Techniques for Non-Invasive Monitoring of Arterial Blood Pressure

imageAgnes S. Meidert1* imageBernd Saugel2
  • 1Department of Anaesthesiology, University Hospital, Ludwig-Maximilians-Universität München, Munich, Germany
  • 2Department of Anesthesiology, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany

Since both, hypotension and hypertension, can potentially impair the function of vital organs such as heart, brain, or kidneys, monitoring of arterial blood pressure (BP) is a mainstay of hemodynamic monitoring in acutely or critically ill patients. Arterial BP can either be obtained invasively via an arterial catheter or non-invasively. Non-invasive BP measurement provides either intermittent or continuous readings. Most commonly, an occluding upper arm cuff is used for intermittent non-invasive monitoring. BP values are then obtained either manually (by auscultation of Korotkoff sounds or palpation) or automatically (e.g., by oscillometry). For continuous non-invasive BP monitoring, the volume clamp method or arterial applanation tonometry can be used. Both techniques enable the arterial waveform and BP values to be obtained continuously. This article describes the different techniques for non-invasive BP measurement, their advantages and limitations, and their clinical applicability.

Background

Monitoring of arterial blood pressure (BP) is a mainstay of hemodynamic monitoring in acutely or critically ill patients. Close monitoring of BP is of great importance to detect and treat hypotension and hypertension early. Both, hypotension and hypertension can impair the function of vital organs, such as the brain (1), the heart (2), and the kidneys (3).

The direct measurement of BP via arterial cannulation is regarded as the clinical reference method (criterion standard). In clinical routine, it is commonly performed during high-risk surgery and in intensive care medicine. The cannulation of an artery, however, can be time-consuming, needs to be done by a trained operator, and is associated—although very rarely (4)—with potential major complications such as embolism, lesion of nerves or vessels, or ischemia. For these reasons, BP is very commonly measured non-invasively.

There are several ways to non-invasively measure BP. Monitoring techniques can be classified according to their ability to measure BP intermittently or continuously (Figure 1). In this article, we describe techniques for non-invasive monitoring of arterial BP and discuss their advantages, limitations, and clinical applicability.

FIGURE 1
www.frontiersin.org

Figure 1. Non-invasive blood pressure monitoring techniques.

Non-Invasive Intermittent Techniques

For intermittent BP measurement, an air-filled occluding cuff can be used that enables BP to be measured either manually or automatically. For all occluding cuffs, the right size is critical for valid measurement (5). Manual measurement of BP by an occluding cuff can be done either by palpation or auscultation (6).

With the palpatory method, an inflatable cuff is wrapped around the upper arm of a patient. The manometer connected to the cuff by a tube shows the pressure applied. The physician feels the radial pulse, inflates the cuff until the brachial artery collapses, and there is no blood flow any more. The pressure at which a pulse can be detected again while deflating the cuff corresponds to the systolic arterial pressure of the patient. This method does not need a stethoscope or any other specific skills or equipment and can also be performed in a noisy environment. However, it only provides the systolic arterial pressure. The auscultatory method is performed in a similar way; after inflation of the cuff to a pressure above the systolic pressure (verified by the vanished radial pulse), the typical Korotkoff sounds can be detected by a stethoscope applied distal of the upper arm cuff during slow deflation. The onset of the sounds corresponds to the patients’ systolic arterial pressure, the last sound at decreasing cuff pressure equals the patients’ diastolic arterial pressure. The advantage of this technique is that it provides the diastolic arterial pressure value, disadvantages include the need for training how to correctly apply this technique and the need of a stethoscope and a quiet environment.

An automated method to measure BP with the help of an occluding cuff employs the oscillometric technique. The cuff is inflated to a preset value automatically. Then, the pressure is gradually being reduced. The pressure wave causes oscillations in the vessel, which can be detected by the cuff. Mean arterial pressure corresponds to the maximum of oscillations (7); an algorithm applied to the change of oscillations sets systolic and diastolic arterial pressure values. These proprietary algorithms differ between manufacturers and are often not publicly available (8). The advantages of oscillometry are mainly the presence of reasonably accurate mean arterial pressure (in normal BP ranges) and the possibility of having an automated tool to determine a patient’s BP at a preset interval. The disadvantages are the overestimation of low and underestimation of high values (9, 10) and the possibility to falsify measurements [e.g., by movement (detected as oscillations) or the patient’s arm resting on the bed] (11).

