Your new experience awaits. Try the new design now and help us make it even better

CASE REPORT article

Front. Med., 14 November 2025

Sec. Precision Medicine

Volume 12 - 2025 | https://doi.org/10.3389/fmed.2025.1710981

This article is part of the Research TopicAdvances in Precision Medicine for Minimally Invasive Treatment of Pelvis/Hip Fractures: Integration of Digital and Intelligent TechnologiesView all 10 articles

Case Report: Robot-assisted sacral fracture reduction with patient-specific finite element planning

Yupeng Ma,Yupeng Ma1,2Junbo GeJunbo Ge1Huanyu HongHuanyu Hong1Tao HuangTao Huang1Yu LiYu Li1Zhengwen SunZhengwen Sun1Tao Sun,
Tao Sun1,2*
  • 1Orthopaedics Department, Yantai Shan Hospital, Yantai, China
  • 2Yantai Key Laboratory for Repair and Reconstruction of Bone & Joint, Yantai Shan Hospital, Yantai, China

Background: Sacral fractures are typically caused by high-energy trauma. They often disrupt the pelvic ring and pose complex anatomical challenges, as the sacrum is surrounded by critical structures—including blood vessels, nerves, and internal organs. Traditional open reduction and internal fixation (ORIF) can restore anatomical alignment but requires extensive tissue exposure. This exposure leads to greater tissue trauma, prolonged recovery, and higher risks of infection, hemorrhage, or nerve damage. For these reasons, minimally invasive surgery (MIS) is preferred. However, MIS demands high technical precision. Robot-assisted fracture reduction (RAFR) systems enhance precision in minimally invasive procedures, while finite element analysis (FEA) optimizes preoperative planning by simulating biomechanics. However, clinical evidence for combining these techniques in complex, multi-injury cases is limited.

Case presentation: A 19-year-old female was admitted to the hospital following high-energy trauma (a fall from height), diagnosed with unstable pelvic fracture (AO C1.3 type), longitudinal sacral fracture (Denis II type with vertical displacement), and multiple concurrent injuries (thoracolumbar fractures, rib fracture, pulmonary contusion, splenic and renal contusions, lumbosacral plexus injury). Preoperative management included supracondylar femoral traction and vital sign stabilization. Preoperative FEA based on the patient’s CT data simulated three internal fixation schemes, showing comparable vertical stability; S1 standard + S2 extended sacroiliac screws were selected to preserve lumbar mobility. The RAFR system was used for surgery: 3D preoperative planning, automatic path design, and intraoperative real-time tracking. Fixation was performed with the selected screws (posterior ring) and an anterior external fixator.

Conclusion: This case illustrates the value of combining FEA and RAFR in treating complex sacral fractures with multiple traumas. It highlights that FEA provides a scientific basis for personalized fixation strategy selection, while RAFR achieves precise, minimally invasive reduction, offering a feasible pathway for personalized, minimally invasive management.

1 Introduction

Sacral fractures, often from high-energy trauma like motor vehicle accidents or falls, frequently disrupt the pelvic ring and may involve the acetabulum, sacroiliac joints, and pubic symphysis. The complex sacral and pelvic anatomy, along with surrounding critical structures (blood vessels, lumbosacral plexus, internal organs), complicates treatment (14). Traditional open reduction and internal fixation (ORIF) restores alignment but brings extensive exposure, tissue trauma, prolonged recovery, and higher risks of infection, hemorrhage, or nerve damage (5), leading to a preference for minimally invasive surgery (MIS) to reduce trauma and speed recovery (6).

MIS for sacral fractures requires high technical skill due to anatomical complexity and individual variations (7). Robot-assisted systems offer stable, precise control for minimally invasive complex reductions, boosting accuracy, safety, and reducing complication (7). Finite element analysis (FEA) simulates stress distribution, bone–implant interactions, and treatment mechanical outcomes, optimizing preoperative plans to improve success and reduce complications.

In sacral fracture management, FEA aids in selecting optimal strategies by simulating mechanical responses (812), and when integrated with robotics, enhances precision via preoperative planning and intraoperative navigation. However, clinical evidence for such combinations—especially in complex, multi-injury cases—is limited.

This case report presents robotic-assisted fracture reduction (RAFR) combined with preoperative FEA in treating an unstable sacral fracture with multiple traumas, highlighting how this approach addresses minimally invasive challenges for personalized, precise management.

