- 1ShenZhen Core Medical Technology CO, LTD, Shenzhen, China
- 2Harbin Institute of Technology, Harbin, China
- 3Chinese Academy of Medical Sciences and Peking Union Medical College Fuwai Hospital, Beijing, China
Background: Continuous-flow ventricular assist devices (VADs) have been widely adopted in clinical practice for the treatment of heart failure, but the effect of their non-pulsatile blood flow on microvascular circulation is still debated. Although VADs with a flatter H-Q curve are known to produce greater pulse pressure (PP), other hemodynamic performances have not been systematically compared and analyzed.
Methods: This study employed a lumped-parameter cardiopulmonary circulation numerical model to compare the hemodynamic responses of two continuous-flow centrifugal pumps: the Corheart 6 (flatter H-Q curve) and the HeartMate 3 (steeper H-Q curve). Comparisons were conducted across four distinct clinical scenarios: left heart failure, right heart failure, myocardial recovery and acute preload shifts. A quantitative assessment focused on arterial PP, peripheral organ perfusion, ventricular unloading, pump suction risk, and pump thrombosis risk.
Results: At the same average pump flow, pumps with a flatter H-Q curve, because of their higher sensitivity to preload, generated higher pump flow pulsatility and greater arterial PP, thereby creating hemodynamic conditions that may theoretically reduce risks associated with flow stasis. However, their ventricular unloading and peripheral organ perfusion were slightly inferior. When pump speed was increased, these pumps achieved ventricular unloading and peripheral organ perfusion comparable to those with steeper H-Q curves while simultaneously yielding even higher arterial PP. In contrast to the static condition, during dynamic events such as acute preload reduction caused by postural changes, VADs with a flatter H-Q curve are better able to maintain systemic perfusion pressure. When applied in right heart failure, right atrium implantation yields superior right ventricular unloading but lower pump flow pulsatility of both pumps.
Conclusion: The findings provide references for VAD developers and clinicians for the optimal design and utilization of blood pumps with different H-Q characteristics.
Introduction
The earliest ventricular assist devices (VADs) employed a pulsatile flow mechanism, mimicking physiological cardiac contraction and relaxation by synchronizing periodic volume changes with heartbeats. However, their substantial size and low reliability limited widespread clinical adoption (Feller et al., 2007; Garatti et al., 2008; Kamdar et al., 2009). In contrast, continuous-flow VADs have become the primary choice for long-term circulatory support due to their smaller size, higher reliability, and fewer adverse events (Caccamo et al., 2011; Teutebe et al., 2020; Cusimano et al., 2021). Nevertheless, some studies have pointed out that their non-pulsatile flow can have detrimental effects on systemic and microvascular circulation, including inhibition of endothelial nitric oxide synthase, which can lead to vascular endothelial dysfunction and increased arterial stiffness (Bartoli et al., 2010; Khambadkone et al., 2003). Under prolonged continuous-flow support, this could elevate the risk of complications such as arteriovenous malformations and gastrointestinal bleeding.
To better approximate normal physiological characteristics, certain continuous-flow VADs have been designed with flatter H-Q curves (Fang P et al., 2023; Motomura et al., 2020), enabling them to generate greater pump flow pulsatility and pulse pressure (PP). Bartoil et al. investigated the hemodynamic response to continuous-flow (CF) and pulsatile-flow (PF) pumps in animal models and demonstrated that CF pumps impair the physiological pulsatility of hemodynamics during ventricular unloading, whereas PF pumps maintain more normative physiological values (Bartoli et al., 2010). The literature has established that reduced physiological pulsatility adversely affects both the macro- and micro-circulation, potentially causing vascular stiffening due to the suppression of endothelial nitric oxide synthase activity (Khambadkone et al., 2003). Several studies have investigated the impact of H-Q curve slope in continuous-flow VADs. Sénage et al. (2014) used an experimental model to simulate and compare the hemodynamic characteristics of an axial-flow pump (HeartMate II, HM II) and a centrifugal pump (VentrAssist, VTA). Their findings showed that, under the same total cardiac output (CO), support with HM II resulted in higher mean arterial pressure, lower left atrial pressure, and higher right atrial pressure, but also carried a higher risk of ventricular suction, which is consistent with the conclusion of study (Giridharan et al., 2015). Eleuteri et al. (2012) investigated differences in myocardial reverse remodeling between axial-flow and centrifugal pumps by measuring changes in left ventricular end-diastolic diameter (LVEDD) before and after implantation. The results indicated that patients supported by axial-flow pumps had smaller LVEDD and greater reductions in the biomarker associated with reverse remodeling. Graefe et al. (2019) found that pumps with flatter H-Q curves achieved higher peak flow rates and thus providing better ventricular unloading during exercise. Telyshev et al. (2019) used a lumped parameter model to compare the hemodynamic effects between VADs. At the same average pump flow, HeartWare demonstrated superior ventricular unloading, evidenced by a smaller ventricular stroke volume. This result differed from Sénage’s findings because HeartWare exhibited a flatter H-Q curve in low-flow regions and a steeper curve in high-flow regions compared to HM II.
