Echo time-dependent observed T1 and quantitative perfusion in chronic obstructive pulmonary disease using magnetic resonance imaging

Simon M. F. Triphan1,2* Marilisa Konietzke1,2,3 Jürgen Biederer1,2,4,5 Monika Eichinger1,2,6 Claus F. Vogelmeier7 Rudolf A. Jörres8 Hans-Ulrich Kauczor1,2,6 Claus P. Heußel1,6 Bertram J. Jobst1,2,6 and Mark O. Wielpütz1,2,6 on behalf of the COSYCONET study group
  • 1Department of Diagnostic and Interventional Radiology, Subdivision of Pulmonary Imaging, University Hospital of Heidelberg, Heidelberg, Germany
  • 2Translational Lung Research Center Heidelberg (TLRC), Member of the German Center for Lung Research (DZL), Heidelberg, Germany
  • 3Boehringer Ingelheim Pharma GmbH and Co. KG, Biberach an der Riß, Germany
  • 4Faculty of Medicine, University of Latvia, Riga, Latvia
  • 5Faculty of Medicine, Christian-Albrechts-Universität zu Kiel, Kiel, Germany
  • 6Department of Diagnostic and Interventional Radiology With Nuclear Medicine, Thoraxklinik at the University Hospital of Heidelberg, Heidelberg, Germany
  • 7Department of Medicine, Pulmonary and Critical Care Medicine, Philipps-University of Marburg (UMR), Member of the German Center for Lung Research (DZL), Marburg, Germany
  • 8Institute and Outpatient Clinic for Occupational, Social and Environmental Medicine, Comprehensive Pneumology Center Munich (CPC-M), Member of the German Center for Lung Research (DZL), University Hospital, Ludwig Maximilians University (LMU) Munich, Munich, Germany

Introduction: Due to hypoxic vasoconstriction, perfusion is interesting in the lungs. Magnetic Resonance Imaging (MRI) perfusion imaging based on Dynamic Contrast Enhancement (DCE) has been demonstrated in patients with Chronic Obstructive Pulmonary Diseases (COPD) using visual scores, and quantification methods were recently developed further. Inter-patient correlations of echo time-dependent observed T1 [T1(TE)] have been shown with perfusion scores, pulmonary function testing, and quantitative computed tomography. Here, we examined T1(TE) quantification and quantitative perfusion MRI together and investigated both inter-patient and local correlations between T1(TE) and quantitative perfusion.

Methods: 22 patients (age 68.0 ± 6.2) with COPD were examined using morphological MRI, inversion recovery multi-echo 2D ultra-short TE (UTE) in 1–2 slices for T1(TE) mapping, and 4D Time-resolved angiography With Stochastic Trajectories (TWIST) for DCE. T1(TE) maps were calculated from 2D UTE at five TEs from 70 to 2,300 μs. Pulmonary Blood Flow (PBF) and perfusion defect (QDP) maps were produced from DCE measurements. Lungs were automatically segmented on UTE images and morphological MRI and these segmentations registered to DCE images. DCE images were separately registered to UTE in corresponding slices and divided into corresponding subdivisions. Spearman’s correlation coefficients were calculated for inter-patient correlations using the entire segmented slices and for local correlations separately using registered images and subdivisions for each TE. Median T1(TE) in normal and defect areas according to QDP maps were compared.

Results: Inter-patient correlations were strongest on average at TE2 = 500 μs, reaching up to |ρ| = 0.64 for T1 with PBF and |ρ| = 0.76 with QDP. Generally, local correlations of T1 with PBF were weaker at TE2 than at TE1 or TE3 and with maximum values of |ρ| = 0.66 (from registration) and |ρ| = 0.69 (from subdivision). In 18 patients, T1 was shorter in defect areas than in normal areas, with the relative difference smallest at TE2.

Discussion: The inter-patient correlations of T1 with PBF and QDP found show similar strength and TE-dependence as those previously reported for visual perfusion scores and quantitative computed tomography. The local correlations and median T1 suggest that not only base T1 but also the TE-dependence of observed T1 in normal areas is closer to that found previously in healthy volunteers than in defect areas.

1 Introduction

Monitoring disease progression and therapy response in chronic obstructive pulmonary disease (COPD) is commonly based on pulmonary function testing (PFT), which provides global measures for lung function impairment (1). Computed tomography (CT) is used for the regional assessment of lung structure and distinction of COPD phenotypes (2). Magnetic resonance imaging (MRI) of the lungs is particularly challenging due to the low proton density in lung parenchyma, respiratory motion and especially the extremely short T2* relaxation time caused by the structure of the lungs (3). MRI was suggested as a radiation-free, non-invasive alternative to PFT and CT and has the additional benefit of techniques for the regional assessment of functional impairment. MRI methods based on image registration and the oscillation of signal intensity allow for the quantification of pulmonary perfusion and ventilation (4, 5). However, arguably the most practical and established approach for functional lung MRI in clinical routine is dynamic contrast enhanced (DCE) MRI. In the lungs, local perfusion is regulated by hypoxic pulmonary vasoconstriction which links perfusion directly to ventilation. Accordingly, the detection of perfusion defects can thus be used to identify abnormal ventilation as long as ventilation/perfusion mismatches can be disregarded. Reduction of the pulmonary vascular bed in emphysema also results in perfusion defects. A visual scoring system for DCE-MRI by radiologist readers has been demonstrated to provide a measure of functional lung abnormality that correlates well with other clinical metrics, including lung function tests in COPD and cystic fibrosis (610).

