Heartbeat detection and personal authentication using a 60 GHz Doppler sensor


Takuma AsanoTakuma AsanoShintaro Izumi
Shintaro Izumi*Hiroshi Kawaguchi
Hiroshi Kawaguchi
  • Architecture Laboratory, Graduate School of Science, Technology and Innovation, Kobe University, Kobe, Japan

Background: Microwave Doppler sensors, capable of detecting minute physiological movements, enable the measurement of biometric information, such as walking patterns, heart rate, and respiration. Unlike fingerprint and facial recognition systems, they offer authentication without physical contact or privacy concerns. This study focuses on non-contact seismocardiography using microwave Doppler sensors and aims to apply this technology for biometric authentication.

Method: We proposed a method for authenticating and identifying heartbeat signals through supervised learning using a conditional variational autoencoder (CVAE). A 60 GHz microwave Doppler sensor was used to capture heartbeat signals, which were processed using a conformer network to detect peaks and segment individual beats. High signal-to-noise ratio waveforms were selected, and time-frequency analysis extracted relevant features. Spectrograms labeled with subject data were input into the CVAE, which encoded subject-specific features into a latent space for authentication.

Results: The proposed heartbeat-based authentication method, validated on 13 subjects, achieved an average balanced accuracy of 97.3% for authentication and an average accuracy of 94.7% for identification. Compared with conventional methods, this approach demonstrated superior performance by effectively encoding subject-specific features while mitigating noise-related challenges.

Conclusion: The proposed method enhanced the feasibility of non-contact heartbeat-based authentication by achieving high accuracy while addressing noise-related challenges. Its application could improve biometric security without compromising user privacy. Further advancements in handling posture variations and scalability are essential for real-world implementation.

1 Introduction

Biometric authentication extracts unique features, such as facial structures and fingerprints, from biometric data to identify individuals. Unlike traditional authentication methods, it does not require users to memorize passwords, enter information, or carry physical identification. Moreover, biometric authentication is more secure because it is difficult to impersonate and less vulnerable to theft. Given these advantages, there is growing demand for new biometric modalities that enhance both usability and security, especially non-contact methods that reduce hygiene risks.

Among various modalities, heart-based biometric authentication has gained attention because it reflects internal physiological characteristics that are difficult to forge and can be measured noninvasively. Individual differences in cardiac muscle thickness, electrical conduction, and vascular elasticity contribute to unique heartbeat patterns (1), and personal identification can be derived by analyzing the amplitude and duration of the P, Q, R, S, and T waves in an electrocardiogram (ECG). Typically, ECGs are recorded by measuring electrical potential differences using electrodes attached to the body surface. Various ECG-based biometric authentication methods have also been proposed. Arteaga-Falconi et al. (2) introduced an ECG-based algorithm that enables authentication by lightly touching a mobile device, pioneering ECG authentication for mobile applications. Choi et al. (3) developed preprocessing and feature extraction techniques to improve authentication robustness against noise in mobile ECG data. Gutta and Cheng (4) proposed a method that simultaneously optimizes feature extraction and classifier design, and Sun et al. (5) utilized individualized autoencoders for personalized ECG authentication.

However, contact-based methods face practical challenges such as the inconvenience of electrode attachment, hygiene concerns, and user discomfort. To address these issues, non-contact techniques have been proposed using microwaves or accelerometer sensors to detect heartbeats or body surface vibrations. Lin et al. (6) proposed a non-contact authentication system utilizing continuous wave radar to sense cardiac motion. Rissacher and Galy (7) used a 2.4 GHz radar system with continuous wavelet transform (CWT) feature extraction, while Okano et al. (8) applied autoregressive (AR) models and spectrogram analysis to heartbeat signals captured by a 24 GHz Doppler sensor. Shi et al. (9) performed personal identification with 24 GHz radar and support vector machines (SVM). Cao et al. (10) extracted micro-Doppler signals using short-time Fourier transform (STFT) and classified them with deep convolutional neural networks (DCNN). Huang et al. (11) achieved continuous authentication based on near-field heart rate signals using an accelerometer sensor, employing a convolutional autoencoder (CAE). Wu et al. (12) combined ultra-wideband (UWB) radar and convolutional neural networks (CNN) for heartbeat-based identification. Hinatsu et al. (13) proposed a biometric authentication method based on measuring subtle displacements of the chest wall using a very high frequency band loop antenna, and Kobayashi et al. (14) introduced a personal identification approach that combines respiratory and heartbeat-derived features measured by a 79 GHz millimeter-wave radar.

