SYSTEMATIC REVIEW article

Front. Hum. Neurosci., 01 September 2025

Sec. Brain Imaging and Stimulation

Volume 19 - 2025 | https://doi.org/10.3389/fnhum.2025.1636115

Methodologies to detect cortico-cortical evoked potentials: a systematic review

  • 1Department of Electrical and Computer Engineering, Maroun Semaan Faculty of Engineering and Architecture, American University of Beirut, Beirut, Lebanon
  • 2Department of Neurosurgery, Mass General Brigham, Harvard Medical School, Boston, MA, United States
  • 3Center for Brain Circuit Therapeutics, Brigham and Women’s Hospital, Harvard Medical School, Boston, MA, United States

Introduction: Cortico-cortical evoked potentials (CCEPs) are electrophysiological responses elicited by direct electrical stimulation of one cortical region and recorded from another, providing insights into functional connectivity and communication pathways between brain areas. However, no consistent standard for defining and measuring CCEPs currently exists.

Methods: We conducted a systematic review of the CCEP literature on detection methods to evaluate commonalities and gaps in methodology. Extracted data included demographics, disease, recording type, montage, recording system, stimulation amplitude and frequency, time window used for epoching around stimulus onset, open access availability, and detection approach.

Results: Of 187 studies undergoing full-text review, over half lacked a description of the CCEP detection method. Specifically, 9.1% utilized visual identification, whereas 49.74% did not explicitly state the method. The remaining 72 studies represented 3,424 patients, of whom 58.3% had sEEG electrodes and most had epilepsy. The most common detection method was threshold-based (68.1%), followed by statistical testing (16.7%) to determine whether CCEPs differed significantly from baseline, data-driven methods (4.1%) that quantify responses after learning from data, and frequency-based approaches (4.1%). Bipolar (48.6%) and single-electrode referential montages (18.1%) were most frequently employed.

Discussion: Current CCEP detection methods lack consensus, with many studies omitting methodological details and relying heavily on threshold-based techniques that assume fixed response shapes. Future research should encourage the use of data-driven approaches, which learn directly from data, offer more robust alternatives, and improve quantification in both clinical and research contexts.

Systematic review registration: https://www.crd.york.ac.uk/PROSPERO/view/CRD42024568261, identifier CRD42024568261.

1 Introduction

Cortical stimulation is a valuable tool for understanding how different brain regions communicate, a fundamental aspect for both neuroscience research and clinical applications. One powerful approach to mapping functional connectivity is cortico-cortical evoked potential (CCEP) recording, which measures how electrical stimulation in one brain region elicits responses in other areas (Prime et al., 2018). Unlike non-invasive imaging techniques like EEG or MRI, which infer connectivity indirectly, CCEPs measure real-time brain network dynamics.

While the physiological basis of cortico-cortical evoked potentials builds on decades of neurophysiological research, the modern concept and clinical application of CCEPs was primarily defined by Matsumoto et al. (2004) who demonstrated this phenomenon during their study on functional connectivity in the human language system using single-pulse electrical stimuli. CCEPs employ low-frequency stimulation (<2 Hz) via stereo-electroencephalography (SEEG) or electrocorticography (ECoG) electrodes and responses are measured at other implanted sites (Prime et al., 2018). CCEPs have been used to map a variety of networks related to language (Matsumoto et al., 2004, 2007), limbic (Enatsu et al., 2015), motor, auditory (Brugge et al., 2003; Howard et al., 2000) and visual systems (Huang et al., 2023). CCEPs are often used clinically in intraoperative brain mapping to preserve essential functional areas during epilepsy or tumor surgeries (Saito et al., 2014; Seidel et al., 2024; Mariani et al., 2021). They have also helped identify seizure onset zones (van’t Klooster et al., 2011) and biomarkers of epileptogenicity (Mouthaan et al., 2016).

Cortico-cortical evoked potentials are thought to comprise several components, including an early (10–30 ms) negative deflection termed “N1” and a later (80–250 ms) slow wave termed “N2” (Keller et al., 2014). More recently, there has been a shift toward using positive/negative deflections (D1/D2) instead of N1/N2 (Feys et al., 2025; Alarcón et al., 2018). Since most studies we included still referred to those responses as N1 and N2 components, we use that terminology for consistency. The definition of a “true” CCEPs (vs. artifacts, volume-conducted potentials, or spurious signals) and the methods for detecting them remain controversial. Many CCEP studies provide little to no detail on their detection methods, and those that do exhibit significant variability. Some rely solely on visual identification of waveforms resembling CCEPs, others use specific amplitude thresholds to define N1 and N2 waves, and some apply statistical analyses to confirm CCEP presence (Matsumoto et al., 2012; Kundu et al., 2020; Valencia et al., 2023). Additionally, most methods rely on pre-defined CCEP shapes, but this may overlook patient- or region-specific variation in response patterns, potentially leading to missed detections or misclassification of responses as artifacts. In summary, no consistent standard for defining and measuring CCEPs currently exists.

In this review, we aimed to comprehensively evaluate and summarize currently used CCEP detection methods. We define cortico-cortical evoked potentials (CCEPs) as “brain” responses elicited by low-frequency electrical stimulation (<2 Hz) that are explicitly assessed for either clinical or statistical significance, either through trial-averaged voltage waveforms or frequency-based markers. Our objectives were to provide clarity on the variety of CCEP detection methods being utilized in the field, to define commonalities and gaps in methodology, and to identify opportunities for standardization.

2 Materials and methods

A systematic review was conducted following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA 2020) guidelines (Liberati et al., 2009). The study protocol was registered with PROSPERO (CRD42024568261).

2.1 Eligibility criteria

The inclusion criteria were human studies, covering all primary studies in English (retrospective, prospective, randomized trials, etc.), with any mention of CCEPs either in the title, abstract or full text. Studies of any disease type, population, patient age and sex were included. Literature reviews, systematic reviews, meta-analyses, guidelines, letters to the editor, commentaries, or trial protocols were excluded.

2.2 Information sources and search strategy

The Medline, Embase, and Web of Science databases were searched using the following keywords: (“cortico cortical evoked potential” or “CCEP” or “single pulse electrical stimulation”). Searches were conducted from database inception through July 11, 2024, using the terms. Full search syntax for each database is provided in Supplementary Material 1. We also performed a backward and forward citation analysis and used each database’s “similar articles” feature.

2.3 Selection process, data extraction, and items

After deduplication using Covidence, three reviewers (TA, AW, MW) independently screened titles and abstracts, followed by full-text review. Discrepancies were resolved by consensus. A PRISMA flow diagram is provided in Figure 1. Studies with no explicit mention of methodology for CCEP quantification were excluded. Also, studies employing high frequency stimulation were excluded, since CCEPs are evoked using low frequency stimulation (Matsumoto et al., 2004). Excluded articles and reasons for exclusion are listed in Supplementary Material 2.

FIGURE 1
Flowchart illustrating the identification of studies via databases and registers. Identification phase: 2,855 records extracted from Pubmed, Embase, and Web of Science; 631 duplicates removed. Screening phase: 2,224 records screened; 1,782 not related. 442 reports sought for retrieval; 254 full texts not retrieved. 187 reports assessed for eligibility; 115 reports excluded due to unspecified methods, visual identification, or language. Inclusion phase: 72 studies included.

Figure 1. PRISMA screening flowchart.

Data collected from studies included recording type (sEEG or ECoG), montage, recording system, stimulation amplitude and frequency, epoch limits, whether the data used was open access, and the detection method. Recording type was categorized based on the terminology used by the original studies, and no reclassification was applied based on surgical technique or electrode geometry. Demographic data like study year, sample size, patient sex, age, and disease type were collected. All data was manually extracted by authors, and any uncertainty or discrepancy was resolved via consensus. The data extracted for each included study is available in Supplementary Material 3.

2.4 Quality assessment

Although this review focused on methodological reporting rather than outcomes, we qualitatively assessed the completeness of methodological descriptions. Studies lacking essential reporting were excluded to maintain interpretive granularity.

2.5 CCEP pre-processing

In this section, we describe some common practices in CCEP pre-processing. A stimulation trial involves recording the CCEP response on one electrode following the delivery of an electrical stimulus through another electrode. The steps preceding CCEP detection usually involve the epoching of the continuous data into the pre- and post-stimulation data relative to the stimulation onset. The stimulation artifact, which lasts around 1–6 ms, is usually excluded from the analysis by discarding, interpolating, or using more sophisticated filters (e.g. Wiener filters) to mitigate its effect (Prime et al., 2018). The data then undergoes normalization relative to the pre-stimulus data across trials and re-referencing followed by filtering for noise reduction (Levinson et al., 2024).

2.6 CCEP detection approaches

Cortico-cortical evoked potential detection methodologies were grouped into four categories: (1) peak detection and thresholding (Mouthaan et al., 2016), which identify a characteristic peak of CCEPs and compare its amplitude to a threshold; (2) statistical tests to classify CCEPs by identifying significant differences between the post- and pre-stimulation periods (Dou et al., 2023); (3) frequency-based methods that detect frequency changes or spectral markers (van den Boom et al., 2024); (4) data-driven methods use the data as the basis to determine the shapes of the evoked responses rather than relying on pre-defined assumptions (Miller et al., 2023).

