Objective: This study constructs a deep learning-based combined algorithm named WaveAttention ResNet (WARN) to investigate the classification accuracy for seven common retinal diseases and the feasibility of AI-assisted diagnosis in this field.
Methods: First, a deep learning-based classification network is constructed. The network is built upon ResNet18, integrated with the Convolutional Block Attention Module (CBAM) and wavelet convolution modules, forming the WARN method for retinal disease classification. Second, the public OCTDL dataset is used to train WARN, which contains classification data for seven retinal disease types: age-related macular degeneration (AMD), diabetic macular edema (DME), epiretinal membrane (ERM), normal (NO), retinal artery occlusion (RAO), retinal vein occlusion (RVO), and vitreomacular interface disease (VID). During this process, ablation experiments and significance tests are conducted on WARN, and comprehensive analyses of various indicators for WARN, ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, and Vision Transformer (ViT) are performed in retinal disease classification tasks. Finally, data provided by Shanxi Eye Hospital are used for testing, and classification results are analyzed.
Results: WARN demonstrates excellent performance on the public OCTDL dataset. Ablation experiments and significance tests confirm the effectiveness of WARN, achieving an accuracy of 90.68%, F1-score of 91.29%, AUC of 97.50%, precision of 93.31%, and recall of 90.68% with relatively short training time. In the dataset from Shanxi Eye Hospital, WARN also performs well, with a recall of 90.85%, precision of 79.94%, and accuracy of 89.18%.
Conclusion: This study fully confirms that the constructed WARN is efficient and feasible for classifying seven common retinal diseases. It further highlights the enormous potential and broad application prospects of AI technology in the field of auxiliary medical diagnosis.
1 Introduction
The accelerating global aging process has made retinal diseases a significant public health issue threatening visual health. Age-related macular degeneration (AMD) has a global prevalence of approximately 8.7%, serving as the primary cause of blindness among individuals over 60 years old in developed countries (1). Diabetic retinopathy affects about 22.27% of diabetic patients, emerging as the main factor for visual impairment in working-age populations (2). Diabetic macular edema (DME), the leading cause of vision loss in diabetic patients, impacts approximately 5.5% of this population (3). Retinal vein occlusion (RVO) has a prevalence of about 0.5% in individuals aged 30 and above, with over 16 million affected people worldwide (4). These diseases are highly disabling, often leading to irreversible vision damage in advanced stages and exacerbating socioeconomic burdens (5).
Optical coherence tomography (OCT) is regarded as the “gold standard” for diagnosing retinal diseases such as AMD (age-related macular degeneration) (6–8), DME (diabetic macular edema) (9–12), and ERM (epiretinal membrane) (13–15). In contrast, fundus photography has certain limitations in diagnosing these conditions. As a non-invasive imaging technique, OCT has become an essential tool for retinal disease diagnosis due to its high resolution, rapid imaging, and good safety. However, the interpretation of OCT images currently relies mainly on manual reading, suffering from issues such as long time consumption and significant subjectivity differences. Additionally, the obvious imbalance between the number of ophthalmologists and the growing demand for examinations further causes diagnostic delays, affecting timely patient treatment. Therefore, developing artificial intelligence (AI)-based auxiliary diagnostic systems is of great significance for improving diagnostic efficiency and consistency and alleviating clinical resource shortages (16).
AI applications in medical image analysis have achieved remarkable progress. In the field of ophthalmology, AI processes complex medical images through deep learning techniques, excavating hidden information to provide comprehensive diagnostic evidence (17, 18). Early studies, such as Xu et al., proposed an improved method for classifying retinal arterioles and venules, providing tools for early disease diagnosis (19). Najeeb et al. used a single-layer CNN to achieve region-of-interest detection in retinal OCT scans (20). With technological advancements, Gao et al. trained a deep CNN with an 88.72% accuracy in grading the severity of diabetic retinopathy (21). Fang et al. proposed a lesion-aware convolutional neural network (LACNN) that simulates the diagnostic thinking of ophthalmologists, enhancing classification efficiency (22).
Despite these advancements, existing studies still have the following limitations:
(1) Most models are based on single-modal data, lacking integration of multi-modal information;
(2) Insufficient classification capabilities for rare diseases or images with multiple lesions;
(3) Limited applicability in clinical environments with constrained computational resources, as model training is time-consuming and updates are restricted.
This study proposes the WaveAttention-ResNet (WARN) model, which enhances feature extraction capabilities by integrating the CBAM attention mechanism and wavelet convolution modules. The goal is to improve the classification accuracy and training performance for seven common retinal diseases, providing a new approach for the widespread clinical application of AI-assisted diagnosis.
2 Methods
2.1 Network
We have successfully constructed a classification method for retinal diseases, namely the wavelet-enhanced ResNet (WaveAttention-ResNet, WARN) based on the Convolutional Block Attention Module (CBAM).
During the process of building this method, we first selected ResNet-18 as the basic framework. The reason for this choice is mainly that ResNet-18 has a concise and efficient structure. This structure can not only effectively avoid the gradient vanishing problem that often occurs in deep networks but also is easy to expand and modify in practical applications, providing great convenience for subsequent optimization work.
In order to further enhance the discriminative ability of the model and enable it to more accurately identify features in the analysis of complex retinal disease images, we skillfully added the CBAM module after the wavelet convolution. The working principle of the CBAM module is to perform attention weighting on the two dimensions of channels and space, enabling the model to focus more on those important features. In this way, it can effectively reduce the interference of background noise on the judgment results and significantly improve the accuracy and reliability of the model analysis.
However, with the increasing complexity of the retinal disease classification task, we found that traditional convolutional operations may not be able to fully capture all the detailed information in the images in some cases. In view of this, we introduced a wavelet convolution module after the CBAM module. The purpose of introducing this module is to make full use of the unique advantages of wavelet transform to further enhance the feature extraction process. Specifically, wavelet transform allows us to conduct a detailed analysis of the images at different scales, so as to better capture the local details and global structures in the images, enabling the model to have a more comprehensive and in-depth understanding of the image information.
As shown in Figure 1, the network we constructed is a deep learning model of wavelet-enhanced ResNet based on CBAM. This model significantly improves the feature extraction ability and the utilization efficiency of spatial-channel information by organically combining the CBAM attention mechanism and the wavelet transform convolution (WTConv2d) module. From the perspective of the overall architecture, this network is composed of multiple ResNet modules, CBAM modules, and WTConv2d modules. Its design concept is to achieve the classification task of retinal disease images efficiently through multi-level feature extraction and enhancement, providing powerful technical support for the accurate diagnosis of retinal diseases.
2.1.1 ResNet-18
ResNet-18, as a classic deep convolutional neural network, was first proposed in 2015 (23). During the training of deep networks, the problems of gradient vanishing and degradation are common and urgent challenges to be solved. ResNet-18 successfully overcame this problem by introducing the Residual Learning mechanism.
The core highlight of ResNet-18 lies in its unique skip connection design. This ingenious design allows the information in the network to be directly transferred from one layer to another without being interfered by non-linear transformations at all. It is precisely because of this design that the network can effectively learn the identity mapping, enabling the successful implementation of deeper networks.
In terms of the overall architecture, ResNet-18 is mainly composed of five parts. Firstly, there is the input layer, which contains a 7 × 7 convolutional layer with a stride of 2. The main function of this convolutional layer is to extract the basic features of the image, and it further reduces the size of the feature map through a max pooling layer.
Following the input layer are four core stages (Stage), each of which is composed of multiple residual blocks, and these residual blocks are exactly the key components of ResNet-18. Inside each residual block, the input data will be processed through two paths respectively. The main path contains two 3 × 3 convolutional layers, and each convolutional layer is followed by a Batch Normalization (BN) operation and a ReLU activation function; the auxiliary path, that is, the skip connection, serves to directly add the input to the output. When the dimensions of the input and output are the same, this operation is very straightforward; however, if the dimensions are different, a 1 × 1 convolution is required to adjust the input dimension to match the output dimension. As the network layers go deeper, the spatial resolution of the feature map gradually decreases, while the number of channels gradually increases. Such changes are conducive to capturing more complex patterns.