2 Case presentation

The patient is a 19-year-old female admitted to the hospital due to high-energy trauma. Physical examination on admission: vital signs were stable, pressure pain in the pelvic region was obvious, lower limb movement was limited, and sensation was reduced (suggesting lumbosacral plexus injury). Imaging examinations (pelvic orthostatic, inlet and outlet X-rays and CT) showed instability of the pelvic ring and a longitudinal fracture of the sacrum (Denis II type) with vertical displacement. The patient had no prior history of chronic illness or surgery. A final diagnosis was made as follows: (1) Pelvic fracture (AO C1.3 type); (2) lumbosacral plexus nerve injury; (3) thoracic vertebral fractures (T5, T8, T12); (4) lumbar vertebral fractures (L1, L2, L3); (5) lumbar transverse process fractures (L1, L2, L3, L5); (6) pelvic effusion; (7) first rib fracture; (8) pulmonary contusion; (9) splenic contusion; (10) renal contusion. Preoperative management: Supracondylar femoral traction was applied due to vertical instability of the pelvis. Stabilization of the patient’s vital signs and symptomatic treatment of multiple injuries throughout the body.

During the preoperative assessment, anteroposterior (AP), inlet, and outlet X-rays of the pelvis were obtained, along with CT scans for a comprehensive evaluation Figure 1. For patients with vertical instability, supracondylar femoral traction was applied. The surgical plan was developed by considering the fracture type, expected reduction outcome, and the feasibility of constructing bony tunnels.

Figure 1
Six medical images labeled A to F. Images A, B, and C are X-rays of a pelvis showing different angles. Image D is a CT scan of the pelvis. Images E and F are 3D reconstructions of the pelvic bones from different perspectives. Each image highlights the pelvic and hip bone structure.

Figure 1. Preoperative imaging studies of the patient’s pelvis were performed using X-rays in the anteroposterior (A), inlet (B), and outlet views (C). The findings suggested a fracture, and CT (D) with three-dimensional reconstruction (E,F) provided more detailed information.

To clarify the chronological sequence of the patient’s clinical management and follow-up, key events—from admission to the 3-month follow-up—are summarized in Table 1. This timeline highlights the alignment between interventions (e.g., preoperative planning, surgery) and corresponding assessments (e.g., imaging, clinical evaluations).

Table 1
www.frontiersin.org

Table 1. Timeline of key clinical events for the patient from admission to 3-month follow-up.

3 Materials and methods

3.1 Overview of the RAFR system

The RAFR system consists of five main components: fracture reduction software, an optical tracking device, a reduction robot, a holding device, and an elastic traction device. The system ensures precise control during the reduction process by real-time tracking of the patient’s pelvis and robotic operations. The holding and elastic traction devices stabilize the pelvis and counteract the restrictive forces from surrounding soft tissues.

3.2 Finite element model establishment

This study utilized CT data (64-slice spiral CT, 1 mm layer thickness) of the patient’s sacral bone L4-L5 vertebral body and pelvis. First, virtual three-dimensional models of the lumbar spine and pelvis were created from DICOM-format CT data using image processing software (Mimics 21.0), with components segmented based on CT grey values. Next, the pelvic 3D model (generated in Mimics) was imported into Design X 2020.0 software for smoothing, ensuring it was suitable for subsequent computations. Ten-node tetrahedral elements were used. The mesh model was then assigned material properties, set as heterogeneous and isotropic. Material properties were assigned to different skeletal regions based on grey values. Using Mimics’ built-in formulas, pelvic grey values were divided into 10 levels, and empirical formulas were assigned based on relevant literature (10, 13, 14). Using the design functions in SpaceClaim 2019 software, models such as intervertebral disks and cylindrical screws were drawn, and then the drawn models were moved to appropriate positions using the move command to simulate fracture reduction and fixation.

3.3 Ligament and muscle modeling and load application

The skeletal and screw mesh models were imported into ANSYS WORKBENCH 2020, and spring-damper elements were used to simulate ligaments and muscles. Material properties and ligament parameters were referenced from previous literature (8, 9, 15, 16). The material properties of the internal fixation were designed as titanium alloy. The sacroiliac joint and pubic symphysis were modeled using constraint constraints, with six degrees of freedom constraints applied to the nodes of both acetabula. A vertical downward force of 500 N was applied to the upper surface of the L4 vertebra. This simulated the weight-bearing effect during upright standing. The pelvic model uses spring elements to simulate the main ligament structures around the pelvis, ensuring joint mobility and stress transmission at the sacroiliac joint (8, 14). Displacement results show that the anterior margin of the sacrum has the greatest mobility, exhibiting a forward and downward movement trend, while both iliac bones exhibit a rotational trend, consistent with literature reports (9, 15). The sacroiliac joint cartilage surface is set as a sliding friction contact (friction coefficient 0.015) (17). The fracture surface is set as a sliding friction contact (friction coefficient 0.3) (18). The screw thread and bone contact surface, the connection between the screw tail and the disc, and the disc and cortical bone surface are all bound connections.