The intermittent nature of BP measurements provided by all the techniques described earlier is a disadvantage they all have in common.

Non-Invasive Continuous Techniques

During the recent years, continuous non-invasive BP monitoring techniques became available that enable a real-time BP curve and numerical BP values to be assessed (just with direct BP measurement).

The continuous non-invasive measurement principles are based on either one of two different techniques, namely arterial applanation tonometry or the volume clamp method. Arterial applanation tonometry is based on the work of Pressman and Newgard (12), who found that a transducer strapped to an artery with a bone underneath, can obtain the arterial pulse wave. The technique has been refined and now is able to assess mean arterial pressure in the radial artery and allows the calculation of diastolic and systolic arterial pressure (e.g., using population-based algorithms) (13). The technique is used in cardiology to assess central vascular pressures (14). The pulse wave obtained by applanation tonometry can be analyzed and bears more information than systolic and diastolic pressure alone. However, these devices are not made for continuous patient monitoring as they have to be hand held by the examiner. A device allowing automated radial artery applanation tonometry is the T-Line system (Tensys Medical, San Diego, CA, USA) (15, 16). The system has been evaluated in various clinical settings (13, 1620).

The second technique for non-invasive continuous BP measurement is called volume clamp method (or vascular unloading technology) based on the work by Penaz et al. (21). The BP is measured at the finger with an inflatable cuff combined with a photodiode. The diameter of the artery in the finger is measured by the photodiode; the pressure in the cuff is adjusted to keep the diameter of the artery constant. From the pressure changes in the cuff, a BP curve can be calculated and transferred to correspond to brachial artery BP. Devices based on this technique are ClearSight (Edwards, Irvine, CA, USA) and CNAP (CNSystems Medizintechnik AG, Graz, Austria).

The continuous non-invasive devices are all sensitive to patient movement; therefore, monitoring of the conscious patient is possible but measurement results need to be checked for plausibility. In case of severe vasoconstriction, peripheral vascular disease, or distorted fingers due to arthritis, clinical experience has shown that it may be difficult to obtain a valid waveform using finger cuffs. Some patients report discomfort from the congestion in venous return from the fingertip where the cuff is place. For this reason, manufacturers recommend to change the cuff to another finger after a certain period of monitoring. In addition, compared to conventional intermittent devices for BP measurement, continuous BP monitoring is relatively expensive.

Are the Techniques Reliable?

Most clinicians ask themselves, whether the non-invasively obtained BP curve shows the “real” BP. Therefore, it is inevitable to discuss the measurement performance in terms of accuracy and precision of the various non-invasive devices. The pressure measured within an artery by means of arterial cannulation is regarded as the reference method of BP measurement. In the absence of direct BP measurement, the auscultatory method with a mercury column is regarded as the “gold standard.” However, as Alpert and colleagues (11) point out, the cuff/stethoscope method itself sometimes differs considerably from intra-arterial pressure. Nonetheless, when evaluating a new non-invasive device using an occluding cuff, reference measurements are performed by the auscultatory method (22). This has led to the common belief that the upper arm cuff measurements represent the “real” BP of a patient. Since the auscultatory method is now widely replaced by devices that engage an oscillometric technology clinicians trust the values produced by the device with the upper arm cuff (23). A survey by Chatterjee and colleagues (23) showed that even in critically ill patients on vasopressors the non-invasive upper arm BP measurement was used to guide therapy by 47% of respondents, although intensivists would be expected to know about the limitations of oscillometric measurement in unstable patients. However, big data base analyses of simultaneous measurements on ICU an OR have demonstrated that the devices using an oscillometric method tend to overestimate hypotensive BP values and to underestimate hypertensive BP values (9, 10). Within the normal BP range, the measurement of mean arterial pressure seems to be sufficiently accurate (9, 10, 24). Studies on the accuracy of oscillometric mean arterial BP in critically ill patients demonstrated that a possible source for inaccuracy lies within the choice of the wrong cuff size (25, 26). However, even when the correctly sized cuff was used, the results still showed clinically unacceptable discrepancy between invasive and non-invasive values (25, 26). Focusing on a possible relationship between obesity and inaccuracy of non-invasive BP measurement, Araghi et al. (27) studied overweight patients in the ICU. The analysis revealed clinically relevant inaccuracy of both, auscultatory and oscillometric, techniques (27), which therefore should not be used to guide therapy in critically ill patients. A similar study compared invasive and oscillometric BP measurement in obese patients undergoing non-cardiac surgery (28). In addition, the same group also examined the cuff position at the forearm of patients. However, oscillometric measurement in these patients in both locations did not allow sufficiently accurate monitoring of BP (28). This leads to the question, whether BP values resulting from oscillometric measurement with other locations for the cuff than the upper arm can be used for guidance of therapy. A single-center study in ICU patients showed acceptable agreement for oscillometric mean arterial BP compared to intra-arterial BP when the cuff was placed at the upper arm, whereas the thigh and ankle location revealed inaccurate values (29). In accordance to these findings, Drake and Hill (30) performed upper arm and ankle measurements during elective cesarean section. The values from the different sites varied considerably; therefore, the oscillometric measurement at the ankle cannot be seen as an alternative to the upper arm (30).

