ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 13 February 2017
Sec. Cardiac Rhythmology
Volume 4 - 2017 | https://doi.org/10.3389/fcvm.2017.00004

Loop Recorder Detected High Rate of Atrial Fibrillation Recurrence after a Single Balloon- or Basket-Based Ablation of Paroxysmal Atrial Fibrillation: Results of the MACPAF Study

imageAlexander Schirdewan1* imageJuliane Herm2,3† imageMattias Roser4 imageUlf Landmesser4 imageMatthias Endres2,3,5,6 imageLydia Koch1 imageKarl Georg Haeusler2,3
  • 1Department of Cardiology, Sana Clinic Lichtenberg, Berlin, Germany
  • 2Department of Neurology, Charité – Universitätsmedizin Berlin, Berlin, Germany
  • 3Center for Stroke Research Berlin, Charité – Universitätsmedizin Berlin, Berlin, Germany
  • 4Department of Cardiology and Pneumology, Charité – Universitätsmedizin Berlin, Berlin, Germany
  • 5German Center for Neurodegenerative Diseases (DZNE), Partner Site, Berlin, Germany
  • 6German Center for Cardiovascular Diseases (DZHK), Partner Site, Berlin, Germany

Purpose: Pulmonary vein isolation (PVI) is an established approach to treat symptomatic non-permanent atrial fibrillation (AF). Detecting AF recurrence after PVI is important, if discontinuation of oral anticoagulation after ablation is considered.

Methods: Patients with symptomatic paroxysmal AF were enrolled in the prospective randomized mesh ablator vs. cryoballoon pulmonary vein (PV) ablation of symptomatic paroxysmal AF study, comparing efficacy and safety of the HD Mesh Ablator® (C.R. Bard, Lowell, MA, USA) and the Arctic Front® (Medtronic, Minneapolis, MN, USA) catheter. Rhythm status post-PVI was closely monitored for 1 year using the implantable loop recorder (ILR) Reveal XT® (Medtronic Minneapolis, MN, USA).

Results: The study was terminated after the first interim analysis due to the inability of the HD Mesh Ablator® to achieve the predefined primary study endpoint, an exit block of all PVs. After a 90-day blanking period, 23 (62.2%) out of 37 study patients (median 63.0 years; 41% females) had at least one episode of AF. AF recurrence was associated with AF episodes during the blanking period {hazard ratios (HR) 5.10 [95% confidence interval (CI) 1.21–21.4]; p = 0.038}, and a common left-sided PV ostium [HR 4.17 (95%CI 1.48–11.8); p = 0.039] but not with catheter type, age, gender, cardiovascular risk profile, or left atrial volume. There was a trend toward AF recurrence in patients without complete PVI of all PV (p = 0.095). Overall, 337 (59.4%) out of 566 ILR-detected episodes represented AF. Comparing patients with AF recurrence to those without, there was no difference in cognitive performance 6 months post-ablation.

Conclusion: Using an ILR, in more than 60% of all patients with paroxysmal AF, a recurrence of AF was detected within 12 months after ablation. In patients with a common PV ostium, the first generation balloon-based catheter is obviously less effective.

Clinical trials: http://Clinicaltrials.gov NCT01061931.

Introduction

Atrial fibrillation (AF) increases stroke risk, impacts quality of life, and is associated with cognitive decline (13). Pulmonary vein isolation (PVI) is now an established therapeutic approach in patients with symptomatic paroxysmal or persistent AF (2). Besides reducing AF-related symptoms, ablation of pulmonary vein (PV) significantly improves quality of life (4).

The ability of PVI to assure continuous sinus rhythm (SR) on the long term is limited (57) but can be improved by repetitive ablation procedures. Reported “success rates” in AF patients strictly depend on AF type, patient characteristics, ablation techniques, and the extent of ECG recording after ablation (810). The implantable loop recorder (ILR) Reveal XT® (Medtronic, Minneapolis, MN, USA) is a promising tool to monitor AF recurrence by detecting paroxysmal AF with a sensitivity of 96% and specificity of 85% (11), and it is more precise to detect AF recurrence after PVI compared to repetitive Holter monitoring (9). Assessment of AF recurrence is of paramount importance in AF patients with low or intermediate stroke risk if discontinuation of oral anticoagulation is considered (12). Moreover, AF recurrence is important to establish the (technical) efficacy and anatomical limitations of new (especially single shot) devices (13).

By implanting the ILR Reveal XT® before PVI within the randomized “mesh ablator vs. cryoballoon pulmonary vein ablation of symptomatic paroxysmal atrial fibrillation” (MACPAF) study, we aimed to evaluate whether the cryoballoon Arctic Front® (Medtronic, Minneapolis, MN, USA) or the radiofrequency-based HD Mesh Ablator® catheter (C.R. Bard, Lowell, MA, USA) is able to achieve continuous SR after a single PVI procedure (14). As reported previously, the Arctic Front® catheter proved to be superior by achieving a bidirectional block, entrance, and “exit block” (EB) of all PVs (15). We here report the ILR data during a 12-month follow-up as well as the impact of AF recurrence on cognitive performance.

