ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 14 October 2022

Sec. Structural Interventional Cardiology

Volume 9 - 2022 | https://doi.org/10.3389/fcvm.2022.1007888

Hemodynamics of transcatheter tricuspid valve replacement with Lux-Valve

    WW

    Wang Wei 1

    LN

    Li Ning 1,2

    NX

    Ning Xiaoping 1

    XZ

    Xu Zhiyun 1

    LB

    Li Bailing 1

    CC

    Cai Chengliang 1

    YF

    Yang Fan 1

    ZG

    Zhou Guangwei 1

    BY

    Bai Yifan 1

    HL

    Han Lin 1*

    QF

    Qiao Fan 1*

    LF

    Lu Fanglin 1*

  • 1. Department of Cardiovascular Surgery, Changhai Hospital, Naval Military Medical University, Shanghai, China

  • 2. Department of Cardiothoracic Surgery, Naval Medical Center of PLA, Shanghai, China

Article metrics

View details

10

Citations

2,8k

Views

1,4k

Downloads

Abstract

Objective:

Transcatheter tricuspid valve intervention (TTVI) has emerged as an alternative treatment option for high-risk and inoperable patients with symptomatic tricuspid regurgitation (TR). However, scarce data in hemodynamic profiles were available on TTVI. In this paper, we attempt to report the hemodynamic profiles of LuX-Valve.

Methods:

30 patients from July 2020 to July 2021 were enrolled in this study. The patient was diagnosed with severe symptomatic TR. The clinical, invasive hemodynamic, and echocardiographic data were collected.

Results:

The surgical success rate was 100%. The cardiac index and stroke volume increased sharply from 2.42(2.27, 2.85) and 47.8(43.6, 62.0) to 3.04 ± 0.63 and 57.2 ± 14.7, respectively. With the elimination of TR and the increase of forward blood flow of the tricuspid valve, the extravascular lung water [798.0 (673.0, 1147.0) vs. 850.3 ± 376.1, P < 0.01] increased subsequently. The peak right atrium pressure decreased after Lux-Valve implantation (21.0 ± 6.4 vs. 19.4 ± 6.5, P < 0.05). On the contrary, the nadir right atrium pressure increased [10.0(8.0, 15.0) vs. 12.0(10.0, 17.0), P < 0.01]. Notably, the right atrium pressure difference dropped sharply from 9.0(5.0, 13.0) to 5.0(4.0, 8.0) after Lux-Valve implantation. There was no significant change in the pulmonary artery pressure. The right atrium volume decreased from 128(83, 188) to 91(67, 167) mL at 1 month and 107(66,157) mL at 6 months. With the remolding of the right heart chamber, the tricuspid annulus diameter shrank significantly from 42.5 ± 5.6 to 36.6 ± 6.3 mm at 1 month and 36.0 (33.0, 38.0) at 6 months.

Conclusion:

Invasive right atrium pressure may act as a potential candidate for TR evaluation and procedural guidance. Elimination of TR by LuX-Valve implantation improves the cardiac output and right atrium pressure and has no significant effect on the pulmonary artery pressure even with the increment of forward blood flow, suggesting the hemodynamic superiority of transcatheter tricuspid valve replacement but needs further study.

What is known?

Compared with conservative medical treatment alone, transcatheter tricuspid valve intervention can significantly reduce the risk of rehospitalization and mortality due to heart failure.

The cardiac function and exercise tolerance were significantly improved during follow-up in severe TR patients after LuX-Valve implantation procedure, suggesting that LuX-Valve system was safe and effective in symptomatic severe TR patients.

What the study adds?

Elimination of TR by LuX-Valve implantation improves the cardiac output and has no significant effect on the pulmonary artery pressure even with the increment of forward blood flow.

The decreased tricuspid annulus diameter and right atrium volume suggest the remolding of right heart after TR elimination.

Invasive right atrium pressure is an important parameter in hemodynamics.

Introduction

Tricuspid regurgitation (TR) is the most common and neglected valvular heart disease of the right heart system. Secondary TR with the characteristics of right heart enlargement and tricuspid annulus (TA) dilation, which arises as a consequence of pulmonary hypertension induced by left-heart valve surgery and atrial fibrillation, takes the predominant position (1). The majority of TR patients are with the manifestation of chronic hepatic and renal insufficiency, coagulation dysfunction, and poor nutritional status on account of long-term right ventricular dysfunction (2). Hence, the mortality and complication risks of redo tricuspid valve surgery are high (3, 4). It is worth mentioning that most patients received diuretic therapy, but the symptoms of cardiac failure were not well controlled. Compared with conservative medical treatment alone, transcatheter tricuspid valve intervention (TTVI) can significantly reduce the risk of rehospitalization and mortality due to heart failure, suggesting the importance of TTVI (5). Based on the aforementioned characteristics, the 2021 European Society of Cardiology guidelines for valvular heart disease for the first time recommends TTVI as a treatment option for severe symptomatic TR patients at IIb level C (6).

