Minimizing contrast agent dosage in CT angiography using a saline chaser with a low trigger threshold


Linghan LiLinghan Li1Xiaojian Zhu
Xiaojian Zhu2*Xunhua Wu

Xunhua Wu3*
  • 1Department of Radiology, Changsha Hospital of Traditional Chinese Medicine (Changsha Eighth Hospital), Changsha, China
  • 2Department of Radiology, Chaling People’s Hospital, Zhuzhou, Hunan, China
  • 3Department of Radiology, Xiangya Hospital, Central South University, Changsha, Hunan, China

Background: Computed tomography angiography (CTA), utilizing iodinated contrast agents, is a first-line diagnostic tool for cardiovascular diseases (CVD). However, the administration of contrast agents introduces potential risks to patients.

Objective: This study aims to design a novel protocol for CTA that integrates a saline chaser strategy and low-threshold triggering and evaluate, and demonstrate its feasibility and practicality minimizing the required contrast agent for imaging.

Methods: A novel protocol for CTA incorporating a saline chaser strategy and low-threshold triggering was developed. Random assignment of eighty patients undergoing CTA examinations divided them into conventional and novel protocol groups. The assessment encompassed iodine intake, radiation dose, image quality, and superior vena cava artefacts in both cohorts.

Results: The novel protocol group exhibited a noteworthy 20% reduction in iodine intake compared to the conventional group (P < 0.05, FDR correction). Notably, Hounsfield units (HU) of the ascending and descending aorta at the T12 level tended to be lower in the novel protocol group (P < 0.05, uncorrected), while parameters like signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) exhibited no significant between-group differences. Compared with conventional group, fewer superior vena cava artefacts were observed in the novel protocol group, and subjective image quality assessment by physicians remained consistent between the two groups (kappa = 0.84, P < 0.01).

Conclusion: The combination of saline chaser strategy with low-threshold triggering in CTA imaging proves a viable approach, significantly curtailing the utilization of iodinated contrast agents, and superior vena cava artefacts.

1 Introduction

Cardiovascular disease (CVD), chiefly characterized by ischemic heart disease and arterial conditions such as stroke, is a significant contributor for global mortality and disability rates (1, 2). Computed tomography angiography (CTA) is a frontline diagnostic tool for assessing arterial health, encompassing both the aorta and coronary arteries (3). Enhancing vascular imaging through CTA necessitates using iodinated contrast agents (4). However, employing these agents poses potential health risks, including Contrast-Associated Acute Kidney Injury (CA-AKI), recognized as an inducer of Acute Kidney Injury (AKI) (5, 6). Notably, CA-AKI accounts for 11% of hospital-related AKI cases, ranking third among causes of iatrogenic kidney injury (7, 8). Consequently, reducing contrast agent dosage and augmenting CTA safety via multifaceted strategies are central tenets of clinical research.

CTA employs various techniques to minimize patient iodine exposure while upholding image quality (9, 10). One such technique involves the saline chaser method, which facilitates the uniform distribution of contrast agents within blood vessels, optimizing vascular opacification (1114). Studies underscore the use of the physiologic saline solution as an optimal supplementary injection, delivered at a rate of 4–5 mL per second, yielding maximal attenuation in the aorta or coronary arteries (15). Another strategic approach involves adjusting CTA trigger thresholds. Multiple studies propose trigger thresholds for aortic CTA protocols within the 150–350 Hounsfield Units (HU) range (16). Innovative research explores real-time monitoring techniques, dynamically adjusting trigger thresholds based on variations in contrast agent concentration within regions of interest (17). This precision tuning mechanism ensures accurate threshold determination, effectively reducing contrast agent dosages while preserving image quality. However, existing clinical research has yet to integrate multiple strategies, and current equipment struggles to integrate the methods mentioned earlier effectively.

Hence, we hypothesize that the synergistic application of a saline chaser strategy and low-threshold triggering can significantly diminish patient iodine exposure without compromising image fidelity. Based on this hypothesis, we developed a novel device termed the “Low-Threshold Trigger with Saline Chaser” and test its function and feasibility in an initial clinical trial. This study aims to comprehensively assess the device's function from a series of indicators including clinical feasibility, iodine intake, radiation dosage, superior vena cava artefact, and vascular image quality.

2 Materials and methods

2.1 Protocol design

The schematic diagram of the novel protocol scanning design and its physical representation is shown in Figure 1. The apparatus comprises four components (Figure 1A). The standard iodinated contrast agent injection unit is the regular contrast agent injection device (Figure 1B). The saline chaser device (Figure 1C) is equipped with a three-way valve, connecting one outlet to the patient, while an additional saline chaser injector, exclusive of the contrast agent injection mode, administers saline solely under the control of the control unit (Figure 1D). Concurrently, the control unit regulates the trigger threshold of the CT machine (Figure 1E). The novel protocol aims to achieve CTA with diminished contrast agent application, upholding image quality. This goal is attained through a straightforward and practical design, incorporating the saline chaser and a low trigger threshold of CTA.