Materials and Methods

Study Design and Study Population

The design of the MACPAF study was previously reported in detail (14, 15). In short, the study’s safety board terminated the study prematurely due to the inability of the HD Mesh Ablator® to achieve the predefined primary study endpoint, an EB of all PVs. Of the intended 108 patients with symptomatic paroxysmal AF (with prior ineffective antiarrhythmic drug treatment, no previous PVI, no unstable structural heart disease), 37 patients were randomized for Arctic Front® or HD Mesh Ablator® catheter ablation. The balloon shaped catheter Arctic Front® (Medtronic, Inc.) uses cryoenergy, whereas the basket shaped catheter device HD Mesh Ablator® (C.R. Bard, Inc.) is based on unique pulsed radiofrequency delivery. In addition, the latter provides the possibility of circumferential mapping, leading to a reduced procedure duration as well as a reduced fluoroscopy time (16).

Study patients underwent PV ablation according to study criteria and were followed up 3, 6, 9, and 12 months afterward. Stable cardiac conditions were ensured before ablation by performing echocardiography, cardiovascular stress test, and (if indicated) coronary angiography. Using MRI or CT datasets, the LA-PVs were segmented to generate 3D LA-PV surface reconstructions. In MACPAF, PV ostial and antral regions were evaluated. A common left-sided ostium was defined as a common ostial circumference with left PVs branching >3 mm away from the common ostium (17).

A comprehensive cognitive testing was carried out before PVI and during the 6 months visit (18), including verbal and non-verbal learning (Rey–Osterrieth complex figure test and the Rey Auditory Verbal Learning Test), short term memory (forward digit-span task), attention and executive function (trail-making test A and B, Stroop test, category and letter fluency, digit-span backward), and reasoning (German Leistungsprüfsystem 50). The ILR Reveal XT® (Medtronic Minneapolis, MN, USA) was implanted after enrollment in 33 (89%) out of 37 study patients to assess AF recurrence after ablation.

Analysis of ECG Data

Atrial fibrillation was defined as absolute arrhythmia without distinguishable P waves lasting longer than 30 s. AF recurrence post-ablation was defined as one AF episode after a 90-day blanking period, verified by a board-certified cardiologist (LK, AS) blinded for patient reported symptoms during follow-up. Rhythm status was monitored using the ILR Reveal XT® (n = 31) or repetitive Holter recordings for at least 24 h (range 1–7 days; n = 6), respectively. The subcutaneously inserted Reveal XT® is able to store up to 49.5 min of automatically or patient activated ECG recording. Holter recording and Reveal XT® data download were scheduled on the 3, 6, 9, and 12 months visit. The duration of available ECG recording was defined as time from first recording of the ILR to the scheduled read-out. AF burden was defined as total duration of all verified AF episodes during ECG recording.

Statistical Analysis

For categorical traits, absolute and relative frequencies were computed. In the case of continuous or quasi-continuous variables with nearly symmetric distribution, the arithmetic mean, SD, minimal and maximal values, otherwise median, quartiles as well as minimal and maximal values were computed. Fisher’s exact test was used to compare proportions for dichotomous outcomes between independent groups or to test independency of two dichotomous traits within a population. The Mann–Whitney U test was applied to analyze not normally distributed variables. In order to analyze predictors of AF recurrence, AF-free survival time was computed. Univariate Cox proportional hazards analysis and log-rank test (Mantel Cox) were used for outcome analyses of AF-free survival. Hazard ratios (HR) and 95% confidence intervals (CI) are presented. The survival function was estimated by the Kaplan–Meier method (product-limit estimator). Due to the limited number of patients, all statistical tests have to be regarded as explorative. p-Values (significance level 0.05) are not adjusted for multiple testing.

Results

Patients’ Baseline Characteristics

The baseline data of the 37 study patients can be found in the Supplementary Material. In short, median age was 63.0 years (IQR 56.5–68.0), 40.5% were females and the median CHA2DS2-VASc score was 2.0 (IQR 1.0–3.0). All patients were available for the follow-up after 180 days; one (2.7%) patient with persisting SR was lost to follow-up afterward. No patient reported a clinically evident ischemic stroke or transient ischemic attack within 1 year post-PVI.