The reliability of non-invasive intermittent BP measurement in patients with arrhythmia has been questioned (31). Two studies have shown recently that there is no relevant difference between oscillometric measurement in patients with or without arrhythmia (32, 33).

For all non-invasive devices that measure BP continuously, numerous validation studies exist (3436). Kim et al. (35) pooled data from various studies comparing non-invasive continuous devices with direct BP, which have been published until 2013 and reported the mean of the differences with its SD. By this approach, they found an overall random-effect pooled bias for mean BP of 3.2 ± 8.4 mmHg. When stratifying the results according to the different measurement technologies described earlier, the analysis for mean BP yielded a bias and SD of 1.3 ± 5.7, 5.5 ± 9.3, and 3.5 ± 6.8 mmHg for the T-Line system, CNAP, and ClearSight, respectively (35). This analysis demonstrated that accuracy and precision of continuous non-invasive devices are not interchangeable with invasive BP measurement. Besides, the group criticized the lack of a recognized standard to define clinical acceptability (35). Vos et al. (37) concluded recently that non-invasive continuous monitoring with ClearSight was interchangeable with monitoring by an oscillometric technique. In their review from 2016, Bartels and colleagues (34) relate these findings to the well-known inaccuracy of oscillometry. The question is whether continuous non-invasive devices need to replace the direct measurement or rather fill the monitoring gap for patients who are insufficiently monitored by intermittent measurements only. Some clinicians find the ability to track changes in BP of the continuous devices particularly helpful in managing patient care.

In the end, the operator has to know about the limitations and pitfalls of any BP measuring technique, both non-invasive and invasive, to select the optimal technology for BP monitoring for the individual patient.

How Should We Measure BP?

The BP monitoring that we use for the individual patient needs to be tailored to the needs of the patients and the clinical setting.

For critically ill patients in the ICU, non-invasive BP monitoring is unlikely to play a big role in the foreseeable future. Although some researchers see the age of total non-invasive BP monitoring dawning (38), in our point of view critically ill patients need frequent arterial blood gas analysis as well as continuous and reliable measurement of absolute BP values.

In the emergency department, it is crucial to identify hemodynamic instability early. Intermittent BP monitoring, however, often is set at quite long intervals (e.g., 15 or 30 min) resulting in missing or only delayed detection of hypotension. There are studies that point out the advantage of continuous monitoring in terms early recognition of deterioration of the patient’s hemodynamic status (39, 40).

For patients undergoing surgical procedures, the appropriate method of BP monitoring needs to be identified considering perioperative cardiovascular risk stratification. There are different types of hypotension during general anesthesia and surgery, e.g., post-induction hypotension, early intraoperative hypotension, and late intraoperative hypotension with different risk factors (41). There is a growing body of evidence that continuous monitoring can be beneficial in terms of BP stability. As Walsh et al. (3) showed even periods of hypotension as short as a few minutes can adversely affect organ function. Therefore, BP measurement should first of all enable the physician to maintain BP stability in the patient. Benes and colleagues (42) have demonstrated that continuous BP measurement helps to keep the BP stable during surgery in beach chair positioning compared to intermittent measurements taken every 5 min. Recently, it has been shown in 160 patients with a history of hypertension that there are significantly less hypotensive episodes during induction of general anesthesia when a continuous method is used instead of intermittent oscillometric measurements every 3 min (43). For patients undergoing planned cesarean section, continuous monitoring helped to detect hypotensive episodes earlier and more often (44).