In this study, we propose a biometric authentication framework that combines a deep generative model, specifically a conditional variational autoencoder (CVAE) (15, 16), with non-contact heartbeat measurement using a 60 GHz microwave Doppler sensor. By employing a CVAE, the latent feature space is structured based on subject IDs, enabling robust classification even under signal variability. The design concept of the proposed authentication system is illustrated in Figure 1.

Figure 1
Diagram illustrating a system using a microwave Doppler sensor to detect heartbeat fluctuations from a seated person. The output wave, shown as a voltage over time graph, undergoes time-frequency analysis into a spectrogram. A CVAE (Conditional Variational Autoencoder) processes the data, passing through an encoder, latent space, and decoder. The latent space is depicted with clusters labeled from zero to twelve, indicating different data groupings.

Figure 1. Concept of the proposed authentication method.

2 Materials and methods

2.1 Microwave Doppler sensor

The time intervals of small displacements on the body surface due to heartbeats strongly correlate with the interbeat intervals in an ECG. Therefore, heartbeats can be detected by measuring these body surface displacements without directly measuring the ECG. In this study, a Doppler sensor utilizing microwaves in the 60 GHz band was used to measure surface displacements.

When microwaves are emitted toward an object, the frequency of the reflected waves shifts based on the object's velocity due to the Doppler effect. The microwave Doppler sensor receives the reflected waves and outputs signals by mixing the transmitted and received waves. If the transmission frequency is f0[Hz], the velocity of the object irradiated by the microwaves is v[m/s], and the speed of light is c[m/s]. The receiving frequency fr[Hz] can be expressed using the Doppler effect, as shown in (Equation 1):

fr=c+vcvf0=f0+2vcvf0f0+2vcf0(cv)(1)

where the frequency shift of the reflected wave relative to the velocity of the object is proportional to the frequency of the transmitted wave. Therefore, higher transmitted frequencies provide higher resolutions. Vibrations on the body surface induced by heartbeats have displacements of <1 mm. Radio waves at 60 GHz are mostly reflected by the body surface; however, this frequency is sufficient to capture the heartbeat component from the body surface.

A Doppler sensor (BGT60LTR11AIP, Infineon Technologies) was used for this experiment, with a sampling rate set to 250 Hz. The measured data were wirelessly transmitted via Bluetooth low energy (LE) and transferred to a personal computer (PC) in real time.

2.2 Measured example of heartbeat

Figure 2 shows the measurement results obtained from the human body (back) using a microwave Doppler sensor. Reference data were measured using a patch-type ECG sensor (SEN0213, Zhiwei Robotics Corp.) attached to both hands and the left leg. The body surface oscillations corresponding to systole and diastole were also measured (Figure 2). The waveform obtained using the microwave Doppler sensor was similar to that obtained using the seismocardiogram (SCG) (17), supporting the hypothesis that a microwave Doppler sensor can measure the SCG without physical contact.

Figure 2
Four graphs illustrate different physiological signals over time in seconds. The Doppler wave graph, in blue, shows frequencies from fifteen to fifty hertz. The second graph, labeled SCG (seismocardiogram), is in green. The third graph, labeled ECG (electrocardiogram), is in red. The fourth graph, labeled PPG (photoplethysmogram), is in light blue. Each graph displays varying voltage levels in millivolts.

Figure 2. Measurement results of heartbeat.

2.3 Measurement and dataset construction

Thirteen subjects (11 males and 2 females, aged 22–24 years) were measured to evaluate the performance of the proposed method under the conditions shown in Figure 3. The measurement duration for each subject was 60 s. The heartbeat was measured while the subjects sat and breathed normally. The measured data were segmented using the method described later, retaining only data with a high signal-to-noise ratio. The frequency region containing the heartbeat component was extracted using a time-frequency analysis. Furthermore, a five-beat time-series moving average was then calculated, followed by data augmentation. This study was approved by the Ethics Committee of the Graduate School of System Informatics, Kobe University (approval number: R03-02). All procedures adhered to the ethical standards of the Institutional and National Research Committee and the 1964 Helsinki Declaration, along with its later amendments or comparable ethical standards.