After the last convolutional stage is completed, the Global Average Pooling (GAP) layer comes into play, which compresses each feature map into a scalar value. Through this operation, the number of parameters is significantly reduced, and the risk of overfitting is also decreased accordingly. Finally, the Fully Connected Layer maps these processed features into the specific category space, thus completing the entire classification task.
Considering the simplicity of the ResNet-18 structure, the convenience in its actual implementation, its relatively short training time, and its certain feature extraction ability, among other factors, we finally decided to choose ResNet-18 as the basic network architecture.
2.1.2 CBAM
CBAM (Convolutional Block Attention Module) (24), as a unique attention mechanism, was originally designed to significantly enhance the feature representation ability in convolutional neural networks (CNNs).
The inspiration for the attention mechanism actually comes from the unique characteristics of the human visual system. As we know, when the human eye observes an image, it does not process the entire image uniformly and indifferently. Instead, according to specific task requirements, it selectively focuses on certain specific regions or features in the image. In the field of deep learning, ingeniously introducing the attention mechanism enables the model to be more sensitive and focused on the truly important parts of the input data, thereby effectively improving the overall performance of the model.
Looking at the working principle of CBAM in detail (as shown in Figure 2), it mainly consists of two key steps: first, Channel Attention, and second, Spatial Attention. The channel attention mechanism focuses on analyzing and weighting different channels of the feature map. By comprehensively considering the importance of the information contained in each channel, it assigns corresponding weights to different channels, enabling the model to highlight those channels that contain key information and thus enhancing the ability to capture important features. The spatial attention mechanism mainly focuses on the spatial dimension of the feature map. It analyzes each position of the feature map to determine which spatial positions contain information that is more crucial for the current task, and then assigns higher weights to these positions, enabling the model to more accurately focus on the important spatial regions in the image and further improving the understanding and expression ability of the image features. These two steps cooperate with each other and complement each other, jointly acting on the convolutional neural network, enabling the network to more efficiently extract and utilize key feature information when processing image data.
In the Channel Attention stage, as shown in Figure 3, the feature maps of each channel are adjusted by calculating the importance of each channel. First, it calculates the results of the max pooling and the average pooling for each channel respectively. Then, a shared Multi-Layer Perceptron (MLP) network is used to calculate the weight of each channel. These weights reflect the importance of each channel. Finally, a Sigmoid function is used to transform these weights into values between 0 and 1, which are then multiplied by the original feature map to enhance or suppress the information of certain channels. The channel attention mechanism is used to highlight the most important channel information in the feature map and suppress the unimportant channels. Specifically, given the input feature map , we first generate two descriptors and through the operations of Global Average Pooling (GAP) and Global Max Pooling (GMP). Both of these two descriptors are C-dimensional vectors. Then, we input these two descriptors into a Multilayer Perceptron (MLP), which consists of a hidden layer and a linear layer. Assuming that the dimension of the hidden layer is C/r (where r is the dimensionality reduction ratio), the output of the MLP can be expressed as:
where the symbol σ represents the Sigmoid activation function. Finally, the channel attention weights are applied to the input feature map X through element-wise multiplication:
In the Spatial Attention stage, as shown in Figure 3, CBAM further calculates the importance of each spatial position. It performs max pooling and average pooling operations on the feature map output by the channel attention module, and then combines these two results through a convolutional layer to finally obtain a spatial attention map. This attention map is also transformed by a Sigmoid function and then applied to the original feature map to strengthen or weaken the features at different spatial positions. The spatial attention mechanism focuses on highlighting the most important spatial regions in the image. Given the feature map X' processed by the channel attention, we first perform max pooling and average pooling operations on each channel to generate two feature maps and . Then, we concatenate these two feature maps together, and generate the spatial attention weights through a 7 × 7 convolutional layer and a Sigmoid activation function.
The final spatially attention-weighted feature map X'’ can be expressed as:
This combination method enables the model to pay more attention to the important regions in the image, while suppressing the background noise and other irrelevant details. When dealing with OCT images, CBAM enhances the model's ability to capture key features by emphasizing them. Specifically, CBAM first evaluates the importance of each feature channel through the channel attention mechanism and assigns higher weights to the more important channels, thereby highlighting the information that is crucial for diagnosis. Secondly, the spatial attention mechanism focuses on identifying which regions in the image are most critical for diagnosis and strengthens the information in these regions, enabling the algorithm to more accurately locate the lesion positions.
2.1.3 Wavelet convolution module
As shown in Figure 4, the core of the wavelet convolution module (25) lies in using wavelet transform to decompose the input feature map into low-frequency (LL) and high-frequency (LH, HL, HH) sub-bands, which respectively represent the main structural information and edge or detail information of the image. This ability of multi-resolution analysis is difficult to achieve by traditional convolutional layers. For the decomposed sub-bands, we adopt the depth convolution operation with small kernels (3 × 3). This can not only effectively reduce the number of parameters but also maintain or even enhance the accuracy of feature extraction. Finally, these processed sub-bands are recombined into a complete output feature map through the inverse wavelet transform (IWT). This method not only reduces the computational cost but also improves the adaptability of the model to features at different scales.
The Wavelet Transform (WT) is a powerful signal processing tool that can decompose an input signal into different frequency components. For a two-dimensional image, we can use the Haar wavelet transform to decompose it into four parts: the low-frequency component , the horizontal high-frequency component , the vertical high-frequency component , and the diagonal high-frequency component . These components respectively correspond to different features of the input signal. The low-frequency component represents the main structural information of the signal, usually containing the overall shape and low-frequency trend of the image. The horizontal high-frequency component captures the edge or texture information in the horizontal direction, the vertical high-frequency component captures the edge or texture information in the vertical direction, and the diagonal high-frequency component captures the detail or noise information in the diagonal direction. By applying different types of filters (low-pass and high-pass), we can decompose the image into different frequency bands, thus achieving different hierarchical expressions of the image information. The specific formulas are as follows:
By performing a convolution operation on the input feature map X" through the above filters and combining it with a downsampling operation with a stride of 2, we can obtain four subband feature maps:
As shown in the figure above. Next, a small convolutional kernel (3 × 3) is applied to each subband feature map for feature extraction. Assuming that the weight tensor of the convolutional kernel is W, the convolution operation can be expressed as:
Finally, these feature maps are reconstructed into a feature map of the original resolution through the Inverse Wavelet Transform (IWT):
This design not only expands the receptive field of the model but also enables it to capture richer frequency information while maintaining a relatively low number of parameters, which helps to improve the classification accuracy.
This method transforms the input image into the wavelet domain, performs convolution operations on different frequency components, and reconstructs the output image using the inverse wavelet transform. Specifically, first apply the Haar wavelet transform to the input imageX, decomposing it into a low-frequency component and three high-frequency components ,, and . Subsequently, lightweight deep convolution operations with small kernels are applied to each frequency component. Then, the inverse wavelet transform (IWT) is used to fuse all convolved components to generate an intermediate output feature map. To preserve the structural information in the original image, the final output is obtained by adding the feature map after inverse wavelet transform to the convolved original input image. This process can be recursively applied to the low-frequency component to achieve multi-level wavelet decomposition, effectively expanding the receptive field without significantly increasing the number of parameters, while retaining image details and structural information.
2.2 Datasets
The newly released OCTDL (26) dataset was selected for this study, featuring remarkable characteristics. It comprehensively covers seven distinct retinal diseases, providing invaluable resources for the diagnosis and in-depth research of ophthalmic disorders.
