- Openwater, San Francisco, CA, United States
Amyloid fibrin(ogen) microclots are misfolded protein aggregates with β-sheet structures that have been associated with Long COVID and numerous thrombo-inflammatory diseases. These microclots persist in circulation and obstruct microvasculature, impair oxygen transport and promote chronic inflammation. Conventional thrombolytic therapies such as recombinant tissue plasminogen activator (rtPA) show limited efficacy against these microclots due to their structure and composition. In this study, we assess the impact of low intensity focused ultrasound (LIFU) stimulation on amyloid microclot fragmentation, the role of cavitation in this process and investigate whether microbubble-assisted ultrasound can enhance their lysis. Amyloid microclot models were generated using freeze-thaw cycles followed by incubation. Microclots were exposed to ultrasound waves at 150, 300, 500 kHz, and 1 MHz under four conditions: ultrasound alone (US), ultrasound with microbubbles (MB + US), ultrasound with rtPA (rtPA + US), and ultrasound with both microbubbles and rtPA (MB + rtPA + US). Low-frequency ultrasound at 150 kHz resulted in a significant clot lysis with up to three-fold reduction in both clot size and the number of large clots. The addition of microbubbles enhanced clot lysis at 150, 300, and 500 kHz. These findings suggest that ultrasound, particularly at 150 kHz is a promising method for amyloid microclot lysis. The combination of ultrasound with microbubbles and rtPA further improved clot fragmentation, rendering it a potential therapeutic tool for conditions like Long COVID.
1 Introduction
Protein misfolding into abnormal amyloid microclots is a defining hallmark of a wide array of human disorders, known as amyloidoses (Almeida and Brito, 2020). Over 50 diseases, including neurodegenerative conditions (e.g., Alzheimer’s, Parkinson’s, Huntington’s diseases), metabolic/endocrine diseases (e.g., Type II diabetes), and systemic amyloidoses (e.g., light-chain amyloidosis), are characterized by pathological amyloid accumulation (Almeida and Brito, 2020; Rambaran and Serpell, 2008; Kell et al., 2025).
In these disorders, soluble proteins misfold and aggregate into insoluble fibrillar microclots where they lead to disruption of tissue architecture and function. These protein aggregates not only can interfere with cellular processes but also lead into disease progression through inducing chronic inflammation and oxidative stress (Almeida and Brito, 2020; Rambaran and Serpell, 2008). Although the composition of amyloid microclots varies in different biological conditions, their common structural feature is the formation of ordered β-across-β architecture (Turner et al., 2023a; Mai et al., 2024). The cross-β core increases the stability of amyloid fibrils, rendering them insoluble and proteolysis-resistant (Rambaran and Serpell, 2008).
Among amyloid-related pathologies, amyloid fibrin(ogen) microclots have recently gained attention due to their role in Long COVID (post-acute COVID-19 syndrome) and cardiovascular diseases. Recent studies have identified amyloid fibrin(ogen) microclots in the blood of patients with Long COVID (Turner et al., 2023a; Kell et al., 2022; Laubscher et al., 2021; Kruger et al., 2022; Pretorius et al., 2021). These amyloid microclots typically range in size from 1 to 200 μm, with a median diameter of approximately 15 µm (Kell et al., 2022; Sakharov et al., 2000). Due to their size, microclots can circulate within the bloodstream and obstruct the microvasculature, disrupting perfusion and impairing oxygen delivery to tissues. As a result, these microclots have been detected as a central pathological factor which are associated with persistent symptoms such as fatigue, breathlessness, and cognitive dysfunction. Further coagulation of such microclots can lead to hypoxia, chronic organ dysfunction and inflammation (Pretorius et al., 2021; Turner et al., 2023b). Proteomic analyses showed that COVID-related microclots contain pro-inflammatory molecules and clotting factors that result in persistent inflammation and resistance to fibrinolysis (Kruger et al., 2022).
Current therapies for Long COVID focus primarily on enzymatic clot dissolution, where fibrinolytic enzymes such as rtPA act as a serine protease to convert plasminogen to plasmin and degrade fibrin networks. Although enzymatic therapies are effective against conventional thrombi, they struggle to address amyloid-rich microclots due to their structural resistance to enzymatic breakdown. This fibrinolysis resistance is further exacerbated in COVID-19 patients by upregulation of α-(2)-antiplasmin which inhibits plasmin activity and limits the effectiveness of rtPA (Kell et al., 2022; Sakharov et al., 2000). Clinical observations confirm this inefficiency, where COVID-19 patients often showed poor response to tPA-mediated thrombolysis (Pretorius et al., 2020).
Another strategy to treat circulating microclots is filtration-based strategies to physically remove them from the bloodstream. Apheresis is reported for severe COVID patients (Jaeger et al., 2022), in which blood is drawn from the patient, passes through an extracorporeal filter, and then returns to the body. However, extracorporeal methods are invasive, resource-intensive, and involve health risks such as bleeding and infection which limit their application to acute conditions or experimental settings. Given the pathological impact and therapy resistance of amyloid microclots, there is a clear need for innovative and non-invasive treatments that can effectively disrupt circulatory amyloid microclots in patients and restore microvascular function.