Figure 3
A person sits in a chair connected to an ECG sensor on their hand. To the side, a mounted microwave Doppler sensor transmits data via Bluetooth LE to a laptop displaying ECG and Doppler wave graphs.

Figure 3. Measurement conditions.

2.4 Preprocessing

The measured time series of the Doppler waves was used to create a dataset for machine learning.

The time-series waveform obtained by the microwave Doppler sensor included various types of noise, such as environmental noise (including ambient and body movement noise), internal noise from the sensor circuit, external noise affecting the sensor. To reduce the impact of noise and respiration, a bandpass filter was applied to allow only specific frequency bands to pass. In this study, a bandpass filter with a passband of 15–50 Hz was applied. This frequency range was carefully selected because, when using a 60 GHz Doppler sensor, heartbeat-induced micro-vibrations are primarily observed within the 15–50 Hz range, whereas respiratory movements mainly appear at frequencies below 10 Hz. Setting the lower cutoff frequency at 15 Hz effectively suppresses respiration components while preserving heartbeat signals. Additionally, the upper cutoff frequency at 50 Hz helps to eliminate high-frequency noise, including 60 Hz power-line interference, thereby improving signal quality for heartbeat detection.

The R-wave peaks of the heartbeats were then extracted from the Doppler waves using a machine learning approach. In this study, we utilized the conformer (18) model to estimate the R-wave peaks of heartbeats. Specifically, Doppler waves measured separately from those used for authentication were used as input features, and the R-wave peaks of the ECG, processed with a 1 Hz high-pass filter, served as the ground truth for training. The trained model was then used to estimate the R-wave peaks from the Doppler waves measured for authentication. Ground-truth data were prepared by assigning a value of one to the R-wave peaks and a value of zero to all the other points. The data were then converted into a triangular waveform format. The conformer is a machine learning model that combines CNN and transformer architectures, primarily used in natural language processing. For the heartbeat peak extraction task using the conformer, data from 13 subjects, with 60 s of data per subject, were used for training. Overlapping processing was applied to compensate for the limited data. The machine-learning model consisted of five conformer blocks with dimensions of 30. Each block included a multihead self-attention module with three attention heads, a convolution module with a kernel size of 31, and a feed-forward module with 512 dimensions. The batch size was set to 512, the number of epochs to 300, the loss function to mean squared error (MSE), the optimizer to Adam, and the learning rate to 0.0001.

After downsampling the waveform of the extracted peak and Doppler wave to 125 Hz, the Doppler wave was segmented based on the peak. The signal-to-noise ratio was then calculated for each segmented Doppler waveform, selecting only those with high signal-to-noise ratios. In this study, the signal-to-noise ratio was calculated using the peak value as the signal and the average value of the Doppler waveform as the noise. Doppler waveforms with a signal-to-noise ratio of four or higher were used for the training data, while those with a signal-to-noise ratio of six or higher were used for the test data. The signal-to-noise ratio was set slightly lower for the training data than for the test data to enhance the data robustness by adding noise. Subsequently, a time-frequency analysis was conducted using a CWT (see Figure 4).

Figure 4
Diagram illustrating the process of extracting R-wave peaks from the Doppler wave using a conformer. Step 1 shows the training and inference processes with Doppler and ECG waveforms; the conformer consists of layer normalization, feed forward networks, convolution, and multi-head self-attention. Step 2 demonstrates Doppler wave splitting and continuous wavelet transform (CWT), with visuals of extracted R-wave peaks, split Doppler waves, and CWT frequency representations.

Figure 4. Dataset construction.

The challenge in non-contact measurements is the isolation of body motion noise. Doppler waves contain noise from body motion and the surrounding environment. However, Doppler shifts due to breathing and body movement are smaller than those caused by heartbeats and can be distinguished using frequency-domain computation. Therefore, we extracted the heartbeat component by using a spectrogram as an input to a neural network.