Assessment of Rhythm Status and AF Recurrence during Follow-up

Due to local infection in two (5.4%) of 33 patients, these ILRs had to be explanted 19 and 35 days after implantation, respectively. Therefore, overall six (16.2%) study patients underwent serial Holter–ECG monitoring [median duration 5.0 days (IQR 1.0–5.0, range 1–7)] during the 1-year follow-up. After the blanking period of 90 days, 23 (62.2%) study patients had an ECG documented episode of AF within the next 270 days. In these 23 patients, the first documented AF episode occurred within 90 and 179 days in 17 (73.9%) patients, within 180 and 269 days in 5 (21.7%), and within 270 and 365 days in 1 patient (4.3%). AF recurrence was detected in 21 patients by using the ILR and in 2 patients by Holter monitoring. Within 90 days post-PVI, 22 (59.5%) of 37 study patients had at least 1 episode of AF. Five (22.7%) of these 22 study patients had documented AF only during the 90-day blanking period and were therefore considered to have continuous SR. Of all 23 study patients with recurrent AF, 13 (56.5%) underwent re-ablation within 1 year after first PVI.

Predictors of AF Recurrence

There was no difference regarding age, sex, or cardiovascular risk profile in patients with or without AF recurrence within 3–12 months post-PVI (Table 1). In addition, there was no impact of absence of any cardiovascular risk factor (“lone AF”) on AF recurrence. However, study patients with a common left-sided PV ostium were significantly more liable to suffer AF recurrence [HR 4.17 (95%CI 1.48–11.8); p = 0.039]. According to Kaplan–Meier analysis, AF-free survival is not related to the used catheter type (p = 0.547; Figure 1). There was a non-significant trend for patients with an EB of all PVs to be in continuous SR at 1-year follow-up [HR 0.46 (95%CI 0.19–1.14); p = 0.095] (Table 1). AF recurrence within the 90-day blanking period was significantly associated with AF recurrence during 90 and 365 days [Table 1; HR 5.10 (95%CI 1.21–21.4); p = 0.038].

TABLE 1
www.frontiersin.org

Table 1. Predictors of atrial fibrillation (AF) recurrence in patients with paroxysmal AF within 90 and 365 days after ablation.

FIGURE 1
www.frontiersin.org

Figure 1. Atrial fibrillation (AF)-free survival after pulmonary vein isolation (PVI) according to (A) achieved exit block of all pulmonary veins (PVs) during first PVI (p = 0.086) and (B) used catheter device (p = 0.547).

Reliability of AF Detection According to the ILR Reveal XT®

The ILR Reveal XT® stored a median of 15.0 (IQR 6.0–63.0; range 1–1,040) episodes of suspected AF per patient and follow-up visit. Median duration of assessable ECG monitoring was 48.1 days (IQR 14.9–84.2) per follow-up. According to the algorithm of the ILR (11), a median AF burden of 0.55% (IQR 0.1–4.8) was present in the automated analysis. Overall, 337 (59.4%) of 566 episodes recorded by the ILR were verified as AF by the involved cardiologists. Frequent premature beats, under and oversensing of R waves, and a high electrical noise factor caused misclassification in the remaining episodes. A median of 9.0 (IQR 4.0–10.0) AF episodes per patient and follow-up was validated. Median duration of assessable Holter monitoring was 48.1 days (IQR 14.9–84.2) per follow-up. Median burden of verified AF in ILR patients was 0.21% (IQR 0–10.0) pre-ablation, 0.0% (IQR 0–4.5) on day 180, and 0.0% (IQR 0–2.0) on day 365, while median burden of automatically detected AF episodes by the ILR was 0.50% (IQR 0.1–9.1) pre-ablation, 0.1% (IQR 0.1–3.8) on day 180, and 0.2% (IQR 0.1–2.1) on day 365 (Figure 2). The Wilcoxon signed-rank test indicated a reduction of the verified AF burden (p = 0.05) as well as the automatically detected AF burden after 365 days (p = 0.034) compared to the pre-ablation period (Figure 2).

FIGURE 2
www.frontiersin.org

Figure 2. Boxplots of (A) automatically detected atrial fibrillation (AF) burden (%) by the implantable loop recorder and (B) computed AF burden of verified AF episodes pre-ablation and during follow-up. p values computed by the Wilcoxon signed-rank test (dotted line pre-ablation vs. day 180 post-ablation and solid line pre-ablation vs. day 365 post-ablation).

Symptomatic AF and Medication during Follow-up

Atrial fibrillation-related symptoms like palpitations and dizziness were reported by 13 (76.5%) of 17 patients with AF recurrence on day 180. In addition, 4 (20%) of 20 patients without AF recurrence reported similar symptoms on day 180. On day 365, such symptoms were reported in 13 (56.5%) of 23 patients with AF recurrence and in 3 (23.1%) of 13 patients without (p = 0.083). In patients with “symptomatic” AF recurrence, recorded AF burden was significantly higher compared to those with “asymptomatic” AF recurrence [median 0.2% (IQR 0.08–0.48) vs. median 2.5% (IQR 0.55–5.28); p = 0.010].

On 1-year follow-up, 11 (31.4%) of all 36 study patients received at least one antiarrhythmic drug (2.9% amiodarone, 11.4% dronedarone, 17.1% other). Only one (9.1%) out of 13 patients without AF recurrence still received antiarrhythmic drugs. In addition, 27 (77.1%) of 36 patients were on oral anticoagulation, while 7 (19.4%) received an antiplatelet agent.