Ultrasound has a long and successful history of treating blood clots in clinical settings, both as a stand-alone therapy and in combination with fibrinolytic enzymes such as rtPA (Školoudík et al., 2025; Draxler et al., 2025). Addition of transcranial ultrasound exposure to IV tPA treatment significantly improved thrombolysis in a phase II trial for acute ischemic stroke, leading to early arterial recanalization and enhanced clinical recovery rates compared to tPA alone (Tsivgoulis and Alexandrov, 2007). In deep vein thrombosis (DVT), ultrasound-assisted catheter thrombolysis outperformed enzymatic thrombolysis alone with high clot clearance and faster treatment (Parikh et al., 2008).
There is a strong theoretical and clinical basis for employing ultrasound to disrupt microclots (Sakharov et al., 2000; Rasouli et al., 2023; Molina et al., 2006; Wang et al., 2022; Ward et al., 2024). Ultrasound in the low-kilohertz to low-megahertz range can induce mechanical phenomena such as acoustic streaming, cavitation, shear forces, and radiation forces that can be used to promote clot fragmentation (Rad et al., 2023). Acoustic streaming generates convective currents and shear stress around the thrombus, enhancing penetration and mixing of fibrinolytic agents into the clot (Rasouli et al., 2023; Bader et al., 2014). Acoustic streaming and acoustic radiation force (ARF) can also physically erode clot structure by pushing fluid through the fibrin mesh, loosening pore spaces and increasing permeation of plasminogen activators into the thrombus (Ward et al., 2024). This streaming action also exposes new fibrin binding sites to plasmin, enhancing enzymatic breakdown (Ward et al., 2024).
Numerous in vitro, animal, and clinical studies report faster and more effective thrombolysis when ultrasound is applied alongside fibrinolytic drugs (Ward et al., 2024). Unlike chemical thrombolysis which relies solely on biochemical mechanisms, ultrasound offers advantage by offering a mechanical mode of action that can help to overcome the fibrinolysis resistance of amyloid clots (Ward et al., 2024). The combination of acoustic streaming and cavitation can physically disrupt and weaken the clot. Prior studies on clots have shown that ultrasound can significantly accelerate clot lysis by mechanical forces (Wang et al., 2022). For example, exposure to 1 MHz ultrasound at ∼2 W/cm2 produced streaming with a shear rate on the order of 40 s−1, roughly doubling the enzymatic lysis rate in vitro (Sakharov et al., 2000; Acconcia et al., 2014). In microbubble-assisted sonothrombolysis, oscillating microbubbles create vigorous microstreaming and localized shear forces that significantly accelerate clot breakdown (Molina et al., 2006). Experimental studies have visually confirmed these effects, showing that stable cavitation bubbles preferentially erode clots and correlate with faster lysis rates compared to enzymatic treatments (Mai et al., 2024). Furthermore, microbubble-assisted sonothrombolysis facilitates cavitation to enhance clot lysis and significantly boost clot dissolution (Molina et al., 2006).
Despite significant progress in sonothrombolysis for large vascular clots, no reported studies have specifically explored ultrasound-mediated lysis of circulating amyloid microclots which are recognized as a key therapeutic target in Long COVID. Here, we hypothesize that applying ultrasound can effectively disrupt amyloid fibrin microclots in vitro via acoustically induced streaming. To test this, we employed a LOC model that enables precise control over microclot exposure conditions, allowing systematic evaluation of the effects of frequency, shear stress, and microbubble presence on clot disruption. Our goal is to identify optimal ultrasound parameters for microclot fragmentation and lay the groundwork for noninvasive strategies to mitigate microvascular dysfunction and coagulopathies in amyloid-related pathologies such as Long COVID.
2 Materials and methods
2.1 Materials
Porcine plasma was obtained from Innovative Research, Inc. (United States; No. IGPCPLANAC500ML) and stored at −20°C until use. Plasma aliquots (1 mL) were used for microclot formation. rtPA was purchased from Sigma-Aldrich (MilliporeSigma, United States; No. T0831) and reconstituted according to the manufacturer’s protocol to achieve a final working concentration of 3 μg/mL. Thioflavin T (ThT) was purchased from Sigma-Aldrich (St. Louis, MO, United States; No. T3516). Vevo MicroMarker® contrast agents (FUJIFILM VisualSonics Corp., Japan; No. 11678) were used as exogenous microbubbles and were diluted in PBS buffer per manufacturer protocol.
2.2 Amyloid microclot formation
Amyloid fibrin(ogen) microclots were generated by subjecting porcine plasma to repeated freeze-thaw cycles, followed by incubation at 37°C. Plasma aliquots (1 mL) underwent 12 freeze-thaw cycles, with each cycle consisting of freezing at −20°C and thawing at 37°C for 2 hours. The aliquots were diluted 1.5 x before the ultrasound treatment.
2.3 Lab on chip device fabrication
The LOC device was designed to replicate a 6 mm diameter popliteal vein, providing a controlled environment for targeted ultrasound exposure of amyloid microclots. The popliteal vein was selected due to its accessibility for transcutaneous ultrasound exposure rendering it a relevant site for potential clinical applications of ultrasound. The device molds were designed using Onshape (version 1.188, PTC Inc., United States) CAD software and fabricated via digital light processing (DLP) technology using a 285 D Printer (CADWorks3D, Canada). One side of the channel was made slightly flat (5.8 mm instead of a full 6 mm circle) to create a smooth surface for bonding the silicone layer. This made it easier to peel off the PDMS and allowed better imaging, while still keeping the overall shape close to that of a popliteal vein. After UV treatment, Momentive RTV615 silicone elastomer (RS Hughes, United States) was cast into the molds and cured following manufacturer-recommended conditions to form the microfluidic channels. The cured elastomer was then bonded onto a 20 µm silicone sheet (Gteek, United Kingdom) using plasma surface treatment.