Images in the dataset were sourced from multiple institutions, including Ural Federal University, Professor Plus Eye Surgery Clinic, and Ural State Medical University. This multi-institutional collaboration ensures the dataset's diversity and representativeness. All images were acquired using an Optovue Avanti RTVue XR device, which utilizes advanced scanning parameters (e.g., dynamic scanning length and high-resolution imaging) to guarantee the quality and detail richness of the collected data.
The annotation process was rigorous and multi-staged: first, seven medical students performed initial annotations; subsequently, two experienced clinical experts independently reviewed the annotations to minimize errors; finally, the clinic director confirmed the annotations to ensure accuracy and consistency.
Comprising 2,064 images from 821 patients (age range: 20–93 years, male-to-female ratio 3:2, mean age 63 years), the dataset covers seven fundus disease types: age-related macular degeneration (AMD), diabetic macular edema (DME), epiretinal membrane (ERM), normal (NO), retinal artery occlusion (RAO), retinal vein occlusion (RVO), and vitreomacular interface disease (VID). Figure 5 illustrates the dataset distribution.
Additionally, Shanxi Eye Hospital provided a pathological image dataset with OCT data acquired by two devices (Heideiberg; VG 200D, Intalight Ltd., China). The dataset includes AMD, DME, ERM, NO, RAO, RVO, and VID (encompassing macular holes and vitreomacular traction syndrome), matching the lesion types in the OCTDL dataset to facilitate comparative analysis.
Annotation was conducted by three experts with over five years of clinical experience in retina, with discrepancies resolved by a senior expert with 20+ years of experience. The 942 annotated images serve as valuable resources, and their distribution is shown in Figure 5.
2.3 Ethical approval
This study was approved by the Ethics Committee of Shanxi Eye Hospital (Approval No.: SXYYLL-KSSC021) and strictly adheres to the principles outlined in the Declaration of Helsinki. Given the retrospective nature of data collection and the use of de-identified images, informed consent was not required.
2.4 Experimental setup
In this experiment, all models were trained with uniform parameters: 100 epochs, batch size of 128, Adam optimizer, learning rate of 0.0001, weight decay of 0.0005, cosine annealing learning rate scheduler, and cross-entropy loss function. The experimental environment comprised an NVIDIA GeForce RTX 4060 GPU, 12th Gen Intel(R) Core(TM) i7-127000 2.10 GHz CPU, Python 3.12, and PyTorch 2.6.0.
2.5 Data processing
The OCTDL dataset was randomly split into training, validation, and test sets at a patient-level ratio of 60:10:30, ensuring that images from each patient belonged to only one subset. This approach prevented data cross-contamination, safeguarded the independence and fairness of model training, validation, and testing, and enhanced generalization to new patient data. To mitigate overfitting and improve model generalization, data augmentation was applied to the training set, including random cropping, horizontal/vertical flipping, color distortion, rotation, translation, and Gaussian blur. Each augmentation method was controlled by application probabilities with rational parameter ranges to avoid excessive distortion.
Additionally, addressing the class imbalance issue, a class-balanced sampling strategy was adopted with a sampling weight decay rate of 0.9. This enabled the model to prioritize minority classes in the early training phase and gradually align with the overall distribution in later stages, thereby enhancing overall classification performance.
2.6 Comparative experiment
To evaluate and compare the performance of different deep learning models in the classification of Optical Coherence Tomography (OCT) images, a series of experiments were designed for comparison, including the networks of ResNet-18, ResNet-50 (23), Swin Transformer v2 (27), EfficientNet (28), and Vision Transformer (ViT) (29).
ResNet-18 and ResNet-50 are based on the residual network architecture. The skip connections are used to overcome the problem of vanishing gradients in deep neural networks. ResNet-18 has only 18 layers, with a simple design and high computational efficiency. It performs excellently in resource-constrained environments or lightweight applications. Liu Y et al. utilized it to develop an Infant Retinal Intelligent Diagnosis System (IRIDS) in combination with a convolutional neural network and a transformer structure. When identifying infant fundus diseases, it outperformed ophthalmologists in several key indicators (30). In the study by Nabrdalik K et al., its backbone model showed excellent accuracy in the binary classification of cardiac autonomic neuropathy (CAN) in diabetic patients (47). S V A et al. used an extended depthwise separable convolutional ResNet (DDSC-RN) with a Support Vector Machine (SVM) classifier to identify retinal biomarkers, achieving a relatively high overall accuracy (31). In contrast, ResNet-50 has 50 layers and adopts a bottleneck structure to reduce the number of parameters while maintaining high accuracy. Although S V A et al. did not directly compare it in the same kind of complex task study, when dealing with tasks that require processing complex patterns, high-resolution images, and a high level of detail, due to this structure, it has a higher accuracy than simple models. Goh JHL et al. developed a model using ResNet50 when detecting referable diabetic retinopathy (DR), providing an important reference for the overall study (32).
Swin Transformer v2 is specifically designed for computer vision. With its hierarchical feature representation and shifted window mechanism, it can efficiently capture information at different scales and has obvious advantages in processing high-resolution images, achieving remarkable results in multiple visual tasks. Li Z et al. constructed a deep learning model with it, achieving a very high average accuracy in the classification of multiple fundus diseases on two independent public datasets (33). Huang C et al. developed a Swin-MCSFNet classifier based on Swin-Transformer for retinal image quality assessment, which performed well (34).
The models of the EfficientNet series use a compound scaling method to balance the network depth, width, and input resolution, improving the performance and computational efficiency. The HDR-EfficientNet method proposed by Abbas Q et al. uses the EfficientNet-V2 network for end-to-end training to identify eye-related diseases. When evaluated on a large enhanced dataset of retinal fundus images, the average area under the curve (AUC) is outstanding (35).
Vision Transformer (ViT) introduced the Transformer architecture from natural language processing into computer vision. It divides an image into patches and operates on them just like processing text sequences. In the study by Goh JHL et al., among the four ViT models for detecting referable DR, the SWIN transformer showed significantly better AUC performance than Convolutional Neural Network (CNN) models in the tests on multiple datasets (32).
Therefore, the reason for choosing ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, and Vision Transformer (ViT) for comparison is that they each have their own characteristics in terms of network architecture, computational efficiency, feature extraction ability, and the processing of information at different scales. They can comprehensively and multi-dimensionally evaluate the performance of different types of models in the task of OCT image classification.
2.7 Evaluation metrics
To comprehensively assess model performance, multiple key metrics were employed for integrated analysis. First, Balanced Accuracy serves as the core criterion for measuring a model's overall classification capability in imbalanced scenarios. By calculating the average recall across all classes, it eliminates the dominance of majority classes in evaluations, fairly reflecting the model's balanced performance across categories. Second, Precision focuses on the proportion of truly positive instances among all positive predictions, a critical metric for reducing false alarm rates. Concurrently, Recall evaluates the model's ability to identify all positive samples, emphasizing its capacity to minimize missed detections. The F1-score, a harmonic mean of Precision and Recall, is specifically designed for imbalanced datasets, while AUC (Area Under the ROC Curve) measures a binary classifier's discriminative power. The ROC curve plots True Positive Rate (TPR/Recall) against False Positive Rate (FPR), and the confusion matrix provides a detailed breakdown of classification results, listing the counts of True Positives (TP), False Positives (FP), False Negatives (FN), and True Negatives (TN). This matrix enables in-depth analysis of model strengths and weaknesses across categories, guiding further optimization.
Integrating multi-dimensional metrics (accuracy, precision, recall, F1-score, AUC, and confusion matrix) ensures a comprehensive and meticulous performance analysis. This approach guarantees that developed models excel not only in overall performance but also demonstrate efficiency and reliability in handling specific classes. Such a multi-level evaluation framework lays a solid foundation for enhancing model performance.
3 Result
3.1 Ablation experiments
To validate the effectiveness of the proposed method, systematic ablation experiments were conducted to analyze classification performance under various model structure combinations. All models used identical random initialization seeds and data augmentation strategies to ensure result reliability and comparability. Using the original ResNet as the baseline model, each experimental group explored the impact of adding different layers of wavelet modules, CBAM modules, and their mixed configurations on model performance. The experimental results are shown in Table 1.