Impact of AF Recurrence on Cognitive Performance

Neuropsychological testing was carried out in 36 (97%) of 37 patients during the 6-month follow-up visit. No patient complained about cognitive impairment during follow-up. Comparing patients with AF recurrence between 90 and 180 days to those without revealed no significant differences regarding all tested domains (Table 2).

TABLE 2
www.frontiersin.org

Table 2. Neuropsychological assessment pre-ablation and 6 months post-ablation according to atrial fibrillation (AF) recurrence.

Discussion

Based on the here reported long-term results of the MACPAF study, the following main findings can be drawn from the study: using an ILR, 62% of all study patients with paroxysmal AF had recurrent AF within 90 days and 1-year post-PVI. AF recurrence during the 90-day blanking period was associated with a fivefold higher risk of AF recurrence. Patients with a common left-sided PV ostium had a fourfold higher risk of AF recurrence after a single ablation procedure, indicating that a first generation balloon- or basket-based catheter is less effective in these patients. There was no apparent impact of AF recurrence on cognitive performance 6 months post-PVI as determined by testing attention, executive functions, short term memory, verbal and non-verbal learning as well as reasoning.

In recent years, multiple innovative technical solutions for left atrial catheter ablation of symptomatic AF have been developed such as the balloon-based Arctic Front® catheter (using cryoenergy) or the basket-based HD Mesh Ablator® catheter (using pulsed radiofrequency), providing the opportunity to induce circumferential scars around the PV ostia (antral aspect) (19). However, the missing ability of the investigated first generation devices to do focal ablation is of practical importance regarding efficacy. The limited device efficacy and the relatively small patient cohort may explain the missing statistical significance regarding AF-free survival which was observed in other studies (20). A statistical trend toward less AF recurrences in patients with EB of all PVs further supports that transmural permanent PVI is an important factor of long-term success.

While these first generation devices obviously did not fit to all distinct PV anatomies, a substantial proportion of patients with paroxysmal AF suffered from mostly asymptomatic AF recurrences after PVI (2123), and MACPAF patients with a common left-sided PV ostium were four times more liable to suffer AF recurrence [HR 4.17 (95%CI 1.48–11.8)]. Our results confirm an early prediction by Ahmed et al. that—based on the analysis of 3D surface reconstructions of LA-PV anatomy from MRI datasets in 101 patients—balloon or basket catheter-based ablation of the common PV antra would be problematic (17). This anatomical variant was found in 14% of all MACPAF patients compared to a prevalence of 2–7% in other ablation cohorts. This probably explains the missing association with AF recurrence in studies using similar ablation catheters (16, 24).

Analyzing the high rate of AF recurrence, the achieved results by the first generation cryoballoon are comparable to radiofrequency systems. The improved engineering of the second-generation cryoballoon leads to substantially higher efficacy also in patients with anatomical PV variants (25, 26). On the other hand, the inefficacy of the basket-based Mesh catheter raised serious concerns regarding the underlying technical concept.

Confirming previous studies (27, 28), patients with AF recurrence during the (well established) 90-day blanking period had a fivefold higher risk for later AF recurrence. However, about 23% of all study patients had documented AF only during the 90-day blanking period and were considered to have persisting SR after PVI. This finding is not in line with a previous study on 35 patients with paroxysmal AF (29), reporting AF recurrence in all 13 patients with AF during the blanking period. In conclusion, we do not think that the so-called blanking period loses its intrinsic meaning. We suggest that (even asymptomatic) patients with documented AF recurrence within 90 days after ablation should undergo intense ECG monitoring if the CHA2DS2-VASc score is ≥2 and long-term oral anticoagulation—despite current guideline recommendations (2, 30, 31)—is not intended.

The method and intensity of rhythm surveillance has become a major topic of post-ablational care. As an example, the randomized RAAFT 2 trial demonstrated a recurrence rate of AF/atrial flutter/atrial tachycardia of 24% by using standard Holter ECGs and a recurrence rate of 47% in the 2-year follow-up by using an additional transtelephonic monitoring (32). Available prospective ILR-studies on AF recurrence after PVI are mostly based on single-center experience, report on a limited number of AF patients, and are heterogeneous regarding AF type, definition of AF recurrence, ablation technique as well as duration of follow-up (9, 10, 29, 32, 33). Subsequently, there are few comparable data to the here reported AF recurrence rate of 62% within 1 year after a single balloon-based PVI of symptomatic paroxysmal AF. Using a similar definition of AF recurrence, Pedrote et al. reported an AF recurrence in 43% of 35 patients with paroxysmal AF within 1 year after a single PVI using radiofrequency (29). A recent propensity score-matched comparison of radiofrequency or cryoballoon ablation in 142 patients with paroxysmal AF reported similar long-term recurrence rates of 44 or 52% after a single ablation procedure, respectively (34). By using a threshold of “percentage of time spent in AF <0.5%” for considering a patient “free” of AF, a recurrence rate of 32% was reported for similar cohorts of patients with paroxysmal AF undergoing a single ablation procedure using radiofrequency (10, 33). In this context, the feasibility of discontinuation of oral anticoagulation within months post-PVI in AF patients with at least moderate stroke risk remains disputable.

