and strain responses, and the anatomical measures of these individual grafts. These measures were simulated by the model employed in the study to adequately predict, based on graft selection, how the reconstructed joint functions in consideration of the geometric dimensions employed when designing these individual grafts.
One main limitation of the literature regarding this topic is the need for more data for the quadriceps tendon graft compared to the other graft sources. Additionally, a 2022 systematic review noted a need for studies investigating the effects of diameter on graft selection (4). Considering this second point, two studies investigating this question found that smaller diameter patellar and hamstring tendon autografts are associated with weaker tensile strength and a greater risk of failure (7). However, neither of these studies investigated how the diameters of these grafts would affect the knee mechanics in its working environment. More importantly, no studies have been conducted to determine how diameter affects the strength of the quadriceps tendon. This can be attributed to the fact that the quadriceps were the latest grafts to be popularized by Orthopaedic surgeons (4). Therefore, this study aimed to investigate how diameter affects the biomechanical properties of the quadriceps graft while also comparing it to the āgold standardā patellar tendon graft (8, 9). We hypothesize that when changing the diameters of the quadriceps and patellar grafts, these grafts will tolerate a significant decrease in diameter before they become too weak to meet mechanical requirements for ACL reconstruction, with the patellar tendon capable of withstanding a more substantial reduction in diameter when compared to the quadriceps. Furthermore, we introduced a controlled Lachman test as a measure of knee stability similar to the passive accessory movement test of the knee performed to identify the integrity of the anterior cruciate ligament (ACL). Finally, we aimed to determine how changes in the dimensions of the grafts affect their interaction with various bony and ligamentous features of the reconstructed joint, with an emphasis placed on the graft-epicondylar interface.
Methods
Patellar and quadriceps autografts mechanical testing
The extensor apparatus was harvested from fresh-frozen, non-irradiated human knees from male and female donors aged 70ā90 without any musculoskeletal pathology. Seven quadriceps tendon-patella-patellar ligament-tibial tuberosity complexes were obtained. Each complex was further divided into two complexes (i) a segment of the quadriceps tendon with the proximal half of the patella attached (QT-P); and (ii) the distal half of the patella, the patellar ligament, and a distal insertion of the tibia (P-PL-T) by performing a transverse cut at the midline of the patella using an oscillating saw following a methodology already used by Staubli (10). The exposed tendon and ligament were wrapped with 0.9% NS-soaked gauze throughout the harvesting and placement procedure. To ensure a proper fixation for the experimental testing, the distal ends of each complex were cemented in a metal box with polyester resin. Stabilizing wires were placed at a perpendicular angle through the longitudinal axis of the tendon complex, and the ends of the stabilizing wires were clamped securely to both sides of the box. This procedure was performed for the P-PL-T for both ends (FigureĀ 1A). For the QT-P complex, a Krackow suture was placed within the free tendon end with 3ā0ā PDS and then secured to the stabilizing wires using multiple suture loops (FigureĀ 1B). The metal boxes were clamped to the materials testing machine (Instron, Instron Corporation, Canton, MA, USA) through a wire loop and a screw. Before testing, 250ā N was applied to secure the sutures. Uniaxial tensile testing was performed at an 18ā mm/min extension rate.
Figure 1
Using a digital caliper, the exposed tendons and ligaments were marked in the thinner cross-section to measure thickness and width with an initial load of 5ā N. The initial length of the tendons and ligaments was an average of 38.5ā mm, measured between cemented extremities for the patellar ligament and between the cemented extremity and the suture for the quadriceps tendon. The width was measured in this minor position, and the two adjacent sections were spaced along one-third of the tendon length. The thickness was measured at three equidistant points on these reference positions. The cross-sectional area was evaluated as the width multiplication and the tissue's average thickness. After testing the entire segment, lateral portions were removed to have two remaining sequential samples of two-thirds and one-third in width. On the central one-third, measurements were taken until failure was reached, defined as either separating the tendon complex from the metal housing or separating the housing container from the arm of the machine.
Stress was determined by dividing the force by the corresponding cross-sectional area measurement (see FigureĀ 6). In addition, the Modulus of the tissues was evaluated using linear regression on the obtained measurements.
Finite element model for the Lachman test
The effective mechanical behavior of the graft and the composition of both material characterization and tensioning are evaluated in a computer-simulated model that best approximates the knee biomechanics. This model was based on the Lachman Test, the most widely used clinical test to determine if the ACL has been torn (11). When building our model, we needed to recreate the structural properties of these grafts by assigning them different inputs for their intrinsic and extrinsic properties according to preselected Young's Modulus values and diameters.