3.4 Internal fixation model design

In this study, standard sacroiliac screws are defined as screws that traverse the fracture line and reach the midline of the sacrum. Extended sacroiliac screws are defined as screws that traverse the fracture line and pass through the contralateral ilium. Three internal fixation models were established, as shown in Figures 2AC: (1) S1 standard sacroiliac screw + S2 extended sacroiliac screw (SDS1EDS2); (2) unilateral L5 segment iliac-lumbar fixation + S1 standard sacroiliac screw (L5SDS1); (3) unilateral L5 segment iliac-lumbar fixation + S2 extended sacroiliac screw (L5EDS2). The lengths and diameters of the lumbar pedicle screws and iliac screws were 45 mm and 6.5 mm, and 70 mm and 7.5 mm, respectively, while the diameter of the sacroiliac screw was 7.3 mm. All materials are made of titanium alloy. A Boolean operation was performed on the three fixation models, and the vertical displacement of each model was recorded.

Figure 2
Illustrations A, B, and C depict pelvic structures with surgical implants. D, E, and F display stress distribution on implants with color scales indicating stress levels from blue (minimal) to red (maximal). Each panel corresponds to varying implant configurations with associated maximum and minimum stress values provided in the legends.

Figure 2. (A) SI standard sacroiliac screw + S2 extended sacroiliac screw (SDS 1 EDS2). (B) Unilateral segment iliac-lumbar fixation + SI standard sacroiliac screw (L5SDS1). (C) Unilateral 1.5 segment iliac-lumbar fixation + S2 extended sacroiliac screw (L5EDS2). (D) Maximum von Mises stress contour map of SDSIEDS2. (E) Maximum von Mises stress contour map of L5SDS1. (F) Maximum von Mises stress contour map of L5EDS2.

4 Results

4.1 Implementation

4.1.1 Preoperative preparation and system configuration

Preoperative CT data were integrated into the robotic system, and the pelvic fracture images were segmented using planning software to construct a reduction model based on mirror symmetry principles (19). The system utilizes an automatic reduction algorithm to perform optimal path planning, which is reviewed and adjusted by the physician to ensure safe and efficient execution (5, 20) (Figure 3).

Figure 3
Figures A, B and C illustrate the process of automated segmentation and three-dimensional reconstruction. Figures D, E and F depict the subsequent computer-assisted surgical planning for pelvic fractures.

Figure 3. The reduction path was fine-tuned based on preoperative planning. Automatic segmentation and 3D reconstruction of pelvic images using a pelvic realignment (A,B,C). Subsequently, computer-aided surgical planning was performed for the pelvic fracture (D,E,F).

4.1.2 Surgical procedure

a. The patient was positioned supine under general anesthesia, with the hip elevated for better access.

b. Following the intraoperative layout, the holding arms were connected, and standard disinfection procedures were carried out.

c. The patient and robotic trackers were securely installed, CBCT data were collected, and image registration was performed.

d. Using navigation drilling, Schantz pins were placed under guidance to ensure pelvic stability.

e. Femoral traction was applied, and the robotic arm moved the pelvis along the planned path to the target position.

f. After achieving reduction, appropriate implants were selected to stabilize the pelvic structure.

g. X-rays were taken before the conclusion of the procedure to verify accurate reduction and correct screw placement.

The surgical procedure is shown in Figure 4.

Figure 4
A. Operating room with robotic surgical equipment. B. Computer screen showing 3D pelvic and leg models. C-G. X-ray images highlighting surgical procedures with instruments. E, H. Patient post-surgery with incisions and bandages.

Figure 4. (A) A connection was established between the five Schantz pins and the holding device, and the femoral condylar traction pin was linked to the elastic traction device. (B) Under real-time 3D navigation, the robotic arm autonomously moved the affected hemipelvis along the pre-planned reduction path, achieving autonomous reduction. (C–H) AII screw channels were verified by fluoroscopy, and the results were satisfactory. The wound appearance was minimally invasive.

