Abstract
Introduction:
Augmented Reality (AR) offers significant potential to enhance surgical precision, improve clinical training, and support intraoperative decision-making. However, developing personalized AR applications for head-mounted displays such as Microsoft HoloLens 2 typically requires advanced technical expertise. Additionally, identifying a robust and accessible method for registering virtual models to the real world remains an open challenge. These limitations hinder the widespread adoption of AR in healthcare, particularly among non-technical users.
Methods:
This work presents EZ-AR, an “easy AR” application for Microsoft HoloLens 2 that automates complex tasks such as configuration and deployment, while ensuring high reliability and smooth interaction. Ultimately, the system enables non-expert users to independently load and visualize personalized 3D models anchored to physical reference markers. The application can load files from local storage or automatically download them from Google Drive in seconds, displaying them in the physical environment using Vuforia or QR code (version 3) tracking. Once loaded, the system operates entirely offline, eliminating any intraoperative dependence on Internet access. The system's performance was assessed in terms of detection range and overall system accuracy under varying surface inclinations and viewing angles.
Results:
A mean projection error of 3.3 ± 2.2 mm was obtained with QR code tracking and 3.9 ± 2.7 mm with Vuforia, both measured at a working distance compatible with typical surgical ergonomics. Further analysis revealed that accuracy is more influenced by viewing angle and surface inclination than by the tracking method itself. Additionally, a usability survey conducted with clinicians confirmed the framework's high acceptability and clinical feasibility.
Discussion:
Overall, EZ-AR provides a robust and accessible framework that simplifies the deployment of head-mounted AR, facilitating its translation and adoption in clinical workflows.
1 Introduction
Augmented Reality (AR) integrates 3D digital content with the physical world (), enhancing medical education (), simulation (), and clinical care (). In surgery, AR improves reproducibility, spatial awareness, and ergonomics compared with traditional methods (). One of the major advantages of AR in surgery is its ability to display relevant 3D virtual information superimposed onto the patient, allowing the clinician to maintain focus on the surgical site while gaining enhanced spatial understanding of the anatomy (; ).
Optical see-through head-mounted displays (OST-HMDs) are often preferred during surgery for their hands-free interaction, enhanced visualization, and ability to maintain an unhindered view of the real surgical environment. This direct view ensures that visual and proprioceptive information remains synchronized (). Despite rapid technological developments in recent years, the Microsoft HoloLens 2 (Redmond, WA, United States) remains the most widely adopted platform in clinical practice. Since its first launch in 2019, Microsoft HoloLens 2 has been featured in over 350 clinical publications1, positioning itself as the leading OST-HMD for surgical applications (). It is particularly employed within the surgical landscape, where guidance based on segmented preoperative models represents its primary application ().
Despite its potential, accurately registering virtual models to the patient’s real anatomy remains a key technical challenge in AR-assisted surgery. Manual alignment (; ; ), SLAM-based anchoring (; ), and external tracking systems (; ) (commonly referred to as outside-in tracking) offer high precision, but are often hindered by drift, hardware complexity, or lack of reproducibility in dynamic clinical environments. Furthermore, existing implementations are frequently limited by closed-source code or poor documentation (; ; ). As an alternative, marker-based, inside-out tracking using visual markers (e.g., ARToolKit, Vuforia, or QR codes) offers a lightweight approach for robust registration. These technologies have been successfully applied in robotic laparoscopic surgeries (), thoracotomy procedures (), complex reconstructive surgeries (), and ultrasound training ().
Despite the accuracy and robustness of these commercial tracking alternatives, developing personalized AR tools often requires advanced programming skills. Furthermore, there is a notable shortage of flexible, user-friendly frameworks to assist developers. As a result, these barriers hinder the development and broader adoption of this transformative technology, limiting its full potential in clinical and research settings. To address these challenges, this work pursues two main objectives:
To develop a lightweight, intuitive, and fully customizable AR application for Microsoft HoloLens 2 that enables non-expert users to independently select, configure, and visualize custom 3D models anchored to a physical reference marker. The application should not require programming skills and should rely on a cloud-based workflow for convenient model storage and retrieval, while also including an alternative local storage pathway to preserve data privacy when required.
To implement and evaluate two alternative marker-based tracking methods, comparing the overall system accuracy achieved with each. Among the commercially available options, we selected Vuforia due to its widespread use in clinical applications; and QR code detection, owing to its native integration into the HoloLens 2 system.
We present EZ-AR, a phonetic play on “easy AR” designed to facilitate AR integration into routine clinical workflows by streamlining personalized surgical planning and intraoperative visualization. This development represents a strategic evolution of our group’s previous research. While our prior studies demonstrated clinical feasibility in specific procedures (such as orthopedic oncology (; ), craniosynostosis (), or microtia reconstruction ()), or focused on hardware characterization (; ), those implementations were bespoke tools requiring significant engineering intervention. This project-specific approach is common in the literature, where independent groups have similarly developed tailored systems, such as the orthopedic application proposed by . EZ-AR builds upon these technical foundations but introduces a critical paradigm shift toward a generalized, open-access framework. By automating deployment through a standardized 3D Slicer-to-HoloLens 2 pipeline, EZ-AR supersedes earlier task-specific systems, enabling non-expert clinicians to autonomously deploy custom AR applications without technical intermediaries.
2 Materials and methods
All components presented in this work, including the AR application, the 3D Slicer module, and the 3D models used for system validation, are freely available in a public GitHub repository at https://github.com/BSEL-UC3M/EZ-AR. Supplementary material referred to in this work is archived on Zenodo at https://zenodo.org/records/20124571.
