Abstract
Introduction: Numerous synthetic, hybrid, and biological grafts and conduits have been deployed to facilitate axonal regeneration across peripheral nerve gaps. Though some strategies have showed promise, larger gaps continue to be an unsolved clinical challenge. Recent evidence suggests that tension-based strategies offer a promising alternative approach to nerve repair. However, whether and to what degree severed peripheral nerves tolerate and accommodate tension, especially in the critical early stages of intervention, is less clear.
Methods: In this study, we evaluated a number of immuno-histochemical outcomes to test the hypothesis that injured rat sciatic nerves accommodate strains of up to 20%, a deformation magnitude that exceeds oft-quoted thresholds for nerve damage. We also assessed the possibility of integrating tension with conduit-based approaches for nerve repair.
Results and Discussion: There were no deficits in axonal, basal laminar, or extracellular matrix morphology with tension, though proximal and distal stumps of nerves in all experimental groups displayed abnormal morphology in proximity to the site of injury. Axons of stretched nerves successfully grew through guidance conduits into the distal stump within 6Ā weeks of repair, thus demonstrating the feasibility of combining tension- and conduit-based regenerative strategies.
Introduction
Traumatic peripheral nerve injuries are a debilitating clinical challenge, resulting in motor dysfunction, sensory loss, and chronic pain. For transected nerves, end-to-end repairs lead to superior clinical outcomes compared to graft-based repairs (Millesi, 1986; ). However, excess tension at the repair site can result in nerve damage and catastrophic repair failure (; Sunderland et al., 2004). Given this concern, surgeons typically use a graft to repair larger (>1Ā cm) nerve gaps (; Singh et al., 2022). Autologous grafts (autografts) are the current gold standard for repairing such gaps, but are limited by donor site morbidity, neuroma, geometric mismatch of the graft with proximal and distal nerve stumps, and limitations on the length and viability of grafts that can be harvested (Schmidt and Leach, 2003; Meek et al., 2005; Selim et al., 2022). Furthermore, autograft efficacy is markedly decreased for gaps >3Ā cm (Selim et al., 2022).
Most alternative strategies for treating severe nerve injury involve the engineering of synthetic, biological, or hybrid conduits or grafts of increasing elegance and complexity [e.g., reviewed in (Schmidt and Leach, 2003; ; Poongodi et al., 2021; ; Selim et al., 2022)]. Such approaches have achieved clinical translation, with acellular human allografts deployed in increasing numbers of clinical trials and large-scale databases (Safa et al., 2020). On the other hand, grafts introduce additional length through which axons must grow before finally reaching the distal stump of the nerve (; ), and though promising for some clinical settings (Rbia et al., 2019), allografts have not achieved autograft efficacy for all clinical contexts (; Peters et al., 2021). Thus, there remains a critical need for new regenerative strategies.
While not conventionally classified as ābiomaterial-ā or āgraft-ā based strategies,ā implantable and non-biological nerve-interfacing biomedical devices represent another approach to nerve regeneration. Peripheral nerve interfaces have long been implanted for neuromodulation or recording (Ray and Maurer, 1975). Correspondingly, neural stimulation principles have been integrated into regenerative technologies (; Maeng et al., 2022). Though less typical, tension-based strategies have also been increasingly incorporated into biomedical devices for peripheral nerve repair. For short gaps, end-to-end repair under tension outperformed a tension-free autograft, based on functional and histological outcomes (; Sunderland et al., 2004; ), confirming that injured nerve stumps can accommodate tension if the repair can be protected (). For larger gaps, nerve-lengthening strategies have been used in rat, rabbit, and primate models. These strategies, which include modified external fixators and screw-based actuators (; ; Saijilafu et al., 2006; Sharula et al., 2010; Yousef et al., 2015), have generally shown positive structural and functional outcomes. We built upon these studies by developing compact, implantable devices that draws the proximal stump linearly towards the distal stump to facilitate more distal nerve reattachment during a subsequent end-to-end repair surgery. This approach yielded comparable or superior outcomes to autografts (; ; ).
Despite promising outcomes, tension-based regenerative approaches largely reflect empirical experimentation. Consequently, we lack insight into whether and to what degree severed peripheral nerves tolerate and accommodate tension, especially in the critical early stages of intervention (; ). In this study, our primary intent was to test the hypothesis that nerves morphologically accommodates strains of up to 20%, a deformation magnitude that exceeds typically cited thresholds for nerve damage to vasculature or neural elements (Wall et al., 1992; ). In addition, we and others hypothesize that there may be benefits to concurrently integrating tension-based strategies with conduit-based solutions (; Zeng et al., 2022). For example, a proximal stump could be stretched into a nerve gap and connected into a conduit, thereby reducing both the effective gap length to be bridged and required conduit material compared to standard tension-free conduits. Thus, a secondary goal of this study was to evaluate the feasibility of a combined tension-conduit based approach.
Materials and methods
Animals
Animal use was approved by the UCSD Institutional Animal Care and Use Committee. Adult male Sprague Dawley rats 12Ā weeks of age (350ā400Ā g) were used. Due to observed autotomy in our initial cohort of Sprague-Dawley rats (), Lewis rats (N = 2ā3 group) of the same age were also used in a subsequent cohort; there were no apparent difference in any measured outcomes (excepting autotomy) across strain, and so data were pooled between strains. A cohort of age-matched controls was also used to provide a frame of reference for parameter values in injured nerves; this avoided any impacts of altered loading on contralateral control limbs.
Nerve injury and device implantation
Sciatic nerve gaps were induced as previously published in a manuscript detailing methods of nerve injury, device fabrication, and device implantation (). Briefly, anesthesia was induced under 5% isoflurane, followed by subcutaneous injection of analgesic (0.05Ā mg/kg buprenorphine) and antibiotics (5Ā mg/kg BaytrilĀ®), and the nerve exposed by separating heads of the biceps femoris. With the hip, knee, and ankle in a neutral position, an 8, 10, or 12Ā mm segment of sciatic nerve was excised proximal to its trifurcation, to create strains of 0%, ā¼10%, and ā¼20% after device implantation (Figure 1A), which imposed a gap of 10Ā mm between the tips of the proximal and distal stump (Figure 1B). Strains on the proximal stump were calculated based on the change of spacing between two painted epineurial markers just proximal to the cuffed region, before and after device implantation ().