4.2 Finite element biomechanical

To assess the stability of the posterior pelvic ring fixation, we measured the displacement at five points (A-E) on the superior sacral surface under a 500 N vertical load applied to the superior endplate of the L4 vertebra. Vertical displacement data for different sacral fixation models, in Table 2. To identify significant differences in vertical displacement between the three models, we used SPSS 26.0 (IBM Corp., United States) to analyze sacral surface displacement data. First, we conducted normality tests on the sacral displacement data, including the Kolmogorov–Smirnov (V) test and Shapiro–Wilk test. These tests yielded p-values > 0.05 for all groups, indicating the data followed a normal distribution. Homogeneity of variance tests were conducted on the vertical displacement data of the three groups, and the results showed that the p-values were greater than 0.05. To further compare sacral displacement among the three groups, an analysis of variance (ANOVA) was performed, and the results showed no significant differences between groups (p > 0.05). Based on the statistical analysis, it was concluded that there were no significant differences in sacral vertical displacement among the SDS1EDS2, L5SDS1, and L5EDS2 groups. The three internal fixation methods were equally effective in terms of vertical stability. The maximum von Mises stress of the internal implants in the three groups was 219.34 MPa, 252.91 MPa, and 197.84 MPa, respectively, as shown in Figures 2DF. From a safety perspective, the maximum stress values of these three internal fixation groups are far below the yield stress of titanium metal (1,050 MPa) (21).

Table 2
www.frontiersin.org

Table 2. Displacement data for the superior surface of the sacrum.

4.3 Follow-up results

The fixation was performed using a standard S1 sacroiliac screw combined with an extended S2 sacroiliac screw. The anterior ring was stabilized with an anterior inferior iliac spine external fixator. Immediate postoperative X-ray showed restoration of the pelvic axis and accurate screw placement. One week postoperatively, CT confirmed fracture reduction meeting the excellent criteria of the Matta standard Figure 5. At the 3-month follow-up, the patient reported no pain, normal hip joint range of motion, partial recovery of lower limb sensation and muscle strength, and imaging demonstrated good fracture healing with no signs of internal fixation loosening or fracture. The patient reported, “I can sit normally now, and walk with double crutches without help. I can also do some light daily activities like making the bed. The numbness in my lower back and thighs has gotten much better, and I’m confident about getting back to my daily life soon.”

Figure 5
Three panels display post-surgery pelvic images. Panel A shows a three-dimensional reconstruction with metal rods inserted in the pelvis. Panel B contains two X-rays with visible implants. Panel C presents two CT scans highlighting the metal rods within the pelvic region.

Figure 5. Postoperative X-ray and 3D CT images demonstrated the patient’s postoperative outcome. (A) Postoperative CT three dimensional reconstruction. (B) Postoperative pelvic X-ray orthopantomogram and entrance position. (C) Postoperative CT plain scan of the pelvis.

5 Discussion

Sacral vertical fractures are a type of posterior pelvic ring injury typically caused by high-energy trauma. These fractures present clinical management challenges, and surgical intervention is often required to restore pelvic ring stability. The key focus of treatment is the reduction and fixation of the posterior pelvic ring (2224). Percutaneous sacroiliac screws have become one of the most commonly used fixation methods due to their minimally invasive nature, reduced blood loss, and lower risk of infection (25). This method is suitable for patients with relatively stable sacral fractures. However, for complex vertical fractures, a single sacroiliac screw may not provide sufficient stability, especially in cases with concomitant horizontal displacement.

Some studies have suggested that dual-plane sacroiliac screw fixation could be an effective strategy for enhancing sacral stability. This approach involves placing screws on two different planes to increase the fixation strength and stability of the fracture fragments. Research indicates that dual-plane fixation offers superior resistance to torsion and shear forces compared to single-plane fixation, particularly in cases of sacral fractures with vertical instability. Additionally, extending the length and varying the angles of the dual-plane screws can further optimize mechanical performance and improve surgical outcomes (8, 26).

Some studies have proposed different views. Schildhauer et al. (27) conducted a biomechanical study comparing early weight-bearing in patients with unstable sacral fractures. The study concluded that triangular fixation provides significantly greater stability for the posterior pelvic ring than sacroiliac screws. Triangular fixation combines lumbopelvic fixation with transverse stabilization of sacral fractures, allowing patients with type C pelvic fractures to immediately engage in full weight-bearing functional exercises (28).

For complex sacral vertical fractures, especially those with a high degree of instability, clinicians have begun exploring the combined use of long sacroiliac screws and iliolumbar fixation. Long sacroiliac screws extend across to the contralateral ilium, providing stabilization not only to the sacrum but also to the entire pelvis. Iliolumbar fixation offers additional support, particularly at the lumbopelvic junction, thereby enhancing the overall treatment outcomes. Studies have shown that this combined approach has distinct advantages in restoring sacral anatomy and maintaining pelvic stability, though it does increase the complexity of the surgery and the technical demands on the surgeon (29, 30).

In this patient case, dual-plane screw fixation and triangular fixation showed comparable efficacy. Dual-plane sacroiliac screw fixation was prioritized because it preserves the patient’s lumbar mobility to the greatest extent possible.