While the use of ILRs is feasible (9, 10, 21, 29, 33) and more effective compared to repetitive Holter monitoring (9), automated recording of AF by the ILR Reveal XT® is prone to false positive results (9). Our data revealed that about 60% of the ILR-detected “AF”-episodes could be verified by board-certified cardiologists. As similarly reported by Kapa et al. (9), main causes of false-positive AF detection by the ILR were frequent atrial and ventricular extrasystoles, undersensing of R waves, oversensing of R and T waves, and a high electrical noise factor. ILR misclassification (41% false-positive) was comparable to Kapa et al. [54% false-positive (9)] and Eitel et al. [38% false-positive (35)] but higher compared to Schmidt et al. [26% false-positive (36)]. While adjusting of the diagnostic algorithm has previously proven to improve specificity (35), further improvements are needed. Despite of being technically outdated by the Reveal LINQ®, socioeconomic reasons are likely to lead to an ongoing use of the Reveal XT®. Next to establishing continuous SR, control of AF-related symptoms is the main goal of ablation therapy. In our cohort, 43% of patients with AF recurrence reported to be free of AF-related symptoms after 12 months. AF burden was significantly reduced in those patients after a single ablation procedure, confirming the concept of single shot devices in principle. This confirms previous studies reporting a reduced rate of symptomatic AF and an improved quality of life even in ablated patients without continuous SR (4, 23).

In addition to AF-related symptoms, cognitive decline has been reported in AF patients (3, 37, 38) as well as in AF patients post-ablation (39). Whether continuous SR post-AF ablation might have a positive impact on cognitive function has not yet been reported. Our results demonstrate that rhythm status had no significant impact on cognitive function over a period of 180 days after ablation. However, large prospective multicenter studies are warranted to clarify this matter.

The study has several strengths but also limitations. While the study design is unique, the premature termination of the study limits the significance of the reported results. Subsequently, we cannot definitively prove that there was no impact of the absence of any cardiovascular risk factor (“lone AF”) on AF recurrence. In addition, PVI-related parameters such as procedure duration or the used catheter type in this randomized study had no impact on AF recurrence, despite of the fact that the HD Mesh Ablator® catheter was inferior to the Arctic Front® catheter in achieving an EB of all PVs (15). Furthermore, MACPAF results cannot be generalized to patients with persistent AF or distinct catheter types. In addition, the vast majority but not all patients were monitored by an ILR, and one patient was lost to follow-up after the 6 months visit.

However, the MACPAF results deliver relevant insights into important practical aspects of ILR-based follow-up performance. Moreover, our findings indicate that first generation balloon-based devices should not be used in patients with a common PV ostium.

Conclusion

Using an ILR in 62% of all study patients with paroxysmal AF a recurrence of AF was detected within 90 days and 12 months after a single balloon-based ablation procedure. While AF burden was diminished after a single ablation procedure in general, about one third of patients with AF recurrence still suffered from symptomatic AF. Detected AF recurrence during the blanking period of 90 days post-ablation renders later AF recurrence fivefold more likely. A common left-sided PV ostium was associated with a fourfold higher risk of AF recurrence, indicating that a single balloon-based ablation procedure is less effective in these patients. Despite the fact that AF recurrence post-ablation had no significant impact on cognitive function 6 months after PVI, our study clearly demonstrates that intensive screening for AF recurrence after ablation is of major importance.

Ethics Statement

The study has been approved by the local Ethics Committee (EA4/087/08). All the study participants gave written informed consent for study participation and scientific use of data.

Author Contributions

AS has made substantial contributions to conception and design, analysis and interpretation of data, and drafted the manuscript. JH has made substantial contributions to analysis and interpretation of data, and drafted the manuscript. LK has made substantial contributions to conception and design, analysis of data, and revised the manuscript critically for important intellectual content. MR, UL, and ME revised the manuscript critically for important intellectual content. KGH has made substantial contributions to conception and design, analysis and interpretation of data, and drafted the manuscript.

Conflict of Interest Statement

AS reports lecture fees and prior study grants by Medtronic, C.R. Bard, and Biotronik. ME reports funding from Bayer and fees paid to the Charité from Amgen, Bayer Healthcare, BI, BMS, EVER, GSK, Pfizer, Novartis, and Sanofi. LK reports lecture fees by Medtronic and a Biotronik-sponsored fellowship. UL received consultant or lecture honoraria from St. Jude. KGH reports lecture fees and study grants by Bayer Healthcare and Sanofi as well as lecture fees from Pfizer and Bristol-Myers Squibb. KGH received advisory board fees from Bayer Healthcare, Pfizer, Medtronic, and Edwards Lifesciences. The remaining authors declare no conflict of interest.