FIGURE 1
Implanted devices were slightly modified from our original devices [Figure 1B; cf. Figure 5 in ()]. Briefly, devices consisted of self-sizing silicone spiral nerve cuffs ā¼0.85Ā mm in diameter [cf. Figure 2 in ()], a telescoping backbone (14-gauge stainless steel hypotube upon a stainless steel rod 1.57Ā mm in diameter) to which the cuffs are attached, and 28-gauge steel wire anchors to secure the device to the underlying nerve bed. Nerve cuffs grasped both nerve stumps, while also distributing clamping loads during proximal stump tensioning. A poly(lactic co-glycolic) acid (PLGA) nerve guidance channel [NGC, cf. Figure 3 in ()] with inner diameter of ā¼1.2ā1.6Ā mm, wall thickness of 0.3ā0.4Ā mm, and length as specified in Figure 1, was also integrated into each device, just distal to the proximal cuff (). NGC dimensions allowed the nerve to rest within its lumen, without compression. All components were affixed to the device backbone using medical grade n-butyl cyanoacrylate adhesive ().
Devices were sterilized using ethylene oxide before use. The backbone of the device was positioned along the original axis of the nerve and anchored to the underlying muscle using 28-gauge steel wire. Proximal and distal nerve stumps were secured in cuffs and in NGC groups, the tips of the nerve stumps were carefully brought to the opening of PLGA NGC by manipulating the epineurium [cf. Figure 5 in ()].
For all animals, anesthesia was maintained at 2%ā2.5% isoflurane throughout the surgery. After device implantation, the incision to muscle and skin were closed by 4-0 VicrylĀ® suture and 3-0 ProleneĀ® monofilament suture, respectively. Rats were provided analgesics and full access to food and water post-op. The quality of device implantations was visually confirmed by an orthopaedic surgeon.
Histological sample preparation
Rats were euthanized via CO2 3 or 6Ā weeks after surgery, to evaluate axonal outgrowth and nerve histomorphometry at early time points after repair. At the terminal time point, the implanted device was retrieved along with at least 8Ā mm of each nerve stump. After noting cuffed regions, nerves were released from the device, pinned on cork at slack length, and snap-frozen in liquid nitrogen-chilled isopentane. Prior to sectioning, segments of interest were embedded in Tissue-TekĀ® O.C.T. Compound (Sakura Finetek, Torrance, CA), and 10 μm cross-sectional slices, sectioned on a cryostat (CM3050S, Leica, Buffalo Grove, IL), were mounted on a glass slide. Cross sections were obtained from 4 regions along the nerve (Figure 1C): 1) far proximal stump (ā¼10Ā mm proximal to the nerve gap); 2) near proximal stump (within the cuff, ā¼3Ā mm proximal to the nerve gap); 3) near distal stump (within the cuff, ā¼3Ā mm distal to the nerve gap); 4) far distal stump (ā¼10Ā mm distal to the original nerve gap).
Immunohistochemistry (IHC), imaging, and image processing
To evaluate degeneration and regeneration of severed nerve stumps, the following primary antibodies, with appropriate anti-mouse or anti-rabbit Alexa FluorĀ®-488 or -594 conjugated secondary antibodies (Invitrogen, Carlsbad, CA), were used: mouse monoclonal anti-rat SMI31 (phosphorylated neurofilament) antibody (Abcam ab24573, Cambridge, MA); rabbit polyclonal anti-laminin antibody (Sigma L9393, St. Louis, MO); mouse monoclonal anti-CD68 (pan-macrophage; Abcam ab31630, Cambridge, MA). All antibodies were diluted 1:200. In addition, Massonās Trichrome stain (Poly Scientific, Bay Shore, NY) was used to identify connective tissue localization. Labeled slides were visualized using a Leica TCS SP5 confocal microscope (Leica microsystems, Buffalo Grove, IL) under 10x/0.6 and 63x/1.4 objectives with excitation and emission filters appropriate for desired fluorophores.
For axon counts and macrophage number, 10x images of SMI 31 antibody-labeled sections were converted to 8-bit binary images in ImageJ (v1.47; NIH, Bethesda, MD), thresholded using Otsuās method, and calculated using the āAnalyze particlesā function following removal of all objects less than 4 pixels as well as large structures such as blood vessels and artifactual labeling at the nerve periphery. Axonal and macrophage density was calculated by dividing the axon or macrophage counts by the cross-sectional area of the intra-epineurial region of the nerve, excluding major blood vessels. For axon diameter measurements, two randomly selected 63x confocal images of SMI31-labeled nerves were thresholded using Otsuās method and processed using a custom MATLAB algorithm (MathWorks, Natick, MA) based on pixels per object. Basal lamina integrity was evaluated from 63x confocal images of laminin antibody-labeled slides. To quantify connectivity and integrity of the basal lamina, we converted 63x images of laminin-labeled nerves to a binary image using Otsuās method following application of a Gaussian smoothing filter (radius 3) on the raw image, to reduce high-frequency noise. Laminin area fraction was calculated as a measure of the overall presence of basal lamina. Laminar compaction was assessed by manually quantifying the ratio of open laminar objects (uncompacted) to total laminar objects (n > 80/section). Connective tissue infiltration and integrity were scored using criteria in Table 1, slightly modified from ().
TABLE 1
| Epineurial integrity |
|---|
| 1. Perfect |
| 2. Intact interior border, fraying outer border |
| 3. Diffuse interior and exterior border, but still visible/dark |
| 4. Diffuse interior ands exterior border, less clear border |
| 5. No epineurium visible |
| Intra-epineurial connective tissue density score |
|---|
| 1. Perfectly clean/normal |
| 2. Slightly less compact inner structure, slight connective tissue |
| 3. Moderate inner structure, moderate connective tissue |
| 4. Heavy connective tissue, light/less organized inner structure |
| 5. Heavy connective tissue; no structural integrity |
Scheme for blind scoring of epineurial integrity and connective tissue infiltration based on trichrome staining.
A rating of 1 for each reflects best quality, and 5 reflects worst quality.
Statistics
Mean parameters among device implantation groups were compared using mixed two-way ANOVA to test effects of region (a repeated measure) and strain. Post hoc comparisons were made using Sidakās test, which accounts for multiple comparisons. Dunnettās test was used to compare far proximal regions at each strain to control values. A minimum sample size of 4 animals per group was required to achieve a power of 0.85 and type I error α of 0.05, given effect sizes of 0.45 calculated using Cohenās d based on pilot (n = 3) means and standard deviations of macrophage density. Power calculations were performed using G*Power 3.1 (). To be conservative, 5ā7 samples per group were analyzed. The variability in sample size within a group resulted from early euthanasia required in some groups due to self-mutilation noted above. n = 7 device implantations were performed for 10% and 20% strain at the 3Ā weeks time point, n = 6 for 0% strain at the 3Ā weeks time point, and n = 5 for each strain at the 6Ā weeks time point.