2.4 Experimental setup
The schematic representation of the LOC platform, transducer arrangement, and acoustic bath setup is shown in Figure 1. The experimental setup consisted of a custom-designed and 3D printed acoustic water bath, which housed the LOC device, a piezoelectric transducer, and supporting components. The transducers were mounted at the bottom of the bath. The LOC chip was mounted on top of the water bath using a custom-designed holder, securing its position at the pre-measured focal plane of the ultrasound transducers to maintain consistent acoustic exposure across all experimental conditions.

Figure 1. (A) Schematic of the experimental platform for ultrasound-mediated amyloid microclot lysis. The LOC platform is placed in an acoustic water bath and exposed to focused ultrasound at 150 kHz, 300, 500 kHz, and 1 MHz. A function generator controls the ultrasound frequency, and an amplifier amplifies the signal. (B) Zoomed-in view of the LOC channel, showing the interaction of ultrasound, microbubbles, and rtPA with microclots. Microbubbles undergo volumetric oscillations in response to acoustic pressure waves, generating streaming effect and shear stress to lyse microclots.
A function generator (DG4162, Rigol Technologies Co., Ltd., China) controlled the ultrasound frequency generated continuous stepped pulse waveform, while an amplifier (Model 1040L, Electronics & Innovation, Ltd., United States) regulated the voltage to the transducers. The bath was filled with degassed water as a coupling medium for ultrasound wave transmission while minimizing acoustic attenuation.
Experiments were conducted under static flow conditions in the LOC platform. Ultrasound was applied at four different frequencies: 150, 300, 500 kHz, 1 MHz. The mechanical index (MI) is defined as MI =
2.5 Microscopy and image analysis
Thioflavin T was used as a fluorescent dye to stain amyloid fibrils. A stock solution of 30 mM Thioflavin T was prepared in phosphate-buffered saline (PBS), pH 7.4. The solution was filtered using a 0.2 µm syringe filter to ensure clarity. Samples were incubated with Thioflavin T at final concentration of 30 μM room temperature for 30 min prior to imaging. Amyloid microclot lysis was evaluated based on clot count reduction and size reduction. The microclots morphology and count were monitored before and after ultrasound exposure, and changes in clot size were analyzed using fluorescence microscopy. This concentration was selected based on previously published studies demonstrating that 30 μM offers an optimal balance between signal intensity and minimal fluorescence quenching (Mai et al., 2024; Sakharov et al., 2000; Xue et al., 2017).
Imaging was conducted using the Countstar BioTech Automated Cell Counter (Model IC1000), which uses epifluorescence microscopy with a fixed-focus optical system and a 5-megapixel CMOS camera. A ×10 objective lens was used to capture the fluorescent microclots, and a GFP filter was used for ThT-labeled samples. Image analysis was performed using segmentation and particle detection plugins in ImageJ software, which measured clot area and count. Equivalent circular diameter (Deq) was calculated from the measured area for each microclot and analyzed in a custom Python script.
2.6 Statistical analysis
All statistical analyses were conducted using GraphPad Prism version 10.4.1. Group comparisons for clot count and average clot diameter were performed using one-way ANOVA. Sidak’s post hoc multiple comparisons test was used to identify pairwise differences between groups. A p-value of less than 0.05 was considered statistically significant.
Each experimental group included three biologically independent samples (n = 3), except for Figure 2, where two replicates (n = 2) were analyzed. For each replicate, three non-overlapping microscopic fields of view were imaged. Every image contained thousands of individual microclots. Clot count and diameter were quantified using ImageJ version 2.16.0 with particle segmentation and custom Python scripts. Diameter was determined using effective diameter to estimate the equivalent size for irregular shapes. Data are reported as mean ± standard deviation (SD).

Figure 2. (A) Schematic representation of amyloid microclot model formation using freeze-thaw cycles. (B) Fluorescence microscopy images of ThT-stained porcine plasma across freeze-thaw cycles. Scale bar is 200 µm. (C) Clot count quantification graph showing the increase in amyloid microclots over cycles.
3 Results and discussion
3.1 Amyloid microclot model formation
Development of a reproducible amyloid fibrin microclot model is essential for investigating their role in clinical disorders including Long COVID, Alzheimer’s, type 2 diabetes (Kell et al., 2025; Turner et al., 2023a). Various in vitro methods have been reported to produce amyloid microclots including drying, heating, exposure to acidic pH, hydrophobic surfaces, and organic solvents (Mai et al., 2024).
In this study, we used the freeze-thaw method and incubation at biological temperature to induce amyloidogenesis. Freeze-thawing results in ice-water interfaces in plasma and introduces structural stress into proteins which could induce protein misfolding and aggregation (Bader et al., 2014). Moreover, the incubation at room temperature can promote amyloidogenesis through oxidative stress and spontaneous fibrinogen misfolding that ultimately enhances the formation of amyloid fibrin microclots. Freeze-thaw-based amyloid formation is a cell and enzyme-free approach. In this method, protein structural modification results from intrinsic protein dynamics rather than external biochemical interactions, offering reproducibility and control over aggregation conditions (Graether et al., 2003; Filippov et al., 2008).