TTVI is in its infancy but with a booming tendency, and gradually becomes an alternative option to minimally invasive surgery (79). Nowadays, TTVI includes leaflet repair, valvuloplasty, heterotopic valve replacement, and orthotropic valve replacement. The approaches include transjugular, transfemoral, and right atrium. However, the concerns for TTVI complications, including low implantation success rate, damage to the surrounding structures of the tricuspid valve (right coronary artery and conduction bundle), and device migration, have been reported in previous studies cohort. Transcatheter tricuspid valve replacement (TTVR) has captured our attention for its merit of eliminating TR instead of degradation of TR. Nevertheless, TTVR is challenging from a technical perspective. At first, the TA is a 3D shape with little calcification, which is insufficient to provide a reliable anchoring zone (10). Secondly, the diameter of the TA changes dynamically with the cardiac cycle, leading to an incomplete fit of the bioprosthesis and the native TA, which may lead to paravalvular leakage. At last, most of the currently reported orthotopic TTVR devices are based on the principle of radial force-dependent, the size of bioprosthesis is unavailable once TA is excessively dilated.

Notably, invasive hemodynamic monitoring has been the cornerstone of surgical management of valvular heart disease. With the popularity of echocardiography, the application of invasive hemodynamic monitors was once limited. However, invasive hemodynamics have been revived with the rise of TTVI recently (11). Exploring hemodynamic changes could not only guide TTVI patient selection and predict patient prognosis, but also deepen the understanding of the pathophysiology of valvular heart disease (12). Previous studies have confirmed that TR elimination after Lux-valve implantation could improve the clinical symptoms, cardiac function, and exercise tolerance of patients (13). However, scarce data in hemodynamic profiles were available on TTVI. In this paper, we attempt to report the hemodynamic changes of LuX-Valve.

Methods

Design and patient enrollment

A total of consecutive 30 patients (11 males) between July 2020 and July 2021 with severe TR were enrolled in this prospective study. All patients who underwent TTVR were with informed consent. The patients were comprehensively evaluated by a multidisciplinary team before surgery and deemed unsuitable for open-heart surgery. The exclusion criteria were listed below: Patients with severe pulmonary hypertension (≥55 mmHg), low left ventricular function (left ventricular ejection fraction <50%), low right ventricular function (tricuspid annular plane systolic excursion (TAPSE) <10 mm or right ventricle fractional area change (FAC) <20%), untreated severe coronary heart disease, coagulation dysfunction, and life expectancy <12 months.

The design of LuX-Valve has been described accurately previously, including a tri-leaflet bioprosthesis, ventricular septal anchor “tongue,” two leaflet-grasping clips, and an atrial disc (1315). Preoperative evaluation of the degree of TR, hemodynamics, and right heart anatomy were achieved by echocardiography, right heart catheterization, and gated cardiac contrast-enhanced CT. Because of the complexity of the anatomical structure of the tricuspid valve complex, preoperative imaging analysis is a key factor for successful implantation. The optimal projection angle and bioprosthesis size were determined by analyzing CT before surgery. The invasive pressure of pulmonary artery, right atrium, and right ventricular were recorded before and after Lux-Valve implantation. The echocardiography data at baseline, 1 month after discharge, and 6 months after discharge were required to collect for all enrolled patients.

Operative procedure

TTVR was performed under general anesthesia in the digital subtraction angiography operating room, and transesophageal echocardiography (TEE) was prepared in advance. The pulse indicator continuous cardiac output (PICCO, PULSION, Germany) was monitored by catheterization of the internal jugular vein and femoral artery. The cardiac output was calculated by thermodilution. The right atrial incision was adopted for the surgical approach. Double-layer 4–0 Prolene purse string suture was used for assisting the implantation of the delivery sheath. TEE was used to guide the implantation of the delivery sheath and LuX-Valve positioning during the operation.

Given the unique anchoring method of LuX-Valve and the periodic changes of the TA with the cardiac cycle, there was no requirement for strict alignment between TA and bioprosthesis plane from our experience. The delivery sheath was adjusted under the guidance of fluoroscopy and TEE to ensure its parallel direction to the interventricular septum for facilitating the fixation of the interventricular septum anchoring component. The bioprosthesis was slowly released with the retrieval of the delivery sheath. The periodical shake of grasping clips could be observed under fluoroscopy once the tricuspid anterior leaflet was hooked. And then, the atrial disc was gradually rebounded. Finally, the ventricular septal anchor “tongue” was secured to the anchoring zone. The time interval from the entry of the delivery sheath into the right atrium and the withdrawal of the sheath out of the right atrium was defined as the device time. As for the TR patients with prior permanent pacemaker implantation, the pacing lead was placed between the bioprosthesis and the native TA after Lux-Valve reimplantation. The hemodynamics was measured immediately before and after Lux-Valve implantation by PICCO. The study design for hemodynamics management of Lux-Valve was shown in Figure 1. Dopamine was used when necessary for inotropic support after surgery. Anticoagulation of warfarin was resumed once pleural fluid drainage was reduced after surgery, and low-molecular-weight heparin bridging anticoagulation was not used in this study. The average time of initiation of warfarin anticoagulation was 2.0 days post-operation. Of note, optimization of intravascular volume was performed during the perioperative period and follow-up.