These studies primarily focused on comparing the impacts of different H-Q curves between axial and centrifugal pumps on key hemodynamic parameters in patients with LHF. However, they did not extensively explore or quantitatively discuss the impact of parameters such as ventricular volume, ventricular pressure, arterial pressure, and pump flow on clinical safety (e.g., suction/thrombosis risk) and effectiveness (e.g., organ perfusion). In addition, VADs are also clinically used for treating right heart failure (RHF). Clinicians sometimes prefer right atrium (RA) implantation over right ventricle (RV) implantation, but clinical outcomes showed no significant differences in pump thrombosis rates or gastrointestinal bleeding (Maynes et al., 2020). However, a comparative analysis of the hemodynamics of these two implantation methods is lacking. Furthermore, the hemodynamic consequences of myocardial recovery and dynamic condition have not been systematically analyzed across different H-Q curve characteristics.
This paper aims to systematically compare and quantitatively evaluate the hemodynamic consequences of two representative, mainstream third-generation magnetic levitation centrifugal VADs—the Corheart 6 (flatter H-Q curve) and the HeartMate 3 (steeper H-Q curve)—across four key clinical application scenarios to inform optimal VAD design and utilization:
1. LHF with left ventricle (LV) implantation.
2. RHF with RA or RV implantation.
3. Myocardial recovery in both LHF and RHF scenarios.
4. Acute preload shifts.
Through a lumped parameter numerical hemodynamic model, this study will quantify the impact of different H-Q slopes on key clinical safety and efficacy indicators, including arterial pulse pressure (PPao), peripheral organ perfusion (MAP), ventricular unloading (ESP, EDV), pump suction risk (EDV, PIVAD), and pump thrombosis risk (PIVAD). This systematic and quantitative comparison is intended to fill the gaps in the existing literature and provide direct reference for current clinical practice.
Methods
Mathematical model of cardiovascular system
A numerical model of the cardiovascular circulation system was employed for simulation experiments, with its construction referencing existing literature (Liu et al., 2020; Korakianitis and Shi, 2006; Hall and Hall, 2011; Westerhof et al., 2009; Pant et al., 2018). These data and waveforms in our article are all consistent with hemodynamic information in the Textbook of Medical Physiology and Guyton and Hall Textbook of Medical Physiology (Liu et al., 2020). As shown in Figure 1, the model consists of the left heart, right heart, systemic circulation, and pulmonary circulation. The arterial system is modeled using the Windkessel model. In this model, all pressure variables are defined as transmural pressures (the difference between internal and external chamber pressures), and the potential physiological effects of pericardial constraint on the heart chambers are not considered.
Figure 1. Lumped parameter models of circulatory system with VAD implantation in different heart chambers. In the numerical model, Dmiv, Daov, Dtiv and Dpav represent mitral, aortic, tricuspid, and pulmonary valves, respectively; R, L and C denote resistance, inertance and compliance, P and Q denote pressure and flow rate, lv, la, rv, and ra denote left ventricle, left atrium, right ventricle, and right atrium; sas, sat, pas, and pat represent the aortic sinus, aorta, pulmonary sinus, and pulmonary artery, respectively; sar, scp, par, and pcp represent systemic arteries, systemic capillaries, pulmonary arterioles, and pulmonary capillaries; and svn and pvn denote systemic veins and pulmonary veins.
The heart is equivalently represented as a chamber with time-varying elastance during the cardiac cycle (Suga and Sagawa, 1974; Karimov et al., 2020; Simaan et al., 2009; Stergiopulos et al., 1996). Ventricular elastance is defined as the ratio of ventricular pressure to volume, which is the reciprocal of capacitance:
In Equations 1, 2, Elv(t) and Erv(t) are the time-varying elastances of the LV and RV, and Clv(t) and Crv(t) are the time-varying compliances of the LV and RV. Plv(t) and Prv(t) and Vlv(t) and Vrv(t) are the pressures and volumes of the LV and RV, respectively. Vlv,0(t) and Vrv,0(t) are theoretical volumes of LV and RV at zero pressure, respectively.