Results
Device implantation
We bridged a 10Ā mm nerve gap following successful device implantation that resulted in a 0% (0.0 ± 0.0), ā¼10% (10.2 ± 0.3%), or ā¼20% (20.2 ± 0.7%) strain in the proximal stump at a location just proximal to the cuffed nerve. The implanted device was retrieved cleanly after three and 6Ā weeks from within an overlying layer of loosely adherent connective tissue [Figure 1D; cf. Figure 6 in ()], and nerves were readily freed from cuffs and the PLGA scaffold for subsequent analysis. On the other hand, 27.8% of all Sprague-Dawley rats (5/18 total) exhibited significant autotomy of the injured limb and were euthanized before the 3-week endpoint. This behavior did not correlate with strain, as 33%, 16.7%, and 33% of animals in 0%, 10%, and 20% strain groups, respectively, were affected. Lewis rats did not exhibit autotomy, consistent with previous data ().
Proximal stump
We examined influences of tension on the regenerating proximal stump, both in intact but mechanically loaded (far proximal) regions, and newly regenerating (near proximal) regions of the proximal stump (Figure 1C). First, we assessed axonal number, density, and diameter based on immunolabeling of phosphorylated neurofilaments (Figure 2; Table 2). Two-way ANOVA indicated no effect of strain or region on axonal number or diameter, but a significant effect of region on axonal density (p < 0.028; Table 2). This regional difference manifest itself most prominently in the 20% strain group, in which far proximal stumps contained a significantly higher density of axons compared to the near proximal region (p < 0.039; Table 2). Axonal density and number in the far proximal (most intact) region were not significantly different from control nerves (density: p > 0.6; number: p > 0.5); however, average axonal diameters for all three strains were significantly smaller than that of control nerves (1.65 ± 0.14; p < 0.02, Dunnettās test).
FIGURE 2
TABLE 2
| Parameter | S/R | FP/NP | 0% | 10% | 20% |
|---|---|---|---|---|---|
| Axon number (#) | ā | FP | 5218 ± 661 | 6446 ± 1,290 | 5491 ± 353 |
| NP | 5064 ± 770 | 5341 ± 556 | 4626 ± 812 | ||
| Axon density (#/μm2) | R | FP | 5 Ć 10ā3 ± 4 Ć 10ā4 | 6 Ć 10ā3 ± 5 Ć 10ā4 | 7 Ć 10ā3 ± 7 Ć 10ā4 |
| NP | 5 Ć 10ā3± 6 Ć 10ā4 | 6 Ć 10ā3± 9 Ć 10ā4 | 5 Ć 10ā3± 4 Ć 10ā4* | ||
| Axon diameter (μm) | ā | FP | 1.34 ± 0.09 | 1.21 ± 0.10 | 1.24 ± 0.03 |
| NP | 1.20 ± 0.13 | 1.05 ± 0.11 | 1.35 ± 0.14 | ||
| Macrophage density (#/μm2) | R | FP | 5 Ć 10ā5 ± 2 Ć 10ā5 | 5 Ć 10ā6 ± 1 Ć 10ā6 | 7 Ć 10ā6 ± 2 Ć 10ā5 |
| NP | 3 Ć 10ā4 ± 9 Ć 10ā5* | 6 Ć 10ā4 ± 2 Ć 10ā4*# | 2 Ć 10ā4 ± 7 Ć 10ā5* | ||
| Epineurial integrity score | R | FP | 1.2 ± 0.2 | 1.2 ± 0.2 | 1.4 ± 0.2 |
| NP | 3.0 ± 0.6* | 2.4 ± 0.2^ | 3.6 ± 0.4* | ||
| Connective tissue infiltration score | R | FP | 1.0 ± 0.0 | 1.4 ± 0.2 | 1.4 ± 0.4 |
| NP | 3.0 ± 0.5* | 2.2 ± 0.2 | 2.8 ± 0.2* | ||
| Laminin area fraction (%) | FP | 42.1 ± 5.0 | 41.3 ± 4.4 | 34.6 ± 6.1 | |
| NP | 37.6 ± 4.8 | 35.7 ± 2.0 | 39.6 ± 7.3 | ||
| Open laminar tubes (%) | R | FP | 88.6 ± 2.3 | 87.0 ± 1.5 | 86.6 ± 2.2 |
| NP | 53.8 ± 3.7* | 54.0 ± 5.1* | 46.0 ± 3.1* |
Summary of mean morphological parameters in proximal stump (Mean+/āSEM).
S/R indicates significant effect of strain (S) and/or region (R) based on 2-way ANOVA. FP, Far proximal; NP, Near proximal. Based on Sidak post-hoc testing: * Significantly different from FP at same strain; ^ Strongly trends towards different from FP at same strain (p < 0.06). # Significantly different from 20% strain in same region. All significant differences are p < 0.05, with exact significance indicated in results text.
We then examined the influence of tension on the inflammatory response and structural environment of the proximal stump. Again, while there was no effect of tension on macrophage density (p = 0.27), a strong regional effect was observed (p < 0.0005). Increased macrophage infiltration was observed at the near proximal stump compared to the far proximal stump for all groups (Figure 3; Table 2), and significantly for the 10% strain group (p < 0.003, Sidak). Consistent with this result, blind scoring of epineurial integrity and intra-epineurial connective tissue infiltration revealed no effect of tension, but a significant effect of region on both epineurial integrity (p < 0.0001) and connective tissue infiltration (p < 0.0002; Figure 4; Tables 1, 2). At both 0% and 20% strain, post-hoc testing revealed decreased epineurial integrity (0%: p < 0.006; 10%: p < 0.06; 20%: p < 0.001; Sidak) and increased connective tissue infiltration (0%: p < 0.002; 20%: p < 0.03; Sidak) in near compared to far proximal regions. The far proximal stump also maintained superior integrity of the basal lamina compared to the near proximal stump. Laminin immunolabeling revealed that though area fractions were regionally similar, there was a significant effect of region on the fraction of open laminar tubes (p < 0.0001), with a higher fraction in far proximal compared to near proximal regions at all three strains (p < 0.0001 for all strains; Figure 5; Table 2). No significant differences were observed in macrophage density, epineurial integrity, connective tissue infiltration, or laminar integrity between the far proximal stump and contralateral control nerves.
FIGURE 3
FIGURE 4
FIGURE 5
Distal stump
We also probed whether tension accelerated or suppressed degenerative outcomes in the distal stump, and whether the structural response varied regionally. There were no effects of tension, favorable or unfavorable, on any inflammatory and morphological parameters in the distal stump. As was the case for the proximal stump, regions closer to the site of injury (near distal) experienced increased connective tissue infiltration and decreased laminar integrity compared to the far distal stump (Figures 6ā8; Table 3). Though there were no regional effects on macrophage density (p = 0.018), there was a significant of region on epineurial integrity (p < 0.0001) connective tissue infiltration (p < 0.0001), and the proportion of open laminar tubes (p < 0.0001), as well as a strong trend towards differences in laminar area fraction (p < 0.07). Post hoc testing revealed significantly higher epineurial integrity (0%: p < 0.005; 10%: p < 0.02; 20%: p < 0.02, Sidak), lower connective tissue infiltration (0%: p < 0.01; 10%: p < 0.004; 20%: p < 0.01, Sidak), and more open laminar tubes (0%: p < 0.0003; 10%: p < 0.03; 20%: p < 0.003, Sidak) in far distal regions compared to near distal regions.