There are various methods for fixing sacral fractures, and selecting the optimal fixation technique is crucial for surgical success. This study integrated preoperative finite element analysis (FEA) to further optimize fixation strategies and surgical approaches, thereby achieving personalized treatment plans. In this patient case, the study focused on two biomechanically stable fixation methods suitable for unilateral sacral vertical fractures (Denis Type II, AO C1.3 type): dual-plane sacroiliac joint screw fixation and sacroiliac joint screw combined with iliac-lumbar fixation. The study designed and compared three fixation strategies, with results showing no statistically significant difference in biomechanical stability between dual-plane fixation and long sacroiliac screw combined with iliac-lumbar fixation. However, dual-plane fixation preserves lumbar mobility. Therefore, for this patient, dual-plane sacroiliac joint screw fixation should be prioritized to ensure a minimally invasive surgical approach.

With advancements in surgical technology, robotic-assisted techniques are increasingly being applied to the surgical management of sacral vertical fractures. Surgical robots improve the precision of screw placement, reduce operational errors, and minimize radiation exposure. However, for complex displaced sacral fractures, inadequate reduction significantly increases the risk of complications associated with sacroiliac screw fixation, making proper reduction a critical intraoperative factor. Recent studies have shown that using robotic systems for sacral fixation enhances surgical safety and accuracy—especially in the reduction and fixation of complex fractures. This study utilized a unique 3D image-guided intelligent RAFR system, which enables truly intelligent, minimally invasive reduction surgery for complex pelvic fractures.

The RAFR (robotic-assisted fracture reduction) system used in this study offers an innovative solution in this context. The RAFR system, based on three-dimensional imaging technology, provides enhanced precision and control for minimally invasive reduction of complex fractures through preoperative planning and intelligent navigation. Compared to traditional surgical methods, this system demonstrates significant advantages, including higher reduction accuracy and a lower rate of complications. In studies involving this RAFR system, 95.5% of 22 patients with unstable pelvic fractures achieved excellent or good reduction outcomes based on the Matta criteria, with the operation process causing no additional damage to patients, demonstrating favorable safety and effectiveness that meet clinical requirements (20).

The primary advantage of the RAFR system lies in its ability to use three-dimensional imaging guidance for reduction planning, minimizing judgment errors inherent in manual operations and enhancing the consistency of surgical quality. This technology allows for the personalized design of the optimal reduction path and target position, ensuring the most precise adjustments to the fracture site during surgery. Additionally, the automation capabilities of the RAFR system significantly reduce operative time, improve surgical efficiency, and provide clearer guidance for the surgeon, thereby lowering the risk of radiation exposure (3135).

This study has several limitations: While this study explored the clinical application of FEA and robot-assisted pelvic reduction and used the patient’s own CT data to construct an FEA model (simulating the patient’s mechanical characteristics), the model can only partially explain features of the patient’s fracture and internal implants. It cannot fully restore the patient’s actual condition or completely represent the patient’s actual biomechanical status. Future research should continue to focus on improving the accuracy of FEA models. Additionally, the RAFR system also has limitations in practical application. The system’s path planning relies on pelvic symmetry, which may not provide sufficient reduction accuracy for patients with bilateral severe comminuted pelvic fractures. This reliance on the symmetry assumption poses challenges when dealing with complex asymmetric fractures. Furthermore, the RAFR system currently lacks the ability to adjust the reduction path in real-time during surgery, which may limit its effectiveness in certain complex cases. Although the RAFR system offers significant clinical advantages in minimally invasive treatment of pelvic fractures, further research is needed to optimize its technical performance and expand its scope of application. Future research directions may include enhancing the system’s automatic path adjustment capabilities, developing technical solutions applicable to a broader range of fracture types, and conducting large-scale randomized controlled trials to further validate its clinical efficacy and safety.

6 Conclusion

This case study demonstratively illustrates the application value of combining finite element analysis with a robot-assisted fracture reduction system in personalized treatment through the management of a complex sacral fracture complicated by multiple injuries. The innovation of this approach lies in the seamless integration of preoperative biomechanical simulation and intraoperative robotic precision: (1) FEA evaluates the biomechanical properties of different fixation schemes, providing a scientific basis for developing individualized surgical plans; (2) the robotic-assisted system uses 3D imaging navigation and precise manipulation to achieve minimally invasive, precise reduction of complex fractures. This effectively overcomes the challenge of insufficient surgical precision in anatomically complex regions—a limitation of traditional surgery. Clinical practice has demonstrated that this technological combination not only enhances the rationality of treatment decisions through preoperative optimization but also reduces trauma and ensures treatment efficacy through intraoperative precise manipulation, without any related complications. It provides a feasible technical pathway for personalized, minimally invasive treatment of complex sacral fractures, and its application potential warrants further exploration and promotion.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Ethics Committee of Yantai Shan Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

YM: Funding acquisition, Writing – original draft, Writing – review & editing. JG: Data curation, Methodology, Writing – original draft. HH: Data curation, Funding acquisition, Methodology, Writing – original draft, Writing – review & editing. TH: Data curation, Methodology, Writing – review & editing. ZS: Data curation, Methodology, Writing – review & editing. YL: Data curation, Writing – original draft. TS: Funding acquisition, Methodology, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. This research was supported by Science & Technology Innovation Development Project of Yantai City, China (No. 2022JCYJ037 and No. 2021YD045 & 2022YD048).