In Equations 3, 4, Emax,lv and Emin,lv are left ventricular end-systolic elastance and left ventricular end-diastolic elastance, respectively, and Emax,rv and Emin,rv are right ventricular end-systolic elastance and right ventricular end-diastolic elastance, respectively. En,lv(tn), En,rv(tn) are normalized ‘double-Hill’ function expressions, which define the steepness and shape of the curve. The same normalized elastance formula is used for both ventricles since the activities of the LV and RV are synchronized.
Valve
The heart valves were simplified as ideal diodes in the model to ensure unidirectional blood flow (Korakianitis and Shi, 2006). While this approach is standard practice in lumped-parameter models, its primary limitation is the inability to simulate valve regurgitation. Nevertheless, this simplification remains effective for assessing the relative hemodynamic impact of VAD support on ventricular unloading. The opening and closing of the aortic valve is controlled by the pressure difference between the LV and the aorta:
where Qaov and Qmiv are the flow rates through the aortic and mitral valves, Plv, Psas, and Pla are the pressures of the LV, aortic sinus, and left atrium, respectively, and the valve opening parameters ARaov and ARmiv switch between 0 and 1:
Ventricular assist device
This study compared the hemodynamic effects of C69 and HM3 (Meissner et al., 2023; Abbott, 2025). As depicted in Figure 2, the C6’s H-Q curve consistently remains flatter than that of the HM3 across the entire operational flow range. The H-Q characteristic curve of the pump is expressed using the following equation. The fitting process was performed using the least squares method, and the coefficient of determination R2 was used to evaluate the goodness of fit to the original data:
Figure 2. H–Q curves of C6 and HM3 continuous-flow centrifugal pumps at different rotational speeds (operating speed ranges: 2200–4300 rpm for C6 (Fang P et al., 2023), 3000–9000 rpm for HM3 (Zayat et al., 2019)).
Where H is the pump head (mmHg), Q is the pump flow rate (L/min), w is the pump rotational speed (rpm), and a, b, and c are coefficients related to the rotary pump and obtained through least-squares fitting. Table 1 presents the fitting coefficients for the characteristic curves of the two pumps.
Simulation cases
The different simulation experiments were conducted in MATLAB Simulink. The ode45 solver was employed for solving the system of equations. The time step was set to 0.001 s. All simulation durations were set to 60 s with the system achieving a stable periodic solution after approximately 10 s. The stable results from the 58 s to the 60 s were selected for analysis across all simulated conditions.
Table 2 lists the key parameters used in the numerical model, all parameters of the hemodynamic model were set according to published literature (Liu et al., 2020; Korakianitis and Shi, 2006; Hall and Hall, 2011; Westerhof et al., 2009; Pant et al., 2018), and identifiability and robustness of the model has been examined, ensuring that the simulated hemodynamic results aligned with physiological characteristics. The R elements represent the microvascular resistance of the vascular system, and the C elements represent the elasticity or volume buffering capacity of the vascular walls. The combined action of the R and C elements constitutes the Windkessel effect, which determines the pulsatility of the arterial pressure waveform. The severity of heart failure (HF) was then modulated by adjusting the maximum ventricular elastance (Emax), with a lower elasticity indicating more severe HF. To isolate the hemodynamic effects of the different H-Q curve characteristics and avoid different volumetric support as a confounding factor, the RPM of both pumps were adjusted to achieve an identical mean pump flow.
Case 1: Left heart failure, VAD inflow cannula implanted in the left ventricle.
The numerical model is established by adjusting Emax,lv from a healthy value of 2.19 mmHg/mL to 1.25 mmHg/mL, with Emax,rv from a healthy value of 0.84 mmHg/mL to 0.55 mmHg/mL. According to the literature (Desai et al., 2011), mild RHF is a common complication of LHF.
Case 2: Right heart failure with healthy left heart, VAD inflow cannula individually implanted in the right ventricle and right atrium.
The numerical model adjusts Emax,rv from a healthy value of 0.84 mmHg/mL to 0.3 mmHg/mL.
Case 3: Myocardial recovery.
The elasticity of ventricular was increased in both Case 1 and Case 2. Specifically, Emax,lv was increased from 1.25 mmHg/mL to 2.0 mmHg/mL with the Emax,rv from 0.55 mmHg/mL to 0.84 mmHg/mL in Case 1 and Emax,rv was increased from 0.3 mmHg/mL to 0.6 mmHg/mL in Case 2, indicating a partial recovery of cardiac function.
Case 4: Acute Preload Change Simulating Postural Hypotension.