Figure 2 illustrates the typical workflow for CCEP pre-processing and detection. The raw data in Figure 2A is first segmented into pre- (−1500 to −5 ms) and post-stimulation (5–1500 ms) excluding the −5 to 5 ms window to remove stimulation artifacts. The data is then filtered, re-referenced, and normalized, resulting in Figure 2B. Figure 2C highlights the heterogeneity of CCEP detection methods by showcasing four approaches. In the first, a CCEP is identified when the average post-stimulation data exceeds 6 times the pre-stimulation standard deviation (van den Boom et al., 2024). In the second method, the root-mean-square (RMS) of the pre- and post-stimulation data is compared using a Wilcoxon rank-sum test (Boulogne et al., 2022; Almashaikhi et al., 2014). For the frequency-based method, the 20–200 ms response is filtered between 10 and 30 Hz, and 20 Hz power is computed using wavelet transform and compared to the mean overall power for significance (van den Boom et al., 2024). Finally, canonical response parametrization (CRP), a data-driven method for CCEP detection, projects the first trial onto the others to obtain projection magnitudes. Trials are classified as CCEPs based on the significance of these magnitudes at the response duration, with canonical shapes visualized (Miller et al., 2023). Further details on CRP are provided in Section “3.3.4 Data-driven methods.”

FIGURE 2
Diagram detailing CCEP detection methods. Section A shows raw data before and after stimulation. Section B presents pre-processed data with pre- and post-stimulation graphs. Section C illustrates four detection methods: 1) Threshold-based, highlighting a cutoff line; 2) Statistical testing, using max-normalized data and rank-sum tests; 3) Frequency-based, employing a wavelet power spectrum; 4) Data-driven, projecting trial data to discover canonical shapes for significant response prediction. Each method is depicted with relevant graphs and annotations.

Figure 2. Cortico-cortical evoked potential pre-processing and detection workflow. (A) Raw data with stimulation artifact shaded in red. (B) Data after epoching, filtering and baseline normalization. (C) Different CCEP detection methods: (1) threshold- based detection, (2) statistical testing-based methods, (3) frequency based methods, and (4) data-driven methods.

3 Results

3.1 Study selection, characteristics and methods

Interest in CCEPs has grown, as reflected by the rising number of related studies (Figures 3A, B). A total of 2,855 studies were screened by title and abstract, with 187 studies undergoing full text review (Figure 1). Of the 187 studies, 9.1% (n = 17) utilized visual identification, while 49.7% (n = 93) did not explicitly mention the detection method. After excluding these studies, 72 remained for review, representing 3,424 patients with a mean age of 30.3 ± 8.3 years (mean ± 1 SD) and an age range of 3–69 years, of whom 1,652 were female (48.2%). Across the 72 studies, CCEPs were recorded with sEEG (n = 42, 58.3%), ECoG (n = 22, 30.6%), or both (n = 8, 11.1%). Most studies focused on patients with epilepsy (n = 67 studies, 93.1%), with some including those with brain tumors (n = 2, 2.8%), or both (n = 3, 4.1%) (Table 1). The stimulation protocols in these studies utilized a mean frequency of 0.7 ± 0.4 Hz (range = 0.06–2 Hz) and amplitude of 5.8 ± 3 mA (range = 0.35–15 mA) to elicit CCEPs. Only 13 studies published their data (comprising 18.1% of papers with a detailed detection method description), with an additional 3 papers sharing their analysis code but not the raw data.

FIGURE 3
Chart A shows the annual number of papers published on CCEPs from 2002 to 2024, with an upward trend reaching 18 papers in 2021. Chart B displays the cumulative number of papers over the same period, steadily increasing to about 170 papers by 2024.

Figure 3. Publication trends in CCEP research. (A) Number of papers published per year surrounding CCEPs. (B) Cumulative number of papers published on CCEPs over time.

TABLE 1
www.frontiersin.org

Table 1. Study characteristics of systematic review, for those reporting their methods.

The reviewed papers fell into one of four categories based on the methods used for CCEP detection. Some papers employed a mix of these methods or outline multiple implementations. We observe that before 2011 most methods were either not explicitly described or rely on visual inspection. The most common methods (n = 49, 68.1%) involved different thresholding or peak detection techniques to identify CCEPs. The second most common method (n = 12, 16.7%) employs statistical tests to determine the prominence of the responses. Data-driven methods are employed in some papers (n = 3, 4.1%), while others (n = 3, 4.1%) use frequency-based methods. The remaining papers (n = 5, 6.9%) includes a combination of methods. A summary of the different subcategories of each detection method and the corresponding studies that utilize them are presented in Table 2.

TABLE 2
www.frontiersin.org

Table 2. Summary of CCEP detection methods grouped by approach.

3.2 Re-referencing montages

The bipolar montage (n = 35, 48.6%) was the most common, followed by several papers where the montage was not explicitly stated (n = 13, 18.1%) and the common average reference (CAR) (n = 4, 5.6%). Some used single-electrode reference with extracranial placements including the skin above the mastoid bone or midline scalp, and intracranial placement using a white matter electrode contact (n = 13, 18.1%). Realizing that bipolar re-referencing affects the polarity of the recording and CAR can introduce bias when many electrodes show an elevated response, some opted for an adjusted CAR (n = 5, 6.8%), where 20% of channels with lowest variance (Valencia et al., 2023; van den Boom et al., 2024) or the mean of the unstimulated channels are utilized (Huang et al., 2023; Miller et al., 2023). Another paper (Campbell et al., 2023) utilized a common median reference (n = 1, 1.4%) that is robust to extreme values introduced by the stimulation artifact (Rolston et al., 2009). One paper utilized both the bipolar and referential montage for comparison (Shahabi et al., 2021).

3.3 Detection methods

3.3.1 Peak and threshold-based detection

The most common method for CCEP detection (n = 49, 68.1%) was based on comparing the amplitude of the post-stimulation response to that of the baseline. Five papers defined amplitude thresholds based on the data and considered responses exceeding them as CCEPs (Yang et al., 2024; Kamali et al., 2020; Hebbink et al., 2019, 2020; Boulogne et al., 2016). Many studies utilized peak detection algorithms to locate characteristic CCEP deflections usually the N1 peak, and then assessed whether its amplitude surpassed a threshold based on the baseline standard deviation (Mouthaan et al., 2016; van den Boom et al., 2024; van Blooijs et al., 2018, van Blooijs et al., 2023a,b, 2024; Nakae et al., 2020; Takeyama et al., 2019; Zhao et al., 2019; Usami et al., 2015, 2017; Yamao et al., 2021; Parvizi et al., 2021; Russo et al., 2021, 2023; Zauli et al., 2022; Sun et al., 2021; Crocker et al., 2021; Luo et al., 2022; Parmigiani et al., 2022; Wu et al., 2023; van’t Klooster et al., 2017; Keller et al., 2011; Filipiak et al., 2021; Cornblath et al., 2023). Boido et al. (2014) employed two different thresholds, peak amplitude and normalized area of the response trace, considering a contact active if either was met.

Another common method defined CCEPs as trials exceeding a z-score threshold after normalization (Hays et al., 2021a,b, 2023a,b; Jedynak et al., 2023, 2024; Lemaréchal et al., 2022; Novitskaya et al., 2020; Silverstein et al., 2020; Trebaul et al., 2018; Usami et al., 2019b; Tóth et al., 2021; Mégevand et al., 2017). Other studies classified trials as CCEPs if the N1 peak-to-peak amplitude (Enatsu et al., 2012, 2013a; 2013b; Yamao et al., 2017) or the RMS of the response exceeded 10% or 20% of the maximum value across electrodes. Additionally, two other studies (Kundu et al., 2020; Campbell et al., 2023) emphasized sustained responses, requiring the post-stimulation amplitude exceeds three times the pre-stimulation median for at least 15 ms.

3.3.2 Statistical testing

Some studies (n = 12, 16.7%) employed statistical tests to determine whether responses differed from baseline. Methods included comparing the average pre- and post-stimulation activity using paired t-tests (Dou et al., 2023) or Welch-tests (Kamada et al., 2017, 2020). Another paper utilized Wilcoxon signed-rank test to compare 1-s pre- and post-stimulation periods and confirmed the CCEP detection by ensuring the amplitude exceeds twice the baseline standard deviation (Boulogne et al., 2022). In another paper, responses were first selected on the basis of visual identification followed by a sliding window that compares each 5 ms section post-stimulation to a 1 s pre-stimulation period using a Wilcoxon test (Almashaikhi et al., 2014).

Mann-Whitney U test was also used to rank the data points and assess the statistical difference between the two groups (Mălîia et al., 2017). CCEPs were also assessed using non-parametric cluster-based testing, where t-statistics were computed at each time point, significant points clustered, and cluster-level statistics formed by summing t-values. Permutation testing then determined statistical significance against the null distribution (Huang et al., 2019). One study analyzed the early, middle, and late response periods of the averaged CCEP using a sign permutation test, randomly multiplying each trial by ±1 to test for significance, since randomization would produce no effect without a CCEP (Shahabi et al., 2021).