The determination of this visual score requires trained radiologists and is simultaneously limited in spatial resolution to the level of lung lobes. It is also severely limited in numerical resolution since scores cannot practically have fine graduations and it also essentially disregards the difference in volume between lung lobes. Quantification of perfusion from DCE-MRI data has been established in other organs and explored in the lungs as well (11, 12). For perfusion quantification, both the calculation of perfusion metrics like pulmonary blood flow (PBF), pulmonary blood volume (PBV) and mean transit time (MTT) as well as the determination of the extent of defects have been shown (13). While these have been examined mainly as global metrics for the entire lung, especially considering the fraction of defects as perfusion defects in percent (QDP), they are also available locally for each image voxel (14).

T1 quantification in the lungs was originally pursued in the context of oxygen-enhanced imaging (15). In these studies, significantly lower baseline lung T1 was found in patients with COPD even without oxygen supplementation, and median T1 was found to correlate with the visual MRI perfusion score (8). In subsequent work, it was further demonstrated that observed lung T1 [T1(TE)] is dependent on the echo time (TE) of the measurement. It was subsequently assumed that T1(TE) reflects different compartments of lung tissue protons inside each imaging voxel (protons in blood and extravascular protons), since the inter-patient correlations between T1(TE) and the MRI morphology and MRI perfusion score as well as other metrics vary with TE in COPD and cystic fibrosis (1618). It was hypothesized that T1(TE) has the potential to differentiate tissue abnormalities such as inflammation or emphysema from perfusion abnormalities on the level of image voxels without the need for contrast material. Previous studies have so far examined the relationship between T1(TE) and DCE-MRI perfusion using only visual MRI perfusion scores on the level of the entire lungs for median T1, and on the level of lobes for visually scored T1 (8, 17, 18). The present work was conducted to further investigate the relationship of T1(TE) with quantitative perfusion metrics on a local level including voxel-wise correlations.

2 Methods

2.1 Patient characteristics

The study included participants of a single study center within the German multicenter COPD cohort study COSYCONET (“COPD and SYstemic consequences-COmorbidities NETwork,” ClinicalTrials.gov Identifier: NCT01245933) (19). The overarching COSYCONET study as well as the part of the imaging-based sub-study [“Image-based structural and functional phenotyping of the COSYCONET cohort using MRI and CT (MR-COPD),” ClinicalTrials.gov Identifier: NCT02629432] described here were approved by the responsible ethics committees of the coordinating centers [Institutional Review Board of the Medical Faculty of the University of University of Marburg (200/09) and of the University of Heidelberg (S-656/2012), Germany]. The participants of the present sub-study gave their written informed consent to undergo extensive clinical assessment including lung function testing, non-contrast CT, and morpho-functional MRI. The patient characteristics are summarized in Table 1. Some patients were included in separate previous reports, which did not assess the analyses made in the present study (13, 18).

TABLE 1
www.frontiersin.org

Table 1. Patient characteristics.

2.2 Magnetic resonance imaging acquisition

All measurements were performed on the same 1.5 T scanner (Magnetom Aera, Siemens Healthineers, Erlangen, Germany). A standardized protocol was employed including T1- and T2-weighted morphological MRI sequences. Of those, the T1-weighted morphological Volumetric Interpolated Breath-hold Examination (VIBE) measurements were used for segmentation. Imaging parameters for those were: Transverse VIBE: TR = 2.61 ms, TE = 1.69 ms, Field of View (FoV) 400 × 300 × 4 mm3, matrix size 320 × 240 × 88. Coronal VIBE: TR = 3.35 ms, TE = 1.63 ms, FoV 400 × 400 × 4 mm3, matrix size 288 × 288 × 56. A keyhole-based 3D gradient echo sequence [Time-resolved angiography With Stochastic Trajectories (TWIST), Siemens Healthineers, Erlangen, Germany], was used for DCE perfusion imaging, resulting in 1.7 s effective time resolution over 20 3D image timepoints. Imaging parameters were: up to 56 slices at 5 mm thickness, matrix size 208 × 256, FoV 366 × 450 mm2, TE 0.76 ms, and flip angle 20°. DCE measurements utilized intravenous contrast injection of 2 mL Gadobutrol 1 mmoL/mL (Gadovist, Bayer Vital, Leverkusen, Germany) injected at a rate of 4 mL/s, followed by a 30 mL 0.9% NaCl chaser. Both VIBE and TWIST were acquired in inspiratory breath-holds.