FIGURE 6
FIGURE 7
FIGURE 8
TABLE 3
| Parameter | S/R | ND/FD | 0% | 10% | 20% |
|---|---|---|---|---|---|
| Macrophage density (#/μm2) | ā | ND | 3 Ć 10ā4 ± 5 Ć 10ā5 | 5 Ć 10ā4 ± 1 Ć 10ā4 | 4 Ć 10ā4 ± 1 Ć 10ā4 |
| FD | 3 Ć 10ā4 ± 7 Ć 10ā5 | 4 Ć 10ā4 ± 1 Ć 10ā4 | 3 Ć 10ā4 ± 1 Ć 10ā4 | ||
| Epineurial integrity | R | ND | 4.0 ± 0.3 | 4.2 ± 0.6 | 4.4 ± 0.4 |
| FD | 1.6 ± 0.2* | 2.2 ± 0.7* | 2.4 ± 0.6* | ||
| Connective tissue infiltration | R | ND | 3.6 ± 0.2 | 3.8 ± 0.6 | 4.0 ± 0.3 |
| FD | 2.4 ± 0.2* | 2.4 ± 0.5* | 2.8 ± 0.4* | ||
| Laminin area fraction | R^ | ND | 35.4 ± 10.1 | 26.4 ± 5.1 | 19.8 ± 4.2 |
| FD | 33.1 ± 6.8 | 43.8 ± 4.9 | 32.7 ± 5.1 | ||
| Open laminar tubes (%) | R | ND | 34.6 ± 3.4 | 39.0 ± 2.8 | 34.6 ± 3.5 |
| FD | 55.5 ± 1.8* | 49.6 ± 1.9* | 49.4 ± 5.4* |
Summary of mean morphological parameters in distal stump (Mean+/āSEM).
S/R indicates significant effect of strain (S) and/or region (R) based on 2-way ANOVA. R^ Strongly trends towards effect of region (p < 0.07) FP, Far proximal; NP, Near proximal. Based on Sidak post-hoc testing: * Significantly different from ND at same strain. All significant differences are p < 0.05, with exact significance indicated in results text.
Axonal outgrowth
Consistent with the lack of adverse effects of strain on axonal, structural, and inflammatory response in proximal or distal stumps, axons successfully regenerated halfway into the PLGA graft at 3Ā weeks in all three experimental groups, and into the distal stump at 6Ā weeks (Figure 9). An accurate quantitative evaluation of regenerative outgrowth was confounded by variability created by the unexpected partial flattening of some PLGA tubes in all three groups. Nevertheless, based on area fractions of regenerating axons within the guidance channels (3Ā weeks) and distal stumps (6Ā weeks), no apparent difference in structural regeneration was observed.
FIGURE 9
Discussion
In this study, we addressed two questions from a morphological perspective: 1) whether tension affects the short-term regenerative response of severed nerves; 2) whether regional differences could be detected in the regenerative response in response to tension. In answering these questions, we also assessed the feasibility of tension-based regenerative strategies in combination with a nerve conduit. Our data indicate that tension did not impair regenerative axonal outgrowth at strains up to 20%. Consistent with such growth, there were also no adverse effects of strains up to 20% on the morphology of axons, basal lamina, or extracellular matrix, suggesting that severed proximal nerve stumps can indeed tolerate and positively adapt to appreciable mechanical loads during regeneration (Figures 2ā5, 9; Table 2). Imposition of tension also did not alter indicators of Wallerian degeneration in the distal stump (Figures 6ā8; Table 3). Several interesting regional differences in structure were observed within regions of the nerve that appeared to be structurally intact, macroscopically. These differences could have important implications for the quality of the regenerative environment when pairing tension-based strategies with conduits or grafts.
Effects of strain on regeneration of severed nerves
Like other tissues that span joints, peripheral nerves exist within a dynamic biomechanical environment. Excessive mechanical loading can profoundly impair peripheral nerve function, with strains above ā¼8%ā12% typically associated with irreversible conduction deficits, structural damage, vascular ischemia, and formation of scar and adhesions (; Millesi, 1981; Millesi, 1986; Ogata and Naito, 1986; Wall et al., 1992; ; Tanoue et al., 1996; ). On the other hand, during joint movement, nerves experience paradoxically high regional strains, in some cases approaching 30% (Wright et al., 2001). Such large deformations likely reflect regional differences in nerve structure and biomechanical properties (Phillips et al., 2004; Mason and Phillips, 2011; Sung et al., 2019), and suggest that nerves can adapt to their mechanical environment (Mueller and Maluf, 2002). Consistent with this perspective, tension has been noted to promote neuronal growth, and is required for proper neuronal function and survival (; Siechen et al., 2009). Cultured neurons also exhibit axonal tension under baseline conditions, and may be dramatically elongated in response to tensile loading, at rates exceeding those achieved by growth cone advancement (; ; Zheng et al., 1991; Pfister et al., 2004). Axonal diameters and biological processes such as organelle and cytoskeletal transport are maintained or, in some cases, increased following stretch, suggesting an active structural and biological response to tension (; ). Similarly, at the tissue level, peripheral nerves accommodate physiological levels of lengthening during organism growth as well as imposed lengthening during surgical procedures such as limb-lengthening surgery (; Simpson et al., 2013). Thus, there is substantial evidence that nerves respond favorably to moderate tension.
Several studies, including a modified version of our device, have demonstrated the feasibility of direct end-to-end repairs under tension (; Sunderland et al., 2004; ; ). In addition, progressive lengthening over days to weeks followed by end-to-end repair also results in positive outcomes (Sharula et al., 2010; ; ; ). However, beyond axonal densities and, occasionally, neuronal geometry, little quantification is provided on the morphological response to tension. Our data complement these studies by providing detailed morphological evidence supporting the absence of adverse effects of strains up to 20% on axonal number and quality as well as a consistent regenerative environment both proximal and distal to the injury site. There appears to be no downside to imposing deformation on proximal stumps exceeding oft-cited 8%ā12% levels thresholds for neural injury (; Millesi, 1981; Millesi, 1986; Ogata and Naito, 1986; Wall et al., 1992; ; Tanoue et al., 1996; ). Maximum strains imposed in this study are similar to those experienced by rat nerves in joint configurations corresponding to maximum physiological strain (). Methodologically, imposition of strains on nerves is enabled by gripping the nerve over a relatively large surface area, such that local stresses are likely to be distributed over this area. Such a feature is likely to protect the nerve from stress concentrations during loading and should be integrated into devices designed for clinical translation. Due to the encapsulation of the nerve and device during repair, it is not possible to assess the degree to which nerves have remodeled to relieve this tension. However, in vitro studies and examinations of uninjured nerves indicate consistency of axonal diameter and nerve function in response to deformation as well as increases in internodal distances (Pfister et al., 2004; Simpson et al., 2013; ). These observations demonstrate active longitudinal and radial growth of nerve fibers. In fact, the lack of adverse effects suggests that maximal thresholds for deformation may exceed 20%.