The reviewer CT and handling Editor declared their shared affiliation, and the handling Editor states that the process nevertheless met the standards of a fair and objective review.

Acknowledgments

The authors appreciate the excellent study assistance of Mrs. R. Brennfleck and Mrs. S. Wenzel (Department of Cardiology and Pneumology, Charité, Berlin). In addition, the authors thank Dr. Ulrike Grittner (Center for Stroke Research Berlin, Charité – Universitätsmedizin Berlin, Germany) for statistical advice.

Funding

The work was supported by funding from the Federal Ministry of Education and Research via the grant Center for Stroke Research Berlin (01 EO 0801). ME receives funding from the DFG (Excellence cluster NeuroCure; SFB TR 43, KFO 247, KFO 213), BMBF (Centre for Stroke Research Berlin), EU (European Stroke Network, WakeUp, Counterstroke), and Corona Foundation.

Supplementary Material

The Supplementary Material for this article can be found online at https://www.frontiersin.org/article/10.3389/fcvm.2017.00004/full#supplementary-material.

Abbreviations

AF, atrial fibrillation; EF, ejection fraction; LA, left atrial; LV, left ventricular; MACPAF study, mesh ablator vs. cryoballoon pulmonary vein ablation of symptomatic paroxysmal atrial fibrillation study; PV(s), pulmonary vein(s); SR, sinus rhythm.

References

1. Ball J, Carrington MJ, Stewart S; SAFETY investigators. Mild cognitive impairment in high-risk patients with chronic atrial fibrillation: a forgotten component of clinical management? Heart (2013) 99:542–7. doi: 10.1136/heartjnl-2012-303182

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, et al. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J (2016) 37:2893–962. doi:10.1093/eurheartj/ehw210

CrossRef Full Text | Google Scholar

3. Knecht S, Oelschläger C, Duning T, Lohmann H, Albers J, Stehling C, et al. Atrial fibrillation in stroke-free patients is associated with memory impairment and hippocampal atrophy. Eur Heart J (2008) 29:2125–32. doi:10.1093/eurheartj/ehn341

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Mantovan R, Macle L, De Martino G, Chen J, Morillo CA, Novak P, et al. Relationship of quality of life with procedural success of atrial fibrillation (AF) ablation and postablation AF burden: substudy of the STAR AF randomized trial. Can J Cardiol (2013) 29:1211–7. doi:10.1016/j.cjca.2013.06.006

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Neumann T, Wójcik M, Berkowitsch A, Erkapic D, Zaltsberg S, Greiss H, et al. Cryoballoon ablation of paroxysmal atrial fibrillation: 5-year outcome after single procedure and predictors of success. Europace (2013) 15:1143–9. doi:10.1093/europace/eut021

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Takigawa M, Takahashi A, Kuwahara T, Okubo K, Takahashi Y, Watari Y, et al. Long-term follow-up after catheter ablation of paroxysmal atrial fibrillation: the incidence of recurrence and progression of atrial fibrillation. Circ Arrhythm Electrophysiol (2014) 7:267–73. doi:10.1161/CIRCEP.113.000471

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Tilz RR, Rillig A, Thum AM, Arya A, Wohlmuth P, Metzner A, et al. Catheter ablation of long-standing persistent atrial fibrillation: 5-year outcomes of the Hamburg Sequential Ablation Strategy. J Am Coll Cardiol (2012) 60:1921–9. doi:10.1016/j.jacc.2012.04.060

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Al-Khatib SM, Allen LaPointe NM, Chatterjee R, Crowley MJ, Dupre ME, Kong DF, et al. Rate- and rhythm-control therapies in patients with atrial fibrillation: a systematic review. Ann Intern Med (2014) 160:760–73. doi:10.7326/M13-1467

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Kapa S, Epstein AE, Callans DJ, Garcia FC, Lin D, Bala R, et al. Assessing arrhythmia burden after catheter ablation of atrial fibrillation using an implantable loop recorder: the ABACUS study. J Cardiovasc Electrophysiol (2013) 24:875–81. doi:10.1111/jce.12141

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Pokushalov E, Romanov A, Corbucci G, Bairamova S, Losik D, Turov A, et al. Does atrial fibrillation burden measured by continuous monitoring during the blanking period predict the response to ablation at 12-month follow-up? Heart Rhythm (2012) 9:1375–9. doi:10.1016/j.hrthm.2012.03.047

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Hindricks G, Pokushalov E, Urban L, Taborsky M, Kuck K-H, Lebedev D, et al. Performance of a new leadless implantable cardiac monitor in detecting and quantifying atrial fibrillation: results of the XPECT trial. Circ Arrhythm Electrophysiol (2010) 3:141–7. doi:10.1161/CIRCEP.109.877852

