ORIGINAL RESEARCH article

Front. Med., 13 January 2026

Sec. Translational Medicine

Volume 12 - 2025 | https://doi.org/10.3389/fmed.2025.1748584

Clinical efficacy study of a spherical nasal vestibular stent

Junxiao Jia&#x;Junxiao Jia1Jikuan Qiu&#x;Jikuan Qiu1Rui LiRui Li1Baoshi FanBaoshi Fan1Yi LuYi Lu2Xiangzhi BaiXiangzhi Bai3Yu Song
Yu Song1*Junxiu Liu
Junxiu Liu1*
  • 1Department of Otolaryngology-Head and Neck Surgery, Peking University First Hospital, Beijing, China
  • 2Image Processing Center, Beihang University, Beijing, China
  • 3Laboratory of Virtual Reality Technology and Systems, Beihang University, Beijing, China

Objective: To design and validate a spherical nasal vestibular stent based on vestibular structural changes for treating nasal obstruction.

Methods: This study enrolled 99 patients with nasal obstruction and confirmed positive findings on anterior rhinoscopy. Pre- and post-dilation sinonasal computed tomography (CT) scans were obtained until symptoms nearly resolved. Three-dimensional (3D) reconstruction was utilized to evaluate anatomical changes, and Spearman correlation analysis was performed to assess the relationship between these changes and visual analog scale (VAS) scores for nasal obstruction. Based on 3D reconstructed models and computational fluid dynamics (CFD) parameter evaluations, a nitinol mesh stent customized to the anatomical characteristics of the nasal vestibule was designed. A single-arm clinical trial in 31 patients was subsequently evaluated the stent using NOSE scores, acoustic rhinometry, and rhinomanometry before and after placement. Adverse events were systematically recorded.

Results: 3D reconstruction showed that changes in nasal vestibule volume before and after dilation correlated with patients’ VAS scores for nasal obstruction. In the clinical trial, the spherical nasal vestibular stent—designed using nasal vestibule volume data—significantly reduced nasal resistance (p < 0.05), increased nasal volume and valve area (p < 0.001), and lowered NOSE scores (p < 0.001). Most patients tolerated the stent well; side effects like dryness and pain were mild.

Conclusion: Based on 3D models from dilated nasal vestibules, this study designed a spherical stent that effectively relieves nasal obstruction with minimal risk, positioning it as a promising non-surgical intervention for clinical application.

Introduction

Nasal obstruction is a common manifestation of nasal pathology, significantly impairing quality of life and psychological well-being. The nasal vestibule accounts for 30–50% of total nasal resistance (1), indicating that structural alterations in this region have a critical impact on nasal ventilation. Anatomical abnormalities involving the nasal vestibule commonly affect both the internal and external nasal valves. The internal nasal valve is anatomically defined by the caudal edge of the upper lateral cartilage, the nasal septum, the lateral nasal wall, and the inferior turbinate (2). Management of internal nasal valve compromise is broadly categorized into surgical and non-surgical interventions. Surgical techniques—including alar batten grafts, lateral crural strut grafts, spreader grafts, butterfly grafts, and suspension sutures (37)—enhance airflow by reinforcing or reconstructing the structural integrity of the internal nasal valve. Non-surgical approaches include external and internal nasal dilators, such as Breathe Right strips (8, 9) and Nozovent silicone stents (10), which function through mechanical expansion of the internal nasal valve and stabilization of the nasal alae (11, 12). However, most current devices primarily target lateral dilation of the internal nasal valve, neglecting the potential advantages of simultaneously enlarging both the external and internal nasal valves—that is, increasing the overall volume of the nasal vestibule to optimize airflow dynamics. Moreover, clinical evidence supporting the efficacy and safety of these over-the-counter devices remains limited.

Based on 3D reconstruction models of the dilated nasal vestibule and CFD parameter analysis, this study proposes a novel spherical stent design for overall volume expansion. A nitinol spherical nasal vestibular stent was developed, and its efficacy and safety in alleviating nasal obstruction were systematically evaluated in clinical trials, offering new insights into non-surgical management.

Method

Study subjects

This study enrolled adult patients from the otolaryngology clinics of Peking University First Hospital and Peking University Third Hospital between July 2021 and October 2024. The protocol was approved by the ethics committees of both institutions.