From a device design perspective, an important distinction between our approach and previous approaches is the use of nerve cuffs to redistribute tension. Suture pullout is cited as a possible mode of failure during nerve traction following reattachment of nerves and other tissues (; ). In contrast, nerve cuffs are more likely to distribute local stress concentrations than sutures, and thereby more predictably and robustly accommodate strain.
Regional differences in regenerative and degenerative environments
An important outcome of our study was the observation of dramatic regional differences within proximal and distal stumps (Figure 10). Far proximal (intact) regions of the ālivingā proximal stump were, as expected, similar in quality and composition to contralateral control nerves. Axon diameters in the proximal stump were slightly smaller for all three strains, likely secondary to the initial dying back following nerve transection (). However, the extracellular matrix and basal lamina maintained their integrity and there were no differences in axonal number and density. Uninjured, cuffed control nerves also showed no apparent changes to underlying structure, confirming that any observed regional differences were not an artifact of compressive neuropathy observed in response to implantation of thicker, larger cuffs (; ).
FIGURE 10
Axon number trended lower in newly regenerated near proximal regions than far proximal regions suggesting that not all injured axons regenerate. This was also reflected in a significant effect of region on axonal density, in which the near proximal region exhibited significantly lower axonal density (Table 2). These effects on axonal quality coincided with substantial differences in the structural quality of the newly regenerated nerve. Indeed, near proximal regions exhibited increased macrophage and connective tissue infiltration as well as reduced epineurial and basal lamina integrity. Such phenomena would be expected at the injury site itself, and thus may reflect either residual effects from the initial injury, or a gradual retrograde flow of inflammatory and fibrotic pathways from the exposed regenerating nerve front, which at 3Ā weeks was several mm from the near proximal region (Figure 9).
In the distal stump, we initially hypothesized that, based on consistent gross macroscopic appearance, nerves would reveal a similar regenerative environment. If anything, based on studies suggesting a poor regenerative environment in the far distal stump, near distal regions might be expected to offer a more favorable regenerative environment (). However, though both regions showed some laminar compaction consistent with Wallerian degeneration, there were dramatic differences in the inflammatory and structural environments between near and far distal stumps. Indeed, near distal stumps displayed increased connective tissue infiltration, a reduced laminar footprint, and reduced epineurial and laminar integrity. As also for the proximal stump, these observations likely reflect a gradual proximal to distal progression of inflammatory and fibrotic processes emerging from the site of nerve transection. Such changes would be expected to have a dramatic negative influence on any axons just entering the distal stump, as they grow towards their targeted end organ. The clinical impacts of this finding are twofold. First, the initial trimming back, or āfreshening,ā of the distal nerve stump by only a few mm during an initial repair, while providing a visually clean interface for securing a graft, may not, in fact, provide a favorable regenerative environment if the internal environment is not similarly āclean.ā This may provide one additional factor contributing to the poor response of a nerve to transection (Schmidt and Leach, 2003; Meek et al., 2005; Selim et al., 2022). Second, if this poor structural quality is indeed a result of the initial injury, though counterintuitive, it may be beneficial to more aggressively trim the distal stump, to encourage axonal outgrowth into a more supportive environment. Tension-based strategies may offset this additional pruning.
Limitations and implications for peripheral nerve regeneration
Our data support the use of nerve lengthening strategies for nerve regeneration. Nerves were successfully stretched using the device, and stably maintained under loading without slippage. Axons also robustly regenerated through an integrated conduit. Given the lack of any adverse phenotypes, even by the most conservative interpretation of the data, lengthened nerves were provided a 2Ā mm (20%) head start in closing a 10Ā mm nerve gap, or a 4Ā mm (40%) head start if assessed based on the length of nerve removed from the gap (12Ā mm vs. 8Ā mm).
While we successfully assessed the quality of regenerating axons and their surrounding structural and inflammatory environment, a limitation of our study is that we did not compare functional regeneration due to our focus on nerve response at early time points. Per a number of prior studies, summarized in (Wood et al., 2011), functional assessments are likely to be most valuable at later time points of 3ā4Ā months. Another limitation of our study was the observed partial collapse of several PLGA conduits mid-graft. While this did not affect the accuracy of our analysis or interpretation of our data, it did increase the variability and preclude the accurate quantification of axonal density and number mid-graft. Though PLGA has been successfully used to fabricate nerve regeneration scaffolds with modest structural and functional recovery (Oh and Lee, 2007; Oh et al., 2008), their biodegradability may render them fragile as well as elicit a local inflammatory response upon fragmentation (). Thus, a more robust, non-tubular conduit such as an autologous graft or an acellular allograft may offer a more structurally sound alternative for longer gaps. Alternately, given the poor quality of the near distal stump noted above, proposed dynamic lengthening would accelerate gap closure, despite the need for follow-up freshening and end to end repair to anastomose clean interfaces. Previous studies have demonstrated the feasibility and potential benefits of lengthening the proximal stump beyond the edge of the distal stump, such that during device removal, nerve stumps are trimmed back and reattached at a single interface, in the absence of a graft (Vaz et al., 2014; ; ). Overall, our work adds to mounting evidence that nerves can capably tolerate strain, and that nerve lengthening offers an attractive and novel approach to peripheral nerve repair.
Statements
Data availability statement
The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.
Ethics statement
The animal study was reviewed and approved by University of California, San Diego Institutional Animal Care and Use Committee.
Author contributions
T-HC and SS designed experiments. T-HC and KV designed and executed experiments. T-HC, EO, JN, KV, RL, and SS analyzed data. T-HC and SS authored the initial manuscript draft. All authors reviewed and edited the manuscript.