The resolution of the time-frequency analysis should be considered carefully. The STFT is a widely used method for time-frequency analysis; however, it struggles to achieve both frequency and time resolution owing to the uncertainty principle of the Fourier transform (19). Therefore, time-frequency analysis was conducted using the CWT, as shown in (Equation 2):

W(a,b)=|a|(1/2)x(t)ψ(tba)dt(2)

where x(t) is the signal to be analyzed, ψ is the mother wavelet, a is the scale parameter corresponding to the frequency to be extracted, and b is the shift parameter corresponding to the time to be analyzed. In this study, the Morlet wavelet with ω0=6 is used as the mother wavelet (20), as shown in (Equation 3):

ψ(t)=π14eiω0tet22(3)

Finally, data augmentation (21) was applied to the spectrograms. This included time shifts, where data points on the time axis were shifted forward and backward by ±2 points; time stretching, where the length on the time axis was altered; mixup (22), which linearly combined two spectrograms; the addition of white noise, a random signal with equal intensity across all frequencies; and the calculation of a moving average over every five beats.

2.5 CVAE model

In this study, we employed a CVAE (15, 16) as the machine learning model for heartbeat-based biometric authentication. The CVAE is a deep generative model proposed by Sohn et al., based on the Variational Autoencoder (VAE) (23), which was originally introduced by Kingma et al.

While the standard VAE learns an unsupervised latent representation of input data, the CVAE extends this framework to support supervised learning by conditioning the latent space on class labels. The VAE encodes input data into latent variables and decodes outputs from compressed representations to reconstruct the original input. In contrast, the CVAE controls the distribution over the latent space based on the labels assigned during training, thereby enabling the model to structure the latent space according to class-specific information.

Let (X,Y)={(x1,y1),,(xN,yN)} represent the measured data and their corresponding class labels, where xiRD denotes the input data and yi0,1,2,,K1 denotes the subject ID. Each data point xi is associated with a latent variable zi. The decoder distribution pθ(x|z) and the prior distribution pψ(z|y) are defined by (Equations 4 and 5):

pθ(x|z)=Bern(μθ(z))(4)
pψ(z|y)=N(μψ(y),σψ2(y))(5)

θ and ψ are hyperparameters of the neural network in (Equations 4 and 5). The posterior distribution is approximated as in (Equation 6). ϕ is a hyperparameter.

qϕ(z|x)=N(μϕ(x),σϕ2(x))(6)

The training objective is to maximize the conditional log-likelihood logpθ,ψ(x|y), which is intractable and instead approximated by maximizing the variational lower bound. This bound is given by the expected log-likelihood of the reconstructed input minus the Kullback–Leibler (KL) divergence between the approximate posterior and prior distributions, as shown in (Equation 7):

logpθ,ψ(x|y)=>qϕ(z|x)logpθ(x,z|y)qϕ(z|x)dz=qϕ(z|x)logpθ(x|z)pψ(z|y)qϕ(z|x)dz=Eqϕ(z|x)[logpθ(x|z)]DKL[qϕ(z|x)||pψ(z|y)]=LCVAE(x,y;θ,ϕ,ψ)(7)

The CVAE model used in this study follows the structure proposed by Sohn et al., with a key modification: the encoder qϕ(z|x,y) is simplified to qϕ(z|x), excluding the label y from the encoder input. This modification allows the encoder to autonomously extract individual-specific features from the input data without explicitly requiring label information during encoding, while the prior remains label-dependent.

The overall model architecture is illustrated in Figure 5. The encoder network predicts the mean and variance of the latent distribution, from which the latent variable z is sampled using the reparameterization trick (23), expressed as z=μ+ϵσ2, where ϵN(0,I) follows a standard normal distribution. Simultaneously, the prior network generates a label-conditioned prior distribution. The KL divergence between the encoder's output and the label-conditioned prior is minimized to structure the latent space, while the decoder reconstructs the input from the latent variable. The overall loss function is the sum of the reconstruction loss, computed using MSE, and the KL divergence term.

Figure 5
Flowchart of a Variational Autoencoder with Encoder and Decoder processes. Input data \\(X\\) is processed by the Encoder to obtain latent variables \\(z_1\\), which undergo a transformation using mean \\(\\mu_\\phi\\), variance \\(\\sigma^2_\\phi\\), and noise \\(\\varepsilon\\). Prior \\(p_\\psi(z|y)\\) links \\(y\\) to another latent variable \\(z_2\\). The Decoder transforms \\(z_2\\) back to output data \\(\\hatX\\). Processes involve Kullback-Leibler Divergence and Mean Squared Error calculations.

Figure 5. Architecture of the proposed model.

Figure 6 presents the detailed network structures of the encoder, prior, and decoder. The encoder consists of a series of two-dimensional convolutional layers followed by max pooling and a feed-forward network (FFN) to predict the mean and variance of the latent distribution. The prior network employs an FFN that maps the label to the latent space parameters. The decoder comprises an FFN followed by two-dimensional transposed convolutional layers to reconstruct the original spectrogram from the latent representation.