Figure 1
Medical research diagram depicting a study on aortic CT imaging. Panel A shows equipment with contrast agent bottles. Panel B illustrates a flowchart for the procedure involving contrast agents, saline solutions, and a CT machine. Panel C displays a saline chaser device. Panel D features a screen with CT settings. Panel E shows a CT scanner setup. Panel F outlines the study methodology, including eligibility criteria, inclusion and exclusion details, and grouping for assessments like contrast dose, radiation dose, and image evaluations.

Figure 1. Schematic and photographic representation of the device. (A) Overall schematic representation of the novel protocol scanning, comprising four main components. (B) Injector for administering regular contrast agent. (C) The saline chaser is regulated by a control system and administered via a three-way valve. (D) Low threshold and saline chaser control device. (E) CT scanning device. (F) The flow chart illustrates the sequence of the study process. Following the inclusion and exclusion process, 80 patients were randomly assigned to two groups in a 1:1 ratio: the conventional group and the novel protocol group. Subsequently, these groups underwent assessments for contrast agent dosage, radiation dosage, and subjective and objective image evaluations.

2.2 Experimental design

This study received approval from the Ethics Committee of Xiangya Hospital (Approval No. 202012237), and informed consent was obtained from all enrolled patients from March 2021 to May 2022. The inclusion criteria encompassed the following: (1) age over 18 years; (2) clinical prescription for CTA; (3) absence of contraindications to CTA; and (4) patient's informed consent. Exclusion criteria encompassed (1) patients unable to breathe spontaneously, (2) pregnant women, and (3) patients who declined participation or were susceptible to examination failure or harm. The study flowchart is presented in Figure 1F.

After enrollment, patients were assigned to either the conventional group or the novel protocol group based on a computer-generated random number list (specifically, using the RAND function in Excel) in a 1:1 ratio.

2.3 Scan protocol and image post-processing

The CTA scans were conducted utilizing a 256-slice Revolution CT scanner manufactured by GE Healthcare. The scan encompassed a range from 5 cm above the apex of the lungs down to the level of the pubic symphysis. The layer thickness and interval parameters were configured at 0.625 mm, and the collimator was finely tuned to 128 mm × 0.625 mm. For contrast enhancement, both study groups were administered injections of a non-ionic contrast agent (350 mg/mL iohexol) via a high-pressure syringe (German Ulrich Missouri XD2001). The syringe was equipped with a 20 G indwelling needle and was inserted into the antecubital vein.

The conventional group scan introduced the contrast agent through the cubital vein. The contrast agent dosage was calculated based on the individual's body weight, with a factor of 1.5 mL per kilogram, and was administered at a flow rate of 4.0 mL/s. The central horizontal aorta at the T8 level was designated as the region of interest (ROI), employing a threshold of 180 CT number, Hounsfield Unit (HU). The image acquisition occurred approximately 5.9 s after attaining the contrast agent's peak concentration.

In the novel protocol group scan, the initial high-pressure injector administered the contrast agent into the cubital vein at a velocity of 4.0 mL/s. The ROI for the central aorta at the T8 level was configured to 60 HU. Upon reaching the peak value, the first injector was halted, and the second high-pressure injector was engaged to infuse 30 mL of normal saline. When the T8 level is observed to trigger the scan, the injection of the contrast agent is stopped, and the saline chaser is started at the same time. The initiation of scanning was delayed by 8 s.

After acquiring thin-slice data for aortic image reconstruction, the data were transmitted to the AW4.7 workstation. The aortic analysis was performed following processing using the software suite integrated with the GE CT scanner. The 3D post-processing encompassed multiplanar reconstruction (MPR), curved planar reconstruction (CPR), maximal intensity projection (MIP), and volume rendering (VR) techniques.

The detailed scanning precedures for different clinical indications could be found in the Supplementary Material.

2.4 Contrast agent dosage and radiation dose assessment

The quantification of contrast agents was articulated using iodine intake as the metric, with iodine intake (g) calculated as the product of iodine contrast agent volume (mL) and contrast agent concentration divided by 1,000. Radiation dosage parameters encompassed the volume CT dose index (CTDIvol, mGy), CT dose-length product (DLP, mGy·cm), and effective radiation dose (ED, mSv), the latter expressed as the product of DLP and a conversion factor denoted as “k”, where k represents the radiation weight conversion factor. The “k” value is 0.017 mSv/(mGy·cm).