Given the limitations of using cadavers, many labs report various values for the biomechanical properties of the individual grafts being tested (12). To balance these variations in the literature, we chose to use a range of Young's Modulus values for both the patellar tendon and quadriceps tendon grafts to average the results generated from our model to gain a more accurate picture of the investigated effects. We chose to include Young's Modulus values from the literature for the patella tendon of 340.0ā MPa (13), 507.4ā MPa (14), 565.9ā MPa (15), and 597.4ā MPa (16), and a value of 263.4ā MPa (15) for the quadriceps tendon.
In choosing diameters to test in our experiment, we followed guidelines in the literature, stating that appropriate graft sizes for ACL reconstruction need to restore up to 85%ā115% of the native ACL total cross-sectional area (17). We chose our largest diameter of 7ā mm based on an experiment that looked at the effect of diameter on the tensile strength of the patellar tendon at diameters of 7, 10, and 15ā mm (18). In line with this past study, we further investigated by starting at the lowest data point used in prior studies and expanding to even smaller diameters. Additionally, this value matches the average diameter of the native ACL mid-substance of 7ā mm (1). The next smallest diameter of 5.34ā mm was chosen based on a value of 5.4ā mm diameter reported for the mid substance of the ACL, based on an analysis of female and older age cadavers (average age of 74.8 years) (19). This would allow us to analyze a potential smaller graft for patients with smaller knee geometry. Finally, we chose our smallest diameter of 3.76ā mm, which would still align with the required reconstructed area of the native ACL, based on a reconstruction of an ACL with a diameter of 5.4ā mm (20).
The knee bone Geometry was reconstructed from CT scan data using a BrightSpeed (GE Medical Systems) scanner with a slice thickness of 0.625ā mm and pixel size of 0.426ā mm. The CT images were then imported to Mimics (Materialise, Leuven, Belgium). The tunnels (shown in FigureĀ 2) for the ACL graft were virtually created at 90° of Flexion using a Boolean subtraction with a cylinder 8ā mm in diameter (21). Its femoral and tibial positions were based on the studies proposed by Bedi et al. and Gadikota et al. (20, 22). An expert surgeon evaluated the axial slices.
Figure 2
Main ligaments were reconstructed following anatomical landmarks on insertion footprints from the literature (23ā28). The meniscus was modeled with variable thickness to match the femoral surface in the relaxed position of the CT scan. Considering the aim of this study, the ACL graft was modeled in the knee-flexed configuration as a cylinder and characterized transversely as isotropic (29). The graft in the model is kept in place by two Softsilk 1.5 fixation titanium screws (Smith & Nephew, Andover, MA) 8ā mm in Diameter (FigureĀ 3A). All other ligaments are modeled as membrane elements reinforced by truss elements (30ā32). The meniscus was modeled as transverse isotropic with an orientation identical to the tibia axis (33). The cartilages (FigureĀ 3B) are created assuming a constant thickness of 2.23ā mm for the patella and 2.14ā mm for the femur with modulus Eā=ā10ā MPa and νā=ā0.45 (34). The cortical bone was modeled with Young's modulus of 17ā GPa (35), while the material properties of the trabecular bone were assigned from the CT data (36ā38).
Figure 3
The model was simulated, imposing an A-P translation of 15.9ā mm, equivalent to the average value reported at the anterior tibial tubercle by Christel et al. (39) for Lachman tests performed on cadaveric specimens (FigureĀ 3C). In addition, six variations in ACL cross-sections from the initial 38.5ā mm2 were implemented, considering cross-section areas of 22.4ā mm2 and 11.10ā mm2 (FigureĀ 4). The boundary conditions were applied to constrain the proximal end of the femur in all degrees of freedom. Additionally, a posterior-anterior displacement was applied to the distal end of the tibia along the projection of the anatomical axis of the femur on the plane sectioning the tibia.
Figure 4
Results
Experimental tensile properties
During mechanical testing, no specimen slippage from the fixation device occurred, and no slippage from the fixation device was observed. Patellar ligaments were generally smaller than quadriceps tendons (126.27 and 135.67ā mm2, respectively). The cross-sectional area of the patellar tendon was reduced by 10.64% of its initial cross-sectional area at L1 (first-third of the length) and approximately 34.61% at L2 (second third of the length). For the quadriceps tendon, the reduction in cross-sectional area for L1 was 9.77%.