Acknowledgments

We thank the patient for permitting us to use his data to complete this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The authors declare that no Gen AI was used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmed.2025.1710981/full#supplementary-material

References

1. Baghel, A, Verma, MK, Sharma, P, Keshav, K, Kumar, A, and Raghav, S. Functional outcomes and quality of life after sacrum fractures managed by either operative or conservative approaches: a case series with one-year follow-up. Cureus. (2024) 16:e59375. doi: 10.7759/cureus.59375

PubMed Abstract | Crossref Full Text | Google Scholar

2. Bandovic, I, Diedring, B, Olson, A, Tan, S, Denisuk, M, Powell, D, et al. Midline sacral fractures: review of the literature. Spartan Med Res J. (2023) 8:38909. doi: 10.51894/001c.38909

PubMed Abstract | Crossref Full Text | Google Scholar

3. Burkhardt, M, Nienaber, U, Krause, J, Pizanis, A, Moersdorf, P, Culemann, U, et al. Complex pelvic traumas: data linkage of the German pelvic injury register and the TraumaRegister DGU®. Unfallchirurg. (2015) 118:957–62. doi: 10.1007/s00113-014-2565-8

PubMed Abstract | Crossref Full Text | Google Scholar

4. Hermans, E, Edwards, MJR, Goslings, JC, and Biert, J. Open pelvic fracture: the killing fracture? J Orthop Surg Res. (2018) 13:83. doi: 10.1186/s13018-018-0793-2

PubMed Abstract | Crossref Full Text | Google Scholar

5. Wu, Z, Dai, Y, and Zeng, Y. Intelligent robot-assisted fracture reduction system for the treatment of unstable pelvic fractures. J Orthop Surg Res. (2024) 19:271. doi: 10.1186/s13018-024-04761-5

PubMed Abstract | Crossref Full Text | Google Scholar

6. Collinge, C, Coons, D, Tornetta, P, and Aschenbrenner, J. Standard multiplanar fluoroscopy versus a fluoroscopically based navigation system for the percutaneous insertion of iliosacral screws: a cadaver model. J Orthop Trauma. (2005) 19:254–8. doi: 10.1097/01.bot.0000151821.79827.fb

PubMed Abstract | Crossref Full Text | Google Scholar

7. Zhao, C, Wang, Y, Wu, X, Zhu, G, and Shi, S. Design and evaluation of an intelligent reduction robot system for the minimally invasive reduction in pelvic fractures. J Orthop Surg Res. (2022) 17:205. doi: 10.1186/s13018-022-03089-2

PubMed Abstract | Crossref Full Text | Google Scholar

8. Zhao, Y, Li, J, Wang, D, Liu, Y, Tan, J, and Zhang, S. Comparison of stability of two kinds of sacro-iliac screws in the fixation of bilateral sacral fractures in a finite element model. Injury. (2012) 43:490–4. doi: 10.1016/j.injury.2011.12.023

PubMed Abstract | Crossref Full Text | Google Scholar

9. Zhao, Y, Zhang, S, Sun, T, Wang, D, Lian, W, Tan, J, et al. Mechanical comparison between lengthened and short sacroiliac screws in sacral fracture fixation: a finite element analysis. Orthop Traumatol Surg Res. (2013) 99:601–6. doi: 10.1016/j.otsr.2013.03.023

PubMed Abstract | Crossref Full Text | Google Scholar

10. Zhang, L, Peng, Y, Du, C, and Tang, P. Biomechanical study of four kinds of percutaneous screw fixation in two types of unilateral sacroiliac joint dislocation: a finite element analysis. Injury. (2014) 45:2055–9. doi: 10.1016/j.injury.2014.10.052

PubMed Abstract | Crossref Full Text | Google Scholar

11. Meiqi, G, Zhe, X, Yifei, L, Penghui, X, Zhen, W, Rui, Z, et al. Finite element analysis of retrograde superior ramus screw of pubis for the treament of pelvic anterior ring fracture. J Orthop Surg Res. (2025) 20:263. doi: 10.1186/s13018-025-05676-5