Patients were included if they presented with persistent unilateral or bilateral nasal obstruction confirmed by a positive anterior rhinoscopy examination, had no history of rhinitis, no upper respiratory tract infection within two weeks prior to enrollment, no long-term use of nasal decongestants, and were not undergoing concurrent treatments for nasal obstruction. Exclusion criteria included pregnancy or lactation, nasal tumors, rhinosinusitis, hematological disorders, nitinol allergy, recent participation in other clinical trials, or any condition deemed unsuitable for the study by the investigators, including cognitive dysfunction that precluded completion of required assessment scales.

A positive anterior rhinoscopy test, similar to the modified Cottle maneuver (13), was defined as the patient reporting significantly improved nasal airflow after manually dilating the nasal valve and anterior naris with an anterior rhinoscope (Figure 1).

Figure 1
A medical illustration showing a person’s nostrils being examined. A gloved hand holds a nasal speculum, gently opening the nasal passage to reveal the inside of the nose.

Figure 1. Anterior rhinoscopy test.

Study protocol

Part 1: 3D reconstruction study before and after nasal vestibular dilation

After informed consent, eligible patients completed a VAS assessment. A sinonasal CT scan was performed under natural conditions (pre-dilation CT). The nasal vestibule was then dilated using an anterior rhinoscope until symptom relief, followed by a second CT scan (post-dilation CT), which was performed while patients self-maintained anterior nasal speculum insertion at a depth and angle that provided subjective relief of nasal obstruction. Prior to CT scan, participants were instructed on the proper technique for inserting and adjusting the speculum to achieve optimal dilation, followed by repeated practice sessions to ensure stability during scanning. DICOM data were imported into Mimics 21.0 for 3D reconstruction. The nasal airway was segmented by thresholding to generate a 3D model, which was exported in STL format for anatomical analysis and CFD validation. CFD were assessed using three key parameters: mean velocity, mean pressure, and mean wall shear stress.

Part 2: Clinical efficacy of the spherical nasal vestibular stent

This part used a single-arm design. Participants were selected from Part 1 on a voluntary basis. Prior to stent placement, patients completed the NOSE questionnaire, acoustic rhinometry, and rhinomanometry. Acoustic rhinometry and rhinomanometry were repeated immediately following stent placement to assess changes in nasal patency and airflow dynamics. Patients wore the stent for at least 8 h daily. Follow-up NOSE assessments were conducted on days 1, 3, and 7. Outcomes included subjective symptoms (NOSE score) and objective measures (rhinomanometry, acoustic rhinometry). The nasal-related adverse reaction questionnaire, non-nasal-related adverse reaction questionnaire, and Western Nasal Dilation Tolerance Scale were employed to assess the safety and tolerability of the nasal stent. Flowchart of this study was shown in Figure 2.

Figure 2
Flowchart detailing a clinical study in three parts. Part 1 involves patient enrollment for nasal dilation assessment with VAS score, CT scans, and 3D reconstruction analysis. Part 2 calculates nasal vestibule volume, converts it to a spherical diameter for stent production, and involves stent manufacturing. Part 3 tests stent efficacy and reaction within a week.

Figure 2. Study flowchart.

Statistical analysis

To evaluate the relationship between nasal vestibular volume changes and symptoms, Spearman correlation analysis was performed between volume change parameters and VAS scores. Paired t-tests compared CFD-derived hydrodynamic parameters before and after dilation; p < 0.05 was considered significant. All analyses used SPSS 27.0, with α = 0.05.

For stent efficacy, sample size was calculated using Simon’s two-stage design. Paired t-tests analyzed NOSE scores, nasal resistance, and volume. Cohen’s d was computed for effect size. Adverse event rates were reported as percentages. All tests used SPSS 27.0; p < 0.05 indicated significance.

Results

Part 1: 3D reconstruction results before and after nasal vestibular dilation

A total of 99 patients (70 males, 29 females; mean age 36.8 ± 11.0 years) were enrolled. 3D models were generated from sinonasal CT scans, and nasal vestibular volume parameters were calculated. Spearman correlation analysis was performed between VAS scores for nasal obstruction and volume change parameters; results are in Table 1. Hydrodynamic parameters, including average flow velocity, pressure difference, and wall shear stress, were computed to assess changes before and after dilation; results are in Figure 3 and Table 2.