As shown in Table 1, the standalone addition of wavelet modules yielded insignificant performance improvements, whereas incorporating CBAM modules led to a gradual performance enhancement with increasing module layers. In the mixed-configuration group, the “ResNet + three-layer wavelet + three-layer CBAM” combination achieved optimal results in accuracy, F1-score, and precision, demonstrating its significant enhancement of model classification performance and validating the effectiveness of the proposed method.
3.2 Statistical significance tests
Paired t-tests were conducted between the ResNet18 baseline model and the WARN model to systematically evaluate differences in five key classification performance metrics. Data from three independent experiments are shown in Table 2, with each experiment recording the performance of ResNet18 and WARN in accuracy (Acc), F1-score, AUC, precision, and recall.
The mean differences, T-statistics, and p-values for each metric are presented in Table 3.

Table 3. Statistical analysis of performance differences: mean differences, T-statistics, and p-values.
The WARN model demonstrated statistically significant improvements in accuracy, F1-score, and recall. Specifically, accuracy increased by an average of 1.86% (p = 0.041), F1-score by 1.30% (p = 0.0083), and recall by 1.86% (p = 0.041), indicating that the wavelet feature extraction and CBAM attention mechanism effectively enhanced the model's ability to identify positive samples and improved overall classification performance. However, the AUC metric showed a minor 0.85% decrease (p = 0.0498), a statistically significant change suggesting potential reductions in discriminant stability across different classification thresholds. Although precision increased by 0.35%, the p-value of 0.512 did not reach statistical significance, indicating unclear effectiveness in reducing false positives.
Pooling data from three experiments, WARN consistently outperformed the ResNet18 baseline model, particularly in identification capability and overall accuracy. While the enhancement strategy positively impacted key performance metrics, the decrease in AUC highlights a need to monitor discriminant stability in practical applications, which could be addressed through further optimization to balance metric performance.
3.3 Experimental results
To systematically assess the classification efficiency of different network architectures, comprehensive metric tests were conducted on ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, Vision Transformer (ViT), and the WARN model. Table 4 details the runtime, accuracy (Acc), F1-score, AUC, precision, and recall of each model on the test set, while Figure 6 visually presents the horizontal comparison trends of core metrics.

Table 4. Experimental results of ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, vision transformer, and WARN.

Figure 6. Experimental results of ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, Vision Transformer, and WARN.
The table presents runtime and classification performance metrics for six deep learning models. In terms of runtime, ResNet-18 was the fastest (1.711 h), while Vision Transformer (ViT) was the slowest (11.42 h). Swin Transformer v2 and ViT demonstrated superior performance across multiple metrics: Swin Transformer v2 achieved the highest accuracy (90.71%) and F1-score (92.20%), whereas ViT ranked first in AUC (98.77%). The WARN model showed a shorter runtime (1.881 h) alongside high accuracy (90.68%), F1-score (91.29%), and recall (90.68%), demonstrating an optimal balance between performance and efficiency. ResNet-50 and EfficientNet exhibited relatively lower metrics, with ResNet-50 having the lowest F1-score (88.22%) and EfficientNet showing slightly lower accuracy (88.75%) compared to other models.
Overall, Swin Transformer v2 and ViT excelled in performance, while WARN stood out for its high performance and fast runtime, making it an efficient and accurate choice. Furthermore, a detailed analysis of the confusion matrix in Figure 7 allows for further identification of specific misclassification patterns by examining the frequency of predictions across categories.

Figure 7. Confusion matrices of ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, Vision Transformer, and WARN.
To assess the classification performance of different vision models in retinal diseases, three core metrics—Precision, Recall, and F1-Score—were calculated based on confusion matrices. The experiment involved six models [ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, Vision Transformer (ViT), and WARN] for classifying seven retinal diseases, including age-related macular degeneration (AMD) and diabetic macular edema (DME). The table below presents detailed evaluation results for each model across disease categories, reflecting their identification accuracy, positive sample capture capability, and comprehensive balance performance. Specific findings are shown in Table 5.

Table 5. The precision, recall, and F1-score of ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, vision transformer, and WARN across various disease categories.
After evaluating the classification performance across multiple retinal disease categories, significant differences in model performance were observed. Overall, Swin Transformer v2 and Vision Transformer (ViT) demonstrated superior performance in Precision, Recall, and F1-Score, particularly in classifying AMD, DME, and VID. In contrast, ResNet-50 and EfficientNet showed moderate to weak performance in most tasks, possibly constrained by model complexity or feature extraction capability.
The proposed WARN method exhibited excellent comprehensive performance across key metrics. Notably, it achieved Precision, Recall, and F1-Score of 0.9737 for VID, highlighting strong discriminative ability. For DME and RAO, WARN's Recall approached or reached perfect scores, indicating exceptional positive sample identification—critical for clinical scenarios demanding low missed diagnosis rates. In RVO classification, while WARN's Recall was relatively low (0.7021), its perfect Precision (1.0) ensured highly reliable predictions, making it suitable for tasks intolerant of misdiagnosis. However, WARN showed notably lower Precision (0.7958) for the normal (NO) category, revealing a tendency to misclassify abnormal images as normal. This underscores the need for future optimization, such as enhancing normal sample learning or introducing stronger discriminative mechanisms.
In summary, WARN demonstrates robust and competitive performance across diverse disease classifications, excelling in Precision and select Recall metrics, which highlights its potential as an auxiliary diagnostic tool. However, improvements are needed to balance Recall and Precision and enhance normal category identification.
3.4 Performance validation — based on retinal disease data from Shanxi Eye Hospital
Table 6 presents the model's performance across different retinal disease classifications using the dataset provided by the Shanxi Eye Hospital. Figure 8 illustrates the WARN model's performance analysis based on the Shanxi Eye Hospital dataset, including a confusion matrix, the number of correctly classified samples, and the accuracy rate for each category.

Table 6. Analysis of WARN's performance based on the dataset from Shanxi Eye Hospital: confusion matrix, correct classification counts, and accuracy rates by category.

Figure 8. Analysis of WARN's performance based on the dataset from Shanxi Eye Hospital: confusion matrix, correct classification counts, and accuracy rates by category.
First, in the age-related macular degeneration (AMD) category, there were 296 total samples, all of which were correctly identified as true positives, resulting in a 100% accuracy rate for the WARN model in this category. Next, the diabetic macular edema (DME) category included 77 samples, of which 53 were correctly identified as true positives, yielding an accuracy rate of 67.95% for the WARN model. In the epiretinal membrane (ERM) category, with 229 total samples, 210 were correctly identified as true positives, leading to a WARN model accuracy of 91.70%. For the normal fundus image (NO) category, there were 108 samples, 92 of which were correctly identified, resulting in an 85.19% accuracy rate for the WARN model. The retinal artery occlusion (RAO) category had 11 samples, 6 of which were correctly identified, giving the WARN model an accuracy rate of 54.55%. The retinal vein occlusion (RVO) category was based on 83 samples, 56 of which were correctly identified, resulting in a 67.47% accuracy rate for the WARN model. Finally, in the vitreomacular interface disease (VID) category, 128 out of 138 total samples were correctly identified as true positives, yielding a WARN model accuracy rate of 92.75%. These data comprehensively detail the WARN model's performance in various retinal disease classifications.
3.5 Grad-CAM
To gain a deeper understanding of the model's decision-making process and validate its reliability in practical applications, we employed Grad-CAM (Gradient-weighted Class Activation Mapping) technique. As an effective visualization method, Grad-CAM generates heatmaps by calculating the gradients of the target class with respect to the feature maps output by the convolutional layers, thereby intuitively demonstrating the regions of the image that the model focuses on when making predictions. Specifically in our study, the last layer of the WARN model was selected as the target layer for Grad-CAM. By applying the Grad-CAM technique to the images in the test set, we generated corresponding heatmaps, which are presented in Figure 9. These heatmaps clearly reveal the model's focus on lesions of different types of retinal diseases. We overall framed the retinal areas with lesions on the images. From the images, it can be seen that the model pays high attention to the lesion areas.