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Wann LS, Curtis AB, January CT, Ellenbogen KA, Lowe JE, Estes NAM, et al. 2011 ACCF/AHA/HRS focused update on the management of patients with atrial fibrillation (updating the 2006 guideline): a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation (2011) 123:104–23. doi:10.1161/CIR.0b013e3181fa3cf4

CrossRef Full Text | Google Scholar

13. Raviele A, Natale A, Calkins H, Camm JA, Cappato R, Ann Chen S, et al. Venice chart international consensus document on atrial fibrillation ablation: 2011 update. J Cardiovasc Electrophysiol (2012) 23:890–923. doi:10.1111/j.1540-8167.2012.02381.x

CrossRef Full Text | Google Scholar

14. Haeusler KG, Koch L, Ueberreiter J, Endres M, Schultheiss H-P, Heuschmann PU, et al. Stroke risk associated with balloon based catheter ablation for atrial fibrillation: rationale and design of the MACPAF study. BMC Neurol (2010) 10:63. doi:10.1186/1471-2377-10-63

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Koch L, Haeusler K, Herm J, Safak E, Fischer R, Malzahn U, et al. Mesh ablator vs. cryoballoon pulmonary vein ablation of symptomatic paroxysmal atrial fibrillation: results of the MACPAF study. Europace (2012) 14:1441–9. doi:10.1093/europace/eus084

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Maagh P, Butz T, Plehn G, Christoph A, Meissner A. Pulmonary vein isolation in 2012: is it necessary to perform a time consuming electrophysical mapping or should we focus on rapid and safe therapies? A retrospective analysis of different ablation tools. Int J Med Sci (2013) 10:24–33. doi:10.7150/ijms.4771

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Ahmed J, Sohal S, Malchano ZJ, Holmvang G, Ruskin JN, Reddy VY. Three-dimensional analysis of pulmonary venous ostial and antral anatomy: implications for balloon catheter-based pulmonary vein isolation. J Cardiovasc Electrophysiol (2006) 17:251–5. doi:10.1111/j.1540-8167.2005.00339.x

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Herm J, Fiebach JB, Koch L, Kopp UA, Kunze C, Wollboldt C, et al. Neuropsychological effects of MRI-detected brain lesions after left atrial catheter ablation for atrial fibrillation: long-term results of the MACPAF study. Circ Arrhythm Electrophysiol (2013) 6:843–50. doi:10.1161/CIRCEP.113.000174

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen S-A, et al. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace (2012) 14:528–606. doi:10.1093/europace/eus027

CrossRef Full Text | Google Scholar

20. Steinwender C, Hönig S, Leisch F, Hofmann R. One-year follow-up after pulmonary vein isolation using a single mesh catheter in patients with paroxysmal atrial fibrillation. Heart Rhythm (2010) 7:333–9. doi:10.1016/j.hrthm.2009.11.013

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Manganiello S, Anselmino M, Amellone C, Pelissero E, Giuggia M, Trapani G, et al. Symptomatic and asymptomatic long-term recurrences following transcatheter atrial fibrillation ablation. Pacing Clin Electrophysiol (2014) 37:697–702. doi:10.1111/pace.12387

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Tondo C, Tritto M, Landolina M, DE Girolamo PG, Bencardino G, Moltrasio M, et al. Rhythm-symptom correlation in patients on continuous monitoring after catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol (2013) 24:154–60. doi:10.1111/jce.12292

CrossRef Full Text | Google Scholar

23. Verma A, Champagne J, Sapp J, Essebag V, Novak P, Skanes A, et al. Discerning the incidence of symptomatic and asymptomatic episodes of atrial fibrillation before and after catheter ablation (DISCERN AF): a prospective, multicenter study. JAMA (2013) 173:149–56. doi:10.1001/jamainternmed.2013.1561

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Di Giovanni G, Wauters K, Chierchia G-B, Sieira J, Levinstein M, Conte G, et al. One-year follow-up after single procedure cryoballoon ablation: a comparison between the first and second generation balloon. J Cardiovasc Electrophysiol (2014) 25:834–9. doi:10.1111/jce.12409

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Aryana A, Singh SM, Kowalski M, Pujara DK, Cohen AI, Singh SK, et al. Acute and long-term outcomes of catheter ablation of atrial fibrillation using the second-generation cryoballoon versus open-irrigated radiofrequency: a multicenter experience. J Cardiovasc Electrophysiol (2015) 26:832–9. doi:10.1111/jce.12695

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Kuck K-H, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun KRJ, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med (2016) 374:2235–45. doi:10.1056/NEJMoa1602014

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Leong-Sit P, Roux J-F, Zado E, Callans DJ, Garcia F, Lin D, et al. Antiarrhythmics after ablation of atrial fibrillation (5A study): six-month follow-up study. Circ Arrhythm Electrophysiol (2011) 4:11–4. doi:10.1161/CIRCEP.110.955393

