An innovative automatic feeding device for preterm infants: promoting the development of sucking ability
Provisionally accepted- 1The Second Affiliated Hospital of Shantou University Medical College, Shantou, China
- 2Shenzhen Children's Hospital, Shenzhen, China
- 3Shenzhen Kaiyang Industrial Co., Ltd, Shenzhen, China
Select one of your emails
You have multiple emails registered with Frontiers:
Notify me on publication
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Preterm infants, particularly those born before 32 weeks of gestation, frequently face challenges in achieving full enteral nutrition due to underdeveloped sucking-swallowing-breathing coordination. Conventional feeding methods, such as the use of indwelling nasogastric tubes, overlook the importance of sucking activity, which is essential for the development of gastrointestinal motility and the secretion of digestive enzymes. To address this issue, we have developed a sucking-rewarded automatic feeding device specifically designed for preterm infants. The device features a specialized pacifier that detects sucking activity and triggers the delivery of a predetermined amount of milk into the stomach via a gastric tube. In addition to promoting sucking-induced satiety, the device continuously monitors sucking waveforms to assess infants' viability and sucking maturity. In a clinical pilot study involving 25 preterm infants, those fed with the device demonstrated a significant increase in intestinal oxygen saturation compared with conventional gavage feeding (p < 0.05). Complementary experiments in 12 newborn beagle puppies showed faster gastric emptying rates (p < 0.01) and elevated gastrointestinal hormone levels (p < 0.05) when using the device. These findings highlight the clinical potential of the proposed device in improving feeding safety, efficiency, and developmental outcomes in preterm infants, and warrant further large-scale clinical trials to validate its long-term efficacy and integration into neonatal care.
Keywords: automatic feeding device, preterm infants, Enteral Nutrition, Gastric Emptying, animal experiments
Received: 26 Aug 2025; Accepted: 24 Nov 2025.
Copyright: © 2025 Luo, Zhao, Lin, Huang, Liu, Xu and Li. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
* Correspondence: Hongping Li
Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.