To verify amyloid formation by freeze-thawing, we employed the Thioflavin T (ThT) dye, which selectively binds to β-sheet-rich amyloid fibrils and enhances fluorescence upon interaction. Fluorescent ThT staining is widely used in amyloid research as a reliable marker for detecting the amyloid transition of fibrinogen, owing to its high specificity for cross-β structures (Turner et al., 2023a; Kell et al., 2022; Laubscher et al., 2021; Pretorius et al., 2016).
Figure 2A shows the fluorescence microscopy images of freeze-thawed plasma samples stained with ThT. In the earlier cycles, there are no visible ThT-labeled amyloid structures. However, green, fluorescent ThT-labeled microclots emerge starting from cycles 7, indicating a gradual amyloid microclot formation. By cycle 8, a marked increase in ThT-tagged microclots can be observed which shows significant amyloid microclot formation and with each cycle more amyloid microclots appear in the sample. Figure 2B shows the quantitative clot counts as a function of cycles, confirming a significant increase in amyloid microclots after cycle 7. This cycle-dependent increase in amyloid microclot counts confirms amyloidogenesis by repeated freeze-thaw cycles and incubation. The formation of microclots with repeated freeze–thaw cycles and incubation are previously reported, showing that freeze-thaw processes generate ice-water interfaces that impose structural stress on plasma proteins and lead to misfolding and aggregation. Fibrinogen, in particular, has been shown to adopt abnormal conformations under these conditions and covert into amyloid-like forms (Graether et al., 2003; Filippov et al., 2008; Cao et al., 2003). This experiment was repeated twice to ensure reproducibility. However, since plasma contains natural proteins, batch-to-batch variations are inherent. As a result, the data presented here is primarily aimed to show the trend of increasing clot formation with successive cycles rather than serving as an exact reference for clot numbers. Although we use a particle detection algorithm to identify discrete microclots, accumulation of submicron amyloid clots can lead to diffuse green fluorescence across the field. These clots are likely too small or diffuse to be detected as distinct particles but still bind Thioflavin T (ThT) and result in a uniform background signal. This background signal may reflect early-stage amyloid formation occurring before the assembly of detectable microclots.
3.2 Effect of ultrasound on amyloid microclots across different frequencies
After amyloid microclot formation, we explore how different ultrasound frequencies impact both microclot morphology and count. The different mechanical interactions of ultrasound with microclots are frequency-dependent, as acoustic forces such as acoustic radiation force and acoustic streaming exhibit varying levels of dominance based on their inherent frequency ranges (Rasouli et al., 2023). The balance between these forces dictates whether ultrasound leads to microclot fragmentation, which is our aim, or instead results in displacement or even aggregation of the microclots. Hence, identifying the optimal frequency and experimental conditions for effective clot disruption is crucial for the potential therapeutic applications of our ultrasound platform. To this end, we tested the effect of different ultrasound frequencies on the lysis of microclots by subjecting them to 150, 300, 500 kHz, and 1 MHz.
This experiment was also repeated in the presence of rtPA, a serine protease used as a thrombolytic agent to dissolve blood clots. rtPA works by binding to fibrin in microclots and then converting plasminogen into plasmin, which subsequently dissolves the clot through a process known as fibrinolysis. We sought to determine whether ultrasound increases the rate of clot dissolution when rtPA is used and whether a possible synergistic effect of the acoustic forces and pharmacological thrombolysis can be observed.
Figures 3A, B show fluorescence microscopy images of ThT-tagged microclots treated with different ultrasound (US) frequencies for US-only and US + rtPA groups. The control samples show a high density of large (over 30 microns) amyloid microclots, which are uniformly distributed across the image. Upon exposure to 150 kHz waves, the microclots exhibit a marked reduction in size. At higher frequencies, however, less fragmentation can be observed, and the clots remain intact.

Figure 3. Effect of ultrasound (US) and ultrasound combined with rtPA (US + rtPA) on microclot lysis at different frequencies. (A, B) Fluorescence microscopy images of microclots under (A) ultrasound alone and (B) ultrasound with rtPA at control, 150, 300, 500 kHz, and 1 MHz. (C, D) Bar graphs showing average clot diameter as a function of ultrasound frequency for (C) US-only and (D) US + rtPA conditions. Clot size is significantly reduced at 150 kHz, with a minor reduction at higher frequencies. (E, F) Clot counts for microclots larger than 30 µm across different conditions, demonstrating a frequency-dependent clot fragmentation. Error bars represent standard deviations. Scale bar is 200 µm.
To quantify microclot lysis under US treatment, image analysis was performed, and the average clot diameter and count of clots above 30 μm at each testing condition were calculated for three samples to ensure statistical robustness. Figures 3C, D show the average clot diameter and Figures 3E, F show the number of clots over 30 μm as a function of ultrasound frequency for both US and US + rtPA samples. In agreement with the microscopy images, a significant reduction in clot size is observed at 150 kHz, while exposure to higher frequencies leads to less pronounced fragmentation. US + rtPA shows a similar trend of reduction in average diameter and large clot count at 150 kHz. For US + rTPA the mean clot diameter decreases from ∼20 µm in the control to ∼7 μm at 150 kHz, confirming significant fragmentation at this frequency. At 300 kHz and higher frequencies there is no statistically significant change in clot size for both US and US + rTPA.