For the young Modulus evaluated for the entire segment, the quadriceps tendon has shown a standard deviation of 76ā MPa, smaller than the 115ā MPa estimated for the patellar tendon. Still, as shown in FigureĀ 5, the range between maximal and minimal values recorded for every strain value was more significant for the quadriceps tendon than for the patellar ligament. We have found an increasing modulus with decreasing cross-section (see TableĀ 1). Specifically, the central one-third graft was a drastic 118% stiffer for the patellar ligament than the entire segment. The quadriceps tendon further demonstrates this trend (FigureĀ 6B, TableĀ 1).
Figure 5
Figure 6
Table 1
| Segment | Patellar ligament | Quadriceps tendon | ||||
|---|---|---|---|---|---|---|
| Portion | Entire segment | Central two third | Central One third | Entire segment | Central Two third | Central portion |
| Cross Section [mm2] | 126.3 | 83.1 | 48.5 | 135.7 | 105.8 | 64.3 |
| Modulus [MPa] | 409.5 | 564.6 | 893.4 | 176.7 | 330.5 | 461.8 |
Young moduli from experiments conducted in the current study.
The Patellar ligament was stiffer than the Quadriceps tendon (FigureĀ 6A). When comparing the central one-third portion of the tissues, the proportion commonly used for grafts, the patellar ligament was substantially stiffer, with a Modulus of 893.4ā MPa, compared to 461.8ā MPa for the quadriceps tendon.
Simulated Lachman test
The average force for the Lachman maneuver was 59.48ā Nā±ā1.82ā N for the patellar tendon graft with a width of 7.00ā mm. For graft sizes of 5.34ā mm and 3.76ā mm, the force was reduced by 21% and 33%, respectively (FigureĀ 7). When used as a graft, the 7ā mm patellar tendon graft was 10.7% stronger than the native ACL. The force calculated using the central portion of the 7ā mm quadriceps tendon as a graft was 5.1% greater than the one produced by the native ACL.
Figure 7
In the cases where the quadricep graft sizes were reduced to 5.34ā mm and 3.76ā mm, the forces were reduced by 27% and 13%. The ANOVA single factor revealed a significant difference between the three graft sizes (pā<ā0.05). In the graft size of 3.76ā mm, the patellar tendon graft (893.4ā MPa) was 234% stiffer than the quadriceps graft (263.4ā MPa), and the force increment was limited to 5.1% (R2ā=ā0.946). For graft sizes of 5.34ā mm and 7.00ā mm, the force increment was evaluated at 8.1% (R2ā=ā0.928) and 9.8% (R2ā=ā0.909), respectively.
We also performed additional tests to investigate the quadriceps tendon (461.8ā MPa) as an alternative to the patellar tendon. The maximum stress for the quadriceps tendon with a graft size of 3.76ā mm was 43.8ā MPa (FigureĀ 8). Increasing the graft dimensions, the maximal stress was significantly reduced to 28.4ā MPa and 27.5ā MPa for the 5.34ā mm and 7.00ā mm graft sizes (FigureĀ 8). Following the intercondylar eminence, the stress 10.88ā MPa was found for the 7.0ā mm graft, whereas the 3.76ā mm graft resulted in 8.5ā MPa maximal stress. A much more pronounced reduction in stress is seen when comparing the stress risers at the femoral condyle bonding part, where we measured 21.1ā MPa and 8.8ā MPa for the 7.00ā mm and the 3.76ā mm grafts. Similar trends in maximum stress were observed for the native ACL and the patellar tendon grafts. However, when comparing values for the maximum stress of the quadriceps tendon to those experienced by the patellar tendon at similar diameters, the patellar tendon grafts in our model showed an average of two times greater maximum stress across relative points of interest on the grafts.
Figure 8
Discussion
We found that within our model, the patellar tendon outperformed the quadriceps at all graft diameters. When averaging the values for graft diameter and Young's Modulus of the respective graft sources, the patellar tendon required a 10.7% greater force relative to the native ACL to induce tibial displacement compared to 5.1% for the quadriceps. This suggests that the patellar tendon yields more robust grafts for ACL reconstruction at smaller graft diameters. Additionally, the maximum stress for the patellar tendon is at least double that of the quadriceps tendon and about 3ā4 times that of the native ACL. This is visualized in FigureĀ 8 by using our modeling results and extracting the maximum stress from the stress response heat map on the tendon. This shows that the patellar tendon better distributes force across the graft and can tolerate more significant stress before deformation.