Table 1
www.frontiersin.org

Table 1. Volumes and related parameters of nasal vestibule.

Figure 3
Visualization comparing wall shear (left) and velocity streamlines (right) using Ansys 2022 R1. Both images feature a color scale indicating intensity, from blue (low) to red (high). Axes markers and scales are included at the bottom, with dimensions provided in meters for wall shear and velocity in meters per second.

Figure 3. CFD analysis of a three-dimensional reconstruction model and streamline illustrations: (A) Before anterior rhinoscopy dilation; (B) After anterior rhinoscopy dilation.

Table 2
www.frontiersin.org

Table 2. Paired t test of CFD outcomes before and after stent placement.

CFD results derived from the 3D models revealed no significant differences in average velocity or pressure (p > 0.05). However, the average wall shear stress decreased significantly after dilation (t = 2.537, p = 0.014).

Part 2: Clinical efficacy of the spherical nasal vestibular stent

In our single-arm clinical trial, the sample size was calculated using Simon’s two-stage design (one-sided α = 0.05, power = 80%), with planned enrollment of 28–42 patients. Ultimately, a total of 31 patients from Part 1 participated in this study on a voluntary basis. The study is registered at Clinicaltrials.gov (NCT05243147).

Subjective symptom evaluation

The NOSE scale was used to assess stent efficacy in improving nasal ventilation. Scores from 31 subjects before and after stent use were compared using a paired t-test; effect size was calculated. Results are shown in Figure 4 and Table 3.

Figure 4
Six line graphs depict changes in nasal symptoms over seven days. Each graph shows a downward trend from day zero to day one, stabilizing thereafter. Graphs illustrate nasal congestion, blockage, trouble breathing, sleeping difficulty due to nasal symptoms, insufficient air intake during exertion, and the total NOSE score, with measurements on vertical axes and days on horizontal axes. Error bars are included.

Figure 4. NOSE score (Day 0 refers to pre-stent placement; days 1, 3, and 7 refer to the 1st, 3rd, and 7th day post-stent placement).

Table 3
www.frontiersin.org

Table 3. Paired t test and effect size of NOSE score before and after stent placement.

The mean baseline NOSE score was 58.55 ± 20.62. Following one day of stent use, the score decreased to 33.68 ± 19.50, further declined to 30.00 ± 16.58 by day three, and remained stable at 30.00 ± 14.39 by day seven. Effect size analysis using Cohen’s d revealed a mean reduction of 26.15 (d = 1.015) on day one, 30.11 (d = 1.212) on day three, and 25.61 (d = 1.037) on day seven.

Objective indicator evaluation

Rhinomanometry and acoustic rhinometry were performed before and after stent placement. Results are presented in Table 4.

Table 4
www.frontiersin.org

Table 4. Results of nasal resistance and acoustic rhinometry changes before and after stent placement.

Adverse event evaluation

Patients completed an adverse event questionnaire at 1, 3, and 7 days after stent use. Incidence is shown in Figure 5.

Figure 5
Four grouped bar charts labeled A, B, C, and D, depict frequencies of issues with nasal devices. Chart A covers comfort issues like pain and distraction. Chart B highlights difficulties in usage and time consumption. Chart C addresses nasal symptoms like rhinorrhea and dry nose. Chart D deals with other symptoms like sleep disturbance and headache. Bars show categories: always, very frequently, about half the time, very rarely, and never. Percentages represent respondent feedback on these experiences.

Figure 5. Adverse event incidence. (A,B) Stent-wearing experience (assessed by the Western Nasal Dilation Tolerance Scale). (C) Nasal-related adverse reactions. (D) Non-nasal-related adverse reactions.