The findings indicate that 150 kHz ultrasound is the most effective frequency for reducing both large clot count and size in both US-only and US + rtPA groups, whereas higher frequencies (300 kHz–1 MHz) result in less pronounced disruption. One potential explanation for clot lysis at 150 kHz is the formation of endogenous microbubbles, which can result in boundary-layer-driven acoustic microstreams (Rasouli et al., 2023). During the 150 kHz experiment, bubbles became visibly present, whereas no such bubble formation was observed at higher frequencies. Plasma and most biological fluids naturally contain gas nuclei that act as precursors for bubble formation under ultrasound exposure. Upon formation, these bubbles oscillate in response to ultrasound actuation and create localized streaming which in turn can enhance clot lysis.
3.3 The role of cavitation in ultrasound-induced clot fragmentation
To investigate whether the mechanism of fragmentation at 150 kHz was directly linked to cavitation and acoustic streaming, we compared the effect of ultrasound exposure on degassed and non-degassed plasma samples. Cavitation requires dissolved gas nuclei to form microbubbles and thus degassing the plasma minimizes endogenous cavitation and bubble formation.
Figure 4 shows the fluorescence microscopy results of microclots in degassed and non-degassed plasma. Clot lysis and reduced micro clot size was observed in the gassed plasma sample after treatment with 150 kHz acoustic waves as shown by the images in Figure 4A. The same ultrasound waves failed to lyse microclots in degassed plasma samples and they remained largely intact. The absence of visible bubble formation in the degassed sample during the test as compared to the non-degassed sample suggests that clot lysis at 150 kHz can be attributed to the formation of endogenous bubbles and their attendant acoustic streaming.

Figure 4. Effect of degassing on ultrasound US-induced microclot lysis at 150 kHz. (A) Fluorescence microscopy images of microclots exposed to 150 kHz ultrasound in non-degassed (left) and degassed (right) conditions. Clot size remains larger in the degassed condition, indicating reduced lysis efficiency. Scale bars represent 100 µm. (B) Quantitative analysis of clot size and count. The left graph shows the average clot diameter across conditions.150 kHz stimulation of non-degassed sample significantly reduced clot size compared to the control, while the degassed condition retains larger clots. The right graph displays the number of clots over 30 μm, showing a significant reduction for non-degassed, whereas degassed conditions result in higher clot retention.
Quantitative data on microclot size after ultrasound exposure is depicted in Figure 4B for both degassed and non-degassed samples. The left graph investigates microclot lysis in degassed vs. non degassed plasma where cavitation effects are inhibited. The y-axis is the mean clot diameter, and the x-axis is frequency. Although the degassed samples showed a minor decrease in clot size it is considered statistically nonsignificant with diameters ranging from ∼17 to ∼14 µm compared to ∼7 µm in non-degassed sample. Moreover, the number of large clots over 30 µm decreased for non-degassed sample to approximately a third of the control while in the degassed sample the variation was not statistically significant. This supports the hypothesis that clot lysis observed at 150 kHz is due to the cavitation, likely through cavitation-induced acoustic streams.
3.4 Effect of microbubbles and rtPA on clot lysis across frequencies
Previous results showed that cavitation is crucial for clot fragmentation at 150 kHz. In this section, we introduce exogenous microbubbles (MBs) into the plasma to promote stable cavitation. We use gas-filled microbubbles ∼1–2 µm in diameter which are coated with a phospholipid shell to render them more stable. In response to acoustic waves, these microbubbles undergo rapid compression and expansion cycles and generate localized fluid motion known as acoustic streams. Unlike endogenous cavitation, which relies on the presence of dissolved gas nuclei that varies between samples, contrast agents provide more consistent and controlled bubble dynamics and acoustic streaming. We hypothesize that the use of exogenous bubble can further increase lysis. In this section we also investigate the impact of MBs in combination with rtPA on clot lysis to assess whether this approach would further enhance enzymatic clot degradation.
Figure 5 shows the fluorescence microscopy images of ThT-stained plasma in the MB and MB + rtPA groups at the tested frequencies. In both groups, clot fragmentation is more pronounced at 150 kHz and 300 kHz, showing a significant reduction in clot size. Fragmentation is also observable with less impact at 500 kHz, while at 1 MHz, the effect diminishes significantly. Clot size analysis shown in Figure 5 further supports these observations and confirms a substantial decrease in clot size. The average clot diameter significantly decreases from ∼18 µm in the control group to ∼7 μm at 150 kHz, ∼8 μm at 300 kHz, and ∼10 μm at 500 kHz, while at 1 MHz, the clot size remained about 10 µm in the MB group. Additionally, the large clot count decreased from ∼450 in the control to ∼70 at 150 kHz, 90 at 300 kHz, 140 at 500 kHz, and 350 at 1 MHz. The results confirm the capability of the ultrasound system in combination with MBs for fragmentation and elimination of large microclots.