2.1 System overview
Our proposal comprises two clearly differentiated components. The primary one is the Microsoft HoloLens 2 headset, which runs the EZ-AR app, specifically designed to operate natively on this device. Upon launch, the application presents a user interface that enables the loading and visualization of up to eight personalized, not predefined, virtual 3D models. The second key component is a custom-developed module within the 3D Slicer platform (). This module provides an intuitive interface for selecting and configuring the 3D models to visualize in the headset. The outputs generated by the 3D Slicer module must be uploaded to a Google Drive folder, from which the headset retrieves them to display in the physical environment. Alternatively, these files can also be uploaded locally to the AR glasses if enhanced privacy is desired or if Internet access is unavailable. Figure 1 shows the architectural overview of the proposed system.
FIGURE 1
Thanks to EZ-AR, users without prior AR development experience can display their own 3D models on Microsoft HoloLens 2. A video tutorial showing how to utilize the system step by step is provided in SuppVideo-1_SetupInstructions.
2.2 3D slicer module
The 3D Slicer module (EZ-AR setup) can be downloaded from the GitHub repository: https://github.com/BSEL-UC3M/EZ-AR. The module is organized into four main sections: Initialization, Positioning, Customize models info and saving, and Save Configuration file. Further details regarding each section and user instructions are provided in the SuppDocument_EZ-AR_InstructionsAndDetails.
This module guides users through a sequential workflow: first, the tracking method to use in the AR application (Vuforia or QR) is selected. Subsequently, the operator can import up to eight virtual 3D models from local storage, utilizing any format compatible with 3D Slicer. This limit of eight 3D models was a design decision to maintain a simple interface with a manageable number of buttons. Using a simple slider-based interface, users can then define the models’ pose relative to the AR reference marker. Finally, they assign a name and display color to each model for visualization in the headset.
The output of the 3D Slicer module is a configuration file containing all these settings. This configuration file, along with the 3D models, must be uploaded to a Google Drive folder with access permissions set to “Anyone with the link”. Microsoft HoloLens 2 subsequently accesses this folder to semi-automatically retrieve, import, and render the virtual content. Alternatively, to comply with strict data privacy protocols, users can bypass the cloud entirely and transfer these files directly to the headset’s local storage via the Windows Device Portal.
2.3 Augmented reality application
The AR application was developed using Unity LTS 2022.3.20f1 LTS, the C# programming language, Visual Studio 2022, and the generic, cross-platform libraries: OpenXR and MRTK (version 2.7.3+). The application is ready for download and use in the GitHub repository mentioned above. Detailed information about the AR application’s functionality and user interface is provided in SuppDocument_EZ-AR_InstructionsAndDetails.
2.4 AR reference markers and tracking
This work proposes using physical AR reference markers to spatially align real and virtual environments. Microsoft HoloLens 2 detects the markers’ pose in space and projects the user-selected 3D virtual models anchored to it. Ultimately, a patient-specific surgical guide could be designed to incorporate a physical support for the AR marker, thereby permitting the automatic registration of virtual anatomical structures to their counterparts in the patient (; ). SuppVideo-2_SampleUseCases shows sample use cases of EZ-AR, including marker alignment with surgical guides for automatic registration, guided object positioning for training and reconstructive surgery, projection of cutting planes or drilling trajectories in hard tissue, and 3D anatomical exploration for surgical planning. The 3D models used in the video were extracted from the Supplementary Data of , available at https://zenodo.org/records/10958624; and from patient 1.3.6.1.4.1.9328.50.4.0001 in , available at https://github.com/MIRACLE-Center/CTPelvic1K. These models are derived from fully de-identified datasets, ensuring that no patient-sensitive information or personal identifiers were involved in the study, in accordance with international data privacy standards. Two alternative tracking methods were implemented and made available within the application: Vuforia and QR code detection. To facilitate understanding and visual distinction, two clearly differentiated AR reference markers were employed (Figure 2).
FIGURE 2
In our implementation, users can choose either of the two tracking methods. Regardless of the chosen alternative, the AR app displays a virtual frame surrounding the AR marker to visually verify the alignment between the virtual and real worlds. That frame transitions from green to red when tracking is lost, serving as an immediate, real-time visual reference for the clinician. In dynamic conditions, this feature allows the clinical team to instantly detect any tracking drift or visual desynchronization caused by sudden movements, providing a continuous qualitative mechanism to verify registration accuracy before proceeding with the interaction. All 3D models used in the validation experiments, including the AR reference markers, are available on GitHub at https://github.com/BSEL-UC3M/EZ-AR.
2.4.1 Vuforia
The Vuforia (SDK) (version 10.25.4, Parametric Technology Corporation Inc.) was integrated into Unity to track predefined image targets in real time without manual camera calibration (; ; ). At runtime, it compares features from a custom, high-quality black-and-white square marker (Figure 2A) against the HoloLens 2 front-facing RGB camera feed to compute the target’s pose.
Despite its ease of implementation and device compatibility, Vuforia operates as a closed-source ‘black box,’ preventing users from fine-tuning its internal image processing pipelines. Additionally, its proprietary nature poses long-term compatibility and support risks, which represents a potential constraint for critical clinical applications.