In the perioperative setting, the likelihood for intraoperative hypotension and the patient’s risk to develop hypoperfusion-induced organ failure should lead to the choice of which BP monitoring to use (Figure 2). We recommend advanced hemodynamic monitoring that allows monitoring of blood flow and fluid responsiveness parameters in the OR for high risk patients undergoing high-risk procedures (45) and in the ICU for patients with severe chronic and acute disease (Figure 2).

FIGURE 2
www.frontiersin.org

Figure 2. Choice of adequate blood pressure monitoring equipment in ICU and OR according to patient and procedural risk (OR) (45) or chronic and acute disease (ICU).

Conclusion

Blood pressure monitoring needs to be tailored to the individual patient. In stable, low-risk patients, intermittent oscillometric BP measurements are usually sufficient. Patients who are at risk for hemodynamic instability should be monitored by continuous BP measurement. Whether continuous non-invasive BP monitoring can improve patient outcome in certain patient collectives or clinical settings (perioperative medicine, emergency medicine) is the subject of the current clinical research. In critically ill patients, we still recommend the continuous invasive BP monitoring with an arterial catheter.

Author Contributions

All authors listed, have made a substantial, direct and intellectual contribution to the work, and approved it for publication.

Conflict of Interest Statement

AM has no conflict of interest. BS collaborates with Pulsion Medical Systems SE (Feldkirchen, Germany) as a member of the medical advisory board and received honoraria for giving lectures and refunds of travel expenses from Pulsion Medical Systems SE. BS received research support from Edwards Lifesciences (Irvine, CA, USA). BS received institutional research grants, unrestricted research grants, and refunds of travel expenses from Tensys Medical Inc. (San Diego, CA, USA). BS received honoraria for giving lectures and refunds of travel expenses from CNSystems Medizintechnik AG (Graz, Austria).

References

1. Bijker JB, Persoon S, Peelen LM, Moons KG, Kalkman CJ, Kappelle LJ, et al. Intraoperative hypotension and perioperative ischemic stroke after general surgery: a nested case-control study. Anesthesiology (2012) 116:658–64. doi:10.1097/ALN.0b013e3182472320

PubMed Abstract | CrossRef Full Text | Google Scholar

2. van Waes JA, van Klei WA, Wijeysundera DN, van Wolfswinkel L, Lindsay TF, Beattie WS. Association between intraoperative hypotension and myocardial injury after vascular surgery. Anesthesiology (2016) 124:35–44. doi:10.1097/ALN.0000000000000922

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Walsh M, Devereaux PJ, Garg AX, Kurz A, Turan A, Rodseth RN, et al. Relationship between intraoperative mean arterial pressure and clinical outcomes after noncardiac surgery: toward an empirical definition of hypotension. Anesthesiology (2013) 119:507–15. doi:10.1097/ALN.0b013e3182a10e26

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care (2002) 6:199–204. doi:10.1186/cc1489

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Petrie JC, O’Brien ET, Littler WA, de Swiet M. Recommendations on blood pressure measurement. Br Med J (1986) 293:611–5. doi:10.1136/bmj.293.6547.611

CrossRef Full Text | Google Scholar

6. Geddes LA, Voelz M, Combs C, Reiner D, Babbs CF. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng (1982) 10:271–80. doi:10.1007/BF02367308

CrossRef Full Text | Google Scholar

7. Ramsey M III. Noninvasive automatic determination of mean arterial pressure. Med Biol Eng Comput (1979) 17:11–8. doi:10.1007/BF02440948

CrossRef Full Text | Google Scholar

8. Smulyan H, Safar ME. Blood pressure measurement: retrospective and prospective views. Am J Hypertens (2011) 24:628–34. doi:10.1038/ajh.2011.22

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Wax DB, Lin HM, Leibowitz AB. Invasive and concomitant noninvasive intraoperative blood pressure monitoring: observed differences in measurements and associated therapeutic interventions. Anesthesiology (2011) 115:973–8. doi:10.1097/ALN.0b013e3182330286

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Lehman LW, Saeed M, Talmor D, Mark R, Malhotra A. Methods of blood pressure measurement in the ICU. Crit Care Med (2013) 41:34–40. doi:10.1097/CCM.0b013e318265ea46

CrossRef Full Text | Google Scholar

11. Alpert BS, Quinn D, Gallick D. Oscillometric blood pressure: a review for clinicians. J Am Soc Hypertens (2014) 8:930–8. doi:10.1016/j.jash.2014.08.014

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Pressman GL, Newgard PM. A transducer for the continuous external measurement of arterial blood pressure. IEEE Trans Biomed Eng (1963) 10:73–81.