T1(TE) mapping was based on an inversion recovery multi-echo 2D ultra-short TE sequence, applied directly before contrast injection (20). Echos at TE1-5 = 70, 500, 1,200, 1,650, and 2,300 μs were acquired split into two separate measurements with TE1, TE3, and TE5 in the first and TE2 and TE4 in the second measurement. Imaging parameters were: One or two coronal slices at 15 mm thickness, matrix size 128 × 128, FoV 500 × 500 mm2, TR = 5 ms, flip angle 6°, and total acquisition time 4 min (1 slice) or 8 min (2 slices). To cover as much lung parenchyma as possible, the first slice was placed through the descending aorta and the second slice 30 mm ventral. For each measurement, a total of 6,000 radial spokes in a golden angle distribution were acquired, divided into blocks starting with an adiabatic inversion pulse followed by 300 radial spokes and separated by 3 s delays. These were then sorted by inversion time (TI), split using a sliding window 120 spokes wide with 60 spokes step-width and reconstructed using a non-uniform Fourier transform implemented in MATLAB (Mathworks, Natick, United States). To achieve ultra-short echo times, pairs of half sinc-pulses with opposing slice-selection gradients, 1,250 μs pulse duration and time-bandwidth-product 2 were employed. For each TE, T1 maps were calculated per voxel from the time-course along TI, based on the UTE images in each slice (21). UTE measurements were acquired during free breathing and gated to expiration using 50% of data, as described in Triphan et al. (20).

2.3 Magnetic resonance imaging analysis

Magnetic resonance images were segmented independently for UTE and DCE measurements: In UTE images, segmentation was performed on images at TE3 using simple region growing (for details, see the online supplement). Since the DCE images are three-dimensional and designed to maximize contrast between tissue with and without contrast agent, these were not segmented directly. Instead, a region growing-based segmentation algorithm was applied to morphological images with a higher resolution and the resulting segmentation masks were then registered to the DCE images, as described previously (13, 22).

For perfusion quantification, an arterial input function was detected automatically and then used to calculate a residual function R(t) by applying a voxel-wise deconvolution with the time-dependent signal in the DCE images (23). Pulmonary blood flow (PBF) maps were derived from R(t) by determining the maximum along the time-course in each individual voxel (24). To calculate defect classification maps and QDP, the following steps were performed: The timepoint of maximum contrast enhancement in the entire lungs (tmax) was determined by taking the mean of R(t) in the entire lung for every timepoint. Otsu’s method was then applied to the values of R(tmax) within the lungs to determine two thresholds (25). Defect classification maps were produced by classifying all voxels below the lower threshold as perfusion defects and the remainder as normal (13). The fraction of defect voxels within the segmented lungs on these maps relative to the total number of segmented voxels was defined as QDP. For all of the following steps, the three partitions of the 3D DCE images that correspond to the location of the UTE slice(s) were determined and averaged to produce a thicker projection image. Note that tmax and the classification thresholds were calculated based on the entire segmented lungs but QDP to be compared to T1(TE) only on these slices corresponding to UTE measurements.

To facilitate local (intra-patient) comparison of T1(TE) and quantitative perfusion maps, two approaches were investigated: (1) DCE images were averaged over their entire time-course and registered onto the corresponding UTE images using Advanced Normalization Tools (ANTs) (26) using the same parameters for all patients. The deformation fields produced by this registration were then applied to the PBF and defect classification maps as well as the lung segmentations derived from morphological imaging to allow for voxel-wise comparisons. An example of these deformation fields is given in Figures 1A,B. (2) On each slice, the left and right lungs were individually subdivided into areas with the following approach: Each lung was subdivided into 10 strips of equal volume vertically and then each of these strips was separated into 10 blocks of equal volume horizontally. By applying this subdivision algorithm on the segmentations of the DCE images and the UTE independently, this approach yields subdivided areas for both different resolutions and different respiratory states, which nevertheless correspond to the same physical volumes in the lungs and thus allow for correlations and comparisons without a registration step. An example of this subdivision is shown in Figures 1C,D.

FIGURE 1
www.frontiersin.org

Figure 1. Example horizontal (A) and vertical (B) deformation field as determined by image registration, scaled in number of voxels. Segmentations with subdivided areas for a DCE image (C) and the corresponding 2D UTE image (D). Areas that correspond to each other in the subdivision are drawn in identical colors.

2.4 Statistical analyses

Inter-patient correlations were calculated for median T1(TE) with median PBF (ρIPBF) and QDP (ρIQDP) within the segmented slices. For each patient, Spearman’s correlation coefficients between T1(TE) and PBF were calculated separately for registered voxels (ρregPBF) and subdivided areas (ρsubPBF), respectively. Registered voxels in UTE were classified as normal perfusion or perfusion defect according to the corresponding classification map from DCE, and the median T1(TE) in each fraction was compared by Wilcoxon rank-sum tests. For subdivided areas, the fraction of defect voxels (i.e., local QDP) in each subdivided area was correlated with median T1(TE) in the corresponding area, analogous to PBF, as the local coefficient ρsubQDP. A value of p <0.05 was considered statistically significant.