The systemic venous compliance Csvn was acutely increased from 20.5 mmHg·s/mL to 30 mmHg·s/mL. Based on previously published literature (Tyberg and Douglas, 1996), this systemic adjustment of the Csvn parameter serves to simulate the redistribution of blood volume between the central and peripheral circulations caused by postural changes (e.g., transition from supine to upright position), thereby inducing an acute change in cardiac preload.
The following variables were evaluated in these test cases:
LVESP, RVESP: Left, right ventricular end-systolic pressure, mmHg.
LVEDP, RVEDP: Left, right ventricular end-diastolic pressure, mmHg.
LVESV, RVESV: Left, right ventricular end-systolic volume, mL.
LVEDV, RVEDV: Left, right ventricular end-diastolic volume, mL.
P-V loop: Ventricular pressure-volume loop.
SPAP: Systolic pulmonary artery pressure, mmHg.
DPAP: Diastolic pulmonary artery pressure, mmHg.
SAP: Systolic aortic pressure, mmHg.
DAP: Diastolic aortic pressure, mmHg.
PPao: Aortic pulse pressure, mmHg, calculated as follows:
PPpa: Pulmonary aortic pulse pressure, mmHg, calculated as follows:
MAP: Mean aortic pressure, mmHg, calculated as follows:
MPAP: Mean pulmonary artery pressure, mmHg, calculated as follows:
CCP: Coronary perfusion pressure, mmHg, using pressure gradient to assess (Nguyen et al., 2018).
QAV: Aortic valve real-time flow, L/min.
QPV: Pulmonary valve real-time flow, L/min.
QAV,AVG: Mean aortic valve flow over one cardiac cycle, L/min, calculated as follows:
where T is the duration of cardiac cycle, s, QVAD is VAD real-time flow, L/min, and QVAD,AVG is mean pump flow over one cardiac cycle, L/min, calculated as follows:
PIVAD is pump flow pulsatility index, physiologically reflects the pulsatility of VAD blood flow and serve as a surrogate marker for native heart contribution and systemic microcirculatory health, calculated as follows:
CO is total cardiac output, L/min, calculated as follows:
Results
Figure 3. LHF, VAD implantation in the ventricle, hemodynamic characteristics under pump support and baseline conditions: (a) Left ventricular pressure-volume loop; (b) Right ventricular pressure-volume loop; (c) Aortic pressure curve; (d) Pulmonary artery pressure curve; (e) Pump flow curve; (f) Left ventricular pressure curve; LHF represents left heart failure baseline condition without VAD support; C6 represents the condition under Corheart 6 support; HM3 represents the condition under HeartMate 3 support.
Case 2. Right heart failure with healthy left heart, VADs individually implanted in the right atrium or right ventricle.
Table 4. Hemodynamic and pump parameters: Baseline and post-RVAD (RA/RV) implantation in RHF with healthy left heart.
Figure 4. RHF with RA/RV VAD implantation, hemodynamic characteristics under pumps support and baseline conditions. (a) left ventricular pressure-volume loop; (b) Right ventricular pressure-volume loop; (c) Aortic pressure curve; (d) Pulmonary artery pressure curve; (e) Pump flow curve; RHF represents RHF baseline condition without VAD support; C6-RA/RV represents the RA/RV under Corheart 6 support; HM3-RA/RV represents the RA/RV under HeartMate 3 support.
Figure 5. Comparison of hemodynamics between C6 and HM3 under acute preload change due to postural shift.
Discussion
Under the same average pump flow, C6, with its flatter H-Q curve, demonstrates greater sensitivity to changes in preload, generating stronger pump flow pulsatility and higher arterial PP(Pulse pressure). These hemodynamic characteristics may partially mitigate the adverse stimuli associated with continuous flow, which have been shown in previous studies to contribute to endothelial dysfunction. Moreover, under normal HR conditions (60–100 bpm), diastole is approximately twice as long as systole. Consequently, with the same average pump flow, the C6 directs more of its flow output toward systole compared to the HM3. As a result, the C6 shows slightly inferior performance in terms of peripheral organ perfusion (evaluated by MAP(Mean aortic pressure)) and ventricular unloading. However, when HR changes significantly, such as during exercise or arrhythmias, this difference may decrease or even be reversed. This is because the proportion of diastole in the cardiac cycle may drop significantly (potentially below 50%). Thus, the overall hemodynamic response primarily depends on the relative duration and actual pump flow within systole and diastole.
In RHF application scenario, a high pump flow can be generated even at low rotational speeds, requiring VAD design to avoid excessive pump flow at minimum operational speed. Further comparison of the same VAD implanted in RV or RA showed that RA implantation achieves better right ventricular unloading due to the higher diastole pump flow. However, compared to RV implantation, RA implantation results in slightly lower PPpa and pump flow pulsatility.