Figure 6
Flowchart of a neural network architecture involving an encoder, decoder, and prior. The encoder, labeled \\(Q(z|x)\\), includes a sequence of Conv2d and MaxPool layers followed by a linear layer transforming input data of size \\(1 \\times 24 \\times 100\\) to \\(z_dim\\). The decoder, labeled \\(P(x|z)\\), reverses this process through linear and Deconv2d layers to reconstruct data. The prior \\(P(z|y)\\) takes a label \\(y\\) and converts it to \\(z_dim\\) through a linear layer, ensuring transformation between data and latent spaces via \\(\\mu\\) and \\(\\sigma\\).

Figure 6. Layer structure details.

2.6 Authentication using CVAE

Biometric authentication was performed using the latent representations generated by the CVAE model. During training, heartbeat spectrograms measured from each subject were encoded into a latent space, where data points with the same subject ID gradually formed distinct clusters. Each cluster's center was calculated as the mean of the latent variables corresponding to that subject's training samples.

The authentication process is depicted in Figure 7. In the inference phase, test spectrograms were encoded into the latent space using the trained encoder. The distance between each encoded representation and the center of each subject's cluster was then calculated (24). In this study, the Mahalanobis distance was used. A test sample was authenticated as belonging to a subject if its Mahalanobis distance to that subject's cluster center was smaller than a predefined threshold; otherwise, it was rejected as an unregistered individual.

Figure 7
Diagram illustrating the training and inference processes of a Conditional Variational Autoencoder (CVAE). The top half shows training with labeled input data fed into an encoder, transforming it into a latent space represented on a graph with mean coordinates for each label. The bottom half shows inference with test data undergoing a similar encoding process. The output is evaluated using Mahalanobis distance to determine authentication by comparing test data to the mean of training data. Each label is illustrated and color-coded on the graph.

Figure 7. Architecture of the proposed model.

This distance-based authentication approach enabled robust and interpretable decision-making in the latent space structured by the CVAE model. The hyperparameters used for training the CVAE model were: batch size of 64, 300 training epochs, Adam optimizer with a learning rate of 0.00001, and a latent space dimension of 50.

3 Results

3.1 Evaluation and method

In this study, 48 s of the measured 60-s data were used for training, and the remaining 12 s were used for performance evaluation. However, the data used in these processes were not used as training data for the conformer-based heartbeat peak extraction.

First, leave-one-out cross-validation (LOOCV) was conducted to evaluate the authentication performance that proved the identity of a specific individual. In this method, the model was trained by excluding the data of one of the 13 subjects. This approach prevented data bias and enabled the assessment of the generalization performance. Of the data from the 12 subjects not excluded by LOOCV, data from one subject were used as a single label for training, whereas the data from the remaining 11 were aggregated into a single label. To balance the data volume, 1/11 of each subject's data was extracted from the remaining 11 to create the training dataset. Finally, the performance in rejecting unregistered subjects was evaluated using inference data from those excluded from LOOCV.

Subsequently, the data of all 13 subjects were trained as separate labels to evaluate the identification performance, which determines which registered individual matched the data. As the identification task involved selecting the subject most closely matched among the registered total, LOOCV was not conducted. For inference, 12 s of data not used in the training for all registered subjects were used to evaluate the performance of correctly identifying registered subjects.

During the inference process, a majority-voting method was employed using multiple consecutive heartbeat waveforms, specifically, majority voting on data from five heartbeats.

The results obtained were used to illustrate the correlation between the false rejection rate (FRR) and false acceptance rate (FAR) when the Mahalanobis distance threshold in the latent space was varied, as shown in Figure 8. Given that these values were correlated, they fluctuated based on the set threshold. The equal error rate (EER) was achieved when the Mahalanobis distance threshold was adjusted such that FRR and FAR were equal. This evaluation was conducted for all the subjects.

Figure 8
Line graph showing FRR, FAR, and EER against threshold values. FRR and FAR approach 1.0 quickly, while EER occurs near the start. The legend distinguishes between regular and 5 data sets using different line styles and markers.

Figure 8. Example of the correlation between FRR and FAR with varying Mahalanobis distance thresholds in the latent space.