The correlation between the stimulation current intensity and post-stimulation response was assessed using Spearman’s correlation to determine responses tied to stimulus intensity (Oane et al., 2020; Arbune et al., 2020). Mălîia et al. (2018) applied this method after ensuring the amplitude exceeds a patient-specific threshold. Similarly, Donos et al. (2016) first identified CCEPs based on the correlation between stimulation intensity and average RMS across trials, then required the trial RMS exceeds a threshold.

Finally, remaining studies focused on detecting prominent peaks for further statistical analysis. One detected the earliest CCEP peak and applied a t-test with permutation testing and temporal cluster correction to assess significance (Krieg et al., 2017), while another utilized the maximum statistic principle with permutation testing (Feys et al., 2024a,2024b).

3.3.3 Frequency-based methods

van den Boom et al. (2024) presented an open-source python package “ER-detect” that employs several methods for CCEP detection. One of which targets the typical early response described as an “S-shape” observed in CCEP data using visual identification. It has a period of approximately 50 ms, exhibiting high power at 20 Hz. The signal between 12 and 90 ms post-stimulation was isolated, filtered (10–30 Hz), and the power was computed using a Hilbert transform, which was compared to a threshold to classify the response. Other methods extracted the broadband gamma signals by filtering (70–170 Hz) the signal. Crowther et al. (2019) used the Hilbert transform to extract the signal amplitude envelope to compute the signal-to-noise ratio (SNR) according to the procedure in Schalk et al. (2007). The SNR was computed by comparing variance across the whole response duration to that of subdivided periods. An SNR > 1 revealed that the signal was modulated by the stimulus, since the variance of the whole signal was greater than the average across the shorter periods. It was then assessed for significance by testing whether the SNR is significantly different from 1 using a randomization test. Usami et al. (2019a) utilized a matching pursuit algorithm to compute the time-frequency (TF) energy of the responses between 100 and 550 ms post-stimulation to obtain cortico-cortical spectral responses (CCSRs), focusing on significant increase in CCSR high-gamma frequencies (80–150 Hz) for further analysis. Another approach by Davis et al. (2018) started with TF decomposition of the data between −0.2 s and 0.2 s centered on the stimulus onset using a multi-taper method. The TF values were used to estimate the inter-trial coherence, a measure of phase alignment across trials, that was tested for significance using a bootstrap technique. Mouthaan et al. (2016) used the Hilbert-Huang Transform (HHT) to obtain a TF matrix from which Event Related Spectral Pertubation (ERSP) images were constructed after averaging across trial epochs per stimulus and response electrode pair. The early responses to stimulation were detected through the significance of spectral perturbations observed in the ERSPs images by a bootsrapping method comparing the deviation to the baseline power.

3.3.4 Data-driven methods

The canonical response parametrization (CRP) method was based on exploring the data to extract the shapes of the CCEP responses labeled canonical responses rather than detecting CCEPs based on predefined shapes or assumptions (Miller et al., 2023). It started by unit-normalizing the trial voltage traces then projecting them on the remaining trials using a semi-normalized dot product while excluding self-projections to obtain a projection magnitude matrix. The temporal structure of the responses was also examined to quantify the length of the significant response by calculating the projection magnitudes while varying the response duration over which the projection is performed. This temporal projection served as a measure of mutual information between the responses, and its peak represented the point after which no additional information is significant, which was used to classify responses as CCEPs by using t-tests to check its significance (Valencia et al., 2023). The “ER-detect” toolbox mentioned in Section “3.3.3 Frequency-based methods” also included detection method based on CRP called “inter-trial similarity,” where the t-values of the right-tailed t-tests, evaluating the projection magnitudes against zero, were compared to a threshold to determine significance (van den Boom et al., 2024). In CRP, the authors also presented a method to parametrize the data into canonical responses and residuals. The canonical shape was obtained by applying linear kernel PCA to identify the eigenvectors of the voltage response matrix where the first one captured the principal component of the data reflecting the most significant features of the response, thus forming the canonical shape. Individual trials Vk(t) in the data were represented as Vk(tkC(t) + εk(t), where C(t) is the canonical response, αk is a scaling factor representing the strength of representation of the canonical shape in trial k, and εk(t) is the residual not explained by C(t). In a previous study (Huang et al., 2023), the authors used basis profile classification (Miller et al., 2021) using a similar methodology to that described earlier to obtain profile curves and classify distinct CCEP shapes across different brain regions.

4 Discussion

This study sought to provide a comprehensive summary of the literature on CCEP detection methodologies. Notable variations in methodologies were revealed, highlighting the need for standardization to ensure reproducibility and clinical interpretability. In many papers, the pre-processing and detection methods are not described in sufficient detail that would enable reproducing the results. Out of the 187 studies assessed for eligibility, 49.7% did not have a method for CCEP detection explicitly stated (n = 93). Also, 9% relied on visual identification (n = 17), which suffers from the bias introduced by the reviewer. There is also considerable variation in methods, even within the same analysis categories. Various metrics are used to measure significance through different statistical tests. Variation in amplitude thresholds for peak detection methods is evident with differing normalization techniques. The results section provides detailed description of the various detection methods. To provide an in-depth evaluation of each method, we discuss the strengths and limitations of each category.

4.1 Strengths and limitations of CCEP detection approaches

Although visual identification was not categorized separately in this review, as our emphasis is on rigorous and reproducible methods, we acknowledge its ease of use in low-resource settings. However, the subjectivity and inter-rater variability inherent in visual analysis limit its reliability and generalizability (van den Boom et al., 2024).

The threshold or peak-based detection methods are widely used due to their simplicity and ease. However, these methods rely on assumptions about the waveform of the CCEP, ignoring the vast heterogeneity in CCEPs, especially in sEEG data (Prime et al., 2018). There is variability in the chosen thresholds and low ones may disregard amplitude variations that are evident in CCEPs leading to an increase in false negatives (Frauscher et al., 2018). Statistical approaches improve upon thresholding by offering a more rigorous method to evaluate post-stimulation deviation from baseline. Techniques such as permutation testing, t-tests, and cluster-based inference increase robustness to noise and reduce subjective bias. Nonetheless, statistical tests may still assume pre-defined CCEP shapes and are sensitive to design choices like test type (parametric vs. non-parametric) and correction for multiple comparisons. Moreover, some assumptions required for these tests like stationarity, normality and linearity in data may not hold for complex neural data (Sugimoto et al., 1977). Frequency-based methods provide valuable insight into the spectral dynamics of CCEP responses, including phase locking and power changes across frequencies. They can reveal neural responses not easily observable in the time domain. However, these methods are sensitive to pre-processing choices, lack standardization, and have introduced conflicting findings when considered in studies extracting spectral biomarkers of epileptogenic zones (Qiang et al., 2025). Data-driven methods like CRP represent a shift toward a more standardized evaluation of CCEPs, since it learns from the data directly without assuming response morphology, accommodating CCEP heterogeneity (Miller et al., 2023). Ultimately, the detection method affects what is deemed significant and thus affects clinical decisions such as surgical planning or language mapping or EZ zone localization.

4.2 Recording montage considerations

One important factor is the effect of the montage on the processing of CCEP data, since it can introduce bias from the stimulated channels and affect the noise level across electrodes. A recent study compared various montages used to re-reference CCEP data recorded using sEEG to examine which montage better localizes electrodes in gray matter (Dickey et al., 2022). The RMS deviation from baseline was obtained after applying either the common average, bipolar, Laplacian montage, or referential montage using a subgaleal electrode placed near the vertex of the head. The Laplacian and bipolar montages were better at localizing CCEPs in gray matter than the CAR or referential montage as measured by the area under the curve for a receiver operating characteristic (Dickey et al., 2022). Another study including 14 patients with sEEG demonstrated that bipolar and Laplacian montages were able to effectively reduce SPES-related signal deflections at extracortical locations, including outside the brain (Mitsuhashi et al., 2020). The bipolar and Laplacian montages help preserve focal signal features and attenuate more distributed signals that are visible in neighboring electrodes like high-amplitude CCEP deflections (Arnulfo et al., 2015). Other montages that have been used are the common median or the adjusted CAR. If the distributed signals that arise from CCEPs are of interest, local re-referencing methods may not be the optimal choice because they may attenuate the signal, distort temporal profiles and lead to phase reversals (Arnulfo et al., 2015; Huang et al., 2024). Thus, the choice of re-referencing montage should be motivated by the signal features of interest, since different montages influence which signals features are preserved.

Some papers employed an adjusted version of the CAR that approximates the common noise while also minimizing the introduced bias from responsive channels by only including 20% of channels with lowest variance or the mean of unstimulated channels. A recent study introduced CAR by least anticorrelation “CARLA” (Huang et al., 2024), an algorithm that first ranks channels based on increasing covariance across trials, where lower covariance channels are considered less responsive. Channels were then iteratively added to the common average starting with the ones with the lowest covariance, and an anti-correlation statistic was calculated at each step between the common average formed by this subset and the re-referenced signals of the remaining channels. The algorithm sought to optimize the selection of the channels to be included in the common average by minimizing the anti-correlation.