Regardless of whether in LHF or RHF, during myocardial recovery scenario, total CO increases despite the pump flow decreases, indicating better organ perfusion. Therefore, in clinical practice, it is essential to monitor the cardiac function of patients receiving VAD support, particularly in cases of RHF. On one hand, PAP elevation must be monitored to prevent potential complications; on the other hand, as right heart function improves and pump flow naturally decreases, anticoagulation strategies should be promptly adjusted to prevent pump thrombosis.
To mitigate hemocompatibility-related complications, numerous studies recommend selecting pumps with flatter H-Q curves to achieve higher PP. Moreover, the enhanced flow pulsatility characteristic—manifested as lower flow during diastole and higher flow during systole—contributes to a reduced risk of suction events and pump thrombosis. A lower thrombosis risk, in turn, allows for reduced anticoagulant dosage, thereby further decreasing non-surgical bleeding events.
Moreover, pumps with flatter H-Q curves exhibit heightened sensitivity to variations in both pre- and after-loads. As a result, they demonstrate slightly inferior ventricular unloading and lower MAP, and more prone to backflow when the pressure difference across the pump increases. Although researches have not yet precisely defined the exact effects of this backflow, careful control of the patient’s blood pressure is recommended when utilizing such pumps. However, these limitations can be mitigated by increasing the pump speed to achieve a higher average flow.
Table 5, Figure 6 illustrates an example from Case 1 where C6 pump speed was increased by 40 rpm. This adjustment resulted in MAP and LVEDV values equivalent to those achieved by HM3. Notably, the PPao remained higher compared to the HM3, while LVESP and LVESV were lower. This demonstrates that the C6, while maintaining greater pulsatility, can achieve organ perfusion and ventricular unloading comparable to pumps with steeper H-Q curves.
Table 5. Hemodynamic and pump parameters of C6 and HM3 under LHF with healthy right heart when MAP is equal.
Figure 6. Comparison of hemodynamics between C6 and HM3 under LHF with healthy right heart, at equal MAP, but different average pump flow.
It should be noted that the relationship between speed elevation and hemodynamic improvement is not linear, as shown in Figure 7. Regardless of the H-Q curve type, the same magnitude of speed adjustment yields varying effects on hemodynamic parameters, depending on the pre-adjustment operating speed.
Figure 7. Left ventricular P-V loop and aortic pressure of C6 and HM3 at different speeds under LHF baseline.
A notable and specific dynamic scenario is the acute preload change caused by postural variation. In this situation, pumps with a flatter H-Q curve demonstrate the capability to generate a greater diastolic flow, thereby maintaining effective perfusion more efficiently during abrupt changes in preload. However, this characteristic simultaneously implies a higher risk of suction. Other dynamic clinical conditions, including physical exercise, acute pain, and transient arrhythmias, may likewise induce abrupt alterations in preload and/or afterload. These conditions will be systematically investigated in future work.
Our findings regarding the pulsatility and unloading performance associated with the H-Q curve are consistent in trend with prior in vitro and computational analyses of earlier VAD devices. However, the contribution of this study extends beyond these qualitative comparisons by providing the first systematic quantitative hemodynamic analysis of the two mainstream third-generation magnetic levitation centrifugal pumps (Corheart 6 and HeartMate 3). Furthermore, this work addresses critical gaps in the existing literature through detailed analyses of clinically relevant scenarios that have previously lacked systematic comparative assessment, including the systematic comparison of RA versus RV implantation modes in RHF scenarios, and the assessment of VAD performance during the myocardial recovery stage. Finally, by quantifying the hemodynamic response of both designs under acute transient conditions (e.g., acute preload shifts), we enhance the understanding of the H-Q curve’s influence on clinical applications.
Limitation
This study has several limitations, which may lead to discrepancies between simulation results and actual clinical outcomes. Firstly, this model does not incorporate any physiological autoregulatory mechanisms, such as the baroreflex, which may affect the absolute predicted values of mean aortic pressure and pulmonary pressure, particularly in highly dynamic scenarios like the myocardial recovery phase. Our analysis therefore focuses on the relative differences and trends in performance between the Corheart 6 and HeartMate 3 VADs under the controlled, fixed baseline conditions, rather than aiming for perfect predictive accuracy of the absolute systemic pressures. Secondly, the current model simplifies the complex fluid dynamics within the pump and the ventricle. The pulsatility index of VAD support (PIVAD) is used as a surrogate for flow-related thrombosis. While effective for comparative purposes, this do not fully capture the detailed shear stress and flow separation effects that contribute to thrombosis and hemolysis in vivo. In future work, we will incorporate physiological feedback model and couple the lumped parameter model with detailed Computational Fluid Dynamics (CFD) simulations in the pump and cardiac chambers. Furthermore, a significant limitation of our HM3 simulation is the exclusion of artificial pulse. The HM3’s artificial pulse is non-synchronous, and the resultant hemodynamic effect would be highly dependent on the heart rate baseline, a complex dynamic variable whose detailed interaction with the control algorithm is not the central focus of this paper.