2.5 Objective evaluation of image quality

The quantitative image quality assessment encompasses specific HUs, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). HU was determined for various anatomical points: the ascending aorta, descending aorta at the T12 level, residual imaging of the superior vena cava, and the left external iliac artery. A region of interest (ROI) measuring 5 mm² was applied. Triplicate measurements were conducted at each designated point, with the standard deviation (SD) of HU taken from the right erector spinae muscle at the same level serving as the background noise reference. ROIs were meticulously positioned to exclude regions with uneven density. The mean value was computed for each corresponding measurement. The aortic SNR was computed using the subsequent formula:

SNR=AorticNoise

And CNR (contrast signal-to-noise ratio):

CNR=(Aortic-Erectorspinae)Noise

2.6 Subjective evaluation of image quality

Two experienced radiologists (LL and LK) employed a double-blind approach to assess aortic and superior vena cava images. They used a subjective 4-point scale to evaluate image quality based on the following criteria:

- 4 points: Excellent image quality, featuring clear 3D reconstruction images, smooth and sharply defined blood vessel edges, and no cross-layer artefacts.

- 3 points: Relatively clear image quality, displaying well-defined blood vessel edges, minimal misalignment artefacts, enabling accurate diagnosis.

- 2 points: Average image quality, characterized by 3D reconstruction artefacts, requiring diagnosis through tomographic combination.

- 1 point: Poor image quality, with substantial artefacts hindering diagnosis.

Scores of 3 and 4 points on the subjective image quality assessment indicated satisfactory diagnostic criteria.


For the assessment of superior vena cava image quality, a 3-point scale was used, focusing primarily on artefact presence within the vein:

- 3 points: No artefacts within the veins.

- 2 points: Few artefacts within the vein, not obstructing observation of peripheral arteries and blood vessels.

- 1 point: Numerous artefacts significantly affecting peripheral artery observation.

2.7 Data analysis

Statistical analysis was conducted using IBM SPSS Statistics for Windows, Version 22.0 (IBM Corp., Armonk, NY, USA). The measurement data were expressed as (X¯±s), and the t-test was employed for inter-group comparison analysis. The count data were presented as frequencies, and the χ² test was used to assess the between-group differences. The agreement between subjective scores was evaluated using Cohen's kappa test, and the kappa values were interpreted as follows: poor agreement (0.00–0.20), fair agreement (0.21–0.40), moderate agreement (0.41–0.60), substantial agreement (0.61–0.80), and almost perfect agreement (0.80–1.00). Statistical significance was set at P < 0.05. To address the issue of multiple comparisons, the False Discovery Rate (FDR) was controlled using the Benjamini-Hochberg (BH) procedure.

3 Results

3.1 Patient groups and comparability

The two patient groups demonstrated strong comparability. All 80 included patients successfully underwent CTA examinations and exhibited favourable procedural tolerance. The conventional group consisted of 33 males and 7 females, with an average age of 57.4 ± 13.2 years and an average body mass index (BMI) of 23.1 ± 3.0 kg/m2. Meanwhile, the novel protocol group comprised 32 males and 8 females, with an average age of 57.3 ± 12.6 years and a BMI of 23.3 ± 4.3 kg/m2. Notably, there were no between-group significant differences in sex, age, age group, height, weight, or BMI, indicating robust comparability (Table 1).

Table 1
www.frontiersin.org

Table 1. Clinical characteristics.

3.2 Clinical indications and key test points

Furthermore, we obtained clinical indications for examination and identified key observation pionts (Table 1). Importantly, no significant differences were observed between the two groups (P = 0.324). The detailed data for each test point is displayed as follows: the abdominal aorta (4 vs. 3), coronary artery (5 vs. 8), iliac artery (1 vs. 0), heart valve (3 vs. 8), and aorta (27 vs. 21). Interestingly, our analysis revealed that over fifty per cent of patients in both groups had previously undergone CTA assessments and even interventional procedures (25 vs. 23, P = 0.410).

3.3 Quantitative and semi-quantitative assessment of image quality

Quantitative image quality did not exhibit any discernible differences between the two groups; however, the device group displayed fewer artefacts in the vena cava. Key indicators for objective assessment encompass HU, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) for specific anatomical regions, including the ascending aorta, descending aorta, left external iliac artery, and superior vena cava at the T12 level (Table 2). No sifnificant differences were observed in the comparison of HU for the left external iliac artery and superior vena cava. Subjects in the patients in conventional group showed a tendency of higher HU values than that in the novel protocol group for the ascending aorta (462.17 ± 36.78 vs. 446.54 ± 32.95, P = 0.049, PFDR corr. = 0.245) and descending aorta at the T12 level (458.62 ± 36.35 vs. 443.06 ± 28.44, P = 0.036, PFDR corr. = 0.245). SNR and CNR results indicated slightly superior measurements in the conventional group for the ascending aorta, descending aorta at the T12 level, and left external iliac artery. However, these differences lacked statistical significance (P > 0.05).