2.4.2 QR tracking
Microsoft HoloLens 2 includes native software that utilizes the device’s stereo cameras to detect QR codes in the scene. While these stereo cameras are excellent for spatial mapping, they operate at lower resolution and have a narrower field of view than the main RGB camera. This tracking alternative is accessible in Unity through an open-source plugin (). This system allows developers to freely access detection parameters and extract 3D world coordinates to anchor virtual models. Unlike Vuforia, this ecosystem-native approach enhances long-term compatibility, eliminates the need for predefined marker content, and dynamically establishes coordinate systems for any valid QR code. This flexibility is highly advantageous for versatile clinical deployments. However, these stereo cameras operate at a lower resolution and narrower field of view than the main RGB camera, making detection highly sensitive to lighting and contrast. A version 3 QR code was employed in this work (Figure 2B).
2.5 System evaluation
2.5.1 Loading time
We measured the time required to load and display virtual 3D models on the HoloLens 2 as the number of models and model file sizes varied. These measurements were carried out under the Primary path (cloud-based workflow) during the initial download from Google Drive and during subsequent loads from the headset’s local storage. Each trial was performed three times.
2.5.2 Detection range
We evaluated both tracking methods by analyzing the detection range as a function of marker size. This is a critical factor for ensuring unobtrusive integration in constrained clinical workspaces. Eight markers per type (Vuforia and QR code), ranging from 3 to 10 cm, were tested to determine the minimum and maximum distances required for reliable tracking. These limits were established by recording the range of recognition and tracking loss over three trials per configuration. This assessment ensures the proposed methods provide a detection range compatible with surgical ergonomics while maintaining minimal marker dimensions.
2.5.3 Overall system accuracy
We analyzed overall system accuracy relative to surface orientation using a custom, 3D-printed PLA platform with inclined planes at 0°, 15°, 30°, 45°, 75°, and 90° (Figure 3).
FIGURE 3
An OptiTrack V120 Duo motion capture system (version 2.2.0, OptiTrack), with sub-millimeter accuracy () was connected to 3D Slicer via PLUS and OpenIGTLink (), and served as the gold standard for all measurements. It simultaneously tracked a pivot-calibrated handheld pointer and the validation platform, registered via point-based registration using the integrated registration cones. One physical copy of each type of AR reference marker (55 mm square) was 3D printed in black and white using a Bambu Lab P1S printer with the AMS unit.
Eight users with varying levels of AR experience (three experts, two intermediates, and three novices) collected the coordinates of fourteen virtual spheres (red Ø 3 mm spheres in) using the pointer over two randomized trials per marker. Participants stood at a fixed distance of 53 cm, regulated by a physical ruler. This operational setup was determined by common ergonomics across all participants and the experimental results obtained in the “Detection range” experiment. Deviations from reference positions were calculated following established protocols (; ; ).
To minimize inherent OST-HMD parallax and perceptual errors, three measures were implemented: First, all participants completed the Microsoft HoloLens 2 eye-tracking calibration protocol. Second, the projected models were small, high-contrast virtual spheres, which facilitated ocular convergence. Third, the physical platform served as a depth cue to reduce vergence-accommodation conflict.
Statistical analysis evaluated differences across AR tracking methods, user proficiency, repetitions, and plane inclinations. In every instance, Shapiro-Wilk tests confirmed a non-normal distribution (p < 0.05), dictating the use of non-parametric methods with a significance level (α) of 0.05. Mann-Whitney U tests compared AR tracking methods and repetitions, while Kruskal–Wallis tests assessed multi-group factors like user variability and plane inclination. All subsequent post hoc analyses were conducted using Conover tests with a consistent Holm-Bonferroni p-value adjustment.
2.5.4 Clinical usability survey
To evaluate the system’s ease of deployment, six volunteer clinicians from orthopedic surgery and traumatology, maxillofacial surgery, and general surgery specialties were asked to independently develop a personalized AR application by following the tutorials and documentation accompanying this manuscript, which are openly available on GitHub and Zenodo (https://github.com/BSEL-UC3M/EZ-AR and https://zenodo.org/records/20124571). Following task completion, their experience was evaluated using a comprehensive questionnaire. This assessment integrated the System Usability Scale (SUS) to measure ease of use and the NASA Task Load Index (NASA-TLX) to evaluate cognitive workload, along with a targeted set of custom questions designed to assess the intuitiveness of the EZ-AR framework. Additionally, each participant’s task completion time was recorded to objectively quantify the time required for clinicians to successfully design their personalized operative application.
3 Results
3.1 Loading time
Table 1 summarizes the average durations for loading and rendering virtual models from both Google Drive and local storage. Cloud-based retrieval was conducted via a stable 1 Gbps Wi-Fi connection. Timings encompass the complete workflow, from the initial “Load Models” command to the final rendering of all assets in their designated positions and colors.
TABLE 1
| Number of models | Total size of all models (KB) | Loading time from google drive, in seconds (mean ± std) | Loading time from local storage, in seconds (mean ± std) |
|---|---|---|---|
| 1 | 59,161 | 24.7 ± 2.0 | 10.5 ± 0.3 |
| 1 | 2,724 | 4.3 ± 0.2 | 0.9 ± 0.1 |
| 1 | 395 | 3.3 ± 0.2 | 0.5 ± 0.1 |
| 3 | 7,633 | 10.9 ± 0.7 | 1.7 ± 0.1 |
| 3 | 24,720 | 14.1 ± 1.0 | 2.9 ± 0.0 |
| 6 | 63,826 | 31.9 ± 1.3 | 8.9 ± 0.2 |
| 8 | 11,835 | 23.0 ± 1.3 | 2.6 ± 0.2 |
Time required to load and display virtual models from either Google Drive or local storage, based on the number of models and the total file size.