4.3 Morphological diversity and signal interpretation

We found that the most common detection methods rely on assumptions about the morphology of the CCEP responses. This can make the analysis process simple and straightforward, but it disregards the variability that is evident in the evoked responses. These methods assume a single morphology for the CCEP response where a sharp deflection called the “N1” peak is expected around 10–30 ms after the stimulus onset and is sometimes followed by a second slower deflection “N2” peak occurring around 80–250 ms from stimulus (Keller et al., 2014). These deflections vary in latency and polarity and are mostly a characteristic of responses recorded using ECoG (Prime et al., 2020), which represent 31.5% of the included studies in this review.

Additionally, we identified differences in the quantification of CCEPs recorded from ECoG compared to sEEG. When using ECoG, a certain morphology, including N1 and N2 peaks, is expected since pyramidal cells are oriented consistently toward the electrode (Prime et al., 2018). However, in sEEG recordings, the depth electrodes sample various anatomical structures and can be located in gray matter, white matter and sometimes cerebrospinal fluid (Prime et al., 2020). This variation in the brain regions sampled and orientation of the cortical layers relative to the electrode likely leads to variation in the observed CCEPs. Indeed, signal amplitude variations have been reported between different anatomical regions during sEEG recordings, along with variations in spectral density distributions (Frauscher et al., 2018). Many studies reported the presence of a wide variety of morphologies and polarities in CCEPs recorded (Prime et al., 2018; Enatsu et al., 2015; Valencia et al., 2023; Miller et al., 2021, 2023; Shine et al., 2017). CCEPs have been used to characterize brain connectivity by quantifying the responses using the strength of the N1 peak or its latency, since it has been tied to direct pathways. However, this approach disregards other response morphologies that are due to indirect propagation or recurrent cortico-subcortico-cortical pathways (Huang et al., 2023; Veit et al., 2021). Thus, it is important to not categorize CCEPs into one expected morphology, since pre-defined forms would fail to capture the significance of various responses and their morphology even after various types of manipulations like temporal scaling or sign-flips (Miller et al., 2023). CCEP recordings from sEEG have been reported to be polymorphic including positive and negative deflections with around three peaks described, the first occurring 10–30 ms after stimulation, the second after 80–250 ms and the third after that (Feys et al., 2024b). The variability in CCEP morphology has also been linked to EZ localization biomarkers.

4.4 Toward standardization and robust quantification

To properly validate the detected CCEPs, it is recommended to utilize a proper referencing montage that minimizes the noise level and bias from stimulated channels, while also considering the anatomical location of the recording electrode primarily in gray matter, as opposed to white matter or CSF. Ensuring across trial reproducibility and dependence on stimulation parameters like amplitude, which has been used to detect significant responses, are all methods that can help decrease misclassification of CCEPs. The heterogeneity observed in the detection methods highlights a clear need for a standardized way to evaluate CCEP responses that is quantitative, automated, reproducible, and can accurately convey the connectivity from the recorded data. One step closer to this standardized method may be the data-driven method described that provides a structured approach to parametrizing CCEP data by inferring the canonical shape from the trials and defining metrics like the response duration that can be used for comparisons across brain regions or patients (Miller et al., 2023). CRP also introduced a method for automated rejection of artifactual trials based on the cross-projection magnitudes. Moreover, CRP can adapt to various modalities (EEG, fMRI, etc.) making the analysis of data from multimodal studies simpler through the integration of electrophysiological and imaging modalities. This consistent response quantification allows for robust connectivity analysis that can be used to understand connectivity patterns in focal epilepsy. It also offers a way to study brain state transitions during anesthesia by identifying response shifts (Yamao et al., 2021). CRP may also be used in analyzing event-related potentials by identifying these structures and their significances in an automated manner with applications in the study of cognition, emotions, sensory functions and brain injuries (Mazzini, 2004).

5 Limitations

This study has several potential limitations. Diseases other than epilepsy (93.1% of included studies) are underrepresented, which may introduce differences in CCEP analysis when considering other diseases or objectives. The absence of CCEP studies in healthy individuals, understandably constrained by ethical considerations, also restricts the ability to define normative baselines for evoked responses. Variability in pre-processing like re-referencing, normalizing, filtering and epoching across studies may also affect how the CCEPs are quantified (Levinson et al., 2024). Different referencing montages (bipolar, common average, adjusted CAR, common median, Laplacian) are used across studies, each influencing signal morphology and noise level. This introduces a confounding factor, especially since the choice of montage affects the detectability of CCEP components and may obscure or distort true responses. Moreover, some studies do not report their montage explicitly, impeding efforts to normalize across datasets. A substantial portion of detection methods assumes the presence of an N1–N2 waveform with fixed latency and morphology. However, this may be appropriate only for certain cortical regions and electrode types. In sEEG recordings, anatomical variability and electrode orientation result in more heterogeneous waveforms, making rigid assumptions about shape or latency misleading. As a result, valuable responses with atypical morphology or longer latencies may be missed or misclassified. Only 13 studies made their raw data available, and only three others shared their code. This limits the ability to benchmark and validate CCEP detection techniques. Furthermore, variability in CCEP amplitude and waveform morphology due to stimulation parameters, anatomical location, and inter-electrode distance is not consistently controlled for or reported across studies. Evidence shows that stimulation levels below 5.5 mA yield more variable CCEPs, while higher intensities produce more stable responses (Kundu et al., 2020). Without standardized current amplitudes, comparisons across studies may reflect stimulation-related variability rather than true neurophysiological differences. Similarly, the distance between stimulation and recording sites inversely affects CCEP amplitude, but many studies do not report or correct for this spatial factor. Regional variability is also evident where hippocampal responses tend to be larger than those from neocortex or white matter (Kundu et al., 2020). Several studies in this review did not explicitly state their method of CCEP detection, illustrating again the need for transparent and reproducible methods for CCEP analysis.

6 Conclusion

Cortico-cortical evoked potentials are a valuable tool for evaluating the dynamic, effective connectivity between brain regions. But there is a high degree of heterogeneity in the methods employed in CCEP detection, making comparisons across studies and reproducibility difficult. Of the slightly less than half of studies that report their methods, most rely on peak detection or threshold-based detection, with some opting for statistical tests or frequency-based methods, all of which are sensitive to referencing montages and stimulation artifacts. Data-driven methods provide a promising avenue for more robust CCEP detection, though have not been widely adopted. By cataloging these diverse methodologies, we aim to foster greater standardization in CCEP research, enhancing comparability, reproducibility, and rigor, ultimately advancing our understanding of functional and structural brain connectivity. Future work in CCEP detection should incorporate data-driven methods for CCEP detection, adopt thorough reporting checklists during methodology and encourage greater contributions toward open-data processing for improved clarity in the field.

Data availability statement

The data analyzed in this study is subject to the following licenses/restrictions: The dataset is not publicly available due to privacy restrictions. Requests to access these datasets should be directed to JR, anJvbHN0b25AYndoLmhhcnZhcmQuZWR1.

Author contributions

TA-S: Methodology, Data curation, Formal analysis, Writing – original draft, Visualization, Conceptualization, Validation, Software, Writing – review & editing, Resources, Investigation. AW: Methodology, Formal analysis, Writing – review & editing, Writing – original draft, Data curation. MW: Data curation, Writing – original draft, Writing – review & editing. AEW: Supervision, Conceptualization, Writing – review & editing, Writing – original draft, Validation. JR: Writing – review & editing, Writing – original draft, Supervision, Validation, Conceptualization.

Funding

The author(s) declare that no financial support was received for the research and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The authors declare that Generative AI was used in the creation of this manuscript. During the preparation of this work the author(s) used ChatGPT in order to improve language and readability. After using this tool/service, the author(s) reviewed and edited the content as needed and take(s) full responsibility for the content of the publication.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fnhum.2025.1636115/full#supplementary-material

References

Alarcón, G., Jiménez-Jiménez, D., Valentín, A., and Martín-López, D. (2018). Characterizing EEG cortical dynamics and connectivity with responses to single pulse electrical stimulation (SPES). Int. J. Neural Syst. 28:1750057. doi: 10.1142/S0129065717500575

Crossref Full Text | Google Scholar

Almashaikhi, T., Rheims, S., Jung, J., Ostrowsky-Coste, K., Montavont, A., De Bellescize, J., et al. (2014). Functional connectivity of insular efferences. Hum. Brain Mapp. 35, 5279–5294. doi: 10.1002/hbm.22549

Crossref Full Text | Google Scholar

Arbune, A., Popa, I., Mindruta, I., Beniczky, S., Donos, C., Daneasa, A., et al. (2020). Sleep modulates effective connectivity: A study using intracranial stimulation and recording. Clin. Neurophysiol. 131, 529–541. doi: 10.1016/j.clinph.2019.09.010

PubMed Abstract | Crossref Full Text | Google Scholar

Arnulfo, G., Hirvonen, J., Nobili, L., Palva, S., and Palva, J. M. (2015). Phase and amplitude correlations in resting-state activity in human stereotactical EEG recordings. Neuroimage 112, 114–127. doi: 10.1016/j.neuroimage.2015.02.031