Discussion

This study aimed to design a universal nasal vestibular stent based on anatomical changes. To establish the rationale, we analyzed the correlation between nasal vestibular volume changes and obstruction symptoms in patients with a positive anterior rhinoscopy test. Using 3D reconstruction from sinonasal CT scans, we measured nasal vestibular volumes. These results indicate that patients with smaller baseline volumes and greater dilation experience more severe symptoms. Thus, the stent should be designed to increase nasal vestibular volume by enhancing local spatial expansion. We also assessed bilateral symmetry effects. No significant correlation was found between symptom scores and absolute volume difference or asymmetry index, suggesting asymmetry is not a primary factor. Therefore, for unilateral stenosis, stent placement may be needed only on the affected side without compromising contralateral airflow.

CFD serves as an objective indicator of nasal ventilation function (14, 15). Unlike our previous study focusing only on the nasal vestibule (16), this analysis included the entire nasal cavity and both sides simultaneously. Average wall shear stress, decreased significantly after dilation, dropping by 0.012 Pa, more than half of pre-dilation levels. Wall shear stress is the tangential frictional force at the wall surface caused by fluid viscosity and near-wall velocity gradient, it is proportional to both viscosity and velocity gradient. This parameter is critical in engineering and biomedical applications and reflects airway resistance in nasal CFD (1719). The marked reduction in wall shear stress indicates improved nasal hydrodynamics and ventilation. Thus, designing a nasal vestibular stent based on volume parameters to relieve obstruction symptoms is feasible.

Currently, clinical nasal vestibular stents fall into two categories: external and internal dilators. Internal types include intranasal stents, nasal clips, and septal stimulators (8, 9, 20). External dilators are limited by visible strips and adhesive-related skin inflammation. Internal dilators are typically made of silicone (9, 12) or stainless steel (21). Silicone may lose elasticity over time, compromising long-term dilation. Metal devices can cause discomfort, poor biocompatibility, and mucosal irritation due to direct contact with nasal tissues. In contrast, this study adopts a novel approach focused on increasing overall nasal vestibular volume. By optimizing material and design, we developed a spherical nitinol stent. Nitinol provides excellent biocompatibility and shape memory. The spherical structure better matches the nasal vestibule’s anatomy, enhancing airflow through volume expansion while minimizing mucosal irritation.

Our prior research (16) shows that spherical dilation offers optimal mechanical support in the nasal vestibule and the stents were shown in Figure 6. The nasal vestibule is narrower at the anterior naris and nasal valve but wider centrally, making a spherical design well-suited to its natural shape. A sphere’s continuous curved surface distributes external forces evenly, as the normal at any point directs toward the center, and fluid pressure on a sphere is uniform. Based on post-dilation volume measurements, we designed a 15.5 mm spherical nasal stent. To balance comfort and ventilation, the stent has an open posterior end, positioning it close to, but not touching the nasal valve mucosa. While it does not directly enlarge the internal nasal valve cross-section, it indirectly improves airflow by supporting lateral alar tissue, effectively increasing both valve area and overall vestibular volume. We selected nitinol for its biocompatibility, corrosion resistance, shape memory, and super elasticity, properties well-established in clinical use (22). Our group previously used woven nitinol stents post-septal surgery, observing strong support, reduced discomfort, and no complications (11). Unlike existing bilateral nasal strips, stents, or clips, which may cause discomfort on the unaffected side in unilateral cases, our stent is worn unilaterally without impairing contralateral airflow.

Figure 6
Illustration of a cylindrical wire mesh device on the left, a photo of the same device with a metallic tip in the center, and on the right, two similar devices inserted into a person's nostrils.

Figure 6. Nasal vestibular stent and its placement within patient’s nasal vestibule.

Subjective symptom improvement was evaluated using the NOSE score. The NOSE score demonstrated a significant reduction from baseline (before stent use) to post-wearing, and remained stable without increase as the duration of wear extended, indicating rapid and substantial relief from nasal obstruction within the first day, with sustained efficacy over time. Effect size analysis using Cohen’s d also confirmed large effects. These results demonstrate that the spherical nasal vestibular stent effectively alleviates nasal obstruction. For objective assessment, rhinomanometry and acoustic rhinometry were used. Following stent placement, nasal resistance during both inhalation and exhalation significantly decreased in both nostrils (p < 0.01), while overall nasal volume and the nasal valve area showed significant increases (p < 0.001). These findings indicate a substantial improvement in nasal ventilation function.