Table 2
www.frontiersin.org

Table 2. Comparison of objective scores of image quality between the two groups (X¯±s).

Semi-quantitative assessment of arterial images displayed substantial agreement between the two groups (kappa = 0.844, P < 0.001), as outlined in Table 3 and Figure 2. Compared with the conventional group, subjects in the novel protocolgroup showed fewer artefacts in superior vena cava, indicative of superior image quality (Table 4 and Figures 2, 3), and our subgroup analyses further supports these results (See Supplementary Tables S1–S4, Supplementary Figures S1–S4). Additionally, interventional procedures displayed no substantial disparity between the groups (25 vs. 23, P = 0.410).

Table 3
www.frontiersin.org

Table 3. A subjective index score of aortic image quality.

Figure 2
CT scans comparing the thoracic images of two patients. Panel A shows images from a patient in the regular group, labeled as Patient 4, with annotations marking certain areas in the scans. Panel B shows images from a patient in the device group, labeled as Patient 5, with similar annotations. Both panels display axial and coronal views, highlighting vascular structures and measurements within the chest cavity.

Figure 2. Comparative analysis of device group and regular group images. (A) Axial and maximum intensity projection views of the superior vena cava in the conventional group were obtained following aortic stent implantation in Patient 4. Noticeable pseudofiltering artefacts were observed in the superior vena cava images. (B) Patient 5, included in the novel protocol group, underwent computed tomography pulmonary angiography (CTPA). The image depicts CTPA findings specific to patients within the novel protocol group. Enhanced imaging of the superior vena cava is evident.

Table 4
www.frontiersin.org

Table 4. A subjective index score of superior vena cava image quality.

Figure 3
CT scans and 3D reconstructions of thoracic arteries in three patients. Panels A, B, C show a regular group patient’s axial view and artery visualization. Panels D, E, F display another patient from the device group with different cross-sections. Panels G, H, I show the third device group patient with emphasis on the aortic arch and descending aorta. Each set highlights anatomical differences and device impact.

Figure 3. Improved image quality of the superior vena Cava in the device group. (A–C) Patient 1, representative of the conventional group, underwent imaging with 80 mL of contrast agents. (A) Axial view of the superior vena cava, (B) Aortic Volume Rendering (VR), and (C) Maximum Intensity Projection (MIP) demonstrate well-developed contrast enhancement. The images are sharp and clear; however, a substantial amount of contrast agent within the superior vena cava interferes with the visualization of the aorta. (D–F) Patient 2, belonging to the novel protocol group, received 60 mL of contrast agents. (D) Sagittal view of the superior vena cava, (E) Aortic VR, and (F) Coronal view exhibit excellent image quality with distinct clarity. (G–I) Patient 3, also from the novel protocol group, was administered 60 mL of contrast agents post-aortic stent implantation. (G) Axial view of the superior vena cava, (H) Aortic VR, and (I) MIP demonstrate high-quality imaging devoid of artefacts while depicting an unobstructed superior vena cava.

3.4 Intergroup analysis of iodine intake and radiation dose

The iodine intake in the conventional group (26.02 ± 1.23 g) was significantly lower than that in the device group (20.81 ± 0.95 g) (P = 0.001), and our subgroup analyses further supports these results (see Supplementary Tables S1–S4, Supplementary Figures S1–S4). Patients in the conventional group received a slightly higher CTDIvol (mGy) value than those in the novel protocol group (8.37 ± 0.56 vs. 8.01 ± 0.76, P = 0.017). It is important to note that despite the reduction in iodine load in the device group, no significant intergroup differences of DLP and ED were observed (P > 0.05), as shown in Table 5.

Table 5
www.frontiersin.org

Table 5. Comparison of radiation dose and iodine load between the two groups of images (X¯±s).

4 Discussion

This is a comprehensive study to use a series clinical and imaging indicators to evaluate the effectiveness and clinical feasibility of our newly developed device for CTA examination. Our quantitative analysis of image quality showed that subjects in the conventional group had a tendency of higher HU values than that in the device group in the ascending aorta and descending aorta at the T12 level. The following semi-quantitative analysis of image quality showed that subjects in the device group showed fewer artefacts in superior vena cava. These findings suggest that the image quality of the device group was at least on par with that of the conventional group. More importantly, our intergroup analysis revealed that subjects in the device group had significantly lower iodine intake than that in the conventional group, indicating that this innovative approach has superiorities in managing iodinated contrast allergies or contrast agents-related biological injury without compromising image quality.