On the other hand, the patellar tendon has its inherent flaws. The patellar tendon was 234% stiffer than the quadriceps tendon. Additionally, past research found that the Young Modulus evaluated for the patellar tendon is approximately two times greater than that of the quadriceps tendon (see Table 2) (15). This is problematic given an association of excessive stiffness and high Young's Modulus with over-constraining of the knee. A study by O'Brien et al. found that the knee experiences more significant constriction with increasing Young's Modulus. In contrast, Suggs et al. found a correlation between graft stiffness and knee constraining at various flexion angles (16, 40). Additionally, it has been shown that an over-constraining effect is a common postoperative outcome of ACL reconstruction. A study by Robertson et al. that investigated postoperative knee stiffness in 100 patients undergoing primary ACL reconstruction with hamstring autografts found an incidence of postoperative knee stiffness of 12% at six months follow-up (40, 41). Even more, one study found that athletes who have undergone ACL reconstruction with quadriceps tendons experience less stiffness when compared to those who received patellar tendon autografts and performed better in specific compound movements such as squatting (42).
Table 2
| Author | StƤubli et al. (10) | Flahiff et al. (13) | Current study | Hashemi et al. (14) | StƤubli et al. (10) | OāBrien et al. (16) | Current study |
|---|---|---|---|---|---|---|---|
| Segment | Quadriceps tendon | Patellar ligament | Quadriceps tendon | Patellar ligament | Patellar ligament | Patellar ligament | Patellar ligament |
| Portion | Central portion (CSS 64.6ā mm2) | Central portion (Width 13.2ā mm) | Central portion (CSS 64.3 mm2) | Central portion (CSS 20.7ā mm2) | Central portion (CSS 34.5ā mm2) | Entire Patellar Ligament (vivo) (CSS 114.8ā mm2) | Central portion (CSS 48.5ā mm2) |
| Ultimate tensile stress [MPa] | 38.0 | 78.4 ā800 to 3,000āN | 62.2 | 58.7 | 69.6 | ā | 68.17 |
| Modulus [MPa] | 263.4 (at 200ā
N) 462.8 (at 800ā N) | 340.0 | 461.8 | 507.4 | 565.9 (at 200ā
N) 811.7 (at 800ā N) | 597.4 | 893.4 |
Young moduli reported in the literature compared to the current study.
Westermann et al. conducted a similar study using a simulated Lachman model to test graft diameters between 5 and 9ā mm (43). However, they designed a graft that restored the joint to ideal proportions in all areas of the knee joint, and that reconstructed the ACL with a graft demonstrating over 150% the strength of the native ACL. In addition, their study showed that anatomically ideal grafts of all sizes could tolerate over 75ā N of force before induced tibial displacement. Based on this, we could compare the strength of the grafts within our model relative to the āgold standardā for ACL reconstruction. We found that the 7ā mm diameter grafts were the only grafts in our study that showed strength comparable to these ideal grafts, with forces of over 55ā N and upwards of -60ā N required to induce tibial displacement. Additionally, the grafts below 7ā mm failed to reach a strength of even 50ā N before tibial displacement. Even more, grafts of these diameters underperformed compared to the native ACL within our model. Given our goal of restoring the native strength of the ACL, grafts below 7ā mm diameters do not approximate the strength of an ideal ACL graft and should not be considered for ACL reconstruction.
Regarding the graft size of 3.76ā mm, we were not surprised by the results collected. We anticipated this graft dimension would not be suitable for ACL reconstruction but chose to include it in our model to validate the diameter at which the grafts fail. Although this graft size is unsuitable for ACL reconstruction, we aimed to understand how small the graft's diameter can be before it is incompatible with ACL reconstruction, making this data point extremely valuable.
Furthermore, using smaller graft diameters within the experimental design allows the model to better estimate how the total area of the grafts affect their biomechanical properties, while maintaining core constants such as width and length. This allowed us to model smaller graft sizes in the context of understating if smaller grafts are suitable for patients with smaller body frames, while only changing a single dimensional variable, namely diameter, which was already the variable under investigation within our study. A more favorable approach to investigating total area would be to adjust the width and diameters of the grafts while keep these variables within ranges that are commonly employed in surgical practice, however this was out of the scope of the research question. We were able to still gauge the importance of area on the strength of smaller grafts by using graft diameters smaller than those commonly used within clinical practice while also using these graft diameters to formally investigate the extent of which diameter affects ACL graft selection.