3 Results

Suitable UTE and DCE measurements were available from 22 patients. In four patients, one of the two measurements comprising the UTE acquisition was mismatched in one slice or insufficient time was available to acquire both slices. Figure 2 shows an example set of base images and parameter maps in one patient.

FIGURE 2
www.frontiersin.org

Figure 2. (A) Original image from DCE MRI, averaged over all timepoints and three slices. (B) The same image, registered to the UTE image. (C) Image derived from inversion recovery UTE that was used as registration target. (D) PBF map after application of the distortion field produced by the image registration, with perfusion defects delineated in blue. (E) T1 map acquired using UTE at TE1 = 70 μs. (F) T1 map at TE2 = 500 μs.

The coefficients of the inter-patient correlation of T1(TE) with median PBF for all corresponding lung slices were ρIPBF1-5 = 0.56, 0.64, 0.59, 0.60, and 0.29 at TE1-5 and the respective correlations of T1(TE) with QDP were ρIQDP1-5 = −0.56, −0.76, −0.63, −0.53, and −0.26. Of these, only the correlations at TE1-4 were significant (at p < 0.02). Table 2 shows the mean local correlation coefficients of T1(TE) with PBF based on both methods for associating UTE with DCE images, as well as the correlation of T1(TE) with QDP. The values of ρregPBF and ρsubPBF for individual patients are also shown in Figure 3 to illustrate the inter-patient variation.

TABLE 2
www.frontiersin.org

Table 2. Mean local correlations.

FIGURE 3
www.frontiersin.org

Figure 3. Spearman’s correlation coefficients for the correlation of T1(TE) with PBF, (A) voxel-wise correlation based on image registration and (B) based on subdivisions, respectively. Each patient is shown with the same color in both graphs.

At TE1 = 70 μs, in 18 out of 22 patients, T1 was shorter in voxels classified as perfusion defect when compared to the rest of the lung voxels classified as normal perfusion in the respective individual. In three patients, no significant difference was found and in one patient, T1 was longer in the defect area. While, as shown in Figure 4A, T1 in normal (non-defect) areas and ΔT1, the relative difference, varied strongly among patients, median T1 in defect areas was shorter at all TEs when considered over all patients. Notably, as shown in Figure 4B, mean ΔT1 was smaller at TE2 than at TE1 and TE3, with ΔT1 = 6.2, 4.0, 6.2, 9.1, and 16.6%, similar to the local correlations observed.

FIGURE 4
www.frontiersin.org

Figure 4. (A) Relative difference of T1 at TE1 = 70 μs between voxels classified as perfusion defect and voxels classified as normal. Each point represents one individual. (B) Mean of individual median T1 over all patients at each TE, according to defect classification.

As demonstrated in Figure 2, visual inspection of parameter maps suggested that stronger local correlations appeared in patients where areas with clearly recognizable perfusion and longer T1 remained. To investigate this observation, the inter-patient correlation between median T1(TE1) and the local correlation coefficients of T1 and PBF was calculated. At TE1 = 70 μs, this was ρregcorr = 0.68 when using registration and ρsubcorr = 0.32 when using subdivided areas.

4 Discussion

In this work, inter-patient correlations of T1(TE) with quantitative perfusion measures showed the strongest correlation coefficient at TE2 = 500 μs. Local voxel-wise correlations instead showed the strongest correlations at TE1 = 70 μs and in fact weaker correlations at TE2 than at TE3 = 1,200 μs. Finally, at TE1, in 18 out of 22 patients, T1 was shorter in voxels classified as defect than in normal areas.

The visual score for perfusion MRI has been established in COPD and cystic fibrosis, and were shown to correlate with clinical metrics (6, 10) and with T1 at conventional echo times (8). Subsequently, it was demonstrated that the inter-patient correlation of median T1(TE) in the entire lungs with this perfusion score and quantitative CT parameters depends on TE. Specifically, the strongest correlation was found at TE = 500 μs (18). This agrees with inter-patient correlations of T1(TE) with quantitative perfusion measures shown in this work which, also gave the strongest correlation at TE2.

Perfusion defects in percent was developed to compensate for the vulnerabilities in the quantification of PBF related to non-linearity of Gadolinium-concentration with MRI signal, artifacts, limited temporal resolution, and low SNR in the lung parenchyma (which is particularly low in diseased areas of the lung parenchyma) (14, 27, 28). This is supported by stronger inter-patient correlations between QDP and the visual perfusion score than were found with visual perfusion scores previously (13). In our present work, the inter-patient correlation between T1(TE) and QDP was also stronger at TE2 and stronger than with PBF. However, while the correlations of T1 with PBF based on subdivision were slightly stronger than those based on image registration, only very weak correlations of T1 with local QDP within subdivisions were found. This is most likely to due to many subdivided areas containing only defect or only normal voxels and thus not describing larger defect areas well. QDP derived from DCE using a different algorithm was found to locally match QDP derived from contrast-free PREFUL MRI (28). QDP derived from T1(TE) (rather than DCE-MRI) was shown to correlate similarly to T1(TE) itself in patients with CF (17), but not in patients in COPD (18). Thus, we did not re-investigate defect classification based on T1(TE) in the present work.