3.2 Metrics

The accuracy of authentication was evaluated using balanced accuracy (BAC), F1-score, and EER, whereas the accuracy of identification was evaluated using ACC.

BAC is a measure used to evaluate classifier performance on unbalanced datasets. It is the average of the true positive rate (TPR) and true negative rate (TNR), which evaluates the prediction accuracy for each class equally.

3.3 Experimental results

The results of calculating the average BAC, F1-score, and EER for each subject during authentication are shown in Table 1. Majority voting was performed using multiple consecutive heartbeat waveforms. Majority voting with five heartbeat datasets resulted in a BAC of 97.3%. For identification, the average ACC for each subject was calculated, with the results presented in Table 2. The majority of the votes using the five heartbeat datasets resulted in an ACC of 94.7%. These results indicate that extending the duration used for authentication and performing inferences based on multiple heartbeat signals can improve the accuracy.

Table 1
www.frontiersin.org

Table 1. Evaluation scores in authentication (macro average).

Table 2
www.frontiersin.org

Table 2. Evaluation scores in identification (macro average).

Additionally, Table 3 presents subject-wise evaluation results, allowing for further analysis of individual differences in performance.

Table 3
www.frontiersin.org

Table 3. Evaluation scores for each subject.

To evaluate the impact of sensor position shifts and changes in heartbeat waveforms over time, two subjects were re-measured at the same location on different days. The results inferred from the re-measured data are shown in Table 4. In this data, the BAC and F1-score decreased by 4.0% and 7.2%, respectively, while the EER increased by 4.8%.

Table 4
www.frontiersin.org

Table 4. Changes in evaluation scores over time.

4 Discussion

Table 5 compares the performance of previous heartbeat-based contactless authentication methods. Prior studies using microwave Doppler sensors (7) demonstrated that feature extraction through time-frequency analysis and K-nearest neighbor search could be applied for authentication. However, the accuracy of these methods was insufficient. In our previous study (8), we used a 24 GHz microwave Doppler sensor for heartbeat extraction. This method relied on AR model-based time–frequency analysis and spectrogram correlation. The sensor was fixed to the front of the chest using a belt; however, body motion noise could not be completely eliminated. Machine learning-based methods such as SVM (6, 9, 10, 13, 14), DCNN (10), and CAE (11) have also been proposed. These methods, however, face limitations such as experimental complexity (6, 14), low accuracy, small sample sizes (9, 10, 13), and restricted applicability (11).

Table 5
www.frontiersin.org

Table 5. Comparison with prior works.

In contrast, the proposed method achieves high authentication and identification performance while offering enhanced usability. Unlike many previous systems, it does not require users to wear sensors or maintain constrained postures. Our system enables high-accuracy authentication without forcing users to fix a sensor to the body with a belt or remain seated in a specific position. Although our experiment was conducted in a seated posture, the non-contact approach allows natural use without burdening the user and is expected to be adaptable to other postures as well. Compared with existing methods requiring controlled experimental setups, our approach has the potential to expand the range of applicable scenarios. This flexibility, combined with high accuracy, is a key strength of the proposed system.

In terms of cost, the 60 GHz Doppler sensor used in this study offers higher frequency and accuracy than lower-frequency alternatives (e.g., 2.4 or 24 GHz), placing it in a mid-range price tier. Some previous studies have employed high-precision but specialized radar systems that may limit widespread adoption. In contrast, our method uses a commercially available radar module that enables non-contact authentication without requiring physical attachment, offering advantages in accessibility, ease of setup, and user experience.

Although the proposed method exhibited high authentication and identification accuracy, a potential limitation is the small sample size. In this study, we proposed a method for non-contact measurement of heartbeat signals using a microwave Doppler sensor and a framework for personal authentication and identification based on these signals. The method was validated using spectrograms from 13 individuals. Results showed that the proposed method achieved an average BAC of 97.3% for authentication and an average ACC of 94.7% for identification. By comparison, Huang et al. (11) reported a BAC of 96.5% and an EER of 3.8%. Our method improved the EER to 2.7%, a 28.9% relative improvement, and increased BAC by 0.8% points. ACC was comparable with the 94.6% reported by Shi et al. (9).