Figure 1

Figure 1

The study design for hemodynamics management of Lux-Valve.

Statistical analysis

Statistical analysis was performed using SPSS V21.0 (Chicago, Illinois, USA). Normally distributed continuous variables were expressed as mean ± standard deviation, and non-normally distributed continuous variables were expressed as median (interquartile range). Categorical variables are expressed as frequencies (proportions). The comparison of normally distributed continuous variables was tested with a paired-sample t-test or two-way ANOVA properly. The comparison of non-normally distributed continuous variables was tested with Wilcoxon or Friedman test properly. P < 0.05 was considered statistically significant.

Results

Baseline

Table 1 presented the baseline data of the TR patients with a mean age of 65.2 years. The Society of Thoracic Surgeons (STS) and CRS scores were 8.966 ± 4.968 and 8.0 (8.0, 8.2). Kansas City Cardiomyopathy Questionnaire (KCCQ) and 6-min walk distance (6MWD) were 49.0 ± 13.0 and 259.3 ± 70.3 m, respectively. 20 patients were classified as NYHA class III. The main complaint of the patients was chest distress in 28 cases, peripheral edema in 23 cases, and abdominal distension with anorexia in 18 cases. Twenty seven patients received regular oral diuretics before surgery, 24 patients had previous left heart valve replacement surgery, including 13 cases of mitral valve replacement, 11 cases of double valve replacement. Five patients received previous permanent pacemaker implantation. The results of preoperative outcomes were shown in Table 2. The average brain natriuretic peptide was 164.9 pg/mL. The electrocardiogram result indicated that 25 patients were with atrial fibrillation. Ascites was identified by abdominal ultrasound in 3 cases. Echocardiography showed all patients were with severe TR with an instantaneous regurgitation volume of 51.7 ± 27.4 mL. The tricuspid annular plane systolic excursion (TAPSE) and fractional area change (FAC) were 15.0(13.0, 18.0) mm and 46.5 ± 6.8%, respectively. The diameter of the TA was 42.5 ± 5.6 mm (Range: 31–55 mm). Right heart catheterization result indicated that the pulmonary systolic and diastolic blood pressures were 40.2 ± 7.6 and 18.7 ± 5.1 mmHg, respectively.

Table 1

Patient (N = 30)
Male11(36.7%)
Age /years65.2 ± 7.9
Height/cm161.6 ± 6.7
Weight/kg57.9 ± 8.9
Body surface area1.62 ± 0.13
STS score8.966 ± 4.968
CRS score8.0(8.0, 8.2)
KCCQ score49.0 ± 13.0
NYHA class
    III20(66.7%)
    IV10(33.3%)
6MWD/m259.3 ± 70.3
Symptoms
   Chest distress28(93.3%)
   Peripheral edema23(76.7%)
   Abdominal distention18(60.0%)
Comorbidities
   Hypertension6(20.0%)
   Diabetes mellitus3(10.0%)
   Coronary artery disease4(13.3%)
   Permanent pacemaker implantation5(16.7%)
   Cerebrovascular accident3(10.0%)
Prior surgery
    MVR13(43.3%)
    DVR11(36.7%)
    PCI1(3.3%)
    CABG1(3.3%)
Medication
  ACEI/ARB3(10.0%)
  Beta blocker4(13.3%)
  Calcium channel blocker2(6.7%)
  Diuretic27(90.0%)

Patients' profile.

6MWD, 6 Min Walk Distance; MVR, Mitral valve replacement; DVR, Double valve replacement; PCI, Percutaneous Coronary Intervention; CABG, Coronary Artery Bypass Grafting; ACEI, Angiotensin Converting Enzyme Inhibitor; ARB, Angiotensin Receptor Blocker.

Table 2

Patient (N = 30)
Laboratory test
  Leukocyte4.75(3.57,5.99)
  Hemoglobin (g/L)122.0(106.0,133.0)
  Platelet144.3 ± 40.8
  Creatinine (μmol/L)77.0 ± 25.1
  Total bilirubin(μmol/L)20.0 ± 9.1
  Direct bilirubin(μmol/L)6.8(4.9, 10.1)
  Albumin(g/L)41.3 ± 3.6
  Brain natriuretic peptide(pg/mL)164.9(103.2, 307.5)
Electrocardiography
  Atrial fibrillation25(83.3%)
  Pacing rhythm5(16.7%)
Ascites3(10.0%)
Pleural effusion0(0%)
Echocardiography
  Left ventricular ejection fraction (%)62.9 ± 8.8
  Transient Tricuspid regurgitation Volume (mL)51.7 ± 27.4
  TAPSE (mm)15.0(13.0,18.0)
  FAC (%)46.5 ± 6.8
  Tricuspid annulus (mm)42.5 ± 5.6
  Flow reversal in the inferior vena cava30(100%)
Right heart catheterization
  Pulmonary artery systolic pressure/mmHg40.2 ± 7.6
  Pulmonary artery diastolic pressure/mmHg18.7 ± 5.1

Preoperative outcomes.