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Providencia R, Defaye P, Lambiase PD, Pavin D, Cebron J-P, Halimi F, et al. Results from a multicentre comparison of cryoballoon vs. radiofrequency ablation for paroxysmal atrial fibrillation: is cryoablation more reproducible? Europace (2016) 19:48–57. doi:10.1093/europace/euw080

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Pedrote A, Arana-Rueda E, García-Riesco L, Sánchez-Brotons J, Durán-Guerrero M, Gómez-Pulido F, et al. Paroxysmal atrial fibrillation burden before and after pulmonary veins isolation: an observational study through a subcutaneous leadless cardiac monitor. J Cardiovasc Electrophysiol (2013) 24:1075–82. doi:10.1111/jce.12190

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Camm AJ, Kirchhof P, Lip GYH, Schotten U, Savelieva I, Ernst S, et al. Guidelines for the management of atrial fibrillation: the task force for the management of atrial fibrillation of the European Society of Cardiology (ESC). Eur Heart J (2010) 31:2369–429. doi:10.1093/eurheartj/ehq278

CrossRef Full Text | Google Scholar

31. Camm AJ, Lip GYH, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Europace (2012) 14:1385–413. doi:10.1093/europace/eus305

CrossRef Full Text | Google Scholar

32. Morillo CA, Verma A, Connolly SJ, Kuck KH, Nair GM, Champagne J, et al. Radiofrequency ablation vs antiarrhythmic drugs as first-line treatment of paroxysmal atrial fibrillation (RAAFT-2): a randomized trial. JAMA (2014) 311:692–700. doi:10.1001/jama.2014.467

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Pokushalov E, Romanov A, Corbucci G, Artyomenko S, Turov A, Shirokova N, et al. Ablation of paroxysmal and persistent atrial fibrillation: 1-year follow-up through continuous subcutaneous monitoring. J Cardiovasc Electrophysiol (2011) 22:369–75. doi:10.1111/j.1540-8167.2010.01923.x

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Knecht S, Sticherling C, von Felten S, Conen D, Schaer B, Ammann P, et al. Long-term comparison of cryoballoon and radiofrequency ablation of paroxysmal atrial fibrillation: a propensity score matched analysis. Int J Cardiol (2014) 176:645–50. doi:10.1016/j.ijcard.2014.06.038

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Eitel C, Husser D, Hindricks G, Frühauf M, Hilbert S, Arya A, et al. Performance of an implantable automatic atrial fibrillation detection device: impact of software adjustments and relevance of manual episode analysis. Europace (2011) 13:480–5. doi:10.1093/europace/euq511

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Schmidt B, Asbach S, Schweika O, Zehender M, Bode C, Faber TS. Atrial fibrillation reduces the atrial impedance amplitude during cardiac cycle: a novel detection algorithm to improve recognition of atrial fibrillation in pacemaker patients. Europace (2007) 9:812–6. doi:10.1093/europace/eum106

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Kwok CS, Loke YK, Hale R, Potter JF, Myint PK. Atrial fibrillation and incidence of dementia: a systematic review and meta-analysis. Neurology (2011) 76:914–22. doi:10.1212/WNL.0b013e31820f2e38

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Santangeli P, Di Biase L, Bai R, Mohanty S, Pump A, Cereceda Brantes M, et al. Atrial fibrillation and the risk of incident dementia: a meta-analysis. Heart Rhythm (2012) 9:1761–8. doi:10.1016/j.hrthm.2012.07.026

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Medi C, Evered L, Silbert B, Teh A, Halloran K, Morton J, et al. Subtle post-procedural cognitive dysfunction after atrial fibrillation ablation. J Am Coll Cardiol (2013) 62:531–9. doi:10.1016/j.jacc.2013.03.073

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: catheter ablation, pulmonary veins, atrial fibrillation, Arctic Front®, HD Mesh Ablator®, ECG monitoring

Citation: Schirdewan A, Herm J, Roser M, Landmesser U, Endres M, Koch L and Haeusler KG (2017) Loop Recorder Detected High Rate of Atrial Fibrillation Recurrence after a Single Balloon- or Basket-Based Ablation of Paroxysmal Atrial Fibrillation: Results of the MACPAF Study. Front. Cardiovasc. Med. 4:4. doi: 10.3389/fcvm.2017.00004

Received: 23 September 2016; Accepted: 18 January 2017;
Published: 13 February 2017

Edited by:

Mehdi Namdar, Hôpitaux Universitaires de Genève, Switzerland

Reviewed by:

Antonio Sorgente, Cleveland Clinic Abu Dhabi, United Arab Emirates
Cheryl Teres, Hôpitaux Universitaires de Genève, Switzerland

Copyright: © 2017 Schirdewan, Herm, Roser, Landmesser, Endres, Koch and Haeusler. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Alexander Schirdewan, a.schirdewan@sana-kl.de

These authors have contributed equally to this work.

Download