Within our study, a few limitations can be identified. Foremost, the cadavers utilized were of a significant age, between 70 and 90. This may have impacted the mechanical properties of experimental characterization. Additionally, the Young's Modulus values found in our study were more significant than those reported in the literature compared to younger cadavers. This finding runs contrary to what has been established in the literature, as it has been well described that the quadriceps undergoes degeneration as the individual ages and should have yielded a weaker response to stress as compared to the quadriceps of younger cadavers tested by other laboratories (44). The different testing conditions employed, namely concerning the load ranges adopted could have impacted the preparation of the grafts, thus explaining why our tissues behaved differentially than those of other researchers. We compensated for this variance by creating a range of Young's Modulus values that would provide us with an average of values to estimate better the actual effect of Young's Modulus on the grafts. Values from the literature were chosen based on studies that followed similar methods to ours. Although we chose to include four Young's Modulus values for the patellar tendon, it is essential to note that only one literature value was used for the quadriceps, given that fewer data exists for this graft source compared to the patellar tendon in the literature.
Next, we used a finite element model to conduct our experiment. We chose to use a model for several key reasons. First, using a model rules out any errors regarding preparing the grafts at their proper diameters. Additionally, variables such as diameter can be easily manipulated within a model, and various results, such as maximum stress and strain across different elements of the reconstructed knee, can be obtained from a single test. This is not attainable when using cadavers. The limitations faced with cadavers are both a lack of accessibility to adequate quantities of cadavers and many inherent flaws that can affect the data collected. These include variables such as how the cadavers were stored, how old the cadavers are, and how the test is designed and carried out. Finally, using a model of the knee in its working environment allowed us to test how diameter affects the mechanics of the joint, something that has yet to be done in prior studies investigating the effects of diameter on ACL grafts.
Currently, no established experimental methodologies exist to validate such interactions between grafts and condyles. Consequently, the conclusions regarding the experimental testing of the tissues have been appropriately separated from the results obtained from the computer model, which overlooks the ACL-bone interaction (45, 46) entirely. The model used within the study primarily serves to investigate the interaction between graft diameter and joint stability but also functions independently to indicate the interaction between the graft and the condyles. Considering this, the measurements taken from the model focused on quantifying the forces distributed across the reconstructed joint at all the anatomical landmarks of the reconstructed joint, with particular consideration of the epicondylar eminence and femoral condyles. Although the validity of this element is limited, the model does quantify the relationship between graft dimensions and the graft-to-condyle interface. This can be used as a preface to further research to investigate this relationship better.
While finite element modeling has its benefits, it is limited in its application and accuracy when compared to research with cadaveric models. Further testing with cadaveric models is highly recommended to conclude what has been found in this study. It would be beneficial to test the functionality of these grafts within cadavers at various diameters with 1ā mm increments, starting at 7ā mm and going up to 10ā mm, to generate finite data to guide surgeons when considering matching functional graft sizes to their patients.
Conclusion
When designing ACL grafts of a smaller area for use with patients that require smaller than traditional grafts, we found that the patellar tendon provided a stronger graft for knee reconstruction at all diameters. This makes the patellar tendon favorable for use as a graft source in patients requiring smaller grafts. This is further corroborated by the point that the quadriceps tendon undergoes a significantly greater degree of degeneration over time as compared the patellar tendon. For traditional graft sizes, this would not be a factor causing surgeons to favor the patellar tendon over the quadriceps tendon, however when maximally confining graft dimensions in the context of our research question, this element becomes increasingly important to consider.
Assuming that a greater emphasis is placed on the utilization of smaller grafts in target populations, it will be increasingly important to choose a graft source that is stronger and retains a greater functional area as the graft ages. Even more, we conclude based on our model that grafts below 7ā mm diameter are unsuitable for reconstruction of the knee. Finally, the interaction we discovered between the condyles and the graft paves the way for future investigations into how graft dimensions can be tailored to the morphometric characteristics of the knee.
Statements
Data availability statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.