TAPSE, Tricuspid Annular Plane Systolic Excursion; FAC, Fractional Area Change.

Perioperative outcome

The perioperative outcome was shown in Table 3. Surgical success was achieved in all patients. The average operation time and device time were 180.0 (140.0, 180.0) and 11.5 ± 4.4 min, respectively. The main valve size used in this study was 28–50 mm. Two cases underwent secondary thoracotomy for hemostasis due to excessive pleural fluid drainage. Additionally, 1 patient underwent redo surgery 10 days after LuX-Valve implantation on account of valve migration and died perioperatively. One patient had moderate paravalvular leakage after valve implantation. There was no occurrence of hemodialysis, new-onset of permanent pacemaker implantation, and prolonged ventilation. The ICU time and in-hospital time were 2.0 (2.0, 2.0) and 24.5 ± 7.8 days, respectively.

Table 3

Patient (N = 30)
Operation time/min180.0(140.0,180.0)
Device time/min11.5 ± 4.4
Lux–valve size
  28–404(13.3%)
  28–5010(33.3%)
  28–554(13.3%)
  30–402(6.7%)
  30–505(16.7%)
  30–555(16.7%)
Postoperative 24h chest drainage volume (mL)95.0(40.0,210.0)
Complications
  Hemodialysis0(0%)
  IABP0(0%)
  Permanent pacemaker implantation0(0%)
  Prolonged Tracheal Intubation (>72h)0(0%)
  Reoperation for bleeding2(6.7%)
  Reoperation for valve migration1(3.3%)
  Paravalvular leakage1(3.3%)
ICU time/day2.0(2.0,2.0)
In–hospital time/day24.5 ± 7.8
Death1(3.3%)

Perioperative outcome.

IABP, Intra Aortic Balloon Pump; ICU, Intensive Care Unit.

Hemodynamic study

Of note, we monitored the hemodynamics during operation by PICCO, and the results were shown in Table 4. The cardiac index and stroke volume increased sharply from 2.42(2.27, 2.85) and 47.8(43.6, 62.0) to 3.04 ± 0.63 and 57.2 ± 14.7, respectively. The stroke index and global ejection fraction after LuX-Valve implantation were significantly higher than before operation. With the elimination of TR and the increase of forward blood flow of the tricuspid valve, the extravascular lung water increased subsequently. The extravascular lung water index and pulmonary vascular permeability index decreased significantly from 16.3 ± 6.7 and 2.7 (1.7, 3.5) to 13.0 (9.0, 20.5) and 2.0 (1.3, 3.4), respectively. In contrast, there was no significant change in the pressure of systemic circulation and pulmonary circulation.

Table 4

Before TTVRAfter TTVR
Central venous pressure14.9 ± 4.514.8 ± 4.7
Heart rate82.1 ± 11.887.7 ± 11.5*
Systolic blood pressure116.3 ± 13.9126.5 ± 22.4*
Diastolic blood pressure63.5 ± 12.166.6 ± 13.2
Cardiac index2.42(2.27,2.85)3.04 ± 0.63***
Stroke Volume47.8(43.6,62.0)57.2 ± 14.7**
SVR1297.0(1025.5,1670.5)1245.0 ± 376.6
SI29.4(26.7,35.5)33.5(30.6,38.4) ***
GEF14.4 ± 4.015.0 ± 3.9*
SVV23.0 (19.5,25.8)17.7 ± 8.9
EVLW798.0 (673.0,1147.0)850.3 ± 376.1 **
GEDI906.0 (759.0, 1030.0)928.0 (866.5, 1016.0) *
EVLWI16.3 ± 6.713.0 (9.0,20.5) **
PVPI2.7 (1.7, 3.5)2.0 (1.3, 3.4) **
PAPS41.1 ± 7.542.6 ± 8.4
PAPD20.0 ± 4.720.8 ± 4.3

PICCO results.

PICCO, pulse indicator continuous cardiac output;TTVR, transcather tricuspid valve replacement; SVR, systemic vascular resistance; GEF, global ejection fraction; SVV, Stroke Volumee Variation; SI, stroke index; EVLW, extravascular lung water; EVLWI, extravascular lung water index; GEDI, global enddiastolic index; PVPI, Pulmonary Vascular Permeability Index.

PAPS, systolic pulmonary artery pressure; PAPD, Diastolic pulmonary artery pressure. *P < 0.05; **P < 0.01; ***P < 0.001.