4 Discussion
This study integrates the CBAM attention mechanism and wavelet convolution based on ResNet to construct a network structure named WARN. Through ablation experiments, significance tests, and comparative experiments, the effectiveness and performance advantages of the proposed model are systematically verified. Compared with previous studies, this work has the following prominent features: First, it is committed to achieving single-shot synchronous classification of seven retinal diseases, significantly improving diagnostic efficiency; second, it uses the latest public OCT image dataset for training, ensuring the model's generalization ability and cutting-edge nature; in addition, it still achieves excellent classification performance with a short training time, demonstrating good practical value. Finally, the model was tested and analyzed on a private dataset provided by the Shanxi Eye Hospital, further verifying its applicability in actual clinical scenarios.
Compared with previous studies, in terms of the complexity of the classification task, most existing studies only focus on the classification of single diseases or up to three to four types of retinal diseases (21, 36–38), while the WARN model proposed in this paper is committed to achieving single-shot synchronous classification of seven common retinal diseases, greatly improving clinical diagnostic efficiency and practicality. Second, in terms of network structure design, existing methods mostly adopt standard convolutional neural network structures, such as VGG, ResNet, or their lightweight variants (20, 39–41), while this paper integrates the CBAM attention mechanism and wavelet convolution module based on ResNet. The introduction of CBAM enables the model to focus on the lesion area and enhance key feature expression (42), while wavelet convolution enhances the multi-scale texture modeling ability of images, helping to capture subtle pathological changes in OCT images, thereby improving overall performance without significantly increasing the number of parameters. Third, in terms of data usage and generalization ability, this paper uses the latest public OCT image dataset for training. Compared with the use of specific datasets in early work (19), it ensures the model's cutting-edge nature and generalization ability. In addition, in terms of model efficiency and practical value, although many studies have achieved high classification accuracy (43, 44), they often rely on long training times or complex network structures, limiting their practical deployment possibilities; while this paper achieves excellent performance in a short training time and has good engineering implementation potential. Finally, in terms of clinical applicability verification, different from most studies that only evaluate model performance based on public datasets, this paper further tests and analyzes the model on a private dataset provided by the Shanxi Eye Hospital, verifying its stability and reliability in real clinical scenarios. In summary, WARN performs well in classification granularity, network structure, training efficiency, and clinical adaptability, providing an efficient solution for intelligent auxiliary diagnosis of retinal diseases.
4.1 Discussion on ablation experiments
To verify the effectiveness of the proposed method, we conducted systematic ablation experiments to compare the classification performance under various model structure combinations. Using the original ResNet as the baseline model, it achieved an accuracy (ACC) of 0.8889, an F1 score of 0.8984, and an AUC value of 0.9867, demonstrating good overall classification ability. On this basis, wavelet transform modules and CBAM attention mechanisms were introduced respectively for comparative experiments. The results show that the wavelet module improves the model's ability to capture image detail features to a certain extent, with the three-layer wavelet structure performing best (ACC = 0.8933, F1 = 0.8925), but performance tends to decline as the number of layers increases, possibly due to redundant information introduction or overfitting. In contrast, the CBAM module significantly improves model performance, especially showing more stable gain effects when stacking multiple layers. The three-layer CBAM structure achieves the optimal AUC value (0.9875), while the four-layer CBAM performs best in ACC (0.9005) and F1 (0.9002), indicating that CBAM can effectively enhance the model's ability to focus on key features.
To further explore the synergistic effects between different modules, we also tested the hybrid structure of wavelet and CBAM. The results show that the “ResNet + three-layer wavelet + three-layer CBAM” combination exhibits the strongest comprehensive performance: accuracy reaches 0.9068, F1 score is as high as 0.9129, precision is 0.9331, and recall is 0.9068. Although its AUC value (0.9750) is slightly lower than that of the pure CBAM model, it outperforms all other configurations in the remaining core indicators, demonstrating the complementary advantages of combining wavelet transform and CBAM attention mechanism in feature extraction and discriminative ability. The effectiveness of the fusion strategy is verified through systematic experiments, and the joint structure of “three-layer wavelet + three-layer CBAM” is confirmed as a promising improvement scheme, so this fusion strategy is selected in subsequent experiments.
4.2 Discussion on significance experiments
By performing paired t-tests between the ResNet18 baseline model and WARN, we systematically evaluated their differences in five key classification performance indicators. The results show that the enhanced model exhibits statistically significant improvements in accuracy (Acc), F1 score, and recall (Recall). Specifically, the accuracy increased by an average of 1.86% (p = 0.041), the F1 score by 1.30% (p = 0.0083), and the recall by 1.86% (p = 0.041). These results indicate that the introduced wavelet feature extraction and CBAM attention mechanism effectively enhance the model's ability to recognize positive samples, thereby improving overall classification performance.
To comprehensively evaluate the performance of WARN in retinal disease classification, we conducted comparative experiments with several current mainstream deep learning models, including ResNet-18, ResNet-50, Swin Transformer v2, EfficientNet, and Vision Transformer (ViT). Evaluation metrics cover runtime (training duration), accuracy (acc%), F1 score (f1%), AUC value (auc%), precision (precision%), and recall (recall%), with all results validated on the same test set to ensure fairness and comparability. In terms of operational efficiency, WARN only requires 1.881 h to complete the training process, slightly higher than ResNet-18 (1.711 h) but significantly better than other complex models such as Swin Transformer v2 (7.707 h) and Vision Transformer (11.42 h). This advantage makes WARN more suitable for practical deployment and clinical applications, especially in resource-constrained or fast-response scenarios.
In terms of classification performance, WARN demonstrates promising results: accuracy (acc) is 90.68%, second only to Swin Transformer v2 (90.71%) and superior to ResNet-50 (89.09%), ResNet-18 (88.89%), and EfficientNet (88.75%); the F1 score is 91.29%, ranking third among all models, slightly lower than Swin Transformer v2 (92.20%) and Vision Transformer (91.84%), indicating its capability to balance precision and recall; precision is 93.31%, second only to Swin Transformer v2 (94.55%), demonstrating high reliability in predicting positive samples; recall is 90.68%, on par with Swin Transformer v2 and significantly better than ResNet-18 (88.89%) and EfficientNet (88.75%), indicating the model's effectiveness in identifying more true positive cases; the AUC value is 97.50%, slightly lower than Vision Transformer (98.77%), Swin Transformer v2 (98.69%), and ResNet-18 (98.67%), but still maintaining strong classification discriminative ability while ensuring efficient training speed.
Comprehensively, although Swin Transformer v2 has slight advantages in indicators such as accuracy, F1 score, and AUC, its high computational cost limits practicality; Vision Transformer, despite the best AUC performance, requires 11 h of training, making it difficult to meet real-time or low-latency requirements. In contrast, WARN achieves high-level performance in all metrics while maintaining low training overhead, reflecting its excellent balance between accuracy and efficiency.
Furthermore, by introducing the wavelet transform and CBAM attention mechanism, WARN can better capture texture details and lesion boundary information in retinal images, thereby enhancing the ability to identify early minor lesions. As a deep learning model integrating wavelet convolution, residual connections, and attention mechanisms, WARN demonstrates superior comprehensive performance in retinal disease classification tasks. It not only outperforms or approaches existing mainstream models in multiple key evaluation indicators but also has significant advantages in training efficiency.