Author contributions
FA is the principal author for the proposed manuscript and was responsible for the proposal and design of the experiment. FA also conducted data analysis and helped in the process of creating the manuscript for the project. GS in coordination with FA helped with the design of the project. Additionally, GS helped in the process of drafting the manuscript. BG worked on data analysis and was most responsible for creating the manuscript for the project. MH and JK helped in the design of the project and helped with data analysis. All authors contributed to the article and approved the submitted version.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisherās note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1.
DuthonVBBareaCAbrassartSFaselJHFritschyDMĆ©nĆ©treyJ. Anatomy of the anterior cruciate ligament. Knee Surg Sports Traumatol Arthrosc. (2006) 14(3):204ā13. 10.1007/s00167-005-0679-9
2.
BrownCHSteinerMECarsonEW. The use of hamstring tendons for anterior cruciate ligament reconstruction. Technique and results. Clin Sports Med. (1993) 12(4):723ā56. 10.1016/S0278-5919(20)30385-9
3.
JanssonKALinkoESandelinJHarilainenA. A prospective randomized study of patellar versus hamstring tendon autografts for anterior cruciate ligament reconstruction. Am J Sports Med. (2003) 31(1):12ā8. 10.1177/03635465030310010501
4.
DiermeierTTishermanRHughesJTulmanMBaum CoffeyEFinkCet alQuadriceps tendon anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc. (2020) 28(8):2644ā56. 10.1007/s00167-020-05902-z
5.
BuerbaRABodenSALesniakB. Graft selection in contemporary anterior cruciate ligament reconstruction. JAAOS Glob Res Rev. (2021) 5(10):e21.00230. 10.5435/JAAOSGlobal-D-21-00230
6.
MaligeABaghdadiSHastMWSchmidtECSheaKGGanleyTJ. Biomechanical properties of common graft choices for anterior cruciate ligament reconstruction: a systematic review. Clin Biomech. (2022) 95:105636. 10.1016/j.clinbiomech.2022.105636
7.
BonielloMRSchwinglerPMBonnerJMRobinsonSPCotterABonnerKF. Impact of hamstring graft diameter on tendon strength: a biomechanical study. Arthrosc J Arthrosc Relat Surg. (2015) 31(6):1084ā90. 10.1016/j.arthro.2014.12.023
8.
Why boneāpatella tendonābone grafts should still be considered the gold standard for anterior cruciate ligament reconstruction | British Journal of Sports Medicine. Available at:https://bjsm.bmj.com/content/43/5/323.long(Cited July 13, 2022).
9.
WoodsTGratzerPF. Effectiveness of three extraction techniques in the development of a decellularized boneāanterior cruciate ligamentābone graft. Biomaterials. (2005) 26(35):7339ā49. 10.1016/j.biomaterials.2005.05.066
10.
StƤubliHUSchatzmannLBrunnerPRincónLNolteLP. Mechanical tensile properties of the quadriceps tendon and patellar ligament in young adults. Am J Sports Med. (1999) 27(1):27ā34. 10.1177/03635465990270011301
11.
van EckCFvan den BekeromMPJFuFHPoolmanRWKerkhoffsGMMJ. Methods to diagnose acute anterior cruciate ligament rupture: a meta-analysis of physical examinations with and without anaesthesia. Knee Surg Sports Traumatol Arthrosc. (2013) 21(8):1895ā903. 10.1007/s00167-012-2250-9
12.
MarieswaranMJainIGargBSharmaVKalyanasundaramD. A review on biomechanics of anterior cruciate ligament and materials for reconstruction. Appl Bionics Biomech. (2018) 2018:e4657824. 10.1155/2018/4657824
13.
FlahiffCMBrooksATHollisJMVander SchildenJLNicholasRW. Biomechanical analysis of patellar tendon allografts as a function of donor age. Am J Sports Med. (1995) 23(3):354ā8. 10.1177/036354659502300319
14.
HashemiJChandrashekarNSlauterbeckJ. The mechanical properties of the human patellar tendon are correlated to its mass density and are independent of sex. Clin Biomech. (2005) 20(6):645ā52. 10.1016/j.clinbiomech.2005.02.008
15.
StaeubliHUAdamOBeckerWBurgkartR. Anterior cruciate ligament and intercondylar notch in the coronal oblique plane: anatomy complemented by magnetic resonance imaging in cruciate ligamentāintact knees. Arthrosc J Arthrosc Relat Surg. (1999) 15(4):349ā59. 10.1016/S0749-8063(99)70051-4
16.
