Your new experience awaits. Try the new design now and help us make it even better

ORIGINAL RESEARCH article

Front. Med.

Sec. Intensive Care Medicine and Anesthesiology

Volume 12 - 2025 | doi: 10.3389/fmed.2025.1538280

Effects of Different Exhalation Valves on CO 2 Rebreathing and Ventilator Performance during Noninvasive Ventilation

Provisionally accepted
Xinyu  LiXinyu Li1Dai  BingDai Bing1Chengguang  ZhouChengguang Zhou2Haijia  HouHaijia Hou1Shengchen  WangShengchen Wang1Xiangrui  LiXiangrui Li1Hongwen  ZhaoHongwen Zhao1Wei  WangWei Wang1Wei  TanWei Tan1*
  • 1Department of Pulmonary and Critical Care Medicine, The First Affiliated Hospital of China Medical University, Shenyang, China
  • 2Shenyang RMS Medical Tech Co., Ltd, No.10-4, Jinhui Street, Hunnan District,, Shenyang, China

The final, formatted version of the article will be published soon.

Background: Noninvasive ventilation (NIV) is widely used to improve oxygenation and reduce carbon dioxide (CO2) retention in patients with respiratory failure. However, it remains unclear whether different types of exhalation valves affect CO2 rebreathing and ventilator performance during NIV.Methods: Three noninvasive ventilators (V60, Flexo, and Stellar150) with single-limb circuits and four different exhalation valves (single-arch, whisper swivel, plateau exhalation, and vented mask valves) were separately connected in series to a lung simulator. CO2 gas was injected from the simulated lung outlet, maintaining the end-expiratory CO2 (PetCO2) at 80 mmHg. Both the CO2 rebreathing volume (CO2REB) and the parameters displayed on the lung simulator and ventilator were recorded under each condition.The mean CO2REB values of the four aforementioned valves were 18.51±2.87, 18.25±2.73, 17.78±2.98, and 14.26±0.92 mL/breath, respectively, with no significant differences among the first three types but all significantly higher than that of the mask valve (all P<0.0001, rate of difference > 10%). Except with the V60 ventilator, some ventilator performance parameters (triggering and control performance) were significantly lower for the plateau valve than for the others, the rate of difference in tidal volume (VT) between the ventilator and the simulated lung exceeded 10% for all exhalation valves (all P<0.01).: Mask valves showed significantly lower CO2 rebreathing than circuit-located valves (single-arch, whisper swivel, and plateau exhalation) in this NIV bench study. The different valves influenced ventilator performance differently, particularly the plateau valve. These findings necessitate further clinical validation in vivo.

Keywords: Noninvasive Ventilation, exhalation valves, CO2 rebreathing, Ventilator performance, Hypercapnia

Received: 02 Dec 2024; Accepted: 04 Jul 2025.

Copyright: © 2025 Li, Bing, Zhou, Hou, Wang, Li, Zhao, Wang and Tan. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

* Correspondence: Wei Tan, Department of Pulmonary and Critical Care Medicine, The First Affiliated Hospital of China Medical University, Shenyang, China

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.