4.3 Discussion on confusion matrix
Through the detailed analysis of the confusion matrix in Figure 7, we can further understand the number of times each category is predicted into various categories, thereby identifying specific types of misclassification issues. WARN shows significant improvements in specific categories (such as ERM and RAO), reflecting the effectiveness of CBAM and wavelet convolution modules—CBAM focuses on key areas through spatial and channel attention, while wavelet convolution expands the receptive field and enhances the capture of local texture and global structure (especially in categories with rich lesion details like ERM and VID). First, in ERM (epiretinal membrane) detection, recall increased from 87.50% in ResNet-18 to 93.06%, a 5.56 percentage point improvement, and the F1 score also rose from 90.68% to 91.78%. This improvement benefits from the wavelet convolution module's more effective capture of local details (such as membranous structures) and the CBAM attention mechanism's weighted optimization of feature maps, reducing confusion with normal samples (NO). Analysis of the confusion matrix shows a significant reduction in misdiagnosis: ResNet-18 had 9 false negatives (FN), while WARN only had 5 FN. In the RAO (retinal artery occlusion) task, due to the scarcity of such samples (only 11 cases), they are easily misdiagnosed as normal samples. However, WARN enhances the ability to extract vascular occlusion features by introducing wavelet convolution, increasing recall from 72.73% to 81.82%, reaching levels comparable to Swin and ViT, and the F1 score leaped from 84.21% to 90.00%. Meanwhile, the number of false negatives decreased from 3 to 2, further verifying the model's enhanced sensitivity to minority categories. For VID (vitreomacular interface disease), although ResNet-18 already had a high recall (97.37%), WARN maintained the same recall while increasing the F1 score from 95.95% to 97.37%. This primarily attributes to the CBAM attention mechanism's priority processing of global features (such as macular traction), effectively reducing the false negative rate. Additionally, while recall rates for AMD and NO categories slightly decreased (96.44%→94.90% and 98.08%→97.44%, respectively), F1 scores remained at high levels (AMD: 97.06%; NO: 86.86%), indicating that the model pays more attention to minority categories in resource allocation, and overall classification performance remains competitive. DME and RVO showed stable performance, with DME maintaining a 100% recall rate. Overall, compared with ResNet-18, WARN excels in key disease categories such as ERM, RAO, and VID, reflecting CBAM's ability to reasonably allocate attention weights and wavelet convolution's advantages in expanding the receptive field and capturing lesion details. The improvement in F1 score indicates that the model is more robust and reliable in clinical practical applications.
4.4 Discussion on hospital experiment results
Using the constructed WARN to test the dataset provided by Shanxi Eye Hospital, the test results are shown in Table 6 and Figure 8, with an accuracy of 89.18%, precision of 79.94%, and recall of 90.85%. For age-related macular degeneration (AMD), the model achieved 100% accuracy, with both the number of true positives (TP) and total samples being 296, indicating the model's excellent recognition ability in this category with almost no false positives. Analysis of misclassification cases in the test set reveals that some DME images were incorrectly classified as AMD (age-related macular degeneration) or ERM (epiretinal membrane), mainly because the lesion features of DME and AMD are very similar in some cases, especially in macular region changes, making it difficult for the model to distinguish between the two diseases. Additionally, some images contain not only DME lesions but also other types of lesions such as ERM, further increasing the complexity of classification. Since these images have multiple lesion features simultaneously, the model may tend to select more obvious or prominent feature categories for classification, leading to errors. For the epiretinal membrane (ERM) category, the model demonstrated high-level accuracy at 91.70%, with TP = 210 and total samples = 229, indicating significant effectiveness in handling this category. The classification accuracy for normal (NO) images was 85.19%, with TP = 92 and total samples = 108, showing good performance but still with room for improvement. For the RAO (retinal artery occlusion) and RVO (retinal vein occlusion) categories, the model performed relatively poorly, with accuracies of 54.55% and 67.47%, respectively. In-depth analysis concludes that one main reason is insufficient sample quantity, which is clearly inadequate for the model to fully learn the features of these diseases. The small sample size directly limits the model's learning ability, causing poor performance when facing new, unseen data. Additionally, the huge difference in sample quantity between different categories in the dataset exacerbates the class imbalance problem, possibly leading the model to pay more attention to categories with more samples during training while ignoring the less-sampled RAO and RVO. Finally, in the vitreomacular interface (VID) category, the model also performed excellently, with an accuracy of 92.75%, TP = 128, and total samples = 138. Through this series of test results, we fully verified the effectiveness and reliability of the constructed network model.
4.5 Limitations and future directions
The application of AI-assisted diagnostic systems in ophthalmic clinical practice holds immense promise. By leveraging deep learning to automatically analyze OCT images, AI can help clinicians rapidly diagnosis for multiple retinal diseases, thereby improving diagnostic efficiency and reducing the incidence of misdiagnosis and missed diagnosis. Especially for dynamic lesions requiring long-term monitoring, such as different stages of retinal vein occlusion, AI can provide continuous and consistent assessment results to assist doctors in formulating treatment plans. Additionally, AI-assisted diagnosis contributes to scientific basis for personalized treatment regimens, enhancing patient management and therapeutic outcomes.
Despite the achievements of this study, several limitations need to be addressed. First, the dataset size is limited, covering only seven common retinal diseases, which undoubtedly restricts the model's application scope. Second, insufficient data diversity is a notable issue—for example, uneven ethnic distribution in samples and single OCT device models—factors that may affect the model's universality and accuracy. Meanwhile, inadequate dataset samples and the presence of multiple lesion types in single OCT images impose higher requirements on accurate annotation and model training, potentially leading to incomplete model learning or bias. These issues highlight the need for practical improvements in future research to enhance model robustness and applicability.
To further improve the diagnostic performance of AI models, a potential direction is integrating multimodal data, such as optical coherence tomography angiography (OCTA). Notably, OCTA provides detailed information on retinal microcirculation, which is crucial for the diagnosis and differentiation of various retinal diseases. Additionally, constructing multimodal models by incorporating clinical metadata (e.g., disease stage) can further enhance classification accuracy (45). By integrating multimodal data, we can obtain more comprehensive pathological information to improve diagnostic accuracy. Furthermore, developing real-time AI systems and seamlessly integrating them into existing clinical workflows (e.g., PACS systems) represents an important future direction. Such systems can not only provide clinicians with immediate diagnostic suggestions but also optimize medical resource allocation, further promoting the practical application of AI in ophthalmology. In follow-up studies, expanding sample sizes through multicenter collaboration—employing federated learning techniques to allow distributed parameter updates for joint model training without sharing raw data—can reduce class imbalance and enhance the model's ability to learn features of various diseases (46). Through continuous technological innovation and interdisciplinary collaboration, we eagerly anticipate AI playing a more significant role in the diagnosis and treatment of retinal diseases in the future.
5 Conclusion
This study developed and validated a deep-learning algorithm, WaveAttention-ResNet (WARN), to enhance the classification accuracy of seven common retinal diseases in optical coherence tomography (OCT) images. By integrating wavelet convolution to expand the model's receptive field and the CBAM attention mechanism to precisely allocate attention weights across spatial and channel dimensions of OCT images, WARN enables effective capture of both local and global features. Experimental results demonstrate that in the public OCTDL dataset, ablation experiments and significance tests confirmed WARN's effectiveness: it achieved 90.68% accuracy, 91.29% F1 score, 97.50% AUC, 93.31% precision, and 90.68% recall with relatively short training time. Additionally, in an independent test dataset from Shanxi Eye Hospital, the model performed robustly, yielding 89.18% accuracy, 90.85% recall, and 79.94% precision. These findings validate WARN's efficiency and feasibility for retinal disease classification. More importantly, this research highlights the non-negligible value and broad prospects of AI technology in auxiliary medical diagnosis. Future work will further optimize model performance and explore its applications in broader medical image analysis tasks to advance diagnostic efficiency.