Crossref Full Text | Google Scholar

Boido, D., Kapetis, D., Gnatkovsky, V., Pastori, C., Galbardi, B., Sartori, I., et al. (2014). Stimulus-evoked potentials contribute to map the epileptogenic zone during stereo-EEG presurgical monitoring. Hum. Brain Mapp. 35, 4267–4281. doi: 10.1002/hbm.22516

Crossref Full Text | Google Scholar

Boulogne, S., Andre-Obadia, N., Kimiskidis, V. K., Ryvlin, P., and Rheims, S. (2016). Cortico-cortical and motor evoked potentials to single and paired-pulse stimuli: An exploratory transcranial magnetic and intracranial electric brain stimulation study. Hum. Brain Mapp. 37, 3767–3778. doi: 10.1002/hbm.23274

PubMed Abstract | Crossref Full Text | Google Scholar

Boulogne, S., Pizzo, F., Chatard, B., Roehri, N., Catenoix, H., Ostrowsky-Coste, K., et al. (2022). Functional connectivity and epileptogenicity of nodular heterotopias: A single-pulse stimulation study. Epilepsia 63, 961–973. doi: 10.1111/epi.17168

PubMed Abstract | Crossref Full Text | Google Scholar

Brugge, J. F., Volkov, I. O., Garell, P. C., Reale, R. A., and Howard, M. A. (2003). Functional connections between auditory cortex on Heschl’s gyrus and on the lateral superior temporal gyrus in humans. J. Neurophysiol. 90, 3750–3763. doi: 10.1152/jn.00500.2003

PubMed Abstract | Crossref Full Text | Google Scholar

Campbell, J. M., Davis, T. S., Anderson, D. N., Arain, A., Inman, C. S., Smith, E. H., et al. (2023). Subsets of cortico-cortical evoked potentials propagate as traveling waves. bioRxiv [Preprint]. doi: 10.1523/JNEUROSCI.1504-24.2025

Crossref Full Text | Google Scholar

Cornblath, E., Lucas, A., Armstrong, C., Greenblatt, A., Stein, J., Hadar, P., et al. (2023). Quantifying trial-by-trial variability during cortico-cortical evoked potential mapping of epileptogenic tissue. Epilepsia 64, 1021–1034. doi: 10.1111/epi.17528

PubMed Abstract | Crossref Full Text | Google Scholar

Crocker, B., Ostrowski, L., Williams, Z., Dougherty, D., Eskandar, E., Widge, A., et al. (2021). Local and distant responses to single pulse electrical stimulation reflect different forms of connectivity. Neuroimage 237:118094. doi: 10.1016/j.neuroimage.2021.118094

PubMed Abstract | Crossref Full Text | Google Scholar

Crowther, L., Brunner, P., Kapeller, C., Guger, C., Kamada, K., Bunch, M., et al. (2019). A quantitative method for evaluating cortical responses to electrical stimulation. J. Neurosci. Methods 311, 67–75. doi: 10.1016/j.jneumeth.2018.09.034

Crossref Full Text | Google Scholar

Davis, T. S., Rolston, J. D., Bollo, R. J., and House, P. A. (2018). Delayed high-frequency suppression after automated single-pulse electrical stimulation identifies the seizure onset zone in patients with refractory epilepsy. Clin. Neurophysiol. 129, 2466–2474. doi: 10.1016/j.clinph.2018.06.021

Crossref Full Text | Google Scholar

Dickey, A. S., Alwaki, A., Kheder, A., Willie, J. T., Drane, D. L., and Pedersen, N. P. (2022). The referential montage inadequately localizes corticocortical evoked potentials in stereoelectroencephalography. J. Clin. Neurophysiol. 39, 412–418. doi: 10.1097/WNP.0000000000000792

PubMed Abstract | Crossref Full Text | Google Scholar

Donos, C., Mălîia, M., Mîndruţă, I., Popa, I., Ene, M., Bălănescu, B., et al. (2016). A connectomics approach combining structural and effective connectivity assessed by intracranial electrical stimulation. Neuroimage 132, 344–358. doi: 10.1016/j.neuroimage.2016.02.054

PubMed Abstract | Crossref Full Text | Google Scholar

Dou, Y., Xia, J., Fu, M., Cai, Y., Meng, X., and Zhan, Y. (2023). Identification of epileptic networks with graph convolutional network incorporating oscillatory activities and evoked synaptic responses. Neuroimage 284:120439. doi: 10.1016/j.neuroimage.2023.120439

PubMed Abstract | Crossref Full Text | Google Scholar

Enatsu, R., Gonzalez-Martinez, J., Bulacio, J., Kubota, Y., Mosher, J., Burgess, R., et al. (2015). Connections of the limbic network: A cortico-cortical evoked potentials study. Cortex 62, 20–33. doi: 10.1016/j.cortex.2014.06.018

Crossref Full Text | Google Scholar

Enatsu, R., Jin, K., Elwan, S., Kubota, Y., Piao, Z., O’Connor, T., et al. (2012). Correlations between ictal propagation and response to electrical cortical stimulation: A cortico-cortical evoked potential study. Epilepsy Res. 101, 76–87. doi: 10.1016/j.eplepsyres.2012.03.004

PubMed Abstract | Crossref Full Text | Google Scholar

Enatsu, R., Matsumoto, R., Piao, Z., O’Connor, T., Horning, K., Burgess, R., et al. (2013b). Cortical negative motor network in comparison with sensorimotor network: A cortico-cortical evoked potential study. Cortex 49, 2080–2096. doi: 10.1016/j.cortex.2012.08.026

PubMed Abstract | Crossref Full Text | Google Scholar

Enatsu, R., Kubota, Y., Kakisaka, Y., Bulacio, J., Piao, Z., O’Connor, T., et al. (2013a). Reorganization of posterior language area in temporal lobe epilepsy: A cortico-cortical evoked potential study. Epilepsy Res. 103, 73–82. doi: 10.1016/j.eplepsyres.2012.07.008

Crossref Full Text | Google Scholar

Feys, O., Schuind, S., Sculier, C., Rikir, E., Legros, B., Gaspard, N., et al. (2025). Dynamics of magnetic cortico-cortical responses evoked by single-pulse electrical stimulation. Epilepsia 66, 503–517. doi: 10.1111/epi.18183

PubMed Abstract | Crossref Full Text | Google Scholar

Feys, O., Wens, V., Rovai, A., Schuind, S., Rikir, E., Legros, B., et al. (2024a). Delayed effective connectivity characterizes the epileptogenic zone during stereo-EEG. Clin. Neurophysiol. 158, 59–68. doi: 10.1016/j.clinph.2023.12.013

Crossref Full Text | Google Scholar

Feys, O., Wens, V., Schuind, S., Rikir, E., Legros, B., De Tiège, X., et al. (2024b). Variability of cortico-cortical evoked potentials in the epileptogenic zone is related to seizure occurrence. Ann. Clin. Transl. Neurol. 11, 2645–2656. doi: 10.1002/acn3.52179

PubMed Abstract | Crossref Full Text | Google Scholar

Filipiak, P., Almairac, F., Papadopoulo, T., Fontaine, D., Mondot, L., Chanalet, S., et al. (2021). Towards linking diffusion MRI based macro- and microstructure measures with cortico-cortical transmission in brain tumor patients. Neuroimage 226:117567. doi: 10.1016/j.neuroimage.2020.117567

PubMed Abstract | Crossref Full Text | Google Scholar

Frauscher, B., von Ellenrieder, N., Zelmann, R., Doležalová, I., Minotti, L., Olivier, A., et al. (2018). Atlas of the normal intracranial electroencephalogram: Neurophysiological awake activity in different cortical areas. Brain 141, 1130–1144. doi: 10.1093/brain/awy035

PubMed Abstract | Crossref Full Text | Google Scholar

Hays, M. A., Smith, R. J., Haridas, B., Coogan, C., Crone, N. E., and Kang, J. Y. (2021b). Effects of stimulation intensity on intracranial cortico-cortical evoked potentials: A titration study. Clin. Neurophysiol. 132, 2766–2777. doi: 10.1016/j.clinph.2021.08.008

PubMed Abstract | Crossref Full Text | Google Scholar

Hays, M. A., Coogan, C., Crone, N. E., and Kang, J. Y. (2021a). Graph theoretical analysis of evoked potentials shows network influence of epileptogenic mesial temporal region. Hum. Brain Mapp. 42, 4173–4186. doi: 10.1002/hbm.25418

PubMed Abstract | Crossref Full Text | Google Scholar

Hays, M., Kamali, G., Koubeissi, M., Sarma, S., Crone, N., Smith, R., et al. (2023a). Towards optimizing single pulse electrical stimulation: High current intensity, short pulse width stimulation most effectively elicits evoked potentials. Brain Stimul. 16, 772–782. doi: 10.1016/j.brs.2023.04.023

PubMed Abstract | Crossref Full Text | Google Scholar

Hays, M., Smith, R., Wang, Y., Coogan, C., Sarma, S., Crone, N., et al. (2023b). Cortico-cortical evoked potentials in response to varying stimulation intensity improves seizure localization. Clin. Neurophysiol. 145, 119–128. doi: 10.1016/j.clinph.2022.08.024