In addition, the invasive pressure of right atrium and right ventricle were recorded during operation (Table 5). The peak right atrium pressure decreased after Lux-Valve implantation (21.0 ± 6.4 vs. 19.4 ± 6.5, P < 0.05). On the contrary, the nadir right atrium pressure increased [10.0(8.0, 15.0) vs. 12.0(10.0, 17.0), P < 0.01]. Notably, the RA pressure difference dropped sharply from 9.0(5.0, 13.0) to 5.0(4.0, 8.0) after Lux-Valve implantation. The volume of right atrium decreased from 128(83,188) mL to 91(67,167) mL at 1 month and 107(66,157) mL at 6 months after elimination of TR and the remodeling of the right heart during follow-up (Table 6). The volume of left ventricle increased significantly with the increment of forward blood flow. The TAPSE decreased significantly from 15.0(13.0, 18.5) to 10.5 ± 3.2 mm after 1 month and 11.0 ± 3.3 mm after 6 months. The FAC and LVEF decreased slightly but without significance. With the remolding of the right heart, the TA diameter shrank significantly from 42.5 ± 5.6 to 36.6 ± 6.3mm at 1 month and 36.0 (33.0, 38.0) at 6 months.

Table 5

Before implantationAfter implantation
Peak RV pressure41.5 ± 7.744.0 ± 9.2
Nadir RV pressure9.4 ± 7.17.0 ± 5.1*
Mean RV pressure21.9 ± 5.422.1 ± 5.0
Peak RA pressure21.0 ± 6.419.4 ± 6.5*
Nadir RA pressure10.0(8.0,15.0)12.0(10.0,17.0)**
Mean RA pressure15.0 ± 4.915.4 ± 5.4
RA pressure difference9.0(5.0,13.0)5.0(4.0,8.0)**
Systolic PAP41.4 ± 7.242.7 ± 8.4
Diastolic PAP19.8 ± 5.020.3 ± 5.2
Mean PAP28.0(24.0,32.0)28.8 ± 6.0

Catheterization results.

RA, right atrium; RV, right ventricle pressure; PAP, pulmonary artery pressure. *P < 0.05; **P < 0.01.

Table 6

Before operation1 month after TTVR6 months after TTVR
RA volume/mL128(83,188)91(67,167)*107(66,157)*
RV volume/mL66(45,94)58.6 ± 21.052(41,64)
LA volume/mL133(106,213)147(104,203)151(107,218)
LV volume/mL90.9 ± 23.8104.8 ± 32.0**103.3 ± 32.0*
Tricuspid annulus /mm42.5 ± 5.636.6 ± 6.3 ***36.0(33.0,38.0) ***
TAPSE/mm15.0(13.0,18.0)10.5 ± 3.2***11.0 ± 3.3***
FAC/%46.5 ± 6.845.1(40.8,48.1)43.2 ± 11.8
LVEF/%62.9 ± 8.857.0(53.5,66.0)57.6 ± 10.3

Echocardiography results.

RA, right atrium; RV, right ventricle; LA, left atrium; LV, left ventricle. TAPSE, Tricuspid Annular Plane Systolic Excursion; FAC, Fractional Area Change; LVEF, Left ventricular ejection fraction. *P < 0.05 compared to before operation; **P < 0.01 compared to before operation; ***P < 0.001 compared to before operation.

Discussion

In this paper, we reported the hemodynamic profiles of LuX-Valve implantation. Elimination of TR by LuX-Valve implantation improves the cardiac output and has no significant effect on the pulmonary artery pressure even with the increment of forwarding blood flow. Our previous studies have demonstrated that TTVR using LuX-Valve system was safe and effective in symptomatic severe TR patients. The cardiac function and exercise tolerance were significantly improved during follow-up (13). Moreover, we, for the first time, proved the feasibility of elimination of TR by Lux-Valve instead of degradation of TR from the hemodynamic aspect.

Prolonged TR leads to hemodynamic abnormalities were verified to be associated with congestive hepatopathy and kidney dysfunction, which was associated with decreased forward cardiac output and circulation perfusion, as well as increased right-sided filling pressure and venous congestion. TR reduction by TTVR device was demonstrated to improve liver function (16). Once TR is eliminated, right atrial pressure should theoretically drop significantly. Instantaneous right atrial pressure changes combined with a marked increase in cardiac output were observed in our study, which may contribute to improved organ function and increased exercise tolerance. Further, intraprocedural invasive right atrial pressures were demonstrated to be associated with TR severity and patient outcomes after transcatheter tricuspid edge-to-edge repair. A lower RA pressure difference was proved with improved outcomes (17). Dannenberg V and colleagues assumed hemodynamic assessment before TTVR was a significant factor for patient prognosis, the logistic regression analysis verified a significant relationship between mean RA pressure and ≥1 grade TR reduction (18). Collectively, a more comprehensive investigation of invasive right atrial pressures may be needed in larger tricuspid TTVI cohorts.