In this study, the final configuration of our apparatus resulted in a reduction of contrast agent dosage by 20% while maintaining imaging quality comparable to the conventional group. Mechanistically, CTDIvol, DLP, and ED are key indicators for assessing CT imaging radiation doses (18). Our findings demonstrate a reduction of about 4% in radiation dose in the equipment group, with no significant differences observed in DLP and ED, indicating consistent radiation conditions during scanning. During the analysis of aortic image quality, aortic HU, SNR, and CNR serve as important objective benchmarks. The results show very small differences in HU, SNR, and CNR, suggesting that our device effectively reduces contrast agent consumption while maintaining image quality following clinical diagnostic standards. Following the low-threshold trigger, contrast agent administration is halted, and 30 mL of saline tracer is administered to ensure sufficient contrast agent volume within the aorta before performing an 8 s delayed scan. Furthermore, increasing the volume of physiological saline injection reduces residual contrast agents within the superior vena cava, minimizing the impact of excessive contrast agents on aortic images and maintaining image quality. We believe that with the application of emerging technologies such as deep learning and personalized contrast agent usage protocols, it is possible to reduce further the dosage of contrast agents and the radiation dose patients receive. This holds the promise of being a progressive approach (19, 20).

Such significant benefits of lower iodine intake, while maintaining comparable image quality, from this newly developed device are crucial for our clinical practice. As is well known, contrast agents-related Acute Kidney Injury (CA-AKI), an important part of iatrogenic renal impairment, is particularly common in CT angiography (CTA) studies (21, 22). It has garnered significant clinical attention due to its high incidence (5%–20%) in patients with risk factors (23). The pathogenesis of CI-AKI is considered to be closely related to the typical hypoxic/toxic injury resulting from changes in renal microcirculation, hypoxia, and reactive oxygen species-mediated cell damage (2325). The impaired renal function in turn increased susceptibility to contrast agents. In addition to the renal impairment, the over usage of iodine contrast agents can cause damage to the thyroid gland (26). A rencent study with large sample size (more than 4 million patients) found that iodine exposure was linked to an increased risk of thyroid dysfunction (27). In some extreme clinical scenarios, patients may be exposed to significantly higher doses of contrast agents and radiation. For example, once patients undergo an interventional operation, they may experience multiple injections of contrast agents, such as multiple interventional operations, postoperative evaluation, and follow-up re-examinations. To solve these clinical issues, we introduces a novel approach that combines a saline chaser with a low-threshold triggering mechanism. This apparatus reduces contrast agent consumption and radiation exposure without compromising image quality, which are prelimilarily validated in our cohort, significantly improving its clinical utility.

While this study did not demonstrate a significant reduction in per-scan effective dose, the novel protocol presents a potential pathway for cumulative dose reduction in clinical practice. By substantially improving first-pass success rate and image quality (as shown in our results), it may reduce the need for repeat scans due to non-diagnostic images. Furthermore, the faster acquisition could decrease the number of required monitoring phases in complex cases. Future studies focusing on workflow and repeat-scan rates are warranted to quantify this potential benefit.

Our newly developed equipment demonstrates significant advantages even in complicated clinical scenarios. For instance, with the increasing ageing population, various forms of CVD necessitate a substantial demand for clinical applications of CTA (28, 29). In this study, we enrolled a cohort consisted of patients with various conditions including post-implantation diagnosis, screening, and assessment for conditions involving the heart, valves, aorta, abdominal aorta, and pulmonary system. As reported in previous studies, The saline chaser strategy or the approach of reducing the CTA trigger threshold aim to minimize contrast agent usage and contrast agents-associated risks (3032). Nevertheless, their demerits could not be ignored. For example, Choi's et al. found that the saline chaser strategy might compromise image quality for small vessel imaging, affecting diagnostic accuracy (11). Another imaging technique study reported that lowering the CTA trigger threshold might elevate image noise and decrease clarity (33). Being different from these techniques, the newly designed equipment in our study ingeniously overcomes such tricky clinical issues. The saline chaser we used here offers more precise control on the contrast agent injection rate and pressure, thereby maintaining image quality. Meanwhile, the low-threshold triggering we adopted in this study can reduce patient's exposure to radiation.