PubMed Abstract | Crossref Full Text | Google Scholar

Hebbink, J., Huiskamp, G., van Gils, S. A., Leijten, F. S., and Meijer, H. G. (2020). Pathological responses to single-pulse electrical stimuli in epilepsy: The role of feedforward inhibition. Eur. J. Neurosci. 51, 1122–1136. doi: 10.1111/ejn.14562

PubMed Abstract | Crossref Full Text | Google Scholar

Hebbink, J., van Blooijs, D., Huiskamp, G., Leijten, F. S., van Gils, S. A., and Meijer, H. G. (2019). A comparison of evoked and non-evoked functional networks. Brain Topogr. 32, 405–417. doi: 10.1007/s10548-018-0692-1

Crossref Full Text | Google Scholar

Howard, M. A., Volkov, I. O., Mirsky, R., Garell, P. C., Noh, M. D., Granner, M., et al. (2000). Auditory cortex on the human posterior superior temporal gyrus. J. Comp. Neurol. 416, 79–92. doi: 10.1002/(sici)1096-9861(20000103)416:1<79::aid-cne6<3.0.co;2-2

Crossref Full Text | Google Scholar

Huang, H., Gregg, N., Ojeda Valencia, G., Brinkmann, B., Lundstrom, B., Worrell, G., et al. (2023). Electrical Stimulation of temporal and limbic circuitry produces distinct responses in human ventral temporal cortex. J. Neurosci. 43, 4434–4447. doi: 10.1523/JNEUROSCI.1325-22.2023

PubMed Abstract | Crossref Full Text | Google Scholar

Huang, H., Ojeda Valencia, G., Gregg, N., Osman, G., Montoya, M., Worrell, G., et al. (2024). CARLA: Adjusted common average referencing for cortico-cortical evoked potential data. J. Neurosci. Methods 407:110153. doi: 10.1016/j.jneumeth.2024.110153

Crossref Full Text | Google Scholar

Huang, Y., Hajnal, B., Entz, L., Fabó, D., Herrero, J., Mehta, A., et al. (2019). Intracortical dynamics underlying repetitive stimulation predicts changes in network connectivity. J. Neurosci. 39, 6122–6135. doi: 10.1523/JNEUROSCI.0535-19.2019

PubMed Abstract | Crossref Full Text | Google Scholar

Jedynak, M., Boyer, A., Chanteloup-Forêt, B., Bhattacharjee, M., Saubat, C., Tadel, F., et al. (2023). Variability of single pulse electrical stimulation responses recorded with intracranial electroencephalography in epileptic patients. Brain Topogr. 36, 119–127. doi: 10.1007/s10548-022-00928-7

PubMed Abstract | Crossref Full Text | Google Scholar

Jedynak, M., Boyer, A., Mercier, M., Chanteloup-Forêt, B., Bhattacharjee, M., Kahane, P., et al. (2024). SEEG electrode shaft affects amplitude and latency of potentials evoked with single pulse electrical stimulation. J. Neurosci. Methods 403:110035. doi: 10.1016/j.jneumeth.2023.110035

Crossref Full Text | Google Scholar

Kamada, K., Kapeller, C., Takeuchi, F., Gruenwald, J., and Guger, C. (2020). Tailor-made surgery based on functional networks for intractable epilepsy. Front. Neurol. 11:73. doi: 10.3389/fneur.2020.00073

PubMed Abstract | Crossref Full Text | Google Scholar

Kamada, K., Ogawa, H., Kapeller, C., Prueckl, R., Hiroshima, S., Tamura, Y., et al. (2017). Disconnection of the pathological connectome for multifocal epilepsy surgery. J. Neurosurg. 129, 1182–1194. doi: 10.3171/2017.6.JNS17452

PubMed Abstract | Crossref Full Text | Google Scholar

Kamali, G., Smith, R., Hays, M., Coogan, C., Crone, N., Kang, J., et al. (2020). Transfer function models for the localization of seizure onset zone from cortico-cortical evoked potentials. Front. Neurol. 11:579961. doi: 10.3389/fneur.2020.579961

Crossref Full Text | Google Scholar

Keller, C. J., Honey, C. J., Mégevand, P., Entz, L., Ulbert, I., and Mehta, A. D. (2014). Mapping human brain networks with cortico-cortical evoked potentials. Philos. Trans. R. Soc. B Biol. Sci. 369:20130528. doi: 10.1098/rstb.2013.0528

Crossref Full Text | Google Scholar

Keller, C., Bickel, S., Entz, L., Ulbert, I., Milham, M., Kelly, C., et al. (2011). Intrinsic functional architecture predicts electrically evoked responses in the human brain. Proc. Natl. Acad. Sci. U.S.A. 108, 10308–10313. doi: 10.1073/pnas.1019750108

PubMed Abstract | Crossref Full Text | Google Scholar

Krieg, J., Koessler, L., Jonas, J., Colnat-Coulbois, S., Vignal, J., Bénar, C., et al. (2017). Discrimination of a medial functional module within the temporal lobe using an effective connectivity model: A CCEP study. Neuroimage 161, 219–231. doi: 10.1016/j.neuroimage.2017.07.061

PubMed Abstract | Crossref Full Text | Google Scholar

Kundu, B., Davis, T., Philip, B., Smith, E., Arain, A., Peters, A., et al. (2020). A systematic exploration of parameters affecting evoked intracranial potentials in patients with epilepsy. Brain Stimul. 13, 1232–1244. doi: 10.1016/j.brs.2020.06.002

Crossref Full Text | Google Scholar

Lemaréchal, J., Jedynak, M., Trebaul, L., Boyer, A., Tadel, F., Bhattacharjee, M., et al. (2022). A brain atlas of axonal and synaptic delays based on modelling of cortico-cortical evoked potentials. Brain 145, 1653–1667. doi: 10.1093/brain/awab362

PubMed Abstract | Crossref Full Text | Google Scholar

Levinson, L., Sun, S., Paschall, C., Perks, K., Weaver, K., Perlmutter, S., et al. (2024). Data processing techniques impact quantification of cortico-cortical evoked potentials. J. Neurosci. Methods 408:110130. doi: 10.1016/j.jneumeth.2024.110130

Crossref Full Text | Google Scholar

Liberati, A., Altman, D., Tetzlaff, J., Mulrow, C., Gotzsche, P. C., Ioannidis, J. P., et al. (2009). The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: Explanation and elaboration. J. Clin. Epidemiol. 62, e1–e34. doi: 10.1016/j.jclinepi.2009.06.006

PubMed Abstract | Crossref Full Text | Google Scholar

Luo, L., Chen, G., Li, S., Wang, J., Wang, Q., and Fang, F. (2022). Distinct roles of theta and gamma rhythms in inter-areal interaction in human visual cortex revealed by cortico-cortical evoked potentials. Brain Stimul. 15, 1048–1050. doi: 10.1016/j.brs.2022.07.056

PubMed Abstract | Crossref Full Text | Google Scholar

Mălîia, M., Donos, C., Barborica, A., Mindruta, I., Popa, I., Ene, M., et al. (2017). High frequency spectral changes induced by single-pulse electric stimulation: Comparison between physiologic and pathologic networks. Clin. Neurophysiol. 128, 1053–1060. doi: 10.1016/j.clinph.2016.12.016

PubMed Abstract | Crossref Full Text | Google Scholar

Mălîia, M., Donos, C., Barborica, A., Popa, I., Ciurea, J., Cinatti, S., et al. (2018). Functional mapping and effective connectivity of the human operculum. Cortex 109, 303–321. doi: 10.1016/j.cortex.2018.08.024

PubMed Abstract | Crossref Full Text | Google Scholar

Mariani, V., Sartori, I., Revay, M., Mai, R., Lizio, D., Berta, L., et al. (2021). Intraoperative corticocortical evoked potentials for language monitoring in epilepsy surgery. World Neurosurg. 151, e109–e121. doi: 10.1016/j.wneu.2021.03.141

PubMed Abstract | Crossref Full Text | Google Scholar

Matsumoto, R., Nair, D. R., Ikeda, A., Fumuro, T., Lapresto, E., Mikuni, N., et al. (2012). Parieto-frontal network in humans studied by cortico-cortical evoked potential. Hum. Brain Mapp. 33, 2856–2872. doi: 10.1002/hbm.21407

PubMed Abstract | Crossref Full Text | Google Scholar

Matsumoto, R., Nair, D. R., LaPresto, E., Bingaman, W., Shibasaki, H., and Lüders, H. O. (2007). Functional connectivity in human cortical motor system: A cortico-cortical evoked potential study. Brain 130, 181–197. doi: 10.1093/brain/awl257

PubMed Abstract | Crossref Full Text | Google Scholar

Matsumoto, R., Nair, D., LaPresto, E., Najm, I., Bingaman, W., Shibasaki, H., et al. (2004). Functional connectivity in the human language system: A cortico-cortical evoked potential study. Brain 127, 2316–2330. doi: 10.1093/brain/awh246