Figure 5. Effect of acoustically stimulated microbubbles (MB) and microbubbles combined with rtPA (MB + rtPA) on microclot lysis at different frequencies. (A, B) Fluorescence microscopy images of microclots under (A) ultrasound and Microbubbles and (B) combination of ultrasound, microbubbles with rtPA at control, 150, 300, 500 kHz, and 1 MHz. (C, D) Bar graphs showing average clot diameter as a function of ultrasound frequency for (C) MB (D) and MB + rtPA conditions. (E, F) Clot counts for microclots larger than 30 µm across different conditions. Error bars represent standard deviations. Scale bar is 200 µm.
A similar trend, with a slightly stronger effect, can be seen in the MB + rtPA conditions, where the average clot diameter decreases from 19 µm in the control to 7 μm at 150 kHz, and the large clot count declines from ∼500 to 50 at 150 kHz. The observed reduction in clot size and count suggests a potential synergistic interaction between ultrasound, microbubbles, and enzymatic lysis, warranting further investigation to optimize this effect. This is particularly relevant given that acoustic streaming is reported to be one of the strongest mechanisms to induce normal advection and enhancing mixing (Rasouli and Tabrizian, 2019). Strong and rapid mixing can accelerate enzymatic clot degradation by facilitating the binding of rtPA to fibrin within the clot, thereby promoting the conversion of plasminogen to plasmin.
These results indicate that microbubble-assisted ultrasound stimulation is more effective for clot fragmentation at lower frequencies (150–500 kHz). The addition of exogenous contrast agents enhances cavitation and acoustic streaming, resulting in more effective clot disruption. Also, in MB added groups the clot fragmentation was observed across multiple frequencies, while in the US-only group, lysis was observed only at 150 kHz. The enhanced efficacy observed at 150–300 kHz aligns with prior studies showing that microbubble volumetric oscillation is markedly stronger in this range. Volumetric expansion ratios can exceed 30-fold at these frequencies, while at 1 MHz, expansion is typically below 2-fold under comparable pressure conditions (Ilovitsh et al., 2018).
The clot lysis in the presence of microbubbles can be attributed to boundary-layer-driven acoustic microstreaming generated by oscillating stable or semi-stable microbubbles which creates strong localized shear forces capable of breaking down the brittle amyloid fibrin(ogen) microclots. This also provides additional targeting specificity to the microclots, as they are more susceptible to fragmentation than other biological particles. Moreover, the streaming-induced mixing may improve enzymatic access by exposing binding sites and facilitating the interaction of rtPA with the microclots (Školoudík et al., 2025). The absence of lysis in degassed conditions supports the role of cavitation-driven streaming in this process. These results emphasize the potential of MB-assisted ultrasound therapy for targeting persistent amyloid microclots in conditions such as Long COVID.
4 Conclusion and future directions
In this study, we showed that ultrasound stimulation is an effective strategy for lysing amyloid fibrin(ogen) microclots, particularly at 150 kHz. Using a LOC model we observed a significant reduction in clot size and large clot count after ultrasound exposure. The lack of fragmentation in degassed plasma confirms that cavitation plays a key role in clot lysis by generating acoustic microstreaming that enhances fibrinolysis. Microbubble contrast agents further improved clot fragmentation through stable cavitation and generating acoustic streams. Our results suggest that the observed lysis effect is primarily governed by the oscillation of microbubbles, which are determined by their physical properties and the applied frequency. While vessel geometry may influence local acoustic conditions, the matching of applied frequency and the resonance characteristics of the microbubbles appears to play a dominant role and indicates that the effect is not critically dependent on the specific dimensions of the LOC model. The combination of ultrasound microbubbles and rtPA (MB + rtPA + US) resulted in the largest reduction in clot size. This suggests that mechanical disruption enhances enzymatic penetration and fibrinolysis. Further characterization methods such as electron microscopy and FTIR spectroscopy could provide deeper insight into fibrillar architecture, molecular conformation, and morphology changes induced by ultrasound. Nevertheless, our findings support the potential of ultrasound-enhanced therapy for amyloid-associated coagulopathies where persistent microclots impair vascular function and suggest ultrasound-based treatments could serve as a non-invasive tool to restore blood flow and improve oxygenation in affected patients.
Prior to advancing toward clinical application, further studies are required to bridge the gap between this proof-of-concept and physiological conditions. While the LOC platform enabled controlled testing of ultrasound-mediated lysis and helped isolate acoustic and biochemical effects, incorporating flow conditions and more anatomically representative vascular geometries will be essential to better simulate the in vivo microenvironment. In addition, quantitative mapping of acoustic streaming and shear forces using methods such as particle image velocimetry (PIV) can clarify the mechanical interactions responsible for clot disruption. Moreover, exploring more frequencies range including 100 kHz could uncover additional insights into threshold behavior and bubble dynamics. Future research should also refine ultrasound parameters for in vivo use, assess safety and efficacy, and explore the synergy between ultrasound and thrombolytic agents in biological systems.
Data availability statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.
Ethics statement
Ethical approval was not required for the studies on animals in accordance with the local legislation and institutional requirements because only commercially available biosamples established cell lines were used.
Author contributions
RR: Conceptualization, Investigation, Writing – original draft. BH: Conceptualization, Supervision, Writing – review and editing. SK: Conceptualization, Supervision, Writing – review and editing.