Several limitations should be addressed in this study. First, the research relies on a single-centre sample, introducing operator variability. Second, although diverse clinical scenarios are explored, broader applicability requires validation through multicenter, large-sample studies. Third, our study was designed as a pilot comparative analysis; thus, while the sample size was sufficient to detect key differences, it was not formally powered based on a pre-specified effect size, nor did it employ a pre-defined non-inferiority margin. A future trial aimed at formally demonstrating non-inferiority would need to prospectively establish such a clinically meaningful margin and calculate the sample size accordingly. The findings from our present work provide essential preliminary data to guide that process. Four, while patient allocation was performed via computer-generated simple randomization, we acknowledge that formal allocation concealment (e.g., sealed envelopes) was not implemented. Future definitive trials would benefit from incorporating such measures to minimize selection bias. Furthermore, the study possesses untapped potential; there is room for further optimization regarding the relationship between the saline chaser's speed, dose, and trigger threshold. However, recognizing these limitations directs future research to enhance and promote this innovative high-pressure injector and low-threshold triggering strategy.

5 Conclusion

Using a saline chaser with a low-threshold trigger for CTA, which involves lower iodine intake without sacrificing image quality, is a viable and promising method for clinical practice.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving humans were approved by Ethics Committee of Xiangya Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

LL: Formal analysis, Writing – original draft, Writing – review & editing. XZ: Formal analysis, Writing – review & editing, Supervision. XW: Conceptualization, Formal analysis, Funding acquisition, Investigation, Writing – original draft, Writing – review & editing.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fradi.2025.1709609/full#supplementary-material

References

1. Knuuti J, Wijns W, Saraste A, Capodanno D, Barbato E, Funck-Brentano C, et al. 2019 ESC guidelines for the diagnosis and management of chronic coronary syndromes the task force for the diagnosis and management of chronic coronary syndromes of the European Society of Cardiology (ESC). Eur Heart J. (2020) 41:407–77. doi: 10.1093/eurheartj/ehz425

PubMed Abstract | Crossref Full Text | Google Scholar

2. Tsao CW, Aday AW, Almarzooq ZI, Alonso A, Beaton AZ, Bittencourt MS, et al. Heart disease and stroke statistics-2022 update: a report from the American Heart Association. Circulation. (2022) 145:E153–639. doi: 10.1161/CIR.0000000000001052

PubMed Abstract | Crossref Full Text | Google Scholar

3. Moss AJ, Williams MC, Newby DE, Nicol ED. The updated NICE guidelines: cardiac CT as the first-line test for coronary artery disease. Curr Cardiovasc Imaging Rep. (2017) 10:1–7. doi: 10.1007/s12410-017-9412-6

Crossref Full Text | Google Scholar

4. Noda Y, Nakamura F, Kawai N, Suzuki R, Matsuo M, Ishihara T, et al. Optimized bolus threshold for dual-energy CT angiography with monoenergetic images: a randomized clinical trial. Radiology. (2021) 300:210102. doi: 10.1148/radiol.2021210102

Crossref Full Text | Google Scholar

5. Sedaghat F, Vadvala HV, Shan A, McMahon MT, Gawande RS. Incidence of contrast-associated acute kidney injury in renal-competent COVID-19 patients undergoing computed chest angiography. J Comput Assist Tomo. (2022) 46:701–6. doi: 10.1097/RCT.0000000000001337

PubMed Abstract | Crossref Full Text | Google Scholar

6. Hoste EAJ, Doom S, De Waele J, Delrue LJ, Defreyne L, Benoit DD, et al. Epidemiology of contrast-associated acute kidney injury in ICU patients: a retrospective cohort analysis. Intensive Care Med. (2011) 37:1921–31. doi: 10.1007/s00134-011-2389-8

PubMed Abstract | Crossref Full Text | Google Scholar

7. Andreucci M, Solomon R, Tasanarong A. Side effects of radiographic contrast media: pathogenesis, risk factors, and prevention. Biomed Res Int. (2014) 2014:741018. doi: 10.1155/2014/872574

PubMed Abstract | Crossref Full Text | Google Scholar

8. Kooiman J, Seth M, Share D, Dixon S, Gurm HS. The association between contrast dose and renal complications post PCI across the continuum of procedural estimated risk. PLoS One. (2014) 9:e90233. doi: 10.1371/journal.pone.0090233

PubMed Abstract | Crossref Full Text | Google Scholar

9. Verfaillie G, Franck C, De Crop A, Beels L, D'Asseler Y, Bacher K. A systematic review and meta-analysis on the radiation dose of computed tomography in hybrid nuclear medicine imaging. EJNMMI Phys. (2023) 10:32. doi: 10.1186/s40658-023-00553-8

PubMed Abstract | Crossref Full Text | Google Scholar

10. Tan SK, Ng KH, Yeong CH, Raja Aman RRA, Mohamed Sani F, Abdul Aziz YF, et al. Personalized administration of contrast medium with high delivery rate in low tube voltage coronary computed tomography angiography. Quant Imag Med Surg. (2019) 9:552–64. doi: 10.21037/qims.2019.03.13