It is also worth noting that in another study of transcatheter tricuspid valve-in-valve therapy for bioprosthetic valve failure, pulmonary artery pressure was increased after valve replacement. In this study, however, pulmonary artery pressure was not significantly changed. This finding led us to focus on hemodynamic studies of Lux-valve. Our results suggest that Lux-valve implantation directly eliminates TR and the increased forward flow does not result in a significant elevation of the pulmonary artery. However, with the increase of forward blood flow, it will inevitably lead to an increase in pulmonary perfusion and left ventricular preload. Therefore, the left ventricular function of TR patients must be in a normal range. For patients with abnormal left ventricular function, the choice of TTVR should be prudent. Additionally, a strengthened diuretic therapy for optimization of intravascular volume was necessary after TTVR since extravascular lung water was elevated as evidenced by PICCO.

As for the significant increase in cardiac stroke volume, it could be explained from the Frank-Starling relationship. Right ventricular stroke volume rise on account of the increase of right ventricular preload. However, a sudden increase in right ventricular filling pressure can lead to decreased compliance of the right ventricular with chronic low right ventricular preload. This may explain the reason why the right ventricular systolic function (TAPSE and FAC) decreased after surgery. Long-term chronic right and left ventricular adaptations after surgery may lead to improved postoperative exercise capacity of TR patients (19). Previous studies have also confirmed that different TTVI devices could significantly improve the clinical symptoms of TR patients, which was demonstrated by the NYHA class, 6MWD, and KCCQ score (20). Additionally, right ventricular remodeling during follow-up, including the decreased TA and the increased TAPSE and FAC level, was verified by echocardiography in previous study (21). However, a decline in TAPSE level was observed in the majority of patients during follow-up in our study. Previous studies have shown that TAPSE has no significant effect on the prognosis of TTVI patients (22). On the contrary, the patient's exercise tolerance and NYHA class during follow-up were significantly improved compared with before operation, which further suggested that TAPSE may not be suitable for assessing right ventricular function. A novel parameter may be needed for right ventricular function assessment in the future.

The application of TTVR devices was in a backward position when compared to other transcatheter repair devices. TTVR may be applicable to a broader indication because of incomplete degrees of TR reduction and functional improvement of the repair devices. In 2017, the GATE bioprosthesis (NaviGate, California, USA) was first reported for clinical application, which was implanted through the transatrial access with 100% technical success, 20% reoperation, and 20% mortality (23, 24). In addition, the EVOQUE system was also used in a compassionate cohort including 25 patients with a technical success rate of 92%. There was no occurrence of intraprocedural mortality, coronary injury, and valve migration, combined with a 100% TR level decrease (25).

TR induced by pacemaker lead could not be neglected in TTVI study, in addition, the TR patients with prior permanent pacemaker implantation were not uncommon. In this study, 5 (16.7%) of 30 cases were with prior permanent pacemaker implantation. The pacing lead was placed between the bioprosthesis and the native TA after Lux-Valve reimplantation. The risk of paravalvular leakage was low on account of the design of atrial disc. Anderson JH et al. identified that the TTVR in the setting of trans-tricuspid valve pacemaker leads without lead extraction or re-replacement can be performed safely with a low risk for complications after analyzing the data from the Valve-in-Valve International Database including 329 cases (26). Similarly, Taramasso M and colleagues verified that TTVI is feasible in selected patients with cardiac implantable electronic device and acute procedural success and short-term clinical outcomes are comparable to those observed in patients without a trans-tricuspid valve lead by analyzing the data from the TriValve registry (27).

In our previous work, we have reported the results of a compassionate multicenter study of Lux-Valve that enrolled 46 TR cases. The surgical success rate was 97.8% with 13.0% in-hospital mortality and 15.2% residual TR (13). 6 cases of 12-month follow-up data after LuX-Valve implantation have also been reported by Sun Z and colleagues (15). Of note, the incidence of paravalvular leakage in this study was 3.3%, which was lower than our previous study, further suggesting the importance of the learning curve in TTVR and emphasizing the importance of a more comprehensive understanding of tricuspid valve anatomy, hemodynamics, and surgical imaging guidance. At last, there are several limitation for our study. At first, the cardiac output was calculated by the thermodilution method in this study, which may lead to underestimation in the presence of significant TR. Secondly, this study was limited by its small cases.

Conclusions

In summary, this study demonstrates that elimination of TR by LuX-Valve implantation improves the cardiac output and right atrium pressure instantaneously and has no significant effect on the pulmonary artery pressure even with the increment of forward blood flow. Additionally, the decreased tricuspid annulus diameter and right atrium volume further verifies the long-term remodeling of right heart after TR elimination. A more comprehensive investigation of invasive right atrial pressures may be needed in larger TTVI cohorts.

Funding

This study was funded by National Natural Science Foundation of China (Grant Nos. 82170376 and 82100383), 234 Panfeng project of Changhai Hospital (Grant No. 2019YXK031), and 2025 Science and Technology Innovation project of Ningbo (Grant No. 2018B10092).

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving human participants were reviewed and approved by Shanghai Changhai Hospital Ethics Committee. The patients/participants provided their written informed consent to participate in this study.