The safety profile and user experience associated with the nasal vestibule stent were also rigorously evaluated. Safety was assessed in terms of both nasal and non-nasal adverse reactions, while subjective tolerance and wearing experience were evaluated using the Western Nasal Dilation Tolerance Scale (21). The Western Nasal Dilation Tolerance Scale is a validated assessment instrument specifically developed for the evaluation of nasal stents in research and clinical practice, designed to capture patient-reported feedback on device use. It primarily assesses comfort during wear and the presence of adverse events, encompassing dimensions such as user experience, ease of use, potential discomfort, and associated complications. Nasal adverse reactions were mostly mild to moderate: nasal dryness occurred in 72.04% of reports (51.61% mild, 20.43% moderate or higher), foreign body sensation or swelling in 70–80%, and minor bleeding in 12.90%, with no active bleeding. Only one case reported severe rhinorrhea and swelling, which resolved after temporary discontinuation. Non-nasal systemic symptoms were rare: 51.61% reported mild to moderate sleep interference, and headache and tearing were infrequent (70.97–83.87% reported no effect), with no serious systemic events. Regarding wear tolerance, 73.18% rarely or never experienced pain, though 26.88% reported discomfort during rest or sleep, and 21.51% expressed concerns about appearance. Most users (84.95–92.47%) found the stent easy and quick to insert, indicating good usability. Overall, the stent showed favorable safety and tolerability, with primarily mild, reversible local reactions. These findings support design improvements, such as surface hydration, and better patient education to enhance adaptation.

This study still has certain limitations in terms of experimental design and aesthetic craftsmanship. From the perspective of experimental design, future research should include more subjects of different races and genders to reduce potential racial and gender biases. At the same time, the sample size should be expanded, the follow-up period should be prolonged, and a randomized controlled trial design should be adopted to directly compare with existing mainstream therapies, thereby providing higher-level evidence-based medical evidence for the clinical application of spherical nasal vestibular stents. In terms of aesthetics and craftsmanship, subsequent work can focus on optimizing the weaving process to improve the appearance design of the stent, promoting the development of personalized customization solutions, and further enhancing wearing comfort and user experience.

Conclusion

The spherical nasal vestibular stent, designed based on nasal vestibular volume parameters, can increase the overall nasal volume and nasal valve area in patients with nasal obstruction, significantly alleviating nasal obstruction symptoms in patients with a positive anterior rhinoscopy test. It demonstrates good safety and can serve as an effective non-surgical treatment option for clinical use.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving humans were approved by this study was approved by the Institutional Review Board of Peking University First Hospital and Peking University Third Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

JJ: Formal analysis, Writing – original draft, Writing – review & editing, Conceptualization, Methodology, Resources, Supervision. JQ: Formal analysis, Writing – original draft, Writing – review & editing, Data curation, Investigation. RL: Data curation, Formal analysis, Writing – original draft. BF: Data curation, Formal analysis, Writing – original draft. YL: Software, Writing – original draft. XB: Supervision, Methodology, Software, Writing – original draft. YS: Conceptualization, Supervision, Writing – review & editing. JL: Supervision, Conceptualization, Funding acquisition, Resources, Writing – review & editing.

Funding

The author(s) declared that financial support was received for this work and/or its publication. Clinical Effectiveness of Oil-Coated Stents versus Bare Metal Stents for Nasal Packing after Septoplasty: A Multicenter, Double-Blind, Randomized Controlled Trial (2024–2025 Peking University Shijijinyuan “Tengyun” Clinical Research Program) TY2025010. Natural Science Foundation of Beijing - Study on the neuroimmune mechanisms of temperature-controlled modulation of nasal autonomic nerve balance in the treatment of allergic rhinitis L252163. National High Level Hospital Clinical Research Funding (Research Achievement Transformation Project of Peking University First Hospital) 2025CX12.

Acknowledgments

We are grateful to all of those with whom we had the pleasure to work during this project.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author BX declared that they were an editorial board member of Frontiers at the time of submission. This had no impact on the peer review process and the final decision.