3.2 Detection range
Figure 4 illustrates the mean operational detection range for each AR tracking method, with shaded bands indicating the average minimum and maximum distances (min-max range) measured experimentally before tracking stability is lost. In both cases, the distance to the marker was considered insufficient or excessive when tracking was completely lost or became too unstable to be considered reliable.
FIGURE 4
While both methods were functional across the 3–10 cm marker range, QR tracking exhibited a significantly narrower operational window than Vuforia. Vuforia also maintained tracking at AR marker inclinations up to 45°, whereas QR codes required a perpendicular perspective, resulting in immediate tracking loss under angular deviations. Consequently, all measurements were taken with a perpendicular orientation. Under these optimal conditions of range and perpendicularity, QR codes provided greater stability in 3D model projection, while Vuforia exhibited consistent, minor jitter.
3.3 Overall system accuracy
We collected data by looking at the validation platform from a perpendicular perspective with both AR reference markers, repeating the procedure by tilting the platform at 45°. To visualize the spatial error distribution, Figure 5 displays density heatmaps generated via Kernel Density Estimation (KDE) with a Gaussian kernel. The probability density function was scaled by the total number of local samples to represent the estimated point concentration in points per mm2. Subplots are grouped by surface orientation and tracking method per perspective, with axes labeled according to anatomical orientations: L-R (left-right), I-S (inferior-superior), and P-A (posterior-anterior). Vuforia data (tilted view) over the 30°-inclined surface was excluded (marked as ‘Not Measured’), due to occlusion at that steep angle. No data could be collected from a tilted view using QR tracking because the system became unstable when the marker was not viewed perpendicularly.
FIGURE 5
The RMS pivot calibration error and the RMSE of the validation platform registration were 0.72 and 0.57 mm, respectively. All eight participants successfully completed the experiments, with no significant differences observed across repeated trials (Mann-Whitney U test). Meaningful differences in overall accuracy were solely related to surface orientation. A post hoc Conover test revealed two main performance clusters: the first included planes inclined at 90°, 75°, and 45°, while the second included those at 30°, 15°, and 0°. No statistically significant differences were found within each cluster, but clear differences emerged between them.
3.4 Clinical usability survey
Technical validation by clinical users proceeded without incidents. Although four of the six volunteers had prior experience using clinical AR applications, none had ever developed one. Despite this, the mean development time per application after reviewing the tutorials was 22 ± 7 min. This timeframe encompasses the entire workflow: model selection, spatial alignment relative to the AR marker, color and label customization, and configuration file transfer for headset deployment. The outcomes of the questionnaire are presented in Figure 6. To facilitate a straightforward, at-a-glance interpretation, scores for negatively phrased questions were inverted during data processing. Consequently, the highest score consistently represents the most favorable outcome (Strongly agree), while the lowest one indicates the most negative response (Strongly disagree).
FIGURE 6
Overall, usability and cognitive workload assessments yielded highly positive feedback. Clinicians unanimously agreed on the system’s simplicity, low physical and mental demand, and high UI functionality. While they all estimated a minimal learning curve and high perceived success for independent development, results indicate that some users still value technical support for deployment, highlighting a standard logistical preference in clinical environments. Regarding clinical applicability, participants unanimously identified orthopedic surgery, oncological surgery, and neurosurgery as the primary beneficiaries of EZ-AR, followed by plastic surgery (four out of six).
4 Discussion
EZ-AR (https://github.com/BSEL-UC3M/EZ-AR) is a fully customizable AR application for HoloLens 2 that enables clinicians without programming expertise to visualize their own personalized 3D models. Users can configure their models and visualization preferences via a dedicated 3D Slicer module for semi-automatic retrieval by the headset (setup details in SuppVideo-1_SetupInstructions). The framework offers a dual-pathway storage approach: cloud storage (via Google Drive) prioritizes accessibility for non-sensitive data, caching assets locally after the initial download. Alternatively, a local pipeline via the Windows Device Portal bypasses the internet for sensitive patient-specific data, ensuring compliance with GDPR and HIPAA. 3D model retrieval times across all tested scenarios were within seconds (Table 1). As illustrated in the system architecture (Figure 1), data transfer is restricted to a one-time initial setup phase. Since assets are stored locally, clinical execution runs entirely offline. This design isolates the local real-time rendering loop from network fluctuations, preventing frame drops or visual desynchronization. The application supports versatile manipulation (visibility, opacity, translation, rotation, scaling) and marker-based tracking for automatic registration with patient anatomy (; ), facilitating virtual planning and intraoperative guidance.
For marker-based tracking, we integrated Vuforia (with widespread clinical adoption) and QR tracking (with native HoloLens 2 support). Vuforia requires prior marker definition, a license, and proprietary support, but evaluation of marker sizes (3–10 cm) showed that it provides a wider detection range and greater user mobility, maintaining an error of 4.2 ± 3.1 mm at a 45° tilt. Conversely, QR tracking requires no prior configuration and ensures native ecosystem integration, but regularly fails under angular deviations due to the stereo camera’s limited field of view (contrary to the statements in the official documentation ()). In both cases, reliable detection was achieved within a 40–55 cm range between the headset and the marker when using square markers measuring 40–55 mm. This range represents a practical compromise: the marker remains small enough to use in surgical environments without obstructing the procedure, while still allowing a comfortable working distance within arm’s reach of the headset wearer.