Author contributions

Experiment design: QF and LF. Data collection, analysis, and paper draft: LN, WW, and NX. Experiment: XZ, LB, CC, YF, BY, HL, QF, LF, and ZG. Paper revision: LF, QF, and HL. Funding: LF and LN. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

    Abbreviations

  • RA

    right atrium

  • RV

    right ventricle

  • TTVI

    transcatheter tricuspid valve intervention

  • TTVR

    transcatheter tricuspid valve replacement

  • TR

    tricuspid regurgitation

  • TA

    tricuspid annulus

  • PICCO

    pulse indicator continuous cardiac output.

References

  • 1.

    PrazFMuraruDKreidelFLurzPHahnRTDelgadoVet al. Transcatheter treatment for tricuspid valve disease. EuroIntervention. (2021) 17:791808. 10.4244/EIJ-D-21-00695

  • 2.

    BeslerCUnterhuberMRommelKPUngerEHartungPvon RoederMet al. Nutritional status in tricuspid regurgitation: implications of transcatheter repair. Eur J Heart Fail. (2020) 22:182636. 10.1002/ejhf.1752

  • 3.

    BuzzattiNIaciGTaramassoMNisiTLapennaEDe BonisMet al. Long-term outcomes of tricuspid valve replacement after previous left-side heart surgery. Eur J Cardiothorac Surg. (2014) 46:71319. 10.1093/ejcts/ezt638

  • 4.

    ZackCJFenderEAChandrashekarPReddyYNVBennettCEStulakJMet al. National trends and outcomes in isolated tricuspid valve surgery. J Am Coll Cardiol. (2017) 70:295360. 10.1016/j.jacc.2017.10.039

  • 5.

    TaramassoMBenfariGvan der BijlPAlessandriniHAttinger-TollerABiascoLet al. Transcatheter versus medical treatment of patients with symptomatic severe tricuspid regurgitation. J Am Coll Cardiol. (2019) 74:29983008. 10.1016/j.jacc.2019.09.028

  • 6.

    VahanianABeyersdorfFPrazFMilojevicMBaldusSBauersachsJet al. Group EESD. 2021 esc/eacts guidelines for the management of valvular heart disease. Eur Heart J. (2022) 43:561632. 10.1093/eurheartj/ehab395

  • 7.

    CammalleriVCarpenitoMBonoMCMegaSUssiaGPGrigioniF. Transcatheter tricuspid valve therapy: from anatomy to intervention. Front Cardiovasc Med. (2021) 8:778445. 10.3389/fcvm.2021.778445

  • 8.

    AsmaratsLPuriRLatibANaviaJLRodes-CabauJ. Transcatheter tricuspid valve interventions: landscape, challenges, and future directions. J Am Coll Cardiol. (2018) 71:293556. 10.1016/j.jacc.2018.04.031

  • 9.

    RussoGTaramassoMPedicinoDGennariMGavazzoniMPozzoliAet al. Challenges and future perspectives of transcatheter tricuspid valve interventions: Adopt old strategies or adapt to new opportunities?Eur J Heart Fail. (2021). 10.1002/ejhf.2398

  • 10.

    AgricolaEAsmaratsLMaisanoFCavalcanteJLLiuSMillaFet al. Imaging for tricuspid valve repair and replacement. JACC Cardiovasc Imaging. (2021) 14:61111. 10.1016/j.jcmg.2020.01.031

  • 11.

    AlkhouliMEleidMFNishimuraRARihalCS. The role of invasive hemodynamics in guiding contemporary transcatheter valvular interventions. JACC Cardiovasc Interv. (2021) 14:253144. 10.1016/j.jcin.2021.08.072

  • 12.

    StockerTJHertellHOrbanMBraunDRommelKPRufTet al. Cardiopulmonary hemodynamic profile predicts mortality after transcatheter tricuspid valve repair in chronic heart failure. JACC Cardiovasc Interv. (2021) 14:2938. 10.1016/j.jcin.2020.09.033

  • 13.

    LuFLAnZMaYSongZGCai CL LiBLet al. Transcatheter tricuspid valve replacement in patients with severe tricuspid regurgitation. Heart (British Cardiac Society). (2021) 107:166470. 10.1136/heartjnl-2020-318199

  • 14.

    LuFQiaoFLvYAnZLiuXCaoPet al. Radial force-independent bioprosthesis for transcatheter tricuspid valve implantation in a preclinical model. Int J Cardiol. (2020) 319:1206. 10.1016/j.ijcard.2020.06.070

  • 15.

    SunZLiHZhangZLiYZhangLXieYet al. Twelve-month outcomes of the lux-valve for transcatheter treatment of severe tricuspid regurgitation. EuroIntervention. (2021) 17:81826. 10.4244/EIJ-D-21-00095

  • 16.