Generative AI statement

The author(s) declared that Generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Locketz, GD, Teo, NW, Walgama, E, Humphreys, IM, and Nayak, JV. The nasal vestibular body: anatomy, clinical features, and treatment considerations. Eur Arch Otorrinolaringol. (2016) 273:777–81. doi: 10.1007/s00405-015-3868-2,

PubMed Abstract | Crossref Full Text | Google Scholar

2. Kovacevic, M, and Wurm, J. Spreader flaps for middle vault contour and stabilization. Facial Plast Surg Clin North Am. (2015) 23:1–9. doi: 10.1016/j.fsc.2014.09.001,

PubMed Abstract | Crossref Full Text | Google Scholar

3. Uebel, CO, and Matta, R. Alar cartilage-an alternative for spreader graft in primary rhinoplasty. Eur J Plast Surg. (2017) 40:417–26. doi: 10.1007/s00238-017-1336-5,

PubMed Abstract | Crossref Full Text | Google Scholar

4. Pedroza, F, Anjos, GC, Patrocinio, LG, Barreto, JM, Cortes, J, and Quessep, SH. Seagull wing graft: a technique for the replacement of lower lateral cartilages. Arch Facial Plast Surg. (2006) 8:396–403. doi: 10.1001/archfaci.8.6.396,

PubMed Abstract | Crossref Full Text | Google Scholar

5. Ulusoy, S, Dinç, ME, Dalğıç, A, Dizdar, D, Avınçsal, MÖ, and Külekçi, M. Effects of spreader grafts on olfactory function in Septorhinoplasty. Aesth Plast Surg. (2016) 40:106–13. doi: 10.1007/s00266-015-0597-2,

PubMed Abstract | Crossref Full Text | Google Scholar

6. Howard, BE, and Madison, CJ. Evolution of the butterfly graft technique: 15-year review of 500 cases with expanding indications. Laryngoscope. (2019) 129:S1–S10. doi: 10.1002/lary.27776,

PubMed Abstract | Crossref Full Text | Google Scholar

7. Miladi, A, McGowan, JW 4th, and Donnelly, HB. Two suturing techniques for the prevention and treatment of nasal valve collapse after Mohs micrographic surgery. Dermatologic Surg. (2017) 43:407–14. doi: 10.1097/DSS.0000000000000975,

PubMed Abstract | Crossref Full Text | Google Scholar

8. Kiyohara, N, Badger, C, Tjoa, T, and Wong, B. A comparison of over-the-counter mechanical nasal dilators: a systematic review. JAMA Facial Plast Surg. (2016) 18:385–9. doi: 10.1001/jamafacial.2016.0291,

PubMed Abstract | Crossref Full Text | Google Scholar

9. Raudenbush, B. Stenting the nasal airway for maximizing inspiratory airflow: internal max-air nose cones versus external breathe right strip. Am J Rhinol Allergy. (2011) 25:249–51. doi: 10.2500/ajra.2011.25.3621,

PubMed Abstract | Crossref Full Text | Google Scholar

10. Gomes, KSP, Suellen, CM, Filho, JE, Caetano, RO, da Silva, CV, de Oliveira, TMD, et al. Nasal dilator and physiological parameters associated to running performance: a systematic review and meta-analysis. J Sports Sci. (2022) 40:2315–26. doi: 10.1080/02640414.2022.2151752,

PubMed Abstract | Crossref Full Text | Google Scholar

11. Santos Ferreira, CH, Dinardi, RR, da Cunha, IC, and Ribeiro de Andrade, C. Nasal function and cardio-respiratory capacity of adolescent with external nasal dilator. Int J Pediatr Otorhinolaryngol. (2020) 139:110430. doi: 10.1016/j.ijporl.2020.110430,

PubMed Abstract | Crossref Full Text | Google Scholar

12. Petruson, B. The importance of improved nasal breathing: a review of the Nozovent nostril dilator. Acta Otolaryngol. (2007) 127:418–23. doi: 10.1080/00016480500417106,

PubMed Abstract | Crossref Full Text | Google Scholar

13. Kim, SJ, Bang, JH, and Lee, KH. Objective parameters for evaluating internal nasal valve compromise: beyond the angle perspective. Clin Exp Otorhinolaryngol. (2024) 17:234–40. doi: 10.21053/ceo.2024.00099,