Under optimal perpendicular conditions at 53 cm, both methods yielded comparable global errors (QR: 3.3 ± 2.2 mm; Vuforia: 3.9 ± 2.7 mm), which align with the 3.1 mm RMSE reported in the literature (). This performance provides an insightful contrast to the tracking benchmarks by (), which reported lower perpendicular absolute errors at 300 mm (0.8 ± 0.2 mm) but massive degradation at a 30° tilt (6.3 ± 0.1 mm). Crucially, our subset of purely perpendicular alignment errors (1.4–1.9 mm with Vuforia and 1.5–1.7 mm with QR, both at 530 mm) incorporates the additional variance of manual pointing and an end-to-end clinical workflow. Thus, our error integrates the optical tracking system’s intrinsic error, calibration and registration residuals, AR projection performance, and user perceptual alignment. Therefore, while experimental setups preclude direct numerical equivalence, our results fall within the variability range reported in (), confirming that EZ-AR achieves a high level of predictability consistent with the intrinsic limits of HoloLens 2 tracking.
Moreover, in OST-HMDs such as the HoloLens 2, display parallax and eye-display calibration introduce variability. To mitigate this, our protocol included individual eye-tracking calibration and small volumetric targets (3 mm spheres) to provide depth cues. Consequently, part of the reported 3–4 mm error is likely driven by human perceptual limitations rather than software instability. Clinically, this combined metric best represents the actual accuracy a surgeon would achieve during a guided procedure.
In real-world terms, user expertise and tracking methods appeared to be non-dominant factors in the overall error variance, suggesting that the vast majority of the deviation is attributable to systemic factors rather than human skill. Descriptively, tracking errors remained consistent across the three experience levels and tracking conditions (QR perpendicular, Vuforia perpendicular, and Vuforia tilted). While the sample size (n = 8) limits formal statistical inference, this preliminary trend suggests that the intuitive interface and mandatory eye-tracking calibration effectively level the learning curve, allowing novices to achieve a precision clinically comparable to that of expert users. Finally, density maps showed circular error distributions at 90°, whereas 75° and 45° inclinations exhibited horizontal elongation (lower vertical error along the I–S axis), consistent with inclined-plane geometry. Conversely, at 30°, 15°, and 0°, L–R errors decoupled from the P–A axis, confirming that depth perception remains the primary limitation for overall accuracy in OST-HMDs (; ).
The use of a rigid, static phantom was a deliberate choice to establish a standardized benchmarking tool across various surfaces and orientations, ensuring generalizable findings independent of a single clinical scenario. Establishing this technical performance baseline in a controlled environment provides a reliable metric for overall accuracy and serves as a direct reference when transferring the EZ-AR system to real-world clinical tasks.
Beyond technical metrics, our clinical survey underscores the framework’s viability. Crucially, the mean deployment time (22 ± 7 min) represents a major efficiency breakthrough compared to traditional clinical AR workflows requiring hundreds of hours of programming (). Regarding workload, our NASA-TLX outcomes mirror the patterns in (), showing equivalent moderate mental and physical demands (mean scores of 29.2 ± 17.2 and 12.5 ± 19.1, respectively, vs. 24.5 ± 1.14 and 13.4 ± 4.6 in their study), minimal stress and frustration levels (12.5 ± 12.5 vs. 15.7 ± 3.4), and a matching high perceived success (75.0 ± 14.4 vs. 61.6 ± 1.1). Similarly, the SUS evaluation confirms EZ-AR’s high usability against (), outperforming their baseline in intended frequency of use (96.7 ± 7.5 vs. 80 ± 16) and functional integration (83.3 ± 13.7 vs. 74 ± 12), while matching low complexity scores (36.7 ± 13.7 vs. 37 ± 13). Furthermore, clinicians unanimously prioritized orthopedic, oncological, and neurosurgical applications as primary uses, positioning EZ-AR as an efficient tool for routine clinical integration, bridging the gap between engineering benchmarks and practical medical deployment.
Lastly, to address the rapid obsolescence and volatile lifecycle of the HMD market, EZ-AR prioritizes architectural resilience by relying on OpenXR and MRTK standards. This ensures cross-platform portability and keeps the workflow adaptable to next-generation devices. Furthermore, despite shifts in the commercial industry, Web of Science bibliometric data indicate that the HoloLens 2 remains at its peak clinical research utility, with a sustained increase in publications. Therefore, providing a standardized framework for this platform addresses the active needs of a vast community of researchers and clinicians currently utilizing this technology.
4.1 Limitations and future work
Despite EZ-AR’s promising evaluation, several limitations remain. First, the application restricts visualization to a maximum of eight 3D models to balance clinical utility with user interface simplicity. Future iterations could incorporate this limit as a user-defined parameter in the configuration file. Second, the user evaluation relies on a small sample size (8 users). Therefore, the similar performance observed across experience levels and tracking methods serves as a descriptive trend rather than formal proof of absolute equivalence. Nevertheless, the consistent error magnitudes suggest that any undetected variation has a negligible impact on practical clinical performance.
Another systemic hurdle stems from the hardware lifecycle and rapid obsolescence in the AR market. Although EZ-AR mitigates this risk by relying on open standards (OpenXR and MRTK) to ensure software portability, future work will focus on validating the workflow on next-generation HMDs as they become commercially accessible.