    KaramNBraunDMehrMOrbanMStockerTJDeseiveSet al. Impact of transcatheter tricuspid valve repair for severe tricuspid regurgitation on kidney and liver function. JACC Cardiovasc Interv. (2019) 12:141320. 10.1016/j.jcin.2019.04.018

  • 17.

    MahowaldMKNishimuraRAPislaruSVMankadSVNkomoVTPadangRet al. Reduction in right atrial pressures is associated with hemodynamic improvements after transcatheter edge-to-edge repair of the tricuspid valve. Cir Cardiovascular Intervent. (2021) 14:e010557. 10.1161/CIRCINTERVENTIONS.121.010557

  • 18.

    DannenbergVKoschutnikMDonàCNitscheCMascherbauerKHeitzingerGet al. Invasive hemodynamic assessment and procedural success of transcatheter tricuspid valve repair-important factors for right ventricular remodeling and outcome. Front Cardiovasc Med. (2022) 9:891468. 10.3389/fcvm.2022.891468

  • 19.

    PatelKSadeghiSAboulhosnJ. Invasive hemodynamic characteristics in patients undergoing transcatheter tricuspid valve-in-valve implantation for treatment of tricuspid stenosis. World J Pediatr Congenit Heart Surg. (2020) 11:4116. 10.1177/2150135120910366

  • 20.

    BocchinoPPAngeliniFVairoAAndreisAFortuniFFranchinLet al. Clinical outcomes following isolated transcatheter tricuspid valve repair: a meta-analysis and meta-regression study. JACC Cardiovasc Interv. (2021) 14:228595. 10.1016/j.jcin.2021.08.050

  • 21.

    LurzPStephan von BardelebenRWeberMSitgesMSorajjaPHausleiterJet al. Transcatheter edge-to-edge repair for treatment of tricuspid regurgitation. J Am Coll Cardiol. (2021) 77:22939. 10.1016/j.jacc.2020.11.038

  • 22.

    KaramNMehrMTaramassoMBeslerCRufTConnellyKAet al. Value of echocardiographic right ventricular and pulmonary pressure assessment in predicting transcatheter tricuspid repair outcome. JACC Cardiovasc Interv. (2020) 13:125161. 10.1016/j.jcin.2020.02.028

  • 23.

    HahnRTGeorgeIKodaliSKNazifTKhaliqueOKAkkocDet al. Early single-site experience with transcatheter tricuspid valve replacement. JACC Cardiovasc Imaging. (2019) 12:41629. 10.1016/j.jcmg.2018.08.034

  • 24.

    NaviaJLKapadiaSElgharablyHHarbSCKrishnaswamyAUnaiSet al. First-in-human implantations of the navigate bioprosthesis in a severely dilated tricuspid annulus and in a failed tricuspid annuloplasty ring. Circ Cardiovasc Interv. (2017) 10:5840. 10.1161/CIRCINTERVENTIONS.117.005840

  • 25.

    FamNPvon BardelebenRSHenseyMKodaliSKSmithRLHausleiterJet al. Transfemoral transcatheter tricuspid valve replacement with the evoque system: A multicenter, observational, first-in-human experience. JACC Cardiovasc Interv. (2021) 14:50111. 10.1016/j.jcin.2020.11.045

  • 26.

    AndersonJHMcElhinneyDBAboulhosnJZhangYRibichiniFEickenAet al. Management and outcomes of transvenous pacing leads in patients undergoing transcatheter tricuspid valve replacement. JACC Cardiovasc Interv. (2020) 13:201220. 10.1016/j.jcin.2020.04.054

  • 27.

    TaramassoMGavazzoniMPozzoliAAlessandriniHLatibAAttinger-TollerAet al. Outcomes of ttvi in patients with pacemaker or defibrillator leads: data from the trivalve registry. JACC Cardiovasc Interv. (2020) 13:55464. 10.1016/j.jcin.2019.10.058

Summary

Keywords

hemodynamics, tricuspid regurgitation, transcatheter tricuspid valve intervention (TTVI), LuX-Valve, transcatheter tricuspid valve replacement

Citation

Wei W, Ning L, Xiaoping N, Zhiyun X, Bailing L, Chengliang C, Fan Y, Guangwei Z, Yifan B, Lin H, Fan Q and Fanglin L (2022) Hemodynamics of transcatheter tricuspid valve replacement with Lux-Valve. Front. Cardiovasc. Med. 9:1007888. doi: 10.3389/fcvm.2022.1007888

Received

31 July 2022

Accepted

29 September 2022

Published

14 October 2022

Volume

9 - 2022

Edited by

Omar Chehab, St Thomas' Hospital, United Kingdom

Reviewed by

Kensuke Hirasawa, Tokyo Medical and Dental University, Japan; Nicola Buzzatti, San Raffaele Hospital (IRCCS), Italy

Updates

Copyright

*Correspondence: Lu Fanglin Qiao Fan Han Lin

†These authors have contributed equally to this work

This article was submitted to Structural Interventional Cardiology, a section of the journal Frontiers in Cardiovascular Medicine

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.