Acknowledgments
We gratefully acknowledge funding from Department of Defense (W81XWH-10-1-0773 and W81XWH-20-1-0510), National Science Foundation (CBET1042522), and the National Skeletal Muscle Research Center at UCSD. We also acknowledge helpful conversations with members of the Neuromuscular Bioengineering Laboratory and Dr. Justin Brown.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisherās note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AbramsR. A.ButlerJ. M.Bodine-FowlerS.BotteM. J. (1998). Tensile properties of the neurorrhaphy site in the rat sciatic nerve. J. Hand Surg. Am.23 (3), 465ā470. 10.1016/s0363-5023(05)80464-2
2
AnavaS.GreenbaumA.JacobE. B.HaneinY.AyaliA. (2009). The regulative role of neurite mechanical tension in network development. Biophys. J.96 (4), 1661ā1670. 10.1016/j.bpj.2008.10.058
3
ArnaoutoglouC. M.SakellariouA.VekrisM.MitsionisG. I.KorompiliasA.IoakimE.et al (2006). Maximum intraoperative elongation of the rat sciatic nerve with tissue expander: Functional, neurophysiological, and histological assessment. Microsurgery26 (4), 253ā261. 10.1002/micr.20236
4
AsriN. A. N.MahatM. M.ZakariaA.SafianM. F.Abd HamidU. M. (2022). Fabrication methods of electroactive scaffold-based conducting polymers for tissue engineering application: A review. Front. Bioeng. Biotechnol.10, 876696. 10.3389/fbioe.2022.876696
5
BazarekS.BrownJ. M.ShahS. B. (2022). Clinical potential of tension-lengthening strategies during nerve repair. Neural Regen. Res.17 (4), 779ā780. 10.4103/1673-5374.322461
6
BerisA. E.NakaK. K.SkopelitouA.KostaI.VragalasV.KonitsiotisS.et al (1996). Functional assessment of the rat sciatic nerve following intraoperative expansion: The effect of recovery duration on behavioural, neurophysiological, and morphological measures. Microsurgery17 (10), 568ā577. 10.1002/(sici)1098-2752(1996)17:10<568:aid-micr7>3.0.co;2-m
7
BhatiaA.DoshiP.KoulA.ShahV.BrownJ. M.SalamaM. (2017). Contralateral C-7 transfer: Is direct repair really superior to grafting?Neurosurg. Focus43 (1), E3. 10.3171/2017.4.focus1794
8
BirchJ. G.SamchukovM. L. (2004). Use of the Ilizarov method to correct lower limb deformities in children and adolescents. J. Am. Acad. Orthop. Surg.12 (3), 144ā154. 10.5435/00124635-200405000-00002
9
BoberB. G.GutierrezE.PlaxeS.GroismanA.ShahS. B. (2015). Combinatorial influences of paclitaxel and strain on axonal transport. Exp. Neurol.271, 358ā367. 10.1016/j.expneurol.2015.06.023
10
BrayD. (1984). Axonal growth in response to experimentally applied mechanical tension. Dev. Biol.102 (2), 379ā389. 10.1016/0012-1606(84)90202-1
11
BrownJ. M.ShahM. N.MackinnonS. E. (2009). Distal nerve transfers: A biology-based rationale. Neurosurg. Focus26 (2), E12. 10.3171/foc.2009.26.2.e12
12
BrownR.PedowitzR.RydevikB.WooS.HargensA.MassieJ.et al (1993). Effects of acute graded strain on efferent conduction properties in the rabbit tibial nerve. Clin. Orthop. Relat. Res.296 (296), 288ā294. 10.1097/00003086-199311000-00046
13
CarrM. M.BestT. J.MackinnonS. E.EvansP. J. (1992). Strain differences in autotomy in rats undergoing sciatic nerve transection or repair. Ann. Plast. Surg.28 (6), 538ā544. 10.1097/00000637-199228060-00008
14
ChereddyK. K.VandermeulenG.PreatV. (2016). PLGA based drug delivery systems: Promising carriers for wound healing activity. Wound Repair Regen.24 (2), 223ā236. 10.1111/wrr.12404
15
ChuangT. H.WilsonR. E.LoveJ. M.FisherJ. P.ShahS. B. (2013). A novel internal fixator device for peripheral nerve regeneration. Tissue Eng. Part C Methods19 (6), 427ā437. 10.1089/ten.tec.2012.0021
16
de AssisA. C. C.ReisA. L. S.NunesL. V.FerreiraL. F. R.BilalM.IqbalH. M. N.et al (2022). Stem cells and tissue engineering-based therapeutic interventions: Promising strategies to improve peripheral nerve regeneration. Cell Mol. Neurobiol. 10.1007/s10571-022-01199-3
17
DennerllT. J.LamoureuxP.BuxbaumR. E.HeidemannS. R. (1989). The cytomechanics of axonal elongation and retraction. J. Cell Biol.109 (6), 3073ā3083. 10.1083/jcb.109.6.3073
18
DunnJ. C.TadlockJ.KlahsK. J.NarimissaeiD.McKayP.NestiL. J. (2021). Nerve reconstruction using processed nerve allograft in the U.S. Military. Mil. Med.186 (5-6), e543āe548. 10.1093/milmed/usaa494
19
FaulF.ErdfelderE.LangA. G.BuchnerA. (2007). G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav. Res. Methods39 (2), 175ā191. 10.3758/bf03193146
20
FloresA. J.LaverniaC. J.OwensP. W. (2000). Anatomy and physiology of peripheral nerve injury and repair. Am. J. Orthop. (Belle Mead NJ)29 (3), 167ā173.