Furthermore, tracking accuracy was assessed under ideal geometric and illumination conditions using a rigid, static platform. While this setup established a reliable baseline for the system’s intrinsic precision, future studies must evaluate performance under non-ideal surgical conditions, including dim lighting, partial occlusions, and curved surfaces. Finally, because this registration assessment was static, it does not fully replicate the conditions of an active operating room. Future research will quantitatively assess Target Registration Error (TRE) under motion using a standardized motion-phantom setup—potentially integrated with robotic arms—to characterize tracking robustness and latency boundaries during continuous surgical maneuvers.
5 Conclusion
This work introduced EZ-AR, a lightweight and intuitive framework for deploying personalized 3D content on Microsoft HoloLens 2 without software development. By combining a user-friendly module in 3D Slicer with a configurable AR application, the system enables clinical users to visualize custom virtual models anchored to physical reference markers using either Vuforia or QR code tracking. Our evaluation confirms comparable overall system accuracy (3–4 mm) at clinical distances for both methods. While QR tracking offers versatility and requires no configuration, Vuforia provides greater flexibility for user movement around the AR reference marker. These results validate EZ-AR as a robust, stable tool for surgical planning and intraoperative visualization, adaptable to diverse clinical constraints and interaction requirements.
Statements
Data availability statement
Supplementary material, including further instructions and sample use cases of the presented EZ-AR app, has been archived on Zenodo and can be accessed at https://zenodo.org/records/20124571. Further inquiries can be directed to the corresponding author.
Ethics statement
The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements because Human participation in this study was limited to a technical assessment involving non-patient volunteers. No personal, medical, or identifiable data were collected, analyzed, or stored. As such, no formal ethics approval was required under our institution’s regulations. All volunteers were fully informed of the nature of the study and provided verbal informed consent prior to their participation. All privacy rights were respected throughout the study.
Author contributions
AP-D-d-l-L: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Resources, Software, Validation, Visualization, Writing – original draft. GA-A: Data curation, Formal Analysis, Investigation, Methodology, Software, Visualization, Writing – review and editing. J-AC-H: Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Writing – review and editing. RP-M: Conceptualization, Project administration, Resources, Supervision, Writing – review and editing. JP: Conceptualization, Formal Analysis, Funding acquisition, Project administration, Resources, Supervision, Writing – review and editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. Research supported by Ministerio de Ciencia, Innovación y Universidades (MCIU/AEI/10.13039/501100011033) and the European Union “NextGenerationEU”/PRTR through projects TED2021-129392B-I00, TED2021-132200B-I00 and PID2023-149604OB-I00; by Comunidad de Madrid through project MAGERIT-CM TEC-2024/COM-44; by the Multiannual Agreement with UC3M through project SAFEDELIVERY-CM-UC3M; and by Universidad Carlos III de Madrid through the Grants for Young Doctors, part of the UC3M Research and Transfer Program (“Proyecto JOVENES PPIT2025” - 2025/00831/001).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. AI tools were used exclusively for linguistic refinement, grammatical correction, and proofreading during the drafting of this manuscript. The authors maintain full accountability for the intellectual content, data interpretation, and final text of the study.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/frvir.2026.1818257/full#supplementary-material
Footnotes
1.^This quantification was obtained using the Citation Report tool in Web of Science (https://www.webofscience.com/wos/woscc/basic-search), by performing a keyword search with “HoloLens 2” and “medicine” in May 2026.
References
1
AkulauskasM.ButkusK.RutkūnasV.BlažauskasT.JegelevičiusD. (2023). Implementation of augmented reality in dental surgery using HoloLens 2: an in vitro study and accuracy assessment. Appl. Sci.13 (14), 8315. 10.3390/app13148315
2
AmirasD.HurkxkensT. J.FigueroaD.PrattP. J.PitrolaB.WaturaC.et al (2021). Augmented reality simulator for CT-guided interventions. Eur. Radiol.31, 8897–8902. 10.1007/s00330-021-08043-0
3
AraujoC. H.AguirreJ. A.PuigL. (2024). 3D printing applications and extended realities in medicine: systematic review. 3D Print. Addit. Manuf.13, 142–156. 10.1089/3dp.2024.0128
4