Google Scholar

13. Meidert AS, Huber W, Hapfelmeier A, Schöfthaler M, Müller JN, Langwieser N, et al. Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care (2013) 28:908–12. doi:10.1016/j.jcrc.2013.06.012

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Nelson MR, Stepanek J, Cevette M, Covalciuc M, Hurst RT, Tajik AJ. Noninvasive measurement of central vascular pressures with arterial tonometry: clinical revival of the pulse pressure waveform? Mayo Clin Proc (2010) 85:460–72. doi:10.4065/mcp.2009.0336

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Saugel B, Dueck R, Wagner JY. Measurement of blood pressure. Best Pract Res Clin Anaesthesiol (2014) 28:309–22. doi:10.1016/j.bpa.2014.08.001

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Dueck R, Goedje O, Clopton P. Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput (2012) 26:75–83. doi:10.1007/s10877-012-9336-2

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W. The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med (2012) 38:1471–7. doi:10.1007/s00134-012-2617-x

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W. Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth (2013) 111:185–90. doi:10.1093/bja/aet025

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Meidert AS, Huber W, Müller JN, Schöfthaler M, Hapfelmeier A, Langwieser N, et al. Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure. Br J Anaesth (2014) 112:521–8. doi:10.1093/bja/aet400

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R. An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia (2008) 63:307–12. doi:10.1111/j.1365-2044.2007.05369.x

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Penaz J, Voigt A, Teichmann W. [Contribution to the continuous indirect blood pressure measurement]. Z Gesamte Inn Med (1976) 31:1030–3.

PubMed Abstract | Google Scholar

22. Association for the Advancement of Medical Instrumentation. American National Standard for Manual, Electronic, or Automated Sphygmomanometers. SP 10. Arlington, VA: Association for the Advancement of Medical Instrumentation (2002).

Google Scholar

23. Chatterjee A, DePriest K, Blair R, Bowton D, Chin R. Results of a survey of blood pressure monitoring by intensivists in critically ill patients: a preliminary study. Crit Care Med (2010) 38:2335–8. doi:10.1097/CCM.0b013e3181fa057f

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Picone DS, Schultz MG, Otahal P, Aakhus S, Al-Jumaily AM, Black JA, et al. Accuracy of cuff-measured blood pressure: systematic reviews and meta-analyses. J Am Coll Cardiol (2017) 70:572–86. doi:10.1016/j.jacc.2017.05.064

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, Woisetschläger C, et al. Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients. Crit Care Med (2000) 28:371–6. doi:10.1097/00003246-200002000-00014

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Bur A, Herkner H, Vlcek M, Woisetschläger C, Derhaschnig U, Delle Karth G, et al. Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients. Crit Care Med (2003) 31:793–9. doi:10.1097/01.CCM.0000053650.12025.1A

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Araghi A, Bander JJ, Guzman JA. Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive? Crit Care (2006) 10:R64. doi:10.1186/cc4411

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Anast N, Olejniczak M, Ingrande J, Brock-Utne J. The impact of blood pressure cuff location on the accuracy of noninvasive blood pressure measurements in obese patients: an observational study. Can J Anaesth (2016) 63:298–306. doi:10.1007/s12630-015-0509-6

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Lakhal K, Macq C, Ehrmann S, Boulain T, Capdevila X. Noninvasive monitoring of blood pressure in the critically ill: reliability according to the cuff site (arm, thigh, or ankle). Crit Care Med (2012) 40:1207–13. doi:10.1097/CCM.0b013e31823dae42

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Drake MJ, Hill JS. Observational study comparing non-invasive blood pressure measurement at the arm and ankle during caesarean section. Anaesthesia (2013) 68:461–6. doi:10.1111/anae.12194