21
ForanI. M.HusseyV.PatelR. A.SungJ.ShahS. B. (2017). Native paraneurial tissue and paraneurial adhesions alter nerve strain distribution in rat sciatic nerves. J. Hand Surg. 43(3):316ā323. 10.1177/1753193417734433
22
GargS. P.HassanA. M.PatelA. A.PerezM. M.StoehrJ. R.KetheeswaranS.et al (2021). Outcomes of tibial nerve repair and transfer: A structured evidence-based systematic review and meta-analysis. J. Foot Ankle Surg.60 (6), 1280ā1289. 10.1053/j.jfas.2021.07.001
23
GeunaS.RaimondoS.RonchiG.Di ScipioF.TosP.CzajaK.et al (2009). Chapter 3: Histology of the peripheral nerve and changes occurring during nerve regeneration. Int. Rev. Neurobiol.87, 27ā46. 10.1016/S0074-7742(09)87003-7
24
GordonT. (2020). Peripheral nerve regeneration and muscle reinnervation. Int. J. Mol. Sci.21 (22), 8652. 10.3390/ijms21228652
25
GrinsellD.KeatingC. P. (2014). Peripheral nerve reconstruction after injury: A review of clinical and experimental therapies. Biomed. Res. Int.2014, 1ā13. 10.1155/2014/698256
26
GuptaR.RowshanK.ChaoT.MozaffarT.StewardO. (2004). Chronic nerve compression induces local demyelination and remyelination in a rat model of carpal tunnel syndrome. Exp. Neurol.187 (2), 500ā508. 10.1016/j.expneurol.2004.02.009
27
GuptaR.NassiriN.HazelA.BathenM.MozaffarT. (2012). Chronic nerve compression alters Schwann cell myelin architecture in a murine model. Muscle Nerve45 (2), 231ā241. 10.1002/mus.22276
28
HaraY.NishiuraY.OchiaiN.SharulaNakajimaY.KubotaS.et al (2012). New treatment for peripheral nerve defects: Reconstruction of a 2?cm, monkey median nerve gap by direct lengthening of both nerve stumps. J. Orthop. Res.30 (1), 153ā161. 10.1002/jor.21476
29
HentzV. R.RosenJ. M.XiaoS. J.McGillK. C.AbrahamG. (1993). The nerve gap dilemma: A comparison of nerves repaired end to end under tension with nerve grafts in a primate model. J. Hand Surg. Am.18 (3), 417ā425. 10.1016/0363-5023(93)90084-g
30
HowarthH. M.AlazizT.NicoldsB.O'ConnorS. (2019). Redistribution of nerve strain enables end-to-end repair under tension without inhibiting nerve regeneration. Neural Regen. Res.14 (7), 1280ā1288. 10.4103/1673-5374.251338
31
HowarthH. M.KadoorA.SalemR.NicoldsB.AdachiS.KanarisA.et al (2019). Nerve lengthening and subsequent end-to-end repair yield more favourable outcomes compared with autograft repair of rat sciatic nerve defects. J. Tissue Eng. Regen. Med.13 (12), 2266ā2278. 10.1002/term.2980
32
HowarthH. M.OrozcoE.LoveringR. M.ShahS. B. (2020). A comparative assessment of lengthening followed by end-to-end repair and isograft repair of chronically injured peripheral nerves. Exp. Neurol.331, 113328. 10.1016/j.expneurol.2020.113328
33
JiangB.ZhangP.YanJ.ZhangH. (2008). Dynamic observation of biomechanic properties of sciatic nerve at the suture site in rats following repairing. Artif. Cells Blood Substit. Immobil. Biotechnol.36 (1), 45ā50. 10.1080/10731190701857777
34
KecheleP. R.BertelliJ. A.DalmarcoE. M.FrodeT. S. (2011). The mesh repair: Tension free alternative on dealing with nerve gaps-experimental results. Microsurgery31 (7), 551ā558. 10.1002/micr.20902
35
LiY.MaZ.RenY.LuD.LiT.LiW.et al (2021). Tissue engineering strategies for peripheral nerve regeneration. Front. Neurol.12, 768267. 10.3389/fneur.2021.768267
36
LoverdeJ. R.OzokaV. C.AquinoR.LinL.PfisterB. J. (2011). Live imaging of axon stretch growth in embryonic and adult neurons. J. Neurotrauma28 (11), 2389ā2403. 10.1089/neu.2010.1598
37
LoverdeJ. R.PfisterB. J. (2015). Developmental axon stretch stimulates neuron growth while maintaining normal electrical activity, intracellular calcium flux, and somatic morphology. Front. Cell Neurosci.9, 308. 10.3389/fncel.2015.00308
38
LundborgG.RydevikB. (1973). Effects of stretching the tibial nerve of the rabbit. A preliminary study of the intraneural circulation and the barrier function of the perineurium. J. Bone Jt. Surg. Br.55 (2), 390ā401. 10.1302/0301-620x.55b2.390
39
MaedaT.HoriS.SasakiS.MaruoS. (1999). Effects of tension at the site of coaptation on recovery of sciatic nerve function after neurorrhaphy: Evaluation by walking-track measurement, electrophysiology, histomorphometry, and electron probe X-ray microanalysis. Microsurgery19 (4), 200ā207. 10.1002/(sici)1098-2752(1999)19:4<200:aid-micr7>3.0.co;2-y
40
MaengW. Y.TsengW-L.LiS.KooJ.HsuehY-Y. (2022). Electroceuticals for peripheral nerve regeneration. Biofabrication14. 10.1088/1758-5090/ac8baa
41
MasonS.PhillipsJ. B. (2011). An ultrastructural and biochemical analysis of collagen in rat peripheral nerves: The relationship between fibril diameter and mechanical properties. J. Peripher Nerv. Syst.16 (3), 261ā269. 10.1111/j.1529-8027.2011.00352.x
42
MeekM. F.CoertJ. H.RobinsonP. H. (2005). Poor results after nerve grafting in the upper extremity: Quo vadis?Microsurgery25 (5), 396ā402. 10.1002/micr.20137
43
MillesiH. (1981). Reappraisal of nerve repair. Surg. Clin. North Am.61 (2), 321ā340. 10.1016/s0039-6109(16)42384-4
44
MillesiH. (1986). The nerve gap. Hand Clin.2 (4), 651ā663. 10.1016/s0749-0712(21)00614-4
45
MuellerM. J.MalufK. S. (2002). Tissue adaptation to physical stress: A proposed "physical stress theory" to guide physical therapist practice, education, and research. Phys. Ther.82 (4), 383ā403. 10.1093/ptj/82.4.383
46
OgataK.NaitoM. (1986). Blood flow of peripheral nerve effects of dissection, stretching and compression. J. Hand Surg. Br.11 (1), 10ā14. 10.1016/0266-7681_86_90003-3
47
OhS. H.KimJ. H.SongK. S.JeonB. H.YoonJ. H.SeoT. B.et al (2008). Peripheral nerve regeneration within an asymmetrically porous PLGA/Pluronic F127 nerve guide conduit. Biomaterials29 (11), 1601ā1609. 10.1016/j.biomaterials.2007.11.036
48
OhS. H.LeeJ. H. (2007). Fabrication and characterization of hydrophilized porous PLGA nerve guide conduits by a modified immersion precipitation method. J. Biomed. Mater Res. A80 (3), 530ā538. 10.1002/jbm.a.30937
49
PetersB. R.WoodM. D.HunterD. A.MackinnonS. E. (2021). Acellular nerve allografts in major peripheral nerve repairs: An analysis of cases presenting with limited recovery. Hand (N Y), 15589447211003175. 10.1177/15589447211003175
50
PfisterB. J.IwataA.MeaneyD. F.SmithD. H. (2004). Extreme stretch growth of integrated axons. J. Neurosci.24 (36), 7978ā7983. 10.1523/jneurosci.1974-04.2004
51
PhillipsJ. B.SmitX.ZoysaN. D.AfokeA.BrownR. A. (2004). Peripheral nerves in the rat exhibit localized heterogeneity of tensile properties during limb movement. J. Physiol.557 (3), 879ā887. 10.1113/jphysiol.2004.061804
52
PoongodiR.ChenY. L.YangT. H.HuangY. H.YangK. D.LinH. C.et al (2021). Bio-scaffolds as cell or exosome carriers for nerve injury repair. Int. J. Mol. Sci.22 (24), 13347. 10.3390/ijms222413347
53
RayC. D.MaurerD. D. (1975). Electrical neurological stimulation systems: A review of contemporary methodology. Surg. Neurol.4 (1), 82ā90.