BaumZ. M. C.LassoA.RyanS.UngiT.RaeE.ZevinB.et al (2019). Augmented reality training platform for neurosurgical burr hole localization. J. Med. Robotics Res.04 (03), 1942001. 10.1142/S2424905X19420017
5
BergerM. F.WinterR.TucaA. C.MichelitschB.SchenkenfelderB.HartmannR.et al (2023). Workflow assessment of an augmented reality application for planning of perforator flaps in plastic reconstructive surgery: game or game changer?. Digit. Health9, 20552076231173554. 10.1177/20552076231173554
6
BirloM.EdwardsP. J. E.ClarksonM.StoyanovD. (2022). Utility of optical see-through head mounted displays in augmented reality-assisted surgery: a systematic review. Med. Image Anal.77, 102361. 10.1016/j.media.2022.102361
7
BourhimE. M.AkhiateA. (2022). “Augmented reality SDK’s: a comparative study,” in Intelligent Systems Design and Applications. Editors AbrahamA.GandhiN.HanneT.HongT.-P.Nogueira RiosT.DingW. (Cham: Springer International Publishing), 559–566. 10.1007/978-3-030-96308-8_52
8
CondinoS.TuriniG.ParchiP. D.ViglialoroR. M.PiolantiN.GesiM.et al (2018). How to build a patient-specific hybrid simulator for orthopaedic open surgery: benefits and limits of mixed-reality using the microsoft HoloLens. J. Healthc. Eng.2018 (1), 5435097. 10.1155/2018/5435097
9
CongX.LiT. (2020). Design and development of virtual medical system interface based on VR-AR hybrid technology. Comput. Math. Methods Med.2020, e7108147. 10.1155/2020/7108147
10
CostaJ. N.Gomes-FonsecaJ.ValenteS.FerreiraL.OliveiraB.TorresH. R.et al (2022). “Ultrasound training simulator using augmented reality glasses: an accuracy and precision assessment study,” in 2022 44th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 4461–4464. 10.1109/EMBC48229.2022.9871406
11
CostaG. M.PetryM. R.MartinsJ. G.MoreiraA. P. G. M. (2024). Assessment of multiple fiducial marker trackers on hololens 2. IEEE Access12, 14211–14226. 10.1109/ACCESS.2024.3356722
12
de OliveiraM. E.DebarbaH. G.LädermannA.ChaguéS.CharbonnierC. (2019). A hand-eye calibration method for augmented reality applied to computer-assisted orthopedic surgery. Int. J. Med. Robotics Comput. Assisted Surg.15 (2), e1969. 10.1002/rcs.1969
13
Díez-MontielA.Pose-Díez-de-la-LastraA.González-ÁlvarezA.SalmerónJ. I.PascauJ.OchandianoS. (2024). Tablet-based augmented reality and 3D printed templates in fully guided microtia reconstruction: a clinical workflow. 3D Print. Med.10 (1), 17. 10.1186/s41205-024-00213-2
14
DoughtyM.GhugreN. R. (2022). Head-mounted display-based augmented reality for image-guided media delivery to the heart: a preliminary investigation of perceptual accuracy. J. Imaging8 (2), 33. 10.3390/jimaging8020033
15
García‐SevillaM.Moreta-MartinezR.García-MatoD.Pose-Diez-de-la-LastraA.Pérez-MañanesR.Calvo-HaroJ. A.et al (2021). Augmented reality as a tool to guide psi placement in pelvic tumor resections. Sensors21 (23), 7824. 10.3390/s21237824
16
HajekJ.UnberathM.FotouhiJ.BierB.LeeS. C.OsgoodG.et al (2018). “Closing the calibration loop: an inside-out-tracking paradigm for augmented reality in orthopedic surgery,” in Presented at the Lecture Notes in Computer Science (Including Subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics), 299–306. 10.1007/978-3-030-00937-3_35
17
KuhlemannI.KleemannM.JauerP.SchweikardA.ErnstF. (2017). “Towards X-ray free endovascular interventions - using HoloLens for on-line holographic visualisation,” in Presented at the Healthcare Technology Letters(Hoboken, NJ: Healthcare Technology Letters). The Institution of Engineering and Technology, 184–187. 10.1049/htl.2017.0061
18
KunzC.MaurerP.KeesF.HenrichP.MarziC.HlaváčM.et al (2020). Infrared marker tracking with the HoloLens for neurosurgical interventions. Curr. Dir. Biomed. Eng.6 (1), 20200027. 10.1515/cdbme-2020-0027
19
LiuP.HanH.DuY.ZhuH.LiY.GuF.et al (2021). Deep learning to segment pelvic bones: large-scale CT datasets and baseline models. Int. J. Comput. Assist. Radiol. Surg.16 (5), 749–756. 10.1007/s11548-021-02363-8
20
Lolambean (2025). QR code tracking overview - mixed reality. Available online at: https://learn.microsoft.com/en-us/windows/mixed-reality/develop/advanced-concepts/qr-code-tracking-overview (Accessed April 29, 2025).
21
MeulsteeJ. W.NijsinkJ.SchreursR.VerhammeL. M.XiT.DelyeH. H. K.et al (2019). Toward holographic-guided surgery. Surg. Innov.26 (1), 86–94. 10.1177/1553350618799552
22
Moreta-MartinezR.García-MatoD.García-SevillaM.Pérez-MañanesR.Calvo-HaroJ.PascauJ. (2018). Augmented reality in computer-assisted interventions based on patient-specific 3D printed reference. Healthc. Technol. Lett.5 (5), 162–166. 10.1049/htl.2018.5072
23
Moreta-MartinezR.Pose-Díez-de-la-LastraA.Calvo-HaroJ. A.Mediavilla-SantosL.Pérez-MañanesR.PascauJ. (2021). Combining augmented reality and 3D printing to improve surgical workflows in orthopedic oncology: smartphone application and clinical evaluation. Sensors21 (4), 1370. 10.3390/S21041370
24
OptiTrack (2026). Duo-3-Trio-3-Spec-Sheet-v2.Pdf. Available online at: https://d111srqycjesc9.cloudfront.net/duo-3-trio-3-spec-sheet-v2.pdf (Accessed May 06, 2026).