PubMed Abstract | Crossref Full Text | Google Scholar

14. Xavier, R, Menger, DJ, de Carvalho, HC, and Spratley, J. An overview of computational fluid dynamics preoperative analysis of the nasal airway. Facial Plast Surg. (2021) 37:306–16. doi: 10.1055/s-0041-1722956,

PubMed Abstract | Crossref Full Text | Google Scholar

15. Plášek, M, Masárová, M, Bojko, M, Komínek, P, Matoušek, P, and Formánek, M. Computational fluid dynamics could enable individualized surgical treatment of nasal obstruction (a preliminary study). Diagnostics. (2022) 12:2642. doi: 10.3390/diagnostics12112642,

PubMed Abstract | Crossref Full Text | Google Scholar

16. Yu, H, Zheng, L, Qiu, J, Wang, J, Xu, Y, Fan, B, et al. Mechanical property analysis and design parameter optimization of a novel nitinol nasal stent based on numerical simulation. Front Bioeng Biotechnol. (2022) 10:1064605. doi: 10.3389/fbioe.2022.1064605,

PubMed Abstract | Crossref Full Text | Google Scholar

17. Lim, ZF, Rajendran, P, Musa, MY, and Lee, CF. Nasal airflow of patient with septal deviation and allergy rhinitis. Vis Comput Ind Biomed Art. (2021) 4:14. doi: 10.1186/s42492-021-00080-2,

PubMed Abstract | Crossref Full Text | Google Scholar

18. Li, L, Zang, H, Han, D, Ramanathan, M Jr, Carrau, RL, and London, NR Jr. Impact of a concha bullosa on nasal airflow characteristics in the setting of nasal septal deviation: a computational fluid dynamics analysis. Am J Rhinol Allergy. (2020) 34:456–62. doi: 10.1177/1945892420905186,

PubMed Abstract | Crossref Full Text | Google Scholar

19. Corda, JV, Shenoy, BS, Ahmad, KA, Lewis, L, Khader, SA, and Zuber, M. Nasal airflow comparison in neonates, infant and adult nasal cavities using computational fluid dynamics. Comput Methods Prog Biomed. (2022) 214:106538. doi: 10.1016/j.cmpb.2021.106538,

PubMed Abstract | Crossref Full Text | Google Scholar

20. Liu, SW, and Lyford-Pike, S. Management of Lateral Wall Insufficiency. Facial Plast Surg. (2023) 39:616–20. doi: 10.1055/s-0043-1773767,

PubMed Abstract | Crossref Full Text | Google Scholar

21. Brandt, MG, Moore, CC, and Doyle, PC. Clinical evaluation of a novel internal nasal dilation stent for the improvement of nasal breathing. Otolaryngol Head Neck Surg. (2008) 138:626–32. doi: 10.1016/j.otohns.2008.01.007,

PubMed Abstract | Crossref Full Text | Google Scholar

22. Alipour, S, Taromian, F, Ghomi, ER, Zare, M, Singh, S, and Ramakrishna, S. Nitinol: from historical milestones to functional properties and biomedical applications. Proc Inst Mech Eng H. (2022) 236:1595–612. doi: 10.1177/09544119221123176,

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: 3D reconstruction, CFD, clinical efficacy, single-arm trial, spherical nasal vestibular stent

Citation: Jia J, Qiu J, Li R, Fan B, Lu Y, Bai X, Song Y and Liu J (2026) Clinical efficacy study of a spherical nasal vestibular stent. Front. Med. 12:1748584. doi: 10.3389/fmed.2025.1748584

Received: 18 November 2025; Revised: 23 December 2025; Accepted: 23 December 2025;
Published: 13 January 2026.

Edited by:

Ajay Bhandari, Indian Institute of Technology Dhanbad, India

Reviewed by:

Russell De Jong, Brooke Army Medical Center, United States
Rama Shankar Choudhary, Manipal Academy of Higher Education, India

Copyright © 2026 Jia, Qiu, Li, Fan, Lu, Bai, Song and Liu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Yu Song, c3lhbmRmQDE2My5jb20=; Junxiu Liu, bGl1anVueGl1c2FueXVhbkBzaW5hLmNvbQ==

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.