54
RbiaN.BulstraL. F.SaffariT. M.HoviusS. E.ShinA. Y. (2019). Collagen nerve conduits and processed nerve allografts for the reconstruction of digital nerve gaps: A single-institution case series and review of the literature. World Neurosurg.127, e1176āe1184. 10.1016/j.wneu.2019.04.087
55
SafaB.JainS.DesaiM. J.GreenbergJ. A.NiacarisT. R.NydickJ. A.et al (2020). Peripheral nerve repair throughout the body with processed nerve allografts: Results from a large multicenter study. Microsurgery40 (5), 527ā537. 10.1002/micr.30574
56
SaijilafuNishiuraY.YamadaY.HaraY.IchimuraH.YoshiiY.et al (2006). Repair of peripheral nerve defect with direct gradual lengthening of the proximal nerve stump in rats. J. Orthop. Res.24 (12), 2246ā2253. 10.1002/jor.20280
57
SchmidtC. E.LeachJ. B. (2003). Neural tissue engineering: Strategies for repair and regeneration. Annu. Rev. Biomed. Eng.5, 293ā347. 10.1146/annurev.bioeng.5.011303.120731
58
SelimO. A.LakhaniS.MidhaS.MosahebiA.KalaskarD. M. (2022). Three-dimensional engineered peripheral nerve: Toward a new era of patient-specific nerve repair solutions. Tissue Eng. Part B Rev.28 (2), 295ā335. 10.1089/ten.teb.2020.0355
59
SharulaHaraY.NishiuraY.SaijilafuKubotaS.OchiaiN. (2010). Repair of the sciatic nerve defect with a direct gradual lengthening of proximal and distal nerve stumps in rabbits. Plast. Reconstr. Surg.125 (3), 846ā854. 10.1097/prs.0b013e3181ccdbd4
60
SiechenS.YangS.ChibaA.SaifT. (2009). Mechanical tension contributes to clustering of neurotransmitter vesicles at presynaptic terminals. Proc. Natl. Acad. Sci. U. S. A.106 (31), 12611ā12616. 10.1073/pnas.0901867106
61
SimpsonA. H.GillingwaterT. H.AndersonH.CottrellD.ShermanD. L.RibchesterR. R.et al (2013). Effect of limb lengthening on internodal length and conduction velocity of peripheral nerve. J. Neurosci.33 (10), 4536ā4539. 10.1523/jneurosci.4176-12.2013
62
SinghV. K.HaqA.TiwariM.SaxenaA. K. (2022). Approach to management of nerve gaps in peripheral nerve injuries. Injury53 (4), 1308ā1318. 10.1016/j.injury.2022.01.031
63
SunderlandI. R.BrennerM. J.SinghamJ.RickmanS. R.HunterD. A.MackinnonS. E. (2004). Effect of tension on nerve regeneration in rat sciatic nerve transection model. Ann. Plast. Surg.53 (4), 382ā387. 10.1097/01.sap.0000125502.63302.47
64
SungJ.SikoraāKlakJ.AdachiS. Y.OrozcoE.ShahS. B. (2019). Decoupled epineurial and axonal deformation in mouse median and ulnar nerves. Muscle Nerve59 (5), 619ā628. 10.1002/mus.26437
65
TanoueM.YamagaM.IdeJ.TakagiK. (1996). Acute stretching of peripheral nerves inhibits retrograde axonal transport. J. Hand Surg. Br.21 (3), 358ā363. 10.1016/s0266-7681(05)80203-7
66
VazK. M.BrownJ. M.ShahS. B. (2014). Peripheral nerve lengthening as a regenerative strategy. Neural Regen. Res.9 (16), 1498ā1501. 10.4103/1673-5374.139471
67
WallE. J.MassieJ.KwanM.RydevikB.MyersR.GarfinS. R. (1992). Experimental stretch neuropathy. Changes in nerve conduction under tension. J. Bone Jt. Surg. Br.74 (1), 126ā129. 10.1302/0301-620x.74b1.1732240
68
WoodM. D.KempS. W.WeberC.BorschelG. H.GordonT. (2011). Outcome measures of peripheral nerve regeneration. Ann. Anat.193 (4), 321ā333. 10.1016/j.aanat.2011.04.008
69
WrightT. W.GlowczewskieF.CowinD.WheelerD. L. (2001). Ulnar nerve excursion and strain at the elbow and wrist associated with upper extremity motion. J. Hand Surg. Am.26 (4), 655ā662. 10.1053/jhsu.2001.26140
70
YousefM. A.DionigiP.MarconiS.CalligaroA.CornagliaA. I.AlfonsiE.et al (2015). Successful reconstruction of nerve defects using distraction neurogenesis with a new experimental device. Basic Clin. Neurosci.6 (4), 253ā264.
71
ZengZ.YangY.DengJ.Saif Ur RahmanM.SunC.XuS. (2022). Physical stimulation combined with biomaterials promotes peripheral nerve injury repair. Bioeng. (Basel)9 (7), 292. 10.3390/bioengineering9070292
72
ZhengJ.LamoureuxP.SantiagoV.DennerllT.BuxbaumR.HeidemannS. (1991). Tensile regulation of axonal elongation and initiation. J. Neurosci.11 (4), 1117ā1125. 10.1523/jneurosci.11-04-01117.1991
Summary
Keywords
peripheral nerve, tensile loading, nerve injury and regeneration, axon, basal lamina, extracellar matrix, PLGA, conduit
Citation
Chuang T-H, Orozco E, Nam JJ, Vaz K, Lovering RM and Shah SB (2023) Injured nerves respond favorably to an integrated tension- and conduit-based regenerative strategy. Front. Front. Biomater. Sci. 2:1041018. doi: 10.3389/fbiom.2023.1041018
Received
10 September 2022
Accepted
18 January 2023
Published
02 February 2023
Volume
2 - 2023
Edited by
Silvia Fare', Politecnico di Milano, Italy
Reviewed by
Vincent Fitzpatrick, Tufts University, United States
Lorenza Draghi, Politecnico di Milano, Italy
Updates
Copyright
Ā© 2023 Chuang, Orozco, Nam, Vaz, Lovering and Shah.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Sameer B. Shah, sbshah@health.ucsd.edu
This article was submitted to Biomaterials Science for Regenerative Therapies, a section of the journal Frontiers in Biomaterials Science
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.