PubMed Abstract | Crossref Full Text | Google Scholar

Mazzini, L. (2004). Clinical applications of event-related potentials in brain injury. Phys. Med. Rehabil. Clin. 15, 163–175. doi: 10.1016/s1047-9651(03)00101-3

PubMed Abstract | Crossref Full Text | Google Scholar

Mégevand, P., Groppe, D., Bickel, S., Mercier, M., Goldfinger, M., Keller, C., et al. (2017). The hippocampus and amygdala are integrators of neocortical influence: A cortico-cortical evoked potential study. Brain Connect. 7, 648–660. doi: 10.1089/brain.2017.0527

PubMed Abstract | Crossref Full Text | Google Scholar

Miller, K. J., Müller, K. R., Valencia, G. O., Huang, H., Gregg, N. M., Worrell, G. A., et al. (2023). Canonical response parameterization: Quantifying the structure of responses to single-pulse intracranial electrical brain stimulation. PLoS Comput. Biol. 19:e1011105. doi: 10.1371/journal.pcbi.1011105

PubMed Abstract | Crossref Full Text | Google Scholar

Miller, K. J., Müller, K.-R., and Hermes, D. (2021). Basis profile curve identification to understand electrical stimulation effects in human brain networks. PLoS Comput. Biol. 17:e1008710. doi: 10.1371/journal.pcbi.1008710

PubMed Abstract | Crossref Full Text | Google Scholar

Mitsuhashi, T., Sonoda, M., Iwaki, H., Luat, A. F., Sood, S., and Asano, E. (2020). Effects of depth electrode montage and single-pulse electrical stimulation sites on neuronal responses and effective connectivity. Clin. Neurophysiol. 131, 2781–2792. doi: 10.1016/j.clinph.2020.09.01

Crossref Full Text | Google Scholar

Mouthaan, B. E., van ’t Klooster, M. A., Keizer, D., Hebbink, G. J., Leijten, F. S. S., Ferrier, C. H., et al. (2016). Single Pulse electrical stimulation to identify epileptogenic cortex: Clinical information obtained from early evoked responses. Clin. Neurophysiol. 127, 1088–1098. doi: 10.1016/j.clinph.2015.07.031

PubMed Abstract | Crossref Full Text | Google Scholar

Nakae, T., Matsumoto, R., Kunieda, T., Arakawa, Y., Kobayashi, K., Shimotake, A., et al. (2020). Connectivity gradient in the human left inferior frontal gyrus: Intraoperative cortico-cortical evoked potential study. Cereb. Cortex 30, 4633–4650. doi: 10.1093/cercor/bhaa065

PubMed Abstract | Crossref Full Text | Google Scholar

Novitskaya, Y., Dümpelmann, M., Vlachos, A., Reinacher, P. C., and Schulze-Bonhage, A. (2020). In vivo-assessment of the human temporal network: Evidence for asymmetrical effective connectivity. Neuroimage 214:116769. doi: 10.1016/j.neuroimage.2020.116769

PubMed Abstract | Crossref Full Text | Google Scholar

Oane, I., Barborica, A., Chetan, F., Donos, C., Maliia, M., Arbune, A., et al. (2020). Cingulate cortex function and multi-modal connectivity mapped using intracranial stimulation. Neuroimage 220:117059. doi: 10.1016/j.neuroimage.2020.117059

PubMed Abstract | Crossref Full Text | Google Scholar

Parmigiani, S., Mikulan, E., Russo, S., Sarasso, S., Zauli, F., Rubino, A., et al. (2022). Simultaneous stereo-EEG and high-density scalp EEG recordings to study the effects of intracerebral stimulation parameters. Brain Stimul. 15, 664–675. doi: 10.1016/j.brs.2022.04.007

PubMed Abstract | Crossref Full Text | Google Scholar

Parvizi, J., Braga, R., Kucyi, A., Veit, M., Pinheiro-Chagas, P., Perry, C., et al. (2021). Altered sense of self during seizures in the posteromedial cortex. Proc. Natl. Acad. Sci. U.S.A. 118:e2100522118. doi: 10.1073/pnas.2100522118

PubMed Abstract | Crossref Full Text | Google Scholar

Prime, D., Rowlands, D., O’Keefe, S., and Dionisio, S. (2018). Considerations in performing and analyzing the responses of cortico-cortical evoked potentials in stereo-EEG. Epilepsia 59, 16–26. doi: 10.1111/epi.13939

Crossref Full Text | Google Scholar

Prime, D., Woolfe, M., Rowlands, D., O’Keefe, S., and Dionisio, S. (2020). Comparing connectivity metrics in cortico-cortical evoked potentials using synthetic cortical response patterns. J. Neurosci. Methods 334:108559. doi: 10.1016/j.jneumeth.2019.108559

PubMed Abstract | Crossref Full Text | Google Scholar

Qiang, Z., Norris, J., Cooray, G., Rosch, R., Miller, K., Hermes, D., et al. (2025). Cortico-cortical evoked potentials: Analytical techniques and emerging paradigms for epileptogenic zone localization. Epilepsia doi: 10.1111/epi.18467 [Epub ahead of print].

Crossref Full Text | Google Scholar

Rolston, J. D., Gross, R. E., and Potter, S. M. (2009). Common median referencing for improved action potential detection with multielectrode arrays. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2009, 1604–1607. doi: 10.1109/IEMBS.2009.5333230

Crossref Full Text | Google Scholar

Russo, S., Mikulan, E., Zauli, F., Sartori, I., Solbiati, M., Furregoni, G., et al. (2023). Neocortical and medial temporal seizures have distinct impacts on brain responsiveness. Epilepsia 64, e118–e126. doi: 10.1111/epi.17580

PubMed Abstract | Crossref Full Text | Google Scholar

Russo, S., Pigorini, A., Mikulan, E., Sarasso, S., Rubino, A., Zauli, F., et al. (2021). Focal lesions induce large-scale percolation of sleep-like intracerebral activity in awake humans. Neuroimage 234:117964. doi: 10.1016/j.neuroimage.2021.117964

PubMed Abstract | Crossref Full Text | Google Scholar

Saito, T., Tamura, M., Muragaki, Y., Maruyama, T., Kubota, Y., Fukuchi, S., et al. (2014). Intraoperative cortico-cortical evoked potentials for the evaluation of language function during brain tumor resection: Initial experience with 13 cases. J. Neurosurg. 121, 827–838. doi: 10.3171/2014.4.JNS131195

PubMed Abstract | Crossref Full Text | Google Scholar

Schalk, G., Kubánek, J., Miller, K., Anderson, N., Leuthardt, E., Ojemann, J., et al. (2007). Decoding two-dimensional movement trajectories using electrocorticographic signals in humans. J. Neural Eng. 4, 264–275. doi: 10.1088/1741-2560/4/3/012

PubMed Abstract | Crossref Full Text | Google Scholar

Seidel, K., Wermelinger, J., Alvarez-Abut, P., Deletis, V., Raabe, A., Zhang, D., et al. (2024). Cortico-cortical evoked potentials of language tracts in minimally invasive glioma surgery guided by Penfield stimulation. Clin. Neurophysiol. 161, 256–267. doi: 10.1016/j.clinph.2023.12.136

PubMed Abstract | Crossref Full Text | Google Scholar

Shahabi, H., Taylor, K., Hirfanoglu, T., Koneru, S., Bingaman, W., Kobayashi, K., et al. (2021). Effective connectivity differs between focal cortical dysplasia types I and II. Epilepsia 62, 2753–2765. doi: 10.1111/epi.17064

PubMed Abstract | Crossref Full Text | Google Scholar

Shine, J., Kucyi, A., Foster, B., Bickel, S., Wang, D., Liu, H., et al. (2017). Distinct patterns of temporal and directional connectivity among intrinsic networks in the human brain. J. Neurosci. 37, 9667–9674. doi: 10.1523/JNEUROSCI.1574-17.2017

PubMed Abstract | Crossref Full Text | Google Scholar

Silverstein, B. H., Asano, E., Sugiura, A., Sonoda, M., Lee, M.-H., and Jeong, J.-W. (2020). Dynamic tractography: Integrating cortico-cortical evoked potentials and diffusion imaging. Neuroimage 215:116763. doi: 10.1016/j.neuroimage.2020.116763

PubMed Abstract | Crossref Full Text | Google Scholar

Sugimoto, H., Ishii, N., Iwata, A., and Suzumura, N. (1977). Stationarity and normality test for biomedical data. Comp. Programs Biomed. 7, 293–304. doi: 10.1016/0010-468x(77)90045-9

PubMed Abstract | Crossref Full Text | Google Scholar

Sun, K., Wang, H., Bai, Y., Zhou, W., and Wang, L. (2021). MRIES: A matlab toolbox for mapping the responses to intracranial electrical stimulation. Front. Neurosci. 15:652841. doi: 10.3389/fnins.2021.652841

PubMed Abstract | Crossref Full Text | Google Scholar

Takeyama, H., Matsumoto, R., Usami, K., Nakae, T., Kobayashi, K., Shimotake, A., et al. (2019). Human entorhinal cortex electrical stimulation evoked short-latency potentials in the broad neocortical regions: Evidence from cortico-cortical evoked potential recordings. Brain Behav. 9:e01366. doi: 10.1002/brb3.1366