PubMed Abstract | Crossref Full Text | Google Scholar

12. Yang, J, Böker, KO, Li, X, Zhou, X, and Lehmann, W. Unilateral Lumbo-pelvic fixation for Denis type I unilateral sacral fracture: a finite element analysis. JOR Spine. (2025) 8:e70022. doi: 10.1002/jsp2.70022

PubMed Abstract | Crossref Full Text | Google Scholar

13. Rho, JY, Hobatho, MC, and Ashman, RB. Relations of mechanical properties to density and CT numbers in human bone. Med Eng Phys. (1995) 17:347–55. doi: 10.1016/1350-4533(95)97314-F

PubMed Abstract | Crossref Full Text | Google Scholar

14. Peng, L, Bai, J, Zeng, X, and Zhou, Y. Comparison of isotropic and orthotropic material property assignments on femoral finite element models under two loading conditions. Med Eng Phys. (2006) 28:227–33. doi: 10.1016/j.medengphy.2005.06.003

PubMed Abstract | Crossref Full Text | Google Scholar

15. Lu, YM, Hutton, WC, and Gharpuray, VM. The effect of fluid loss on the viscoelastic behavior of the lumbar intervertebral disc in compression. J Biomech Eng. (1998) 120:48–54. doi: 10.1115/1.2834306

PubMed Abstract | Crossref Full Text | Google Scholar

16. Faizan, A, Sairyo, K, Goel, VK, Biyani, A, and Ebraheim, N. Biomechanical rationale of ossification of the secondary ossification center on apophyseal bony ring fracture: a biomechanical study. Clin Biomech (Bristol). (2007) 22:1063–7. doi: 10.1016/j.clinbiomech.2007.04.012

PubMed Abstract | Crossref Full Text | Google Scholar

17. Hu, P, Wu, T, Wang, HZ, Qi, XZ, Yao, J, Cheng, XD, et al. Biomechanical comparison of three internal fixation techniques for stabilizing posterior pelvic ring disruption: a 3D finite element analysis. Orthop Surg. (2019) 11:195–203. doi: 10.1111/os.12431

PubMed Abstract | Crossref Full Text | Google Scholar

18. Hartensuer, R, Grüneweller, N, Lodde, MF, Evers, J, Riesenbeck, O, and Raschke, M. The S2-alar-iliac screw for pelvic trauma. Z Orthop Unfall. (2021) 159:522–32. doi: 10.1055/a-1190-5987

PubMed Abstract | Crossref Full Text | Google Scholar

19. Ead, MS, Westover, L, Polege, S, McClelland, S, Jaremko, JL, and Duke, KK. Virtual reconstruction of unilateral pelvic fractures by using pelvic symmetry. Int J Comput Assist Radiol Surg. (2020) 15:1267–77. doi: 10.1007/s11548-020-02140-z

PubMed Abstract | Crossref Full Text | Google Scholar

20. Zhao, C, Cao, Q, Sun, X, Wu, X, Zhu, G, and Wang, Y. Intelligent robot-assisted minimally invasive reduction system for reduction of unstable pelvic fractures. Injury. (2023) 54:604–14. doi: 10.1016/j.injury.2022.11.001

PubMed Abstract | Crossref Full Text | Google Scholar

21. Zhang, E, Li, S, Ren, J, Zhang, L, and Han, Y. Effect of extrusion processing on the microstructure, mechanical properties, biocorrosion properties and antibacterial properties of Ti-cu sintered alloys. Mater Sci Eng C Mater Biol Appl. (2016) 69:760–8. doi: 10.1016/j.msec.2016.07.051

PubMed Abstract | Crossref Full Text | Google Scholar

22. Denis, F, Davis, S, and Comfort, T. Sacral fractures: an important problem. Retrospective analysis of 236 cases. Clin Orthop Relat Res. (1988) 227:67–81.

PubMed Abstract | Google Scholar

23. Taguchi, T, Kawai, S, Kaneko, K, and Yugue, D. Operative management of displaced fractures of the sacrum. J Orthop Sci. (1999) 4:347–52. doi: 10.1007/s007760050115

PubMed Abstract | Crossref Full Text | Google Scholar

24. Mehta, S, Auerbach, JD, Born, CT, and Chin, KR. Sacral fractures. J Am Acad Orthop Surg. (2006) 14:656–65. doi: 10.5435/00124635-200611000-00009

PubMed Abstract | Crossref Full Text | Google Scholar

25. Griffin, DR, Starr, AJ, Reinert, CM, Jones, AL, and Whitlock, S. Vertically unstable pelvic fractures fixed with percutaneous iliosacral screws: does posterior injury pattern predict fixation failure? J Orthop Trauma. (2003) 17:399–405. doi: 10.1097/00005131-200307000-00001