PubMed Abstract | Crossref Full Text | Google Scholar

11. Kim DJ, Kim TH, Kim SJ, Kim DP, Oh CS, Ryu YH, et al. Saline flush effect for enhancement of aorta and coronary arteries at multidetector CT coronary angiography. Radiology. (2008) 246:110–5. doi: 10.1148/radiol.2453061953

PubMed Abstract | Crossref Full Text | Google Scholar

12. Masuda T, Funama Y, Nakaura T, Imada N, Sato T, Yamashita Y, et al. Delivering the saline chaser via a spiral flow-generating tube improves arterial enhancement for computed tomography angiography of the lower extremities. J Comput Assist Tomogr. (2015) 39:962–8. doi: 10.1097/RCT.0000000000000292

PubMed Abstract | Crossref Full Text | Google Scholar

13. Husarik DB, Bashir MR, Weber PW, Nichols EB, Howle LE, Merkle EM, et al. Contrast-enhanced magnetic resonance angiography: first-pass arterial enhancement as a function of gadolinium-chelate concentration, and the saline chaser volume and injection rate. Invest Radiol. (2012) 47:121–7. doi: 10.1097/RLI.0b013e3182300603

PubMed Abstract | Crossref Full Text | Google Scholar

14. Marin D, Nelson RC, Guerrisi A, Barnhart H, Schindera ST, Passariello R, et al. 64-Section multidetector CT of the upper abdomen: optimization of a saline chaser injection protocol for improved vascular and parenchymal contrast enhancement. Eur Radiol. (2011) 21:1938–47. doi: 10.1007/s00330-011-2139-x

PubMed Abstract | Crossref Full Text | Google Scholar

15. Washio H, Ohira S, Kanayama N, Wada K, Karino T, Komiyama R, et al. Effect of a saline flush technique for head and neck imaging in dual-energy CT: improvement of image quality and perivenous artefact reduction using virtual monochromatic imaging. Clin Radiol. (2019) 74:805–12. doi: 10.1016/j.crad.2019.06.010

PubMed Abstract | Crossref Full Text | Google Scholar

16. Kristiansen CH, Thomas O, Tran TT, Roy S, Hykkerud DL, Sanderud A, et al. Halved contrast medium dose in lower limb dual-energy computed tomography angiography-a randomized controlled trial. Eur Radiol. (2023) 33:6033–44. doi: 10.1007/s00330-023-09575-3

PubMed Abstract | Crossref Full Text | Google Scholar

17. Baliyan V, Shaqdan K, Hedgire S, Ghoshhajra B. Vascular computed tomography angiography technique and indications. Cardiovasc Diagn Ther. (2019) 9:S14–27. doi: 10.21037/cdt.2019.07.04

PubMed Abstract | Crossref Full Text | Google Scholar

18. Plocque A, Mitri C, Lefèvre C, Tabary O, Touqui L, Philippart F. Should we interfere with the interleukin-6 receptor during COVID-19: what do we know so far? Drugs. (2023) 83:1–36. doi: 10.1007/s40265-022-01803-2

PubMed Abstract | Crossref Full Text | Google Scholar

19. Wang K, Tan F, Zhu Z, Kong L. Exploring changes in depression and radiology-related publications research focus: a bibliometrics and content analysis based on natural language processing. Front Psychiatry. (2022) 13:978763. doi: 10.3389/fpsyt.2022.978763

PubMed Abstract | Crossref Full Text | Google Scholar

20. Ananthakrishnan L, Kay FU, Zeikus EA, Chu ES, Chang J, Barr JD, et al. What the baby formula and medical contrast material shortages have in common: insights and recommendations for managing the iodinated contrast media shortage. Radiol Cardiothorac Imaging. (2022) 4:e220101. doi: 10.1148/ryct.220101

PubMed Abstract | Crossref Full Text | Google Scholar

21. Weisbord SD, Palevsky PM, Kaufman JS, Wu H, Androsenko M, Ferguson RE, et al. Contrast-associated acute kidney injury and serious adverse outcomes following angiography. J Am Coll Cardiol. (2020) 75:1311–20. doi: 10.1016/j.jacc.2020.01.023

PubMed Abstract | Crossref Full Text | Google Scholar

22. Davenport MS, Perazella MA, Yee J, Dillman JR, Fine D, McDonald RJ, et al. Use of intravenous iodinated contrast media in patients with kidney disease: consensus statements from the American College of Radiology and the National Kidney Foundation. Radiology. (2020) 294:660–8. doi: 10.1148/radiol.2019192094