Conclusion
This study employed a numerical model of the circulatory system to investigate the hemodynamic characteristics of VADs with varying H-Q curve slopes across four distinct clinical scenarios. The findings indicate that under conditions of equal average pump flow, pumps with flatter H-Q curve exhibit lower diastolic pump flow and higher systolic pump flow. This generates greater pump flow pulsatility and a higher arterial PP, which can help reduce the risk of vascular malformations, as well as pump suction and pump thrombosis. However, their overall ventricular unloading and organ perfusion is slightly diminished. From a computational perspective, the results suggest that when utilizing pumps with flatter H-Q curves, maintaining a slightly higher average pump flow target may be beneficial to ensure adequate organ perfusion and ventricular unloading while preserving pulsatility.
In contrast to the static condition, during dynamic events such as acute preload reduction caused by postural changes, pumps with flatter H-Q curve exhibit improved preservation of systemic perfusion pressure, albeit at the expense of a greater reduction in LVEDV.
When applying the two types of H-Q curve pumps in RHF scenarios, the characteristic hemodynamic changes observed with RV implantation are analogous to those seen with LV implantation. Conversely, with RA implantation, both pump types exhibit similar hemodynamic performance due to the minimal preload variations. Comparing the results of implanting the same VAD in RA versus RV, RA implantation yields superior right ventricular unloading. However, this comes at the cost of lower pulmonary artery PP and reduced pump flow pulsatility. Furthermore, as cardiac function recovers from HF, an increase in PAP and a decrease in pump flow are observed. Consequently, meticulous attention to anticoagulation management is crucial.
Data availability statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.
Author contributions
YY: Writing – review and editing, Conceptualization, Writing – original draft. ZW: Writing – original draft, Data curation. ZX: Writing – original draft, Data curation. SY: Writing – original draft, Writing – review and editing, Conceptualization. LZ: Conceptualization, Writing – review and editing, Writing – original draft.
Funding
The author(s) declared that financial support was received for this work and/or its publication. The study was supported by funding from the Key Technology Research and Development Project for the Miniaturized Centrifugal Magnetic Levitation Artificial Heart, with project number KOTD20210811090222040. The project is part of the Shenzhen High-Level Talent Innovation and Entrepreneurship Program, which is granted by Shenzhen Municipal Bureau of Science, Technology and Innovation.
Conflict of interest
Authors YY, ZX, and SY were employed by ShenZhen Core Medical Technology CO, LTD.
The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
Abbott (2025). HeartMate 3™ left ventricular assist system instructions for use. Available online at: https://manuals.eifu.abbott.
Bartoli C. R., Giridharan G. A., Litwak K. N., Sobieski M., Prabhu S. D., Slaughter M. S., et al. (2010). Hemodynamic responses to continuous versus pulsatile mechanical unloading of the failing left ventricle. ASAIO J. 56 (5), 410–416. doi:10.1097/MAT.0b013e3181e7bf3c
Caccamo M., Eckman P., John R. (2011). Current state of ventricular assist devices. Curr. Heart Fail Rep. 8 (2), 91–98. doi:10.1007/s11897-011-0050-z
Cusimano V., Di Molfetta A., Ferrari G. (2021). Designing an active valvulated outflow conduit for a continuous-flow left ventricular assist device to increase pulsatility: a simulation study. ASAIO J. 67 (5), 529–535. doi:10.1097/MAT.0000000000001255
Desai R. V., Meyer P., Ahmed M. I., Mujib M., Adamopoulos C., White M., et al. (2011). Relationship between left and right ventricular ejection fractions in chronic advanced systolic heart failure: insights from the BEST trial. Eur. J. Heart Fail 13 (4), 392–397. doi:10.1093/eurjhf/hfq206