Funding
The author(s) declare that no financial support was received for the research and/or publication of this article.
Acknowledgments
The authors would like to thank Dr. Muhammad Zubair for his contributions to the measurement and characterization of the focal length of the ultrasound transducers.
Conflict of interest
Authors RR, BH, and SK were employed by Openwater.
Generative AI statement
The author(s) declare that Generative AI was used in the creation of this manuscript. Generative AI was used to help with copy editing, formatting, grammar correction, and improving the clarity of the text including the language of the abstract, introduction, and figure captions. All scientific content, data interpretation, and conclusions were developed and validated by the authors.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
Acconcia, C., Leung, B. Y. C., Manjunath, A., and Goertz, D. E. (2014). Interactions between individual ultrasound-stimulated microbubbles and fibrin clots. Ultrasound Med. Biol. 40, 2134–2150. doi:10.1016/j.ultrasmedbio.2014.03.008
Almeida, Z. L., and Brito, R. M. M. (2020). Structure and aggregation mechanisms in amyloids. Molecules 25, 1195. doi:10.3390/molecules25051195
Bader, K. B., Gruber, M. J., and Holland, C. K. (2014). Shaken and stirred: mechanisms of ultrasound-enhanced thrombolysis. Ultrasound Med. Biol. 41, 187–196. doi:10.1016/j.ultrasmedbio.2014.08.018
Cao, E., Chen, Y., Cui, Z., and Foster, P. R. (2003). Effect of freezing and thawing rates on denaturation of proteins in aqueous solutions. Biotechnol. Bioeng. 82, 684–690. doi:10.1002/bit.10612
Draxler, D. F., Brodard, J., Ho, H., Chalkou, K., Turovskij, E., Keragala, C. B., et al. (2025). The individual fibrinolytic capacity predicts the efficacy of ultrasound-assisted catheter-directed thrombolysis in patients with acute pulmonary embolism. J. Thrombosis Haemostasis 23, 1416–1427. doi:10.1016/j.jtha.2024.12.043
Filippov, A. V., Suleimanova, A. V., Grobner, G., and Antsutkin, O. N. (2008). Effect of freezing on amyloid peptide aggregation and self-diffusion in an aqueous solution. Colloid J. 70, 501–506. doi:10.1134/s1061933x08040157
Graether, S. P., Slupsky, C. M., and Sykes, B. D. (2003). Freezing of a fish antifreeze protein results in amyloid fibril formation. Biophys. J. 84, 552–557. doi:10.1016/s0006-3495(03)74874-7
Ilovitsh, T., Ilovitsh, A., Foiret, J., Caskey, C. F., Kusunose, J., Fite, B. Z., et al. (2018). Enhanced microbubble contrast agent oscillation following 250 kHz insonation. Sci. Rep. 8, 16347. doi:10.1038/s41598-018-34494-5
Jaeger, B. R., Arron, H. E., Kalka-Moll, W. M., and Seidel, D. (2022). The potential of heparin-induced extracorporeal LDL/fibrinogen precipitation (H.E.L.P.)-apheresis for patients with severe acute or chronic COVID-19. Front. Cardiovasc Med. 9, 1007636. doi:10.3389/fcvm.2022.1007636
Kell, D. B., Pretorius, E., and Zhao, H. (2025). A direct relationship between ‘blood stasis’ and fibrinaloid microclots in chronic, inflammatory, and vascular diseases, and some traditional natural products approaches to treatment. Pharm. (Basel) 18, 712. doi:10.3390/ph18050712
Kell, D. B., Laubscher, G. J., and Pretorius, E. (2022). A central role for amyloid fibrin microclots in long COVID/PASC: origins and therapeutic implications. Biochem. J. 479, 537–559. doi:10.1042/bcj20220016
Kruger, A., Vlok, M., Turner, S., Venter, C., Laubscher, G. J., Kell, D. B., et al. (2022). Proteomics of fibrin amyloid microclots in long COVID/post-acute sequelae of COVID-19 (PASC) shows many entrapped pro-inflammatory molecules that may also contribute to a failed fibrinolytic system. Cardiovasc Diabetol. 21, 190. doi:10.1186/s12933-022-01623-4
Laubscher, G. J., Lourens, P. J., Venter, C., Kell, D. B., and Pretorius, E. (2021). TEG®, microclot and platelet mapping for guiding early management of severe COVID-19 coagulopathy. J. Clin. Med. 10, 5381. doi:10.3390/jcm10225381
Mai, X., Hu, Y., Wu, Z., Guo, X., Dong, M., Jia, L., et al. (2024). Self-assembly of human fibrinogen into microclot-mimicking antifibrinolytic amyloid fibrinogen particles. ACS Appl. Bio Mater 8, 825–834. doi:10.1021/acsabm.4c01651
Molina, C. A., Ribo, M., Rubiera, M., Montaner, J., Santamarina, E., Delgado-Mederos, R., et al. (2006). Microbubble administration accelerates clot lysis during continuous 2-MHz ultrasound monitoring in stroke patients treated with intravenous tissue plasminogen activator. Stroke 37, 425–429. doi:10.1161/01.str.0000199064.94588.39