OāBrienTDReevesNDBaltzopoulosVJonesDAMaganarisCN. Mechanical properties of the patellar tendon in adults and children. J Biomech. (2010) 43(6):1190ā5. 10.1016/j.jbiomech.2009.11.028
17.
OfferhausCAlbersMNagaiKArnerJWHƶherJMusahlVet alIndividualized anterior cruciate ligament graft matching: in vivo comparison of cross-sectional areas of hamstring, patellar, and quadriceps tendon grafts and ACL insertion area. Am J Sports Med. (2018) 46(11):2646ā52. 10.1177/0363546518786032
18.
CooperDEDengXHBursteinALWarrenRF. The strength of the central third patellar. Tendon graft: a biomechanical study. Am J Sports Med. (1993) 21(6):818ā24. 10.1177/036354659302100610
19.
MunetaTTakakudaKYamamotoH. Intercondylar notch width and its relation to the configuration and cross-sectional area of the anterior cruciate ligament: a cadaveric knee study. Am J Sports Med. (1997) 25(1):69ā72. 10.1177/036354659702500113
20.
BediAAltchekDW. The āfootprintā anterior cruciate ligament technique: an anatomic approach to anterior cruciate ligament reconstruction. Arthroscy. (2009) 25(10):1128ā38. 10.1016/j.arthro.2009.03.008
21.
KopfSForsytheBWongAKTashmanSIrrgangJJFuFH. Transtibial ACL reconstruction technique fails to position drill tunnels anatomically in vivo 3D CT study. Knee Surg Sports Traumatol Arthrosc. (2012) 20(11):2200ā7. 10.1007/s00167-011-1851-z
22.
GadikotaHRSimJAHosseiniAGillTJLiG. The relationship between femoral tunnels created by the transtibial, anteromedial portal, and outside-in techniques and the anterior cruciate ligament footprint. Am J Sports Med. (2012) 40(4):882ā8. 10.1177/0363546511434276
23.
BassoOJohnsonDPAmisAA. The anatomy of the patellar tendon. Knee Surg Sports Traumatol Arthrosc. (2001) 9(1):2ā5. 10.1007/s001670000133
24.
Espregueira-Mendesda SilvaMV. Anatomy of the lateral collateral ligament: a cadaver and histological study. Knee Surg Sports Traumatol Arthrosc. (2006) 14(3):221ā8. 10.1007/s00167-005-0681-2
25.
FeyenHVan OpstalNBellemansJ. Partial resection of the PCL insertion site during tibial preparation in cruciate-retaining TKA. Knee Surg Sports Traumatol Arthrosc. (2013) 21(12):2674ā9. 10.1007/s00167-012-1997-3
26.
LindMMenhertFPedersenAB. Incidence and outcome after revision anterior cruciate ligament reconstruction: results from the danish registry for knee ligament reconstructions. Am J Sports Med. (2012) 40(7):1551ā7. 10.1177/0363546512446000
27.
SalimRSalzlerMJBerginMAZhengLCareyREKfuriMet alFluoroscopic determination of the tibial insertion of the posterior cruciate ligament in the sagittal plane. Am J Sports Med. (2015) 43(5):1142ā6. 10.1177/0363546514568277
28.
ChahlaJMoatsheGDeanCSLaPradeRF. Posterolateral corner of the knee: current concepts. Arch Bone Jt Surg. (2016) 4(2):97ā103.
29.
WeissJAMakerBNGovindjeeS. Finite element implementation of incompressible, transversely isotropic hyperelasticity. Comput Methods Appl Mech Eng. (1996) 135(1ā2):107ā28. 10.1016/0045-7825(96)01035-3
30.
AdouniMShirazi-AdlAShiraziR. Computational biodynamics of human knee joint in gait: from muscle forces to cartilage stresses. J Biomech. (2012) 45(12):2149ā56. 10.1016/j.jbiomech.2012.05.040
31.
MarouaneHShirazi-AdlAAdouniMHashemiJ. Steeper posterior tibial slope markedly increases ACL force in both active gait and passive knee joint under compression. J Biomech. (2014) 47(6):1353ā9. 10.1016/j.jbiomech.2014.01.055
32.
MesfarWShirazi-AdlA. Biomechanics of the knee joint in flexion under various quadriceps forces. Knee. (2005) 12(6):424ā34. 10.1016/j.knee.2005.03.004
33.