The re-measurement experiment, which simulated temporal variation in sensor conditions, showed a modest decline in performance (BAC −4.0%, F1-score −7.2%, EER +4.8%). This degradation appears acceptable for practical use, as the envisioned application scenario assumes a fixed device installation (e.g., in a specific room or vehicle) without repositioning after initial calibration. Under this assumption, the ability to authenticate previously enrolled users after a time interval remains sufficient for real-world deployment.

Regarding the number of subjects, we agree that evaluating generalizability with a larger and more diverse population is scientifically important. However, the intended use scenario for this system assumes deployment in relatively small communities, such as research groups, internal organizational teams, or households. In such cases, optimizing the model for a limited user base is both practical and effective. Therefore, we believe the current sample size remains meaningful from a practical perspective. Nonetheless, we plan to expand the subject pool in future work to investigate scalability and robustness further.

Although remote authentication is not the primary focus of this study, the proposed method assumes that the target user remains relatively stationary. We recognize that achieving robust authentication under light movement conditions would enhance the system's practicality. We consider this an important direction for future research aimed at improving the robustness and versatility of the approach.

Future work should not only examine the impact of increasing the number of subjects on authentication accuracy, but also develop methods that account for variations in biometric signals due to changes in posture and physical condition. Improving robustness to postural variation remains a key challenge, and algorithms capable of adapting to such changes must be established. For example, an authentication system should function reliably even when the user is leaning forward or resting against a chair. Furthermore, it is essential to maintain stable authentication performance under conditions involving light body movement. Achieving this will require advances in signal processing and machine learning algorithms to extract consistent and invariant heartbeat components across diverse measurement conditions.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by the Ethics Committee of Kobe University. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

TA: Conceptualization, Investigation, Methodology, Writing – original draft, Writing – review & editing. SI: Conceptualization, Investigation, Methodology, Writing – original draft, Writing – review & editing. HK: Conceptualization, Investigation, Methodology, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. This research was partially supported by the Japan Science and Technology Agency (JST) under Grant Number JPMJPF2115.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Hall JE. Guyton and Hall Textbook of Medical Physiology E-book. Philadelphia, PA: Elsevier Health Sciences (2015).

Google Scholar

2. Arteaga-Falconi JS, Al Osman H, El Saddik A. ECG authentication for mobile devices. IEEE Trans Instrum Meas. (2016) 65:591–600. doi: 10.1109/TIM.2015.2503863

Crossref Full Text | Google Scholar

3. Choi HS, Lee B, Yoon S. Biometric authentication using noisy electrocardiograms acquired by mobile sensors. IEEE Access. (2016) 4:1266–73. doi: 10.1109/ACCESS.2016.2548519

Crossref Full Text | Google Scholar

4. Gutta S, Cheng Q. Joint feature extraction and classifier design for ECG-based biometric recognition. IEEE J Biomed Health Inform. (2016) 20:460–8. doi: 10.1109/JBHI.2015.2402199

PubMed Abstract | Crossref Full Text | Google Scholar

5. Sun L, Zhong Z, Qu Z, Xiong N. PerAE: an effective personalized autoencoder for ECG-based biometric in augmented reality system. IEEE J Biomed Health Inform. (2022) 26:2435–46. doi: 10.1109/JBHI.2022.3145999

PubMed Abstract | Crossref Full Text | Google Scholar

6. Lin F, Song C, Zhuang Y, Xu W, Li C, Ren K. Cardiac scan: a non-contact and continuous heart-based user authentication system. Proceedings of the 23rd Annual International Conference on Mobile Computing and Networking (2017). p. 315–28. doi: 10.1145/3117811.3117839

Crossref Full Text | Google Scholar

7. Rissacher D, Galy D. Cardiac radar for biometric identification using nearest neighbour of continuous wavelet transform peaks. Proc. ISBA 2015; Hong Kong (2015). p. 1–6. doi: 10.1109/ISBA.2015.7126356

Crossref Full Text | Google Scholar

8. Okano T, Izumi S, Kawaguchi H, Yoshimoto M. Non-contact biometric identification and authentication using microwave Doppler sensor. Proc. BioCAS 2017; Torino, Italy (2017). p. 1–4. doi: 10.1109/BIOCAS.2017.8325160

Crossref Full Text | Google Scholar

9. Shi K, Will C, Weigel R, Koelpin A. Contactless person identification using cardiac radar signals. Proc. I2MTC 2018; Houston, Texas, USA (2018). p. 1–6. doi: 10.1109/I2MTC.2018.8409645