Data availability statement
The datasets presented in this article are not readily available because the data supporting the findings of this study are available upon reasonable request. Interested researchers should contactemhhbmdsajIwMDRAMTYzLmNvbQ==to submit requests. Requests to access the datasets should be directed toemhhbmdsajIwMDRAMTYzLmNvbQ==.
Ethics statement
The studies involving humans were approved by Medical Ethics Committee of Shanxi Eye Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants' legal guardians/next of kin because The data used have been approved by the Medical Ethics Committee of Shanxi Eye Hospital. Since it is impossible to obtain written consent for this retrospective case picture study, and the pictures used in the study have been de-identified, covering up the patients' private information. Moreover, privacy protection training has been provided for the personnel of the cooperating parties, and measures such as clearly defining the scope of data use have been taken, which can ensure the security of patients' information.
Author contributions
BG: Conceptualization, Writing – original draft, Data curation, Investigation, Methodology, Software, Validation, Visualization, Writing – review & editing, Formal analysis. DW: Conceptualization, Data curation, Investigation, Methodology, Software, Validation, Visualization, Writing – original draft, Writing – review & editing, Formal analysis. RZ: Data curation, Formal analysis, Investigation, Writing – original draft, Resources. JH: Data curation, Formal analysis, Investigation, Writing – original draft, Resources. WL: Data curation, Formal analysis, Investigation, Writing – original draft, Resources. YW: Formal analysis, Methodology, Supervision, Writing – original draft, Project administration. XY: Formal analysis, Writing – original draft, Project administration. LZ: Conceptualization, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Writing – original draft, Writing – review & editing, Data curation, Supervision.
Funding
The author(s) declare that financial support was received for the research and/or publication of this article. This project was supported by the Doctoral Fund of Shanxi Eye Hospital (B202404); Fund Program for the Scientific Activities of Selected Returned Overseas Professionals in Shanxi Province (20250064); Support for Young Top Talents of “Sanjin Talents” (SJYC2024455); Central Government-Guided Local Science and Technology Development Fund (YDZJSX2025D081).
Acknowledgments
We gratefully acknowledge the financial support from the following institutions and programs: Doctoral Fund of Shanxi Eye Hospital (Grant No. B202404); Fund Program for the Scientific Activities of Selected Returned Overseas Professionals in Shanxi Province (Grant No. 20250064); Support for Young Top Talents under the “Sanjin Talents” Program (Grant No. SJYC2024455); Central Government-Guided Local Science and Technology Development Fund (Grant No. YDZJSX2025D081).
Conflict of interest
BG, DW and XY were employed by Netchina Huaxin Technology Co., Ltd.
The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declare that no Generative AI was used in the creation of this manuscript.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Wong WL, Su X, Li X, Cheung CM, Klein R, Cheng CY, et al. Global prevalence of age-related macular degeneration and disease burden projection for 2020 and 2040: a systematic review and meta-analysis. Lancet Glob Health. (2014) 2(2):e106–16. doi: 10.1016/S2214-109X(13)70145-1
2. Teo ZL, Tham YC, Yu M, Chee ML, Rim TH, Cheung N, et al. Global prevalence of diabetic retinopathy and projection of burden through 2045: systematic review and meta-analysis. Ophthalmology. (2021) 128(11):1580–91. doi: 10.1016/j.ophtha.2021.04.027
3. Im J, Jin YP, Chow R, Yan P. Prevalence of diabetic macular edema based on optical coherence tomography in people with diabetes: a systematic review and meta-analysis. Surv Ophthalmol. (2022) 67(4):1244–51. doi: 10.1016/j.survophthal.2022.01.009
4. Flaxel CJ, Adelman RA, Bailey ST, Fawzi A, Lim JI, Vemulakonda GA, et al. Retinal vein occlusions preferred practice pattern®. Ophthalmology. (2020) 127(2):P288–320. doi: 10.1016/j.ophtha.2019.09.029
5. Xiao W, Chen X, Yan W, Zhu Z, He M. Prevalence and risk factors of epiretinal membranes: a systematic review and meta-analysis of population-based studies. BMJ Open. (2017) 7(9):e014644. doi: 10.1136/bmjopen-2016-014644
6. Damian I, Nicoară SD. SD-OCT biomarkers and the current Status of artificial intelligence in predicting progression from intermediate to advanced AMD. Life (Basel). (2022) 12(3):454. doi: 10.3390/life12030454
7. Burchard CV, Roider J, Kepp T. Analysis of OCT scanning parameters in AMD and RVO. Diagnostics (Basel). (2024) 14(5):516. doi: 10.3390/diagnostics14050516
8. Sotoudeh-Paima S, Jodeiri A, Hajizadeh F, Soltanian-Zadeh H. Multi-scale convolutional neural network for automated AMD classification using retinal OCT images. Comput Biol Med. (2022) 144:105368. doi: 10.1016/j.compbiomed.2022.105368
9. Munk MR, Somfai GM, de Smet MD, Donati G, Menke MN, Garweg JG, et al. The role of intravitreal corticosteroids in the treatment of DME: predictive OCT biomarkers. Int J Mol Sci. (2022) 23(14):7585. doi: 10.3390/ijms23147585
10. Tripathi A, Kumar P, Mayya V, Tulsani A. Generating OCT B-scan DME images using optimized generative adversarial networks (GANs). Heliyon. (2023) 9(8):e18773. doi: 10.1016/j.heliyon.2023.e18773
11. Wang X, Tang F, Chen H, Cheung CY, Heng PA. Deep semi-supervised multiple instance learning with self-correction for DME classification from OCT images. Med Image Anal. (2023) 83:102673. doi: 10.1016/j.media.2022.102673
12. Alsaih K, Lemaitre G, Rastgoo M, Massich J, Sidibé D, Meriaudeau F. Machine learning techniques for diabetic macular edema (DME) classification on SD-OCT images. Biomed Eng Online. (2017) 16(1):68. doi: 10.1186/s12938-017-0352-9
13. Tang Y, Gao X, Wang W, Dan Y, Zhou L, Su S, et al. Automated detection of epiretinal membranes in OCT images using deep learning. Ophthalmic Res. (2023) 66(1):238–46. doi: 10.1159/000525929
14. Matoba R, Morizane Y. Epiretinal membrane: an overview and update. Jpn J Ophthalmol. (2024) 68(6):603–13. doi: 10.1007/s10384-024-01127-6
15. Stevenson W, Prospero Ponce CM, Agarwal DR, Gelman R, Christoforidis JB. Epiretinal membrane: optical coherence tomography-based diagnosis and classification. Clin Ophthalmol. (2016) 10:527–34. doi: 10.2147/OPTH.S97722
16. De Fauw J, Ledsam JR, Romera-Paredes B, Nikolov S, Tomasev N, Blackwell S, et al. Clinically applicable deep learning for diagnosis and referral in retinal disease. Nat Med. (2018) 24(9):1342–50. doi: 10.1038/s41591-018-0107-6
17. Wei W, Anantharanjit R, Patel RP, Cordeiro MF. Detection of macular atrophy in age-related macular degeneration aided by artificial intelligence. Expert Rev Mol Diagn. (2023) 23(6):485–94. doi: 10.1080/14737159.2023.2208751
18. Lim JI, Rachitskaya AV, Hallak JA, Gholami S, Alam MN. Artificial intelligence for retinal diseases. Asia Pac J Ophthalmol (Phila). (2024) 13(4):100096. doi: 10.1016/j.apjo.2024.100096
19. Xu X, Ding W, Abràmoff MD, Cao R. An improved arteriovenous classification method for the early diagnostics of various diseases in retinal image. Comput Methods Programs Biomed. (2017) 141:3–9. doi: 10.1016/j.cmpb.2017.01.007
20. Najeeb S, Sharmile N, Khan MS, Sahin I, Islam MT, Hassan Bhuiyan MI. Classification of retinal diseases from OCT scans using convolutional neural networks. 2018 10th International Conference on Electrical and Computer Engineering (ICECE); Dhaka, Bangladesh (2018). p. 465–8. doi: 10.1109/ICECE.2018.8636699
21. Gao Z, Li J, Guo J, Chen Y, Yi Z, Zhong J. Diagnosis of diabetic retinopathy using deep neural networks. IEEE Access. (2019) 7:3360–70. doi: 10.1109/ACCESS.2018.2888639
22. Fang L, Wang C, Li S, Rabbani H, Chen X, Liu Z. Attention to lesion: lesion - aware convolutional neural network for retinal optical coherence tomography image classification. IEEE Trans Med Imaging. (2019) 38(8):1959–70. doi: 10.1109/TMI.2019.2898414
23. He K, Zhang X, Ren S, Sun J. Deep residual learning for image recognition. Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition (2016). p. 770–8.