PubMed Abstract | Crossref Full Text | Google Scholar

23. Hossain M, Costanzo E, Cosentino J, Patel C, Vachharajani T. Contrast-induced nephropathy: pathophysiology, risk factors, and prevention. Saudi J Kidney Dis Transpl. (2018) 29:1. doi: 10.4103/1319-2442.225199

PubMed Abstract | Crossref Full Text | Google Scholar

24. Butt T, Lehti L, Apelqvist J, Gottsäter A, Acosta S. Contrast-associated acute kidney injury in patients with and without diabetes mellitus undergoing computed tomography angiography and local thrombolysis for acute lower limb ischemia. Vasc Endovasc Surg. (2022) 56:151–7. doi: 10.1177/15385744211051503

PubMed Abstract | Crossref Full Text | Google Scholar

25. Azzalini L, Spagnoli V, Ly HQ. Contrast-induced nephropathy: from pathophysiology to preventive strategies. Can J Cardiol. (2016) 32:247–55. doi: 10.1016/j.cjca.2015.05.013

PubMed Abstract | Crossref Full Text | Google Scholar

26. Hsieh M-S, Chiu C-S, Chen W-C, Chiang J-H, Lin S-Y, Lin M-Y, et al. Iodinated contrast medium exposure during computed tomography increase the risk of subsequent development of thyroid disorders in patients without known thyroid disease a nationwide population-based, propensity score-matched, longitudinal follow-up study. Medicine. (2015) 94:e2279. doi: 10.1097/MD.0000000000002279

PubMed Abstract | Crossref Full Text | Google Scholar

27. Inoue K, Guo R, Lee ML, Ebrahimi R, Neverova NV, Currier JW, et al. Iodinated contrast administration and risks of thyroid dysfunction: a retrospective cohort analysis of the U.S. veterans health administration system. Thyroid. (2023) 33:230–8. doi: 10.1089/thy.2022.0393

PubMed Abstract | Crossref Full Text | Google Scholar

28. Carroll BJ, Schermerhorn ML, Manning WJ. Imaging for acute aortic syndromes. Heart. (2020) 106:182–9. doi: 10.1136/heartjnl-2019-314897

PubMed Abstract | Crossref Full Text | Google Scholar

29. An H, Xie R, Ge Y, Wang T. Progress of CT aortic angiography combined with coronary artery in the evaluation of acute aortic syndrome. Front Cardiovasc Med. (2022) 9:1036982. doi: 10.3389/fcvm.2022.1036982

PubMed Abstract | Crossref Full Text | Google Scholar

30. Cademartiri F, Palumbo A, Maffei E, La Grutta L, Runza G, Pugliese F, et al. Diagnostic accuracy of 64-slice CT in the assessment of coronary stents. Radiol Med. (2007) 112:526–37. doi: 10.1007/s11547-007-0159-z

PubMed Abstract | Crossref Full Text | Google Scholar

31. de Monyé C, Cademartiri F, de Weert TT, Siepman DAM, Dippel DWJ, van Der Lugt A. Sixteen-detector row CT angiography of carotid arteries: comparison of different volumes of contrast material with and without a bolus chaser. Radiology. (2005) 237:555–62. doi: 10.1148/radiol.2372040653

PubMed Abstract | Crossref Full Text | Google Scholar

32. Cademartiri F, Luccichenti G, Marano R, Runza G, Midiri M. Use of saline chaser in the intravenous administration of contrast material in non-invasive coronary angiography with 16-row multislice computed tomography. Radiol Med. (2004) 107:497–505.15195012

PubMed Abstract | Google Scholar

33. Kanematsu M, Goshima S, Miyoshi T, Kondo H, Watanabe H, Noda Y, et al. Whole-body CT angiography with low tube voltage and low-concentration contrast material to reduce radiation dose and iodine load. Am J Roentgenol. (2014) 202:W106–16. doi: 10.2214/AJR.13.10720

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: computed tomography angiography, contrast-induced nephropathy, image quality, iodine contrast agent, saline chaser

Citation: Li L, Zhu X and Wu X (2026) Minimizing contrast agent dosage in CT angiography using a saline chaser with a low trigger threshold. Front. Radiol. 5:1709609. doi: 10.3389/fradi.2025.1709609

Received: 20 September 2025; Revised: 13 December 2025;
Accepted: 29 December 2025;
Published: 26 January 2026.

Edited by:

Zhicheng Peng, University of Pennsylvania, United States

Reviewed by:

Yuanhao Zhao, University of Washington, United States
Xinchi Xie, University of Southern California, United States

Copyright: © 2026 Li, Zhu and Wu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Xunhua Wu, Mjc5MzI4NTI3QHFxLmNvbQ==; Xiaojian Zhu, MzgwNTYzMDJAcXEuY29t

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.