25
OunA.HagerdornN.ScheidegerC.ChengX. (2024). Mobile devices or head-mounted displays: a comparative review and analysis of augmented reality in healthcare. IEEE Access12, 21825–21839. 10.1109/ACCESS.2024.3361833
26
PalumboA. (2022). Microsoft HoloLens 2 in medical and healthcare context: state of the art and future prospects. Sensors22 (20), 7709. 10.3390/s22207709
27
PieperS.HalleM.KikinisR. (2004). “3D slicer; 3D slicer,” in 2004 2nd IEEE International Symposium on Biomedical Imaging: Nano to Macro (IEEE Cat No. 04EX821). 10.1109/ISBI.2004.1398617
28
Pose-Díez-De-la-LastraA.Moreta-MartinezR.García-SevillaM.García-MatoD.Calvo-HaroJ. A.Mediavilla-SantosL.et al (2022). HoloLens 1 vs. HoloLens 2: improvements in the new model for orthopedic oncological interventions. Sensors22 (13), 4915. 10.3390/s22134915
29
Pose-Díez-de-la-LastraA.García-SevillaM.TappA.TousidonisM.Darriba-AllesJ. V.LinguraruM. G.et al (2025). Microsoft HoloLens 2 vs. tablet-based augmented reality and 3D printing for fronto-orbital reconstruction of craniosynostosis: a case study. 3D Print. Med.11 (1), 13. 10.1186/s41205-025-00251-4
30
PrattP.IvesM.LawtonG.SimmonsJ.RadevN.SpyropoulouL.et al (2018). Through the HoloLensTM looking glass: augmented reality for extremity reconstruction surgery using 3D vascular models with perforating vessels. Eur. Radiol. Exp.2 (1), 0–6. 10.1186/s41747-017-0033-2
31
PTC (2023). Vuforia enterprise augmented reality (AR) software. Available online at: https://www.ptc.com/en/products/vuforia (Accessed May 03, 2023).
32
QianL.DeguetA.KazanzidesP. (2018). ARssist: augmented reality on a head-mounted display for the first assistant in robotic surgery. Healthc. Technol. Lett.5 (5), 194–200. 10.1049/htl.2018.5065
33
RaeE.LassoA.HoldenM. S.MorinE.LevyR.FichtingerG. (2018). “Neurosurgical burr hole placement using the microsoft HoloLens,” in Medical Imaging 2018: Image-Guided Procedures, Robotic Interventions, and Modeling (Bellingham, Washington: Society of Photo-Optical Instrumentation Engineers), 190–197. 10.1117/12.2293680
34
RollandJ. P.FuchsH. (2000). Optical Versus video see-through head-mounted displays in medical visualization. Presence Teleoperators and Virtual Environ.9 (3), 287–309. 10.1162/105474600566808
35
TangR.MaL.XiangC.WangX.LiA.LiaoH.et al (2017). Augmented reality navigation in open surgery for hilar cholangiocarcinoma resection with hemihepatectomy using video-based in situ three-dimensional anatomical modeling: a case report. Medicine96 (37), e8083. 10.1097/MD.0000000000008083
36
TokudaJ.FischerG. S.PapademetrisX.YanivZ.IbanezL.ChengP.et al (2009). OpenIGTLink: an open network protocol for image-guided therapy environment. Int. J. Med. Robotics Comput. Assisted Surg.5 (4), 423–434. 10.1002/RCS.274
37
UngureanuD.BogoF.GallianiS.SamaP.DuanX.MeekhofC.et al (2020). HoloLens 2 research mode as a tool for computer vision research. arXiv:2008.11239. 10.48550/arXiv.2008.11239
38
UppotR. N.LagunaB.McCarthyC. J.De NoviG.PhelpsA.SiegelE.et al (2019). Implementing virtual and augmented reality tools for radiology education and training, communication, and clinical care. Radiology291 (3), 570–580. 10.1148/radiol.2019182210
39
VassalloR.RankinA.ChenE. C. S.PetersT. M. (2017). “Hologram stability evaluation for microsoft HoloLens,” in Medical Imaging 2017: Image Perception, Observer Performance, and Technology Assessment (Bellingham, Washington: Society of Photo-Optical Instrumentation Engineers), 295–300. 10.1117/12.2255831
40
YoonJ.ChenR. E.KimE. J.AkinduroO. O.KerezoudisP.HanP. K.et al (2018). Augmented reality for the surgeon: systematic review. International Journal Medical Robotics + Computer Assisted Surgery MRCAS14 (4), e1914. 10.1002/RCS.1914
Summary
Keywords
augmented reality, computer-assisted interventions, marker-based tracking, microsoft HoloLens 2, surgical planning
Citation
Pose-Díez-de-la-Lastra A, Arce-Alonso G, Calvo-Haro J-A, Pérez-Mañanes R and Pascau J (2026) EZ-AR: an accessible AR application for multiple model visualization and real-time tracking in healthcare settings. Front. Virtual Real. 7:1818257. doi: 10.3389/frvir.2026.1818257
Received
26 February 2026
Revised
09 June 2026
Accepted
17 June 2026
Published
15 July 2026
Volume
7 - 2026
Edited by
Georgios Tsoulfas, Aristotle University of Thessaloniki, Greece
Reviewed by
Muhammad Awais Sattar, Luleå University of Technology, Sweden
Anant Vemuri, Olympus Europa SE & Co. KG, Germany
Nova Eka Diana, YARSI University, Indonesia
Updates
Copyright
© 2026 Pose-Díez-de-la-Lastra, Arce-Alonso, Calvo-Haro, Pérez-Mañanes and Pascau.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Alicia Pose-Díez-de-la-Lastra, apose@ing.uc3m.es
ORCID: Alicia Pose-Díez-de-la-Lastra, orcid.org/0000-0003-1802-2884; Gemma Arce-Alonso, orcid.org/0009-0000-0241-2146; José-Antonio Calvo-Haro, orcid.org/0000-0001-5144-7572; Rubén Pérez-Mañanes, orcid.org/0000-0003-4051-2759; Javier Pascau, orcid.org/0000-0003-1484-731X
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.