Unlike the inter-patient correlations, the local correlations found in the present study were highest at TE1 and higher at TE3 than at TE2. This can be attributed to lung areas more affected by disease having not only shorter T1 but also exhibiting a different dependence of T1 on TE. This is supported by the behavior of T1 in areas classified as perfusion defect or normal, when averaged over the patient collective: There, the average difference between these areas decreased from TE1 to TE2 and then increased again at TE3. This behavior becomes clearer when considering the data published previously comparing patients with COPD to healthy volunteers (18). There, the TE-dependence of T1 was much more gradual and observable at longer TEs in healthy volunteers, while the observed T1 in patients with COPD did not notably increase further after TE = 500 μs, which also meant that the separation between these groups was smallest at TE = 500 μs. Assuming that the TE-dependence of T1 in areas classified as normal through perfusion is closer to that in healthy volunteers and in defect areas closer to that in the patient group, this is consistent with the results seen in this work. It should however be noted that in the present collective areas classified as normal are still quite abnormal in T1 compared to values observed in healthy volunteers. Thus, our results show that both base T1 and its TE-dependence differ in areas of normal and defective perfusion. This further supports the hypothesis that it may be possible to separate the effects of local tissue changes and perfusion on T1 and investigate both with a single measurement. However, this is not yet possible with the current approach.

A large variation of local correlations was observed, with the strongest observed local correlations reaching the same magnitude as the inter-patient correlations. The overall visual impression of T1 and PBF parameter maps was that strong correlations were mainly present in patients with some volume of remaining functional lung tissue. In those cases, a notable spread of perfusion values exists that can be correlated in the first place. As such, we assumed that low local correlations might indicate relatively uniformly severely damaged lungs and thus also short T1. In order to confirm this assumption, we calculated the inter-patient correlation of median T1(TE1) with local correlation coefficients in each patient. That this correlation was significant supports the assertion that weak local correlations were primarily found in patients with more damaged lungs. However, this correlation was much weaker when based on subdivided areas. Since regions with long T1 are unlikely to coincide with the borders of subdivided areas, they can be assumed to describe these less well than the direct voxel-wise correlations. Similarly, in cases where a perfusion deficit was present all over the lung with corresponding short T1, these borders may matter less. At the same time, subdivided areas were less vulnerable to outlier values produced by quantification at very low SNR due to reduced proton density in severe emphysema. In further research, it may be worthwhile to investigate this in milder or presymptomatic cases of COPD, where a larger spread of values to correlate could be investigated. Additionally, MRI in general struggles in patients with severe emphysema simply due to lack of proton signal, though we might surmise that T1 mapping at TE1 = 70 μs should be better able to cope. However, in this work we were limited by the possibilities provided by DCE imaging.

Since in COPD patients, the TE-dependence of T1 is shifted to shorter TEs in comparison to healthy subjects, a tight spacing of TEs was necessary to depict the effects of interest in this work (16, 18): The TE-dependence of local and inter-patient correlations discussed here was visible at TE1 = 70 μs and TE2 = 500 μs, but it is not possible to measure T1 at both TEs in one TR due to the time required for the read dephasing gradient in between. Accordingly, the acquisition had to be split into two sets of echoes, which were implemented in separate measurements. This had the additional problem that the dc-signal used for respiratory gating is not necessarily aligned between those two measurements in case the patient changes how deep they breathe between them. Additionally, the two measurements could be misaligned due to operator error, making the entire slice unusable. After this study was concluded, we redesigned the sequence such that different sets of TEs could be acquired in alternating successive TRs which has neither of these disadvantages while still providing the required tight spacing of TEs.

The 4D MRI used for perfusion quantification was a single measurement, which meant that if the contrast injection was administered too early or late or the patient holding their breath for unsufficient time, it was entirely unusable for quantification. Since the UTE measurement was not part of the primary study protocol, we could only plan two slices and did not have time to retry in case of mistakes. However, these were separately acquired and each split into two measurements, as stated above. In cases where one slices was unusable due to operator error in at least one such sub-measurement, we decided to nevertheless use the remaining slice since the total number of patient measurements was so small. Thus, the UTE images being based on 2D slices was a major limitation of this work: a reliable 3D implementation of the multi-echo inversion recovery UTE would allow for a better quality of registration and give better coverage of the lungs, which would also provide improved comparability with perfusion images.

We investigated the approach to derive local correlations from subdivided areas in addition to image registration to avoid reliance on the registration algorithm, as it is not practically possible to guarantee a successful registration: The quality of image registration limits the comparability and thus voxel-wise correlation of parameter maps. The approach using subdivided areas was designed as an alternative that avoids this problem by increasing the size of spatial units to potentially correct for some imprecision in registration but only with the limitation that it is based on the assumption that both lungs have equal volume. This was implemented in order to be able to use the same number and pattern of subdivisions in each patient. In future work, the accuracy of local correlations could be increased by generating a specific subdivision pattern for each patient, possibly in 3D, which employs different number of subdivisions for each lung. Furthermore, the two segmentations may be subtly different beyond the issue of comparing two- with three-dimensional data. Finally, a relevant limitation for both approaches was that DCE images were acquired in inspiratory breath-holds, while UTE images were acquired during free breathing and reconstructed to expiration retrospectively.