Eleuteri K. L., Soleimani B., Brehm C., Stephenson E., Pae W., el-Banayosy A. (2012). 237. reverse remodeling (RR) following rotary blood pump implantation: is there a difference between axial (AFP) and centrifugal flow pumps (CFP)? J. Heart Lung Transpl. 31, S86. doi:10.1016/j.healun.2012.01.244
Fang P., Du J., Boraschi A., Bozzi S., Redaelli A., Schmid Daners M., et al. (2022). Insights into the low rate of in-pump thrombosis with the HeartMate 3: does the artificial pulse improve washout? Front. Cardiovasc Med. 9, 775780. doi:10.3389/fcvm.2022.775780
Fang P., Yang Y., Wei X., Yu S. (2023). Preclinical evaluation of the fluid dynamics and hemocompatibility of the corheart 6 left ventricular assist device. Artif. Organs 47 (6), 951–960. doi:10.1111/aor.14498
Feller E. D., Sorensen E. N., Haddad M., Pierson R. N., Johnson F. L., Brown J. M., et al. (2007). Clinical outcomes are similar in pulsatile and nonpulsatile left ventricular assist device recipients. Ann. Thorac. Surg. 83, 1082–1088. doi:10.1016/j.athoracsur.2006.10.034
Garatti A., Bruschi G., Colombo T., Russo C., Lanfranconi M., Milazzo F., et al. (2008). Clinical outcome and bridge to transplant rate of left ventricular assist device recipient patients: comparison between continuous-flow and pulsatile-flow devices. Eur. J. Cardiothorac. Surg. 34, 275–280. doi:10.1016/j.ejcts.2008.02.019
Giridharan G. A., Koenig S. C., Soucy K. G., Choi Y., Pirbodaghi T., Bartoli C. R., et al. (2015). Left ventricular volume unloading with axial and centrifugal rotary blood pumps. Asaio J. 61 (3), 292–300. doi:10.1097/MAT.0000000000000201
Graefe R., Henseler A., Körfer R., Meyns B., Fresiello L. (2019). Influence of left ventricular assist device pressure-flow characteristic on exercise physiology: assessment with a verified numerical model. Int. J. Artif. Organs 42 (9), 490–499. doi:10.1177/0391398819846126
Hall J. E., Hall M. E. (2011). Guyton and hall textbook of medical physiology. Philadelphia, PA: Elsevier.
Kamdar F., Boyle A., Liao K., Colvin-adams M., Joyce L., John R. (2009). Effects of centrifugal, axial, and pulsatile left ventricular assist device support on end-organ function in heart failure patients. J. Heart Lung Transpl. 28, 352–359. doi:10.1016/j.healun.2009.01.005
J. H. Karimov, R. C. Starling, and K. Fukamachi (2020). Mechanical support for heart failure: current solutions and new technologies. 1st ed. (Springer).
Khambadkone S., Li J., de-Leval M. R., Cullen S., Deanfield J. E., Redington A. N. (2003). Basal pulmonary vascular resistance and nitric oxide responsiveness late after Fontan-type operation. Circulation 107 (25), 3204–3208. doi:10.1161/01.CIR.0000074210.49434.40
Korakianitis T., Shi Y. (2006). Numerical simulation of cardiovascular dynamics with healthy and diseased heart valves. J. Biomech. 39 (11), 1964–1982. doi:10.1016/j.jbiomech.2005.06.016
Liu H., Liu S., Ma X., Zhang Y. (2020). A numerical model applied to the simulation of cardiovascular hemodynamics and operating condition of continuous-flow left ventricular assist device. Math. Biosci. Eng. 17 (6), 7519–7543. doi:10.3934/mbe.2020384
Maynes E. J., O'Malley T. J., Patel P., Prochno K. W., Wood C. T., Choi J. H., et al. (2020). Right atrial versus right ventricular HeartWare HVAD position in patients on biventricular HeartWare HVAD support: a systematic review. Artif. Organs 44 (9), 926–934. doi:10.1111/aor.13675
Meissner F., Eichelkraut D., Schimmel M., Maier S., Vestner H., Schoen M., et al. (2023). Impact of an accessory for left ventricular assist devices on device flow and pressure head in vitro. Bioengineering 10, 486. doi:10.3390/bioengineering10040486
Montalto A., Piazza V., Albi F., Gherli R., Contento C., Palermo A., et al. (2019). Pump speed optimization in patients implanted with the HeartMate 3 device. Transplant. Proc. 51, 206–209. doi:10.1016/j.transproceed.2018.02.210
Motomura T., Karimov J., Fukamachi K. (2020) “EVAHEART 2 left ventricular assist system: a hemocompatible centrifugal pump with physiological pulsatility,” in Mechanical support for heart failure. Springer, 535–544.