PubMed Abstract | Crossref Full Text | Google Scholar

26. Ma, Y, Zhao, Y, Zou, D, Dong, S, Sun, X, Cheng, G, et al. Breakage risk of different sacroiliac screws in unilateral sacral fractures a finite element analysis. BMC Musculoskelet Disord. (2022) 23:953. doi: 10.1186/s12891-022-05898-2

PubMed Abstract | Crossref Full Text | Google Scholar

27. Schildhauer, TA, Ledoux, WR, Chapman, JR, Henley, MB, Tencer, AF, and Routt, MLJr. Triangular osteosynthesis and iliosacral screw fixation for unstable sacral fractures: a cadaveric and biomechanical evaluation under cyclic loads. J Orthop Trauma. (2003) 17:22–31. doi: 10.1097/00005131-200301000-00004

PubMed Abstract | Crossref Full Text | Google Scholar

28. Josten, C, Schildhauer, TA, and Muhr, G. Therapy of unstable sacrum fractures in pelvic ring. Results of of osteosynthesis with early mobilization. Chirurg. (1994) 65:970–5.

Google Scholar

29. Ma, Y, Huang, T, Liu, W, Hong, H, Zhao, Y, Lin, J, et al. Biomechanical effects of S1 sacroiliac screws versus S2 sacroiliac screws on sacroiliac screws combined with a lumbar iliac fixation in the treatment of vertical sacral fractures: a biomechanical finite element analysis. BMC Musculoskelet Disord. (2023) 24:750. doi: 10.1186/s12891-023-06884-y

PubMed Abstract | Crossref Full Text | Google Scholar

30. Ma, Y, Zhao, Y, Hong, H, Huang, T, and Li, Y. Biomechanical comparison of four triangular osteosynthesis fixations for unilateral vertical sacral fractures. Sci Rep. (2023) 13:4424. doi: 10.1038/s41598-023-31418-w

PubMed Abstract | Crossref Full Text | Google Scholar

31. Ron, O, Joskowicz, L, Milgrom, C, and Simkin, A. Computer-based periaxial rotation measurement for aligning fractured femur fragments from CT: a feasibility study. Comput Aided Surg. (2002) 7:332–41. doi: 10.3109/10929080209146522

PubMed Abstract | Crossref Full Text | Google Scholar

32. Nakajima, Y, Tashiro, T, Sugano, N, Yonenobu, K, Koyama, T, Maeda, Y, et al. Fluoroscopic bone fragment tracking for surgical navigation in femur fracture reduction by incorporating optical tracking of hip joint rotation center. IEEE Trans Biomed Eng. (2007) 54:1703–6. doi: 10.1109/TBME.2007.900822

PubMed Abstract | Crossref Full Text | Google Scholar

33. Koo, TK, and Papuga, MO. A computer aided method for closed reduction of diaphyseal tibial fracture using projection images: a feasibility study. Comput Aided Surg. (2009) 14:45–57. doi: 10.3109/10929080903016102

PubMed Abstract | Crossref Full Text | Google Scholar

34. Du, D, Liu, Z, Omori, S, Kurita, M, Tomita, T, Sugamoto, K, et al. Computer-aided parachute guiding system for closed reduction of diaphyseal fractures. Int J Med Robot. (2014) 10:325–31. doi: 10.1002/rcs.1533

PubMed Abstract | Crossref Full Text | Google Scholar

35. Kausch, L, Thomas, S, Kunze, H, Privalov, M, Vetter, S, Franke, J, et al. Toward automatic C-arm positioning for standard projections in orthopedic surgery. Int J Comput Assist Radiol Surg. (2020) 15:1095–105. doi: 10.1007/s11548-020-02204-0

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: sacral fracture, robot-assisted fracture reduction (RAFR), finite element analysis (FEA), minimally invasive surgery, pelvic trauma

Citation: Ma Y, Ge J, Hong H, Huang T, Li Y, Sun Z and Sun T (2025) Case Report: Robot-assisted sacral fracture reduction with patient-specific finite element planning. Front. Med. 12:1710981. doi: 10.3389/fmed.2025.1710981

Received: 22 September 2025; Accepted: 31 October 2025;
Published: 14 November 2025.

Edited by:

Yu Wang, Beihang University, China

Reviewed by:

Yunning Wang, Beihang University, China
Sun Yujie, Yantai Yuhuangding Hospital, China

Copyright © 2025 Ma, Ge, Hong, Huang, Li, Sun and Sun. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Tao Sun, c3VudDUxOUAxNjMuY29t

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.