As noted above, the approach discussed here relies entirely on using local perfusion as an indicator of lung function. However, even in early COPD stages, ventilation/perfusion (V/Q) mismatches have been reported, suggesting a blocked hypoxic pulmonary vasoconstriction mechanism to inhibit inflammation (29, 30). Nevertheless, there remains a significant association between ventilation and perfusion in COPD patients and correlation of the extent of V/Q mismatches with disease severity was relatively modest (31). To incorporate the effect of local V/Q mismatches, an additional direct measure of ventilation could be combined with perfusion imaging techniques. While non-contrast enhanced lung perfusion imaging methods have been introduced that also depict ventilation, it should be noted that the 2D UTE sequence used in this work was initially developed for oxygen-enhanced imaging since both T1 and T2* (which was not considered in this work but can be quantified from the same measurement) depend on the presence of molecular oxygen (dissolved in tissue and as gas in the alveoli, respectively) (4, 20). An additional acquisition while supplying subjects with oxygen could thus be used to determine local ventilation directly and thus investigate V/Q mismatches.

In conclusion, we have shown inter-patient correlations of T1(TE) with quantitative perfusion measures that are comparable to those described previously for T1(TE) with semi-quantitative visual perfusion scores from perfusion MRI and quantitative CT parameters. Local (intra-patient) correlations were much weaker on average, but the strongest correlations found reached a similar scale as the inter-patient correlation coefficients. Importantly, the TE-dependency of the local correlation coefficients implies that not only base T1 but also its TE-dependency in normal areas is closer to the behavior previously found in healthy volunteers than in defect areas. That the correlation coefficients themselves correlate with overall average T1 implies that weak local correlations would be observed in patients with few remaining areas of well-functioning lung tissue and thus low contrasts between local observations. Accordingly, it may be interesting to investigate DCE perfusion and T1(TE) in a patient cohort with less severe, earlier disease, both individually and in combination.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Institutional Review Board of the Medical Faculty of the University of Marburg (200/09) and of the University of Heidelberg (S-656/2012). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

ST: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Software, Validation, Visualization, Writing – original draft, Writing – review & editing. MK: Conceptualization, Investigation, Methodology, Software, Writing – original draft, Writing – review & editing. JB: Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Supervision, Writing – review & editing. ME: Conceptualization, Supervision, Writing – review & editing. CV: Conceptualization, Funding acquisition, Project administration, Resources, Writing – review & editing. RJ: Conceptualization, Funding acquisition, Project administration, Resources, Writing – review & editing. H-UK: Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Writing – review & editing. CH: Conceptualization, Methodology, Project administration, Writing – review & editing. BJ: Conceptualization, Investigation, Methodology, Writing – review & editing. MW: Conceptualization, Investigation, Methodology, Project administration, Supervision, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article. This project was supported by grants from the German Federal Ministry of Education and Research (01GI0884 and 82DZL004B1). Contrast medium for MRI was sponsored by Bayer Vital GmbH, Leverkusen, Germany. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. For the publication fee we acknowledge financial support by Heidelberg University within the funding programme open access publications.

Acknowledgments

We thank all patients for their participation in this study. We also thank the MRI technologists at the Thoraxklinik at the University of Heidelberg for their assistance in performing the MRI examinations.

Conflict of interest

MK was an employee of Boehringer Ingelheim.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmed.2023.1254003/full#supplementary-material

SUPPLEMENTARY FIGURE 1 | Relative difference of T1 at TE1 to TE5 between voxels classified as perfusion defect and voxelsclassified as normal. Each point represents one individual. This corresponds to Figure 4A at all TE.

References

1. Hashimoto, N, Wakahara, K, and Sakamoto, K. The importance of appropriate diagnosis in the practical Management of Chronic Obstructive Pulmonary Disease. Diagnostics. (2021) 11. doi: 10.3390/diagnostics11040618

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Lynch, DA, Austin, JHM, Hogg, JC, Grenier, PA, Kauczor, H-U, Bankier, AA, et al. CT-definable subtypes of chronic obstructive pulmonary disease: a statement of the Fleischner society. Radiology. (2015) 277:192–205. doi: 10.1148/radiol.2015141579

CrossRef Full Text | Google Scholar

3. Hatabu, H, Alsop, DC, Listerud, J, Bonnet, M, and Gefter, WB. T2* and proton density measurement of normal human lung parenchyma using submillisecond echo time gradient echo magnetic resonance imaging. Eur J Radiol. (1999) 29:245–52. doi: 10.1016/S0720-048X(98)00169-7

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Bauman, G, Puderbach, M, Deimling, M, Jellus, V, Chefd'hotel, C, Dinkel, J, et al. Non-contrast-enhanced perfusion and ventilation assessment of the human lung by means of fourier decomposition in proton MRI. Magn Reson Med. (2009) 62:656–64. doi: 10.1002/mrm.22031