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Sykes D, Dewar R, Mohanaruban K, Donovan K, Nicklason F, Thomas DM, et al. Measuring blood pressure in the elderly: does atrial fibrillation increase observer variability? BMJ (1990) 300:162–3. doi:10.1136/bmj.300.6718.162

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Lakhal K, Ehrmann S, Martin M, Faiz S, Réminiac F, Cinotti R, et al. Blood pressure monitoring during arrhythmia: agreement between automated brachial cuff and intra-arterial measurements. Br J Anaesth (2015) 115:540–9. doi:10.1093/bja/aev304

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Lakhal K, Martin M, Ehrmann S, Faiz S, Rozec B, Boulain T. Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia. J Clin Monit Comput (2017). doi:10.1007/s10877-017-0067-2

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Bartels K, Esper SA, Thiele RH. Blood pressure monitoring for the anesthesiologist: a practical review. Anesth Analg (2016) 122:1866–79. doi:10.1213/ANE.0000000000001340

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, et al. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology (2014) 120:1080–97. doi:10.1097/ALN.0000000000000226

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Ameloot K, Palmers PJ, Malbrain ML. The accuracy of noninvasive cardiac output and pressure measurements with finger cuff: a concise review. Curr Opin Crit Care (2015) 21:232–9. doi:10.1097/MCC.0000000000000198

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Vos JJ, Poterman M, Mooyaart EA, Weening M, Struys MM, Scheeren TW, et al. Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia. Br J Anaesth (2014) 113:67–74. doi:10.1093/bja/aeu091

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Lakhal K, Ehrmann S, Boulain T. Non-invasive blood pressure monitoring in the critically ill: time to abandon the intra-arterial catheter? Chest (2017). doi:10.1016/j.chest.2017.10.030

CrossRef Full Text | Google Scholar

39. Wagner JY, Prantner JS, Meidert AS, Hapfelmeier A, Schmid RM, Saugel B. Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department. Scand J Trauma Resusc Emerg Med (2014) 22:8. doi:10.1186/1757-7241-22-8

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, et al. Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med (2011) 29:782–9. doi:10.1016/j.ajem.2011.05.014

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Südfeld S, Brechnitz S, Wagner JY, Reese PC, Pinnschmidt HO, Reuter DA, et al. Post-induction hypotension and early intraoperative hypotension associated with general anaesthesia. Br J Anaesth (2017) 119:57–64. doi:10.1093/bja/aex127

PubMed Abstract | CrossRef Full Text | Google Scholar

42. Benes J, Simanova A, Tovarnicka T, Sevcikova S, Kletecka J, Zatloukal J, et al. Continuous non-invasive monitoring improves blood pressure stability in upright position: randomized controlled trial. J Clin Monit Comput (2015) 29:11–7. doi:10.1007/s10877-014-9586-2

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Meidert AS, Nold JS, Hornung R, Paulus AC, Zwissler B, Czerner S. The impact of continuous non-invasive arterial blood pressure monitoring on blood pressure stability during general anaesthesia in orthopaedic patients: a randomised trial. Eur J Anaesthesiol (2017) 34(11):716–22. doi:10.1097/EJA.0000000000000690

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, et al. Detection of hypotension during caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth (2012) 109:413–9. doi:10.1093/bja/aes224

PubMed Abstract | CrossRef Full Text | Google Scholar

45. Pearse RM, Harrison DA, James P, Watson D, Hinds C, Rhodes A, et al. Identification and characterisation of the high-risk surgical population in the United Kingdom. Crit Care (2006) 10:R81. doi:10.1186/cc5122

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: blood pressure monitoring, perioperative monitoring, non-invasive blood pressure, arterial pressure, oscillometry, applanation tonometry, vascular unloading technique

Citation: Meidert AS and Saugel B (2018) Techniques for Non-Invasive Monitoring of Arterial Blood Pressure. Front. Med. 4:231. doi: 10.3389/fmed.2017.00231

Received: 18 October 2017; Accepted: 04 December 2017;
Published: 08 January 2018

Edited by:

Samir G. Sakka, Witten/Herdecke University, Germany

Reviewed by:

Vsevolod V. Kuzkov, Northern State Medical University, Russia
Andres Zorrilla-Vaca, University of Valle, Colombia

Copyright: © 2018 Meidert and Saugel. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Agnes S. Meidert, agnes_meidert@web.de

Download