YaoJKuangGMWongDWCNiuWXZhangMFanYB. Influence of screw length and diameter on tibial strain energy density distribution after anterior cruciate ligament reconstruction. Acta Mech Sin. (2014) 30:241ā9. 10.1007/s10409-014-0027-8
34.
ÅuczkiewiczPDaszkiewiczKWitkowskiWChróÅcielewskiJZarzyckiW. Influence of meniscus shape in the cross sectional plane on the knee contact mechanics. J Biomech. (2015) 48(8):1356ā63. 10.1016/j.jbiomech.2015.03.002
35.
SamitierGMarcanoAIAlentorn-GeliECugatRFarmerKWMoserMW. Failure of anterior cruciate ligament reconstruction. Arch Bone Jt Surg. (2015) 3(4):220ā40.
36.
HelgasonBTaddeiFPĆ”lssonHSchileoECristofoliniLVicecontiMet alA modified method for assigning material properties to FE models of bones. Med Eng Phys. (2008) 30(4):444ā53. 10.1016/j.medengphy.2007.05.006
37.
KellerTS. Predicting the compressive mechanical behavior of bone. J Biomech. (1994) 27(9):1159ā68. 10.1016/0021-9290(94)90056-6
38.
SchileoEBalistreriLGrassiLCristofoliniLTaddeiF. To what extent can linear finite element models of human femora predict failure under stance and fall loading configurations?J Biomech. (2014) 47(14):3531ā8. 10.1016/j.jbiomech.2014.08.024
39.
ChristelPSAkgunUYasarTKarahanMDemirelB. The contribution of each anterior cruciate ligament bundle to the lachman test: a cadaver investigation. J Bone Joint Surg Br. (2012) 94(1):68ā74. 10.1302/0301-620X.94B1.26562
40.
SuggsJWangCLiG. The effect of graft stiffness on knee joint biomechanics after ACL reconstructionāāa 3D computational simulation. Clin Biomech. (2003) 18(1):35ā43. 10.1016/S0268-0033(02)00137-7
41.
RobertsonGAJColemanSGSKeatingJF. Knee stiffness following anterior cruciate ligament reconstruction: the incidence and associated factors of knee stiffness following anterior cruciate ligament reconstruction. Knee. (2009) 16(4):245ā7. 10.1016/j.knee.2008.12.014
42.
WangBZhongJLXuXHShangJLinNLuHD. Incidence and risk factors of joint stiffness after anterior cruciate ligament reconstruction. J Orthop Surg. (2020) 15(1):175. 10.1186/s13018-020-01694-7
43.
WestermannRWWolfBRElkinsJM. Effect of acl reconstruction graft size on simulated lachman testing: a finite element analysis. Iowa Orthop J. (2013) 33:70ā7.
44.
KannusPJózsaL. Histopathological changes preceding spontaneous rupture of a tendon. A controlled study of 891 patients. JBJS. (1991) 73(10):1507. 10.2106/00004623-199173100-00009
45.
MoglKEShirazi-AdlA. Biomechanics of passive knee joint in drawer: load transmission in intact and ACL-deficient joints. Knee. (2003) 10(3):265ā76. 10.1016/S0968-0160(02)00135-7
46.
KiaMSchaferKLipmanJCrossMMaymanDPearleAet alA multibody knee model corroborates subject-specific experimental measurements of low ligament forces and kinematic coupling during passive flexion. J Biomech Eng. (2016) 138(5):051010. 10.1115/1.4032850
Summary
Keywords
ACL, autograft, patellar tendon, quadriceps tendon, diameter, finite element modeling, Lachman test
Citation
Amirouche F, Solitro GF, Gligor BZ, Hutchinson M and Koh J (2023) Investigating the effect of autograft diameter for quadriceps and patellar tendons use in anterior cruciate ligament reconstruction: a biomechanical analysis using a simulated Lachman test. Front. Surg. 10:1122379. doi: 10.3389/fsurg.2023.1122379
Received
12 December 2022
Accepted
12 September 2023
Published
11 October 2023
Volume
10 - 2023
Edited by
Paphon Sa-ngasoongsong, Mahidol University, Thailand
Reviewed by
Qichang Mei, Ningbo University, China Soroush Baghdadi, Childrenās Hospital of Philadelphia, United States
Updates
Copyright
Ā© 2023 Amirouche, Solitro, Gligor, Hutchinson and Koh.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Farid Amirouche amirouch@uic.edu
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.