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Bauman, G, and Bieri, O. Matrix pencil decomposition of time-resolved proton MRI for robust and improved assessment of pulmonary ventilation and perfusion. Magn Reson Med. (2017) 77:336–42. doi: 10.1002/mrm.26096

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Eichinger, M, Optazaite, D-E, Kopp-Schneider, A, Hintze, C, Biederer, J, Niemann, A, et al. Morphologic and functional scoring of cystic fibrosis lung disease using MRI. Eur J Radiol. (2012) 81:1321–9. doi: 10.1016/j.ejrad.2011.02.045

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Wielpütz, MO, Puderbach, M, Kopp-Schneider, A, Stahl, M, Fritzsching, E, Sommerburg, O, et al. Magnetic resonance imaging detects changes in structure and perfusion, and response to therapy in early cystic fibrosis lung disease. Am J Respir Crit Care Med. (2014) 189:956–65. doi: 10.1164/rccm.201309-1659OC

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Jobst, BJ, Triphan, SMF, Sedlaczek, O, Anjorin, A, Kauczor, HU, Biederer, J, et al. Functional lung MRI in chronic obstructive pulmonary disease: comparison of T1 mapping, oxygen-enhanced T1 mapping and dynamic contrast enhanced perfusion. PLoS One. (2015) 10:e0121520. doi: 10.1371/journal.pone.0121520

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Stahl, M, Wielpütz, MO, Graeber, SY, Joachim, MC, Sommerburg, O, Kauczor, H-U, et al. Comparison of lung clearance index and magnetic resonance imaging for assessment of lung disease in children with cystic fibrosis. Am J Respir Crit Care Med. (2017) 195:349–59. doi: 10.1164/rccm.201604-0893OC

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Wielpütz, MO, Eichinger, M, Wege, S, Eberhardt, R, Mall, MA, Kauczor, H-U, et al. Mid-term reproducibility of chest MRI in adults with clinically stable cystic fibrosis and chronic obstructive pulmonary disease. Am J Respir Crit Care Med. (2019) 200:103–7. doi: 10.1164/rccm.201812-2356LE

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Hatabu, H, Tadamura, E, Levin, DL, Chen, Q, Li, W, Kim, D, et al. Quantitative assessment of pulmonary perfusion with dynamic contrast-enhanced MRI. Magnet Reson Med. (1999) 42:1033–8. doi: 10.1002/(sici)1522-2594(199912)42:6<1033:aid-mrm7>3.0.co;2-7

CrossRef Full Text | Google Scholar

12. Ohno, Y, Hatabu, H, Murase, K, Higashino, T, Kawamitsu, H, Watanabe, H, et al. Quantitative assessment of regional pulmonary perfusion in the entire lung using three-dimensional ultrafast dynamic contrast-enhanced magnetic resonance imaging: preliminary experience in 40 subjects. J Magn Reson Imaging. (2004) 20:353–65. doi: 10.1002/jmri.20137

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Schiwek, M, Triphan, SMF, Biederer, J, Weinheimer, O, Eichinger, M, Vogelmeier, CF, et al. Quantification of pulmonary perfusion abnormalities using DCE-MRI in COPD: comparison with quantitative CT and pulmonary function. Eur Radiol. (2022) 32:1879–90. doi: 10.1007/s00330-021-08229-6

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Konietzke, M, Triphan, SMF, Eichinger, M, Bossert, S, Heller, H, Wege, S, et al. Unsupervised clustering algorithms improve the reproducibility of dynamic contrast-enhanced magnetic resonance imaging pulmonary perfusion quantification in muco-obstructive lung diseases. Front Med. (2022) 9:1022981. doi: 10.3389/fmed.2022.1022981

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Jakob, PM, Wang, T, Schultz, G, Hebestreit, H, Hebestreit, A, and Hahn, D. Assessment of human pulmonary function using oxygen-enhanced T1 imaging in patients with cystic fibrosis. Magn Reson Med. (2004) 51:1009–16. doi: 10.1002/mrm.20051

CrossRef Full Text | Google Scholar

16. Triphan, SMF, Jobst, BJ, Breuer, FA, Wielpütz, MO, Kauczor, H-U, Biederer, J, et al. Echo time dependence of observed T1 in the human lung. J Magn Reson Imaging. (2015) 42:610–6. doi: 10.1002/jmri.24840

CrossRef Full Text | Google Scholar

17. Triphan, SMF, Stahl, M, Jobst, BJ, Sommerburg, O, Kauczor, H-U, Schenk, J-P, et al. Echo time-dependence of observed lung T1 in patients with cystic fibrosis and correlation with clinical metrics. J Magn Reson Imaging. (2020) 52:1645–54. doi: 10.1002/jmri.27271

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Triphan, SMF, Weinheimer, O, Gutberlet, M, Heußel, CP, Vogel-Claussen, J, Herth, F, et al. Echo time-dependent observed lung T1 in patients with chronic obstructive pulmonary disease in correlation with quantitative imaging and clinical indices. J Magn Reson Imaging. (2021) 54:1562–71. doi: 10.1002/jmri.27746