Parikh, S., Motarjeme, A., McNamara, T., Raabe, R., Hagspiel, K., Benenati, J. F., et al. (2008). Ultrasound-accelerated thrombolysis for the treatment of deep vein thrombosis: initial clinical experience. J. Vasc. Interv. Radiol. 19, 521–528. doi:10.1016/j.jvir.2007.11.023
Pretorius, E., Mbotwe, S., Bester, J., Robinson, C. J., and Kell, D. B. (2016). Acute induction of anomalous and amyloidogenic blood clotting by molecular amplification of highly substoichiometric levels of bacterial lipopolysaccharide. J. R. Soc. Interface 13, 20160539. doi:10.1098/rsif.2016.0539
Pretorius, E., Venter, C., Laubscher, G. J., Lourens, P. J., Steenkamp, J., and Kell, D. B. (2020). Prevalence of readily detected amyloid blood clots in ‘unclotted’ Type 2 Diabetes Mellitus and COVID-19 plasma: a preliminary report. Cardiovasc Diabetol. 19, 193. doi:10.1186/s12933-020-01165-7
Pretorius, E., Vlok, M., Venter, C., Bezuidenhout, J. A., Laubscher, G. J., Steenkamp, J., et al. (2021). Persistent clotting protein pathology in Long COVID/Post-Acute Sequelae of COVID-19 (PASC) is accompanied by increased levels of antiplasmin. Cardiovasc Diabetol. 20, 172. doi:10.1186/s12933-021-01359-7
Rad, I. J., Chapman, L., Tupally, K. R., Veidt, M., Al-Sadiq, H., Sullivan, R., et al. (2023). A systematic review of ultrasound-mediated drug delivery to the eye and critical insights to facilitate a timely path to the clinic. Theranostics 13, 3582–3638. doi:10.7150/thno.82884
Rambaran, R. N., and Serpell, L. C. (2008). Amyloid fibrils: abnormal protein assembly. Prion 2, 112–117. doi:10.4161/pri.2.3.7488
Rasouli, M. R., and Tabrizian, M. (2019). An ultra-rapid acoustic micromixer for synthesis of organic nanoparticles. Lab. Chip 19, 3316–3325. doi:10.1039/c9lc00637k
Rasouli, R., Villegas, K. M., and Tabrizian, M. (2023). Acoustofluidics – changing paradigm in tissue engineering, therapeutics development, and biosensing. Lab. Chip 23, 1300–1338. doi:10.1039/d2lc00439a
Sakharov, D. V., Hekkenberg, R. T., and Rijken, D. C. (2000). Acceleration of fibrinolysis by high-frequency ultrasound: the contribution of acoustic streaming and temperature rise. Thromb. Res. 100, 333–340. doi:10.1016/s0049-3848(00)00319-4
Školoudík, D., Hrbáč, T., Kovář, M., Beneš, V., Fiedler, J., Branca, M., et al. (2025). Sonolysis during carotid endarterectomy: randomised controlled trial. BMJ 388, e082750. doi:10.1136/bmj-2024-082750
Tsivgoulis, G., and Alexandrov, A. V. (2007). Ultrasound-enhanced thrombolysis in acute ischemic stroke: potential, failures, and safety. Neurotherapeutics 4, 420–427. doi:10.1016/j.nurt.2007.05.012
Turner, S., Khan, M. A., Putrino, D., Woodcock, A., Kell, D. B., and Pretorius, E. (2023b). Long COVID: pathophysiological factors and abnormalities of coagulation. Trends Endocrinol. Metab. 34, 321–344. doi:10.1016/j.tem.2023.03.002
Turner, S., Laubscher, G. J., Khan, M. A., Kell, D. B., and Pretorius, E. (2023a). Accelerating discovery: a novel flow cytometric method for detecting fibrin(ogen) amyloid microclots using long COVID as a model. Heliyon 9, e19605. doi:10.1016/j.heliyon.2023.e19605
Wang, Z., Pan, Y., Huang, H., Zhang, Y., Li, Y., Zou, C., et al. (2022). Enhanced thrombolysis by endovascular low-frequency ultrasound with bifunctional microbubbles in venous thrombosis: in vitro and in vivo study. Front. Bioeng. Biotechnol. 10, 965769. doi:10.3389/fbioe.2022.965769
Ward, R. E., Martinez-Correa, S., Tierradentro-García, L. O., Hwang, M., and Sehgal, C. M. (2024). Sonothrombolysis: state-of-the-art and potential applications in children. Children 11. doi:10.3390/children11010057
Keywords: microclots, ultrasound, fibrinolysis, long COVID, lab on a chip, LIFU
Citation: Rasouli R, Hartl B and Konecky SD (2025) Low-intensity ultrasound lysis of amyloid microclots in a lab-on-chip model. Front. Bioeng. Biotechnol. 13:1604447. doi: 10.3389/fbioe.2025.1604447
Received: 04 April 2025; Accepted: 11 June 2025;
Published: 30 June 2025.
Edited by:
Pablo Taboada, University of Santiago de Compostela, SpainReviewed by:
Feng Shen, Beijing University of Technology, ChinaAbraham Samuel Finny, Waters Corporation, United States
Antonio Topete, University of Guadalajara, Mexico
Copyright © 2025 Rasouli, Hartl and Konecky. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Reza Rasouli, cmV6YUBvcGVud2F0ZXIuY2M=