PubMed Abstract | Crossref Full Text | Google Scholar

Tóth, E., Bokodi, V., Somogyvári, Z., Maglóczky, Z., Wittner, L., Ulbert, I., et al. (2021). Laminar distribution of electrically evoked hippocampal short latency ripple activity highlights the importance of the subiculum in vivo in human epilepsy, an intraoperative study. Epilepsy Res. 169:106509. doi: 10.1016/j.eplepsyres.2020.106509

PubMed Abstract | Crossref Full Text | Google Scholar

Trebaul, L., Deman, P., Tuyisenge, V., Jedynak, M., Hugues, E., Rudrauf, D., et al. (2018). Probabilistic functional tractography of the human cortex revisited. Neuroimage 181, 414–429. doi: 10.1016/j.neuroimage.2018.07.039

PubMed Abstract | Crossref Full Text | Google Scholar

Usami, K., Milsap, G., Korzeniewska, A., Collard, M., Wang, Y., Lesser, R., et al. (2019b). Cortical responses to input from distant areas are modulated by local spontaneous alpha/beta oscillations. Cereb. Cortex 29, 777–787. doi: 10.1093/cercor/bhx361

PubMed Abstract | Crossref Full Text | Google Scholar

Usami, K., Korzeniewska, A., Matsumoto, R., Kobayashi, K., Hitomi, T., Matsuhashi, M., et al. (2019a). The neural tides of sleep and consciousness revealed by single-pulse electrical brain stimulation. Sleep 42:zsz050. doi: 10.1093/sleep/zsz050

PubMed Abstract | Crossref Full Text | Google Scholar

Usami, K., Matsumoto, R., Kobayashi, K., Hitomi, T., Matsuhashi, M., Shimotake, A., et al. (2017). Phasic REM transiently approaches wakefulness in the human cortex-a single-pulse electrical stimulation study. Sleep 40:zsx077. doi: 10.1093/sleep/zsx077

PubMed Abstract | Crossref Full Text | Google Scholar

Usami, K., Matsumoto, R., Kobayashi, K., Hitomi, T., Shimotake, A., Kikuchi, T., et al. (2015). Sleep modulates cortical connectivity and excitability in humans: Direct evidence from neural activity induced by single-pulse electrical stimulation. Hum. Brain Mapp. 36, 4714–4729. doi: 10.1002/hbm.22948

PubMed Abstract | Crossref Full Text | Google Scholar

Valencia, G. O., Gregg, N. M., Huang, H., Lundstrom, B. N., Brinkmann, B. H., Pal Attia, T., et al. (2023). Signatures of electrical stimulation driven network interactions in the human limbic system. J. Neurosci. 43, 6697–6711. doi: 10.1523/JNEUROSCI.2201-22.2023

PubMed Abstract | Crossref Full Text | Google Scholar

van Blooijs, D., Blok, S., Huiskamp, G. J., Van Eijsden, P., Meijer, H. G., and Leijten, F. (2024). The effect of propofol on effective brain networks. Clin. Neurophysiol. 161, 222–230. doi: 10.1016/j.clinph.2024.01.012

PubMed Abstract | Crossref Full Text | Google Scholar

van Blooijs, D., Leijten, F. S., van Rijen, P. C., Meijer, H. G., and Huiskamp, G. J. (2018). Evoked directional network characteristics of epileptogenic tissue derived from single pulse electrical stimulation. Hum. Brain Mapp. 39, 4611–4622. doi: 10.1002/hbm.24309

PubMed Abstract | Crossref Full Text | Google Scholar

van Blooijs, D., van den Boom, M., van der Aar, J., Huiskamp, G., Castegnaro, G., Demuru, M., et al. (2023a). Developmental trajectory of transmission speed in the human brain. Nat. Neurosci. 26, 537–541. doi: 10.1038/s41593-023-01272-0

PubMed Abstract | Crossref Full Text | Google Scholar

van Blooijs, D., van der Stoel, M., Huiskamp, G., Demuru, M., Ramsey, N., and Leijten, F. (2023b). Local cortical network stimulation as a concept for focal epilepsy treatment. medRxiv [Preprint]. doi: 10.1101/2023.10.30.23297463

Crossref Full Text | Google Scholar

van den Boom, M. A., Gregg, N. M., Valencia, G. O., Lundstrom, B. N., Miller, K. J., van Blooijs, D., et al. (2024). ER-detect: A pipeline for robust detection of early evoked responses in BIDS-iEEG electrical stimulation data. bioRxiv [Preprint]. doi: 10.1101/2024.01.09.574915

PubMed Abstract | Crossref Full Text | Google Scholar

van’t Klooster, M. A., van Klink, N. E. C., van Blooijs, D., Ferrier, C. H., Braun, K. P. J., Leijten, F. S. S., et al. (2017). Evoked versus spontaneous high frequency oscillations in the chronic electrocorticogram in focal epilepsy. Clin. Neurophysiol. 128, 858–866. doi: 10.1016/j.clinph.2017.01.017

PubMed Abstract | Crossref Full Text | Google Scholar

van’t Klooster, M. A., Zijlmans, M., Leijten, F. S., Ferrier, C. H., Van Putten, M. J., and Huiskamp, G. J. (2011). Time–frequency analysis of single pulse electrical stimulation to assist delineation of epileptogenic cortex. Brain 134, 2855–2866. doi: 10.1093/brain/awr211

PubMed Abstract | Crossref Full Text | Google Scholar

Veit, M., Kucyi, A., Hu, W., Zhang, C., Zhao, B., Guo, Z., et al. (2021). Temporal order of signal propagation within and across intrinsic brain networks. Proc. Natl. Acad. Sci. U.S.A. 118:e2105031118. doi: 10.1073/pnas.2105031118

PubMed Abstract | Crossref Full Text | Google Scholar

Wu, T., Kaboodvand, N., McGinn, R., Veit, M., Davey, Z., Datta, A., et al. (2023). Multisite thalamic recordings to characterize seizure propagation in the human brain. Brain 146, 2792–2802. doi: 10.1093/brain/awad121

PubMed Abstract | Crossref Full Text | Google Scholar

Yamao, Y., Matsumoto, R., Kunieda, T., Nakae, T., Nishida, S., Inano, R., et al. (2021). Effects of propofol on cortico-cortical evoked potentials in the dorsal language white matter pathway. Clin. Neurophysiol. 132, 1919–1926. doi: 10.1016/j.clinph.2021.04.021

PubMed Abstract | Crossref Full Text | Google Scholar

Yamao, Y., Suzuki, K., Kunieda, T., Matsumoto, R., Arakawa, Y., Nakae, T., et al. (2017). Clinical impact of intraoperative CCEP monitoring in evaluating the dorsal language white matter pathway. Hum. Brain Mapp. 38, 1977–1991. doi: 10.1002/hbm.23498

PubMed Abstract | Crossref Full Text | Google Scholar

Yang, B., Zhao, B., Li, C., Mo, J., Guo, Z., Li, Z., et al. (2024). Localizing seizure onset zone by a cortico-cortical evoked potentials-based machine learning approach in focal epilepsy. Clin. Neurophysiol. 158, 103–113. doi: 10.1016/j.clinph.2023.12.135

PubMed Abstract | Crossref Full Text | Google Scholar

Zauli, F., Del Vecchio, M., Russo, S., Mariani, V., Pelliccia, V., d’Orio, P., et al. (2022). The web of laughter: Frontal and limbic projections of the anterior cingulate cortex revealed by cortico-cortical evoked potential from sites eliciting laughter. Philos. Trans. R. Soc. Lond. B Biol. Sci. 377:20210180. doi: 10.1098/rstb.2021.0180

PubMed Abstract | Crossref Full Text | Google Scholar

Zhao, C., Liang, Y., Li, C., Gao, R., Wei, J., Zuo, R., et al. (2019). Localization of epileptogenic zone based on cortico-cortical evoked potential (CCEP): A feature extraction and graph theory approach. Front. Neuroinform. 13:31. doi: 10.3389/fninf.2019.00031

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: cortico-cortical evoked potentials, single pulse electrical stimulation, CCEP detection, SEEG, ECoG, epilepsy, functional connectivity

Citation: Al-Sadek T, Wadhwa A, Wadhwa M, Warren AEL and Rolston JD (2025) Methodologies to detect cortico-cortical evoked potentials: a systematic review. Front. Hum. Neurosci. 19:1636115. doi: 10.3389/fnhum.2025.1636115

Received: 20 June 2025; Accepted: 18 August 2025;
Published: 01 September 2025.

Edited by:

Napoleon R. Torres Martinez, Commissariat À l’Energie Atomique et Aux Energies Alternatives (CEA), France

Reviewed by:

Eishi Asano, Wayne State University, United States
Odile Feys, Hopital de la Timone, France

Copyright © 2025 Al-Sadek, Wadhwa, Wadhwa, Warren and Rolston. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Tamara Al-Sadek, dG9zMDVAbWFpbC5hdWIuZWR1

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.