24. Woo S, Park J, Lee JY, Kweon IS. Cbam: convolutional block attention module. Proceedings of the European Conference on Computer Vision (ECCV) (2018). p. 3–19
25. Finder SE, Amoyal R, Treister E, Freifeld O. Wavelet convolutions for large receptive fields. European Conference on Computer Vision; Cham: Springer Nature Switzerland (2024). p. 363–80.
26. Kulyabin M, Zhdanov A, Nikiforova A, Stepichev A, Kuznetsova A, Ronkin M, et al. Octdl: optical coherence tomography dataset for image-based deep learning methods. Sci Data. (2024) 11(1):365. doi: 10.1038/s41597-024-03182-7
27. Liu Z, Hu H, Lin Y, Yao Z, Xie Z, Wei Y, et al. Swin transformer v2: scaling up capacity and resolution. Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition (2022). p. 12009–19.
28. Tan M, Le Q. Efficientnet: rethinking model scaling for convolutional neural networks. International Conference on Machine Learning; PMLR (2019). p. 6105–14.
29. Dosovitskiy A, Beyer L, Kolesnikov A, Weissenborn D, Zhai X, Unterthiner T, et al. An image is worth 16×16 words: Transformers for image recognition at scale. arXiv [Preprint]. arXiv:2010.11929 (2020).
30. Liu Y, Xie H, Zhao X, Tang J, Yu Z, Wu Z, et al. Automated detection of nine infantile fundus diseases and conditions in retinal images using a deep learning system. EPMA J. (2024) 15(1):39–51. doi: 10.1007/s13167-024-00350-y
31. S V A, G DB, Raman R. Automatic identification and severity classification of retinal biomarkers in SD - OCT using dilated depthwise separable convolution ResNet with SVM classifier. Curr Eye Res. (2024) 49(5):513–23. doi: 10.1080/02713683.2024.2303713
32. Goh JHL, Ang E, Srinivasan S, Lei X, Loh J, Quek TC, et al. Comparative analysis of vision transformers and conventional convolutional neural networks in detecting referable diabetic retinopathy. Ophthalmol Sci. (2024) 4(6):100552. doi: 10.1016/j.xops.2024.100552
33. Li Z, Han Y, Yang X. Multi-fundus diseases classification using retinal optical coherence tomography images with swin transformer V2. J Imaging. (2023) 9(10):203. doi: 10.3390/jimaging9100203
34. Huang C, Jiang Y, Yang X, Wei C, Chen H, Xiong W, et al. Enhancing retinal fundus image quality assessment with swin - transformer - based learning across multiple color - spaces. Transl Vis Sci Technol. (2024) 13(4):8. doi: 10.1167/tvst.13.4.8
35. Abbas Q, Daadaa Y, Rashid U, Sajid MZ, Ibrahim MEA. HDR-EfficientNet: a classification of hypertensive and diabetic retinopathy using optimize EfficientNet architecture. Diagnostics (Basel). (2023) 13(20):3236. doi: 10.3390/diagnostics13203236
36. Perdomo O, Rios H, Rodríguez FJ, Otálora S, Meriaudeau F, Müller H, et al. Classification of diabetes - related retinal diseases using a deep learning approach in optical coherence tomography. Comput Methods Programs Biomed. (2019) 178:181–9. doi: 10.1016/j.cmpb.2019.06.016
37. Kim J, Tran L. Retinal disease classification from OCT images using deep learning algorithms. 2021 IEEE Conference on Computational Intelligence in Bioinformatics and Computational Biology (CIBCB); Melbourne, Australia (2021). p. 1–6. doi: 10.1109/CIBCB49929.2021.9562919
38. Asif S, Amjad K, Qurrat-Ul-Ain . Deep residual network for diagnosis of retinal diseases using optical coherence tomography images. Interdiscip Sci. (2022) 14(4):906–16. doi: 10.1007/s12539-022-00533-z
39. Alqudah AM. AOCT - NET: a convolutional network automated classification of multiclass retinal diseases using spectral - domain optical coherence tomography images. Med Biol Eng Comput. (2020) 58(1):41–53. doi: 10.1007/s11517-019-02066-y
40. Singh R, Sharma N, Chauhan R, Dangi S, Gupta R. VGG 16 Pre - trained model for early detection of retinal diseases. 2023 3rd International Conference on Smart Generation Computing, Communication and Networking (SMART GENCON); Bangalore, India (2023). p. 1–5. doi: 10.1109/SMARTGENCON60755.2023.10442010
41. Vishal SM, Khurana S. Hybrid convolutional neural network and support vector machine approaches for enhanced retinal disease classification. 2024 3rd International Conference for
42. Xu L, Wang L, Cheng S, Li Y. MHANet: a hybrid attention mechanism for retinal diseases classification. PLoS One. (2021) 16(12):e0261285. doi: 10.1371/journal.pone.0261285
43. He J, Wang J, Han Z, Ma J, Wang C, Qi M. An interpretable transformer network for the retinal disease classification using optical coherence tomography. Sci Rep. (2023) 13(1):3637. doi: 10.1038/s41598-023-30853-z
44. Zhao J, Zhu J, He J, Cao G, Dai C. Multi - label classification of retinal diseases based on fundus images using ResNet and transformer. Med Biol Eng Comput. (2024) 62(11):3459–69. doi: 10.1007/s11517-024-03144-6
45. Esteva A, Chou K, Yeung S, Naik N, Madani A, Mottaghi A, et al. Deep learning-enabled medical computer vision. NPJ Digit Med. (2021) 4(1):5. doi: 10.1038/s41746-020-00376-2
46. Singh P, Singh M, Singh R, Singh N. Federated Learning: Challenges, Methods, and Future Directions. Federated Learning for IoT Applications (2022).
47. Nabrdalik K, Irlik K, Meng Y, Kwiendacz H, Piaśnik J, Hendel M, et al. Artificial intelligence based classification of cardiac autonomic neuropathy from retinal fundus images in patients with diabetes: the Silesia diabetes heart study. Cardiovasc Diabetol. (2024) 23(1):296. doi: 10.1186/s12933-024-02367-z
Keywords: deep learning, retinal diseases, ResNet18, CBAM, wavelet convolution
Citation: Guo B, Wang D, Zhang R, Hou J, Liu W, Wu Y, Yang X and Zhang L (2025) WaveAttention-ResNet: a deep learning-based intelligent diagnostic model for the auxiliary diagnosis of multiple retinal diseases. Front. Radiol. 5:1608052. doi: 10.3389/fradi.2025.1608052
Received: 8 April 2025; Accepted: 14 July 2025;
Published: 29 July 2025.
Edited by:
Yanwu Xu, Baidu, ChinaReviewed by:
Huihui Fang, Nanyang Technological University, SingaporeKarthik Karunakaran, Prediscan Medtech, India
Phayung Meesad, King Mongkut’s University of Technology North Bangkok, Thailand
Copyright: © 2025 Guo, Wang, Zhang, Hou, Liu, Wu, Yang and Zhang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Lijuan Zhang, emhhbmdsajIwMDRAMTYzLmNvbQ==
†These authors have contributed equally to this work