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Karch, A, Vogelmeier, C, Welte, T, Bals, R, Kauczor, H-U, Biederer, J, et al. The German COPD cohort COSYCONET: aims, methods and descriptive analysis of the study population at baseline. Respir Med. (2016) 114:27–37. doi: 10.1016/j.rmed.2016.03.008

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Triphan, SMF, Breuer, FA, Gensler, D, Kauczor, H-U, and Jakob, PM. Oxygen enhanced lung MRI by simultaneous measurement of T1 and T2* during free breathing using ultrashort TE. J Magn Reson Imaging. (2015) 41:1708–14. doi: 10.1002/jmri.24692

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Deichmann, R, and Haase, A. Quantification of T1 values by SNAPSHOT-FLASH NMR imaging. J Magn Reson Imaging. (1992) 96:608–12. doi: 10.1016/0022-2364(92)90347-A

CrossRef Full Text | Google Scholar

22. Kohlmann, P, Strehlow, J, Jobst, B, Krass, S, Kuhnigk, J-M, Anjorin, A, et al. Automatic lung segmentation method for MRI-based lung perfusion studies of patients with chronic obstructive pulmonary disease. Int J Comput Assist Radiol Surg. (2015) 10:403–17. doi: 10.1007/s11548-014-1090-0

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Kohlmann, P, Laue, H, Krass, S, and Peitgen, H-O (2011). “Fully-automatic determination of the arterial input function for dynamic contrast-enhanced pulmonary MR imaging” in MIUA.

Google Scholar

24. Sourbron, S, Dujardin, M, Makkat, S, and Luypaert, R. Pixel-by-pixel deconvolution of bolus-tracking data: optimization and implementation. Phys Med Biol. (2007) 52:429–47. doi: 10.1088/0031-9155/52/2/009

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Otsu, N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern. (1979) 9:62–6. doi: 10.1109/TSMC.1979.4310076

CrossRef Full Text | Google Scholar

26. Avants, BB, Tustison, N, and Song, G. Advanced normalization tools (ANTS). Insight J. (2009) 2:1–35. doi: 10.54294/uvnhin

CrossRef Full Text | Google Scholar

27. Woodhouse, N, Wild, JM, Paley, MNJ, Fichele, S, Said, Z, Swift, AJ, et al. Combined helium-3/proton magnetic resonance imaging measurement of ventilated lung volumes in smokers compared to never-smokers. J Magn Reson Imaging. (2005) 21:365–9. doi: 10.1002/jmri.20290

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Kaireit, TF, Voskrebenzev, A, Gutberlet, M, Freise, J, Jobst, B, Kauczor, H-U, et al. Comparison of quantitative regional perfusion-weighted phase resolved functional lung (PREFUL) MRI with dynamic gadolinium-enhanced regional pulmonary perfusion MRI in COPD patients. J Magn Reson Imaging. (2019) 49:1122–32. doi: 10.1002/jmri.26342

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Agustí, AG, Barberá, JA, Roca, J, Wagner, PD, Guitart, R, and Rodriguez-Roisín, R. Hypoxic pulmonary vasoconstriction and gas exchange during exercise in chronic obstructive pulmonary disease. Chest. (1990) 97:268–75. doi: 10.1378/chest.97.2.268

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Hoffman, EA, and Chon, D. Computed tomography studies of lung ventilation and perfusion. Proc Am Thorac Soc. (2005) 2:492–506. doi: 10.1513/pats.200509-099DS

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Rodríguez-Roisin, R, Drakulovic, M, Rodríguez, DA, Roca, J, Barberà, JA, and Wagner, PD. Ventilation-perfusion imbalance and chronic obstructive pulmonary disease staging severity. J Appl Physiol. (2009) 106:1902–8. doi: 10.1152/japplphysiol.00085.2009

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: magnetic resonance imaging, functional lung imaging, T1 mapping, lung T1, chronic obstructive pulmonary disease, dynamic contrast enhancement, quantitative perfusion

Citation: Triphan SMF, Konietzke M, Biederer J, Eichinger M, Vogelmeier CF, Jörres RA, Kauczor H-U, Heußel CP, Jobst BJ and Wielpütz MO (2024) Echo time-dependent observed T1 and quantitative perfusion in chronic obstructive pulmonary disease using magnetic resonance imaging. Front. Med. 10:1254003. doi: 10.3389/fmed.2023.1254003

Received: 06 July 2023; Accepted: 08 December 2023;
Published: 05 January 2024.

Edited by:

Jonathan P. Dyke, Cornell University, United States

Reviewed by:

Marcel Gutberlet, Hannover Medical School, Germany
Nicolau Beckmann, Novartis Institutes for BioMedical Research, Switzerland
Roberto Carbone, University of Genoa, Italy

Copyright © 2024 Triphan, Konietzke, Biederer, Eichinger, Vogelmeier, Jörres, Kauczor, Heußel, Jobst and Wielpütz. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Simon M. F. Triphan, simon.triphan@uni-heidelberg.de

Download