Nascimbene A., Bark D., Smadja D. M. (2024). Hemocompatibility and biophysical interface of left ventricular assist devices and total artificial hearts. Blood 143 (8), 661–672. doi:10.1182/blood.2022018096
Nguyen T., Do H., Pham T., Vu L. T., Zuin M., Rigatelli G. (2018). Left ventricular dysfunction causing ischemia in patients with patent coronary arteries. Perfusion 33 (2), 115–122. doi:10.1177/0267659117727826
Pant S., Corsini C., Baker C., Hsia T. Y., Pennati G., Vignon-Clementel I. E. (2018). A lumped parameter model to study atrioventricular valve regurgitation in stage 1 and changes across stage 2 surgery in single ventricle patients. IEEE Trans. Biomed. Eng. 65, 2450–2458. doi:10.1109/TBME.2018.2797999
Sénage T., Février D., Michel M., Pichot E., Duveau D., Tsui S., et al. (2014). A mock circulatory system to assess the performance of continuous-flow left ventricular assist devices (LVADs): does axial flow unload better than centrifugal LVAD? ASAIO J. 60 (2), 140–147. doi:10.1097/MAT.0000000000000045
Simaan M. A., Ferreira A., Chen S. H., Antaki J. F., Galati D. G. (2009). A dynamical state space representation and performance analysis of a feedback-controlled rotary left ventricular assist device. IEEE Trans. Control Syst. Technol. 17, 15–28. doi:10.1109/tcst.2008.912123
Stergiopulos N., Meister J. J., Westerhof N. (1996). Determinants of stroke volume and systolic and diastolic aortic pressure. Am. J. Physiol. 270 (6 Pt 2), H2050–H2059. doi:10.1152/ajpheart.1996.270.6.H2050
Suga H., Sagawa K. (1974). Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle. Circ. Res. 35, 117–126. doi:10.1161/01.res.35.1.117
Telyshev D., Petukhov D., Selishchev S. (2019). Numerical modeling of continuous-flow left ventricular assist device performance. Int. J. Artif. Organs 42 (11), 611–620. doi:10.1177/0391398819852365
Teuteberg J. J., Cleveland J. C., Jr, Cowger J., Higgins R. S., Goldstein D. J., Keebler M., et al. (2020). The society of thoracic surgeons INTERMACS 2019 annual report: the changing landscape of devices and indications. Ann. Thorac. Surg. 109 (3), 649–660. doi:10.1016/j.athoracsur.2019.12.005
Tyberg J. V., Douglas R. H. (1996). Orthostatic hypotension and the role of changes in venous capacitance. Med. Science Sports Exercise 28 (Suppl. l), S29–S31. doi:10.1097/00005768-199610000-00029
Uriel N., Adatya S., Malý J., Kruse E., Rodgers D., Heatley G., et al. (2017). Clinical hemodynamic evaluation of patients implanted with a fully magnetically levitated left ventricular assist device (heartMate 3). J. Heart Lung Transplant. 36 (1), 28–35. doi:10.1016/j.healun.2016.07.008
Westerhof N., Lankhaar J. W., Westerhof B. E. (2009). The arterial windkessel. Med. Biol. Eng. Comput. 47 (2), 131–141. doi:10.1007/s11517-008-0359-2
Keywords: heart FaiIure, hemodynamic, H-Q curve, lumped parameter model, ventricle assist device
Citation: Yang Y, Wang Z, Xie Z, Yu S and Zou L (2026) Hemodynamic consequences of pressure-flow curve gradient variations in continuous-flow ventricular assist devices. Front. Physiol. 16:1730883. doi: 10.3389/fphys.2025.1730883
Received: 23 October 2025; Accepted: 26 December 2025;
Published: 12 January 2026.
Edited by:
Ahsan H Khandoker, Khalifa University, United Arab EmiratesReviewed by:
Emanuele Gasparotti, Gabriele Monasterio Tuscany Foundation (CNR), ItalyMauricio Felippi de Sá Marchi, Estácio de Sá University, Brazil
Copyright © 2026 Yang, Wang, Xie, Yu and Zou. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: ShunzhouYu, eXVzaHVuemhvdUBjb3JldGVjaG1lZC5jb20=; Liang Zou, em91bGlhbmc0ODM2NzQ0QDE2My5jb20=
†ORCID: Yuzhuo Yang, orcid.org/0000-0003-3844-5518; Zhenyu Wang, orcid.org/0009-0009-0259-9591; Zipeng Xie, orcid.org/0009-0001-1424-7456; Liang Zou, orcid.org/0009-0005-9739-074X
‡These authors have contributed equally to this work
Yuzhuo Yang1†‡