Crossref Full Text | Google Scholar

10. Cao P, Xia W, Li Y. Heart ID: human identification based on radar micro-Doppler signatures of the heart using deep learning. Remote Sens (Basel). (2019) 11:1220. doi: 10.3390/rs11101220

Crossref Full Text | Google Scholar

11. Huang Y, Qiu M, Chen L, Peng Z, Zhang Q, Wu K. NF-heart: a near-field non-contact continuous user authentication system via ballistocardiogram. Proc ACM Interact Mob Wearable Ubiquitous Technol. (2023) 7:1–24. doi: 10.1145/3580851

Crossref Full Text | Google Scholar

12. Wu S, Sakamoto T, Oishi K, Sato T, Inoue K, Fukuda T, et al. Person-specific heart rate estimation with ultra-wideband radar using convolutional neural networks. IEEE Access. (2019) 7:168484–94. doi: 10.1109/ACCESS.2019.2954294

Crossref Full Text | Google Scholar

13. Hinatsu S, Wada S. Biometric authentication using chest wall displacement recorded by VHF-band loop antenna. IEEE Access. (2024) 12:196475–87. doi: 10.1109/ACCESS.2024.3521013

Crossref Full Text | Google Scholar

14. Kobayashi H, Tanaka Y, Sakamoto T. Individual identification using radar-measured respiratory and heartbeat features. IEEE Access. (2024) 12:190972–87. doi: 10.1109/ACCESS.2024.3517668

Crossref Full Text | Google Scholar

15. Kingma DP, Rezende DJ, Mohamed S, Welling M. Semi-supervised learning with deep generative models. Proc. NIPS 2014; Montreal, Canada (2014). p. 3581–9

Google Scholar

16. Sohn K, Lee H, Yan X. Learning structured output representation using deep conditional generative models. Proc. NIPS 2015; Montreal, Canada (2015). p. 3483–91

Google Scholar

17. Mora N, Cocconcelli F, Matrella G, Ciampolini P. Detection and analysis of heartbeats in seismocardiogram signals. Sensors. (2020) 20:1670. doi: 10.3390/s20061670

PubMed Abstract | Crossref Full Text | Google Scholar

18. Gulati A, Qin J, Chiu CC, Parmar N, Zhang Y, Yu J, et al. Conformer: convolution-augmented transformer for speech recognition (2020).

19. Cohen L. Time-Frequency Analysis. Upper saddle River, NJ: Prentice Hall (1995).

Google Scholar

20. Torrence C, Compo GP. A practical guide to wavelet analysis. Bull Am Meteorol Soc. (1998) 79:61–78. doi: 10.1175/1520-0477(1998)079%3C0061:APGTWA%3E2.0.CO;2

Crossref Full Text | Google Scholar

21. Shorten C, Khoshgoftaar TM. A survey on image data augmentation for deep learning. J Big Data. (2019) 6:60. doi: 10.1186/s40537-019-0197-0

Crossref Full Text | Google Scholar

22. Zhang H, Cisse M, Dauphin YN, Lopez-Paz D. Mixup: beyond empirical risk minimization. Proc. ICLR 2018; Vancouver, BC, Canada (2017).

Google Scholar

23. Kingma DP, Welling M. Auto-encoding variational Bayes. Proc. ICLR 2014; Banff, AB, Canada (2013).

Google Scholar

24. Zhu Q, Zhang R. A classification supervised auto-encoder based on predefined evenly-distributed class centroids (2019).

Keywords: biometric authentication, conditional variational autoencoder (CVAE), heartbeat, microwave Doppler sensor, non-contact measurement

Citation: Asano T, Izumi S and Kawaguchi H (2025) Heartbeat detection and personal authentication using a 60 GHz Doppler sensor. Front. Digit. Health 7:1570144. doi: 10.3389/fdgth.2025.1570144

Received: 3 February 2025; Accepted: 4 August 2025;
Published: 21 August 2025.

Edited by:

Chen Wenxi, University of Aizu, Japan

Reviewed by:

Tooraj Nikoubin, The University of Texas at Dallas, United States
Eko Sakti Pramukantoro, Brawijaya University Hospital, Indonesia

Copyright: © 2025 Asano, Izumi and Kawaguchi. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Shintaro Izumi, c2hpbkBjczI4LmNzLmtvYmUtdS5hYy5qcA==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.